Abstract
Background
With the emergence of photon-counting CT, ultrahigh-resolution (UHR) imaging can be performed without dose penalty. This study aims to directly compare the image quality of UHR and standard resolution (SR) scan mode in femoral artery angiographies.
Methods
After establishing continuous extracorporeal perfusion in four fresh-frozen cadaveric specimens, photon-counting CT angiographies were performed with a radiation dose of 5 mGy and tube voltage of 120 kV in both SR and UHR mode. Images were reconstructed with dedicated convolution kernels (soft: Body-vascular (Bv)48; sharp: Bv60; ultrasharp: Bv76). Six radiologists evaluated the image quality by means of a pairwise forced-choice comparison tool. Kendall’s concordance coefficient (W) was calculated to quantify interrater agreement. Image quality was further assessed by measuring intraluminal attenuation and image noise as well as by calculating signal-to-noise ratio (SNR) and contrast-to-noise ratios (CNR).
Results
UHR yielded lower noise than SR for identical reconstructions with kernels ≥ Bv60 (p < 0.001). UHR scans exhibited lower intraluminal attenuation compared to SR (Bv60: 406.4 ± 25.1 versus 418.1 ± 30.1 HU; p < 0.001). Irrespective of scan mode, SNR and CNR decreased while noise increased with sharper kernels but UHR scans were objectively superior to SR nonetheless (Bv60: SNR 25.9 ± 6.4 versus 20.9 ± 5.3; CNR 22.7 ± 5.8 versus 18.4 ± 4.8; p < 0.001). Notably, UHR scans were preferred in subjective assessment when images were reconstructed with the ultrasharp Bv76 kernel, whereas SR was rated superior for Bv60. Interrater agreement was high (W = 0.935).
Conclusions
Combinations of UHR scan mode and ultrasharp convolution kernel are able to exploit the full image quality potential in photon-counting CT angiography of the femoral arteries.
Relevance statement
The UHR scan mode offers improved image quality and may increase diagnostic accuracy in CT angiography of the peripheral arterial runoff when optimized reconstruction parameters are chosen.
Key points
• UHR photon-counting CT improves image quality in combination with ultrasharp convolution kernels.
• UHR datasets display lower image noise compared with identically reconstructed standard resolution scans.
• Scans in UHR mode show decreased intraluminal attenuation compared with standard resolution imaging.
Graphical Abstract
Supplementary Information
The online version contains supplementary material available at 10.1186/s41747-023-00398-x.
Keywords: CT angiography, Femoral arteries, Photon-counting computed tomography (CT), Small pixel effect, Ultrahigh resolution
Background
Over the past decade, conventional energy-integrating detector CT has been advanced to its presently conceivable physical limits in terms of dose efficiency and spatial resolution [1]. Particularly with respect to pixel size, constructional restrictions essentially prevent further improvement [2]. Since incident photons are first converted into light before being measured by the diode as a detector signal, opaque septa are necessary to prevent crosstalk between the sensors. Further reduction in pixel size would lead to an increase in dead space, severely limiting dose efficiency [3]. To increase the spatial resolution nevertheless, most conventional scanners employ an aperture-narrowing comb or grid filter positioned in front of the detector and thus behind the patient for ultrahigh resolution (UHR) imaging. In consequence, photons proportional to the incoming x-ray beam are absorbed without contributing to the actual image despite passing through the patient and thus increasing the patient dose [4, 5]. Even a recently released CT scanner equipped with a 0.25 × 160-row detector and detector-sided UHR filter could not overcome the dose disadvantage of UHR imaging entirely [6].
The latest photon-counting detector CT (PCD-CT) technology, however, offers the option of a dose-neutral UHR mode due to the novel detector design without the need for a radiopaque interlayer [7, 8]. Since the conversion of photons into electrical energy occurs in the semiconductor material without the intermediate step of conversion to visible light, septa are no longer required. Hereby, smaller subpixel readout becomes feasible resulting in an effective pixel size of 150 × 176 µm in the isocenter and thereby a relevant increase in spatial resolution [3, 7, 9, 10]. While 4 subpixels are read out together as one unit in standard resolution (SR) mode — a process known as 2 × 2 binning with a resulting z-collimation of 140 rows × 0.4 mm detector pixels — the subpixels are read out separately to acquire images in UHR mode. Due to the resulting increase in data volume, however, the z-collimation is limited to 120 rows × 0.2 mm [7, 11–13]. For cardiac [14–19], pulmonary [20, 21], and musculoskeletal imaging [22–25], several advantages of the UHR scan mode have already been demonstrated. Benefits for imaging of the vascular periphery can be surmised from these data, especially superior delineation of narrow caliber or severely calcified vessels. While this is expected to increase the diagnostic accuracy and confidence, primarily in the lower leg arteries, the actual extent of this effect in clinical routine is yet unclear.
