Current and future developments of synthetic computed tomography generation for radiotherapy

Wouter van Elmpt a,, Vicki Trier Taasti a, Kathrine Røe Redalen b
PMCID: PMC10696097  PMID: 38058591

The literature on generation of synthetic computed tomography (CT) based on magnetic resonance imaging (MRI) or cone-beam CT (CBCT) has been rapidly increasing over the last years [1], [2], [3], [4]. Recently, also commercial solutions for synthetic CT generation have been introduced, advancing adaptive radiotherapy workflows [5], [6], [7], [8], [9], [10], [11]. The interest in MRI-only workflows with synthetic CTs replacing the planning CT is also growing [12].

With the clinical introduction of synthetic CTs, there has been a shift in the literature. While the first papers in this field generally focused on image-based assessments only and mainly reported on CT number accuracy [13], [14], recent papers are focusing on evaluating the entire chain of the radiotherapy process [15], [16]. Linked to this, the time it takes for these algorithms to perform the corrections also becomes an important part of the evaluation. Besides image quality and geometrical correctness, a dose/volume-based evaluation by (re)calculation of treatment plans on synthetic CT scans is currently seen as one of the standard methods to quantify the accuracy of a synthetic CT generation [17], [18].

With the aim of establishing a patient-specific method for quality assurance, Dal Bello et al. [15] in this ESTRO 2023 paper highlight collection compared four commonly used strategies to create a synthetic CT from the MRI; ranging from a very simplistic assumption of overriding the entire patient inside the body contour with water to advanced correction methods based on deformable image registration (DIR) or deep-learning based neural networks. Creating a separate synthetic CT based on an independent neural network was found to provide an efficient and accurate validation method; however, for patients with metal implants the MR-based synthetic CTs were less reliable.

In another ESTRO 2023 highlight paper, Texier et al. stressed the need for using multi-center data also in the training step of the deep-learning models [16]. They performed an investigation using data from four different centers to create a synthetic CT model for MRI. Here the generalizability was poor if the model was applied on scans from a different site that was not included in the model. This is important to realize when robust deep-learning synthetic CT models are intended to be used across multiple MRI scanners, protocols, or institutes.

For CBCT based adaptive therapy, the synthetic CT has also shown to be a powerful tool in the evaluation of treatment quality. Typically, the image quality of CBCT is insufficient for accurate dose calculation. Synthetic CT approaches have shown to be able to allow for dosimetric assessment of treatment planning for example in the lung [19], head-and-neck [20], or simulating prostate motion [21]. Also, novel approaches using the diagnostic CT in combination with the CBCT in a palliative setting has been explored [22].

However, to be able to safely use synthetic CTs in an (online) adaptive workflow, it is important to ensure that the synthetic CTs represent the geometry of the patient correctly. With frequently used DIR-based algorithms, this geometrical accuracy could be investigated and steered by the regularization of the deformation vector field. Currently, many CBCT-based synthetic CT generation methods are based on deep-learning which are trained on both planning CT and CBCT scans which may not represent the same anatomy of the patient. It should therefore be evaluated if the synthetic CT method only improves the image quality of the CBCT without introducing anatomical differences. In a study by de Hond et al., it was found that among three deep-learning synthetic CT methods for CBCT conversion the method giving the best image quality performance, based on mean absolute error of CT numbers, was different from the method giving the best anatomical performance, based on organ-at-risk volumes and average surface distance [23].

Furthermore, for moving tumors the treatment plan is often based on a 4D-CT, whereas in the treatment room typically only 2D or 3D imaging is available. Deep-learning methods have been proposed to generate 3D-sCTs based on 2D radiographs [24], or even to generate 4D-sCTs based on 3D-CBCTs [25]. The integration in a clinical routine workflow is however not yet fully investigated.

For proton therapy, MRI-only treatment planning is also advantageous in order to bypass the MR-CT co-registration and to benefit from the improved soft tissue contrast. However, MRI alone cannot provide stopping-power ratio (SPR) information. Recognizing that dual-energy CT can estimate SPR better than single-energy CT, Liu et al. [26] demonstrated a new method using deep-learning to create synthetic dual-energy CT from MRI to calculate SPR. Hence, it is likely that synthetic CTs will find their way into both photon- and proton therapy in the years to come, closely linked with development of novel artificial intelligence algorithms.

