. 2023 Oct 5;28(40):2300054. doi: 10.2807/1560-7917.ES.2023.28.40.2300054

External quality assessment of orthohantavirus and lymphocytic choriomeningitis virus molecular detection and serology in Europe, 2021

Mert Erdin 1,*, Kamelia R Stanoeva 2,*, Ramona Mögling 2, Miša Korva 3, Nataša Knap 3, Katarina Resman Rus 3, Cristina Domingo 4, Johan HJ Reimerink 2, Ankje de Vries 2, Hussein Alburkat 1, Mira Utriainen 1, Céline M Gossner 5, Tarja Sironen 1,6, Tatjana Avšič-Županc 3,**, Chantal BEM Reusken 2,**, Olli Vapalahti 1,6,7,**
PMCID: PMC10557384  PMID: 37796441

Abstract

Background

Rodent-borne viruses such as orthohantaviruses and arenaviruses cause considerable disease burden with regional and temporal differences in incidence and clinical awareness. Therefore, it is important to regularly evaluate laboratory diagnostic capabilities, e.g. by external quality assessments (EQA).

Aim

We wished to evaluate the performance and diagnostic capability of European expert laboratories to detect orthohantaviruses and lymphocytic choriomeningitis virus (LCMV) and human antibody response towards orthohantaviruses.

Methods

We conducted an EQA in 2021; molecular panels consisted of 12 samples, including different orthohantaviruses (Seoul, Dobrava-Belgrade (DOBV), Puumala (PUUV) and Hantaan orthohantavirus), LCMV and negative controls. Serological panels consisted of six human serum samples reactive to PUUV, DOBV or negative to orthohantaviruses. The EQA was sent to 25 laboratories in 20 countries.

Results

The accuracy of molecular detection of orthohantaviruses varied (50‒67%, average 62%) among 16 participating laboratories, while LCMV samples were successfully detected in all 11 participating laboratories (91-100%, average 96%). The accuracy of serological diagnosis of acute and past orthohantavirus infections was on average 95% among 20 participating laboratories and 82% in 19 laboratories, respectively. A variety of methods was used, with predominance of in-house assays for molecular tests, and commercial assays for serological ones.

Conclusion

Serology, the most common tool to diagnose acute orthohantavirus infections, had a high accuracy in this EQA. The molecular detection of orthohantaviruses needs improvement while LCMV detection (performed in fewer laboratories) had 95% accuracy. Further EQAs are recommended to be performed periodically to monitor improvements and challenges in the diagnostics of rodent–borne diseases.

Keywords: Hantaviruses, Lymphocytic choriomeningitis virus, External Quality Assessment, molecular diagnostics, serology, Europe


Key public health message.

What did you want to address in this study?

Emerging viruses newly appear or increase in frequency and may cause outbreaks or even pandemics. Orthohantaviruses and lymphocytic choriomeningitis virus, carried by rodents, are of public health concern as emerging viruses. We (Emerging Viral Disease Expert Laboratory Network) aimed to evaluate how well the European expert laboratories can detect these viruses.

What have we learnt from this study?

Detection of orthohantavirus genetic material by molecular methods requires improvement. Lymphocytic choriomeningitis virus molecular detection is relatively good. Detecting orthohantavirus infections by measuring human antibodies raised against them is at satisfactory level, especially for diagnosing acute cases.

What are the implications of your findings for public health?

Orthohantaviruses and lymphocytic choriomeningitis virus are important causes for possibly preventable human infections. Being able to accurately diagnose them may help general surveillance, early detection of outbreaks, and localisation of risk areas, where the infections occur. That in turn can lead to better and timely public health measures and prevention of these diseases.

Introduction

Rodent-borne viruses, such as orthohantaviruses and arenaviruses, cause considerable disease burden with regional and temporal differences in incidence and clinical awareness and present potential for outbreaks. Spillover events between various rodent species or from host rodents to humans are frequent [1], and preparedness is essential. One of the main aspects of preparedness for sporadic infections or outbreaks is the ability to diagnose them efficiently and accurately, and hence diagnostic methods need to be evaluated regularly. While two external quality assessments (EQAs) have been performed on orthohantavirus serology in Europe in the past 20 years [2,3], no EQA on molecular diagnostics has been performed for neither orthohantaviruses nor for lymphocytic choriomeningitis virus (LCMV).

