. 2023 Dec 13;2023:8907315. doi: 10.1155/2023/8907315

In-Hospital and 1-Year Clinical Results from the French Registry Using Polymer-Free Sirolimus-Eluting Stents in Acute Coronary Syndrome and Stable Coronary Artery Disease

Quentin Landolff 1,, Marine Quillot 2, Fabien Picard 3, Patrick Henry 4, Georgios Sideris 4,5, Olivier Bizeau 6, Christophe Piot 7, Bernard Jouve 8, Jérôme Rischner 9, Mourad Mejri 10, Claude Charmasson 11, Raphael Lasserre 12, Hervé Pouliquen 13, Thierry Joseph 14, Jacques Monsegu 15, Bernard Karsenty 16, Victoria Martin Yuste 17, Nicolas Richet 18, Guy Lapeyre 19, Fabrizio Beverelli 20, Farzin Beygui 21, René Koning 1
PMCID: PMC10733033  PMID: 38125031

Abstract

Objectives

The aim of this postmarket clinical study was to assess the safety and efficacy of the latest generation polymer-free sirolimus-eluting stents (PF-SES) in an all-comers population comparing outcomes in stable coronary artery disease (CAD) versus acute coronary syndrome (ACS) in France.

Background

The efficacy and safety of the first-generation PF-SES have already been demonstrated by randomized controlled trials and “all-comers” observational studies.

Methods

For this all-comers observational, prospective, multicenter study, 1456 patients were recruited in 22 French centers. The primary endpoint was target lesion revascularization (TLR) rate at 12 months and secondary endpoints included major adverse cardiac events (MACE) and bleeding.

Results

895 patients had stable CAD and 561 had ACS. At 12 months, 2% of patients had a TLR, with similar rates between stable CAD and ACS (1.9% vs 2.2%, p = 0.7). The overall MACE rate was 5.2% with an expected higher rate in patients with ACS as compared to those with stable CAD (7.3% vs 3.9%, p = 0.007). The overall bleeding event rate was 4.5%, with similar rates in stable CAD as compared to ACS patients (3.8% vs 5.6%, p = 0.3). Dual antiplatelet therapy (DAPT) interruptions prior to the recommended duration occurred in 41.7% of patients with no increase in MACE rates as compared to patients who did not prematurely interrupt DAPT (3.9% vs 6.1%, p = 0.073).

Conclusions

The latest generation PF-SES is associated with low clinical event rates in these all-comers patients. There was a high rate of prematurely terminated DAPT, without any effect on MACE at 12 months. This trial is registered with NCT03809715.

1. Introduction

Most commercially available coronary drug-eluting stents (DES) have polymer coatings, whether durable or resorbable, to fix the drug and slow down its elution. Some DES, such as the latest generation PF-SES Coroflex® ISAR NEO (B. Braun) device, an improved version of its predecessor the first-generation PF-SES Coroflex® ISAR DES which was extensively studied in the ISAR 2000 registry, do not have polymers.

The effectiveness and clinical safety of this concept (polymer-free) have already been demonstrated in two large-scale studies [1, 2]. New clinical data have confirmed the effectiveness of this device [36].

The objective of this postmarket clinical study was to assess the safety and effectiveness endpoints of the latest generation PF-SES in an unselected large patient population comparing outcomes in stable coronary artery disease (CAD) and acute coronary syndrome (ACS) in a French cohort.

2. Methods

2.1. Study Design and Patient Population

This study is an all-comers observational, nonrandomized, prospective, multicenter, postmarket registry (ClinicalTrials.gov Identifier: NCT03809715). A total of 1456 nonconsecutive patients of which 895 patients had stable CAD and 561 patients had ACS including 205 with ST-elevation myocardial infarction (STEMI) were recruited in 22 French centers.

The inclusion criteria were as follows: patients who were at least 18 years of age with significant coronary lesions (>50% diameter stenosis) and patients fulfilling the standard recommendations for percutaneous coronary intervention (PCI) based on the last European Society of Cardiology (ESC) recommendations. Exclusion criteria were as follows: intolerance to sirolimus and/or probucol, allergy to the coating's components, pregnancy and lactation, complete occlusion of the treated vessel due to a failed recanalization, cardiogenic shock, hemorrhagic diathesis or another disorder such as gastrointestinal ulceration or cerebral circulatory disorders which restrict the use of platelet aggregation inhibitor therapy and anticoagulation therapy, emergency cardiac surgery decision after myocardial infarction, patients with an ejection fraction of <30%, culprit coronary artery reference diameter <2.00 mm, and treatment of the left coronary stem.

