Skip to main content
Journal of Applied Clinical Medical Physics logoLink to Journal of Applied Clinical Medical Physics
. 2023 Nov 13;24(12):e14103. doi: 10.1002/acm2.14103

Validation of the geometric equivalent field concept in total scatter factor calculations, for half‐, quarter‐ and off‐isocenter asymmetric square fields

Mohammad Samir Hmodi 1,2, Majeda Nahili 1, Ioannis A Tsalafoutas 3,, Bassam Saad 2,4, Ali Hasan 2, Ousamah Anjak 5, Karlos Shamout 2
PMCID: PMC10691636  PMID: 37957949

Abstract

Objective

Monitor unit (MU) verification for any symmetric or asymmetric field is performed using a total scatter factor (Scp), that is calculated based on the geometric equivalent square field (GESF) concept. In this study, we measured the Scp of various asymmetric square fields (ASFs) and their respective GESFs.

Methods

Square half‐fields (SHFs), square quarter‐fields (SQFs) and square off‐isocenter fields (SOFs), with sizes ranging from 3×3 cm2 to 20×20 cm2 were created, by varying the collimator jaws of two Varian iX Linacs (6/18 and 6/23 MV). A semi‐flex ion chamber was used to measure Scp at a depth of 10 cm within a water phantom, at the effective field center (EFC) of all ASFs, and at the isocenter (IC) of their respective GESFs. The later Scp values were corrected by the off‐axis ratio [OAR(r)] of the 40×40 cm2 field size, where r is the distance between EFC and IC.

Results

The results show that the Scp (EFC) is independent of the type of the ASF (SHF, SQF, or SOF) and no significant difference exists between the 18 and 23 MV beams. Compared with the Scp (IC), the Scp (EFC) increased with increasing r, by up to 2% and 4% for 18/23 and 6 MV, respectively.

Conclusions

The GESF concept provides acceptable accuracy (< 2%) for the calculation of Scp of the ASFs used in most clinical situations (except from SOF with EFC at large r), and thus can be used in MU verification calculations.

Keywords: asymmetric square fields, equivalent fields, monitor unit verification, off‐axis ratio, total scatter factor

1. INTRODUCTION

The total scatter factor (Scp), also referred to as output factor in water, has been always an important parameter for the calculation of monitor units (MUs) in radiotherapy, as a function of the treatment field (TF) size. Two different approaches have been used to calculate the Scp. In the first approach, which is also applicable in irregular TFs, Scp is derived by the product of two factors, Sc and Sp. 1 , 2 Sc is the head or collimator scatter factor, which is related to the collimator opening dimensions. Sc accounts for the increase in output observed with increased field size, attributed to the increased amount of radiation scattered by the collimator jaws that reaches the phantom (or the patient). Sp is the phantom scatter factor, which is related to the amount of scatter produced within the irradiated volume of the phantom/patient. For the Sp calculations it is taken into account that the presence of blocks or a multi‐leaf collimator (MLC), modify the dimensions of the TF and therefore the amount of scattered radiation produced within the phantom/patient. In the second approach, the Scp is taken directly from tables, for different field sizes (collimator dimensions) and beam energies, 3 and incorporates both the head and phantom scatter. However, this approach is applicable in manual MU calculations for regular treatment fields only (when no blocks are used) or for a quick cross‐check of the MU calculated by a treatment planning system (TPS).

TFs may be either symmetric or asymmetric, depending on the horizontal distances of the collimator jaws to the isocenter (IC). TFs are usually described using the notation (y1,y2,x1,x2), where y1, y2, x1 and x2 are the respective collimator jaw positions (in cm) determined at the level of the IC. Asymmetric fields are commonly used in radiotherapy, and regarding treatment planning, there are three special categories of interest (for single‐isocenter techniques), according to the location of the IC with respect to the TF: (a) The IC is centered at an edge of the TF, as in the case of the half‐beams used in the treatment of breast, 4 as shown in Figure 1a, head and neck, 5 and craniospinal tumors, 6 , 7 (b) The IC is located at a corner of the TF as in the case of a quarter‐beams used for the treatment of the chest wall in breast cancer cases, 4 , 8 as shown in Figure 1b, c) The IC is outside TF, in special cases like the off‐isocenter‐beams are used for the treatment of the supraclavicular area and the boost field used for breast tumors, 9 as shown in Figure 1c and d, respectively.

FIGURE 1.

FIGURE 1

Examples of clinical uses of asymmetric fields are given. (a) Supraclavicular field using a half‐beam. (b) Tangential fields using quarter‐beams. (c) Supraclavicular field using off‐ isocenter beam. (d) Boost field using off‐ isocenter beams. Green marks indicate the EFC. EFC, effective field center.

In all special categories mentioned above, the IC is partially or completely blocked, so MUs are calculated at the effective field center (EFC) rather than the IC. The distance between EFC and IC is called off‐axis distance (r). In MU calculations based on the geometric equivalent square field (GESF) concept, the use of dosimetric quantities like Scp, PDD, and TMR determined using symmetric fields will lead to significant errors, should the off‐axis ratio (OAR) value of the maximum field size (i.e., 40 × 40 cm2) at the location of the EFC is not taken into account. This problem has been studied for square half‐fields (SHFs) only, and the errors observed were up to 3% regarding the output factors (Sc and Sp) calculation, 10 and 7% regarding dose calculations within the phantom/patient. 11

On the other hand, previous studies have shown that the Scp of an asymmetric field, corrected by the OAR value at the location of EFC, are approximately equal to the corresponding values measured at the IC of their GESF, and in this case differences in MU calculations are less than 1%. 12 , 13 , 14 , 15 , 16 , 17 Thus, several mathematical formulas were developed to improve the accuracy of calculations for asymmetric fields based on the GESF concept, 12 , 18 , 19 , 20 , 21 the simplest of which is the area‐perimeter ratio. 22 According to these formulas, in order to calculate the Scp for asymmetric fields, the Sc and Sp of their GESF can be used. 3 , 23 , 24 , 25 , 26

In the referenced studies this conclusion has been validated for almost all Varian Linacs, such as Clinac 4/100 with 4 MV beam, 600C with 6 MV beam, 2100C with energies 6 and 10 MV, 2100CD with energies 6 and 15 MV, and the Linacs 2100C, 2300C/D, Clinac‐DHX (with energies 6 and 18MV), but no relevant data exist for Varian Clinac IX accelerators, or the 23 MV beam energy. Most important, few studies have measured the Scp of half‐fields, while fewer studies have measured the Scp of the off‐isocenter and quarter‐fields. 11 , 17 Moreover, there are no detailed data regarding the Scp measurements.

The aim of this study is to directly measure the Scp of Varian Clinac IX accelerators for the half‐, quarter‐ and off‐isocenter fields and compare it with the values measured at the respective GESFs. This study will be limited to asymmetric square fields (ASFs), for which the effective equivalent field is the same as that of symmetric square field (SSF) whose Scp are tabulated and therefore no approximation is needed. For example, using the notation (y1,y2,x1,x2) described above, a SHF with dimensions (0, 10, 5, 5), a square quarter‐field (SQF) with dimensions (0, 10, 0, 10) and a square off‐isocenter field (SOF) with dimensions (−10, 20, 5, 5) have all a respective GESF size of 10 × 10 cm2. 21 , 22

2. MATERIALS AND METHODS

2.1. Linear accelerators (LINAC)

This study was performed in two Varian IX Linacs (Varian Medical Systems, Palo Alto, California, USA). The first Linac (L1) emits 6 and 23 MV energy photon beams, while the second Linac (L2) emits 6 and 18 MV energy photon beams. For each Linac, the dimensions of the treatment field are determined by two pairs of secondary collimators jaws, the upper jaw (y1, y2) and the lower jaw (x1, x2). It should be noted that x1 and y1 can move beyond the IC, for a distance up to −2 and −10 cm, respectively.

