A Potential Pitfall and Clinical Solutions in Surface-Guided Deep Inspiration Breath Hold Radiation Therapy for Left-Sided Breast Cancer

Chuan Zeng a,1, Qiyong Fan a,1, Xiang Li a, Yulin Song a, Licheng Kuo a, Michalis Aristophanous a, Laura I Cervino a, Linda Hong a, Simon Powell b, Guang Li a,
PMCID: PMC10692299  PMID: 38047221

Abstract

Purpose

Deep inspiration breath hold (DIBH) is an effective technique to spare the heart in treating left-sided breast cancer. Surface-guided radiation therapy (SGRT) is increasingly applied in DIBH setup and motion monitoring. Patient-specific breathing behavior, either thoracically driven or abdominally driven (A-DIBH), should be unaltered, online identified, and monitored accordingly to ensure reproducible heart-sparing treatment.

Methods and Materials

Sixty patients with left-sided breast cancer treated with SGRT were analyzed: 20 A-DIBH patients with vertical chest elevation (VCE ≤ 5 mm) were prospectively identified, and 40 control patients were retrospectively and randomly selected for comparison. At simulation, both free-breathing (FB) and DIBH computed tomography (CT) were acquired, guided by a motion surrogate placed around the xiphoid process. For SGRT treatment setups, the region of interest (ROI) was defined on the CT chest surface, and the surrogate-based setup was a backup. For all 60 patients, the VCE was measured as the average of the FB-to-DIBH elevations at the breast and xiphoid process, together with abdominal elevation. In the 40-patient control group, A-DIBH patients (VCE ≤ 5 mm) were identified. Of the 20 A-DIBH patients, 10 were treated with volumetric modulated arc therapy plans, and 10 patients were treated with tangent plans. Clinical DIBH plans were recalculated on FB CT to compare maximum dose (DMax), 5% of the maximum dose (D5%), mean dose (DMean), and V30Gy, V20Gy, and V5Gy of the heart and lungs and their significance.

Results

In the 20 A-DIBH patients, VCE = 3 ± 2 mm, surrogate motion (9 ± 6 mm), and abdomen motion of 14 ± 5 mm are found. Heart dose reduction from FB to DIBH is significant (P < .01): ∆DMax = –8.4 ± 9.8 Gy, ∆D5% = –2.4 ± 4.4 Gy, and ∆DMean = –0.6 ± 0.9 Gy. Six out of 40 control patients (15%) are found to have VCE ≤ 5 mm.

Conclusions

A-DIBH (VCE ≤ 5 mm) patient population is significant (15%), and they should be identified in the SGRT workflow and monitored accordingly. A new abdominal ROI or an abdominal surrogate should be used instead of the conventional chest-only ROI. Patient-specific DIBH should be preserved for higher reproducibility to ensure heart sparing.

Introduction

Deep inspiration breath hold (DIBH) has been used for radiation therapy of patients with left-sided breast cancer for about 2 decades to spare the heart as vertical chest wall elevation (VCE) at DIBH separates the heart from the chest wall in the supine position due to gravity and/or inferior shift due to diaphragmatic pull. Clinical data have shown that DIBH is very effective in reducing heart toxicity from breast irradiation,1,2 even for some patients who may not reproduce DIBH completely at treatment.2,3 Alternatively, the prone treatment can be used to spare the heart and lungs, but it is difficult to cover local lymph nodes, so it may not be suitable for treating patients with nodal involvement.2 Initially, a surrogate-guided technique was developed using a surrogate box placed near the xiphoid process to monitor the chest elevation, such as a real-time positioning management (RPM, Varian Medical Systems, Palo Alto, CA) system or its newer version, respiratory gating for scanners (RGSC). However, the variation of RPM or RGSC box placement throughout a multifractional radiation therapy treatment course introduces uncertainty in monitoring DIBH levels. To eliminate the surrogate placement uncertainty, surface-guided radiation therapy (SGRT) has been developed and implemented in the clinic for left-breast DIBH treatment.4, 5, 6 Although the region of interest (ROI) defined for breast SGRT setup and motion monitoring may vary, the chest wall surface has been used to achieve DIBH for treatment.7,8 A popular ROI includes the breast and surrounding area with a unique landscape,8,9 namely a chest-only ROI. Furthermore, the SGRT technique allows aligning of the ipsilateral arm and chin first at free-breathing (FB) before DIBH, minimizing the breast deformation for partial breast irradiation and reproducing the position of the axillary and supraclavicular lymph nodes for patients with locally advanced breast cancer.8,10 Moreover, recent advances in SGRT have promised to remove the need for tattoos in conventional patient setups using room lasers for patient alignment, including left-sided DIBH breast treatment.11, 12, 13 Therefore, the SGRT technique has been increasingly applied in DIBH radiation therapy for left-sided breast cancer.

To ensure a reproducible DIBH, offline patient training can be applied, facilitating DIBH simulation and treatment for better compliance and reproducibility,9,14 and online visual coaching seems necessary so that patients can consistently control the extent of DIBH with guidance.4,15 Thoracic-driven DIBH (T-DIBH) and abdominal-driven DIBH (A-DIBH) have been recently studied and brought to the attention of clinicians that both offer heart sparing.16, 17, 18 Because the training/coaching approach will alter some patients’ native breathing behavior, training must be reinforced, and patients’ participation via visual coaching is required to ensure consistent T-DIBH or A-DIBH from simulation to the entire multifractional treatment. Although this is a widely studied and adopted method, it has its shortcomings, including higher clinical workload and the likelihood of breathing pattern changes,15 namely falling back to the natural behavior. In addition, changing patients’ native breathing behavior may be a challenge for some patients, as a recent study showed that 20 of 120 patients (17%) cannot distinguish between T-DIBH and A-DIBH at simulation,18 so training such patients may not be feasible. Furthermore, when an ROI is drawn on the chest surface only as the default ROI, it implies that the patients are going through T-DIBH predominantly using the intercoastal muscles with sufficient chest elevation. This default ROI may not apply to monitor A-DIBH as it does not cover the abdominal surface.

Alternatively, the actual patient-specific breathing behavior can be preserved in the DIBH treatment by identifying a patient as thoracic-breathing or abdominal-breathing early in the clinical workflow (the simulation-planning-treatment process) using the VCE measure, as well as providing a simple, adaptive DIBH monitoring strategy to T-DIBH or A-DIBH patients. As a patient's natural breathing behavior is more reproducible, it would lead to a more consistent treatment outcome with potentially less demand for daily maintenance. This rationale forms the hypothesis of the study to pursue and test, aiming at a feasible alternative clinical approach.

