Agreement of intraocular pressure measurement with Corvis ST, non-contact tonometer, and Goldmann applanation tonometer in children with ocular hypertension and related factors

Hou-Gang Li 1, Yan-Hui Chen 1,, Fang Lin 2, Si-Yu Li 1, Qing-Hua Liu 1, Chun-Ge Yin 1, Xi-Yue Chen 1, Xin-Jie Zhang 1, Yue Qu 1, Yan-Nian Hui 3,
PMCID: PMC10559031  PMID: 37854370

Abstract

AIM

To access the agreement of intraocular pressure (IOP) values obtained from biomechanically corrected tonometer [Corvis ST (CST)], non-contact tonometer (NCT), and Goldmann applanation tonometer (GAT) in children with NCT measured-IOP (NCT-IOP) values of 22 mm Hg or more, and related factors.

METHODS

A total of 51 eyes with NCT-IOP≥22 mm Hg in children aged 7 to 14y were examined and IOP was measured by CST, NCT, and GAT. Based on GAT measured IOP (GAT-IOP), ocular hypertension (OHT) group (≥22 mm Hg, 24 eyes) and the non-OHT group (<22 mm Hg, 27 eyes) were defined. We compared the agreement of the three measurements, i.e., CST measured IOP (CST-IOP), GAT-IOP, and NCT-IOP, and further analyzed the correlation between the differences in tonometry readings, central corneal thickness (CCT), axial length (AL), optic disc rim volume, and age.

RESULTS

Compared with the OHT group, thicker CCT, larger rim volume, and higher differences between NCT-IOP and GAT-IOP, were found in the non-OHT group. The differences between CST-IOP and GAT-IOP were lower than the differences between NCT-IOP and GAT-IOP in both groups. The mean differences in CST-IOP and GAT-IOP were 1.26 mm Hg (95% limit of agreement ranged from 0.1 to 2.41 mm Hg, OHT group) and 1.20 mm Hg (95% limit of agreement ranged from -0.5 to 3.00 mm Hg, non-OHT group), and the mean differences in NCT and GAT were 3.90 mm Hg (95% limit of agreement ranged from -0.19 to 9.70 mm Hg, OHT group) and 6.00 mm Hg (95% limit of agreement ranged from 1.50 to 10.50 mm Hg, non-OHT group). The differences between CST-IOP and GAT-IOP were not related to CCT, age, and AL in both groups; while the differences between NCT-IOP and GAT-IOP were related to CCT in the OHT group (r=0.93, P<0.001) and to CCT and AL in the non-OHT group (r=0.66, P<0.001, r=-0.81, P<0.001).

CONCLUSION

The accuracy of NCT in the diagnosis of pediatric OHT is low. The agreement of CST-IOP and GAT-IOP was significantly higher in children with and without OHT than in those with NCT-IOP and GAT-IOP. Therefore, CST can be used as a good alternative for IOP measurement in children. The impacts of CCT and AL on NCT measurement need to be fully considered when managing childhood IOP.

Keywords: ocular hypertension, children, non-contact tonometer, Goldmann applanation tonometer, Corvis ST, corneal biomechanics, central corneal thickness, axial length

INTRODUCTION

Ocular hypertension (OHT), defined as intraocular pressure (IOP) of >21 mm Hg and the absence of optic disc damage or glaucomatous visual field results, is common in children ten years old or younger[1][2]. Accurate IOP measurement is critical in diagnosing and managing pediatric OHT. Although children share the same range of normal IOP as adults, the physiology and pathology of OHT in children differ markedly from that in adults, resulting in a more complex condition than in adults. Most pediatric OHTs were detected by non-contact tonometer (NCT) during the visual examination or optical correction in outpatient. The disadvantages of NCT included that the accuracy showed a significant decrease in eyes with higher true IOP or thicker central corneal thickness (CCT)[3]. Moreover, NCT was vulnerable to the effects of long eyelashes and frequent blinking[4]. Consequently, pediatric OHT diagnosed on NCT measured IOP (NCT-IOP) alone can lead to overdiagnosis and overtreatment.

As a gold standard for IOP measurement, Goldmann applanation tonometer (GAT) must touch the eye and requires anesthesia and fluorescein drops. As previously reported, although GAT can achieve accurate IOP values compared with NCT, children often do not cooperate fully with this invasive procedure[5]. In addition, CCT, corneal biomechanics, and axial length (AL) could interfere with GAT readings[6][8].