When y1, y2 x1, x2, are all equal, the TF is SSF, as shown in Figure 2a. A half‐field is obtained by setting one of these jaws to position 0, and therefore there are four possible SHF sets (y1 = 0, or y2 = 0, or x2 = 0, or x1 = 0), which have their EFC located either on the X‐ or Y‐axis, as shown in Figure 2b. A quarter‐field is obtained by setting one jaw of each pair equal to 0 and there are four possible SQF sets (y1 = x2 = 0, y2 = x2 = 0, y1 = x1 = 0 or y2 = x1 = 0), which have their EFCs located on the diagonal axes, as shown in Figure 2c. Finally, two SQF sets were studied by setting y1 = −5 cm and y1 = −10 cm, which have their EFCs located on the positive Y‐axis, as shown in Figures 2d and e, respectively. The distance (r) between IC and EFC in quarter‐fields was calculated using Equation (1).

r=xi22+yi22 (1)

where i = 1 and/or 2

FIGURE 2.

FIGURE 2

Examples of the geometry of the studied treatment fields are given. Red and green marks indicate the IC and EFC locations, respectively, and r is the distance between them. For SHF (b) and SQF (c) there are four different EFC locations (marked with 1, 2, 3 and 4), for the different field size sets. EFC, effective field center; IC, isocenter; SHF, square half‐field.

2.2. Water phantom and ionization chamber description

A MP3 water phantom 600 × 500 × 408 mm3 (PTW, Freiburg, Germany) was used, connected with TANDEM dual‐channel electrometer, TBA control unit, and hand pendant. Its proper positioning was fine‐tuned using three laser sources and an inclinometer (Tajima, SLANT 100). To measure the charge, a semi‐flex chamber (PTW 31010, 0.125 cc) connected to a PTW UNIDOS electrometer was used. The chamber was located at a depth of 10 cm within the water phantom and the distance of the x‐ray source focus to the water's surface was adjusted to 90 cm and the gantry angle was kept constant at 0°. For each measurement, the ionization chamber was irradiated to 100 MU and the resulting charge (R) value was recorded by the electrometer.

2.3. Scp: Measurements and calculations

Scp for SSFs [Scp(IC)]

The ionization chamber was placed at the IC of all SSFs which are the respective GESF of one or more of the ASFs studied, that is, 3 × 3, 5 × 5, 10 × 10, 15 × 15, and 20 × 20 cm2, plus the 40 × 40 cm2. The Scp (IC) was calculated from Equation (2).

ScpIC=RICX×YRIC10×10cm2 (2)

where RIC(X × Y) and RIC(10 × 10 cm2) are the measured charges R at the IC for the SSFs (X = Y) and the reference field (10 × 10 cm2), respectively.

Scp for ASFs [Scp(EFC)]

Using the handheld pendant, the ionization chamber was moved at distance (r) to the EFC for all the ASFs studied, which are:

  1. SHFs: 3 × 3 cm2 (r = 1.5 cm), 5 × 5 cm 2 (r = 2.5 cm), 10 × 10 cm2 (r = 5 cm), 15 × 15 cm2 (r = 7.5 cm), 20 × 20 cm2 (r = 10 cm).

  2. SQFs: 3 × 3 cm2 (r = 2.12 cm), 5 × 5 cm 2 (r = 3.54 cm), 10 × 10 cm2 (r = 7.07 cm), 15 × 15 cm2 (r = 10.61 cm), 20 × 20 cm2 (r = 14.14 cm).

  3. SOFs with y1 = −5 cm: 3 × 3 cm2 (r = 6.5 cm), 5 × 5 cm2 (r = 7.5 cm), 10 × 10 cm2 (r = 10 cm), 15 × 15 cm2 (r = 12.5 cm).

  4. SOFs with y1 = −10 cm: 3 × 3 cm2 (r = 11.5 cm), 5 × 5 cm2 (r = 12. 5 cm), 10 × 10 cm2 (r = 15 cm).

The Scp(EFC) was calculated from Equation (3), which is similar to the head scatter (symbolized as Hs or Sc) Equations, defined in previous studies. 12 , 13 , 14

ScpEFC=REFC(y1,y2,x1,x2,r)RIC(10×10cm2)×OARr (3)

where REFC (y1, y2, x1, x2, r) and RIC (10 × 10 cm2) are the charges measured at the location of EFC and IC, respectively, and OAR(r) is the OAR measured in a water phantom for a point at a perpendicular distance r cm from the central axis for the maximum field size (40 × 40 cm2), 27 , 28 , 29 , 30 as defined in Equation (4).

OARr=Rr(40×40cm2)RIC(40×40cm2) (4)

where Rr (40 × 40 cm2) is the measured charge at a distance r from the IC and RIC (40×40 cm2) is the measured charge at the IC for 40 × 40 cm2 field size. In Figure 3 are shown the off‐axis points that correspond to the EFCs of the ASFs used in this study. For the measurement of Rr, the largest distance (r) from the IC was 15 cm, being within 80% of the largest field size and away from the field edges. 28 , 31 , 32

FIGURE 3.

FIGURE 3

The off‐axis points that correspond to EFCs points of the ASFs used in this study are shown. The red marks on the orthogonal and blue marks on diagonal axes denote the EFCs for the SHFs and SQFs, respectively. The green marks on the y2 axis denote the EFCs for the SOFs. ASFs, asymmetric square fields; EFC, effective field center; SHFs, square half‐fields; SOFs, square off‐isocenter fields; SQFs, square quarter‐fields.

To clarify the order of measurements and calculations made in this study, it must be stressed that before performing any measurements with ASFs, the values of the charge RIC(x, y) of all SSFs (which are the GESF of one or more of the ASFs studied, plus the 40 × 40 cm2) were first recorded, with the ion chamber positioned at the IC and irradiation of the phantom with 100 MU. For Scp(EFC) measurements with ASFs, an example of the measurements performed will be given for the SQF 5 × 5 cm2. The collimator jaws were first adjusted at (5, 0, 5, 0) and the chamber was moved at the respective EFC, which is on the diagonal (at distance r = 3.54 cm), the phantom was irradiated using 100 MU, and the value of the charge REFC(5, 0, 5, 0, r = 3.5 cm) was recorded (to get the numerator of Equation 3). Then, with the chamber kept the same place, the collimator jaws were adjusted to get the maximum SSF 40 × 40 cm2, and the reading Rr(40 × 40 cm2) was recorded for an irradiation with 100 MU (to get the numerator of Equation 4). The same procedure was followed for the measurements with all ASFs studied.

For the cases of SHFs and SQFs, the Scp(EFC) was calculated (using Equations 3 and 4), at four clinically possible locations having the same distance r from the IC, as shown in Figures 2b and c, respectively. The final Scp(EFC) was defined as the average of the four values as shown in Equation (5). This consideration in calculating Scp(EFC) is similar to the diagonal normalized flatness (FDN) and OAR calculations defined in previous studies. 31 , 33 , 34

Scp¯EFC=i=14ScpEFCi4 (5)

The difference (Df%) between Scp(IC) and Scp(EFC) was calculated using Equation (6).

Df%=[ScpEFCScpIC]ScpIC×100% (6)

For each Linac and ASF, at every EFC point located at a distance r to the IC, the measurements of both beam energies were performed, to reduce the uncertainty regarding chamber position.

The uncertainty of the Scp measurements was calculated as the standard deviation (STDEV.P) of the three measurements for the SSFs and SOFs including the respective OAR measurements, and the SD of the four averages of the measurements of the four alternative quadrants for the SHFs and SQFs using Equation (7).

ΔScp=STDEV.P[Scp1,Scp2,,ScpN]N (7)

where N is the number of repeated readings. N = 3 when calculating Scp for SSFs and SOFs, and N = 4 when calculating Scp for SHFs and SQFs.

3. RESULTS

More than 400 measurements were made in the two Linacs to calculate OAR(r) and then Scp at various points located on one horizontal plane parallel to the water phantom surface at a depth of 10 cm. These points represent the EFCs of the ASFs and the IC of the SSF. All measurements and calculations are listed in Tables 1, 2, 3, 4. These tables display the comparison of the individual Scp values calculated for all ASFs (SHFs, SQFs, and SOFs) with their respective GESFs.

TABLE 1.