Orthohantaviruses (genus Orthohantavirus, family Hantaviridae) are widely distributed in Eurasia. Dobrava-Belgrade orthohantavirus (DOBV), Seoul orthohantavirus (SEOV) and Hantaan orthohantavirus (HTNV) cause haemorrhagic fever with renal syndrome, while Puumala orthohantavirus (PUUV) causes a milder form called nephropathia epidemica [4]. Distribution of these clinically important orthohantaviruses varies partly depending on the reservoir host distribution. The main reservoir hosts of the orthohantaviruses in Europe are yellow-necked mice (Apodemus flavicollis, distributed throughout most of Europe) and striped field mice (Apodemus agrarius, Central and Eastern Europe) for different lineages of DOBV, black rats and brown rats (Rattus rattus/norvegicus, throughout the globe) for SEOV and bank voles (Myodes glareolus, throughout Europe) for PUUV [4-6]. Transmission of orthohantaviruses to humans may occur through inhalation of virus-containing aerosols of rodent excreta or direct contact with the reservoir hosts [4]. In 2020, 1,647 cases [7] of orthohantavirus infections were notified in the European Union/European Economic Area (EU/EEU) countries, most of them in Finland (1164 cases), Germany (229 cases) and Sweden (61 cases). In the light of overlapping distribution of host reservoirs, differences in clinical severity, heterologous vs homologous antibody responses and epidemiological patterns, it is valuable to identify the species of the causative orthohantavirus.

Lymphocytic choriomeningitis virus (genus Mammarenavirus, family Arenaviridae) is an arenavirus with emerging potential due to the worldwide distribution of its host, the house mouse (Mus musculus) [8]. Transmission to humans occurs via inhalation of house mouse secreta (mostly nasal secretions) or direct contact with the infected host animal [8]. The clinical presentation varies from an asymptomatic infection to severe neurological and congenital disease [9-11]. Because of the variety of symptoms and possibly high proportion of mild or asymptomatic infections, detection of LCMV and diagnosis of the infection in humans are seldom considered and therefore underdiagnosed [8]. Thus, it is of importance to demonstrate the use of and assess the diagnostic methods.

The aim of this study was to evaluate the performance and diagnostic capability of European expert laboratories (in the EU/EEA countries and the EU pre-accession countries) to detect orthohantaviruses and LCMV with molecular approaches and human antibody response towards orthohantaviruses by performing an EQA.

Methods

Participants

In August 2021, member laboratories of the European Emerging Viral Diseases Expert Laboratory Network (EVD-LabNet) were invited to register for this EQA or to forward the invitation to competent laboratories in their countries. A choice between receiving a molecular and serological panel, or both, was offered upon EQA registration.

Preparation of the molecular panel

The molecular panel consisted of 12 samples: seven samples with four different orthohantaviruses (one sample with HTNV, two samples with SEOV, two samples with DOBV and two samples with PUUV), two samples with LCMV and three samples were negative. The panel composition is presented in Table 1. Virus stock solutions were prepared by growing LCMV Armstrong (007v-EVA02708), HTNV 76-118 (007v-EVA02761), SEOV R22 (007v-EVA00922), DOBV 907/5 (007v-EVA00807) and PUUV Cg18-20 (007v-EVA00809) in Vero E6 cells (African green monkey epithelial cells, clone E6) in Dulbecco’s modified Eagle’s medium with high glucose and GlutaMAX supplement (Thermo Fisher Scientific, 61965026), supplemented with 4% FBS (Euroclone, ECS0180L) at 37°C, 5% CO2 atmosphere and 95% relative humidity. Viruses were grown for 7 (LCMV), 13 (HTNV), 14 (SEOV and DOBV) and 15 days (PUUV), respectively. All cell culture experiments were performed in Biosafety Level 3 facilities. After incubation, the cell culture supernatants were collected and centrifuged twice at 4°C (10 min at 3200 × g and 5 min at 20,800 × g) in an Eppendorf 5804R centrifuge (Eppendorf, Hamburg, Germany). Supernatants containing viruses were heat and gamma irradiation inactivated (60°C, 1 h). After inactivation, final viral stock quantification was performed as described below. Undiluted viral concentrations were: HTNV 7.10 × 102 copies/mL, SEOV 1.90 × 103 copies/mL, PUUV 4.80 × 103 copies/mL, DOBV 1.70 × 104 copies/mL and LCMV 2.40 × 105 copies/mL. Subsets of the inactivated supernatant were diluted as indicated in Table 1.

Table 1. Orthohantavirus and lymphocytic choriomeningitis virus molecular panel composition and performance of participating laboratories (n = 16) of external quality assessment, Europe, 2021.

Sample ID Virus Strain GenBank Dilution Submissions Laboratories Correct results False negative Inconclusive
Submissions Laboratories
n %
1 HTNV 76-118 NC005222 Undiluted 17 13 11 65 9 5 1
4 SEOV R22 AF488707 Undiluted 15 14 10 67 8 4 1
9 SEOV R22 AF488707 Undiluted 15 14 11 73 9 3 1
6 DOBV 907/5 L41916a Undiluted 18 14 8 44 8 9 1
7 DOBV 907/5 L41916a 1:10 18 14 7 39 7 9 2
10 PUUV CG1820 M63194 Undiluted 20 15 11 55 10 9 0
12 PUUV CG1820 M63194M 1:10 20 15 10 50 10 80 2
3 LCMV ARM 53b M20869 1:10 11 11 11 100 11 0 0
8 LCMV ARM 53b M20869 1:1000 11 11 10 91 11 1 0
Controls (n = 3) NEG Not applicable 153 16 135 88 16 2b 16

DOBV: Dobrava-Belgrade orthohantavirus; HTNV: Hantaan orthohantavirus; ID: identification; LCMV: lymphocytic choriomeningitis virus; NEG: negative; PUUV: Puumala orthohantavirus; SEOV: Seoul orthohantavirus.

aThe DOBV 907/5 isolate, used in this EQA gives 99-100% nucleotide similarity with the DOBV 3970/87 isolate, submitted to GenBank.

bFalse positive reported for negative control samples.

