Abstract
The mechanical properties of polyurethane dental aligners have been studied in an oral environment at 37°C and subjected to thermal cycling between 5°C and 55°C for long periods of time at different mechanical stresses. The aim is to determine the efficacy of the orthodontic aligner at different stress levels, the effect of thermal cycling with therapy time on tooth position correction. Sixty aligners with the same design were studied applying tensions of 0, 3 and 30 N and determining the deformation at different times from 1 to 760 h. Half of these aligners were subjected to stresses submerged in artificial saliva at 37°C and the other half were subjected to thermal cycles between 2°C and 55°C in salivary medium. Deformation was determined using a high-resolution stereo magnifier and ImageJ image analysis software. Water adsorption by the polyurethane was determined at the different test times. The results showed that in the unloaded aligners there is no appreciable deformation, but with thermal cycling there is a light shrinkage of the aligner due to the semi-crystallization process (ordering of polymeric chains) of the polyurethane. When applying loads of 3 and 30 N, creep curves with constant deformation transition zones can be seen. The transition zones decrease as the applied mechanical load increases. In addition, the significant effect of thermal cycling on the reduction of the transition zone of the aligners has been demonstrated. The transition zones are optimal for dental correction as constant stresses are exerted for tooth movement. The effect of thermal cycling shortens the constant deformation zone and reduces tooth alignment time. It was observed that the absorption of water in the aligner is constant after 1 h of immersion and does not exceed 0.4% by weight of absorbed water.
Keywords: polymers; orthodontic appliances, removable; clear aligner appliances; mechanical properties; creep
1 Introduction
Orthodontic systems of clear aligners and retainers combined with digital planning systems turn out to be the present and future of orthodontics. Effective orthodontic therapy necessitates a thorough understanding of the materials for our patients, without adequate knowledge of the materials used in our patients, their properties, and biocompatibility. Currently, the digital manufacturing method uses CAD-CAM (Computer Aided Design Computer Aided Manufacturing) technology and digital workflow protocol. The digital image is obtained either through an intraoral scanner or indirectly through buccal impressions in high-quality polyvinyl siloxane (PVS) material which are then digitally scanned to provide a digital model (.stl). Teeth movements will be planned with the digital dental model using digital CAD platforms and generating virtual sequential models with the teeth in the desired positions, planned for each stage. Printing 3D models of each virtual configuration through CAM technology involves incorporating subtractive (milling) or additive (3D printing) manufacturing techniques (Hartshorne and Wertheimer, 2022).
The performance of aligners depends on the composition of the material used for their manufacture, being directly dependent on the manufacturing process. The most widely used is the conventional vacuum thermoforming method that includes molding the thermoplastic material into physical models (Tartaglia et al., 2021). Thermoplastic Polyurethane (TPU), composed mainly of di- and tri-isocyanates and polyols, is another extremely versatile polymer with advantages such as excellent mechanical and elastomeric characteristics, chemical and abrasion resistance. TPU is the material that was one of the first to be used and has been evolving to this day. One of its advantages is that when subjected to a load, TPU changes shape but can recover, when the load is removed it elongates and recovers due to its flexibility. The material also features high tear and fracture strength (Bichu et al., 2022).
The changes suffered by orthodontic materials when they are introduced into an electrolytic medium such as the oral cavity, as well as the effect they suffer due to masticatory stress and other environmental changes, such as temperature, have been extensively studied by numerous authors in order to evaluate the effectiveness of the materials and the effect of the degradation of its components on the health of patients with orthodontic appliances. Most of these studies are carried out analyzing the effect on the components of fixed multibracketts appliances in different brands and materials (Pascual et al., 1999; Manero et al., 2003; Gil et al., 2004; Djeu et al., 2005; Keim et al., 2014; Abbate et al., 2015; Han, 2015; Li et al., 2015; Hennessy et al., 2016; Gu et al., 2017; Lanteri et al., 2018). With the increase in orthodontic treatment with aligners, it is necessary to enhance the studies in which the composition of its material is analyzed when it is introduced into the oral cavity and the degradation of its components based on various environmental stimuli to which they are subjected in the oral cavity (Eliades et al., 1999; Schuster et al., 2004; Eliades et al., 2007; Eliades et al., 2009; Gracco et al., 2009; Rossini et al., 2015; Halimi et al., 2016; Buxadera-Palomero et al., 2017; Eliades and Bisphenol, 2017; Yi et al., 2018; Francisco et al., 2022).
Invisaling® system clear aligners are thermoplastic polymers based on polyurethanes, aromatic multilayer thermoplastic polyurethane, methylphenyl diisocyanate and 1,6-hexanediol plus additives (Raghavan et al., 2017). They are characterized by having higher values of hardness and elastic modulus and index, but lower values of creep resistance compared to other materials also used as aligners. Thermoplastics have excellent aesthetics and adaptability to different shapes that are widely used for orthodontic treatment. Since Kesling in 1945 produced the first aligner made of elastic polymer (Shalish et al., 2012), there has been an important development in materials and especially in CAD-CAM manufacturing processes, thermofabrication, 3D printing, etc. that have been fundamental to improve orthodontic treatments (Kesling, 1945; Elkholy et al., 2015; Gomez et al., 2015). In 2013, Invisalign® began using SmartTrack as a material for its aligners, a multilayer polymer that provides light and constant force and superior control over tooth movement (Simon et al., 2014).
Although there are numerous studies on the biological behavior of clear aligners with different thermoplastic polymeric materials, there is no much evidence on their mechanical properties (Align Technology, 2023). Due to this lack of evidence, it is important to carry out research which analyze the behavior of clear aligners in the oral environment, and the influence of thermal cycles. The hypothesis of this research was that polyurethanes, being thermoplastic materials, would produce elongation in the aligner by exerting constant tension on its walls. Furthermore, the effect of temperature can influence the movement of polymeric chains, affecting the correction of dental position. The aim is to analyze in an artificial mouth the influence of the temperature and masticatory forces on the behavior of the clear aligner characteristics.
