Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2023 Dec 19.
Published in final edited form as: J Heart Lung Transplant. 2022 Jul 16;41(12):1716–1726. doi: 10.1016/j.healun.2022.07.003

Hemodynamic reserve predicts early right heart failure after LVAD implantation

Jacob M Read a, Nnamdi I Azih a, Carli J Peters c, Vikram Gurtu d, Julie K Vishram-Nielsen d, Stephen P Wright d, Ana Carolina Alba d, Mathew J Gregoski e, Nicole A Pilch f, Steven Hsu g, Michael V Genuardi c, Chakradhari Inampudi b, Gregory R Jackson b, Nicholas Pope h, Lucas P Witer h, Arman Kilic h, Brian A Houston b, Susanna Mak d, Edo Y Birati c,i, Ryan J Tedford b
PMCID: PMC10729844  NIHMSID: NIHMS1948037  PMID: 35934606

Abstract

BACKGROUND:

Early right heart failure (RHF) remains a major source of morbidity and mortality after left ventricular assist device (LVAD) implantation, yet efforts to predict early RHF have proven only modestly successful. Pharmacologic unloading of the left ventricle may be a risk stratification approach allowing for assessment of right ventricular and hemodynamic reserve.

METHODS:

We performed a multicenter, retrospective analysis of patients who had undergone continuous-flow LVAD implantation from October 2011 to April 2020. Only those who underwent vasodilator testing with nitroprusside during their preimplant right heart catheterization were included (n = 70). Multivariable logistic regression was used to determine independent predictors of early RHF as defined by Mechanical Circulatory Support–Academic Research Consortium.

RESULTS:

Twenty-seven patients experienced post-LVAD early RHF (39%). Baseline clinical characteristics were similar between patients with and without RHF. Patients without RHF, however, achieved higher peak stroke volume index (SVI) (30.1 ± 8.8 vs 21.7 ± 7.4 mL/m2; p < 0.001; AUC: 0.78; optimal cut-point: 22.1 mL/m2) during nitroprusside administration. Multivariable analysis revealed that peak SVI was significantly associated with early RHF, demonstrating a 16% increase in risk of early RHF per 1 ml/m2 decrease in SVI. A follow up cohort of 10 consecutive patients from July 2020 to October 2021 resulted in all patients being categorized appropriately in regards to early RHF versus no RHF according to peak SVI.

CONCLUSION:

Peak SVI with nitroprusside administration was independently associated with post-LVAD early RHF while resting hemodynamics were not. Vasodilator testing may prove to be a strong risk stratification tool when assessing LVAD candidacy though additional prospective validation is needed.

Keywords: stroke volume index, ventricular reserve, left ventricular assist device, hemodynamics, pulmonary hypertension


Left ventricular assist devices (LVADs) are increasingly utilized to improve survival and quality of life in patients with advanced heart failure.1,2 Unfortunately, early right heart failure (RHF) after LVAD implantation is a common complication associated with significant morbidity and mortality.35 Preimplant prediction of early RHF remains a challenge at least in part due to difficulties in properly assessing right ventricular (RV) function.610 The gold standard to assess RV function relies on multibeat pressure volume relations to determine RV-pulmonary arterial (PA) coupling, yet these measures are expensive, technically difficult and not feasible for routine clinical practice.11 Alternatively, measures of RV reserve present as an appealing option as it is technically easier to assess. Recent data suggest measurements of RV reserve are tightly associated with RV-PA coupling,12 and therefore, assessing RV reserve may represent an opportunity for better prediction of post-LVAD early RHF compared with traditional resting evaluations.

Reserve measures are usually performed with exercise or inotropic agents like dobutamine. This can be problematic in potential LVAD recipients who may already require inotropic agents or have substantial exertional limitations. Administration of nitroprusside acutely lowers systemic afterload, lowers pulmonary vascular resistance (PVR), and increases left ventricular (LV) stroke volume thereby pharmacologically unloading the LV in a simulacrum of mechanical support devices.13 However, if the RV lacks “functional” reserve and is unable to maintain LV preload, stroke volume augmentation will be limited. We therefore, hypothesized that the assessment of hemodynamic reserve in response to vasodilator administration and the associated changes in stroke volume may be helpful to predict early RHF post LVAD implantation.

Methods

After IRB approval, we performed a multicenter retrospective analysis of patients who had undergone continuous-flow LVAD implantation from October 2011 to April 2020. The three centers included were the Medical University of South Carolina (2016–2020), the University of Pennsylvania (2017–2020) and the University Health Network in Toronto (2011–2020). Only patients who underwent clinically indicated vasodilator testing with nitroprusside within one year of LVAD implant as part of their advanced heart failure therapy evaluation were included (n = 70). Vasodilator testing was performed as part of an advanced therapy evaluation when pulmonary vascular resistance ≥2.5 WU (n = 62) or in the setting of significantly elevated pulmonary arterial wedge pressure (PAWP) and systolic pulmonary artery pressure (SPAP; n = 8).14,15 Relevant demographics, clinical characteristics and hemodynamics at baseline and at peak vasodilation were recorded. Right heart failure was defined using the new 2020 Mechanical Circulatory Support––Academic Research Consortium (MCS-ARC) definition and included both early acute and early postimplant RHF (Supplemental Table 1). Stroke volume index was calculated as (cardiac output/heart rate)/body surface area. PA elastance (Ea) was calculated by systolic PA pressure divided by stroke volume. After determining the optimal cut-point for peak SVI to predict RHF, an additional 10 consecutive patients who received vasodilator testing prior to LVAD implant from July 2020 to October 2021 were also analyzed.