The primary endpoint was the accumulated target lesion revascularization (TLR) rate at 12 months. Patients needed to be symptomatic with proof of ischemia for TLR. In addition, onsite quantitative coronary angiography (QCA) or “angiographic eyeballing” was used to determine at least a 70% diameter stenosis at the initial angioplasty site. The secondary endpoints during hospitalization were as follows: major adverse cardiac events (MACE) including target lesion revascularization (TLR), myocardial infarction (MI) (defined by a new event documented with elevated cardiac enzymes during the follow-up; elevated enzyme levels during the hospital stay are not considered as a new event), and cardiac death (defined by a proven cardiac death), and those at 12 months were as follows: MACE including TLR, MI, all-cause death, and definite and probable stent thrombosis.

Clinical and angiographic patient data were collected prospectively and recorded in a pseudonymized fashion (only the hospital could identify the patient), to meet all requirements regarding the updated general data protection regulation. Written informed consent was obtained from all patients before the procedure. The study will be presented to at least one ethics committee. National requirements for obtaining several ethics committees' opinions or to oblige the physician to obtain ethical counseling from the responsible ethics committee will be considered appropriate. In France, for our registry, the study complies with the Helsinki Declaration, in its most recent version, and was approved by the French National Agency for the Safety of Medicines and Health Products.

2.2. Stent Design, Procedure, and Follow-Up

The latest generation PF-SES Coroflex® ISAR NEO (B. Braun), without polymer, was therefore the device used in our study. Instead of using a polymer, a nonpolymer coating technology is employed: the excipient probucol to gradually release the drug and the active drug sirolimus on the abluminal side only, which is a macrolide and potent cytostatic inhibitor of smooth muscle proliferation. Its antiproliferative effects have been proven in a plethora of studies [7]. In addition, the stent backbone has ultrathin struts (55–65 µm), one of the thinnest struts thickness in the market. The differences between these two devices are mainly the modified stent architecture, i.e., ring and slightly widened connectors, increasing radio-opacity and subsequent visualization of the second-generation PF-SES as well as radial strength (50% increase) with less late recoil [8] (Figure 1).

Figure 1.

Figure 1

Comparison of two polymer-free sirolimus-eluting stents (PF-SES) with differences in stent architecture (top panel: Coroflex® ISAR and bottom panel: Coroflex® ISAR NEO, B. Braun) [4].

PCI was performed according to current clinical practice standards. Stent implantation procedures were carried out in compliance with approved indications as part of the manufacturer's CE certification. The implantation of the study device was recommended in an unselected patient population with stable CAD or ACS in France.

Dual antiplatelet therapy (DAPT) of 6 months for stable CAD and 12 months for ACS was recommended, but the final decisions concerning DAPT duration, type of P2Y12 receptor inhibitor, and the use of glycoprotein IIb/IIIa inhibitors were left to the physician's discretion [9].

2.3. Study Supervision, Data Management, and Definitions

The study was initiated by B. Braun. The scientific study committee was responsible for the development of the protocol and the writing of the manuscript. The committee had unrestricted access to all study data.

An independent clinical events committee, provided with all necessary and available data, adjudicated all major cardiovascular events and protocol endpoints. Procedural success was defined as a visually assessed <30% diameter stenosis without coronary dissection.

2.4. Statistical Analysis

Statistical analyses were conducted using SPSS version 23 (IBM, Munich, Germany).

Deviations from the study plan were assessed as protocol violations if at least one of the inclusion or exclusion criteria was not fulfilled. All patients who received at least one device were included in the intention-to-treat analysis. All patients who received at least one device without any protocol violation were included in the per-protocol analysis.

Patient data were identified by the patient number assigned during data entry, the study center, the time of treatment, and the time of recording.

The standard procedures for the comparison of variables and outcomes between groups (e.g., study center, country, and age) were planned as follows: Fisher's exact test for binary variables, χ2 test for categorical variables in k × 2 tables with k > 2, and Wilcoxon–Mann–Whitney U test or Student's t-test or ANOVA (analysis of variance) for continuous variables. In the case of small cell expectations occurring in the majority of the test situations, data of ordinal scale were to be analyzed by means of the U test instead of the χ2 test.

All statistical tests were two-tailed with the prespecified significance level of α = 5%.

The statistical evaluation of the study data was organized by B. Braun Vascular Systems, Berlin organization.