The commercial synthetic CT solutions often originate from developments of in-house scripts, finally being translated to clinical innovation. It is yet another example of how physics-driven research result in new technological solutions for the benefit of further personalization and improvements of cancer treatment. It can be anticipated that the current hurdles with standardization and validation will be solved. At the moment, the preferred validation is ideally based on time-consuming collection of data from multiple institutions to make the models robust. In the future, either federated learning solutions, or other novel ways could be envisioned through an integrated approach of creating unlimited privacy-safe synthetic images for the purpose of validation.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

  • 1.Boulanger M., Nunes J.-C., Chourak H., Largent A., Tahri S., Acosta O., et al. Deep learning methods to generate synthetic CT from MRI in radiotherapy: A literature review. Phys Med. 2021;89:265–281. doi: 10.1016/j.ejmp.2021.07.027. [DOI] [PubMed] [Google Scholar]
  • 2.Spadea M.F., Maspero M., Zaffino P., Seco J. Deep learning based synthetic-CT generation in radiotherapy and PET: A review. Med Phys. 2021;48:6537–6566. doi: 10.1002/mp.15150. [DOI] [PubMed] [Google Scholar]
  • 3.Giacometti V., Hounsell A.R., McGarry C.K. A review of dose calculation approaches with cone beam CT in photon and proton therapy. Phys Med. 2020;76:243–276. doi: 10.1016/j.ejmp.2020.06.017. [DOI] [PubMed] [Google Scholar]
  • 4.Taasti V.T., Klages P., Parodi K., Muren L.P. Developments in deep learning based corrections of cone beam computed tomography to enable dose calculations for adaptive radiotherapy. Phys Imaging Radiat Oncol. 2020;15:77–79. doi: 10.1016/j.phro.2020.07.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Sibolt P., Andersson L.M., Calmels L., Sjöström D., Bjelkengren U., Geertsen P., et al. Clinical implementation of artificial intelligence-driven cone-beam computed tomography-guided online adaptive radiotherapy in the pelvic region. Phys Imaging Radiat Oncol. 2021;17:1–7. doi: 10.1016/j.phro.2020.12.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Byrne M., Archibald-Heeren B., Hu Y., Teh A., Beserminji R., Cai E., et al. Varian ethos online adaptive radiotherapy for prostate cancer: Early results of contouring accuracy, treatment plan quality, and treatment time. J Applied Clin Med Phys. 2022:23:e13479. doi: 10.1002/acm2.13479. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Kisling K., Keiper T.D., Branco D., Kim G.G., Moore K.L., Ray X. Clinical commissioning of an adaptive radiotherapy platform: Results and recommendations. J Applied Clin Med Phys. 2022;23 doi: 10.1002/acm2.13801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Palacios M.A., Gerganov G., Cobussen P., Tetar S.U., Finazzi T., Slotman B.J., et al. Accuracy of deformable image registration-based intra-fraction motion management in Magnetic Resonance-guided radiotherapy. Phys Imaging Radiat Oncol. 2023;26 doi: 10.1016/j.phro.2023.100437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Van de Schoot A.J.A.J., Van den Wollenberg W., Carbaat C., De Ruiter P., Nowee M.E., Pos F., et al. Evaluation of plan quality in radiotherapy planning with an MR-linac. Phys Imaging Radiat Oncol. 2019;10:19–24. doi: 10.1016/j.phro.2019.04.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Palmér E., Karlsson A., Nordström F., Petruson K., Siversson C., Ljungberg M., et al. Synthetic computed tomography data allows for accurate absorbed dose calculations in a magnetic resonance imaging only workflow for head and neck radiotherapy. Phys Imaging Radiat Oncol. 2021;17:36–42. doi: 10.1016/j.phro.2020.12.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lerner M., Medin J., Jamtheim Gustafsson C., Alkner S., Siversson C., Olsson L.E. Clinical validation of a commercially available deep learning software for synthetic CT generation for brain. Radiat Oncol. 2021;16:66. doi: 10.1186/s13014-021-01794-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Bird D., Henry A.M., Sebag-Montefiore D., Buckley D.L., Al-Qaisieh B., Speight R. A Systematic Review of the Clinical Implementation of Pelvic Magnetic Resonance Imaging-Only Planning for External Beam Radiation Therapy. Int J Radiat Oncol Biol Phys. 2019;105:479–492. doi: 10.1016/j.ijrobp.2019.06.2530. [DOI] [PubMed] [Google Scholar]
