Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study

Maryam Darami 1, Babak Mahmoudian 2, Michael Ljungberg 3, Jalil Pirayesh Islamian 4,
PMCID: PMC10581757  PMID: 37854088

Abstract

Objectives  Collimators have a significant role in image quality and detectability in single-photon emission computed tomography (SPECT) imaging. Using an appropriate alloy that effectively absorbs scattered photons, without induced secondary x-rays, and with proper rigidity and weight may provide an effective approach to the image improvement that conventionally collimators made of lead (Pb).

Materials and Methods  A Siemens E.CAM SPECT imaging system equipped with low-energy high-resolution (LEHR) collimator was simulated by the Simulating Medical Imaging Nuclear Detectors Monte Carlo program. Experimental and simulated data were compared based on a 2-mm 99m Tc point source in an acrylic cylindrical Deluxe phantom (Data Spectrum, Inc). Seven types of tungsten (W) alloys (Wolfmet), with W content from 90 to 97% by weight, were then used as collimator materials of the simulated system. Camera parameters, such as energy- and spatial resolution, image contrast, and collimator-related parameters, such as fraction of septal penetration, scatter-to-primary ratios, and percentage of induced secondary x-rays, due to interactions in the collimator, were evaluated.

Results  Acceptable conformity was found for the simulated and experiment systems in terms of energy spectra, 10.113 and 10.140%, full width at half-maximum (FWHM) of the point spread function (PSF) curves, 8.78 and 9.06 mm, sensitivity, 78.46 and 78.34 cps/MBq, and contrast in images of 19.1 mm cold spheres in the Deluxe phantom, 79.17 and 78.97%, respectively. Results on the parameters of the simulated system with LEHR collimator made from the alloys showed that the alloy consisting of 90% W, 6% nickel, and 4% copper provided an FWHM of 8.76 mm, resulting in a 0.2% improvement in spatial resolution. Furthermore, all the Wolfmet collimators showed a 48% reduction in the amount of X-rays production compared to the Pb.

Conclusion  A Wolfmet LEHR collimator, made by a combination of W (90%), Ni (6%), and Cu (6%) provides a better image quality and detectability compared to the Pb.

Keywords: Wolfmet, collimator, image quality, SPECT, SIMIND, tungsten

Introduction

Single-photon emission computed tomography (SPECT) is a substantial tomographic functional imaging modality used to detect and characterize a variety of diseases, including cancers and infections. 1 2 SPECT is a cost-effective imaging modality 3 4 5 that improves diagnostic accuracy by facilitating three-dimensional information of the radiopharmaceutical distribution in in vivo and is also a preferable imaging approach due to radioisotope availability. However, SPECT images can be strongly influenced by penetration and scattering in the collimator walls and related X-ray emissions. Meanwhile, optimization of these effects may improve image quality and lesion detectability. 6 7 The purpose of a collimator is to let only those photons that propagate in an appropriate direction along a collimator hole to pass through and interact in the scintillation crystal but stop others. 8 9 However, not all photons will be stopped by the collimator walls. In order to improve collimator performance, the collimator absorption efficiency 10 11 can be increased by using materials of high atomic numbers and mass density. Conventional nuclear medicine collimators are fabricated from lead (Pb) with small amounts of antimony (Sb). Pb is considered as effective shielding material for impinging photons because of its high density (11.35 g/cm 3 ) and atomic number ( Z  = 82). 12 13 In spite of other properties such as low cost and easy access, it is toxic, nonrigid, and produce secondary ionizing radiations of relatively high energies; 14 15 Pb may not be a proper collimator material of imaging systems, at least when it is used on its own.

