The coronavirus disease 2019 (COVID-19) pandemic has caused disruption in health systems all over the world. A considerable number of patients with COVID-19 need mechanical ventilation.1, 2 It is therefore anticipated that there will be a surge in the number of patients requiring respiratory support and there may be a shortfall in the number of devices available.
Increasing numbers of patients with chronic respiratory failure are being treated at home with small mechanical ventilators.3 Such ventilators are suitable for both airway stabilisation and ventilator support in patients with diseases ranging from chronic obstructive pulmonary disease to neuromuscular disease (eg, patients with spinal injury and patients with motor neuron disease). Given that the number of such ventilators is considerable, the use of these machines to help ventilate patients with COVID-19 is appealing. In this case series, we report how ventilators designed for domiciliary support of patients with chronic respiratory failure may be used for invasive support in patients with COVID-19 needing mechanical ventilation.
Methods
We studied a cohort of consecutive ventilated patients in COVID-19-dedicated intensive care units (ICUs) at San Raffaele Scientific Institute, Milan, Italy. In a proof- of-concept study, patients receiving invasive ventilatory support (at any stage of the disease) were transitioned for 3 hours from the ICU ventilator to ventilators designed for support of patients with chronic respiratory insufficiency (the intervention period) and then back to ICU ventilators. Arterial blood gases were collected hourly during this period. In addition to descriptive analysis, we measured the coefficient of variation within each subject, calculated as the standard deviation divided by the mean.
Two devices were tested: the VEMO 150 (EOVE, VitalAire, Italy) and VIVO 55 (Breas, MedicAir, Italy) ventilators. The system used for each of the devices is described below. Supplementary oxygen was supplied at low pressure through the purpose-built inlet on both machines.
System for the VEMO 150
Outflow port → Oximeter → Heat and moisture exchanger (HME) → In tube → Y pieces → HME → Tracheal tube and back to the inflow port (Figure 1).
Figure 1.
Systems used for the two continuous positive airway pressure devices tested: VEMO 150 (EOVE, VitalAire, Italy) and VIVO 55 (Breas, MedicAir, Italy) ventilators
HME = heat and moisture exchanger.
System for the VIVO 55
Outflow port → Oximeter → HME→ In tube → Exhalation valve → HME → Tracheal tube (Figure 1).
Results
Over 10 days, from 28 March to 7 April 2020, seven patients were enrolled in this case series, including one patient with few measurements. Ventilatory variables are shown in Table 1.
Table 1.
Clinical characteristics of the patients
| Patients |
|||||||
|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | |
| Clinical characteristics | |||||||
| Age (years) | 47 | 69 | 49 | 49 | 76 | 54 | 65 |
| Gender | Male | Female | Male | Female | Male | Male | Female |
| Use of NMBA | Yes | No | Yes | No | No | Yes | No |
| Haemodynamic impairment | No | No | No | No | No | No | No |
| Tracheostomy | No | Yes | No | No | No | No | No |
| Ventilator* | |||||||
| Beginning | Servo-i | Servo-i | EVITA 4 | EVITA 4 | Servo-i | EVITA 4 | Servo-i |
| Hour 1 | VEMO 150 | VIVO 50 | VEMO 150 | VIVO 50 | VIVO 50 | VEMO150 | |
| Hour 2 | VEMO 150 | VIVO 50 | VEMO 150 | VIVO 50 | VIVO 50 | VEMO150 | |
| Hour 3 | VEMO 150 | VIVO 50 | VIVO 50 | VIVO 50 | VEMO 150 | VEMO150 | |