The biomechanically corrected tonometer (Corvis ST, CST) is a noncontact tonometer. CST could provide a biomechanically corrected IOP value, independent of corneal biomechanics and CCT[9]. Numerous studies have demonstrated that CST-measured IOP (CST-IOP) provides more stable IOP readings in normal eyes[10] and eyes with “weaker” cornea compared with other tonometers[6],[11][12]. And the CST-IOP showed a high agreement with GAT measured IOP (GAT-IOP) in eyes with healthy corneas[10]. Furthermore, an ex vivo experiment indicated no significant difference between CST-IOP and true IOP[13].

Given the limitations of NCT and GAT, it is thus of great importance to assess the accuracy of CST in pediatric OHT. Although GAT, NCT, and CST have been widely used in clinical practice for IOP measurement, little was known about the agreement of the three tonometries in pediatric OHT. We aimed to analyze the agreement of NCT-IOP, GAT-IOP, and CST-IOP in children with an NCT-IOP of 22 mm Hg or more and to evaluate the factors associated with differences in three tonometries. These results may help to explore the utility of CST in pediatric OHT.

SUBJECTS AND METHODS

Ethical Approval

This cross-sectional study was performed in the Ophthalmology Hospital of Tangshan from November 2021 to May 2022. A total of 42 children (aged 7 to 14) were enrolled. All participants' parents signed the informed consent. This study was approved by the Ethics Committee of the Ophthalmology Hospital of Tangshan and complied with the Declaration of Helsinki (approval number: YKYY-LL-2022-22).

The inclusion criteria included an untreated NCT-IOP≥22 mm Hg in at least one eye with normal anterior segment structure and best-corrected visual acuity (BCVA) of 20/25 or better, without visual field defects or optic nerve damage. If both eyes were eligible, both were selected.

This study's exclusion criteria included a history of eye surgery, eye trauma, contact lens wear, corneal astigmatism of 2 D or more, strabismus, congenital eye disease, and topical or systemic corticosteroid application within three months.

All enrolled children underwent comprehensive ophthalmic examinations, including slit-lamp biomicroscopy, BCVA, refractive error, and spherical equivalent (phoropter, VT-10; Topcon, Japan), AL, CCT, and corneal curvature measurement (IOL Master 700, Carl Zeiss Meditec, Jena, Germany), at least two reliable standard automated perimetry with the Octopus (Haag-Streit International, Koeniz, Switzerland; G1 TOP test strategy). Reliable visual field defined as a false-negative rate of <25%, a false-positive rate of <25%, and fixation losses of <20%, and reliability factor not greater than 15%.

IOP Measurement Procedures

All IOP measurements were performed from 2:00 to 5:00 p.m. Children were emotionally stable and avoided blinking. The order of IOP measurement was NCT (Topcon, Tokyo, Japan), CST (Oculus, Wetzlar, Germany), and GAT (AT 20, Carl Zeiss, Germany) with a 10-min interval between three tonometry. NCT was performed with three or more measurements per eye, three values with fluctuation of 3 mm Hg or less were chosen, and the average value was named the NCT-IOP. CST was performed at least twice per eye with an interval of >2min between examinations, and the mean values were named CST-IOP. GAT was performed under topic anesthesia, three consecutive measurements of each eye were taken, and the mean values were named GAT-IOP. Three experienced technicians who were blinded to the patient's condition performed NCT, CST, and GAT.

Stereo Digital Photographs Examination

Stereo digital photographs of the fundus were acquired with a Kowa Nonmyd WX 3D (2D/3D non-mydriatic retinal camera, Kowa, Japan). Vertical cup-to-disc ratio, optic disc area, rim area, and rim volume were automatically measured with built-in software. Corneal curvature and spherical equivalent were used to adjust the ocular magnification.