The molecular EQA panel was tested by two independent laboratories before distribution to the participants. For detection of the orthohantaviruses, one expert laboratory used multiple S-segment targeting real-time reverse transcription (RT)-PCR assays [12,13] while the other expert laboratory used the RT-PCR methods as described by Kramski et al. [13]. The two expert laboratories tested detection of LCMV by an in-house S-segment targeting real-time PCR, developed on the basis of alignment of different LCMV S-segment sequences. Briefly, the reaction mix consisted of TaqManFast Virus 1-Step Master Mix (Applied Biosystems), 0.2 µM primers (LCMV S: 5'-gggATCCTAggCTTTTTggAT-3' and LCMV As: 5'-gCCAATAATgACAATZgTTgAT-3') and 0.3 µM probe (LCMV P 5'-FAM-CCTCAAACATTgTCACAATCTgACCCAT-TAMRA-3') used with the following temperature protocol: 50°C for 5 min, 95°C for 20 s, 40 cycles at 95°C for 3 s and 60°C for 30 s. Testing by the expert laboratories yielded consistent results.

Preparation of the serology panel

The serology panel consisted of six human serum samples. Two samples were reactive to PUUV, two to DOBV with commercial enzyme immunoassay (Reagena, Toivala, Finland) and two negative control samples did not contain reactive antibodies against orthohantaviruses. All samples were heat inactivated (56°C, 1 h) before aliquoting. The panel composition is presented in Table 2. For each virus, a sample with both IgM and IgG antibodies, thus representing an acute infection and a sample with IgG but no IgM antibodies, representing a past infection, were included. Prior to shipping to participants, two expert laboratories tested the serology panel using both commercial enzyme immunoassays and in-house immunofluorescence assays. One expert laboratory used both a commercial enzyme immunoassay for DOBV and PUUV (Reagena, Toivala, Finland) and an in-house immunofluorescence assay [14]. The other expert laboratory used a commercial indirect immunofluorescence assay (Hantavirus Mosaic 1 biochips, Euroimmun, Lübeck, Germany).

Table 2. Orthohantavirus serology panel composition and performance of participating laboratories (n = 20) of external quality assessment, Europe, 2021.

Sample ID Virus Infection phase Antibody type Submissions Correct results False results Inconclusive results Cross-reactionsa Sensitivity (%)b
1 PUUV Past infection IgM-/IgG+ 20 12 2 3 3 75
3 DOBV Acute infection IgM+/IgG+ 19 14 0 1 4 95
4 DOBV Past infection IgM-/IgG+ 19 12 0 2 5 90
6 PUUV Acute infection IgM+/IgG+ 20 17 1 0 2 95
2 NEG Not applicable 55 55 0 0 0 100
5 NEG Not applicable 55 55 0 0 0 100

DOBV: Dobrava-Belgrade orthohantavirus; ID: identification; NEG: negative; PUUV: Puumala orthohantavirus.

aThese cross-reactive results were one of the following: heterologous detections of different sero-group orthohantaviruses, homologous detection of same sero-group but different orthohantavirus species, between acute infection (IgM + /IgG + ) and past infection (IgM−/IgG + ) samples.

bSensitivity (%) calculated by using both total number of correct results and total number of cross-reaction results.

Unfortunately, we could not produce serology panel samples for LCMV due to shortage of available sample material necessary for testing of the panel and supply for all participating laboratories.

Distribution of the panels

All samples of the molecular panel were freeze-dried, and serum samples of the serology panel were lyophilised. The EQA was dispatched at ambient temperature in November 2021 to the EQA participants, including accompanying instructions and if requested, customs documentation.

Result submission and evaluation

We collected EQA results via online submission forms (separate molecular and serology forms), hosted on the EU-survey platform. Participants received instructions and links to access the forms via email. Brief instructions were also provided within the results submission forms. We asked participants to submit methodological information, outcomes of each assay used for the diagnosis and the final (diagnostic) result per panel sample. The results for the molecular and serology panel were analysed separately. We used the number of correct results in the total number of samples in each panel to calculate the accuracy.

Microsoft Excel Office 365 and R versions 4.1.3 to 4.3.0/R studio were used for data processing and analysis. Maps were created using the ECDC Map Maker tool (EMMa).

Results

Participation in the external quality assessment

In total, twenty-five laboratories from 20 EU countries or EU pre-accession countries participated in the EQA. Seventeen laboratories opted to receive both panels, while five and three laboratories requested only the molecular or serology panel respectively. In terms of submitted EQA responses, sixteen laboratories (out of 22 received panels) participated in the molecular panel (Figure 1A). All these 16 laboratories used one or more orthohantavirus-specific detection methods, and 11 laboratories used a molecular assay targeting LCMV (Figure 1B). Assays targeting PUUV showed the highest number of different methods, while SEOV had the lowest number.