2 Materials and methods
The aim is to determine the creep of the aligner, i.e., the increase in deformation when applying a constant force for long periods of time. This mechanical behavior simulates the effect of stresses created on the aligner in the mouth over time and can be determined when the aligner is active. For this purpose, different forces will be applied by means of a spring calibrated with a constant force, determining the elongation increases at different times. The tests will be carried out at a constant temperature and thermocycled. In this way it will be possible to determine the effect of the variations of the temperatures on the mechanical creep behavior of the aligners (Morresi et al., 2014).
An in vitro, experimental transversal research was carried out. Due to the in vitro design of the study, the approval of the Ethics Committee was not necessary. According to the calculation of experiments, a total of 60 identical clear aligners were studied. These aligners were manufactured by Align Technologies, Inc. (Tempe, Arizona, USA) with polyurethane 1,6-hexanedial methylene diphenyl diisocyanate. The thickness of the clear aligners was 0.72 mm.
Thirty samples were placed in an artificial saliva container at constant temperature of 37°C. The chemical composition of artificial saliva can be seen in Table 1 (Murakami et al., 1983). Thermal thermocycling was applied to 30 random samples for 600 cycles (1,200 immersions) considering in this case 20 temperature changes in the patient’s mouth for 30 days. For the thermocycling study, equipment manufactured in the dental laboratory of the International University of Catalunya (UIC, Sant Cugat del Vallés, Spain) was used to control the thermal cycles by immersion in a thermostatic bath with a controlled temperature at 2°C for 30 s and another 30 s in a bath at 55°C (Figure 1).
TABLE 1.
Chemical composition | g/dm3 |
---|---|
K2HPO4 | 0.20 |
KCI | 1.20 |
KSCN | 0.33 |
Na2HPO4 | 0.26 |
NaCl | 0.70 |
NaHCO3 | 1.50 |
Urea | 1.50 |
Lactic acid | Until pH = 6.7 |
Mechanical tests were carried out at 0 N, 3 N and 30 N applied to the first left molar (3.6). We studied 10 inserts for each were studied. The force is exerted with a calibrated spring that exerts tensile stress on the mold walls as can be seen in Figure 2. The springs are calibrated by a highly sensitive Adamel Lhomargy dynamometer (Adamel Lhomargy ST2507-E23, Saint Baldolph, France). These stresses simulate the stresses exerted by the tooth on the template. Elongations caused by applied force and/or thermal cycling were determined by observing the samples under a high-precision stereo magnification Q-star high-resolution system using ImageJ software for measurements (Bethesda Q-2022-ST18965, Maryland, USA). Two marks are made on the aligner with a diamond-tipped indenter at two points. The sample is positioned in such a way that there can be no movement and it is ensured that the electronic source is always at the same angle. ImageJ is used to determine the difference between points in the unloaded state, obtaining the distance between the points automatically, which corresponds to the initial distance.
ImageJ is the world’s fastest pure Java image processing program. It can filter a 2048 × 2048 image in 0.1 s. That’s 40 million pixels per second. 8-bit grayscale or indexed color, 16-bit unsigned integer, 32-bit floating-point and RGB color. Tools are provided for zooming (1:32 to 32:1) and scrolling images. All analysis and processing functions work at any magnification factor. Create rectangular, elliptical or irregular area selections. Create line and point selections. Edit selections and automatically create them using the wand tool. Draw, fill, clear, filter or measure selections. Save selections and transfer them to other images. Supports smoothing, sharpening, edge detection, median filtering and thresholding on both 8-bit grayscale and RGB color images. Interactively adjust brightness and contrast of 8, 16 and 32-bit images. Measure lengths in this case and standard deviation. Calibrate using density standards. Split a 32-bit color image into RGB or HSV components. Merge 8-bit components into a color image. Convert an RGB image to 8-bit indexed color. Apply pseudo-color palettes to grayscale images (Vallet-Regí et al., 2005).
The aligner is subjected to a constant force and the distance at different test times is determined. With the new distance the software calculates the elongation. The sensitivity of the distances calculated by the equipment and the software was 0.001 mm and has been used in different research studies that require high precision (Vallet-Regí et al., 2005; Godoy-Gallardo et al., 2016a; Herrero-Climent et al., 2023). Measurement times were: 0, 1, 2, 5, 10, 20, 24, 48, 72, 96, 120, 144, 168, 216, 240, 264, 268, 312, 336, 384, 432, 480, 528, 576, 624, 672, 720 h. The 720 h correspond to the hours of a month, the maximum application time. Elongation was calculated by determining the difference between initial separation between two points of the aligner and the distance between the same points after the experimental time has elapsed. The points are always the same and were identified by a small incision in the surrounding area (Herrero-Climent et al., 2023).
The water absorption of the clear aligners at different times without thermal cycling was studied using 20 mm × 20 mm samples. Samples were stored in a desiccator until constant weight (Initial weight = W1) was reached and then immersed in artificial saliva at 37°C for 1 min. After the different times they are dried with a cloth and weighed (Second weight = W2). They are weighed on a balance with a sensitivity of 0.00001 g (Sartorius x1000, Barcelona, Spain) and the water absorption (WA) is calculated according to the formula (Padrós et al., 2020):
Data were statistically analyzed using Student's t-tests with a 99.5% significance level (p < 0.005) and asympotics significance and Turkey multiple comparison tests to assess any statistically significant differences between sample groups. All statistical analyzes were performed using MinitabTM software (Minitab version 13.0, Minitab Inc., USA).
