Abstract
Background
Point-of-care tests (POCT) for haemoglobin are increasingly used to guide intraoperative transfusion. However, their accuracy compared to central laboratory tests is unknown. The objective was to perform a systematic review and meta-analysis of method comparison studies assessing the accuracy of POCT versus central laboratory haemoglobin tests in patients undergoing surgery.
Methods
Electronic databases were searched from inception to April 2020 (updated August 2023). Any methodological approach comparing haemoglobin measurements between POCT and central laboratory in patients undergoing surgery under anaesthesia in the operating room were included. Data abstraction was guided by PRISMA and risk of bias was assessed by QUADAS-2. Data were extracted independently and in duplicate by two reviewers. Outcomes included mean differences between POCT and central laboratory haemoglobin with associated standard deviations and 95% limits of agreement (LOA).
Results
Of 3057 citations, 34 studies were included (n = 2427, 6857 paired measurements). Several devices were compared (pulse co-oximetry, n = 25; HemoCue, n = 10; iSTAT, n = 6; blood gas analysers, n = 10; haematology analyser, n = 2). Median sample size was 41 patients, and 11 studies were funded by device manufacturers. Fifteen of 34 studies had low risk of bias. Pooled mean differences (95% LOA) were: pulse co-oximeters 2.3 g/l (−25.2–29.8), HemoCue −0.3 g/l (−11.1–10.5), iSTAT −0.3 g/l (−8.4–7.8) and blood gas analysers −2.6 g/l (−17.8–12.7).
Conclusion
All POCT examining intraoperative haemoglobin measurement yielded pooled mean difference LOAs larger than the allowable limit difference of ±4 g/dl. Intraoperative haemoglobin measured by POCT should not be considered interchangeable with central laboratory values and caution is necessary when using these tests to guide intraoperative transfusion.
Transfusions during surgery are commonly guided by point-of-care devices measuring haemoglobin values but their accuracy is unknown. This systematic review and meta-analysis of 32 studies shows that all point-of-care tests measuring intraoperative haemoglobin yielded pooled mean differences and limits of agreements beyond the allowable limit difference. Thus, haemoglobin measured by point-of-care devices during surgery should not be considered interchangeable with central laboratory values and caution is necessary when using these devices to guide intraoperative transfusions.
Introduction
Red blood cell (RBC) transfusions are common in surgery and may account for 27–44% of all transfused RBC units in the hospital1. Transfusions can be life-saving, but also carry risks such as allergic and transfusion reactions, transfusion-associated acute lung injury and transfusion-associated circulatory overload, and have been associated with transfusion-related immunomodulation, which can potentially lead to worse perioperative and long-term oncologic outcomes in surgical patients2–4. Lastly, they are an expensive and limited resource, estimated to cost up to €696 per unit5.
It is well established that haemoglobin measurement plays a central role in any decision to transfuse RBCs. A recent systematic review of clinical practice guidelines providing transfusion recommendations revealed that of 10 guidelines, eight recommended transfusing based on haemoglobin values6–9. A 2016 Cochrane systematic review of studies guiding transfusions identified 31 trials that involved haemoglobin measurements as a trigger for transfusion10. Lastly, a recent survey of Canadian anaesthesiologists reported intraoperative haemoglobin levels to be the most important parameter for transfusion decision-making—more important than blood loss or haemodynamics11.
Haemoglobin can be assessed by several methods. Considered to be the gold standard and part of the complete blood count, the haemoglobincyanide (HiCN) method uses an internationally accepted reference calibrator and provides a measured haemoglobin concentration12. However, this process is time-consuming and increasingly less useful to guide intraoperative transfusion in the context of acute bleeding. More recently, point-of-care testing devices for haemoglobin (POCT-Hb) have evolved and become the current standard of care during surgery. These devices are relatively simple to use and yield results within seconds to minutes, leading to greater clinical use in the operating room.
There are several classes of POCT-Hb. The first type chemically converts haemoglobin found in the blood sample to azide-haemoglobin, which is then measured by absorption photometry. This technology can yield a haemoglobin value from 10 µl of whole blood in 10–60 s13. A commonly used device in this category is HemoCue (HemoCue AB, Angelholm, Sweden). A second method provides a calculated haemoglobin value based on the conductometric method. Using 65–100 µl of whole blood, it calculates haemoglobin in 120 s by multiplying the haematocrit (hct) by a proportionality constant (Hb (g/l) = hct (%) × 3.4)14. A commonly used device in this category is iSTAT (Abbott Laboratories, Abbott Park, IL, USA). A third common method is pulse oximetry. This technology uses 12 or more wavelengths of light to measure total haemoglobin, and allows for non-invasive continuous monitoring15. Devices in this category include Masimo Radical-7 (Masimo Corp., Irvine, CA, USA).
POCT-Hb devices have been validated with static haemoglobin values, such as in healthy blood donors16–18, or in non-operative settings, such as the intensive care unit or emergency room19–21. The use of POCT-Hb in the intraoperative setting, where haemoglobin levels can change quickly due to bleeding and rapid fluid shifts from concurrent intravenous fluid administration22, is relatively untested. There are also few studies assessing POCT-Hb devices within the critical zone of potential transfusion of 60–100 g/l, highlighting a major criticism of existing evaluations23–25. The relationship between POCT-Hb and transfusion decisions has been emphasized by multiple other authors23,26–30.
