Abstract
Introduction:
The use of a waterpipe to smoke tobacco has emerged as a popular trend in the United States. Waterpipe smoking establishments have had an increasing presence in the U.S., despite smoke-free air legislation. Dangers of waterpipe smoking have been documented, but less data has been gathered about the waterpipe café itself. This project sought to determine a waterpipe-specific calibration factor (CF) for measuring waterpipe aerosol, and field-test this CF by conducting surveillance on the existing waterpipe cafés of western and central New York.
Methods:
Nine laboratory-controlled experiments were conducted to determine a waterpipe-specific CF. In the lab, two TSI SidePak AM510 Personal Aerosol Monitors and two sampling trains for gravimetric PM2.5 sampling were present during waterpipe smoking sessions (lasting 1–3 hours). Indoor air quality was assessed in 7 waterpipe cafés in three counties of New York, and real-time measurements of particulate matter (PM2.5) and carbon monoxide (CO) were obtained.
Results:
Results from the 9 controlled waterpipe experiments determined a calibration factor of 0.38 (SD 0.08), which should be used to convert SidePak measurements to true PM2.5 measurements. When applying the CF to the measurements taken in the 7 public waterpipe venues, the mean PM2.5 concentration was 515 micrograms per cubic meter (SD=338.8) while the mean ambient CO was 20.5 parts per million (SD=18.3). The mean active smoking density was 2.41 waterpipes per 100 cubic meters of air. The PM2.5 levels increased with increasing active smoking density (rho=0.68, p=0.09).
Conclusions:
Applying the waterpipe-specific CF for the SidePak, 0.38, allowed for field assessments to be conducted in locations with waterpipe smoke to determine accurate particle exposure concentrations. The concentrations of both particulate matter and carbon monoxide were above established air quality standards and therefore increase the health risks of both patrons and workers of these establishments.
Keywords: waterpipe, particulate matter, PM2.5, calibration factor, indoor air quality
1. Introduction
Burning tobacco products are major emitters of respirable suspended particles that are less than 2.5 microns in diameter (PM2.5). These particulates are easily inhaled deep into the lungs and are known to cause numerous adverse cardiovascular and respiratory health effects (Akl et al., 2010; Montazeri et al., 2016; Waziry et al., 2016). In laboratory tests, waterpipe, or hookah, use results in emissions of a wide range of known toxic and carcinogenic compounds including ultrafine to fine particulate matter (Fiala et al., 2012; Zhou et al., 2017), aldehydes (Al Rashidi et al., 2008; Shihadeh et al., 2012), carcinogenic polyaromatic hydrocarbons (Jacob et al., 2011, 2013; Shihadeh and Saleh, 2005), and carbon monoxide (Shihadeh and Saleh, 2005; Shihadeh et al., 2012). When comparing the emissions of cigarettes to waterpipes, the emissions from a single waterpipe use session exceed the emissions from a person smoking two cigarettes within the same time period (Daher et al., 2010). Waterpipes produce four times the acetaldehyde, 27 times the formaldehyde (Al Rashidi et al., 2008), and 38 times the yield of benzo[a]pyrene (Sepetdjian et al., 2008). Carbon monoxide (CO) is emitted at high rates as well, with CO from a single waterpipe session matching the CO of ten cigarette smokers (Daher et al., 2010).
Dangers of waterpipe use are being increasingly documented and include significant associations with lung cancer, respiratory disease, low-birth weight babies, and periodontal disease (Akl et al., 2010; Montazeri et al., 2017; Waziry et al., 2017). Reports of carbon monoxide poisoning associated with waterpipe smoking are also being increasingly documented in the literature (Hojer and Enghag, 2011; Lim et al., 2009; Turkmen et al., 2011). Despite these health hazards associated with hookah use, the number of waterpipe cafés has increased in the United States (Noonan, 2010). Several studies have sought to characterize the levels of particulate matter in these hookah café environments (Cobb et al., 2013; Fiala et al., 2012; Torrey et al., 2015), with mean PM2.5 levels ranging from 374µg/m3 in Virginia cafés (Cobb et al., 2012) to 1419µg/m3 in cafés in Toronto, Canada (Zhang et al., 2015).