3 Results
Figure 3 shows the elongation values with the test time for the aligner appliance unloaded and loaded at 3 and 30N with and without thermal cycling.
The water absorption results of the artificial saliva give constant values from 5 h onwards, as can be seen in Figure 4.
4 Discussion
It can be observed that the samples that do not undergo loading and without thermal cycling, the elongations are practically negligible, and no changes in the shape of the template are observed. Samples without loading but which are thermally cycled show shrinkage of the template after 550 h of cycling. This shortening of the distances may be due to a partial chain folding process, called crystallisation, which decreases the distances.
The polyurethanes used are composed of two main polymeric segments: hard and soft chains. These two types of segments have different chain dynamics with repulsive interaction towards each other (Huang et al., 2012). The hard chains, with high extent of hydrogen bonding, have higher glass transition temperature (Tg) compared to soft chains and act as a physical cross-linker of soft segments. The repulsive interaction between hard and soft segments results in a decrease in crystallization of the soft segments (Lu et al., 2017). Salvekar et al. (Salvekar et al., 2017) determined for polyols composed of two different chains, their final physical and thermodynamic properties depend on the extent of phase separation, and their relative composition. To control the extent of phase separation for two different polyol chains, there are different procedures such as copolymerization with different structures, from random to block copolymers, and directed self-assembly of homopolymers with block copolymers of the same chemistry as homopolymers (Ge et al., 2012).
Some polyurethanes present shape memory property is due to the sensitivity of polyurethane to stimuli to retain an induced shape at a constant low temperature for a time: it also exhibits the ability to recover the original shape after removing the external stimulus force in the reheating process (Xie and Rousseau, 2009; Wu et al., 2017). These polymers are known for their ability to recover shape while maintaining constant stresses. In orthodontics, low and constant tensions in tempo are optimal to cause tooth movement; this same behavior is achieved with superelastic NiTi orthodontic wires (Gil and Planell, 1993; Gil et al., 2012; Gil et al., 2013; Bellini et al., 2016; Gil et al., 2018). The justification for the high shape resilience of polyetherimides is the presence of hard domains that act as physical cross-linkers to prevent chain slippage (Wu et al., 2017). The soft chains are responsible for fixing the temporary shape over a range of temperature (Tg) or crystallization temperature (Tc) (Ge et al., 2012; Wu et al., 2017). Thermal cycling below the Tg temperature will allow chain folding during heating, which reduces the dimensions of the aligner.
The materials with 3N mechanical loading show an initial deformation of the template due to the application of the load at time zero. Afterwards, a slight increase in elongation is observed with a very flat transition zone. This plateau is ideal for orthodontic treatment as the stress on the tooth does not cause an increase in elongation. At longer times, an increase in elongation is observed due to the creep of the polymeric material (Lombardo et al., 2012; Ihssen et al., 2019; Papadopoulou et al., 2019). This behavior is accentuated when thermal cycles are applied to the insole. At times longer than 600 h, a more intense increase in elongation is observed.
The insoles subjected to 30 N undergo a greater elongation than in the case of 3 N and a plateau suitable for orthodontic therapy can also be seen. After 500 h, a slight change in the slope of the deformation rate is observed. When subjected to this 30 N load and the material is thermally cycled, the plateau is reduced and from 450 h of treatment the elongation increases considerably. At high loads, as in this case, the thermal cycling has an intensifying effect on creep, which is why the treatment times under these conditions have to be shorter.
As can be seen, the aligners follow the typical creep graph, which has three stages (Espinar-Escalona et al., 2013; Iijima et al., 2015; Lombardo et al., 2015; Dalaie et al., 2021).
The first stage corresponds to a sudden increase in deformation when mechanical load is applied.
The second stage is the plateau zone which is the zone where the mechanical force exerted on the teeth is practically constant. It is called the transition phase and is the most suitable force for the correction of tooth positions. Constant, low intensity forces are the most suitable for tissue accommodation and the correct apposition and deposition of the bone that causes orthodontic movements.
The third stage is a stage of gross growth where the aligner may have lost thickness due to constant tension, which has produced a rearrangement of the polymeric chains causing a sudden increase in deformation. This stage is ineffective for any tooth movement and the aligner has ceased to perform its function.
This creep treatment occurs because polyurethane is a thermoplastic polymer, i.e., the temperature causes the polymer chains to align and can modify their shape and therefore the stresses exerted on the teeth or molars. These movements of polymer chains are produced by the effect of temperature. When the polyurethane is not mechanically loaded, the effect of temperature is the folding of chains in the so-called pseudocrystalline zones. This chain folding causes a decrease in volume, and it is for this reason that the aligner at 0N with thermal cycling up to 55°C shrinks (Lombardo et al., 2017; Dalaie et al., 2021).
The chain folding process does not occur when the polyurethane is subjected to mechanical stress as it prevents folding due to the directionality of the force exerted. For samples subjected to mechanical stress, thermal cycling facilitates the movement of the polymer chains in the direction of the stress, causing a faster creep of the aligner and reducing the plateau of the aligners, thus impairing tooth movement (Hodge et al., 1996; Godoy-Gallardo et al., 2016b; Jaggy et al., 2020).
For a given load and level of deformation, a material can be considered as symmetrical if the creep strain does not vary over time while an asymmetrical material presents a change in creep strain rate. Previous studies reported the asymmetrical behavior of polyurethane (Kasgoz, 2021), which is related, to the general chemical nature of the polymer, for example, including interactions among the polymer chains over time, chemical bonds and others. At 25°C, this grade of polyurethane displayed an asymmetric nature at only 30 N of load, which is an extreme condition, not commonly achieved when in service. In this case, an increase in the elongation rate can be observed during the test run.