In this study, we prospectively identified 20 patients with left-sided breast cancer with A-DIBH, namely VCE ≤ 5 mm between FB and DIBH surfaces, measured the heart separation from the chest wall, compared the difference between FB and DIBH, and provided feasible solutions to monitor patients with A-DIBH properly using an abdominal ROI or the RPM/RGSC inferior to the xiphoid process. To demonstrate the worst-case-scenario dosimetric consequence for incorrect A-DIBH motion monitoring, the clinical DIBH plans were recalculated on FB computed tomography (CT) for comparison. Finally, to assess the scale of this clinical issue, we retrospectively and randomly selected 40 DIBH patients with left-breast cancer and quantified the percentage of abdominal-breathing patients (VCE ≤ 5 mm with A-DIBH), illustrating the potential clinical effect .

Methods and Materials

Selection of A-DIBH and control DIBH patients from SGRT treatments

In this study, we identified 20 DIBH patients with left-sided breast cancer with A-DIBH or VCE ≤ 5 mm between FB and DIBH CT surfaces prospectively in the clinical planning and plan checking process, raising the concern about the inability of the chest-only ROI to distinguish such patients’ DIBH from FB owing to limited VCE. Clinically, planners and plan checkers were called upon to report any patients with breast cancer for DIBH SGRT with VCE ≤ 5 mm and changed the DIBH monitoring method from AlignRT (v6.2/v6.3, VisionRT, London, UK) to RPM/RGSC (Varian Medical Systems, Palo Alto, CA). Among the 20 eligible patients, 10 were treated with volumetric modulated arc therapy (VMAT) for locally advanced breast cancer, and the other 10 were treated with opposite tangent fields. Nine of the 10 tangent plans were 2 fields treating early stage cancer, and 1 was 4 fields (opposite tangents plus supraclavicular field and posterior axillary boost field) treating locally advanced disease. These patients were analyzed both anatomically and dosimetrically to provide clinically feasible solutions.

To assess the scope of the clinical issue of A-DIBH, another 40 patients with left-sided breast cancer were randomly selected from our AlignRT-guided DIBH database as a control patient group. The body surface elevations at the chest and abdomen were measured to identify the A-DIBH patients, compare the difference in heart position at the anterior and lateral edges between FB and DIBH, and assess the distribution of VCE among the patients. The motion of the surrogate placed near the xiphoid process was assessed by locating the optical reflectors on the surrogate box (n = 2 for RPM and n = 4 for RGSC), measuring their vertical motions, and averaging them to represent the surrogate motion. The difference between the SGRT and surrogate-guided DIBH-monitoring techniques was analyzed.

Measurements of external surface elevation based on FB and DIBH simulation CTs

For each of the 60 DIBH patients, 3 sets of external surface elevation were measured based on the differences between FB and DIBH simulation CT scans at the center of the breast, the xiphoid process, and the abdomen (the maximal elevation point inferior to the xiphoid process), as shown in Fig. 1. The average of the first 2 measurements (at the breast and xiphoid process) was used to estimate the VCE value within the ROI, which should be above the average ROI motion as the upper chest usually moved less. The maximal abdominal elevation was measured on the CT image sets, usually near the inferior border of the field of view (FOV), which was about 10 to 20 cm inferior to the xiphoid process.

Figure 1.

Figure 1

Illustration of measurement of vertical body surface elevation at (A) the breast, (B) the xiphoid process, and (C) the abdomen. The default chest-only ROI (solid yellow rectangle) for surface-guided radiation therapy setup in free-breathing and thoracic-driven deep inspiration breath hold (DIBH) motion monitoring, and an alternative abdomen ROI (dotted yellow triangle) for abdominal-driven DIBH motion monitoring are shown. The vertical chest elevation of the ROI is estimated as the average of those at points A and B. The respiratory gating for scanners (length = 7.4 cm) is visible in the DIBH computed tomography (light gray) with superimposed free-breathing body contour (green). Abbreviations: DIBH = deep inspiration breath hold; ROI = region of interest.

Additionally, the surrogate motion (RPM or RGSC), their placement location, and the default gating upper and lower thresholds were measured. The vertical motion of the RPM or RGSC box (Fig. 1B, C) was the average of all reflectors on the surrogate (2 for RPM and 4 for RGSC), and the default gating thresholds were determined from the breathing curve at simulation and recorded in the ARIA system (Varian Medical Systems, Palo Alto, CA). This motion may or may not be sufficient depending on the placement location. For all 60 patients, the VCE at the chest ROI, the surrogate motion, and the maximal abdominal elevation inferior to the xiphoid process were measured. The abdominal elevation was examined to determine whether a better box placement location existed with a larger vertical motion, especially for patients with limited VCE, or how much the abdominal surface should be covered by a second alternative ROI for A-DIBH motion monitoring (see Fig. 1).

Plan dosimetric evaluation of heart and lung sparing from FB to DIBH

When treating A-DIBH patients with VCE ≤ 5 mm, SGRT would fail to distinguish whether a patient was in DIBH or FB due to the clinical tolerance of ±3 mm/±3° configured for the AlignRT system in 6° of freedom. In the worst-case scenario, the patient may be treated in FB status completely as opposed to planned DIBH status, resulting in little gain in sparing the organs at risk (OARs), such as the heart and lungs. To quantify the dosimetric effect, we compared the doses to the heart and the lungs between the clinical DIBH plan and recalculated the FB plan with the same beams from the DIBH plan, simulating the aforementioned worst clinical scenario. Several dosimetric parameters were evaluated for the heart (mean dose [DMean], maximum dose [DMax], 5% of the maximum dose [D5%], V20 Gy, V30 Gy) and both lungs (V20 Gy, V10 Gy, V5 Gy, DMean). These doses were compared between the DIBH and FB plans to quantify OAR sparing among the abdominal-breathing patients, the statistical significance, and explanation of the sparing effects based on the shape and location of the OARs, most importantly the heart.

Results

Different anterior body elevations from FB to DIBH for A-DIBH breast patients

For abdominal-breathing patients, the heart position shift (inferiorly) and shape change (elongated) owing to the pull from the diaphragm in A-DIBH provides the mechanism for heart sparing in both VMAT and tangent planning, as shown in Fig. 2. Although the external VCE is small, the heart position and shape changes from FB to A-DIBH occur in the abdominal-breathing patients at various extents, allowing for heart sparing when DIBH is achieved.

Figure 2.

