Abstract
PURPOSE
This study aimed to investigate whether the accuracy of intraoral scanners is influenced by different scanning strategies in an in vitro setting, through a systematic review and meta-analysis.
MATERIALS AND METHODS
This review was conducted in accordance with the PRISMA 2020 standard. The following PICOS approach was used: population, tooth impressions; intervention, the use of intraoral scanners with scanning strategies different from the manufacturer’s instructions; control, the use of intraoral scanners following the manufacturers’ requirements; outcome, accuracy of intraoral scanners; type of studies, in vitro. A comprehensive literature search was conducted across various databases including Embase, SciELO, PubMed, Scopus, and Web of Science. The inclusion criteria were based on in vitro studies that reported the accuracy of digital impressions using intraoral scanners. Analysis was performed using Review Manager software (version 5.3.5; Cochrane Collaboration, Copenhagen, Denmark). Global comparisons were made using a standardized mean difference based on random-effect models, with a significance level of α = 0.05.
RESULTS
The meta-analysis included 15 articles. Digital impression accuracy significantly improved under dry conditions (P < 0.001). Moreover, trueness and precision were enhanced when artificial landmarks were used (P ≤ 0.02) and when an S-shaped pattern was followed (P ≤ 0.01). However, the type of light used did not have a significant impact on the accuracy of the digital intraoral scanners (P ≥ 0.16).
CONCLUSION
The accuracy of digital intraoral scanners can be enhanced by employing scanning processes using artificial landmarks and digital impressions under dry conditions.
Keywords: Accuracy, Computer-Aided Design, Digital impression, Intraoral scanner, Precision, Dental impression technique, Computer-assisted diagnosis
INTRODUCTION
Conventional impression procedures have traditionally been used to capture the three-dimensional (3D) geometry of dental tissues. However, these methods have limitations due to the potential for errors in analogue impression material volume and the subsequent expansion of dental stone, often necessitating additional dental laboratory services.1,2,3 In response to these challenges, intraoral scanners (IOS) have been developed for dental clinicians.4,5
Critical to any lifelong dental restoration is the accuracy of the impression, which determines the marginal fit and overall quality of the final restoration.5,6 It’s important to note that the accuracy of a measurement method should adhere to the International Organization for Standardization (ISO) guidelines, which encompass both trueness and precision (ISO 5725-1).7 Trueness relates to the systematic error or measurement bias between the target object and the reference object, indicating the agreement between an accepted reference value and the mean value derived from multiple test outcomes. In contrast, precision accounts for the random errors when a procedure is repeated.8,9,10 Prior research has reported precision and trueness values for conventional impressions for full-arch models in the range of 13 - 61 mm and 20 - 55 mm, respectively, while digital impressions using IOSs have shown corresponding values of 31 - 60 mm and 40 - 59 mm.11,12
Integrating an IOS into a dental practice can enhance treatment quality and the patient experience while serving as a valuable marketing tool.13 Therefore, it’s imperative for clinicians to understand IOS technology to efficiently integrate it into chairside dentistry.5 These IOS devices offer various scanning settings, and scanning procedures can be customized to suit operator preferences.14 All IOS systems generate 3D models by “stitching” together multiple images captured from different angles, with the success of this merging process closely tied to the operator’s steadiness during scanning.15
However, it’s essential to acknowledge that IOSs have limitations, which may restrict their use in certain clinical situations. IOSs typically require a well-defined finish line geometry for accurate scanning. Additionally, deep vertical preparations can impede proper light penetration for precise digital reconstruction, and environmental factors such as humidity and ambient lighting can affect the quality of the scanned surface.16,17,18,19
As of now, there is not a single IOS device recognized as the gold standard for the accuracy of digital impressions in fabricating dental restorations. Therefore, the aim of this study is to investigate whether in vitro, the accuracy of IOSs can be enhanced through alternative techniques through a systematic review and meta-analysis. The null hypothesis for this review posits that no difference in accuracy will be observed when employing alternative techniques in contrast to the manufacturer’s guidelines.
MATERIALS AND METHODS
This paper was achieved in agreement with the PRISMA 2020 guidelines.20 The protocol and important documents regarding to this review were recorded at Open Science Framework (https://osf.io/gf7b4/). The succeeding PICOS approach was executed: population, tooth impression; intervention, the use of digital (IOS) with strategies different to those suggested by the company; control, the use of digital (IOS) concurring to the manufacturers’ instructions; outcome, the accuracy (trueness and precision); type of studies, in vitro findings. A research question was performed: Does the use of strategies different to those recommended by the manufacturer improve the accuracy of IOS when taking a tooth impression?
1. Literature search
A search of the literature was accompanied until January 12th, 2023. The following databases were investigated: Embase, Scielo, PubMed, Scopus, and Web of Science. The search strategy and the keywords used in PubMed are outlined in Table 1; these were adapted for use in other databases. After completing the search, all identified articles were imported into the Rayyan QCRI mobile application.21
Table 1. Search strategy used in PubMed.
Search number | Keywords |
---|---|
#1 | Intraoral scanner OR Intraoral scanning device OR Digital impression scanner |
#2 | Accuracy OR Precision OR Trueness |
#3 | # 1 AND #2 |
2. Study selection
Two experienced researchers (CECS and RB) individually evaluated the titles and abstracts of all the articles. Selection was based on the following inclusion criteria: (1) in vitro studies reporting the accuracy of digital impressions using IOS, (2) evaluated the accuracy of quadrant or full-arch impressions, (3) inclusion of a control group using IOS following the manufacturer’s guidelines, and (4) availability of mean and standard deviation (SD) data. Review papers, case reports, case series, and documents published in languages other than English were excluded. In cases where the necessary information was lacking, the full article was thoroughly reviewed to ascertain eligibility. Additionally, the reference lists of selected articles were manually searched to identify any potentially overlooked manuscripts during the initial search, while adhering to the specified criteria. Any discrepancies in article eligibility were resolved through the judgment of a third researcher (LH).
3. Data Extraction
Relevant data from the selected articles were extracted using Microsoft Office Excel 2019 (Microsoft Corporation, Redmond, WA, USA). This information encompassed the study’s publication year, the IOS used, the tested strategies or techniques, and the primary results. In cases of missing or unclear data, the corresponding researchers of the studies were contacted via email to acquire the necessary information. Any unreceived information after a two-week waiting period was omitted from this review.
4. Quality Assessment
The risk of bias (RoB) of the manuscripts included in this review was independently assessed by two participants of the review (LH and RB) according to the evaluation of the following parameters: specimen randomization, single operator, control group, operator blinded, standardized specimens, and sample size calculation. If the authors reported the parameter, the study received a “√” for that parameter. In the case of not reporting the information, such parameter received a “Χ”. The RoB was classified according to the sum of “√” answers received: 1 to 2 denoted a high bias, 3 to 4 medium, and 5 to 6 showed a low RoB.
5. Statistical analysis
The meta-analysis was performed using the Review Manager software (version 5.3.5; The Cochrane Collaboration, Copenhagen, Denmark). A random-effect model was used for the analyses, and pooled-effect estimates were obtained by comparing the standardized mean difference of the trueness and precision of the digital models obtained following the manufacturers’ recommendations against the other strategies. A P-value < .05 was considered statistically significant. The heterogeneity and inconsistency were measured using the Cochran Q test and the I2 test.
RESULTS
A total of 3268 documents were initially retrieved from the various databases. After removing duplicates entries, 2628 articles were assessed by reading the title and abstract. Subsequently, 92 studies persisted for full text examination. Among these, 37 studies were excluded for the following reasons: (23) did not include comparisons between different scanning techniques,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44 (7) were clinical studies,45,46,47,48,49,50,51 (5) full-text manuscripts were inaccessible,52,53,54,55,56 (1) did not assess accuracy on teeth,57 and (1) did not use any digital IOS.58 Ultimately, 55 documents were included in the qualitative analysis. From these, 40 articles were further excluded from the meta-analysis due to the following reasons: (15) had dissimilar test conditions compared to other studies,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73 (10) lacked available mean and standard deviation data,6,74,75,76,77,78,79,80,81,82 (9) did not have other studies available for comparison,14,15,58,83,84,85,86,87,88 (3) did not include a control group,89,90,91 and (3) the accuracy could not be determined.16,92,93 Consequently, 15 articles remained for the quantitative analysis.94,95,96,97,98,99,100,101,102,103,104,105,106,107,108 A flowchart of the study selection process agreeing to the PRISMA statement is presented in Figure 1.
Fig. 1. PRISMA 2020 flow diagram of study identification.
Table 2 provides an overview of the characteristics of the studies included in the qualitative analysis. These studies evaluated various IOSs, and among the different techniques tested, the most commonly examined factors included scanning speed, scanning pattern, tip size, illuminance and color temperature of ambient light, position and operator, presence of humidity, scanning resolution, software version used, use of artificial markers, scanning distance, model stitching, and the utilization of scanning-aid materials.