The tests were performed at thermal cycles to simulate an accelerated aged condition for this grade of polyurethane and were stipulated according to previous studies, where better ageing responses were obtained at this level of temperature (55°C) (Jia et al., 2011). As seen on Figure 3, under 30 N and thermal cycling accelerated ageing conditions all specimens tend to present an increase of elongation. Creep strain rate results in both tested temperatures did not show a creep fracture point, which demonstrate that in the aforementioned conditions, the material used did not fail even after the lengthy duration of the test. However, the aligner losses the activation to move the teeth.
The modeling of the creep behavior of polyurethanes must take into consideration two contributions: the viscoelastic behavior and the long-term behavior, i.e., the influence of time. The modeling of viscoelasticity integrates the structure-properties interactions, the share of elastic and viscous deformation in the total deformation and the relaxation coefficient. Many viscoelastic models have been developed and used by various authors in order to define the relationship between viscoelastic structure and polyurethane characteristics (Wu et al., 1994; Findley and Davis, 2013).
The best fitting model is the four-element Burger model (Wu et al., 1994; Jhao et al., 2022; Cui and Clyne, 2023), which is a simple combination of the Maxwell and Kelvin-Voigt models.
Under the condition of linear deformation, the total strain of a viscoelastic solid may be given as the sum of immediate elastic deformation of Maxwell spring (εM1), delayed elastic deformation of Maxwell dashpot (εM2) and viscous deformation of Kelvin unit (εK) (Eq. 1).
(1) |
Considering the combination of the Maxwell and Kelvin-Voight elements in the Burger model, the mathematical representation of the Burger model may be given as follows;
(2) |
where EM and ηM are the modulus and viscosity of the Maxwell spring and dashpot, respectively; EK and ηK are the modulus and viscosity of the Kelvin spring and dashpot, respectively, and σ is the applied stress value.
From the values in Figure 4, it can be confirmed that water absorption by the polyurethane is constant from the first hours of placement, reaching values that do not exceed 0.4% wt of the polyurethane composition. Therefore, the mechanical behavior studied is not affected by water absorption but depends fundamentally on the composition of the polymer, the stresses applied and the thermal cycling of the aligner (Mima et al., 1983; Skaik et al., 2019; Cui and Clyne, 2023).
Several studies have been carried out analyzing different materials of invisible aligners for orthodontics, observing creep phenomena for all of them. Cianci C. et al. (Cianci et al., 2020) studied the compression behavior under cyclic loading of Polyethylene terephthalate-glycol (PET-G) -one of the most famous and used thermoplastic- and used as invisible aligners. It was observed for the first day of testing of the aligners that stiffening effects occur while cyclic loading progresses. It was also observed that stiffening decreases during the non-loading time between two successive sessions of test while stiffening effects are observed again when a new set of cycles is applied to the aligners. One of the results obtained was that in the case of testing with saliva saliva an higher stress recovery is observed between two subsequent loading sessions; moreover, the hygrothermal environment showed a contribution to reduce the stress accumulation effect during the test. In our case we have observed that the influence of the thermal cycles accelerates the creep processes reducing the transition straight line which is the desired one to provoke the activation of tooth alignment.
Other investigations (Lombardo et al., 2017; Albertini et al., 2022) carried out for different aligner materials show behaviors like those observed in our results, although the tests have been carried out in short periods of time (24 h) in all the polymers released a significant amount of stress during the 24-h period. Stress release was greater during the first 8 h, reaching a plateau that generally remained constant. The single-layer materials, F22 Aligner polyurethane (Sweden & Martina, Due Carrare, Padova, Italy) and Duran polyethylene terephthalate glycol-modified (SCHEU, Iserlohn, Germany), exhibited the greatest values for both absolute stress and stress decay speed. The double-layer materials, Erkoloc-Pro (Erkodent, Pfalzgrafenweiler, Germany) and Durasoft (SCHEU), exhibited very constant stress release, but at absolute values up to four times lower than the single-layer samples tested. Albertini et al. (Albertini et al., 2022) conducted studies at longer periods of time at constant temperature determining that all the materials that we tested showed a rapidity of stress decay during the first few hours of application, before reaching a plateau phase. The polyurethane showed the greatest level of final stress, with relatively constant stress release during the entire 15-day period. Further research after in vivo aging is necessary in order to study the real aligners’ behavior during orthodontic treatment. These results show similar behaviors to those obtained without thermal cycling and as they suggest could be influenced by thermal cycling in salivary medium as in this work, we have been able to complete the research of Albertini et a (Albertini et al., 2022).
It has been possible to observe the consequences of the structural changes of semi-crystallization with the volume contractions of the aligners when the material is not subjected to mechanical loading. These results could be compared with the work of Condo et al. (CondoPazzini et al., 2018). These authors characterized the samples by means of Fourier transform infrared spectroscopy, micro-Raman spectroscopy, X-ray diffraction, tensile and indentation strength test. The results showed an increase of crystalline fraction producing an increase of hardness and hyper-plasticity.
This research work has several limitations as we have tried to simulate the oral environment in which the aligners work. However, chewing stress levels have not been considered, as well as the bacterial environment to which the aligner may be subjected. One of the limitations of this study is the determination of dental forces and occlusal forces. A good model is the study by Duanmu Zheng et al. (Duanmu et al., 2021) where they developed a biomechanical model for the analysis of occlusal stresses. Future work could use this model to develop the behavior of aligners under the conditions described in this contribution. In any case, the mechanical behavior of the aligners has been verified and the hypothesis put forward in the study has been confirmed.