Statistical analysis

Statistical analyses were completed using SPSS version 25 (IBM corporation, Armonk, NY). Means and standard deviations were calculated for all continuous variables; frequencies and percentages for all categorical variables. Variables with continuous outcomes differences were assessed using independent samples t-tests with early postimplant RHF versus no early post-implant RHF serving as the independent groups. All continuous outcomes were examined using Levene’s Test for equality of variances. If significant variance differences were found between groups, results were reported based on adjusted degrees of freedom using the Welch-Satterthwaite method. Categorical variables were assessed using X2 analyses or Fischer’s exact tests as appropriate.

A total of 50 patient covariates were screened for significant correlation with early RHF using univariable logistic regression models to ensure individual covariates had adequate statistical power for inclusion. Following univariable logistic regression models, 2 sets of multivariable logistic models were completed. In the first multivariable model (i.e., liberal model), any covariate with a univariable p value ≤0.10 was entered into a multivariable logistic regression model to determine independent predictors of early RHF. The exception to this criterion was to exclude cardiac index (CI) and PA Ea because of the strong significant correlation (collinearity) with SVI. SVI was selected over CI and PA Ea based on stronger Wald coefficient, odds ratio (OR), and p-value with the early RHF outcome. In the second multivarianble regression model (conservative model), any covariate with a univariable p value ≤0.05 was entered into a multivariable logistic regression model to determine independent predictors of early RHF. Logistic regression results are presented as OR with corresponding p values. P values ≤0.05 were considered statistically significant. Finally, area under the curve (AUC) was calculated with a receiver operating characteristic (ROC) curve using predicted probabilities from the logistic regression model with early post implant RHF status to demonstrate an aggregate of the true positive and false positive rates.

Results

Of the 70 patients included, 27 patients developed early RHF (39%) after LVAD implantation, including 14 patients who required right ventricular assist device (RVAD) support (20%). The mean age of the cohort was 49 ± 14 years with the majority of patients being men (77%) and with nonischemic cardiomyopathy (74%). Early RHF differed by centers, device type, and time between RHC and LVAD implant. There were no other significant difference in demographics, preimplant clinical characteristics, RHF Risk Scores,1618 relevant laboratory values, or echocardiographic assessment of LV, RV, or valvular function (Table 1). Institution specific data is shown in Supplemental Table 2.

Table 1.

Demographics and Clinical Characteristics.

Entire cohort (N = 70) No early RHF (N = 43) Early RHF (N = 27) p-value

Facility 0.006
 MUSC(n) 28 18 [42] 10 [37]
 Penn (n) 25 10 [23] 15 [56]
 Toronto (n) 17 15 [35] 2 [7]
Gender 0.493
 Male (n) 54 32 [74] 22 [81]
Race 0.63
 Caucasian (n) 34 23 [53] 11 [41]
 Black (n) 30 17 [40] 13 [48]
 Asian (n) 1 1 [2] 0[0]
 Hispanic (n) 2 1 [2] 1 [4]
 Other(n) 3 1 [2] 2 [7]
BMI (kg/m2) 30 ± 7 29 ± 7 31 ± 7 0.374
Age (years) 49 ± 14 50 ± 15 49 ± 12 0.952
Prior Sternotomy (n) 6 4 [9] 2 [7] 1.00
Pulmonary Vasodilator during RHC (n) 7 3 [7] 4 [15] 0.417
Time between RHC and LVAD implant (days) 61 ± 87 80 ± 99 29 ± 99 0.007
Implant Strategy 0.654
 Sternotomy (n) 66 40 [93] 26 [96]
 Thoracotomy (n) 4 3 [7] 1 [4]
Device Type 0.025
 HVAD (n) 14 8 [19] 6 [22]
 HM2 (n) 14 13 [30] 1 [4]
 HM3(n) 42 22 [51] 20 [74]
Heart Failure Etiology 0.596
 Ischemic (n) 18 12 [28] 6 [22]
 Non-Ischemic (n) 52 31 [72] 21 [78]
INTERMACS Profile 0.084
 Profile 1 (n) 10 4 [9] 6 [22]
 Profile 2 (n) 16 7 [16] 9 [33]
 Profile 3 (n) 30 21 [49] 9 [33]
 Profile 4–7 (n) 14 11 [26] 3 [11]
Temporary Mechanical Support at Implant 0.059
 IABP (n) 9 5 [12] 4 [15]
 Impella (n) 2 0[0] 2 [7]
 ECMO at Implant (n) 4 1 [2] 3 [11]
Echo parameters
 EDD LV (cm) 7.0 ± 0.9 6.9 ± 0.9 7.1 ± 1.0 0.606
 LV Ejection Fraction (%) 18 ± 7 18 ± 6 17 ± 7 0.405
 TAPSE (mm) 16 ± 5 16 ± 6 15 ± 5 0.388
 TAPSE/PASP (mm/mmHg) 0.32 ± 0.19 0.33 ± 0.17 0.32 ± 0.22 0.901
 Severe RV Dysfunction (n) 12 4 [9] 8 [30] 0.755
 Severe Mitral Regurgitation (n) 24 12 [28] 12 [44] 0.199
 Severe RV Dilation (n) 11 8 [19] 3 [11] 0.511
 Severe Tricuspid Regurgitation (n) 7 6 [14] 1 [4] 0.238
Penn-Columbia Risks Score 0.503
 Low Risk (n) 39 24 [56] 15 [56]
 Medium Risk (n) 17 12 [28] 5 [19]
 High Risk (n) 14 7 [16] 7 [26]
Euromacs RV Risk Score 3.2 ± 1.9 2.9 ± 1.9 3.7 ± 1.9 0.107
Kormos RHF Risk Score 1.1 ± 1.9 0.9 ± 1.7 1.6 ± 2.2 0.154
CRITT Score 1.6 ± 0.9 1.5 ± 1.0 1.8 ± 0.8 0.163
Lab Values
 Hemoglobin (g/dL) 11.7 ± 2.2 11.7 ± 2.3 11.5 ± 2.2 0.671
 Sodium (mmol/L) 135 ± 5 136 ± 4 134 ± 6 0.345
 Creatinine (mg/dL) 1.5 ± 0.6 1.4 ± 0.5 1.7 ± 0.8 0.11
 Total Bilirubin (mg/dL) 1.4 ± 1.5 1.5 ± 1.8 1.4 ± 0.2 0.871

