Skip to main content
The Canadian Journal of Infectious Diseases & Medical Microbiology = Journal Canadien des Maladies Infectieuses et de la Microbiologie Médicale logoLink to The Canadian Journal of Infectious Diseases & Medical Microbiology = Journal Canadien des Maladies Infectieuses et de la Microbiologie Médicale
. 2023 Dec 18;2023:4441338. doi: 10.1155/2023/4441338

Radiology Department: A Potential Source of Multidrug-Resistant Microorganisms: A Cross-Sectional Study at Tertiary Hospital, Palestine

Zena Odeh 1,2, Safaa Abatli 3, Mohammad Qadi 3,
PMCID: PMC10749721  PMID: 38146333

Abstract

Introduction

Globally, healthcare facilities face a great challenge in the form of hospital-acquired infections (HAIs). Aside from the morbidity and mortality they cause, these illnesses are also extremely costly. Research on infection transmission in the medical field has been considerable, but not so much in the radiology department.

Aim

This study aims to identify the presence of multidrug-resistant (MDR) microbes on surfaces that are frequently touched in computed tomography (CT), magnetic resonance imaging (MRI), ultrasound (US), plain X-ray examination rooms, and portable radiography that are susceptible to contamination as well as to investigate the potential dangers of contracting MDR organisms to patients and healthcare providers. Materials and Method. In this study, 160 swab samples were collected from the radiology department during May and June 2022. Samples were obtained from 80 predefined surfaces twice within and outside of CT and MRI examination rooms as well as from US and plain X-ray machines and portable X-ray machines. Samples were taken at 7:00 a.m. using cotton swabs following the regular cleaning procedure. Bacterial colony-forming units (CFUs) per square centimeter were calculated after swabbing a 100 cm2 surface.

Results

Nearly all of the surfaces tested had bacterial CFUs. The highest contamination rate was found on keyboards ranging from (1.2–8) CFU/cm2, the sides of patient tables (1.2–20) CFU/cm2, knee coil (2.4–3) CFU/cm2, and patient leg supports (1.2–8) CFU/cm2. A noticeable increase in the contamination was noticed in June compared to May, and this was consistent with the increase in the number of isolated patients in the hospital, the workload in the radiology department, and the number of patients referred to the hospital. In our study, none of the examined sites showed contamination with MDR Gram-negative bacteria such as extended-spectrum beta-lactamases producing Enterobacterales (ESPL) or Carbapenemase-producing Enterobacterales (CPE). On the other hand, methicillin-resistant Staphylococcus (MRS), vancomycin-resistant Staphylococcus (VRS), and vancomycin-resistant Enterococcus (VRE) were detected.

Conclusion

All of the radiology department equipment and sites could be a source of bacterial infection including MDR, so the obligatory and committed disinfection protocol must be revised and implemented in the morning and between patients.

1. Introduction

These days, the importance of infection control and prevention in health care centers cannot be overstated. Hospitals and health care centres often have surfaces contaminated with microorganisms, which can lead to serious healthcare-associated infections (HAIs) affecting both patients and health care workers. While it is well established that HAIa can cause serious diseases, there are still gaps in our understanding of how these bacteria are transmitted to patients. This gap underscores the critical importance of strict hygiene protocols and infection control measures within hospital and health care centre settings to reduce the risk of HAIs [1].

The radiology department and the use of different radiological equipment, including computed tomography (CT), magnetic resonance imaging (MRI), ultrasound (US), plain X-rays, and portable X-ray machines, have a vital role in diagnosing and monitoring different conditions [2, 3]. Cross-contamination is a critical concern in this area since patients are transferred to the radiology department from various clinics and departments with a wide spectrum of medical conditions. These patients' susceptibility to illness is already high due to their underlying medical conditions. Because these patients may come into direct or indirect contact with health care workers (HCWs) and contaminated surfaces and equipment, the risk of nosocomial infection transmission among them increases, making them vulnerable to HAIs [3, 4].

In recent years, the incidence of HAIs in radiology departments has witnessed a concerning increase. These infections jeopardize the safety of both HCWs and patients, further complicating radiographic investigation or intervention [5, 6].

HAIs are commonly referred to as nosocomial infections, which are defined as infections acquired within 48 hours of admission to a healthcare facility [6, 7]. In contrast to community-acquired infections (CAIs), these infections usually occur as a result of pathogens taking advantage of patients whose normal defenses against infection are compromised [3]. Extensive literature has highlighted the Centres for Disease Control and Prevention's (CDC) classification of nosocomial infection sites into 13 types with up to 50 infection sites [8]. These infections put hospitalized patients at a big risk, leading to significant morbidity and mortality, particularly in low-income countries rather than in high-income countries around the world [4, 7, 9].

High-touch surfaces in patient rooms, such as bed controls, bed rallies, call buttons, and bedside tray tables, represent a critically important multidrug-resistant organism (MDRO) reservoir and increase the risk of acquisition by other patients, visitors, and hospital staff who are exposed to them. The presence of MDR bacteria within hospital settings poses a significant threat that extends to the selection of empiric antibiotics that target MDR bacteria. This can perpetuate a vicious cycle of increasing antimicrobial resistance [2, 10, 11].

Several studies have demonstrated that disinfection plays a significant role in controlling microbial carriage in people who are not harboring MDROs. Beyond disinfection, there are several essential infection prevention methods, such as environmental cleaning, hand hygiene, contact precautions, and active screening. Disinfection offers a universal solution by safeguarding both MDRO carriers and noncarriers [12].

Disinfection is becoming increasingly significant in most hospitals because the number of patients harboring MDROs asymptomatically is growing over time [13]. Increased research on MDROs persistence in the hospital environment and subsequent transmission has recently resulted in a greater emphasis on hospital environmental hygiene [1416].

Drug-resistant and MDROs include methicillin-resistant Staphylococcus aureus (MRSA), methicillin-resistant Staphylococcus epidermis (MRSE), vancomycin-resistant Staphylococcus aureus (VRSA), vancomycin-resistant Enterococcus (VRE), extended-spectrum beta-lactamases producing Enterobacterales (ESPL), carbapenemase-producing Enterobacterales (CPE), carbapenem-resistant Acinetobacter baumannii (CR-AB), and carbapenem-resistant Pseudomonas aeruginosa (CR-PA) [5, 16, 17].