Table 2. Summary of findings.
Study and year | Intraoral scanner used | Strategy or technique tested | Main results |
---|---|---|---|
Al-ibrahim, 202183 | inEos X5 (Dentsply-Sirona) | Scanning speed. | Different scanning speeds did not affect the precision. However, the trueness was affected between slow and standard scanning speed. |
3Shape D2000 (3Shape) | |||
An, 202295 | Emerald (Planmeca) | Scanning speed. | Small scanner tip adversely influenced trueness and precision. Fast scanning speed and S-shaped scan strategy has lower precision. |
Scanning patterns. | |||
Alenezi, 202292 | ARCTICA AutoScan (KaVo) | Scanning patterns. | Straight and zigzag scan patterns had marginal and internal gaps that were clinically recognized. |
Donmez, 202297 | TRIOS 3 (3Shape) | Scanning patterns. | After linear scan of complete arch, the rotational scanning of the palate did not develop any supplementary advantage for accuracy. |
Ender, 201312 | Lava COS (3MESPE) | Scanning patterns. | Full arch dental impressions were feasible with a high accuracy, if satisfactory scan approaches were implemented. |
Cerec Bluecam (Sirona Dental Systems) | |||
Cadent iTero (Cadent Ltd.) | |||
Feng, 202186 | TRIOS 3 (3Shape) | Scanning patterns. | The accuracy of IOS differed for full-arch from maxillary to mandibulary archs. |
Gavounelis, 202298 | i500 (Medit) | Scanning patterns. | The scan strategy affected the accuracy of com- plete-arch impressions. |
Kim, 202263 | TRIOS 3 (3Shape) | Scanning patterns. | The scan origin location have an influence on the accuracy of IOS. |
Latham, 201976 | Planmeca Emerald (Planmeca) | Scanning patterns. | The trueness and precision for some IOS were affected by scan pattern. |
TRIOS 3 (3Shape) | |||
iTero Element (Align Technology) | |||
CEREC Omnicam (Dentsply Sirona) | |||
Li, 2022102 | TRIOS 3 (3Shape) | Scanning patterns. | For complete-arch implant rehabilitation, the scan display meaningfully affected the speed and the accuracy of a digital impression. |
Carestream CS 3600 (Carestream) | |||
Mai, 202293 | i700 (Medit) | Scanning patterns. | The segmental scan by means of two scan parts seems to be efficient for full-arch intraoral scans. |
Pattamavilai, 202271 | Virtuo Vivo (Dental Wings Inc) | Scanning patterns. | Scanning pattern affected the trueness of the IOS. |
TRIOS 3 (3Shape) | |||
True Definition (3M ESPE) | |||
Kanjanasavitree, 2022100 | Trios (3Shape) | Scanning patterns. | Scanning patterns and artificial landmarks had influence on the accuracy of the IOS. |
Diker, 202161 | Trios 3 (3Shape) | Scanning patterns. | The accuracy was influenced by the scanning sequence of the different IOS tested. |
iTero Element 2 (Align Technology) | |||
CEREC Omnicam (Dentsply Sirona) | |||
Emerald (Planmeca) | |||
Virtuo Vivo (Dental Wings) | |||
Diker, 202159 | TRIOS 3 (3Shape) | Scanning patterns. | The scanning sequence affected the accuracy of the IOS tested. |
iTero Element 2 (Align Technology Inc) | |||
Cerec Omnicam (Dentsply Sirona) | |||
Planmeca Emerald (Plan-meca Oy) | |||
Cerec Primescan (Dentsply Sirona) | |||
Virtuo Vivo (Dental Wings Inc) | |||
Diker, 202060 | Trios 3 (3Shape) | Scanning patterns. | The scanning sequence affected the accuracy of digital impressions of IOS used. |
iTero Element (Align Technology Inc) | |||
Cerec Omnicam (Dentsply-Sirona) | |||
Planmeca Emerald (Planmeca) | |||
Cerec Primescan (Dentsply-Sirona) | |||
Virtuo Vivo (Dental Wings Inc | |||
Zarone, 202082 | TRIOS 3 (3Shape) | Scanning patterns. | Buccopalatal technique showed superior trueness and precision than palatobuccal strategy for the wrinkled typodont scans. |
Arakida, 201894 | True Definition (3M ESPE Dental Products) | Ambient light conditions. | For an appropriate digital impression, the 3900 K and 500 lux conditions were efficient for lighting condition. |
Ochoa-López, 202269 | TRIOS 3 (3Shape) | Ambient light conditions. | Ambient light affected the accuracy of IOSs evaluated. |
Primescan (Dentsply Sirona). | |||
iTero Element 5D (Align Technology) | |||
i500 (Medit), i700 (Medit) | |||
CS3600 (Carestream) | |||
CS3700 (Carestream | |||
Wesemann, 202081 | TRIOS 3 (3shape) | Ambient light conditions | The accuracy and scanning time of IOS was affected by the ambient light. |
Cerec Omnicam iTero Element (Dentsply Sirona) | |||
iTero Element (Align Technology) | |||
CS 3600 (Carestream Dental, Triangulation) | |||
Planmeca Emerald (Planmeca) | |||
Aadva IOS (GC Europe) | |||
Revilla-León, 201916 | iTero Element (Cadent Ltd) | Ambient light conditions. | The conditions of ambient light influenced the accuracy of the IOSs used in this study. |
CEREC Omnicam (Dentsply, Sirona) | |||
TRIOS 3 (3Shape) | |||
Revilla-León, 202080 | iTero Element (Cadent LTD) | Ambient light conditions. | Variations in ambient scanning light condition significantly affect the mesh quality |
Omnicam (Cerec-Sirona) | |||
TRIOS 3 (3Shape). | |||
Revilla-León, 2021107 | TRIOS 3 (3Shape) | Ambient light conditions. | 1000-lux of illumination light condition was suggested in an attempt to increase the scanning accuracy of the IOS. |
The chair light must be prevented. | |||
Cakmak, 202285 | ATOS 5 (GOM) | Software version. | Deviations in implant scan body scans could differ if different software version were used. |
TRIOS 3 (3Shape) | |||
Haddadi, 201815 | ATOS 5 (GOM) | Software version. | The accuracy of an IOS was influenced by the software version. |
Cerec Omnicam (Dentsply Sirona) | |||
Peroz, 202179 | TRIOS 3 (3shape) | Software version | The operator, the inspection software, and the mesh density displayed no influence on the true ness of the result. |
Chen, 202196 | Trios 3 (3Shape) | Presence of humidity. | The presence of liquid on the tooth surface might influence the accuracy of the IOS. |
Primescan (Dentsply-Sirona) | |||
Goómez-Polo, 202299 | TRIOS 3 (3Shape) | Presence of humidity. | By drying of the surface scanned, there is a rise in the IOS accuracy. |
Rapone, 2020106 | CS 3600 (Carestream Dental) | Presence of humidity. | Humidity affected the accuracy of IOS. |
TRIOS 3 (3Shape) | |||
CEREC Omnicam (Dentsply Sirona) | |||
You, 202291 | Trios 3 (3Shape) | Presence of humidity and variations in occlusal force during intraoral scanning. | A moist cavity surface without visually evident salivary contamination is acceptable. |
Song, 202090 | CS3600 (Carestream Dental) | Presence of humidity. | The presence of artificial saliva influenced the scanned images. |
i500 (Medit Corp) | |||
Trios3 (3Shape) | |||
Omnicam (Dentsply Sirona) | |||
Mizumoto, 201967 | TRIOS 3 (3Shape) | Scanning strategies. | Scanning techniques and scan bodies affected the accuracy of the IOS evaluated in this study. |
Mizumoto, 2019b77 | Trios3 (3Shape) | Scanning strategies. | The accuracy of digital scans of edentulous maxillary arch was alike regardless of the stitching procedure. |
Motel, 202078 | TRIOS 3 (3Shape) | Scanning strategies. | Scan strategies affected the quality of digital impressions. |
Müller, 201668 | TRIOS Pod scanner (3Shape) | Scanning strategies. | The strategy suggested by the company demonstrated the highest precision and trueness of full-arch scans. |
Oh, 202070 | i500 (Medit) | Scanning strategies. | The segmental method for scanning the area improved the accuracy of IOS. |
TRIOS 3 (3Shape) | |||
Lopes, 202264 | CEREC Primescan (Dentsply Sirona) | Scanning strategies. | Scan strategies affected the accuracy, however, implant angulation did not influence the accuracy of IOS. |
Mandelli, 201865 | 3M True Definition Scanner (3M ESPE) | Scanning strategies. | No stitching strategy showed less precision when compared to the stitching strategy. |
Medina-Sotomayor, 201866 | Trios (3Shape) | Scanning strategies. | The scanning strategy affected the accuracy of the IOS evaluated. |
iTero (Align Technology Inc) | |||
Cerec AC Omnicam (Dentsply Sirona) | |||
True Definition (3M ESPE) | |||
Passos, 2019105 | Omnicam (Dentsply Sirona) | Scanning strategies. | The linear experimental group showed the greatest scanning time for IOS tested. |
