Abstract
Background
Motion artifacts in planning computed tomography (CT) for transcatheter aortic valve implantation (TAVI) can potentially skew measurements required for procedural planning. Whether such artifacts may affect safety or efficacy has not been studied.
Methods
We conducted a retrospective analysis of 852 consecutive patients (mean age, 82 years; 47% women) undergoing TAVI-planning CT at a tertiary care center. Two independent observers divided CTs according to the presence of motion artifacts at the annulus level (Motion vs. Normal group). Endpoints included surrogate markers for inappropriate valve selection: annular rupture, valve embolization or misplacement, need for a new permanent pacemaker, paravalvular leak (PVL), postprocedural transvalvular gradient, all-cause death.
Results
Forty-six (5.4%) patients presented motion artifacts on TAVI-planning CT (Motion group). These patients had more preexisting heart failure, moderate-severe mitral regurgitation, and atrial fibrillation. Interobserver variability of annular measurement (Normal vs. Motion group) did not differ for mean annular diameter but was significantly different for perimeter and area. Presence of motion artifacts on planning CT did not affect the prevalence of PVL (≥moderate PVL 0% vs. 2.5% p = 0.5), mean transvalvular gradient (6±3 mmHg vs 7±5 mmHg, p = 0.1), or the need for additional valve implantation (0% vs. 2.8%, p = 0.6). One annular rupture occurred (Normal group). Pacemaker implantation, procedural duration, hospital stay, 30-day outcomes, and all-cause mortality did not differ between the groups.
Conclusions
Motion artifacts on planning CT were found in about 5% of patients. Measurements for valve selection were possible without the need for repeat CT, with mean diameter-derived annulus measurement being the most accurate. Motion artifacts were not associated with worse outcomes.
Keywords: Aortic stenosis, Computed tomography, Motion artifacts, Preprocedural planning, TAVI
Graphical abstract
Introduction
Computed tomography (CT) is the gold standard for preprocedural planning in patients scheduled for transcatheter aortic valve implantation (TAVI).1 CT is used for visualization and measurement of the annulus, the left ventricular outflow tract, the sinus of Valsalva, origin of the coronary ostia, and the iliofemoral arteries. Furthermore, the perpendicular valve planes (implant views) are identified. Precise annular measurements lead to improved outcomes including fewer paravalvular leaks (PVLs).
Despite the widespread use of multidetector CT and electrocardiogram (ECG) triggering, motion artifacts (or pulsation artifacts) in the annular region are encountered quite frequently in clinical practice.2 Patients with a greater likelihood of presenting this artifact are those with atrial fibrillation and those unable to follow commands or hold their breath.
We hypothesized that motion artifacts may reduce the precision and/or reproducibility of annular measurements required for valve selection and may result in worse hemodynamic (PVLs, valvular gradient) and clinical outcomes.
Methods
Between June 2012 and December 2021, a total of 852 consecutive patients with severe aortic stenosis underwent TAVI at the Heart Center Lucerne. All patients underwent planning CT and were discussed in the interdisciplinary heart team. All patients were enrolled in the prospective SwissTAVI registry, which was approved by the local ethics committee (NCT01368250).3, 4, 5 All patients provided written informed consent for the procedure and for prospective data acquisition and follow-up examinations.
Planning Computed Tomography
TAVI-planning CT was obtained in all patients with a Flash Dual Source scanner, Siemens Somatom Definition Flash (Siemens Medical, Erlangen, Germany) with ECG triggering in retrospective mode as previously described.6,7 The phase between 30% and 70% of the heart rate was recorded and reconstruction was performed during systole as well as diastole. The complete data set was transferred to a dedicated postprocessing workstation, measurements and analyses were performed with a commercially available software (Merlin, Phoenix-PACS, Freiburg, Germany). Analyses, measurements for access site and valve selection (aortic annulus, left ventricular outflow tract, aortic root and sinotubular junction dimentions and coronary height) were performed using a multiplanar double-oblique reconstruction.