In addition to Pb, other materials that have been used or proposed as collimator materials are tungsten (W), gold (Au), uranium (U), and platinum (Pt). However, because of substantial costs, Pb and W are by far the most commonly used materials. 16

In a study by Rajaee et al, on the effect of Au, Pb, U, and W as a high-energy-general-purpose parallel-hole collimator material on image quality, they demonstrated that the spatial resolution improved as the atomic number increased but with a cost of the system sensitivity. 10 Lee et al have shown from Monte Carlo simulations that the use of U provided about 3.19, 4.19, and 8.01% higher spatial resolution compared to Au, W, and Pb, although the system sensitivity of Pb was determined 4.25, 6.53, and 10.28% compared to the W, Au, and U, respectively. 17 In 2013, Weng et al, showed that imaging systems with a W parallel-hole collimator and cadmium zinc telluride (CZT) detectors compared to using Pb with a conventional NaI(Tl) detector have reduced the amount of septal penetration, image noise, and sharpness, and they concluded that the reduction may be attributed to W linear attenuation coefficient and also better energy resolution of CZT detector. 18 In another study by Lee et al, it was shown that the spatial resolution of U, with the same sensitivity, was better than Pb, W, and Au. 11 In this regard, it was suggested that the Pb–Sb alloy with a ratio of 98 and 2%, compared to the common Pb collimator, improves spatial resolution and energy resolution about 1.3 and 3%, respectively, with equal sensitivity. 11 They have also reported that the Pb–Sb alloy collimator improved the contrast and detection capability of the simulated cold and hot liver lesions of an NURBs based cardiac and torso (NCAT) Human Torso Phantom compared to the Pb alone.

In studies on metal alloys as pinhole collimator materials, Peterson et al used Rose's metal alloy, an alloy consisting of bismuth (Bi), Pb, and tin (Sn) with fractional weights of 50, 28, and 22%, respectively, in preclinical small‐animal imaging. Their studied parameters included penetration-to-scatter and penetration-to-total component ratios, system sensitivity, and the spatial resolution for two radionuclides, 99m Tc and 125 I. 19 The authors concluded that Rose's metal is an alternative to standard materials not only for low-energy photon imaging but also for medium-energy applications that require low-cost, flexible pinhole configurations and designs, and that can tolerate a degraded spatial resolution compared to Au and W.

Due to its high density of 19.26 g/cm 3 and the atomic number of 74, W exhibits excellent absorption behavior against photon radiation such as X-rays and γ-radiation. 20 In an experimental study on shielding features of W–Cu, with weighting percentages of W85 wt%–Cu15 wt % and W75 wt%–Cu25 wt% composites, at low energy 0.266 MeV, Dong et al have found highest values of linear attenuation coefficient, 2.11 and 4.33 cm −1 for W75 wt%–Cu25 wt% and W85 wt%–Cu15 wt%, while they found the lowest values with high-energy 1.25 MeV x-rays, 0.44 and 0.48 cm −1 for W75 wt%–Cu25 wt% and W85 wt%–Cu15 wt%, respectively, so it is concluded that W–Cu composites are suitable to be used in several applications of radiation protection. 21

Kaur et al presented in a review article the applicability of metallic alloys including W, Cu, Fe, Ni, and Pb-based alloys in γ-rays shielding and expressed this in terms of mass-attenuation coefficient (μ m ), effective atomic number (Z eff ), electron density (Ne), mean-free path (mfp), half-value layer (HVL), tenth-value layer (TVL), exposure buildup factor, and energy exposure buildup factor, and they concluded that mass-attenuation coefficient values of Pb-based alloys and W-based alloys with lowest values of mfp, HVL, and TVL are comparable because they present high atomic numbers in a large weight percentage ratio than the other constituent elements. However, both the melting point and cost factor are high for W-based alloys. 22 Murty et al evaluated Z eff of different W–Cu alloys by exposure to photons with an energy range of 60–400 keV. They extended the study to also check the validation of the mixture rule and any energy dependency of effective atomic number in the energy range of 100 to 1,400 keV. Both studies showed no energy dependence of Z eff . 23 24

W alloy performs close to Au in terms of stopping power, with a better trade-off between performance and cost with 131 I imaging. 25 In 2019, Nguyen et al compared the performance of Pb, W, Au, and depleted U pinhole collimators in terms of spatial resolution on EXIRAD-3D, a newly autoradiography technique based on a highly focusing multipinhole collimator that achieves micronresolution SPECT for cryo-cooled tissue samples, for different radioisotopes including 111 In (171 keV and 245 keV), 99m Tc (140 keV), 201 Tl (71 keV), and 125 I (27 keV). 26 They found that the image contrast-to-noise ratio (CNR) was improved with a higher atomic number in a range from 1.9 to 36.6%. They suggested W collimator to be a good choice to use for a wide ranges of SPECT radioisotopes. 26

