Augmented Reality-Assisted Spine Surgery: An Early Experience Demonstrating Safety and Accuracy with 218 Screws

Fenil R Bhatt 1, Lindsay D Orosz 2,, Anant Tewari 2, David Boyd 3, Rita Roy 2, Christopher R Good 1, Thomas C Schuler 1, Colin M Haines 1, Ehsan Jazini 1
PMCID: PMC10556900  PMID: 35000409

Abstract

Study Design

Prospective cohort study.

Objectives

In spine surgery, accurate screw guidance is critical to achieving satisfactory fixation. Augmented reality (AR) is a novel technology to assist in screw placement and has shown promising results in early studies. This study aims to provide our early experience evaluating safety and efficacy with an Food and Drug Administration-approved head-mounted (head-mounted device augmented reality (HMD-AR)) device.

Methods

Consecutive adult patients undergoing AR-assisted thoracolumbar fusion between October 2020 and August 2021 with 2 -week follow-up were included. Preoperative, intraoperative, and postoperative data were collected to include demographics, complications, revision surgeries, and AR performance. Intraoperative 3D imaging was used to assess screw accuracy using the Gertzbein-Robbins (G-R) grading scale.

Results

Thirty-two patients (40.6% male) were included with a total of 222 screws executed using HMD-AR. Intraoperatively, 4 (1.8%) were deemed misplaced and revised using AR or freehand. The remaining 218 (98.2%) screws were placed accurately. There were no intraoperative adverse events or complications, and AR was not abandoned in any case. Of the 208 AR-placed screws with 3D imaging confirmation, 97.1% were considered clinically accurate (91.8% Grade A, 5.3% Grade B). There were no early postoperative surgical complications or revision surgeries during the 2 -week follow-up.

Conclusions

This early experience study reports an overall G-R accuracy of 97.1% across 218 AR-guided screws with no intra or early postoperative complications. This shows that HMD-AR-assisted spine surgery is a safe and accurate tool for pedicle, cortical, and pelvic fixation. Larger studies are needed to continue to support this compelling evolution in spine surgery.

Keywords: augmented reality, thoracolumbar fusion, complications, revisions, fixation, pedicle screw, cortical screw

Introduction

Accurate pedicle screw placement is crucial to achieving stable posterior fixation and avoiding complications in spinal surgery. There is growing interest in the use of augmented reality (AR) navigation in spine surgery as an alternative to the current approaches: traditional navigation, robotic-guidance, fluoroscopic-guidance, and freehand.1-7 Augmented reality systems are capable of projecting patient anatomy directly onto the surgical field, eliminating the need for the surgeon to break line-of-sight with the patient while operating. Prior AR studies in spine surgery have shown potential in decreasing intraoperative radiation exposure to operating room staff. 8 Several versions of AR are described in the literature, including systems that incorporate a head-mounted device (HMD) in the surgical workflow. 9 To date, only one HMD-AR system has received FDA approval, granted in December 2019.

There are limited in vivo studies demonstrating the safety, efficacy, and accuracy of HMD-AR navigation when used in spine surgery. Prior studies have shown that AR-guided navigation systems are simple to integrate into modern operating rooms with limited disruptions in traditional workflow and higher screw placement accuracy than traditional freehand techniques.9-12 The purpose of this study is to provide an early experience safety profile and accuracy rating of HMD-AR-guided posterior instrumentation in a cohort of patients undergoing thoracolumbar spinal fusion by 3 fellowship-trained spine surgeons novice to AR technology. We hypothesized that intra- and postoperative complication rates, revision rates, and screw accuracy would be comparable to traditional freehand, fluoroscopic-guided, and robotic-guided techniques without a learning curve.

Methods

Study Design and Patient Selection

This is an IRB-approved prospective observational study of adult patients (age 18–76) undergoing AR-navigated thoracolumbar fusion surgery between October 2020 and August 2021 at a multi-surgeon, single institution (Advarra Pro00042720, 5/16/2020). Surgeries were performed by 3 board certified spine surgeons using the only FDA-approved AR system for spine surgery, xvision-Spine (XVS) system (Augmedics, Ltd, Philadelphia, PA). All 3 surgeons in this study were experienced with other guidance techniques, including fluoroscopy, navigation and robotics, and were able to begin utilizing the HMD-AR system after only 2–3 hours of training with cadavers and demonstration models. All patients signed informed consent and a notice of privacy practices prior to surgery.