Statistical Analyses

SPSS 25.0 and MedCalc (version 19.3.0, Ostend, Belgium) were used for data analysis. The normal distribution of data was assessed using the K-S single sample test. For normal data, data were presented as mean±SD or percentages, and differences between groups were compared using the independent sample t-test or Chi-square test. For nonnormal data, variables were represented as medians (upper quartile, lower quartile), and differences between groups were compared using the Mann-Whitney U test. The difference between GAT-IOP, CST-IOP, and NCT-IOP in each group were compared by repeated measurement ANOVA with Bonferroni post hoc test. Bland-Altman plots assessed the agreement between GAT-IOP, CST-IOP, and NCT-IOP. Pearson correlation coefficient (r) analyzed the factors associated with the difference between GAT-IOP, NCT-IOP, and CST-IOP. P<0.05 was considered statistically significant.

RESULTS

Initially, we enrolled 80 eyes (42 children) with NCT-IOP of 22 mm Hg or more. Of them, GAT measurement was performed in 51 eyes. Finally, fifty-one eyes performed GAT, CST, and NCT were divided into two groups: the OHT group (with GAT-IOP≥22 mm Hg) and the non-OHT group (with GAT-IOP<22 mm Hg).

Demographic and Clinical Characteristics of Subjects in Two Groups

Fifty-one eyes (30 patients) completed the GAT measurement, with a GAT success rate of 63.75%. Mean age was 9.88±2.01y, median AL 23.81 mm, male 27.45%, right eyes 47.06%. Mean CST-IOP was 21.70±2.75 mm Hg, with median NCT-IOP 24.90 mm Hg, and median GAT-IOP 21.00 mm Hg. The relationship between age and IOP values (values obtained by GAT, CST, and NCT) was nonlinear. Peaks in CST-IOP and GAT-IOP occurred at age 8 years old and NCT-IOP at age 11, followed by the decrease in IOP with the age increasing in all three tonometries (Figures 1, 2).

Figure 1. Distribution of IOP stratified by age and tonometer in all children.

Figure 1

IOP: Intraocular pressure; NCT: Non-contact tonometry; CST: Biomechanically corrected tonometry (Corvis ST); GAT: Goldmann applanation tonometry.

Figure 2. Scatter plots of IOP measured by biomechanically corrected tonometry (Corvis ST), non-contact tonometry and Goldmann applanation tonometry in the total children (A), OHT group (B), and non-OHT group (C).

Figure 2

OHT: Ocular hypertension group; NCT: Non-contact tonometry; GAT: Goldmann applanation tonometry; CST: Corvis ST.

There was no significant difference in age, AL, sex, disc area, rim area, and the cup-to-disc ratio between the OHT group and non-OHT group (all P>0.05). The rim volume and CCT were significantly lower in the OHT group than in the non-OHT group (P<0.05; Table 1).

Table 1. Differences of characteristics between the OHT group and non-OHT group.

Characteristics OHT group (n=24) Non-OHT group (n=27) P
Age, y 9 (8, 10.5) 9 (9, 13) 0.058a
AL, mm 23.86±1.07 23.76±0.62 0.688b
Male, n (%) 18 (75.0) 19 (70.4) 0.712c
CST, mm Hg 23.93±1.52 19.73±1.97 <0.001b
NCT, mm Hg 26 (23.30, 29.45) 24.70 (23.30, 25.10) 0.077a
GAT, mm Hg 22 (22.00, 23.00) 19 (18.00, 20.00) <0.001a
CCT, µm 546 (535.25, 613.25) 598 (578.00, 619.00) 0.016a
Disc area, mm2 2.37 (1.95, 3.08) 3.09 (2.26, 3.16) 0.089a
Rim area, mm2 1.90 (1.51, 2.25) 2.10 (1.51, 2.28) 0.677a
Rim volume, mm3 0.24 (0.16, 0.26) 0.38 (1.17, 0.42) 0.049a
C/D 0.50 (0.37, 0.57) 0.53 (0.53, 0.56) 0.325a

OHT: Ocular hypertension; AL: Axial length; CST: Corvis ST; NCT: Non-contact tonometry; GAT: Goldmann applanation tonometry; CCT: Central corneal thickness; C/D: Vertical cup-to-disc ratio. aMann-Whitney U test was used to compare the significance between groups for non-normal distribution parameters, and the results were documented as median (lower and upper quartile). bIndependent sample t-test or cChi-square test was used to determine the significance between groups for normal distribution parameters, and the results were documented as mean±SD.

