Early removal of swabs can accelerate laboratory work and enhance quality by decreasing contamination and errors in downstream procedures. This article aims to compare the traditional testing workflow in a diagnostic laboratory (red), with manual swab removal and aliquoting vs a swab-out laboratory workflow (blue).
Keywords: swabbing, SARS-CoV-2 detection, viral load, cycle threshold (Ct)
Abstract
Removing the swab after collection can speed up diagnosis and improve the quality of laboratory procedures. This study investigates the impact of swab removal on SARS-CoV-2 detection in clinical specimens with a focus on high Cycle threshold (Ct) samples (Cts≥32). The method assessed pairs of SARS-CoV-2 samples mimicking combined throat and nose swabs and tested them on two real-time-PCR platforms; the Applied Biosystems 7500 and the Abbott Alinity. Swab removal did not significantly affect detection rates of SARS-CoV-2 samples with Ct values<32, regardless of the PCR platform. However, reduced reproducibility was seen at the endpoint limit of detection of the platforms, which meant that fewer samples with Ct values≥32 were detected in the swab removal group.
Data Summary
A supplementary file is available with the online version of this article. All the supporting data discussed in this article can be found at https://github.com/azulzorzoli/swab_article_repository.
Impact Statement.
Most automated Real-Time PCR testing platforms require swabs to be removed from the tubes before analysis, a time-consuming task performed by laboratory staff. The COVID-19 pandemic encouraged diagnostic laboratories to explore ways to reduce contamination and errors while improving the speed of testing. Among the alternatives, laboratories considered removing the swab immediately after specimen collection to simplify the patient-to-result path, but the impact on performance has received limited attention in the literature. This article investigates the effect of removing the swab before specimen transport and addresses the critical issue of identifying samples with low viral load. Particular attention is given to the ability of different swab-collection methods to detect SARS-CoV-2 specimens with Cycle threshold (Ct) values ≥32.
By comparing ‘swab-in’ and ‘swab-out’ methods on identical samples, this work provides valuable insights into how sample collection techniques impact the sensitivity of viral detection. The research also includes a platform performance comparison, making it relevant for laboratory practice.
The study suggests that retaining the swab after sample collection may enhance the detection rate for weakly positive samples (Ct ≥32). Although the results are preliminary and would benefit from a larger dataset, they contribute to the ongoing discussions in diagnostic laboratories about how to interpret the results from specimens with low viral load.
Introduction
Correctly identifying positive samples, particularly in healthcare settings, is crucial in guiding public health measures and mitigating the spread of viral diseases [1, 2]. Specimens with low viral loads show high Cycle threshold (Ct) results in real-time PCR tests [3, 4]. These weak positives can be seen in patients shedding residual nucleic acid from a previous infection, but also because of poor swab collection technique. High Ct samples may also occur in pre-symptomatic or asymptomatic individuals.
Greene [5] and Stevenson [6] discussed sample collection strategies, mainly using swab removal as a potential pre-analytical strategy. Early removal of swabs can accelerate laboratory work and enhance quality by decreasing contamination and errors in downstream procedures. However, little attention has been given to the potential effect of swab-collection methods in detecting specimens with low viral load. This work compares swab-in and swab-out methods on identical samples in parallel, focusing on specimens with Ct values≥32. A platform performance comparison was also included in this study using the Applied Biosystems 7500 and the Abbott Alinity.
Methods and results
Thirty-one positive and twenty-seven negative SARS-CoV-2 testing samples mimicking combined throat and nose swabs were generated from serial dilutions of clinical specimens in Viral PCR Sample Solution (VPSS), the transport inactivation media routinely used in the laboratory. For the ‘swab-in’ samples, the swab was swirled in VPSS for 5 s, the shaft was broken, and the tube closed. For the ‘swab-out’ specimens, the swab was submerged into VPSS, swirled five times, squeezed against the tube’s wall and then discarded. Samples were incubated overnight at room temperature to simulate collection-to-reception time. The sample sets were tested through A) a partially automated workflow using an Easymag for nucleic acid extraction followed by the Applied Biosystems 7500 and the Altona RealStar SARS-CoV-2 RT-PCR Kit 1.0, or B) the fully automated platform Alinity m from Abbot.
Table 1 shows the average Cycle threshold (Ct) or Copy Number (CN) values measured for each sample group for each platform. Of note, Ct and CN values are not directly comparable across different platforms. CN numbers are often lower than Ct values, with the first estimating the number of gene or genomic region copies and the second counting the cycles needed to amplify a target gene to a detectable level. Table S1, available in the online version of this article contains all platforms and methods’ individual Ct/CN values.