In a recent study, cleaning surfaces by using a sodium lauryl sulfate-based detergent can prevent MRSA transmission in health-care settings and reduce the risk of surface contamination at hospitals in general and radiology departments in particular [18]. In a previous investigation involving bacterial detection, samples were obtained from CT equipment, and the results indicated that the CT wrap was the most contaminated with germs, prompting the development of a novel cleaning procedure [6].

The real pandemic outbreak of COVID-19, which killed several million people and had major global economic consequences, is an indication that much more work is necessary to tackle infectious diseases and the growing global issue of antibiotic resistance [19]. In the case of hospitalized patients, the case-fatality rate linked with bacteremia ranges from 35% to 50% and is usually associated with MDR Gram-negative bacteria such as ESBL and CRE [20].

Globally, prevalent carbapenemases are found in Enterobacteriaceae, including Klebsiella pneumoniae carbapenemases (KPC) [21]. K. pneumoniae, which produces KPC, is a worldwide threat [22]. In a recent study, bacterial contamination was found in almost all selected areas in the radiology department, including MRI and CT equipment; fortunately, no MDROs such as MRSA, ESPL, or CPE were detected [7].

The spread of MDROs and HAIs presents a major challenge to the global healthcare community especially developing countries, as they have much higher risks of HCAIs with a ratio of 20 : 1 as compared to developed countries [23].

This cross-sectional study was conducted at a tertiary hospital in Palestine, where we investigated the presence of multidrug-resistant organisms on highly touched surfaces in the radiology department as well as the potential danger of patients and healthcare workers contracting multidrug-resistant organisms. Swab samples were taken from the commonly hand-touched sites in the department of radiology and cultured at a microbiology lab. Antimicrobial resistance was tested.

2. Materials and Methods

2.1. Study Design

A cross-sectional study design was carried over at the radiology department at Tertiary Hospital, Palestine.

2.2. Study Population

The commonly hand-touched sites were assessed according to the previous studies [7], and an adapted flow chart was created. Briefly, swab samples targeted the surfaces inside and outside: plain X-ray, CT, MRI, US, and portable X-ray instruments, in addition to examination rooms, as shown in Table 1.

Table 1.

Commonly hand-touched sites.

MRI CT CANON CT SEMENS Interventional US US for general use X-ray Portable X-ray primax Portable X-ray care stream
Canter of patient Table 1 Head pillow Head support Linear probe Linear probe X-ray cassette X-ray cassette X-ray cassette
Head support RT side edge of patient table RT side edge of patient table Curve linear probe Curve linear probe Wall Bucky control panel Touch screen Touch screen
Head coil 1 Centre of patient table Centre of patient table Key board of US US keyboard Wall Bucky Hands of machine Hand of machine
Anaesthesia machine Injector control panel Another table in the CT room Touch screen of US Small touch screen Touch screen of X-ray tube
RT side gantry control panel CT keyboard Leg support Table in control room Prep medication table
X-ray cassette RT side gantry control panel RT side gantry control panel Door of examination room Centre of patient table
LT side gantry control panel LT side gantry control panel Table of US tools RT side edge of patient table
Centre of gantry Centre of gantry Centre of gantry Hand of US LT side edge of patient table
Ear support (headphone) Surface of emergency trolley Prep medication table Head pillow Key board in control room
Surface coil 1 Control room keyboard Keyboard in control room Mouse in the control room Mouse
RT side edge of patient Table 1 LT side edge of the patient table LT side edge of the patient table Keyboard in control room X-ray keyboard
Surface coil 2 Injector touch screen Another table in the CT room Large touch screen
Head coil 2 CT mouse CT mouse Patient table
Knee coil Mouse in the control room
Centre of patient Table 2
RT side edge of patient Table 2
LT side edge of patient Table 2
Leg support 1
Leg support 2

2.3. Study Time

The sample swabs were collected in two separate time periods: the first one was in May 2022 and the other one was in June 2022. The same sites were investigated, and these sites are the commonly hand-touched sites surface described in Table 1.

2.4. Study Sample and Settings

160 swab samples were collected as described in Table 1 to cover the 80 determined commonly hand-touched sites in May and June. Each swab sample covered an area of 100 cm2 (around 10 cm  10 cm). Samples were taken in each month, every other day.

Sampling was carried out with a sterile swab. The sterile swab was presoaked in sterile normal saline and inserted directly after swabbing into a falcon tube with 2 ml phosphate buffer saline (PBS) and transported immediately to the microbiology lab at the Faculty of Medicine and Health Sciences at An-Najah National University.

2.5. Bacterial Culture and Detection

Each tube was vortexed for one minute, and 100 µl were cultured into each of the following agar plates:

  1. Blood agar.

  2. Chocolate agar.

  3. MacConkey agar.

  4. Mannitol salt agar (MSA).

  5. MSA + Oxacillin (6 µg/ml)⟶Methicillin-resistant Staphylococcus (MRS) detection.

  6. MacConkey agar + Meropenem (1 µg/ml)⟶ Carbapenem-resistant Enterobacteriaceae (CRE) detection.

  7. MacConkey agar + Cefotaxime (1 µg/ml)⟶ Extended Spectrum Beta Lactamase (ESBL) detection.

  8. Bile esculin + Vancomycin (6 µg/ml)⟶ Vancomycin-resistant Enterococcus (VRE) detection.

Antibiotics used in this study were purchased from Sigma Aldrich, while the media were purchased from Oxoid. Media with antibiotics were prepared as described in the Clinical and Laboratory Standard Institute (CLSI) 2021 and as described in the literature [24, 25].

All plates were incubated at 37 degree/24 hours aerobically, while chocolate agar plates were incubated in 5% CO2.

American Type Culture Collection (ATCC) strains (E. coli ATCC 25922 and S. aureus ATCC 25923) and clinically confirmed strains (MRSA, ESBL, CRE and VRE) were used as controls for the prepared media in each preparation.

