Abstract
Background: Physiological pacing has gained significant interest due to its potential to achieve optimal hemodynamic response. This study aimed to assess left ventricular performance in terms of electrical parameters, specifically QRS duration and mechanical performance, evaluated as myocardial work. We compared His Bundle Pacing (HBP) and Left Bundle Branch Area Pacing (LBBAP) to evaluate their effects. Methods: Twenty-four patients with class I or IIa indications for pacing were enrolled in this study, with twelve patients undergoing HBP implantation and another twelve undergoing LBBAP implantation. A comprehensive analysis of myocardial work was conducted. Results: Our findings indicate that there were no major differences in terms of spontaneous and HBP activation in myocardial work, except for global wasted work (217 mmHg% vs. 283 mmHg%; p 0.016) and global work efficiency (87 mmHg% vs. 82 mmHg%; p 0.049). No significant differences were observed in myocardial work between spontaneous activation and LBBAP. Similarly, no significant differences in myocardial work were found between HBP and LBBAP. Conclusions: Both pacing modalities provide physiological ventricular activation without significant differences when compared to each other. Moreover, there were no significant differences in QRS duration between HBP and LBBAP. However, LBBAP demonstrated advantages in terms of feasibility, as it achieved better lead electrical parameters compared to HBP (threshold@0.4 ms 0.6 V vs. 1 V; p = 0.045—sensing 9.4 mV vs. 2.4 mV; p < 0.001). Additionally, LBBAP required less fluoroscopy time (6 min vs. 13 min; p = 0.010) and procedural time (81 min vs. 125 min; p = 0.004) compared to HBP.
Keywords: physiological pacing, conduction system pacing, his bundle pacing, left bundle branch area pacing
1. Clinical Perspective
What Is Known: His Bundle Pacing (HBP) and Left Bundle Branch Area Pacing (LBBAP) have been recognized as more physiological alternatives to traditional right ventricular pacing. LBBAP has shown greater feasibility compared to HBP, although direct comparison data between the myocardial work in HBP and LBBAP are limited. What the Study Adds: Our study contributes to the existing knowledge by demonstrating that both HBP and LBBAP provide physiological ventricular activation, with no significant differences observed between the two pacing modalities in terms of myocardial work and QRS duration. However, LBBAP showcased advantages, such as reduced need for fluoroscopy, shorter procedural time, and improved electrical parameters. These findings further support the potential of LBBAP as a favorable pacing option.
2. Background
Historically, dual-chamber pacing emerged as a superior alternative to single-chamber pacing due to its ability to maintain physiological atrioventricular synchrony whenever possible. However, the detrimental effects of mid- and long-term right ventricular (RV) pacing [1,2,3,4] have underscored the necessity for an alternative pacing site to achieve more physiological ventricular activation. Cardiac resynchronization therapy (CRT) using biventricular pacing partially addressed this issue [5]. Nonetheless, biventricular pacing still results in non-physiological activation, as it originates from the epicardium and propagates through the myocardium instead of the conduction system. In recent years, conduction system pacing (CSP) has garnered substantial attention and adoption in clinical practice [6,7]. Initially, the main focus of CSP was His Bundle Pacing (HBP) [8,9]. However, Left Bundle Branch Area Pacing (LBBAP) has emerged as a promising alternative, surpassing many of the limitations associated with HBP in terms of feasibility, electrical parameters, and device settings [10,11,12]. However, no direct comparison data between myocardial work in HBP and LBBAP are available. In this study, we aimed to investigate the effects of HBP and LBBAP on left ventricular (LV) performance in a subset of patients with class I or IIa indications for pacemaker (PM) implantation.
3. Methods
This study is derived from the EMPATHY study [13], which is a prospective, single-center cohort study conducted at the Cardiology Unit of Azienda Ospedaliero-Universitaria di Ferrara, Italy. The study enrolled consecutive patients who underwent HBP or LBBAP. The study protocol was registered on ClinicalTrials.gov (NCT05222672) and received approval from the local ethics committee. Informed consent was obtained from all patients.
The inclusion criteria for this study were as follows: (1) class I or IIa indication for pacemaker implantation, based on the European guidelines [14]; (2) age ≥ 18 years; and (3) signed written informed consent. The exclusion criteria included (1) an inability to provide informed consent, (2) pregnancy, (3) severe mitral or aortic valve disease, and (4) left ventricular ejection fraction (LVEF) < 35%.
The primary endpoint of this study was to compare LV myocardial work during spontaneous ventricular activation (SVA) with HBP or LBBAP. The secondary endpoint was to compare the change in LV myocardial work between stimulated activation with HBP and LBBAP and SVA.
3.1. Implantation
3.1.1. His Bundle Pacing
The index procedure involved the implantation of a pacemaker and simultaneous three-dimensional electroanatomical Mapping (3D-EAM). The right ventricle and the His bundle area were mapped non-fluoroscopically using a high-density mapping catheter, which was inserted via the femoral vein. The pacing leads were inserted via the left cephalic or axillary vein and positioned using a combination of 3D-EAM and fluoroscopy [13].