3. Results

3.1. Patient Demographic Data, Lesion Morphologies, and Procedural Details

From November 2018 to December 2020, 1456 patients were treated for 2079 coronary lesions with 2110 DES in 22 centers in France. The mean age of patients was 66.2 ± 11.3 years, and 74.2% of patients were men. Diabetes mellitus was found in 22.1% of patients in the ACS subgroup versus 28.9% of those in the stable CAD subgroup (p = 0.003). ACS accounted for 38.5% of PCI indications including STEMI for 14.1% and stable CAD for 61.5% (Table 1).

Table 1.

Patient demographic data, lesion morphologies, and procedural details of Coroflex® ISAR NEO (B. Braun).

Patient demographics
Variable All patients Stable CAD ACS p value
Number of patients 1456 895 561
205 STEMI
Number of lesions 2079 1302 777
Number of DES 2110 1323 787
DES per patient 1.40 ± 0.66 1.41 ± 0.67 1.38 ± 0.66 0.461
Age (years) 68.2 ± 11.3 68.4 ± 10.3 67.9 ± 12.9 0.384
Male gender 1080 (74.2%) 693 (77.4%) 387 (69.0%) <0.001
Diabetes
 Insulin-dependent 103 (7.1%) 77 (8.6%) 26 (4.6%) 0.003
 Non-insulin-dependent 280 (19.2%) 182 (20.3%) 98 (17.5%)
Hypertension 902 (62.0%) 571 (63.8%) 331 (59.0%) 0.067
History of smoking 437 (30.0%) 254 (28.4%) 183 (32.6%) 0.086
Renal insufficiency
 Stage 1 to 3a GFR > 60 ml/min/1.73 m2 1401 (96.3%) 864 (96.5%) 537 (95.7%) 0.760
 Stage 3b to 5 GFR < 60 ml/min/1.73 m2 55 (3.7%) 31 (3.5%) 24 (4.3%)
Dialysis dependence 20 (1.4%) 13 (1.5%) 7 (1.2%) 0.744

Treated lesions were complex (B2 or C, according to the American Heart Association/American College of Cardiology classification) in 39.3% of patients in the ACS subgroup versus 31.4% of those in the stable CAD subgroup (p < 0.001). The mean number of DES per patient was 1.40 ± 0.66 (Table 2).

Table 2.

Lesion and procedural characteristics.

Variable All patients Stable CAD ACS pvalue

Lesion morphologies
Number of patients 1456 895 561
205 STEMI
Number of lesions 2079 1302 777
Number of DES used 2110 1323 787
Target vessel lesions
 LAD 921 (44.3%) 579 (44.5%) 342 (44.0%) 0.057
 CX 521 (25.1%) 339 (26.0%) 182 (23.4%)
 RCA 634 (30.5%) 384 (29.5%) 250 (32.2%)
 Graft 3 (0.1%) 0 (0.0%) 3 (0.4%)
Thrombus burden 138 (6.6%) 13 (1.0%) 125 (16.1%) <0.001
Total occlusion 276 (13.3%) 77 (5.9%) 199 (25.6%) <0.001
Chronic total occlusion 45 (2.2%) 36 (2.8%) 9(1.2%) 0.015
Diffuse vessel disease 640 (30.8%) 416 (32.0%) 224 (28.8%) 0.136
Calcification 364 (17.5%) 234 (18.0%) 130 (16.7%) 0.471
Ostial lesion 164 (7.9%) 104 (8.0%) 60 (7.7%) 0.828
Bifurcations 341 (16.4%) 221(17.0%) 120 (15.4%) 0.362
Severe tortuosity 87 (4.2%) 55 (4.2%) 32 (4.1%) 0.907
De novo 2027 (97.5%) 1269 (97.5%) 758 (97.6%) 0.605
BMS-ISR 14 (0.7%) 10 (0.8%) 4 (0.5%)
DES-ISR 22 (1.1%) 15 (1.2%) 7 (0.9%)
Unknown stent ISR 16 (0.8%) 8 (0.6%) 8 (1.0%)

AHA/ACC type lesions
 A 197 (9.5%) 144 (11.1%) 53 (6.8%) <0.001
 B1 1169 (56.2%) 749 (57.5%) 420 (54.1%)
 B2 409 (19.7%) 228 (17.5%) 181 (23.3%)
 C 304 (14.6%) 181 (13.9%) 123 (15.8%)
Reference diameter (mm) 2.88 ± 0.47 2.84 ± 0.45 2.93 ± 0.48 <0.001
Lesion length 17.3 ± 7.5 17.3 ± 7.6 17.3 ± 7.5 0.915
Degree of stenosis (%) 83.4 ± 12.5 80.9 ± 12.5 87.7 ± 11.3 <0.001

