Abstract
Background
The commercially available design of a three-dimensional (3D)–printed titanium (3D-Ti) cage can be divided into two types according to the presence of a window: a cage with a window that allows filling of bone graft materials and a non-window cage for stand-alone use. This prospective observational case series study aimed to explore the clinical feasibility of using a non-window type 3D-Ti cage in cases of combined window and non-window cage implantation. Furthermore, we evaluated the bone in growth patterns of non-window cages and their correlation with published fusion grading systems.
Methods
A total of 31 consecutive patients who underwent single-level posterior lumbar interbody fusion surgery were included. Two 3D-Ti cages with different designs were inserted: a non-window cage on the left side and a window cage on the right side. Radiographic fusion was defined by the segmental angle between flexion and extension radiographs (F-E angle) and cage bridging bone (CBB) scores on computed tomography. The association between the F-E angle and osteointegration scoring system including the surface osteointegration ratio (SOR) score was analyzed.
Results
Radiographic fusion was achieved in 27 of 31 patients (87%) at 12 months postoperatively. Among the non-window cages, 23 of 31 (74.2%) had fair SOR scores, while 19 of 31 (61.3%) window cages had fair intra-cage CBB scores. The higher the SOR score was, the smaller the flexion-extension angle (SOR 0 vs. SOR 1: 6.30° ± 2.43° vs. 1.95° ± 0.99°, p < 0.001; SOR 0 vs. SOR 2: 6.03° ± 2.43° vs. 0.99°± 0.74°, p < 0.001).
Conclusions
The clinical feasibility of using a non-window 3D-Ti cage during lumbar interbody fusion might be acceptable. Furthermore, a newly suggested fusion criterion for the use of the non-window cage, the SOR score, showed a significant association with the published fusion grading systems, demonstrating its feasibility in determining interbody fusion in lumbar spinal surgery.
Keywords: Titanium, Printing, Lumbar surgery, Spinal fusion, Osseointegration
Lumbar interbody fusion has become one of the most commonly performed procedures for degenerative disease of the lumbar spine,1) showing favorable clinical outcomes. To achieve desirable results, solid fusion of the operated segment is crucial.2) Since Hacker et al.3) successfully reported using an interbody cage, various materials and designs for interbody cages have been introduced, including stainless steel, polyetheretherketone (PEEK), titanium alloy, and carbon fibers, to provide early mechanical stability and a fusion bed for the operated segments.
Among these materials, PEEK and titanium are the most widely used materials. PEEK has a similar elastic modulus to the vertebral body, which could decrease the stress-shielding effect on the operated segment.4) However, its poor biocompatibility inhibits bone growth on the cage surface.5,6) In contrast, titanium alloy has superior biocompatibility, promotes osteoblast migration, and results in bone formation on the surface.7,8) Still, its high elastic modulus compared with the vertebral body remains a concern.9,10)
Recently, three-dimensional (3D) printers have allowed the generation of a porous structured titanium cage, which minimizes the disadvantages of titanium alloy including its high elastic modulus and metal artifact on computed tomography (CT) and magnetic resonance imaging while maintaining its biocompatibility.11) The commercially available design of a 3D-printed titanium (3D-Ti) cage can be divided into two types according to the presence of a window: a cage with a window that allows filling of bone graft materials and a non-window cage for standalone use. In the case of a window cage, the grafted material in the cage void can induce new bone bridging, which is not possible with non-window cages.
To our knowledge, no studies have prospectively evaluated the bone ingrowth patterns between 3D-Ti window and non-window cages. We hypothesized that 3D-Ti cages with and without a window will provide equally favorable radiographic outcomes. Therefore, we aimed to evaluate the clinical feasibility of using a 3D-Ti non-window cage in a case series involving combined implantation of window and non-window cages before conducting a randomized trial and to evaluate the radiographic fusion patterns of non-window cages and their correlation with published fusion grading systems.