For His bundle pacing, an active fixation lead (SelectSecure 3830, Medtronic, Minneapolis, MN, USA, or Solia S, Biotronik, Berlin, Germany) was placed using a non-deflectable sheath (C315, Medtronic, Minneapolis, MN, USA, or Selectra 3D, Biotronik, Berlin, Germany). The position of the His bundle lead was confirmed using unipolar and bipolar intracardiac electrograms, with standard criteria employed to determine selective and non-selective His capture [15]. The intraprocedural unipolar capture threshold at 0.4 ms during the implant procedure was used to define the pacing threshold. In all patients, a backup right ventricular lead was implanted, and an atrial lead was added, if necessary, based on the pacing indication (Figure 1). Antibiotic prophylaxis and antithrombotic drugs were administered following the protocols of the center and international guidelines [16,17].
3.1.2. Left Bundle Branch Area Pacing
Before the procedure, the thickness of the basal interventricular septum and the presence of a septal scar were assessed. The pacing leads were inserted via the left cephalic or axillary vein. The active fixation 3830 lumenless lead (Medtronic, Minneapolis, MN, USA) and the non-deflectable C315His sheath, as well as the Biotronik Solia 60 stylet-driven lead (Biotronik, Berlin, Germany) and Selectra 3D, were used as the pacing leads and delivery guiding catheters.
The procedure began by advancing the guiding catheter into the right ventricle (RV) over a J wire. In the left anterior oblique (LAO) view (30–40°), contrast injection via the guiding catheter was conducted to delineate the septum. After counterclockwise rotation of the guiding catheter to achieve a perpendicular orientation to the septum, a pace map was conducted by exposing the tip of the pacing lead [18]. The aim was to obtain a QRS morphology in lead II that was more positive than in lead III, aVr and aVl discordant, and W’ pattern with a notch at the nadir of QRS complex in V1.
The screwing technique varied depending on the type of lead used. For lumenless leads, 3–5 rapid rotations were applied based on the thickness of the ventricular septum while keeping the guiding catheter in position. For stylet-driven leads, the first screw was exposed, and after a pacing check, 2–3 rapid rotations similar to lumenless leads were performed. As the screw advanced into the interventricular septum, the paced QRS morphology exhibited a right bundle branch block (RBBB) pattern, with the notch at the nadir of the QRS in lead V1 shifting to the end of the QRS, indicating successful LBBAP (Figure 2). To summarize, LBBAP capture was confirmed by observing the transition from a “w” pattern (observed with the catheter lying on the right side of the interventricular septum) to a right bundle branch block pattern with the onset of an R’ (when the catheter was screwed in the septum) in lead V1, the development of a terminal S wave in V6, and the presence of an LBB potential on the unipolar EGM. A fast LV activation time in V6 < 75 ms was also aimed for [19]. The myocardial current of injury (COI) was recorded in the electrogram (EGM). In cases where perforation into the LV occurred, the lead was repositioned at a different location. An atrial lead was implanted when needed based on the pacing indication. Antibiotic prophylaxis and antithrombotic drugs were administered following the center’s protocols [16,17].
During the procedure, the pacing threshold was measured in the unipolar configuration at 0.4 ms, and the impedance was also determined in the unipolar configuration. The final R-wave amplitude was measured in the bipolar configuration. Successful LBBAP was considered when a deep septal lead position was achieved, and the paced QRS complex included a terminal R/r wave in lead V1, indicating a delay in right ventricle activation. In rare cases, a QS configuration (lack of terminal R) in V1 was accepted if a terminal R/r wave in lead V1 appeared during programmed stimulation or other features indicating LBBAP, as described below, were present.
3.2. Imaging and Electrocardiogram
After the procedure, all patients underwent echocardiographic evaluation to assess various parameters. In patients with intrinsic rhythm, we performed the echocardiogram during spontaneous ventricular activation, setting the pacemaker to VVI mode at 30 bpm. For those without intrinsic rhythm, we analyzed previous echocardiographic exams conducted during intrinsic rhythm during the previous six months. In all cases, the echocardiogram was also performed during stimulated rhythm, with the pacemaker set to VVI or DDD mode, depending on whether it was a single or dual-chamber device and the presence of sinus rhythm or atrial fibrillation. The pacing rate was adjusted to achieve overdrive pacing. Basic information, such as LV volume (in mL), LV ejection fraction (LVEF) (in %), atrial volume (in mL), and the degree of valve diseases were collected. Global longitudinal strain (GLS) and myocardial work (MW) were analyzed as part of the evaluation. MW is a novel echocardiographic technique that utilizes speckle tracking analysis to estimate left ventricular performance by measuring the area under the pressure–strain loop curve derived from GLS and blood pressure [20,21,22]. All echocardiographic examinations were performed using the GE Vivid E9 with M5S transducers, and GLS and MW analyses were conducted offline using EchoPAC software V.202 (GE Healthcare, Chicago, IL, USA). The echocardiographic parameters were assessed according to international standards [23]. The collected MW parameters included:
Global constructive work (GCW): the sum of work performed during myocardial shortening in systole and myocardial lengthening during isovolumetric diastole;
Global wasted work (GWW): the sum of work performed by myocardial lengthening in systole and myocardial shortening during isovolumetric diastole;
Global work index (GWI): the work performed throughout systole, specifically between mitral valve closure and opening;
Global work efficiency (GWE): expressed as the percentage ratio of GCW to the sum of GCW and GWW.
