Graphical abstract
Keywords: Portable telecentric digital holographic microscopy, Holographic microscopy, Quantitative phase imaging, Portable microscope
Highlights
-
•
In comparison to commercially available DHM systems, P-TDHM is fully modular, allowing users to replace and customize individual modules to meet their specific experimental requirements.
-
•
The telecentric design removes the need for complex optics.
-
•
Users can easily switch between in-line and off-axis modes of P-TDHM, depending on their specific experimental needs.
Abstract
We present the design of a low-cost, portable telecentric digital holographic microscope (P-TDHM) that utilizes off-the-shelf components. We describe the system’s hardware and software elements and evaluate its performance by imaging samples ranging from nano-printed targets to live HeLa cells, HEK293 cells, and Dolichospermum via both in-line and off-axis modes. Our results demonstrate that the system can acquire high quality quantitative phase images with nanometer axial and sub-micron lateral resolution in a small form factor, making it a promising candidate for resource-limited settings and remote locations. Our design represents a significant step forward in making telecentric digital holographic microscopy accessible and affordable to the broader community.
Specifications table
Hardware name | Portable Telecentric Digital Holographic Microscopy |
Subject area |
|
Hardware type | Imaging tools |
Closest commercial analog | HoloMonitor |
Open source license | CC BY-NC-SA |
Cost of hardware | $6000 USD |
Source file repository | https://zenodo.org/record/8206220 |
1. Hardware in context
Digital holographic microscopy (DHM) is a bright field imaging technique that uses a single exposure to obtain a three-dimensional (3D) quantitative phase image [1]. Its high temporal resolution, limited only by the camera acquisition speed, has had an enormous impact in the fields of biology [2], [3] and micro-electromechanical systems (MEMS) [4], [5]. In-line DHM utilizes diffractive propagation methods for real-time particle [6] and cell tracking [7]. On the other hand, off-axis DHM is well-suited for real-time monitoring of sample dynamics [8], [9]. Telecentric digital holographic microscopy (TDHM) is an extension of DHM that compensates for spherical aberration in the optical objectives by adding a tube lens, applicable to both in-line and off-axis DHM. This design optically mitigates high-order phase aberrations in the hologram[18], effectively reducing wavefront distortions introduced by the microscope objective [10], [11], [12]. It bypasses the computation cost and image quality deterioration through the elimination of numerical compensations, such as polynomial fitting [19].
Current commercial off-axis DHM systems are expensive (> $30000 USD) and bulky (benchtop size). Systems such as the T-1000 (LynceeTec Inc.) and HO-DHM-UT01 (Holmarc, Inc.) incorporate piezo stages and control boxes to optimize interference fringes and suppress coherent noise [13]. However, these electromechanical devices increase system complexity and contribute to their high cost, which can be a barrier to use. The HoloMonitor M4 (PHI Inc.) is a phase imaging system small enough (310 x 180 x 85 mm) to fit in an incubator [14]. However, its portability comes into question for in situ monitoring as it requires the use of various peripherals, such as a laser driver, during operation [15]. Additionally, the HoloMonitor’s non-telecentric design limits its lateral resolution, making it less suitable for high-resolution applications [16], [17].
This paper presents a portable TDHM (P-TDHM) for imaging of transparent and translucent samples containing both in-line and off-axis functionalities (Fig. 1), with a supporting software pipeline. With the help of a mathematical optimization model [20], we simplified the optics construction and eliminated the need for electromechanical systems for hologram optimization. As a result, we have now developed a customizable DHM design that incorporates 3D-printed parts and features a user-friendly Graphical User Interface (GUI).
Fig. 1.
Configuration of the proposed P-TDHM. (a) is the picture of the proposed system enclosed by acrylic panels, comprising three distinct layers (b). The system light path is shown in (c).
2. Hardware description
The system is comprised of three layers, each housing specific optical components, as shown in Fig. 1. The laser and interference layers manipulate and modulate the laser beam into interference fringes, respectively. The support layer incorporates an optical path-matching mechanism to compensate for objective-induced spherical aberrations by matching the focal length of the tube lens. Using reconstruction algorithms [21], P-TDHM can digitally refocus on the target without the need for physical adjustments [22]. The design allows for an x-y stage should that be needed for sample positioning. As currently designed, P-TDHM has dimensions of 30 × 29 × 22 cm and weighs 5.2 kg, including its casing. The P-TDHM can switch between in-line and off-axis operation through the use of a beam block slider.
P-TDHM can achieve sub-wavelength axial resolution, depending on the signal-to-noise ratio of the camera and sample clarity. The system's theoretical lateral resolution is determined by the numerical aperture of the objective lens and the laser wavelength. The resulting field of view only depends on the magnification and the sensor size because the optic field is greater than that of the camera sensor. Based on our hardware setup, the user can replace the objective with other objectives as necessary to meet the image requirements of the experiment, as seen in Table 1.
Table 1.
Objectives with respective theoretical lateral resolution and field of view.
Objectives (Magnification / NA) |
Lateral Resolution (μm) | Field of View (μm) |
---|---|---|
4X / 0.1 | 3.9 | 1387 x 1850 |
10X / 0.25 | 1.6 | 555 x 740 |
20X / 0.35 | 1.2 | 277 x 370 |
20X / 0.75 | 0.52 | 277 x 370 |
60X / 0.9 | 0.44 | 92 x 133 |
100X / 0.95 | 0.41 | 55 x 74 |
P-TDHM uses a single exposure to generate 3D phase maps or 3D volumes for position tracking and imaging. As such, the acquisition speed is solely dependent on the camera sensor. The camera can be replaced to achieve the desired speed and sensor resolution [20]. P-TDHM uses a custom software application developed using PyQt5 in Python 3 to provide users with a graphical interface for adjusting the DHM parameters as well as viewing, processing, and saving images. The software is based on a series of extensible Standard Operation Procedures (SOPs) developed for the in-line and off-axis modes of P-TDHM. P-TDHM is fully customizable and 3D-printable, which significantly reduces material expenditure and enhances the system's flexibility. Given its portability and flexibility, P-TDHM is ideally suited for field studies and on-site imaging.