Procedural data
Radial access 1349 (92.7%) 825 (92.3%) 524 (93.4%) 0.423
Predilation per lesion 1221 (58.7%) 766 (58.8%) 455 (58.6%) 0.902
DES per patient 1.40 ± 0.66 1.41 ± 0.67 1.38 ± 0.66 0.461
DES diameter (mm) 2.87 ± 0.46 2.83 ± 0.45 2.81 ± 0.47 <0.001
DES length (mm) 20.6 ± 8.0 20.4 ± 8.0 21.0 ± 8.0 0.083
DES inflation pressure (atm) 14.4 ± 2.4 14.3 ± 2.4 14.7 ± 2.5 0.001
Overall technical success per stent 2094 (99.2%) 1308 (98.9%) 787 (99.9%) 0.010

Most procedures were carried out using radial approach (92.7%). No data concerning the use of intracoronary imaging are available. Angiographic procedural success for all lesions treated was achieved in 99.2% of patients (Table 2).

The remaining baseline data are summarized in Tables 1 and 2.

3.2. Medical Therapy

All patients were treated with acetylsalicylic acid. There was no preloading in 50.7% of patients. Prior to angioplasty, patients were treated with clopidogrel in 26.4% of cases, ticagrelor in 21.3% of cases, and prasugrel in 1.6% of cases. After angioplasty, patients were treated with clopidogrel in 59.2% of cases, ticagrelor in 37.9% of cases, and prasugrel in 2.9% of cases. DAPT duration was inferior to 6 months in 5.8% of patients in the stable CAD population and 6–12 months in 87.7% of those in the stable CAD population. DAPT duration was superior to 12 months in 7.9% of those in the ACS population. Time to discharge was 2.5 ± 4.9 days. DAPT interruption occurred in 42.2% of patients. The reasons for DAPT interruption were on physician's advice (38.5%), due to complications (1.5%), patient's own decision (0.4%), and unknown causes (1.3%) (Table 3).

Table 3.

Medical therapy characteristics.

Comedication
Variable All patients Stable CAD ACS p value
Number of patients 1456 895 561
Patients with follow-up 1340 (92.0%) 836 (93.4%) 504 (89.8%) 0.014
Preloading
 No preloading 738 (50.7%) 523 (58.4%) 215 (38.3%) <0.001
 Clopidogrel 385 (26.4%) 289 (32.3%) 96 (17.1%)
 Ticagrelor 310 (21.3%) 74 (8.3%) 236 (42.1%)
 Prasugrel 23 (1.6%) 9 (1.0%) 14 (2.5%)
DAPT maintenance after angioplasty
 Clopidogrel 862 (59.2%) 665 (74.3%) 197 (35.1%) <0.001
 Ticagrelor 552 (37.9%) 212 (23.7%) 340 (60.6%)
 Prasugrel 42 (2.9%) 18 (2.0%) 24 (4.3%)
Anticoagulation
 No anticoagulation 1379 (94.7%) 850 (95.0%) 529 (94.3%) 0.180
 Vitamin K antagonist 21 (1.4%) 15 (1.7%) 6 (1.1%)
 NOAC 54 (3.7%) 30 (3.4%) 24 (4.3%)
 Others 2 (0.1%) 0 (0.0%) 2 (0.4%)
DAPT duration
 1 month 34 (2.3%) 27 (3.0%) 7 (1.3%) <0.001
 2 months 5 (0.3%) 1 (0.1%) 4 (0.7%)
 3 months 50 (3.4%) 22 (2.5%) 28 (5.0%)
 4 months 2 (0.1%) 1 (0.1%) 1 (0.2%)
 5 months 2 (0.1%) 1 (0.1%) 1 (0.2%)
 6–9 months 445 (30.6%) 364 (40.8%) 81 (14.5%)
 10–12 months 810 (55.7%) 419 (46.9%) 391 (69.8%)
 >12 months 101 (7.0%) 57 (6.4%) 44 (7.9%)
 Unknown 4 (0.3%) 1 (0.1%) 3 (0.5%)
Time to discharge in days 2.5 ± 4.9 1.8 ± 4.1 3.6 ± 5.8 <0.001
Time to follow-up or event in months 12.4 ± 0.7 12.4 ± 0.6 12.4 ± 0.9 0.513
DAPT interruption (<6 months for stable CAD, <12 months for ACS) 566 (42.2%) 454 (54.3%) 320 (63.5%) 0.001
DAPT interruption
 ≤6 months 107 (8.0%) 70 (8.4%) 37 (7.3%) 0.005
 >6 months 449 (33.5%) 305 (36.5%) 144 (28.6%)
 No interruption 784 (58.5%) 461 (55.1%) 323 (64.1%)
Reasons for DAPT interruption
 On physician's advice 515 (38.5%) 352 (42.1%) 163 (32.3%) 0.001
 Due to complications 20 (1.5%) 8 (1.0%) 12 (2.4%)
 Patient's own decision 6 (0.4%) 4 (0.5%) 2 (0.4%)
 Not interrupted 781 (58.3%) 458 (54.8%) 323 (64.1%)
 Unknown 18 (1.3%) 14 (1.7%) 4 (0.8%)

Table 3 summarizes the remaining data.