A twelve-lead ECG was performed and analyzed as well. Echocardiography and ECG evaluations were conducted in different conditions, including SVA and during HBP or LBBAP, respectively. The measurements for HBP and LBBAP were obtained during pacing at a fixed rate in DDD mode with an optimized AV delay or in VVI mode, depending on whether the patients were in sinus rhythm or atrial fibrillation
3.3. Statistical Analysis
Continuous variables were expressed as mean ± standard deviation when normally distributed, as estimated using the Shapiro–Wilk test, or as median and interquartile range. Not all variables were normally distributed. Categorical variables were expressed as numbers and percentages.
Differences between repeated measurements during SVA and HBP or LBBAP were assessed using the Wilcoxon rank sum test. Differences between independent measurements in HBP and LBBAP were assessed using the Mann–Whitney U test. In order to adjust for baseline parameters during SVA, the comparison of myocardial work during HBP and LBBAP was not made using relative values calculated as the difference between stimulated and spontaneous parameter divided by the spontaneous one [example: dGCWLBBAP = (GCWLBBAP − GCWSVA)/GCWSVA] and multiplied for 100 to obtain percentage change.
p values < 0.05 were considered statistically significant. The statistical analysis was performed using STATA, version 15.0 (StataCorp LLC, College Station, TX, USA).
4. Results
A total of twenty-four patients were enrolled in this study, with four of them (two in each group) being female. The patients were matched for comorbidities and implant characteristics. The baseline clinical characteristics, procedural data, and electrical parameters are summarized in Table 1. The median age was 79 years (IQR 73–85) in the HBP group and 81 years (IQR 73–85) in the LBBAP group, with no significant difference between the groups. There were no significant differences between the groups in terms of the main cardiovascular risk factors, such as arterial hypertension, dyslipidemia, and diabetes (Table 1).
Table 1.
HBP (12 Patients) |
LBBAP (12 Patients) |
p-Value | |
---|---|---|---|
Baseline characteristics | |||
Age (years) | 79 (73–85) | 81 (73–85) | 0.88 |
Female sex | 2 (17) | 2 (17) | 1.00 |
BMI (kg/m2) | 25.1 (23.4–28.3) | 24.7 (22.3–29.1) | 0.69 |
CAD | 4 (33) | 2 (17) | 0.64 |
Atrial fibrillation | 7 (58) | 5 (42) | 0.68 |
Diabetes mellitus | 0 (0) | 4 (33) | 0.093 |
Hypertension | 10 (83) | 9 (75) | 1.00 |
Dyslipidemia | 6 (50) | 8 (67) | 0.68 |
Procedural data | |||
Procedure duration (min) | 125 (120–140) | 81 (70–120) | 0.004 |
Fluoroscopy time (min) | 13 (9–21) | 6 (5–11) | 0.010 |
Selective HPB capture | 6 (50) | - | - |
Electrical parameters | |||
Lead impedance (ohm) | 509 (447–568) | 718 (609–795) | 0.002 |
Lead sensing (mV) | 2.4 (1.8–4.8) | 9.4 (8–13.5) | <0.001 |
Lead threshold (V@0.4 ms) | 1.0 (0.5–2.5) | 0.6 (0.4–0.8) | 0.045 |
EF (%) | 51 (42–63) | 55 (51–56) | 0.750 |
GLS (%) | −14 (−19–−11) | −15 (−18–−10) | 0.410 |
Left atrial volume (mL/m2) | 41 ± 10 | 39 ± 13 | 0.721 |
Furthermore, no significant differences were observed in the major cardiovascular comorbidities that could affect the implantation procedure. Specifically, there were no significant differences in the history of atrial fibrillation (AF) (58% in the HBP group vs. 42% in the LBBAP group; p = 0.68) or coronary artery disease (CAD) (33% in the HBP group vs. 17% in the LBBAP group; p = 0.64). Baseline LVEF was also comparable between the two groups (51% in the HBP group vs. 55% in the LBBAP group; p = 0.75).
Both global longitudinal strain (GLS) and left atrial volume were similar between groups (for details, see Table 1).