Total scatter factors (Scp) for symmetric (SSF), half (SHF), quarter (SQF) and off‐isocenter (SOF) fields of 6 MV photon beam generated by Linac 1. r(cm): The distance between IC and EFC. OAR(r): Off‐axis ratio. Ratio 1 = R(ASF)/R(SSF). Ratio 2 = Scp(ASF)/Scp(SSF). The Scp, Scp¯, Df %, and ΔScp were calculated from Equations (3), (5), (6), and (7), respectively.

Jaw setting (cm)
Field size (cm2) Field shape y1 y2 x1 x2 Area (cm2) r (cm) Charge R (nc) Ratio 1 OAR(r) Scp Ratio 2
Scp¯±ΔScp
Df¯±ΔDf%
3 × 3 SSF 1.5 1.5 1.5 1.5 9 0 2.426 1.000 1 0.827 1.000 0.827 ± 0.001 0.0
SHF 0 3 1.5 1.5 9 1.5 2.438 1.005 1.003 0.828 1.002 0.830 ± 0.001 0.4 ± 0.2
3 0 1.5 1.5 9 1.5 2.451 1.010 1.007 0.830 1.003
1.5 1.5 0 3 9 1.5 2.456 1.012 1.005 0.833 1.007
1.5 1.5 3 0 9 1.5 2.435 1.004 1.002 0.828 1.002
SQF 3 0 3 0 9 2.12 2.454 1.012 1.006 0.831 1.006 0.831 ± 0.000 0.4 ± 0.1
3 0 0 3 9 2.12 2.467 1.017 1.012 0.831 1.005
0 3 0 3 9 2.12 2.46 1.014 1.01 0.830 1.004
0 3 3 0 9 2.12 2.442 1.007 1.003 0.830 1.004
SOF −5 8 1.5 1.5 9 6.5 2.49 1.026 1.016 0.835 1.010 0.835 ± 0.001 1.0 ± 0.2
−10 13 1.5 1.5 9 11.5 2.488 1.017 1.01 0.840 1.015 0.840 ± 0.001 1.5 ± 0.2
5 × 5 SSF 2.5 2.5 2.5 2.5 25 0 2.615 1.000 1 0.891 1.000 0.891 ± 0.001 0.0
SHF 0 5 2.5 2.5 25 2.5 2.643 1.011 1.005 0.896 1.006 0.898 ± 0.001 0.8 ± 0.1
5 0 2.5 2.5 25 2.5 2.661 1.018 1.009 0.899 1.009
2.5 2.5 0 5 25 2.5 2.674 1.023 1.013 0.900 1.009
2.5 2.5 5 0 25 2.5 2.643 1.011 1.003 0.898 1.008
SQF 5 0 5 0 25 3.54 2.669 1.021 1.009 0.902 1.012 0.901 ± 0.001 1.1 ± 0.1
5 0 0 5 25 3.54 2.694 1.030 1.018 0.902 1.012
0 5 0 5 25 3.54 2.677 1.024 1.014 0.900 1.010
0 5 5 0 25 3.54 2.65 1.013 1.004 0.900 1.009
SOF −5 10 2.5 2.5 25 7.5 2.706 1.035 1.016 0.908 1.019 0.908 ± 0.001 1.9 ± 0.1
−10 15 2.5 2.5 25 12.5 2.71 1.036 1.007 0.917 1.029 0.917 ± 0.001 2.9 ± 0.2
10 × 10 SSF 5 5 5 5 100 0 2.934 1.000 1 1.000 1.000 1.00 ± 0.001 0.0
SHF 0 10 5 5 100 5 2.997 1.021 1.015 1.006 1.006 1.010 ± 0.0.001 1.0 ± 0.2
10 0 5 5 100 5 3.026 1.031 1.022 1.009 1.009
5 5 0 10 100 5 3.042 1.037 1.026 1.011 1.011
5 5 10 0 100 5 2.993 1.020 1.007 1.013 1.013
SQF 10 0 10 0 100 7.07 3.018 1.029 1.014 1.014 1.014 1.012 ± 0.002 1.2 ± 0.2
10 0 0 10 100 7.07 3.052 1.040 1.033 1.007 1.007
0 10 0 10 100 7.07 3.045 1.038 1.027 1.011 1.011
0 10 10 0 100 7.07 3.009 1.026 1.01 1.015 1.015
SOF −5 15 5 5 100 10 3.03 1.033 1.014 1.018 1.018 1.018 ± 0.001 1.8 ± 0.1
−10 20 5 5 100 15 3.048 1.039 0.994 1.045 1.045 1.045 ± 0.001 4.5 ± 0.1
15 × 15 SSF 7.5 7.5 7.5 7.5 225 0 3.12 1.000 1 1.063 1.000 1.063 ± 0.001 0.0
SHF 0 15 7.5 7.5 225 7.5 3.23 1.035 1.016 1.084 1.019 1.076 ± 0.003 1.2 ± 0.3
15 0 7.5 7.5 225 7.5 3.208 1.037 1.023 1.069 1.005
7.5 7.5 0 15 225 7.5 3.254 1.043 1.034 1.073 1.009
7.5 7.5 15 0 225 7.5 3.192 1.023 1.007 1.080 1.016
SQF 15 0 15 0 225 10.6 3.221 1.032 1.01 1.087 1.022 1.082 ± 0.002 1.7 ± 0.3
15 0 0 15 225 10.6 3.265 1.046 1.033 1.077 1.013
0 15 0 15 225 10.6 3.253 1.043 1.028 1.079 1.014
0 15 15 0 225 10.6 3.199 1.025 1.005 1.085 1.020
SOF −5 20 7.5 7.5 225 12.5 3.233 1.036 1.007 1.094 1.029 1.094 ± 0.001 2.9 ± 0.1
20 × 20 SSF 10 10 10 10 400 0 3.244 1 1 1.106 1.000 1.106 ± 0.001 0.0
SHF 0 20 10 10 400 10 3.35 1.033 1.014 1.126 1.018 1.127 ± 0.002 1.9 ± 0.3
20 0 10 10 400 10 3.369 1.039 1.018 1.128 1.020
10 10 0 20 400 10 3.394 1.046 1.033 1.120 1.013
10 10 20 0 400 10 3.327 1.026 1.001 1.133 1.025
SQF 20 0 20 0 400 14.1 3.361 1.036 1.001 1.144 1.035 1.138 ± 0.003 2.9 ± 0.4
20 0 0 20 400 14.1 3.412 1.052 1.029 1.130 1.022
0 20 0 20 400 14.1 3.401 1.048 1.023 1.133 1.025
0 20 20 0 400 14.1 3.352 1.033 0.998 1.145 1.035

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

TABLE 2.

Total scatter factors (Scp) for symmetric (SSF), half (SHF), quarter (SQF) and off‐isocenter (SOF) fields of 23 MV photon beam generated by Linac 1. r (cm): The distance between IC and EFC. OAR(r): Off‐axis ratio. Ratio 1 = R(ASF)/R(SSF). Ratio 2 = Scp(ASF)/Scp(SSF). The Scp, Scp¯, Df %, and ΔScp were calculated from Equations (3), (5), (6), and (7), respectively.