Definition of Motion Artifact and Classification
Motion artifact was defined as a double contour of the annulus, making clear-cut targets for measurement of diameter/perimeter/area difficult. If motion artifacts were present during the first scan, the scan was repeated up to 2 additional times aiming to obtain artifact-free images. If 3 sets of images presented motion artifacts, no further scan was performed. These patients were classified in the Motion artifact group (46, 5.4% of the cohort). Importantly, if the patient had any set of images free of motion artifact, they were classified in the Normal group. In the Normal group, 120 patients (14.2% of the cohort) presented at least one set of CT images with motion artifacts at the annular level.
CTs were analyzed by 2 independent TAVI operators with extensive experience in CT analysis and sizing of transcatheter heart valves (THVs). The classification was binary according to the presence or absence of motion artifact at the annular level. Motion artifacts at other anatomical locations did not qualify since they do not affect sizing measurements for THV.
Blinded Assessment of Interobserver Variability
Two TAVI operators with extensive experience in CT planning independently assessed mean annular diameter and annular perimeter of a randomly selected cohort of 20 CTs without motion artifact and 20 with motion artifact presented in random order. Random selection was performed with Google random number generator (https://www.google.com/search?q=random+number).
Study Endpoints
In order to assess an inadvertent oversizing or undersizing of the valve, surrogate parameters of an inappropriate valve choice were assessed. Annular rupture and new permanent pacemakers served as a surrogate for oversizing. Need for additional valve implantation (as a consequence of valve embolization or “pop-out”), PVL, and elevated mean transvalvular gradient on predischarge echocardiography served as surrogate parameters of undersizing. All-cause death during long-term follow-up was compared between groups.
All events, defined according to the updated definitions of the Valve Academic Research Consortium (VARC-3),8 were adjudicated by an independent clinical events committee in a prospective fashion.
Statistical Analysis
Continuous variables are presented as mean ± SD or median (interquartile range) as appropriate, categorical variables as n (%). When normally distributed, continuous variables were compared with Student's t-test; if not fitting a normal distribution, the Wilcoxon rank sum test was used. The Fisher exact test was used for comparisons of categorical variables. Event-free survival rates were estimated using the Kaplan-Meier method and compared with the log-rank test. Univariate and multivariable Cox regression analyses were performed to identify predictors for mortality and hospitalization for heart failure during follow-up. Variables with a univariate p value <0.1 were included in the multivariable model. A two-tailed p value <0.05 was considered statistically significant. Statistical analyses were performed with STATA (Version 13, StataCorp, College Station, TX). Interobserver agreement was assessed with correlation coefficients and Bland-Altman plots, differences in interobserver variability between Motion and Normal group (comparison of slopes and intercepts) using Prism 9 (Version 9.0.2, GraphPad Software, San Diego, CA).
Results
Baseline characteristics are shown in Table 1. The mean age was 81 ± 6 years, 47% of patients were female. Motion artifacts were present in 46 patients (5.4% of the cohort). Patients with motion artifacts more commonly had prior myocardial infarction (19.6% vs. 9.4% p = 0.039), atrial fibrillation (45.7% vs. 19.0%, p < 0.001), mitral regurgitation ≥ moderate (14.0% vs. 3.6%, p = 0.007), higher NT-pro-BNP (2631 (1069-6223) pg/mL vs. 1428 (571-3770) pg/mL, p = 0.02), and had a lower LVEF (51 ± 15% vs. 56 ± 13%, p = 0.01).
Table 1.