In the HBP group, pacing indications were as follows: seven patients (58%) for II- or III-degree atrioventricular block, two patients (17%) for brady–tachy syndrome, two patients (17%) for symptomatic brady–atrial fibrillation, and one patient (8%) for sick sinus syndrome. In the LBBAP group, the indications were: eight patients (67%) for II- or III-degree atrioventricular block, one patient (8%) for brady–tachy syndrome, two patients (17%) for symptomatic brady–atrial fibrillation and one patient (8%) for sick sinus syndrome.
Selective HBP was obtained in six patients (50%). LBBAP was feasible in all the patients. The procedural duration was significantly longer in the HBP group compared to the LBBAP group (125 min; IQR 120–140 vs. 81 min; IQR 70–120; p 0.004). Similarly, intraprocedural fluoroscopy time was higher in the HBP group compared to the LBBAP group (13 min; IQR 9–21 vs. 6 min; IQR 5–11; p 0.01). Electrical lead parameters were assessed after implantation. Lead impedance was significantly lower in the HBP group compared to the LBBAP group (509 ohm; IQR 447–568 vs. 718 ohm; IQR 608–795; p 0.002). Lead bipolar sensing was significantly lower in the HBP group compared to the LBBAP group (2.4 mV; IQR 1.8–4.8 vs. 9.5 mV; IQR 8–13.5; p < 0.001).
The lead unipolar threshold was significantly higher in the HBP group compared to the LBBAP group (1.0 V; IQR 0.5–2.5 vs. 0.6 V; IQR 0.4–0.8; p 0.045).
ECG and SVA Myocardial Work
The baseline data for electrocardiographic and myocardial work are presented in Table 2. The spontaneous QRS duration did not show a statistically significant difference between the HBP and LBBAP groups (106 ms; IQR 88–140 vs. 115 ms; IQR 90–132, p = 0.76). Similarly, the paced QRS duration did not exhibit a statistically significant difference between the HBP and LBBAP groups (124 ms; IQR 98–140 vs. 128 ms; IQR 118–136, p = 0.62).
Table 2.
HBP (12 Patients) |
LBBAP (12 Patients) |
p-Value | |
---|---|---|---|
Spontaneous QRS duration (ms) | 106 (88–140) | 115 (90–132) | 0.76 |
Paced QRS duration (ms) | 124 (98–140) | 128 (118–136) | 0.62 |
Spontaneous GWI (mmHg%) | 1120 (902–1763) | 1548 (1072–2273) | 0.094 |
Spontaneous GCW (mmHg%) | 1648 (1044–2152) | 2089 (1705–2594) | 0.052 |
Spontaneous GWW (mmHg%) | 217 (125–249) | 137 (105–286) | 0.670 |
Spontaneous GWE (mmHg%) | 87 (80–90) | 93 (83–95) | 0.065 |
No significant differences were observed in any of the myocardial work indexes between the HBP and LBBAP groups.
In the first part of the analysis, a comparison was made between SVA and paced ventricular activation in the HBP group (Table 3). No significant differences were found in the GWI (1110 mmHg% IQR 902–1763 in SVA vs. 1020 mmHg% IQR 822–1969 in HBP; p 0.534) and GCW (1648 mmHg% IQR 1044–2152 in SVA vs. 1505 mmHg% IQR 1151–2133 in HBP; p 0.075) between the two types of activation.
Table 3.
HBP (12 Patients) | |||
---|---|---|---|
Spontaneous | Paced | p-value | |
QRS duration (ms) | 106 (88–140) | 124 (98–140) | 0.75 |
GWI (mmHg%) | 1110 (902–1763) | 1020 (822–1969) | 0.534 |
GCW (mmHg%) | 1648 (1044–2152) | 1505 (1151–2133) | 0.075 |
GWW (mmHg%) | 217 (125–249) | 283 (205–354) | 0.016 |
GWE (mmHg%) | 87 (80–90) | 82 (73–90) | 0.049 |
LBBAP (12 patients) | |||
Spontaneous | Paced | p-value | |
QRS duration (ms) | 115 (90–132) | 128 (118–136) | 0.05 |
GWI (mmHg%) | 1548 (1072–2273) | 1545 (1363–1961) | 0.374 |
GCW (mmHg%) | 2089 (1705–2594) | 2320 (2073–2530) | 0.929 |
GWW (mmHg%) | 137 (105–286) | 264 (195–341) | 0.091 |
GWE (mmHg%) | 93 (83–95) | 87 (83–92) | 0.109 |
However, GWW was significantly higher in HBP ventricular activation compared to SVA (217 mmHg% IQR 125–249 in SVA vs. 283 mmHg% IQR 205–354 in HBP; p-value = 0.016), and GWE was significantly lower in HBP ventricular activation compared to SVA (87% IQR 80–90 in SVA vs. 82% IQR 73–90 in paced rhythm; p 0.049).