To investigate the UHR mode in direct comparison with the SR mode in photon-counting CT angiography (CTA) of the femoral arteries, this study introduces a cadaveric in vitro model with continuous extracorporeal perfusion for objective and subjective image quality assessment.
Methods
Human cadaver in vitro phantom
This study was performed in accordance with applicable law and the local ethics committee approved the experimental design. To establish the in vitro model, 4 cadaveric specimens (2 men, 2 women) were obtained from the local institute of anatomy. Body donors had previously provided informed consent to the use of their cadavers for research and teaching purposes. Of note, 3 of 4 cadavers were used for experiments in a previous study [26]. The femoral vessels of each specimen were continuously perfused with a 13 mg/ml iodine contrast medium solution to achieve a targeted intraluminal attenuation of 400 HU with a standard 120-kV scan in SR mode [27]. Perfusion was established via vascular accesses in the groin and below the knee by means of a pump circuit and examined under flux to simulate a standard pelvic-leg CTA. Details of the experimental setup and the preparation of the cadaveric specimens can be found elsewhere [28].
Scan protocol and image reconstruction
Upper leg runoff CTA were performed using a clinical PCD-CT (Naeotom Alpha; Siemens Healthcare GmbH, Forchheim, Germany). CT scans were acquired with a consistent radiation dose of 5 mGy and a tube voltage of 120 kV, resulting in an effective tube current of 63 mAs. Both SR (z-collimation 144 × 0.4 mm) and UHR mode (120 × 0.2 mm) were utilized without repositioning of specimens. The pitch factor was set to 0.9 for UHR and 0.4 for SR examinations to match the scan duration between acquisition modes. Of note, the helical pitch is a dimensionless factor depending on the collimation width, which is 57.6 mm for SR versus 24 mm for UHR mode. Rotation time was set to 0.5 s.
For each leg runoff, individual polyenergetic reconstructions were prepared with a small field of view measuring 150 mm and a slice thickness/increment of 1.0 mm, respectively. Three different vascular convolution kernels were applied to reformat the images, listed in increasing order of sharpness: Body-vascular-(Bv)48 (soft; ρ50 = 5.40 lp/cm), Bv60 (sharp; ρ50 = 8.79 lp/cm), and Bv76 (ultrasharp; ρ50 = 16.47 lp/cm). A 5122 pixel matrix was employed for Bv48 and Bv60, while the matrix for the Bv76 contained 7682 pixels according to the default scanner settings. In each case, a 4th-generation iterative reconstruction algorithm at strength level 3 was employed (QIR, Siemens). Since the lowest energy threshold was set to 20 keV by default, images only incorporate x-ray quanta above said threshold in order to eliminate electronic noise.
Image quality analysis
Four arterial levels (proximal, middle, distal superficial femoral artery, and popliteal artery) were evaluated. Intraluminal attenuation, as well as the attenuation of the surrounding muscle and subcutaneous fat tissue, were measured using standardized regions of interest with a size of 30 mm2. Standard deviation of signal attenuation within the unenhancing lipid tissue was deemed representative of image noise. A radiologist with 5 years of experience in CTA imaging performed all measurements using clinical PACS software (Merlin; Phönix-PACS, Freiburg, Germany).
Signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated as and . Image stacks were compared using an in-house programmed forced-choice pairwise comparison tool. Axial-oriented image slices of both legs at the mid-level of the superficial femoral artery were presented, resulting in 48 images and 120 one-on-one comparisons. Six radiologists, including three residents with 1 to 5 years and three board-certified radiologists with 7 to 8 years of clinical training, independently evaluated the luminal assessability and delineation of calcifications using a certified diagnostic monitor (RadiForce RX660; EIZO, Hakusan, Japan) while blinded to any protocol-related information. Data were processed in a Jupyter Notebook environment with Python version 3.8.15 using a Bradley-Terry Model [29] to allow for hierarchical ranking of the compared subgroups, ranging from 1 (best) to 6 (worst) in terms of subjective image quality.