Differences of CST-IOP, NCT-IOP and GAT-IOP

The differences between CST-IOP and GAT-IOP were lower than those between NCT-IOP and GAT-IOP in both groups (all P<0.001). The difference between CST-IOP and GAT was insignificant between the OHT and non-OHT groups (P=0.800). The differences between NCT-IOP and GAT-IOP and the differences between CST-IOP and NCT-IOP significantly differed between both groups (all P<0.05; Table 2).

Table 2. Intra- and intergroup comparisons between the differences in three tonometry readings.

Items CST-GAT, mm Hg NCT-GAT, mm Hg CST-NCT, mm Hg CST-GAT vs NCT-GAT CST-GAT vs CST-NCT NCT-GAT vs CST-NCT
Total 1.23±0.76 5.00±2.80 -3.77±2.69 <0.001a <0.001a <0.001a
OHT (n=24) 1.26±0.59 3.91±2.94 -2.65±2.76 <0.001a <0.001a <0.001a
Non-OHT (n=27) 1.21±0.17 5.98±0.44 -4.77±0.43 0.001a 0.001a <0.001a
t 0.255 -2.800 3.301 - - -
P 0.800b 0.007b 0.004b - - -

OHT: Ocular hypertension; CST: Corvis ST; NCT: Non-contact tonometry; GAT: Goldmann applanation tonometry. aComparison was performed using the repeated measurements analysis of variance with Bonferroni corrected. bIndependent sample t-test was used to determine the significance between OHT group and non-OHT group.

mean±SD

Agreement of IOP Measurement Among Three Devices

The Bland-Altmann analysis plots illustrated the agreement and the difference in IOP among devices (Figure 3). CST showed better agreement with GAT than NCT did in the OHT group (mean IOP difference: 1.26 and 3.90 mm Hg, respectively, P<0.001), in non-OHT group (mean IOP difference: 1.20 and 6.00 mm Hg, respectively, P<0.001), and in all subjects (mean IOP difference: 1.23 and 5.00 mm Hg, respectively, P<0.001).

Figure 3. Bland-Altman scatter plot showing the agreement among different IOP measurement devices.

Figure 3

The Bland-Altmann analysis plots illustrate the agreement of IOP between devices in total groups (A, D, and G), OHT group (B, E, and H), and non-OHT group (C, F, and I). Dotted lines show 95% limits of agreement; IOP: Intraocular pressure; OHT: Ocular hypertension group; CST: Biomechanically corrected tonometry (Corvis ST); NCT: Non-contact tonometry; GAT: Goldmann applanation tonometry.

Correlations Between the IOP Difference, CCT, AL, Optic Disc Rim Volume, and Age

Pearson correlation coefficients evaluated the influence of CCT, AL, optic disc rim volume, and age on the IOP difference between three tonometry reading in each group (Table 3).

Table 3. Correlation of CCT, AL, optic disc rim volume, and age with the differences in three tonometry readings.

Items Total
OHT group
Non-OHT group
CCT AL Age Optic disc rim volume CCT AL Age Optic disc rim volume CCT AL Age Optic disc rim volume
CST-GAT 0.23 (0.101) -0.06 (0.675) -0.27 (0.058) -0.27 (0.052) 0.37 (0.076) 0.24 (0.269) -0.19 (0.370) -0.12 (0.570) 0.24 (0.227) -0.38 (0.052) -0.30 (0.135) -0.34 (0.084)
NCT-GAT 0.85 (<0.001) -0.06 (0.673) 0.12 (0.404) -0.05 (0.742) 0.93 (<0.001) 0.35 (0.092) -0.03 (0.899) 0.05 (0.810) 0.66 (<0.001) -0.81 (<0.001) 0.033 (0.87) -0.41 (0.036)
CST-NCT -0.82 (<0.001) 0.05 (0.748) -0.20 (0.160) -0.03 (0.845) -0.91 (<0.001) -0.33 (0.121) -0.01 (0.956) -0.08 (0.706) -0.58 (0.001) 0.69 (<0.001) -0.15 (0.447) 0.29 (0.149)

r: Pearson correlation coefficient; OHT: Ocular hypertension; CCT: Central corneal thickness; AL: Axial length; CST: Corvis ST; NCT: Non-contact tonometry; GAT: Goldmann applanation tonometry.

r (P)

The difference between CST-IOP and GAT-IOP was not associated with age, AL, optic disc rim volume, and CCT in all subjects and two groups (all P>0.05). The difference between NCT-IOP and GAT-IOP was positively correlated with CCT in all subjects (r=0.85, P<0.001) and in the OHT group (r=0.93, P<0.001) and associated with CCT (r=0.66, P<0.001), AL (r=-0.81, P<0.001), and optic disc rim volume (r=-0.41, P<0.036) in the non-OHT group. The difference between CST-IOP and NCT-IOP was correlated with CCT in all subjects (r=-0.82, P<0.001) and in the OHT group (r=-0.91, P<0.001) and associated with CCT (r=-0.58, P=0.001) and AL (r=0.69, P<0.001) in the non-OHT group.

