Simultaneous tensile and shear measurement of the human cornea in vivo using S0- and A0-wave optical coherence elastography

Guo-Yang Li 1,†,#, Xu Feng 1,, Seok-Hyun Yun 1,2,*
PMCID: PMC10872441  NIHMSID: NIHMS1952041  PMID: 38101555

Abstract

Understanding corneal stiffness is valuable for improving refractive surgery, detecting corneal abnormalities, and assessing intraocular pressure. However, accurately measuring the elastic properties, specifically the tensile and shear moduli that govern mechanical deformation, has been challenging. To tackle this issue, we have developed guided-wave optical coherence elastography that can simultaneously excite and analyze symmetric (S0) and anti-symmetric (A0) elastic waves in the cornea at around 10 kHz frequencies, enabling us to extract tensile and shear properties from measured wave dispersion curves. We verified the technique using elastomer phantoms and ex vivo porcine corneas and investigated the dependence on intraocular pressure using acoustoelastic theory that incorporates corneal tension and a nonlinear constitutive tissue model. In a pilot study involving six healthy human subjects aged 31 to 62, we measured shear moduli (Gzx) of 94±20 kPa (mean±standard deviation) and tensile moduli (Exx) of 4.0±1.1 MPa at central corneas. Our preliminary analysis of age-dependence revealed contrasting trends: −8.3±4.5 kPa/decade for shear and 0.30±0.21 MPa/decade for tensile modulus. This OCE technique has the potential to become a highly useful clinical tool for the quantitative biomechanical assessment of the cornea.

Keywords: Cornea, nonlinear anisotropy, optical coherence elastography, tissues in vivo

Graphical Abstract

graphic file with name nihms-1952041-f0001.jpg

1. Introduction

The mechanical properties of the cornea play an essential role in establishing biomechanical homeosis with intraocular pressure (IOP) and maintaining normal corneal shapes. Measuring corneal biomechanics is significant in various aspects of corneal health and disease management, including the development of diagnostic metrics for early detection of keratoconus, monitoring corneal crosslinking, and accurately predicting corneal shapes after refractive surgeries.

The stromal tissue of the cornea has a lamellar microstructure consisting of the collagen fiber network and interfibrillar matrix [13]. This unique microstructure make the cornea mechanically anisotropic and approximately transverse isotropic [4]. The tensile modulus along the corneal plane is predominantly influenced by the tensile modulus of the collagen fibrils, whereas the out-of-plane shear modulus is largely dictated by the shear modulus of the extrafibrillar matrix. Therefore, measurements for both tensile and shear moduli would be essential for an accurate characterization of corneal biomechanics.

However, current techniques have severe limitations in measuring tensile and shear moduli in vivo. Mechanical tools, such as stress-strain tests and torsional tests, are invasive and not easily configurable for in vivo measurements. Two commercial instruments, namely the Ocular Response Analyzer (Reichert) [5] and Corvis ST (Oculus) [6, 7], provide phenomenological, biomechanical indices through the inverse analysis of corneal deformation against air-puffs but do not explicitly generate elastic modulus information. Corneal indentation can, in principle, measure tensile modulus, but with compromised accuracy [8, 9]. Brillouin microscopy can measure longitudinal modulus with high resolution [10, 11], but this elastic property is not directly related to tensile and shear properties required to describe corneal deformation. Ultrasound elastography [12] and optical coherence elastography (OCE) [1317] have been applied to the cornea in vivo. To date, these techniques have employed almost exclusively the antisymmetric (A0) elastic wave, whose velocity is predominantly governed by out-of-plane shear modulus [18]. In-plane tensile moduli of the cornea has been estimated from the velocity dispersion [18] or the displacement profile [19] of the A0 wave, but with relatively large fitting uncertainties.

In this work, we demonstrate a noninvasive elastography method to quantify both the tensile and shear moduli of the cornea using guided elastic waves. We employ a high-frequency OCE technique to simultaneously generate symmetric (S0) as well as antisymmetric A0 guided waves in the cornea and measure their propagation speeds. We adopt an acoustoelastic wave model that incorporates IOP-induced tension, tissue anisotropy, and nonlinearities. We then obtain tensile and shear moduli directly by fitting the model to the measured dispersion curves of the S0 and A0 waves. After testing this technique using phantoms and porcine corneas, we apply it to in vivo human corneas.