Jaw setting (cm)
Field size (cm2) Field shape y1 y2 x1 x2 Area (cm2) r(cm) Charge R (nc) Ratio 1 OAR(r) Scp Ratio 2
Scp¯±ΔScp
Df¯±ΔDf%
3 × 3 SSF 1.5 1.5 1.5 1.5 9 0 2.877 1.000 1 0.829 1.000 0.829 ± 0.001 0.0
SHF 0 3 1.5 1.5 9 1.5 2.867 0.997 0.995 0.830 1.002 0.831 ± 0.001 0.3 ± 0.2
3 0 1.5 1.5 9 1.5 2.937 1.021 1.015 0.833 1.006
1.5 1.5 0 3 9 1.5 2.923 1.016 1.014 0.830 1.002
1.5 1.5 3 0 9 1.5 2.887 1.003 1.002 0.830 1.001
SQF 3 0 3 0 9 2.12 2.934 1.020 1.013 0.834 1.007 0.832 ± 0.001 0.4 ± 0.2
3 0 0 3 9 2.12 2.966 1.031 1.025 0.833 1.006
0 3 0 3 9 2.12 2.903 1.009 1.008 0.829 1.001
0 3 3 0 9 2.12 2.88 1.001 0.998 0.831 1.003
SOF −5 8 1.5 1.5 9 6.5 2.974 1.034 1.03 0.832 1.004 0.832 ± 0.001 0.4 ± 0.2
−10 13 1.5 1.5 9 11.5 3.024 1.051 1.038 0.839 1.013 0.839 ± 0.001 1.3 ± 0.2
5 × 5 SSF 2.5 2.5 2.5 2.5 25 0 3.186 1.000 1 0.918 1.000 0.918 ± 0.001 0.0
SHF 0 5 2.5 2.5 25 2.5 3.202 1.005 1.005 0.918 1.000 0.920 ± 0.001 0.3 ± 0.2
5 0 2.5 2.5 25 2.5 3.299 1.035 1.032 0.921 1.003
2.5 2.5 0 5 25 2.5 3.286 1.031 1.027 0.922 1.004
2.5 2.5 5 0 25 2.5 3.235 1.015 1.013 0.920 1.002
SQF 5 0 5 0 25 3.54 3.304 1.037 1.028 0.926 1.009 0.925 ± 0.001 0.8 ± 0.2
5 0 0 5 25 3.54 3.369 1.057 1.046 0.928 1.011
0 5 0 5 25 3.54 3.291 1.033 1.026 0.924 1.007
0 5 5 0 25 3.54 3.254 1.021 1.015 0.923 1.006
SOF −5 10 2.5 2.5 25 7.5 3.307 1.038 1.032 0.923 1.006 0.923 ± 0.001 0.6 ± 0.1
−10 15 2.5 2.5 25 12.5 3.359 1.054 1.038 0.932 1.016 0.932 ± 0.001 1.6 ± 0.1
10 × 10 SSF 5 5 5 5 100 0 3.472 1.000 1 1.000 1.000 1.00 ± 0.001 0.0
SHF 0 10 5 5 100 5 3.565 1.027 1.022 1.005 1.005 1.008 ± 0.002 0.8 ± 0.3
10 0 5 5 100 5 3.648 1.051 1.043 1.007 1.007
5 5 0 10 100 5 3.656 1.053 1.048 1.005 1.005
5 5 10 0 100 5 3.6 1.037 1.022 1.015 1.015
SQF 10 0 10 0 100 7.07 3.633 1.046 1.035 1.011 1.011 1.009 ± 0.001 0.9 ± 0.1
10 0 0 10 100 7.07 3.691 1.063 1.055 1.008 1.008
0 10 0 10 100 7.07 3.637 1.048 1.039 1.008 1.008
0 10 10 0 100 7.07 3.59 1.034 1.025 1.009 1.009
SOF −5 15 5 5 100 10 3.643 1.049 1.039 1.010 1.010 1.010 ± 0.001 1.0 ± 0.1
−10 20 5 5 100 15 3.658 1.054 1.033 1.020 1.020 1.020 ± 0.001 2.0 ± 0.1
15 × 15 SSF 7.5 7.5 7.5 7.5 225 0 3.601 1.000 1 1.037 1.000 1.037 ± 0.001 0.0
SHF 0 15 7.5 7.5 225 7.5 3.736 1.037 1.032 1.043 1.005 1.044 ± 0.001 0.7 ± 0.1
15 0 7.5 7.5 225 7.5 3.806 1.057 1.048 1.046 1.009
7.5 7.5 0 15 225 7.5 3.812 1.059 1.054 1.042 1.004
7.5 7.5 15 0 225 7.5 3.735 1.037 1.029 1.045 1.008
SQF 15 0 15 0 225 10.6 3.813 1.059 1.042 1.054 1.016 1.049 ± 0.002 1.2 ± 0.2
15 0 0 15 225 10.6 3.867 1.074 1.062 1.049 1.011
0 15 0 15 225 10.6 3.807 1.057 1.05 1.044 1.007
0 15 15 0 225 10.6 3.769 1.047 1.033 1.051 1.013
SOF −5 20 7.5 7.5 225 12.5 3.797 1.054 1.038 1.054 1.016 1.054 ± 0.001 1.6 ± 0.1
20 × 20 SSF 10 10 10 10 400 0 3.683 1 1 1.061 1.000 1.061 ± 0.001 0.0
SHF 0 20 10 10 400 10 3.863 1.049 1.039 1.071 1.010 1.073 ± 0.001 1.2 ± 0.1
20 0 10 10 400 10 3.929 1.067 1.054 1.074 1.012
10 10 0 20 400 10 3.945 1.071 1.059 1.073 1.011
10 10 20 0 400 10 3.864 1.049 1.035 1.075 1.014
SQF 20 0 20 0 400 14.1 3.92 1.064 1.039 1.087 1.024 1.082 ± 0.002 2.0 ± 0.2
20 0 0 20 400 14.1 3.965 1.077 1.06 1.077 1.016
0 20 0 20 400 14.1 3.927 1.066 1.049 1.078 1.016
0 20 20 0 400 14.1 3.885 1.055 1.032 1.084 1.022

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

TABLE 3.

Total scatter factors (Scp) for symmetric (SSF), half (SHF), quarter (SQF) and off‐isocenter (SOF) fields of 6 MV photon beam generated by Linac 2. r (cm): The distance between IC and EFC. OAR(r): Off‐axis ratio. Ratio 1 = R(ASF)/R(SSF). Ratio 2 = Scp(ASF)/Scp(SSF). The Scp, Scp¯, Df %, and ΔScp were calculated from Equations (3), (5), (6), and (7), respectively.