The TSI SidePak AM510 Personal Aerosol Monitor (TSI, Inc., Shoreview, MN, USA) is a light-scattering instrument or laser photometer that has proven useful in measuring exposure to cigarette smoke to inform debates over smoke-free air policies and to evaluate the effectiveness of these policies (Connolly et al., 2009; Hyland et al., 2008; Jones et al., 2006; Klepeis et al., 2007; Lee et al., 2008; Lee et al., 2009; Liu et al., 2011; Liu et al., 2010; Maziak et al., 2008; Schneider et al., 2008; Schoj et al., 2010). A protocol for using the SidePak to measure secondhand smoke has been developed and used in over 65 countries around the world (Hyland et al., 2008; Travers, 2006). The SidePak is a portable, battery-operated device using a built-in sampling pump that continuously measures PM2.5. The sampling pump draws air inside and the particulate matter in the air scatters the light from an internal laser. The amount of light scattered is measured and converted into the mass concentration of fine particles (PM2.5). The conversion from light scattered to mass concentration is dependent on the type of particles in the air (e.g. their size and density). Therefore, the SidePak needs to be calibrated for the specific type of aerosol measured. A custom calibration factor of 0.29–0.32 has been determined for the SidePak when measuring cigarette-specific second hand smoke (SHS) to ensure accurate measurements (Jiang et al., 2011; Lee et al., 2007; Travers, 2008b).
Currently, there is no validated TSI SidePak calibration factor for waterpipe smoke, and therefore, one major limitation of previous studies assessing waterpipe aerosol is the inconsistent application of a calibration factor specific to waterpipe smoke. For example, in the five previously published studies we are aware of reporting on waterpipe café aerosol, two of them (Cobb et al., 2012; Zhang et al., 2015) apply the cigarette-specific calibration factor (0.32) to their data, one uses a calibration factor of 0.6 (Torrey et al., 2015), and two do not use a calibration factor at all (calibration factor of 1.0) (Fiala et al., 2012; Zhou et al., 2015). There is then a 3-fold potential difference in the accuracy of reported particulate matter concentrations in these studies. This confusion regarding the application of an aerosol-specific calibration factor when using light-scattering instruments can result in very misleading results. For example, one paper in the literature underestimated true cigarette secondhand smoke PM2.5 levels by 88% by not applying an appropriate calibration factor (Goodman et al., 2007; Travers and Lee, 2008). The lack of a consistent application of a waterpipe-specific calibration factor limits the validity and generalizability of findings. Therefore, the aims of this study were to determine a waterpipe-specific calibration factor for the TSI SidePak by comparing it to reference methods, and apply it to real-time data collected in waterpipe cafés in western and central New York State.
2. Materials and methods
2.1 Calibration factor experiments
The study carried out 9 controlled experiments where participants smoked a single waterpipe (hookah) in a manner of their choosing (ad libidum) in an enclosed room (40 m3) for 1–3 hour sessions. Quick light charcoal was used and the bowl was packed with approximately 12 grams of shisha. Two TSI SidePak AM510 Personal Aerosol Monitors and two pumps with filters for gravimetric PM2.5 sampling were present during smoking.
SidePaks were equipped with 2.5 micron impactors to measure PM2.5. SidePak flow rates were calibrated to 1.7 L/min as required for the 2.5 micron impactor. The device calibration factor was set at 1.0. SidePaks were zero calibrated with a HEPA filter prior to each session. In the 6 experiments with two SidePaks running side-by-side, the mean difference in unadjusted PM2.5 levels was 6.4%. In these experiments the mean of the two SidePaks was used for analysis. The other three experiments used data from a single SidePak due to equipment failure.