Figure 2

Two examples of heart elongation and position change for abdominal-driven deep inspiration breath hold (DIBH) in patients with little vertical chest wall elevation (VCE). (A) Patient #1 (mastectomy; VCE = 1 mm; volumetric arc therapy; heart maximum dose reduction of 4.6 Gy) on a flat breast board, and (B) patient #16 (implant; VCE = 4 mm; tangents; heart maximum dose reduction of 27.8 Gy) on a tilted board. Both axial and sagittal views are shown. The heart contour is pink in DIBH and green in free-breathing. Body contour is brown in DIBH and green in free-breathing. The green cross mark on the left chest wall is the treatment isocenter. Abbreviations: DIBH = deep inspiration breath hold; VCE = vertical chest wall elevation.

Table 1 tabulates all 20 patients with left-sided breast cancer with little VCE at A-DIBH. The first 10 patients with the locally advanced disease are treated with VMAT, whereas the other 10 patients are mostly at the early stage and treated with tangent plans. The vertical elevations at the chest surface and abdominal surface are listed, together with the motion of the surrogate (RPM or RGSC) that is visible in the simulation FB and DIBH CT images. Although the VCE (averaged at the breast and xiphoid process) is very small (3 ± 2 mm), the motion of the surrogate is large enough (10 ± 6 mm) near the xiphoid process due to the finite sizes of the surrogates (RPM: 34 mm and RGSC: 74 mm along the superior-inferior direction). The elevation at the abdomen (15 ± 6 mm) is even larger and more sensitive in monitoring A-DIBH.

Table 1.

Anterior body surface elevation and surrogate motion for 20 patients with left-sided breast cancer with abdominal-driven deep inspiration breath hold and little vertical chest elevation

Chest surface elevation (mm)
Surrogate motion (mm)
Abdomen surface elevation (mm)
Patient Age Breast status Plan tech Breast* Xiphoid Device Motion Location Motion Location
1 44 M V 0.0 1.6 RPM 5.0 61.6 7.5 136.5
2 61 L V 0.4 2.7 - - - 14.1 76.1
3 47 M V –0.6 4.1 RPM 25.0 96.7 28.1 129.6
4 47 I V 5.3 4.9 RGSC 13.0 59.4 18.2 135.9
5 86 M V 4.9 3.6 RPM 19.7 117.2 20.0 100.0
6 52 I V 0.8 1.1 RPM 1.3 108.4 13.0 129.5
7 75 L V 3.9 5.7 RPM 5.3 61.8 11.1 111.1
8 57 M V 5.4 4.9 RPM 9.3 24.0 7.7 77.2
9 34 M V 4.5 5.3 RGSC 5.2 99.0 11.1 111.2
10 40 I V 5.0 5.3 RPM 14.8 123.1 9.9 99.2
11 46 L T 1.0 3.7 RGSC 12.5 117.3 13.0 128.2
12 74 L T 1.6 0.6 RGSC 8.0 120.1 20.7 142.7
13 75 L T 2.1 4.3 RPM 4.0 138.3 9.7 94.8
14 65 L T –2.5 –3.5 RGSC 4.7 86.5 11.7 142.7
15 59 L T 5.2 4.3 RGSC 7.3 77.3 15.1 104.4
16 37 I T 3.1 5.9 RPM 10.6 62.6 21.5 156.3
17 34 L T 3.7 3.3 RPM 3.5 58.1 12.4 110.9
18 65 L T 3.6 2.8 RGSC 3.6 81.3 12.2 133.6
19 61 L T 4.3 4.5 RGSC 10.2 163.4 12.4 141.9
20 65 L T 3.7 2.9 RGSC 6.1 83.2 10.3 99.3
AVG 56 - - 2.8 3.4 - 8.9 91.5 14.0 118.1
STD 15 - - 2.3 2.2 - 6.0 33.8 5.2 22.8

Abbreviations: AVG = average; I = implant; L = lumpectomy; M = mastectomy; RGSC = respiratory gating for scanners with the box longitudinal length of 74 mm; RPM = real-time positioning management with the box longitudinal length of 34 mm; STD = standard deviation; T = tangents; V = volumetric arc therapy; VCE = mean vertical chest elevation between the breast isocenter and the xiphoid process. Plan Tech = planning technique.

The isocenter within the left breast is used to measure the vertical elevation.

The locations are from the xiphoid process to either the inferior end of the surrogate box or the abdominal point where the elevation is the largest.

Plan dosimetry comparison between FB and DIBH for heart and lung sparing

The DIBH plans and FB plans for the 20 patients are compared in pairs regarding heart dose. The dose reduction from FB to DIBH is significant (P < .05) in ∆DMax = –8.4 ± 9.8 Gy, ∆D5% = –2.4 ± 4.4 Gy, and ∆DMean = –0.6 ± 0.9 Gy. Although the dose reduction in the ipsilateral lung is also observed, it is trivial and not clinically significant (<1% on average). For the 10 VMAT plans, heart dose-sparing results from FB to DIBH are shown in Table 2, and for the 10 tangent plans, heart dose sparing is even more pronounced (∆DMax = –11.6 ± 10.8 Gy; P = .008) in Table 3. This is because the dose falloff from in-field to out-of-field in tangent plans is sharper than in VMAT plans. The 10 VMAT patients and a 4-field patient (#16) have locally advanced breast cancer (most with mastectomy or implant) and are treated with breast and regional nodal irradiation. The remaining 9 tangent (2F) patients have early stage breast cancer with lumpectomy. There is 1 VMAT case and 1 tangent case with the heart dose slightly higher in A-DIBH than in FB, but both are below the average heart dose in their VMAT or tangent groups, respectively.

Table 2.

Heart dose sparing based on VMAT plan dosimetry comparison between FB and DIBH for patients with left-sided breast cancer with abdominal-driven breathing and little vertical chest elevation (≤5 mm)

Planning
FB (heart, cGy)
DIBH (heart, cGy)
Heart sparing (cGy)
Patient Plan Rx (cGy) DMean DMax D5% DMean DMax D5% DMean DMax D5%
1 VMAT 200 × 25 379 2326 954 370 1871 958 –10 –455 4
2 VMAT 200 × 25 626 4180 1687 489 3378 1096 –137 –802 –591
3 VMAT 200 × 25 701 6005 2096 461 3362 898 –240 –2643 –1198
4 VMAT 200 × 25 733 4536 2192 485 4196 1439 –249 –340 –753
5 VMAT 266 × 16 330 1891 733 289 1401 588 –41 –490 –145
6 VMAT 200 × 25 824 4937 2291 685 3939 1799 –139 –998 –492
7 VMAT 200 × 25 383 3287 1045 353 2894 926 –30 –393 –119
8 VMAT 200 × 25 450 3637 907 425 2900 811 –25 –737 –96
9 VMAT 200 × 25 528 5176 1157 496 4287 1137 –32 –889 –20
10 VMAT 200 × 25 431 3564 965 437 3647 1024 5 84 59
AVG 539 3954 1402 449 3188 1068 –90 –766 –335
STD 172 1276 601 107 952 339 95 731 411
P value .02 .01 .03

Abbreviations: AVG = average; D5% = 5% of the maximum dose; DIBH = deep inspiration breath hold; DMax = maximum dose; DMean = mean dose; FB = free breathing; Rx = prescription; STD = standard deviation; VMAT = volumetric modulated arc therapy.