DISCUSSION

The incidence of OHT is very high in children[2]. IOP in children with OHT may be as high as 28 to 36 mm Hg[14][15]. In the present study, the maximal NCT, GAT, and CST were 33.00, 26.00, and 28.1 mm Hg, respectively. The peak IOP of CST and GAT occurred at age 8y and NCT at age 11y. IOP measurement is central to the management of glaucoma and OHT in children. Children often do not cooperate fully for GAT due to the need for contact with the cornea, which may hinder its practical application. Compared to GAT, CST measurement is more easily administered in children without contact with the cornea. Similar to previous studies[16], the success rate of GAT measurement in the current research was 63.75% (51 eyes); meanwhile, the success rate of CST measurement was 100%. In all 51 eyes with NCT-IOP ≥22 mm Hg, 27 (52.94%) were false OHT, the accuracy of NCT in the diagnosis of pediatric OHT was only 47.06%, using GAT-IOP as a reference.

The result of OHT Treatment Study[17] indicated that OHT patients with thin CCT were more likely to develop open-angle glaucoma. Thin CCT represents an increase in lamina cribrosa compliance, contributing to the deformation of lamina cribrosa under high IOP[18]. Our study found that CCT in the OHT group was significantly thinner than in the non-OHT group (546 vs 578 µm). Compared with the non-OHT group, rim volume was lower in the OHT group, despite the absence of significant glaucomatous optic nerve damage in all enrolled eyes. This result suggests that elevated IOP may impact the structure of optic nerve head[19][20]. Glaucoma may result in irreversible visual impairment in children, which is responsible for 5% of blindness in children worldwide[21][22]. A decrease in IOP is the only reliable treatment strategy for pediatric glaucoma[23]. Accordingly, accurate measurement of IOP and optic nerve assessment should be considered in children, although less than 10% of untreated OHT progress to glaucoma after 5-10y[21],[24].

In addition, CCT thickening can lead to an increase in eyeball stiffness which might contribute to IOP overestimation, which increases the difference between NCT and GAT[25][26]. Consistent with the previous study, our result indicated that the difference between NCT and GAT in the non-OHT group was more significant than in the OHT group (P=0.007). Moreover, the difference between NCT and GAT showed a strong positive correlation with CCT.

Notably, it has also been suggested that elevated IOP can also cause corneal edema via a pressure gradient, ultimately resulting in increased CCT[27]. Therefore, the relationship between CCT and IOP in children remains uncertain[28]. It appears progressively more evident that the true IOP is likely unpredictable with linear correction formulas for CCT. At the same time, the corneal biomechanics might likely contribute, together with CCT, to the tonometric artifact[29]. Compared to GAT and NCT, CST adequately accounted for corneal biomechanical properties and provided a more comprehensive performance for IOP assessment[6],[11],[13]. In a study by Halkiadakis et al[30] no correlation was found between CST-IOP and CCT in eyes with OHT, primary open angle glaucoma, or normal eyes. In the present study, CST-IOP and GAT-IOP showed better consistency in the OHT group, non-OHT group, and all subjects compared with NCT-IOP. The difference between CST-IOP and GAT-IOP was not significant between the two groups and was unrelated to CCT. These results were in agreement with Hong et al's study[31].

IOP increase may play a critical role in myopia development[32][34]. The association of elevated IOP and elongated AL or an increased myopic refractive error was found in children[1],[35][36]. In the current research, the difference between NCT-IOP and GAT-IOP and the difference between NCT-IOP and CST-IOP negatively correlated with AL in the non-OHT group. This result may be due to a positive correlation between AL and GAT-IOP or CST-IOP, which reduces the difference between NCT-IOP and GAT-IOP or CST-IOP. Interestingly, there was no significant relationship between AL and the difference between NCT-IOP and GAT-IOP in the OHT group. We cannot explain this result. It might be due to a small sample in our study. Several studies indicated that increased AL was associated with elevated IOP[1],[35][37], although longitudinal analysis suggested that IOP had essentially no relationship with myopia progression in school children[32][34]. Moreover, previous results suggested that increased scleral compliance due to AL prolongation may cause slightly lower IOP due to biased tonometry[34]. The relationship between IOP and AL is very confounded and needs further exploration.

