Abstract
In order to evaluate the reliability of the ID ICP-MS method for the measurement of magnesium, zinc, and copper in human serum, we investigated the traceability, precision, trueness, and uncertainty of the method. This method traces the contents of magnesium, zinc, and copper in human serum to the standard materials NIST SRM3131a, SRM3168a, and SRM3114 respectively, thus completing the traceability to SI unit. The repeatability of this method for measuring magnesium, zinc, and copper in the human serum reference material GBW09152 was found to be 0.2%, 0.7%, and 0.6% (n = 9), respectively. The measurement, when employed to measure the magnesium, zinc, and copper in standard materials, had caused a maximum deviation of less than 0.88%, 1.35%, and 1.15%, respectively. The measurement results are within the stated uncertainty range of standard materials. The expanded uncertainties were 0.2 mg·kg−1, 0.04 mg·kg−1, and 0.08 mg·kg−1 (K = 2) for magnesium, zinc, and copper, respectively. Therefore, this method has high trueness, good reproducibility, and simple operation and is suitable for tracing the values of magnesium, zinc, and copper in human serum.
1. Introduction
Isotope dilution mass spectrometry, which has many advantages over conventional methods, is the only authoritative method that can directly provide trace and ultratrace values [1]. In common analysis methods, the accuracy of the results may be affected by the loss of the elements to be tested in the process of sample pretreatment, matrix effects, and instrument signal drift [2–5]. Isotope dilution mass spectrometry is the method which measures only the isotope abundance ratio in the sample. The abundance ratio becomes a constant value when the concentration of the isotopic spike is added to the sample and reaches the equilibrium with the absence of external contamination, and the sample loss in the process of sample separation and concentration does not affect the abundance ratio [6]. Isotope dilution method can greatly eliminate interference and errors caused by sample pretreatment and has thus been confirmed by the Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology (CCQM) as one of the five methods with absolute measurement properties [7, 8]. This isotope dilution mass spectrometry originated from the UK Government Chemist Laboratory (GCL) [9, 10]. We have established isotope dilution mass spectrometry for the detection of potassium and selenium in human serum using this principle [11]. Compared with conventional isotope dilution mass spectrometry, it has the advantage of not having to calibrate the concentration of the isotopic spike during the standardization process. It can determine the chemical composition in serum only by means of measuring the selected pair of isotope abundance ratios, thus reducing the requirements on the performance of the mass spectrometry instrument [12–17].
Trace elements of magnesium, zinc, and copper in the human body are closely related to people's health [18–22]. Currently, there is no reference method for measuring zinc and copper in serum by using isotope dilution mass spectrometry in the Joint Committee on Traceability on Laboratory Medicine (JCTLM) list. Only the measurement of the concentration of magnesium in serum by using traditional isotope dilution mass spectrometry is available [23–25]. A precise method for analyzing magnesium, zinc, and copper in human serum by using two-step isotope dilution mass spectrometry was established in this laboratory. The repeatability of this method for measuring magnesium, zinc, and copper in the human serum reference material GBW09152 was found to be 0.2%, 0.7%, and 0.6% (n = 9), respectively. The expanded uncertainties were 0.2 mg·kg−1, 0.04 mg·kg−1, and 0.08 mg·kg−1 (K = 2) for magnesium, zinc, and copper, respectively. With the quality of easy operation, high trueness, and good reproducibility, this method is suitable as a reference method for measuring magnesium, zinc, and copper elements in human serum.