Baseline characteristics
| Baseline characteristic | All (n = 852) | Motion artifact (n = 46) | Normal (n = 806) | p value |
|---|---|---|---|---|
| Age, years | 81 ± 6 | 82 ± 5 | 81 ± 6 | 0.568 |
| Female sex | 393 (46.1%) | 20 (43.5%) | 373 (46.3%) | 0.76 |
| BMI, kg/m2 | 27.1 ± 5.1 | 28.2 ± 4.4 | 27.3 ± 5.1 | 0.13 |
| Hypertension | 768 (79.6%) | 35 (76.1%) | 643 (79.8%) | 0.55 |
| Diabetes | 194 (22.8%) | 10 (21.7%) | 184 (22.8%) | 0.86 |
| Peripheral arterial disease | 136 (16.0%) | 5 (10.9%) | 131 (16.3%) | 0.41 |
| Coronary artery disease | 439 (51.5%) | 26 (56.5%) | 413 (51.2%) | 0.55 |
| Prior myocardial infarction | 85 (10.0%) | 9 (19.6%) | 76 (9.4%) | 0.039 |
| Prior PCI | 165 (19.4%) | 9 (19.6%) | 156 (19.4%) | 1.0 |
| Prior CABG | 52 (6.1%) | 0 (0%) | 52 (6.5%) | 0.11 |
| Heart failure | 233 (27.4%) | 14 (30.4%) | 219 (27.2%) | 0.61 |
| Prior stroke/TIA | 93 (10.9%) | 3 (6.5%) | 90 (11.2%) | 0.47 |
| Atrial fibrillation | 174 (20.4%) | 21 (45.7%) | 153 (19.0%) | <0.001 |
| eGFR, ml/min | 55 ± 25 | 56 ± 25 | 55 ± 22 | 0.57 |
| COPD | 118 (13.9%) | 3 (6.5%) | 115 (14.3%) | 0.19 |
| Aortic valve area, cm2 | 0.76 ± 0.18 | 0.76 ± 0.14 | 0.76 ± 0.18 | 0.96 |
| Mean gradient, mmHg | 46 ± 16 | 41 ± 16 | 46 ± 16 | 0.05 |
| LVEF (%) | 56 ± 13 | 51 ± 15 | 56 ± 13 | 0.01 |
| Mitral regurgitation ≥ moderate | 33 (34.2%) | 6 (14.0%) | 26 (3.6%) | 0.007 |
| NT-pro-BNP, pg/mL | 1472 (590-3843) | 2631 (1069-6223) | 1428 (571-3770) | 0.02 |
| EuroSCORE II, % | 5.0 ± 4.4 | 4.5 ± 3.9 | 5.0 ± 4.5 | 0.52 |
| STS PROM, % | 4.5 ± 0.41 | 4.2 ± 0.5 | 4.5 ± 0.1 | 0.68 |
Notes. Bold values indicate statistical significance (p < 0.05).
Abbreviations: BMI, body mass index; CABG, coronary artery bypass grafting; COPD, chronic obstructive pulmonary disease; eGFR, estimated glomerular filtration rate; LVEF, left ventricular ejection fraction; NT-pro-BNP, N-terminal pro brain natriuretic peptide; PCI, percutaneous coronary intervention; STS PROM, society of thoracic surgeons Predicted Risk of Morbidity and Mortality; TIA, transient ischemic attack.
Figure 1 shows an example of a motion artifact at the annulus level on TAVI-planning CT (Figure 1a), with a normal CT for comparison (Figure 1b).
Figure 1.
Examples of planning CT. (a) Motion artifacts (∗) affecting the ascending aorta, aortic root, annulus (middle panel). (b) Normal CT scan.
Abbreviation: CT, computed tomography.
Interobserver Variability
The resulting TAVI-planning measurements in a subset of randomly selected patients (motion artifact n = 20 and normal n = 20) independently assessed by 2 operators are depicted in Figure 2. Measurements comprised the mean annular diameter derived by the long and short diameter (Figure 2a), annular perimeter (Figure 2b), and annular area (Figure 2c). The 2 independent operators achieved similar agreement with a numerically lower r2-value in the motion artifact group: the mean annulus diameter r2 was 0.82 and 0.85 for motion artifact and normal CT, respectively (Figure 2a), perimeter r2 0.84 and 0.93 for motion artifact and normal CT, respectively (Figure 2b), and area r2 0.72 and 0.91 for motion artifact and normal CT, respectively (Figure 2c). The statistical difference was further analyzed with slope and intercept comparison: Motion and Normal group did not have a different interobserver variability for mean diameter (p = 0.4), but interobserver variability differed significantly for perimeter (p = 0.002) and area (p = 0.01), Figure 3.
Figure 2.