Abbreviations: BMI, body mass index; Echo, echocardiography; ECMO, extracorpeal membrane oxygenation; EDD, end-diastolic diameter; HM2, Heartmate 2; HM3, Heartmate 3; HVAD, Heartware Ventricular Assist Device; IABP, Intra-aortic balloon pump; LV, left ventricle; LVAD, left ventricular assist device; MUSC, Medical University of South Carolina; PASP, pulmonary artery systolic pressure; Penn, University of Pennsylvania; RHC, right heart cathertization; RHF, right heart failure; RV, right ventricle; TAPSE, tricuspid annular plane systolic excursion; Toronto, University of Toronto.

Values are presented as mean ± standard deviation unless otherwise indicated; [ ] = percentage.

Hemodynamics

Table 2 summarizes preoperative baseline hemodynamics, hemodynamics measured at peak vasodilator infusion, and changes in hemodynamics from peak to baseline. At baseline, lower SVI, higher PA Ea, higher mean PA pressure, and higher heart rate were statistically associated with early RHF. However, absolute differences of these values between groups were relatively modest. During nitroprusside administration (mean 1.7 ± 1.6 mcg/kg/min), both groups realized similar reductions in systemic mean arterial pressure, pulmonary pressures, and PAWP. Compared to the group with early RHF, those without RHF achieved a higher peak SVI (30.1 ± 8.8 vs 21.7 ± 7.4 mL/m2; p < 0.001), higher peak CI (2.39 ± 0.62 vs 1.97 ± 0.36 L/min/m2; p = 0.001), lower PA Ea (0.83 ± 0.31 vs 1.19 ± 0.52 mmHg/mL; p = 0.002), and lower heart rate (82 ± 17 vs 96 ± 22; p = 0.004). The change in CI and SVI was also higher in the group without RHF, despite no significant difference in change in RV afterload (PVR or PA Ea).

Table 2.

Hemodynamic Measurements

Entire cohort (N = 70) [mean ± SD] No early RHF (N = 43) [mean ± SD] Early RHF (N=27) [mean ± SD] p-value

Baseline hemodynamics
MAP (mmHg) 83 ± 9 82 ± 9 85 ± 8 0.185
PAPM (mmHg) 41 ± 7 40 ± 7 43 ± 7 0.044
CI (L/min/m2) 1.65 ± 0.36 1.66 ± 0.36 1.62 ± 0.36 0.669
RAP (mmHg) 11 ± 5 10 ± 4 12 ± 6 0.134
PAWP (mmHg) 25 ± 5 24 ± 5 25 ± 6 0.305
RAP:PAWP 0.45 ± 0.19 0.43 ± 0.18 0.49 ± 0.21 0.223
PVR (Wood units) 5.2 ± 2.2 4.9 ± 2.1 5.6 ± 2.3 0.248
SVI (mL/m2) 19 ± 6 21 ± 6 17 ± 5 0.01
Pulmonary Ea (mmHg/mL) 1.72 ± 0.64 1.58 ± 0.56 1.94 ± 0.70 0.021
Pca (mL/mmHg) 1.34 ± 0.68 1.40 ± 0.78 1.24 ± 0.50 0.342
PAPi(mmHg) 3.9 ± 3.0 4.3 ± 3.5 3.2 ± 1.9 0.155
SVR (dynes/sec/cm−5) 1807 ± 477 1796 ± 458 1824 ± 514 0.82
HR (bpm) 89 ± 20 83 ± 17 99 ± 22 0.001
Delta: Peak - Baseline
MAP (mmHg) −14 ± 8 −14 ± 9 −14 ± 8 0.754
PAPM (mmHg) −11 ± 7 −11 ± 7 −12 ± 8 0.472
CI (L/min/m2) 0.58 ± 0.50 0.73 ± 0.52 0.35 ± 0.37 0.001
PAWP (mmHg) −7 ± 6 −8 ± 5 −7 ± 7 0.655
PVR (Wood units) −2.0 ± 2.1 −2.1 ± 2.0 −1.8 ± 2.3 0.662
SVI (mL/m2) 7.5 ± 7.2 9.2 ± 7.5 4.7 ± 6.0 0.011
Pulmonary Ea (mmHg/mL) −0.76 ± 0.48 −0.75 ± 0.41 −0.76 ± 0.59 0.98
HR (bpm) −2 ± 10 −1 ± 7 −3 ± 13 0.51
Peak Vasodilator Hemodynamics
Nitroprusside dose (mcg/kg/min) 1.7 ± 1.6 1.8 ± 1.9 1.7 ± 1.2 0.816
MAP (mmHg) 69 ± 9 67 ± 10 71 ± 8 0.121
PAPM (mmHg) 30 ± 7 29 ± 6 31 ± 7 0.179
CI (L/min/m2) 2.23 ± 0.57 2.39 ± 0.62 1.97 ± 0.36 0.001
PAWP (mmHg) 17 ± 6 16 ± 6 18 ± 7 0.243
PVR (Wood units) 3.1 ± 1.4 2.9 ± 1.3 3.4 ± 1.6 0.152
SVI (mL/m2) 27 ± 9 30 ± 9 22 ± 7 <0.001
Pulmonary Ea (mmHg/mL) 0.97 ± 0.44 0.83 ± 0.31 1.19 ± 0.52 0.002
HR (bpm) 88 ± 20 82 ± 17 96 ±22 0.004

Abbreviations: CI, cardiac index; Ea, effective arterial elastance; HR, heart rate; MAP, mean arterial pressure; PAPi, pulmonary arterial pulsitility index; PAPM, mean pulmonary arterial pressure; PAWP, pulmonary arterial wedge pressure; Pca, pulmonary arterial compliance; PVR, pulmonary vascular resistance; RAP, right atrial pressure; SD, standard deviation; SVI, stroke volume index; SVR, systemic vascular resistance.

