Technical note: Impact of Dose Voxel Kernel (DVK) values on dosimetry estimates in 177Lu and 90Y Radiopharmaceutical Therapy (RPT) applications

Rachele Danieli 1,2,3, Daniele Pistone 4,5,6,*, Jonathan Tranel 7, Francesca Botta 8, Carlos Uribe-Munoz 9,10, Davide Raspanti 11, Francesc Salvat 12, Scott J Wilderman 13, Manuel Bardiès 14,15, Ernesto Amato 4,5,16, Yuni K Dewaraja 17, Marta Cremonesi 18
PMCID: PMC10843484  NIHMSID: NIHMS1929862  PMID: 37712869

Abstract

Background:

Radiopharmaceutical Therapy (RPT) is an increasingly adopted modality for treating cancer. There is evidence that the optimization of the treatment based on dosimetry can improve outcomes. However, standardization of the clinical dosimetry workflow still represents a major effort. Among the many sources of variability, the impact of using different Dose Voxel Kernels (DVKs) to generate absorbed dose (AD) maps by convolution with the time-integrated activity (TIA) distribution has not been systematically investigated.

Purpose:

This study aims to compare DVKs and assess the differences in the ADs when convolving the same TIA map with different DVKs.

Methods:

DVKs of 3×3×3 mm3 sampling - nine for 177Lu, nine for 90Y - were selected from those most used in commercial/free software or presented in prior publications. For each voxel within a 11×11×11 matrix, the coefficient of variation (CoV) and the percentage difference between maximum and minimum values (% maximum difference) were calculated. The total absorbed dose per decay (SUM), calculated as the sum of all the voxel values in each kernel, was also compared. Publicly available quantitative SPECT images for two patients treated with 177Lu-DOTATATE and PET images for two patients treated with 90Y-microspheres were used, including organs at risk (177Lu: kidneys; 90Y: liver and healthy liver) and tumors’ segmentations. For each patient, the mean AD to the volumes of interest (VOIs) was calculated using the different DVKs, the same TIA map and the same software tool for dose convolution, thereby focusing on the DVK impact. For each VOI, the % maximum difference of the mean AD between maximum and minimum values was computed.

Results:

The CoV (% maximum difference) in voxels of normalized coordinates [0,0,0], [0,1,0] and [0,1,1] were 5%(21%), 9%(35%) and 10%(46%) for the 177Lu DVKs. For the case of 90Y, these values were 2%(9%), 4%(14%) and 4%(16%). The CoV (% maximum difference) for SUM was 9%(33%) for 177Lu, and 4%(15%) for 90Y. The variability of the mean tumor and organ AD was up to 19% and 15% in 177Lu-DOTATATE and 90Y-microspheres patients, respectively.

Conclusions:

This study showed a considerable AD variability due exclusively to the use of different DVKs. A concerted effort by the scientific community would contribute to decrease these discrepancies, strengthening the consistency of AD calculation in RPT.

Keywords: Dose-Voxel-Kernel, 90Y, 177Lu, patient-specific dosimetry, radionuclide therapy

Introduction and background

In Radiopharmaceutical Therapy (RPT), three-dimensional (3D) voxel-level internal dosimetry is receiving increasing attention, driven by the interest in estimating not only the average absorbed dose (AD) to volumes of interest (VOIs), but also the AD non-uniformity due to the non-uniform spatial distribution of activity. The starting point for 3D voxel-level dosimetry is 3D voxel-level activity maps, derived from quantitative tomographic images (i.e. PET or SPECT scans) acquired at single or multiple time-points. Following voxel-level time-integration techniques, a time-integrated activity (TIA) map can be computed and, by means of one of the approaches described below, converted into a 3D AD map1.

To date, three different approaches are available for AD map calculation2:

  1. Local Energy Deposition (LED), which assumes the energy released in a voxel of the TIA map to be absorbed within the same voxel. LED typically includes the contribution of charged particles and ignores the one of photons;

  2. Convolution of a Dose Voxel Kernel (DVK) with a TIA map, where a DVK is defined as a 3D voxel matrix representing the mean AD to a target voxel per decay event in the central source voxel, both of which are contained in an infinite homogeneous medium. DVKs are sometimes referred to also as Voxel S-factors (VSFs) or Voxel S-Values (VSVs);

  3. Direct Monte Carlo (MC) simulation of the radionuclide decays that uses the distribution from a TIA map and tracks all the transport and interactions between the emitted radiation and the patient’s body where different materials are modeled from CT imaging.