Gravimetric PM2.5 was determined using two Leland Legacy Sampling pumps running at 10 L/min and connected to SKC Personal Environmental Monitors (PEM) consisting of a single-stage impactor loaded with pre-weighed 37mm PTFE filters. Large aerosol particles (>2.5 microns) tend to be filtered by the single-stage impactor and collected on the greased ring, while the remaining smaller particles (PM2.5) are collected on the pre-weighed filter paper. Flow rates were calibrated before and after each experiment and were always within 5% of 10 L/min. After sampling was complete, the filter papers were manually removed and sent back to the certified laboratory (Galson Laboratories, Syracuse, NY) to be post-weighed. Duplicate gravimetric samples were obtained for all experiments and the mean of the two was used for all analyses. The mean difference between duplicates was 9.3%. Blanks were included with 25% of samples and all were negative for significant mass changes. Gravimetric PM2.5 concentrations were calculated by dividing the weight of particulate matter collected by the volume of air sampled.
A Kanomax Piezobalance Dust Monitor was also used as a secondary reference instrument in one experiment. The piezobalance is not as portable, rugged, or sensitive as the TSI SidePak. It also has less time resolution and requires near constant user attention. However, the piezobalance can more directly measure the mass of airborne particles and is therefore not subject to the limitations of having to calibrate it for the aerosol in question, as with the light-scattering SidePak. The piezobalance was fitted with a 2.5 micron impactor. Particles enter the piezobalance, become electrically charged and deposit on the piezo-crystal. The total mass of the deposited particulates alters the piezo-crystal’s frequency. Since the change in frequency is proportional to the mass of the particulates, the actual weight of the particulates is obtained.
In addition to the nine waterpipe experiments, nine positive control experiments were conducted with burning cigarettes (Marlboro Reds). Cigarettes were smoked by hand, using a 60cc syringe roughly following ISO standard smoking regimen 3308. All sampling methods were run identically to that of the waterpipe experiments.
2.2 Air quality monitoring in field:
Indoor air quality was assessed in seven hookah bars in three cities (from the different counties) in central and western New York State. These seven venues represented all the hookah establishments that allow indoor smoking in the 3 cities sampled. One venue was excluded from the sample as it only allowed smoking out-of-doors, and there was no smoking taking place (even outdoors) at this establishment due to the season of the study.
The TSI SidePak AM510 Personal Aerosol Monitor was used to sample and record the levels of respirable suspended particles that are less than 2.5 microns in diameter. The SidePak was calibrated to a flow rate of 1.7 liters per minute and was zero-calibrated prior to each use by attaching a HEPA filter according to instructions established by the Air Pollution Exposure Research Laboratory at Roswell Park Cancer Institute (Travers, 2008a). The SidePak uses a built-in sampling pump to continuously draw in air and was set to a thirty-second log interval, which averages the previous 30 one-second measurements. The calibration factor of 0.38 was applied to the raw data retrieved from the SidePak to account for the properties of waterpipe smoke. The calibration factor was determined from the experiments in first part of this study. The data points were then averaged to provide a mean PM2.5 concentration for each establishment.
Sampling was discreet in order to not disturb the patrons’ typical behavior. The SidePak was placed in a handbag with the sampling tube sticking out to conceal the device and minimize its sound. A minimum of 30 minutes was spent in each venue. The number of waterpipes burning, the number of people per waterpipe, and use of other tobacco products were recorded every 15 minutes during sampling. Room dimensions were also measured using a Laser Distance Meter (Leica Geosystems DISTO E7300).
Measurements of indoor carbon monoxide levels were taken using a Q-TRAK Indoor Air Quality Monitor (TSI, Inc., Shoreview, MN, USA). The Q-TRAK measures carbon monoxide (CO), carbon dioxide (CO2), temperature, and humidity. The equipment was set to a thirty-second log interval, and sampling procedures were the same as those described for the SidePak.
Field notes were used to document the presence of the use of other tobacco products, the products sold in the venue (tobacco products, food products, etc.), the type of waterpipe used (single hose versus multiple hose; vented versus non-vented), the type of charcoal used, and the behaviors of patrons in the café.