W alloys, including Wolfmet, do not only have a very high mass density, but are very good in attenuating ionizing radiation. 27 This makes the alloys ideal for shielding proposes in nuclear medicine or nuclear industry. Fig. 1 demonstrates the photon absorption characteristics of Wolfmet HE 395, a composite of W (95%)–Ni (3.5%)–Fe (1.5%), and other shielding materials. The TVL is plotted as a function of photon energy assuming narrow beam exposure condition. The data were supplied by the National Physical Laboratory. 27 HE 395 and Pb have shown TVLs of 24.3 and 38.3 mm, against 60 Co gamma rays (1,173.2 and 1,332.5 keV), with HVL thickness of 7.0 and 11.7 mm, respectively.

Fig. 1.

Fig. 1

Comparative absorption data (10th value layer) of water, concrete, aluminum, iron, lead, Wolfmet ® HE 395, as a function of the energy of incident narrow-beam photons. Data supplied by the National Physical Laboratory. 27

Here, we evaluated the effects of Wolfmet tungsten alloys, as the LEHR collimator material, on the functional parameters and the image quality of an SPECT system.

Material and Methods

We used the Simulating Medical Imaging Nuclear Detectors (SIMIND) Monte Carlo program (developed by Prof. Michael Ljungberg, Lund University, Lund, Sweden) to simulate a Siemens E.CAM SPECT camera (Siemens Medical Solutions, Erlangen, Germany). 28 29 In brief, each of the two camera heads consisted of a removable low-energy high-resolution (LEHR) collimator, a thin NaI(Tl) scintillation crystal, a light-guide, and an array of photomultiplier tubes (PMTs). The geometry of LEHR collimator was included parallel hexagonal holes defined as 1.11-mm flat-to-flat hole diameter, 24.05 mm hole length, and 0.16-mm septal thickness. The NaI (Tl) crystal was rectangular in shape with the dimension of 59.1 × 44.5 × 0.95 cm. The light yield was 40k photons/MeV with a peak emission spectrum at 415 nm. 7 29 The emitted light-photons from the various interactions in the NaI (Tl) crystal were collected by 59 PMTs where 53 of them had a diameter of 7.6 cm and the other six had a 5.1 cm diameter. The photocathode was of the bialkali type with a quantum efficiency of approximately 30%. The light guide with coupling grease ensured an effective light collecting and a good optical coupling between the scintillating crystal and PMTs to reduce nonlinearity and so contribute to spatial resolution. To consider the contribution of various structures attached to back of the crystal on backscattering of the emitted photons, mainly with high-energy gamma rays, a single slab of Pyrex was also substituted and included in the simulation. 30

Experimental data were obtained to validate the SIMIND simulations and these were based on scanning (1) a 2 mm 99m Tc point source with 3.7 MBq activity located at a distance of 10 cm from the detector, and (2) a clear Acrylic Plexiglass Deluxe ECT (model ECT/DLX/P) water-filled cylindrical phantom with cold spheres and rods ( Table 1 ) in a hot-background using 370 MBq 99m Tc. 31 The energy window was defined as 126 to 154 keV.

Table 1. Specification of the Deluxe phantom.

Interior cylinder dimensions 216 mm ø x 186 mm
Cylinder wall thickness 3.2 mm
Cylinder volume 6.75 L
Cylinder volume with inserts 6.1 L
Cold rods insert height 88 mm
Sphere center from base plate 127 mm

Based on NEMA specifications on quality control procedures, the system spatial and energy resolution were determined from planar images of 99m Tc point source acquiring at a distance of 10 cm from the collimator surface and center of the field of view. Matrix size was 128 × 128 matrix size with 0.27 mm pixel size, and 10 million counts per projection were collected. 32 33 Additionally, the system sensitivity (cps/MBq) specified for a point source at 25 cm from the surface of the detector.

