Life cycle and post-mortem ingrowth patterns of a leadless pacemaker system

Hermann Blessberger 1,2,, Daniel Kiblboeck 3,4, Claudia Grosse 5,6, Petar Noack 7,8, Jakob Ebner 9,10, Jakob Boetscher 11,12, Julian Maier 13,14, Stefan Schwarz 15,16, Christian Reiter 17,18, Alexander Fellner 19,20, Michael Lichtenauer 21, Joerg Kellermair 22,23, Juergen Kammler 24,25,26, Karim Saleh 27,28, Clemens Steinwender 29,30,31,b
PMCID: PMC10869213  PMID: 38242703

What’s new?

  • About 45% of our patients with a leadless pacemaker system (LPS) died from cardiac causes. In our study there were no signs of pacemaker-induced arrhythmic death.

  • Encapsulation of the LPS varied and was not time-dependent.

  • The histological analysis of eight post-mortem tissue samples revealed a fibrous capsule with varying degrees of lymphocyte-rich inflammation adjacent to the LPS body without any signs of endothelialization.

Leadless pacemaker systems (LPS) have become an established therapy in selected patients. However, there is still debate about the feasibility and method of device retrieval after battery depletion or in cases of systemic endocarditis where device infection cannot be ruled out, especially years after initial implantation.1 To date, data are still limited on how long patients wear their LPS, what causes their death, and the extent to which LPS become overgrown with cardiac tissue over time. Overgrowth may affect the feasibility of retrieval and electrical device performance. In the future, it may become possible to precisely control tissue coverage by applying particular cell- or fibrosis-repellent surface coatings or treatments.2

In this retrospective single-centre cohort study, all patients who had undergone a Micra™ LPS (Medtronic Inc., Minneapolis, USA) implantation at our hospital were included. The study was approved by the local ethics committee. Patient survival status was determined by contacting local registration authorities, and the cause of death was obtained from the Austrian National Statistical Institute and by reviewing autopsy reports. The ‘service period’ was defined as the time from LPS implantation to either the last follow-up or the patient’s death. A thorough pathohistological examination including immunohistological stains was performed by two experienced pathologists when post-mortem myocardial tissue blocks containing the LPS were available.

The study cohort comprised 283 patients (LPS implant between December 2013 and July 2020, average age 79.2 years, 36.4% female), of whom 71 (25.1%) died during a median follow-up of 2.7 (IQR: 1.5–4.4, range: 0–7.1) years. The indications for leadless pacing were atrial fibrillation with slow conduction (41.3%), high-grade AV block (33.6%), sick sinus syndrome with a low anticipated ventricular pacing rate (14.5%), and others (10.6%). The median age at death was 85.9 years (IQR: 79.9–89.7, range: 61.8–99.1 years). The median service period was 1.9 years (IQR: 1.1–3.1, range: 0–5.7 years). Death was due to cardiac in 32 (45.1%) and non-cardiac conditions in 39 (54.9%) patients. Cardiac causes comprised: acute coronary syndrome (n = 6), coronary heart disease (n = 11), heart failure (n = 2), ischaemic cardiomyopathy (n = 2), aortic stenosis (n = 1), and other cardiac diseases (n = 10). Non-cardiac conditions included different types of cancer (n = 12), renal disease (n = 6), neurological conditions (n = 5), chronic obstructive pulmonary disease (n = 1), infections other than device infection (n = 6), as well as others (n = 9). No death was classified as device-related or as arrhythmic death that might have been associated with leadless pacing. During follow-up, LPS dysfunction was a rare event (three cases of raised pacing thresholds, all within three months after implant), and two LPS could be successfully retrieved (each about one month after the initial implantation). No LPS infections were noted in our cohort, and no relapses were detected in patients with LPS implantation after previous lead extraction due to device infection. The histological analysis of eight (11.3%) post-mortem tissue samples (service period: 18–2083 days, age 75–87 years) revealed a fibrous capsule with varying degrees of lymphocyte-rich inflammation adjacent to the LPS body and its tines (Table 1). The macroscopic encapsulation of the pacemaker was complete in two cases and partial (not covering the tail with the retrieval knob) in the other cases, and not associated with the length of the service period (Figure 1). The maximum thickness of the fibrous cast ranged between 465.1 µm and 1195.8 µm. CD34 immunohistochemistry showed that no endothelial cells covered the surface of the collagenous cast ‘membrane’. A brush lavage from the surface of the LPS in two patients (#6 and #7) confirmed this finding.

Table 1.