Data Collection

Preoperative patient demographic information, Charlson Comorbidity Index (CCI), and smoking status, was collected using electronic questionnaires. Intraoperative data was collected using case report forms (CRFs) and included: complications, radiation time/dose, number of screws revised due to misplacement, estimated blood loss (EBL), surgical time, and adverse events. In-hospital data included: postoperative complications, revisions, and length of stay. Postoperative data collected at 2 weeks included: medical and surgical complications and revision rates.

Intraoperative 3D imaging was performed following instrumentation. The AR-placed screws were graded for accuracy using the highly cited Gertzbein-Robbins (G-R) grading scale by two fellowship-trained spine surgeons and one independent neuroradiologist. 13 A modified G-R grading scale was used to grade S2AI screw accuracy, which reflects screw deviation from the ideal trajectory; a screw breach was defined as perforation of cortical bone resulting in any portion of the screw protruding outside the confines of the pelvis: Grade A screws had 0 mm breach, grade B screws had less than 3 mm breach, grade C screws had 3–6 mm breach, and grade D screws had greater than 6 mm breach.14,15

All analyses were performed using IBM SPSS V 27.0 (Armonk, New York). Descriptive statistics are presented as frequencies (percentages) or means and standard deviations.

Results

Patient Characteristics

Thirty-two patients (40.6% male) with a mean age of 50.9 ± 15.0 years undergoing AR-assisted posterior thoracolumbar instrumented fusion were selected for inclusion. Mean BMI was 30.3 ± 4.9 kg/m2, mean CCI score was 0.3 ± 0.5, and 4 (12.5%) patients were active nicotine users. The most common preoperative diagnoses were stenosis (31.3%) and lumbar post-laminectomy syndrome (28.1%). All patients failed nonoperative treatments prior to surgery, had a mean visual analog scale of 6.0 ± 2.1 for pain, and 24 (75.0%) used narcotic medication for pain control within 1 year prior to surgery. 20 patients (62.5%) had a history of prior spine surgery, including lumbar: fusion, arthroplasty, and decompression (Table 1).

Table 1.

Patient Characteristics.

Characteristic Value
N 32
Male sex 13 (40.6)
Mean age (years) 50.9 ± 15.0
Mean CCI score 0.3 ± 0.5
ASA score: 2–3 31 (96.9)
Mean BMI (kg/m2) 30.3 ± 4.9
Mean DEXA lowest T-score −0.8 ± 1.2
Nicotine use 4 (12.5)
Mean preoperative VAS pain (0–10) 6.0 ± 2.1
Failed nonoperative treatments 32 (100.0)
Taking opioids for spine pain in last year 24 (75.0)
Primary indication
 Deformity 6 (18.8)
 Instability 5 (15.6)
 Post-laminectomy syndrome 9 (28.1)
 Pseudoarthrosis 2 (6.3)
 Stenosis 10 (31.3)
Previous lumbar spine surgery 20 (62.5)
 Arthroplasty 1 (3.1)
 Decompression 8 (25.0)
 Fusion 16 (50.0)

Values represent the number of patients (%) or mean ± SD.

CCI Charlson Comorbidity Index, ASA American Society of Anesthesiologists; BMI Body mass index, DEXA Dual-energy x-ray absorptiometry; VAS Visual analog scale.

Surgical Details

The majority of patients (84.4%) underwent staged circumferential fusion. Posterior approach was midline (open) in 19 (59.4%) patients and percutaneous (MIS) in 13 (40.6%) patients. 261 total screws were placed, of which 222 (85.1%) were initially executed using AR-navigation. Of these, 4 (1.8%) stimulated < 10 milliampules (mA) during pedicle screw stimulation and were removed for revision. Two screws were revised using AR and the remaining 2 were revised freehand, resulting in an overall surgical accuracy of 98.2%. Thirty-nine (14.9%) non-AR-assisted screws were placed using freehand or fluoroscopic-guidance techniques. This was done at the surgeon’s discretion in instances where additional intraoperative 3D imaging would have been required to place screws; in order to reduce potential radiation exposure to the patient and expedite the case the surgeon opted not to utilize AR for screw placement. Augmented reality technology was not abandoned or converted to another navigation system. 3D imaging radiation dose to place screws using AR-navigation was on average 576.8 ± 368.8 mGycm. Average fluoroscopy time was available for 27 patients, with a mean time of 25.7 ± 29.8 seconds, of which 19 (70.4%) had a mean radiation dose of 0.3 ± 0.4 mGym2. Of the 208 (79.7%) AR-placed screws scanned with 3D imaging intraoperatively following instrumentation: 187 (89.9%) were pedicle, 17 (8.2%) were cortical, and 4 (1.9%) were S2A1 screws. (Table 2).