Direct MC radiation transport is considered, at least in principle, the most accurate and personalized approach for internal dosimetry as it takes into account the heterogeneity of the tissue density and composition. Because MC simulations require relevant computational resources and expertise, however, LED and convolution algorithms are more frequently adopted for clinical dosimetry3.

DVKs are generally calculated exploiting MC simulations4. However, different MC codes and different processes - e.g. direct MC simulation, MC-based volume integration of a dose-point kernel (DPK), which is the radial distribution of the AD per decay event due to an isotropic point source located inside a virtually infinite and homogeneous absorbing medium, and MC-derived analytical methods - have been adopted by different authors to generate the actual DVKs from simulations. Even if some sets of DVKs have been compared to each other5, to our knowledge no specific investigation of the impact of using different DVKs on the AD calculation has been carried out yet in a systematic manner.

The aim of this work is to assess the differences in the ADs when convolving the same TIA map with different DVKs. This has been done using a set of 18 commonly adopted or recently developed DVKs for 177Lu and 90Y and using imaging data from patients treated with 177Lu-DOTATATE and with 90Y-microspheres.

Materials and methods

Dose voxel kernels

We used a total of 18 different DVKs with details given in Tables 1 (for 177Lu) and 2 (for 90Y). For 177Lu, nine DVKs with voxel sizes of 3×3×3mm3 and matrix sizes ranging from 11×11×11 to 275×275×275 were compared. Among them, five were generated by direct MC simulation (Lanconelli6, Pistone7, Salvat, Tranel and UMICH DPM), two were derived from DPKs (Graves8 and Vergara9), and two were computed based on analytical methods (Amato10 and Pistone_AM7). For 90Y, nine DVKs with voxel sizes of 3×3×3 mm3 and matrix sizes from 11×11×11 to 17×17×17 were selected. Six of the selected 90Y DVKs were computed by direct MC simulation (Bolch11, Lanconelli6, Pistone, Salvat, Tranel and UMICH DPM), two starting from a DPK (Graves8 and Vergara9) and one with an analytical method (Amato10). Note that because 90Y is a pure beta emitter (i.e. there are no gamma emissions) and given that the range of its beta particles in tissue (maximum range in soft tissue: 11 mm) is much shorter than gammas, the kernel matrix size of 90Y that accounts for a complete energy deposition of the emitted particles is smaller than for 177Lu.

Table 1:

Details about the computation methods and setting adopted for all the 177Lu DVKs employed in the present study.

177Lu DVKs Matrix
size
Material
composition
Material density Decay data Beta spectrum Calculation method MC code for DVK or DPK generation Number of simulated particles Energy/range cutoff Surrounding medium
Amato 10 11×11×11 Soft tissue 1.04 g/cm3 Radar17 Radar*17 Analytical method Geant4 9.1 10^8 10 μm Infinite volume
Graves 8 275×275×275 Water 1.00 g/cm3 NNDC18 Radar17 MC volume integration of DPKs MCNP 6.2 10^6 per emission type 1 keV Sphere of 2 m radius for DPK
Lanconelli 6 11×11×11 Soft tissue 1.04 g/cm3 Radar17§ Radar**17 Direct MC simulation DOSXYZnrc (EGSnrc) 2.5*10^7 1 keV Cubical region with side 16.5 cm
Pistone 7 15×15×15 Soft tissue 1.03 g/cm3 Radar17 Radar17 Direct MC simulation GAMOS 6.2.0 10^9 100 μm Cubical region with side 50 cm
Pistone_AM 7 15×15×15 Soft tissue 1.03 g/cm3 Radar17 Radar17 Analytical method GAMOS 6.2.0 10^9 100 μm Cubical region with side 50 cm
Salvat 11×11×11 Soft tissue 1.00 g/cm3 Nucleide19 García-Toraño et al.20 Direct MC simulation PENELOPE (version 2018)21 3*10^8 5 keV (electrons), 1 keV (photons) Sphere of 20 cm radius
Tranel 15×15×15 Water 1.00 g/cm3 ENSDF22 ENSDF22 Direct MC simulation GATE 9.0 10^8 10 μm Cubical region with side 1 m
Vergara 9 11×11×11 Soft tissue 1.04 g/cm3 ENSDF22 ENSDF22 MC volume integration of DPKs GATE (version not specified) 10^8 1 μm Cubical region with side 2.1 m
UMICH DPM 267×267×267 Soft tissue 1.00 g/cm3 NuDat 3.023 BetaShape24 Direct MC simulation Dose Planning Method (DPM)25 10^8 (electrons), 4*10^10 (photons) 20 keV (electrons), 4 keV (photons) Cubical region with side 267×3mm

Published within this paper

DVK computed by MC volume integration of DPKs in water, converted into soft tissue DVK by the Cross approximation26.