2. Theory of Guided Elastic Waves

2.1. Mechanical model of the cornea

The deformation of tissue in response to mechanical stress is governed by its tensile and shear properties. The stroma dominates the overall stiffness of the cornea, accounting for approximately 90% of its total thickness [20]. The stroma consists of 200–300 lamellar layers containing collagen fibrils aligned along the layers (Fig. 1A), which are typically orthogonally stacked between adjacent lamellas [21]. The collagen fibers bear the tensile stress induced by the IOP [22, 23]. The in-plane tensile stress σ can be estimated using the Young-Laplace equation:

σ=IOP×R/(2h), (1)

where R and h denote the radius and thickness of the cornea, respectively (Fig. 1B).

Fig. 1. Mechanical model for elastic wave motion in the cornea.

Fig. 1.

(A) (left) OCT image of a healthy human volunteer and (right) electron micrographs of a porcine corneal tissue showing the microstructure of the stroma layer. (B) Static load of the cornea in physiological condition and the elastic wave motion. The fundamental guided modes are asymmetric (A0) and symmetric (S0), respectively. Boxes illustrate specific tissue deformations involved in the wave motion. (C) The coordinate system used in the model. (D) Stress σ and stretch ratio λ of the cornea under different IOP levels. The normal physiological range of the IOP is typically 10–20 mmHg (gray area). (E) Theoretical dispersion relations of the A0 and S0 modes at zero IOP (solid lines) and 15 mmHg (dashed lines). The markers show the results obtained from finite element analysis (FEA). (F) Effect of the IOP on the phase velocities of A0 (10 kHz) and S0 (4 kHz) modes for a weakly anisotropic material with R=7.5 mm, h=0.53 mm, μ=30kPa, k1=50kPa, k2=200, and κ=0.

Given the corneal microstructure, it is reasonable to adopt the Holzapfel-Gasser-Ogden (HGO) model [24], which has been widely used for mechanical modeling of arterial walls. The strain energy function of the HGO model can be written as:

W=μ2I1-3+k1k2i=12ek2κI1-3+(1-3κ)I4i-12-1, (2)

where μ ,k1 ,k2 and κ are constitutive parameters. μ represents the initial shear modulus, k1 represents the anisotropic tensile response, and k2 describes the nonlinear stiffening effect of collagen fibrils at large strain. κ characterizes the dispersion of collagen fibrils. 0κ1/3, where κ=0 is for unidirectionally aligned fibrils and κ=1/3 corresponds to randomly distributed fibrils. According to previous studies [2526], the collagen fibrils of central corneas are approximately aligned in two orthogonal directions along the superior-inferior and nasal-temporal axes, giving κ values lower than 0.13. In our study, we assumed κ=0 for simplicity. The influence of corneal curvature on wave speed is negligible (see Supplementary Fig. S1). Therefore, we can use Cartesian coordinates x1,x2,x3 as shown in Fig. 1C. The axes of the collagen fibrils are denoted by M=(1,0,0)T and M'=(0,0,1)T. The invariants in Eq. (2), I1,I41 and I42, can be defined by M, M', and the deformation gradient F:I1=trFTF, I41=(FM)(FM), I42=FM'FM [24]. The Cauchy stress tensor can be determined using σij=FiIW/FjI-pδij, where p is a Lagrange multiplier for material incompressibility, δij is the Kronecker delta, and the Einstein summation convention is used for I{1,2,3}. For biaxial stress σ, which corresponds to in-plane corneal tension, the deformation strain tensor is F=diagλ,λ-2,λ, where λ is the stretch ratio along x1 and det(F)=1 due to the incompressibility of tissue. λ is obtained from

σ=F1IWF1I-F2IWF2I. (3)

To estimate corneal tension, we consider a typical corneal radius of curvature (R) of 7.5 mm and a corneal thickness (h) of 0.53 mm. For illustration, we considered previously reported values of μ=60kPa, k1=50kPa, k2=200, and κ=0 [25], although we later find that this k1 value is underestimated for human corneas. Figure 1D illustrates the variations of stress σ and stretch ratio λ for a range of IOP from 0 to 40 mmHg. Due to the nonlinear mechanical properties, the stiffness of the cornea increases with tensile strain, and the stretch ratio λ exhibits a decreasing slope as IOP increases.