Finally, all the identified multidrug-resistant isolates were confirmed as such through antibiotic susceptibility testing, following the CLSI protocols.

2.6. Vancomycin Resistance

The screening method was used to detect vancomycin resistance. In short, bile esculin media was prepared with a concentration of 6 µg/ml; further confirmation was approached through E-test. E test was applied for those strains that needed to be screened for vancomycin sensitivity: MRS detected using MSA + oxacillin and colonies grown on bile esculin agar with vancomycin.

2.7. Contamination Rate Calculation

Simply put each commonly encountered site in our study was swabbed with a sterile, premoistened PBS swab, targeting a 100 cm2 (10 cm  10 cm) area. After swabbing, the swab was placed in an already prepared falcon tube containing 2 ml sterile PBS, vortexed well, and 100 µl was transferred and cultivated on the prepared media's surface.

Following the incubation period, colonies on each plate were counted, and the findings were given as CFU/cm2, taking the swabbed area and volume of PBS buffer in the tubes into account (calculation CFU/cm2: counted colonies number  0.2), as shown in Table S3.

2.8. Ethical Approval

Ethical approval was taken from the institutional review board (IRB) at An-Najah National University and An-Najah National University Hospital.

3. Results

3.1. Bacterial Detection

Bacterial growth was nearly detected in all targeted sites. Regarding samples collected in May, bacterial growth was detected in 49/80 sites. Samples collected in June showed growth in 52/80 sites. Interestingly, in total, 60/80 sites showed bacterial growth as shown in Tables S1 and S2.

In the radiology department as well as all other inanimate objects, Gram-positive bacteria are predominant in and out of CT, MRI, plain X-ray, US rooms, and portable X-ray, while Gram-negative was only detected in four sites, as shown in Tables S1 and S2.

3.2. Total Bacterial Contamination

Regarding bacterial contamination ≥1 CFUs/cm2, in May, out of 80 sites examined, 21 sites (26%) had contamination, while in June, 20 sites (25%) were considered contaminated. In total, 29 sites showed contamination of ≥1 CFUs/cm2, as shown in Table 2.

Table 2.

Sites with the contamination rate of ≥1 CFU/cm2, either on blood or on chocolate agar.

Sample number Sample source May experiment June experiment
Contamination rate chocolate agar Contamination rate blood agar Contamination rate chocolate agar Contamination rate blood agar
2 MRI head support 0.6 2 0.4 0.2
4 Head pillow of CT CANON 0.6 0 2.2 0.4
6 Centre of patient table CT CANON 1.6 0.4 8.8 10
11 Keyboard of CT CANON 2 0.6 6 4
14 LT side gantry control panel CT CANON 0.4 1 0 0
15 Primax X-ray portable touch screen 0 0.6 1 6
25 Keyboard of US for general use 1.2 1.4 2.4 1.2
30 RT side of MRI patient table 4.8 1 0 0
33 Keyboard in CT CANON control room 3 0 0.6 1.6
34 CT Siemen's keyboard control room 4 8 6 6
44 Keyboard of interventional US 1 0.6 0.2 0
48 Patient leg support of CT semen's 0.2 0 0.6 1.2
51 Large touch screen for US for general use 0 0 12.8 0.4
52 Table of US for general use in control room 4.8 1.8 1.6 1.4
56 Hand of US for general use 1 0.2 0.8 0.4
58 Mouse of US for general use in control room 2.2 0.8 0.4 0.2
59 Keyboard in control room of US for general use 3.4 0.6 1 0.2
60 RT side of X-ray patient table 0.2 0 2.4 0.8
63 Hand of carestearm portable 8 0.2 0 0.8
64 Carestream touchscreen portable 0.2 0.2 5.2 5
66 LT side of patient table CT CANON 8 8 1.6 1.2
67 Touch screen of injecter in CT CANON control room 0.2 0 1.8 1.6
72 Knee coil 2.4 3 0.6 0.4
73 Centre of MRI patient Table 2 1.4 0 2.8 2.4
74 RT side of MRI patient Table 2 2.8 4.8 3 0.6
75 LT side of MRI patient Table 2 5.8 4.2 8.4 6
77 Keyboard of X-ray in control room 3.2 3.4 6 8
79 Patient leg support 1 2.2 2.4 0.4 0.4
80 Patient leg support 2 2 3.4 8 5.4

3.3. Gram-Positive: Staphylococcus (MSA-Growth)

Concerning contamination with bacteria that can grow on MSA which is mainly Staphylococcus, in May, out of 80 sites examined, 9 sites had a contamination rate of ≥1 CFUs/cm2, while in June, 18 sites were considered contaminated. In total, 19 sites showed contamination of ≥1 CFUs/cm2, as shown in Table 3.

Table 3.

Sites contaminated with bacteria that can grow on MSA agar during two months.

Sample number/month Sample source Mannitol salt agar (CFUs/cm2)
4/May and June Head pillow of CT CANON 0.2/0.4
5/June RT side of patient table CT CANON 0.2
6/May and June Centre of patient table CT CANON 1.6/4
7/May Head coil MRI 1 0.2
11/June Keyboard of CT CANON 3.4
14/May LT side gantry control panel CT CANON 0.2
15/June Primax X-ray portable touch screen 4
24/May and June Table of prep medication in X-ray room 0.2/0.2
25/May and June Keyboard of US for general use 1/1.4
26/May Patient table of US for general use 0.2
27/June Ear plugs of MRI 0.8
29/June MRI surface coil 1 0.2
30/May RT side of MRI patient table 2.2
31/June LT side of MRI patient Table 1 3
32/June Trolley emergency in CT CANON room 0.4
33/June Keyboard in CT CANON control room 1.8
34/June CT Siemens keyboard control room 6
35/June CT Siemens mouse control room 0.4
36/June CT Siemens inside gantry 0.2
38/June RT side edge of CT patient table 0.2
39/June CT Siemens head support 0.6
44/May Keyboard of interventional US 0.4
45/May Touch screen of interventional US 0.2
48/May and June Patient leg support of CT Siemens 0.6/0.2
49/June Linear probe for US general use 1.2
50/June Curve linear probe for US for general use 0.4
52/May and June Table of US for general use in control room 1.8/1.6
54/May and June Small touch screen for US general use 0.2/0.2
55/May and June Patient table tools in US for general use room 0.6/0.4
56/May and June Hand of US machine for general use 0.6/1
57/June Patient head pillow in US for general use room 0.6
58/May Mouse of US for general use in control room 0.6
59/May and June Keyboard in control room of US for general use 1.6/1
60/June RT side of X-ray patient table 0.8
62/June Keyboard in control room of X-ray 0.4
63/June Hand of Carestream portable 0.2
64/June Carestream touch screen portable 0.2
66/May and June LT side of patient table CT CANON 3.6/1.4
67/June Touch screen of injector in CT CANON control room 1.6
72/June Knee coil 0.8
73/June Centre of MRI patient Table 2 1.2
74/May and June RT side of MRI patient Table 2 2.2/1.6
75/May and June LT side of MRI patient Table 2 1.2/4.4
77/May and June Keyboard of X-ray in control room 0.6/5.6
78/June Mouse of X-ray in control room 0.8
79/May and June Patient leg support MRI 1 1.2/0.4
80/May and June Patient leg support MRI 2 0.2/5.4