The heart dose reduction is significant (P < .05) with ∆DMax = –7.7 ± 7.3 Gy, ∆D5% = –3.4 ± 4.1 Gy, and ∆DMean = –0.9 ± 1.0 Gy from FB to DIBH.

Table 3.

Heart dose sparing based on tangent plan dosimetry comparison between free-breathing (FB) and deep inspiration breath hold (DIBH) for patients with left-sided breast cancer with abdominal-driven breathing and little vertical chest elevation (≤5 mm)

Planning
FB (heart, cGy)
DIBH (heart, cGy)
Heart sparing (cGy)
Patient Plan Rx (cGy) DMean DMax D5% DMean DMax D5% DMean DMax D5%
11 2F 265 × 16 64 544 190 56 374 161 –8 –170 –29
12 2F 265 × 16 65 1425 234 62 1108 231 –2 –317 –3
13 2F 520 × 5 49 2251 176 51 574 187 3 –1677 12
14 2F 265 × 16 75 3189 232 73 1622 252 –2 –1567 20
15 2F 265 × 16 145 3969 434 92 1321 297 –53 –2648 –137
16 4F 200 × 25 411 4788 1940 209 2010 557 –202 –2778 –1383
17 2F 375 × 8 38 1677 119 42 2059 125 4 382 6
18 2F 265 × 16 100 3109 289 81 1535 239 –19 –1574 –50
19 2F 520 × 5 611 3720 1731 476 2638 886 –135 –1082 –845
20 2F 265 × 17 254 3165 699 253 3014 723 –1 –151 24
AVG - - 181 2784 604 139 1626 366 –42 –1158 –239
STD - - 191 1295 672 138 839 262 71 1082 481
P value - - - - - - - - .10 .008 .15

Abbreviations: 2F = 2 fields (opposite tangent fields); 4F = 4 fields(opposite tangents, supraclavicular field, and posterior axillar boost field); AVG = average; D5% = 5% of the maximum dose; DIBH = deep inspiration breath hold; DMean = mean dose; FB = free breathing; Rx = prescription; STD = standard deviation.

The heart-sparing DMax = –11.6 ± 10.8 Gy (P = .008) from FB to DIBH.

It is worthwhile to note that VMAT plans have relatively higher heart doses (DMax, D5%, and DMean) than the tangent plans because multiple beam angles are applied, and some exit through the heart, leading to higher heart doses even though VMAT plans are optimized to spare the heart as much as possible.19 More importantly, the treatment target for the VMAT plan is usually larger, covering local lymph nodes, including the internal mammary node; therefore, this is also responsible for the higher heart dose. For these reasons, heart dose distribution in DIBH VMAT plans contains a large low-dose region and is very different from that of tangent plans. For instance, for DMean = 5 Gy, a tangent plan would need a considerable amount of the heart inside the beam field to raise the average between the full prescribed dose and near-zero background dose. For DIBH tangent plans, the heart is always blocked by the multileaf collimators (MLC), as shown in Fig. 3. However, the same MLC setting may not block the heart effectively in FB, leading to a much higher DMax as the anterior-lateral portion of the heart may fall inside the opposite tangent fields.

Figure 3.

Figure 3

(A) Illustration of heart blocking (patient #11) in the DIBH tangent plan using multileaf collimators, which (B) may not block the heart in FB, resulting in the anterior-lateral portion of the heart inside the tangent fields and causing a much higher heart maximum dose, at or near the full prescribed dose. Abbreviations: DIBH = deep inspiration breath hold; FB = free breathing.

Percentage of A-DIBH patients among the control group

The population of native abdominal-driven breathing patients is 6 of the 40 randomly selected control patients, about 15%, as shown in Table 4. Both the VCE (for SGRT ROI) and the motion of the surrogate (RPM/RGSC) are provided. Although the surrogate is placed around the xiphoid process, its motion represents a mix of both chest and abdominal motions due to its finite size (RPM: 3.4 cm and RGSC: 7.4 cm in the superior-inferior direction). Figure 4 shows the distributions of the chest ROI vertical motion based on the VCE values and surrogate motion data, suggesting the difference between surface-guided and surrogate-guided DIBH procedures. Within the FOV of FB and DIBH CT images, the most pronounced abdominal motion is also measured, indicating if the abdomen moves greater than the chest surface. Figure 5 provides a proposed clinical workflow to identify A-DIBH patients as early as simulation and apply a new abdomen ROI or switch to RPM/RGSC for A-DIBH monitoring.

Table 4.

Body surface elevations and surrogate motion at different locations among the 40 randomly selected patients with left-sided breast cancer DIBH in the control measurement

