Abstract
Type 1 diabetes mellitus (T1D) is an autoimmune-related disease resulting in insulin dependency, treated with insulin injection via pen devices or continuous subcutaneous insulin infusion (CSII). Face-to-face instruction for managing insulin injection and dosing and machine-to-device troubleshooting are required early to initiate CSII from insulin injections. Thus, T1D individuals may encounter significant barriers to pen devices or CSII introduction if they live in remote rural areas. In this regard, intermittently scanned continuous glucose monitoring (isCGM) can share visualized glucose profiles via a cloud-platform-based system, offering the potential as an effective tool in telemedicine. Herewith, we report two cases of subjects with T1D living in remote rural areas whose CSII was safely introduced in outpatient settings with the aid of cloud-platform-based isCGM and a video-meeting tool. They showed improved glucose profiles after CSII initiation. Even under the coronavirus disease 2019 (COVID-19) pandemic, the telemedicine system enabled healthcare providers to monitor glucose profiles and confirm device procedures of CSII. We emphasize the usefulness of online instruction with cloud-platform-based isCGM for introducing CSII in cases with barriers to healthcare access, particularly during the COVID-19 pandemic.
Keywords: covid-19, telemedicine, intermittently scanned continuous glucose monitoring, insulin pump, type 1 diabetes
Introduction
Type 1 diabetes mellitus (T1D) is a disease mediated by autoimmune triggers destroying pancreatic beta cells with exogenous insulin dependency, the cause of which is still unclear. Generally, there are two methods of exogenous insulin supplementation: multiple daily injections (MDI) using pen devices and continuous subcutaneous insulin infusion (CSII) [1] using insulin pumps. In MDI, basal insulin for modulating fasting blood glucose and short-acting insulin for modulating postprandial blood glucose were administered via pen devices several times a day. However, insulin-dose adjustments have limited flexibility in MDI because once injected, basal insulin cannot be adjusted for diurnal glucose changes, such as due to the dawn phenomenon [2] and during or after exercise. Otherwise, with CSII, basal insulin can be set as "units per hour" anytime daily. Thus, based on the guidelines, CSII is recommended [2] as the first therapeutic option for persons with T1D, and unstable glucose levels could be improved with CSII with precise and personalized dosing. An automated insulin delivery system (AID) [3] also enables automatically modulated insulin doses in response to glucose levels, though it is more expensive than CSII.
Otherwise, the insulin pump introduction has several limitations, including healthcare access for frequent outpatient visits, mainly in early-stage for accustoming mechanical troubles and economic problems in the countries where CSII is self-funded [4]. Rural people in Japan have many difficulties with healthcare access because Japan is a mountain- or island-rich country, and approximately 10% of the population lives in hard-to-move areas [5]. Therefore, they could not attend to outpatients frequently and felt demotivated to introduce CSII by their circumstances. Thus, in Japan, the initiation of CSII has generally been performed on an inpatient basis for sufficient education and training in emergencies, unlike abroad [6].
Intermittently scanned continuous glucose monitoring (isCGM) is a new technology for glucose management [7], displaying trends in glucose profiles. Increased or decreased glucose levels are indicated via time-above range (TAR; glucose levels >180 mg/dL), time-in-range (TIR; glucose levels 70-180 mg/dL), and time-below-range (TBR; glucose levels <70 mg/dL). Although hemoglobin A1c (HbA1c) is still the gold standard for assessing glucose control, these parameters have recently attracted attention as novel indicators, leading to meaningful benefits in the quality of life and personalized glycemic management [7]. In addition, the recent-implemented cloud-platform-based system Libre View® (Abbott Diabetes Care Inc., CA, USA) enables persons with diabetes, their families, and healthcare professionals to access glucose trends online for remote diabetes management even under the coronavirus disease 2019 (COVID-19) pandemic [8]. This cloud-based system provides a contactless share of glucose levels to healthcare providers and outpatient subjects. Moreover, recent advances in video-meeting tools enable interactive conversations via the Internet between healthcare providers and subjects in outpatient settings. Otherwise, although there are some reports about the usefulness of telemedicine in T1D subjects from the viewpoint of improved medical expenses and treatment motivation [9] or glucose profiles [10] from abroad, only a few reports show the usefulness of telemedicine for the subjects with diabetes in Japan, such as a report about type 2 diabetes [11] and there is no previous report using these telemedicine systems for CSII introduction as far as we know. Herein, we report the first case report using telemedicine systems to introduce CSII in the COVID-19 pandemic.