Below is the specification of P-TDHM system:
P-TDHM Parameters | Details |
---|---|
Dimensions | 30 cm × 29 cm × 22 cm |
Camera (BFS-U3-120S4M-CS) | Pixel size:1.85 μm, Frame rate: 31 fps, Resolution:4000 x 3000, ADC: 10-bit, 12-bit |
Objective (magnification / NA) | Olympus SLMPlan 20x / 0.35 |
Resolution | Lateral: 1.2 μm, Axial: 50 nm |
Field of View Dimensions | 277 × 370 μm |
Cost of Hardware | $6000 |
Design files
Design files summary
Design file name | File type | Open source license | Location of the file |
---|---|---|---|
1st_layer_Beam_spliter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Beam_support | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Laser_module | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Extra_support | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_ND_filter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Obj_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Objective_holder | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Ref_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Sample_holder | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Pinhole_adapter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_layer_Cube_adapter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_Beam_splitter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_Laser_module | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_ND_filter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_Obj_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_Objective_holder | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
1st_cover_Ref_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_Beamspander | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_Cam_Cube | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_ND_filter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_Obj | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
2nd_layer_support_ND | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Beamspander_back | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Beamspander_BackCover | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Beamspander_front | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Cam_cube | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_ND_filter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Top_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_Tube_lens | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_layer_TubeLens_cover | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_cover_cube | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_cover_mirror | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_cover_NDfilter | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
3rd_cover_tube | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Bot_long_down | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Bot_long_up | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Bot_short_left | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Bot_short_right | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Side_back_left | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Side_back_right | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Side_front_left | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Side_front_right | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Top_long_up | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Top_long_down | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Top_short_left | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Box_Top_short_left | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Cover_Cam (acrylic) | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Cover_front (acrylic) | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Cover_ND_side (acrylic) | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Cover_Sample_side (acrylic) | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
Cover_Top&Bot (X2) (acrylic) | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
BeamBlock_in_line | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
BeamBlock_off_axis | 3D Model(.stl) | CC BY-NC-SA | https://zenodo.org/record/8206220 |
DHMViewer | Compressed Python package (. zip) | GNU General Public License v3 (GPLv3) | https://zenodo.org/record/8206220 |
Laser module: The pinhole unit improves the spatial coherence property of the illumination source, and the beam expander increases the field of view.
ND filter: The ND filter adjusts the intensity of the object and reference beam, ensuring they are matched to achieve optimal interference contrast [23], [24].
First layer beam splitter: The beam splitter divides the illumination beam into the object and reference beams.
Mirrors: Mirrors alter the direction of the beam. Together, they utilize vertical space to create a 3D optical path, resulting in a smaller system for portability.
Sample holder module: The module holds samples for imaging. Users can redesign this component to meet their experimental requirements. A box of size 1.2″ x 2.3.
x 0.85″ has been reserved for the custom sample holder designs. Please refer to the assembling files for best fit with respect to the optical path. The focus plane of the objective we listed is 6.8 mm away from the objective side.
Objective holder module: The module holds microscope objectives (SLMPlan 20x/0.35). Users can customize this part to fit their desired objective.
Layer cover modules: The layer covers separate the different layers and allow the modules to function independently. They provide the flexibility for users to redesign each module without affecting other layers.
Second layer supports: The second layer provides support to the third layer and maintains the correct focal distance for the tube lens.
Third layer beam splitter: The beam splitter merges the object and reference beam, resulting in the superposition of the two beams on the camera sensor. The mirror in the reference path allows precise control of the interference fringe to optimize P-TDHM image quality [20].
Third layer beam expander back & cover: The beam expander helps to match the size of the reference beam with the size of the objective beam. The residual spherical aberration caused by the microscope objective can be compensated for using the linear stage.
Box and acrylic boards: These components protect the system from its surroundings.
Bill of materials
Bill of materials summary
Designator | Component | Quantity | Unit cost [USD] | Total cost [USD] | Source of materials | Material type |
---|---|---|---|---|---|---|
SLMPlan 20x/0.35 | Objective | 1 | $1175.00 | $1175.00 | https://www.spachoptics.com/olympus-slmplan-20x-objective-lens-p/olympus-slmplan-20x.htm | Non-specific |
BFS-U3-120S4M-CS | CMOS camera | 1 | $491.00 | $491.00 | https://www.flir.ca/products/blackfly-s-usb3/?model = BFS-U3-120S4M-CS | Non-specific |
CPS635 | Laser diode module | 1 | $98.45 | $98.45 | https://www.thorlabs.com/thorproduct.cfm?partnumber = CPS635 | Non-specific |