3. Results
3.1 Calibration factor experiments
Results from the 9 controlled waterpipe experiments yielded a mean calibration factor of 0.38 (SD 0.08, range 0.23 to 0.50) to convert SidePak measurements to true PM2.5 measurements. Good linear correlation (R2 = 0.91) between the SidePak and gravimetric method can be seen across a wide range of concentrations (72 – 1365 µg/m3) in Figure 1. Positive control cigarette experiments yielded a mean cigarette-specific calibration factor for the SidePak of 0.29, the same as previously demonstrated by multiple research groups (Jiang et al., 2011; Lee et al., 2007; Travers, 2008b). In the cigarette experiments, the number of cigarettes smoked by hand was varied to increase the range of expected PM2.5 concentrations. The waterpipe experiments were limited by having a single person smoke the waterpipe ad libidum, and therefore have a smaller PM2.5 range. Across all controlled experiments the temperature ranged from 23.9°C to 26.0°C and the relative humidity ranged from 36% to 58.6%. No adjustments for relative humidity were made as we do not expect humidity below 60% to have a significant impact on the results (Wu, et al., 2005). Applying the 0.38 calibration factor to the SidePak data, seen in Figure 2, showed excellent quantitative and qualitative agreement between the adjusted SidePak data and piezobalance.
3.2 Air quality monitoring in field
An average of 49.75 minutes was spent on data collection at each venue, with a range of 37 to 64 minutes. Table 1 displays the mean characteristics of each venue. Mean PM2.5 concentration ranged from 18 µg/m3 to 1137 µg/m3 and the overall mean for these venues was 515.0 µg/m3 (SD=338.8 µg/m3). All PM2.5 levels are adjusted using the 0.38 waterpipe-specific calibration factor. Ambient carbon monoxide ranged from 1.5 ppm to 53.4 ppm with a mean of 20.5 ppm (SD=18.3 ppm). Real-time data for PM2.5 and carbon monoxide concentrations in each venue sampled can be viewed in Figure 3.
Table 1.
Venue Number | Vented WP | Quick-light charcoal used | Shared hose (Total number hoses) | Shared mouthpiece | Volume (m3) | Mean number of people in venue | Mean number of WPs burning | Mean number of people per WP | Mean number of OTPs burning | WP + OTP Smoker Density* | Mean PM2.5 level (µg/m3) | Mean CO level (ppm) |
---|---|---|---|---|---|---|---|---|---|---|---|---|
| ||||||||||||
1 | Yes | Yes | Yes (1) | Yes | 204 | 35.0 | 10.0 | 3.3 | 0.0 | 4.90 | 617.0 | 37.4 |
2† | Yes | Yes | No (2) | No | 500 | 32.2 | 11.0 | 3.0 | 1.9 | 2.58 | 851.0 | 20.9 |
3 | Yes | Yes | No (2) | No | 143 | 21.0 | 4.3 | 4.8 | 2.0 | 4.39 | 593.0 | 10.7 |
4 | No | Yes | No (2) | No | 306 | 18.3 | 4.7 | 2.8 | 3.0 | 2.52 | 934.0 | 15.0 |
5 | No | Yes | Yes (1) | No | 357 | 6.0 | 1.5 | 2.0 | 0.0 | 0.42 | 18.0 | 1.5 |
6 | No | No | Yes (1) | No | 476 | 20.0 | 6.0 | 3.3 | 0.7 | 1.40 | 436.0 | 52.4 |
7 | Yes | Yes | No (2) | No | 229 | 4.0 | 1.5 | 3.0 | 0.0 | 0.66 | 158.0 | 5.5 |
| ||||||||||||
Overall mean: | 57.1% | 85.7% | 42.9% | 14.3% | 316 | 19.5 | 5.6 | 3.2 | 1.1 | 2.41 | 515.0 | 20.5 |
data averaged across three observations
WP + OTP Smoker Density = Mean number of waterpipes plus other tobacco products per 100 m3 of venue volume.