Figure 1.

Participating laboratories of the external quality assessment of orthohantaviruses and lymphocytic choriomeningitis virus, Europe, 2021 (n = 25)

DOBV: Dobrava-Belgrade orthohantavirus; HTNV: Hantaan orthohantavirus; LCMV: lymphocytic choriomeningitis virus; PUUV: Puumala orthohantavirus; SEOV: Seoul orthohantavirus.

For molecular detection of LCMV, all participating laboratories used one method. The lower part of the graph represents the number of methods used by participant laboratories for each orthohantavirus species.

Figure 1

Twenty of the 25 laboratories (all sent out serology panels) participated in the serology panel (Figure 1A). All these 20 laboratories used IgM assays for detecting acute infection (IgM and IgG positive) and there were a total of 27 submissions of IgM results, while 19 laboratories tested the panel samples also for previous or recent infection (positive IgG and negative IgM), adding up to a total of 27 submissions of IgG results from these 19 laboratories (Figure 1B).

Performance of participant laboratories

The proportion of correct results for molecular detection of orthohantaviruses varied between 50% and 67% and between 91% and 100% for LCMV (Figure 2A). The accuracy was around 62% for orthohantaviruses and 96% for LCMV (Figure 2C). Accuracy corresponds to the capability of laboratories to give right diagnosis to each sample (Table 1).

Figure 2.

Results of detection of orthohantaviruses and lymphocytic choriomeningitis virus and 2021 EQA in Europe

DOBV: Dobrava-Belgrade orthohantavirus; HTNV: Hantaan orthohantavirus; LCMV: lymphocytic choriomeningitis virus; PUUV: Puumala orthohantavirus; SEOV: Seoul orthohantavirus.

Figure 2

Accuracy of orthohantavirus serology for acute infection was 95%, while it was around 82% for past infection (Figure 2C, Table 2). While the accuracy of the serology for acute orthohantavirus infection of DOBV and PUUV was 95%, the accuracy for past infection was 75% for PUUV and 90% for DOBV (Figure 2B). The results for each sample are presented in Table 2.

Results of the molecular panel

Eight different automated systems were used for nucleic acid extraction by 10 laboratories and three different manual nucleic acid extraction methods by six laboratories. More details on the methods for nucleic acid extractions can be found in Supplementary Table S3.

All participants of the molecular EQA tested the panel samples for at least one orthohantavirus species. There were 23 submissions from 16 laboratories with some laboratories using and reporting results of more than one method (Table 3): 16 laboratories used published in-house assays, three laboratories used commercial assays and seven laboratories used unpublished in-house or own-design assays. There was further variety in which orthohantavirus and which segment (mostly L and S) the submitted methods targeted. The total percentage of correct positive results of all applied methods for detection of orthohantaviruses was 59% (Table 3). The detailed list of orthohantavirus molecular methods used and results reported is available in Supplementary Table S1.

Table 3. Results and summary of methods used for detection of orthohantaviruses and lymphocytic choriomeningitis virus in the molecular panel of the external quality assessment, Europe, 2021.

Method Number of laboratories Target Correct positive /expected positivesa False negative/expected positivesa False positive/expected negativesa Inconclusivea
Orthohantaviruses (7 samples)
In-house PCR [12,13,15-20] 5 Pan-Hanta L-segment 9/35 26/35 0/25 1/60
3 Hantavirus S-segment 14/21 7/21 0/15 2/36
12 Virus-specific S-segment 28/35 6/35 4/109 3/144
3 Virus-specific unknown gene targetb 9/10 0/10 0/26 1/36
Commercial PCR assays 2 HTNV 0/1 1/1 2/11 0/12
Hantavirus generic unknown gene target 5/7 2/7 0/5 0/12
Sequencing Illumina NextSeq 1 Unbiased HTS approach 3/7 4/7 0/5 0/12
Total (%) 58.62 39.66 3.06 2.24
Lymphocytic choriomeningitis virus (2 samples)
In-house PCR [21-25] 1 Pan-Arena L-segment 2/2 0/2 0/10 0/12
6 LCMV S-segment 12/12 0/12 0/60 0/72
2 LCMV L-segment 3/4 1/4 2/20 0/24
1 LCMV unknown gene targetb 2/2 0/2 0/10 0/12
Sequencing 1 LCMV sequencing 2/2 0/2 0/10 0/12
Total (%) 95.45 4.55 1.81 0

HTNV: Hantaan orthohantavirus; HTS: high-throughput sequencing; LCMV: lymphocytic choriomeningitis virus.

aCorrect positive results and false negative were calculated by taking into account the number of panel samples containing the target virus. In contrast, inconclusive results and false positive were calculated using the total number of samples (n = 12) and expected number of negative samples. The number of expected results varies depending on the used methods.

bLaboratories that did not provide any protocol for molecular detection were classified under unknown.

