Abstract
During endoscopic ultrasound‐guided biliary drainage (EUS‐BD), there is a risk for bile leakage until stent deployment, which can result in severe peritonitis, particularly when passing a drainage stent becomes challenging despite tract dilation. There is no established method or dedicated device to optimize EUS‐BD. Therefore, we have developed a novel stent deployment technique using the tapered sheath dilator. To address the safety and technical aspects of the EUS‐BD technique, we retrospectively analyzed 11 consecutive patients who underwent EUS‐BD using the tapered sheath dilator. The procedure involved the insertion of a guidewire, followed by mechanical dilation using the tapered sheath dilator. Subsequently, the inner catheter was removed and drainage stents (up to 6 Fr in diameter) were deployed through the outer sheath. We found a 100% technical success rate for tract dilation and stent deployment; moreover, all patients achieved clinical success. The median time for dilation was 40 s (range, 8–198), whereas the median time from dilation to stent deployment was 10 min (range, 6–19). Notably, no cases of bile leakage or peritonitis were observed. In conclusion, the use of the integrated device for tract dilation and stent delivery system might provide a safe and straightforward technique for drainage stenting during EUS‐BD.
Keywords: biliary drainage, bile leakage, Endosheather, EUS‐BD, tapered sheath dilator
INTRODUCTION
Endoscopic ultrasound‐guided biliary drainage (EUS‐BD) is a highly effective treatment option for various conditions, including obstructive jaundice, cholangitis, and cholecystitis. 1 , 2 However, the deployment of a drainage stent through a needle tract poses a significant challenge, even after dilation. This prolonged interval between dilation and stent deployment can lead to bile leakage through the dilated tract, potentially resulting in infected biloma or severe peritonitis. Despite numerous attempts to address this issue, a standardized method has yet to be established.
Therefore, we have developed a novel stent deployment technique using an endoscopic tapered sheath (EndoSheather; Piolax, Inc.) 3 , 4 ; this device serves as a dilation tool and a delivery system, allowing for mechanical dilation of the needle tract and the smooth insertion of a drainage stent through its indwelling outer sheath, which effectively bridges the gap to the target space. This eliminates the need to remove the dilation device and can prevent bile leakage in EUS‐BD.
PROCEDURE OR TECHNIQUE
Procedure
EUS‐BD was performed in all patients in a prone position using a linear echoendoscope (GF‐UCT260; Olympus Medical Systems) and CO2 insufflation. The bile duct or gallbladder was punctured under EUS guidance using a 19‐gauge needle and a small amount of contrast medium was injected to confirm accurate placement. A 0.025‐inch guidewire was inserted, followed by the mechanical dilation of the needle tract using a novel guide sheath with a tapered tip of the inner catheter (EndoSheather; Figure 1). Subsequently, the inner catheter was removed, whereas the outer sheath remained in place inside the bile duct or gallbladder. The outer sheath with a Y‐connector attachment facilitated effective aspiration of bile juice and injection of contrast medium to ensure accurate stent positioning. Finally, drainage stents (up to 6 Fr in diameter) were selected based on the specific requirements of each case and deployed through the outer sheath (Figures 2 and 3, and Video S1). The stents used in our study included fully covered self‐expandable metal stents with a 5.9‐Fr (8 mm × 8 cm/12 cm; HANAROSTENT Biliary Full Cover Benefit; Boston Scientific Co.), an uncovered self‐expandable metal stents with a 5.7‐Fr (8 mm × 6 cm; BileRush Selective; Piolax) and 5.4‐Fr (8 mm × 6 cm; Zeo stent V; ZEON Medical Inc.), and a plastic stent with a 6‐Fr (4/7/10 cm; Zimmon double pig‐tail stent; Cook Medical) delivery systems.
Patients and methods
We retrospectively analyzed 11 consecutive patients who underwent EUS‐BD using a tapered sheath dilator at Nagoya City University Hospital between August 2021 and March 2023. The indications for the procedure were determined at the discretion of the endoscopists. The review board of the Nagoya City University Graduate School of Medical Sciences (approval no. 60–22–0052) approved the study protocol; informed consent was obtained from all patients.
The primary endpoint was to determine the incidence rate of early adverse events, including bile leakage and peritonitis, within 2 weeks. Especially, bile leakage was based on the findings around the puncture site of computed tomography on the day following the EUS‐BD. In addition to this finding, a diagnosis of peritonitis can be made with fever (>38°C), abdominal pain with peritoneal irritation signs (not caused by pancreatitis or perforation), or an elevated inflammatory response (an increase from baseline). Adverse events were classified according to the American Society for Gastrointestinal Endoscopy lexicon. 5 Besides, we assessed the technical feasibility and efficacy of the procedure as secondary endpoints. Technical success was defined as successful tract dilation and subsequent stent deployment. Clinical success for cholestasis was determined by a reduction > 50% or normalization of the plasma bilirubin within 14 days, following the Tokyo criteria 2014, 6 and for cholecystitis, it was defined as the resolution of clinical symptoms and laboratory findings associated with cholecystitis. Furthermore, we measured the duration of dilation, the time interval between dilation and stent deployment, and the overall procedure time, defined as the time from the first puncture to stent deployment.