Statistics
Data is presented as mean ± standard deviation unless differently specified. Statistical analyses were conducted with specialized software (DATAtab e.U., Graz, Austria). Continuous variables are reported as mean ± standard deviation. Significance was determined at an alpha level of p < 0.05. Objective image parameters were evaluated with Bonferroni-corrected t-tests. Subjective image quality ratings were compared using Friedman tests, and preference ranks were obtained by fitting a Bradley-Terry Model for the entire dataset and for each rater individually [29]. Bump chart diagrams and Boxplots were generated for visualization purposes. Inter-reader agreement was measured using Kendall’s concordance coefficient (W).
Results
Intraluminal attenuation remained constant between scan modes regardless of the reconstruction kernel used (p > 0.999). However, the intraluminal attenuation of scans performed in UHR mode was lower than in the SR mode despite identical tube voltage settings (e.g., Bv60: 406.4 ± 25.1 HU versus 418.1 ± 30.1 HU; p < 0.001). Image noise increased with sharper reconstruction kernels in all comparisons (e.g., concerning intraluminal measurement, SR: Bv48 11.2 ± 2.1 HU versus Bv76 75.4 ± 16.2 HU). However, the image noise associated with sharp and ultrasharp reconstructions (Bv60/76) was significantly lower when using the UHR instead of the SR scan mode for image acquisition (e.g., for intraluminal measurement, UHR: Bv76 50.2 ± 11.7 HU versus SR: 75.4 ± 16.2 HU; p < 0.001). Intraluminal signal attenuation and standard deviation thereof are summarized in Fig. 1, while a comprehensive display is provided in Table 1. Figure 2 contains representative axial image slices acquired in UHR and SR mode and reconstructed with the three different convolution kernels.
Table 1.
Scan Mode/Kernel | Attlum [HU] | SD Attlum [HU] | SD Attmusc [HU] | SD Attfat [HU] | SNR | CNR | |
---|---|---|---|---|---|---|---|
SR | Bv48 | 419.9 ± 25.6 | 11.2 ± 2.1 | 11.1 ± 1.8 | 11.1 ± 1.1 | 38.3 ± 4.7 | 33.7 ± 4.4 |
Bv60 | 418.1 ± 30.1 | 21.8 ± 3.6 | 23.5 ± 12.7 | 21.8 ± 5.2 | 20.9 ± 5.3 | 18.4 ± 4.8 | |
Bv76 | 421.9 ± 27.9 | 75.4 ± 16.2 | 73.9 ± 12.7 | 69.2 ± 15.9 | 6.6 ± 1.7 | 5.8 ± 1.5 | |
UHR | Bv48 | 404.5 ± 24.6 | 10.1 ± 2.4 | 10.7 ± 1.4 | 10.8 ± 1.1 | 38.0 ± 4.8 | 33.4 ± 4.4 |
Bv60 | 406.4 ± 25.1 | 17.2 ± 3.4 | 19.5 ± 3.8 | 17.2 ± 4.2 | 25.9 ± 6.4 | 22.7 ± 5.8 | |
Bv76 | 401.9 ± 26.1 | 50.2 ± 11.7 | 49.9 ± 8.9 | 45.6 ± 11.6 | 9.7 ± 2.6 | 8.5 ± 2.3 |
Attfat Attenuation of fat tissue, Attlum Intraluminal attenuation, Attmusc Attenuation of muscle tissue, CNR Contrast-to-noise-ratio, HU Hounsfield units, SD Standard deviation, SNR Signal-to-noise-ratio, SR Standard resolution, UHR Ultrahigh resolution
CNR in SR and UHR mode did not significantly differ when the soft convolution kernel Bv48 was used for image reconstruction (UHR: 33.4 ± 4.4 versus SR: 33.7 ± 4.4; p = 0.722). However, CNR was significantly higher for UHR acquisitions when the sharp kernel Bv60 (UHR: 22.7 ± 5.8 versus SR: 18.4 ± 4.8; p < 0.001) or the ultrasharp kernel Bv76 were employed (UHR: 8.5 ± 2.3 versus SR: 5.8 ± 1.5; p < 0.001). A similar tendency was observed for SNR, which was significantly higher for reconstructions of UHR data with the sharp and ultrasharp kernels (p < 0.001, each), while no significant difference was observed for reconstructions with the soft kernel (p = 0.799). CNR and SNR results are summarized in Table 1 and visualized in Fig. 3.