METHODS
Study Design and Participants
This study was a prospective case series conducted at three institutes and was approved by each Institutional Review Board (No. 2305-001-19468). A total of 31 consecutive patients, who signed informed consents and planned to undergo single-level posterior lumbar interbody fusion (PLIF) surgery for degenerative lumbar disease between July 2018 and January 2020, were included. The inclusion criteria were as follows: (1) patients aged ≥ 18 years and ≤ 85 years and (2) patients with available stress radiographs and multiaxial reconstructed CT scans obtained 12 months postoperatively. Patients with medical conditions that could inhibit the fusion process, including malignancy, infection, and other metabolic diseases, were excluded.
Surgical Procedure
Three spine surgeons from three different institutions (KSS, DGC, and JJY) performed PLIF with identical procedures. After disc preparation, an equal amount of 6 mL of extra-cage bone graft, consisting of a half-mixed combination of local autologous bone from laminectomy and allograft, was used to fill the anterior intervertebral space bilaterally. Subsequently, 3D-Ti cages (Genoss, Suwon, Korea) with different designs were inserted: a non-window cage was used on the left side, and a window cage filled with local autobone was used on the right side (Fig. 1).
Fig. 1. Illustration of the surgical method using bilateral interbody three-dimensional–printed titanium cages with different graft compositions and bone graft. Prior to the cages being inserted, an equal amount of 6 mL of bone graft material, consisting of a half-mixed combination of autologous local bone and allograft chip bone, was placed in the anterior intervertebral space bilaterally.
Determination of Fusion Criteria and Radiologic Evaluation
Radiographic fusion was defined on both stress radiographs and CT 12 months postoperatively according to the agreement of two spine surgeons (DWH and CWJ). Fusion in stress radiographs was defined when the following criteria were fulfilled: (1) no cage subsidence more than 3 mm and (2) less than 3° difference in the segmental angle between flexion and extension radiographs (F-E angle) at 12 months postoperatively.12,13)
On multiaxial reconstructed CT, we used the following scoring system, which was previously reported,12) to investigate the bridging bone patterns with the two different titanium cages. For the window cage, we used the intra-cage bridging bone (InCBB) score and extra-cage bridging bone score. This system evaluates the new bone formation pattern intra- and extra-cage and assigns 0 to 2 points according to the degree of completion of bone bridging (grade 0: no bridging; grade 1: incomplete bridging or bridging at the superior or inferior endplate with a clear radiolucent line; grade 2: complete bridging). The higher the sum of these scores was, the more firm and stable fusion was achieved in the segment.12)
For the non-window cage, we developed a surface osteointegration ratio (SOR) score that evaluates the osteointegration ratio between the cage’s surface and the endplates on sagittal CT images. When a translucent line was visible between the cage and the endplate in more than 50% of the sagittal images, zero points were assigned. When more than 50% of the endplate was integrated with the cage without a translucent line, 1 point was assigned. The scores for the upper and lower endplates were summed for a maximum score of 2 (Fig. 2).
Fig. 2. Newly introduced radiographic osteointegration evaluation system for the non-window type cage. The surface osteointegration ratio (SOR) score assesses the level of osteointegration and translucency between the cage surface and each upper and lower endplate. One point is assigned to each endplate when more than 50% of the endplate is integrated with the cage without translucency, for a maximum score of 2. The left image shows a SOR score of 0 point, as osteointegration is less than 50% on both the upper and lower endplates. The middle image shows a SOR score of 1 point, as osteointegration is 50% or more on the upper endplate but less than 50% on the lower endplate. The right image shows a SOR score of 2 points, as osteointegration is 50% or more on both the upper and lower endplates.
