Abstract
As the mechanism for worse prognosis after cardiac resynchronization therapy (CRT) upgrades in heart failure patients with RVP dependence (RVP-HF) has clinical implications for patient selection and CRT implementation approaches, this study’s objective was to evaluate prognostic implications of cardiac magnetic resonance (CMR) findings and clinical factors in 102 HF patients (23.5% female, median age 66.5 years old, median follow-up 4.8 years) with and without RVP dependence undergoing upgrade and de novo CRT implants. Compared with other CRT groups, RVP-HF patients had decreased survival (p = 0.02), more anterior late-activated LV pacing sites (p = 0.002) by CMR, more atrial fibrillation (p = 0.0006), and higher creatinine (0.002). CMR activation timing at the LV pacing site predicted post-CRT LV functional improvement (p < 0.05), and mechanical activation onset < 34 ms by CMR at the LVP site was associated with decreased post-CRT survival in a model with higher pre-CRT creatinine and B-type natriuretic peptide (AUC 0.89; p < 0.0001); however, only the higher pre-CRT creatinine partially mediated (37%) the decreased survival in RVP-HF patients. In conclusion, RVP-HF had a distinct CMR phenotype, which has important implications for the selection of LV pacing sites in CRT upgrades, and only chronic kidney disease mediated the decreased survival after CRT in RVP-HF.
Keywords: cardiac magnetic resonance, heart failure, cardiac resynchronization therapy, pacemakers, natriuretic peptides
1. Introduction
Between 1993 and 2009, the number of patients implanted with permanent pacemakers increased by 55.6% in the United States, and 2.9 million patients overall received permanent pacemakers during this period [1]. Although conduction system pacing is increasing in popularity, RV pacing (RVP) is still more commonly implemented using active or passive fixation leads implanted in the ventricular septum, typically near the RV apex, which leads to the need for cardiac resynchronization therapy [2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20] in patients who develop heart failure [21]. Unfortunately, several studies have demonstrated worse survival after CRT upgrades for heart failure associated with RVP (RVP-HF) compared with survival in patients undergoing de novo CRT implants [22,23], although this finding has not been consistent across all studies [24,25]. Chronic kidney disease and atrial fibrillation in patients with CRT upgrades have been suggested as possible explanations for this difference in prognosis [22,23].
As most cardiac implantable electronic devices are now MR-conditional [26,27], and cardiac magnetic resonance (CMR) has been shown to provide important prognostic data in patients with heart failure [28,29], including those undergoing CRT [30,31], we designed a clinical study to evaluate differences in clinical and CMR parameters with respect to prognosis in patients undergoing CRT defibrillator upgrades for RVP-HF, CRT defibrillator upgrades for other indications, and de novo CRT defibrillator implants. The hypothesis was that selected CMR and clinical findings would be different in these patient groups and provide a mechanism for the decreased survival observed in multiple cohorts of patients undergoing CRT upgrade implants compared with those undergoing de novo CRT implants [22,23].
2. Materials and Methods
This study was approved by the Institutional Review Board for Human Subjects Research at the University of Virginia and included the enrollment of patients in different cohorts. Inclusion criteria for Group 1 required that patients were undergoing a de novo CRT defibrillator (CRT-D), and inclusion criteria for CRT upgrade patients required that they were undergoing an upgrade to a CRT defibrillator system with a coronary venous LV pacing lead for either RVP-HF (Group 2) or another accepted indication (Group 3). Other inclusion criteria for all patients were that they met published criteria for CRT [32] and were willing to have a CMR prior to the CRT procedure. The GFR had to be 45 cc/min/m2 before CRT to have a gadolinium-based contrast agent (GBCA) for research CMR at our institution, and the GBCA was withheld in the case of a lower GFR at the time of the scan. Exclusion criteria included contraindications for MRI, persistent atrial fibrillation without third-degree AV block, and a PVC burden greater than 15%. Patients were enrolled between 2011 and 2021.
Prior to CRT device implantation, clinical characteristics for all patients were collected through intake forms and cross-referenced with electronic health records. These included demographics, medications, comorbid conditions, the QRS duration from the 12-lead electrocardiogram, and relevant laboratory findings, as described in Table 1. Clinical parameters were also integrated as the Seattle Heart Failure Model score, as previously described [33].
Table 1.
All (N = 102) | De Novo CRT (N = 50) |
CRT Upgrade with RVP (N = 29) |
CRT Upgrade w/o RVP (N = 23) |
p Value (All Groups) |
p Value (Group 1 vs. 2) |
|
---|---|---|---|---|---|---|
Demographics | ||||||
Age, years | 66.5 (58.3–72.9) | 65.4 (60.4–70.8) | 74.0 (64.0–79.0) | 62.0 (56.0–70.0) | 0.02 | 0.03 |
BMI, kg/m2 | 28.3 (24.2–33.2) | 28.0 (24.0–31.7) | 28.2 (23.3–33.1) | 29.1 (25.4–35.5) | 0.8 | |