*

merge of monoenergetic e- DVKs, binning: 25 keV

**

merge of monoenergetic e- DVKs, binning: 10 keV

§

Characteristic x-rays and mono-energetic excluded

Comparison of DVKs

For each voxel in a 11×11×11 matrix, the coefficient of variation (CoV) – defined as the standard deviation divided by the mean - and the percentage difference between the maximum and the minimum values across all the selected DVKs normalized by the minimum (% maximum difference) were computed. In addition, normalized DVKs were generated by dividing each DVK value by the average value obtained with all the DVKs. The total absorbed dose per decay (SUM), calculated as the sum of all the voxel values in each kernel, was also compared. Given a DVK with matrix size N×N×N, SUM was computed as:

SUM=0,0,0+i=iN2i,0,0+j=1N20,j,0+k=1N20,0,k+i=1Nj=1N4i,j,0+i=1Nk=1N4i,0,k+j=1Nk=1N40,j,k+i=1Nj=1Nk=1N8i,j,k

All the DVKs used in the study are available in the Supplemental Material.

Comparison of ADs on Patient Data

Quantitative post-therapy 177Lu SPECT/CT images of two patients treated with 177Lu-DOTATATE12 and 90Y PET/CT images of two patients treated with 90Y-microspheres13 were used. These data are publicly available in the University of Michigan Deep Blue Data Repository.12,14 The 177Lu-DOTATATE data is the same used in the Society of Nuclear Medicine 177Lu Dosimetry Challenge.14,15 Segmentations of organs at risk (OARs) - kidneys16 for the two 177Lu-DOTATATE patients, and whole and normal liver for the two 90Y-microspheres patients – and tumors were also provided 14,15. For each patient, the AD maps were calculated using the different DVKs coupled with the same dosimetry software, MIM SurePlan MRT (MIM Software, USA), thereby focusing on the impact specifically of the DVK. Before performing the dosimetry, all the selected DVKs were manually imported into the software. For 177Lu, SPECT images acquired at four different time-points were registered and segmentations were propagated from the reference time-point (24 h) to the others. Next, 177Lu time-activity curves generated for each VOI were fitted to a sum of exponentials defined according to the Akaike Information Criterion. After the time-integration at the VOI level, 177Lu TIA maps – one for each patient - were derived from the reference activity map by scaling the activity in each voxel by the ratio between the TIA computed at the VOI level to the total activity in the VOI corresponding to the voxel considered. For 90Y, instead, a time-integration at the voxel-level considering only the physical decay was performed, as is standard because the microspheres do not redistribute. The resulting SPECT or PET-derived TIA maps, resampled to cubic voxel of 3 mm size, were converted into AD maps by means of a Fast Fourier Transform convolution with each of the selected DVKs. Finally, the AD maps were divided by a CT-based density map – also resampled to the DVK resolution – and manually multiplied by the mass density of the medium in which the DVKs were calculated to account for the difference between the latter and the actual tissue density. Excluding this last step, there was no manual intervention in the steps leading to the calculation of AD. For OARs and tumors, mean AD, D10 and D90 (i.e. the minimum AD to 10% and 90% of the volume, respectively) were computed. The average of the mean AD, D10 and D90 obtained with all the DVKs were calculated as well as the CoV and the % maximum difference. To further illustrate the variability among different DVKs, normalized mean AD were generated by dividing each mean AD by the average from all ADs.

Results

Comparison of DVKs

Figure 1 and 2 (top panels) show the selected 177Lu and 90Y DVKs as a function of the distance from the source voxel (intended as the distance from the geometric center of the source voxel, here and in the following), respectively. For 177Lu, the source voxel of the DVK contributes to almost all its energy deposition, the maximum beta particle range in tissue being 2 mm. For 90Y, in comparison, the contribution at voxels away from the source is more significant, due to the higher beta particle energy corresponding to a maximum range in tissue of 11 mm. Figure 1 and 2 (bottom panels) show the normalized DVKs as function of the distance from the source voxel. The CoV and the % maximum difference are plotted in Supplemental Fig S1.

Figure 1:

Figure 1:

Top panel - 177Lu DVKs as a function of the distance from the source voxel. Bottom panel - normalized 177Lu DVKs as a function of the distance from the source voxel. For better visualization, the values at different voxels but corresponding to the same distance have been averaged.

Figure 2:

Figure 2:

Top panel - 90Y DVKs as a function of the distance from the source voxel. Bottom panel - normalized 90Y DVKs as a function of the distance from the source voxel (part of data from Bolch are out of range). For better visualization, the values at different voxels but corresponding to the same distance have been averaged.