Interobserver variability. Correlation (left) and Bland-Altman plots (right) for interobserver variability for Annulus in long axis (a), annular perimeter (b) and annular area (c). In Bland-Altman plots bias is shown with a dashed line, 95% limits of agreement with dotted lines.
Figure 3.
Slope comparison of interobserver variability between Motion and Normal group.
Bland-Altman plots showed several differences. Mean diameter motion artifact: bias −0.5 mm, 95% CI (−2.7 to 1.7), normal: bias −0.6 mm, 95% CI (−2.3 to 1.2), (Figure 2a). Perimeter motion artifact: bias −1.6 mm, 95% CI (−8.3 to 5.2), normal: bias −2.5 mm, 95% CI (−7.1 to 2.1), (Figure 2b). Area motion artifact: bias −33.9 mm2, 95% CI (−114.1 to 46.3), normal: bias −33.2 mm2, 95% CI (−96.1 to 29.8), (Figure 2c). We interpret these differences as not clinically relevant, especially mean diameter (bias difference of 0.1 mm between Normal and Motion). Additionally, there was no consistent bias toward oversizing or undersizing when motion artifacts were present. Both operators measured an annular size between 2 valve sizes (as suggested by the IFU of Evolut Pro +, Medtronic, Dublin, Ireland) in 1 case in the motion artifact group (5%) and 1 case in the normal group (5%).
Procedural Characteristics and Outcomes
There were no significant differences in procedural characteristics (Table 2). Presence of motion artifacts was not associated with a difference in type of valve implanted, nor was it associated with need for a second valve (due to valve embolization or “pop-out”) or with prolonged procedure time. The in-hospital outcomes were comparable in patient with and without motion artifacts; in particular, no difference was present in regards to mean transvalvular gradient or PVLs on predischarge echocardiography. Patients with motion artifacts did have a higher rate of PVLs. A graphical depiction of the prevalence of PVLs on discharge echocardiography is present in Figure 4. Thirty-day outcomes including implantation of a new pacemaker did not differ between the 2 groups (Table 2).
Table 2.
Procedural characteristics, in-hospital and 30-d outcomes
| Procedural characteristics | All (n = 852) | Motion artifact (n = 46) | Normal (n = 806) | p value |
|---|---|---|---|---|
| Implanted valve | 0.38 | |||
| Sapien 3 | 151 (17.7%) | 11 (23.9%) | 140 (17.4%) | |
| Corevalve/Evolut R/PRO | 275 (32.3%) | 14 (30.4%) | 261 (32.4%) | |
| ACURATE neo/neo 2 | 333 (39.1%) | 19 (41.3%) | 314 (37.0%) | |
| Allegra | 93 (10.9%) | 2 (4.4%) | 91 (11.3%) | |
| Procedural duration, min | 48 (38-60) | 51 (42-59) | 48 (38-60) | 0.31 |
| Implantation of a second valve | 21 (2.4%) | 0 (0%) | 21 (2.6%) | 0.63 |
| In-hospital outcomes | ||||
| Mean gradient, mmHg | 7.3 ± 4.5 | 6.3 ± 3.0 | 7.4 ± 4.6 | 0.11 |
| Aortic valve area, cm2 | 2.1 ± 0.5 | 2.1 ± 0.6 | 2.1 ± 0.5 | 0.57 |
| Paravalvular leak ≥ moderate | 19 (2.2%) | 0 (0%) | 19 (2.4%) | 0.62 |
| Median duration of hospitalization, days | 6(3-8) | 5(3-10) | 6(3-8) | 0.95 |
| 30-d outcomes | ||||
| Vascular complications | 0.26 | |||
| None | 761 (89.3%) | 39 (84.8%) | 722 (89.6%) | |
| Minor vascular complication | 34 (4.0%) | 4 (8.7%) | 30 (3.7%) | |
| Major vascular complication | 57 (6.7%) | 3 (6.5%) | 54 (6.7%) | |
| Bleeding | 0.25 | |||
| None | 769 (90.3%) | 40 (87.0%) | 729 (90.5%) | |
| Minor bleeding | 23 (2.7%) | 2 (4.4%) | 21 (2.6%) | |
| Major bleeding | 35 (4.1%) | 1 (2.2%) | 34 (4.2%) | |
| Life-threatening bleeding | 25 (2.9%) | 3 (6.5%) | 22 (2.7%) | |
| Implantation of a new permanent pacemaker | 71 (8.3%) | 6 (13.0%) | 65 (8.1%) | 0.27 |
| Any stroke | 29 (3.4%) | 1 (2.2%) | 28 (3.5%) | 1.0 |
| Annular rupture | 1 (0.1%) | 0 (0%) | 1 (0.1%) | 1.0 |
| All-cause mortality | 12 (1.4%) | 0 (0%) | 12 (1.5%) | 1.0 |
Figure 4.