Image contrast and spatial resolution were evaluated by simulating the SPECT imaging of a digitized cylindrical Deluxe phantom. 31 In brief, SPECT projections along the axis of rotation were calculated for a situation with the phantom centrally positioned at 15 cm distance from the collimator surface using a matrix size of 128 × 128 with 0.39 cm pixel size, 128 projection views in a 360 degrees clockwise rotation with one million counts per projection. An OS-EM reconstruction algorithm was used to reconstruct the SPECT projections with eight iterations and four subsets. Compensation for photon attenuation was included. The spatial resolution and image contrast were determined visually from the smallest recognizable rods and 19.1 mm spheres in the reconstructed images, respectively. 7

Simulations were conducted using the conventional LEHR collimator design with two Pb materials and multiple Wolfmet tungsten alloys ( Table 2 ).

Table 2. Details on the Wolfmet alloys components as a LEHR collimator 35 .

Collimator material Composition (weight %)
Lead Pb (100%)
Lead antimony Pb (98%), Sb (2%)
Tungsten W (100%)
W_ha190 W (90%), Ni (6%), Cu (4%)
W_ha193 W (93%), Ni (3.5%), Cu (3.5%)
W_ha195 W (95%), Ni (2.5%), Cu (2.5%)
W_he390 W (90%), Ni (5%), Fe (5%)
W_he395 W (95%), Ni (3.5%), Fe (1.5%)
W_he397 W (97%), Ni (1.5%), Fe (1.5%)
W_he3925 W (92.5%), Ni (3.8%), Fe (3.8%)

Abbreviation: LEHR, low-energy high-resolution.

Results

A comparison of measured and simulated energy pulse-height distributions yields an acceptable conformity in terms of the functional and image parameters ( Fig. 2 ).

Fig. 2.

Fig. 2

Energy pulse-height distribution obtained from the 99m Tc point source (140 keV) simulated by SIMIND ( blue ) and from measurement ( red ).

The calculated energy resolution (FWHM) from the 99m Tc point source from simulation and measurement were calculated to 10.11 and 10.14%, respectively. The spatial resolution (FWHM) at 10 cm source to collimator distance was calculated to 8.78 and 9.06 mm for simulation and measurement and the system sensitivity was determined to be 78.46 and 78.34 cps/MBq, respectively.

Contrast measures from the cold sphere images of cylindrical Deluxe phantom were calculated according to the following equation:

graphic file with name 10-1055-s-0043-1771287-fo2350006-1.gif

where M sp and M cy correspond to the minimum pixel value of the cold spheres and the maximum pixel value in the Deluxe phantom cylinder, respectively. The results on contrast values of cold bodies in background of the Deluxe phantom cylinder are presented in Table 3 .

Table 3. Results on the calculated image contrast of Deluxe phantom cold spheres.

System Diameter of the Deluxe phantom cold spheres (mm)
9.5 12.7 15.9 19.1 25.4 31.8
Measurement 53.24 59.79 69.44 78.97 87.39 92.47
Simulation 51.92 56.48 68.89 79.17 83.03 89.21

Profiles on image uniformity, without attenuation correction as with the experimental, obtained from images of the Deluxe phantom are presented in Fig. 3 for the experiment and simulated SPECT imaging systems.

Fig. 3.

Fig. 3

Profiles of image uniformity obtained from mages of Deluxe phantom SPECT from measurement ( blue ) and simulation ( red ).

Results on the functional parameters of the simulated SPECT system equipped with 10 collimator materials from the 99m Tc point source simulations are presented in Table 4 .

Table 4. Functional parameters of the simulated SPECT system from the 99m Tc point source scanning with 10 different collimator materials .

Material a Sensitivity
(cps/MBq)
Efficiency Spatial resolution (mm)
Window Spectrum FWHM FWTM
Pb 78.459 0.7388 0.9336 8.780 15.52
Pb_Sb 78.635 0.7390 0.9334 8.807 15.58
W 77.233 0.7534 0.9323 8.751 15.42
W_ha190 78.266 0.7531 0.9321 8.762 15.50
W_ha193 77.927 0.7536 0.9324 8.771 15.51
W_ha195 77.824 0.7531 0.9325 8.781 15.48
W_he390 78.388 0.7535 0.9322 8.800 15.52
W_he395 77.629 0.7537 0.9322 8.766 15.48
W_he397 77.533 0.7535 0.9324 8.772 15.48
W_he3925 78.041 0.7531 0.9322 8.791 15.50

Abbreviation: FWHM, full width at half-maximum.

a

Pb (lead), Sb (antimony), W (tungsten).