Vertical chest elevation (mm)
Measured surrogate motion between FB and DIBH cts (mm)
Default gating threshold (mm)*
Abdomen motion
Patient Plan Breast Xiphoid Type Motion Location Lower Upper Greater?
21 T 16.6 13.0 RGSC 18.1 Chest 11.0 16.0 Y
22 T 8.5 10.1 RGSC 10.5 Chest 6.5 11.5 N
23 V 5.0 3.8 RPM 7.0 Chest 5.1 10.1 Y
24 T 12.6 12.0 RGSC 24.8 Chest 8.0 13.0 Y
25 V 13.1 13.0 RGSC 21.9 Chest 11.7 16.7 Y
26 T 16.1 16.6 RPM 25.3 Chest 27.6 32.6 M
27 V 13.2 11.5 RPM 8.5 Chest 13.2 18.2 Y
28 T 17.8 15.1 RPM 16.7 Chest 15.1 19.9 N
29 T 5.3 4.1 RGSC 25.6 Chest 26.9 31.9 N
30 T 11.8 15.5 RPM 25.9 Chest 27.1 32.1 N
31 T 10.4 8.2 RPM 13.9 Chest 12.6 17.6 Y
32 T 7.1 6.8 RGSC 16.5 Chest 8.0 13.0 Y
33 V 8.0 7.9 RPM 18.0 Chest 9.6 14.6 N
34 V 5.4 4.9 RPM 13.3 Chest 13.5 18.5 Y
35 V 4.8 4.9 RGSC 4.1 Chest 3.8 8.8 Y
36 T 10.5 5.7 RGSC 16.1 ABD 17.0 21.9 Y
37 V 6.8 10.0 RPM 24.5 Chest 18.5 22.5 M
38 T 16.8 17.3 RGSC 22.1 Chest 13.1 18.1 N
39 T 10.4 11.4 RPM 9.8 Chest 8.4 13.4 M
40 V 10.1 5.4 RGSC 9.3 Chest 8.7 13.7 N
41 V 19.9 15.3 RGSC 20.1 Chest 8.0 13.0 Y
42 V 18.4 15.2 RGSC 18.1 Chest 14.7 19.7 M
43 T 12.9 13.0 RPM 12.1 Chest 11.8 16.8 M
44 V 16.3 5.3 RGSC 10.4 ABD 8.0 13.0 Y
45 T 11.3 11.3 RGSC 28.2 Chest 20.0 25.0 Y
46 V 10.5 7.4 RGSC 18.6 Chest 11.8 16.8 M
47 T 11.3 10.8 RPM 10.4 Chest 15.6 20.6 Y
48 T 11.6 8.9 RGSC 9.5 Chest 9.7 14.7 M
49 T 12.2 10.6 RPM 10.3 Chest 7.3 12.3 N
50 T 21.1 20.2 RGSC 24.6 ABD 22.0 27.0 N
51 V 5.0 5.8 RPM 13.9 ABD 16.0 21.0 M
52 V 14.3 11.4 RGSC 8.3 Chest 6.1 11.1 Y
53 V 6.5 9.6 RPM 10.0 Chest 10.6 15.6 Y
54 V 9.6 6.3 RPM 5.8 Chest 8.0 13.0 M
55 V 6.0 6.8 RPM 7.3 Chest 6.2 11.2 Y
56 T 16.8 11.6 RGSC 20.6 Chest 16.0 21.0 M
57 T 6.0 6.9 RGSC 17.4 Chest 15.0 20.0 Y
58 T 4.0 6.0 RGSC 11.0 Chest 7.5 12.5 Y
59 V 14.4 15.8 RPM 21.3 Chest 20.1 25.1 Y
60 T 24.4 26.8 RGSC 25.6 Chest 25.5 30.5 N

Abbreviations: ABD = abdomen; DIBH = deep inspiration breath hold; FB = free breathing; Y = yes, M = maybe, N = no; T = tangents; V = volumetric modulated arc therapy, RGSC = respiratory gating for scanners with the box longitudinal length of 74 mm; RPM = real-time positioning management with the box longitudinal length of 34 mm .

The default DIBH gating window is determined through the RPM or RGSC system.

M is indicated for a larger abdominal motion due to insufficient ABD in the field of view of FB and DIBH computed tomography.

Patients 23, 29, 34, 35, 51, and 58 have vertical chest elevation ≤5 mm.

Six patients are identified as abdominally-driven breathing individuals with little vertical chest elevation (≤5 mm).

Figure 4.

Figure 4

Histogram of chest surface elevation and surrogate motion and threshold in 40 control breast deep inspiration breath hold patients. The vertical elevation at (A) the breast ISO and (B) the xiphoid process has a significantly higher population of small values compared with (C) surrogate motion and (D) the default gating threshold. Even though the surrogate was placed around the xiphoid, it senses both thoracic and abdominal motion due to its finite size (RPM: 3.4 cm; RGSC: 7.4 cm in the superior-inferior direction). Abbreviations: ISO = isocenter, PRM = real-time positioning management, RGSC = respiratory gating for scanners.

Figure 5.

Figure 5

A proposed clinical workflow for online identification of A-DIBH patients and adaptation of an alternative ROI for surface-guided radiation therapy setup or use of RPM/RGSC as the backup method. Four opportunities to identify A-DIBH patients: (1) at simulation, (2) at planning and (3) plan checking, and (4) at the first treatment fraction. It is worthwhile to note that the chest-only ROI can be used for FB setup for both groups of patients: only from FB to DIBH does setup require a different motion monitoring technique. Abbreviations: A-DIBH = abdominal-driven DIBH; DIBH = deep inspiration breath hold; FB = free breathing; RGSC = respiratory gating for scanners; ROI = region of interest; RPM = real-time positioning management; T-DIBH = thoracic-driven DIBH; VCE = vertical chest wall elevation.

Discussion

Identification of A-DIBH patients early in the clinical workflow

As abdominal-driven breathing patients account for about 15% (6 of 40) of the control group, a sizable portion of the patient population, these patients must be identified early in the clinical workflow, as shown in Fig. 5. This finding is consistent with a previous report of 27% A-DIBH in 22 patients studied,16 suggesting that the population of A-DIBH patients is significant and must be treated accordingly. First, at simulation, it is the first opportunity to learn how a patient breathes into DIBH. After a brief patient training on how to perform DIBH near their maximum capacity without interference with the patient's native breathing behavior, therapists should check the vertical surface elevation at the xiphoid process and central abdomen before CT scans using a ruler and the horizontal room laser. Therefore, the abdominal-driven breathing patients can be identified before the CT scan, and the surrogate box should be placed interior to the xiphoid process with the maximum vertical motion. Furthermore, the information should be passed to the planning and treatment teams to use an appropriate motion monitoring technique for A-DIBH treatment.

Second, at the beginning of the planning process, the dosimetrist should check VCE when checking the alignment between the FB and DIBH CTs, and if VCE ≤ 5 mm, appropriate action should be taken, which is discussed in the next section. For tangent plans, a physician is often setting an MLC block to avoid the heart in the beam's eye view, and MLC motion is optimized to achieve a uniform dose in the midline of breast separation in the beam's eye view.20,21 For VMAT planning, the heart is always set as the primary OAR (an avoidance structure) in plan optimization to achieve clinical goals. Because of a sharp dose falloff outside of the planned target volume in tangent fields or VMAT plan, a small heart-shape deformation (elongation) or position change (toward the inferior direction) due to the diaphragmatic pull at A-DIBH may lead to a significant reduction of DMax (Tables 2 and 3, although DMean reduction (<1 Gy) may not be clinically significant.1 The dosimetrist must assess if the surrogate position at the simulation is optimal to monitor the DIBH (motion >5 mm); otherwise, an inferior shift of the box may be warranted in the patient's setup instruction. Alternatively, the default chest-only ROI should be altered to include or shift to the central abdominal area till the inferior border within the CT FOV.