Case presentation
Case 1
A 36-year-old man with an eight-year history of T1D living approximately 1.5 hours by car from our hospital had been treated with basal-bolus MDI (insulin to carbohydrate ratio [CIR], 1 U:10-10-10 g; insulin sensitivity factor [ISF], 60 mg/dL/unit; target glucose levels <120 mg/dL) with insulin lispro and degludec 12-14 U. His endogenous insulin secretion was decreased (when the fasting plasma glucose [FPG] was 120 mg/dL, the fasting C-peptide immunoreactivity [CPR] and immunoreactive insulin [IRI] was 0.3 ng/dL and 0.6 mU/mL), and anti-glutamic acid decarboxylase (GAD) antibodies were positive (279.1 U/mL). His HbA1c level was >10.0% for several years and remained above the target level (<7.0%) shown in Japanese guidelines [1] or the American and European diabetes association [2] before the initiation of CSII. He used isCGM for four years to manage hypoglycemia, mainly due to excessive physical activities carrying heavy merchandise during his work at a drugstore. Owing to the fear of unexpected hypoglycemia, he was willing to maintain high blood glucose levels during the daytime (Figure 1), which resulted in more elevated glucose levels with TAR and TIR of 74% and 23% than the recommended percentage (<25%, >70%) in consensus statements [2,12], respectively (Table 1).
Table 1. Summary of laboratory data and isCGM indicators before and after CSII initiation.
After CSII initiation, all glucose-related parameters improved. In particular, the TIR and TBR were increased and decreased, respectively.
CSII - continuous subcutaneous insulin injection; HbA1c - hemoglobin A1c; TAR - time-above range; TIR - time-in range; TBR - time-below range; %CV - percent coefficient of variation
| Case | Laboratory data | Before CSII | One month after CSII | Four months after CSII |
| Case 1 | Mean glucose level (mg/dL) | 250 | 177 | 205 |
| HbA1c | 10.5 | 8 | 7.9 | |
| TAR (%) | 74 | 47 | 52 | |
| TIR (%) | 23 | 51 | 47 | |
| TBR (%) | 3 | 2 | 1 | |
| %CV | 40.8 | 37.3 | 40.5 | |
| Case 2 | Mean glucose level (mg/dL) | 172 | 193 | 163 |
| HbA1c | 7.9 | 7.9 | 7.1 | |
| TAR (%) | 40 | 49 | 37 | |
| TIR (%) | 54 | 50 | 60 | |
| TBR (%) | 6 | 1 | 3 | |
| %CV | 40.6 | 39.4 | 34.5 |
Figure 1. Examples of glucose profiles in both cases.
Examples of glucose profiles on isCGM of cases 1 and 2 are shown. Arrows are the typical timing, significantly increased or decreased glucose fluctuation.
isCGM - intermittently scanned continuous glucose monitoring; AGP - ambulatory glucose profile; CSII - continuous subcutaneous insulin infusion
Case 2
A 46-year-old woman with an 11-year history of T1D living approximately one hour by car from our hospital had been treated with basal-bolus MDI (CIR 1 U: 10-10-10 g; ISF of 50 mg/dL/unit; target glucose levels <120 mg/dL) with insulin lispro and glargine 12 U. Her endogenous insulin secretion was almost under detection limit (when the FPG was 149 mg/dL, the fasting CPR and IRI was <0.02 ng/dL and 0.6 mU/mL), and anti-GAD antibodies were positive (141.3 U/mL). Her HbA1c level was around 7-8% for several years. Notably, she often experienced hypoglycemia as an increased TBR of approximately 6-8% before the initiation of CSII using isCGM for several years. Especially, unexpected hypoglycemia occurred mainly late at night and sometimes in the daytime due to increased physical activity while at work in a factory (Figure 1) with TAR, TIR, and TBR of 40%, 54%, and 6%, respectively (Table 1).