5 Conclusion
Polyurethane dental aligners have characteristic creep curves with transition zones for loads of 3 and 30 N. These plateaus, which are optimal for dental alignment, are reduced by the effect of the load and thermal cycling. It was found that the water absorption in the inserts is constant after 1 h and reaches values of 0.4% by weight of the polyurethane. It has also been confirmed that thermal cycling in the insoles without mechanical stress is susceptible to shrinkage due to semi-crystallisation processes of the polymeric material.
Acknowledgments
The authors are grateful to the Universidad Alfonso X el sabio y el Banco de Santander for the administrative and technical support.
Funding Statement
The author(s) declare financial support was received for the research, authorship, and/or publication of this article. The work was supported by the Spanish Government and the Ministry of Science and Innovation of Spain by research projects CONCEPTO PDC 2022-133628-C22 (co-funded by the European Regional Development Fund (ERDF), a way to build Europe).
Data availability statement
The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.
Author contributions
PC-L: Investigation, Project administration, Writing–original draft. PA-B: Methodology, Project administration, Visualization, Writing–original draft. AM-V: Conceptualization, Investigation, Writing–review and editing. MP-C: Investigation, Methodology, Writing–review and editing. JG: Conceptualization, Investigation, Writing–original draft. JA: Data curation, Funding acquisition, Software, Writing–review and editing.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- Abbate G. M., Caria M. P., Montanari P., Mannu C., Orrù G., Caprioglio A., et al. (2015). Periodontal health in teenagers treated with removable aligners and fixed orthodontic appliances. J. Orofac. Orthop. 76 (3), 240–250. 10.1007/s00056-015-0285-5 [DOI] [PubMed] [Google Scholar]
- Albertini P., Mazzanti V., Mollica F., Pellitteri F., Palone M., Lombardo L. (2022). Stress relaxation properties of five orthodontic aligner materials: a 14-day in-vitro study. Bioengineering 9, 349. 10.3390/bioengineering9080349 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Align Technology (2023). Inc. INVISALIGNt presenta smart-track. Available in: http://invisalign.com.au/doctor/doc/folletos/SmartTrack_brochure.pdf August 8, 2023.
- Bellini H., Moyano J., Gil J., Puigdollers A. (2016). Comparison of the superelasticity of different nickel–titanium orthodontic archwires and the loss of their properties by heat treatment. J.Mater.Sci. Mater. Med. 27, 158. 10.1007/s10856-016-5767-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bichu Y. M., Alwafi A., Liu X., Andrews J., Ludwig B., Bichu A. Y., et al. (2022). Advances in orthodontic clear aligner materials. Bioact. Mater 20 (22), 384–403. 10.1016/j.bioactmat.2022.10.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buxadera-Palomero J., Calvo C., Torrent-Camarero S., Gil F. J., Mas-Moruno C., Canal C., et al. (2017). Biofunctional polyethylene glycol coatings on titanium: an in vitro-based comparison of functionalization methods. Colloids Surfaces B Biointerfaces 152, 367–375. 10.1016/j.colsurfb.2017.01.042 [DOI] [PubMed] [Google Scholar]
- Cianci C., Pappalettera G., Renna G., Casavola C., Laurenziello M., Battista G., et al. (2020). Mechanical behavior of PET-G tooth aligners under cyclic loading. Front. Mater. 7. 10.3389/fmats.2020.00104 [DOI] [Google Scholar]
- CondoPazzini R. L., Cerroni L., Pasquantonio G., Lagana G., Pecora A., Mussi V., et al. (2018). Mechanical properties of “two generations” of teeth aligners: change analysis during oral permanence. Dent. Mater. 37 (5), 835–842. 10.4012/dmj.2017-323 [DOI] [PubMed] [Google Scholar]
- Cui Y., Clyne T. W. (2023). Creep of particle and short fibre reinforced polyurethane rubber. Mech. Time-Depend Mater 27, 45–58. 10.1007/s11043-021-09524-x [DOI] [Google Scholar]
- Dalaie K., Fatemi S. M., Ghaffari S. (2021). Dynamic mechanical and thermal properties of clear aligners after thermoforming and aging. Prog. Orthod. 28 (1), 15. 10.1186/s40510-021-00362-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Djeu G., Shelton C., Maganzini A. (2005). Outcome assessment of Invisalign and traditional orthodontic treatment compared with the American Board of Orthodontics objective grading system. Am. J. Orthod. Dentofac. Orthop. 128 (3), 292–298. 10.1016/j.ajodo.2005.06.002 [DOI] [PubMed] [Google Scholar]
- Duanmu Z., Liu L., Deng Q., Ren Y., Wang M. (2021). Development of a biomechanical model for dynamic occlusal stress analysis. Int. J. Oral Sci. 13, 29. 10.1038/s41368-021-00133-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Eliades T., Bisphenol A. (2017). Bisphenol A and orthodontics: an update of evidence-based measures to minimize exposure for the orthodontic team and patients. Am. J. Orthod. Dentofac. Orthop. 152 (4), 435–441. 10.1016/j.ajodo.2017.08.004 [DOI] [PubMed] [Google Scholar]