Clinical Outcome Measurement
The clinical outcomes were evaluated using patient-reported outcome (PRO) measures including the leg pain numeric rating scale (NRS), the 12-item short form survey (SF-12), the Oswestry Disability Index (ODI), and the EuroQoL-5 dimension (EQ-5D). The ODI14) is a self-administered questionnaire that measures “back-specific function” on a 9-item scale with six response categories each. The EQ-5D is a five-dimensional health state classification; the five dimensions are mobility, self-care, usual activities, pain/discomfort, and anxiety/depression. An EQ-5D “health state” is defined by selecting one level from each dimension. The EQ-5D preference-based measure can be regarded as a continuous outcome scored on a 0 to 1.00 scale, with 1.00 indicating “full health” and 0 representing death. These data were collected preoperatively and reassessed at 3, 6, and 12 months postoperatively. In 12 domains, the SF-1215) evaluates a patient’s physical, social, and mental health. It consists of two components including physical and mental parts that summarize a patient’s health status. The population mean for the SF-12 scale is 50, with a standard deviation of 10. A higher score indicates a healthier condition.
Statistical Analysis
In calculating the sample size, the primary endpoint of this study was the proportion of endplate-cage surface osteointegration in the fusion group and the anticipated proportion of osteointegration in the fusion group at 12 months postoperatively was 90%. With a two-sided 95% confidence interval of 12%, 25 fusion segments were needed. Assuming a fusion rate with the 3D-Ti cage of 90%,16) a total of 28 patients were needed. Considering a dropout rate of 10%, we enrolled 32 patients in this study. G*Power software version 3.1.9.7 (Düsseldorf, Germany) was used to calculate the sample size. The continuous variables between the fusion and non-fusion groups were analyzed using independent t-tests. Differences in the F-E angle according to the SOR score were compared using analysis of variance. Simple linear regression analysis was performed to evaluate the correlation between the CBB and SOR scores. In addition, a subgroup analysis was performed within the fusion group to compare the bone formation patterns between the window and non-window cage sides. All statistical analyses were performed using IBM SPSS ver. 23.0 (IBM Corp., Armonk, NY, USA).
RESULTS
Patient Characteristics and Surgical Outcomes
A total of 31 patients with a mean age of 74.2 ± 6.9 years were enrolled in the study. Baseline demographic parameters and the PRO results were compared according to the fulfillment of fusion criteria (Table 1). Fusion criteria were fulfilled in 27 of 31 patients (87%) at 12 months postoperatively. Age, sex, operation level, and cage height were not significantly different between the fusion and non-fusion groups (Table 1). All PRO measures including the ODI, leg pain NRS, SF-12, and EQ-5D scores were improved at 12 months postoperatively compared with preoperative values, and there were no significant differences between the fusion and non-fusion groups (Table 1).
Table 1. Descriptive Analysis of Clinical Characteristics in the Present Study.
Variable | Fused | Non-fused | Total | p-value | |
---|---|---|---|---|---|
Number of patients | 27 (87) | 4 (13) | 31 | - | |
Age (yr) | 73.8 ± 6.9 | 77.2 ± 6.3 | 74.2 ± 6.9 | 0.353 | |
Female sex | 20 (74) | 2 (5) | 22 (71) | 0.689 | |
Operation level | 0.431 | ||||
L2–3 | 1 (3.7) | 0 | 1 (3.2) | ||
L4–5 | 11 (40.7) | 3 (75.0) | 14 (45.2) | ||
L5–S1 | 15 (55.6) | 1 (25.0) | 16 (51.6) | ||
Cage height (mm) | 9.3 ± 0.6 | 9.0 ± 0.8 | 9.3 ± 0.6 | 0.341 | |
ODI | |||||
Preoperative | 0.505 ± 0.159 | 0.565 ± 0.174 | 0.512 ± 0.159 | 0.554 | |
Postoperative 12 mo | 0.149 ± 0.112 | 0.198 ± 0.149 | 0.155 ± 0.156 | 0.568 | |
NRS | |||||
Preoperative | 7.8 ± 1.3 | 8.0 ± 1.6 | 7.8 ± 1.3 | 0.840 | |
Postoperative 12 mo | 1.8 ± 1.4 | 3.3 ± 2.2 | 2.0 ± 1.6 | 0.289 | |
SF-12 | |||||
Preoperative | 33.0 ± 10.8 | 24.0 ± 3.9 | 31.9 ± 10.6 | 0.112 | |
Postoperative 12 mo | 42.1 ± 8.1 | 41.0 ± 10.5 | 42.0 ± 8.3 | 0.800 | |
EQ-5D | |||||
Preoperative | 0.420 ± 0.219 | 0.432 ± 0.126 | 0.422 ± 0.208 | 0.876 | |
Postoperative 12 mo | 0.808 ± 0.124 | 0.722 ± 0.257 | 0.797 ± 0.144 | 0.555 |
Values are presented as number (%) or mean ± standard deviation.