Eleven of 16 laboratories participated in the LCMV molecular panel (Table 3). Among the 11 laboratories testing for LCMV, five used in-house assays previously published, one used sequencing, one an unpublished own-design molecular assay and four did not give information on the method applied. Assays targeting LCMV S-segment were used by five laboratories and L-segment by three. The percentage of correct positive results reported for LCMV was 95% (Table 3). The detailed list of LCMV molecular methods used and results reported is available in Supplementary Table S1.

Results of the serology panel

Thirteen laboratories used enzyme immunoassays (EIA) for IgM testing, six used immunoblotting (IB) and four used both immunochromatography (IC) and immunofluorescence assays (IFA). While EIAs targeted generic orthohantaviruses or orthohantavirus antigenic pools, IBs, ICs and IFAs targeted specific orthohantavirus species. The percentage of correct results reported for serology of acute orthohantavirus infection was 94% (at assay level). The proportion of cross-reactions, defined as a positive result between orthohantaviruses from different serogroups or an IgM negative/IgG positive sample reported IgM positive, was 15% (Table 2, Table 4). More details can be found in Supplementary Table S2. Testing for IgG was mainly performed by EIA: 13 of the 19 participating laboratories. Eight laboratories used IFA and six used IB. As with IgM assays, IgG assays also targeted orthohantavirus antigen pools in EIAs and specific orthohantavirus species in IFAs and IBs. In total, 91% of the results for serology of past orthohantavirus infection were correct (at assay level). The total proportion of cross-reactions, meaning a positive result caused by orthohantaviruses from a different serogroup (PUUV-like vs. DOBV-like), or a heterologous virus within the group (DOBV vs SEOV) was 11% (Table 4). A detailed list of serology methods used and results reported is given in Supplementary Table S2.

Table 4. Results and summary of methods for determination of acute and past infection of orthohantaviruses in the serology panel of the external quality assurance, Europe, 2021.

Method Assays Number of laboratories Target Correct resultsa False-negative resultsa Cross-reactive resultsa,b Inconclusive resultsa
Acute infection (IgM)
EIA Commercial 1 Hantavirus sero-group or virus specific 1/1 0/1 0/4 0/4
11 Hantavirus generic 21/22 1/22 8/44 9/44
In-house [26] 1 Hantavirus sero-group or virus specific 2/2 0/2 1/4 0/4
ICA Commercial 4 4/4 0/4 3/16 2/16
IB Commercial 6 12/12 0/12 1/24 4/24
IFA Commercial 3 4/6 0/6 3/12 2/12
In-house 1 2/2 0/2 0/4 0/4
Total (%) 93.87 2.04 14.81 15.74
Past infection (IgG)
EIA Commercial 1 Hantavirus sero-group or virus specific 2/2 0/2 0/4 0/4
11 Hantavirus generic 41/44 1/44 0/44 2/44
In-house [25] 1 Hantavirus sero-group or virus specific 2/2 0/2 2/4 0/4
IB Commercial 6 22/24 0/24 2/24 0/24
IFA Commercial 3 8/12 0/12 2/12 2/12
In-house 5 9/10 0/10 6/20 1/16
Total (%) 89.36 1.06 11.11 4.80

EIA: enzyme Immunoassay; IB: immunoblot assay; ICA: immunochromatography assay; IFA: immunofluorescence assay.

aThe number of correct results, inconclusive results and false-negative results were calculated using only the total number of target samples (n = 91). In contrast, cross-reactive results were calculated by including all samples that might potentially give a positive result (n = 108; number of submissions × number of potentially reactive samples for one submission = total number of potentially reactive samples).

bThese cross-reactive results were either heterologous detections of different sero-group orthohantaviruses or between acute infection (IgM + /IgG + ) and past infection (IgM-/IgG + ) samples.

Discussion

The results of the serological part of this EQA performed among 20 expert laboratories in Europe, suggest that serological diagnostics for acute orthohantavirus infection was performed with 95% accuracy. Serological methods are the most common means to diagnose an orthohantavirus infection in Europe. However, the molecular part of this EQA showed that sensitivity of the molecular diagnostics for orthohantaviruses should be improved.

In the molecular EQA for orthohantaviruses, correct results were obtained in 62% of positive samples. One reason for this suboptimal performance might be the widespread use of pan-hantavirus targeting primers [15]; the molecular detection sensitivity for orthohantaviruses of these primers was low. Even though this method shows good performance for higher viral loads and is particularly useful for detection of new or previously unknown orthohantaviruses from rodents, it did not provide good results for diagnostic purposes in this EQA. Interestingly, the study also suggests that methods targeting S-segment instead of L-segment of orthohantaviruses performed better.

Even though LCMV cases might be unrecognised or unreported, the LCMV molecular EQA gave 96% correct results. Similar to orthohantaviruses, LCMV S-segment is the one target that participants preferred. One false negative detection for LCMV was from a method targeting the L-segment. However, notably, not all participating laboratories tested for LCMV (11 out of 16) and given the global and pan-European circulation of LCMV, all expert laboratories should be alert about LCMV presence and have the capability to diagnose LCMV infections.