We focus on the elastic waves guided in the cornea. In our OCE experiments, the vibration amplitudes of the elastic waves are only around 100 nm. Thus, the wave motion can be considered an incremental, linear perturbation on the prestressed configuration of the cornea, governed by the wave equation:

α4ψx14+2β4ψx12x22+γ4ψx24=ρ4ψx12t2+4ψx22t2, (4)

where ρ is the density, ψ is related to the displacement components induced by wave motion (u1 and u2 as u1=ψ/x2 and u2=-ψ/x1. The coefficients α,β, and γ are determined by the strain energy function and λ (see Refs. [27, 28] and Supplementary Note 1). The upper boundary (epithelium) interfacing with the air is stress-free, while the lower boundary (endothelium) is in contact with the aqueous humor. By solving Eq. (4) together with the linear acoustic equation of fluid (aqueous humor) and considering the appropriate boundary conditions, we obtain a secular equation that describes the wave velocity dispersion of the guided waves in the cornea (Supplementary Note 1).

2.2. The guided waves and elastic moduli

According to the Lamb wave theory [29, 30], the plate geometry of the cornea supports two fundamental guided waves with symmetric (S0) and antisymmetric (A0) motions, respectively. The S0 mode is an extensional, dilatational wave with its velocity largely governed by tensile modulus along the propagation direction. The A0 mode is a flexural, bending wave with its velocity governed by out-of-plane shear modulus. We use Gzx and Exx to denote the out-of-plane shear modulus and in-plane tensile modulus, respectively, under the prestress condition set by the corneal tension. Under a stress-free condition (i.e., zero corneal tension, σ=0), Gzx reduces to an intrinsic shear modulus Gzxσ=0(=μ), and Exx reduces to intrinsic Young’s modulus Exxσ=0=4μ+4k1 (see Supplementary Note 2). For isotropic materials k1=0,Exxσ=0 is equal to 4μ. We introduce a ratio, Exx/4Gzx as an anisotropy index. For isotropic and stress-free materials, the ratio is equal to 1, but Exx/4Gzx1 is expected for highly anisotropic materials, such as the cornea.

Figure 1E shows the theoretical dispersion curves of the A0 and S0 waves as a function of frequency for two different IOP levels of 0 and 15 mmHg, respectively. At zero frequency, the velocity of the A0 wave is zero (even at 15 mmHg due to the presence of the aqueous humor). At 0 mmHg IOP, the S0 wave speed in the low frequency limit is equal to Exxσ=0/ρ. (see Supplementary Note 2). For the highly anisotropic corneal tissue, the plane shear modulus Exxσ=0 is nearly equivalent to uniaxial tensile modulus or Young’s modulus. On the other hand, in the high frequency regime where the wavelengths become shorter than the corneal thickness, both the A0 and S0 velocities converge to plateaus. At zero tension, the asymptotic velocities are close to μ/ρ (see Supplementary Note 2). The corneal tension increases the high-frequency asymptotic plateaus through its effects on the material properties and acoustic deformation (see Supplementary Note 2).

For illustration, in Fig. 1F we plot the phase velocities of the A0 mode at 10 kHz and the S0 mode at 4 kHz as a function of the IOP level for the modestly anisotropic material parameters (μ= 60 kPa, k1=50kPa, k2=200). We find that the phase velocity of A0 mode varies approximately linearly with IOP with a slope of ~0.04 m/smmHgr2>0.99. On the other hand, the S0 wave speed is nearly insensitive to IOPs lower than 10 mmHg, but increases with IOPs greater than 15 mmHg with a slope of ~0.91 m/smmHgr2>0.99. The S0 wave is more sensitive than the A0 wave to IOP in the physiological range because of the large exponential nonlinearity of tensile elasticity against the corneal tension. It should be noted that the tension affects the wave velocities not only through the nonlinear material properties but also directly through the tension-induced restoring forces on the wave displacements. The acoustoelastic analysis using α,β, and γ effectively decomposes the two mechanisms and allows us to extract the elastic moduli from the wave velocities (Supplementary Note 1).