Table 1.
Platform |
Method |
Ct/CN <32 Group |
Ct/CN ≥32 Group |
---|---|---|---|
ABI (E gene) |
Swab-in |
Average Ct=29.31 |
Average Ct=36.52 |
Range: 25.27–31.92 |
Range: 32.14–38.25 |
||
Swab-out |
Average Ct=28.84 |
Average Ct=36.35 |
|
Range: 25.42–31.15 |
Range: 32.08–39.49 |
||
Alinity m |
Swab-in |
Average CN=28.05 |
Average CN=35.67 |
Range: 21.86–31.85 |
Range: 32.12–38.80 |
||
Swab-out |
Average CN=27.84 |
Average CN=35.56 |
|
Range: 21.93–31.00 |
Range: 32.03–40.83 |
Fig. 1 summarises the agreement between the two swab methodologies for each platform and Ct/CN group. The results show no discordant results for samples with strong or moderately strong viral loads (Ct <32), but several high-Ct (≥32) samples were not detected. The expected variation in sample collection resulted in some samples failing detection using the swab-in method but being positive after removing the swab. These inconsistencies were interpreted cautiously, with only the positive samples in both methods or positive in swab-in and negative in swab-out, being considered for analysis. Our data analysis strategy considered samples positive only when both replicates were detected. Despite this, a paired t-test revealed no statistical differences between the means of the swab-in vs swab-out samples regardless of the swab group tested (strongly positive and moderately positive, with Ct <32, and weakly positive with Ct ≥32) and the platform used (Figs S1 and S2, Tables S2 and S4).
The datasets were analysed for each platform individually, comparing the expected results (swab-in) to the ones resulting from the swab removal. Table 2 details the Positive Percentage Agreement (PPA), Negative Percentage Agreement (NPA) and Percentage Overall Agreement (POA) results when comparing the swab-in vs swab-out method. ABI performed slightly better than Alinity when comparing the overall results (POA 90.9 % vs 87.9 %). When testing samples with Ct <32, these two platforms showed no detrimental effect, with PNA, NPA and POA of 100.00 %. However, lower PPA and PNA were seen for the Ct ≥32 groups, with ABI performing better than the Alinity platform (PPA 77.3 % vs 65.0 %). The contingency tables in the supplementary materials describe these results in more detail (Table S3).
Table 2.
Group |
Platform |
PPA |
NPA |
POA |
no. of positive samples tested |
no. of negative samples tested |
---|---|---|---|---|---|---|
All samples |
Alinity |
81.6 % (CI 95 %=66.6–90.8 %) |
96.4 % (CI 95 %=82.3–99.4 %) |
87.9 % (CI 95 %=77.9–93.7 %) |
38 |
28 |
CN <32 |
Alinity |
100.0 % (CI 95 %=82.4–100.0 %) |
100.0 % (CI 95 %=87.5–100.0 %) |
100.0 % (CI 95 %=92.1–100.0 %) |
18 |
27 |
CN ≥32 |
Alinity |
65.0 % (CI 95 %=43.3–91.9 %) |
96.4 % (CI 95 %=82.3–99.4 %) |
83.3 % (CI 95 %=70.9–91.3 %) |
20 |
28 |
All samples |
ABI |
83.3 % (CI 95 %=66.4–92.7 %) |
97.2 % (CI 95 %=85.8–99.5 %) |
90.9 % (CI 95 %=81.6–95.8 %) |
30 |
36 |
Ct <32 |
ABI |
100.0 % (CI 95 %=67.6–100.0 %) |
100.0 % (CI 95 %=90.1–100.0 %) |
100.0 % (CI 95 %=91.8–100.0 %) |
8 |
35 |
Ct ≥32 |
ABI |
77.3 % (CI 95 %=56.6–89.9 %) |
97.2 % (CI 95 %=85.8–99.5 %) |
89.7 % (CI 95 %=79.2–95.2 %) |
22 |
36 |
The COVID-19 pandemic saw the development of new tests for detecting SARS-CoV-2 and a plethora of information describing assays, platforms, and sample collection strategies [7, 8]. In this work, we evaluated if discarding the swab after sample collection would increase the risk detection rate, specifically for weakly positive samples (Ct ≥32). We saw no impact in samples with strongly positive and moderately positive viral loads (here defined as samples with Cts<32) but we recorded a lower detection rate after swab removal in samples with Ct values above 32.
