Abstract
Portable artificial lung systems are under development, but there are few technologies available that adjust the carbon dioxide (CO₂) removal in response to changes in patient metabolic needs. Our work describes the second generation of a CO₂ based portable servoregulation system that automatically adjusts CO2 removal in artificial lungs (ALs).
Four adult sheep (68±14.3 kg) were used to test the servoregulator. The servoregulator controlled air sweep flow through the lung to meet a target exhaust gas CO2 (tEGCO2) level in normocapnic and hypercapnic (PaCO2 >60mmHg) conditions at varying flow rates (0.5–1.5 L/min) & at tEGCO2 levels of 10, 20 & 40 mmHg.
In hypercapnic sheep, average post-AL blood pCO2 values were 22.4±3.6 mmHg for tEGCO2 of 10 mmHg, 28.0±4.1 mmHg for tEGCO2 of 20 mmHg and 40.6±4.8 mmHg for tEGCO2 of 40 mmHg. The controller successfully and automatically adjusted the sweep gas flow to rapidly (<10 min) meet the target tEGCO2 level when challenged with changes in inlet blood flow or target EGCO2 levels for all animals.
These in vivo data demonstrate an important step toward portable artificial lungs that can automatically modulate CO2 removal and allow for substantial changes in patient activity or disease status in ambulatory applications.
Keywords: Servoregulation, Portable Artificial Lung, CO2 Removal
Introduction
Both end-stage lung disease (ESLD) and acute respiratory distress syndrome (ARDS) have a profound impact on the healthcare system, each affecting more than 100,000 Americans per year.1–3 Artificial lung (AL) and extracorporeal membrane oxygenation (ECMO) based CO2 removal promptly relieves the symptoms of ESLD and can serve as a bridge to recovery by enabling the lung to heal in acute cases and as a bridge to transplant in chronic cases.4–7
Traditionally, femoral cannulation for ECMO immobilizes patients, but with the advent of dual lumen internal jugular vein cannulation for veno-venous (VV) ECMO, ambulation is possible.8 The central sport model can be used via innominate or subclavian artery cannulation. This provides the ability for physical therapy and ambulation while patients await lung transplantation.9 Ambulatory ECMO systems with the AL, blood pump, controller, battery, and oxygen cylinder on a wheeled pole enable limited mobility and exercise.10–12 Because ambulatory ECMO provides a more active physiotherapy for patients, this has led to decreased length of stay in post-transplant ICU, leading to lower costs for the hospital and patient.13 Additionally, ambulatory ECMO has been shown to have improved outcomes in critically ill patients.14 Despite these promising results, many patients are still restricted to ICU care due to several logistical challenges that arise from developing an ambulatory ECMO program.15
A CO2 removal system that automatically regulates CO2 to meet dynamic physiological needs can simultaneously minimize logistical issues of an ambulatory ECMO program, while improving patient stability. Traditional ECMO systems risk removing insufficient CO2 during periods of activity (resulting in dyspnea and increased minute ventilation) or excessive CO2 during periods of inactivity (resulting in hypocapnia and blood alkalosis).16,17 Without CO2 regulation, patient comfort, activity, and quality of life will thus be severely limited when used for long-term ambulatory ECMO.18 The Quantum Ventilation Lite is a recently developed semi portable ECMO control system that can make automatic ventilatory changes but has not yet been tested on patients in the current literature.19
Our lab previously developed a subclavian arterio-venous (AV) access method for an ovine model. The published data show that varying sweep flow can drastically impact CO2 removal and sweep flows up to 15 L/min can significantly reduce PaCO2 from 74 to 49 mmHg using a blood flow of 1.0 L/min.20 Based on this established principle, we developed and performed in vitro and in vivo testing of the first-generation benchtop servoregulation system (Gen 1) that automatically adjusts CO2 removal in ALs under various metabolic conditions. It does this by modulating sweep gas flow rate based on exhaust gas CO2 partial pressure (EGCO2) as a representation for arterial partial pressure of arterial CO2 (PaCO2).21 The Gen 1 system rapidly and effectively adjusted to maintain the target EGCO2, which was reflected in the PaCO2.