In the second part of the analysis, a comparison was made between SVA and paced ventricular activation in the LBBAP group (Table 3). No differences were found in myocardial work parameters between SVA and paced rhythm. In particular, GWI was 1548 (1072–2273) mmHg% in SVA vs. 1545 (1363–1961) mmHg% in paced rhythm with p-value = 0.374; GCW was 2089 (1705–2594) mmHg% in SVA vs. 2320 (2073–2530) mmHg% in paced rhythm with p-value = 0.929; GWW was 137 (105–286) mmHg% in SVA vs. 264 (195–341) mmHg% in paced rhythm with p-value = 0.091; GWE was 93 (83–95)% in SVA vs. 87 (83–92)% in paced rhythm with p-value = 0.109 (Figure 2).
Relative differences between SVA and paced ventricular activation of all myocardial work indexes and paced QRS duration did not show significant differences between the LBBAP and HBP groups. The relative differences in GWI, GCW, GWW, GWE, and paced QRS duration were comparable between the two groups (Table 4).
Table 4.
HBP (12 Patients) |
LBBAP (12 Patients) |
p-Value | |
---|---|---|---|
dQRS (%) | 0 (−7.5–+42.8) | +16.4 (−4.9–+32.0) | 0.54 |
dGWI (%) | +0.1 (−18.4–+36.4) | −6.3 (−22.6–+24.4) | 0.67 |
dGCW (%) | 12.3 (−2.8–+16.2) | −0.9 (−10.2–+6.2) | 0.25 |
dGWW (%) | +66.4 (+8.4–+91.7) | +113.3 (−5.3–+218.2) | 0.53 |
dGWE (%) | −2.4 (−16.7–+1.1) | −5.3 (−9.7–+1.1) | 0.77 |
dQRS: relative difference in QRS [(QRSpaced − QRSspontaneus) × 100/QRSspontaneus]; dGWI: relative difference in global work index [(GWIpaced − GWIspontaneus) × 100/GWIspontaneus]; GCW: global constructive work [(GCWpaced − GCWspontaneus) × 100/GCWspontaneus]; GWW: global wasted work [(GWWpaced − GWWspontaneus) × 100/GWWspontaneus]; GWE global work efficiency [(GWEpaced − GWEspontaneus) × 100/GWEspontaneus].
5. Discussion
Conduction system pacing, which includes HBP and LBBAP, is a physiological pacing modality that stimulates the myocardium through specialized conduction fascicles [6]. It aims to avoid the electrical and mechanical dyssynchrony caused by the traditional RV Apical Pacing and its associated detrimental effects.
In our study, we assessed the impact of HBP and LBBAP on left ventricular performance, focusing on electrical and mechanical synchronization, in a subgroup of patients with a class I or IIa indication for pacemaker implantation. Our main findings are as follows:
-
(1)
There were no statistically significant differences in terms of myocardial work index and global constructive work between spontaneous and paced activation in both LBBAP and HBP groups;
-
(2)
The relative difference in all myocardial work parameters between sinus and paced activation did not show statistically significant differences between the two groups.
Spontaneous and paced QRS durations were not statistically different in either group;
-
(3)
The relative difference between spontaneous and paced QRS durations did not exhibit statistically significant differences between the HBP and LBBAP groups.
Myocardial work (MW) is an emerging tool in studying myocardial mechanics [20]. Unlike traditional parameters, such as ejection fraction, MW incorporates both deformation and load, providing additional information on cardiac performance. It is also more effective in quantitatively assessing mechanical synchrony and efficiency compared to speckle-tracking imaging [21,22]. Furthermore, MW offers a better evaluation of synchronization than electrocardiographic features, such as QRS duration [24]. In our study, we found no significant differences in the global work index and global constructive work between spontaneous and paced ventricular activation in both groups. HBP exhibits a significantly higher global wasted work and lower global work efficiency compared to SVA, likely attributed to the considerable percentage of non-selective HBP. In our study group, selective HBP was achieved in half of the patients, aligning with the previous literature. [25]. Furthermore, the comparison of relative differences between spontaneous and paced ventricular activation in all myocardial work parameters did not reveal any statistically significant differences between the two groups. Additionally, in the LBBAP group, global wasted work and global work efficiency were comparable between spontaneous and paced rhythm. The EMPATHY study [13] demonstrated that HBP had similar myocardial performance to SVA and superior ventricular efficiency compared to RV Apical Pacing. A recent study by Wang et al. [26] showed that LBBAP resulted in more effective myocardial work than RVP.
The duration of surface QRS has commonly been used as an approximate measure of electrical synchronization [27]. In our study, we found no significant differences in QRS duration between spontaneous and paced ventricular activation in both the HBP and LBBAP groups and between paced activation in the two groups. These results align with previous studies [8,9,10,13] that have reported similar findings regarding QRS duration in relation to different pacing modalities.