Results on the collimator-related parameters, including septal penetration, geometric photons, Compton scatter in collimator, and the ratio of the scattered-to-primary photons in the photopeak energy windows, besides, the percentage of the collimator-induced x-rays are shown in Table 5 . The parameters were calculated by SIMIND during the simulation according to the related system, isotope, and phantom settings and, hence, were presented as data in the result files.

Table 5. Radiation interaction events of a 2 mm 99m Tc point source with 3.7 MBq activity with the simulated SPECT system depend on collimator materials .

Collimator
material
Penetration
(%)
Primary
(%)
Scatter
(%)
c X-Ray
(%)
Scatter/Total
×10 −5
AC a WEW b AC WEW AC WEW AC Total
Pb 6.21 4.62 87.62 93.95 2.20 1.43 3.96 535
Pb_Sb 6.29 4.73 87.57 93.84 2.21 1.44 3.93 536
W 3.88 3.42 91.15 94.16 1.81 1.44 2.05 538
W_ha190 4.79 4.27 92.93 95.60 1.97 1.57 2.09 536
W_ha193 4.54 4.02 91.44 94.45 1.94 1.53 2.09 538
W_ha195 4.38 3.87 91.60 94.62 1.91 1.51 2.11 536
W_he390 4.80 4.27 91.12 94.17 1.98 1.56 2.10 535
W_he395 4.29 3.83 91.76 94.67 1.87 1.88 2.07 536
W_he397 4.15 3.69 91.90 94.82 1.86 1.48 2.09 536
W_he3925 4.57 4.07 91.37 94.40 1.95 1.54 2.10 535

Abbreviation: SPECT, single-photon emission computed tomography.

a

Photons after collimator.

b

Photons within energy windows.

c

X-ray in collimator was equal zero for all collimators.

Table 6 provides the measures of image contrast of the Deluxe phantom produced from the simulated SPECT system with different collimator materials. Reconstructed images of the cold spheres using the collimators are presented in Fig. 4 .

Table 6. The results on the SPECT image contrast of 19.1 mm cold sphere of Deluxe phantom with different collimator materials.

Collimator material Pb Pb_Sb W W_ha190 W_ha193 W_ha195 W_he390 W_he395 W_he397 W_he3925
contrast (%) 79.17 79.25 77.97 79.00 80.49 79.15 78.76 78.76 80.33 80.18

Abbreviation: SPECT, single-photon emission computed tomography.

Fig. 4.

Fig. 4

Reconstructed images of the SIMIND simulated SPECT of cold spheres (slice 56) of Deluxe phantom by with the LEHR collimator made from lead (Pb), lead–antimony (Pb–sb), tungsten (w), and Wolfmet alloys. SPECT projections along the axis of rotation were calculated for a situation with the phantom centrally positioned at 15 cm distance from the collimator surface using a matrix size of 128 × 128 with 0.39 cm pixel size. 128 projection views in a FO clockwise rotation with 1M counts per projection. A OSEM reconstruction algorithm was used to reconstruct the SPECT projections with eight iterations and four subsets. Compensation for photon attenuation was included.

Discussion

The purpose of the present study was to investigate the effects of W alloys as collimator material of SPECT imaging system on functional parameters of the system including spatial resolution, sensitivity, and so image quality by SIMIND Monte Carlo simulation.

In this regard, seven alloys of Wolfmet tungsten were used in the simulation of the potential benefits of LEHR W collimators for the current SPECT imaging system ( Table 2 ). According to the results, W alloys collimators demonstrated less septal penetration, scatter radiation, and also a better primary ratio compared to the Pb which produces secondary ionizing radiations with relatively high energies especially at whole the energy spectra ( Table 5 ). Furthermore, tungsten with a high density and atomic number of 19.26 g/cm 3 and 74, respectively, may exhibit excellent absorption behavior against the photon radiations and hence on image contrast ( Table 6 ).