Though functional and effective, the previous system lacked a unitary enclosure, was wall-powered, utilized various large components, used desiccant dehumidifiers which required frequent replacements, and incorporated a loud volumetric pump. These characteristics preclude its use in portable applications. However, it successfully proved the sweep gas volume and EGCO2 servoregulation method which the portable “Gen 2” model described in this paper uses. We have recently developed a portable Gen 2 version of this control system that weighs less than 2 kg and can operate for up to 12 hours on a single battery charge.22 This manuscript describes the use of this portable Gen 2 model in vivo with 4 sheep to assess its ability to successfully modulate EGCO2 in vivo under hypercapnic conditions.
Methods
Servoregulation Principle
A diagram of the CO2 based sweep gas servoregulation controller is shown in Figure 1. Using a Proportional-Integral-Derivative (PID) feedback controller, the system modulates pump-driven sweep air flow through an AL in order to attain a desired EGCO2. The PID feedback controller operates in a continuous cycle, reading the current value of EGCO2 from the CO2 sensor and comparing it to the user-specified target EGCO2 value (tEGCO2). If the tEGCO2 is less than the current EGCO2, power to the air blower is increased, which increases sweep gas flow through the AL. If the tEGCO2 is greater than the current EGCO2, the controller decreases power to the air pump, thus reducing the sweep gas flow through the AL. The proportional, integrative, and differential parameters determine exact pump power based on target and measured EGCO2 values. Properly tuned PID controllers are capable of responding rapidly to changes in equilibrium and can do so with minimal oscillation.
Figure 1. Flow path diagram of servoregulator system and components.

Blue arrow is exhaust and gray arrow is sampled gas to ambient air.
Flow Path and System Design
The components of the servoregulator system were selected to minimize size, power consumption, and noise with a view to facilitating integration into future ambulatory and wearable AL systems (Figure 1).
A WM7040–24V blower (Haoson, China) takes in ambient air, providing up to 20 L/min against the resistance of the AL and circuit. The blowers individually measure 70×70×40 mm in size and offer a quiet operation (~45 dBA). Only one blower is used at a time, with check valves to ensure unidirectional air flow. The blower-driven air passes through a flow sensor (Sensirion, Germany SFM3400-AW) before being routed into the AL. A differential pressure sensor (Honeywell, Charlotte, NC ABP2) is connected across the AL to monitor gas-side pressure. The exhaust gases leaving the AL are routed through a water trap and then expire into the atmosphere.
Side-stream capnography was used to measure EGCO2. The sampling circuit includes a TCS Micropump, (England, D250BLZ-V) to sample the exhaust gases. To remove moisture, the sampled gases are passed through a moisture trap (CO2 Meter Inc., Ormand beach, FL CM-0103), and 24 inches of tubing (Nafion, Lakewood, NJ, ME-110–24BB). The dried air then enters the CO2 sensor assembly (heated to increase dew point) containing a CO2 sensor (SprintIR-W20%, Gas Sensing Solutions Ltd., Scotland) before being exhausted into the ambient atmosphere.
Hollow fiber ALs typically accumulate moisture at the outlet of fiber bundles when moisture from plasma diffuses across the microporous membrane. Some of that moisture condenses and if enough condensation forms it can affect the CO2 clearance ability of the lung over time.23 To address this issue, the Gen 2 servoregulation system continuously monitors the fluidic resistance of the gas side of the AL. When the resistance has significantly increased, indicating moisture accumulation on the gas side of the AL, the system automatically generates a high flow pulse of sweep gas to flush out any accumulated moisture. Prior to the described animal study, the Gen 2 servoregulation system was tested in vitro with blood and water to verify operation, functionality, and reliability.22
Animal Preparation
In vivo experiments were conducted on 4 adult sheep (68±14.3 kg; Figure 2). All sheep were fasted the night prior to surgery. All experiments were conducted in accordance with the National Institutes of Health’s Guide for the Care and Use of Laboratory Animals. This study was approved by the University of Michigan Institutional Animal Care and Use Committee (IACUC). All sheep were sedated with propofol 6 mg/kg through a hoof vein catheter prior to endotracheal tube intubation, followed with general inhalational anesthesia with isoflurane (1–3%) and ventilated under volume-controlled ventilation. The sheep were paralyzed with vecuronium 0.01–0.04 mg/kg/hr as minute ventilation was modified during the study. The sheep were kept under anesthesia for 10–12 hours per procedure for completion of the study. The sheep were cannulated directly into the subclavian artery and vein to create an arterio-venous (AV) shunt to attach the Novalung iLA oxygenator (Novalung iLA Membrane Ventilator, Xenios AG, Heilbronn Germany, 0.5–7.0 L/min). Once it was attached to the lung, the servoregulator controller was connected to the gas side of the artificial lung and mounted on a cart adjacent to the animal. An arterial line inserted in the femoral artery and a Swan-Ganz catheter in the jugular vein allowed for hemodynamic monitoring, arterial blood sampling, and ease of fluid administration intraoperatively. Heparin was titrated to achieve an activated clotting time (ACT) of 275–375 s (Hemochron, Netherlands).