2. Materials and Methods
2.1. Materials and Reagents
The laboratory is a class 100,000 cleanroom. The experimental water was provided by a water purification system: Milli-Q Advantage (Millipore, USA). The nitric acid (Ultrapure-BVIII) and hydrochloric acid (MOS grade) used in the experiment were produced by the Beijing Institute of Chemical Reagents, China. The microanalytical balance used for sample weighing was Mettler Toledo XS205 (Switzerland), and the sample analysis was performed on an ICP mass spectrometer: ELAN DRC-e (PerkinElmer, USA). The standard materials used for method tracing were magnesium (Mg) standard solution SRM3131a (NIST, USA), copper (Cu) standard solution SRM3114 (NIST, USA), and zinc (Zn) standard solution SRM3168a (NIST, USA). The isotope spikes used in the method were 25Mg (assay: 97%), 65Cu (assay: 99%), and 67Zn (assay: 94%), concentrated isotopes from the Oak Ridge National Laboratory, USA. Trueness verification was performed by using inorganic components in frozen human serum GBW09152 (National Institute of Metrology, China), ERM-DA120a (LGC, UK) and electrolytes in frozen human serum SRM956d (NIST, USA).
2.2. Instrument Parameters
The instrument parameters for this research are listed in Table 1.
Table 1.
Instrument parameters.
| Parameter | Mg | Cu | Zn |
|---|---|---|---|
| ICP RF power (W) | 1100 | 1100 | 1100 |
| Gas flows (L/min) | 0.97 | 0.97 | 0.97 |
| Lens voltages (V) | 6.75 | 6.75 | 6.75 |
| Analog stage voltage (V) | −1700 | −1700 | −1700 |
| Pulse stage voltage (V) | 900 | 900 | 900 |
| Dwell time per AMU | 2 ms | 2 ms | 2 ms |
| Scan mode | Peak hopping | Peak hopping | Peak hopping |
| Sweeps/reading | 25 | 25 | 25 |
| Readings/replicate | 15 | 20 | 25 |
| Replicates | 15 | 20 | 25 |
| Cell Gas A: CH4 (mL/min) | 1.4 | 0 | 0.45 |
| RP q: Ar (mL/min) | 0.8 | 0.25 | 0.25 |
2.3. Methods
-
(1)
A two-step dilution of NIST SRM3131a was performed by using the weighing method in a ratio of approximately 1 : 20, and a three-step dilution of NIST SRM3168a and NIST SRM3114 in a ratio of approximately 1 : 20. The final dilution must be prepared on the day of the experiment.
-
(2)
The concentrated isotopic metal chips of 25Mg, 65C, and 67Zn were dissolved with BVIII grade nitric acid and diluted to an appropriate concentration with ultrapure water, followed by a two-step dilution on the day of the experiment.
-
(3)
A mixed solution of magnesium standard solution and 25Mg isotope diluent, a mixed solution of zinc standard solution and 67Zn isotope diluent, and a mixed solution of copper standard solution and 65Cu isotope diluent were prepared, respectively, by using the weighing method. A mixed solution of the serum sample and 25Mg isotope diluent, a mixed solution of the serum sample and 67Zn isotope diluent, and a mixed solution of the serum sample and 65Cu isotope diluent were also prepared, respectively, by using the weighing method. The isotopic ratio (24Mg/25Mg or 66Zn/67Zn or 63Cu/65Cu) in the mixture solution was close to 1, and at the same time, the cps signal intensities of the corresponding isotopes in the mixture of the standard solution and the isotope diluent as well as in the mixture of the serum sample and the isotope diluent were close.
-
(4)
Solutions of magnesium, copper, and zinc with appropriate concentrations were prepared on the day of the experiment so that the signal intensity of 24Mg, 66Zn, and 63Cu in the solution is consistent with that of the corresponding isotopes in the mixed solution in (3).
-
(5)
Solutions of 25Mg, 67Zn, and 65Cu diluents with appropriate concentrations were prepared on the day of the experiment so that the signal intensity of the isotopes in the solution is consistent with that of the corresponding isotopes in the mixed solution in (3).