3.4. Gram-Positive: MRS

Regarding the surface contamination with MRS, six sites in May and 13 sites in June showed growth on MSA + oxacillin, which means MRS is suspected to be present. Later on, all suspected colonies were confirmed as MRS after subculture, and cefoxitin resistance was demonstrated by the disk diffusion method. Out of 13 sites, only one site had a contamination rate ≥1 CFUs/cm2 in June and none in May. However, the site with a contamination rate ≥1 CFUs/cm2 is the centre of the patient table of the CT Canon with 1.4 CFUs/cm2, as shown in Table 4.

Table 4.

Sites contaminated with MRS on mannitol salt agar with oxacillin.

Sample number/month Sample source Mannitol salt agar with oxacillin (CFUs/cm2)
4/June Head pillow of CT CANON 0.4
5/June RT side of patient table CT CANON 0.2
6/May and June Centre of patient table CT CANON 0.2/1.4
11/June Keyboard of CT CANON 0.2
15/June Primax X-ray portable touch screen 0.4
34/May CT Siemens keyboard control room 0.4
55/June Patient table tools in US for general use room 0.4
59/June Keyboard in control room of US for general use 0.2
66/May and June LT side of patient table CT CANON 0.6/0.6
72/May Knee coil 0.2
74/May RT side of MRI patient Table 2 0.2
75/June LT side of MRI patient Table 2 0.6
79/May Patient leg support 1 0.4

3.5. Gram-Positive: Vancomycin-Resistant Staphylococcus (VRS)

After MRS detected and confirmed, all MRS isolates were tested for vancomycin sensitivity using E-test. In May, no VRS was detected, and all isolates were found to be vancomycin sensitive Staphylococcus. Surprisingly, in June, 7 out of 13 detected MRS were confirmed as VRS, namely, samples from head pillow of CT Canon, centre of patient table CT Canon, right side gantry control panel CT Canon, primax X-ray portable touch screen, patient table tools in US for general use room, keyboard in control room of US for general use, and left side of MRI patient Table 2.

3.6. Gram-Positive: VRE

VRE were detected in five of the examined sites, one site in May with a contamination rate <1 CFUs/cm2, while the other four times were in June, with a contamination rate of ≥1 CFUs/cm2, as shown in Table 5.

Table 5.

Sites harbors VRE in the radiology department, on bile esculin agar with vancomycin.

Sample number/month Sample source Bile esculin agar with vancomycin (CFUs/cm2)
4/June Head pillow of CT CANON 6
6/June Centre of patient table CT CANON 8
15/June Primax X-ray portable touch screen 2
26/May Patient table of US for general use 0.2
67/June Touch screen of injector in CT CANON control room 3

3.7. Gram-Negatives

Our results showed that only four Gram-negative bacteria were isolated from four different sites during the two cohorts done in May and June 2022. Three of these contaminated sites have contamination rate of <1 CFUs/cm2 during May, while the fourth contaminated site (centre of patient table–CT cannon) has a contamination rate of 1.4 CFUs/cm2 and this contamination was detected in June, as shown in Tables S1, S2 and 6.

Table 6.

Samples with growth of Gram-negative bacteria.

Sample number/month Sample source CFU/cm2
6/June Centre of patient table CT CANON 1.4
7/May Head coil MRI 1 0.2
27/May Ear plug of MRI 0.2
75/May LT side of MRI patient Table 2 0.2

3.8. Gram-Negative: ESBL and CRE

None of the equipment sites from the radiology department that were tested showed growth of ESBL and CRE, according to the data in Tables S1 and S2.

3.9. Contamination Rate

Because of the increasing number of patients and an increased number of referred patients in June, the contamination rate average of 1.3 CFUs/cm2 is greater than in May 0.79 CFUs/cm2, as shown in S1 and S2 Tables, while for the study period (May and June), no differences were noticed in the disinfectant materials used to disinfect the surfaces, the time, and frequency at which the surfaces were cleaned.

Growth conditions affect the contamination rate, as the data shown in Table S3, the contamination rate on chocolate agar was higher than the contamination rate on blood agar for almost all sites during the two cohorts. Chocolate agar has a higher contamination rate than blood agar mostly because chocolate agar contains lysed red blood cells with better growth for fastidious organisms and due to the fact that chocolate agar plates were incubated in an anaerobic environment [26].

The results showed a contamination median value greater than 3 CFU/cm2 from seven common surface sites tested in the CT, MRI, US, and plain X-ray; centre and sides of the examination table X-ray patient's tables, knee coils, MRI patient's leg support, and all of the radiology machine keyboards.

Alarming findings reveal that the highest contamination rate was found in the CT Canon's core of the patient table and the large touch screen of the US for general use, with 10 CFUs/cm2 and 12.8 CFUs/cm2, respectively. Another alarming piece of data shows the high contamination rate on the right and left side of MRI patient (Table 2) MRI patient leg support 2, keyboard in the X-ray control room, left side of patient table of CT Canon, CT Siemen's keyboard in the control room, and keyboard and table of US for general use; in both cohorts, as shown in Table S3.