HBP is a technically challenging procedure that requires time and the use of a backup right ventricular pacing lead [18]. The success rate of HBP implantation varies in the literature [25,28,29], but it generally improves with increasing experience in both HBP and LBBAP [10]. While early studies reported longer fluoroscopy and procedural times for HBP, recent experiences have shown reduced times, although still longer than traditional RVP procedures [25]. In our study population, procedural and fluoroscopy times were longer in the HBP group than in the LBBAP group (125 vs. 80, p = 0.004; 13 vs. 6, p = 0.01, respectively), despite using 3D-EAM during HBP.
Another important aspect to consider is the type of capture achieved by HBP. It can be either selective, capturing only the His bundle tissue, or non-selective, involving fusion capture of the His bundle and adjacent myocardium [8]. The type of capture depends on the pacing location, surrounding atrial or ventricular tissue, and pacing output amplitude [25]. Achieving selective HBP was not always straightforward in our study population, as it was achieved in only six patients (50%), which is in line with the previous literature [25].
In contrast, LBBAP is defined as capturing the subendocardial area on the left side of the interventricular septum, with or without simultaneous conduction system capture. It encompasses techniques such as Left Bundle Branch Pacing (LBBP), Left Fascicular Pacing (LFP), and Left Ventricular Septal Pacing (LVSP) [18]. Consequently, LBBAP is more predictable and easier to achieve compared to HBP, with a higher implant success rate [10,12,19]. Another advantage of LBBAP is that the lead is positioned deep in the ventricular septal myocardium, providing backup septal pacing in case of left bundle branch capture loss due to more distal conduction system disease.
The electrical performance of the implanted ventricular leads differed significantly between the two groups, with the HBP group exhibiting higher impedance, lower sensing, and higher unipolar threshold compared to the LBBAP group. These findings emphasize the clinical advantages of LBBAP over HBP, as LBBAP consistently exhibits superior electrical parameters.
Limitations
One of the limitations of our study is the relatively small sample size, with only 24 patients. However, given the scarcity of data comparing HBP and LBBAP in terms of myocardial performance, our findings provide valuable insights. Another limitation is that our study was conducted at a single center, which may introduce a potential limitation in terms of generalizability. However, this limitation also eliminates the potential bias associated with variations in expertise between different centers. Additionally, the assessment of mechanical synchrony was dependent on the quality of echocardiography images. Nevertheless, the use of speckle-tracking technology has been shown to reduce this limitation, as it has a feasibility rate of over 90% [30].
6. Conclusions
Both His Bundle Pacing and Left Bundle Branch Area Pacing offer physiological ventricular activation with optimal electrical and mechanical synchrony. However, LBBAP emerges as a more favorable option due to its greater feasibility, reduced need for fluoroscopy, and shorter procedural time, effectively addressing the limitations associated with HBP.
Author Contributions
Conceptualization, M.M., M.B. and F.V.; methodology, M.M., M.B. and F.V.; software, G.A. and N.B.; validation, G.A., N.B. and F.V.; formal analysis, G.A. and N.B.; investigation, G.A., M.B., N.B. and F.V.; resources, M.M., M.B., C.B., M.D.R. and F.V.; data curation, G.A. and N.B.; writing—original draft preparation, G.A. and N.B.; writing—review and editing, M.M., M.B. and F.V.; visualization, M.M., M.B., C.B., M.D.R. and F.V.; supervision, M.B.; project administration, M.M. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
This study was approved by our local Ethics Committee named “Comitato Etico di Area Vasta Emilia Centro” in 20 May 2021 and by our Internal Review Board. Approval Code: 457/2021/Oss/AOUFe.
Informed Consent Statement
Informed consent was obtained from all subjects.
Data Availability Statement
Data is available on request.