- (6)
In this formula, Cz is the concentration of the standard solution, mY is the mass of the enriched isotope added to the serum sample, mX is the mass of the serum sample added to the mixture of the serum sample and the enriched isotope, mYc is the mass of the enriched isotope added to the standard solution, and mzc is the mass of the standard solution added to the mixture of the standard solution and the enriched isotope. Rz is the isotope ratio of 24Mg/25Mg or 66Zn/67Zn or 63Cu/65Cu in the standard solution, and RY is the isotope ratio of 24Mg/25Mg (66Zn/67Zn or 63Cu/65Cu) in the enriched isotope. RB is the isotope ratio of 24Mg/25Mg (66Zn/67Zn or 63Cu/65Cu) in the mixture of serum sample and the enriched isotope, and RBc the isotope ratio of 24Mg/25Mg (66Zn/67Zn or 63Cu/65Cu) in the mixture of standard solution and the enriched isotope. CB is a blank in the measurement process.
It was found that the fluctuation of the isotope ratio in the standard solution or the enriched isotope had little impact on the final results, while the measurement fluctuation of the isotope ratio in the mixed solution would lead to changes in the final results. Therefore, the mixed solution of the standard solution and the enriched isotope and the mixed solution of the serum sample and the enriched isotope must be alternately measured six times so as to reduce the error introduced by instrument measurement drift.
3. Results and Discussion
3.1. Determination of Instrument Measurement Conditions
In this experiment, the dynamic reaction cell mode of inductively coupled plasma mass spectrometry (ICP-MS) was used to eliminate interference. When analyzing magnesium, interference from NaH was severe, and interference from Ca++ and LiO may also exist. The condition optimization was aiming to obtain the best analysis effect when analyzing 24Mg, and the appropriate flow rate of the reaction gas CH4 and argon. When analyzing zinc, interference from SO2, ClO2, and ArP may exist, and interference on 67Zn was more likely to occur. The condition optimization was aimed at obtaining the best analysis effect when analyzing 67Zn and the appropriate flow rate of the reaction gas CH4 and argon. When analyzing copper, interference from SO2 and PO2 may exist, while all these interferences can be ignored in actual analysis. Therefore, this experiment analyzed copper in blood under standard mode.
3.2. Deduction of Signal Background in the Experimental Method
Solutions of serum magnesium with 6 concentration gradients ranging from 0.5 μg/ml to 9 μg/ml were measured. The response signals of 24Mg and 25Mg are linearly related to the concentration range of magnesium (see Figures 1(a) and 1(b)), with a linear correlation coefficient (r) of 0.99999. However, the ratio of 24Mg response signal to 25Mg response signal was not constant but gradually tended to be constant with the increase of concentration (see Figure 1(c)). This is because the straight lines in Figures 1(a) and 1(b) have a nonzero intercept, indicating the presence of a blank response signal for 24Mg and 25Mg. As the concentration becomes smaller, the relative proportion of the response signal becomes higher, and the impact on the ratio of 24Mg to 25Mg response signal is more significant. Therefore, this blank response signal cannot be ignored. After subtracting the blank response signal from the measured impact signal, a constant value is obtained for the ratio of the 24Mg to 25Mg·g response signals. The same is true for the determination of 66Zn/67Zn and 63Cu/65Cu, so in this experiment, the response signals used are all values after deducting the blank signal.
Figure 1.

(a) Relationship between the response signal of 24Mg and magnesium concentration; (b) relationship between the response signal of 25Mg and magnesium concentration; (c) relationship between the ratio of 24Mg to 25Mg response signals and sample concentration.
3.3. Effect of Solution Reaction System on Measurement Precision
It was found in the experiment that using a 0.02% hydrochloric acid system can improve the stability of 66Zn/67Zn measurement. Therefore, the determination of zinc in serum was indeed carried out in a 0.02% hydrochloric acid system.