ODI: Oswestry Disability Index, NRS: numeric rating scale, SF-12: 12-item short form survey, EQ-5D: EuroQoL-5 dimension.
F-E Angle and SOR Score with the Non-window Cage
The F-E angle in the group with SOR scores of 1 and 2 was significantly lower than that in the group with an SOR score of 0; in contrast, there was no significant difference in the F-E angle between the group with an SOR score of 1 and the group with an SOR score of 2 (SOR 0 vs. SOR 1: 6.30° ± 2.43° vs. 1.95° ± 0.99°, p < 0.001; SOR 0 vs. SOR 2: 6.03° ± 2.43° vs. 0.99° ± 0.74°, p < 0.001; SOR 1 vs. SOR 2: 1.95° ± 0.99° vs. 1.80°± 2.07°, p = 0.248) (Fig. 3).
Fig. 3. Association between the flexion-extension (F-E) angle and surface osteointegration ratio (SOR) score. The F-E angle in the group with SOR scores of 1 and 2 was significantly lower than that in the group with an SOR score of 0. Values are presented as mean ± standard deviation. *p < 0.05.
Correlation of Bridging Bone Scoring System and SOR Score System
The correlation between the CBB and SOR scores was evaluated using simple linear regression analysis. The regression analysis showed that the SOR score was statistically significantly positively correlated with the bridging bone scoring systems (R2 = 0.356, p < 0.001) (Table 2).
Table 2. Univariate Regression Analysis for Osteointegration Scoring Systems of the Communicating Bridging Bone Score.
Univariate regression analysis for bridging bone score | |||
---|---|---|---|
β (95% confidence interval) | R 2 | p-value | |
SOR score | 2.407 (1.806-3.001) | 0.356 | < 0.001 |
SOR: surface osteointegration ratio. Systems of the Communicating Bridging Bone Score
Intra-cage Osteointegration Patterns of Two Types of Cages
The cages that scored 2 points in the InCBB score, SOR score, or had a positive CB sign were classified into the fair intra-cage bone formation group. Among non-window cages, 23 out of 31 (74.2%) cages showed fair SOR scores (Fig. 4). On the other hand, among window cages, 19 out of 31 (61.3%) cages showed fair InCBB scores, indicating fair new bone formation in the void (Fig. 4).
Fig. 4. Comparison of the intra-cage new bone formation ratio of the two types of cages. Among non-window cages, 23 out of 31 cages (74.2%) showed fair surface osteointegration ratio (SOR) scores, while among the window cages, 19 of 31 cages (61.3%) showed fair intra-cage bridging bone scores, indicating fair new bone formation in the void.
DISCUSSION
The present study demonstrated that patients who underwent PLIF with a non-window 3D-Ti cage without bone graft material on one side showed favorable radiologic and clinical outcomes. Furthermore, the osteointegration scoring system that evaluates bone growth around the 3D-Ti cage, the SOR, showed favorable agreement with the known published radiographic fusion criteria.12)
The overall fusion rate in the present study was 87.1%, which is consistent with the findings of previous studies,17,18) reporting a fusion rate of 80%–86% at 1 year postoperatively. Furthermore, the results from CT scans confirmed that even without bone graft materials, sufficient endplate-cage intersurface bone ingrowth was achieved on the non-window cage side. Arts et al.19) reported that a 3D-Ti cage accelerated the fusion process in the early postoperative period even without a bone graft or biomaterial compared with a PEEK cage in cervical fusion surgery. This result is in accordance with the present study, which demonstrated a consistent surface osteointegration score compared with the window-type cage side with a favorable interbody fusion rate.