Despite the lack of evidence validating their use in the operative setting, POCT-Hb devices have become ubiquitous to guide intraoperative transfusion decisions. The aim of this study is to perform a systematic review and meta-analysis of method comparison studies assessing the accuracy of POCT-Hb compared to central laboratory testing in patients undergoing surgery.
Methods
Information sources
This study was registered with PROSPERO (CRD42021233103). Reporting of this review was guided by the PRISMA statement (Table S1).
A systematic search was designed by an information specialist (R.S.). The search included EMBASE (1947 to August 2023), Ovid MEDLINE (1946 to August 2023) and EBM Reviews—Cochrane Central Register of Controlled Trials (August 2023).
Search strategy
The full electronic search strategy was peer reviewed and conducted following the Peer Review of Electronic Search Strategies (PRESS) guidelines31. The search was not limited by language or patient population. Grey literature was included in the form of conference abstracts. References of included articles were reviewed manually for other relevant studies. Finally, the list of included references was circulated to a small group of experts in anaesthesiology and transfusion medicine to identify any additional missing studies. The search strategy is reported in the Supplementary Methods.
Eligibility criteria
Study participants were required to be patients undergoing any surgery under general or neuraxial anaesthesia in an operating room. Studies that included data from other clinical settings were also included if data provided for the intraoperative period were reported separately. Studies of interest were those that compared haemoglobin values provided by POCT and the reference central laboratory standard. The gold standard was defined as the HiCN test method with blood samples collected in ethylenediaminetetraacetic acid (EDTA) vacuum collection tubes and processed by a haematology analyser in a central laboratory. POCT devices included any non-invasive measurements via pulse co-oximetry, via occlusion spectroscopy and via transcutaneous reflection. This review also included invasive measurements via absorption photometry, calculations via conductivity or via blood gas analysers. Sampling between the POCT device and central laboratory must have been taken concurrently or analysed from the same sample.
Studies that did not account for within-individual correlation between successive measurements (that is repeated haemoglobin measurements taken from the same patient but analysed as separate data points) were eligible for inclusion, but were accounted for in the analysis. Given differences in transfusion practice, neonatal populations were excluded. Studies that assessed the accuracy of POCT-Hb versus central lab haemoglobin in settings outside the operating room were excluded, such as endoscopy or minor bedside procedures. Finally, studies in which only the haemoglobin mean difference was provided, without the standard deviation (s.d.) or 95% limits of agreement (LOA), were excluded if those values could not be provided by the authors or derived from other data points or graphs.
Effect measures
The primary outcome was the bias or mean difference (MD) between POCT-Hb minus central lab haemoglobin measurements and its s.d. If not provided in the text, 95% LOA intervals were also calculated using the following formula: 95% LOA = mean difference ± 1.96 × s.d.
Data collection process
Articles identified through the search strategy were imported into Covidence (Covidence, Melbourne, Australia), an online citation manager for systematic reviews32. Title, abstract and full-text screening were performed independently and in duplicate by two reviewers (H.A., T.L.). Authors were contacted for any uncertainties regarding eligibility. At both stages of review, any discrepancies were documented, discussed and adjudicated by the senior author (G.M.). Google Translate was used to translate non-English or non-French articles. Reasons for exclusion were documented and reported in the PRISMA flow diagram (Fig. 1).
Data items
A standard data extraction form was created using Covidence, which was then exported into a Microsoft Excel spreadsheet (Microsoft Corporation, Redmond, WA, USA). Information gathered comprised study characteristics such as year of publication, location, funding source, sample size and study design and patient characteristics including age, sex and weight. Operative characteristics including type of surgery, anaesthesia, duration, blood loss, transfusions, fluid balance and operative interventions were also recorded. Finally, POCT device, central lab analyser, number of paired measurements, timing of samples and sampling site (arterial, venous or capillary) were also documented. Data extraction was performed independently by two reviewers (H.A., T.L.) and authors were contacted for any uncertainties.
Statistical analysis
The MD, 95% LOA and s.d. were extracted from each study. Confidence bands of LOA were estimated using a random effects model and robust variance estimation (RVE), as previously described by Williamson et al.33. For studies that did not account for within-individual correlation for repeated measurements, the s.d. was adjusted to provide an RVE. Confidence bands of LOA are presented by the lower 95% limit for the lower value of the LOA and the upper 95% limit for the upper value of the LOA. The pooled estimate of MD, 95% LOA and confidence bands of LOA were estimated using the approach described by Tipton and Shuster34. The correlation coefficient (r) and kappa coefficient (κ) were also extracted. Heterogeneity was determined by the Chi square test (significance level 0.05) combined with the I2 statistic. Possible sources of heterogeneity were investigated with predefined subgroup analyses including funding source, blood loss, haemoglobin range and sampling site. A sensitivity analysis was performed based on risk of bias. Meta-analysis of each outcome was performed using R (version 4.0.2) in R Studio (version 2022.07.2 build 576), using the base and stats packages35.