Paravalvular leak on discharge echocardiography.
After a median follow-up of 17 (8-39) months, all-cause mortality did not differ between patients with normal CT or motion artifacts at the annular level (hazard ratio (HR) 1.0, 95% CI 0.6-1.9, p = 0.9, Figure 5).
Figure 5.
Clinical outcomes. Kaplan-Meier curves for all-cause mortality.
Univariate and multivariable predictors for all-cause mortality are shown in Table 3. In multivariable analysis, female sex, heart failure, atrial fibrillation, and kidney function remained significant predictors, with heart failure and atrial fibrillation being the strongest predictors (HR 1.71 [1.24-2.37] and HR 1.30[1.05-1.78], respectively, Table 3).
Table 3.
Univariate and multivariable predictors for all-cause mortality
| Variable | Univariate HR (95% CI) | p value | Multivariable HR (95% CI) | p value |
|---|---|---|---|---|
| Motion artifact | 1.01 (0.55-1.87) | 0.952 | - | |
| Age, per year | 1.03 (1.01-1.06) | 0.011 | 1.01 (0.99-1.04) | 0.297 |
| Female sex | 0.71 (0.54-0.95) | 0.021 | 0.62 (0.46-0.84) | 0.002 |
| Diabetes | 1.23 (0.90-1.70) | 0.192 | - | |
| Hypertension | 1.15 (0.80-1.66) | 0.453 | - | |
| Prior stroke/TIA | 1.38 (0.94-2.02) | 0.100 | - | |
| Coronary artery disease | 0.97 (0.74-1.28) | 0.840 | - | |
| Heart failure | 2.27 (1.71-3.01) | <0.001 | 1.71 (1.24-2.37) | 0.001 |
| Atrial fibrillation | 1.71 (1.28-2.30) | <0.001 | 1.30 (1.05-1.78) | 0.01 |
| Peripheral arterial disease | 1.69 (1.23-2.33) | 0.001 | 1.19 (0.82-1.70) | 0.335 |
| COPD | 1.26 (0.87-1.83) | 0.219 | - | |
| eGFR (ml/min) | 0.98 (0.98-0.99) | <0.001 | 0.99 (0.98-0.99) | 0.030 |
| STS PROM | 1.08 (1.05-1.12) | <0.001 | 1.02 (0.98-1.07) | 0.200 |
Notes. Bold values indicate statistical significant (p < 0.05).
Abbreviations: COPD, chronic obstructive pulmonary disease; eGFR, estimated glomerular filtration rate; STS PROM, society of thoracic surgeons Predicted Risk of Morbidity and Mortality; TIA, transient ischemic attack.
Discussion
We asked the question: “do motion artifacts at the annulus level during TAVI-planning CT affect measurement for valve selection and outcomes?”. In 852 consecutive patients undergoing TAVI, 5.4% presented motion artifacts on TAVI-planning CT (Motion group); the main distinguishing feature was a greater proportion of atrial fibrillation, which is known to interfere with ECG triggering during CT scan. Moreover, patients in the Motion group had a lower LVEF, a higher proportion of mitral regurgitation ≥ moderate, and higher NT-pro-BNP, all of which are associated with atrial fibrillation and furthermore could have led to more difficulties in breath-holding during image acquisition, possibly contributing to motion artifacts.