Figure 2. In vivo servoregulator testing.

Schematic diagram of circuit set up with cannula placement on animal. Gas circuit is marked in light blue and blood circuit marked in red.
Ventilation and Servoregulation
The servoregulator was initially tested in these sheep at 0.5 and 1.0 L/min AL blood flow rates with tEGCO2 levels of 10, 20, & 40 mmHg while ventilated physiologically (7 mL/kg) and at a rate of 12 breaths/minute to maintain normocapnia (goal PaCO₂ of 35–45 mmHg). The minute ventilation was then reduced to 30–40% of baseline to induce hypercapnia (goal PaCO2 60–90 mmHg) while maintaining appropriate oxygenation. The blood flow rate through the AL circuit was set at 0.5, 1.0 & 1.5 L/min, which was controlled using an adjustable Hoffman clamp. The servoregulation system was subsequently tested for tEGCO2 levels of 10, 20, & 40 mmHg at these blood flow rates and the response of the PID controller was measured.
Figure 3 depicts how the hypercapnic conditions were created and blood gases were measured throughout the animal experiments. For each test setting (“therapy” in Figure 3), the servoregulator remained on until PaCO2 stabilized (~20 min), then steady-state measurements were taken for that setting. Then the servoregulator was turned off and the animal was permitted to return to hypercapnia (“washout” in Figure 3) before the next condition was tested.
Figure 3. Schematic for hypercapnic conditions.

X axis is time duration of prep, blue areas indicate when therapy was administered, with time duration above. Note 60 minutes of baseline conditions before initiation of therapy.
Data Collection
Animal hemodynamics were recorded throughout the duration of the prep including heart rate (HR), central venous pressure (CVP), cardiac output (CO), and temperature. Every 15 minutes, tidal volume, respiratory rate, and minute ventilation were obtained as well as arterial blood gases (pH, partial pressure of CO2 and O2, and electrolytes). Artificial lung data included: pre (inlet) and post (outlet) blood gases, exhaust CO2, blood flow, and pre(inlet) and post (outlet) blood pressure. ACT was obtained hourly with our ACT goal as stated above.
Results
All 4 surgeries and post-surgical evaluations occurred without any major incidents. No studies were required to terminate early due to lack of animal stability. In all studies, the servoregulator performed appropriately without any significant issues.
When the controller was challenged with modifications to inlet blood flow or tEGCO2 levels, it was able to adjust the sweep gas flow in all animals to rapidly (<10 min) achieve the specified EGCO2 level. Figure 4 shows operation of the servoregulator system during three washout/therapy sessions in a single hypercapnic sheep for tEGCO2 values of 10 (left), 20 (middle), and 40 (right) mmHg and an AL blood flow of 1 L/min. Data for all four hypercapnic animals were combined in Figures 5 and 6. Increasing the tEGCO2 resulted in a decrease in sweep gas flow rate, CO2 removal, and electrical power consumption of the system (Figure 5). The CO2 removal rate was 46.89±9.6 versus 57.77±8.4 mL/min at .5L/min with a tEGCO2 of 20 mmHg and 58.3±3.52 versus 106.7±11.37 mL/min at1.0 L/min with a tEGCO2 of 20 mmHg.
Figure 4. Servoregulator function in single sheep.

This figure represents servoregulator system function during three washout/therapy sessions in a single hypercapnic sheep for tEGCO2 values of 10 (left), 20 (middle), and 40 (right) mmHg and an AL blood flow of 1 L/min. X axis = Time, blue line = exhaust CO2, dotted line = target exhaust CO2, Red line = pump flow rate, red dot = blood CO2 pre oxygenator, green star = blood CO2 post oxygenator.
Figure 5. Combined results n=4 in hypercapnic conditions.