Study risk of bias assessment
The quality of included studies was assessed according to the revised Quality Assessment of Diagnostic Studies (QUADAS-2) guidelines36. This tool consists of four domains including patient selection, index test, reference standard, and flow and timing. The risk of bias is assessed for each domain in addition to applicability for the first three domains. The questionnaire was adopted from Kim et al.37 and tailored to this review, reported in the Supplementary Methods. If answers to all signalling questions for a domain were ‘yes’, then risk of bias for that domain was judged low. If two or more domains were deemed ‘high’ risk of bias, then the overall assessment for the study was judged ‘high’ risk of bias. Quality assessment was performed by three reviewers (H.A., T.L., R.G.) independently and any disagreements were discussed and adjudicated by the senior author (G.M.).
Results
Extent of evidence identified
Results from the search strategy are shown in Fig. 1. A total of 3000 de-duplicated citations were identified for title and abstract review. A total of 155 citations were eligible for full-text review, of which 34 studies were included in the systematic review.
Characteristics of included studies
Study characteristics are reported in Table S2. Among the included studies, 25 compared pulse co-oximetry devices to central lab (n = 1110 patients, 4059 paired measurements)26–28,38–58, nine compared HemoCue (n = 525 patients, 1962 paired measurements)26,27,39,44,57,59–63, six compared iSTAT (n = 146 patients, 294 paired measurements)19,28,40,47,59,64 and 10 compared blood gas analysers (n = 821 patients, 3381 paired measurements)26,28,39,41,49,50,57,60,63,65. Two studies compared different POCT haematology analysers (n = 295 patients, 553 paired measurements)66,67 and were not meta-analysed. There were 27 full-text papers and seven conference abstracts. Median publication date was 2014 (range 1991–2023). Median sample size was 41 (range 6–348). Eleven studies were funded by device manufacturers, one was funded by the U.S. military, three were university-funded, 10 were not funded and 10 did not report funding sources. Participants in 25 studies received a general anaesthetic, one under spinal, and eight were not reported. Eight studies included cardiac surgery, seven neurosurgery, two transplant, two orthopaedic surgery, two gynaecological surgery, two urology, one vascular and 12 were not specified. Nine studies quantified intraoperative blood loss (mean range 100–3400 ml) and three reported blood transfusions (mean range 0–750 ml). Three studies reported mean difference of POCT-Hb devices within the critical transfusion zone of 60–100 g/l.
All included studies reported haemoglobin MD, 28 reported 95% LOA and 17 reported s.d. of the MD. Six studies accounted for within-individual correlation between successive measurements26,27,40,52,53,65.
Risk of bias assessment
The methodological quality of included studies is shown in Table 1, based on the revised QUADAS-2 guidelines. In total, 15 of 34 studies were assessed to have low risk of bias. Patient selection and flow and timing were assessed to have the highest risk of bias across all studies, while index test and reference standard had the lowest risk. Only six studies reported a sampling method (that is, random, consecutive or convenience sample) and only four papers reported on their types of study design (that is, superiority, equivalence, inferiority).
Table 1.
Meta-analysis
Table 2 shows the pooled analysis of each class of POCT device and its subgroup analyses. If the included studies had insufficient data to perform subgroup analyses, they were not reported.
Table 2.
Analysis group | Number of studies | Number of patients | Bias (g/l) | Standard deviation of bias | Between-study heterogeneity in bias (I2) | 95% LOA (g/l) |
---|---|---|---|---|---|---|
Pulse Co-Ox | 25 | 1110 | 2.3 | 1.15 | 0.57 | (−25.2,29.8) |
Pulse Co-Ox (funded) | 8 | 423 | 2.1 | 1.17 | 0.98 | (−31.0,35.1) |
Pulse Co-Ox (unfunded) | 17 | 687 | 2.2 | 1.14 | 0.18 | (−22.1,26.6) |
Pulse Co-Ox (blood loss > 1 litre) | 5 | 305 | −2.5 | 0.90 | 0.80 | (−27.9,22.9) |
HemoCue | 10 | 525 | −0.3 | 0.48 | 0.06 | (−11.1,10.5) |
HemoCue (arterial) | 5 | 314 | 0.1 | 0.41 | 0.07 | (−9.6,9.87) |
HemoCue (capillary) | 5 | 228 | 1.4 | 0.93 | 0.12 | (−18.5,21.2) |
HemoCue (venous) | 2 | 90 | −1.2 | 0.51 | 0.19 | (−14.5,12.1) |
iSTAT | 6 | 146 | −0.3 | 0.32 | 0.06 | (−8.4,7.8) |
Blood gas analyser | 10 | 821 | −2.6 | 0.51 | 0.32 | (−17.8,12.7) |
Low risk of bias (Pulse Co-Ox) | 13 | 675 | 1.9 | 1.20 | 0.97 | (−29.2,32.9) |
POCT, point-of-care tests; LOA, limits of agreement.