- Eliades T., Eliades G., Watts D. C. (1999). Structural conformation of in vitro and in vivo aged orthodontic elastomeric modules. Eur. J. Orthod. 21 (6), 649–658. 10.1093/ejo/21.6.649 [DOI] [PubMed] [Google Scholar]
- Eliades T., Gioni V., Kletsas D., Athanasiou A., Eliades G. (2007). Oestrogenicity of orthodontic adhesive resins. Eur. J. Orthod. 29 (4), 404–407. 10.1093/ejo/cjm040 [DOI] [PubMed] [Google Scholar]
- Eliades T., Pratsinis H., Athanasiou A. E., Eliades G., Kletsas D. (2009). Cytotoxicity and estrogenicity of Invisalign appliances. Am. J. Orthod. Dentofac. Orthop. 136 (1), 100–103. 10.1016/j.ajodo.2009.03.006 [DOI] [PubMed] [Google Scholar]
- Elkholy F., Panchaphongsaphak T., Kilic F., Schmidt F., Lapatki B. G. (2015). Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J. Orofac. Orthop. 76 (6), 460–475. 10.1007/s00056-015-0307-3 [DOI] [PubMed] [Google Scholar]
- Espinar-Escalona E., Llamas-Carreras J. M., Barrera-Mora J. M., Abalos-Lasbrucci C., Gil-Mur F. J. (2013). Effect of temperature on the orthodontic clinical applications of NiTi closed-coil springs. Med. Oral Patol. Oral Cir. Bucal 18 (4), e721–e724. 10.4317/medoral.19073 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Findley W. N., Davis F. A. (2013). Creep and relaxation of nonlinear viscoelastic materials. North Holand. London: Courier Corporation. [Google Scholar]
- Francisco I., Paula A. B., Ribeiro M., Marques F., Travassos R., Nunes C., et al. (2022). The biological effects of 3D resins used in orthodontics: a systematic review. Bioeng. (Basel) 3 (1), 15. 10.3390/bioengineering9010015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ge Q., Yu K., Ding Y., Qi H. J. (2012). Prediction of temperature-dependent free recovery behaviors of amorphous shape memory polymers. Soft Matter 8 (43), 11098–11105. 10.1039/C2SM26249E [DOI] [Google Scholar]
- Gil F. J., Cenizo M., Espinar E., Rodriguez A., Rúperez E., Manero J. M. (2013). NiTi superelastic orthodontic wires with variable stress obtained by ageing treatments. Mater. Lett. 104, 5–7. 10.1016/j.matlet2013.03.135 [DOI] [Google Scholar]
- Gil F. J., Espinar E., Llamas J. M., Manero J. M., Ginebra M. P. (2012). Variation of the superelastic properties and nickel release from original and reused NiTi orthodontic archwires. J. Mech. Beh. Biomed. Mater. 6, 113–119. 10.1016/j.jmbbm.2011.11.005 [DOI] [PubMed] [Google Scholar]
- Gil F. J., Planell J. A. (1993). Aplicaciones biomédicas del titanio y aleaciones. Biomecánica. 1, 34–43. [Google Scholar]
- Gil F. J., Rupérez E., Velasco E., Aparicio C., Manero J. M. (2018). Mechanism of fracture of NiTi superelastic endodontic rotary instruments. J. Mater. Sci. Mater. Med. 29 (131), 131. 10.1007/s10856-018-6140-7 [DOI] [PubMed] [Google Scholar]
- Gil F. J., Solano E., Peña J., Engel E., Mendoza A., Planell J. A. (2004). Microstructural, mechanical and citotoxicity evaluation of different NiTi and NiTiCu shape memory alloys. J. Mater Sci. Mater Med. 15, 1181–1185. 10.1007/s10856-004-5953-8 [DOI] [PubMed] [Google Scholar]
- Godoy-Gallardo M., Guillem-Marti J., Sevilla P., Manero J. M., Gil F. J., Rodriguez D. (2016b). Anhydride-functional silane immobilized onto titanium surfaces induces osteoblast cell differentiation and reduces bacterial adhesion and biofilm formation. Mater Sci. Eng. C Mater Biol. Appl. 59, 524–532. 10.1016/j.msec.2015.10.051 [DOI] [PubMed] [Google Scholar]
- Godoy-Gallardo M., Manzanares-Céspedes M. C., Sevilla P., Nart J., Manzanares N., Manero J. M., et al. (2016a). Evaluation of bone loss in antibacterial coated dental implants: an experimental study in dogs. Mater Sci. Eng. C Mater Biol. Appl. 69, 538–545. 10.1016/j.msec.2016.07.020 [DOI] [PubMed] [Google Scholar]
- Gomez J. P., Peña F. M., Martínez V., Giraldo D. C., Cardona C. I. (2015). Initial force systems during bodily tooth movement with plastic aligners and composite attachments: a three-dimensional finite element analysis. Angle Orthod. 85 (3), 454–460. 10.2319/050714-330.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gracco A., Mazzoli A., Favoni O., Conti C., Ferraris P., Tosi G., et al. (2009). Short-term chemical and physical changes in invisalign appliances. Aust. Orthod. J. 25 (1), 34–40. [PubMed] [Google Scholar]
- Gu J., Tang J. S., Skulski B., Fields H. W., Beck F. M., Firestone A. R., et al. (2017). Evaluation of Invisalign treatment effectiveness and efficiency compared with conventional fixed appliances using the Peer Assessment Rating index. Am. J. Orthod. Dentofac. Orthop. 151 (2), 259–266. 10.1016/j.ajodo.2016.06.041 [DOI] [PubMed] [Google Scholar]
- Halimi A., Benyahia H., Bahije L., Adli H., Azeroual M. F., Zaoui F. (2016). A systematic study of the release of bisphenol A by orthodontic materials and its biological effects. Int. Orthod. 14 (4), 399–417. 10.1016/j.ortho.2016.10.005 [DOI] [PubMed] [Google Scholar]
- Han J. Y. (2015). A comparative study of combined periodontal and orthodontic treatment with fixed appliances and clear aligners in patients with periodontitis. J. Periodontal Implant Sci. 45 (6), 193–204. 10.5051/jpis.2015.45.6.193 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hartshorne J., Wertheimer M. B. (2022). Emerging insights and new developments in clear aligner therapy: a review of the literature. AJO-DO Clin. Companion. 2 (4), 311–324. 10.1016/j.xaor.2022.04.009 [DOI] [Google Scholar]