BS004 | 0.5″ beam splitter | 1 | $183.01 | $183.01 | https://www.thorlabs.com/thorproduct.cfm?partnumber = BS004 | Non-specific |
BS031 | 2″ beam splitter | 1 | $526.86 | $526.86 | https://www.thorlabs.com/thorproduct.cfm?partnumber = BS031 | Non-specific |
AC127-019-A-ML | Beam expander | 1 | $87.35 | $87.35 | https://www.thorlabs.com/thorproduct.cfm?partnumber = AC127-019-A-ML | Non-specific |
AC254-030-A-ML | 1 | $116.46 | $116.46 | https://www.thorlabs.com/thorproduct.cfm?partnumber = AC254-030-A-ML | Non-specific | |
P50D | 50 μm pinhole | 1 | $73.41 | $73.41 | https://www.thorlabs.com/thorproduct.cfm?partnumber = P50D | Non-specific |
LMR1 | Pinhole mount | 1 | $15.69 | $15.69 | https://www.thorlabs.com/thorproduct.cfm?partnumber = LMR1 | Non-specific |
M-DS25-X | Linear stage | 1 | $124.00 | $124.00 | https://www.newport.com/p/M-DS25-X | Aluminum |
NDC-25C-4 M-A | ND filter | 1 | $506.90 | $506.90 | https://www.thorlabs.com/thorproduct.cfm?partnumber = NDC-25C-4M | Non-specific |
NDC-50C-2 M-A | ND filter | 1 | $530.19 | $530.19 | https://www.thorlabs.com/thorproduct.cfm?partnumber = NDC-50C-2M-A | Non-specific |
AC127-019-A-ML | Beam expander | 1 | $87.35 | $87.35 | https://www.thorlabs.com/thorproduct.cfm?partnumber = AC127-019-A-ML | Non-specific |
AC254-250-A-ML | 1 | $108.70 | $108.70 | https://www.thorlabs.com/thorproduct.cfm?partnumber = AC254-250-A-ML | Non-specific | |
AC508-180-A-ML | Tube lens | 1 | $151.96 | $151.96 | https://www.thorlabs.com/thorproduct.cfm?partnumber = AC508-180-A-ML | Non-specific |
7T184-13 | Linear stage | 1 | $277.08 | $277.08 | https://www.standa.lt/products/catalog/translation_rotation?item = 254 | Aluminum |
KS05 | 0.5″ mirror mount | 4 | $80.14 | $320.56 | https://www.thorlabs.com/thorproduct.cfm?partnumber = KS05 | Aluminum |
BB05 | 0.5″ mirror | 4 | $53.06 | $212.24 | https://www.thorlabs.com/thorproduct.cfm?partnumber = BB05-E01 | Non-specific |
U50-PL | 1″ mirror mount | 5 | $118.00 | $590.00 | https://www.newport.com/p/U50-PL | Aluminum |
BB1-E02 | 1″ mirror | 5 | $77.35 | $386.75 | https://www.thorlabs.com/thorproduct.cfm?partnumber = BB1-E02 | Non-specific |
UPA1 | 1″ mirror adapter | 5 | $17.30 | $86.50 | https://www.newport.com/p/UPA1 | Non-specific |
5. Build instructions
All the parts were 3D printed using PLA + filament with 15 % infill and three shells on a RAISE3D Pro2 3D printer using files sliced by IdeaMaker. The layer height was set to 0.2 mm. The support varied depending on the parts themselves (Table Design Files Summary), as printing parameters depend on the 3D printer. The complete assembly file can be found in the online repository (https://zenodo.org/record/8206220). The screw types are indicated in the table:
Screws | Specifications | Quantities |
---|---|---|
① | #6–32 x 1/2″ UNC Hex Socket Head Cap | 90 |
② | #8–32 x 3/8″ UNC Hex Socket Head Cap | 8 |
③ | #8–32 x 1/2″ UNC Hex Socket Head Cap | 3 |
④ | M2 x 3 mm Hex Socket Head Cap | 6 |
⑤ | M2 x 5 mm Hex Socket Head Cap | 6 |
Laser Safety Notice
The build includes a Class IIIa laser diode module (4.5 mW, 635 nm). Do not power the laser during assembly. Please wear appropriate PPE during optical alignment.
First layer assembly (Fig. 2).
Fig. 2.
First layer assembly visual guide. The first layer expands the illumination beam size and splits the laser into object beam and reference beam.
1. Mount the laser, pinhole, linear stage (M-DS25-X), mirror, and beam splitter on the printed part (1st_layer_Laser_module.stl)
Alignment: Turn on the laser. Align the laser beam to go through the pinhole by adjusting the linear stage.
2. Cover the laser module with (1st_cover_Laser_module.stl) and mount the mirror and ND filter on (1st_layer_ND_filter.stl). Connect it to the enclosed laser module with (1st_layer_Beam_support.stl).
Connect the (1st_layer_Beam_spliter.stl and 1st_layer_Ref_mirror.stl) to the existing body and mount the mirror and the beam splitter
Alignment: Turn on the laser. Align the laser beam to propagate along the optical path and go through the center of the beam splitter.
4. Mount the mirror on (1st_layer_Ref_mirror.stl). Then enclose the body with (1st_cover_Beam_splitter.stl and 1st_cover_Ref_mirror.stl).
5. Mount objective and mirrors on (1st_layer_Obj_mirror.stl and 1st_layer_Objective_holder.stl) separately, then connect them with screws.
6. Mount (1st_layer_Extra_support.stl) on the parts from step 5, then connect to the main body. After aligning the object beam, enclose the first layer.
Second and third layer assembly (Fig. 3).
Fig. 3.
Second and third layer visual guide. The second layer supports the third layer and provides a proper focal distance for the tube lens. The third layer adjusts the interference angle to obtain high quality holograms.
7. Mount the second layer parts, shown in the design files summary, on the first layer individually and install the lens (AC127-019-A-ML, AC508-180-A-ML).
8. Mount the FLIR camera (3rd_layer_Cam_cube.stl).
9. Mount the reference mirror (3rd_layer_Top_mirror.stl).
10. Mount most parts of the third layer on the existing body, including the 2″ beam splitter (BS031), mirror, and 2″ lens (AC254-250-A-ML).
11. Mount the ND filter (3rd_layer_ND_filter.stl).
12. Mount the mirror (3rd_cover_tube.stl).
13. Enclose the object beam and beam splitter modules. Turn on the laser and camera, aligning the object beam located in the center of the camera. Mount the ND filter on the main body (3rd_layer_ND_filter.stl).
14. Enclose the reference beam path. Turn on the laser and camera, adjusting the manual stage and reference mirror until the desired Fourier spectrum is obtained
Coverage assembly (Fig. 4).
Fig. 4.
Coverage visual guide. The coverage isolates the optic system from the environment.
15. Mount the parts together with the black acrylic boards. (Box_Bot_x.stl)
16. Connect to the main body and mount the parts with black acrylic boards. Be careful to avoid damaging the laser cable when mounting the side piece near the camera. (Box_Side_x.stl).
17. Enclose the instrument using (Box_Top_x.stl) and the acrylic board
Routine Maintenance
Due to vibration and use, P-TDHM may occasionally need to be re-aligned. Alignment requires opening the external panels to reach the mirror mounts at the corners. Adjust the ND filters and ensure the intensity of the object beam is approximately twice that of the reference beam [23], [24].
6. Operation instructions
Software Installation and Setup
System requirements: 64-bit Windows 7 or later, macOS 10.14 or later, Ubuntu Linux 18.04 or later, with Python 3.7 or later, a monitor with 1280x800 or higher resolution, and a computer with a USB3 port.
1. Download and install the latest Python3 (Version 3.7 or above) release. (3.7.13 and 3.8.5 has been tested).
2. Download and extract the package DHMViewer into the desired directory.
3. Open a terminal in the DHMViewer directory, install the dependencies using pip command. The dependencies include Matplotlib, NumPy, Pillow, PyQt5, Scikit_image, and Tifffile.