Jaw setting (cm)
Field size (cm2) Field shape y1 y2 x1 x2 Area (cm2) r(cm) Charge R (nc) Ratio 1 OAR(r) Scp Ratio 2
Scp¯±ΔScp
Df¯±ΔDf%
3 × 3 SSF 1.5 1.5 1.5 1.5 9 0 2.5 1.000 1 0.828 1.000 0.828 ± 0.001 0.0
SHF 0 3 1.5 1.5 9 1.5 2.529 1.012 1.005 0.833 1.007 0.831 ± 0.001 0.4 ± 0.2
3 0 1.5 1.5 9 1.5 2.529 1.012 1.01 0.829 1.002
1.5 1.5 0 3 9 1.5 2.518 1.007 1.003 0.831 1.004
1.5 1.5 3 0 9 1.5 2.53 1.012 1.008 0.831 1.004
SQF 3 0 3 0 9 2.12 2.551 1.020 1.01 0.836 1.010 0.835 ± 0.001 0.8 ± 0.2
3 0 0 3 9 2.12 2.529 1.012 1.008 0.831 1.004
0 3 0 3 9 2.12 2.538 1.015 1.004 0.837 1.011
0 3 3 0 9 2.12 2.526 1.010 1.003 0.834 1.007
SOF −5 8 1.5 1.5 9 6.5 2.58 1.032 1.017 0.840 1.015 0.840 ± 0.001 1.5  ± 0.2
−10 13 1.5 1.5 9 11.5 2.578 1.031 1.013 0.843 1.018 0.843 ± 0.001 1.8 ± 0.2
5 × 5 SSF 2.5 2.5 2.5 2.5 25 0 2.694 1.000 1 0.892 1.000 0.892 ± 0.001 0.0
SHF 0 5 2.5 2.5 25 2.5 2.726 1.012 1.006 0.897 1.006 0.897 ± 0.001 0.5 ± 0.2
5 0 2.5 2.5 25 2.5 2.739 1.017 1.014 0.894 1.003
2.5 2.5 0 5 25 2.5 2.722 1.010 1.005 0.897 1.005
2.5 2.5 5 0 25 2.5 2.745 1.019 1.012 0.898 1.007
SQF 5 0 5 0 25 3.54 2.755 1.023 1.016 0.898 1.007 0.898 ± 0.001 0.7 ± 0.2
5 0 0 5 25 3.54 2.74 1.017 1.011 0.897 1.006
0 5 0 5 25 3.54 2.747 1.020 1.01 0.90 1 1.010
0 5 5 0 25 3.54 2.73 1.013 1.009 0.896 1.004
SOF −5 10 2.5 2.5 25 7.5 2.788 1.035 1.017 0.908 1.018 0.908 ± 0.001 1.8 ± 0.1
−10 15 2.5 2.5 25 12.5 2.794 1.037 1.009 0.917 1.028 0.917 ± 0.001 2.8 ± 0.2
10 × 10 SSF 5 5 5 5 100 0 3.02 1.000 1 1.000 1.000 1.00 ± 0.001 0.0
SHF 0 10 5 5 100 5 3.087 1.022 1.015 1.007 1.007 1.008 ± 0.000 0.8 ± 0.1
10 0 5 5 100 5 3.111 1.030 1.022 1.008 1.008
5 5 0 10 100 5 3.086 1.022 1.015 1.007 1.007
5 5 10 0 100 5 3.118 1.032 1.023 1.009 1.009
SQF 10 0 10 0 100 7.07 3.141 1.040 1.026 1.014 1.014 1.010 ± 0.001 1.0 ± 0.2
10 0 0 10 100 7.07 3.115 1.031 1.023 1.008 1.008
0 10 0 10 100 7.07 3.098 1.026 1.016 1.010 1.010
0 10 10 0 100 7.07 3.108 1.029 1.02 1.009 1.009
SOF −5 15 5 5 100 10 3.119 1.033 1.016 1.017 1.017 1.017 ± 0.001 1.7 ± 0.1
−10 20 5 5 100 15 3.138 1.039 0.998 1.041 1.041 1.041 ± 0.001 4.1 ± 0.1
15 × 15 SSF 7.5 7.5 7.5 7.5 225 0 3.216 1.000 1 1.065 1.000 1.065 ± 0.001 0.0
SHF 0 15 7.5 7.5 225 7.5 3.327 1.035 1.017 1.083 1.017 1.076 ± 0.003 1.1 ± 0.3
15 0 7.5 7.5 225 7.5 3.305 1.028 1.022 1.071 1.006
7.5 7.5 0 15 225 7.5 3.303 1.027 1.021 1.071 1.006
7.5 7.5 15 0 225 7.5 3.338 1.038 1.024 1.079 1.014
SQF 15 0 15 0 225 10.6 3.351 1.042 1.022 1.086 1.020 1.083 ± 0.002 1.7 ± 0.3
15 0 0 15 225 10.6 3.327 1.035 1.023 1.077 1.011
0 15 0 15 225 10.6 3.342 1.039 1.016 1.089 1.023
0 15 15 0 225 10.6 3.317 1.031 1.016 1.081 1.015
SOF −5 20 7.5 7.5 225 12.5 3.335 1.037 1.009 1.094 1.028 1.094 ± 0.001 2.8 ± 0.1
20 × 20 SSF 10 10 10 10 400 0 3.345 1 1 1.108 1.000 1.108 ± 0.001 0.0
SHF 0 20 10 10 400 10 3.453 1.032 1.016 1.125 1.016 1.125 ± 0.003 1.6 ± 0.3
20 0 10 10 400 10 3.475 1.039 1.025 1.123 1.014
10 10 0 20 400 10 3.451 1.032 1.02 1.120 1.011
10 10 20 0 400 10 3.485 1.042 1.018 1.134 1.023
SQF 20 0 20 0 400 14.1 3.504 1.048 1.013 1.145 1.034 1.140 ± 0.003 2.9 ± 0.3
20 0 0 20 400 14.1 3.481 1.041 1.019 1.131 1.021
0 20 0 20 400 14.1 3.496 1.045 1.013 1.143 1.032
0 20 20 0 400 14.1 3.475 1.039 1.009 1.140 1.030

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

TABLE 4.

Total scatter factors (Scp) for symmetric (SSF), half (SHF), quarter (SQF) and off‐isocenter (SOF) fields of 18MV photon beam generated by Linac 2. r (cm): The distance between IC and EFC. OAR(r): Off‐axis ratio. Ratio 1 = R(ASF)/R(SSF). Ratio 2 = Scp(ASF)/Scp(SSF). The Scp, Scp¯, Df %, and ΔScp were calculated from Equations (3), (5), (6), and (7), respectively.

Jaw setting (cm)
Field size (cm2) Field shape y1 y2 x1 x2 Area (cm2) r(cm) Charge R (nc) Ratio 1 OAR(r) Scp Ratio 2
Scp¯±ΔScp
Df¯±ΔDf%
3 × 3 SSF 1.5 1.5 1.5 1.5 9 0 2.937 1.000 1 0.832 1.000 0.832 ± 0.001 0.0
SHF 0 3 1.5 1.5 9 1.5 2.98 1.015 1.005 0.840 1.010 0.838 ± 0.001 0.6 ± 0.2
3 0 1.5 1.5 9 1.5 2.964 1.009 1.003 0.838 1.006
1.5 1.5 0 3 9 1.5 2.928 0.997 0.996 0.833 1.001
1.5 1.5 3 0 9 1.5 2.995 1.020 1.012 0.839 1.008
SQF 3 0 3 0 9 2.12 3.001 1.022 1.013 0.840 1.009 0.839 ± 0.001 0.8 ± 0.2
3 0 0 3 9 2.12 2.948 1.004 1 0.836 1.004
0 3 0 3 9 2.12 2.971 1.012 1.002 0.840 1.010
0 3 3 0 9 2.12 3.02 1.028 1.019 0.840 1.009
SOF −5 8 1.5 1.5 9 6.5 3.055 1.040 1.03 0.841 1.010 0.841 ± 0.001 1.0 ± 0.2
−10 13 1.5 1.5 9 11.5 3.109 1.059 1.04 0.847 1.018 0.847 ± 0.001 1.8 ± 0.2
5 × 5 SSF 2.5 2.5 2.5 2.5 25 0 3.233 1.000 1 0.916 1.000 0.916 ± 0.001 0.0
SHF 0 5 2.5 2.5 25 2.5 3.308 1.023 1.015 0.924 1.008 0.922 ± 0.001 0.7 ± 0.2
5 0 2.5 2.5 25 2.5 3.285 1.016 1.011 0.921 1.005
2.5 2.5 0 5 25 2.5 3.263 1.009 1.001 0.924 1.008
2.5 2.5 5 0 25 2.5 3.327 1.029 1.024 0.921 1.005
SQF 5 0 5 0 25 3.54 3.36 1.039 1.03 0.925 1.009 0.924 ± 0.001 0.9 ± 0.2
5 0 0 5 25 3.54 3.309 1.024 1.017 0.922 1.006
0 5 0 5 25 3.54 3.317 1.026 1.018 0.924 1.008
0 5 5 0 25 3.54 3.375 1.044 1.032 0.927 1.012
SOF −5 10 2.5 2.5 25 7.5 3.37 1.042 1.033 0.925 1.009 0.925 ± 0.001 0.9 ± 0.1
−10 15 2.5 2.5 25 12.5 3.416 1.057 1.038 0.933 1.018 0.933 ± 0.001 1.8 ± 0.1
10 × 10 SSF 5 5 5 5 100 0 3.528 1.000 1 1.000 1.000 1.00 ± 0.001 0.0
SHF 0 10 5 5 100 5 3.645 1.033 1.027 1.006 1.006 1.006 ± 0.001 0.6 ± 0.2
10 0 5 5 100 5 3.64 1.032 1.026 1.006 1.006
5 5 0 10 100 5 3.607 1.022 1.02 1.002 1.002
5 5 10 0 100 5 3.685 1.045 1.036 1.008 1.008
SQF 10 0 10 0 100 7.07 3.692 1.046 1.038 1.008 1.008 1.008 ± 0.001 0.8 ± 0.2
10 0 0 10 100 7.07 3.647 1.034 1.028 1.006 1.006
0 10 0 10 100 7.07 3.652 1.035 1.028 1.007 1.007
0 10 10 0 100 7.07 3.7 1.049 1.038 1.010 1.010
SOF −5 15 5 5 100 10 3.713 1.052 1.042 1.010 1.010 1.010 ± 0.001 1.0 ± 0.1
−10 20 5 5 100 15 3.726 1.056 1.034 1.021 1.021 1.021 ± 0.001 2.1 ± 0.1
15 × 15 SSF 7.5 7.5 7.5 7.5 225 0 3.655 1.000 1 1.036 1.000 1.036 ± 0.001 0.0
SHF 0 15 7.5 7.5 225 7.5 3.809 1.042 1.033 1.045 1.009 1.045 ± 0.001 0.9 ± 0.1
15 0 7.5 7.5 225 7.5 3.811 1.043 1.032 1.047 1.010
7.5 7.5 0 15 225 7.5 3.778 1.034 1.027 1.043 1.006
7.5 7.5 15 0 225 7.5 3.846 1.052 1.041 1.047 1.011
SQF 15 0 15 0 225 10.6 3.875 1.060 1.044 1.052 1.016 1.049 ± 0.001 1.3 ± 0.2
15 0 0 15 225 10.6 3.826 1.047 1.038 1.045 1.008
0 15 0 15 225 10.6 3.841 1.051 1.039 1.048 1.011
0 15 15 0 225 10.6 3.88 1.062 1.046 1.051 1.015
SOF −5 20 7.5 7.5 225 12.5 3.861 1.056 1.038 1.054 1.018 1.054 ± 0.001 1.8  ± 0.1
20 × 20 SSF 10 10 10 10 400 0 3.737 1 1 1.059 1.000 1.059 ± 0.001 0.00
SHF 0 20 10 10 400 10 3.933 1.052 1.042 1.070 1.010 1.074 ± 0.001 1.4 ± 0.2
20 0 10 10 400 10 3.941 1.055 1.037 1.077 1.017
10 10 0 20 400 10 3.905 1.045 1.033 1.071 1.012
10 10 20 0 400 10 3.965 1.061 1.045 1.075 1.015
SQF 20 0 20 0 400 14.1 3.988 1.067 1.041 1.086 1.025 1.082 ± 0.002 2.2 ± 0.2
20 0 0 20 400 14.1 3.951 1.057 1.039 1.078 1.018
0 20 0 20 400 14.1 3.958 1.059 1.039 1.080 1.019
0 20 20 0 400 14.1 3.997 1.070 1.044 1.085 1.024