| Finish | Servo-i | Servo-i | EVITA 4 | EVITA 4 | Servo-i | EVITA 4 | Servo-i |
| Ventilatory mode† | |||||||
| Beginning | AutoFlow | PSV | AutoFlow | PSV | PSV | AutoFlow | PCV |
| Hour 1 | VCV | PSV | VCV | PSV | PSV | PCV | |
| Hour 2 | VCV | PSV | VCV | PSV | PSV | PCV | |
| Hour 3 | VCV | PSV | PSV | PSV | VCV | PCV | |
| Finish | AutoFlow | PSV | AutoFlow | PSV | PSV | AutoFlow | PCV |
| Tidal volume (mL) | |||||||
| Beginning | 480 | 400 | 380 | 340 | 620 | 440 | 420 |
| Hour 1 | 480 | 420 | 380 | 350 | 540 | 410 | |
| Hour 2 | 480 | 405 | 380 | 360 | 500 | 380 | |
| Hour 3 | 480 | 410 | 370 | 550 | 440 | 420 | |
| Finish | 480 | 400 | 380 | 380 | 640 | 440 | 420 |
| Respiratory rate (mpm) | |||||||
| Beginning | 26 | 22 | 26 | 28 | 25 | 28 | 35 |
| Hour 1 | 26 | 22 | 26 | 26 | 24 | 35 | |
| Hour 2 | 26 | 22 | 26 | 26 | 24 | 35 | |
| Hour 3 | 26 | 22 | 25 | 24 | 28 | 35 | |
| Finish | 26 | 22 | 26 | 28 | 22 | 28 | 35 |
| Pmax (cmH2O) | |||||||
| Beginning | 26 | 20 | 33 | 14 | 18 | 31 | 34 |
| Hour 1 | 26 | 20 | 33 | 14 | 18 | 34 | |
| Hour 2 | 27 | 20 | 32 | 14 | 18 | 34 | |
| Hour 3 | 26 | 20 | 14 | 18 | 26 | 34 | |
| Finish | 27 | 20 | 33 | 14 | 18 | 30 | 34 |
| PEEP (cmH2O) | |||||||
| Beginning | 10 | 8 | 15 | 8 | 12 | 10 | 8 |
| Hour 1 | 10 | 8 | 15 | 8 | 12 | 8 | |
| Hour 2 | 10 | 8 | 15 | 8 | 12 | 8 | |
| Hour 3 | 10 | 8 | 8 | 12 | 10 | 8 | |
| Finish | 10 | 8 | 15 | 8 | 12 | 10 | 8 |
| Fio2 (%) | |||||||
| Beginning | 70% | 40% | 70% | 50% | 65% | 60% | 80% |
| Hour 1 | 70% | 40% | 70% | 53% | 68% | 85% | |
| Hour 2 | 70% | 40% | 72% | 52% | 68% | 86% | |
| Hour 3 | 70% | 40% | 52% | 67% | 48% | 83% | |
| Finish | 70% | 40% | 100% | 50% | 65% | 50% | 80% |
| Oxygen flow (L/min) | |||||||
| Beginning | |||||||
| Hour 1 | 16 | 8 | 16 | 12 | 20 | 15 | 18 |
| Hour 2 | 16 | 8 | 16 | 12 | 20 | 16 | |
| Hour 3 | 16 | 8 | 12 | 20 | 15 | 15 | |
| Finish | |||||||
| Pao2 (mmHg) | |||||||
| Beginning | 96.5 | 84.0 | 74.4 | 79.2 | 61.9 | 93.0 | 62.4 |
| Hour 1 | 89.4 | 87.9 | 79.0 | 75.2 | 77.0 | 58.1 | |
| Hour 2 | 91.7 | 80.4 | 69.7 | 85.8 | 79.7 | 59.6 | |
| Hour 3 | 85.8 | 94.1 | 78.1 | 85.2 | 84.3 | 53.3 | |
| Finish | 102.6 | 75.5 | 61.0 | 71.5 | 79.6 | 100.5 | 56.7 |
| Paco2 (mmHg) | |||||||
| Beginning | 58.1 | 44.3 | 58.4 | 65.4 | 35.5 | 52.3 | 65.8 |
| Hour 1 | 55.5 | 37.5 | 62.9 | 64.4 | 34.5 | 63.7 | |
| Hour 2 | 46.5 | 43.9 | 72.5 | 67.6 | 36.0 | 57.0 | |
| Hour 3 | 50.4 | 44.3 | 70.6 | 37.1 | 59.2 | 59.9 | |
| Finish | 57.0 | 37.3 | 64.4 | 65.3 | 38.6 | 53.2 | 45.1 |
| pH | |||||||
| Beginning | 7.40 | 7.34 | 7.37 | 7.32 | 7.43 | 7.34 | 7.40 |
| Hour 1 | 7.42 | 7.35 | 7.32 | 7.34 | 7.44 | 7.43 | |
| Hour 2 | 7.44 | 7.34 | 7.31 | 7.31 | 7.42 | 7.43 | |
| Hour 3 | 7.42 | 7.34 | 7.32 | 7.43 | 7.29 | 7.44 | |
| Finish | 7.40 | 7.39 | 7.33 | 7.34 | 7.42 | 7.27 | 7.47 |
| Bicarbonate (mmol/L) | |||||||
| Beginning | 32.5 | 22.9 | 32.9 | 32.9 | 23.4 | 27.6 | 40.5 |
| Hour 1 | 33.4 | 20.6 | 31.9 | 34.2 | 23.0 | 41.5 | |
| Hour 2 | 29.9 | 22.5 | 35.5 | 33.1 | 23.1 | 37.9 | |
| Hour 3 | 30.3 | 23.1 | 35.5 | 23.8 | 27.8 | 40.1 | |
| Finish | 32.5 | 22.1 | 33.4 | 34.9 | 23.4 | 23.9 | 39.1 |
Fio2 = inspired fraction of oxygen; mpm = movements per minute; NMBA = neuromuscular blocking agent; Paco2 = arterial partial pressure of carbon dioxide; Pao2 = arterial partial pressure of oxygen; PEEP = positive end-expiratory pressure; Pmax = maximum airway pressure; PSV = pressure support ventilation; VCV = volume controlled ventilation.