Primescan (Dentsply Sirona) | |||
Stefanelli, 202173 | i500 (Medit) | Scanning strategies. | Newer tips appears to be effective for improving the accuracy of IOS. |
Oh, 2020b104 | Trios 3 (3Shape) | Scanning-aid agents. | The liquid type agent gave more accuracy for an IOS. |
Oh, 2021103 | i500 (Medit) | Scanning-aid agents. | The use of a scanning-aid material shortened the scanning time. |
TRIOS (3shape) | |||
Oh, 202289 | i500 (Medit) | Scanning-aid agents. | To obtain more accurate scanning images, the use of liquid agent was proved to be efficient. |
Kim, 201962 | TRIOS 3 (3Shape) | Scanning distance. | A difference was found between the accuracy of the scan distance and the accuracy of the IOS. |
CS 3500 (Carestream) | |||
PlanScan (Planmeca) | |||
Rotar, 202272 | i700 (Medit) | Scanning distance. | Increased scanning distances might lessen the accuracy of a digital impression. |
Kim, 2016101 | CS3500 (Carestream) | Artificial landmarks. | The accuracy of IOS was improved by alumina landmarks. |
Cerec Omnicam (Dentsply Sirona) | |||
TRIOS 3 (3Shape) | |||
Tao, 2020108 | TRIOS 3 (3Shape) | Artificial landmarks. | Resin markers placed on the hard palate of edentulous maxillary model might increase the precision of the IOS. |
Chiu, 202014 | TRIOS 3 (3Shape) | Scan resolution. | A high-resolution mode of the software obtained more data and this could not certainly advantage the scanner accuracy. |
Hayama, 20186 | Model non specified (Carestream) | Head scanner size. | Larger scanning head may improve the accuracy of removable partially edentulous denture. |
Kurz, 201587 | CEREC Omnicam (Dentsply Sirona) | Powder-free system. | The powder-free tested system could be used safely to scan different material surfaces. |
Revilla-Leoón, 202288 | TRIOS 4 (3Shape) | Ambient temperature changes. | Increasing the ambient temperature has a superior effect on the IOS accuracy compared with a decrease ambient temperature. |
Arcuri, 201974 | Trios3 (3Shape) | Scanbody material, position and operator | The use of a PEEK scan body achieved higher outcomes. Operator did not demonstrate influence on the accura |
Baek, 202284 | i500 (Medit) | Superimposing the custom abutment library data. | The accuracy of IOS was improved by means of superimposition of a titanium custom abutment with a prescanned custom abutment collection data. |
Results from the meta-analysis are described in Figures 2, 3, 4, 5, 6. Figure 2 illustrates the impact of humidity on the accuracy of digital intraoral scanners. The analysis reveals that the presence of humidity negatively affected the accuracy of digital IOSs (P < .001).
Fig. 2. Forest plot of the accuracy of intraoral scanners when used under wet or dry conditions.
Fig. 3. Forest plot of the precision (A) and trueness (B) of intraoral scanners when used with artificial landmarks.
Fig. 4. Forest plot of the precision (A) and trueness (B) of intraoral scanners when used under different illumination conditions.
Fig. 5. Forest plot of the precision (A) and trueness (B) of intraoral scanners when used under different scanning patterns.
Fig. 6. Forest plot of the precision (A) and trueness (B) of intraoral scanners when used with scanning-aid materials.
In Figure 3, the meta-analysis results on the precision and trueness of digital IOSs when used in conjunction with artificial landmarks are presented. Both precision (P = .0009) and trueness (P < .001) improved when artificial landmarks were employed.
The impact of ambient light conditions on the precision and trueness of IOSs is shown in Figure 4. The meta-analysis results indicate that these parameters were not significantly affected when the digitalization process was performed under room light or zero-light conditions (P = .27; P = .16).
Figure 5 illustrates the effect of the scanning pattern on the precision and trueness of IOSs. The analysis indicates that differences in measurements were more pronounced when an S-shaped pattern was used (P = .0007). Conversely, discrepancies between a master model and the digital scanner were more substantial when a linear pattern was employed (P = .04).
Finally, the forest plot of the accuracy of IOSs when used in conjunction with scanning-aid materials is presented in Figure 6. The results reveal that only precision is affected by the use of scanning-aid materials (P < .001).
Table 3 provides an analysis of the Risk of Bias (RoB) in the articles included in the qualitative review. Most of the articles were categorized as having a medium RoB. The categories that most manuscripts did not meet included the assessment of a single operator, operator blinding, and sample size calculation.
Table 3. Risk of bias analysis.
Study | Specimens randomization | Single Operator | Control group | Operator blinded | Standardized specimens | Sample size calculation | Risk of bias |
---|---|---|---|---|---|---|---|
Al-ibrahim, 2021 | Χ | √ | √ | Χ | √ | Χ | Medium |
Alenezi, 2022 | Χ | √ | √ | Χ | √ | Χ | Medium |
An, 2022 | Χ | Χ | √ | Χ | √ | Χ | High |
Arakida, 2018 | Χ | Χ | √ | Χ | √ | Χ | High |
Arcuri, 2019 | √ | Χ | √ | √ | √ | Χ | Medium |
Baek, 2022 | Χ | √ | √ | Χ | √ | Χ | Medium |
Çakmak, 2022 | Χ | Χ | √ | Χ | √ | Χ | High |
Chen, 2021 | Χ | √ | √ | Χ | √ | √ | Medium |
Chiu, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Diker, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Diker, 2021 | Χ | √ | √ | Χ | √ | √ | Medium |
Diker, 2021b | Χ | Χ | √ | Χ | √ | √ | Medium |
Donmez, 2021 | Χ | √ | √ | Χ | √ | Χ | Medium |
Ender, 2013 | Χ | Χ | √ | Χ | √ | Χ | High |
Feng, 2021 | Χ | Χ | √ | Χ | √ | √ | Medium |
Gavounelis, 2022 | Χ | Χ | √ | Χ | √ | √ | Medium |
Gómez-Polo, 2022b | √ | √ | √ | Χ | √ | √ | Low |
Haddadi, 2018 | Χ | √ | √ | Χ | √ | Χ | Medium |
Hayama, 2018 | Χ | √ | √ | Χ | √ | Χ | Medium |
Kanjanasavitree, 2022 | √ | Χ | √ | Χ | √ | Χ | Medium |
Kim, 2016 | Χ | Χ | √ | Χ | √ | Χ | High |
Kim, 2019 | √ | √ | √ | Χ | √ | Χ | Medium |
Kim, 2022 | Χ | Χ | √ | Χ | √ | √ | Medium |
Kurz, 2015 | Χ | √ | √ | Χ | √ | Χ | Medium |
Latham, 2019 | √ | √ | √ | Χ | √ | √ | Low |
Li, 2022 | Χ | √ | √ | Χ | √ | Χ | Medium |
Lopes, 2022 | Χ | √ | √ | Χ | √ | √ | Medium |
Yen Mai, 2022 | Χ | √ | √ | √ | √ | Χ | Medium |
Mandelli, 2018 | Χ | Χ | √ | Χ | √ | Χ | High |
Medina-Sotomayor, 2018 | Χ | Χ | √ | Χ | √ | √ | Medium |
Mizumoto, 2019 | √ | √ | √ | Χ | √ | Χ | Medium |
Mizumoto, 2019b | Χ | √ | √ | Χ | √ | Χ | Medium |
Motel, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Müller, 2016 | Χ | Χ | √ | Χ | √ | Χ | High |
Ochoa-López, 2022 | Χ | √ | √ | √ | √ | √ | Low |
Oh, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Oh, 2020b | Χ | Χ | √ | Χ | √ | Χ | High |
Oh, 2021 | Χ | Χ | √ | Χ | √ | Χ | High |
Oh, 2022 | Χ | Χ | √ | Χ | √ | Χ | High |
Passos, 2019 | √ | √ | √ | Χ | √ | Χ | Medium |
Pattamavilai, 2022 | √ | √ | √ | Χ | √ | Χ | Medium |
Peroz, 2021 | Χ | √ | √ | Χ | √ | Χ | Medium |
Rapone, 2020 | Χ | √ | √ | Χ | √ | Χ | Medium |
Revilla-León, 2019 | Χ | Χ | √ | Χ | √ | Χ | High |
Revilla-León, 2020 | Χ | Χ | √ | Χ | √ | √ | Medium |
Revilla-León, 2021 | Χ | Χ | √ | Χ | √ | Χ | High |
Revilla-León, 2022b | Χ | Χ | √ | Χ | √ | Χ | High |
Rotar, 2022 | Χ | √ | √ | Χ | √ | Χ | Medium |
Song, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Stefanelli, 2021 | Χ | √ | √ | Χ | √ | Χ | Medium |
Tao, 2020 | Χ | Χ | √ | Χ | √ | Χ | High |
Wesemann, 2020 | Χ | Χ | √ | Χ | √ | √ | Medium |
You, 2022 | Χ | Χ | √ | Χ | √ | Χ | High |
Zarone, 2020 | √ | Χ | √ | Χ | √ | Χ | Medium |
√=YES and Χ= NO
DISCUSSION
This systematic review and meta-analysis assessed the accuracy of IOSs under various conditions, including humidity, light illuminance, the use of artificial markers, scanning patterns, and scanning-aid materials. The results revealed that certain techniques enhanced the accuracy of IOSs, while others had no significant impact. As a result, the initial hypothesis, which posited that no variation in IOS accuracy would occur when using alternative methods contrary to the manufacturer’s instructions, was only partially accepted.