The third opportunity for identification of A-DIBH patients is at the second independent plan checking process in which a senior physicist checks plan quality and approves it for treatment delivery. The anterior VCE must be checked together with FB DIBH CT alignment. Similarly, the plan checker should work with the planner to change the patient setup instruction to provide appropriate instruction for DIBH monitoring. In fact, most of the A-DIBH patients were identified by the plan checkers in this study.

The last opportunity is at the first fraction of treatment, in which the therapists who conduct the DIBH SGRT treatment (default procedure) should check the ROI surface difference after FB setup using the DIBH reference. Whereas in FB, the real-time motion monitoring will show the shift of the FB from DIBH; if a little vertical shift (VCE ≤ 5 mm) is identified using the default chest-only ROI, the therapists should inform the planner and plan checker to resolve this clinical issue first before treatment. The action tolerance threshold of 3 mm is usually kept unchanged unless a patient had difficulties reproducing the depth of DIBH after several trials. Also, a near-miss event should be reported via the institutional reviewing board for assessment by a dedicated clinical safety reading group.

Clinical solutions for proper monitoring of A-DIBH patients

The surrogate-guided and surface-guided motion monitoring indicators of DIBH are not equivalent, as they monitor different parts of the anterior surface and, thus, sense different motions. Therefore, their differences must be characterized thoroughly to avoid any potential pitfalls in routine clinical operations. As shown in Fig. 4, although the xiphoid process is included in both motion monitoring techniques, different results may occur, especially for A-DIBH patients. In the 40 control patients in this study, RPM/RGSC is still effective (motion >5 mm) in 95% of cases to monitor DIBH, including the abdominal-driven breathing patients, as it covers part of the upper abdominal surface due to its finite length (RPM: 3.4 cm and RGSC: 7.4 cm). However, the SGRT approach with a chest-only ROI (Fig. 1) is not suitable for A-DIBH as the ROI does not sense the abdominal motion. Therefore, caution must be paid when adopting the SGRT technology for left-sided breast DIBH treatment so that most breathing behaviors are covered.

Here, 2 feasible clinical solutions are suggested to cover abdominal-breathing patients without substantial changes to the existing clinical practice. First, within the SGRT approach, the ROI should be altered to cover the central part of the abdominal surface, which is correlated (r > 0.6) with the diaphragm motion.22 The soft abdominal surface is not supported by the rib cage, which is moving with the chest. To ensure significant motion detection in SGRT for abdominal-driven breathing patients, a second ROI could be created on the central abdominal surface to sense the abdominal surface motion, distinguishing A-DIBH from FB, whereas the default chest-only ROI can still be used for FB setup (Fig. 5). In fact, a recent study on automatic ROI definition using deep learning reported that the learned ROI includes the abdominal surface in 5 of 10 tested patients.23 Second, it is an option to switch the motion monitoring technique from SGRT to RPM/RGSC, as the DIBH scan at simulation is guided by the surrogate in our institution and used as the backup method in case the AlignRT system is down. However, the surrogate position may need to be shifted inferiorly, as shown in Fig. 1 and Table 1, in which 8 of 20 patients (40%) have a small surrogate motion (4 ± 1 mm) but higher abdominal motion (11 ± 2 mm) inferior to the clinical surrogate position. It is worthwhile to note that studies have shown similar setup accuracy between surface-guided and surrogate-guided procedures, whereas SGRT has the advantage of timesaving in the process.5,24 Clinically, as long as A-DIBH patients are identified before treatment (Fig. 5), the abdominal-driven breathing patients will be treated correctly with expected heart sparing, as shown in Tables 2 and 3.

Alternatively, but with a substantial change in the current clinical practice, a more intensive patient training and visual coaching program could be introduced to uniformly train all patients to go through chest-driven breathing and ensure T-DIBH throughout the treatment. This means that the native breathing behavior of A-DIBH patients must be consistently altered, which was previously reported.6,14 However, it is necessary to conduct daily visual feedback guidance, perhaps reinforced training, to allow patients to follow the trained breathing behavior throughout the multifractional treatment, so patients would not fall back to their native breathing behavior. In our clinics, patients are only instructed with a few brief DIBH practices with no intention to change their native behavior before CT scans at simulation, similar to another report,9 and guided to perform daily DIBH with audio feedback during SGRT treatment. The purpose of the initial training is to allow patients to get used to the DIBH process and reach their maximal capacity. Since patients’ native breathing behavior is not altered, the DIBH treatment tends to be more consistent and reproducible.

Heart sparing for abdominal-driven breathing patients in SGRT DIBH treatment

Recently, 3 comparison studies on the benefit of heart sparing by training patients to perform T-DIBH and A-DIBH versus FB at CT simulation used 2 planning methods: 3-dimensional conformal radiation therapy and intensity modulated radiation therapy.16, 17, 18 Our results in dosimetry and population of A-DIBH patients are consistent with theirs, showing heart sparing benefit of A-DIBH. By comparing the dosimetry of the plans, both Zhao et al16 and Matsumoto et al18 concluded that significantly lower doses to the heart in A-DIBH than in T-DIBH, although Hirata et al17 found similar heart doses in both DIBHs, much lower than FB, confirming the benefit in heart sparing for abdominal-driven breathing patients with breast cancer.

Zhao et al16 reported 27% native A-DIBH patients (6 of 22), which is consistent with our finding of 15% (6 of 40), suggesting a significant patient population that must be treated differently. However, our approach to handling A-DIBH clinically is fundamentally different from theirs, as we believe the patient's native breathing behavior is more reproducible and should not be altered. In fact, Matsumoto et al18 reported that among 120 patients, 20 could not follow the training to properly perform T-DIBH or A-DIBH in CT simulation, so they were excluded from that study, and only 100-patient results were reported.18 In reality, such events could occur in the clinic at simulation and/or treatment, resulting in these patients’ exclusion from DIBH treatment; otherwise, such uncertainty has to be accepted for DIBH treatment. Although A-DIBH may lower the heart dose more than T-DIBH, the number of A-DIBH patients is significantly smaller than T-DIBH, so changing most patients’ breathing behavior (73%-85%) could introduce substantial uncertainty, which could diminish the slight benefit of heart sparing in A-DIBH. After all, the previous reports support our clinical approach that patients’ native behavior should be preserved for higher DIBH reproducibility, and training and coaching patients to go through an unfamiliar DIBH process may pose a substantial challenge, as 17% of patients (20 of 120) could not follow the guidance18 and the tendency of falling back to patients’ natural breathing behavior.