Note: WP = waterpipe; OTP = other tobacco product; Calibration factor of 0.38 applied to all PM2.5 data
Field notes were collected on a variety of behaviors and are also summarized in Table 1. On average, a waterpipe is shared among 3 persons. Almost half (four out of seven) of the cafés offer patrons their own hose while most (six out of seven) provide participants with their own mouthpiece. All of the venues except for one used quick-light charcoal; the one that did not reported using a “natural” charcoal made of coconut shells that was kept lit on electric coils in the back room. Just over half (57.1%) of the venues provided a waterpipe that was vented, meaning the waterpipe had a purge valve, which allows users to clear smoke from the base.
Of the seven establishments, three identified as a hookah lounge and smoke shop, two identified as a hookah lounge and restaurant (prepared food on premises) and two identified as primarily a hookah lounge. All venues had at least a minimal quantity of shisha available for take-home sale. Patrons smoking a tobacco product (cigarette or little cigar) other than hookah were observed in four of the seven establishments and the smell of marijuana was noted in one venue. All venues provided a menu to select a brand and flavor of shisha; none of these menus provided warnings about the consequences of tobacco smoke.
4. Conclusions
Using the SidePak with the factory calibration setting of 1.00 overestimated gravimetrically-determined PM2.5 levels from waterpipe smoke. Applying the waterpipe-specific calibration factor for the SidePak, 0.38, allowed for field assessments to be conducted in locations with waterpipe smoke to determine accurate particle exposure concentrations. Light-scattering has several advantages over older, traditional methods of particle assessment: it measures and provides real-time data, is cost-effective, battery-operated and portable, and allows for discreet data collection, without affecting the patrons in the environment.
Limitations to this work include the relatively small convenience sample of publicly available hookah venues, though at the time of data collection, this sample accounted for all hookah cafés allowing indoor hookah smoking across three counties.
Another potential limitation and challenge is patrons’ concurrent use of other tobacco products. While bars have historically allowed the use of cigarettes, it was not common to have to take into account the use of other smoked products (tobacco or otherwise) when conducting cigarette air monitoring or other surveillance activities. In this sample of waterpipe cafés, the use of other combustible products was common (4 out of 7), and protocols will have to be developed to properly take this into account. While we suspect the bulk of the PM measured resulted from waterpipe usage, the fact remains that in many of the locations sampled there was likely additional particulate contributions from other burning tobacco products. Further studies should do surveillance to identify exactly what products are being used in waterpipe cafés and examine the impact of mixtures of particulate sources on the device calibration factor.
Waterpipe establishments have had an increasing presence in the United States, and accurate PM2.5 levels determined with the SidePak in public waterpipe venues were high and pose a serious health risk as they exceed existing public health and worker exposure limits. The mean PM2.5 (µg/m3) was well above the cutoff for hazardous air according to the United States EPA’s Air Quality Index. When outdoor air quality reaches levels above 250 µg/m3 (“hazardous”), people are advised to restrict outdoor physical activity while people with heart/lung disease, older adults, and children are advised to remain indoors and keep physical activity levels low (U.S. Environmental Protection Agency, 2013). Additionally, the mean PM2.5 of the waterpipe cafés in this sample is higher than those observed in bars that allowed cigarette smoking. Measurements taken before Clean Indoor Air Laws went into effect suggest an average of 412 µg/m3 in bars or restaurants where smoking was allowed (Center for Disease Control and Prevention, 2004).
With the appropriate calibration factor, the SidePak can now be used by researchers and public health practitioners to gather accurate data on levels of particulate air pollution exposure resulting from waterpipe smoking. Accurate exposure data is a powerful tool to inform debates over smoke-free air policies and to evaluate these policies.
Funding Statement:
This work was supported by Roswell Park Cancer Institute and National Cancer Institute (NCI) grant P30CA016056. The content is solely the responsibility of the authors and does not necessarily represent the official view of the National Institutes of Health.