Table 2.

Surgical Details.

Variable Value
Staged circumferential fusion 27 (84.4)
PIF approach
 Midline (open) 19 (59.4)
 Percutaneous (MIS) 13 (40.6)
PIF implants (total) 261 (100.0)
 Cortical screws 17 (6.5)
 Pedicle screws 239 (91.6)
 S2AI screws 5 (1.9)
AR PIF implants (imaged for accuracy grading) 208 (79.7)
 Cortical screws 17 (8.2)
 Pedicle screws 187 (89.9)
 S2AI screws 4 (1.9)
Posterior decompression 25 (78.1)
Revision status
 Revision decompression 2 (6.3)
 Revision fusion 6 (18.8)
Neuromonitoring/pedicle screw stimulation 32 (100.0)
Skin-to-skin time (hrs) 3.6 ± 1.7
Estimated blood loss (mL) 224.0 ± 332.5
Blood transfusion 3 (9.4)
Adverse events 0 (0.0)
Radiation
 Mean total fluoroscopy time (sec) 25.7 ± 29.8
 Mean total fluoroscopy dose (mGycm2) 0.3 ± 0.4
 Mean total 3D imaging dose (mGycm) 576.8 ± 368.8
Instrumentation placement
 Free-hand / fluoroscopic-guided 43 (16.4)
 Augmedics xvision-Spine system 218 (83.5)
 AR abandoned 0 (0.0)
AR total executed screws 222 (85.1)
 No. accurate screws 218 (98.2)
 No. misplaced screws 4 (1.8)

Values represent the number of patients (%) or mean ± SD.

PIF Posterior instrumented fusion, MIS Minimally invasive surgery, No. Number.

Accuracy Grading and Complication and Revision Rates

A total of 208 AR-placed screws were graded for accuracy. 198 (95.2%) thoracolumbar screws were assigned a G-R grade A or B. All 4 (100%) S2A1 screws were assigned a modified G-R grade of A (Table 3).

Table 3.

Accuracy Grading.

Variable Value
Total AR screws graded for accuracy 208
 Gertzbein–Robbins grade A 187 (89.9)
 Gertzbein–Robbins grade B 11 (5.3)
 Gertzbein–Robbins grade C 6 (2.9)
 Gertzbein–Robbins grade D 0 (.0)
 Gertzbein–Robbins grade E 0 (.0)
 Modified G-R (S2AI) grade A 4 (1.9)
 Modified G-R (S2AI) grades B-E 0 (.0)

Values represent the number of patients (%) or mean ± SD.

AR: Augmented reality, G-R: Gertzbein–Robbins, S2AI: S2-alar-iliac.

There were no intraoperative or postoperative complications during the hospital stay. At the 2-week postoperative visit, there were 3 medical complications: 1 (3.1%) DVT and 2 (6.3%) cases of pneumonia. No patients required revision surgery at any point up to the 2 -week follow-up (Table 4).

Table 4.

Complications and Revision Rates.

Variable Value
Hospitalization
 Surgical complications 0 (.0)
 Medical complications 0 (.0)
 Revision surgery 0 (.0)
 Length of stay (days) 4.1 ± 1.6
2 weeks postoperatively
 Surgical complications 0 (.0)
 Medical complications 3 (9.4)
  Deep vein thrombosis 1 (3.1)
  Pneumonia 2 (6.3)
 Revision surgery 0 (.0)

Values represent the number of patients (%) or mean ± SD.

Discussion

To the best of our knowledge, this study has the largest sample size to date reporting complication and accuracy rates for AR-assisted posterior instrumented thoracolumbar fusions using a HMD. We report an overall screw accuracy rate of 97.1% (Gertzbein-Robbins Grade A or B) for 208 AR-placed thoracic, lumbosacral, and pelvic screws. While there were 3 medical complications, no intraoperative or postoperative surgical complications occurred, and no patients required early revision surgery within a 2 -week period.