X axis = target exhaust CO2, Y axis = Outlet blood PCO2, CO2 removal rate, Sweep flow rate, Power consumption (clockwise). Each color represents a different blood flow rate (.5, 1, 1.5 L/min).
Figure 6. pH and CO2 level at blood flow .5, 1 and 1.5 L/min.

X axis depicts target exhaust CO2, Left Y axis represents blood CO2 and Right Y axis represents blood pH levels. Blue and red are baseline and steady state respectively.
Increasing blood flow through the AL increased sweep gas flow rate, CO2 removal, and electrical power consumption.
The servoregulator further decreased animal arterial blood PaCO2 and increased pH in hypercapnic animals under all test conditions (Figure 6). The average pre-device pH was 7.14±.007 while post-device was 7.43±0.08 in our hypercapnic models. This demonstrates normalization of post-device pH correlating with adequate CO2 removal.
In normocapnic sheep, CO2 removal was lower than in the hypercapnic animals. At 1.0 L/min and tEGCO2 of 20 mmHg, blood PaCO2 decreased from 106.7±11.4 mmHg to 58.3±3.52 mmHg (p<0.01). This trend was consistent at 0.5 L/min with a target exhaust gas of 20 mmHg, where blood PaCO2 decreased from 57.8 ±8.4 to 46.9±4.7 (p=0.14) mmHg in normocapnic sheep.
Discussion
The advancement in this servoregulator system provides a crucial step in the advancement of an ambulatory artificial lung that can make real-time adjustments to changes in blood CO2 levels (patient metabolic demands). As in the previous generation of this system, this servoregulator was able to effectively adjust CO2 removal rapidly to meet the tEGCO2 value. In a normocapnic animal, the servoregulator functioned appropriately and did not induce a state of hypocapnia for tEGCO2 values of 20 and 40 mmHg. In addition, in hypercapnic animals, the servoregulator was able to remove higher levels of CO2 successfully. Both the arterial pCO2 as well as the pH levels reflected this function.
Membrane lungs are very efficient at removing CO2 which can lead to hypocapnia. Not only does this device aid in regulating hypercapnia but it can also prevent hypocapnic events using servoregulation. Our primary target population consists of COPD patients using this therapy as a bridge to transplantation. The ECCOR study discussed (20) was specifically for use in acutely decompensated COPD patients whereas we are looking to use this device in chronic patients. In a systematic review and meta-analysis in 2022 of the ECCOR device, there was no major difference in the prognosis of patients with and without this treatment, but it did reduce PaCO2, increased pH and improved the PaO2/FiO2 ratio in both ARDS and COPD patients.24 This is important in caring for chronic patients as a bridge to transplantation. We do not foresee our device changing prognosis but improving quality of life, reducing metabolic changes seen by hypo and hypercapnia and increasing ease of breathing. Additionally, some of the complications seen with ECCOR were related to bleeding due to vascular access issues. We plan to place these devices in a more controlled setting and use a vascular graft which could potentially reduce the risk of access complications. Work by Combes et al. describes that when ECCOR was initially studied, there was no statistically significant difference in mortality, but an increased number of patients died which led to the decreased usage of the device.25 Since then, there have been changes to these devices and they are more frequently used to reduce the potential damage from mechanical ventilation and can facilitate lower tidal volumes.26,27 The SUPERNOVA pilot trial also demonstrated feasibility and safety of this device in reducing tidal volume without serious adverse events.28 ECCOR has also been studied in end stage cystic fibrosis patients with 1 year survival at 80%.29
This model of servoregulator is improved from its previous model in several different parameters. It is a more compact system with an integrated battery pack allowing for improved portability. Additionally, this model has an automated sweep gas flushing mechanism to reduce condensation which can decrease CO2 clearance efficiency.
We envision that this servoregulator will function in the clinical setting to allow for moment-to-moment adjustments in CO2 removal based off the patient condition and current clinical status. If a patient is active, naturally their CO2 will rise and require an increase in the CO2 removal from the servoregulator. This system has demonstrated its ability to function in that application by increasing sweep gas and CO2 clearance to sustain the appropriate target exhaust gas CO2 level.
Conclusion
These in vivo data demonstrate an important step toward the development of ambulatory and wearable artificial lungs that can effectively modulate CO2 removal and thereby allow for significant fluctuations in patient activity or disease status. In future work, we plan to further miniaturize the system so that it is wearable and aim to perform longer term animal studies. Additionally, to achieve approval for clinical use, we will use standard clinical device protocol for the hardware and software development of this device.