Conflicts of Interest
The authors declare no conflict of interest.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Sharma A.D., Rizo-Patron C., Hallstrom A.P., O’Neill G.P., Rothbart S., Martins J.B., Roelke M., Steinberg J.S., Greene H.L., DAVID Investigators Percent right ventricular pacing predicts outcomes in the DAVID trial. Heart Rhythm. 2005;2:830–834. doi: 10.1016/j.hrthm.2005.05.015. [DOI] [PubMed] [Google Scholar]
- 2.Sweeney M.O., Hellkamp A.S., Ellenbogen K.A., Greenspon A.J., Freedman R.A., Lee K.L., Lamas G.A. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003;107:2932–2937. doi: 10.1161/01.CIR.0000072769.17295.B1. [DOI] [PubMed] [Google Scholar]
- 3.Wilkoff B.L., Cook J.R., Epstein A.E., Greene H.L., Hallstrom A.P., Hsia H., Kutalek S.P., Sharma A., Chamber D., VVI Implantable Defibrillator Trial Investigators Dual chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: The dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA. 2002;288:3115–3123. doi: 10.1001/jama.288.24.3115. [DOI] [PubMed] [Google Scholar]
- 4.Nahlawi M., Waligora M., Spies S.M., Bonow R.O., Kadish A.H., Goldberger J.J. Left ventricular function during and after right ventricular pacing. J. Am. Coll. Cardiol. 2004;44:1883–1888. doi: 10.1016/j.jacc.2004.06.074. [DOI] [PubMed] [Google Scholar]
- 5.Herweg B., Welter-Frost A., Vijayaraman P. The evolution of cardiac resynchronization therapy and an introduction to conduction system pacing: A conceptual review. Europace. 2021;23:496–510. doi: 10.1093/europace/euaa264. [DOI] [PubMed] [Google Scholar]
- 6.Vijayaraman P., Bordachar P., Ellenbogen K.A. The Continued Search for Physiological Pacing: Where Are We Now? J. Am. Coll. Cardiol. 2017;69:3099–3114. doi: 10.1016/j.jacc.2017.05.005. [DOI] [PubMed] [Google Scholar]
- 7.Togashi I., Sato T. Conduction system pacing: Current status and prospects. J. Cardiol. 2023;81:413–419. doi: 10.1016/j.jjcc.2023.01.011. [DOI] [PubMed] [Google Scholar]
- 8.Vijayaraman P., Chung M.K., Dandamudi G., Upadhyay G.A., Krishnan K., Crossley G., Campbell K.B., Lee B.K., Refaat M.M., Saksena S., et al. His Bundle Pacing. J. Am. Coll. Cardiol. 2018;72:927–947. doi: 10.1016/j.jacc.2018.06.017. [DOI] [PubMed] [Google Scholar]
- 9.Catanzariti D., Maines M., Manica A., Angheben C., Varbaro A., Vergara G. Permanent His-bundle pacing maintains long-term ventricular synchrony and left ventricular performance, unlike conventional right ventricular apical pacing. Europace. 2012;15:546–553. doi: 10.1093/europace/eus313. [DOI] [PubMed] [Google Scholar]
- 10.Li Y., Chen K., Dai Y., Li C., Sun Q., Chen R., Gold M.R., Zhang S. Left bundle branch pacing for symptomatic bradycardia: Implant success rate, safety, and pacing characteristics. Heart Rhythm. 2019;16:1758–1765. doi: 10.1016/j.hrthm.2019.05.014. [DOI] [PubMed] [Google Scholar]
- 11.Su L., Wang S., Wu S., Xu L., Huang Z., Chen X., Zheng R., Jiang L., Ellenbogen K.A., Whinnett Z.I., et al. Long-Term Safety and Feasibility of Left Bundle Branch Pacing in a Large Single-Center Study. Circ. Arrhythmia Electrophysiol. 2021;14:e009261. doi: 10.1161/CIRCEP.120.009261. [DOI] [PubMed] [Google Scholar]
- 12.Ponnusamy S.S., Arora V., Namboodiri N., Kumar V., Kapoor A., Vijayaraman P. Left bundle branch pacing: A comprehensive review. J Cardiovasc. Electrophysiol. 2020;31:2462–2473. doi: 10.1111/jce.14681. [DOI] [PubMed] [Google Scholar]
- 13.Malagù M., Vitali F., Massafra R.F., Cardelli L.S., Pavasini R., Guardigli G., Rapezzi C., Bertini M. Three-Dimensional Electroanatomical Mapping and Myocardial Work Performance during Spontaneous Rhythm, His Bundle Pacing and Right Ventricular Pacing: The EMPATHY Study. J. Cardiovasc. Dev. Dis. 2022;9:377. doi: 10.3390/jcdd9110377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Glikson M., Nielsen J.C., Kronborg M.B., Michowitz Y., Auricchio A., Barbash I.M., Barrabés J.A., Boriani G., Braunschweig F., Brignole M., et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur. Heart J. 2021;42:3427–3520. doi: 10.1093/eurheartj/ehab364. [DOI] [PubMed] [Google Scholar]
- 15.Vijayaraman P., Dandamudi G., Zanon F., Sharma P.S., Tung R., Huang W., Koneru J., Tada H., Ellenbogen K.A., Lustgarten D.L. Permanent His bundle pacing: Recommendations from a Multicenter His Bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm. 2018;15:460–468. doi: 10.1016/j.hrthm.2017.10.039. [DOI] [PubMed] [Google Scholar]