RESULTS
Table 1 presents the baseline clinical characteristics of all patients. The study included four male and seven female patients, with a median age of 73 (range, 52–97) years. Among the patients, seven had malignancies, such as pancreatic cancer, whereas four had benign conditions, such as cholecystitis. EUS‐BD was an alternative management approach to address challenging situations where conventional transpapillary techniques were unacceptable. The drainage targets were the bile duct in eight (72.7%) and the gallbladder in three (27.2%) patients. The objectives of drainage were to control jaundice in six (54.5%) and infection in six (54.5%) patients.
TABLE 1.
Characteristic | Patients (n = 11) |
---|---|
Gender (male/female) | 4/7 |
Age (years), median (range) | 73 (52–97) |
Primary biliary disease | |
Malignant (n = 7) | |
Pancreatic cancer | 4 |
Bile tract cancer | 2 |
Gastric cancer | 1 |
Benign (n = 4) | |
Cholecystitis | 3 |
Anastomosis stricture | 1 |
Target to drainage | |
Bile duct | 8 |
Gallbladder | 3 |
Purpose of drainage | |
Improvement of jaundice | 6 * |
Infection control | 6 * |
Presence of cholangitis before drainage | 4 |
Total bilirubin (mg/dL), median (range) | 1.3 (0.5–13.0) |
One case is duplicated.
Table 2 presents the details of each EUS‐BD. Among the patients, EUS‐guided hepaticogastrostomy (EUS‐HGS) was performed in four, EUS‐guided antegrade stenting (EUS‐AS) in three, EUS‐guided gallbladder drainage (EUS‐GBD) in three, and EUS‐guided hepaticoduodenostomy (EUS‐HDS) in one. Regarding tract dilation, the technical success rate was 100% (11/11), with a median dilation time of 40 s (range, 8–198). Subsequent stent placement was successful in all patients (11/11), with a median time from dilation to stent deployment of 10 min (range, 6–19). The median procedure time was 23 min (range, 14–95). Furthermore, clinical success was achieved in all patients (11/11) who presented with jaundice and cholangitis.
TABLE 2.
# | Disease | Procedure | Technical outcome | Puncture site | Diameter of the bile duct (mm) | Length of the hepatic parenchyma (mm) † | Time for dilation (s) |
---|---|---|---|---|---|---|---|
1 | Pancreatic cancer | EUS‐HGS | Success | B3 | 5.0 | 20 | 40 |
2 | Hilar cholangiocarcinoma | EUS‐HGS | Success | B3 | 4.0 | 35 | 20 |
3 | Pancreatic cancer | EUS‐HGS | Success | B3 | 4.5 | 21 | 56 |
4 | Anastomosis stricture | EUS‐HGS | Success | B3 | 4.0 | 14 | 8 |
5 | Hilar cholangiocarcinoma | EUS‐HDS | Success | B6 | 5.1 | 13 | 11 |
6 | Pancreatic cancer | EUS‐AS | Success | B3 | 4.7 | 20 | 110 |
7 | Gastric cancer | EUS‐AS | Success | B3 | 2.0 | 17 | 12 |
8 | Pancreatic cancer | EUS‐AS | Success | B3 | 4.0 | 25 | 9 |
9 | Acute cholecystitis | EUS‐GBD | Success | Gallbladder | ‐ | ‐ | 141 |
10 | Acute cholecystitis | EUS‐GBD | Success | Gallbladder | ‐ | ‐ | 169 |
11 | Acute cholecystitis | EUS‐GBD | Success | Gallbladder | ‐ | ‐ | 198 |
# | Clinical outcome | Stents used (diameter × length) | Time from dilation to stent deployment (min) | Any symptoms after the procedure | Early adverse event |
---|---|---|---|---|---|
1 | Success | 8 mm × 120 mm FCSEMS | 10 | ‐ | ‐ |
2 | Success | 8 mm × 120 mm FCSEMS | 10 | ‐ | ‐ |
3 | Success | 8 mm × 120 mm FCSEMS | 10 | Fever | Cholangitis |
4 | Success | 8 mm × 120 mm FCSEMS | 7 | ‐ | ‐ |
5 | Success | 6 Fr × 100 mm DPPS | 6 | ‐ | ‐ |
6 | Success | 8 mm × 60 mm UCSEMS (/8 mm 12 cm FCSEMS) ⁑ | 10 | ‐ | ‐ |
7 | Success | 8 mm × 60 mm UCSEMS (/7 Fr plastic stent 14 cm) ⁑ | 7 | ‐ | ‐ |
8 | Success | 8 mm × 80 mm FCSEMS (/8 mm 10 cm FCSEMS) ⁑ | 8 | ‐ | ‐ |
9 | Success | 8 mm × 80 mm FCSEMS | 12 | ‐ | ‐ |
10 | Success | 8 mm × 80 mm FCSEMS | 10 | ‐ | ‐ |
11 | Success | 8 mm × 120 mm FCSEMS | 19 | ‐ | ‐ |
Abbreviations: DPPS, double pigtail plastic stent; EUS‐AS, EUS‐guided antegrade stenting; EUS‐GBD, EUS‐guided gallbladder drainage; EUS‐HDS, EUS‐guided hepaticoduodenostomy; EUS‐HGS, EUS‐guided hepaticogastrostomy; FCSEMS, fully‐covered self‐expanded metallic stent; UCSEMS, uncovered self‐expanded metallic stent.