Pulse co-oximeter
Meta-analysis of 25 studies assessing pulse co-oximeters revealed an MD (95% LOA) of 2.3 (−25.2 to 29.8) g/l and high heterogeneity (I2 = 57%; Fig. 2). When subgroup analyses were performed for industry funding (n = 8), non-industry funding (n = 15) or blood loss greater than 1 litre (n = 5), the MDs (95% LOA) were 2.1 (−31.0 to 35.1) g/l, 2.6 (−25.0 to 30.1) g/l, and −2.5 (−27.9 to 22.9) g/l, respectively. Two studies reported MDs within the critical transfusion zone of 60–100 g/l; thus, a subgroup was not performed. Sensitivity analysis of pulse co-oximetry studies at low risk of bias (n = 12) revealed an MD (95% LOA) of 1.9 (−29.2 to 32.9) g/l (Fig. S1).
HemoCue
Meta-analysis of 10 studies assessing HemoCue revealed an MD (95% LOA) of −0.3 (−11.1 to 10.5) g/l and low heterogeneity (I2 = 6%; Fig. 3). Subgroup analyses of arterial, capillary and venous samples demonstrated MDs (95% LOA) of 0.1 (−9.6 to 9.8) g/l, 1.4 (−18.5 to 21.2) g/l and −1.2 (−14.5 to 12.1) g/l, respectively.
iSTAT
Meta-analysis of six studies assessing iSTAT revealed an MD (95% LOA) of −0.3 (−8.4 to 7.8) g/l and low heterogeneity (I2 = 6%; Fig. 4).
Blood gas analysers
Meta-analysis of 10 studies assessing blood gas analysers revealed an MD (95% LOA) of −2.6 (−17.8 to 12.7) g/l and low heterogeneity (I2 = 32%; Fig. S2).
Discussion
In this systematic review, the accuracy of point-of-care haemoglobin devices compared to central laboratory measurements in patients undergoing surgery was assessed. Most studies compared pulse co-oximetry to central laboratory measurements, with fewer studies examining HemoCue, iSTAT and blood gas analysers. Of the 34 included studies, none examined occlusion spectroscopy or transcutaneous reflection spectroscopy. Almost a third of studies were industry-funded and only a minority quantified blood loss. Only three studies compared devices within the critical transfusion zone of 60–100 g/l. Overall, the accuracies of the reported devices are inadequate to guide intraoperative transfusion decision-making. The sample size and number of available studies for each POCT device were low and further high-quality prospective accuracy studies are warranted.
To the authors' knowledge, this is the first systematic review and meta-analysis that investigates method comparison studies of POCT devices versus central laboratory haemoglobin measurements specifically in the operative setting. Given that haemoglobin values play a central role in transfusion decision-making in surgery, it is imperative to evaluate the accuracy of POCT devices as they are increasingly used in lieu of formal central laboratory assays. The current analyses demonstrate that the pooled bias (g/l) of pulse co-oximetry, HemoCue, iSTAT and blood gas analysers was 2.3, −0.3, −0.3 and −2.6, respectively. The bias indicates the difference of the mean error above and below the reference measurement but does not report the magnitude of the error in each direction. Thus, clinically, the bias alone would be insufficient to compare methods. Rather, the 95% LOAs provide the interval within which 95% of the differences between haemoglobin measurements by the two methods are expected to lie68,69. In this context, the 95% LOA is more clinically relevant to assess agreement between methods. The pooled 95% LOA (g/l) for pulse co-oximeters, HemoCue, iSTAT and blood gas analysers were −25.2 to 29.8, −11.1 to 10.5, −8.4 to 7.8 and −17.8 to 12.7, respectively. These intervals are much larger than the allowable difference of ±4 g/l defined by the Institute of Quality Management in Healthcare (IQMH)70. Alternatively, when using an allowable difference of ±10 g/l, as previously argued by Morey and colleagues24, the 95% LOA of iSTAT falls within this range; however, only six studies (n = 146 patients) were included in this meta-analysis, of which four had high risks of bias. As such, haemoglobin values measured by POCT devices should not be considered interchangeable with central laboratory values and abundant caution is needed when using these devices to guide transfusion decisions in the operating room.
Other systematic reviews and meta analyses of the accuracy of POCT devices measuring haemoglobin have been reported21,37,71,72, but all have significant differences compared to this review. Shabaninejad and colleagues71 included 28 studies comparing Radical-7 pulse co-oximetry to central laboratory measurements in the operative setting and demonstrated a bias (95% LOA) of 2.7 (−4.4 to −1.0) g/l. However, their meta-analysis included other POCT devices such as iSTAT and blood gas analysers as reference measurements. The accuracy of these devices has not been validated for use in surgery and therefore these were not considered an acceptable reference comparator in the current review. Further, Shabaninejad et al.’s review was limited to Radical-7 pulse co-oximetry, and sensitivity analyses based on study quality and risk of bias were not presented. In 2020, Zortea and colleagues72 included eight studies comparing haemoglobin values measured by non-invasive techniques versus central laboratory. Their sensitivity analysis of surgical patients in four studies revealed a mean overall difference of 0.02 (95% c.i. −0.43 to 0.47). However, this group included patients outside of the operative setting and used POCT devices such as blood gas analysers as reference standards. In 2015, Hiscock and colleagues21 published a meta-analysis of 39 studies comparing pulse co-oximetry and HemoCue to central laboratory haemoglobin measurements and reported a bias (95% LOA) of −0.3 (−30 to 29 g/l) and 0.8 (−13 to 14 g/l), respectively. Their results demonstrate that Masimo pulse co-oximetry devices have lower precision and wider 95% LOA compared to HemoCue, which is consistent with the current analysis; however, their review included primarily non-operative data. Further, Kim and colleagues37 analysed 32 studies comparing non-invasive haemoglobin measurements (Masimo, OrSense) to central laboratory testing. More specifically, a subgroup analysis of 13 studies conducted in the perioperative setting demonstrated a bias (95% LOA) of 3.9 (−22.1 to 29.8 g/l). Again, this subgroup included patients outside of the operative setting. Lastly, none of the other reviews addressed within-individual correlation between successive measurements. In the current paper, six studies reported repeated measures. For the remaining 26 studies, the standard deviation was adjusted to provide a robust variance estimation.