Titanium alloy is a biocompatible material that can promote bone growth and has been used for decades in the orthopedic and spinal surgery fields.20) The elastic modulus of earlier titanium alloy cages was much stronger than that of bone.21) Therefore, cage subsidence has been reported due to a stress-shielding effect in the fusion segment.22) By comparison, a 3D-Ti cage decreases the elastic modulus, which prevents the stress-shielding effect.23) However, it remains controversial whether the void structure within the cage for bone graft materials affects the structural stability of a 3D-printed cage. For this reason, a non-window type cage has been used, but in this case, there is controversy regarding whether bone graft material should be placed in the cage because it may be detrimental for interbody fusion. However, from the present study, we confirmed that sufficient bone growth in the endplate-cage intersurface could be achieved on the non-window cage side. These results suggest that favorable interbody fusion can be achieved even with a non-window titanium cage alone without bone graft material while providing sufficient structural stability.
The ideal interbody cage must be rigid enough to maintain stability but preferably have a similar Young’s modulus to that of bone to prevent subsidence and stress-shielding.24) Titanium alloy cages have excellent biocompatibility and mechanical stability.7,8) Still, due to the difference in the elastic modulus between titanium and cortical bone, a higher cage subsidence rate than with a PEEK cage remains a concern.9,10) Several factors are reportedly associated with cage subsidence following interbody fusion including poor endplate bone quality, narrow and tall cage, and endplate injury during disc preparation.25,26,27,28,29,30,31) Therefore, careful disc preparation to avoid endplate injury, especially during the use of a 3D-Ti cage, is a crucial factor for preventing cage subsidence (Fig. 5).
Fig. 5. Representative case for cage subsidence. (A) A 64-year-old male patient who underwent L4–5 posterior lumbar interbody fusion surgery. (B, C) Cage subsidence was confirmed 1 year postoperatively on simple radiographs and multiaxial computed tomography scans.
The feasibility of using fusion criteria to assess bridging bone is a concern that remains in spinal surgery using a titanium cage. The Bridwell interbody fusion grading system is the most well-known tool used to evaluate fusion rates in spinal surgery; still, due to metal artifacts associated with titanium alloy, concerns about the feasibility of its use remain.32) Recently, our team suggested a scoring system12) that evaluates intra-cage and extra-cage osteogenesis patterns on CT scans and reported its usefulness in determining interbody fusion in lumbar fusion surgery. In the present study, we have presented the SOR score as a novel grading system to evaluate the endplate-cage intersurface osteointegration (Fig. 2). The results demonstrated that the SOR score was significantly correlated with the bridging bone score and with the F-E angle on dynamic radiographs (Table 2 and Fig. 3). In this context, the SOR scoring system with the 3D-Ti cage seems to be clinically useful for determining segment fusion in lumbar spinal surgery. In addition, previous research by Segi et al.33) reported that patients with a positive trabecular remodeling sign had lower rates of instrument failure and pseudoarthrosis. In our study, the communicating bridging sign, which indicates trabecular remodeling, (Fig. 6) was also observed in patients who achieved firm fusion. Considering this, we believe that the communication bridging sign could be a clinical determinant for evaluating fusion in non-window 3D-Ti cages.
Fig. 6. The communicating bridging (CB) sign on sagittal computed tomography images (white arrows) is considered positive when newly generated bone passes through the internal structure of the cage itself with trabecular remodeling signal.
On the comparison of the intra-cage new bone formation ratio of the two types of cages, 23 cases (74.2%) with the non-window cage had fair SOR scores, while only 19 cases (61.3%) with the window cage side had fair InCBB scores (Fig. 4). Although direct comparison of the two types of cages in the current study design is difficult, these results could demonstrate that the non-window type cage can achieve sufficient surface osteointegration similar to the window type cage. Furthermore, these results support the clinical applicability of the SOR scoring system for evaluating fusion with the non-window 3D-Ti cage.