Regardless of scan mode, sharp reconstructions were deemed superior to images generated with soft kernels. The UHR images reconstructed with the ultrasharp kernel Bv76 received the highest observer ratings, followed by SR images reconstructed with the same kernel. While a measurable improvement in CNR/SNR was recorded for UHR mode acquisitions when the sharp kernel Bv60 was used for image reconstruction, the raters generally preferred the SR mode. When the soft reconstruction kernel Bv48 was used, no preference was observed between SR and UHR mode. Inter-reader agreement was high (W = 0.935) with individual rankings visualized in form of a bump chart in Fig. 4. Ranking differences were significant in all cases (p < 0.001).
Discussion
This experimental study on four continuously perfused human cadaver models evaluated the influence of standard and ultrahigh-resolution scan mode on image quality in photon-counting CT angiographies of the femoral arteries dependent on the reconstruction kernel used under standardized conditions. Our results indicate the superiority of a combination of ultrahigh-resolution scan mode and ultrasharp image reconstruction based on quantitative assessment and multiobserver analysis in a pairwise forced-choice setup.
A recent investigation was able to demonstrate the dose reduction and image quality optimization potential of PCD-CT in SR mode compared to energy-integrating detector CT using similar scanner settings [28]. Of note, the authors stated that the SR scan mode was deliberately used instead to achieve maximal comparability. While most radiation dose-defining settings (e.g., tube voltage, effective mAs) were chosen in accordance with this aforementioned study, we consciously selected a single radiation dose level of 5 mGy for the present work, which was deemed appropriate for clinical use and is below the applicable national guideline value. In contrast to prior studies on CTA of the vascular periphery, this research, to the authors’ best knowledge, is the first to provide direct side-by-side comparisons between UHR and SR scan mode in a fully standardized in vitro environment.
Notably, the choice of reconstruction kernel had no significant influence on intraluminal attenuation. However, somewhat surprisingly, the mean intraluminal attenuation at the same tube voltage (120 kV) was recorded to be significantly lower in UHR mode than in SR mode. Potential dilution effects during the cadaver perfusion, e.g., due to extravasation of contrast medium, could be ruled out in a supplementary experiment, in which cylindrical sample tubes holding different concentrations of iodine solution were scanned with the same acquisition settings (Supplemental Fig. S1). For this reason, a physical cause of the observed difference is conceivable, but more studies are needed to evaluate this effect further.
Consistent with the literature on cardiac PCD-CT, overall image noise increased with higher spatial frequency [14, 16, 18]. However, in accordance with a previous study by Leng et al. [10], this increase in noise could be significantly reduced by scanning in UHR mode. Even though SNR and CNR were numerically lowest for reconstructions with the ultrasharp Bv76 kernel, observers preferred this post-processing setting, especially if the reconstructed data was acquired in UHR mode. This finding is generally in line with recent musculoskeletal imaging studies, which also postulated the best subjective results for the sharpest investigated kernels [30, 31]. Within each scan mode, images reconstructed with sharper kernels were subjectively rated better in the forced-choice comparison setup. This tendency may be explained by a more detailed visualization of the vessel wall and thus superior delineation of calcified and noncalcified plaques. Also allowing for reduced image noise, image acquisition in UHR mode holds potential to increase the diagnostic accuracy of CTA, particularly in narrow-caliber and/or circularly calcified vessels, e.g., in advanced peripheral arterial disease [32, 33]. In that regard, promising results have already been demonstrated in cardiac CT [14–17] and phantom studies [19, 34, 35]. While SR mode was preferred over UHR mode for reconstructions with the sharp Bv60 kernel, we attribute this observation to the image impression more closely resembling the familiar output of EID-CT scanners used in our department. Although CTA is usually focused on the depiction of contrasted vessels, it must be mentioned that the assessability of the adjacent soft tissues is reduced with increasing spatial frequency of the convolution kernel. Therefore, reconstructions with more than one kernel may be necessary for a comprehensive evaluation of the vascular runoff and surrounding soft tissue. In that regard, a sharp kernel (e.g., Bv64) may serve as a compromise [16]. Since the pixel matrix also has considerable influence on the image noise level [36], the vendor-specific default settings were used in the present study.