3.3. Clinical Outcomes at 12 Months

Follow-up was obtained in 92% of the patients at 12.4 ± 0.7 months (Table 4).

Table 4.

Clinical outcomes.

Clinical outcomes
Variable All patients Stable CAD ACS p value
Number of patients 1456 895 561
205 STEMI
Patients with follow-up 1340 (92.0%) 836 (93.4%) 504 (89.8%) 0.014
Time to follow-up or event in months 12.4 ± 0.7 12.4 ± 0.6 12.4 ± 0.9 0.142
In-hospital MACE 11 (0.8%) 4 (0.4%) 7 (1.2%) 0.086
In-hospital TLR 2 (0.1%) 1 (0.1%) 1 (0.2%) 0.739
 Re-PCI 1 (0.1%) 1 (0.1%) 0 (0.0%) 0.428
 CABG 1 (0.1%) 0 (0.0%) 1 (0.2%) 0.206
In-hospital MI 6 (0.4%) 1 (0.1%) 5 (0.9%) 0.024
In-hospital cardiac death 3 (0.2%) 2 (0.2%) 1 (0.2%) 0.853
12-month accumulated MACE 70 (5.2%) 33 (3.9%) 37 (7.3%) 0.007
Accumulated MACE interrupted DAPT 22 (3.9%) 9 (2.4%) 13 (7.2%) 0.073
Accumulated MACE no DAPT 48 (6.1%) 24 (5.2%) 24 (7.4%)
Accumulated TLR 27 (2.0%) 16(1.9%) 11 (2.2%) 0.735
 Re-PCI 25 (1.9%) 15 (1.8%) 10 (2.0%) 0.803
 CABG 3 (0.2%) 1 (0.1%) 2 (0.4%) 0.298
Accumulated MI 15 (1.1%) 6 (0.7%) 9 (1.8%) 0.072
Accumulated all-cause death 33 (2.5%) 14 (1.7%) 19 (3.8%) 0.017
Accumulated stent thrombosis 8 (0.6%) 4 (0.5%) 4 (0.8%) 0.468
Accumulated bleeding complications postdischarge
 No bleeding 1280 (95.5%) 804 (96.2%) 476 (94.4%) 0.330
 BARC 1–3 56 (4.2%) 30 (3.6%) 26 (5.2%)
 BARC >3 4 (0.3%) 2 (0.2%) 2 (0.4%)

With the exception of expected higher rates of MI in the ACS subgroup (0.9% vs 0.1%, p = 0.024), no significant difference was found between the ACS and the stable CAD subgroups at hospital discharge regarding the rates of MACE, TLR, and cardiac death (Table 4).

The rates of accumulated overall TLR and definite and probable stent thrombosis were low: 2% and 0.6%, respectively. As expected, rates of overall MACE (7.3% vs 3.9%, p = 0.007) and all-cause death (3.8% vs 1.7%, p = 0.017) were higher in patients in the ACS subgroup than those in the stable CAD subgroup. However, rates of overall TLR, definite and probable stent thrombosis, and overall bleeding were similar in both stable CAD and ACS settings (Table 4).

There was a DAPT interruption prior to the recommended time in 42.2% of cases, without a significant increase in the rate of MACE in all patients: 3.9% in the case of DAPT interruption versus 6.1% in the case of no DAPT interruption (p = 0.073). There was no bleeding in 96.4% of patients in the “no DAPT interruption” group and 93.2% of patients in the “DAPT interruption” group, BARC 1 in 1.2% and 3.2%, BARC 2 in 0.7% and 2.4%, BARC 3 in 0.9% and 0.6%, BARC 5 in 0.1% and 0.6% (unknown in 0.6% and 0%) (p = 0.001) (Tables 3 and 4).

The remaining baseline data are summarized in Table 4.

4. Discussion

Our data showed that the latest generation PF-SES with ultrathin struts is associated with high procedural success rates and low clinical event rates during hospitalization and at 12 months in all-comers patients, with moderate to highly complex lesions. There was a high rate of premature interruption of DAPT, without any effect on MACE or bleeding at 12 months, in this population with mostly low to intermediate complex lesions, highlighting the safety profile of this latest generation PF-SES.