4. Discussions

The findings of our study indicate that Gram-positive bacteria were more detected in the radiology department than Gram-negative ones, and this was expected. This result is consistent with previous investigations that have found that Gram-positive were more common than Gram-negative bacteria on inanimate surfaces in the radiology department [7]. It was demonstrated that Gram-positive bacteria have a stronger potential for surviving on inanimate surfaces and environment [27]. Furthermore, Gram-positive bacteria also make up a significant portion of the skin's microbiota [28]. In short, the higher occurrence of Gram-positive bacteria in the radiology department can be attributed to their survival characteristics and their presence on the skin.

In our study, the MRSA contamination rate on the inert surface was relatively high 28/160 (17.5%). On the other hand, in Sweden, swabs were taken from the bore, table, and wrap of two quaternary care inpatient CT scanners; the wrap was the most contaminated item on a CT scanner, and the prevalence of MRSA was significantly low [29]. However, in another study in Ireland, from 125 samples collected from the radiology department, MRSA was detected from one sample only, bore in the MRI gantry [30].

However, in earlier research on cassettes and lead aprons carried out in radiology departments across the United Kingdom, there was no evidence of MRSA [2]. In our research, no MRS was detected from the MRI gantry, but the other 27 samples taken from different sites of the radiology department's equipment were positive for MRS. MRSA was present on an X-ray cassette that had been utilized in the operation room [31]. In addition, our investigation showed no evidence of MRS and bacterial contamination on the X-ray cassette.

In general, the sample area makes up only a small portion of the overall surfaces, which may reduce the sensitivity of the test when attempting to identify resistant bacteria that are present in low numbers. In addition, the purpose of this investigation was to identify any contamination on the surfaces inside and outside the radiology examination rooms and equipment and determine whether they were more likely to be contaminated. In the majority of the CT, X-ray, and US examination rooms both inside and outside, as well as on the patient tables of the MRI machines, the keyboards contained a noticeable bacterial contamination rate that ranged 1–8 CFU/cm2.

This has additionally been shown to be the case in other research, as they showed that work station sites in the radiology department are 64.3% (9 of 14) contaminated with S. aureus and 21.4% (3 of 14) are contaminated with enteric organisms [7, 32]. It is probable that this is because medical staff members do not adequately disinfect their hands after dealing with the patient within the examination room or that they do not regularly disinfect the keyboards and patient tables. Both of these factors contribute to the spread of infection. There has been a lot of research on how important it is to practice good hand hygiene in order to prevent the spread of infection [33].

The simplest, most effective, and least expensive strategy to prevent the spread of microorganisms is to practice strict hand cleanliness. In our study, a large number of CFU/cm2 was found in most cases on the LT side of MRI patient (Table 2) the centre of patient table CT CANON, and the large touch screen for US for general use with 8.4, 10, and 12.8 CFUs/cm2; respectively. One possible explanation is that the patients' clothing had been in contact with their bodies for at least 15 minutes and HCWs hands when dealing with patients during examination. In addition, the examination tables' sides of MRI, CT, and X-ray had contamination rates ranging from 1–8.4 CFU/cm2. This may be due to the fact that patients sit on the examination table with their skin in close contact with the side of the examination table. The examination table's sides and centre, as well as the MRI knee coil and patient legs support, are most likely not thoroughly disinfected. In general, a large number of surfaces showed bacterial contamination that was higher than the permissible limit of one CFU/cm2. Meanwhile, other studies showed low numbers of CFUs/cm2 on the side of the tunnel of the MRI camera in the radiology department [7] and a decrease in bacterial growth presence in the magnetic field [34, 35].

The surfaces may not have been cleaned thoroughly enough, and this may be a cause of the infection. This could be the result of, for instance, an insufficient amount of staff education on infection management, or it could be because the cleaning is not effective. When deciding on a cleaning procedure, there are many factors to take into consideration. It should be efficient, but at the same time, it should not have any negative effects on either humans or the environment, and it should not be too expensive [36].

Alterations to cleaning procedures and the kind of materials used have varying effects on certain infections. A cleaning solution containing hydrogen peroxide is excellent against bacteria and viruses, but it is harmful to humans and cannot be used for continuous cleaning [37].

Self-disinfecting surfaces covered with copper and silver have also been studied, and this has been demonstrated to minimize HAI. For the pathogen, efficiency, the environment, and the economy to all benefit from disinfection procedures, further research is needed [38].

In the radiology department, fortunately, we detected a low number of CFU/cm2 approximately near zero CFU/cm2 on the side of the Siemens CT gantry, patient table of X-ray, head coil MRI, surface coils of MRI, and probes for interventional ultrasonography compared to other sites in the MRI examination room, the side of the X-ray patient table, and probes of US for general use. Since patients contact this location practically every time they use these machines, it is highly unlikely that it is disinfected more regularly than other parts of the apparatus.

Concerning the effect that magnetic fields have on the number of bacteria present, additional research needs to be carried out. A previous study investigated the potential effect of a rotating magnetic field against bacteria through cell wall disruption and changes in morphology [39].

We were able to detect substantial differences in the contamination between months of May and June in the radiology department. In June, patient isolation, workload, and the number of referrals for patients were all higher than in May. As a reflection of that, the average contamination rate from all sites in June was 1.3 CFUs/cm2, while in May, it was 0.79 CFUs/cm2. Moreover, six sites showed contamination with MRS in May compared to 13 sites in June, with only one site in June having a contamination rate of ≥1 CFUs/cm2. Surprisingly, 7 sites showed contamination with VRS, and all were in June. Regarding contamination with Gram-negative bacteria, a contamination rate of ≥1 CFUs/cm2 was detected only at one site in June. In addition to that, VRE was detected at one site in May with a contamination rate of <1 CFUs/cm2, while it was detected at four sites in June, with a contamination rate of ≥1 CFUs/cm2.