Values are presented as mean ± standard deviation unless otherwise indicated.

Association between clinical and hemodynamic characteristics and early RHF

Univariable and multivariable predictors of early RHF are shown in Table 3. In both liberal and conservative multivariable models, only baseline mean PA pressure and peak SVI remained significant, even when considering baseline SVI. ROC curves for SVI at baseline, delta, and peak are shown in Figure 1, with the AUC being the highest (0.780) for peak SVI. The optimal cut point was 22.1 mL/m2, which had a 56% sensitivity, 84% specificity, 75% negative predictive value, and 68% positive predictive value for predicting postimplant early RHF (Figure 2). Multivariable analysis revealed that peak SVI was significantly associated with early RHF, demonstrating a 16% increase in risk of early RHF per 1 ml/m2 decrease in SVI. Peak SVI, but not baseline or delta SVI, was also associated with early RHF when defined strictly by postoperative RVAD use (Supplemental Figure 1, p = 0.027; AUC: 0.693). Peak SVI also remained predictive of early RHF when limiting our cohort to only those with RHC within 6 months of VAD implant (n = 61, AUC: 0.757; p = 0.001), HeartMate 3 implants (n = 43; AUC: 0.832; p < 0.001), or those patients who did not require RVAD implant (n = 54; AUC: 0.800, p = 0.001). Cardiopulmonary bypass time (93 ± 39 vs 113 ± 50 min, p = 0.096) and intraoperative transfusions (1.1 ± 2.1 vs 2.2 ± 2.7, p = 0.089) were numerically higher in those who developed RHF, though neither reached statistical significance.

Table 3.

Univariable and Multivariable Predictors of Right Heart Failure

Characteristic Univariable model OR|95% CI|p-value Liberal model OR|95% CI|p-value Conservative model OR|95°% CI|p-value

Facility
 MUSC Reference Reference
 Penn (2.70)|0.888–8.21| p = 0.080 (1.35)|.264–6.93| p = 0.716
 Toronto (0.24)|0.045–1.27| p = 0.093 (.832)|.104–6.65| p = 0.863
Race
 White reference
 Black (1.60)|0.577–4.43| p = 0.367
 Hispanic (2.09)|0.119–36.63| p = 0.614
 Asian (0.00)|0.000–0.00| p = 1.00
 Other (4.18)|0.341–51.24| p = 0.263
Gender
 Female Reference
 Male (1.51)|0.461–4.96| p = 0.495
Age (0.999)|0.966–1.03| p = 0.951
Device Type
 HM3 Reference Reference Reference
 HM2 (0.085)|0.010-.706| p = 0.023 (0.095)|0.007–1.26| p = 0.074 (0.081)|0.008–0.815| p = 0.033
 HVAD (0.825)|0.244–2.79| p = 0.757 Reference Reference
Heart Failure Etiology
 Non-Ischemic reference
 Ischemic (0.738)|0.239–2.28| p = 0.597
 INTERMACS Profile (0.514)|0.298–0.885| p = 0.016 (0.745)|0.318–1.75| p = 0.499 (0.707)|0.339–1.48| p = 0.356
Temporary Mechanical Support at Implant
 No Reference Reference
 Yes (3.2)|.921–11.12| p = 0.067 (1.03)|0.162–6.57| p = 0.972
Echo parameters
 EDD LV (cm) (1.15)|0.688 –1.91| p = 0.601
 LV Ejection Fraction (%) (0.968)|0.897 – 1.04| p = 0.401
 TAPSE (mm) (0.959)|0.872 –1.05| p = 0.384
Severe RV Dysfunction
 No Reference
 Yes (0.761)|0.205–2.82| p = 0.683
Severe RV Dilation
 No reference
 Yes (0.547)|0.132–2.27| p = 0.406
Severe Mitral Regurgitation
 No reference
 Yes (2.07)|0.753–5.68| p = 0.159
Severe Tricuspid Regurgitation
 No Reference
 Yes (0.240)|0.027 –2.12| p = 0.199
Baseline Hemodynamics
 MAP (mmHg) (1.04)|0.982–1.10| p = 0.186
 PAPM (mmHg) (1.08)|1.00–1.16| p = 0.051 (1.12)|1.00–1.24| p = 0.049 (1.12)|1.00–1.25| p = 0.046
 CI (L/min/m2) (.738)|.187–2.91| p = 0.664
 RAP (mmHg) (1.08)|.975–1.19| p = 0.143
 PAWP (mmHg) (1.05)|.957–1.15| p = 0.301
 RAP:PAWP (4.85)|0.380–61.89| p = 0.224
 PVR (Wood units) (1.15)|0.911–1.44| p = 0.245
 SVI (mL/m2) (0.882)|0.798–0.975| p = 0.014 (1.07)|0.916–1.25| p = 0.397 (1.06)|0.912–1.24| p = 0.436
 Pulmonary Ea (mmHg/mL)a (2.61)|1.11–6.12| p = 0.028
 Pca (mL/mmHg) (0.644)|0.254–1.64| p = 0.355
 PAPi (mmHg) (0.863)|0.699–1.06| p = 0.167
 HR(BPM)a (1.04)|1.01–1.07| p = 0.003
Peak Hemodynamics
 MAP (mmHg) (1.05)|0.987–1.12| p = 0.126
 PAPM (mmHg) (1.05)|0.977–1.14| p = 0.179
 CI (L/min/m2)a (0.169)|0.049-.579| p = 0.005
 PAWP (mmHg) (1.05)|0.972–1.14| p = 0.211
 PVR (Wood units) (1.29)|0.909–1.83| p = 0.154
 SVI (mL/m2) (0.864)|0.794-.939| p < 0.001 (0.852)|0.756–0.959| p = 0.008 (0.848)|0.754–0.955| p = 0.006
 Pulmonary Ea (mmHg/mL)a Risk Score (9.05)|2.19–37.41| p = 0.002
 Penn Columbia Risk Score (Low Risk) Reference
 Penn Columbia Risk Score (Medium Risk) (0.667)|0.196–2.27| p = 0.517
 Penn Columbia Risk Score (High Risk) (1.60)|0.468–5.48| p = 0.454
 Euromacs RV Risk Score (1.24)|0.952–1.61| p = 0.111
 Kormos RHF Risk Score (1.20)|0.928–1.56| p = 0.164
 CRITT Score Lab Values (1.47)|0.854–2.54| p = 0.164
Lab Values
 Hemoglobin (g/dL) (0.952)|0.764–1.19| p=.666
 Sodium (mmol/L) (0.953)|.862–1.05| p = 0.341
 Creatinine (mg/dL) (1.88)|.838–4.23| p = 0.126
 Total Bilirubin (mg/dL) (0.973)|0.704–1.34| p = 0.869
Days Between RHC and LVAD implant (0.990)|0.981–0.999| p = 0.033 (0.999)|0.988–1.01| p = 0.815 (0.998)|0.987–1.01| p = 0.719