Head-mounted AR devices have been shown in preliminary studies to be effective during cannulation and pedicle screw placement.16-18 HMD-AR has also been found to have excellent pilot hole placement rates in both novice and experienced surgeons, as well as significantly reduced breach rates in AR-placed screws compared to screws placed freehand in a laboratory setting. 19 In our in vivo study, only 4 out of 222 planned AR screws were considered misplaced as defined by pedicle screw stimulation of less than 10 mA. In two separate instances, a slight medial breach was determined by intraoperative CT imaging, these screws were removed and revised using the same HMD-AR technology. Figure 1 depicts an example of a slight medial breach that was revised using HMD-AR. Two other screws were found to have similar medial or lateral breaches and revised freehand at the surgeon’s discretion. Overall, these results coupled with the prior experience of the surgeons demonstrate that HMD-AR technology is highly effective in real-world scenarios without a significant learning curve using the technology.

Figure 1.

Figure 1.

(a) Axial O-arm image prior to instrumentation used for AR registration demonstrating small right L2 pedicle (b) Axial O-arm image during instrumentation demonstrating slightly medial right L2 pedicle screw (c) Axial O-arm image at the completion of instrumentation demonstrating right L2 pedicle screw repositioned laterally.

As spine surgery has evolved from utilizing traditional freehand and fluoroscopic-guided techniques to 3D imaging technologies such as navigation, robotics, and AR navigation, this paper aims to contribute our initial experience with and evaluation of AR technology. A recent meta-analysis in patients aged 4–22 found an overall complication rate of 5.84% for pedicle screws placed using various techniques: conventional freehand, modified freehand, and image-guided. 20 In the MIS reFRESH study by Good et al., at 1 year postop, the Hazard Ratios for complications and revision surgeries was 5.8 and 11.0 times higher, respectively, in the fluoroscopic-guided group than the robotic-guided group. 21 The retrospective analysis by Yahanda et al. showed excellent results in their cohort of 9 patients undergoing HMD-AR-guided percutaneous placement of 63 screws with no intra- or postoperative complications or revision surgeries up to the most recent follow up (which was not specifically defined, but was at least beyond the hospitalization period). 22 Similarly, we did not encounter any surgical complications or early revision surgeries in the intraoperative or immediate postoperative period. Our results compare favorably to the current literature and suggest that AR-assisted surgery has a low complication rate that is comparable to the current methods for screw placement assistance.

Studies evaluating pedicle screw accuracy often use the Gertzbein–Robbins (G-R) grading scale and consider G-R grades A or B clinically accurate, represented by a pedicle screw with 2 mm of a cortical breach or less.22,23 Several studies have attempted to characterize the accuracy of screws placed using AR technology. In a cadaveric study, Burstrom et al. reported a screw accuracy rate ranging from 97.4% to 100% based on screw diameter and Yahanda et al. reported 100% accuracy in vivo (63 screws) using the G-R grading system.22,23 In a larger cohort of 20 patients and 253 AR-guided screws using a monitor display, Elmi-Terander et al. reported a G-R accuracy rate of 94.1%. 24 Our unique study reports a similar screw accuracy rate (97.1%) with 3 spine surgeons using a head-mounted display and executing pedicle, cortical, and pelvic fixation. This is comparable to that reported in the literature for robotic-guided systems, where accuracy ranges from 97.9% to 100%, suggesting AR technology is a compelling alternative.25-28

The main limitation of this study is the 2-week length of follow-up. Although the primary aim of this paper was to measure intraoperative screw placement accuracy a secondary aim was to report on early complications and revision surgeries which can be affected by duration of follow-up. While our sample size is certainly the largest to date, larger multi-center studies will further characterize screw accuracy under a variety of conditions with different operators. The authors of this study intend to report further follow-up data to determine the long term complication and revision rates for patients undergoing AR-assisted screw placement.

Conclusion

This study reports on the safety and accuracy of 218 HMD-AR-guided screws in 32 patients with an overall Gertzbein–Robbins accuracy of 97.1%. There were no surgical complications or revision surgeries during the 2 -week postoperative period, suggesting that AR-guided screw placement is safe and accurate. While the results of our study show that AR is a compelling evolution in navigation for screw placement, further multi-center studies with larger sample sizes have greater potential to identify superiority to traditional techniques.

Footnotes

The author(s) declared following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: CRG, has received consulting fees from K2M, Stryker, Mazor and is on an advisory board for Stryker, Augmedics, has received royalties from K2M and has stock in Augmedics and NSite. EJ, has received consulting fees from Stryker and Precision Spine. CMH, has received consulting fees from Globus Medical, and Spineart. The authors have no other conflicts of interest to declare.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

Research Ethics Statement: This study was approved by a centralized IRB, Advarra, on May 16th (Pro00042720). This study was conducted in accordance with the 1964 Helsinki Declaration, its amendments, and other equivalent ethical standards. All study participants or their legally authorized representative signed written informed consent, included in the Notice of Privacy Practices, for study enrollment and publication prior to surgery.