Based on our results, we can say that reducing contamination effectively in a healthcare setting, such as a radiology department, requires a comprehensive approach involving staff education, proper cleaning and disinfection protocols, and ongoing monitoring. We emphasize that hospital staff must be educated and trained regularly on proper hand hygiene techniques. We also propose implementing a regular cleaning schedule for all surfaces within the radiology department to ensure that cleaning includes high-touch areas like doorknobs, keyboards, phones, and equipment surfaces [4043].

Another technique for infection control is the use of disposable barriers (e.g., plastic covers) on equipment surfaces that come into direct contact with patients. These can be changed between each patient, reducing the risk of cross-contamination [44].

Because patients and family members are a significant part of the contamination equation, we encourage educating patients about the importance of hand hygiene and covering coughs and sneezes to reduce the risk of spreading infections within the department [45].

We also encourage conducting regular audits and assessments of cleaning and infection control practices to ensure that the whole team has a clear understanding of the hand hygiene importance and infection prevention precautions [46, 47].

4.1. Limitations

In general, it is important to note that the sampled area in our study represents only a fraction of the entire surfaces being studied at a specific period of time. This limitation could potentially reduce the test's sensitivity and may restrict the generalizability of our results, especially when trying to detect resistant bacteria that may be present in small quantities.

Our study observes significant variability in contamination rates between different months, which raises questions about the consistency of factors affecting contamination levels. Factors such as patient workload, isolation practices, and referral rates can fluctuate and complicate the interpretation of results.

5. Conclusion

There is an ongoing debate all around the world regarding whether or not the setting of a hospital contributes to the spread of HAIs. However, there is evidence from research that supports the concept that hospitals can operate as a crucial reservoir of numerous nosocomial infections in a variety of environments. These environments include surfaces, medical equipment, and water systems.

In this investigation, it was concluded that

  1. Radiology department could be a source of healthcare-acquired infection. Gram-positive bacteria were the most present bacteria and multidrug-resistant bacteria were detected from various sites with a contamination rate which exceeded the limit of 1 CFU/cm2 for bacterial contamination.

  2. The relatively high methicillin-resistant Staphylococcus contamination rate observed in this study highlights the importance of regular monitoring for Staphylococcus contamination in radiology departments. It emphasizes the need for stringent infection control practices to prevent the spread of methicillin-resistant Staphylococcus within healthcare facilities.

  3. Increase in the work load, referred and isolated patients which were proportional to the increase in the contamination rate, the presence of Gram-negative, and multidrug-resistant bacteria.

  4. Surface cleaning and disinfecting must frequently focus on keyboards in the radiology department, examination patient table sides and centres, knee coil, US machine, and patient legs support in particular. Proper hand hygiene can also help reduce the risk of bacterial transmission within examination rooms.

In summary, our research provides practical insights that can guide infection control practices in radiology departments. It highlights the need for rigorous cleaning and disinfection protocols, awareness of specific bacterial strains like methicillin-resistant Staphylococcus, and further exploration of innovative solutions to minimize bacterial contamination and reduce the risk of infections in healthcare settings.

Acknowledgments

We would like to acknowledge the Faculty of Medicine and Health Sciences and Faculty of Graduate Studies at the An-Najah National University (https://www.najah.edu) for facilitating the accomplishment of the current study. We are also thankful to the An-Najah National University Hospital for taking part in this study. This project was funded by the An-Najah National University, as it was a master thesis for a student in the master's “Infectious Diseases Prevention and Control” program.

Data Availability

Data used in this study will be available from the corresponding author upon reasonable request.

Conflicts of Interest

The authors declare that there are no conflicts of interest.

Authors' Contributions

MQ conceptualized and designed the study and did literature search, lab work, reviewing, and manuscript writing. ZA conceptualized the study and did literature search, lab work, analysis, and wrote the manuscript. SA revised and wrote the manuscript. All authors have read and approved the final version of the manuscript for submission.

Supplementary Materials

Supplementary Materials

Table S1: sample source and number of colonies on each plate during May. Table S2: sample source and number of colonies on each plate during June. Table S3: calculated contamination rate (CFUs/cm2) during the two months on both blood and chocolate agar from each sample source.