For 177Lu, CoV (% maximum difference) in voxels of normalized coordinates [0,0,0] (central voxel), [0,0,1] (3 mm from the center) and [0,1,1] (4.2 mm from the center) - which are the voxels giving the maximum contribution to the AD - were 5% (21%), 9% (35%), 10% (46%), respectively. For SUM, results were 9% (33%). Maximum differences were attributable to Lanconelli/Vergara (source voxel), Amato/UMICH DPM (voxels [0,0,1] and [0,1,1]) and Lanconelli/UMICH DPM (SUM). For 90Y, CoV (% maximum difference) results were 2% (9%), 4% (14%) and 4% (16%) at voxels of coordinates [0,0,0], [0,0,1] and [0,1,1], respectively; 4% (15%) in SUM. In this case, Amato/Graves provides the maximum difference in the source voxel, Lanconelli/Amato in the two diagonal voxels and Lanconelli/Vergara in SUM.

Comparison of ADs on Patient Data

For each patient, mean AD, D10 and D90 obtained with each selected DVK are provided in the Supplemental Tables S14. Average, CoV and % maximum difference of mean AD, D10 and D90 to kidneys (177Lu), liver/healthy liver (90Y) and tumors (177Lu and 90Y) are given in Table 3. Normalized mean AD are shown in Figure 3. For each VOI, maximum differences were attributable to Lanconelli/Vergara for both 177Lu and 90Y patients.

Table 3:

Average, CoV and % maximum difference of mean AD, D10 and D90 to OARs and tumors in patients treated with a) 177Lu and b) 90Y.

a) 177Lu DOTATATE : Patient 4 177Lu DOTATATE : Patient 6
Right Kidney Left Kidney Tumor 1 Tumor 2 Right Kidney Left Kidney Tumor 1 Tumor 2 Tumor 3 Tumor 4
Mean AD
Average (Gy) 2.63 2.71 19.80 28.50 5.39 3.43 3.19 30.89 3.09 3.73
CoV 5% 5% 5% 5% 5% 5% 5% 5% 5% 5%
% maximum difference 18% 19% 18% 19% 18% 19% 19% 19% 18% 19%
D10
Average (Gy) 3.58 3.66 32.74 41.43 7.58 5.19 4.09 39.08 4.39 6.56
CoV 5% 5% 5% 5% 5% 5% 5% 5% 5% 5%
% maximum difference 18% 18% 19% 19% 19% 19% 19% 19% 18% 19%
D90
Average (Gy) 1.56 1.64 7.84 13.93 2.78 1.79 2.28 21.96 1.70 1.28
CoV 5% 5% 5% 5% 5% 5% 5% 5% 5% 5%
% maximum difference 18% 18% 18% 19% 19% 18% 19% 19% 18% 18%
b) 90Y microspheres: Patient 26 90Y microspheres: Patient 56
Liver Healthy liver Tumor 1 Tumor 2 Tumor 3 Tumor 4 Liver Healthy liver Tumor 1 Tumor 2
Mean AD
Average (Gy) 67.57 64.68 220.04 450.38 63.52 673.57 52.59 48.39 432.42 107.22
CoV 4% 4% 4% 4% 4% 4% 4% 4% 4% 4%
% maximum difference 15% 15% 14% 14% 14% 14% 15% 15% 14% 14%
D10
Average (Gy) 180.16 172.58 328.39 727.69 122.25 1051.03 118.56 114.67 681.62 161.56
CoV 4% 4% 4% 4% 4% 4% 4% 4% 4% 4%
% maximum difference 15% 15% 14% 14% 13% 14% 15% 15% 14% 13%
D90
Average (Gy) 0.46 0.44 119.44 176.70 13.83 293.86 0.32 0.32 230.99 40.64
CoV 4% 4% 4% 4% 4% 4% 4% 4% 4% 4%
% maximum difference 14% 15% 15% 17% 17% 16% 16% 16% 16% 17%

Figure 3:

Figure 3:

Mean AD from each kernel divided by the average value from all ADs

Discussion

There is increasing evidence that personalized dosimetry could play an important role in the optimization of RPT treatments29. Multicenter trials are therefore essential to explore AD-effect relationships allowing for that optimization30. The clinical dosimetry workflow, however, consists of different tasks, each of which can be performed in multiple ways. This lack of standardization prevents drawing conclusions from multicenter trials because of the discrepancies that might occur in the data among centers. The standardization of the clinical dosimetry workflow following RPT, therefore, still represents a major challenge for physicists in the field15.