Histological features of eight autopsy cases

Case number #1 #23 #3 #4 #5 #6 #7 #8
Baseline characteristics and macroscopic features
Age at death (years) 76.7 83.3 75.7 75.6 86.1 86.3 87.4 83.9
Sex Male Male Female Male Male Female Male Male
Indication Afib with slow conduction Afib with slow conduction Afib with slow conduction SSS Afib with slow conduction AVB III AVB III AVB III
 Length of service period in years (days) 1.6 (576) 1.0 (355) 1.9 (697) 1.1 (403) 0.6 (228) 5.7 (2083) 0.05 (18) 4.1 (1498)
Cause of death Myocardial infarction Renal disease, pulmonary fibrosis Ovarian cancer Liver cirrhosis Renal disease Ischaemic cardiomyopathy Pulmonary disease, conventional device infection with sepsis Prostate cancer
Macroscopic tissue coverage of LPS In part Total Total In part In part In part In part In part
Neointima No No No No No No No No
Location of LPS Apical Apical Septal Apical Septal Apical Septal Septal
Histological features of the cast
Fibrosis ++ ++ +/++ ++ ++ ++ + ++
Inflammatory infiltrate ++ ++/+++ + ++ ++ + −/+ +++
CD3 pos. lymphocytes ++ +++ + ++ ++ + +/− +++
CD20 pos. lymphocytes −/+ + −/+ −/+ + + −/+ +
Neutrophils + −/+ −/+ −/+
CD34 staining
Fibrin exudates + ++ +++ +++ ++ ++
Vascular proliferation ++ +++ +/− +/++ +++ + −/+ +
Haemosiderin ++ ++ −/+ +++
Maximal thickness (µm) 974.3 916.1 485.8 931.2 922.1 1195.8 465.1 823.4
Miscellaneous None Focal myocardial scarring next to LPS None None Amyloidosis, focal myocardial scarring next to LPS Small calcifications None Focal myocardial scarring next to LPS

+, mild; ++, moderate; +++, marked; −, not present.

LPS, leadless pacemaker system; SSS, sick sinus syndrome.

Figure 1.

Figure 1

Macroscopic ingrowth patterns and microscopic views of the tissue surrounding the Micra™ LPS. (A) Cast totally covering the LPS (Patient #3, service period 697 days). (B) Cast only partially covering the LPS (Patient #1, service period 576 days). Stained sections of the tissue covering the device (LPS-tissue interface) with a deep layer of dense inflammatory cells (lymphocytes) covered by a capsule of collagenous tissue (cast, C and D). (E and F) Tissue reaction at the fixation tines. (C) Haematoxylin/eosin stain. Fibrous cast capsule with small blood vessels and lymphocytes. (D) Immunohistological stain for CD3 (T-lymphocytes). (E) Haematoxylin/eosin stain. Fibrous tissue reaction surrounding a fixation tine. (F) Elastica-Van-Gieson stain. Collagen red, myocardium yellow. LPS, leadless pacemaker system.

Our indications for LPS therapy were in agreement with current guidelines.4,5 Compared with previously published data, the initial implantation success rate (99.6% in our cohort vs. 99.2%6), the 30-day complication rate (3.5% in our cohort vs. 2.9%6), and the long-term device-related complication rate (4.6% in our cohort vs. 4.6%7) were within the same range. Despite the significantly longer follow-up in our study (median 2.7, up to 7.1 years), all-cause mortality in our cohort (25.1%) was lower than in the Micra™ CED study (34%), which had a follow-up of only two years.7 This may be explained—at least in part—by a lower burden of comorbidities in our set of patients, as reflected by a lower Charlson index (4.5 ± 1.6 points in our cohort vs. 5.1 ± 3.47). However, the proportion of causes of death in our cohort (cardiac: 45.1%, non-cardiac: 54.9%) was quite comparable to the 12-month results of the Micra™ Transcatheter Pacing Study (cardiac: 41.6%, non-cardiac: 55.8%, unknown: 2.6%).6 In both our cohort and in the Micra™ LPS 12-month follow-up study, one death occurred during the index stay that was not related to the LPS implantation procedure (due to sepsis in our cohort).

The fibrotic reaction seen in our histopathological analyses is most likely due to direct implantation trauma and mechanical irritation of the right ventricular wall and trabeculae by the LPS.8,9 It has been shown in vitro that IL-1β and IL-6, which act on fibroblasts and endothelial cells, are important for mediating this tissue reaction.10 In our cohort, maximum cast thickness and extent of the macroscopic overgrowth were unpredictable and not associated with the length of the service period. Conversely, they probably depend more on the individual predisposition to scar formation, which is underlined by our observation that the LPS is either already fully encapsulated one year after implantation3 or is not yet completely covered by tissue after 5.7 years. In a porcine model, after 12 weeks, the extent of fibrosis was significantly less with the Micra™ LPS than with conventional Medtronic™ 5076 leads.11 Micra™ retrievals described in the literature were feasible up to four years after implantation but were technically demanding in some cases.12–15 We could not identify endothelial lining with CD34 staining at the device-tissue interface. Such an ‘acellular interface’ was also found in an analysis of collagen-rich connective tissue surrounding standard pacemaker leads.8 In addition to the fibrotic cast, we found T-cell-rich inflammation with fibrin exudates, haemosiderin, and vascular proliferation adjacent to the LPS, indicating wound healing, scarring, and an individual reaction to foreign material.

Contributor Information

Hermann Blessberger, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Daniel Kiblboeck, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Claudia Grosse, Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria; Institute of Clinical Pathology and Molecular Pathology, Kepler University Hospital, Linz, Austria.