Verification of the simulation with the conventional LEHR collimator confirmed the results of Azarm et al 29 34 and Toossi et al. 29 Meanwhile, according to the obtained results, there were 1.40 and 1.18% reduction in system sensitivity with W_he397 alloy, compared to Pb–Sb and Pb alone ( Table 4 ), respectively. Azarm et al, in a study on the effects of collimator material on spatial and energy resolution, have shown that the common collimator made of 98% Pb and 2% Sb could provide a better spatial and energy resolution, 7.68 mm and 9.87%, compared to 7.74 mm and 10.14% of Pb alone, respectively. 34 Our study also showed that Pb–Sb collimator improved the spatial resolution by 0.31% compared to Pb, and W_ha190 alloy further improved to about 0.2% compared to Pb and Sb. It must be mentioned that all the studied alloys showed a reduction in the penetration fraction, collimator-induced X-rays, and increased geometrical fraction compared to Pb. However, Collimator with W alone provides better results for the parameters compared to the studied materials ( Table 5 ). In this regard, results on the evaluation of pinhole collimator materials for micronresolution ex vivo SPECT by Nguyen et al 2019 indicate that using materials with higher photon attenuation including U, Au, W, and Pb yields images with better CNR for the 125 I, 201 Tl, 99m Tc, and 111 In isotopes with improvements ranging from 1.9 to 36.6%. 26 They have suggested W collimator as a good choice for a wide range of SPECT isotopes, so our results also confirmed their findings. In a similar manner with photon-stopping power in terms of photon attenuation, increasing the linear attenuation coefficient was demonstrated to have potential effects on the improvement of the spatial resolution and, hence, on isotopic image quality, for example, image contrast, resolution, and detectability. 11 17 It should be noted that linear attenuation coefficient is also influenced by the beam energy. 21 22 Regarding image contrast, the value obtained for the 19.1 mm cold sphere inside the Deluxe phantom increased by 1.32, 1.16, and 1.01 with W_ha193, W_he397, and W_he3925 compared to the Pb, respectively ( Table 6 and Fig. 4 ). One explanation could be the penetration rate reduction.

Finally, the emergence of new production techniques, such as direct 3-D printing of metals or “cold casting” of W-composite materials, will open new possibilities for the fabrication of more tailored collimator designs that would be impossible or very expensive to construct by conventional methods of fabrication including casting hot Pb (microcast) or folding Pb foil (microlinear).

Conclusion

The present study showed that the use of a potential LEHR collimator made from Wolfmet tungsten alloys could a better spatial resolution, improved contrast, and image quality in general when comparing conventional Pb and Pb–Sb collimators.

Acknowledgments

The authors wish to thank Steve Jeffery, Business Development Manager of WOLFMET ® , for the information about the Wolfmet alloys.

Funding Statement

Funding This work was financially supported by a grant from the Medical Radiation Sciences Research Team, Tabriz University of Medical Sciences, Tabriz, Iran (Grant no # 65077).

Conflicts of Interest None declared.

Authors' Contribution

M.D. was responsible for writing—original draft, visualization, methodology, investigation, data curation, and conceptualization. B.M was responsible for writing—review and editing, supervision, formal analysis, and conceptualization. M.L. was responsible for providing materials for SIMIND simulation, review, and editing. J.P.I. was responsible for writing—review and editing, supervision, funding acquisition, and conceptualization.