Many commercial and in-house software packages are now available for the computation of the AD and comparisons among them have been performed31. Such comparisons, however, usually do not allow a characterization of the variability associated to a single task within the clinical dosimetry workflow, for example the variability associated with the use of different algorithms for image registration or with the method for integrating the time-activity curve. Each software, in fact, offers different and limited solutions to perform the various tasks. The identification of the major sources of variability to be reduced in view of a standardization requires each of them to be isolated and analyzed separately from all the others. In this context, we aim to study the variability introduced when using different DVKs for the computation of the AD via convolution, while being fully aware that this may not be among the major sources of variability affecting the AD.

Our comparison showed that differences in the AD due only to the use of different DVKs are not negligible, with maximum values at the VOI level of 19% and 15% in patients treated with 177Lu-DOTATATE and 90Y-microspheres, respectively. These maximum differences were associated to the same pair of DVKs (Lanconelli/Vergara), perhaps indicating that these DVKs might be less accurate than other DVKs included in this comparison. Observed discrepancies were similar in different patients and VOIs (Figure 3).

Several factors, presented in Table 1, can contribute to the observed differences among the considered DVKs: the calculation method (e.g. direct MC simulation vs MC-based volume integration of DPKs), the nuclear data (e.g. RADAR spectra vs ICRP Publication 107), the definition of material density and composition (e.g. water vs soft tissue) or the MC code (e.g. Geant4 vs Penelope) and simulation parameters, such as the number of primaries, the energy/range cut-off and the physical models. Differences in the kernel size should be also considered. The analysis of these factors and of their impact on the accuracy of DVKs is beyond the scope of this note, as our study was designed to assess variability and not accuracy.

A limitation of this study is that DVKs for only two different radioisotopes (177Lu and 90Y) were included. The inclusion of radionuclides with higher gamma emission probability and energy, e.g. 131I, would have been of interest to investigate the impact of different DVKs on the AD far from the source. Moreover, for each of the included radioisotopes, we considered patients treated with the same type of RPT (peptide receptor radionuclide therapy and radioembolization for 177Lu and 90Y, respectively), which means similar activity distributions. Finally, the results presented in this study are valid for a specific voxel size (3mm). When different voxel sizes are considered, higher or lower differences in the AD calculation can be in principle observed. Based on some limited comparisons already made, however, we can state that for voxel sizes other than 3 mm (range 2–4 mm) differences in the DVKs and mean ADs are generally of the same magnitude as those reported in this study. Exceptionally, however, AD differences up to 50% were found when including 177Lu Graves’ DVKs for sizes other than 3 mm, although we are aware that recently corrections have been made after the conclusion of our study [https://zenodo.org/record/7596345#.ZBLouHaZOUk]. We therefore recommend always benchmarking the DVK chosen for each specific voxel size, even for already published datasets. As example, the few investigated DVKs for sizes other than 3 mm (2.4 mm and 4 mm) are plotted in Supplemental Fig S23.

The RPT dosimetry community should be made aware of the magnitude of the differences in DVKs and the associated variability in patient AD estimates. Since the use of different DVKs is only one of the many possible sources of variability affecting the AD calculation, in fact, the final variability could be considerable and potentially impair the possibility to identify dose-effect relationships from multicenter data. Additional investigations are ongoing to assess the robustness of the various methods for DVK computation and to generate a standardized set of DVKs for the AD calculation.

Conclusions

This study shows substantial differences between DVKs from different sources that propagate into not negligible variability (up to 19%) in patient-specific AD due exclusively to the use of different DVKs. Among the many sources of variability affecting the calculation of the AD, the variability introduced by using different DVKs can be easily reduced by standardization, strengthening the consistency of AD calculation in RPT.

Supplementary Material

Supinfo
Lu177_Amato_11x11x11_3mm.csv
Lu177_Lanconelli_11x11x11_2.4mm.csv
Lu177_Lanconelli_11x11x11_3mm.csv
Lu177_Lanconelli_11x11x11_4mm.csv
Lu177_Pistone_11x11x11_3mm.csv
Lu177_PistoneAM_11x11x11_3mm.csv
Lu177_Salvat_11x11x11_3mm.csv
Lu177_Tranel_15x15x15_3mm.csv
Lu177_UMICH_DPM_201x201x201_4.0mm.csv
Lu177_Graves_275x275x275_3mm.csv
Lu177_UMICH_DPM_267x267x267_3mm.csv
Lu177_Vergara_11x11x11_2.4mm.csv
Lu177_Vergara_11x11x11_3mm.csv
Lu177_Vergara_11x11x11_4.0mm.csv
Y90_Amato_11x11x11_3mm.csv
Y90_Bolch_11x11x11_3mm.csv
Y90_Graves_13x13x13_3mm.csv
Y90_Lanconelli_11x11x11_2.4mm.csv
Y90_Lanconelli_11x11x11_3mm.csv
Y90_Lanconelli_11x11x11_4mm.csv
Y90_Pistone_11x11x11_3mm.csv
Y90_Salvat_11x11x11_3mm.csv
Y90_Tranel_15x15x153mm.csv
Y90_UMICH_DPM_13x13x13_4.0mm.csv
Y90_UMICH_DPM_17x17x17_3mm.csv
Y90_UMICH_DPM_21x21x21_2.4mm.csv
Y90_Vergara_11x11x11_2.4mm.csv
Y90_Vergara_11x11x11_3mm.csv
Lu177_UMICH_DPM_333x333x333_2.4mm.csv
Y90_Vergara_11x11x11_4.0mm.csv