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

Based on these results, there are no significant differences in Scp values calculated for two 6 MV energies for L1 and L2, as illustrated in Figure 4. The same was true for 18 and 23 MV, except probably for the field 3 × 3 cm2 off‐isocenter fields, where the difference was about 1%, as shown in Figure 5. This confirms the accuracy of the calculation methodology followed and its compatibility for both accelerators, so that the average Scp values can be adopted for the two 6 MV beams and respectively for the 18/23 MV x‐ray beam energies.

FIGURE 4.

FIGURE 4

Scp differences Df % between the L1 and the L2 for the 6 MV energy for all studied fields. Scp, scatter factor.

FIGURE 5.

FIGURE 5

Scp difference Df % between the beam energies of 18 MV (L1) and 23 MV (L2), for all studied fields. Scp, scatter factor.

Tables 1, 2, 3, 4 show that the uncertainty associated with the Scp calculation is small with maximum error being 0.3% and 0.2% for energies 6 and 18/23 MV, respectively. It was noted from Figures 4 and 5 that the maximum uncertainty when studying the difference in the Scp between the two accelerators (considering respective TF comparisons) reached 0.5% and 0.3% for 6 and 18/23 MV, respectively. On the other hand, it was found that the estimated uncertainty in calculating the difference in Scp between ASFs and SSFs is less than 1% and ranges from 0.1% to 0.3% and from 0.1% to 0.2%, as shown in Table 5 and Figure 6, for energies 6 and 18/23 MV respectively.

TABLE 5.

The Scp differences between ASFs (SHFs, SHFs and SOFs) and their equivalent fields SSFs or GESFs for the 6 and 18/23 MV photon energies. Df¯%: Average Df % values for every two equal energies from Tables 1, 2, 3, 4. The superscripts in parenthesis for the 10×10 fields (6 MV) indicate the distance r between the IC and the EFC. With bold are denoted differences from 1% to 2% and with bold italics differences above 2%.

Df¯±ΔDf%: ASFs & SSFs (GESF)for 6 MV Df¯±ΔDf%: ASFs & SSFs (GESF) for 18/23 MV
SOFs SOFs
Field size (cm2) SHFs SQFs −5 cm −10 cm SHFs SQFs −5 cm −10 cm
3 × 3 0.4 ± 0.2 0.7 ± 0.2 1.2 ± 0.2 1.7 ± 0.2 0.4 ± 02 0.6 ± 0.2 0.7 ± 0.2 1.5 ± 0.2
5 × 5 0.7 ± 0.1 0.9 ± 0.2 1.8 ± 0.1 2.8  ±  0.2 0.4 ± 0.2 0.8 ± 0.2 0.7 ± 0.1 1.7 ± 0.1
10 × 10 0.9(5 cm) ± 0.2 1.1 (7.1 cm)  ± 0.2 1.7 (10 cm)  ± 0.1 4.3 (15 cm)  ±  0.1 0.7 ± 0.2 0.8 ± 0.1 1.0 ± 0.1 2.1  ±  0.1
15 × 15 1.1 ± 0.3 1.7 ± 0.3 2.8  ±  0.1 0.8 ± 0.1 1.2 ± 0.2 1.7 ± 0.1
20 × 20 1.8 ± 0.3 2.9 ± 0.3 1.3 ± 0.2 2.1 ± 0.2

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; GESF, geometric equivalent square field; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

FIGURE 6.

FIGURE 6

The average Scp differences (Df%) between ASFs and GESF for the 6 MV and 18/23 MV photon beam energies. (a) SHF. (b) SQF. (c) and (d) SOF with y1 = −5 and y1 = −10 cm, respectively. ASFs, asymmetric square fields; GESF, geometric equivalent square field; Scp, scatter factor; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field.

Table 5 shows the Scp differences between ASFs and their GESFs for the 6 and 18/23 MV photon energies, which are also represented graphically in Figure 6. From Table 5 and Figure 6 it becomes evident that the Scp differences between ASF and their respective GESF, increase gradually with the increasing field size (i.e., with increasing Table's 5 column number). Also, the more asymmetric the field is (SHF → SQF → SOF) the larger the more the increase is (i.e., with increasing Table's 5 row number), as the more asymmetric the field is the larger the distance r of its geometric center (EFC) to the IC is. However, these differences reduce with increasing beam energy. Consequently, there are two factors that may affect the increase of Scp(EFC), and it can be either both or one of them. The first factor is the method of the jaw setting (SHF, SQF or SOF) and the second factor is the off‐axis distance (r).

The effect of the method of jaw setting can be investigated by comparing directly the ratio of the measured charge values at the center of two different ASF shapes (e.g., the SOF with dimensions 15 × 15 cm2 and 5 × 5 cm2) which have the same EFC (e.g., R(15×15cm2,r=12.5cm)R(5×5cm2,r=12.5cm)) with the measured charge ratio of their GESFs in the IC (e.g.,R(15×15cm2,r=0cm)R(5×5cm2,r=0cm)). Table 6 shows that the two ratios are almost equal at all studied energies and the difference between them is less than 0.05%. Since the OAR(r) used to correct the charge measurements is the same for all ASFs having the EFC at same distance r, the Scp(EFC) is independent of the asymmetric field size or the location of the field margin from the IC for these studied cases. Therefore, this implies that the main factor contributing to the Scp increase is the off‐axis distance (r) of the EFC. Indeed, in Figure 7 are shown the Scp differences between ASFs and GESFs (expressed as their ratio) for the 6 and 18/23 MV with relation to the EFC's r values. This figure shows that the average Scp ratios increase gradually with r and that the average values of 6 MV data, are in almost all cases larger than the respective 18/23 MV data. For 18/23 MV, the differences are within or close to 2% but for 6 MV, the differences exceed 4%.

TABLE 6.

The field‐size effect test on Scp (EFC). The results were extracted from Tables 1, 2, 3, 4.