ORCID iDs

Fenil R. Bhatt https://orcid.org/0000-0001-9864-8515

Lindsay D. Orosz https://orcid.org/0000-0002-7189-7746

References

  • 1.Burström G, Persson O, Edström E, Elmi-Terander A. Augmented reality navigation in spine surgery: A systematic review. Acta Neurochir. 2021;163(3):843-852. doi: 10.1007/s00701-021-04708-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Yoo JS, Patel DS, Hrynewycz NM, Brundage TS, Singh K. The utility of virtual reality and augmented reality in spine surgery. Ann Transl Med. 2019;7(S5):S171. doi: 10.21037/atm.2019.06.38. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Carl B, Bopp M, Saß B, Nimsky C. Microscope-based augmented reality in degenerative spine surgery: Initial experience. World Neurosurg. 2019;128:e541-e551. doi: 10.1016/j.wneu.2019.04.192. [DOI] [PubMed] [Google Scholar]
  • 4.Auloge P, Cazzato RL, Ramamurthy N, et al. Augmented reality and artificial intelligence-based navigation during percutaneous vertebroplasty: A pilot randomised clinical trial. Eur Spine J. 2020;29(7):1580-1589. doi: 10.1007/s00586-019-06054-6. [DOI] [PubMed] [Google Scholar]
  • 5.Abe Y, Sato S, Kato K, et al. A novel 3D guidance system using augmented reality for percutaneous vertebroplasty. J Neurosurg Spine. 2013;19(4):492-501. doi: 10.3171/2013.7.SPINE12917. [DOI] [PubMed] [Google Scholar]
  • 6.Carl B, Bopp M, Saß B, Pojskic M, Voellger B, Nimsky C. Spine surgery supported by augmented reality. Glob Spine J. 2020;10(suppl_2):41S-55S. doi: 10.1177/2192568219868217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Yuk FJ, Maragkos GA, Sato K, Steinberger J. Current innovation in virtual and augmented reality in spine surgery. Ann Transl Med. 2021;9(1):94. doi: 10.21037/atm-20-1132. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Edström E, Burström G, Omar A, et al. Augmented reality surgical navigation in spine surgery to minimize staff radiation exposure. Spine (Phila Pa 1976). 2020;45(1):E45-E53. doi: 10.1097/BRS.0000000000003197. [DOI] [PubMed] [Google Scholar]
  • 9.Dennler C, Bauer DE, Scheibler AG, et al. Augmented reality in the operating room: A clinical feasibility study. BMC Musculoskelet Disord. 2021;22(1):1-9. doi: 10.1186/s12891-021-04339-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Elmi-Terander A, Burström G, Nachabé R, et al. Augmented reality navigation with intraoperative 3D imaging vs fluoroscopy-assisted free-hand surgery for spine fixation surgery: A matched-control study comparing accuracy. Sci Rep. 2020;10(1):1-8. doi: 10.1038/s41598-020-57693-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Edström E, Burström G, Nachabe R, Gerdhem P, Terander AE. A novel augmented-reality-based surgical navigation system for spine surgery in a hybrid operating room: Design, workflow, and clinical applications. Oper Neurosurg. 2020;18(5):496-502. doi: 10.1093/ons/opz236. [DOI] [PubMed] [Google Scholar]
  • 12.Carl B, Bopp M, Saß B, Voellger B, Nimsky C. Implementation of augmented reality support in spine surgery. Eur Spine J. 2019;28(7):1697-1711. doi: 10.1007/s00586-019-05969-4. [DOI] [PubMed] [Google Scholar]
  • 13.Gertzbein S, Robbins S. Accuracy of pedicular screw placement in vivo. Spine. 1990;15(1):11-14. doi: 10.1097/00007632-199001000-00004. [DOI] [PubMed] [Google Scholar]
  • 14.Shillingford JN, Laratta JL, Park PJ, et al. Human versus robot: A propensity-matched analysis of the accuracy of free hand versus robotic guidance for placement of S2 Alar-Iliac (S2AI) screws. Spine. 2018;43(21):E1297-E1304. doi: 10.1097/BRS.0000000000002694. [DOI] [PubMed] [Google Scholar]