Table2:

Details about the computation methods and setting adopted for all the 90Y DVKs employed in the present study.

90Y DVKs Matrix
size
Material
composition
Material density Decay data Beta spectrum Calculation method MC code for DVK or DPK generation Number of simulated particles Energy/range cutoff Surrounding medium
Amato 10 11×11×11 Soft tissue 1.04 g/cm3 Radar17 Radar*17 Analytical method Geant4 9.1 10^8 10 μm Infinite volume
Bolch 11 11×11×11 Soft tissue 1.04 g/cm3 Eckerman et al.27,28 Not available Direct MC simulation EGS4 Not available Not available Cubical region with side 24.3 cm
Graves 8 13×13×13 Water 1.00 g/cm3 NNDC18 Radar17 MC volume integration of DPKs MCNP 6.2 10^6 per emission type 1 keV Sphere of 2 m radius for DPK
Lanconelli 6 11×11×11 Soft tissue 1.04 g/cm3 Radar17 § Radar**17 Direct MC simulation DOSXYZnrc (EGSnrc) 2.5*10^7 1 keV Cubical region with side 16.5 cm
Pistone 15×15×15 Soft tissue 1.03 g/cm3 ENSDF 22 ENSDF 22 Direct MC simulation GAMOS 6.2.0 10^9 100 μm Cubical region with side 50 cm
Salvat 11×11×11 Soft tissue 1.00 g/cm3 Nucleide19 García-Toraño et al.20 Direct MC simulation PENELOPE (version 2018)21 2*10^8 5 keV (electrons), 1 keV (photons) Sphere of 20 cm radius
Tranel 15×15×15 Water 1.00 g/cm3 ENSDF22 ENSDF22 Direct MC simulation GATE 9.0 10^8 10 μm Cubical region with side 1 m
Vergara 9 11×11×11 Soft tissue 1.04 g/cm3 ENSDF22 ENSDF22 MC volume integration of DPKs GATE (version not specified) 10^8 1 μm Cubical region with side 2.1 m
UMICH DPM 17×17×17 Soft tissue 1.00 g/cm3 NuDat 3.023 BetaShape24 Direct MC simulation DPM25 10^9 20 keV (electrons), 4 keV (photons) Cubical region with side 17×3mm

Published within this paper

DVK computed by MC volume integration of DPKs in water, converted into soft tissue DVK by the Cross approximation26.

*

merge of monoenergetic e- DVKs, binning: 25 keV

**

merge of monoenergetic e- DVKs, binning: 10 keV

§

Characteristic x-rays and mono-energetic excluded

Acknowledgements

The authors acknowledge David Mirando for reviewing some comparisons. Yuni Dewaraja acknowledges grant support from R01CA240706 and R01EB022075 awarded by the National Institute of Health, USA. This work was partially supported by the Italian Ministry of Health with Ricerca Corrente and 5×1000 funds.

Footnotes

Conflict of Interest

Yuni Dewaraja is a consultant for MIM Software.