Measurements required for valve selection were possible in all patients with motion artifacts. When assessed by 2 independent operators, a randomly selected sample of n = 20 patients showed that no major discrepancy in measurements was present using mean diameter derived from long and short axis, but significant differences were present in the interobserver variability for perimeter and area. Bland-Altman plots did not show a consistent bias toward either oversizing or undersizing in the Motion group. Importantly, Bland-Altman plots demonstrated that the absolute difference in sizing is unlikely to be clinically relevant: for mean diameter, the difference was −0.5 mm (−2.7 to 1.7 mm) for Motion and −0.6 mm, (−2.3 to 1.2 mm) for Normal, meaning that the bias between the Motion and Normal group is 0.1 mm, therefore unlikely to sway the valve choice toward a larger (or smaller) THV. Similarly, the difference in bias for perimeter between the Motion and Normal group was 0.9 mm and 0.7 mm2 for area. Interobserver variability in our sample is similar to that reported previously,9 with mean diameter being one of the most motion-artifact “resistant.” In a previous study, mean diameter and area were the parameters with the best performance in terms of interobserver variability.10
The impact of inadvertent inclusion of an artifact-affected or “blurry” part of the annulus is likely to be more pronounced when perimeter or area measurements are performed because more of the “blurry” part might be included as opposed to long/short perimeter derived measurements where only a line might cross into the “blurry” part of the annulus. This is one possible explanation why long/short axis derived mean diameter performs better in patients with motion artifacts.
Implantation of a transcatheter heart valve was technically possible in all patients regardless of the presence of motion artifacts on planning CT, without significant differences in procedural characteristics. Surrogate parameters of an inadequate valve selection did not show any signal toward potential harm. Surrogate parameters for inadvertent oversizing (annular rupture, higher pacemaker-implantation rate), as well as surrogate parameters for inadvertent undersizing (valve embolization or pop-out, PVLs and elevated transvalvular gradients) were similar in patients with and without motion artifacts on TAVI-planning CT. In addition, all-cause mortality during follow-up did not differ between the 2 groups. Univariate and multivariable predictors for all-cause mortality did not include motion artifact and seemed to be tied to patients’ baseline comorbidities.
Taken together, our data suggest that (i) in the presence of motion artifact, it is possible to perform accurate, reproducible measurement for valve selection, (ii) long and short axis derived mean annular diameter may be used instead of perimeter or area in such patients, (iii) there is no signal toward higher procedural complications in patients with motion artifacts, (iv) the hemodynamic result is comparable to patients without motion artifact, and (v) there is no survival disadvantage during long-term follow-up.
Limitations
There are several limitations that need to be considered when interpreting the results from our study. First, all limitations inherent to the retrospective, single-center nature of the study apply. Second, the number of patients with motion artifacts is low. This might partly be due to the use of second-generation dual source “flash” scanners, which have a higher temporal resolution than single source scanners and therefore lead to less artifacts. This, of course, limits the generalizability of our findings. Third, this study represents consecutive patients at a single institution, where although 2 operators and 1 radiologist perform the measurements independently for every patient, the program director is involved in every patient’s measurement and valve selection process: this lowers the degree of operator-to-operator variability in the measurement process required for valve selection by introducing a certain continuity and standard in measurement and valve selection. This limits the generalizability to other institutions or programs with several operators performing valve selection independently.
Furthermore, it is well-known that several other factors in addition to annular measurements play a role in valve choice (type of valve as well as valve size): extent and degree of valve calcification, left vetricular outflow tract calcification, leaflet symmetry, coronary height, sinotubular junction diameter, and presence of a “horizontal aorta” being the most prominent. These factors might have affected valve choice, and in the setting of a “borderline measurement” might have “swayed” sizing toward a smaller or larger size. Correction for these factors was not possible in our study.
Conclusions
Motion artifacts on TAVI-planning CT affect about 5% of patients. The presence of motion artifacts allows accurate and reproducible valve selection using long and short axis derived mean annular diameter and does not affect procedural, hemodynamic, or clinical outcomes. These findings suggest that the presence of motion artifact does not warrant repeat CT.
Ethics Statement
The research reported adhered to the local and institutional ethical guidelines.
Funding
The authors have no funding to report.