Sources of Funding Statement
This work was in part supported by U.S. National Institutes of Health grant R21HL140995 and Department of Veterans Affairs Merit Review award I01RX003114. The contents do not represent the views of the United States Government or the U.S. Department of Veterans Affairs.
IRB/Animal Care Committee number and approval
Extracorporeal Support protocol PRO00009705 was approved by the University of Michigan Institutional Animal Care and Use Committee on August 10, 2020 and expires on August 10, 2023.
Footnotes
Conflict of Interest
The authors have no conflicts of interest to disclose.
References
- 1.American Lung Association. Trends in COPD (Chronic Bronchitis and Emphysema): Morbidity and Mortality. [Chicago, Ill.]: Epidemiology and Statistics Unit, Research and Health Education Division, 2013. [Google Scholar]
- 2.Kochanek KD, Murphy SL, Xu JQ, Tejada-Vera B. Deaths: Final data for 2014. Natl Vital Stat Rep 65 no. 4, 2016. [PubMed] [Google Scholar]
- 3.American Lung Association. Lung Health & Diseases. Learn About ARDS, 2016 [Google Scholar]
- 4.Zwischenberger JB, Conrad SA, Alpard SK, Grier LR, Bidani A: Percutaneous Extracorporeal Arteriovenous CO 2 Removal for Severe Respiratory Failure. Ann Thorac Surg 68: 181–187, 1999. [PubMed] [Google Scholar]
- 5.Diaz-Guzman E, Hoopes CW, Zwischenberger JB: The Evolution of Extracorporeal Life Support as a Bridge to Lung Transplantation. ASAIO J 59: 3–10, 2013. [DOI] [PubMed] [Google Scholar]
- 6.Hoopes CW, Kukreja J, Golden J, Davenport DL, Diaz-Guzman E, Zwischenberger JB: Extracorporeal membrane oxygenation as a bridge to pulmonary transplantation. J Thorac Cardiovasc Surg 145: 862–868, 2013. [DOI] [PubMed] [Google Scholar]
- 7.Mangi AA, Mason DP, Yun JJ, Murthy SC, Pettersson GB: Bridge to lung transplantation using short-term ambulatory extracorporeal membrane oxygenation. J Thorac Cardiovasc Surg 140: 713–715, 2010. [DOI] [PubMed] [Google Scholar]
- 8.Garcia JP, Iacono A, Kon ZN, Griffith BP: Ambulatory extracorporeal membrane oxygenation: a new approach for bridge-to-lung transplantation. J Thorac Cardiovasc Surg 139: e137–e139, 2010. doi: 10.1016/j.jtcvs.2009.12.021 [DOI] [PubMed] [Google Scholar]
- 9.Chicotka S, Rosenzweig EB, Brodie D, Bacchetta M: The “Central Sport Model”: Extracorporeal Membrane Oxygenation Using the Innominate Artery for Smaller Patients as Bridge to Lung Transplantation. ASAIO J 63: e39–e44, 2017. doi: 10.1097/MAT.0000000000000427 [DOI] [PubMed] [Google Scholar]
- 10.Matheis G, Novosel E, Hammer T: Final report summary - AMBULUNG (Ambulatory Bio-Artificial Lung), 2016. Available at: https://cordis.europa.eu/project/id/304932/reporting.