Weight, kg | 85.3 (75.3–103.0) | 82.6 (73.6–99.8) | 91.6 (73.9–103.0) | 93.8 (80.2–103.9) | 0.5 | |
Female | 24 (23.5) | 15 (30.0) | 3 (10.3) | 6 (26.1) | 0.1 | |
NYHA Heart Failure Class | <0.0001 | <0.0001 | ||||
II | 35 (34.3) | 3 (6.0) | 18 (62.1) | 14 (60.9) | ||
III | 67 (65.7) | 47 (94.0) | 11 (37.9) | 9 (39.1) | ||
Race | 0.5 | |||||
Black | 16 (15.7) | 6 (12.0) | 5 (17.2) | 5 (21.7) | ||
White/Other | 86 (84.3) | 44 (88.0) | 24 (82.8) | 18 (78.3) | ||
SHFM Score | 0.38 (0.005–0.81) | 0.55 (0.22–0.89) | 0.19 (−0.01–0.79) | 0.2 (−0.015–0.41) | 0.1 | |
Comorbid Conditions | ||||||
Ischemic Cardiomyopathy | 49 (48.0) | 23 (46.0) | 11 (37.9) | 15 (65.2) | 0.1 | |
Atrial Fibrillation | 33 (32.4) | 11 (22.0) | 18 (62.1) | 4 (17.4) | 0.0004 | 0.0006 |
Chronic Kidney Disease | 36 (35.3) | 14 (28.0) | 14 (48.3) | 8 (34.8) | 0.2 | |
Prior CABG | 27 (26.5) | 11 (22.0) | 7 (24.1) | 9 (39.1) | 0.3 | |
Medications | ||||||
Beta-Blocker | 96 (94.1) | 49 (98.0) | 25 (86.2) | 22 (95.7) | 0.07 | |
ACE Inhibitor/ARB | 85 (83.3) | 46 (92.0) | 19 (65.5) | 20 (87.0) | 0.01 | 0.005 |
Loop Diuretic | 79 (77.5) | 40 (80.0) | 18 (62.1) | 21 (91.3) | 0.04 | 0.1 |
Statin | 67 (65.7) | 28 (56.0) | 19 (65.5) | 20 (87.0) | 0.04 | 0.5 |
Laboratory Studies, Vital Signs & Exercise Testing | ||||||
Systolic BP, mmHg | 118.5 (102.0–130.0) | 118.0 (108.0–129.8) | 127.0 (110.0–138.0) | 107.0 (96.5–120.0) | 0.01 | 0.2 |
Sodium, mEq/L | 138.0 (136.3–140.0) | 138.0 (137.0–139.8) | 139.0 (138.0–141.0) | 137.0 (136.0–139.0) | 0.1 | |
Creatinine, mg/dL | 1.2 (0.96–1.4) | 1.05 (0.9–1.3) | 1.3 (1.2–1.6) | 1.1 (0.95–1.4) | 0.004 | 0.002 |
Hemoglobin, g/dL | 13.3 (12.3–14.5) | 13.6 (12.3–14.4) | 13.1 (12.3–14.5) | 13.1 (12.4–14.6) | 0.9 | |
GFR, mL/min/1.72 m2 | 64.1 (51.5–82.6) | 72.2 (59.1–87.0) | 58.0 (43.0–74.0) | 63.0 (54.0–77.0) | 0.02 | 0.006 |
BNP, pg/mL | 272.0 (137.3–752.8) | 210.5 (118.3–699.0) | 272.0 (124.0–879.0) | 304.0 (262.0–642.0) | 0.2 | |
Peak VO2, mL/kg/min | 0.075 (−0.88–1.4) | 14.3 (12.2–16.1) | 14.4 (13.4–14.8) | 14.4 (12.7–14.6) | 0.9 | |
CMR/Echocardiography
Assessment Parameters |
||||||
LVEF, % | 23.7 (17.4–29.8) | 23.2 (18.9–28.0) | 29.0 (20.0–35.0) | 19.0 (16.5–26.0) | 0.02 | 0.06 |
LVEDVI, mL/m2 | 127.9 (104.8–165.0) | 133.7 (112.2–170.2) | 102.0 (77.4–115.0) | 152.0 (126.2–183.5) | <0.0001 | 0.0002 |
LVESVI, mL/m2 | 97.2 (73.6–127.0) | 101.2 (84.9–134.9) | 71.3 (49.9–90.4) | 117.0 (95.9–148.5) | <0.0001 | 0.0005 |
RVEF, % | 36.1 (26.0–45.5) | 36.1 (24.4–48.7) | 37.5 (32.0–44.6) | 35.1 (27.8–40.1) | 0.7 | |
RVEDVI, mL/m2 | 65.8 (55.3–83.7) | 66.0 (53.5–81.2) | 65.3 (55.5–86.9) | 65.8 (57.3–79.3) | 0.9 | |
RVESVI, mL/m2 | 40.2 (32.7–55.2) | 42.0 (31.8–57.7) | 38.7 (31.6–52.8) | 39.6 (38.1–50.3) | 0.9 | |
LGE Presence | 55 (53.9) | 27 (54.0) | 14 (48.3) | 14 (60.9) | 0.7 | |
CURE-SVD | 0.57 (0.45–0.72) | 0.60 (0.45–0.78) | 0.55 (0.50–0.69) | 0.55 (0.37–0.67) | 0.4 | |
Scar at LV Free Wall | 29 (28.4) | 15 (30.0) | 5 (17.2) | 9 (39.1) | 0.2 | |
Electrical Parameters | ||||||
Paced QRS | 31 (30.4) | 0 (0.0) | 29 (100.0) | 0 (0.0) | <0.0001 | <0.0001 |
QRS, ms | 161.0 (147.0–180.0) | 160.0 (146.3–180.0) | 179.0 (164.0–200.0) | 152.0 (136.0–168.0) | 0.0004 | 0.002 |
QLV, ms | 130.0 (98.0–154.3) | 112.5 (81.3–140.0) | 160.0 (150.0–172.0) | 108.0 (99.0–127.0) | <0.0001 | <0.0001 |
QLV/QRS Ratio | 0.78 (0.67–0.89) | 0.74 (0.58–0.81) | 0.89 (0.84–0.96) | 0.76 (0.69–0.79) | <0.0001 | <0.001 |
LBBB | 67 (65.7) | 48 (96.0) | 0 (0) | 16 (69.6) | <0.0001 | <0.0001 |
RBBB | 7 (6.9) | 3 (6.0) | 0 (0) | 3 (13.0) | 0.5 | |
TOS at LV Lead, ms | 87.5 (69.0–114.8) | 106.0 (72.3–118.0) | 86.0 (66.0–107.0) | 78.0 (68.0–96.5) | 0.06 | 0.1 |
TOS at Latest Activated LV Segment, ms |
119.5 (101.0–136.0) | 124.0 (97.3–136.0) | 112.0 (100.0–134.0) | 119.0 (104.0–141.0) | 0.7 | |
Response Measures
at 6 Months Post-CRT |
||||||
Fractional Change in LVESVI |
−0.2 (−0.32–−0.08) | −0.18 (−0.36–−0.023) | −0.21 (−0.36–−0.09) | −0.2 (−0.265–−0.135) | 0.7 | |
BNP, pg/mL | 172.5 (63.8–442.8) | 132.5 (51.8–451.8) | 172.5 (70.0–526.0) | 195.0 (150.5–238.0) | 0.6 | |
Change in Peak VO2, mL/kg/min |
0.075 (−0.88–1.37) | 0.40 (−1.5–2.3) | −0.26 (−0.9–0.4) | 0.075 (−0.03–0.3) | 0.3 | |
Survival Status at 4 Years | 0.2 | |||||
Alive | 83 (81.4) | 44 (88.0) | 21 (72.4) | 18 (78.2) | ||
Dead | 19 (18.6) | 6 (12.0) | 8 (27.6) | 5 (21.8) |
Values are median (interquartile range) or n (%). ACE = angiotensin-converting enzyme; ARB = angiotensin receptor blocker; BMI = body mass index; BNP = B-type natriuretic peptide; BP = blood pressure; CABG = coronary artery bypass graft; CURE-SVD = circumferential uniformity ratio estimate with singular value decomposition; GFR = glomerular filtration rate; LBBB = left bundle branch block; LGE = late gadolinium enhancement; LVEDVI = left ventricular end-diastolic volume index; LVEF = left ventricular ejection fraction; LVESVI = left ventricular end-systolic volume index; NYHA = New York Heart Association; QLV = QRS-LV electrogram time; RBBB = right bundle branch block; RVEDVI = right ventricular end-diastolic volume index; RVEF = right ventricular ejection fraction; RVESVI = right ventricular end-systolic volume index; SHFM = Seattle Heart Failure Model; TOS = time to the onset of circumferential shortening.