Based on the circumstances above, we proposed the usage of CSII for better glycemic management, although AID was refused for economic reasons. They also refused hospitalization, a typical CSII introduction course in Japan, for the initial adjustment of the CSII owing to the fear of infection during the COVID-19 pandemic and the long-distance traveling difficulties. Therefore, we proposed an alternative plan as CSII with isCGM. First, we assessed the subjects' patterns of glucose trends before CSII initiation via isCGM (Figure 1). After that, according to these trends, we initiated CSII using MiniMedTM 670G (Medtronic, Northridge, CA, USA) in both of the cases with insulin lispro at the basal insulin rate of 0.5 U/hour for 24 h (case 1: presented flat glucose trend) and 0.45 U/hour at 11 PM-1 AM, 0.50 U/hour at 1 AM-5 AM, 0.45 U/hour at 5 AM-7 AM, 0.50 U/hour at 7 AM-11 PM (case 2: presented decreased glucose levels mainly late at night) with same CIR and ISF before initiation of CSII over isCGM. Subsequently, we offered him an online meeting (Zoom Video Productions Inc., San Jose, CA) three days after the face-to-face CSII initiation to confirm glucose trend via isCGM and any troubleshooting or insulin-loading procedure when needed to exchange insulin syringes and other supplies. During the meeting, we observed the actual procedure of filling an insulin reservoir and inserting an infusion set (Figure 2), which was appropriate and safe. Therefore, although we initially planned another online procedure checkup, we did not conduct an online follow-up. Concurrently, we discussed the glucose trends and adjusted the bolus insulin settings and basal insulin rate based on data from the cloud-platform-based system LibreView® (Figure 1). Though both cases showed hypoglycemia on the first day of CSII initiation, they showed relatively well-controlled glucose trends (Figure 1). Based on these data we proposed correction of CSII programs (case1: 0.50 U/hour at 0 AM-8 AM, 0.525 U/hour at 8 AM-10 AM, 0.475 U/hour at 10 AM-5 PM and 0.5 U/hour at 5 PM to 12 PM with CIR 8-10-10 g/U; case 2: 0.45 U/hour at 11 PM-1 AM, 0.50 U/hour at 1 AM-5 AM, 0.45 U/hour at 5 AM-10 AM, 0.40 U/hour at 10 AM-7 PM and 0.50 at 7 PM-11 PM with CIR 10-10-9 g/U). We also adjusted the temporary basal rate late before the workup for exercise-induced hypoglycemia.
Figure 2. The telemedicine image with video-meeting tools sharing isCGM data and procedure.
Desktop image of telemedicine in the present case. We confirmed the insulin pump procedure (left side) through a video meeting tool (Zoom Video Productions Inc, San Jose, CA) and glucose trends using LibreView® (Abbott Diabetes Care Inc., CA, USA), which is displayed on the right side. A video-cropped image is obtained at the time of exchanging insulin syringes.
isCGM - intermittently scanned continuous glucose monitoring
After that, they continued face-to-face outpatient visits approximately every month and showed decreased HbA1c TBR and increased TAR (Table 1). In particular, they adequately used temporary basal late based on their predicted exercise quantity. Additionally, case 1 was infected with COVID-19 in Japan's seventh wave of the COVID-19 pandemic, nine months after the initiation of CSII. At that time, we again conducted telemedicine follow-up and were able to follow his glucose trends. We advised him to increase basal late via cloud-platform-based isCGM in a remote area.
Discussion
This report describes two Japanese cases with T1D in whom CSII (MiniMedTM 670G) was successfully initiated using a combination of a video-meeting tool and cloud-platform-based isCGM.
The CGM system has shown significant advances recently. The updated recommendation from the American Diabetes Association [13] noted that MDI or CSII combined with CGM leads to better HbA1c levels and less hypoglycemia in youths and adults with T1D than those without T1D. Insulin delivery systems, including several types of insulin pumps, have been developed over the past few decades. Among them, AID is more expensive than using CSII solely, even with isCGM in Japanese self-funded insurance systems, and some subjects could select CSII (with isCGM) solely instead of AID in Japan. However, CSII is still not widely spread enough in Japan [14], despite its usefulness with isCGM [15]. Insulin pump therapy has mechanical procedure challenges and potential risks to users, resulting in unintended discontinuation of insulin due to kinking of the catheter and inappropriate syringe priming. Therefore, CSII is generally initiated during hospitalization in Japan to provide mechanical trouble education and careful titration of insulin doses [6], unlike other countries introducing it in outpatient settings. However, if subjects live in a place with barriers to healthcare access and elaborate follow-up, initiation of insulin pump therapy becomes difficult, especially during the COVID-19 pandemic. Thus, a consensus has not yet been established regarding selecting appropriate insulin and glucose monitoring systems. Currently, these devices are introduced individually based on socioeconomic status, ethnicity, and educational background [16].