Abbreviations: CI, cardiac index; Ea, effective arterial elastance; Exp(B), exponentiation of B coefficient; EDD, end-diastolic diameter; HM2, Heartmate 2; HM3, Heartmate 3; HR, heart rate; HVAD, Heartware Ventricular Assist Device; LV, left ventricle; LVAD, left ventricular assist device; MAP, mean arterial pressure; MCS, mechanical circulatory support; OR, odds ratio; PAPi, pulmonary arterial pulsatility index; PAPM, mean pulmonary artery pressure; PAWP, pulmonary arterial wedge pressure; Pca, pulmonary arterial compliance; Penn, University of Pennsylvania; PVR, pulmonary vascular resistance; RAP, right atrial pressure; RHC, right heart catheterization; RHF, right heart failure; RV, right ventricle; SVI, stroke volume index; TAPSE, tricuspid annular plane systolic excursion; Toronto, University of Toronto.

Liberal Model includes univariable covariates ≤ 0.10, Conservative model includes univariable covariates ≤ 0.05.

a

CI, Pulmonary Ea, HR, and SVI have significant collinearity so only SVI was included in the multivariable models.

Figure 1.

Figure 1

ROC curve of SVI association with early RHF. ROC curve for stroke volume index at baseline, delta, and peak values to predict MCS-ARC defined early right heart failure after LVAD implantation. *Abbreviations: AUC=area under the curve; LVAD=left ventricular assist device; RHF=right heart failure; ROC=receiver operating characteristic; SVI=stroke volume index.

Figure 2.

Figure 2

SVI after nitroprusside predicts early RHF. Population pyramid dichotomized at SVI of 22.1 mL/m2. The x-axis details the number of patients whose SVI is above or below 22.1. The y axis details if the patient did or did not develop RHF. Percentages for each quadrant is depicted with the total percentage representing the entire cohort. The majority with SVI <22.1 mL/m2 developed early RHF, with a specificity of 84%, sensitivity of 56%, negative predictive value of 75%, and positive predictive value of 68%. *Abbreviations: LVAD=left ventricular assist device; RHF=right heart failure; SVI=stroke volume index.

Follow up cohort

Three of the 10 patients included in the follow up cohort developed RHF. Clinical and hemodynamic characteristics of this cohort are found in Supplemental Table 3. Only peak SVI was different between those who did and did not develop RHF (p = 0.017). All of the patients were categorized appropriately according to the previously established peak SVI cut-point of 22.1 mL/m2.

LVAD implants without vasodilator testing

Supplemental table 4 outlines the demographics and clinical characteristics of the 191 excluded patients who received LVAD therapy over the study period but did not receive pre-implant vasodilator testing. Many characteristics were similar. This group was slightly older, had a higher proportion of INTERMACS 1 profile, and less commonly had severe mitral regurgitation. In terms of preimplant hemodynamics, the non-vasodilator group had similar degrees of right and left heart dysfunction but lower, albeit still elevated, RV afterload. One hundred forty-seven of the 191 patients (77%) met at least one of the criteria for vasodilator testing.

Discussion

The major findings of our study can be summarized as follows: (1) Early postimplant RHF was common in our cohort, (2) baseline clinical characteristics, echocardiographic and hemodynamic measures were not strongly and independently predictive of early RHF, and (3) hemodynamics at the peak of nitroprusside vasodilator administration, particularly peak SVI, was a good discriminator of those who did and did not develop early postimplant RHF. These findings suggest that vasodilator testing and acute pharmacologic LV unloading may be useful in risk stratification prior to LVAD implantation.