The accuracy of IOSs decreased when used in humid conditions, such as saliva contamination, leading to increased discrepancies compared to dry conditions. It’s worth noting that while IOSs offer advantages like improved diagnostic efficacy, reduced patient discomfort, simplified clinical procedures, and time savings when taking optical impressions, they are susceptible to factors like saliva secretion, humidity, and intraoral temperature variations, which can affect accuracy.71,106,109,110,111,112 Initially, saliva negatively affects digital impressions by washing out the contrasting powder applied in some IOS. Specifically, saliva can negatively impact digital impressions, causing the scanner to misinterpret the geometry due to the presence of saliva on the tooth’s surface.87 In addition to this, a previous study demonstrated that the measurement of deviations in ‘saliva samples’ is much higher than the clinically adequate cut-off value of 120 microns.113
When comparing zero-light conditions to room light, the accuracy of IOSs in terms of precision and trueness remained unaffected. Determining the ideal lighting condition for accurate scans remains a subject of debate, as different recommendations exist for dental operatory lighting conditions.107 Recommendations regarding the optimal operating light in an office of a dental operatory are varied. The European Standard for Illumination (EN 12464) suggest 500 lux as general lighting, 1000 lux in medical or examination rooms, and 10000 lux inside the mouth.114 Revilla-León et al.107 established that the lighting condition must be selected with reference to the specific IOS system used. Arakida et al.94 studied the effect of illuminance and the color temperature of ambient light on the precision, trueness, and the scanning time of a digital impression and determined that 500 lux with 3900 K was the most suitable lighting condition for digital impressions. Further, another previous study115 deduced that the total absence of an external light delivers optimum outcomes. Overall, each scanner must have its own specifications.
In a clinical environment, an IOS might not deliver accurate scans mainly because of excess afterimages caused by the presence of movable tissues like the tongue or the frenum.108 To solve this issue, the use of resin markers on the palate surface was proposed.116,117 This claim matched the results of this analysis, since the use of artificial markers influenced both the precision and trueness of IOSs. In cases involving fully edentulous arches or implant overdenture frameworks, obtaining accurate digital impressions can be challenging due to mobile tissue.118,119 Additionally, Flügge et al.,120 conveyed that the accuracy of scanned areas decreased when significant distances separated dental implant scanbodies. Furthermore, the presence of stable attached gingiva or palate wasn’t enough to eliminate complications in stitching scans obtained from IOSs due to the absence of clear anatomical landmarks.121 Clinicians recommend the use of artificial markers to overcome these challenges, and this study confirmed that the quality of scan data improved when artificial markers were used.101
For the influence of the scanning pattern on the precision and trueness of IOSs, the discrepancies between different measurements of IOSs were higher when a s-shaped pattern was used. On the other hand, the discrepancies between a master model and the digital scanner were higher when a linear pattern was used. The precision of the digital model relies on the starting point of the scan area as demonstrated previously.122 In contrast, another study stating that the starting position of the area scanned does not influence accuracy and highlighted that the rotational and vertical movements of the IOSs head must be diminished, as an alteration of the direction might disturb an image-stitching procedure. Additionally, a report70 determined that the vertical rotation of the IOS must be prevented. Previous manuscripts.66,105 established numerous IOSs and discovered that the scan strategy influences the accuracy differently, depending on the data capturing method of each scanner. The scan strategy is strictly linked to the image merging software; if the scanner movement has severe changes in orientation or too fast, the stitching procedure might be compromised.5,15 Scanning pattern that include a rotation of the IOS head may be delicate to the examiner. In all clinical scenarios, a careful scanning without a time limit is proposed, and trained operator should follow the recommendation of the manufacturer for better accuracy.
Finally, when the accuracy of IOSs is used in conjunction with scanning-aid materials, the results showed that only the precision is affected. It’s worth noting that most IOSs in the dental market are powder-free, eliminating the need for scanning sprays. However, the clinical performance of these systems is limited to short-distance indications.12,123 Besides, even though it is for a short distance, in some clinical situations, a challenge could occur when the clinician aims to obtain a reliable and a precise data for the required parts in the narrow and deep parts of the prepared teeth and prostheses using metallic materials as there is a light reflection.124 Actually, the IOSs used in dental field are particularly sensitive to the transparency and glossiness of the scanned zone.125,126 To address this limitation, researchers have explored the use of powder coatings to enhance opacity and eliminate reflections. Nevertheless, when the dentists use powder-type agents on teeth surface, making a uniform layer of powder by spraying is difficult, as the quantity of applied powder is importantly affected by the presence of saliva, the skills of operators, and the presence of tongue, adding the space between the dental arches.127 Thus, the powder necessities to be applied as thin as possible, and the spraying time must be as short as possible.107 Liquid-type scanning-aid materials with brush techniques have also been proposed to create thinner and more uniform layers compared to powder sprays, improving accuracy.104 The results from this study propose that scanning-aid materials might be used efficiently to attain full-arch scan data.
Several limitations should be considered in this systematic review and meta-analysis. First, clinical studies were not included, limiting the generalizability of these findings to clinical scenarios. Additionally, the exclusion of some articles from the meta-analysis due to non-standardized scanning procedures affected the overall robustness of the analysis. Standardization in research methods is essential to facilitate meaningful comparisons in future meta-analyses. Lastly, it’s essential to emphasize the importance of continuous training and education for clinicians and lab technicians to maintain a high clinical standard in the rapidly evolving field of digital dentistry.
CONCLUSION
Within the scope of this study, it can be concluded that several techniques, different from those recommended by the manufacturers of intraoral digital scanners, have been proposed. In order to improve the accuracy of these devices, this systematic review recommends taking digital impressions in dry conditions and using artificial landmarks, especially in cases without intraoral landmarks, such as edentulous patients.