The current review has several limitations. Results should be interpreted carefully as 18 of 34 studies were assessed to have high overall risk of bias. It is also noteworthy that 11 studies were funded by device manufacturers. The accuracy of the POCT device may be overestimated as industry-funded studies that show a larger difference between the POCT device and the central laboratory may be less likely to be published. In addition, a high level of heterogeneity was identified in the pulse co-oximetry group, likely owing in part to different operation types, blood loss, and intraoperative interventions such as cardiopulmonary bypass and acute normovolaemic haemodilution, which may affect haemoglobin measurements. Only three studies reported blood transfusions41,46,55. In the study of De Rosa, transfusion was performed if haemoglobin was <80 g/l in healthy patients or <90 g/l in patients with cardiac disease or active bleeding. In Gupta et al., blood transfusion was at the discretion of the anaesthesiologist and parameters were not specified. No patients received allogeneic blood transfusions in Saito et al.
Intraoperative decision-making for RBC transfusions is complex and is not based on a robust evidence base. A 2021 systematic review of clinical practice guidelines for intraoperative RBC transfusions identified 10 guidelines73. However, recommendations were highly variable and data were extrapolated from non-operative settings. Further, none provided recommendations on the most appropriate method for haemoglobin measurement, although all implied that transfusions should be guided at least in part by haemoglobin/haematocrit triggers, thresholds and/or targets. No guideline discussed the role of point-of-care haemoglobin testing.
This review suggests that POCT haemoglobin devices are insufficiently accurate to be used interchangeably with central laboratory haemoglobin testing in the operating room. This finding is particularly important as it pertains to transfusion decision-making, and inaccurate haemoglobin measurements could lead to over- or under-transfusion, both of which can lead to significant patient harm. Further prospective accuracy data are required to compare the accuracy of POCT haemoglobin devices in the operative setting, principally within the critical transfusion zone of 60–100 g/l, as well as to determine their ability to appropriately guide transfusion.
Supplementary Material
Contributor Information
Hilalion (San) Ahn, Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada.
Tori Lenet, Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada; Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
Richard W D Gilbert, Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada.
Ranjeeta Mallick, Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
Julie L V Shaw, Department of Pathology and Laboratory Medicine, University of Ottawa, Ottawa, ON, Canada.
Dean A Fergusson, Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
Daniel I McIsaac, Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada; Department of Anesthesiology & Pain Medicine, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada.
Guillaume Martel, Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada; Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
Funding
No specific funding was received for this project.
Disclosure
The authors declare no conflicts of interest.
Supplementary material
Supplementary material is available at BJS Open online.
Data availability
The authors declare that the data supporting the findings of this study are available within the article and its supplementary files.
References
- 1. Shehata N, Forster A, Lawrence N, Rothwell DM, Fergusson D, Tinmouth A et al. Changing trends in blood transfusion: an analysis of 244,013 hospitalizations. Transfusion 2014;54:2631–2639 [DOI] [PubMed] [Google Scholar]
- 2. Madjdpour C, Spahn DR. Allogeneic red blood cell transfusions: efficacy, risks, alternatives and indications. Br J Anaesth 2005;95:33–42 [DOI] [PubMed] [Google Scholar]
- 3. Vamvakas EC, Blajchman MA. Transfusion-related immunomodulation (TRIM): an update. Blood Rev 2007;21:327–348 [DOI] [PubMed] [Google Scholar]
- 4. Whitson BA, Huddleston SJ, Savik K, Shumway SJ. Risk of adverse outcomes associated with blood transfusion after cardiac surgery depends on the amount of transfusion. J Surg Res 2010;158:20–27 [DOI] [PubMed] [Google Scholar]
- 5. Shander A, Hofmann A, Gombotz H, Theusinger OM, Spahn DR. Estimating the cost of blood: past, present, and future directions. Best Pract Res Clin Anaesthesiol 2007;21:271–289 [DOI] [PubMed] [Google Scholar]