3. Methods

3.1. OCE system

We used a custom-built, swept-source phase-sensitive optical coherence tomography (OCT) imaging system [13]. The OCT system has a center wavelength of 1300 nm and a bandwidth of 80 nm at an A-line rate of 43.2 kHz, offering an axial resolution of ~ 15 μm. The illumination power on the cornea is below 10 mW in compliance with the ANSI Z136.1–2014 safety standard. Trigger pulses generated from a narrowband fiber-optic Bragg grating (i.e., lambda trigger) ensures time synchronization for the OCT data acquisition, beam scanning, and wave generation. The displacements of the guided waves were measured using the method previously described [13, 30]. Briefly, continuous pure tone stimuli were used with frequencies varied from 2 to 16 kHz with an interval of 2 kHz. At each transverse location of the optical beam, 172 A-lines were acquired, which constitutes a single M-scan data. A total of 96 transverse positions along the sample are scanned. The data acquisition time was about 0.4 s at each frequency.

We extracted displacement profiles over time t, and then performed a Fourier transform to move the data from time t domain to frequency f domain. The phase and amplitude of the mechanical vibration at the driving frequency were determined. This lock-in detection effectively reduces noises and eliminates any phase changes by subject’s motion in other frequencies outside the lock-in bandwidth (250 Hz) around the excitation frequency. To calculate the surface wave speeds, we obtained 1-dimensional displacement profiles from the top surface of the sample, and applied a Fourier transform to measure the wavenumber k of each wave. Two most prominent peaks in the kx domain were identified using a global search algorithm and assigned to the A0 and S0 waves. Only one primary peak associated with the A0 wave could appear at low frequencies. This analysis in the spatial frequency domain is critical to remove other higher-order modes especially at high frequencies [31]. The phase velocity is determined by v=2πf/k. To minimize the effect from residual corneal vibration from the prior frequency, the first A-line in the M-scan was ignored in post-processing. The surface ripple artifacts were eliminated in post data processing using the approach detailed in Ref [32].

The noise equivalent amplitude sensitivity was measured to be ~4 nm per A-line, or 19 pm/Hz, for an optical SNR of 30 dB at the surface of sample. With the 172 A-line M-mode lock-in integration (over 4 ms or 250 Hz), the effective sensitivity becomes 300 pm. The ultimate OCE sensitivity is further enhanced by the spatial analysis over 96 locations. Without wave attenuation, the maximum sensitivity would be ~ 31 pm. The typical vibration amplitude used in the experiments was larger than 20 nm. The resulting high signal to noise ratio allows the wavenumber to be determined with high precision (<< 1%).

3.2. Optimization of the wave-excitation probe

To generate both the A0 and S0 waves in the cornea, we utilized a custom-designed vibrating probe consisting of a piezoelectric transducer (PZT) and a probe tip, as depicted in Fig. 2B. We tested different tilt angles of the probe (α=0, 15, and 30 degrees) and found that the highest A0 amplitude at α=0 deg and the lowest A0 amplitude at α=30 deg. On the other hand, the S0 amplitude was the maximum at α=15 deg (Supplementary Fig. S3B). Considering that S0 wave dispersion data above 10 kHz are critical, we selected the tilt angle of 15 degrees and employed this configuration for all subsequent experiments.

Fig. 2. Verification using elastomer phantoms.

Fig. 2.

(A) Schematic of the experimental setup including the contact piezoelectric probe with a tilt angle α between the vibration direction and the surface normal. (B) Picture of the contact probe. (C) Picture of the PDMS sheet phantom on an artificial anterior chamber. (D) Displacement maps at 4 kHz and 12 kHz (IOP = 0 mmHg). (E) Corresponding surface displacement profiles. Only the real parts are shown. (F) Fourier transformations of the surface displacement profiles to measure the dispersion relations. The S0 mode is clearly identified at 6 –16 kHz. (G) Phase velocity dispersion relations and the theoretical model. (H) Effect of IOP on the A0 wave velocity at 16 kHz. The line represents the theoretical fit based on the neo-Hookean model k1=0.

Using the Hertz model of elastic contact, we calculated the contact length of the tip on the corneal surface to be ~ 520 μm. To verify this, we put ink on the tip of the probe and measured the length of the ink transfer mark to a phantom. Our result was approximately 500 μm in good agreement with the theoretical prediction. According to the Hertz model, the indentation depth of the probe tip onto the corneal tissue should be around 100 μm.