In this case report, we conducted safe and effective CSII initiations using an online meeting tool and the cloud-platform-based isCGM in the subjects residing in a rural area and demonstrated the resulting long-term improvement in HbA1c and TIR similar to the previous report showing CSII advantage in T1D subjects [15]. Similarly, the effectiveness of the online pump training program in introducing the AID has been reported [17,18]. Although these technologies seem very useful and practical in terms of insulin pump education, additional careful follow-up with healthcare professionals is necessary after the initiation of insulin pumps since the mechanism or procedure of insulin pumps is becoming more complicated, and the Food and Drug Administration classifies insulin pumps as class II (moderate-risk) devices [19]. Furthermore, given that the CSII does not have a high- or low-glucose alert system, we should be more careful about the risk of CSII initiation than AID. Under such conditions, a telemedicine approach using a video-based meeting tool and cloud-platform-based isCGM could solve mechanical- or procedure-related glycemic excursions in the early stages after insulin pump initiation in a remote area with barriers to healthcare access. Moreover, the telemedicine approach is reported to improve motivations for glycemic control in subjects with diabetes in remote areas from hospitals [9,20].
Additionally, the chances of using telemedicine are increasing due to the COVID-19 pandemic. Telemedicine with cloud-platform-based isCGM shows efficacy and safety outcomes equal to face-to-face outpatient settings in T1D subjects [10]. The usefulness of telemedicine with CGM in different situations for T1D subjects requires further study.
Conclusions
This report presents unique cases of CSII initiation using an online meeting tool and cloud platform-based isCGM. There are several economic and geographical barriers to introducing insulin pumps, especially in Japan. Notably, during the COVID-19 pandemic, the telemedicine approach for the subjects wondering about the initiation of CSII might be a solid tool to support their safe and effective outpatient CSII introduction. A randomized study revealing the efficacy and safety of telemedicine for the initiation and follow-up of T1D subjects with insulin pumps would be of great interest.
Acknowledgments
We thank the patients for agreeing to publish this case report. We also thank Yuichiro Matsuo, Yoshitomo Naka (Medtronic Japan, Tokyo, Japan), Hiroyuki Takahashi, Kenichi Takaya (Abbott Diabetes Care Inc., CA, USA), and all the medical staff at the Ohta Nishinouchi Hospital for supporting the online follow-up of the subjects. Additionally, the authors thank Editage for providing grammar and spelling checks.
The authors have declared that no competing interests exist.
Author Contributions
Concept and design: Haremaru Kubo
Acquisition, analysis, or interpretation of data: Haremaru Kubo, Kazuhiro Sugimoto
Drafting of the manuscript: Haremaru Kubo
Critical review of the manuscript for important intellectual content: Haremaru Kubo, Kazuhiro Sugimoto
Supervision: Kazuhiro Sugimoto
Human Ethics
Consent was obtained or waived by all participants in this study
References
- 1.Japanese clinical practice guideline for diabetes 2019. Araki E, Goto A, Kondo T, et al. J Diabetes Investig. 2020;11:1020–1076. doi: 10.1111/jdi.13306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.The management of type 1 diabetes in adults. A consensus report by the American Diabetes Association (ADA) and the European Association for the Study of Diabetes (EASD) Holt RI, DeVries JH, Hess-Fischl A, et al. Diabetes Care. 2021;44:2589–2625. doi: 10.2337/dci21-0043. [DOI] [PubMed] [Google Scholar]
- 3.Open-source automated insulin delivery in type 1 diabetes. Burnside MJ, Lewis DM, Crocket HR, et al. N Engl J Med. 2022;387:869–881. doi: 10.1056/NEJMoa2203913. [DOI] [PubMed] [Google Scholar]