Most of the serological methods for acute infection used were commercial assays, that in general lack the ability to differentiate between responses to orthohantaviruses at the species level. Such differentiation is challenging unless laboratories use species-specific assays or use molecular methods in parallel. However, knowledge of the orthohantavirus species may not be essential for the clinical diagnosis as long as orthohantavirus infection is confirmed and both PUUV and DOBV-related antigens are used to cover the serological spectrum in Europe. The assays were accurate yielding correct results in 94% of positive samples. In a previous EQA from 2012, the percentage of correct IgM results was 62% [2]. The proportion of cross-reactive and inconclusive results was higher among the IgM assays, and this, we assume, occurred mainly due to weak or at-the-threshold signal in pooled commercial assays. Fortunately, most of the participating laboratories that performed these pooled assays, also used confirmatory assays, such as IB or IFA, and this finally led to accurate result of panel samples (see Diagnostic Conclusions). In summary, compared with the previous EQAs, there has been considerable improvement in the diagnosis of acute orthohantavirus infections via serological assays. However, due to relatively low numbers of participating laboratories, the findings cannot be extrapolated directly to the larger diagnostic community.

A high proportion of the assays used for orthohantavirus IgG detection were commercial assays, not designed to differentiate between the different orthohantavirus species. However, in-house assays were used more often for IgG detection than for IgM detection. Although percentages of false-negative results and cross-reactive results were lower in IgG assays than IgM assays, the number of total correct positive results was also lower than for IgM assays. The main reason for this is that IgM negative and IgG positive samples were slightly reactive in IgM assays, which led some laboratories to report their final conclusions incorrectly for past infection samples as "inconclusive”, thus decreasing the rate of accuracy of past infection serology in general. Furthermore, while the percentage of correct results for IgG in this EQA is 89%, it was 88% in the previous EQA in 2010–2011 [2]. Consequently, there has been no apparent change in the rate of IgG detection of orthohantaviruses in European expert laboratories over the past decade. Furthermore, the participating laboratories, performing orthohantavirus and LCMV diagnostics, were few. Moreover, not all participants opted to receive both molecular and serology panels, outlining differences in the array of diagnostics methods available across Europe.

Conclusion

External quality assessments are useful for assessment diagnostic capability and accuracy, and it is important to perform EQAs periodically. Here we targeted methods to detect and diagnose rodent-borne virus infections. This EQA demonstrated that serological diagnosis, which comprises almost all diagnostics used for acute human orthohantavirus infections occurring in Europe, is at an acceptable level with 95% accuracy. However, the performance of molecular detection methods for use in clinical diagnostics of orthohantaviruses could be improved, for example through introduction and use of S-segment targeted or orthohantavirus species specific assays. While molecular methods to detect LCMV were performing well, the number of participating laboratories with this capacity was limited.

Ethical statement

Not applicable, the study involved no human subjects nor animals.

Funding statement

This work was financed by the European Centre for Disease Prevention and Control (ECDC) under specific contract No 1 ECD.11886 ID 23729 implementing Framework contract No ECDC/2020/010. For providing access to the virus strains used, we thank the EVA GLOBAL consortium (funded by the European Union’s Horizon 2020 research and innovation programme under grant agreement No. 871029).

Acknowledgements

We thank all the EQA participants: Control of Communicable Diseases Department, Unit of Laboratory Surveillance of Viral Emergent Diseases, Institute of Public Health, Tirana, Albania; Institute of Virology, University of Veterinary Medicine Vienna, Vienna, Austria; Center for Virology, Medical University of Vienna, Vienna, Austria; Clinical Reference Laboratory, ITG, Institute of Tropical Medicine, Antwerpen, Belgium; National Reference Laboratory of Vector-borne Pathogens, NCIPD, National Center of Infectious and Parasitic Disease, Sofia, Bulgaria; Research Unit, University Hospital for Infectious Diseases “Dr. Fran Mihaljević”, Zagreb, Croatia; Virus Department, Statens Serum Institut, Copenhagen, Denmark; CNR des Hantavirus, Institut Pasteur, Paris, France; Institute of Virology Charite, Universitätsmedizin Berlin, Berlin, Germany; Department of Microbiology, Aristotle University of Tessaloniki, Tessaloniki, Greece; Scientific Department, Army Medical Center, Rome, Italy; National Institute for Infectious Diseases Lazzaro Spallanzani IRCCS, Rome, Italy; Microbiology and Virology Laboratory, Infectious Diseases Department, Amedeo di Savoia Hospital – ASL Città di Torina, Torino, Italy; Istituto Superiore di Sanità, Rome, Italy; Dipartimento di Medicina Molecolare, Università di Padova, Padova, Italy; Department of Serology and Zoonosis, National Institute of Public Health, Prishtine, Kosovo; National Microbiology Reference Laboratory, Riga East University Hospital, Riga, Latvia; Département de Microbiologie, Laboratoire National de Santé, Dudelange, Luxembourg; Laboratory of Virology, Institute of Public Health, Skopje, North Macedonia; Molecular Diagnostics, Pathology Department, Mater Dei Hospital, Msida, Malta; Clinical Virology, Erasmus MC, Rotterdam, the Netherlands; Center for Vectors and Infectious Diseases Research, National Institute of Health, Lisbon, Portugal; Virologicky ustav, Biomedical Research Center of the Slovak Academy of Science, Bratislava, Slovakia; Laboratoria de Arbovirus, Instituto Salud Carlos III, Madrid, Spain; Department of Microbiology, Public Health Agency of Sweden, Solna, Sweden.