References
- 1.Chen LC, Xu ZQ. Innovative 3D dental measurement for tooth model restoration. Key Eng Mater. 2005;295:145–150. [Google Scholar]
- 2.Park HS, Shah C. Development of high speed and high accuracy 3D dental intra oral scanner. Proced Eng. 2015;100:1174–1181. [Google Scholar]
- 3.Ali A. Accuracy of digital impressions achieved from five different digital impression systems. Dentistry. 2015;5:1–6. [Google Scholar]
- 4.Duret F. Toward a new symbolism in the fabrication of prosthetic design. Cah Prothese. 1985;13:65–71. [PubMed] [Google Scholar]
- 5.Richert R, Goujat A, Venet L, Viguie G, Viennot S, Robinson P, Farges JC, Fages M, Ducret M. Intraoral scanner technologies: a review to make a successful impression. J Healthc Eng. 2017;2017:8427595. doi: 10.1155/2017/8427595. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hayama H, Fueki K, Wadachi J, Wakabayashi N. Trueness and precision of digital impressions obtained using an intraoral scanner with different head size in the partially edentulous mandible. J Prosthodont Res. 2018;62:347–352. doi: 10.1016/j.jpor.2018.01.003. [DOI] [PubMed] [Google Scholar]
- 7.ISO 5725-1. Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions. Geneva; Switzerland: International Standards Organization (ISO); 1994. [Google Scholar]
- 8.Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J. 2009;28:44–56. doi: 10.4012/dmj.28.44. [DOI] [PubMed] [Google Scholar]
- 9.Atieh MA, Ritter AV, Ko CC, Duqum I. Accuracy evaluation of intraoral optical impressions: a clinical study using a reference appliance. J Prosthet Dent. 2017;118:400–405. doi: 10.1016/j.prosdent.2016.10.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Gan N, Xiong Y, Jiao T. Accuracy of intraoral digital impressions for whole upper jaws, including full dentitions and palatal soft tissues. PLoS One. 2016;11:e0158800. doi: 10.1371/journal.pone.0158800. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Mehl A. Full arch scans: conventional versus digital impressions-an in vitro study. Int J Comput Dent. 2009;12:11–28. [PubMed] [Google Scholar]
- 12.Ender A, Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent. 2013;109:121–128. doi: 10.1016/S0022-3913(13)60028-1. [DOI] [PubMed] [Google Scholar]
- 13.Kihara H, Hatakeyama W, Komine F, Takafuji K, Takahashi T, Yokota J, Oriso K, Kondo H. Accuracy and practicality of intraoral scanner in dentistry: A literature review. J Prosthodont Res. 2020;64:109–113. doi: 10.1016/j.jpor.2019.07.010. [DOI] [PubMed] [Google Scholar]
- 14.Chiu A, Chen YW, Hayashi J, Sadr A. Accuracy of CAD/CAM digital impressions with different intraoral scanner parameters. Sensors (Basel) 2020;20:1157–1166. doi: 10.3390/s20041157. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Haddadi Y, Bahrami G, Isidor F. Effect of software version on the accuracy of an intraoral scanning device. Int J Prosthodont. 2018;31:375–376. doi: 10.11607/ijp.5781. [DOI] [PubMed] [Google Scholar]
- 16.Revilla-León M, Jiang P, Sadeghpour M, Piedra-Cascón W, Zandinejad A, Özcan M, Krishnamurthy VR. Intraoral digital scans: Part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners. J Prosthet Dent. 2020;124:575–580. doi: 10.1016/j.prosdent.2019.06.004. [DOI] [PubMed] [Google Scholar]
- 17.Carneiro Pereira AL, Souza Curinga MR, Melo Segundo HV, da Fonte Porto Carreiro A. Factors that influence the accuracy of intraoral scanning of total edentulous arches rehabilitated with multiple implants: A systematic review. J Prosthet Dent. 2023;129:855–862. doi: 10.1016/j.prosdent.2021.09.001. [DOI] [PubMed] [Google Scholar]
- 18.Jivanescu A, Faur AB, Rotar RN. Can dental office lighting intensity conditions influence the accuracy of intraoral scanning? Scanning. 2021;2021:9980590. doi: 10.1155/2021/9980590. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Son K, Lee KB. Effect of finish line locations of tooth preparation on the accuracy of intraoral scanners. Int J Comput Dent. 2021;24:29–40. [PubMed] [Google Scholar]
- 20.Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, Shamseer L, Tetzlaff JM, Akl EA, Brennan SE, Chou R, Glanville J, Grimshaw JM, Hróbjartsson A, Lalu MM, Li T, Loder EW, Mayo-Wilson E, McDonald S, McGuinness LA, Stewart LA, Thomas J, Tricco AC, Welch VA, Whiting P, Moher D. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. Int Surg J. 2021;88:105906–105915. doi: 10.1016/j.ijsu.2021.105906. [DOI] [PubMed] [Google Scholar]
- 21.Ouzzani M, Hammady H, Fedorowicz Z, Elmagarmid A. Rayyan-a web and mobile app for systematic reviews. Syst rev. 2016;5:1–10. doi: 10.1186/s13643-016-0384-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Çakmak G, Donmez MB, Akay C, de Paula MS, Mangano FG, Abou-Ayash S, Yilmaz B. Effect of measurement techniques and operators on measured deviations in digital implant scans. J Dent. 2023;130:104388–104394. doi: 10.1016/j.jdent.2022.104388. [DOI] [PubMed] [Google Scholar]
- 23.González de Villaumbrosia P, Martínez-Rus F, García-Orejas A, Salido MP, Pradíes G. In vitro comparison of the accuracy (trueness and precision) of six extraoral dental scanners with different scanning technologies. J Prosthet Dent. 2016;116:543–550.e1. doi: 10.1016/j.prosdent.2016.01.025. [DOI] [PubMed] [Google Scholar]
- 24.Donmez MB, Marques VR, Çakmak G, Yilmaz H, Schimmel M, Yilmaz B. Congruence between the meshes of a combined healing abutment-scan body system acquired with four different intraoral scanners and the corresponding library file: An in vitro analysis. J Dent. 2022;118:103938–103945. doi: 10.1016/j.jdent.2021.103938. [DOI] [PubMed] [Google Scholar]
- 25.Elter B, Diker B, Tak Ö. The trueness of an intraoral scanner in scanning different post space depths. J Dent. 2022;127:104352–104357. doi: 10.1016/j.jdent.2022.104352. [DOI] [PubMed] [Google Scholar]
- 26.Gómez-Polo M, Piedra-Cascón W, Methani MM, Quesada-Olmo N, Farjas-Abadia M, Revilla-León M. Influence of rescanning mesh holes and stitching procedures on the complete-arch scanning accuracy of an intraoral scanner: An in vitro study. J Dent. 2021;110:103690–103698. doi: 10.1016/j.jdent.2021.103690. [DOI] [PubMed] [Google Scholar]
- 27.Hategan SI, Ionel TF, Goguta L, Gavrilovici A, Negrutiu ML, Jivanescu A. Powder and powder-free intra-oral scanners: digital impression accuracy. Prim Dent J. 2018;7:40–43. doi: 10.1177/205016841800700207. [DOI] [PubMed] [Google Scholar]
- 28.Ioshida M, Muñoz BA, Rios H, Cevidanes L, Aristizabal JF, Rey D, Kim-Berman H, Yatabe M, Benavides E, Alvarez MA, Volk S, Ruellas AC. Accuracy and reliability of mandibular digital model registration with use of the mucogingival junction as the reference. Oral Surg Oral Med Oral Pathol Oral Radiol. 2019;127:351–360. doi: 10.1016/j.oooo.2018.10.003. [DOI] [PubMed] [Google Scholar]
- 29.Jeon JH, Choi BY, Kim CM, Kim JH, Kim HY, Kim WC. Three-dimensional evaluation of the repeatability of scanned conventional impressions of prepared teeth generated with white- and blue-light scanners. J Prosthet Dent. 2015;114:549–553. doi: 10.1016/j.prosdent.2015.04.019. [DOI] [PubMed] [Google Scholar]
- 30.Kim YH, Han SS, Choi YJ, Woo CW. Linear accuracy of full-arch digital models using four different scanning methods: an in vitro study using a coordinate measuring machine. Appl Sci. 2020;10:2741–2752. [Google Scholar]
- 31.Lee B, Nam NE, Shin SH, Lim JH, Shim JS, Kim JE. Evaluation of the trueness of digital implant impressions according to the implant scan body orientation and scanning method. Appl Sci. 2021;11:3027–3038. [Google Scholar]
- 32.Marques VR, Çakmak G, Yilmaz H, Abou-Ayash S, Donmez MB, Yilmaz B. Effect of scanned area and operator on the accuracy of dentate arch scans with a single implant. J Clin Med. 2022;11:4125–4137. doi: 10.3390/jcm11144125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Nedelcu RG, Persson AS. Scanning accuracy and precision in 4 intraoral scanners: an in vitro comparison based on 3-dimensional analysis. J Prosthet Dent. 2014;112:1461–1471. doi: 10.1016/j.prosdent.2014.05.027. [DOI] [PubMed] [Google Scholar]