In our radiation therapy clinics, DIBH breast treatment is decided by an attending physician at simulation if DIBH can help to achieve extra heart-chest wall separation compared with FB based on CT simulation. For patients with early stage disease, both tangent DIBH and prone treatment planning are viable options, whereas for patients with late-stage disease, VMAT or tangents with supraclavicular field and posterior axillary boost field at DIBH are often used. In tangents planning, a heart MLC block is usually drawn by the attending physician to achieve heart sparing (Fig. 3), whereas in VMAT planning, the heart is set as the primary OAR to spare.

For the 20 A-DIBH patients, the dosimetric comparison of FB and DIBH plans using the same technique shows that the most significant OAR sparing is the heart dose reduction in DMax and D5% (P < .05). Although the <1 Gy reduction of DMean is also statistically significant, clinically it may not be significant as the risk of long-term heart ischemic event is reduced by only <20% with a DMean reduction of <1 Gy.1 For tangent plans, the heart ∆DMean = –1.4 ± 1.4 Gy is already low at DIBH (Table 3) by using the MLC heart blocking. The most significant heart dose reduction is ∆DMax = –11.6 ± 10.8 Gy (P = .01), which reflects high-dose volume reduction of the heart away from the tangential beams. Therefore, the left anterior descending (LAD) coronary artery and left ventricle (LV) can be effectively spared from FB to DIBH.25,26 In a 2013 report on the risk of ischemic heart disease from radiation therapy by Derby et al,1 the increase of the DMean in tangent treatment is directly related to the high-dose heart volume, which may include the LAD and LV.

VMAT planning tends to produce a higher DMean (4.5 ± 1.1 Gy) at DIBH due to beam exits and internal mammary node coverage, but heart toxicity is manageable.19,27 The 1-Gy heart mean dose reduction may only result in a 7.4% reduction of heart ischemic events, according to the 2013 report.1 However, the report based on patients with left-sided breast cancer treated with tangent plans may not apply to the VMAT plan, as it has a very different dose distribution from a tangent plan. For instance, the heart DMax in tangent treatment at FB is almost constant near the full prescribed dose if a part of the heart is inside the beam fields; therefore, only the DMean, a volume-weighted average dose between the full dose and negligible background dose in a tangent plan, may have a linear relationship with the heart ischemic events. In VMAT planning, the local lymph nodes, including the internal mammary node, are also the targets, and multiple beams are used that may exit from the heart, leading to an inevitably higher low-dose volume, even though the inverse planning tries to minimize the heart dose (as the primary OAR) through optimization. Therefore, although the DMean is higher in the VMAT DIBH plan, and the volume that receives the full prescribed dose is small, the heart ischemic event is clinically manageable,6,14 different from the report.1

From FB to DIBH, because of the elongated heart shape and inferiorly shifted heart position, the anterior surface of the heart moves more inferiorly and/or medially by a few mm or even a few cm (Fig. 2); therefore, the increased separation effectively reduces the heart DMax by 7.7 ± 7.3 Gy (P < .05) and D5% by 3.4 ± 4.1 Gy. In our clinics, LAD and LV are not routinely contoured as avoidance structures so they cannot be evaluated quantitatively. However, as the LAD and LV are located on the anterior and lateral side of the heart and, thus, tend to receive the highest radiation dose, the DMax reduction often correlates with dose reduction to the LAD and LV. Therefore, DIBH is still an effective approach to treating abdominal-driven breathing patients with left-sided breast cancer with a significant heart-sparing benefit, even when a small VCE ≤ 5 mm is observed.

Last, it should be noted that the dosimetric study reported here presents the worst-case scenario as the patient is unlikely to be in full FB status throughout the entire treatment session despite ineffective SGRT monitoring due to small VCE, which cannot distinguish FB from A-DIBH. Therefore, more realistic setups would be random distribution over the course of multifractional treatment, and the dosimetric effect would likely be smaller than depicted in this work. The dosimetric comparison aimed to show that (1) A-DIBH can spare the heart, and (2) it is important to identify A-DIBH patients so they can be treated optimally with proper heart sparing. Our results are consistent with the previous studies,16, 17, 18 demonstrating the lowered heart dose by A-DIBH compared with FB. In the future, we plan to investigate actual heart doses between T-DIBH and A-DIBH patients using daily 2DkV data that are prescribed and acquired for VMAT patients with locally advanced disease.

Conclusion

A significant amount of DIBH patients (15%) with left-sided breast cancer in the control group of this study are native abdominal-driven breathing patients with little VCE (≤5 mm) from FB to A-DIBH. Clinically, abdominal-driven breathing patients can be benefitted from DIBH treatment, but they must be identified early in the clinical SGRT workflow during simulation, planning, plan checking, or at the latest the first fraction treatment so that appropriate motion monitoring techniques can be applied to ensure reproducible A-DIBH for heart-sparing. Immediate clinical solutions should be either to create a second ROI to include the central abdominal surface for DIBH setup in SGRT or to switch the motion monitoring technique to RPM/RSGC with a possible inferior placement shift when applicable. With these suggested solutions, patients with left-sided breast cancer with native abdominal-driven breathing behavior can still benefit from DIBH treatment with heart sparing.

Disclosures

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgments

This research is in part supported by the MSK Cancer Center Support Grant/Core Grant (P30 CA008748). The authors are grateful to Bo Zhao, Wei Lu, Ellen Yorke, Feifei Li, and Jeonghoon Park for their help in identifying and collecting patient data from the radiation oncology clinics.

Footnotes

Sources of support: This research is in part supported by the Memorial Sloan-Kettering Cancer Center support grant/core grant (P30 CA008748).

The data used in this study are available upon request after anonymization of patients’ identifiers.