This study has several limitations. First, the study population was limited in size, and the follow-up duration was relatively short. While we elucidated the sample size calculation process in the Materials and Methods section, we acknowledge that a cohort of 31 cases may not provide sufficient statistical power. However, we believe that this pilot study, which aims to assess the feasibility of non-window cages rather than directly comparing window and non-window cages on a 1 : 1 basis, can have clinical significance with the current design. Second, in the current study design, a direct comparison of two cages is difficult because the two types of cages were inserted in the same segment, potentially influencing each other’s cage-endplate fusion. This is because if the cage-endplate fusion is first achieved on one side cage, the motion on the other side is reduced and the fusion process can be facilitated, which can bias the accurate evaluation. However, it should be considered that this study was a preliminary study to evaluate the clinical feasibility of using a 3D-Ti non-window type cage before performing a randomized trial. Despite these limitations, our study has some strengths in that the intra-cage bone formation patterns of the window and non-window type 3D-Ti cages were descriptively compared. A further randomized prospective study involving an objective comparative analysis with the window cage is being conducted.
In conclusion, the present study demonstrated that the clinical feasibility of using a non-window 3D-Ti cage for lumbar interbody fusion might be acceptable. Furthermore, the osteointegration tendency of the cage indicated that a 3D non-window cage itself could act as fusion bed without any bone graft materials in the cage void. Furthermore, a newly suggested fusion criterion for non-window cages, the SOR score, showed a significant association with the bridging bone score and F-E angle on dynamic radiographs, demonstrating its feasibility in determining interbody fusion in lumbar spinal surgery.
Footnotes
CONFLICT OF INTEREST: No potential conflict of interest relevant to this article was reported.
References
- 1.Yavin D, Casha S, Wiebe S, et al. Lumbar fusion for degenerative disease: a systematic review and meta-analysis. Neurosurgery. 2017;80(5):701–715. doi: 10.1093/neuros/nyw162. [DOI] [PubMed] [Google Scholar]
- 2.Buser Z, Brodke DS, Youssef JA, et al. Synthetic bone graft versus autograft or allograft for spinal fusion: a systematic review. J Neurosurg Spine. 2016;25(4):509–516. doi: 10.3171/2016.1.SPINE151005. [DOI] [PubMed] [Google Scholar]
- 3.Hacker RJ, Cauthen JC, Gilbert TJ, Griffith SL. A prospective randomized multicenter clinical evaluation of an anterior cervical fusion cage. Spine (Phila Pa 1976) 2000;25(20):2646–2655. doi: 10.1097/00007632-200010150-00017. [DOI] [PubMed] [Google Scholar]
- 4.Vadapalli S, Sairyo K, Goel VK, et al. Biomechanical rationale for using polyetheretherketone (PEEK) spacers for lumbar interbody fusion-A finite element study. Spine (Phila Pa 1976) 2006;31(26):E992–E998. doi: 10.1097/01.brs.0000250177.84168.ba. [DOI] [PubMed] [Google Scholar]
- 5.Noiset O, Schneider YJ, Marchand-Brynaert J. Fibronectin adsorption or/and covalent grafting on chemically modified PEEK film surfaces. J Biomater Sci Polym Ed. 1999;10(6):657–677. doi: 10.1163/156856299x00865. [DOI] [PubMed] [Google Scholar]
- 6.De Bartolo L, Morelli S, Bader A, Drioli E. The influence of polymeric membrane surface free energy on cell metabolic functions. J Mater Sci Mater Med. 2001;12(10-12):959–963. doi: 10.1023/a:1012857031409. [DOI] [PubMed] [Google Scholar]