Length of the hepatic parenchyma was measured from the bile duct to the periphery of the liver on the EUS image.
(/), the stent placed for the dilated needle tract in EUS‐guided antegrade stenting.
Regarding the primary endpoint of early adverse events, non‐occlusion moderate cholangitis, necessitating 5 days of conservative treatment, occurred in one patient (9.1%) on the day after the procedure. Meanwhile, no cases of bile leakage or peritonitis were observed.
DISCUSSION
EUS‐BD has gained increasing importance in cases where conventional endoscopic retrograde cholangiopancreatography has been unsuccessful, particularly when access to the papilla is difficult due to duodenal obstruction or surgically altered anatomy. 7 , 8 Over the past decade, significant technical advancements and a growing body of clinical experience have been observed in the field of EUS‐BD. However, several important factors need to be addressed in terms of safety and technical aspects to optimize EUS‐BD.
The EUS‐BD involves puncture, guidewire placement, dilation of the needle tract, and stent deployment. Each of these steps is critical to the overall success of the procedure; particular attention should be given to the dilation of the needle tract, and stent deployment in terms of safety. Following the dilation, there is a risk for bile juice leakage until stent deployment, which can lead to infected biloma or severe peritonitis.
Several studies have investigated strategies to prevent bile leakage during EUS‐BD. Kawakubo et al. 9 demonstrated that the use of a covered metal stent is more effective than plastic stents in preventing bile leakage after EUS‐HGS; however, this strategy alone does not reduce bile leakage during the EUS‐BD. Ishiwatari et al. 10 reported that aspirating >10 mL bile before stent deployment during EUS‐HGS can reduce the risk for adverse events, including bile leakage and cholangitis. Similarly, Yamamoto et al. 11 suggested that maintaining a distance of ≥ 2.5 cm between the intrahepatic bile duct and the liver periphery during EUS‐HGS can reduce the incidence of bile peritonitis. However, regardless of the method used, a certain degree of bile leakage can still occur following tract dilation.
The development of a safer and more reliable modality for EUS‐BD is of utmost importance to promote its widespread adoption, particularly among novice interventional endoscopists. Tract dilation techniques vary across different facilities, including the use of an ultra‐tapered mechanical dilator, balloon catheter, or electrocautery dilator. 12 Honjo et al. reported that the use of mechanical dilators in EUS‐HGS has resulted in fewer adverse events and a high technical success rate. 13 However, despite tract dilation, passing a drainage stent, whether metal or plastic, through a needle tract can be challenging; this can be attributed to insertional axis deviation or inadequate tract dilation. Consequently, the procedure time may be prolonged and additional over‐dilation using an alternative device may be needed, resulting in an increased risk for adverse events.
Park et al. 14 emphasized the effectiveness of the one‐step metal stent placement technique using a tapered metal tip; this technique has been recommended to address safety and technical concerns. Although this technique demonstrated a one‐step technical success rate of 88% (14/16) without the need for additional dilation, tract dilation was still required in two patients. Therefore, we developed an improved stenting technique using an integrated device for tract dilation and a stent delivery system.
The novel endoscopic tapered sheath (EndoSheather) was initially designed to acquire biopsy samples to diagnose biliary strictures. 4 This device facilitates the smooth insertion of biopsy forceps to the targeted lesion during endoscopic retrograde cholangiopancreatography. It also features a tapered tip on the inner catheter and a closely fitting outer sheath, which ensures excellent pushability and serves as a dilation device. In addition, the outer sheath with a mesh braided structure provides optimal kink resistance, which makes it suitable for use as a delivery system. Therefore, the utilization of the tapered sheath for dilation and stent delivery during EUS‐BD can effectively address safety and technical concerns. Our study demonstrated a 100% technical success rate for tract dilation and stent deployment; moreover, all patients achieved clinical success. Notably, no cases of bile leakage or peritonitis were observed.