“pip install -r./configuration.txt”
4. Download and Install the Latest Spinnaker Full SDK from FLIR official website (https://www.flir.ca/support-center/iis/machine-vision/downloads/spinnaker-sdk-and-firmware-download/). We tested SpinnakerSDK_FULL_3.0.0.118_x64.exe and SpinnakerSDK_FULL_2.7.0.128_x64.exe.
Intended Workflow
-
1.
Connect the DC power for the laser module and plug the FLIR camera's USB cable into the computer's USB3 port.
-
2.
Select the image mode (in-line or off-axis) by sliding the beam blocker to the respective positions.
-
3.
Launch the Spinnaker SpinView software, select camera interface and start capturing.
-
4.
Fine-tune the ND filters to optimize intensity of object and reference beams [23], [24] with the help of the GUI exposure views. The intensity ratio between object and reference beams can be 1 ∼ 1.5.
-
5.
Place the samples on the sample holder properly and slide into the microscope body.
-
6.
Captures the background and hologram images with the capture software manufacturer, saving them to two separate directories.
-
7.
Launch the DHMViewer program, set the desired operation mode and optical parameters.
-
8.
Specify the locations of background and holograms files in the DHMViewer user interface.
-
9.
Optionally specify a Region of Interest (ROI) to be processed.
-
10.
Choose whether to save the processed image, respective file directory, and image types. Select the range of images to be processed.
-
11.
Process (and save) the images, optionally saving the current system configuration.
-
12.
To obtain reconstructed z-stacks for a given off-axis hologram image, repeat steps (7) to (12) and switch between the off-axis and in-line operation modes.
Software Operation
Launch the DHMViewer program by double clicking dhm_viewer.py or running the Python command in the terminal. (“Python3./dhm_viewer.py”) When the launcher window appears, seen in (Fig. 6a.), click on the desired mode of operation (off-axis or in-line) to switch to the appropriate SOP mode. The main window should appear (Fig. 5.). From the left side panel, click the dropdown menu to expand the details of each step, and click the red proceed button.
-
1.
Optionally load an “.ini” configuration file from the file menu to jump to step 4 (Fig. 7a.). If needed, click the menu items in the menu bar to switch to another DHM mode.
-
2.
Set the parameters matching the desired system configuration and camera sensor specification (Fig. 8a.).
-
3.
Specify the file location of the background and the directory locations of the hologram image(s). Enter the address by hand or use the file browser by clicking the “Load” button. The background and the hologram will be loaded onto the viewer (Fig. 8b.).
-
4.
Optionally specify the Region of Interest. Only the content inside ROI will be processed. Click the checkbox “Edit ROI”, then follow the messages above the viewer to select ROI on screen (Fig. 9.). One can also shuffle the spin-box and click the “Update image” button at the top of the viewer to view the entire range of images to check the ROI's relative position (Fig. 8c.)
-
5.
Check the “save image” checkbox should the processed images need to be saved. In off-axis mode, select the image type to be saved. Select the range of images to be processed by the start and stop points by the spin-boxes. Click the “peek” button to view the selected start/stop point. Should anything be saved, select a file directory (Fig. 8d.).
-
6.
Click the process image button to process the selected images. A progress bar, a “Pause” button, and an “End task” button should appear on top of the viewer. Click the respective button to pause the operation or to end the processing and go back to save settings. On the side panel, click the dropdown menu to select the viewing image type during processing; clicking the “Live save” button will save the viewer's currently displayed content (Fig. 8e.). In in-line mode, when “Reconstructed volume” is selected as the viewing type, each slice of the reconstructed hologram can be viewed after processing and saving is complete.
Fig. 6.
DHMViewer menu actions. When the user first launches the program, the launcher window shall appear as in (a). By performing a mouse click action as indicated by the red arrow marked “1″ in (a), the user will observe the SOP of the respective operation mode (”off-axis“) selected in the Main Window. Clicking the dropdown menu pointed by the red arrow ”2″ in (b) results in the expansion of the “Set System Parameter” step of the SOP in (c). Clicking the “Confirm and Go to Next” button as indicated by the red arrow “3″ will lead to the expansion of the next step (”Load Images“ as in Fig. 5.) and the simultaneous closing of the current step (”Set System Parameter“). Nevertheless, every previously set dropdown menu can be reopened for setting changes. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 5.
DHMViewer main window. The user interface contains a side panel containing SOP items (left), a display and graphics toolbar (top right), and a graphics viewer (right).
Fig. 7.
DHMViewer program menu items. The “DHM Type” mode selector can be used to select the desired SOP for either the in-line or the off-axis mode, visible in (b). The “Load Configuration Receipt” menu item is visible in (a) at the start of the execution of a SOP. The additional menu option of “Dump Configuration Receipt” is available after all the parameter configurations have been completed, in Step 4 of the SOP, as shown in (c).
Fig. 8.
DHMViewer graphical user interface, Side panel with the “Off-axis TDHM” mode SOP. The “Set System Parameter” step of the SOP in (a) includes objective magnification and refractive index. The “Load Images” step in (b) consists of input fields and file handling interfaces for selecting the background and hologram image files. The “Set Region of Interest” step in (c) employs an interactive rectangular region selection to be performed on the graphics viewer by click-drag action. The “Set Saving Parameters” step in (d) includes input fields for saving indices in image sequence, saving image type and file handling interfaces. The “Process and Save Images” step in (e) allows format options for processed images to be viewed during processing. (f) illustrates a height map image for a representative target specified in the Image Validation section, displayed in DHMViewer.
Fig. 9.
DHMViewer graphical user interface, graphics viewer with “Set Region of Interest” step in SOP. By enabling the “Enable ROI Settings” checkbox in Fig. 8c., an interactive rectangular region selection can be performed on the graphics viewer by click-drag action in (a). By Clicking the “Save Current ROI” option in Fig. 8c., the ROI is saved and the yellow bounding box in (b) results. Sample: Nano-printed resin target. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Once every dropdown menu has been checked, user is able to click save the configuration to save all current settings into a “.ini” file in the desired directory (Fig. 7c.).