Disclosure Statement
Mathias Wolfrum serves as a proctor for Biosensors. Matthias Bossard has received speaker fees from Abbott Vascular, Amgen, Astra Zeneca, Bayer, Daichii-Sankyo, Mundipharma, and SIS Medical. Stefan Toggweiler serves as a proctor and consultant for Abbott, Biosensors, Boston Scientific, and Medtronic, as a consultant for Shockwave, Teleflex, Medira, atHeart, and Veosource, has received institutional research grants from Boston Scientific and Fumedica AG, and holds equity in Hi-D Imaging. The other authors declare no conflict of interest.
Acknowledgments
Central Illustration was created using Servier Medical Art (www.servier.com).
References
- 1.Blanke P., Weir-McCall J.R., Achenbach S., et al. Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/Transcatheter aortic valve replacement (TAVR): an expert consensus document of the society of cardiovascular computed tomography. JACC Cardiovasc Imaging. 2019;12:1–24. doi: 10.1016/j.jcmg.2018.12.003. [DOI] [PubMed] [Google Scholar]
- 2.Roos J.E., Willmann J.K., Weishaupt D., Lachat M., Marincek B., Hilfiker P.R. Thoracic aorta: motion artifact reduction with retrospective and prospective electrocardiography-assisted multi-detector row CT. Radiology. 2002;222:271–277. doi: 10.1148/radiol.2221010481. [DOI] [PubMed] [Google Scholar]
- 3.Stortecky S., Heg D., Tueller D., et al. Infective endocarditis after transcatheter aortic valve replacement. J Am Coll Cardiol. 2020;75:3020–3030. doi: 10.1016/j.jacc.2020.04.044. [DOI] [PubMed] [Google Scholar]
- 4.Tomii D., Okuno T., Praz F., et al. Potential candidates for transcatheter tricuspid valve intervention after transcatheter aortic valve replacement: predictors and prognosis. JACC Cardiovasc Interv. 2021;14:2246–2256. doi: 10.1016/j.jcin.2021.07.030. [DOI] [PubMed] [Google Scholar]
- 5.Stortecky S., Franzone A., Heg D., et al. Temporal trends in adoption and outcomes of transcatheter aortic valve implantation: a SwissTAVI Registry analysis. Eur Heart J Qual Care Clin Outcomes. 2019;5:242–251. doi: 10.1093/ehjqcco/qcy048. [DOI] [PubMed] [Google Scholar]
- 6.John D., Buellesfeld L., Yuecel S., et al. Correlation of device landing zone calcification and acute procedural success in patients undergoing transcatheter aortic valve implantations with the self-expanding CoreValve prosthesis. JACC Cardiovasc Interv. 2010;3:233–243. doi: 10.1016/j.jcin.2009.11.015. [DOI] [PubMed] [Google Scholar]
- 7.Brinkert M., De Boeck B., Stampfli S.F., et al. Predictors of paravalvular leak following implantation of the ACURATE neo transcatheter heart valve: the PREDICT PVL study. Open Heart. 2020;7:e001391. doi: 10.1136/openhrt-2020-001391. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Kappetein A.P., Head S.J., Genereux P., et al. Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document. J Am Coll Cardiol. 2012;60:1438–1454. doi: 10.1016/j.jacc.2012.09.001. [DOI] [PubMed] [Google Scholar]
- 9.Le Couteulx S., Caudron J., Dubourg B., et al. Multidetector computed tomography sizing of aortic annulus prior to transcatheter aortic valve replacement (TAVR): variability and impact of observer experience. Diagn Interv Imaging. 2018;99:279–289. doi: 10.1016/j.diii.2017.12.007. [DOI] [PubMed] [Google Scholar]
- 10.Gurvitch R., Webb J.G., Yuan R., et al. Aortic annulus diameter determination by multidetector computed tomography: reproducibility, applicability, and implications for transcatheter aortic valve implantation. JACC Cardiovasc Interv. 2011;4:1235–1245. doi: 10.1016/j.jcin.2011.07.014. [DOI] [PubMed] [Google Scholar]