3.4. Process Blank (LOB) and Detection Limits (LOD)
While preparing the mixture of serum sample and the enriched isotope, an appropriate amount of 25Mg (67Zn or 65Cu) solution was taken into the blank sample tube as a process blank, so that the 24Mg/25Mg (66Zn/67Zn or 63Cu/65Cu) in the process blank was approximately equal to 2. The process blank was determined along with the samples in the same batch. The process blanks for magnesium, zinc, and copper in serum were 0.5 mg/kg, 0.09 mg/kg, and 0.010 mg/kg, respectively. When the confidence interval of 95% was determined, the detection limits of magnesium, zinc, and copper in serum were 0.7 mg/kg, 0.11 mg/kg, and 0.016 mg/kg, respectively.
3.5. Method Precision
The method precision was the relative standard deviation of the measurement results of 6 bottles of human serum reference materials. The experiments were carried out in 2 consecutive days, with 3 bottles each day, and 3 parallel for each bottle. The results are shown in Table 2. The precision for magnesium, zinc, and copper in different concentrations of human serum reference materials was lower than 0.3%, 0.9%, and 0.6%, respectively.
Table 2.
Repeatability of magnesium, zinc, and copper in human serum by ID ICP-MS.
| Mg (mg·kg−1) | Zn (mg·kg−1) | Cu (mg·kg−1) | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Level 1 | Level 2 | Level 3 | Level 1 | Level 2 | Level 3 | Level 1 | Level 2 | ||
| Day 1 | 1# | 18.0 | 23.3 | 28.7 | 0.625 | 0.831 | 1.597 | 1.06 | 1.61 |
| 18.0 | 23.3 | 28.7 | 0.627 | 0.834 | 1.583 | 1.06 | 1.62 | ||
| 18.1 | 23.2 | 28.9 | 0.646 | 0.835 | 1.610 | 1.06 | 1.62 | ||
| 2# | 18.0 | 23.3 | 28.8 | 0.631 | 0.834 | 1.597 | 1.07 | 1.60 | |
| 18.0 | 23.3 | 28.8 | 0.627 | 0.825 | 1.594 | 1.06 | 1.62 | ||
| 18.0 | 23.2 | 29.0 | 0.634 | 0.832 | 1.595 | 1.06 | 1.62 | ||
| 3# | 18.0 | 23.3 | 28.7 | 0.628 | 0.818 | 1.607 | 1.06 | 1.61 | |
| 18.0 | 23.3 | 28.7 | 0.634 | 0.819 | 1.599 | 1.06 | 1.63 | ||
| 18.0 | 23.3 | 28.9 | 0.625 | 0.826 | 1.596 | 1.06 | 1.61 | ||
|
| |||||||||
| Day 2 | 4# | 18.0 | 23.3 | 28.7 | 0.644 | 0.828 | 1.601 | 1.05 | 1.60 |
| 18.0 | 23.2 | 28.7 | 0.634 | 0.827 | 1.606 | 1.05 | 1.60 | ||
| 18.0 | 23.3 | 28.7 | 0.633 | 0.825 | 1.616 | 1.05 | 1.60 | ||
| 5# | 18.0 | 23.2 | 28.7 | 0.634 | 0.827 | 1.599 | 1.05 | 1.61 | |
| 18.0 | 23.3 | 28.7 | 0.632 | 0.830 | 1.596 | 1.06 | 1.60 | ||
| 18.0 | 23.3 | 28.7 | 0.629 | 0.826 | 1.600 | 1.06 | 1.62 | ||
| 6# | 18.0 | 23.3 | 28.7 | 0.628 | 0.823 | 1.606 | 1.06 | 1.60 | |
| 18.0 | 23.2 | 28.7 | 0.633 | 0.830 | 1.591 | 1.06 | 1.59 | ||
| 18.0 | 23.3 | 28.8 | 0.632 | 0.825 | 1.602 | 1.06 | 1.60 | ||
|
| |||||||||
| s | 0.03 | 0.03 | 0.09 | 0.006 | 0.005 | 0.008 | 0.005 | 0.010 | |
|
| |||||||||
| Avg | 18.0 | 23.3 | 28.8 | 0.632 | 0.828 | 1.600 | 1.06 | 1.61 | |
|
| |||||||||
| CV | 0.2% | 0.1% | 0.3% | 0.9% | 0.6% | 0.5% | 0.4% | 0.6% | |
3.6. Method Trueness
Magnesium, zinc, and copper in standard substances NIST956D, ERM-DA120a, and GBW09152 were analyzed by using isotope dilution mass spectrometry. Parallel analysis was conducted three times a day for three consecutive days, and the results were good, as shown in Table 3.