There are two significant limitations in this study. First, the relationship between IOP, CCT, AL, and age has been challenging to resolve because of its cross-sectional nature. Although optic nerve structure analysis was performed in the current study, the long-term ocular course of children with OHT may provide additional insight regarding the relationship of IOP, AL, corneal material properties, and optic nerve structure. Second, small sample sizes may lead to selective bias in parameter measurement and incorrect estimation of the association between IOP and other ocular parameters.

In conclusion, the agreement of CST-IOP and GAT-IOP was significantly better than NCT-IOP and GAT-IOP in children with NCT-IOP of 22 mm Hg or more. Moreover, no significant relationship was found between the difference between CST-IOP and GAT-IOP and CCT, AL, and age in all groups. Both CCT thickening and AL shorting might contribute to the overestimation of NCT measurement. For children with NCT-IOP≥22 mm Hg, who cannot tolerate GAT, CST measurements can be used as an alternative.

Acknowledgments

Authors' contributions: Chen YH had full access to all the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. Concept and design: Chen YH and Hui YN. Acquisition, analysis, or interpretation of data: Li HG, Lin F, Li SY, Liu QH, Yin CG, Chen XY, Zhang XJ, Qu Y.

Foundation: Supported by Medical Science Research Project of Hebei Province in 2023 (No.20231818).

Conflicts of Interest: Li HG, None; Chen YH, None; Lin F, None; Li SY, None; Liu QH, None; Yin CG, None; Chen XY, None; Zhang XJ, None; Qu Y, None; Hui YN, None.