- Hennessy J., Garvey T., Al-Awadhi E. A. (2016). A randomized clinical trial comparing mandibular incisor proclination produced by fixed labial appliances and clear aligners. Angle Orthod. 86 (5), 706–712. 10.2319/101415-686.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herrero-Climent M., Punset M., Molmeneu M., Brizuela A., Gil J. (2023). Differences between the fittings of dental prostheses produced by CAD-CAM and laser sintering processes. J. Funct. Biomater. 14, 67. 10.3390/jfb14020067 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hodge R. M., Edward G. H., Simon G. P. (1996). Water absorption and states of water in semicrystalline poly(vinyl alcohol) films. Polymer 37 (8), 1371–1376. 10.1016/0032-3861(96)81134-7 [DOI] [Google Scholar]
- Huang W. M., Zhao Y., Wang C. C., Ding Z., Purnawali H., Tang C., et al. (2012). Thermo/chemo-responsive shape memory effect in polymers: a sketch of working mechanisms, fundamentals and optimization. J. Polym. Res. 19 (9), 9952. 10.1007/s10965-012-9952-z [DOI] [Google Scholar]
- Ihssen B. A., Willmann J. H., Nimer A., Drescher D. (2019). Effect of in vitro aging by water immersion and thermocycling on the mechanical properties of PETG aligner material. J. Orofac. Orthop. 80 (6), 292–303. 10.1007/s00056-019-00192-8 [DOI] [PubMed] [Google Scholar]
- Iijima M., Kohda N., Kawaguchi K., Muguruma T., Ohta M., Naganishi A., et al. (2015). Effects of temperature changes and stress loading on the mechanical and shape memory properties of thermoplastic materials with different glass transition behaviours and crystal structures. Eur. J. Orthod. 37 (6), 665–670. 10.1093/ejo/cjv013 [DOI] [PubMed] [Google Scholar]
- Jaggy F., Zinelis S., Polychronis G., Patcas R., Schätzle M., Eliades G., et al. (2020). ATR-FTIR analysis and one-week stress relaxation of four orthodontic aligner materials. Mater. (Basel) 13 (8), 1868. 10.3390/ma13081868 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jhao Y.-S., Ouyang H., Yang F., Lee S. (2022). Thermo-mechanical and creep behaviour of polylactic acid/thermoplastic polyurethane blends. Polymers 14, 5276. 10.3390/polym14235276 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jia Y., Peng K., Gong X., Zhang Z. (2011). Creep and recovery of polypropylene/carbon nanotube composites. Int. J. Plasticity 27, 1239–1251. 10.1016/j.ijplas.2011.02.004 [DOI] [Google Scholar]
- Kasgoz A. (2021). Mechanical, tensile creep and viscoelastic properties of thermoplastic polyurethane/polycarbonate blends. Fibers Polym. 22, 295–305. 10.1007/s12221-021-0113-z [DOI] [Google Scholar]
- Keim R. G., Gottlieb E. L., Vogels D. S., Vogels P. B. (2014). 2014 JCO study of orthodontic diagnosis and treatment procedures, Part 1: results and trends. J. Clin. Orthod. 48 (10), 607–630. [PubMed] [Google Scholar]
- Kesling H. D. (1945). The philosophy of the tooth positioning appliance. Am. J. Orthod. Oral Surg. 31, 297–304. 10.1016/0096-6347(45)90101-3 [DOI] [Google Scholar]
- Lanteri V., Farronato G., Lanteri C., Caravita R., Cossellu G. (2018). The efficacy of orthodontic treatments for anterior crowding with Invisalign compared with fixed appliances using the Peer Assessment Rating Index. Quintessence Int. 49 (7), 581–587. 10.3290/j.qi.a40511 [DOI] [PubMed] [Google Scholar]
- Li W., Wang S., Zhang Y. (2015). The effectiveness of the Invisalign appliance in extraction cases using the the ABO model grading system: a multicenter randomized controlled trial. Int. J. Clin. Exp. Med. 15 (5), 8276–8282. [PMC free article] [PubMed] [Google Scholar] [Retracted]
- Lombardo L., Arreghini A., Maccarrone R., Bianchi A., Scalia S., Siciliani G. (2015). Optical properties of orthodontic aligners--spectrophotometry analysis of three types before and after aging. Prog. Orthod. 16, 41. 10.1186/s40510-015-0111-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lombardo L., Marafioti M., Stefanoni F., Mollica F., Siciliani G. (2012). Load deflection characteristics and force level of nickel titanium initial archwires. Angle Orthod. 82 (3), 507–521. 10.2319/032511-213.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lombardo L., Martines E., Mazzanti V., Arreghini A., Mollica F., Siciliani G. (2017). Stress relaxation properties of four orthodontic aligner materials: a 24-hour in vitro study. Angle Orthod. 87 (1), 11–18. 10.2319/113015-813.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lu H., Wang X., Yu K., Huang W. M., Yao Y., Leng J. (2017). A phenomenological formulation for the shape/temperature memory effect in amorphous polymers with multi-stress components. Smart Mater Struct. 26 (9), 095011. 10.1088/1361-665x/aa77b3 [DOI] [Google Scholar]
- Manero J. M., Gil F. J., Padrós E., Planell J. A. (2003). Applications of environmental scanning electron microscopy (ESEM) in biomaterials field. Microsc. Res. Tech. 61, 469–480. 10.1002/jemt.10358 [DOI] [PubMed] [Google Scholar]