7. Validation and characterization
Image validation
The 3D imaging was validated using nano-additive manufacture targets produced by two-photon polymerization (Nanoscribe GmbH, PPGT2) of IP-Dip2 positive photoresist on glass. [25]. After photopolymerization, the polymer has a refractive index of 1.547. The targets are produced using a Nanoscribe Photonic Professional GT2 two-photon polymerization system. This system uses a 100 fs, 80 MHz pulsed laser with a 780 nm wavelength through a 63x focusing objective, which is rastered across the surface in a designated pattern. The target comprised three 40 µm wide, 250 nm high steps. Scanning electron microscope (Hitachi, SU7000) and atomic force microscope (Asylum Research, Cypher) images of the target are shown in Fig. 10a. and 10b, respectively. Fig. 10c. shows the target as imaged by TDHM and Fig. 10d. shows good agreement in both step height and lateral dimensions between the TDHM and AFM images profiles. The result shows that the height detection differences between AFM and P-TDHM was less than 50 nm.
Fig. 10.
DHM image validation with nano-printed target imaged by (a) SEM, (b) AFM, (c) TDHM. (d) Corresponding section plot of the image result obtained by AFM and TDHM.
The resulting topography (Optical Path Length (OPL)) is theoretically described by [26],
Where λ is the wavelength of the illumination, is the thickness distribution of the sample, ns and nm represent the refractive index of the sample and media, respectively.
3D tracking (in-line mode)
The in-line beam blocker functions by obstructing the reference beam of P-TDHM thereby establishing the classic in-line TDHM configuration, which is well-suited for volumetric sample tracking. In the absence of the reference beam, the field of view can be regarded as an interference pattern between the original laser field and the image field, which is scattered by the sample. Without the reference field, this configuration offers superior image contrast. To validate its tracking ability, imaging of 4.5 µm diameter polymer particles freely suspended in phosphate-buffered solution (PBS) was performed. As shown in Fig. 11 and Fig. 12, the hologram was generated by subtracting the background from the acquired original image. The Angular Spectrum method [27] was then employed to reconstruct the hologram 200 times at 1 μm increments.
Fig. 11.
Schematic of in-line reconstruction. The plane which the objective focuses on is called the image plane, and the plane which the target is located is known as the sample plane. Reconstruction is performed using the Angular Spectrum method.
Fig. 12.
Principle of 3D particle tracking for in-line mode. After subtracting the background, the hologram was reconstructed 200 times with 1 μm steps using the Angular Spectrum method. The focused image was 150 μm away from the image plane.
The results of the reconstruction (Fig. 11 and Fig. 12) indicate that the sample plane was determined to be 150 μm away from the image plane. The reconstructed images can be evaluated via its Fourier spectrum with micrometer accuracy [21]. The Angular Spectrum method is employed to reconstruct the entire field of view. Set in in-line mode, P-TDHM system can simultaneously track all particles in 3D with a single-shot hologram, provided the diffraction patterns can be clearly captured.
Numerical refocusing (off-axis mode)
The reconstruction algorithm can also be executed in the off-axis mode of the proposed system. Due to the impact of the reference wave, the axial resolution of its reconstruction may not be as high as that of the in-line mode, but it is still sufficient for numerical refocusing of the optic field. As shown in Fig. 13, the unfocused hologram could be reconstructed back into its focus plane numerically.
Fig. 13.
Auto-refocusing results of HeLa cells. With the same reconstruction algorithm used in Fig. 12, the image of HeLa cells was reconstructed six times in different depth positions. This proves that the reconstruction algorithm can also be used to numerically refocus.
Long-Term Imaging
To establish the validity of P-TDHM for long term imaging, HEK293 cells were continuously imaged for 33 h. The HEK293T (ATCC CRL-3216) cells were cultured in high glucose (4.5 g/L) Dulbecco's Modified Eagle's medium (Gibco), supplemented with 10 % fetal bovine serum (Gibco) and penicillin/streptomycin (Gibco; 100U/mL penicillin, 100 µg/mL streptomycin). The cells were harvested five hours prior to imaging and transferred into 35 mm culture dishes for further analysis. Following the instructions described previously, the dish with cells was imaged in room temperature directly without adding any extra nutrients, as shown in Fig. 14.
Fig. 14.
Migration of living HEK239T cells (ATCC CRL-3216) in PBS, imaged by P-TDHM. The entire sequence spans 33 h. The unit of the height bar is micrometer.
Cell growth and migration continued for 15 h with the nutrient, before the collapse of the aggregation after 20 h. Quantitative phase images (QPI) represent the product of sample thickness and its refractive index, which closely relate to the cell dry mass [28], and is theoretically not impacted by transients in illumination power. Additionally, reconstruction algorithms such as the Angular Spectrum method allow P-TDHM to digitally refocus the images. These attributes suggest that TDHM is better suited for long-term imaging of cellular dynamics than conventional brightfield microscopy.
In-field Imaging
Ultimately, the compact design and user-friendly software provides P-TDHM with the in-field imaging capabilities. These distinctive traits collectively point to the potential of P-TDHM as a tool for advancing phytoplankton studies. The significance of monitoring phytoplankton spans industries, research, and natural resource management, as emphasized by [29].
In a preliminary study, P-TDHM was tested on-site in a remote location.(See Fig. 15) The study was conducted on surface water samples procured during a bloom event on July 11, 2023, along the shoreline of Cayuga Lake, USA. P-TDHM successfully captured detailed images of Dolichospermum, the dominant phytoplankton in cyanobacterially harmful algal blooms. This bloom event was later reported by Community Science Institute (CSI), USA, and the dominance of Dolichospermum was confirmed [30].
Fig. 15.
In-field imaging of Dolichospermum using P-TDHM. The sample is from Meyers Park in Cayuga Lake, USA. (a) In-field imaging with P-TDHM; (b) sampling location. Dolichospermum imaged by (c) bright field and (d) P-TDHM, the unit of the height bar is micrometers.
P-TDHM’s inherent stability over fluctuating imaging conditions, such as light sources and focus adjustments, contribute to the method's consistency and reliability in phytoplankton monitoring. Additionally, P-TDHM imaging is not impacted by illumination intensity transients, so it is more reliable to analyze harmful algal blooms from different time and locations. The depth information provided by P-TDHM also offers an additional dimension for more comprehensive analysis.