- 16.Steffel J., Collins R., Antz M., Cornu P., Desteghe L., Haeusler K.G., Oldgren J., Reinecke H., Roldan-Schilling V., Rowell N., et al. 2021 European Heart Rhythm Association Practical Guide on the Use of Non-Vitamin K Antagonist Oral Anticoagulants in Patients with Atrial Fibrillation. Europace. 2021;23:1612–1676. doi: 10.1093/europace/euab065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Malagù M., Vitali F., Brieda A., Cimaglia P., De Raffele M., Tazzari E., Musolino C., Balla C., Serenelli M., Cultrera R., et al. Antibiotic prophylaxis based on individual infective risk stratification in cardiac implantable electronic device: The PRACTICE study. Europace. 2022;24:413–420. doi: 10.1093/europace/euab222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Burri H., Jastrzebski M., Cano Ó., Čurila K., de Pooter J., Huang W., Israel C., Joza J., Romero J., Vernooy K., et al. EHRA clinical consensus statement on conduction system pacing implantation: Endorsed by the Asia Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS), and Latin American Heart Rhythm Society (LAHRS) Europace. 2023;25:1208–1236. doi: 10.1093/europace/euad043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Huang W., Chen X., Su L., Wu S., Xia X., Vijayaraman P. A beginner’s guide to permanent left bundle branch pacing. Heart Rhythm. 2019;16:1791–1796. doi: 10.1016/j.hrthm.2019.06.016. [DOI] [PubMed] [Google Scholar]
- 20.Papadopoulos K., Tok Ö.Ö., Mitrousi K., Ikonomidis I. Myocardial Work: Methodology and Clinical Applications. Diagnostics. 2021;11:573. doi: 10.3390/diagnostics11030573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Boe E., Skulstad H., Smiseth O.A. Myocardial work by echocardiography: A novel method ready for clinical testing. Eur. Heart J. Cardiovasc. Imaging. 2019;20:18–20. doi: 10.1093/ehjci/jey156. [DOI] [PubMed] [Google Scholar]
- 22.Russell K., Eriksen M., Aaberge L., Wilhelmsen N., Skulstad H., Remme E.W., Haugaa K.H., Opdahl A., Fjeld J.G., Gjesdal O., et al. A novel clinical method for quantification of regional left ventricular pressure–strain loop area: A non-invasive index of myocardial work. Eur. Heart J. 2012;33:724–733. doi: 10.1093/eurheartj/ehs016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Lang R.M., Badano L.P., Mor-Avi V., Afilalo J., Armstrong A., Ernande L., Flachskampf F.A., Foster E., Goldstein S.A., Kuznetsova T., et al. Recommendations for cardiac chamber quantification by echocardiography in adults: An update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. Eur. Heart J. Cardiovasc. Imaging. 2015;16:233–271. doi: 10.1093/ehjci/jev014. [DOI] [PubMed] [Google Scholar]
- 24.Ghio S., Constantin C., Klersy C., Serio A., Fontana A., Campana C., Tavazzi L. Interventricular and intraventricular dys- synchrony are common in heart failure patients, regardless of QRS duration. Eur. Heart J. 2004;25:571–578. doi: 10.1016/j.ehj.2003.09.030. [DOI] [PubMed] [Google Scholar]
- 25.Abdelrahman M., Subzposh F.A., Beer D., Durr B., Naperkowski A., Sun H., Oren J.W., Dandamudi G., Vijayaraman P. Clinical Outcomes of His Bundle Pacing Compared to Right Ventricular Pacing. J. Am. Coll. Cardiol. 2018;71:2319–2330. doi: 10.1016/j.jacc.2018.02.048. [DOI] [PubMed] [Google Scholar]
- 26.Wang P., Yang L., Zheng S., Mai J., Wei Y., Liu Y., Deng B., Lv H., Chen Y., Qiu Q. Left bundle branch pacing on mechanical synchrony and myocardial work in bradycardia patients. Int. J. Cardiovasc. Imaging. 2023;39:369–378. doi: 10.1007/s10554-022-02742-5. [DOI] [PubMed] [Google Scholar]
- 27.Chen K., Li Y., Dai Y., Sun Q., Luo B., Li C., Zhang S. Comparison of electrocardiogram characteristics and pacing parameters between left bundle branch pacing and right ventricular pacing in patients receiving pacemaker therapy. Europace. 2019;21:673–680. doi: 10.1093/europace/euy252. [DOI] [PubMed] [Google Scholar]
- 28.Sharma P.S., Dandamudi G., Naperkowski A., Oren J.W., Storm R.H., Ellenbogen K.A., Vijayaraman P. Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice. Heart Rhythm. 2015;12:305–312. doi: 10.1016/j.hrthm.2014.10.021. [DOI] [PubMed] [Google Scholar]
- 29.Keene D., Arnold A.D., Jastrzębski M., Burri H., Zweibel S., Crespo E., Chandrasekaran B., Bassi S., Joghetaei N., Swift M., et al. His bundle pacing, learning curve, procedure characteristics, safety, and feasibility: Insights from a large international observational study. J. Cardiovasc. Electrophysiol. 2019;30:1984–1993. doi: 10.1111/jce.14064. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Bertini M., Ng A.C., Antoni M.L., Nucifora G., Ewe S.H., Auger D., Marsan N.A., Schalij M.J., Bax J.J., Delgado V. Global longitudinal strain predicts long-term survival in patients with chronic ischemic cardiomyopathy. Circ. Cardiovasc. Imaging. 2012;5:383–391. doi: 10.1161/CIRCIMAGING.111.970434. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data is available on request.