- 7.Chong E, Pelletier MH, Mobbs RJ, Walsh WR. The design evolution of interbody cages in anterior cervical discectomy and fusion: a systematic review. BMC Musculoskelet Disord. 2015;16:99. doi: 10.1186/s12891-015-0546-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Rao PJ, Pelletier MH, Walsh WR, Mobbs RJ. Spine interbody implants: material selection and modification, functionalization and bioactivation of surfaces to improve osseointegration. Orthop Surg. 2014;6(2):81–89. doi: 10.1111/os.12098. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Cabraja M, Oezdemir S, Koeppen D, Kroppenstedt S. Anterior cervical discectomy and fusion: comparison of titanium and polyetheretherketone cages. BMC Musculoskelet Disord. 2012;13:172. doi: 10.1186/1471-2474-13-172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Chen Y, Wang X, Lu X, et al. Comparison of titanium and polyetheretherketone (PEEK) cages in the surgical treatment of multilevel cervical spondylotic myelopathy: a prospective, randomized, control study with over 7-year follow-up. Eur Spine J. 2013;22(7):1539–1546. doi: 10.1007/s00586-013-2772-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Li P, Jiang W, Yan J, et al. A novel 3D printed cage with microporous structure and in vivo fusion function. J Biomed Mater Res A. 2019;107(7):1386–1392. doi: 10.1002/jbm.a.36652. [DOI] [PubMed] [Google Scholar]
- 12.Lee J, Lee DH, Jung CW, Song KS. The significance of extra-cage bridging bone via radiographic lumbar interbody fusion criterion. Global Spine J. 2023;13(1):113–121. doi: 10.1177/2192568221993097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Larsen JM, Rimoldi RL, Capen DA, Nelson RW, Nagelberg S, Thomas JC., Jr Assessment of pseudarthrosis in pedicle screw fusion: a prospective study comparing plain radiographs, flexion/extension radiographs, CT scanning, and bone scintigraphy with operative findings. J Spinal Disord. 1996;9(2):117–120. [PubMed] [Google Scholar]
- 14.Fairbank JC, Pynsent PB. The Oswestry Disability Index. Spine (Phila Pa 1976) 2000;25(22):2940–2952. doi: 10.1097/00007632-200011150-00017. [DOI] [PubMed] [Google Scholar]
- 15.Godil SS, Parker SL, Zuckerman SL, Mendenhall SK, McGirt MJ. Accurately measuring the quality and effectiveness of cervical spine surgery in registry efforts: determining the most valid and responsive instruments. Spine J. 2015;15(6):1203–1209. doi: 10.1016/j.spinee.2013.07.444. [DOI] [PubMed] [Google Scholar]
- 16.Massaad E, Fatima N, Kiapour A, Hadzipasic M, Shankar GM, Shin JH. Polyetheretherketone versus titanium cages for posterior lumbar interbody fusion: meta-analysis and review of the literature. Neurospine. 2020;17(1):125–135. doi: 10.14245/ns.2040058.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Brantigan JW, Steffee AD, Lewis ML, Quinn LM, Persenaire JM. Lumbar interbody fusion using the Brantigan I/F cage for posterior lumbar interbody fusion and the variable pedicle screw placement system: two-year results from a Food and Drug Administration investigational device exemption clinical trial. Spine (Phila Pa 1976) 2000;25(11):1437–1446. doi: 10.1097/00007632-200006010-00017. [DOI] [PubMed] [Google Scholar]
- 18.Madan S, Boeree NR. Outcome of posterior lumbar interbody fusion versus posterolateral fusion for spondylolytic spondylolisthesis. Spine (Phila Pa 1976) 2002;27(14):1536–1542. doi: 10.1097/00007632-200207150-00011. [DOI] [PubMed] [Google Scholar]
- 19.Arts M, Torensma B, Wolfs J. Porous titanium cervical interbody fusion device in the treatment of degenerative cervical radiculopathy: 1-year results of a prospective controlled trial. Spine J. 2020;20(7):1065–1072. doi: 10.1016/j.spinee.2020.03.008. [DOI] [PubMed] [Google Scholar]
- 20.Palm WJ, 4th, Rosenberg WS, Keaveny TM. Load transfer mechanisms in cylindrical interbody cage constructs. Spine (Phila Pa 1976) 2002;27(19):2101–2107. doi: 10.1097/00007632-200210010-00005. [DOI] [PubMed] [Google Scholar]