- 6. Baker L, Park L, Gilbert R, Ahn H, Martel A, Davies A et al. Guidelines on the intraoperative transfusion of red blood cells: a systematic review. BMJ Open 2019;9:e029684. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Turgeon AF, Fergusson DA, Doucette S, Khanna MP, Tinmouth A, Aziz A et al. Red blood cell transfusion practices amongst Canadian anesthesiologists: a survey. Can J Anaesth 2006;53:344–352 [DOI] [PubMed] [Google Scholar]
- 8. Brunskill SJ, Millette SL, Shokoohi A, Pulford EC, Doree C, Murphy MF et al. Red blood cell transfusion for people undergoing hip fracture surgery. Cochrane Database Syst Rev 2015;4:CD009699. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Wilkinson KL, Brunskill SJ, Doree C, Trivella M, Gill R, Murphy MF. Red cell transfusion management for patients undergoing cardiac surgery for congenital heart disease. Cochrane Database Syst Rev 2014;2:CD009752. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Carson JL, Stanworth SJ, Roubinian N, Fergusson DA, Triulzi D, Doree C et al. Transfusion thresholds and other strategies for guiding allogeneic red blood cell transfusion. Cochrane Database Syst Rev 2016;10:CD002042. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Bennett S, Ayoub A, Tran A, English S, Tinmouth A, McIsaac DI et al. Current practices in perioperative blood management for patients undergoing liver resection: a survey of surgeons and anesthesiologists. Transfusion 2018;58:781–787 [DOI] [PubMed] [Google Scholar]
- 12. Shah VB, Shah BS, Puranik GV. Evaluation of non cyanide methods for hemoglobin estimation. Indian J Pathol Microbiol 2011;54:764–768 [DOI] [PubMed] [Google Scholar]
- 13. Sanchis-Gomar F, Cortell-Ballester J, Pareja-Galeano H, Banfi G, Lippi G. Hemoglobin point-of-care testing: the HemoCue system. J Lab Autom 2013;18:198–205 [DOI] [PubMed] [Google Scholar]
- 14. Papadea C, Foster J, Grant S, Ballard SA, Cate JC, Southgate WM et al. Evaluation of the i-STAT portable clinical analyzer for point-of-care blood testing in the intensive care units of a university children’s hospital. Ann Clin Lab Sci 2002;32:231–243 [PubMed] [Google Scholar]
- 15. Cannesson M, Talke P. Recent advances in pulse oximetry. F1000 Med Rep 2009;1:66. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Baart AM, de Kort WLAM, van den Hurk K, Pasker-de Jong PCM. Hemoglobin assessment: precision and practicability evaluated in the Netherlands—the HAPPEN study. Transfusion 2016;56:1984–1993 [DOI] [PubMed] [Google Scholar]
- 17. Bahadur S, Jain S, Jain M. Estimation of hemoglobin in blood donors: a comparative study using HemoCue and cell counter. Transfus Apher Sci 2010;43:155–157 [DOI] [PubMed] [Google Scholar]
- 18. Radtke H, Polat G, Kalus U, Salama A, Kiesewetter H. Hemoglobin screening in prospective blood donors: comparison of different blood samples and different quantitative methods. Transfus Apher Sci 2005;33:31–35 [DOI] [PubMed] [Google Scholar]
- 19. Steinfelder-Visscher J, Teerenstra S, Gunnewiek JMTK, Weerwind PW. Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients. J Extra Corpor Technol 2008;40:57–60 [PMC free article] [PubMed] [Google Scholar]
- 20. Dolscheid-Pommerich RC, Dolscheid S, Grigutsch D, Stoffel-Wagner B, Graeff I. Comparability of point-of-care versus central laboratory hemoglobin determination in emergency patients at a supra-maximal care hospital. PLoS One 2016;11:e0166521. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Hiscock R, Simmons S, Carstensen B, Gurrin L. Comparison of Massimo Pronto-7 and HemoCue Hb 201+ with laboratory haemoglobin estimation: a clinical study. Anaesth Intensive Care 2014;42:608–613 [DOI] [PubMed] [Google Scholar]
- 22. Wu P, Morey TE, Harris NS, Gravenstein N, Rice MJ. Intravenous fluids cause systemic bias in a conductivity-based point-of-care hematocrit meter. Anesth Analg 2012;114:314–321 [DOI] [PubMed] [Google Scholar]
- 23. Johnson M, Marwick PC, Coetzee JF. Comparison of hemoglobin measurements by 3 point-of-care devices with standard laboratory values and reliability regarding decisions for blood transfusion. Anesth Analg 2019;131:640–649 [DOI] [PubMed] [Google Scholar]
- 24. Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg 2011;113:89–91 [DOI] [PubMed] [Google Scholar]
- 25. Rice MJ, Gravenstein N, Morey TE. Noninvasive hemoglobin monitoring: how accurate is enough? Anesth Analg 2013;117:902–907 [DOI] [PubMed] [Google Scholar]
- 26. Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth 2013;111:946–954 [DOI] [PubMed] [Google Scholar]
- 27. Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology 2011;115:548–554 [DOI] [PubMed] [Google Scholar]