Footnotes
Declaration of Interests: No conflicts of interest for MJT, JAK, LV
REFERENCES
- Akl EA, Gaddam S, Gunukula SK, Honeine R, Jaoude PA, Irani J, 2010. The effects of waterpipe tobacco smoking on health outcomes: a systematic review. International journal of epidemiology 39, 834–857. [DOI] [PubMed] [Google Scholar]
- Al Rashidi M, Shihadeh A, Saliba NA, 2008. Volatile aldehydes in the mainstream smoke of the narghile waterpipe. Food and chemical toxicology : an international journal published for the British Industrial Biological Research Association 46, 3546–3549. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Center for Disease Control and Prevention, 2004. Indoor air quality in hospitality venues before and after implementation of a clean indoor air law--Western New York, 2003. MMWR Morb Mortal Wkly Rep 53, 1038–1041. [PubMed] [Google Scholar]
- Cobb CO, Vansickel AR, Blank MD, Jentink K, Travers MJ, Eissenberg T, 2012. Indoor air quality in Virginia waterpipe cafes. Tob Control. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cobb CO, Vansickel AR, Blank MD, Jentink K, Travers MJ, Eissenberg T, 2013. Indoor air quality in Virginia waterpipe cafes. Tobacco control 22, 338–343. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Connolly GN, Carpenter CM, Travers MJ, Cummings KM, Hyland A, Mulcahy M, Clancy L, 2009. How smoke-free laws improve air quality: a global study of Irish pubs. Nicotine & tobacco research : official journal of the Society for Research on Nicotine and Tobacco 11, 600–605. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Daher N, Saleh R, Jaroudi E, Sheheitli H, Badr T, Sepetdjian E, Al Rashidi M, Saliba N, Shihadeh A, 2010. Comparison of carcinogen, carbon monoxide, and ultrafine particle emissions from narghile waterpipe and cigarette smoking: Sidestream smoke measurements and assessment of second-hand smoke emission factors. Atmospheric environment (Oxford, England : 1994) 44, 8–14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fiala SC, Morris DS, Pawlak RL, 2012. Measuring indoor air quality of hookah lounges. American journal of public health 102, 2043–2045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goodman P, Agnew M, McCaffrey M, Paul G, Clancy L, 2007. Effects of the Irish smoking ban on respiratory health of bar workers and air quality in Dublin pubs. American journal of respiratory and critical care medicine 175, 840–845. [DOI] [PubMed] [Google Scholar]
- Hojer J, Enghag M, 2011. Carbon monoxide poisoning caused by water pipe smoking. Clinical toxicology (Philadelphia, Pa.) 49, 702–703. [DOI] [PubMed] [Google Scholar]
- Hyland A, Travers MJ, Dresler C, Higbee C, Cummings KM, 2008. A 32-country comparison of tobacco smoke derived particle levels in indoor public places. Tobacco control 17, 159–165. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jacob P 3rd, Abu Raddaha AH, Dempsey D, Havel C, Peng M, Yu L, Benowitz NL, 2011. Nicotine, carbon monoxide, and carcinogen exposure after a single use of a water pipe. Cancer epidemiology, biomarkers & prevention : a publication of the American Association for Cancer Research, cosponsored by the American Society of Preventive Oncology 20, 2345–2353. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jacob P 3rd, Abu Raddaha AH, Dempsey D, Havel C, Peng M, Yu L, Benowitz NL, 2013. Comparison of nicotine and carcinogen exposure with water pipe and cigarette smoking. Cancer epidemiology, biomarkers & prevention : a publication of the American Association for Cancer Research, cosponsored by the American Society of Preventive Oncology 22, 765–772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jiang RT, Acevedo-Bolton V, Cheng KC, Klepeis NE, Ott WR, Hildemann LM, 2011. Determination of response of real-time SidePak AM510 monitor to secondhand smoke, other common indoor aerosols, and outdoor aerosol. Journal of Environmental Monitoring 13, 1695–1702. [DOI] [PubMed] [Google Scholar]