References

  • 1.Lee S J, Park H J. Single photon emission computed tomography (SPECT) or positron emission tomography (PET) imaging for radiotherapy planning in patients with lung cancer: a meta-analysis. Sci Rep. 2020;10(01):14864. doi: 10.1038/s41598-020-71445-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Sadremomtaz A, Telikani Z. Validation and optimization studies of small animal SPECT using GATE Monte Carlo simulation. Nucl Instrum Methods Phys Res A. 2019;915:94–101. [Google Scholar]
  • 3.Taherparvar P, Sadremomtaz A. Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT. Australas Phys Eng Sci Med. 2018;41(01):31–39. doi: 10.1007/s13246-017-0607-6. [DOI] [PubMed] [Google Scholar]
  • 4.Wang Y. Development and applications of high-sensitivity and high-resolution fully three-dimensional SPECT imaging techniques using two different collimator designs. The University of North Carolina at Chapel Hill; 2004. [Google Scholar]
  • 5.Baghani H R. Image quality parameters in brain imaging with fan-beam collimator: a Monte Carlo study on radiation scattering effects. Radiological Phys Technol. 2019;12(02):194–200. doi: 10.1007/s12194-019-00514-w. [DOI] [PubMed] [Google Scholar]
  • 6.McQuaid S J, Southekal S, Kijewski M F, Moore S C. Joint optimization of collimator and reconstruction parameters in SPECT imaging for lesion quantification. Phys Med Biol. 2011;56(21):6983–7000. doi: 10.1088/0031-9155/56/21/014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Islamian J P, Toossi M T, Momennezhad M, Zakavi S R, Sadeghi R, Ljungberg M. Monte Carlo study of the effect of collimator thickness on T-99m source response in single photon emission computed tomography. World J Nucl Med. 2012;11(02):70–74. doi: 10.4103/1450-1147.103419. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Khorshidi A, Ashoor M, Hosseini S H, Rajaee A. Estimation of fan beam and parallel beam parameters in a wire mesh design. J Nucl Med Technol. 2012;40(01):37–43. doi: 10.2967/jnmt.111.089904. [DOI] [PubMed] [Google Scholar]
  • 9.Khorshidi A, Ashoor M, Hosseini S H, Rajaee A. Evaluation of collimators' response: round and hexagonal holes in parallel and fan beam. Prog Biophys Mol Biol. 2012;109(03):59–66. doi: 10.1016/j.pbiomolbio.2012.03.003. [DOI] [PubMed] [Google Scholar]
  • 10.Rajaee A, Shahriari M, Kamali AA, Hosseini S. Simulation study of the influence of collimator material on image quality improvement for high energy photons in nuclear medicine using MCNP code. J Theor Appl Phys. 2011;4(04):13–18. [Google Scholar]
  • 11.Lee Y-J, Kim D-H, Kim H-J. The effect of high-resolution parallel-hole collimator materials with a pixelated semiconductor SPECT system at equivalent sensitivities: Monte Carlo simulation studies. J Korean Phys Soc. 2014;64:1055–1062. [Google Scholar]
  • 12.Gopal S, Sanjeevaiah B. A method to determine the γ-ray attenuation coefficients. Nucl Instrum Methods. 1973;107(02):221–225. [Google Scholar]
  • 13.Rezaei-Ochbelagh D, Azimkhani S. Investigation of gamma-ray shielding properties of concrete containing different percentages of lead. Appl Radiat Isot. 2012;70(10):2282–2286. doi: 10.1016/j.apradiso.2012.06.020. [DOI] [PubMed] [Google Scholar]
  • 14.Kobayashi S, Hosoda N, Takashima R. Tungsten alloys as radiation protection materials. Nucl Instrum Methods Phys Res A. 1997;390(03):426–430. [Google Scholar]
  • 15.Directive E. Restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS) Off J Eur Communities. 2013;46:19–23. [Google Scholar]
  • 16.Van Audenhaege K, Van Holen R, Vandenberghe S, Vanhove C, Metzler S D, Moore S C. Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging. Med Phys. 2015;42(08):4796–4813. doi: 10.1118/1.4927061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Lee Y-J, Ryu H-J, Lee S-W, Park S-J, Kim H-J. Comparison of ultra-high-resolution parallel-hole collimator materials based on the CdTe pixelated semiconductor SPECT system. Nucl Instrum Methods Phys Res A. 2013;713:33–39. [Google Scholar]
  • 18.Weng F, Bagchi S, Huang Q, Seo Y. Design Studies of a CZT-based detector combined with a pixel-geometry-matching collimator for SPECT imaging. Paper presented at: 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC); 2013. [DOI] [PMC free article] [PubMed]