- 34.Park JM. Comparative analysis on reproducibility among 5 intraoral scanners: sectional analysis according to restoration type and preparation outline form. J Adv Prosthodont. 2016;8:354–362. doi: 10.4047/jap.2016.8.5.354. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Park JH, Hwang CJ, Choi YJ, Houschyar KS, Yu JH, Bae SY, Cha JY. Registration of digital dental models and cone-beam computed tomography images using 3-dimensional planning software: Comparison of the accuracy according to scanning methods and software. Am J Orthod Dentofacial Orthop. 2020;157:843–851. doi: 10.1016/j.ajodo.2019.12.013. [DOI] [PubMed] [Google Scholar]
- 36.Passos L, Meiga S, Brigagão V, Neumann M, Street A. Digital impressions’ accuracy through “cut-out-rescan” and “data exchange by over scanning” techniques in complete arches of two intraoral scanners and CAD/CAM software. J Prosthodont Res. 2022;66:509–513. doi: 10.2186/jpr.JPR_D_20_00089. [DOI] [PubMed] [Google Scholar]
- 37.Pedroche LO, Bernardes SR, Leão MP, Kintopp CC, Correr GM, Ornaghi BP, Gonzaga CC. Marginal and internal fit of zirconia copings obtained using different digital scanning methods. Braz Oral Res. 2016;30:e113–e120. doi: 10.1590/1807-3107BOR-2016.vol30.0113. [DOI] [PubMed] [Google Scholar]
- 38.Pereira ALC, Segundo HVM, Júnior LCA, Germano AR, Carreiro ADFP. Accuracy between intraoral and extraoral scanning: Three-dimensional deviation and effect of distance between implants from two scanning methods. J Indian Prosthodont Soc. 2022;22:279–287. doi: 10.4103/jips.jips_108_22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Pozzi A, Arcuri L, Lio F, Papa A, Nardi A, Londono J. Accuracy of complete-arch digital implant impression with or without scanbody splinting: an in vitro study. J Dent. 2022;119:104072–104079. doi: 10.1016/j.jdent.2022.104072. [DOI] [PubMed] [Google Scholar]
- 40.Ren X, Son K, Lee KB. Accuracy of proximal and occlusal contacts of single implant crowns fabricated using different digital scan methods: an in vitro study. Materials (Basel) 2021;14:2843. doi: 10.3390/ma14112843. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Revilla-León M, Zeitler JM, Barmak AB, Kois JC. Accuracy of the 3-dimensional virtual patient representation obtained by using 4 different techniques: An in vitro study. J Prosthet Dent. 2022;1(22):S0022–3913(22)00342. doi: 10.1016/j.prosdent.2022.05.016. [DOI] [PubMed] [Google Scholar]
- 42.Virard F, Millet C, Bienfait A, Ducret M. Improving digital scans by using a peripheral custom tray: a dental technique. J Prosthet Dent. 2021;126:173–177. doi: 10.1016/j.prosdent.2020.05.022. [DOI] [PubMed] [Google Scholar]
- 43.Ye J, Wang S, Wang Z, Liu Y, Sun Y, Ye H, Zhou Y. Comparison of the dimensional and morphological accuracy of three-dimensional digital dental casts digitized using different methods. Odontology. 2023;111:165–171. doi: 10.1007/s10266-022-00736-2. [DOI] [PubMed] [Google Scholar]
- 44.Zhang YJ, Qiao SC, Qian SJ, Zhang CN, Shi JY, Lai HC. Influence of different factors on the accuracy of digital impressions of multiple implants: an in vitro study. Int J Oral Maxillofac Implants. 2021;36:442–449. doi: 10.11607/jomi.8532. [DOI] [PubMed] [Google Scholar]
- 45.Camcı H, Salmanpour F. Effect of saliva isolation and intraoral light levels on performance of intraoral scanners. Am J Orthod Dentofacial Orthop. 2020;158:759–766. doi: 10.1016/j.ajodo.2020.03.022. [DOI] [PubMed] [Google Scholar]
- 46.Gajbhiye V, Banerjee R, Jaiswal P, Chandak A, Radke U. Comparative evaluation of three gingival displacement materials for efficacy in tissue management and dimensional accuracy. J Indian Prosthodont Soc. 2019;19:173–179. doi: 10.4103/jips.jips_285_18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Imburgia M, Kois J, Marino E, Lerner H, Mangano FG. Continuous scan strategy (CSS): a novel technique to improve the accuracy of intraoral digital impressions. Eur J Prosthodont Restor Dent. 2020;28:128–141. doi: 10.1922/EJPRD_2105Imburgia14. [DOI] [PubMed] [Google Scholar]
- 48.Koseoglu M, Kahramanoglu E, Akin H. Evaluating the effect of ambient and scanning lights on the trueness of the intraoral scanner. J Prosthodont. 2021;30:811–816. doi: 10.1111/jopr.13341. [DOI] [PubMed] [Google Scholar]
- 49.Lim SW, Hwang HS, Cho IS, Baek SH, Cho JH. Registration accuracy between intraoral-scanned and cone-beam computed tomography-scanned crowns in various registration methods. Am J Orthod Dentofacial Orthop. 2020;157:348–356. doi: 10.1016/j.ajodo.2019.04.031. [DOI] [PubMed] [Google Scholar]
- 50.Moon YG, Lee KM. Comparison of the accuracy of intraoral scans between complete-arch scan and quadrant scan. Prog Orthod. 2020;21:36. doi: 10.1186/s40510-020-00337-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Nedelcu R, Olsson P, Thulin M, Nyström I, Thor A. In vivo trueness and precision of full-arch implant scans using intraoral scanners with three different acquisition protocols. J Dent. 2023;128:104308–104319. doi: 10.1016/j.jdent.2022.104308. [DOI] [PubMed] [Google Scholar]
- 52.Althubaitiy R, Sambrook R, Weisbloom M, Petridis H. The accuracy of digital implant impressions when using and varying the material and diameter of the dental implant scan bodies. Eur J Prosthodont Restor Dent. 2022;30:305–313. doi: 10.1922/EJPRD_2367Althubaitiy09. [DOI] [PubMed] [Google Scholar]
- 53.Denneulin T, Rignon-Bret C, Ravalec G, Tapie L, Bouter D, Wulfman C. Accuracy of complete-arch implant digital scans: effect of scanning protocol, number of implants, and scan body splinting. Int J Prosthodont. 2023;36:219–227. doi: 10.11607/ijp.7332. [DOI] [PubMed] [Google Scholar]
- 54.Fang Y, Fang JH, Jeong SM, Choi BH. A Technique for digital impression and bite registration for a single edentulous arch. J Prosthodont. 2019;28:e519–e523. doi: 10.1111/jopr.12786. [DOI] [PubMed] [Google Scholar]
- 55.Osman RB, Alharbi NM. Influence of scan technology on the accuracy and speed of intraoral scanning systems for the edentulous maxilla: An in vitro study. J Prosthodont. 2022 Dec 26; doi: 10.1111/jopr.13633. Epub ahead of print. [DOI] [PubMed] [Google Scholar]
- 56.Reich S, Yatmaz B, Raith S. Do “cut out-rescan” procedures have an impact on the accuracy of intraoral digital scans? J Prosthet Dent. 2021;125:89–94. doi: 10.1016/j.prosdent.2019.11.018. [DOI] [PubMed] [Google Scholar]
- 57.Garikano X, Amezua X, Iturrate M, Solaberrieta E. Evaluation of repeatability of different alignment methods to obtain digital interocclusal records: An in vitro study. J Prosthet Dent. 2022:S0022-3913(22)00498-X. doi: 10.1016/j.prosdent.2022.07.014. Epub ahead of print. [DOI] [PubMed] [Google Scholar]
- 58.Bratos M, Bergin JM, Rubenstein JE, Sorensen JA. Effect of simulated intraoral variables on the accuracy of a photogrammetric imaging technique for complete-arch implant prostheses. J Prosthet Dent. 2018;120:232–241. doi: 10.1016/j.prosdent.2017.11.002. [DOI] [PubMed] [Google Scholar]
- 59.Diker B, Tak Ö. Accuracy of six intraoral scanners for scanning complete-arch and 4-unit fixed partial dentures: an in vitro study. J Prosthet Dent. 2022;128:187–194. doi: 10.1016/j.prosdent.2020.12.007. [DOI] [PubMed] [Google Scholar]
- 60.Diker B, Tak Ö. Comparing the accuracy of six intraoral scanners on prepared teeth and effect of scanning sequence. J Adv Prosthodont. 2020;12:299–306. doi: 10.4047/jap.2020.12.5.299. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Diker B, Tak Ö. Accuracy of digital impressions obtained using six intraoral scanners in partially edentulous dentitions and the effect of scanning sequence. Int J Prosthodont. 2021;34:101–108. doi: 10.11607/ijp.6834. [DOI] [PubMed] [Google Scholar]
- 62.Kim MK, Kim JM, Lee YM, Lim YJ, Lee SP. The effect of scanning distance on the accuracy of intra-oral scanners used in dentistry. Clin Anat. 2019;32:430–438. doi: 10.1002/ca.23334. [DOI] [PubMed] [Google Scholar]
- 63.Kim EY, Wada J, Sakamoto K, Ishioka Y, Arai Y, Murakami N, Yamazaki T, Hayama H, Utsumi M, Inukai S, Wakabayashi N. Effect of scanning origin location on data accuracy of abutment teeth region in digital impression acquired using intraoral scanner for removable partial denture: a preliminary in vitro study. J Clin Med. 2022;11:7392. doi: 10.3390/jcm11247392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Lopes D, Nishyama R, Steagall W, Jr, Tamaki R, Neto PT. Impact of different scan strategies and implant angulation on impression accuracy of full arch multiple implant: an in vitro study. Braz Dent Sci. 2022;25:e3006 [Google Scholar]