- 4.Psychosocial aspects of continuous subcutaneous insulin infusion in children with type 1 diabetes in Egypt; a limited resources country perspective. El Samahy MH, Salah NY, Abdeen MS, Falastin BR. Diabetol Metab Syndr. 2022;14:82. doi: 10.1186/s13098-022-00853-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Perceptions of residents among rural communities with medical group practice in Japan. Manabe T, Sawada T, Kojo T, Iguchi S, Haruyama S, Maeda T, Kotani K. Int J Environ Res Public Health. 2019;16:5124. doi: 10.3390/ijerph16245124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Changes in glycemic control and quality of life in pediatric type 1 diabetics with continuous subcutaneous insulin infusion of insulin aspart following multiple daily injection therapy. Kawamura T, Urakami T, Sugihara S, Kim HS, Mochizuki M, Amamiya S. Clin Pediatr Endocrinol. 2008;17:39–47. doi: 10.1297/cpe.17.39. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Continuous glucose monitoring and metrics for clinical trials: an international consensus statement. Battelino T, Alexander CM, Amiel SA, et al. Lancet Diabetes Endocrinol. 2023;11:42–57. doi: 10.1016/S2213-8587(22)00319-9. [DOI] [PubMed] [Google Scholar]
- 8.Clinical feasibility of remote intermittently scanned continuous glucose monitoring in coronavirus disease 2019 patients with and without diabetes during dexamethasone therapy. Shimizu N, Hayashi A, Ito S, et al. Endocr J. 2022;69:597–604. doi: 10.1507/endocrj.EJ21-0540. [DOI] [PubMed] [Google Scholar]
- 9.Reducing the cost of diabetes care with telemedicine, smartphone, and home monitoring. Kesavadev J, Mohan V. J Indian Inst Sci. 2023;103:231–242. doi: 10.1007/s41745-023-00363-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.COVID-19 pandemic lockdown in young people with type 1 diabetes: Positive results of an unprecedented challenge for patients through telemedicine and change in use of continuous glucose monitoring. Salabelle C, Ly Sall K, Eroukhmanoff J, et al. Prim Care Diabetes. 2021;15:884–886. doi: 10.1016/j.pcd.2021.06.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Glycemic control, renal progression, and use of telemedicine phone consultations among Japanese patients with type 2 diabetes mellitus during the COVID-19 pandemic: retrospective cohort study. Sankoda A, Nagae Y, Waki K, et al. JMIR Diabetes. 2023;8:0. doi: 10.2196/42607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Battelino T, Danne T, Bergenstal RM, et al. Diabetes Care. 2019;42:1593–1603. doi: 10.2337/dci19-0028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.7. Diabetes technology: standards of medical care in diabetes-2022. American Diabetes Association Professional Practice Committee. Diabetes Care. 2022;45:0–112. [Google Scholar]
- 14.The percentage of continuous subcutaneous insulin infusion usage among adult type 1 diabetes mellitus patients in Japan: a cross-sectional study at national hospital organization hospitals. Murata T, Aoki Y, Kato Y, et al. J Diabetes Sci Technol. 2017;11:1055–1056. doi: 10.1177/1932296817697330. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Clinical value of flash glucose monitoring in patients with type 1 diabetes treated with continuous subcutaneous insulin infusion. Moreno-Fernandez J, Pazos-Couselo M, González-Rodriguez M, Rozas P, Delgado M, Aguirre M, Garcia-Lopez JM. Endocrinol Diabetes Nutr. 2018;65:556–563. doi: 10.1016/j.endinu.2018.04.003. [DOI] [PubMed] [Google Scholar]
- 16.RT-CGM in conjunction with CSII vs MDI in optimizing glycaemic control in T1DM: systemic review and meta-analysis. William J, McCluskey J, Gleeson N. Endocrinol Diabetes Metab. 2022;5:0. doi: 10.1002/edm2.324. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Successful initiation of hybrid closed-loop system using virtual pump training program in a teenager with type 1 diabetes previously treated with multiple daily injections. Petrovski G, Campbell J, Almajali D, Al Khalaf F, Hussain K. J Diabetes Sci Technol. 2021;15:1394–1398. doi: 10.1177/1932296820950753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Early and sustained increase in time in range 1 year after initiation of hybrid close loop therapy via telemedicine in type 1 diabetes patients. Gómez AM, Henao D, Parra D, et al. Acta Diabetol. 2023;60:943–949. doi: 10.1007/s00592-023-02051-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Insulin pump risks and benefits: a clinical appraisal of pump safety standards, adverse event reporting and research needs. A joint statement of the European Association for the Study of Diabetes and the American Diabetes Association Diabetes Technology Working Group. Heinemann L, Fleming GA, Petrie JR, Holl RW, Bergenstal RM, Peters AL. Diabetologia. 2015;58:862–870. doi: 10.1007/s00125-015-3513-z. [DOI] [PubMed] [Google Scholar]
- 20.Application of telemedicine in diabetes care: the time is now. Aberer F, Hochfellner DA, Mader JK. Diabetes Ther. 2021;12:629–639. doi: 10.1007/s13300-020-00996-7. [DOI] [PMC free article] [PubMed] [Google Scholar]