  • 19.Peterson M, Strand S E, Ljungberg M. Using Rose's metal alloy as a pinhole collimator material in preclinical small-animal imaging: a Monte Carlo evaluation. Med Phys. 2015;42(04):1698–1709. doi: 10.1118/1.4914423. [DOI] [PubMed] [Google Scholar]
  • 20.Handtrack D, Tabernig B, Kestler H, Pohl P, Glatz W, Sigl L. Tungsten heavy alloys for collimators and shieldings in the X-ray diagnostics. Paper presented at: Proceedings of 18th PLANSEE seminar on refractory metals and hard materials; 2013.
  • 21.Dong M, Tishkevich D, Hanfi M et al. WCu composites fabrication and experimental study of the shielding efficiency against ionizing radiation. Radiat Phys Chem. 2022;200:110175. [Google Scholar]
  • 22.Kaur T, Sharma J, Singh T. Review on scope of metallic alloys in gamma rays shield designing. Prog Nucl Energy. 2019;113:95–113. [Google Scholar]
  • 23.Murty V R, Winkoun D P, Devan K R.Effective atomic numbers for W/Cu alloy using transmission experiments Appl Radiat Isot 200053(4–5):945–948. [DOI] [PubMed] [Google Scholar]
  • 24.Murty VR.Effective atomic numbers for W/Cu alloy for total photon attenuation Radiat Phys Chemi 200471(3–4):667–669. [Google Scholar]
  • 25.Auer B, De Beenhouwer J, Kalluri K, Goding J C, Furenlid L R, King M A. Preliminary investigation of design parameters of an innovative multi-pinhole system dedicated to brain SPECT imaging. Paper presented at: 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC); 2018.
  • 26.Nguyen M P, Goorden M C, Kamphuis C, Beekman F J. Evaluation of pinhole collimator materials for micron-resolution ex vivo SPECT. Phys Med Biol. 2019;64(10):105017. doi: 10.1088/1361-6560/ab1618. [DOI] [PubMed] [Google Scholar]
  • 27.Wolfmet. Wolfmet tungesten alloys radiation shielding data sheet. Accessed April 3, 2022 at:https://www.wolfmet.com/wp-content/uploads/2017/01/Wolfmet_Radiation_Shielding2.pdf2017.
  • 28.Ljungberg M, Strand S-E. A Monte Carlo program for the simulation of scintillation camera characteristics. Comput Methods Programs Biomed. 1989;29(04):257–272. doi: 10.1016/0169-2607(89)90111-9. [DOI] [PubMed] [Google Scholar]
  • 29.Bahreyni Toossi M T, Islamian J P, Momennezhad M, Ljungberg M, Naseri S H. SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys. 2010;35(01):42–47. doi: 10.4103/0971-6203.55967. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Islamian J P, Toossi M T, Momennezhad M, Zakavi S R, Sadeghi R.Monte Carlo study of the effect of backscatter material thickness on 99mTc source response in single photon emission computed tomography Iran J Med Phys 201310(1–2):69–77. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Pirayesh I J, Bahreyni T MT, Momennezhad M, Naseri S, Ljungberg M. Simulation of a quality control Jaszczak phantom with SIMIND Monte Carlo and adding the phantom as an accessory to the program. Iran J Med Phys. 2012;9(02):135–140. [Google Scholar]
  • 32.Blust J. Gamma camera acceptance testing: the first quality control. J Nucl Med Technol. 1994;22(02):58–61. [Google Scholar]
  • 33.EANM Physics Committee ; EANM Working Group on Nuclear Medicine Instrumentation Quality Control . Busemann Sokole E, Płachcínska A, Britten A, Lyra Georgosopoulou M, Tindale W, Klett R. Routine quality control recommendations for nuclear medicine instrumentation. Eur J Nucl Med Mol Imaging. 2010;37(03):662–671. doi: 10.1007/s00259-009-1347-y. [DOI] [PubMed] [Google Scholar]
  • 34.Azarm A, Islamian J P, Mahmoudian B, Gharepapagh E. The effect of parallel-hole collimator material on image and functional parameters in SPECT imaging: A SIMIND Monte Carlo study. World J Nucl Med. 2015;14(03):160–164. doi: 10.4103/1450-1147.163242. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Wolfmet. Wolfmet tungsten alloys technical information. Accessed April 3, 2022 at:https://www.wolfmet.com/wp-content/uploads/2017/01/Wolfmet_Tungsten_Alloys_Technical_Information1.pdf..2019.

Articles from World Journal of Nuclear Medicine are provided here courtesy of Thieme Medical Publishers