Before the CRT procedure, patients underwent cardiac magnetic resonance (CMR) on a 1.5 T scanner. The MRI protocol included cine imaging, late gadolinium enhancement for myocardial scar detection, and strain imaging with displacement encoding with stimulated echoes (DENSE). Late gadolinium enhancement (LGE) was performed ten minutes after the injection of 0.1 mmol/kg of gadoterate meglumine (Dotarem), and a wideband pulse (bandwidth of 3.8 kHz) was added as needed [34] to minimize off-resonance artifacts in patients with cardiac implantable electronic devices at the time of the scan. Volumetric and scar analyses of CMR images were performed using suiteHEART Software v5.1.0 (Neosoft, Pewaukee, WI, USA).
Regional circumferential strain was acquired using cine DENSE [35] in short-axis planes at basal, mid-ventricular, and apical levels using a temporal resolution of 17 ms, pixel size of 2.8 × 2.8 mm2, and slice thickness of 8 mm. Displacement was encoded in two orthogonal directions, and a spiral k-space trajectory was used with 6 interleaves per image. Other parameters included: field of view = 350 × 350 mm2, displacement encoding frequency ke = 0.1 cycles/mm, flip angle = 15°, and echo time = 1.9 ms. DENSE images were analyzed using custom software. In a small number of patients in whom off-resonance artifacts were present on DENSE imaging, circumferential strain (Ecc) was derived using a convolutional neural network (Strain-Net) designed by our group to predict regional strain from cine imaging and validated in a recently published study demonstrating more accurate regional strain assessments compared with commercially available feature tracking [36].
The CURE-SVD, a robust and validated mechanical dyssynchrony parameter (range: 0–1; 1 = greatest synchrony) was derived from Ecc as previously reported [37,38]. Regional mechanical activation was derived from CMR mechanical activation maps, which were created as previously reported for all patients from a regional Ecc matrix [30,37,38]. This matrix was derived from midwall Ecc based on cardiac segment and phase, and an active contour guided by the Ecc gradient was used to automatically detect the mechanical activation time field, which was defined as the time to the onset of circumferential shortening [38]. Intuitively, this is the time to the detection of a negative slope in a regional Ecc curve, as a negative slope of the Ecc curve indicates myocardial shortening. In order to calculate mechanical activation at the left ventricular pacing (LVP) site, the lead location was mapped to an LV segment (American Heart Association 17-segment model) based on procedural fluoroscopy using previously described methods [30,39,40,41].
Echocardiography was used to calculate the fractional change in left ventricular end-systolic volume index (LVESVI-FC) as: (LVESVIPost-CRT − LVESVIPre-CRT)/LVESVIPre-CRT; negative values are favorable, and −0.15 is often used as a cutoff for response in the literature when there is a need to convert the LVESVI-FC to a binary response indicator [31]. Cardiopulmonary exercise testing measured peak VO2 before and after CRT, and measurements of BNP were used to consider the neurohormonal axis in CRT response.
Patients were followed for a median of 4.8 years for all-cause mortality up until 1 April 2023 through a review of the electronic health record. All patients enrolled were followed at the University of Virginia Health System, and their vital status was maintained in the electronic health record.
All statistical analyses were performed using R and Python. Statistical tests were performed to identify differences in baseline characteristics between the three upgrade/de novo CRT groups. Kruskal–Wallis tests were used to compare continuous variables, as not all data were normally distributed, and Dunn’s post hoc nonparametric test was used for pairwise comparisons. Chi-square tests and Fisher’s exact tests were used to compare differences in discrete variables. Reverse Kaplan–Meier curves stratified by cohort group were constructed, and the log-rank p-values were calculated. Median follow-up was determined based on the reverse Kaplan–Meier analysis. Logistic regression with least absolute shrinkage and selection operator for L1-regularization (LASSO logistic regression) was performed using the glmnet package to identify the predictors of survival. The tuning parameter λ was used based on expert recommendations [42] to eliminate all but four covariates. The model was then derived using logistic regression in a 70% training set from the cohort, and prediction accuracy was evaluated based on the model derived in the training set in a 30% test cohort. Survival at 4 years was chosen as the outcome for this logistic regression model based on the median follow-up time. Receiver operating characteristic (ROC) curves were plotted for performance in the model in the combined training and test sets with exact Pearson/Clopper 95% confidence intervals at three points of interest. A nomogram was also generated for the overall logistic regression model with four covariates for 4-year survival. These covariates were also evaluated in Cox proportional hazards models, for which hazard ratios, 95% confidence intervals, and p-values were reported.
Mediation analysis was performed to determine the extent to which the effect of an independent variable (the device group in this study) was mediated by predictors such as impaired renal function. The proposed mediator (creatinine) was regressed on the independent variable using linear regression, and survival/survival time was regressed on the mediator using a parametric survival model with a Gaussian distribution. The average causal mediation effect (ACME) was determined as the product of the regression coefficient for the effect of the independent variable on the mediator in the linear regression model and the regression coefficient for the effect of the mediator on survival time in 1000 bootstrapped samples. The method then facilitated the calculation of 95% confidence intervals for the mediator effect. This mediator effect was compared with the effect of the independent variable on survival/survival time.
3. Results
3.1. Baseline Characteristics and Response Measures of Entire Patient Cohort
The baseline characteristics for the 102 patients (median age 66.5 years old with interquartile range [IQR] 58.3 to 72.9; 23.5% female) are shown in Table 1.
During a median follow-up of 4.8 years (IQR 4.2 to 5.0 years), 26 (25.5%) patients died. The 4-year death rate was 18.7% (19/102). Group 2 patients were older (p = 0.03), had a higher frequency of atrial fibrillation (p = 0.0006), had greater pre-CRT QRS durations (p = 0.002), and had higher creatinine values (p = 0.002). BNP values were not significantly different across groups, although there was a trend for greater BNP in Group 2 vs. Group 1 (Figure 1).
3.2. Mechanical and Electrical Activation at the Left Ventricular Pacing Site by Upgrade/De Novo Group
With respect to CMR structural findings, in comparison with Group 1, Group 2 patients had a significantly smaller baseline LVESVI (p = 0.0005), a smaller baseline LVEDVI (p = 0.0002), and greater LVEF (p = 0.06) (Figure 2).