References

  • 1.Sgouros G, Hobbs RF. Dosimetry for radiopharmaceutical therapy. Semin Nucl Med 2014;44(3):172–178. doi: 10.1053/j.semnuclmed.2014.03.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bardies M, Vergara Gil A. ABSORBED DOSE CALCULATION. In: International Atomic Energy Agency (IAEA), ed. Dosimetry for Radiopharmaceutical Therapy [IAEA Preprint] Vienna (Austria); 2022:154–175. http://inis.iaea.org/search/search.aspx?orig_q=RN:53037189. [Google Scholar]
  • 3.Della Gala G, Bardiès M, Tipping J, Strigari L. Overview of commercial treatment planning systems for targeted radionuclide therapy. Phys medica PM an Int J devoted to Appl Phys to Med Biol Off J Ital Assoc Biomed Phys 2021;92:52–61. doi: 10.1016/j.ejmp.2021.11.001 [DOI] [PubMed] [Google Scholar]
  • 4.Auditore L, Pistone D, Amato E, Italiano A. Monte Carlo methods in nuclear medicine. In: Signore A, ed. Nuclear Medicine and Molecular Imaging Elsevier; 2022:587–606. doi: 10.1016/B978-0-12-822960-6.00136-8 [DOI] [Google Scholar]
  • 5.Pacilio M, Lanconelli N, Lo MS, et al. Differences among Monte Carlo codes in the calculations of voxel S values for radionuclide targeted therapy and analysis of their impact on absorbed dose evaluations. Med Phys 2009;36(5):1543–1552. doi: 10.1118/1.3103401 [DOI] [PubMed] [Google Scholar]
  • 6.Lanconelli N, Pacilio M, Meo S Lo, et al. A free database of radionuclide voxel S values for the dosimetry of nonuniform activity distributions. Phys Med Biol 2012;57(2):517–533. doi: 10.1088/0031-9155/57/2/517 [DOI] [PubMed] [Google Scholar]
  • 7.Pistone D, Auditore L, Italiano A, Baldari S, Amato E. An analytic model to calculate Voxel S-Values for 177Lu. Published online 2022. doi: 10.1088/2057-1976/ac997e [DOI] [PubMed]
  • 8.Graves S, Tiwari A, Merrick M, et al. Accurate resampling of radial dose point kernels to a Cartesian matrix for voxelwise dose calculation Published online December 2020. doi: 10.5281/ZENODO.3827983 (Version 1.1 at https://zenodo.org/record/7596345#.ZD1DPnbP2KJ) [DOI]
  • 9.Vergara Gil A OpenDoseDVKData https://gitlab.com/opendose/opendosedvkdata. Accessed September 9, 2022.
  • 10.Amato E, Minutoli F, Pacilio M, Campennì A, Baldari S. An analytical method for computing voxel S values for electrons and photons. Med Phys 2012;39:6808–6817. doi: 10.1118/1.4757912 [DOI] [PubMed] [Google Scholar]
  • 11.Bolch WE, Bouchet LG, Robertson JS, et al. MIRD pamphlet no. 17: The dosimetry of nonuniform activity distributions - Radionuclide S values at the voxel level. J Nucl Med 1999;40(1). [PubMed] [Google Scholar]
  • 12.Dewaraja YK, Van BJ. Lu-177 DOTATATE Anonymized Patient Datasets: Multi-Time Point Lu-177 SPECT/CT Scans. Univ Michigan - Deep Blue Data Published online 2021. doi: 10.7302/0n8e-rz46 [DOI]
  • 13.Van BJ, Dewaraja YK. Y-90 PET/CT & SPECT/CT and Corresponding Contours Dataset 31JULY2020. Univ Michigan - Deep Blue Data Published online 2020. doi: 10.7302/pf4m-vn04 [DOI]
  • 14.Uribe C, Peterson A, Van B, et al. An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 1: Design and Early Results of the SNMMI Dosimetry Challenge. J Nucl Med 2021;62(Supplement 3):36S–47S. doi: 10.2967/jnumed.121.262748 [DOI] [PubMed] [Google Scholar]
  • 15.Uribe C, Brosch-Lenz J, Peterson A, et al. Variability in Dosimetry Calculations: an Analysis of the Results Submitted to the SNMMI Lu-177 Dosimetry Challenge. J Nucl Med 2022;63(supplement 2):2351. [Google Scholar]
  • 16.Dewaraja YK, Van BJ. Lu-177 DOTATATE Anonymized Patient Datasets: Lesion and Organ Volumes of Interest. Univ Michigan - Deep Blue Data Published online 2021. doi: 10.7302/vhrh-qg23 [DOI]
  • 17.RAdiation Dose Assessment Resource (RADAR) - The Decay Data http://www.doseinfo-radar.com/RADARDecay.html. Accessed September 9, 2022.