RV functional reserve response to LV unloading

Prediction of early RHF after LVAD remains challenging with most risk scores, and prediction tools only show modest discrimination in validation cohorts.19 Indeed, 4 published risk models included in our analysis were not predictive of RHF nor were echocardiographic or previously identified hemodynamic values like pulmonary artery pulsatility index and RAP:PAWP ratio. Both perioperative and postoperative factors clearly contribute to postoperative RHF, ones that are either not present or predictable preoperatively.20,21 However, limitations related to preoperative assessment of RV function also play a role. Recent evidence suggests measures of RV reserve may be more closely associated with the gold standard of RV functional assessment (RV-PA coupling) than resting measures. In our study, nitroprusside, a potent systemic afterload reducer, predictably unloaded the LV and lowered LV filling pressures which in turn reduced pulmonary pressures and PVR. Stroke volume increased in the majority of patients, with a disparate response between those who ultimately did and did not develop early RHF after LVAD implantation. This disparate response may relate to a lack of RV functional reserve, and specifically, the inability to maintain sufficient LV preload as systemic afterload is reduced. It is important to note that in our cohort PA pressures, PVR, and pulmonary Ea had nearly identical reductions in those who developed early RHF and those who did not. This suggests that the disparate changes in SVI were not merely a reflection of differences in RV afterload but rather differences in RV contractility to handle increasing venous return. Indeed, a recent study by Kremer and colleagues suggested RV volume loading may be able to uncover impaired RV-PA coupling and lack of RV reserve.22 Importantly, pharmacologic unloading of the LV and the subsequent RV response are hemodynamically similar to what would occur with mechanical unloading with LVAD implantation. Whether testing the response of the RV to temporary LV mechanical support would also be predictive of early RHF is unknown.

The definition of early RHF in our study was based on the new MCS-ARC adverse events definitions.23 This definition is similar to the 2013 INTERMACS criteria for severe and severe-acute RHF which has been associated with poor outcomes after LVAD.24 In our study, all subjects declared to have early RHF met both definitions. Peak SVI remained predictive of early RHF even if post-implant RHF was defined more narrowly as the need for RVAD. RHF was common with 39% of subjects meeting the broader criteria for early RHF and 20% requiring RVAD support within the first 30 days. While relatively high rates of early RHF are similar to recent reports of centrifugal devices,2527 center specific thresholds for RVAD implantation likely also played a role with 8 of the 14 RVAD implantations planned at time of LVAD implant. Importantly, and given the retrospective nature of this study, the decision to use a planned RVAD was likely not related to vasodilator testing results. A subgroup analysis excluding all patients with RVAD implant also exhibited consistent results. Additionally, most but not all subjects had combined post- and pre-capillary PH; thus, higher RV afterload may have contributed to the high rates of early RHF. However, PVR has not been found to be a strong predictor of early RHF in other studies.11,16

RV functional reserve in prediction of early RHF

Two prior studies have hinted that measures of RV reserve may be useful in predicting early RHF after LVAD implantation. In one small study of 6 patients with resting RV dysfunction by echocardiography, the RV response to dobutamine infusion (changes in estimated SPAP and tricuspid annular plane systolic excursion) was modestly different between those subjects who did and did not develop early RHF.28 Although certainly dobutamine or another inotropic agent could be used to assess RV functional reserve, many higher risk patients are already on inotropic agents at the time of LVAD evaluation. In this cohort, 56% of subjects were already on an inotrope at time of RHC. Inotropic agents may also not unload the LV to the same extent as nitroprusside. These two factors may therefore limit the utility of such assessments. Another more recent manuscript by Gonzalez et al. compared the predictive ability of baseline hemodynamics to those when the patient’s heart failure was “optimized” prior to implant.29 Optimization included a combination of diuretic therapy, inotropic agents and vasodilators. The authors concluded that the optimized hemodynamics, specifically the optimized PA pulsatility index, was additive to prediction models of post-LVAD early RHF. Similar to our study, the most optimized (peak) hemodynamic values, rather than the change from baseline to optimized, were more predictive of early RHF.

The AUC for peak stroke volume index in our study fell into the acceptable to good range (0.78), with 2 subgroup analyses (HM3 only; RHF excluding RVAD use) being slightly higher. As a comparison the AUC for the Euromacs RHF risk score was 0.7 in the derivation and 0.67 in the validation cohort.16 There is clearly still progress to make in risk prediction of RHF. Given this persistent uncertainly, using these risk stratification tools in shared-decision making discussions and perhaps to identify high risk individuals who may benefit from additional perioperative support may be more appropriate than restricting LVAD candidacy altogether.

Limitations

There are several limitations of this work which merit further discussion. First, this was a retrospective analysis, and only subjects who had a clinical indication for vasodilator testing were included. Therefore, the cohort is somewhat biased in that only patients with elevated PVR ≥2.5 woods or significantly elevated PAWP and SPAP underwent vasodilator testing. A comparison cohort of patients who underwent LVAD implant without vasodilator testing had generally similar characteristics, but lower, albiet still elevated, RV afterload. It remains unknown if this strategy would be helpful in predicting early RHF in those without a pre-capillary component to their PH or in those with a more modest baseline PAWP and SPAP. However, the latter hemodynamic profile would be uncommon in LVAD candidates as those with normal or near normal PAWP may not be optimal candidates for LVAD therapy as additional LV unloading would be unlikely to improve heart failure symptoms.30 While we believe the multicenter nature of our study is an overall strength, the incidence of early RHF was lower in one center compared to the other two centers. This difference was not maintained in multivariable modeling suggesting that patient characteristics rather than center specific differences played a more significant role. Further, we completed both a liberal and conservative multivariable model approach to examine the tradeoffs between statistical power (i.e., increased in the liberal model but this also increased chances of a type 1 error) and a conservative model (i.e., more parsimonious and less likely to be over fit); across both models peak SVI remained significant. However, we acknowledge the limitation of sample size and need for additional external validation with future prospective studies. Finally, we did not have measures of right atrial pressure during nitroprusside infusion which limits our ability to calculate changes in right heart hemodynamics.