Table 3.
Analysis of standard reference material by two-way ID ICP-MS.
| Mg (mg·kg−1) | Zn (mg·kg−1) | Cu (mg·kg−1) | |||||||
|---|---|---|---|---|---|---|---|---|---|
| NIST956d | GBW | ERM | GBW | ERM | GBW | ||||
| Level 1 | Level 2 | Level 3 | 09152 | -DA120a | 09152 | -DA120a | 09152 | ||
| Day 1 | Test 1 | 34.98 | 22.91 | 10.49 | 20.59 | 0.649 | 1.156 | 1.126 | 1.064 |
| Test 2 | 34.84 | 22.70 | 10.38 | 20.54 | 0.664 | 1.148 | 1.131 | 1.069 | |
| Test 3 | 34.60 | 22.72 | 10.41 | 20.61 | 0.660 | 1.140 | 1.126 | 1.073 | |
|
| |||||||||
| Day 2 | Test 1 | 35.08 | 22.51 | 10.37 | 20.56 | 0.653 | 1.140 | 1.145 | 1.082 |
| Test 2 | 34.60 | 22.95 | 10.43 | 20.63 | 0.652 | 1.156 | 1.140 | 1.077 | |
| Test 3 | 34.66 | 22.83 | 10.51 | 20.52 | 0.650 | 1.144 | 1.143 | 1.080 | |
|
| |||||||||
| Day 3 | Test 1 | 34.74 | 22.59 | 10.36 | 20.50 | 0.665 | 1.147 | 1.131 | 1.069 |
| Test 2 | 34.67 | 22.66 | 10.25 | 20.60 | 0.658 | 1.137 | 1.130 | 1.067 | |
| Test 3 | 35.04 | 22.73 | 10.28 | 20.56 | 0.651 | 1.156 | 1.135 | 1.072 | |
|
| |||||||||
| Avg | 34.80 | 22.73 | 10.39 | 20.57 | 0.656 | 1.147 | 1.13 | 1.072 | |
|
| |||||||||
| cv | 0.5% | 0.6% | 0.8% | 0.2% | 0.9% | 0.7% | 0.6% | 0.6% | |
|
| |||||||||
| Certified values | 34.96 ± 0.24 | 22.83 ± 0.16 | 10.30 ± 0.08 | 20.75 ± 0.44 | 0.658 ± 0.033 | 1.132 ± 0.056 | 1.130 ± 0.033 | 1.085 ± 0.044 | |
|
| |||||||||
| Bias (%) | 0.46% | 0.42% | −0.84% | 0.88% | 0.33% | −1.35% | −0.37% | 1.15% | |
3.7. Uncertainty Evaluation
In this research, the uncertainty caused by factors such as the experimental reagents, the samples, the laboratory environments, the solution preparation, the instrument measurement, and the data processing has been evaluated as the source of uncertainty in the measurement process. It can be concluded that by evaluating the uncertainty of each parameter in formula (1), the uncertainty caused by each factor in the measurement process can be fully included. Each parameter in formula (1) is an independent parameter, and the uncertainty uc(y) related to measurement is calculated as follows:
| (2) |
The formula of the sensitivity coefficient (∂f/∂xi) of each parameter in formula (1) is as follows:
| (3) |
Therefore,
| (4) |
This research has evaluated the uncertainty of measurement and Table 4 is the source of uncertainty for measuring the parameters. The uncertainty of the measurement of Type A is the experimental standard deviation of the 6 repeated instrument measurements, taking the worst result in the experiment as the evaluation data. The uncertainty of the measurement of Type B is the uncertainty of solution preparation, which is synthesized from the uncertainty resulting from the electronic balance calibration and the uncertainty resulting from weighing.