Petar Noack, Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria; Institute of Clinical Pathology and Molecular Pathology, Kepler University Hospital, Linz, Austria.

Jakob Ebner, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Jakob Boetscher, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Julian Maier, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Stefan Schwarz, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Christian Reiter, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Alexander Fellner, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Michael Lichtenauer, Department of Internal Medicine II, Paracelsus Medical University, Salzburg, Austria.

Joerg Kellermair, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Juergen Kammler, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria; Department of Internal Medicine II, Paracelsus Medical University, Salzburg, Austria.

Karim Saleh, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria.

Clemens Steinwender, Department of Cardiology, Kepler University Hospital, Krankenhausstraße 9, 4020 Linz, Austria; Medical Faculty of the Johannes Kepler University Linz, Altenberger Straße 64, 4040 Linz, Austria; Department of Internal Medicine II, Paracelsus Medical University, Salzburg, Austria.

Data availability

The data presented in this study will be shared upon reasonable request to the corresponding author.

References

  • 1. Koay A, Khelae S, Wei KK, Muhammad Z, Mohd Ali R, Omar R. Treating an infected transcatheter pacemaker system via percutaneous extraction. HeartRhythm Case Rep 2016;2:360–2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Heitz J, Plamadeala C, Muck M, Armbruster O, Baumgartner W, Weth Aet al. Femtosecond laser-induced microstructures on Ti substrates for reduced cell adhesion. Appl Phys A 2017;123:734. [Google Scholar]
  • 3. Kypta A, Blessberger H, Lichtenauer M, Steinwender C. Complete encapsulation of a leadless cardiac pacemaker. Clin Res Cardiol 2016;105:94. [DOI] [PubMed] [Google Scholar]
  • 4. Boersma LV, El-Chami M, Steinwender C, Lambiase P, Murgatroyd F, Mela Tet al. Practical considerations, indications, and future perspectives for leadless and extravascular cardiac implantable electronic devices: a position paper by EHRA/HRS/LAHRS/APHRS. Europace 2022;24:1691–708. [DOI] [PubMed] [Google Scholar]
  • 5. Steinwender C, Lercher P, Schukro C, Blessberger H, Prenner G, Andreas Met al. State of the art: leadless ventricular pacing: a national expert consensus of the Austrian Society of Cardiology. J Interv Card Electrophysiol 2020;57:27–37. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JMet al. Long-term performance of a transcatheter pacing system: 12-month results from the Micra Transcatheter Pacing Study. Heart Rhythm 2017;14:702–9. [DOI] [PubMed] [Google Scholar]
  • 7. El-Chami MF, Bockstedt L, Longacre C, Higuera L, Stromberg K, Crossley Get al. Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. Eur Heart J 2022;43:1207–15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Keiler J, Schulze M, Sombetzki M, Heller T, Tischer T, Grabow Net al. Neointimal fibrotic lead encapsulation—clinical challenges and demands for implantable cardiac electronic devices. J Cardiol 2017;70:7–17. [DOI] [PubMed] [Google Scholar]
  • 9. Stokes K, Anderson J, McVenes R, McClay C. The encapsulation of polyurethane-insulated transvenous cardiac pacemaker leads. Cardiovasc Pathol 1995;4:163–71. [DOI] [PubMed] [Google Scholar]
  • 10. Edlinger C, Paar V, Tuscher T, Jirak P, Motloch LJ, Kammler Jet al. Potential mechanisms of endothelialisation in individuals implanted with a leadless pacemaker systems: an experimental in vitro study. J Electrocardiol 2019;55:72–7. [DOI] [PubMed] [Google Scholar]
  • 11. Bonner M, Eggen M, Haddad T, Sheldon T, Williams E. Early performance and safety of the Micra transcatheter pacemaker in pigs. Pacing Clin Electrophysiol 2015;38:1248–59. [DOI] [PubMed] [Google Scholar]
  • 12. Kiani S, Merchant FM, El-Chami MF. Extraction of a 4-year-old leadless pacemaker with a tine-based fixation. HeartRhythm Case Rep 2019;5:424–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Bhatia NK, Kiani S, Merchant FM, Delurgio DB, Patel AM, Leon ARet al. Life cycle management of Micra transcatheter pacing system: data from a high-volume center. J Cardiovasc Electrophysiol 2021;32:484–90. [DOI] [PubMed] [Google Scholar]
  • 14. Dar T, Akella K, Murtaza G, Sharma S, Afzal MR, Gopinathannair Ret al. Comparison of the safety and efficacy of Nanostim and Micra transcatheter leadless pacemaker (LP) extractions: a multicenter experience. J Interv Card Electrophysiol 2020;57:133–40. [DOI] [PubMed] [Google Scholar]
  • 15. Kawasaki M, Yamada T, Watanabe T, Furukawa Y, Fukunami M. Extraction of a “floating leadless pacemaker”: a case report. HeartRhythm Case Rep 2020;6:457–9. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data presented in this study will be shared upon reasonable request to the corresponding author.


Articles from Europace are provided here courtesy of Oxford University Press