In-field imaging operations with P-TDHM only require the microscope body and a personal computer. With a custom-designed flow chamber and appropriate algorithms, P-TDHM can be used for particle analysis akin to the FlowCan system (Yokogawa Fluid Imaging Technologies, Inc) with the additional capability of 3D volumetric imaging. Similarly, the user could incorporate waterproof panelling in the design of P-TDHM to perform in-situ, automated and submersible imaging, as demonstrated by FlowCytobot (McLANE Research Laboratories, Inc).
P-TDHM’s inherent flexibility enables many other applications in the field of biology, physics [31], and medical diagnosis [32], [33]. For example, it can be introduced to assist medical diagnosis in blood tests, retinal imaging, and dental imaging. P-TDHM can also be used for airborne particle and aerosol detection to monitor pollutants in air [34], [35]. Additionally, with a proper design of the sample holder with microfluids similar to [36], the proposed system could work as tomographic cytometry as well [37].
Concluding remarks
The portable telecentric digital holographic microscope (P-TDHM) described herein retains the well-established and comprehensive capabilities of conventional TDHMs but is built in a compact lightweight form factor. Compared to commercial alternatives, the cost-effective nature of P-TDHM confers a high degree of accessibility to a broad range of potential end-users.
Capabilities
-
•
High-resolution imaging. The axial resolution of P-TDHM is not limited by conventional diffraction limit.
-
•
Quantitative phase imaging. P-TDHM provides quantitative phase information of the sample, which can be used to measure a sample’s refractive index and thickness.
-
•
Real-time imaging. Volume acquisition is limited only by the acquisition speed of the camera.
-
•
Long-term monitoring. Theoretically, the recording range of the T-DHM is unlimited.
-
•
Portability.
Limitations
-
•
Limited imaging depth. P-TDHM requires the illuminating laser to pass through the sample and interfere with the reference beam. As such, the performance of P-TDHM will be limited by the sample’s refractive index and thickness.
-
•
Vibration sensitive. The axial resolution of P-TDHM reaches tens of nanometers, so the environmental vibration would impact the beam interference and the image quality. Thus, users should keep it stable during the imaging, like adding rubber isolation underneath.
Ethics statements
This work does not involve the use of human and animal subjects.
CRediT authorship contribution statement
Lei Jin: Conceptualization, Formal analysis, Investigation, Methodology, Software, Visualization, Writing – original draft, Writing – review & editing, Validation. Ziyang Yu: Software, Investigation, Writing – original draft. Aaron Au: Conceptualization, Investigation, Methodology, Resources, Visualization, Writing – review & editing. Peter Serles: Investigation, Resources, Writing – review & editing. Nan Wang: Data curation, Investigation, Writing – review & editing. Jeremy T. Lant: Resources, Writing – review & editing. Tobin Filleter: Resources, Supervision. Christopher M. Yip: Funding acquisition, Supervision, Writing – review & editing, Resources.
Declaration of competing interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
The authors would like to thank the Centre for Research and Application in Fluidic Technologies (CRAFT) under the Canada Foundation for Innovation/Ontario Research Fund #36442 for access to the Nanoscribe Photonic Professional GT2, the Vanier Canada Graduate Scholarship (PS), and the Natural Sciences and Engineering Research Council of Canada (CMY: grant number RGPIN-2015-043).
Biographies
Lei Jin is a Ph.D. candidate in the Yip Lab in the Institute of Biomedical Engineering & Donnelly Center at the University of Toronto, Canada. His academic pursuits include holographic 4D imaging, optic design & simulation, and advanced biomedical imaging. Lei has served as a peer-reviewer for the Optica publishing group, Optical Society of America. He received his M.A.Sc. from Tianjin University, and his B.E.Sc. from Hefei University of Technology.
ZiyangYu is a Ph.D. student in the Institute of Biomedical Engineering at the University of Toronto. His research focuses on smart, machine learning-accelerated biomedical imaging systems. He earned a B.A.Sc in Computer Engineering from the University of Toronto in 2023.
Aaron Au obtained his B.A.Sc. in Nanotechnology Engineering from the University of Waterloo in 2014 and his M.A.Sc. in Chemical Engineering & Applied Chemistry from the University of Toronto in 2018. He has since been employed as a lab technician in Professor Christopher Yip’s lab.
PeterSerles is a Ph.D. Candidate, Vanier Scholar, and Connaught PhD Fellow in the Department of Mechanical & Industrial Engineering at the University of Toronto. Peter’s research interests are in nanoscale additive manufacturing, nanomechanics, and nanoscale device design. He has authored more than 20 peer-reviewed papers in top international journals such as Nature, Science Advances, and Advanced Materials and has held research positions with the National Research Council of Canada, Rice University, and the Institut National des Sciences Appliquées de Lyon. He received his M.A.Sc. from the University of Toronto, and his B.E.Sc. from Western University where he won the Canadian Society for Mechanical Engineers Gold Medal.
Nan Wang is a PhD candidate in the Richardson Lab within the Department of Civil and Environmental Engineering at Cornell University, USA. Her academic pursuits center on the multifaceted domain of cyanobacterial Harmful Algal Bloom monitoring and underlying mechanisms. Prior to her doctoral studies, Nan Wang earned her B.A.Sc. degree from Nankai University, China, in 2019.
JeremyLant is a postdoctoral fellow in Aaron Wheeler's lab at the University of Toronto. Jeremy's affiliations are the Terrence Donnelly Centre for Molecular and Cellular Biology Research and the Department of Chemistry. Jeremy's research focuses on automated bio-assays, mammalian cell culture, and single cell preparation for -omics. Jeremy holds a PhD and B.M.Sc. degree in biochemistry from the Schulich School of Medicine and Dentistry at the University of Western Ontario
TobinFilleter is currently a Professor and the Associate Chair of Graduate Studies in the Department of Mechanical & Industrial Engineering at the University of Toronto. Prior to joining the MIE department at U of T, Dr. Filleter was a postdoctoral research fellow in the Department of Mechanical Engineering at Northwestern University (2009-2012). Dr. Filleter received a BSc (Eng.) in Engineering Physics from Queen’s University (2003) and PhD in Physics from McGill University (2009). During his PhD Dr. Filleter also spent time in Germany as a visiting scientist at the INM-Leibniz Institute for New Materials. Professor Filleter’s research interests are in nanomechanics of materials. Specific areas of research include nanotribology, mechanics of 2D materials, nanocomposites, and non-destructive testing. He has authored papers in many top international journals including Nature, Nature Materials, Science Advances, and Nature Communications. He is the recipient of several major awards including the Erwin Edward Hart Professorship, CSME I.W. Smith Award, and Ontario Early Researcher Award.