References

  • 1.Facciolà A., Pellicanò G. F., Visalli G., et al. The role of the hospital environment in the healthcare-associated infections: a general review of the literature. European Review for Medical and Pharmacological Sciences . 2019;23(3):1266–1278. doi: 10.26355/eurrev_201902_17020. [DOI] [PubMed] [Google Scholar]
  • 2.Boyle H., Strudwick R. M. Do lead rubber aprons pose an infection risk? Radiography . 2010;16(4):297–303. doi: 10.1016/j.radi.2010.03.002. [DOI] [Google Scholar]
  • 3.Giacometti M., Gualano M. R., Bert F., et al. Microbiological contamination of radiological equipment. Acta Radiologica . 2014;55(9):1099–1103. doi: 10.1177/0284185113512481. [DOI] [PubMed] [Google Scholar]
  • 4.Picton-Barnes D., Pillay M., Lyall D. A systematic review of healthcare-associated infectious organisms in medical radiation science departments. Healthcare . 2020;8(2):p. 80. doi: 10.3390/healthcare8020080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Jia H., Li W., Hou T., et al. The attributable direct medical cost of healthcare associated infection caused by multidrug resistance organisms in 68 hospitals of China. BioMed Research International . 2019;2019:7. doi: 10.1155/2019/7634528.7634528 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Ilyas F., Burbridge B., Babyn P. Health care-associated infections and the radiology department. Journal of Medical Imaging and Radiation Sciences . 2019;50(4):596–606.e1. doi: 10.1016/j.jmir.2019.07.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Palmqvist C., Samuelsson A., Fröding I., Giske C. Surface contamination of CT and MRI equipment—a potential source for transmission of hospital-acquired infections. Journal of Radiology Nursing . 2019;38(4):254–260. doi: 10.1016/j.jradnu.2019.09.002. [DOI] [Google Scholar]
  • 8.Khan H. A., Ahmad A., Mehboob R. Nosocomial infections and their control strategies. Asian Pacific Journal of Tropical Biomedicine . 2015;5(7):509–514. doi: 10.1016/j.apjtb.2015.05.001. [DOI] [Google Scholar]
  • 9.Muradali D., Gold W. L., Phillips A., Wilson S. Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study. American Journal of Roentgenology . 1995;164(6):1521–1524. doi: 10.2214/ajr.164.6.7754907. [DOI] [PubMed] [Google Scholar]
  • 10.Elliott T. M., Hurst C., Doidge M., Hurst T., Harris P. N., Gordon L. G. Unexpected benefit of COVID-19 hospital restrictions: reduction in patients isolating with multidrug resistant organisms after restrictions were lifted. Infection, Disease and Health . 2022;27(1):10–14. doi: 10.1016/j.idh.2021.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kim J. S., Kim H. S., Park J. Y., et al. Contamination of X-ray cassettes with methicillin-resistant Staphylococcus aureus and methicillin-resistant Staphylococcus haemolyticus in a radiology department. Annals of laboratory medicine . 2012;32(3):206–209. doi: 10.3343/alm.2012.32.3.206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Larrosa M. N., Almirante B. Isolation strategy for controlling the spread of multidrug-resistant organisms: is this still an essential option in hospitals? Infectious Diseases and Clinical Microbiology . 2021;39(9):425–428. doi: 10.1016/j.eimc.2021.04.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Huang S. S. Chlorhexidine-based decolonization to reduce healthcare-associated infections and multidrug-resistant organisms (MDROs): who, what, where, when, and why? Journal of Hospital Infection . 2019;103(3):235–243. doi: 10.1016/j.jhin.2019.08.025. [DOI] [PubMed] [Google Scholar]
  • 14.Blanco N., O’Hara L. M., Harris A. D. Transmission pathways of multidrug-resistant organisms in the hospital setting: a scoping review. Infection Control and Hospital Epidemiology . 2019;40(4):447–456. doi: 10.1017/ice.2018.359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Saliba R., Ghelfenstein-Ferreira T., Lomont A., et al. Risk factors for the environmental spread of different multidrug-resistant organisms: a prospective cohort study. Journal of Hospital Infection . 2021;111:155–161. doi: 10.1016/j.jhin.2021.01.029. [DOI] [PubMed] [Google Scholar]
  • 16.Chia P. Y., Sengupta S., Kukreja A., Sl Ponnampalavanar S., Ng O. T., Marimuthu K. The role of hospital environment in transmissions of multidrug-resistant gram-negative organisms. Antimicrobial Resistance and Infection Control . 2020;9(1):p. 29. doi: 10.1186/s13756-020-0685-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Anderson D. J., Chen L. F., Weber D. J., et al. Enhanced terminal room disinfection and acquisition and infection caused by multidrug-resistant organisms and Clostridium difficile (the Benefits of Enhanced Terminal Room Disinfection study): a cluster-randomised, multicentre, crossover study. The Lancet . 2017;389(10071):805–814. doi: 10.1016/s0140-6736(16)31588-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Petti S., Polimeni A., Dancer S. J. Effect of disposable barriers, disinfection, and cleaning on controlling methicillin-resistant Staphylococcus aureus environmental contamination. American Journal of Infection Control . 2013;41(9):836–840. doi: 10.1016/j.ajic.2012.09.021. [DOI] [PubMed] [Google Scholar]
  • 19.Taggar G., Attiq Rehman M., Boerlin P., Diarra M. Molecular epidemiology of carbapenemases in enterobacteriales from humans, animals, food and the environment. Antibiotics . 2020;9(10):p. 693. doi: 10.3390/antibiotics9100693. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Quispe A. M., Soza G., Pons M. J. Multidrug resistance and its association with Enterobacteriales and age among pregnant Peruvian women with- bacteremia. The Journal of Infection in Developing Countries . 2020;14(12):1402–1409. doi: 10.3855/jidc.12569. [DOI] [PubMed] [Google Scholar]
  • 21.van Duin D., Doi Y. The global epidemiology of carbapenemase-producing Enterobacteriaceae. Virulence . 2017;8(4):460–469. doi: 10.1080/21505594.2016.1222343. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Cury A. P., Girardello R., Duarte A. J. D. S., Rossi F. KPC-producing Enterobacterales with uncommon carbapenem susceptibility profile in Vitek 2 system. International Journal of Infectious Diseases . 2020;93:118–120. doi: 10.1016/j.ijid.2020.01.016. [DOI] [PubMed] [Google Scholar]
  • 23.Hefzy E. M., Wegdan A. A., Abdel Wahed W. Y. Hospital outpatient clinics as a potential hazard for healthcare associated infections. Journal of Infection and Public Health . 2016;9(1):88–97. doi: 10.1016/j.jiph.2015.06.015. [DOI] [PubMed] [Google Scholar]
  • 24.Kumar S., Vyas A. Utilization of MacConkey-Meropenem screening agar for the detection of carbapenem resistanant Enterobacteriaceae in a tertiary care hospital. 2018. https://api.semanticscholar.org/CorpusID:215762188 .