- Jones SC, Travers MJ, Hahn EJ, Robertson H, Lee K, Higbee C, Hyland A, 2006. Secondhand smoke and indoor public spaces in Paducah, Kentucky. J Ky Med Assoc 104, 281–288. [PubMed] [Google Scholar]
- Klepeis NE, Ott WR, Switzer P, 2007. Real-time measurement of outdoor tobacco smoke particles. Journal of the Air & Waste Management Association (1995) 57, 522–534. [DOI] [PubMed] [Google Scholar]
- Lee K, Hahn EJ, Pieper N, Okoli CT, Repace J, Troutman A, 2008. Differential impacts of smoke-free laws on indoor air quality. Journal of environmental health 70, 24–30, 54. [PubMed] [Google Scholar]
- Lee K, Hahn EJ, Riker C, Head S, Seithers P, 2007. Immediate impact of smoke-free laws on indoor air quality. Southern medical journal 100, 885–889. [DOI] [PubMed] [Google Scholar]
- Lee K, Hahn EJ, Robertson HE, Lee S, Vogel SL, Travers MJ, 2009. Strength of smoke-free air laws and indoor air quality. Nicotine & tobacco research : official journal of the Society for Research on Nicotine and Tobacco 11, 381–386. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lim BL, Lim GH, Seow E, 2009. Case of carbon monoxide poisoning after smoking shisha. International journal of emergency medicine 2, 121–122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu R, Hammond SK, Hyland A, Travers MJ, Yang Y, Nan Y, Feng G, Li Q, Jiang Y, 2011. Restaurant and bar owners’ exposure to secondhand smoke and attitudes regarding smoking bans in five Chinese cities. International journal of environmental research and public health 8, 1520–1533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu RL, Yang Y, Travers MJ, Fong GT, O’Connor RJ, Hyland A, Li L, Nan Y, Feng GZ, Li Q, Jiang Y, 2010. A cross-sectional study on levels of second-hand smoke in restaurants and bars in five cities in China. Tobacco control 19 Suppl 2, i24–29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maziak W, Ali RA, Fouad MF, Rastam S, Wipfli H, Travers MJ, Ward KD, Eissenberg T, 2008. Exposure to secondhand smoke at home and in public places in Syria: a developing country’s perspective. Inhalation toxicology 20, 17–24. [DOI] [PubMed] [Google Scholar]
- Montazeri Z, Nyiraneza C, El-Katerji H, Little J, 2016. Waterpipe smoking and cancer: systematic review and meta-analysis. Tob Control. [DOI] [PubMed] [Google Scholar]
- Montazeri Z, Nyiraneza C, El-Katerji H, Little J, 2017. Waterpipe smoking and cancer: systematic review and meta-analysis. Tobacco control 26, 92–97. [DOI] [PubMed] [Google Scholar]
- Noonan D, 2010. Exemptions for hookah bars in clean indoor air legislation: a public health concern. Public health nursing (Boston, Mass.) 27, 49–53. [DOI] [PubMed] [Google Scholar]
- Schneider S, Seibold B, Schunk S, Jentzsch E, Potschke-Langer M, Dresler C, Travers MJ, Hyland A, 2008. Exposure to secondhand smoke in Germany: air contamination due to smoking in German restaurants, bars, and other venues. Nicotine & tobacco research : official journal of the Society for Research on Nicotine and Tobacco 10, 547–555. [DOI] [PubMed] [Google Scholar]
- Schoj V, Sebrie EM, Pizarro ME, Hyland A, Travers MJ, 2010. Informing effective smokefree policies in Argentina: air quality monitoring study in 15 cities (2007–2009). Salud publica de Mexico 52 Suppl 2, S157–167. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sepetdjian E, Shihadeh A, Saliba NA, 2008. Measurement of 16 polycyclic aromatic hydrocarbons in narghile waterpipe tobacco smoke. Food and chemical toxicology : an international journal published for the British Industrial Biological Research Association 46, 1582–1590. [DOI] [PubMed] [Google Scholar]