The primary reason for the recorded noise reduction in UHR mode lies presumably in the “small pixel effect,” a result of the smaller detector pixel size and therefore downscaled focal spot in fan direction [37]. On the reconstruction side, in order to realize a certain degree of sharpness, the convolution kernel dampens the enhancement of high frequencies, resulting in a reduced noise increase. On one side, this phenomenon can be exploited to produce scans with a lower radiation dose while maintaining the same noise level compared to SR data. On the other side, the small pixel effect allows for noise reduction with the same radiation dose [7, 38]. Due to the predominantly elderly population receiving pelvic-leg CTA, better image quality may be preferred over dose reduction in these patients. Presumably, more precise radiological assessment can lead to a reduction of potentially unnecessary invasive examinations in clinical routine. In this context, virtual low keV reconstructions facilitating contrast agent reduction are also of high interest and warrant further studies.
For the present investigation, image reconstructions were deliberately performed with a slice thickness of 1 mm. In principle, a lower slice thickness can be selected, but the diagnostic gain is questionable due to increasing image noise. In support, a patient study by Milos et al. [39] comparing PCD-CT of the lung in UHR mode with different convolution kernels showed that image noise increases significantly with decreasing slice thickness. While this study did not provide a control group with scans in SR mode, it can be assumed that the observed indirect relationship between image noise and slice thickness is at least equally pronounced in SR mode. With regard to the present work, the relative noise reduction in UHR mode may be even more pronounced at lower reconstructed slice thicknesses, benefitting especially the diagnostic evaluation of the lower leg runoff.
UHR scan mode should next be evaluated with respect to diagnostic accuracy, especially for small vessel diameters (e.g., in the lower leg) and in the presence of high-grade calcifications. Based on a recent cardiac CT study by Rajendran et al. [40], the postulated advantages of UHR examinations may also be translatable to arteriosclerotic alterations in peripheral vessels. While additional comparative studies are mandated, first publications recommend the use of UHR scan mode in principle for all PCD-CT examinations [41]. In the context of smaller scan volumes, e.g., in cardiac or musculoskeletal imaging, this statement may be valid, but the considerably increased raw data file size of UHR compared to SR acquisitions is certainly relevant for larger scan volumes at this time [7]. Of note, file size has become a pertinent cost factor both in terms of computation time and energy consumption, which is to be seen critically considering climate change and the overall need for “greener” medicine [42].
In addition to being a single-center, single-scanner investigation, some limitations of this study must be mentioned. First, due to the experimental and laborious study design, only 8 lower extremities of 4 body donors were examined. With the current design of the perfusion model, only large-lumen femoral arteries could be examined, however, the largest diagnostic benefit of UHR scan mode can certainly be expected for the narrow caliber arteries of the lower leg. Second, to increase comparability and provide a clinically applicable imaging setup, we employed a single tube voltage of 120 kV and constant CT dose index of 5.0 mGy, while reconstructing all data with a slice thickness/increment of 1.0 mm. To maintain realistic scan durations among acquisition modes, the spiral pitch factor had to be increased for UHR studies. Third, the implications of using lower tube voltages were not investigated. Similarly, spectral post-processing applications, such as virtual monoenergetic imaging, were not in the scope of the present study. Further studies are necessary in that regard.
To conclude, in photon-counting CT angiographies of the femoral arteries, reconstructing data with an ultrasharp convolution kernel is necessary to exploit the full image quality potential of the scanner’s ultrahigh-resolution scan mode.
Supplementary Information
Abbreviations
- Bv
Body vascular (convolution kernel)
- CNR
Contrast-to-noise ratio
- CTA
Computed tomography angiography
- PCD-CT
Photon-counting detector computed tomography
- SNR
Signal-to-noise ratio
- SR
Standard resolution
- UHR
Ultrahigh resolution
Authors’ contributions
PG designed the study, analyzed all data, and prepared the manuscript. RH programmed and supervised the pairwise comparison software. VH, HH, RH, ASK, PP, and AMA performed the image quality assessment. HH, VH, PF, AMA, SL, DP, and JPG supported the preparation of the manuscript and figures. PG and PF performed statistical analyses and revised the manuscript. TAB, ASK, PF, HH, and JPG contributed to the preparation of the manuscript and provided quality control. SE, DP, and SL prepared the cadaveric specimens. JPG supervised the study. All authors read and approved the final manuscript.