Table 4.
Sources of uncertainty in the determination.
| Sources of uncertainty | Mg | Zn | Cu | |||
|---|---|---|---|---|---|---|
| GBW09152 | GBW09152 | GBW09152 | ||||
| Value (xi) | u c (xi) | Value (xi) | u c (xi) | Value (xi) | u c (xi) | |
| Type A uncertainties | ||||||
| RY | 0.014 | 0.000 | 0.034 | 0.000 | 0.004 | 0.000 |
| Rz | 6.667 | 0.021 | 6.208 | 0.014 | 2.107 | 0.003 |
| RBc | 0.920 | 0.002 | 1.041 | 0.003 | 1.080 | 0.004 |
| RB | 0.885 | 0.003 | 1.004 | 0.003 | 1.049 | 0.002 |
| CB (mg·kg−1) | 0.5 | 0.3 | 0.09 | 0.05 | 0.010 | 0.005 |
| Type B uncertainties | ||||||
| Cz (mg·kg−1) | 19.398 | 0.003 | 1.290 | 0.002 | 1.008 | 0.002 |
| mzc (mg) | 3.0 × 102 | 0.040 | 2.6 × 102 | 0.040 | 5.4 × 102 | 0.040 |
| mYc (mg) | 5.5 × 102 | 0.040 | 2.0 × 102 | 0.040 | 4.0 × 102 | 0.040 |
| mY (mg) | 3.4 × 102 | 0.040 | 200 × 102 | 0.040 | 4.0 × 102 | 0.040 |
| mx (mg) | 2.0 × 102 | 0.040 | 300 × 102 | 0.040 | 4.8 × 102 | 0.040 |
| Combined types A and B | — | 0.25 | — | 0.045 | — | 0.011 |
| Degrees of freedom (Veff) | 8 | 8 | 8 | |||
| Coverage factor (k) | — | 2 | — | 2 | — | 2 |
| Measured value (mg·kg−1) | 20.57 | — | 1.147 | — | 1.072 | — |
| Expanded uncertainty (U(), k = 2) (mg·kg−1) | — | 0.2 | — | 0.04 | — | 0.08 |
4. Conclusions
This isotope dilution mass spectrometry established in our laboratory with the quality of easy operation, high trueness, and good reproducibility was used to accurately analyze the contents of magnesium, zinc, and copper in human serum. In the isotope dilution mass spectrometry determination process, there is no need to measure the accurate amount of isotope, the enriched isotope, which reduces the requirements for mass spectrometry instruments [26, 27]. This method is suitable as a reference method for assigning values of magnesium, zinc, and copper in human serum.
Acknowledgments
This work was supported by the Science and Technology Research and Development Project of Langfang, China (no. 2023011027).
Data Availability
The data used to support the findings of this study are included within the article.
Conflicts of Interest
The authors declare that there are no conflicts of interest regarding the publication of this article.