In addition, we would like to thank Rémi N. Charrel (AMU), Lance D. Presser (RIVM), Tamás Bakonyi (ECDC), as well as all the staff members who provided laboratory, administrative and other technical support at RKI, University of Ljubljana, University of Helsinki and RIVM.

Supplementary Data

Supplementary Material

Conflict of interest: None declared.

Authors’ contributions: Conceptualisation: CBEMR, OV, TS; Methodology, Data Analysis: ME, KRS, RM; Validation: TAZ, MK, NK, KRR, CDC, JHJH, AdV, HA, MU; Visualisation: ME, KRS; Project Administration: KRS, RM, CMG; Supervision: TS, TAZ, CBEMR, OV; Resources: OV, CBEMR, CMG; Writing – Original Draft: ME, KRS, RM; Writing – Reviewing and Editing: ME, KRS, RM, TS, OV, CMG, CBEMR, MK, NK, KRR, CDC, JHJR, AdV, HA, MU, TAZ.

References

  • 1. Williams EP, Spruill-Harrell BM, Taylor MK, Lee J, Nywening AV, Yang Z, et al. Common themes in zoonotic spillover and disease emergence: lessons learned from bat- and rodent-borne RNA viruses. Viruses. 2021;13(8):1509. 10.3390/v13081509 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Escadafal C, Avšič-Županc T, Vapalahti O, Niklasson B, Teichmann A, Niedrig M, et al. Second external quality assurance study for the serological diagnosis of hantaviruses in Europe. PLoS Negl Trop Dis. 2012;6(4):e1607. 10.1371/journal.pntd.0001607 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Biel SS, Donoso Mantke O, Lemmer K, Vaheri A, Lundkvist A, Emmerich P, et al. Quality control measures for the serological diagnosis of hantavirus infections. J Clin Virol. 2003;28(3):248-56. 10.1016/S1386-6532(03)00009-X [DOI] [PubMed] [Google Scholar]
  • 4. Vaheri A, Strandin T, Hepojoki J, Sironen T, Henttonen H, Mäkelä S, et al. Uncovering the mysteries of hantavirus infections. Nat Rev Microbiol. 2013;11(8):539-50. 10.1038/nrmicro3066 [DOI] [PubMed] [Google Scholar]
  • 5. Forbes KM, Sironen T, Plyusnin A. Hantavirus maintenance and transmission in reservoir host populations. Curr Opin Virol. 2018;28:1-6. 10.1016/j.coviro.2017.09.003 [DOI] [PubMed] [Google Scholar]
  • 6. Kruger DH, Figueiredo LTM, Song JW, Klempa B. Hantaviruses--globally emerging pathogens. J Clin Virol. 2015;64:128-36. 10.1016/j.jcv.2014.08.033 [DOI] [PubMed] [Google Scholar]
  • 7.European Centre for Disease Prevention and Control (ECDC). Hantavirus Infection. Annual Epidemiological Report for 2020.Stockholm: ECDC, 2023.Available from: https://www.ecdc.europa.eu/en/publications-data/hantavirus-infection-annual-epidemiological-report-2020 [Google Scholar]
  • 8. Vilibic-Cavlek T, Savic V, Ferenc T, Mrzljak A, Barbic L, Bogdanic M, et al. Lymphocytic choriomeningitis-emerging trends of a neglected virus: a narrative review. Trop Med Infect Dis. 2021;6(2):88. 10.3390/tropicalmed6020088 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Delaine M, Weingertner AS, Nougairede A, Lepiller Q, Fafi-Kremer S, Favre R, et al. Microcephaly caused by lymphocytic choriomeningitis virus. Emerg Infect Dis. 2017;23(9):1548-50. 10.3201/eid2309.170775 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Mathur G, Yadav K, Ford B, Schafer IJ, Basavaraju SV, Knust B, et al. High clinical suspicion of donor-derived disease leads to timely recognition and early intervention to treat solid organ transplant-transmitted lymphocytic choriomeningitis virus. Transpl Infect Dis. 2017;19(4):e12707. 10.1111/tid.12707 [DOI] [PubMed] [Google Scholar]
  • 11. Folk S, Steinbecker S, Windmeyer J, Macneil A, Campbell S, Rollin PE. Lymphocytic choriomeningitis with severe manifestations, Missouri, USA. Emerg Infect Dis. 2011;17(10):1973-4. 10.3201/eid1710.110911 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Aitichou M, Saleh SS, McElroy AK, Schmaljohn C, Ibrahim MS. Identification of Dobrava, Hantaan, Seoul, and Puumala viruses by one-step real-time RT-PCR. J Virol Methods. 2005;124(1-2):21-6. 10.1016/j.jviromet.2004.10.004 [DOI] [PubMed] [Google Scholar]