EVITA 4 (Soma Tech), Servo-i (Getinge), VEMO 150 (EOVE), VIVO 55 (Breas).
AutoFlow (Dräger).
The arterial partial pressure of oxygen (Pao2) values at the beginning of the intervention ranged from 61.9 mmHg to 96.5 mmHg and, at the end, from 56.7 mmHg to 102.6 mmHg (Table 1 and Figure 2). From the beginning and during the intervention period, the variation in Pao2 values ranged from 4.9% to 13.1%, with the highest increase being from 61.9 mmHg to 79.7 mmHg and the highest decrease from 96.5 mmHg to 85.8 mmHg.
Figure 2.
Temporal changes in arterial partial pressure of oxygen (Pao2)
Change in Pao2 at the beginning using intensive care unit (ICU) ventilators, then hourly for 3 hours using ventilators designed for support of patients with chronic respiratory insufficiency, and at the end returning to ICU ventilators.
No patient developed worse hypoxaemia during the intervention period. Ventilators designed for support of patients with chronic respiratory insufficiency were able to offer a fraction of inspired oxygen (Fio2) level from 40% to 86%, with oxygen flows ranging from 8 L/min to 20 L/min.
Similarly, Arterial partial pressure of carbon dioxide (Paco2) (coefficient of variation, 3.0–9.3%) and pH (coefficient of variation, 0.0–0.3%) levels remained stable during the test, and a considerably increase in the Paco2 level (from 58.4 mmHg to 72.5 mmHg) was observed in only one patient (Table 1). The patient developed fever and hypoxia unrelated to the intervention, which persisted even after return to the ICU ventilator.
Discussion
In this case series, seven critically ill patients with COVID-19 were ventilated with ventilators designed for support of patients with chronic respiratory insufficiency during a short period. As assessed by hourly arterial blood gases, Pao2, Paco2 and pH levels were kept constant during the intervention and only one patient developed worsening of the Paco2 level in the setting of ventilation unrelated to the intervention but related to high fever and clinical deterioration.
This proof-of-concept study has several limitations. First, this was a small case series that included no controls. Second, it is unclear if a longer period of ventilation would result in similar findings, although the findings for the first 3 hours are encouraging. Third, ventilation was provided by two specific ventilators; whether a different ventilator would have been associated with different outcomes cannot be determined.
Conclusion
In this uncontrolled case series of seven critically ill patients with COVID-19 and under mechanical ventilation, the use of ventilators designed for support of patients with sleep apnoea and/or chronic respiratory insufficiency was safe and did not result in worsening of hypoxaemia or hypercapnia. The limited sample size and study design preclude a definitive statement about the potential effectiveness of this treatment.
Competing interests
None declared.
References
- 1.Arentz M., Yim E., Klaff L., et al. Characteristics and outcomes of 21 critically ill patients with COVID-19 in Washington State. JAMA. 2020 doi: 10.1001/jama.2020.4326. [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Yang X., Yu Y., Xu J., et al. Clinical course and outcomes of critically ill patients with SARS-CoV-2 pneumonia in Wuhan, China: a single-centered, retrospective, observational study. Lancet Respir Med. 2020 doi: 10.1016/S2213-2600(20)30079-5. [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]