Conclusions

In summary, peak SVI during nitroprusside infusion was significantly associated with post-LVAD early RHF. These findings suggest vasodilator testing and acute pharmacologic LV unloading may be useful in risk stratification prior to LVAD implantation in patients with elevated pulmonary pressures and should be prospectively studied.

Supplementary Material

Figures
Tables

Acknowledgments

We wish to acknowledge our mechanical circulatory support teams who help us care for this complex patient population.

Financial disclosures

No grants, contracts, or other financial support was received for this manuscript.

Abbreviations:

CI

cardiac index

Ea

elastance

MCS-ARC

Mechanical Circulatory Support–Academic Research Consortium

PAWP

pulmonary artery wedge pressure

RHC

right heart catheterization

RHF

right heart failure

RVAD

right ventricular assist device

SVI

stroke volume index

Footnotes

Disclosures

All authors report no direct conflicts of interest related to this manuscript. Dr. Tedford reports general disclosures to include consulting relationships with Medtronic, Abbott, Aria CV Inc., Arena Pharmaceuticals, Acceleron, Itamar, Edwards LifeSciences, Eidos Therapeutics, Lexicon Pharmaceuticals, Gradient and United Therapeutics. Dr. Tedford is on a steering committee for Acceleron and Abbott as well as a research advisory board for Abiomed. He also does hemodynamic core lab work for Actelion and Merck. Dr. Birati reports general disclosures to include research support paid to the University of Pennsylvania by Impulse Dynamics and Medtronic Inc. Dr. Houston reports general disclosures to include consulting relationships with Medtronic and Bioventrix. He has received research grant funding from Medtronic. Dr. Jackson reports general disclosures to include speaking honorarium from Respicardia, Inc. Dr. Inampudi reports general disclosures to include consulting relationship with Abbott. Dr. Kilic reports general disclosures to include consulting relationship with Abiomed and medical advisor board for Medtronic Inc. Dr. Genuardi reports general disclosures to include consulting relationship with Respicardia, and non-financial compensation from Abbott.

Supplementary materials

Supplementary material associated with this article can be found in the online version at https://doi.org/10.1016/j.healun.2022.07.003.