  • 13. Kramski M, Meisel H, Klempa B, Krüger DH, Pauli G, Nitsche A. Detection and typing of human pathogenic hantaviruses by real-time reverse transcription-PCR and pyrosequencing. Clin Chem. 2007;53(11):1899-905. 10.1373/clinchem.2007.093245 [DOI] [PubMed] [Google Scholar]
  • 14. Avsic-Zupanc T, Petrovec M, Furlan P, Kaps R, Elgh F, Lundkvist A. Hemorrhagic fever with renal syndrome in the Dolenjska region of Slovenia--a 10-year survey. Clin Infect Dis. 1999;28(4):860-5. 10.1086/515185 [DOI] [PubMed] [Google Scholar]
  • 15. Klempa B, Fichet-Calvet E, Lecompte E, Auste B, Aniskin V, Meisel H, et al. Hantavirus in African wood mouse, Guinea. Emerg Infect Dis. 2006;12(5):838-40. 10.3201/eid1205.051487 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Weidmann M, Schmidt P, Vackova M, Krivanec K, Munclinger P, Hufert FT. Identification of genetic evidence for dobrava virus spillover in rodents by nested reverse transcription (RT)-PCR and TaqMan RT-PCR. J Clin Microbiol. 2005;43(2):808-12. 10.1128/JCM.43.2.808-812.2005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Bowen MD, Gelbmann W, Ksiazek TG, Nichol ST, Nowotny N. Puumala virus and two genetic variants of Tula virus are present in Austrian rodents. J Med Virol. 1997;53(2):174-81. [DOI] [PubMed] [Google Scholar]
  • 18. Pang Z, Li A, Li J, Qu J, He C, Zhang S, et al. Comprehensive multiplex one-step real-time TaqMan qRT-PCR assays for detection and quantification of hemorrhagic fever viruses. PLoS One. 2014;9(4):e95635. 10.1371/journal.pone.0095635 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Papa A, Johnson AM, Stockton PC, Bowen MD, Spiropoulou CF, Alexiou-Daniel S, et al. Retrospective serological and genetic study of the distribution of hantaviruses in Greece. J Med Virol. 1998;55(4):321-7. [DOI] [PubMed] [Google Scholar]
  • 20. Niskanen S, Jääskeläinen A, Vapalahti O, Sironen T. Evaluation of real-time RT-PCR for diagnostic use in detection of Puumala virus. Viruses. 2019;11(7):661. 10.3390/v11070661 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Bowen MD, Peters CJ, Nichol ST. Phylogenetic analysis of the Arenaviridae: patterns of virus evolution and evidence for cospeciation between arenaviruses and their rodent hosts. Mol Phylogenet Evol. 1997;8(3):301-16. 10.1006/mpev.1997.0436 [DOI] [PubMed] [Google Scholar]
  • 22. Cordey S, Sahli R, Moraz ML, Estrade C, Morandi L, Cherpillod P, et al. Analytical validation of a lymphocytic choriomeningitis virus real-time RT-PCR assay. J Virol Methods. 2011;177(1):118-22. 10.1016/j.jviromet.2011.06.018 [DOI] [PubMed] [Google Scholar]
  • 23. Vieth S, Drosten C, Lenz O, Vincent M, Omilabu S, Hass M, et al. RT-PCR assay for detection of Lassa virus and related Old World arenaviruses targeting the L gene. Trans R Soc Trop Med Hyg. 2007;101(12):1253-64. 10.1016/j.trstmh.2005.03.018 [DOI] [PubMed] [Google Scholar]
  • 24. Fornůsková A, Hiadlovská Z, Macholán M, Piálek J, de Bellocq JG. New perspective on the geographic distribution and evolution of lymphocytic choriomeningitis virus, Central Europe. Emerg Infect Dis. 2021;27(10):2638-47. 10.3201/eid2710.210224 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Coulibaly-N’Golo D, Allali B, Kouassi SK, Fichet-Calvet E, Becker-Ziaja B, Rieger T, et al. Novel arenavirus sequences in Hylomyscus sp. and Mus (Nannomys) setulosus from Côte d’Ivoire: implications for evolution of arenaviruses in Africa. PLoS One. 2011;6(6):e20893. 10.1371/journal.pone.0020893 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lee HW, Calisher C, Schmaljohn CS. Manual of hemorrhagic fever with renal syndrome and hantavirus pulmonary syndrome. Seoul: WHO Collaborating Center for Virus Reference and Research (Hantaviruses); 1998. p. 250. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material

Articles from Eurosurveillance are provided here courtesy of European Centre for Disease Prevention and Control

RESOURCES