- Mima M., Iwamoto R., Yoshikawa S. (1983). Highly deacetylated chitosan and its properties. J. Appl. Polym. Sci. 28 (6), 1909–1917. 10.1002/app.1983.070280607 [DOI] [Google Scholar]
- Morresi A. L., D'Amario M., Capogreco M., Gatto R., Marzo G., D'Arcangelo C., et al. (2014). Thermal cycling for restorative materials: does a standardized protocol exist in laboratory testing? A literature review. J. Mech. Behav. Biomed. Mater 29, 295–308. 10.1016/j.jmbbm.2013.09.013 [DOI] [PubMed] [Google Scholar]
- Murakami H., Kawai T., Hisada K., Mori S., Hasegawa J. (1983). Fluid behavior of saliva in dental materials science. Aichi Gakuin Daigaku Shigakkai Shi 21 (4), 702–707. [PubMed] [Google Scholar]
- Padrós R., Punset M., Molmeneu M., Velasco A. B., Herrero-Climent M., Rupérez E., et al. (2020). Mechanical properties of CoCr dental-prosthesis restorations made by three manufacturing processes. Influence of the microstructure and topography. Metals 10, 788. 10.3390/met10060788 [DOI] [Google Scholar]
- Papadopoulou A. K., Cantele A., Polychronis G., Zinelis S., Eliades T. (2019). Changes in roughness and mechanical properties of Invisalign® appliances after one- and two-weeks use. Mater. (Basel) 12 (15), 2406. 10.3390/ma12152406 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pascual B., Gurruchaga M., Ginebra M. P., Gil F. J., Planell J. A., Goñi I. (1999). Influence of the modification of P/L ratio on a new formulation of acrylic bone cement. Biomaterials 20 (5), 465–474. 10.1016/S0142-9612(98)00192-6 [DOI] [PubMed] [Google Scholar]
- Raghavan A. S., Pottipalli Sathyanarayana H., Kailasam V., Padmanabhan S. (2017). Comparative evaluation of salivary bisphenol A levels in patients wearing vacuum-formed and Hawley retainers: an in-vivo study. Am. J. Orthod. Dentofac. Orthop. 151 (3), 471–476. 10.1016/j.ajodo.2016.07.022 [DOI] [PubMed] [Google Scholar]
- Rossini G., Parrini S., Castroflorio T., Deregibus A., Debernardi C. L. (2015). Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod. 85 (5), 881–889. 10.2319/061614-436.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Salvekar A. V., Huang W. M., Xiao R., Wong Y. S., Venkatraman S. S., Tay K. H., et al. (2017). Water-Responsive shape recovery induced buckling in biodegradable photo-cross-linked poly(ethylene glycol) (PEG) hydrogel. Acc. Chem. Res. 50 (2), 141–150. 10.1021/acs.accounts.6b00539 [DOI] [PubMed] [Google Scholar]
- Schuster S., Eliades G., Zinelis S., Eliades T., Bradley T. G. (2004). Structural conformation and leaching from in vitro aged and retrieved Invisalign appliances. Am. J. Orthod. Dentofac. Orthop. 126 (6), 725–728. 10.1016/j.ajodo.2004.04.021 [DOI] [PubMed] [Google Scholar]
- Shalish M., Cooper-Kazaz R., Ivgi I., Canetti L., Tsur B., Bachar E., et al. (2012). Adult patients' adjustability to orthodontic appliances. Part I: a comparison between Labial, Lingual, and Invisalign™. Eur. J. Orthod. 34(6):724–730. 10.1093/ejo/cjr086 [DOI] [PubMed] [Google Scholar]
- Simon M., Keilig L., Schwarze J., Jung B. A., Bourauel C. (2014). Treatment outcome and efficacy of an aligner technique--regarding incisor torque, premolar derotation and molar distalization. BMC Oral Health 11 (14), 68. 10.1186/1472-6831-14-68 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Skaik A., Li Wei X., Abusamak I., Iddi I. (2019). Effects of time and clear aligner removal frequency on the force delivered by different polyethylene terephthalate glycol-modified materials determined with thin-film pressure sensors. Am. J. Orthod. Dent. Orthop. 155 (1), 98–107. 10.1016/j.ajodo.2018.03.017 [DOI] [PubMed] [Google Scholar]
- Tartaglia G. M., Mapelli A., Maspero C., Santaniello T., Serafin M., Farronato M., et al. (2021). Direct 3D printing of clear orthodontic aligners: current state and future possibilities. Mater. (Basel) 14 (7), 1799. 10.3390/ma14071799 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vallet-Regí M., Román J., Padilla S., Doadrio J. C., Gil F. J. (2005). “Bioactivity and mechanical properties of SiO2-CaO-P2O5 glass-ceramics”. J. Mater. Chem. 15, 1353–1359. 10.1039/b415134h [DOI] [Google Scholar]
- Wu X. L., Huang W. M., Lu H. B., Wang C. C., Cui H. P. (2017). Characterization of polymeric shape memory materials. J. Polym. Eng. 37 (1), 1–20. 10.1515/polyeng-2015-0370 [DOI] [Google Scholar]
- Wu Y. K., Lodoen G. A., Anderson J. M., Baer E., Hiltner A. (1994). Creep of a poly(etherurethane urea) in an oxidative environment. J. Biomed. Mater Res. 28 (4), 515–522. 10.1002/jbm.820280414 [DOI] [PubMed] [Google Scholar]
- Xie T., Rousseau I. A. (2009). Facile tailoring of thermal transition temperatures of epoxy shape memory polymers. Polym. Guildf. 50 (8), 1852–1856. 10.1016/j.polymer.2009.02.035 [DOI] [Google Scholar]
- Yi J., Xiao J., Li Y., Li X., Zhao Z. (2018). External apical root resorption in non-extraction cases after clear aligner therapy or fixed orthodontic treatment. J. Dent. Sci. 13 (1), 48–53. 10.1016/j.jds.2017.09.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.