A careful methodology was used to collect paired specimen samples to reduce inconsistencies, but methodological differences may have still impacted the results presented here. Issues with this collection methodology can introduce errors that confound the interpretation of clinical results. For instance, low viral load samples might result from poorly collected swab specimens. However, they might also reflect the initial stages of an infectious process or the residual viral RNA shedding from individuals who are no longer infectious [9, 10]. Although this article does not aid in clinical interpretation, it can be used to make reporting decisions for samples with low viral loads.
Many laboratories have long-standing protocols for rejecting empty containers and only processing those with a retained swab to minimise processing errors. Those pathways could be revised in light of these results, as specimens with Ct <32 are unaffected by the swab removal. By doing so, laboratories can ensure the processing of more specimens and provide high-quality results to their patients and healthcare providers.
Our data suggest that retaining the swab could help detect weakly positive samples. However, a larger data set would be required to reach a definitive conclusion. It is possible that the reduction in the assay’s ability to consistently detect low viral loads, which occurs when working at the platform’s detection levels, compounded by the expected variations in the swirling technique, is leading to the observed decrease in the detection rate.
Supplementary Data
Funding information
This research did not receive any specific grant from funding agencies in the public, commercial, or nor-for-profit sectors.
Acknowledgements
We thank Imogen Johnston-Menzies and Ahlam Darasi for their assistance with the laboratory testing of samples. Thanks to Nikki Hanley and the rest of the NHS Lothian Virology team for their help with this project. Thanks to Noha ElSakka for critically reviewing this manuscript.
Author contributions
A.Z.: conceptualisation, methodology, investigation, data curation, and writing of the original draft. S.B., A.M.: analysis and editing of the original draft. G.M.: conceptualisation and data curation. R.G.: Analysis and review of the manuscript. K.T.: conceptualisation, editing of the original draft and the overall supervision of the study.
Conflicts of interest
The author(s) declare that there are no conflicts of interest.
Footnotes
Abbreviations: CN, copy number; Ct, cycle threshold; NPA, negative percentage agreement; POA, percentage overall agreement; PPA, positive percentage agreement; VPSS, viral PCR sample solution.
Two supplementary figures and four supplementary tables are available with the online version of this article.
References
- 1.Public Health England Evaluations, validations and verifications of diagnostic tests. UK Standards for Microbiology Investigations. 2017;Q1:1–45. vol. [Google Scholar]
- 2.Cevik M, Tate M, Lloyd O, Maraolo AE, Schafers J, et al. SARS-CoV-2, SARS-CoV, and MERS-CoV viral load dynamics, duration of viral shedding, and infectiousness: a systematic review and meta-analysis. Lancet Microbe. 2021;2:e13–e22. doi: 10.1016/S2666-5247(20)30172-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Puhach O, Meyer B, Eckerle I. SARS-CoV-2 viral load and shedding kinetics. Nat Rev Microbiol. 2023;21:147–161. doi: 10.1038/s41579-022-00822-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ehret R, Breuer S, Dhein J, Reinhardt B, Obermeier M. Clinical evaluation of the automated abbott realtime SARS-CoV-2, alinity m SARS-CoV-2, and alinity m Resp-4-Plex assays. J Virol Methods. 2022;299:114338. doi: 10.1016/j.jviromet.2021.114338. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Greene DN, Matthys T, Lockwood CM. Swab-free transport as an optimized preanalytical workflow for SARS-CoV-2 amplification. J Appl Lab Med. 2021;6:606–613. doi: 10.1093/jalm/jfaa197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Stevenson LG, deFilippi CR. Swab-free transport as an optimized preanalytical workflow for SARS-CoV-2 molecular testing. J Appl Lab Med. 2021;6:580–583. doi: 10.1093/jalm/jfaa242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Davies E, Farooq HZ, Brown B, Tilston P, McEwan A, et al. An overview of SARS-CoV-2 molecular diagnostics in Europe. Clin Lab Med. 2022;42:161–191. doi: 10.1016/j.cll.2022.02.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Mosscrop L, Watber P, Elliot P, Cooke G, Barclay W, et al. Evaluation of the impact of pre-analytical conditions on sample stability for the detection of SARS-CoV-2 RNA. J Virol Methods. 2022;309:114607. doi: 10.1016/j.jviromet.2022.114607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.van Kampen JJA, van de Vijver DAMC, Fraaij PLA, Haagmans BL, Lamers MM, et al. Duration and key determinants of infectious virus shedding in hospitalized patients with coronavirus disease-2019 (COVID-19) Nat Commun. 2021;12:267. doi: 10.1038/s41467-020-20568-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.He X, Lau EHY, Wu P, Deng X, Wang J, et al. Temporal dynamics in viral shedding and transmissibility of COVID-19. Nat Med. 2020;26:1491–1493. doi: 10.1038/s41591-020-1016-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.