- 65.Mandelli F, Gherlone EF, Keeling A, Gastaldi G, Ferrari M. Full-arch intraoral scanning: comparison of two different strategies and their accuracy outcomes. J Osseointegr. 2018;10:65–74. [Google Scholar]
- 66.Medina-Sotomayor P, Pascual MA, Camps AI. Accuracy of four digital scanners according to scanning strategy in complete-arch impressions. PloS one. 2018;13:e0202916. doi: 10.1371/journal.pone.0202916. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Mizumoto RM, Alp G, Özcan M, Yilmaz B. The effect of scanning the palate and scan body position on the accuracy of complete-arch implant scans. Clin Implant Dent Relat Res. 2019;21:987–994. doi: 10.1111/cid.12821. [DOI] [PubMed] [Google Scholar]
- 68.Müller P, Ender A, Joda T, Katsoulis J. Impact of digital intraoral scan strategies on the impression accuracy using the TRIOS Pod scanner. Quintessence Int. 2016;47:343–349. doi: 10.3290/j.qi.a35524. [DOI] [PubMed] [Google Scholar]
- 69.Ochoa-López G, Cascos R, Antonaya-Martín JL, Revilla-León M, Gómez-Polo M. Influence of ambient light conditions on the accuracy and scanning time of seven intraoral scanners in complete-arch implant scans. J Dent. 2022;121:104138–104146. doi: 10.1016/j.jdent.2022.104138. [DOI] [PubMed] [Google Scholar]
- 70.Oh KC, Park JM, Moon HS. Effects of scanning strategy and scanner type on the accuracy of intraoral scans: A new approach for assessing the accuracy of scanned data. J Prosthodont. 2020;29:518–523. doi: 10.1111/jopr.13158. [DOI] [PubMed] [Google Scholar]
- 71.Pattamavilai S, Ongthiemsak C. Accuracy of intraoral scanners in different complete arch scan patterns. J Prosthet Dent. 2022:S0022-3913(22)00046-4. doi: 10.1016/j.prosdent.2021.12.026. Epub ahead of print. [DOI] [PubMed] [Google Scholar]
- 72.Rotar RN, Faur AB, Pop D, Jivanescu A. Scanning distance influence on the intraoral scanning accuracy-an in vitro study. Materials (Basel) 2022;15:3061–3068. doi: 10.3390/ma15093061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Stefanelli LV, Franchina A, Pranno A, Pellegrino G, Ferri A, Pranno N, Di Carlo S, De Angelis F. Use of intraoral scanners for full dental arches: could different strategies or overlapping software affect accuracy? Int J Environ Res Public Health. 2021;18:9946–9959. doi: 10.3390/ijerph18199946. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Arcuri L, Pozzi A, Lio F, Rompen E, Zechner W, Nardi A. Influence of implant scanbody material, position and operator on the accuracy of digital impression for complete-arch: A randomized in vitro trial. J Prosthodont Res. 2020;64:128–136. doi: 10.1016/j.jpor.2019.06.001. [DOI] [PubMed] [Google Scholar]
- 75.Ender A, Mehl A. Influence of scanning strategies on the accuracy of digital intraoral scanning systems. Int J Comput Dent. 2013;16:11–21. [PubMed] [Google Scholar]
- 76.Latham J, Ludlow M, Mennito A, Kelly A, Evans Z, Renne W. Effect of scan pattern on complete-arch scans with 4 digital scanners. J Prosthet Dent. 2020;123:85–95. doi: 10.1016/j.prosdent.2019.02.008. [DOI] [PubMed] [Google Scholar]
- 77.Mizumoto RM, Yilmaz B, McGlumphy EA, Jr, Seidt J, Johnston WM. Accuracy of different digital scanning techniques and scan bodies for complete-arch implant-supported prostheses. J Prosthet Dent. 2020;123:96–104. doi: 10.1016/j.prosdent.2019.01.003. [DOI] [PubMed] [Google Scholar]
- 78.Motel C, Kirchner E, Adler W, Wichmann M, Matta RE. Impact of different scan bodies and scan strategies on the accuracy of digital implant impressions assessed with an intraoral scanner: an in vitro study. J Prosthodont. 2020;29:309–314. doi: 10.1111/jopr.13131. [DOI] [PubMed] [Google Scholar]
- 79.Peroz S, Spies BC, Adali U, Beuer F, Wesemann C. Measured accuracy of intraoral scanners is highly dependent on methodical factors. J Prosthodont Res. 2022;66:318–325. doi: 10.2186/jpr.JPR_D_21_00023. [DOI] [PubMed] [Google Scholar]
- 80.Revilla-León M, Subramanian SG, Att W, Krishnamurthy VR. Analysis of different illuminance of the room lighting condition on the accuracy (trueness and precision) of an intraoral scanner. J Prosthodont. 2021;30:157–162. doi: 10.1111/jopr.13276. [DOI] [PubMed] [Google Scholar]
- 81.Wesemann C, Kienbaum H, Thun M, Spies BC, Beuer F, Bumann A. Does ambient light affect the accuracy and scanning time of intraoral scans? J Prosthet Dent. 2021;125:924–931. doi: 10.1016/j.prosdent.2020.03.021. [DOI] [PubMed] [Google Scholar]
- 82.Zarone F, Ruggiero G, Ferrari M, Mangano F, Joda T, Sorrentino R. Comparison of different intraoral scanning techniques on the completely edentulous maxilla: an in vitro 3-dimensional comparative analysis. J Prosthet Dent. 2020;124:762.e1–762.e8. doi: 10.1016/j.prosdent.2020.07.017. [DOI] [PubMed] [Google Scholar]
- 83.Al-Ibrahim I, Keeling A, Osnes C. The effect of operator scanning speed on the trueness and precision of full-arch digital impressions captured in vitro using an intraoral scanner. J Osseoint. 2021;13(4 Supplement):S265–S270. [Google Scholar]
- 84.Baek YW, Lim YJ, Kim MJ, Kwon HB. Effect of custom abutment data superimposition on the accuracy of implant abutment level scanning: an in vitro study. J Prosthet Dent. 2022:S0022-3913(22)00176-7. doi: 10.1016/j.prosdent.2022.03.008. [DOI] [PubMed] [Google Scholar]
- 85.Cakmak G, Marques VR, Donmez MB, Lu WE, Abou-Ayash S, Yilmaz B. Comparison of measured deviations in digital implant scans depending on software and operator. J Dent. 2022;122:104154. doi: 10.1016/j.jdent.2022.104154. [DOI] [PubMed] [Google Scholar]
- 86.Feng CW, Hung CC, Wang JC, Lan TH. Accuracy of different head movements of intraoral scanner in full arch of both maxilla and mandible. Appl Sci. 2021;11:8140–8152. [Google Scholar]
- 87.Kurz M, Attin T, Mehl A. Influence of material surface on the scanning error of a powder-free 3D measuring system. Clin Oral Investig. 2015;19:2035–2043. doi: 10.1007/s00784-015-1440-5. [DOI] [PubMed] [Google Scholar]
- 88.Revilla-León M, Gohil A, Barmak AB, Gómez-Polo M, Pérez-Barquero JA, Att W, Kois JC. Influence of ambient temperature changes on intraoral scanning accuracy. J Prosthet Dent. 2022:S0022-3913(22)00061-0. doi: 10.1016/j.prosdent.2022.01.012. [DOI] [PubMed] [Google Scholar]
- 89.Oh HS, Lim YJ, Kim B, Kim MJ, Kwon HB, Baek YW. Effect of scanning-aid agents on the scanning accuracy in specially designed metallic models: a laboratory study. PLoS One. 2022;17:e0267742–e0267751. doi: 10.1371/journal.pone.0267742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Song J, Kim M. Accuracy on scanned images of full arch models with orthodontic brackets by various intraoral scanners in the presence of artificial saliva. Biomed Res Int. 2020;2020:2920804. doi: 10.1155/2020/2920804. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.You R, Han F, Sun Z, Xie H, Chen C. Evaluation of intraoral optical scanning accuracy for fine structure reconstruction and occlusal records of inlay/onlay preparations. Int J Comput Dent. 2022;25:257–265. doi: 10.3290/j.ijcd.b2599691. [DOI] [PubMed] [Google Scholar]
- 92.Alenezi A, Yehya M, Alkhodary M. Effect of full arch two scanning techniques on the accuracy of overdenture conventional and CAD/CAM Co-Cr bars. Saudi Dent J. 2022;34:553–564. doi: 10.1016/j.sdentj.2022.08.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Mai HY, Mai HN, Lee CH, Lee KB, Kim SY, Lee JM, Lee KW, Lee DH. Impact of scanning strategy on the accuracy of complete-arch intraoral scans: a preliminary study on segmental scans and merge methods. J Adv Prosthodont. 2022;14:88–95. doi: 10.4047/jap.2022.14.2.88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Arakida T, Kanazawa M, Iwaki M, Suzuki T, Minakuchi S. Evaluating the influence of ambient light on scanning trueness, precision, and time of intra oral scanner. J Prosthodont Res. 2018;62:324–329. doi: 10.1016/j.jpor.2017.12.005. [DOI] [PubMed] [Google Scholar]