REFERENCES

  • 1.Jiang WJ, Wu JF, Hu YY, et al. Intraocular pressure and associated factors in children: the Shandong children eye study. Invest Ophthalmol Vis Sci. 2014;55(7):4128. doi: 10.1167/iovs.14-14244. [DOI] [PubMed] [Google Scholar]
  • 2.Krag S, Larsen D, Albertsen BK, Glerup M. Risk of ocular hypertension in children treated with systemic glucocorticoid. Acta Ophthalmol. 2021;99(8):e1430–e1434. doi: 10.1111/aos.14820. [DOI] [PubMed] [Google Scholar]
  • 3.Brusini P, Salvetat ML, Zeppieri M. How to measure intraocular pressure: an updated review of various tonometers. J Clin Med. 2021;10(17):3860. doi: 10.3390/jcm10173860. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Gandhi PD, Gürses-Özden R, Liebmann JM, Ritch R. Attempted eyelid closure affects intraocular pressure measurement. Am J Ophthalmol. 2001;131(4):417–420. doi: 10.1016/s0002-9394(00)00802-3. [DOI] [PubMed] [Google Scholar]
  • 5.Uzlu D, Akyol N, Türk A, Oruç Y. A comparison of three different tonometric methods in the measurement of intraocular pressure in the pediatric age group. Int Ophthalmol. 2020;40(8):1999–2005. doi: 10.1007/s10792-020-01375-8. [DOI] [PubMed] [Google Scholar]
  • 6.Bao FJ, Huang W, Zhu R, et al. Effectiveness of the goldmann applanation tonometer, the dynamic contour tonometer, the ocular response analyzer and the corvis ST in measuring intraocular pressure following FS-LASIK. Curr Eye Res. 2020;45(2):144–152. doi: 10.1080/02713683.2019.1660794. [DOI] [PubMed] [Google Scholar]
  • 7.Bilgeç MD, Atalay E, Sözer Ö, et al. The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes. Int Ophthalmol. 2020;40(4):849–857. doi: 10.1007/s10792-019-01248-9. [DOI] [PubMed] [Google Scholar]
  • 8.Ma JN, Wang Y, Hao WT, Jhanji V. Comparative analysis of biomechanically corrected intraocular pressure with corneal visualization Scheimpflug technology versus conventional noncontact intraocular pressure. Int Ophthalmol. 2020;40(1):117–124. doi: 10.1007/s10792-019-01159-9. [DOI] [PubMed] [Google Scholar]
  • 9.Ye C, Yu M, Lai G, Jhanji V. Variability of corneal deformation response in normal and keratoconic eyes. Optom Vis Sci. 2015;92(7):e149–e153. doi: 10.1097/OPX.0000000000000628. [DOI] [PubMed] [Google Scholar]
  • 10.Ramm L, Herber R, Spoerl E, et al. Intraocular pressure measurement using ocular response analyzer, dynamic contour tonometer, and scheimpflug analyzer corvis ST. J Ophthalmol. 2019;2019:3879651. doi: 10.1155/2019/3879651. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Chen YH, Rong HA, Liu W, Liu GH, Du B, Jin CC, Wei RH. Agreement of corrected intraocular pressure values between corvis ST and pentacam in patients with keratoconus, subclinical keratoconus, and normal cornea. Cornea. 2021;40(11):1426–1432. doi: 10.1097/ICO.0000000000002707. [DOI] [PubMed] [Google Scholar]
  • 12.Ziaei M, Gokul A, Vellara H, et al. Measurement of in vivo biomechanical changes attributable to epithelial removal in keratoconus using a noncontact tonometer. Cornea. 2020;39(8):946–951. doi: 10.1097/ICO.0000000000002344. [DOI] [PubMed] [Google Scholar]
  • 13.Eliasy A, Chen KJ, Vinciguerra R, et al. Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST. Exp Eye Res. 2018;175:98–102. doi: 10.1016/j.exer.2018.06.013. [DOI] [PubMed] [Google Scholar]
  • 14.Fan DSP, Chiu TYH, Congdon N, et al. Measurement of intraocular pressure with pressure phosphene tonometry in children. J Pediatr Ophthalmol Strabismus. 2011;48(3):167–173. doi: 10.3928/01913913-20100618-02. [DOI] [PubMed] [Google Scholar]
  • 15.Kageyama M, Hirooka K, Baba T, Shiraga F. Comparison of ICare rebound tonometer with noncontact tonometer in healthy children. J Glaucoma. 2011;20(1):63–66. doi: 10.1097/IJG.0b013e3181d12dc4. [DOI] [PubMed] [Google Scholar]
  • 16.Feng CS, Jin KW, Yi K, Choi DG. Comparison of intraocular pressure measurements obtained by rebound, noncontact, and goldmann applanation tonometry in children. Am J Ophthalmol. 2015;160(5):937–943.e1. doi: 10.1016/j.ajo.2015.07.029. [DOI] [PubMed] [Google Scholar]
  • 17.Brandt JD, Beiser JA, Kass MA, Gordon MO. Central corneal thickness in the ocular hypertension treatment study (OHTS) Ophthalmology. 2001;108(10):1779–1788. doi: 10.1016/s0161-6420(01)00760-6. [DOI] [PubMed] [Google Scholar]
  • 18.Leske MC. Factors for glaucoma progression and the effect of treatment. Arch Ophthalmol. 2003;121(1):48. doi: 10.1001/archopht.121.1.48. [DOI] [PubMed] [Google Scholar]