  • 25.Brown D. F. J., Walpole E. Evaluation of selective and enrichment media for isolation of glycopeptide-resistant enterococci from faecal specimens. Journal of Antimicrobial Chemotherapy . 2003;51(2):289–296. doi: 10.1093/jac/dkg076. [DOI] [PubMed] [Google Scholar]
  • 26.Calatrava E. Other Streptococcus species and Enterococcus. In: Rezaei N., editor. Encyclopedia of Infection and Immunity . Oxford, UK: Elsevier; 2022. pp. 529–541. [Google Scholar]
  • 27.Lemmen S. W., Häfner H., Zolldann D., Stanzel S., Lütticken R. Distribution of multi-resistant Gram-negative versus Gram-positive bacteria in the hospital inanimate environment. Journal of Hospital Infection . 2004;56(3):191–197. doi: 10.1016/j.jhin.2003.12.004. [DOI] [PubMed] [Google Scholar]
  • 28.Chiller K., Selkin B. A., Murakawa G. J. Skin microflora and bacterial infections of the skin. Journal of Investigative Dermatology-Symposium Proceedings . 2001;6(3):170–174. doi: 10.1046/j.0022-202x.2001.00043.x. [DOI] [PubMed] [Google Scholar]
  • 29.Childress J., Burch D., Kucharski C., Young C., Kazerooni E. A., Davenport M. S. Bacterial contamination of CT equipment: use of ATP detection and culture results to target quality improvement. Academic Radiology . 2017;24(8):923–929. doi: 10.1016/j.acra.2017.01.022. [DOI] [PubMed] [Google Scholar]
  • 30.Shelly M. J., Scanlon T. G., Ruddy R., Hannan M. M., Murray J. G. Meticillin-resistant Staphylococcus aureus (MRSA) environmental contamination in a radiology department. Clinical Radiology . 2011;66(9):861–864. doi: 10.1016/j.crad.2011.05.002. [DOI] [PubMed] [Google Scholar]
  • 31.Tugwell J., Maddison A. Radiographic markers–a reservoir for bacteria? Radiography . 2011;17(2):115–120. doi: 10.1016/j.radi.2010.10.005. [DOI] [Google Scholar]
  • 32.Duszak R., Lanier B., Tubbs J. A., Ogilvie M., Thompson-Jaeger S. Bacterial contamination of radiologist workstations: results of a pilot study. Journal of the American College of Radiology . 2014;11(2):176–179. doi: 10.1016/j.jacr.2013.05.017. [DOI] [PubMed] [Google Scholar]
  • 33.Forrester L. A., Bryce E. A., Mediaa A. K. Clean Hands for Life™: results of a large, multicentre, multifaceted, social marketing hand-hygiene campaign. Journal of Hospital Infection . 2010;74(3):225–231. doi: 10.1016/j.jhin.2009.10.027. [DOI] [PubMed] [Google Scholar]
  • 34.Fojt L., Strašák L., Vetterl V., Šmarda J. Comparison of the low-frequency magnetic field effects on bacteria Escherichia coli, Leclercia adecarboxylata and Staphylococcus aureus. Bioelectrochemistry . 2004;63(1-2):337–341. doi: 10.1016/j.bioelechem.2003.11.010. [DOI] [PubMed] [Google Scholar]
  • 35.Strašák L., Vetterl V., Šmarda J. Effects of low-frequency magnetic fields on bacteria Escherichia coli. Bioelectrochemistry . 2002;55(1-2):161–164. doi: 10.1016/s1567-5394(01)00152-9. [DOI] [PubMed] [Google Scholar]
  • 36.Dancer S. J. How do we assess hospital cleaning? A proposal for microbiological standards for surface hygiene in hospitals. Journal of Hospital Infection . 2004;56(1):10–15. doi: 10.1016/j.jhin.2003.09.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Rock C., Small B. A., Thom K. A. Innovative methods of hospital disinfection in prevention of healthcare-associated infections. Current Treatment Options in Infectious Diseases . 2018;10(1):65–77. doi: 10.1007/s40506-018-0153-0. [DOI] [Google Scholar]
  • 38.Esolen L. M., Thakur L., Layon A. J., et al. The efficacy of self-disinfecting bedrail covers in an intensive care unit. American Journal of Infection Control . 2018;46(4):417–419. doi: 10.1016/j.ajic.2017.09.026. [DOI] [PubMed] [Google Scholar]
  • 39.Woroszyło M., Ciecholewska-Juśko D., Junka A., et al. Rotating magnetic field increases β-lactam antibiotic susceptibility of methicillin-resistant Staphylococcus aureus strains. International Journal of Molecular Sciences . 2021;22(22) doi: 10.3390/ijms222212397.12397 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Zeng C., Liu H., Jiang Y., et al. How to reduce the exposure risk of medical staff from SARS-CoV-2 by reducing environmental contamination: experience from designated hospitals in China. Frontiers in Public Health . 2022;10 doi: 10.3389/fpubh.2022.963999.963999 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Fernando S. A., Gray T. J., Gottlieb T. Healthcare-acquired infections: prevention strategies. Internal Medicine Journal . 2017;47(12):1341–1351. doi: 10.1111/imj.13642. [DOI] [PubMed] [Google Scholar]
  • 42.Thomas R. E., Thomas B. C., Conly J., Lorenzetti D. Cleaning and disinfecting surfaces in hospitals and long-term care facilities for reducing hospital- and facility-acquired bacterial and viral infections: a systematic review. Journal of Hospital Infection . 2022;122:9–26. doi: 10.1016/j.jhin.2021.12.017. [DOI] [PubMed] [Google Scholar]
  • 43.Qadi M., Khayyat R. Microbes on the mobile phones of healthcare workers in Palestine: identification, characterization, and comparison. Canadian Journal of Infectious Diseases and Medical Microbiology . 2021;2021:9. doi: 10.1155/2021/8845879.8845879 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Rutala W. A., Weber D. J. A review of single-use and reusable gowns and drapes in health care. Infection Control and Hospital Epidemiology . 2001;22(4):248–257. doi: 10.1086/501895. [DOI] [PubMed] [Google Scholar]
  • 45.Rogers S. Engaging patients and family members in better hand hygiene practices: a teaching hospital’s challenge. Healthcare Quarterly . 2013;16(4):27–31. [PubMed] [Google Scholar]
  • 46.Bryce E. A., Scharf S., Walker M., Walsh A. The infection control audit: the standardized audit as a tool for change. American Journal of Infection Control . 2007;35(4):271–283. doi: 10.1016/j.ajic.2006.05.293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Hay A. Audit in infection control. Journal of Hospital Infection . 2006;62(3):270–277. doi: 10.1016/j.jhin.2005.09.008. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Materials

Table S1: sample source and number of colonies on each plate during May. Table S2: sample source and number of colonies on each plate during June. Table S3: calculated contamination rate (CFUs/cm2) during the two months on both blood and chocolate agar from each sample source.

Data Availability Statement

Data used in this study will be available from the corresponding author upon reasonable request.


Articles from The Canadian Journal of Infectious Diseases & Medical Microbiology = Journal Canadien des Maladies Infectieuses et de la Microbiologie Médicale are provided here courtesy of Wiley

RESOURCES