  • 13.Emami H., Dong M., Nejad-Davarani S.P., Glide-Hurst C.K. Generating synthetic CTs from magnetic resonance images using generative adversarial networks. Med Phys. 2018;45:3627–3636. doi: 10.1002/mp.13047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Yuan N., Dyer B., Rao S., Chen Q., Benedict S., Shang L., et al. Convolutional neural network enhancement of fast-scan low-dose cone-beam CT images for head and neck radiotherapy. Phys Med Biol. 2020;65 doi: 10.1088/1361-6560/ab6240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Dal Bello R., Lapaeva M., La Greca S.-E., Wallimann P., Günther M., Konukoglu E., et al. Patient-specific quality assurance strategies for synthetic computed tomography in magnetic resonance-only radiotherapy of the abdomen. Phys Imaging Radiat Oncol. 2023;27 doi: 10.1016/j.phro.2023.100464. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Texier B., Hémon C., Lekieffre P., Collot E., Tahri S., Chourak H., et al. MR-to-CT synthesis using 2D cycle-GAN with multicenter learning. Phys Imaging. Radiat Oncol. 2023:100511. doi: 10.1016/j.phro.2023.100511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Thing R.S., Nilsson R., Andersson S., Berg M., Lund M.D. Evaluation of CBCT based dose calculation in the thorax and pelvis using two generic algorithms. Phys Med. 2022;103:157–165. doi: 10.1016/j.ejmp.2022.10.012. [DOI] [PubMed] [Google Scholar]
  • 18.Wyatt J.J., Kaushik S., Cozzini C., Pearson R.A., Petit S., Capala M., et al. Comprehensive dose evaluation of a Deep Learning based synthetic Computed Tomography algorithm for pelvic Magnetic Resonance-only radiotherapy. Radiother Oncol. 2023;184 doi: 10.1016/j.radonc.2023.109692. [DOI] [PubMed] [Google Scholar]
  • 19.Taasti V.T., Hattu D., Peeters S., Van der Salm A., Van Loon J., De Ruysscher D., et al. Clinical evaluation of synthetic computed tomography methods in adaptive proton therapy of lung cancer patients. Phys Imaging Radiat Oncol. 2023;27 doi: 10.1016/j.phro.2023.100459. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Allen C., Yeo A.U., Hardcastle N., Franich R.D. Evaluating synthetic computed tomography images for adaptive radiotherapy decision making in head and neck cancer. Phys Imaging Radiat Oncol. 2023;27 doi: 10.1016/j.phro.2023.100478. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Feng S.-Q., Brouwer C.L., Korevaar E.W., Vapiwala N., Kang-Hsin Wang K., Deville C., et al. Dose evaluation of inter- and intra-fraction prostate motion in extremely hypofractionated intensity-modulated proton therapy for prostate cancer. Phys Imaging Radiat Oncol. 2023;27 doi: 10.1016/j.phro.2023.100474. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kejda A., Quinn A., Wong S., Lowe T., Fent I., Gargett M., et al. Evaluation of the clinical feasibility of cone-beam computed tomography guided online adaption for simulation-free palliative radiotherapy. Phys Imaging Radiat Oncol. 2023;28 doi: 10.1016/j.phro.2023.100490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.De Hond Y.J.M., Kerckhaert C.E.M., Van Eijnatten M.A.J.M., Van Haaren P.M.A., Hurkmans C.W., Tijssen R.H.N. Anatomical evaluation of deep-learning synthetic computed tomography images generated from male pelvis cone-beam computed tomography. Phys Imaging Radiat Oncol. 2023;25 doi: 10.1016/j.phro.2023.100416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Loÿen E., Dasnoy-Sumell D., Macq B. Patient-specific three-dimensional image reconstruction from a single X-ray projection using a convolutional neural network for on-line radiotherapy applications. Phys Imaging Radiat Oncol. 2023;26 doi: 10.1016/j.phro.2023.100444. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Schmitz H., Thummerer A., Kawula M., Lombardo E., Parodi K., Belka C., et al. ScatterNet for projection-based 4D cone-beam computed tomography intensity correction of lung cancer patients. Phys Imaging Radiat Oncol. 2023;27 doi: 10.1016/j.phro.2023.100482. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Liu R., Lei Y., Wang T., Zhou J., Roper J., Lin L., et al. Synthetic dual-energy CT for MRI-only based proton therapy treatment planning using label-GAN. Phys Med Biol. 2021;66 doi: 10.1088/1361-6560/abe736. [DOI] [PubMed] [Google Scholar]

Articles from Physics and Imaging in Radiation Oncology are provided here courtesy of Elsevier

RESOURCES