  • 18.National Nuclear Data Center. http://www.nndc.bnl.gov/index.jsp. Accessed March 1, 2018.
  • 19.NUCLEIDE. http://www.nucleide.org/DDEP_WG/DDEPdata.htm. Accessed October 12, 2022.
  • 20.García-Toraño E, Peyres V, Bé MM, Dulieu C, Lépy MC, Salvat F. Simulation of decay processes and radiation transport times in radioactivity measurements. Nucl Instruments Methods Phys Res Sect B Beam Interact with Mater Atoms 2017;396:43–49. doi: 10.1016/j.nimb.2017.02.002 [DOI] [Google Scholar]
  • 21.Agency NE. PENELOPE 2018: A Code System for Monte Carlo Simulation of Electron and Photon Transport; 2019. doi: 10.1787/32da5043-en [DOI]
  • 22.Evaluated Nuclear Structure Data File https://www.nndc.bnl.gov/ensdf/. Accessed October 12, 2022.
  • 23. National Nuclear Data Center NuDat (Nuclear Structure and Decay Data). NuDat 3.
  • 24.Mougeot X Reliability of usual assumptions in the calculation of β and ν spectra. Phys Rev C 2015;91(5):55504. doi: 10.1103/PhysRevC.91.055504 [DOI] [Google Scholar]
  • 25.Van B, Dewaraja YK, Niedbala JT, et al. Experimental validation of Monte Carlo dosimetry for therapeutic beta emitters with radiochromic film in a 3D-printed phantom. Med Phys Published online August 2022. doi: 10.1002/mp.15926 [DOI] [PMC free article] [PubMed]
  • 26.Cross WG. Variation of beta dose attenuation in different media. Phys Med Biol 1968;13(4):611–618. doi: 10.1088/0031-9155/13/4/310 [DOI] [PubMed] [Google Scholar]
  • 27.Eckerman KF, Westfall RJ, Ryman JC, Cristy M. Availability of nuclear decay data in electronic form, including beta spectra not previously published. Health Phys 1994;67(4):338–345. doi: 10.1097/00004032-199410000-00004 [DOI] [PubMed] [Google Scholar]
  • 28.Eckerman KF, Westfall RJ, Ryman JC, Cristy M. Nuclear Decay Data Files of the Dosimetry Research Group United States; 1993. http://inis.iaea.org/search/search.aspx?orig_q=RN:25039816. [Google Scholar]
  • 29.Garin E, Tselikas L, Guiu B, et al. Personalised versus standard dosimetry approach of selective internal radiation therapy in patients with locally advanced hepatocellular carcinoma ( DOSISPHERE-01 ): a randomised , multicentre , open-label phase 2 trial. Lancet Gastroenterol Hepatol 2020;1253(20):1–13. doi: 10.1016/S2468-1253(20)30290-9 [DOI] [PubMed] [Google Scholar]
  • 30.Taprogge J, Wadsley J, Miles E, Flux GD. Recommendations for Multicentre Clinical Trials Involving Dosimetry for Molecular Radiotherapy. Clin Oncol (R Coll Radiol) 2021;33(2):131–136. doi: 10.1016/j.clon.2020.12.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Mora-Ramirez E, Santoro L, Cassol E, et al. Comparison of commercial dosimetric software platforms in patients treated with 177Lu-DOTATATE for peptide receptor radionuclide therapy. Med Phys 2020;47(9):4602–4615. doi: 10.1002/mp.14375 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supinfo
Lu177_Amato_11x11x11_3mm.csv
Lu177_Lanconelli_11x11x11_2.4mm.csv
Lu177_Lanconelli_11x11x11_3mm.csv
Lu177_Lanconelli_11x11x11_4mm.csv
Lu177_Pistone_11x11x11_3mm.csv
Lu177_PistoneAM_11x11x11_3mm.csv
Lu177_Salvat_11x11x11_3mm.csv
Lu177_Tranel_15x15x15_3mm.csv
Lu177_UMICH_DPM_201x201x201_4.0mm.csv
Lu177_Graves_275x275x275_3mm.csv
Lu177_UMICH_DPM_267x267x267_3mm.csv
Lu177_Vergara_11x11x11_2.4mm.csv
Lu177_Vergara_11x11x11_3mm.csv
Lu177_Vergara_11x11x11_4.0mm.csv
Y90_Amato_11x11x11_3mm.csv
Y90_Bolch_11x11x11_3mm.csv
Y90_Graves_13x13x13_3mm.csv
Y90_Lanconelli_11x11x11_2.4mm.csv
Y90_Lanconelli_11x11x11_3mm.csv
Y90_Lanconelli_11x11x11_4mm.csv
Y90_Pistone_11x11x11_3mm.csv
Y90_Salvat_11x11x11_3mm.csv
Y90_Tranel_15x15x153mm.csv
Y90_UMICH_DPM_13x13x13_4.0mm.csv
Y90_UMICH_DPM_17x17x17_3mm.csv
Y90_UMICH_DPM_21x21x21_2.4mm.csv
Y90_Vergara_11x11x11_2.4mm.csv
Y90_Vergara_11x11x11_3mm.csv
Lu177_UMICH_DPM_333x333x333_2.4mm.csv
Y90_Vergara_11x11x11_4.0mm.csv