The proportion with the favorable finding of mechanical activation greater than 34 ms at the LVP site was similar in all three groups (p = 0.31 vs. Group 1 and p = 0.80 vs. Group 3). (The cutoff of 34 ms was identified as the optimal multiple of 17 ms [DENSE temporal resolution] based on the Youden index to maximize sensitivity and specificity for 4-year survival.) As shown in Table 2 and Figure 3, Group 2 patients had a prominent anterior shift of late-activated sites compared with Group 1 (48% vs. 10% anterior, 42% vs. 52% anterolateral, and 10% vs. 38% inferolateral; p = 0.0002) with a similar distribution of LV free wall scar (30% vs. 17%, p = 0.2).
Table 2.
De Novo CRT (N = 50) |
CRT Upgrade with RVP (N = 29) |
CRT Upgrade w/o RVP (N = 23) |
p Value (All Groups) |
p Value (Group 1 vs. 2) |
|
---|---|---|---|---|---|
LV Lead Location | |||||
Longitudinal | 0.4 | ||||
Basal | 9 (18.0) | 7 (24.1) | 9 (39.1) | ||
Mid-ventricular | 29 (58.0) | 17 (58.6) | 11 (47.8) | ||
Apical | 12 (24.0) | 5 (17.2) | 3 (13.0) | ||
Circumferential | 0.6 | ||||
Anterior | 8 (16.0) | 1 (3.4) | 2 (8.7) | ||
Anteroseptal | 0 (0.0) | 0 (0.0) | 0 (0.0) | ||
Inferoseptal | 0 (0.0) | 0 (0.0) | 0 (0.0) | ||
Inferior | 1 (2.0) | 0 (0.0) | 1 (4.3) | ||
Inferolateral | 12 (24.0) | 7 (24.1) | 5 (21.7) | ||
Anterolateral | 29 (58) | 21 (72.4) | 15 (65.2) | ||
Location of Latest
Mechanical Activation |
|||||
Longitudinal | <0.0001 | ||||
Basal | 14 (28.0) | 25 (86.2) | 16 (69.6) | <0.0001 | <0.0001 |
Mid-ventricular | 31 (62.0) | 4 (13.8) | 6 (26.1) | <0.0001 | <0.0001 |
Apical | 5 (10.0) | 0 (0.0) | 1 (4.3) | 0.6 | |
Circumferential | 0.003 | ||||
Anterior | 5 (10.0) | 14 (48.3) | 9 (39.1) | 0.0002 | 0.0002 |
Anteroseptal | 0 (0.0) | 0 (0.0) | 0 (0.0) | ||
Inferoseptal | 0 (0.0) | 0 (0.0) | 0 (0.0) | ||
Inferior | 1 (2.0) | 1 (3.4) | 1 (4.3) | 0.8 | |
Inferolateral | 17 (34.0) | 2 (6.9) | 7 (30.4) | 0.02 | 0.007 |
Anterolateral | 27 (54) | 12 (41.4) | 6 (26.1) | 0.08 |
Values are n (%).
In separate linear regression models in the combined cohort of de novo CRT and CRT upgrade patients, a longer Q-LV time was associated with a greater improvement in the LVESVI after CRT (p = 0.020) after adjustment for CURE-SVD (p = 0.0085), which was also a significant predictor of this short-term outcome. LV free wall scar with LGE was associated with this response indicator alone (p = 0.057), but not after adjustment for either the Q-LV, the CURE-SVD, or both. For the outcome of improvement (lowering) in BNP post-CRT with adjustment for the pre-CRT value, both CURE-SVD (p = 0.037) and mechanical activation ≥ 34 ms at the LV lead implant site (p = 0.042) were significant predictors in a multivariable linear regression model. Both the Q-LV (p = 0.043) and CMR mechanical activation (p = 0.0039) were independently associated with survival time after CRT in a multivariable Cox proportional hazards regression model; however, only the mechanical activation predictor was associated with survival time after CRT after adjustment for pre-CRT BNP. Lead placement in anterior, anterolateral, posterolateral, or mid-apical-lateral AHA segments did not have an additional association in any of these models. No interactions with group assignments were noted in any of the models.
3.3. All-Cause Mortality by Group
In a Cox proportional hazards model, Group 2 patients had worse survival compared with Group 1 patients (HR 3.26 [95% CI 1.18–9.04], p = 0.023) (Figure 4A) but similar survival compared with Group 3 patients (Figure 4B). Comparing Group 1 and Group 2 patients, the six-month LVESVI-FC post-CRT (p = 0.006), six-month BNP post-CRT (p = 0.002), and six-month change in peak VO2 post-CRT (p = 0.02) were associated with long-term survival after adjustment for CRT group (p = 0.0002, p < 0.0001, and p = 0.002, respectively, in three separate models).
3.4. L1-Regularization for Selection of Covariates Associated with Survival
A plot of the LASSO regression coefficient values as a function of the tuning parameter (λ) is shown in Figure 5. Using a tuning parameter of 0.08, the following four parameters were identified for survival prediction: pre-CRT BNP, pre-CRT creatinine, history of CABG surgery, and a CMR activation time at the LV lead implant site ≥ 34 ms. Although Group 2 was associated with decreased 4-year survival, the other four covariates had more significant associations in the LASSO regression model. Of note, although patients in Group 2 were older, age was not a significant predictor of survival (p = 0.32).
3.5. Training and Test Sets with Standard Logistic Regression
The cohort was then split into a 70% training set and a 30% test set, and a standard logistic regression model with these four covariates was derived in just the training set. Predictions from this model in the test set were compared with actual events, and the predictions had an accuracy of 93.3% for predicting 4-year survival in the test set.
3.6. Receiver Operating Characteristic Analysis and Nomogram
Receiver operating characteristic analysis from a logistic regression model with these four covariates in all 102 patients demonstrated a model AUC of 0.89 (p < 0.0001) (Figure 6A). The model is shown in Table S1. For clinical application, a nomogram to estimate 4-year survival based on this logistic regression model for 4-year survival is provided in Figure 7. For reference, a separate logistic regression model with two parameters, (1) the Seattle Heart Failure Model (SHFM) score substituted for the clinical covariates (BNP, creatinine, and prior CABG surgery) and (2) CMR activation time at the LV lead implant site ≥ 34 ms, had an AUC of 0.77 for 4-year survival (Figure 6B).