Asymmetric jaw setting (cm) Ratio (Asym1/Asym2; at same EFC and different shape): Ratio (Sym1. GESF/Sym2.GESF; at IC) [Df %]
Field size (cm2) y1 y2 x1 x2 r 6 MV (L1) 6 MV (L2) 23 MV (L1) 18 MV (L2)
20 × 20 0 20 10 10 10 1.1056:1.1057 1.1071:1.1076 1.0604:1.0608 1.0593:1.0592
10 × 10 5 15 5 5 10 [−0.0043] [−0.0479] [−0.0360] [0.0010]
15 × 15 0 15 7.5 7.5 7.5 1.1936:1.1931 1.1933:1.1938 1.1297:1.1303 1.1303:1.1305
5 × 5 5 10 2.5 2.5 7.5 [0.0442] [−0.0365] [−0.0471] [−0.0232]
15 × 15 5 20 7.5 7.5 12.5 1.1930:1.1931 1.1936:1.1938 1.1304:1.1303 1.1303:1.1305
5 × 5 10 15 2.5 2.5 12.5 [−0.0107] [−0.0113] [0.0123] [−0.0230]

Abbreviations: ASF, asymmetric square fields; EFC, effective field center; GESF, geometric equivalent square field; IC, isocenter; SHF, square half‐field; SOF, square off‐isocenter field; SQF, square quarter‐field; SSF, symmetric square field.

FIGURE 7.

FIGURE 7

Average Scp difference between ASFs and GESFs for the 6 and 18/23 MV energies. ASFs, asymmetric square fields; GESF, geometric equivalent square field; Scp, scatter factor.

4. DISCUSSION

The flattening filter (FF) has a Gaussian‐like shape, and as a result the beam energy of the central ray passing through IC is relatively larger than the rays away from the IC, and thus more penetrating. As a result, OAR(r) profiles at small depths are not flat around the central axis and present their maximum values at a distance from the IC, depending on the field size (as shown in Figure 8). 16 Daryoush et al. 35 confirmed that the 18 MV energy photons scatter preferentially in the forward direction, while the 6 MV energy photons scatter in the forward and sides direction with a relative predominance in the forward direction of 0.5%. On the other hand, the Scp increases with decreasing energy for larger field size due to the increasing phantom scatter contribution (Sp). In asymmetric fields, the EFC shifts to the sides and thus the lateral scatter increases with increasing r due to decreasing energy, 16 as long as the measurement is performed away from the beam edges. This agrees with the results shown in Table 5, Figure 6 and Figure 7.

FIGURE 8.

FIGURE 8

OAR(r) profiles at dmax for 23 MV. OAR, off‐axis ratio.

Khan et al. 13 and John et al. 2 measurements confirmed that calculating the Scp based on the corrections using the OAR(r) 40 × 40 cm2 field, leads to errors in the output factors greater than 5% when r > 10 cm. In this research, the errors were less, especially at the energy 18 MV, even though the distance r was up to 15 cm. Possibly, this is because in the previous studies a homogeneous field in all directions was considered (i.e., no difference between left and right, or up and down field halves), and OAR(r) values were derived from the cross‐beam or diagonal profile only, regardless of the direction of the EFC.

Das et al. 36 who studied the applicability of the GESF concept for rectangular fields ranging from 0.5 × 0.5 cm2 to 5 × 5 cm2, exhibited that differences in field output factors exist between linacs from two different vendors, because of the different design of the collimators. Such differences were observed in our study when comparing Scp values between two Varian Linacs, especially for 3 × 3 cm2 fields (see Figures 4 and 5). However, in the Das et al. 36 study, no asymmetric fields were included.

In Figure 6c there is a jump in Df% for the 5 × 5 cm2 field size in the case of SOF (y1 = −5 cm) for the 6 MV beams only, which was not observed in Figures 6a,b,d. Since this jump was observed in both Linacs, it may be assumed that is attributed to the design of the FF at the measurement point.

Jackson et al. 23 confirmed that the differences in MU between manual and TPS calculations for breast radiotherapy increase when these calculations require OAR(r) corrections, and compared 6 MV and Cobalt beams (the OAR correction is neglected for Co‐60) under similar conditions while neglecting influencing factors such as tissue heterogeneities in both calculations. Also, computerized investigations proved that these differences are small at the IC and large away from it under the same conditions. 3 , 24 , 37 In addition, Ian et al. 25 , 26 proved, after correcting the missing tissue in breast, that these differences increase with the decreasing energy of the photon beam.

Part of the remaining unresolved difference in MUs can be explained by the fact that the equivalent field in the TPS is defined as the symmetric field in which the PDD of its central axis has the same properties as the PDD in the effective axis of the asymmetric field, and this also applies to the Scp. This means that the effect of the OAR(r) is included in the TPS calculations, whereas in manual calculations the OAR(r) effect is only corrected for PDD, TPR or TMR and is not taken into account for correcting the output factor or total Scp which are computed directly from the GESF.

In breast radiotherapy with a single‐isocenter technique, after correction of the missing tissue, the GESF for the half‐ or quarter‐ fields does not exceed 10 × 10 cm2 and also (r) does not exceed 7 cm in most clinical cases, 25 , 26 , 38 so the difference is within 1% (Table 5). This also applies to cases of the brain and head & neck radiotherapy. With respect to craniospinal radiation, the width of the lower spine half‐field is proportional to cover the vertebral bodies only, and therefore r and GESF are within the acceptable limits. For more complex clinical cases when radiotherapy is without half‐beam blocking where the EFC can be 10 cm away from the IC and the GESF to the supraclavicular field (off‐isocenter field) is up to 10 × 10 cm2, 9 the agreement is within 2% (Table 5).

In this paper, we have calculated the differences in total Scp between different ASFs and their GESFs, and the results can be used for manual MU check calculations of three‐dimensional conformal radiotherapy (3D‐CRT) plans to improve their accuracy. Based on these findings, it is preferable in treatment plans to have the EFC as close as possible to IC, to keep the difference Df% as low as possible, especially for the breast radiotherapy when chest wall is longer than 20 cm (single‐isocenter technique without the use of half beam blocks), 9 as in the case previously shown in Figure 1c.

5. CONCLUSION

The results of this study suggest that in cases where the IC is partially or completely blocked, the Scp(EFC) is dependent on the distance of the measuring point (i.e., the EFC) from the IC but independent of the method of jaw setting (half‐, quarter‐ and off‐isocenter fields). For all measurements, the Scp(EFC) was larger than Scp(IC). This increase was up to 2% and 4% for energies 23/18 and 6 MV respectively, but generally it was within 2% for most clinical cases. In this study we have provided detailed information about Scp(EFC) measurements and calculations which are tabulated in Tables 1, 2, 3, 4 and can be utilized as correction factors to improve the accuracy of quick, manual verifications of MU in the 3D‐CRT treatment plans, which is still used in some developing countries.

AUTHOR CONTRIBUTIONS

Mohammad Samir Hmodi, Majeda Nahili and Ousamah Anjak conceived this project. Mohammad Samir Hmodi and Ali Hasan designed the experiments and performed the measurements. Mohammad Samir Hmodi and Karlos Shamout wrote the draft manuscript. Majeda Nahili, Ioannis A Tsalafoutas and Mohammad Hmodi analyzed the data, interpreted the results, and revised the manuscript. All authors have approved the manuscript's final version.

CONFLICT OF INTEREST STATEMENT

The authors have no conflict of interest to declare.

ACKNOWLEDGMENTS

This study was supported by Damascus University and Tishreen University Hospital (no funding). Open Access funding was provided by the Qatar National Library.