  • 19.Wang YX, Panda-Jonas S, Jonas JB. Optic nerve head anatomy in myopia and glaucoma, including parapapillary zones alpha, beta, gamma and delta: Histology and clinical features. Prog Retin Eye Res. 2021;83:100933. doi: 10.1016/j.preteyeres.2020.100933. [DOI] [PubMed] [Google Scholar]
  • 20.Chen YH, Hui YN. Commentary review on peripapillary morphological characteristics in high myopia eyes with glaucoma: diagnostic challenges and strategies. Int J Ophthalmol. 2021;14(4):600–605. doi: 10.18240/ijo.2021.04.18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kang JM, Tanna AP. Glaucoma. Med Clin N Am. 2021;105(3):493–510. doi: 10.1016/j.mcna.2021.01.004. [DOI] [PubMed] [Google Scholar]
  • 22.Bouhenni RA, Ricker I, Hertle RW. Prevalence and clinical characteristics of childhood glaucoma at a tertiary care children's hospital. J Glaucoma. 2019;28(7):655–659. doi: 10.1097/IJG.0000000000001259. [DOI] [PubMed] [Google Scholar]
  • 23.Javed A, Arshad MO, Nasir J, et al. Comparison between the values of intra-ocular pressure measured by Goldmann applanation tonometer and Air-Puff non-contact tonometer and their relationship with central corneal thickness. J Pak Med Assoc. 2022;72(1):149. doi: 10.47391/JPMA.11-1978. [DOI] [PubMed] [Google Scholar]
  • 24.Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK, 2nd, Wilson MR, Gordon MO. The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120(6):701–713. doi: 10.1001/archopht.120.6.701. [DOI] [PubMed] [Google Scholar]
  • 25.Mark HH. Corneal curvature in applanation tonometry. Am J Ophthalmol. 1973;76(2):223–224. doi: 10.1016/0002-9394(73)90164-5. [DOI] [PubMed] [Google Scholar]
  • 26.Kohlhaas M. Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol. 2006;124(4):471. doi: 10.1001/archopht.124.4.471. [DOI] [PubMed] [Google Scholar]
  • 27.Cheng X, Petsche SJ, Pinsky PM. A structural model for the in vivo human cornea including collagen-swelling interaction. J R Soc Interface. 2015;12(109):20150241. doi: 10.1098/rsif.2015.0241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Ferreira CCM, Tavares IM. Intraocular pressure and central corneal thickness in full-term newborns. Arquivos Brasileiros De Oftalmol. 2017;80(5):313–316. doi: 10.5935/0004-2749.20170076. [DOI] [PubMed] [Google Scholar]
  • 29.Manni G, Oddone F, Parisi V, Tosto A, Centofanti M. Progress in Brain Research. Amsterdam: Elsevier; 2008. Intraocular pressure and central corneal thickness; pp. 25–30. [DOI] [PubMed] [Google Scholar]
  • 30.Halkiadakis I, Gryparis A, Markopoulos I, Konstadinidou V, Zintzaras E, Tzakos M. Evaluation of Corvis ST tonometer with the updated software in glaucoma practice. Int J Ophthalmol. 2022;15(3):438–445. doi: 10.18240/ijo.2022.03.11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Hong JX, Xu JJ, Wei AJ, et al. A new tonometer—the corvis ST tonometer: clinical comparison with noncontact and goldmann applanation tonometers. Invest Ophthalmol Vis Sci. 2013;54(1):659. doi: 10.1167/iovs.12-10984. [DOI] [PubMed] [Google Scholar]
  • 32.Manny RE, Deng L, Crossnoe C, Gwiazda J. IOP, myopic progression and axial length in a COMET subgroup. Optom Vis Sci. 2008;85(2):97–105. doi: 10.1097/OPX.0b013e3181622633. [DOI] [PubMed] [Google Scholar]
  • 33.Manny RE, Mitchell GL, Cotter SA, Jones-Jordan LA, Kleinstein RN, Mutti DO, Twelker JD, Zadnik K. Intraocular pressure, ethnicity, and refractive error. Optom Vis Sci. 2011;88(12):1445–1453. doi: 10.1097/OPX.0b013e318230f559. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Li SM, Iribarren R, Li H, et al. Intraocular pressure and myopia progression in Chinese children: the Anyang Childhood Eye Study. Br J Ophthalmol. 2019;103(3):349–354. doi: 10.1136/bjophthalmol-2017-311831. [DOI] [PubMed] [Google Scholar]
  • 35.Wang PY, Chen SD, Liu YM, et al. Lowering intraocular pressure: a potential approach for controlling high myopia progression. Invest Ophthalmol Vis Sci. 2021;62(14):17. doi: 10.1167/iovs.62.14.17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Zhang DY, Wang LY, Jin L, Wen YY, Zhang XH, Zhang LY, Zhu H, Wang ZY, Yu X, Xie C, Tong JP, Shen Y. A review of intraocular pressure (IOP) and axial myopia. J Ophthalmol. 2022;2022:1–10. doi: 10.1155/2022/5626479. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Juliano J, Burkemper B, Lee J, et al. Longer axial length potentiates relationship of intraocular pressure and peripapillary vessel density in glaucoma patients. Invest Ophthalmol Vis Sci. 2021;62(9):37. doi: 10.1167/iovs.62.9.37. [DOI] [PMC free article] [PubMed] [Google Scholar]