- 95.An H, Langas EE, Gill AS. Effect of scanning speed, scanning pattern, and tip size on the accuracy of intraoral digital scans. J Prosthet Dent. 2022:S0022-3913(22)00326-2. doi: 10.1016/j.prosdent.2022.05.005. [DOI] [PubMed] [Google Scholar]
- 96.Chen Y, Zhai Z, Li H, Yamada S, Matsuoka T, Ono S, Nakano T. Influence of liquid on the tooth surface on the accuracy of intraoral scanners: an in vitro study. J Prosthodont. 2022;31:59–64. doi: 10.1111/jopr.13358. [DOI] [PubMed] [Google Scholar]
- 97.Donmez MB, Çakmak G, Atalay S, Yilmaz H, Yilmaz B. Trueness and precision of combined healing abutment-scan body system depending on the scan pattern and implant location: an in-vitro study. J Dent. 2022;124:104169–104179. doi: 10.1016/j.jdent.2022.104169. [DOI] [PubMed] [Google Scholar]
- 98.Gavounelis NA, Gogola CC, Halazonetis DJ. The effect of scanning strategy on intraoral scanner’s accuracy. Dent J (Basel) 2022;10:123–133. doi: 10.3390/dj10070123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Gómez-Polo M, Ortega R, Sallorenzo A, Agustín-Panadero R, Barmak AB, Kois JC, Revilla-León M. Influence of the surface humidity, implant angulation, and interimplant distance on the accuracy and scanning time of complete-arch implant scans. J Dent. 2022;127:104307–104316. doi: 10.1016/j.jdent.2022.104307. [DOI] [PubMed] [Google Scholar]
- 100.Kanjanasavitree P, Thammajaruk P, Guazzato M. Comparison of different artificial landmarks and scanning patterns on the complete-arch implant intraoral digital scans. J Dent. 2022;125:104266–104273. doi: 10.1016/j.jdent.2022.104266. [DOI] [PubMed] [Google Scholar]
- 101.Kim JE, Amelya A, Shin Y, Shim JS. Accuracy of intraoral digital impressions using an artificial landmark. J Prosthet Dent. 2017;117:755–761. doi: 10.1016/j.prosdent.2016.09.016. [DOI] [PubMed] [Google Scholar]
- 102.Li Z, Huang R, Wu X, Chen Z, Huang B, Chen Z. Effect of scan pattern on the accuracy of complete-arch digital implant impressions with two intraoral scanners. Int J Oral Maxillofac Implants. 2022;37:731–739. doi: 10.11607/jomi.9248. [DOI] [PubMed] [Google Scholar]
- 103.Oh HS, Lim YJ, Kim B, Kim MJ, Kwon HB, Baek YW. Influence of scanning-aid materials on the accuracy and time efficiency of intraoral scanners for full-arch digital scanning: an in vitro study. Materials (Basel) 2021;14:2340–2351. doi: 10.3390/ma14092340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Oh HS, Lim YJ, Kim B, Kim WH, Kim MJ, Kwon HB. Influence of applied liquid-type scanning-aid material on the accuracy of the scanned image: an in vitro experiment. Materials (Basel) 2020;13:2034–2045. doi: 10.3390/ma13092034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Passos L, Meiga S, Brigagão V, Street A. Impact of different scanning strategies on the accuracy of two current intraoral scanning systems in complete-arch impressions: an in vitro study. Int J Comput Dent. 2019;22:307–319. [PubMed] [Google Scholar]
- 106.Rapone B, Palmisano C, Ferrara E, Di Venere D, Albanese G, Corsalini M. The accuracy of three intraoral scanners in the oral environment with and without saliva: a comparative study. Appl Sci. 2020;10:7762–7781. [Google Scholar]
- 107.Revilla-León M, Jiang P, Sadeghpour M, Piedra-Cascón W, Zandinejad A, Özcan M, Krishnamurthy VR. Intraoral digital scans-Part 1: Influence of ambient scanning light conditions on the accuracy (trueness and precision) of different intraoral scanners. J Prosthet Dent. 2020;124:372–378. doi: 10.1016/j.prosdent.2019.06.003. [DOI] [PubMed] [Google Scholar]
- 108.Tao C, Zhao YJ, Sun YC, Heng MD, Xie QF, Pan SX. Accuracy of intraoral scanning of edentulous jaws with and without resin markers. Chin J Dent Res. 2020;23:265–271. doi: 10.3290/j.cjdr.b867887. [DOI] [PubMed] [Google Scholar]
- 109.Lee KM. Comparison of two intraoral scanners based on three-dimensional surface analysis. Prog Orthod. 2018;19:1–7. doi: 10.1186/s40510-018-0205-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Grassi FR, Rapone B, Scarano Catanzaro F, Corsalini M, Kalemaj Z. Effectiveness of computer-assisted anesthetic delivery system (sta™) in dental implant surgery: a prospective study. Oral Implantol (Rome) 2017;10:381–389. doi: 10.11138/orl/2017.10.4.381. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Tsirogiannis P, Reissmann DR, Heydecke G. Evaluation of the marginal fit of single-unit, complete-coverage ceramic restorations fabricated after digital and conventional impressions: A systematic review and meta-analysis. J Prosthet Dent. 2016;116:328–335.e2. doi: 10.1016/j.prosdent.2016.01.028. [DOI] [PubMed] [Google Scholar]
- 112.Ahlholm P, Sipilä K, Vallittu P, Jakonen M, Kotiranta U. Digital versus conventional impressions in fixed prosthodontics: a review. J Prosthodont. 2018;27:35–41. doi: 10.1111/jopr.12527. [DOI] [PubMed] [Google Scholar]
- 113.Ferrari M, Keeling A, Mandelli F, Lo Giudice G, Garcia-Godoy F, Joda T. The ability of marginal detection using different intraoral scanning systems: A pilot randomized controlled trial. Am J Dent. 2018;31:272–276. [PubMed] [Google Scholar]
- 114.European Lightening Standard. EN 12464-1. Light and lighting - Lighting of work places - Part 1: Indoor work places. . 2011. [Google Scholar]
- 115.Blanco D, Fernández P, Cuesta E, Suárez CM. Influence of ambient light on the quality of laser digitized surfaces; Proceedings of the world congress on engineering, vol. 1; 2008. pp. 32–37. [Google Scholar]
- 116.Fang JH, An X, Jeong SM, Choi BH. Digital intraoral scanning technique for edentulous jaws. J Prosthet Dent. 2018;119:733–735. doi: 10.1016/j.prosdent.2017.05.008. [DOI] [PubMed] [Google Scholar]
- 117.Chebib N, Kalberer N, Srinivasan M, Maniewicz S, Perneger T, Müller F. Edentulous jaw impression techniques: An in vivo comparison of trueness. J Prosthet Dent. 2019;121:623–630. doi: 10.1016/j.prosdent.2018.08.016. [DOI] [PubMed] [Google Scholar]
- 118.Patzelt SB, Vonau S, Stampf S, Att W. Assessing the feasibility and accuracy of digitizing edentulous jaws. J Am Dent Assoc. 2013;144:914–920. doi: 10.14219/jada.archive.2013.0209. [DOI] [PubMed] [Google Scholar]
- 119.Andriessen FS, Rijkens DR, van der Meer WJ, Wismeijer DW. Applicability and accuracy of an intraoral scanner for scanning multiple implants in edentulous mandibles: a pilot study. J Prosthet Dent. 2014;111:186–194. doi: 10.1016/j.prosdent.2013.07.010. [DOI] [PubMed] [Google Scholar]
- 120.Flügge TV, Schlager S, Nelson K, Nahles S, Metzger MC. Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofacial Orthop. 2013;144:471–478. doi: 10.1016/j.ajodo.2013.04.017. [DOI] [PubMed] [Google Scholar]
- 121.Deferm JT, Schreurs R, Baan F, Bruggink R, Merkx MAW, Xi T, Bergé SJ, Maal TJJ. Validation of 3D documentation of palatal soft tissue shape, color, and irregularity with intraoral scanning. Clin Oral Investig. 2018;22:1303–1309. doi: 10.1007/s00784-017-2198-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Anh JW, Park JM, Chun YS, Kim M, Kim M. A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: effects of tooth irregularity and scanning direction. Korean J Orthod. 2016;46:3–12. doi: 10.4041/kjod.2016.46.1.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Khraishi H, Duane B. Evidence for use of intraoral scanners under clinical conditions for obtaining full-arch digital impressions is insufficient. Evid Based Dent. 2017;18:24–25. doi: 10.1038/sj.ebd.6401224. [DOI] [PubMed] [Google Scholar]
- 124.Huang MY, Son K, Lee KB. Effect of distance between the abutment and the adjacent teeth on intraoral scanning: an in vitro study. J Prosthet Dent. 2021;125:911–917. doi: 10.1016/j.prosdent.2020.02.034. [DOI] [PubMed] [Google Scholar]
- 125.Burde AV, Dudea D, Cuc S, Moldovan M, Campian RS. Three-dimensional evaluations of the coating thickness of two optical conditioning scanning sprays. Mater Plast. 2016;53:65–67. [Google Scholar]
- 126.Quaas S, Rudolph H, Luthardt RG. Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. J Dent. 2007;35:903–908. doi: 10.1016/j.jdent.2007.08.008. [DOI] [PubMed] [Google Scholar]
- 127.Dehurtevent M, Robberecht L, Béhin P. Influence of dentist experience with scan spray systems used in direct CAD/CAM impressions. J Prosthet Dent. 2015;113:17–21. doi: 10.1016/j.prosdent.2014.07.006. [DOI] [PubMed] [Google Scholar]