- 28. Maslow A, Bert A, Singh A, Sweeney J. Point-of-care hemoglobin/hematocrit testing: comparison of methodology and technology. J Cardiothorac Vasc Anesth 2016;30:352–362 [DOI] [PubMed] [Google Scholar]
- 29. Morey TE, Gravenstein N, Rice MJ. Assessing point-of-care hemoglobin measurement: be careful we don’t bias with bias. Anesth Analg 2011;113:1289–1291 [DOI] [PubMed] [Google Scholar]
- 30. Broderick AJ. Point-of-care haemoglobin measurement—state of the art or a bleeding nuisance? Anaesthesia 2015;70:1225–1229 [DOI] [PubMed] [Google Scholar]
- 31. McGowan J, Sampson M, Salzwedel DM, Cogo E, Foerster V, Lefebvre C. PRESS Peer Review of Electronic Search Strategies: 2015 guideline statement. J Clin Epidemiol 2016;75:40–46 [DOI] [PubMed] [Google Scholar]
- 32. Covidence Systematic Review Software. Melbourne, Australia: Veritas Health Innovation; Available from: www.covidence.org [Google Scholar]
- 33. Williamson PR, Lancaster GA, Craig JV, Smyth RL. Meta-analysis of method comparison studies. Stat Med 2002;21:2013–2025 [DOI] [PubMed] [Google Scholar]
- 34. Tipton E, Shuster J. A framework for the meta-analysis of Bland–Altman studies based on a limits of agreement approach. Stat Med 2017;36:3621–3635 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. R Core Team . R: A Language and Environment for Statistical Computing. Vienna, Austria: R Foundation for Statistical Computing; Available from: https://www.R-project.org [Google Scholar]
- 36. Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 2011;155:529–536 [DOI] [PubMed] [Google Scholar]
- 37. Kim SH, Lilot M, Murphy LSL, Sidhu KS, Yu Z, Rinehart J et al. Accuracy of continuous noninvasive hemoglobin monitoring: a systematic review and meta-analysis. Anesth Analg 2014;119:332–346 [DOI] [PubMed] [Google Scholar]
- 38. Adel A, Awada W, Abdelhamid B, Omar H, Abd El Dayem O, Hasanin A et al. Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses. J Clin Monit Comput 2018;32:1025–1031 [DOI] [PubMed] [Google Scholar]
- 39. Applegate II RL, Applegate PM, Cannesson M, Peiris P, Ladlie BL, Torp K. Multicenter comparison of three intraoperative hemoglobin trend monitoring methods. J Clin Monit Comput 2020;34:883–892 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb station. Am J Surg 2011;201:592–598 [DOI] [PubMed] [Google Scholar]
- 41. De Rosa RC. Accuracy of continuous monitoring of haemoglobin concentration after in-vivo adjustment. Angiol Vasc Surg 2020;5:1–6 [Google Scholar]
- 42. Desebbe O, Faulcon C, Henaine R, Tran L, Koffel C, Delannoy B et al. Tissue hemoglobin monitoring is unable to follow variations of arterial hemoglobin during transitions from pulsatile to constant flow in cardiac surgery. J Cardiothorac Vasc Anesth 2014;28:668–673 [DOI] [PubMed] [Google Scholar]
- 43. Kayhan GE, Colak YZ, Sanli M, Ucar M, Toprak HI. Accuracy of non-invasive hemoglobin monitoring by pulse CO-oximeter during liver transplantation. Minerva Anestesiol 2017;83:485–492 [DOI] [PubMed] [Google Scholar]
- 44. Frasca D, Mounios H, Giraud B, Boisson M, Debaene B, Mimoz O. Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss. Anaesthesia 2015;70:803–809 [DOI] [PubMed] [Google Scholar]
- 45. Gill H, Navaratnarajah M, Tallach E, Fernandez E, Sury M. Assessment of a noninvasive hemoglobin monitor (Masimo SpHb) in infants and small children undergoing craniofacial surgery. Pediatr Anesth 2012;22:916–917 [Google Scholar]
- 46. Gupta N, Kulkarni A, Bhargava A, Prakash A, Gupta N. Utility of non-invasive haemoglobin monitoring in oncosurgery patients. Indian J Anaesth 2017;61:543. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Jou F, Kurth C, Beckman E, Istaphanous G. Absolute and trend accuracy of continuous and noninvasive hemoglobin in pediatric surgery patients. Anesth Analg 2010;110:S1–S52021551425 [Google Scholar]
- 48. Kamel MM, Hasanin A, Nawar B, Mostafa M, Jacob VF, Elhadi H et al. Evaluation of noninvasive hemoglobin monitoring in children with congenital heart diseases. Pediatr Anesth 2020;30:571–576 [DOI] [PubMed] [Google Scholar]
- 49. Kapoor I, Mahajan C, Mamuliya R, Prabhakar H. Assessment of accuracy of continuous noninvasive versus invasive method of hemoglobin estimation in patients undergoing pituitary surgery. J Neuroanaesth Crit Care 2018;5:168–172 [Google Scholar]
- 50. Khanna P, Rajagopalan V, Singh G, Prabhakar H. A comparison of non-invasive versus invasive methods of haemoglobin estimation in patients undergoing intracranial surgery. South Afr J Anaesth Analg 2014;20:160–163 [Google Scholar]
- 51. Marques NR, Kramer GC, Voigt RB, Salter MG, Kinsky MP. Trending, accuracy, and precision of noninvasive hemoglobin monitoring during human hemorrhage and fixed crystalloid bolus. Shock 2015;44:45–49 [DOI] [PubMed] [Google Scholar]