References
- 1.Zhao M.-T. The characteristics of isotope dilution mass spectrometry. Journal of Chinese Mass Spectrometry Society . 2004;25:167–168. [Google Scholar]
- 2.Grote-Koska D., Klauke R., Brand K., Schumann G. Reference measurement procedure for the determination of electrolytes in human blood via ICP-OES measurement. Metrologia . 2018;55(2):pp245–253. doi: 10.1088/1681-7575/aaaa3f. [DOI] [Google Scholar]
- 3.Liang S., Wang X., Wu H., Sun H. Determination of 9 heavy metal elements in sediment by icp-ms using microwave digestion for sample preparation. Spectroscopy and Spectral Analysis . 2012;32(3):809–812. [PubMed] [Google Scholar]
- 4.Thienpont L. M., Van Nuwenborg J. E., Reinauer H., Stöckl D. Validation of candidate reference methods based on ion chromatography for determination of total sodium, potassium, calcium and magnesium in serum through comparison with flame atomic emission and absorption spectrometry. Clinical Biochemistry . 1996;29(6):501–508. doi: 10.1016/s0009-9120(96)00090-2. [DOI] [PubMed] [Google Scholar]
- 5.Barbagallo R. P., Boley N., Holcombe G., et al. Production and certification of four frozen human serum certified reference materials containing creatinine and electrolytes. Annals of Clinical Biochemistry . 2008;45(2):160–166. doi: 10.1258/acb.2007.007126. [DOI] [PubMed] [Google Scholar]
- 6.Schaumlöffela D., Łobińskiab R. Isotope dilution technique for quantitative analysis of endogenous trace element species in biological systems. International Journal of Mass Spectrometry . 2005;242(2-3):217–223. doi: 10.1016/j.ijms.2004.11.020. [DOI] [Google Scholar]
- 7.Kessler A. Mass spectrometry–a key technique for traceability in clinical chemistry. TrAC, Trends in Analytical Chemistry . 2016;84:74–79. doi: 10.1016/j.trac.2016.03.017. [DOI] [Google Scholar]
- 8.Ng S. Y., Zou A., Sim L. P., et al. High accuracy measurements of trace elements in drinking water by exact matching isotope dilution mass spectrometry (IDMS) International Journal of Mass Spectrometry . 2012;321-322:19–24. doi: 10.1016/j.ijms.2012.05.008. [DOI] [Google Scholar]
- 9.Sargent M., Harte R., Harrington C. Guidelines for Achieving High Accuracy in Isotope Dilution Mass Spectrometry (IDMS) London, UK: RSC Analytical Methods Committee; 2002. [Google Scholar]
- 10.Van Nevel L., Aregbe Y., Taylor P. D. P. CCQM-K14 Key Comparison Calcium Amount Content in Serum . Geel, Belgium: Metrologia; 2004. Internal IRMM report number GE/R/IM/17/03. [Google Scholar]
- 11.Yang J., Chi L., Li S. High-accuracy determination of potassium and selenium in human serum by two-step isotope dilution ICPMS. Journal of Analytical Methods in Chemistry . 2021;2021:9. doi: 10.1155/2021/5533631.5533631 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Vogeser M., Seger C. Mass spectrometry methods in clinical diagnostics –state of the art and perspectives. TrAC, Trends in Analytical Chemistry . 2016;84:1–4. doi: 10.1016/j.trac.2016.04.018. [DOI] [Google Scholar]
- 13.Li S., Wang J. Measurement of calcium in human serum by dynamic reaction cell and two-way ID–ICP–MS. Chemical Analysis and Meterage . 2015;24:12–15. [Google Scholar]
- 14.Zhao M.-T., Wang J., Lu B., Ji J., Zhang Y. Application of isotope dilution mass spectrometry to research of certified reference materials. Accreditation and Quality Assurance . 2002;7(3):111–114. doi: 10.1007/s00769-001-0429-8. [DOI] [Google Scholar]
- 15.Boulyga S. F., Klötzli U., Stingeder G., Prohaska T. Optimization and application of ICPMS with dynamic reaction cell for precise determination of 44Ca/40Ca isotope ratios. Analytical Chemistry . 2007;79(20):7753–7760. doi: 10.1021/ac0711790. [DOI] [PubMed] [Google Scholar]