Funding
Open Access funding enabled and organized by Projekt DEAL. PG (Z-02CSP/18) and JPG (Z-3BC/02) were financially supported by the Interdisciplinary Center of Clinical Research Würzburg. The current investigation was further supported by the German Research Foundation. This publication was supported by the Open Access Publication Fund of the University of Wuerzburg and the DEAL project.
Availability of data and materials
The datasets generated and/or analyzed during the current study are not publicly available but are available from the corresponding author on reasonable request. Due to the nature of this research, participants of this study did not agree to their data to be shared in a public repository.
Declarations
Ethics approval and consent to participate
This study was performed in accordance with applicable law after obtaining approval from the institutional review board of the University of Würzburg (protocol number: 20220413 01). Body donors had previously provided written informed consent to the use of their cadavers for research and teaching purposes.
Consent for publication
Not applicable.
Competing interests
Three of the authors (JPG, ASK, and TAB) receive speaker honoraria from Siemens Healthcare GmbH. The other authors declare no relationships with any companies related to the matter of the article.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.McCollough CH, Leng S, Yu L, et al. Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology. 2015;276:637–653. doi: 10.1148/radiol.2015142631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lell MM, Wildberger JE, Alkadhi H, et al. Evolution in computed tomography: the battle for speed and dose. Invest Radiol. 2015;50:629–644. doi: 10.1097/RLI.0000000000000172. [DOI] [PubMed] [Google Scholar]
- 3.Willemink MJ, Persson M, Pourmorteza A, et al. Photon-counting CT: technical principles and clinical prospects. Radiology. 2018;289:293–312. doi: 10.1148/radiol.2018172656. [DOI] [PubMed] [Google Scholar]
- 4.Flohr TG, Stierstorfer K, Süss C, et al. Novel ultrahigh resolution data acquisition and image reconstruction for multi-detector row CT. Med Phys. 2007;34:1712–1723. doi: 10.1118/1.2722872. [DOI] [PubMed] [Google Scholar]
- 5.Esquivel A, Ferrero A, Mileto A, et al. Photon-counting detector CT: key points radiologists should know. Korean J Radiol. 2022;23:854–865. doi: 10.3348/kjr.2022.0377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Kawashima H, Ichikawa K, Takata T, et al. Technical note: performance comparison of ultra-high-resolution scan modes of two clinical computed tomography systems. Med Phys. 2020;47:488–497. doi: 10.1002/mp.13949. [DOI] [PubMed] [Google Scholar]
- 7.Lell M, Kachelrieß M. Computed tomography 2.0: new detector technology, ai, and other developments. Invest Radiol. 2023;58:587–601. doi: 10.1097/RLI.0000000000000995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Leng S, Yu Z, Halaweish A, et al. Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system. J Med Imaging (Bellingham) 2016;3:43504. doi: 10.1117/1.JMI.3.4.043504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Meloni A, Frijia F, Panetta D et al. (2023) Photon-counting computed tomography (PCCT): technical background and cardio-vascular applications. Diagnostics (Basel) 13. 10.3390/diagnostics13040645 [DOI] [PMC free article] [PubMed]
- 10.Leng S, Rajendran K, Gong H, et al. 150-μm spatial resolution using photon-counting detector computed tomography technology: technical performance and first patient images. Invest Radiol. 2018;53:655–662. doi: 10.1097/RLI.0000000000000488. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Alkadhi H, Euler A. The future of computed tomography: personalized, functional, and precise. Invest Radiol. 2020;55:545–555. doi: 10.1097/RLI.0000000000000668. [DOI] [PubMed] [Google Scholar]
- 12.Wildberger JE, Alkadhi H. New horizons in vascular imaging with photon-counting detector CT. Invest Radiol. 2023;58:499–504. doi: 10.1097/RLI.0000000000000957. [DOI] [PubMed] [Google Scholar]