- 52. Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg 2011;112:858–863 [DOI] [PubMed] [Google Scholar]
- 53. Patino M, Schultz L, Hossain M, Moeller J, Mahmoud M, Gunter J et al. Trending and accuracy of noninvasive hemoglobin monitoring in pediatric perioperative patients. Anesth Analg 2014;119:920–925 [DOI] [PubMed] [Google Scholar]
- 54. Perez A, Bisbe E, Alvarez J, Lopez R, Fau M, Escolano F. Preliminary clinical evaluation of a continuous and noninvasive hemoglobin monitor (Masimo pulse CO-oximeter) in surgical patients. Transfus Altern Transfus Med 2009;11(s1):1–40 [Google Scholar]
- 55. Saito J, Kitayama M, Oishi M, Kudo T, Sawada M, Hashimoto H et al. The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood. J Anesth 2015;29:29–34 [DOI] [PubMed] [Google Scholar]
- 56. Zeng R, Liu H, Li H, Xu X, Shangguan W, Li Y et al. Abstract PR282: accuracy of noninvasive and continuous hemoglobin measurements in Chinese children. Anesth Analg 2016;123:364–365 [Google Scholar]
- 57. Divatia J, Pradhan S, Verma P. Haemoglobin estimation by non-invasive pulse co-oximetry or arterial blood co-oximetry are inferior to arterial hemocuetm estimation during intra-operative haemorrhage. Anaesth Intensive Care 2020;48:30 [Google Scholar]
- 58. Gurskaia V, Pulkina O, Ivanov V, Savvina I. Intraoperative evaluation of non-invasive hemoglobin in children with craniosynostosis. Anesth Analg 2021;133:1222 [Google Scholar]
- 59. McNulty SE, Torjman M, Grodecki W, Marr A, Schieren H. A comparison of four bedside methods of hemoglobin assessment during cardiac surgery. Anesth Analg 1995;81:1197–1202 [DOI] [PubMed] [Google Scholar]
- 60. Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol 2007;39:10–17 [PMC free article] [PubMed] [Google Scholar]
- 61. Richards NA, Boyce H, Yentis SM. Estimation of blood haemoglobin concentration using the HemoCue® during caesarean section: the effect of sampling site. Int J Obstet Anesth 2010;19:67–70 [DOI] [PubMed] [Google Scholar]
- 62. Sfez M, Desruelle P, Pennecot GF, Cohen-Bacrie P. [The importance of the preoperative measurement of hemoglobin concentration. Evaluation of a colorimetric method (Hemocue)]. Cah Anesthesiol 1991;39:247–251 [PubMed] [Google Scholar]
- 63. Spielmann N, Mauch J, Madjdpour C, Schmugge M, Weiss M, Haas T. Accuracy and precision of hemoglobin point-of-care testing during major pediatric surgery. Int J Lab Hematol 2012;34:86–90 [DOI] [PubMed] [Google Scholar]
- 64. Ng WL, Short TG, Gunn KN, Fuge GS, Slon B. Accuracy and reliability of the i-STAT point-of-care device for the determination of haemoglobin concentration before and after major blood loss. Anaesth Intensive Care 2014;42:495–499 [DOI] [PubMed] [Google Scholar]
- 65. Carabini LM, Navarre WJ, Ault ML, Bebawy JF, Gupta DK. A comparison of hemoglobin measured by co-oximetry and central laboratory during major spine fusion surgery. Anesth Analg 2015;120:60–65 [DOI] [PubMed] [Google Scholar]
- 66. Despotis GJ, Saleem R, Bigham M, Barnes P. Clinical evaluation of a new, point-of-care hemocytometer. Crit Care Med 2000;28:1185–1190 [DOI] [PubMed] [Google Scholar]
- 67. Jonsson H, Larsson A, Semenas E, Soderberg E, Lipcsey M. Point of care analysis of hematology in the operating theater—a prospective observational study of accuracy and feasibility. Clin Lab 2023;69:2. [DOI] [PubMed] [Google Scholar]
- 68. Altman DG, Bland JM. Measurement in medicine: the analysis of method comparison studies. Statistician 1983;32:307 [Google Scholar]
- 69. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135–160 [DOI] [PubMed] [Google Scholar]
- 70. Johnston A, Bourner G, Martin T, McFarlane A, Good D, Padmore R et al. Guidance for quality control practices and precision goals for CBCs based on IQMH patterns-of-practice survey. Int J Lab Hematol 2019;41:15–22 [DOI] [PubMed] [Google Scholar]
- 71. Shabaninejad H, Ghadimi N, Sayehmiri K, Hosseinifard H, Azarfarin R, Gorji HA. Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis. J Anesth 2019;33:441–453 [DOI] [PubMed] [Google Scholar]
- 72. Zortea T, da Silva Wizbicki DP, Madeira K, Ambrosio PG, de Souza ROB, Duraes ESM. [Noninvasive hemoglobin monitoring in clinical trials: a systematic review and meta-analysis]. Braz J Anesthesiol 2020;70:388–397 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Baker L, Park L, Gilbert R, Ahn H, Martel A, Lenet T et al. Intraoperative red blood cell transfusion decision-making: a systematic review of guidelines. Ann Surg 2021;274:86–96 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The authors declare that the data supporting the findings of this study are available within the article and its supplementary files.