Christopher Yip received his B.A.Sc. in Chemical Engineering & Applied Chemistry from the University of Toronto in 1988 and his Ph.D. in Chemical Engineering from the University of Minnesota in 1996. He joined the Department of Chemical Engineering & Applied Chemistry in 1997 and was promoted to Associate Professor in 2002 and to Full Professor in 2007. He is a faculty member with the Department of Chemical Engineering & Applied Chemistry, the Department of Biochemistry and the Institute of Biomaterials and Biomedical Engineering. He is a Principal Investigator with the Donnelly Centre for Cellular & Biomolecular Research at the University of Toronto. He was appointed the Dean of the Faculty of Applied Science & Engineering on July 2, 2019, after serving two years as Associate Vice-President, International Partnerships in the University's Office of the Vice President, International. He is a Fellow of the American Association for the Advancement of Science (2009), a Fellow of the Engineering Institute of Canada (2014) and held a Tier II Canada Research Chair in Molecular Imaging (2000-2010).
References
- 1.M.K. Kim, Principles and techniques of digital holographic microscopy Terms of Use: http://spiedl.org/terms Principles and techniques of digital holographic microscopy, (2010). doi: 10.1117/6.0000006.
- 2.Marquet P., Depeursinge C., Magistretti P.J. Exploring neural cell dynamics with digital holographic microscopy. Annu. Rev. Biomed. Eng. 2013;15:407–431. doi: 10.1146/ANNUREV-BIOENG-071812-152356. [DOI] [PubMed] [Google Scholar]
- 3.Doblas A., Roche E., Ampudia-Blasco F.J., Martínez-Corral M., Saavedra G., Garcia-Sucerquia J. Diabetes screening by telecentric digital holographic microscopy. J. Microsc. 2016;261:285–290. doi: 10.1111/JMI.12331. [DOI] [PubMed] [Google Scholar]
- 4.Pagliarulo V., Russo T., Miccio L., Ferraro P. Numerical tools for the characterization of microelectromechanical systems by digital holographic microscopy. J. Micro/Nanolithogr. MEMS MOEMS. 2015;14 doi: 10.1117/1.jmm.14.4.041314. [DOI] [Google Scholar]
- 5.Coppola G., Ferraro P., Iodice M., De Nicola S., Finizio A., Grilli S. A digital holographic microscope for complete characterization of microelectromechanical systems. Meas. Sci. Technol. 2004;15:529. doi: 10.1088/0957-0233/15/3/005. [DOI] [Google Scholar]
- 6.Sheng J., Malkiel E., Katz J. Digital holographic microscope for measuring three-dimensional particle distributions and motions. Appl. Opt. 2006;45:3893–3901. doi: 10.1364/AO.45.003893. [DOI] [PubMed] [Google Scholar]
- 7.Langehanenberg P., Ivanova L., Bernhardt I., Ketelhut S., Vollmer A., Dirksen D., Georgiev G., von Bally G., Kemper B. Automated three-dimensional tracking of living cells by digital holographic microscopy. J. Biomed. Opt. 2009;14 doi: 10.1117/1.3080133. [DOI] [PubMed] [Google Scholar]
- 8.Y. Emery, E. Cuche, F. Marquet, N. Aspert, P. Marquet, J. Kühn, T. Colomb, F. Montfort, F. Charrière, C. Depeursinge, P. Debergh, R. Conde, Digital Holographic Microscopy (DHM) for metrology and dynamic characterization of MEMS and MOEMS, Mems, Moems, and Micromachining II. 6186 (2006) N1860–N1860. doi: 10.1117/12.660029.
- 9.Witte S., Plauşka A., Ridder M.C., van Berge L., Mansvelder H.D., Groot M.L. Short-coherence off-axis holographic phase microscopy of live cell dynamics. Biomed. Opt. Express. 2012;3:2184. doi: 10.1364/BOE.3.002184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sánchez-Ortiga E., Ferraro P., Martínez-Corral M., Saavedra G., Doblas A. Digital holographic microscopy with pure-optical spherical phase compensation. J. Opt. Soc. Am. A Opt. Image Sci. Vis. 2011;28:1410. doi: 10.1364/JOSAA.28.001410. [DOI] [PubMed] [Google Scholar]
- 11.Doblas A., Sánchez-Ortiga E., Martínez-Corral M., Saavedra G., Garcia-Sucerquia J. Accurate single-shot quantitative phase imaging of biological specimens with telecentric digital holographic microscopy. J. Biomed. Opt. 2014;19 doi: 10.1117/1.JBO.19.4.046022. [DOI] [PubMed] [Google Scholar]
- 12.Mikuła M., Kozacki T., Józwik M., Kostencka J. Accurate shape measurement of focusing microstructures in Fourier digital holographic microscopy. Appl. Opt. 2018;57:A197. doi: 10.1364/AO.57.00A197. [DOI] [PubMed] [Google Scholar]
- 13.Yu H., Jia S., Dong J., Huang D., Xu S. Phase curvature compensation in digital holographic microscopy based on phase gradient fitting and optimization. J. Opt. Soc. Am. A Opt. Image Sci. Vis. 2019;36:D1. doi: 10.1364/JOSAA.36.0000D1. [DOI] [PubMed] [Google Scholar]
- 14.Zhang Y., Judson R.L. Evaluation of holographic imaging cytometer holomonitor M4® motility applications. Cytometry A. 2018;93:1125–1131. doi: 10.1002/CYTO.A.23635. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Sebesta M., Egelberg P.J., Langberg A., Lindskov J.-H., Alm K., Janicke B. HoloMonitor M4: holographic imaging cytometer for real-time kinetic label-free live-cell analysis of adherent cells. SPIE. 2016;9718 doi: 10.1117/12.2216731. [DOI] [Google Scholar]
- 16.Sirico D.G., Miccio L., Wang Z., Memmolo P., Xiao W., Che L., Xin L., Pan F., Ferraro P. Compensation of aberrations in holographic microscopes: main strategies and applications. Appl. Phys. B. 2022;128:1–19. doi: 10.1007/S00340-022-07798-8/FIGURES/6. [DOI] [Google Scholar]
- 17.Y. Liu, Z. Wang, J. Huang, Recent Progress on Aberration Compensation and Coherent Noise Suppression in Digital Holography, Applied Sciences 2018, Vol. 8, Page 444. 8 (2018) 444. doi: 10.3390/APP8030444.