- Shihadeh A, Saleh R, 2005. Polycyclic aromatic hydrocarbons, carbon monoxide, “tar”, and nicotine in the mainstream smoke aerosol of the narghile water pipe. Food and chemical toxicology : an international journal published for the British Industrial Biological Research Association 43, 655–661. [DOI] [PubMed] [Google Scholar]
- Shihadeh A, Salman R, Jaroudi E, Saliba N, Sepetdjian E, Blank MD, Cobb CO, Eissenberg T, 2012. Does switching to a tobacco-free waterpipe product reduce toxicant intake? A crossover study comparing CO, NO, PAH, volatile aldehydes, “tar” and nicotine yields. Food and chemical toxicology : an international journal published for the British Industrial Biological Research Association 50, 1494–1498. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Torrey CM, Moon KA, Williams DA, Green T, Cohen JE, Navas-Acien A, Breysse PN, 2015. Waterpipe cafes in Baltimore, Maryland: Carbon monoxide, particulate matter, and nicotine exposure. Journal of exposure science & environmental epidemiology 25, 405–410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Travers MJ (2006) Global Air Monitoring Study Indoor Air Monitoring Protocol. Roswell Park Cancer Institute. Roswell Park Cancer Institute, Buffalo, NY. Available from: http://www.tobaccofreeair.org/downloads/GAMS%20Air%20Monitoring%20Protocol%20June%202006.pdf [Google Scholar]
- Travers MJ, 2008a. SidePak Aerosol Monitor Manual. Air Quality Monitor: Technical assistance on monitor preparation and use; and collecting and extracting data for analysis., in: Roswell Park Cancer Institute; (Ed.), Buffalo, NY, pp. 1–48. [Google Scholar]
- Travers MJ, 2008b. Smoke-Free Air Policy: Changing What’s in the Air and in the Body, Social and Preventive Medicine. University at Buffalo, SUNY, Buffalo, NY. [Google Scholar]
- Travers MJ, Lee K, 2008. Particulate air pollution in Irish pubs is grossly underestimated. American journal of respiratory and critical care medicine 177, 236–237; author reply 237–238. [DOI] [PubMed] [Google Scholar]
- Turkmen S, Eryigit U, Sahin A, Yeniocak S, Turedi S, 2011. Carbon monoxide poisoning associated with water pipe smoking. Clinical toxicology (Philadelphia, Pa.) 49, 697–698. [DOI] [PubMed] [Google Scholar]
- U.S. Environmental Protection Agency, 2013. National ambient air quality standards (NAAQS): Particulate Matter (PM) Standards. [Google Scholar]
- Waziry R, Jawad M, Ballout RA, Al Akel M, Akl EA, 2017. The effects of waterpipe tobacco smoking on health outcomes: an updated systematic review and meta-analysis. International journal of epidemiology 46, 32–43. [DOI] [PubMed] [Google Scholar]
- Wu CF, Delfino RJ, Floro JN, Samimi BS, Quintana PJ, Kleinman MT, Liu LJ, 2005. Evaluation and quality control of personal nephelometers in indoor, outdoor, and personal environments. Journal of Exposure Analysis and Environmental Epidemiology 15(1), 99–110. [DOI] [PubMed] [Google Scholar]
- Zhang B, Haji F, Kaufman P, Muir S, Ferrence R, 2015. ‘Enter at your own risk’: a multimethod study of air quality and biological measures in Canadian waterpipe cafes. Tobacco control 24, 175–181. [DOI] [PubMed] [Google Scholar]
- Zhou S, Behrooz L, Weitzman M, Pan G, Vilcassim R, Mirowsky JE, Breysee P, Rule A, Gordon T, 2017. Secondhand hookah smoke: an occupational hazard for hookah bar employees. Tobacco control 26, 40–45. [DOI] [PubMed] [Google Scholar]
- Zhou S, Weitzman M, Vilcassim R, Wilson J, Legrand N, Saunders E, Travers M, Chen LC, Peltier R, Gordon T, 2015. Air quality in New York City hookah bars. Tobacco control 24, e193–198. [DOI] [PMC free article] [PubMed] [Google Scholar]