Our study had several limitations. First, it was a retrospective study conducted at a single center. Second, it had a small sample size. Third, there are areas for improvement in the EUS‐BD technique, particularly regarding the selection of stents. Future advancements in stent devices are expected to enhance the applicability of this technique. Despite these limitations, our study provides valuable insights into the safe and effective stenting in the EUS‐BD. However, a prospective randomized controlled trial involving a larger sample size is warranted to further validate our findings.
In conclusion, the use of the integrated device for tract dilation and stent delivery might provide a safe and straightforward technique for drainage stenting during EUS‐BD. Our innovative approach could contribute to the development of optimal methods for EUS‐BD.
CONFLICT OF INTEREST STATEMENT
None.
Supporting information
ACKNOWLEDGMENTS
We thank all physicians who supported the EUS‐BD procedure at the Nagoya City University Hospital.
REFERENCES
- 1. Will U, Thieme A, Fueldner F, Gerlach R, Wanzar I, Meyer F. Treatment of biliary obstruction in selected patients by endoscopic ultrasonography (EUS)‐guided transluminal biliary drainage. Endoscopy 2007; 39: 292–295. [DOI] [PubMed] [Google Scholar]
- 2. Guzmán‐Calderón E, Chacaltana A, Díaz R, Li B, Martinez‐Moreno B, Aparicio JR. Head‐to‐head comparison between endoscopic ultrasound guided lumen apposing metal stent and plastic stents for the treatment of pancreatic fluid collections: A systematic review and meta‐analysis. J Hepatobiliary Pancreat Sci 2022; 29: 198–211. [DOI] [PubMed] [Google Scholar]
- 3. Matsumori T, Uza N, Shiokawa M et al. Mapping biopsy for bile duct cancer using a novel device delivery system. Endoscopy 2021; 54: E217–219. [DOI] [PubMed] [Google Scholar]
- 4. Matsumori T, Uza N, Okada H et al. Innovative method for the diagnosis of bile duct lesions using a novel tapered‐tip sheath system to facilitate biliary biopsies. Gastrointest Endosc 2023; 98: 43–50.e1. [DOI] [PubMed] [Google Scholar]
- 5. Cotton PB, Eisen GM, Aabakken L et al. A lexicon for endoscopic adverse events: Report of an ASGE workshop. Gastrointest Endosc 2010; 71: 446–454. [DOI] [PubMed] [Google Scholar]
- 6. Isayama H, Hamada T, Yasuda I et al. TOKYO criteria 2014 for transpapillary biliary stenting. Dig Endosc 2015; 27: 259–264. [DOI] [PubMed] [Google Scholar]
- 7. Giovannini M, Moutardier V, Pesenti C, Bories E, Lelong B, Delpero JR. Endoscopic ultrasound‐guided bilioduodenal anastomosis: A new technique for biliary drainage. Endoscopy 2001; 33: 898–900. [DOI] [PubMed] [Google Scholar]
- 8. Meng FS, Zhang ZH, Ji F. Therapeutic role of endoscopic ultrasound in pancreaticobiliary disease: A comprehensive review. World J Gastroenterol 2015; 21: 12996–13003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Kawakubo K, Isayama H, Kato H et al. Multicenter retrospective study of endoscopic ultrasound‐guided biliary drainage for malignant biliary obstruction in Japan. J Hepatobiliary Pancreat Sci 2014; 21: 328–334. [DOI] [PubMed] [Google Scholar]
- 10. Ishiwatari H, Satoh T, Sato J et al. Bile aspiration during EUS‐guided hepaticogastrostomy is associated with lower risk of postprocedural adverse events: A retrospective single‐center study. Surg Endosc 2021; 35: 6836–6845. [DOI] [PubMed] [Google Scholar]
- 11. Yamamoto Y, Ogura T, Nishioka N et al. Risk factors for adverse events associated with bile leak during EUS‐guided hepaticogastrostomy. Endosc Ultrasound 2020; 9: 110–115. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Ogura T, Higuchi K. Endoscopic ultrasound‐guided hepaticogastrostomy: Technical review and tips to prevent adverse events. Gut Liver 2021; 15: 196–205. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Honjo M, Itoi T, Tsuchiya T et al. Safety and efficacy of ultra‐tapered mechanical dilator for EUS‐guided hepaticogastrostomy and pancreatic duct drainage compared with electrocautery dilator (with video). Endosc Ultrasound 2018; 7: 376–382. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Park DH, Lee TH, Paik WH et al. Feasibility and safety of a novel dedicated device for one‐step EUS‐guided biliary drainage: A randomized trial. J Gastroenterol Hepatol 2015; 30: 1461–1466. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.