At 12 months, our study confirmed previous large-scale studies demonstrating the safety and effectiveness endpoints of the latest generation PF-SES with low rates of TLR (2.0%), MACE (5.2%), MI (1.1%), and all-cause death (2.5%). Definite and probable stent thrombosis occurred in 0.6% of cases at 1 year. In the ISAR 2000 registry, studying the safety and effectiveness endpoints of first- and latest generation PF-SES, including in special cases such as left main angioplasty (which was an exclusion criterion in our study), 7000 patients had a 9-month low TLR and MACE rate (2.2% and 4.4%). MACE was significantly increased in the ACS population (6.3%) and left main lesions (6.7%). Probable or possible stent thrombosis occurred in 0.7% of cases [10]. These favorable results must be weighed by other studies such as the Australian registry using new-generation DES showing at 12 months an even lower mortality (1.26%), rate of TLR (1.16%), and rate of MACE (1.78%) [11]. Similarly, the French registry using biolimus-eluting stent showed at 12 months even lower TLR (1.7%), all-cause death (1.5%), MI (1.4%), and definite and probable stent thrombosis (0.4%) [12].

In our study, regarding ACS including STEMI (36.5%), the results showed at 1 year a low rate of TLR (2.2%), MI (1.8%), all-cause death (3.8%), and definite and probable stent thrombosis (0.8%). These very favorable results appeared to be better than those found with second-generation DES, e.g., with everolimus-eluting stents at 1 year. In a study consisting of 1899 patients, the rates were reported for TLR (3.0%), MI (2.4%), all-cause death (7.4%), and definite or probable stent thrombosis (1.6%) [13]. On the contrary, these very favorable results appeared to be worse than those found with thin-strut-durable-polymer-everolimus- and zotarolimus-eluting stents at 5 years in a study consisting of 1547 patients with very low TLR rate (2.0%), MI (5.3%), all-cause death (4.8%), and definite and probable stent thrombosis (2.1%) [14].

The strut thickness of PF-SES, investigated in our study (55–65 µm), corresponds to one of the thinnest struts currently in stent design technology. It is well known that the thickness of the struts increases the stent's thrombogenicity and restenosis. Reducing struts' thickness could reduce thrombogenicity and vascular inflammation and improve the clinical impact of the devices. This could explain the good results found in our study. These results are confirmed at 1 year by a recent meta-analysis which showed that second-generation DES with ultrathin struts (<70 µm) had better outcomes than thicker second-generation DES in terms of target lesion failure (TLF) [6]. The long-term benefits of ultrathin struts DES have been confirmed by two recent large meta-analyses [7, 8]. The first meta-analysis, pooled data from 18 publications from 10 randomized trials, showed a reduction of 12% in TLF at 2 years and 19% at 3 years after ultrathin strut stent implantation [7]. The second meta-analysis showed the same data with the use of ultrathin struts DES after an average follow-up of 2.5 years with a 15% reduction in the rate of TLF essentially linked to the 25% reduction of TLR, without any difference in terms of MACE [8].

The pathogenesis of delayed re-endothelialization after DES implantation is not fully elucidated. It appears that the presence of some polymers may induce an inflammatory response that delays vascular healing [15]. The development of polymer-free DES was therefore motivated by the intention to obtain healing comparable to that of bare metal stent (BMS), leading to both low restenosis and low late thrombosis rates. No data concerning the use of intracoronary imaging in our registry are available, as interventional cardiologists were not asked to fill in this index in the database. The randomized FRIENDLY OCT study showed that strut re-endothelialization was significantly higher for the latest generation PF-SES as compared to a sirolimus-eluting stent with biodegradable polymer. There was no significant difference in terms of cardiovascular events at 12 months [5]. These results suggest faster vascular healing with the latest generation PF-SES associated with potential clinical benefits.

Using clopidogrel rather than ticagrelor had no effect on MACE, bleeding, and thrombotic events in stable CAD or ACS patients [16, 17]. As observed in our study, compared with 6 months of DAPT, 3–6 months of DAPT did not increase the risk of cardiovascular death, myocardial infarction, target vessel revascularization (TVR), definite and probable stent thrombosis, and Bleeding Academic Research Consortium (BARC) 1 year after PF-SES implantation [18, 19]. In contrast to our study, DAPT interruption prior to the recommended time could lead to an increase in clinical event rates (MACE, TLR, MI, accumulated mortality) [20].