3.7. Kaplan–Meier Analysis and Cox Proportional Hazards Analysis
In a Cox proportional hazards model with 26 death events, the pre-CRT creatinine, the pre-CRT BNP, CABG history, and mechanical activation ≥ 34 ms were all associated with survival in the multivariable model (Table S2). Creatinine had the greatest association with survival in this model with a hazard ratio HR of 5.78 per mg/dL (p < 0.0001). Baseline BNP (p = 0.012), CABG (p = 0.020), and CMR mechanical activation time ≥ 34 ms at the LV pacing site (p = 0.040) were also significant predictors in the model. Reverse Kaplan–Meier curves are shown in Figure 8 for baseline BNP stratified by the median (A), baseline creatinine stratified by the median (B), prior CABG (C), and mechanical activation ≥ 34 ms (D). For reference, the corresponding survival model with the SHFM score and CMR mechanical activation time ≥34 ms at the LV pacing site is shown in Table S3. The hazard ratio for the SHFM in this model was HR 2.81 per unit increase in SHFM score (p = 0.0059).
3.8. Mediation Effects
The effect of a CRT upgrade with RVP-HF relative to de novo CRT on the likelihood of survival was partially mediated by the pre-CRT creatinine. The regression coefficient for the total effect of the CRT upgrade with RVP-HF on survival time using parametric survival regression with a Gaussian distribution was −2.302 (p = 0.034). As Figure 9 illustrates, the causal mediation effect was (0.361) × (−2.37) = −0.855. The bootstrapped 95% confidence interval for the mediator effect was −1.89 to −0.19 (p = 0.008), and the bootstrapped confidence interval for the total effect was −3.85 to −0.25 (p = 0.034). There was a 37% mediation effect of creatinine for the RVP-HF group effect on survival time (−0.855/−2.302 = 0.371). Other candidate baseline characteristics and response measures did not have a significant average causal mediation effect.
4. Discussion
This study offers a novel presentation of differences in prognosis, CMR structural findings, comorbidities, neurohormonal activation, renal function, and electromechanical differences at the LV pacing site among patients undergoing CRT upgrades with and without baseline RV pacing dependence versus de novo CRT patients. An important clinical finding is that baseline renal function and levels of neurohormones are the predominant drivers of survival after both de novo CRT implants and CRT upgrades with a significant contribution from whether late CMR mechanical activation is achieved at the left ventricular (LV) pacing site. These findings also present a roadmap to identify prognosis in patients with CRT upgrades, and a nomogram is provided for this purpose. Moreover, the worse survival in upgrade RVP-HF patients in this cohort was mediated significantly by a comorbid condition, specifically renal impairment, in the upgrade RVP-HF cohort. Other comorbid conditions more common in CRT upgrade patients for RVP-HF in this cohort, such as atrial fibrillation, were found not to mediate the decreased survival as impaired kidney function did.
With respect to prognosis in CRT upgrades, prior studies have yielded partially conflicting results. While some studies found similar overall survival in de novo and upgrade CRT patients [24,25], others found that upgrade CRT patients had worse survival [22,23]. CRT upgrade patients were noted to have an increased prevalence of atrial fibrillation and more impaired kidney function, as we found in the present cohort. Furthermore, CRT upgrade patients with RVP-HF in our cohort had smaller baseline LV volumes compared with the CRT upgrade patients having other indications and de novo CRT patients. In many ways, upgrade patients without RVP-HF looked more like de novo patients than upgrade patients with RVP-HF, which highlights the importance of distinguishing upgrade patients with and without RV pacing dependence.
It is remarkable that the Budapest Upgrade CRT randomized trial showed a clinical benefit for upgrades to CRT devices in patients with RV pacing dependence [43]. The findings in our study are consistent with those from the Budapest Upgrade CRT trial because the comparison groups in our study for the patients with RV pacing dependence and CRT upgrades were patients with either de novo CRT implants or patients with CRT upgrades not dependent on RV pacing. In contrast, the control group for the Budapest Upgrade CRT trial consisted of patients with RV pacing dependence who did not receive CRT upgrades through randomization.
With respect to implant findings and prognosis, LV free wall sites having later mechanical and electrical activation timing represent the areas in the LV that are most dysfunctional and typically undergoing the most stretch in early systole, such that pacing these sites is likely to have the most beneficial effect on LV function. In this cohort, we found the electrical activation at LVP site, mechanical activation at the LVP site, and overall mechanical dyssynchrony based on the CURE-SVD mechanical dyssynchrony index were prognostically important across de novo CRT and upgrade groups. The finding that electrical and mechanical activation at the LVP site both contribute favorably to response and survival in this mixed cohort highlights the importance of achieving both late electrical and late mechanical activation to achieve optimal CRT response rates.
Another remarkable finding in this cohort is the anterior shift in late-activated sites observed for CRT upgrade patients with RVP-HF. This can be understood intuitively by the fact the RV lead position is typically in the area of seven o’clock on the LV short-axis views, and the anterior LV at the one o’clock position is the most distant from the RV pacing lead along the circumference of LV short axis, thus representing the most late-activated site. This anterior shift in late activation in RVP-HF has important implications for the CRT implant procedure in patients with CRT upgrades with RV pacing in this area. Although the LV pacing site chosen was at the discretion of the operator in this study, and this study was not a prospective interventional trial of CMR guidance for the selection of the LVP site, the prognostic importance of mechanical and electrical late activation at the LVP site in this cohort suggests that CRT upgrade patients with RVP-HF may have better results if operators target more anterolateral sites, and more anterior sites may even be good pacing targets in this scenario. In many patients, the anterior or anterolateral LV may be easier to access from the coronary sinus and may be less likely to have phrenic nerve capture; however, operators implanting CRT in patients with LBBB typically strive to achieve a more posterolateral LV lead position if there is not scar in this area. Furthermore, sometimes coronary venous pacing options are limited, and an anterior lead position may be the only one that works.
Limitations
There were several limitations of this study. Although a cohort of 102 patients with curated CMR, clinical, laboratory, response, and survival data with a median follow-up over 4 years distributed over three groups is considered to be reasonably large and sufficient for the mechanistic study presented, a larger cohort would have facilitated additional approaches based on neural networks and other machine learning methods; however, it is reassuring that the main model had excellent accuracy in a test set when derived in a training set. Second, although patients were on maximally tolerated medical therapy appropriate at the time of enrollment, angiotensin-neprilysin inhibitors and sodium-glucose cotransporter-2 inhibitors have since become widely used medications for heart failure. While this is an inherent limitation for any long-term study because new medications may well be introduced during the five years of follow-up, the impact on the results of this study is likely very modest, as sodium-glucose cotransporter-2 inhibitors decrease heart failure and other cardiac events with minimal impact on survival [25], and heart failure risk models such as the SHFM have been shown to perform equally well in studies of angiotensin-neprilysin inhibitors [26] compared with other cohorts. Third, although conduction system pacing patients were not included in this analysis, the indication in heart failure with reduced ejection fraction is just now evolving, and the approach of incorporating CMR and neurohormonal/clinical data together with the statistical methodology presented in this paper can also be applied to future cohorts with conduction system pacing.