- 11.Kotloff RM, Thabut G: Lung transplantation. Am J Respir Crit Care 184: 159–171, 2011. [DOI] [PubMed] [Google Scholar]
- 12.Christie JD, Edwards LB, Kucheryavaya AY, et al. : The registry of the international society for heart and lung transplantation: 29th adult lung and heart-lung transplant report - 2012. J Heart Lung Transplant 31: 1073–1086, 2012. [DOI] [PubMed] [Google Scholar]
- 13.Yusen RD, Christie JD, Edwards LB: The registry of the international society for heart and lung transplantation: Thirtieth adult lung and heart-lung transplant report - 2013; Focus theme: Age. J Heart Lung Transplant 32: 965–978, 2013. [DOI] [PubMed] [Google Scholar]
- 14.Bain JC, Turner DA, Rehder KJ, et al. : Economic Outcomes of Extracorporeal Membrane Oxygenation With and Without Ambulation as a Bridge to Lung Transplantation. Respir Care 61: 1–7, 2016. [DOI] [PubMed] [Google Scholar]
- 15.Turner DA, Cheifetz IM, Rehder KJ, et al. : Active rehabilitation and physical therapy during extracorporeal membrane oxygenation while awaiting lung transplantation: A practical approach*. Crit Care Medicine 39: 2593–2598, 2011. [DOI] [PubMed] [Google Scholar]
- 16.Karagiannidis C, Windisch W, Bein T: Rapid Changes in Arterial Carbon Dioxide Levels Caused by Extracorporeal Membrane Oxygenation. The Temptation of a Fascinating Technology. Am J Respir Crit Care Med 201(12): 1466–1468, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Patel S, Miao JH, Yetiskul E, Anokhin A, Majmundar SH: Physiology, Carbon Dioxide Retention. [Updated 2022 Jan 4]. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2022. Jan-. [PubMed] [Google Scholar]
- 18.Bein T, Weber-Carstens S, Apfelbacher C, et al. : The quality of acute intensive care and the incidence of critical events have an impact on health-related quality of life in survivors of the acute respiratory distress syndrome - a nationwide prospective multicenter observational study. Ger Med Sci 18: Doc01, 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.The Quantum Ventilation Lite. (2020). Spectrum Medical. Retrieved November 19, 2022, from https://www.spectrummedical.com/quantum-perfusion-for-the-icu-cath-lab-and-transport/quantum-perfusion-icu-transport/quantum-ventilation-system
- 20.Witer LJ, Howard RA, Trahanas JM, et al. : Large animal model of pumpless arteriovenous extracorporeal CO2 removal using room air via subclavian vessels. ASAIO J 62: 110–113, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Potkay JA, Thompson AJ, Toomasian J, Lynch W, Bartlett RH, Rojas-Peña A: Toward a Servoregulation Controller to Automate CO2 Removal in Wearable Artificial Lungs. ASAIO J 68: 698–706, 2022. doi: 10.1097/MAT.0000000000001551 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Shaikh N, Zhang A, Jenter J, et al. : A Portable Servoregulation Controller to Automate CO2 Removal in Artificial Lungs. Bioengineering 9: 593, 2022. doi: 10.3390/bioengineering9100593 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Fang Y, Novak PJ, Hozalski RM, Cussler EL, Semmens MJ: Condensation studies in gas permeable membranes..J Membr Sci 231: 47–55, 2004. doi: 10.1016/j.memsci.2003.10.039 [DOI] [Google Scholar]
- 24.Zhu Y, Zhen W, Zhang X, et al. : Extracorporeal Carbon Dioxide Removal in Patients with Acute Respiratory Distress Syndrome or Chronic Obstructive Pulmonary Disease: A Systematic Review and Meta-Analysis [published online ahead of print, 2022 Aug 29]. Blood Purif 2022;1–11. doi: 10.1159/000525983 [DOI] [PubMed] [Google Scholar]
- 25.Combes A, Brodie D, Aissaoui N, et al. : Extracorporeal carbon dioxide removal for acute respiratory failure: a review of potential indications, clinical practice and open research questions. Intensive Care Med 48(10):1308–1321, 2022. doi: 10.1007/s00134-022-06796-w [DOI] [PubMed] [Google Scholar]
- 26.Hager DN, Krishnan JA, Hayden DL, et al. : Tidal volume reduction in patients with acute lung injury when plateau pressures are not high. Am J Respir Crit Care Med 172:1241–1245 30, 2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Terragni PP, Del Sorbo L, Mascia L, et al. : Tidal volume lower than 6 ml/kg enhances lung protection: role of extracorporeal carbon dioxide removal. Anesthesiology 111:826–835, 2009. [DOI] [PubMed] [Google Scholar]
- 28.Combes A, Fanelli V, Pham T, et al. : Feasibility and safety of extracorporeal CO2 removal to enhance protective ventilation in acute respiratory distress syndrome: the SUPERNOVA study. Intensive Care Med 45:592–600, 2019. [DOI] [PubMed] [Google Scholar]
- 29.Tipograf Y, Salna M, Minko E, et al. : Outcomes of extracorporeal membrane oxygenation as a bridge to lung transplantation. Ann Thorac Surg 107:1456–146, 2019. [DOI] [PubMed] [Google Scholar]