References

  • 1.Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, et al. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med 2001;345(20):1435–43. [DOI] [PubMed] [Google Scholar]
  • 2.Kadakia S, Moore R, Ambur V, Toyoda Y. Current status of the implantable LVAD. Gen Thorac Cardiovasc Surg 2016;64(9):501–8. [DOI] [PubMed] [Google Scholar]
  • 3.Kavarana MN, Pessin-Minsley MS, Urtecho J, Catanese KA, Flannery M, Oz MC, et al. Right ventricular dysfunction and organ failure in left ventricular assist device recipients: a continuing problem. Ann Thorac Surg 2002;73(3):745–50. [DOI] [PubMed] [Google Scholar]
  • 4.Takeda K, Naka Y, Yang JA, Uriel N, Colombo PC, Jorde UP, et al. Outcome of unplanned right ventricular assist device support for severe right heart failure after implantable left ventricular assist device insertion. J Heart Lung Transplant 2014;33(2):141–8. [DOI] [PubMed] [Google Scholar]
  • 5.Baxter RD, Tecson KM, Still S, Collier JDG, Felius J, Joseph SM, et al. Predictors and impact of right heart failure severity following left ventricular assist device implantation. J Thorac Dis 2019;11(Suppl 6):S864–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Fukamachi K, McCarthy PM, Smedira NG, Vargo RL, Starling RC, Young JB. Preoperative risk factors for right ventricular failure after implantable left ventricular assist device insertion. Ann Thorac Surg 1999;68(6):2181–4. [DOI] [PubMed] [Google Scholar]
  • 7.Drakos SG, Janicki L, Horne BD, Kfoury AG, Reid BB, Clayson S, et al. Risk factors predictive of right ventricular failure after left ventricular assist device implantation. Am J Cardiol 2010;105(7):1030–5. [DOI] [PubMed] [Google Scholar]
  • 8.Matthews JC, Koelling TM, Pagani FD, Aaronson KD. The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol 2008;51(22):2163–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Boegershausen N, Zayat R, Aljalloud A, Musetti G, Goetzenich A, Tewarie L, et al. Risk factors for the development of right ventricular failure after left ventricular assist device implantation-a single-centre retrospective with focus on deformation imaging. Eur J Cardiothorac Surg 2017;52(6):1069–76. [DOI] [PubMed] [Google Scholar]
  • 10.Wagner T, Bernhardt AM, Magnussen C, Reichenspurner H, Blankenberg S, Grahn H. Right heart failure before LVAD implantation predicts right heart failure after LVAD implantation - is it that easy? J Cardiothorac Surg 2020;15(1):113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Muslem R, Ong CS, Tomashitis B, Schultz J, Ramu B, Craig ML, et al. Pulmonary Arterial Elastance and INTERMACS-Defined Right Heart Failure Following Left Ventricular Assist Device. Circ Heart Fail 2019;12(8):e005923. [DOI] [PubMed] [Google Scholar]
  • 12.D’Alto M, Pavelescu A, Argiento P, Romeo E, Correra A, Di Marco GM, et al. Echocardiographic assessment of right ventricular contractile reserve in healthy subjects. Echocardiography 2017;34(1):61–8. [DOI] [PubMed] [Google Scholar]
  • 13.Opasich C, Cioffi G, Gualco A. Nitroprusside in decompensated heart failure: what should a clinician really know? Curr Heart Fail Rep 2009;6(3):182–90. [DOI] [PubMed] [Google Scholar]
  • 14.Mehra MR, Kobashigawa J, Starling R, Russell S, Uber PA, Parameshwar J, et al. Listing criteria for heart transplantation: International Society for Heart and Lung Transplantation guidelines for the care of cardiac transplant candidates–2006. J Heart Lung Transplant 2006. Sep;25(9):1024–42. [DOI] [PubMed] [Google Scholar]
  • 15.Costard-Jäckle A, Fowler MB. Influence of preoperative pulmonary artery pressure on mortality after heart transplantation: testing of potential reversibility of pulmonary hypertension with nitroprusside is useful in defining a high risk group. J Am Coll Cardiol 1992. Jan;19 (1):48–54. [DOI] [PubMed] [Google Scholar]
  • 16.Soliman OII, Akin S, Muslem R, Boersma E, Manintveld OC, Krabatsch T, et al. Derivation and Validation of a Novel Right-Sided Heart Failure Model After Implantation of Continuous Flow Left Ventricular Assist Devices: The EUROMACS (European Registry for Patients with Mechanical Circulatory Support) Right-Sided Heart Failure Risk Score. Circ 2018;137(9):891–906. [DOI] [PubMed] [Google Scholar]
  • 17.Kiernan MS, Grandin EW, Brinkley M Jr, Kapur NK, Pham DT, Ruthazer R, et al. Early Right Ventricular Assist Device Use in Patients Undergoing Continuous-Flow Left Ventricular Assist Device Implantation: Incidence and Risk Factors From the Interagency Registry for Mechanically Assisted Circulatory Support. Circ Heart Fail 2017;10 (10):e003863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Atluri P, Goldstone AB, Fairman AS, MacArthur JW, Shudo Y, Cohen JE, et al. Predicting right ventricular failure in the modern, continuous flow left ventricular assist device era. Ann Thorac Surg 2013;96 (3):857–63. discussion 863–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Kalogeropoulos AP, Kelkar A, Weinberger JF, Morris AA, Georgiopoulou VV, Markham DW, et al. Validation of clinical scores for right ventricular failure prediction after implantation of continuous-flow left ventricular assist devices. J Heart Lung Transplant 2015;34 (12):1595–603. [DOI] [PubMed] [Google Scholar]
  • 20.Frankfurter C, Molinero M, Vishram-Nielsen JKK, Foroutan F, Mak S, Rao V, et al. Predicting the Risk of Right Ventricular Failure in Patients Undergoing Left Ventricular Assist Device Implantation: A Systematic Review. Circ Heart Fail 2020;13(10):e006994. [DOI] [PubMed] [Google Scholar]
  • 21.Houston BA, Shah KB, Mehra MR, Tedford RJ. A new “twist” on right heart failure with left ventricular assist systems. J Heart Lung Transplant 2017;36(7):701–7. [DOI] [PubMed] [Google Scholar]
  • 22.Kremer N, Rako Z, Douschan P, et al. , Unmasking right ventricular-arterial uncoupling during fluid challenge in pulmonary hypertension [e-pub ahead of print], J Heart Lung Transplant, 41, 2022, 345–355, doi: 10.1016/j.healun.2021.11.019, accessed August 1, 2022. [DOI] [PubMed] [Google Scholar]
  • 23.Kormos RL, Antonides CFJ, Goldstein DJ, Cowger JA, Starling RC, Kirklin JK, et al. Updated definitions of adverse events for trials and registries of mechanical circulatory support: A consensus statement of the mechanical circulatory support academic research consortium. J Heart Lung Transplant 2020;39 (8):735–50. [DOI] [PubMed] [Google Scholar]
  • 24.Nochioka K, Querejeta Roca G, Claggett B, Biering-Sørensen T, Matsushita K, Hung CL, et al. Right Ventricular Function, Right Ventricular-Pulmonary Artery Coupling, and Heart Failure Risk in 4 US Communities: The Atherosclerosis Risk in Communities (ARIC) Study. JAMA Cardiol 2018;3(10):939–48. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Saeed D, Muslem R, Rasheed M, Caliskan K, Kalampokas N, Sipahi F, et al. Less invasive surgical implant strategy and right heart failure after LVAD implantation. J Heart Lung Transplant 2021;40(4):289–97. [DOI] [PubMed] [Google Scholar]
  • 26.Teuteberg JJ, Cleveland JC Jr, Cowger J, Higgins RS, Goldstein DJ, Keebler M, et al. The Society of Thoracic Surgeons Intermacs 2019 Annual Report: The Changing Landscape of Devices and Indications. Ann Thorac Surg 2020. Mar;109(3):649–60. [DOI] [PubMed] [Google Scholar]
  • 27.Rivas-Lasarte M, Kumar S, Derbala MH, Ferrall J, Cefalu M, Rashid SMI, et al. Prediction of right heart failure after left ventricular assist implantation: external validation of the EUROMACS right-sided heart failure risk score. Eur Heart J Acute Cardiovasc Care 2021. Oct 1;10 (7):723–32. [DOI] [PubMed] [Google Scholar]
  • 28.Deswarte G, Kirsch M, Lesault PF, Trochu JN, Damy T. Right ventricular reserve and outcome after continuous-flow left ventricular assist device implantation. J Heart Lung Transplant 2010;29 (10):1196–8. [DOI] [PubMed] [Google Scholar]
  • 29.Gonzalez MH, Wang Q, Yaranov DM, Albert C, Wolski K, Wagener J, et al. Dynamic Assessment of Pulmonary Artery Pulsatility Index Provides Incremental Risk Assessment for Early Right Ventricular Failure after Left Ventricular Assist Device. J Card Fail 2021;27 (7):777–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Molina EJ, Shah P, Kiernan MS, Cornwell WK 3rd, Copeland H, Takeda K, et al. The Society of Thoracic Surgeons Intermacs 2020 Annual Report. Ann Thorac Surg 2021;111(3):778–92. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Figures
Tables

RESOURCES