5. Conclusions
In conclusion, RVP-HF had a distinct CMR phenotype, which has important implications for the selection of LV pacing sites in CRT upgrades, and only chronic kidney disease mediated the decreased survival after CRT in RVP-HF.
Supplementary Materials
The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/jcdd10100409/s1, Table S1: Supplemental Logistic Regression Model; Table S2: Supplemental Cox Proportional Hazards Regression Model #1; Table S3: Supplemental Cox Proportional Hazards Regression Model #2.
Author Contributions
Conceptualization, D.J.B., P.J.A.O., F.H.E. and K.C.B.; methodology, D.J.B., P.J.A.O., F.H.E. and K.C.B.; validation, D.J.B., P.J.A.O., F.H.E., K.C.B. and Y.W.; formal analysis, D.J.B., P.J.A.O., F.H.E., K.C.B., Y.W., M.A., X.G. and F.L.M.; data curation, D.J.B., P.J.A.O., F.H.E., K.C.B., X.G. and F.L.M.; writing—original draft preparation, D.J.B., P.J.A.O., F.H.E. and K.C.B.; writing—review and editing, D.J.B., P.J.A.O., Y.W., F.L.M., M.A., X.G., R.M., A.D., N.M., O.J.M., J.M.M., P.K.M., W.C.L., S.M., A.R.P., F.H.E. and K.C.B. visualization, D.J.B., P.J.A.O. and K.C.B. supervision, K.C.B. and F.H.E.; project administration, K.C.B.; funding acquisition, K.C.B. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review Board for Human Subject Research at the University of Virginia (protocol code IRB-HSR 19967, approved most recently on 28 March 2023).
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study. Written informed consent has been obtained from the patient(s) to publish this paper if applicable.
Data Availability Statement
The data for this study will be posted on GitHub.
Conflicts of Interest
Bilchick has received research funding and performed consulting for Medtronic and has received research funding from Siemens for studies related to CRT in the past two years. Mangrum has received research grant support from Boston Scientific, CardioFocus, and Abbott. Mason has received consulting fees and honoraria from Medtronic and consulting fees from Boston Scientific. Abdi is employed by Siemens Healthineers. Mehta has been a speaker for Boston Scientific and has received research funding from Medtronic. Patel has research grants from GE Healthcare, Siemens Healthineers, Neosoft, and Circle CVI. Levy is a consultant to Medtronic and Impulse Dynamics, steering committee member for Cardiac Dimensions and Respircardia, and clinical endpoint committee member for GUIDE-HF (CardioMEMS, Abbott), Beckman Coulter, ANTHEM-HF (LivaNova), and SOLVE-CRT (EBR Systems, Inc). The University of Washington CoMotion holds the copyright to the Seattle Heart Failure Model.
Funding Statement
This research was funded by NHLBI grants R21 HL140445 (PI: Bilchick), R01 HL159945 (PI: Bilchick), and R01 HL147104 (PI: Epstein).
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Greenspon A.J., Patel J.D., Lau E., Frisch D., Ho R., Pavri B., Ochoa J., Kurtz S. Trends in permanent pacemaker implantation in the United States from 1993 to 2009: Increasing complexity of patients and procedures. J. Am. Coll. Cardiol. 2012;60:1540–1545. doi: 10.1016/j.jacc.2012.07.017. [DOI] [PubMed] [Google Scholar]
- 2.Bristow M.R., Feldman A.M., Saxon L.A. Heart failure management using implantable devices for ventricular resynchronization: Comparison of Medical Therapy, Pacing, and Defibrillation in Chronic Heart Failure (COMPANION) trial. J. Card. Fail. 2000;6:276–285. doi: 10.1054/jcaf.2000.9501. [DOI] [PubMed] [Google Scholar]
- 3.Butter C., Auricchio A., Stellbrink C., Fleck E., Ding J., Yu Y., Huvelle E., Spinelli J., on behalf of the Pacing Therapy for Chronic Heart Failure II (PATH-CHF-II) Study Group Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients. Circulation. 2001;104:3026–3029. doi: 10.1161/hc5001.102229. [DOI] [PubMed] [Google Scholar]
- 4.Auricchio A., Stellbrink C., Sack S., Block M., Vogt J., Bakker P., Huth C., Schöndube F., Wolfhard U., Böcker D., et al. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J. Am. Coll. Cardiol. 2002;39:2026–2033. doi: 10.1016/S0735-1097(02)01895-8. [DOI] [PubMed] [Google Scholar]
- 5.Abraham W.T., Fisher W.G., Smith A.L., Delurgio D.B., Leon A.R., Loh E., Kocovic D.Z., Packer M., Clavell A.L., Hayes D.L., et al. Cardiac resynchronization in chronic heart failure. N. Engl. J. Med. 2002;346:1845–1853. doi: 10.1056/NEJMoa013168. [DOI] [PubMed] [Google Scholar]
- 6.Higgins S.L., Hummel J.D., Niazi I.K., Giudici M.C., Worley S.J., A Saxon L., Boehmer J.P., Higginbotham M.B., De Marco T., Foster E., et al. Cardiac resynchronization therapy for the treatment of heart failure in patients with intraventricular conduction delay and malignant ventricular tachyarrhythmias. J. Am. Coll. Cardiol. 2003;42:1454–1459. doi: 10.1016/S0735-1097(03)01042-8. [DOI] [PubMed] [Google Scholar]
- 7.Young J.B., Abraham W.T., Smith A.L., Leon A.R., Lieberman R., Wilkoff B., Canby R.C., Schroeder J.S., Liem L.B., Hall S., et al. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: The MIRACLE ICD Trial. JAMA. 2003;289:2685–2694. doi: 10.1001/jama.289.20.2685. [DOI] [PubMed] [Google Scholar]
- 8.Bristow M.R., Saxon L.A., Boehmer J., Krueger S., Kass D.A., De Marco T., Carson P., DiCarlo L., DeMets D., White B.G., et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N. Engl. J. Med. 2004;350:2140–2150. doi: 10.1056/NEJMoa032423. [DOI] [PubMed] [Google Scholar]
- 9.Ellenbogen K.A. Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy. 5th ed. Elsevier; Amsterdam, The Netherlands: 2017. [Google Scholar]
- 10.Cleland J.G., Daubert J.C., Erdmann E., Freemantle N., Gras D., Kappenberger L., Tavazzi L. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N. Engl. J. Med. 2005;352:1539–1549. doi: 10.1056/NEJMoa050496. [DOI] [PubMed] [Google Scholar]