- 13.Zsarnóczay E, Varga-Szemes A, Emrich T, et al. Characterizing the heart and the myocardium with photon-counting CT. Invest Radiol. 2023;58:505–514. doi: 10.1097/RLI.0000000000000956. [DOI] [PubMed] [Google Scholar]
- 14.Mannil M, Hickethier T, von Spiczak J, et al. Photon-counting CT: high-resolution imaging of coronary stents. Invest Radiol. 2018;53:143–149. doi: 10.1097/RLI.0000000000000420. [DOI] [PubMed] [Google Scholar]
- 15.von Spiczak J, Mannil M, Peters B, et al. Photon counting computed tomography with dedicated sharp convolution kernels: tapping the potential of a new technology for stent imaging. Invest Radiol. 2018;53:486–494. doi: 10.1097/RLI.0000000000000485. [DOI] [PubMed] [Google Scholar]
- 16.Si-Mohamed SA, Boccalini S, Lacombe H, et al. Coronary CT angiography with photon-counting CT: first-in-human results. Radiology. 2022;303:303–313. doi: 10.1148/radiol.211780. [DOI] [PubMed] [Google Scholar]
- 17.Rajagopal JR, Farhadi F, Richards T, et al. Evaluation of coronary plaques and stents with conventional and photon-counting CT: benefits of high-resolution photon-counting CT. Radiol Cardiothorac Imaging. 2021;3:e210102. doi: 10.1148/ryct.2021210102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Gassenmaier T, Petri N, Allmendinger T, et al. Next generation coronary CT angiography: In vitro evaluation of 27 coronary stents. Eur Radiol. 2014;24:2953–2961. doi: 10.1007/s00330-014-3323-6. [DOI] [PubMed] [Google Scholar]
- 19.Zsarnoczay E, Fink N, Schoepf UJ, et al. Ultra-high resolution photon-counting coronary CT angiography improves coronary stenosis quantification over a wide range of heart rates - a dynamic phantom study. Eur J Radiol. 2023;161:110746. doi: 10.1016/j.ejrad.2023.110746. [DOI] [PubMed] [Google Scholar]
- 20.Remy-Jardin M, Hutt A, Flohr T, et al. Ultra-high-resolution photon-counting CT imaging of the chest: a new era for morphology and function. Invest Radiol. 2023;58:482–487. doi: 10.1097/RLI.0000000000000968. [DOI] [PubMed] [Google Scholar]
- 21.Si-Mohamed S, Boccalini S, Rodesch P-A, et al. Feasibility of lung imaging with a large field-of-view spectral photon-counting CT system. Diagn Interv Imaging. 2021;102:305–312. doi: 10.1016/j.diii.2021.01.001. [DOI] [PubMed] [Google Scholar]
- 22.Grunz J-P, Huflage H, Heidenreich JF, et al. Image quality assessment for clinical cadmium telluride-based photon-counting computed tomography detector in cadaveric wrist imaging. Invest Radiol. 2021;56:785–790. doi: 10.1097/RLI.0000000000000789. [DOI] [PubMed] [Google Scholar]
- 23.Conrads N, Grunz JP, Huflage H, et al. Ultrahigh-resolution computed tomography of the cervical spine without dose penalty employing a cadmium-telluride photon-counting detector. Eur J Radiol. 2023;160:110718. doi: 10.1016/j.ejrad.2023.110718. [DOI] [PubMed] [Google Scholar]
- 24.Patzer TS, Kunz AS, Huflage H et al. (2023) Ultrahigh-resolution photon-counting CT in cadaveric fracture models: spatial frequency is not everything. Diagnostics (Basel) 13. 10.3390/diagnostics13101677 [DOI] [PMC free article] [PubMed]
- 25.Rajendran K, Baffour F, Powell G, et al. Improved visualization of the wrist at lower radiation dose with photon-counting-detector CT. Skeletal Radiol. 2023;52:23–29. doi: 10.1007/s00256-022-04117-2. [DOI] [PubMed] [Google Scholar]
- 26.Gruschwitz P, Hartung V, Kleefeldt F, et al. Photon-counting versus energy-integrating detector CT angiography of the lower extremity in a human cadaveric model with continuous extracorporeal perfusion. Invest Radiol. 2023 doi: 10.1097/RLI.0000000000000982. [DOI] [PubMed] [Google Scholar]
- 27.Sandfort V, Choi Y, Symons R, et al. An optimized test bolus contrast injection protocol for consistent coronary artery luminal enhancement for coronary CT angiography. Acad Radiol. 2020;27:371–380. doi: 10.1016/j.acra.2019.05.003. [DOI] [PubMed] [Google Scholar]