Hmodi MS, Nahili M, Tsalafoutas IA, et al. Validation of the geometric equivalent field concept in total scatter factor calculations, for half‐, quarter‐ and off‐isocenter asymmetric square fields. J Appl Clin Med Phys. 2023;24:e14103. 10.1002/acm2.14103

REFERENCES

  • 1. Sanz D, Nelli F. The separation of the head and phantom scatter components from a phase space description. Phys Med Biol. 2004;21(18):4259‐4275. [DOI] [PubMed] [Google Scholar]
  • 2. Gibbons JP, Antolak JA, Followill DS, et al. Monitor unit calculations for external photon and electron beams: report of the AAPM Therapy Physics Committee Task Group No. 71. Med Phys. 2014;41(3):031501. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Mayivaganan A, Athiyaman H, Chougule A, Kumar HS. Verification of monitor unit calculations for eclipse treatment planning system by in‐house developed spreadsheet. Iran J Med Phys. 2018;15(2):87‐100. [Google Scholar]
  • 4. Kagkiouzis J, Platoni K, Kantzou I, et al. Review of the three‐field techniques in breast cancer radiotherapy. JBUON. 2017;22(3):599‐605. [PubMed] [Google Scholar]
  • 5. Saw CB, Krishna KW, Enke CA, Hussey DH. Dosimetric evaluation of abutted fields using asymmetric collimators for treatment of head and neck. Int J Radiat Oncol Biol Phys. 2000;47(3):821‐824. [DOI] [PubMed] [Google Scholar]
  • 6. Kathpal M, Rihan D. Use of a noncoplanar half‐beam block on the lower spinal field to decrease the maximum bowel and cumulative dose in craniospinal irradiation. Int J Radiat Oncol Biol Phys. 2013;84(3):S290. [Google Scholar]
  • 7. Goswami B, Jain RK, Yadav S, et al. Dosimetric comparison of integral dose for different techniques of craniospinal irradiation. J Radiother Pract. 2000;20(3):345‐350. [Google Scholar]
  • 8. Romeo N. A new isocentric technique for exact geometric matching in the radiotherapy of the breast and ipsilateral supraclavicular fossa using dual asymmetric jaws. Physica Med. 2012;28(4):281‐287. [DOI] [PubMed] [Google Scholar]
  • 9. Zhang T, Dilworth JT, Marina O, Chen P, Benedetti L, Liu Q. A three‐field monoisocentric inverse breast treatment planning technique without half‐beam blocking. J Appl Clin Med Phys. 2015;16(5):246‐258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Varatharaj C, Ravikumar M, Supe S, Sathiyan S, Ganesh K, Arunkumar T. Dosimetric investigation of dual energy photon beams with asymmetric collimator jaws. Pol J Med Phys Eng. 2008;14(1):33‐45. [Google Scholar]
  • 11. Salk J. A simple formalism for calculation and verification of dose in asymmetric x‐ray fields. Department of Radiotherapy, University of Ulm, Germany. Accessed July 22, 2023. https://citeseerx.ist.psu.edu/doc/10.1.1.517.1295 [Google Scholar]
  • 12. Shin R, Li A, Chu H, Hsu W. Calculation of head scatter factors at isocenter or at center of field for any arbitrary jaw setting. Med Phys. 1999;26(4):506‐511. [DOI] [PubMed] [Google Scholar]
  • 13. Khan F, Gerbi B, Deibel F. Dosimetry of asymmetric x‐ray collimators. Med Phys. 1986;13(6):936‐941. [DOI] [PubMed] [Google Scholar]
  • 14. Cadman P. A dosimetric investigation of scatter conditions for dual asymmetric collimators in open fields. Med Phys. 1995;22(4):457‐463. [DOI] [PubMed] [Google Scholar]
  • 15. Millin AE, Smith CW. A beam profile generation algorithm for wedged half‐beam blocked asymmetric fields. Phys Med Biol;39(1):63‐73. [DOI] [PubMed] [Google Scholar]
  • 16. Tenhunen M, Lahtinen T. Relative output factors of asymmetric megavoltage beams. Radiother Oncol. 1994;32(3):226‐231. [DOI] [PubMed] [Google Scholar]
  • 17. Araki F, Ikeda R, Moribe N, et al. Dose calculation for asymmetric photon fields with independent jaws and multi‐leaf collimators. Med Phys. 2000;27(2):340‐345. [DOI] [PubMed] [Google Scholar]
  • 18. Jin HY, Zhou LH, Wen Lv Q, Chen CM, Chen GJ. A simplified method for the calculation of equivalent squares of irregular photon fields. Conf Proc IEEE Eng Med Biol Soc. 2005:2005:7091‐7094. 10.1109/IEMBS.2005.1616140 [DOI] [PubMed] [Google Scholar]
  • 19. McCurdy B, Pistorius P. Determination of equivalent photon fields through integrated 1D convolution kernels. Phys Med Biol. 1999;44(12):2971‐2985. [DOI] [PubMed] [Google Scholar]
  • 20. Vadash P, Bjärngard B. An equivalent‐square formula for head scatter factors. Med Phys. 1993;20(3):733‐734. [DOI] [PubMed] [Google Scholar]
  • 21. Tahmasebi Birgani MJ, Chegeni N, Zabihzadeh M, Hamzian N. An analytical method to calculate equivalent fields to irregular symmetric and asymmetric photon fields. Med Dosim. 2014;39(1):54‐59. [DOI] [PubMed] [Google Scholar]
  • 22. Sterling T, Perry H, Katz L. Automation of radiation treatment planning. Br J Radiol. 1964;37:544‐550. [DOI] [PubMed] [Google Scholar]
  • 23. Chan J, Russell D, Peters VG, Farrell TJ. Comparison of monitor unit calculations performed with a 3D computerized planning system and independent ‘‘hand’’ calculations: results of three years' clinical experience. Appl Clin Med Phys. 2002;3(4):293‐301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Slassi N, Ouabi H, El Khayati N. Comparison of an in‐house developed monitor unit double‐check program for 3D conformal radiation therapy and treatment planning system verification. J Radiother Pract. 2019;18(3):1‐11. [Google Scholar]
  • 25. Kay I, Dunscombe P. Verifying monitor unit calculations for tangential breast fields. J Appl Clin Med Phys. 2006;7(2):50‐57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Kay I, Meyer T. Verifying monitor unit calculations for tangential whole breast fields in three‐dimensional planning. J Appl Clin Med Phys. 2008;9(1):47‐53. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Nailon WH, Welsh D, McDonald K, et al. EPID‐based in vivo dosimetry using Dosimetry Check™: overview and clinical experience in a 5‐yr study including breast, lung, prostate, and head and neck cancer patients. J Appl Clin Med Phys. 2019;20(1):6‐16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Cho W, Kielar KN, Mok E, et al. Multisource modeling of flattening filter free (FFF) beam and the optimization of model parameters. Med Phys. 2011;38(4):1931‐1942. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Vargas‐Verdesoto MX, Álvarez‐Romero JT. Determination and verification of a 2D pencil‐beam kernel for a radiosurgery system with cones. Med Dosim. 2013;38(2):215‐220. [DOI] [PubMed] [Google Scholar]
  • 30. Healy B, Murry R. Testing of a treatment planning system with beam data from IAEA TECDOC 1540. J Med Phys. 2011;36(2):107‐110. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Gao S, Balter PA, Rose M, Simon WE. A comparison of methods for monitoring photon beam energy constancy. J Appl Clin Med Phys. 2016;17(6):242‐253. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Gibbons J. A system of dosimetric calculations. Khan's The Physics of Radiation Therapy. 6th ed. Wolters Kluwer Health; 2020:344‐345. [Google Scholar]
  • 33. Goodall S, Harding N, Simpson J, Alexander L, Morgan S. Clinical implementation of photon beam flatness measurements to verify beam quality. J Appl Clin Med Phys. 2015;16(6):340‐345. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Gao S, Balter PA, Rose M, Simon WE. Measurement of changes in linear accelerator photon energy through flatness variation using an ion chamber array. Med Phys. 2013;40(4):042101. [DOI] [PubMed] [Google Scholar]
  • 35. Sheikh‐Bagheri D, Rogers DW. Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code. Med Phys. 2002;29(3):391‐402. [DOI] [PubMed] [Google Scholar]
  • 36. Das IJ, Dogan SK, Gopalakrishnan M, Ding GX, Longo M, Franscescon P. Validity of equivalent square field concept in small field dosimetry. Med Phys. 2022;49(6):4043‐4055. [DOI] [PubMed] [Google Scholar]
  • 37. Sellakumar P, Arun C, Sanjay S, Ramesh S, et al. Comparison of monitor units calculated by radiotherapy treatment planning system and an independent monitor unit verification software. Physica Med. 2011;27(1):21‐29. [DOI] [PubMed] [Google Scholar]
  • 38. Jinno S, Tachibana H, Moriya S, et al. A multi‐institutional study of independent calculation verification in inhomogeneous media using a simple and effective method of heterogeneity correction integrated with the Clarkson method. J Radiat Res (Tokyo). 2018;59(4):490‐500. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley

RESOURCES