- 11.Sutton M.G., Plappert T., Hilpisch K.E., Abraham W.T., Hayes D.L., Chinchoy E. Sustained reverse left ventricular structural remodeling with cardiac resynchronization at one year is a function of etiology: Quantitative Doppler echocardiographic evidence from the Multicenter InSync Randomized Clinical Evaluation (MIRACLE) Circulation. 2006;113:266–272. doi: 10.1161/CIRCULATIONAHA.104.520817. [DOI] [PubMed] [Google Scholar]
- 12.Shimano M., Tsuji Y., Yoshida Y., Inden Y., Tsuboi N., Itoh T., Suzuki H., Muramatsu T., Okada T., Harata S., et al. Acute and chronic effects of cardiac resynchronization in patients developing heart failure with long-term pacemaker therapy for acquired complete atrioventricular block. Europace. 2007;9:869–874. doi: 10.1093/europace/eum119. [DOI] [PubMed] [Google Scholar]
- 13.Gilliam F.R., 3rd, Kaplan A.J., Black J., Chase K.J., Mullin C.M. Changes in heart rate variability, quality of life, and activity in cardiac resynchronization therapy patients: Results of the HF-HRV registry. Pacing Clin. Electrophysiol. 2007;30:56–64. doi: 10.1111/j.1540-8159.2007.00582.x. [DOI] [PubMed] [Google Scholar]
- 14.Lindenfeld J., Feldman A.M., Saxon L., Boehmer J., Carson P., Ghali J.K., Anand I., Singh S., Steinberg J.S., Jaski B., et al. Effects of cardiac resynchronization therapy with or without a defibrillator on survival and hospitalizations in patients with New York Heart Association class IV heart failure. Circulation. 2007;115:204–212. doi: 10.1161/CIRCULATIONAHA.106.629261. [DOI] [PubMed] [Google Scholar]
- 15.De Marco T., Wolfel E., Feldman A.M., Lowes B., Higginbotham M.B., Ghali J.K., Wagoner L., Kirlin P.C., Kennett J.D., Goel S., et al. Impact of cardiac resynchronization therapy on exercise performance, functional capacity, and quality of life in systolic heart failure with QRS prolongation: COMPANION trial sub-study. J. Card. Fail. 2008;14:9–18. doi: 10.1016/j.cardfail.2007.08.003. [DOI] [PubMed] [Google Scholar]
- 16.Linde C., Abraham W.T., Gold M.R., Sutton M.S.J., Ghio S., Daubert C., REVERSE (REsynchronization reVErses Remodeling in Systolic left vEntricular dysfunction) Study Group Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J. Am. Coll. Cardiol. 2008;52:1834–1843. doi: 10.1016/j.jacc.2008.08.027. [DOI] [PubMed] [Google Scholar]
- 17.Moss A.J., Hall W.J., Cannom D.S., Klein H., Brown M.W., Daubert J.P., Estes N.A.M., III, Foster E., Greenberg H., Higgins S.L., et al. Cardiac-resynchronization therapy for the prevention of heart-failure events. N. Engl. J. Med. 2009;361:1329–1338. doi: 10.1056/NEJMoa0906431. [DOI] [PubMed] [Google Scholar]
- 18.Tang A.S., Wells G.A., Talajic M., Arnold M.O., Sheldon R., Connolly S., Hohnloser S.H., Nichol G., Birnie D.H., Sapp J.L., et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N. Engl. J. Med. 2010;363:2385–2395. doi: 10.1056/NEJMoa1009540. [DOI] [PubMed] [Google Scholar]
- 19.Ellenbogen K.A., Gold M.R., Meyer T.E., Lozano I.F., Mittal S., Waggoner A.D., Lemke B., Singh J.P., Spinale F.G., Van Eyk J.E., et al. Primary results from the SmartDelay determined AV optimization: A comparison to other AV delay methods used in cardiac resynchronization therapy (SMART-AV) trial: A randomized trial comparing empirical, echocardiography-guided, and algorithmic atrioventricular delay programming in cardiac resynchronization therapy. Circulation. 2010;122:2660–2668. doi: 10.1161/circulationaha.111.027284. [DOI] [PubMed] [Google Scholar]
- 20.Theuns D.A., Smith T., Hunink M.G., Bardy G.H., Jordaens L. Effectiveness of prophylactic implantation of cardioverter-defibrillators without cardiac resynchronization therapy in patients with ischaemic or non-ischaemic heart disease: A systematic review and meta-analysis. Europace. 2010;12:1564–1570. doi: 10.1093/europace/euq329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Curtis A.B., Worley S.J., Chung E.S., Li P., Christman S.A., St. John Sutton M. Improvement in Clinical Outcomes with Biventricular Versus Right Ventricular Pacing: The BLOCK HF Study. J. Am. Coll. Cardiol. 2016;67:2148–2157. doi: 10.1016/j.jacc.2016.02.051. [DOI] [PubMed] [Google Scholar]
- 22.Schwertner W.R., Behon A., Merkel E.D., Tokodi M., Kovács A., Zima E., Osztheimer I., Molnár L., Király Á., Papp R., et al. Long-term survival following upgrade compared with de novo cardiac resynchronization therapy implantation: A single-centre, high-volume experience. Europace. 2021;23:1310–1318. doi: 10.1093/europace/euab059. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Vamos M., Erath J.W., Bari Z., Vagany D., Linzbach S.P., Burmistrava T., Israel C.W., Duray G.Z., Hohnloser S.H. Effects of Upgrade Versus De Novo Cardiac Resynchronization Therapy on Clinical Response and Long-Term Survival: Results from a Multicenter Study. Circ. Arrhythm. Electrophysiol. 2017;10:e004471. doi: 10.1161/CIRCEP.116.004471. [DOI] [PubMed] [Google Scholar]
- 24.Wokhlu A., Rea R.F., Asirvatham S.J., Webster T., Brooke K., Hodge D.O., Wiste H.J., Dong Y., Hayes D.L., Cha Y.-M. Upgrade and de novo cardiac resynchronization therapy: Impact of paced or intrinsic QRS morphology on outcomes and survival. Heart Rhythm. 2009;6:1439–1447. doi: 10.1016/j.hrthm.2009.07.009. [DOI] [PubMed] [Google Scholar]
- 25.Kanai M., Yagishita D., Shoda M., Ejima K., Hagiwara N. Clinical outcomes of upgrade to versus de novo cardiac resynchronization therapy in mild heart failure patients with atrioventricular block. J. Cardiol. 2022;79:6–14. doi: 10.1016/j.jjcc.2021.07.012. [DOI] [PubMed] [Google Scholar]