- 16.Simpson L. A., Hearn R., Merson S., Catterick T. A comparison of double-focusing sector field ICP-MS, ICP-OES and octopole collision cell ICP-MS for the high-accuracy determination of calcium in human serum. Talanta . 2005;65(4):900–906. doi: 10.1016/j.talanta.2004.08.012. [DOI] [PubMed] [Google Scholar]
- 17.Yu L. L., Davis W. C., Nuevo Ordonez Y., Long S. E. Fast and accurate determination of K, Ca, and Mg in human serum by sector field ICP-MS. Analytical and Bioanalytical Chemistry . 2013;405(27):8761–8768. doi: 10.1007/s00216-013-7320-4. [DOI] [PubMed] [Google Scholar]
- 18.Dent A., Selvaratnam R. Measuring magnesium-Physiological, clinical and analytical perspectives. Clinical Biochemistry . 2022;105-106:1–15. doi: 10.1016/j.clinbiochem.2022.04.001. [DOI] [PubMed] [Google Scholar]
- 19.Tuttle A., Fitter S., Hua H., Moussavi K. The effects of magnesium coadminstration during treatment of hypokalemia in the emergency department. Journal of Emergency Medicine . 2022;63(3):399–413. doi: 10.1016/j.jemermed.2022.06.007. [DOI] [PubMed] [Google Scholar]
- 20.Riondato J., Vanhaecke F., Moens L., Dams R. Determination of trace and ultratrace elements in human serum with a double focusing magnetic sector inductively coupled plasma mass spectrometer. Journal of Analytical Atomic Spectrometry . 1997;12(9):933–937. doi: 10.1039/a701652b. [DOI] [Google Scholar]
- 21.Zhang L., Shao J., Tan S.-W., Ye H.-P., Shan X.-Y. Association between serum copper/zinc ratio and lung cancer: a systematic review with meta-analysis. Journal of Trace Elements in Medicine & Biology . 2022;74 doi: 10.1016/j.jtemb.2022.127061.127061 [DOI] [PubMed] [Google Scholar]
- 22.Ghazizadeh H., Yaghooti-Khorasani M., Darroudi S., et al. Evaluation of the association between the Healthy Eating Index and the level of serum and dietary intake of copper and zinc. Obesity Medicine . 2020;19 doi: 10.1016/j.obmed.2020.100277.100277 [DOI] [Google Scholar]
- 23.Yan Y., Zhang C., Zhang J., et al. Measurement of serum sodium and magnesium by inductively coupled plasma mass spectrometry with aluminum as internal standard. Clinical Laboratory . 2016;62:719–725. doi: 10.7754/clin.lab.2015.150140. [DOI] [PubMed] [Google Scholar]
- 24.Kramer U., Kress M., Reinauer H., Spannagl M., Kaiser P. Candidate Reference Measurement Procedures for Chloride, Potassium, Sodium, Calcium, Magnesium, and Lithium by Inductively Coupled Plasma (Isotope Dilution) Sector Field Mass Spectrometry (ICP-(ID) SFMS) in Serum. Clinical Laboratory . 2013;59:1017–1029. doi: 10.7754/clin.lab.2012.120902. [DOI] [PubMed] [Google Scholar]
- 25.Dewitte K., Stöckl D., Van de Velde M., Thienpont L. M. Evaluation of intrinsic and routine quality of serum total magnesium measurement. Clinica Chimica Acta . 2000;292(1-2):55–68. doi: 10.1016/s0009-8981(99)00203-x. [DOI] [PubMed] [Google Scholar]
- 26.Xu C., Ding Y., Leung H. W., Tong B. M. K., Shin R. Y., Lee T. K. Development of high accuracy methods for the certification of calcium, iron, magnesium and potassium in human serum. Journal of Trace Elements in Medicine & Biology . 2017;40:61–66. doi: 10.1016/j.jtemb.2016.12.007. [DOI] [PubMed] [Google Scholar]
- 27.Vassileva E., Wysocka I., Betti M. Reference measurements for cadmium, copper, mercury, lead, zinc and methyl mercury mass fractions in scallop sample by isotope dilution inductively coupled plasma mass spectrometry. Microchemical Journal . 2014;116:197–205. doi: 10.1016/j.microc.2014.05.011. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data used to support the findings of this study are included within the article.