References

  • 1.Darby SC, Ewertz M, McGale P, et al. Risk of ischemic heart disease in women after radiotherapy for breast cancer. N Engl J Med. 2013;368:987–998. doi: 10.1056/NEJMoa1209825. [DOI] [PubMed] [Google Scholar]
  • 2.Desai N, Currey A, Kelly T, Bergom C. Nationwide trends in heart-sparing techniques utilized in radiation therapy for breast cancer. Adv Radiat Oncol. 2019;4:246–252. doi: 10.1016/j.adro.2019.01.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Harry T, Rahn D, Semenov D, et al. Cardiac dosimetric evaluation of deep inspiration breath-hold level variances using computed tomography scans generated from deformable image registration displacement vectors. Med Dosim. 2016;41:22–27. doi: 10.1016/j.meddos.2015.06.004. [DOI] [PubMed] [Google Scholar]
  • 4.Cervino LI, Gupta S, Rose MA, Yashar C, Jiang SB. Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy. Phys Med Biol. 2009;54:6853–6865. doi: 10.1088/0031-9155/54/22/007. [DOI] [PubMed] [Google Scholar]
  • 5.Lu W, Li G, Hong L, et al. Reproducibility of chestwall and heart position using surface-guided versus RPM-guided DIBH radiotherapy for left breast cancer. J Appl Clin Med Phys. 2023;24:e13755. doi: 10.1002/acm2.13755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Reitz D, Walter F, Schonecker S, et al. Stability and reproducibility of 6013 deep inspiration breath-holds in left-sided breast cancer. Radiat Oncol. 2020;15:121. doi: 10.1186/s13014-020-01572-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Laaksomaa M, Sarudis S, Rossi M, et al. AlignRT® and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed? J Appl Clin Med Phys. 2019;20:97–104. doi: 10.1002/acm2.12553. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Li G, Lu W, O'Grady K, et al. A uniform and versatile surface-guided radiotherapy procedure and workflow for high-quality breast deep-inspiration breath-hold treatment in a multi-center institution. J Appl Clin Med Phys. 2022;23:e13511. doi: 10.1002/acm2.13511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Kugele M, Edvardsson A, Berg L, Alkner S, Andersson Ljus C, Ceberg S. Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy. J Appl Clin Med Phys. 2018;19:25–38. doi: 10.1002/acm2.12214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Li G. Advances and potentials of optical surface imaging in radiotherapy. Phys Med Biol. 2022;67:16TR02. doi: 10.1088/1361-6560/ac838f. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Jimenez RB, Batin E, Giantsoudi D, et al. Tattoo free setup for partial breast irradiation: A feasibility study. J Appl Clin Med Phys. 2019;20:45–50. doi: 10.1002/acm2.12557. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Sueyoshi M, Olch AJ, Liu KX, Chlebik A, Clark D, Wong KK. Eliminating daily shifts, tattoos, and skin marks: Streamlining isocenter localization with treatment plan embedded couch values for external beam radiation therapy. Pract Radiat Oncol. 2019;9:e110–e117. doi: 10.1016/j.prro.2018.08.011. [DOI] [PubMed] [Google Scholar]
  • 13.Rigley J, Robertson P, Scattergood L. Radiotherapy without tattoos: Could this work? Radiography. 2020;26:288–293. doi: 10.1016/j.radi.2020.02.008. [DOI] [PubMed] [Google Scholar]
  • 14.Schonecker S, Walter F, Freislederer P, et al. Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the Catalyst™/Sentinel™ system for deep inspiration breath-hold (DIBH) Radiat Oncol. 2016;11:143. doi: 10.1186/s13014-016-0716-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Al-Hallaq HA, Cervino L, Gutierrez AN, et al. AAPM task group report 302: Surface-guided radiotherapy. Med Phys. 2022;49:e82–e112. doi: 10.1002/mp.15532. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Zhao F, Shen J, Lu Z, et al. Abdominal DIBH reduces the cardiac dose even further: A prospective analysis. Radiat Oncol. 2018;13:116. doi: 10.1186/s13014-018-1062-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Hirata K, Narabayashi M, Hanai Y, et al. Comparison of thoracic and abdominal deep inspiration breath holds in whole-breast irradiation for patients with left-sided breast cancer. Breast Cancer. 2021;28:1154–1162. doi: 10.1007/s12282-021-01259-4. [DOI] [PubMed] [Google Scholar]
  • 18.Matsumoto Y, Kunieda E, Futakami N, et al. Dose and organ displacement comparisons with breast conservative radiotherapy using abdominal and thoracic deep-inspiration breath-holds: A comparative dosimetric study. J Appl Clin Med Phys. 2023;24:e13888. doi: 10.1002/acm2.13888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Kuo L, Ballangrud AM, Ho AY, Mechalakos JG, Li G, Hong L. A VMAT planning technique for locally advanced breast cancer patients with expander or implant reconstructions requiring comprehensive postmastectomy radiation therapy. Med Dosim. 2019;44:150–154. doi: 10.1016/j.meddos.2018.04.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Hong L, Hunt M, Chui C, et al. Intensity-modulated tangential beam irradiation of the intact breast. Int J Radiat Oncol Biol Phys. 1999;44:1155–1164. doi: 10.1016/s0360-3016(99)00132-7. [DOI] [PubMed] [Google Scholar]
  • 21.Chui CS, Hong L, Hunt M, McCormick B. A simplified intensity modulated radiation therapy technique for the breast. Med Phys. 2002;29:522–529. doi: 10.1118/1.1460875. [DOI] [PubMed] [Google Scholar]
  • 22.Song Y, Zhai X, Liang Y, Zeng C, Mueller B, Li G. Evidence-based region of interest (ROI) definition for surface-guided radiotherapy (SGRT) of abdominal cancers using deep-inspiration breath-hold (DIBH) J Appl Clin Med Phys. 2022;23:e13748. doi: 10.1002/acm2.13748. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Chen H, Chen M, Lu W, et al. Deep-learning based surface region selection for deep inspiration breath hold (DIBH) monitoring in left breast cancer radiotherapy. Phys Med Biol. 2018;63 doi: 10.1088/1361-6560/aaf0d6. [DOI] [PubMed] [Google Scholar]
  • 24.Penninkhof J, Fremeijer K, Offereins-van Harten K, et al. Evaluation of image-guided and surface-guided radiotherapy for breast cancer patients treated in deep inspiration breath-hold: A single institution experience. Tech Innov Patient Support Radiat Oncol. 2022;21:51–57. doi: 10.1016/j.tipsro.2022.02.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Gaal S, Kahan Z, Paczona V, et al. Deep-inspirational breath-hold (DIBH) technique in left-sided breast cancer: various aspects of clinical utility. Radiat Oncol. 2021;16:89. doi: 10.1186/s13014-021-01816-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Song J, Tang T, Caudrelier JM, et al. Dose-sparing effect of deep inspiration breath hold technique on coronary artery and left ventricle segments in treatment of breast cancer. Radiother Oncol. 2021;154:101–109. doi: 10.1016/j.radonc.2020.09.019. [DOI] [PubMed] [Google Scholar]
  • 27.Ho AY, Ballangrud A, Li G, et al. Long-term pulmonary outcomes of a feasibility study of inverse-planned, multibeam intensity modulated radiation therapy in node-positive breast cancer patients receiving regional nodal irradiation. Int J Radiat Oncol Biol Phys. 2019;103:1100–1108. doi: 10.1016/j.ijrobp.2018.11.045. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Advances in Radiation Oncology are provided here courtesy of Elsevier

RESOURCES