This large French multicenter (22 centers) observational study has some limitations: it is a registry, not a randomized trial, since the PF-SES was not compared to another DES in a control group. The overall low event rate could be due in part to the population included with mostly low to intermediate complex lesions (e.g., A and B1 lesion types in more than two-thirds of cases). There was no learning curve for the latest generation PF-SES implantation, as it was similar to other contemporary DES. The duration of DAPT was left to the operator's judgment at that time; however, it did not affect TLR, MACE, and bleeding at 12 months.

5. Conclusion

The latest generation PF-SES with ultrathin struts is associated with high procedural success rates and low ischemic or bleeding events during the hospital stay and at 12-month follow-up. This all-comers patient population with stable CAD or ACS was also partly treated in moderate to highly complex lesions. There was a high rate of prematurely terminated DAPT, with no effect on MACE or bleeding at 12 months, in this population with mostly low to intermediate complex lesions, highlighting the highly favorable security profile of the DES.

Acknowledgments

The authors thank all investigators who participated in the registry, Florian Krackhardt, Inigo Martinez-Luengas, and Mark Rosenberg for their contribution to the independent clinical events committee, Matthias Waliszewski and Audrey Michel (B. Braun) for providing the data required to write the manuscript, and Naomi Stone (native professional and English speaker) for her help in editing the manuscript.

Data Availability

Access to the clinical data is restricted for two reasons. Ethics approval was provided for the planned analyses only. Additional pooling of data with other sources was excluded. Moreover, the data set is proprietary by definition.

Conflicts of Interest

Quentin Landolff, Marine Quillot, Fabien Picard, Georgios Sideris, Raphael Lasserre, and René Koning report consulting fees from B. Braun. The other authors have no conflicts of interest to declare.

Authors' Contributions

All authors listed meet the authorship criteria according to the latest guidelines of the International Committee of Medical Journal Editors. All authors are in agreement with the manuscript.