- 21.Smit TH, Muller R, van Dijk M, Wuisman PI. Changes in bone architecture during spinal fusion: three years follow-up and the role of cage stiffness. Spine (Phila Pa 1976) 2003;28(16):1802–1809. doi: 10.1097/01.BRS.0000083285.09184.7A. [DOI] [PubMed] [Google Scholar]
- 22.van Dijk M, Smit TH, Sugihara S, Burger EH, Wuisman PI. The effect of cage stiffness on the rate of lumbar interbody fusion: an in vivo model using poly(l-lactic Acid) and titanium cages. Spine (Phila Pa 1976) 2002;27(7):682–688. doi: 10.1097/00007632-200204010-00003. [DOI] [PubMed] [Google Scholar]
- 23.Chen C, Hao Y, Bai X, et al. 3D printed porous Ti6Al4V cage: Effects of additive angle on surface properties and biocompatibility; bone ingrowth in Beagle tibia model. Mater Des. 2019;175:107824 [Google Scholar]
- 24.Tan JH, Cheong CK, Hey HW. Titanium (Ti) cages may be superior to polyetheretherketone (PEEK) cages in lumbar interbody fusion: a systematic review and meta-analysis of clinical and radiological outcomes of spinal interbody fusions using Ti versus PEEK cages. Eur Spine J. 2021;30(5):1285–1295. doi: 10.1007/s00586-021-06748-w. [DOI] [PubMed] [Google Scholar]
- 25.Rentenberger C, Okano I, Salzmann SN, et al. Perioperative risk factors for early revisions in stand-alone lateral lumbar interbody fusion. World Neurosurg. 2020;134:e657–e663. doi: 10.1016/j.wneu.2019.10.164. [DOI] [PubMed] [Google Scholar]
- 26.Okano I, Jones C, Rentenberger C, et al. The association between endplate changes and risk for early severe cage subsidence among standalone lateral lumbar interbody fusion patients. Spine (Phila Pa 1976) 2020;45(23):E1580–E1587. doi: 10.1097/BRS.0000000000003668. [DOI] [PubMed] [Google Scholar]
- 27.Chatham LS, Patel VV, Yakacki CM, Dana Carpenter R. Interbody spacer material properties and design conformity for reducing subsidence during lumbar interbody fusion. J Biomech Eng. 2017;139(5):0510051–0510058. doi: 10.1115/1.4036312. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Lambrechts MJ, Heard J, D’Antonio N, et al. A comparison of radiographic alignment between bilateral and unilateral interbody cages in patients undergoing transforaminal lumbar interbody fusion. Asian Spine J. 2023 May 25; doi: 10.31616/asj.2022.0316. [Epub] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Fujibayashi S, Takemoto M, Ishii K, et al. Multicenter prospective study of lateral lumbar interbody fusions using bioactive porous titanium spacers without bone grafts. Asian Spine J. 2022;16(6):890–897. doi: 10.31616/asj.2021.0354. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Kim YH, Ha KY, Kim YS, et al. Lumbar interbody fusion and osteobiologics for lumbar fusion. Asian Spine J. 2022;16(6):1022–1033. doi: 10.31616/asj.2022.0435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Kaliya-Perumal AK, Soh TL, Tan M, Oh JYL. Early postoperative loss of disc height following transforaminal and lateral lumbar interbody fusion: a radiographic analysis. Asian Spine J. 2022;16(4):471–477. doi: 10.31616/asj.2021.0109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Nemoto O, Asazuma T, Yato Y, Imabayashi H, Yasuoka H, Fujikawa A. Comparison of fusion rates following transforaminal lumbar interbody fusion using polyetheretherketone cages or titanium cages with transpedicular instrumentation. Eur Spine J. 2014;23(10):2150–2155. doi: 10.1007/s00586-014-3466-9. [DOI] [PubMed] [Google Scholar]
- 33.Segi N, Nakashima H, Shinjo R, et al. Trabecular bone remodeling as a new indicator of osteointegration after posterior lumbar interbody fusion. Global Spine J. 2022 Apr 12; doi: 10.1177/21925682221090484. [Epub] [DOI] [PMC free article] [PubMed] [Google Scholar]