  • 15.Laratta JL, Shillingford JN, Meredith JS, Lenke LG, Lehman RA, Gum JL. Robotic versus freehand S2 alar iliac fixation: In-depth technical considerations. J Spine Surg. 2018;4(3):638-644. doi: 10.21037/jss.2018.06.13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Gibby JT, Swenson SA, Cvetko S, Rao R, Javan R. Head-mounted display augmented reality to guide pedicle screw placement utilizing computed tomography. Int J Comput Assist Radiol Surg. 2019;14(3):525-535. doi: 10.1007/s11548-018-1814-7. [DOI] [PubMed] [Google Scholar]
  • 17.Liebmann F, Roner S, von Atzigen M, et al. Pedicle screw navigation using surface digitization on the Microsoft HoloLens. Int J Comput Assist Radiol Surg. 2019;14(7):1157-1165. doi: 10.1007/s11548-019-01973-7. [DOI] [PubMed] [Google Scholar]
  • 18.Elmi-Terander A, Nachabe R, Skulason H, et al. Feasibility and accuracy of thoracolumbar minimally invasive pedicle screw placement with augmented reality navigation technology. Spine. 2018;43(14):1018-1023. doi: 10.1097/BRS.0000000000002502. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Dennler C, Jaberg L, Spirig J, et al. Augmented reality-based navigation increases precision of pedicle screw insertion. J Orthop Surg Res. 2020;15(1):1-8. doi: 10.1186/s13018-020-01690-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.De Vega B, Navarro AR, Gibson A, Kalaskar DM. Accuracy of pedicle screw placement methods in pediatrics and adolescents spinal surgery: A systematic review and meta-analysis. Glob Spine J. Published online. 2021. doi: 10.1177/21925682211003552. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Good CR, Orosz L, Schroerlucke SR, et al. Complications and revision rates in minimally invasive robotic-guided versus fluoroscopic-guided spinal fusions: The MIS ReFRESH prospective comparative study. Spine. 2021;46:1661-1668. doi: 10.1097/brs.0000000000004048. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Yahanda AT, Moore E, Ray WZ, Pennicooke B, Jennings JW, Molina CA. First in-human report of the clinical accuracy of thoracolumbar percutaneous pedicle screw placement using augmented reality guidance. Neurosurg Focus. 2021;51(2):E10. doi: 10.3171/2021.5.focus21217. [DOI] [PubMed] [Google Scholar]
  • 23.Burström G, Nachabe R, Persson O, Edström E, Elmi Terander A. Augmented and virtual reality instrument tracking for minimally invasive spine surgery: A feasibility and accuracy study. Spine (Phila Pa 1976). 2019;44(15):1097-1104. doi: 10.1097/BRS.0000000000003006. [DOI] [PubMed] [Google Scholar]
  • 24.Elmi-Terander A, Burström G, Nachabe R, et al. Pedicle screw placement using augmented reality surgical navigation with intraoperative 3D imaging: A first in-human prospective cohort study. Spine (Phila Pa 1976). 2019;44(7):517-525. doi: 10.1097/BRS.0000000000002876. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Joseph JR, Smith BW, Liu X, Park P. Current applications of robotics in spine surgery: A systematic review of the literature. Neurosurg Focus. 2017;42(5):E2. doi: 10.3171/2017.2.FOCUS16544. [DOI] [PubMed] [Google Scholar]
  • 26.Huntsman KT, Ahrendtsen LA, Riggleman JR, Ledonio CG. Robotic-assisted navigated minimally invasive pedicle screw placement in the first 100 cases at a single institution. J Robot Surg. 2020;14(1):199-203. doi: 10.1007/s11701-019-00959-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Van Dijk JD, Van Den Ende RPJ, Stramigioli S, Köchling M, Höss N. Clinical pedicle screw accuracy and deviation from planning in robot-guided spine surgery: Robot-guided pedicle screw accuracy. Spine (Phila Pa 1976). 2015;40(17):E986-E991. doi: 10.1097/BRS.0000000000000960. [DOI] [PubMed] [Google Scholar]
  • 28.Jiang B, Pennington Z, Zhu A, et al. Three-dimensional assessment of robot-assisted pedicle screw placement accuracy and instrumentation reliability based on a preplanned trajectory. J Neurosurg Spine. 2020;33(4):519-528. doi: 10.3171/2020.3.SPINE20208. [DOI] [PubMed] [Google Scholar]

Articles from Global Spine Journal are provided here courtesy of SAGE Publications