- 28.Gruschwitz P, Hartung V, Kleefeldt F, et al. Continuous extracorporeal femoral perfusion model for intravascular ultrasound, computed tomography and digital subtraction angiography. PLoS ONE. 2023;18:e0285810. doi: 10.1371/journal.pone.0285810. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Bradley RA, Terry ME. Rank analysis of incomplete block designs: I. The method of paired comparisons. Biometrika. 1952;39:324–345. doi: 10.2307/2334029. [DOI] [Google Scholar]
- 30.Booij R, Kämmerling NF, Oei EHG, et al. Assessment of visibility of bone structures in the wrist using normal and half of the radiation dose with photon-counting detector CT. Eur J Radiol. 2023;159:110662. doi: 10.1016/j.ejrad.2022.110662. [DOI] [PubMed] [Google Scholar]
- 31.Kämmerling N, Sandstedt M, Farnebo S, et al. Assessment of image quality in photon-counting detector computed tomography of the wrist - an ex vivo study. Eur J Radiol. 2022;154:110442. doi: 10.1016/j.ejrad.2022.110442. [DOI] [PubMed] [Google Scholar]
- 32.Kosmala A, Weng AM, Schmid A, et al. Dual-energy CT angiography in peripheral arterial occlusive disease: diagnostic accuracy of different image reconstruction approaches. Acad Radiol. 2022;29(Suppl 4):S59–S68. doi: 10.1016/j.acra.2020.10.028. [DOI] [PubMed] [Google Scholar]
- 33.Klink T, Wilhelm T, Roth C, et al. Dual-energy CTA in patients with symptomatic peripheral arterial occlusive disease: study: Studie über die diagnostische Genauigkeit und limitierende Faktoren (dual-energy CTA in patients with symptomatic peripheral arterial occlusive disease: study of diagnostic accuracy and impeding factors) Rofo. 2017;189:441–452. doi: 10.1055/s-0043-101526. [DOI] [PubMed] [Google Scholar]
- 34.Decker JA, O'Doherty J, Schoepf UJ, et al. Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility. Eur Radiol. 2023;33:2469–2477. doi: 10.1007/s00330-022-09283-4. [DOI] [PubMed] [Google Scholar]
- 35.Petritsch B, Petri N, Weng AM, et al. Photon-Counting Computed Tomography for Coronary Stent Imaging. In Vitro Evaluation of 28 Coronary Stents. Invest Radiol. 2021;56:653–660. doi: 10.1097/RLI.0000000000000787. [DOI] [PubMed] [Google Scholar]
- 36.Hata A, Yanagawa M, Honda O, et al. Effect of Matrix Size on the Image Quality of Ultra-high-resolution CT of the Lung: Comparison of 512 × 512, 1024 × 1024, and 2048 × 2048. Acad Radiol. 2018;25:869–876. doi: 10.1016/j.acra.2017.11.017. [DOI] [PubMed] [Google Scholar]
- 37.Hsieh SS, Leng S, Rajendran K, et al. Photon Counting CT: Clinical Applications and Future Developments. IEEE Trans Radiat Plasma Med Sci. 2021;5:441–452. doi: 10.1109/TRPMS.2020.3020212. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Huflage H, Hendel R, Kunz AS, et al. Investigating the Small Pixel Effect in Ultra-High Resolution Photon-Counting CT of the Lung. Invest Radiol. 2023 doi: 10.1097/RLI.0000000000001013. [DOI] [PubMed] [Google Scholar]
- 39.Milos R-I, Röhrich S, Prayer F, et al. Ultrahigh-Resolution Photon-Counting Detector CT of the Lungs: Association of Reconstruction Kernel and Slice Thickness With Image Quality. AJR Am J Roentgenol. 2023;220:672–680. doi: 10.2214/AJR.22.28515. [DOI] [PubMed] [Google Scholar]
- 40.Rajendran K, Petersilka M, Henning A, et al. First Clinical Photon-counting Detector CT System: Technical Evaluation. Radiology. 2022;303:130–138. doi: 10.1148/radiol.212579. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Klein L, Dorn S, Amato C, et al. Effects of Detector Sampling on Noise Reduction in Clinical Photon-Counting Whole-Body Computed Tomography. Invest Radiol. 2020;55:111–119. doi: 10.1097/RLI.0000000000000616. [DOI] [PubMed] [Google Scholar]
- 42.Heye T, Knoerl R, Wehrle T, et al. The Energy Consumption of Radiology: Energy- and Cost-saving Opportunities for CT and MRI Operation. Radiology. 2020;295:593–605. doi: 10.1148/radiol.2020192084. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets generated and/or analyzed during the current study are not publicly available but are available from the corresponding author on reasonable request. Due to the nature of this research, participants of this study did not agree to their data to be shared in a public repository.