- 26.Nazarian S., Hansford R., Rahsepar A.A., Weltin V., McVeigh D., Ipek E.G., Kwan A., Berger R.D., Calkins H., Lardo A.C., et al. Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices. N. Engl. J. Med. 2017;377:2555–2564. doi: 10.1056/NEJMoa1604267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Russo R.J., Costa H.S., Silva P.D., Anderson J.L., Arshad A., Biederman R.W., Boyle N.G., Frabizzio J.V., Birgersdotter-Green U., Higgins S.L., et al. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N. Engl. J. Med. 2017;376:755–764. doi: 10.1056/NEJMoa1603265. [DOI] [PubMed] [Google Scholar]
- 28.McCrohon J.A., Moon J.C.C., Prasad S.K., McKenna W., Lorenz C., Coats A., Pennell D. Differentiation of heart failure related to dilated cardiomyopathy and coronary artery disease using gadolinium-enhanced cardiovascular magnetic resonance. Circulation. 2003;108:54–59. doi: 10.1161/01.CIR.0000078641.19365.4C. [DOI] [PubMed] [Google Scholar]
- 29.Senthilkumar A., Majmudar M.D., Shenoy C., Kim H.W., Kim R.J. Identifying the etiology: A systematic approach using delayed-enhancement cardiovascular magnetic resonance. Heart Fail. Clin. 2009;5:349–367. doi: 10.1016/j.hfc.2009.02.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Bilchick K.C., Kuruvilla S., Hamirani Y.S., Ramachandran R., Clarke S.A., Parker K.M., Stukenborg G.J., Mason P., Ferguson J.D., Moorman J.R., et al. Impact of mechanical activation, scar, and electrical timing on cardiac resynchronization therapy response and clinical outcomes. J. Am. Coll. Cardiol. 2014;63:1657–1666. doi: 10.1016/j.jacc.2014.02.533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Bivona D.J., Tallavajhala S., Abdi M., Oomen P.J., Gao X., Malhotra R., Darby A.E., Monfredi O.J., Mangrum J.M., Mason P.K., et al. Machine learning for multidimensional response and survival after cardiac resynchronization therapy using features from cardiac magnetic resonance. Heart Rhythm. O2. 2022;3:542–552. doi: 10.1016/j.hroo.2022.06.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Epstein A.E., DiMarco J.P., Ellenbogen K.A., Estes N.M., III, Freedman R.A., Gettes L.S., Gillinov A.M., Gregoratos G., Hammill S.C., Hayes D.L., et al. 2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: A report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J. Am. Coll. Cardiol. 2013;61:e6–e75. doi: 10.1016/j.jacc.2012.11.007. [DOI] [PubMed] [Google Scholar]
- 33.Levy W.C., Mozaffarian D., Linker D.T., Sutradhar S.C., Anker S.D., Cropp A.B., Anand I., Maggioni A., Burton P., Sullivan M.D., et al. The Seattle Heart Failure Model: Prediction of survival in heart failure. Circulation. 2006;113:1424–1433. doi: 10.1161/CIRCULATIONAHA.105.584102. [DOI] [PubMed] [Google Scholar]
- 34.Do D.H., Eyvazian V., Bayoneta A.J., Hu P., Finn J.P., Bradfield J.S., Shivkumar K., Boyle N.G. Cardiac magnetic resonance imaging using wideband sequences in patients with nonconditional cardiac implanted electronic devices. Heart Rhythm. 2018;15:218–225. doi: 10.1016/j.hrthm.2017.10.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Mangion K., Carrick D., Carberry J., Mahrous A., McComb C., Oldroyd K.G., Eteiba H., Lindsay M., McEntegart M., Hood S., et al. Circumferential strain predicts major adverse cardiovascular events following an acute ST-segment-elevation myocardial infarction. Radiology. 2019;290:329–337. doi: 10.1148/radiol.2018181253. [DOI] [PubMed] [Google Scholar]
- 36.Wang Y., Sun C., Ghadimi S., Auger D.C., Croisille P., Viallon M., Mangion K., Berry C., Haggerty C.M., Jing L., et al. StrainNet: Improved Myocardial Strain Analysis of Cine MRI by Deep Learning from DENSE. Radiol. Cardiothorac. Imaging. 2023;5:e220196. doi: 10.1148/ryct.220196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Ramachandran R., Chen X., Kramer C.M., Epstein F.H., Bilchick K.C. Singular value decomposition applied to cardiac strain from MR imaging for selection of optimal cardiac resynchronization therapy candidates. Radiology. 2015;275:413–420. doi: 10.1148/radiol.14141578. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Auger D.A., Bilchick K.C., Gonzalez J.A., Cui S.X., Holmes J.W., Kramer C.M., Salerno M., Epstein F.H. Imaging left-ventricular mechanical activation in heart failure patients using cine DENSE MRI: Validation and implications for cardiac resynchronization therapy. J. Magn. Reson. Imaging. 2017;46:887–896. doi: 10.1002/jmri.25613. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Parker K.M., Bunting E., Malhotra R., Clarke S.A., Mason P., Darby A.E., Kramer C.M., Salerno M., Holmes J.W., Bilchick K.C. Postprocedure mapping of cardiac resynchronization lead position using standard fluoroscopy systems: Implications for the nonresponder with scar. Pacing Clin. Electrophysiol. 2014;37:757–767. doi: 10.1111/pace.12344. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Albertsen A.E., Nielsen J.C., Pedersen A.K., Hansen P.S., Jensen H.K., Mortensen P.T. Left ventricular lead performance in cardiac resynchronization therapy: Impact of lead localization and complications. Pacing Clin. Electrophysiol. 2005;28:483–488. doi: 10.1111/j.1540-8159.2005.40066.x. [DOI] [PubMed] [Google Scholar]
- 41.Rabbat M.G., Kwong R.Y., Heitner J.F., Young A.A., Shanbhag S.M., Petersen S.E., Selvanayagam J.B., Berry C., Nagel E., Heydari B., et al. The Future of Cardiac Magnetic Resonance Clinical Trials. JACC Cardiovasc. Imaging. 2022;15:2127–2138. doi: 10.1016/j.jcmg.2021.07.029. [DOI] [PubMed] [Google Scholar]
- 42.Tibshirani R. Regression Shrinkage and Selection via the Lasso. J. R. Stat. Soc. 1996;58:267–288. doi: 10.1111/j.2517-6161.1996.tb02080.x. [DOI] [Google Scholar]
- 43.Merkely B., Hatala R., Wranicz J.K., Duray G., Földesi C., Som Z., Németh M., Goscinska-Bis K., Gellér L., Zima E., et al. Upgrade of right ventricular pacing to cardiac resynchronisation therapy in heart failure: A randomised trial. Eur. Heart J. 2023:ehad591. doi: 10.1093/eurheartj/ehad591. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data for this study will be posted on GitHub.