- 18.E. Sánchez-Ortiga, A. Doblas, M. Martínez-Corral, G. Saavedra, P. Ferraro, Digital holographic microscopy with pure-optical spherical phase compensation, JOSA A, Vol. 28, Issue 7, Pp. 1410-1417. 28 (2011) 1410–1417. doi: 10.1364/JOSAA.28.001410. [DOI] [PubMed]
- 19.T. Colomb, E. Cuche, F. Charrière, J. Kühn, N. Aspert, F. Montfort, P. Marquet, C. Depeursinge, Automatic procedure for aberration compensation in digital holographic microscopy and applications to specimen shape compensation, Applied Optics, Vol. 45, Issue 5, Pp. 851-863. 45 (2006) 851–863. doi: 10.1364/AO.45.000851. [DOI] [PubMed]
- 20.Jin L., Yu Z., Au A., Yip C. Practical approach for optimizing off-axis telecentric digital holographic microscope design. Appl. Opt. 2022;61 doi: 10.1364/AO.476308. [DOI] [PubMed] [Google Scholar]
- 21.P. Langehanenberg, B. Kemper, G. von Bally, Autofocus algorithms for digital-holographic microscopy, Biophotonics 2007: Optics in Life Science (2007), Paper 6633_13. (2007) 6633_13. doi: 10.1364/ECBO.2007.6633_13.
- 22.Langehanenberg P., von Bally G., Kemper B. Autofocusing in digital holographic microscopy. 3D Res. 2011;2:1–11. doi: 10.1007/3DRES.01(2011)4/METRICS. [DOI] [Google Scholar]
- 23.Carl D., Kemper B., Wernicke G., Von Bally G. Parameter-optimized digital holographic microscope for high-resolution living-cell analysis. Appl. Opt. 2004;43:6536–6544. doi: 10.1364/AO.43.006536. [DOI] [PubMed] [Google Scholar]
- 24.Trujillo C., Castañeda R., Piedrahita-Quintero P., Garcia-Sucerquia J. Automatic full compensation of quantitative phase imaging in off-axis digital holographic microscopy. Appl. Opt. 2016;55:10299. doi: 10.1364/AO.55.010299. [DOI] [PubMed] [Google Scholar]
- 25.Serles P., Haché M., Tam J., Maguire A., Li T., Wang G., Sebastian K., Lou J., Jia C., Ajayan P.M., Howe J., Zou Y., Filleter T. Mechanically robust pyrolyzed carbon produced by two photon polymerization. Carbon N y. 2023;201:161–169. doi: 10.1016/J.CARBON.2022.09.016. [DOI] [Google Scholar]
- 26.Goodman J.W. 3rd ed. Roberts and Company Publishers; 2005. Introduction to Fourier optics. [Google Scholar]
- 27.C. Mann, M. Kim, C.J. Mann, M.K. Kim, Quantitative phase-contrast microscopy by angular spectrum digital holography, Https://Doi.Org/10.1117/12.645412. 6090 (2006) 50–57. doi: 10.1117/12.645412.
- 28.Kastl L., Isbach M., Dirksen D., Schnekenburger J., Kemper B. Quantitative phase imaging for cell culture quality control. Cytometry A. 2017;91:470–481. doi: 10.1002/cyto.a.23082. [DOI] [PubMed] [Google Scholar]
- 29.Owen B.M., Hallett C.S., Cosgrove J.J., Tweedley J.R., Moheimani N.R. Reporting of methods for automated devices: A systematic review and recommendation for studies using FlowCam for phytoplankton. Limnol. Oceanogr. Methods. 2022;20:400–427. doi: 10.1002/lom3.10496. [DOI] [Google Scholar]
- 30.Community Science Institute Database, (n.d.). http://database.communityscience.org/hab_events/450 (accessed August 16, 2023).
- 31.Yu X., Hong J., Liu C., Kim M.K. Review of digital holographic microscopy for three-dimensional profiling and tracking. Opt. Eng. 2014;53 doi: 10.1117/1.oe.53.11.112306. [DOI] [Google Scholar]
- 32.Y.-S. Choi, S.-J. Lee, Three-dimensional volumetric measurement of red blood cell motion using digital holographic microscopy, 2009. [DOI] [PubMed]
- 33.O’Connor T., Santaniello S., Javidi B. COVID-19 detection from red blood cells using highly comparative time-series analysis (HCTSA) in digital holographic microscopy. Opt. Express. 2022;30:1723. doi: 10.1364/oe.442321. [DOI] [PubMed] [Google Scholar]
- 34.D. Pal, Y. Nazarenko, T.C. Preston, P.A. Ariya, Advancing the science of dynamic airborne nanosized particles using Nano-DIHM, Communications Chemistry 2021 4:1. 4 (2021) 1–17. doi: 10.1038/s42004-021-00609-9. [DOI] [PMC free article] [PubMed]
- 35.D. Pal, M. Amyot, C. Liang, P.A. Ariya, Real-time 4D tracking of airborne virus-laden droplets and aerosols, Communications Engineering 2023 2:1. 2 (2023) 1–15. doi: 10.1038/s44172-023-00088-x.
- 36.F. Merola, P. Memmolo, L. Miccio, R. Savoia, M. Mugnano, A. Fontana, G. D’Ippolito, A. Sardo, A. Iolascon, A. Gambale, P. Ferraro, Tomographic flow cytometry by digital holography, Light: Science & Applications 2017 6:4. 6 (2016) e16241–e16241. doi: 10.1038/lsa.2016.241. [DOI] [PMC free article] [PubMed]
- 37.Pirone D., Xin L., Bianco V., Miccio L., Xiao W., Che L., Li X., Memmolo P., Pan F., Ferraro P. Identification of drug-resistant cancer cells in flow cytometry combining 3D holographic tomography with machine learning. Sens Actuators B Chem. 2023;375 doi: 10.1016/J.SNB.2022.132963. [DOI] [Google Scholar]