References

  • 1.Kufner S., Sorges J., Mehilli J., et al. ISAR-TEST-5 investigators. Randomized trial of polymer-free sirolimus- and probucol-eluting stents versus durable polymer zotarolimus-eluting stents: 5-year results of the ISAR-TEST-5 trial. JACC: Cardiovascular Interventions . 2016;9(8):784–792. doi: 10.1016/j.jcin.2016.01.009. [DOI] [PubMed] [Google Scholar]
  • 2.Krackhardt F., Kočka V., Waliszewski M. W., et al. Polymer-free sirolimus-eluting stents in a large-scale all-comers population. Open Heart . 2017;4(2):p. 592. doi: 10.1136/openhrt-2017-000592. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Otaegui I., Pérez de Prado A., Massotti M., et al. Intrapatient randomization to study strut coverage in polymer-free versus biodegradable-polymer sirolimus-eluting stent implantations. Journal of the American College of Cardiology: Cardiovascular Interventions . 2020;13(7):899–900. doi: 10.1016/j.jcin.2019.11.020. [DOI] [PubMed] [Google Scholar]
  • 4.Bangalore S., Toklu B., Patel N., Feit F., Stone G. W. Newer-generation ultrathin strut drug-eluting stents versus older second-generation thicker strut drug-eluting stents for coronary artery disease. Circulation . 2018;138(20):2216–2226. doi: 10.1161/circulationaha.118.034456. [DOI] [PubMed] [Google Scholar]
  • 5.Hussain Y., Gaston S., Kluger J., et al. Long term outcomes of ultrathin versus standard thickness second-generation drug eluting stents: meta-analysis of randomized trials. Catheterization and Cardiovascular Interventions . 2022;99(3):563–574. doi: 10.1002/ccd.29866. [DOI] [PubMed] [Google Scholar]
  • 6.Madhavan M. V., Howard J. P., Naqvi A., et al. Long-term follow-up after ultrathin vs. conventional 2nd-generation drug-eluting stents: a systematic review and meta-analysis of randomized controlled trials. European Heart Journal . 2021;42(27):2643–2654. doi: 10.1093/eurheartj/ehab280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Fontes I., Barrow D. Du sirolimus au stent Cypher®: lesétapes d’une victoire contre la restenose. Annales de Cardiologie et d’Angeiologie . 2004;53(1):S3–S8. doi: 10.1016/s0003-3928(04)90001-7. [DOI] [PubMed] [Google Scholar]
  • 8.Nuruddin A. A. B., Ahmad W. A. W., Waliszewski M., et al. Impact of coronary stent architecture on clinical outcomes: do minor changes in stent architecture really matter? Cardiol Therapeutics . 2021;10(1):175–187. doi: 10.1007/s40119-020-00204-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Valgimigli M., Bueno H, Byrne R. A., et al. ESC Scientific Document Group; ESC Committee for Practice Guidelines (CPG); ESC National Cardiac Societies. 2017 ESC focused update on dual antiplatelet therapy in coronary artery disease developed in collaboration with EACTS: the Task Force for dual antiplatelet therapy in coronary artery disease of the European Society of Cardiology (ESC) and of the European Association for Cardio-Thoracic Surgery (EACTS) European Heart Journal . 2018;39(3):213–260. doi: 10.1093/eurheartj/ehx419. [DOI] [PubMed] [Google Scholar]
  • 10.Krackhardt F., Kočka V., Waliszewski M., et al. Unrestricted use of polymer-free sirolimus eluting stents in routine clinical practice. Medicine (Baltimore) . 2020;99(8) doi: 10.1097/md.0000000000019119.19119 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Eccleston D. S., Chowdhury E., Rafter T., et al. Long-term outcomes of contemporary percutaneous coronary intervention with the xience drug-eluting stent: results from a multicentre Australian registry. Journal of Clinical Medicine . 2022;12(1):p. 280. doi: 10.3390/jcm12010280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Maupas E., Lipiecki J., Levy R., et al. Safety and efficacy outcomes of 3rd generation DES in an all‐comer population of patients undergoing PCI: 12‐month and 24‐month results of the e‐Biomatrix French registry. Catheterization and Cardiovascular Interventions . 2017;90(6):890–897. doi: 10.1002/ccd.27081. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Yoshikawa S., Ashikaga T., Miyazaki T., Kurihara K., Hirao K. Long-term efficacy and safety of everolimus-eluting stent implantation in Japanese patients with acute coronary syndrome: five-year real-world data from the tokyo-MD PCI study. Journal of Interventional Cardiology . 2019;2019:13. doi: 10.1155/2019/3146848.3146848 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jaguszewski M., Dörig M., Frangieh A. H., et al. Safety and efficacy profile of bioresorbable-polylactide-polymer-biolimus-A9-eluting stents versus durable-polymer-everolimus- and zotarolimus-eluting stents in patients with acute coronary syndrome. Catheterization and Cardiovascular Interventions: Official Journal of the Society for Cardiac Angiography and Interventions . 2016;88(6):E173–E182. doi: 10.1002/ccd.26617. [DOI] [PubMed] [Google Scholar]
  • 15.Byrne R. A., Joner M., Kastrati A. Polymer coatings and delayed arterial healing following drug-eluting stent implantation. Minerva Cardioangiologica . 2009;57(5):567–584. [PubMed] [Google Scholar]
  • 16.Krackhardt F., Waliszewski M., Kočka V., et al. Real-world dual antiplatelet therapy following polymer-free sirolimus-eluting stent implantations to treat coronary artery disease. Cardiovascular Drugs and Therapy . 2020;34(3):335–344. doi: 10.1007/s10557-020-06963-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Krackhardt F., Waliszewski M., Kočka V., et al. Correction to: real-world dual antiplatelet therapy following polymer-free sirolimus-eluting stent implantations to treat coronary artery disease. Cardiovascular Drugs and Therapy . 2022;36(6):1257–1258. doi: 10.1007/s10557-021-07290-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Jin U., Seo K.-W., Yang H.-M., et al. Efficacy and safety of 3 versus 6 Months of dual-antiplatelet therapy in patients implanted with a Coroflex ISAR stents: a prospective, multicenter, randomized clinical trial. Journal of Invasive Cardiology . 2022;34(9):E653–E659. doi: 10.25270/jic/21.00330. [DOI] [PubMed] [Google Scholar]
  • 19.Han J.-K., Hwang D., Yang S., et al. Comparison of 3- to 6-month versus 12-month dual antiplatelet therapy after coronary intervention using the contemporary drug-eluting stents with ultrathin struts: the HOST-IDEA randomized clinical trial. Circulation . 2023;147(18):1358–1368. doi: 10.1161/circulationaha.123.064264. [DOI] [PubMed] [Google Scholar]
  • 20.Krackhardt F., Waliszewski M., Wan Ahmad W. A., et al. Polymer-free sirolimus-eluting stent use in Europe and Asia: ethnic differences in demographics and clinical outcomes. PLoS One . 2020;15(1) doi: 10.1371/journal.pone.0226606.226606 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Access to the clinical data is restricted for two reasons. Ethics approval was provided for the planned analyses only. Additional pooling of data with other sources was excluded. Moreover, the data set is proprietary by definition.