3.3. Statistical Analysis

We report the wave speed data as mean ± standard deviation (std) for the phantom, ex vivo porcine corneas, and in vivo human corneas. To highlight the inter-sample variation of the porcine samples, we report the data as boxplots indicating the median, the 25th percentile (Q1) and the 75th percentile (Q3). The whisker extends from the hinge to the largest or smallest value no further than 1.5 times the interquartile range of the hinge. The 95% confidence interval, mean and SD were calculated for the in vivo human data across the age span.

4. Results

4.1. Validation using phantoms

To validate the OCE method, we conducted experiments using a 0.45-mm thick polydimethylsiloxane (PDMS) sheet, which is an isotropic material. The PDMS sheet was mounted in an artificial anterior chamber, as illustrated in Fig. 2C. Initially, no pressure or in-plane stress was applied to the PDMS sheet. Displacement maps and surface displacement profiles were obtained at 4 kHz and 12 kHz, as shown in Figs. 2D and 2E, respectively. In the wavenumber domain (Fig. 2F), two distinct peaks were observed. The peak at the lower wavenumber corresponded to the S0 mode with a higher phase velocity, while the peak at the higher wavenumber corresponded to the A0 mode with a lower velocity. The A0 wave was clearly detected across all frequencies ranging from 2 to 16 kHz. However, the S0 wave was reliably identified only at frequencies of 6 kHz and above and was not consistently observed at 2 and 4 kHz due to the relatively short contact length of the probe tip (0.5 mm) compared to the S0 half wavelengths (~3.5 mm at 4 kHz). The corresponding phase velocity dispersion curves are depicted in Fig. 2G, showing good agreement between the experimental data and our theoretical model. For this tension-free, isotropic PDMS phantom, we determined the intrinsic shear modulus Gzxσ=0=μ=184kPa and Exxσ=0=736kPa. These values satisfy the expected relationship Exxσ=0=4μ, which holds only for isotropic materials. We then cut the sample into small pieces and performed uniaxial tensile tests using a commercial mechanical machine (see Supplementary Fig. S4). By curve fitting of the strain-stress curve we measured a quasi-static shear modulus of 172 kPa. This value is quite close to 184 kPa obtained from OCE at around 10 kHz frequencies.

To investigate the effect of intraocular pressure (IOP), we increased the IOP from 0 to 40 mmHg by attaching a water column with controlled height to the artificial anterior chamber. Figure 2H presents the A0 wave velocity at 16 kHz as a linear function of IOP, demonstrating a linear slope of 0.02 m/s⋅mmHg, or 0.2% increase per mmHg. This slope aligns with the theoretical expectation for a neo-Hookean model of PDMS μ=184kPa,k1=0.

4.2. Porcine corneas ex vivo

To further validate the OCE method, we conducted experiments using fresh porcine eye globes ex vivo, where the IOP was controlled using a saline water column, as shown in Fig. 3A. Displacement maps were obtained at frequencies of 2 kHz, 4 kHz, and 6 kHz at an IOP of 5 mmHg, as depicted in Fig. 3B. Corresponding displacement profiles at the cornea surface are presented in Fig. 3C. In the wavenumber domain, both the A0 and S0 modes were clearly identified, with the S0 mode becoming dominant at higher frequencies. By fitting the measured dispersion curves over frequency (Fig. 3E), we determined an in situ out-of-plane shear modulus of Gzx=9.0±0.6kPa and an in situ in-plane tensile modulus of Exx=216±22kPa. The ratio Exx/4Gzx is 6, indicating a significant tensile-to-shear anisotropy in the corneal tissue. Figure 3F displays the wave velocities measured at 2 kHz and 10 kHz from seven porcine eye globe samples. The mean phase velocities for the A0 and S0 modes at 2 kHz were 2.18 ± 0.08 m/s and 17.9 ± 6.1 m/s (Mean ± SD), respectively, while the A0 wave velocity at 10 kHz was 3.46 ± 0.19 m/s.

Fig. 3. Simultaneous measurement of the A0 and S0 waves in ex vivo porcine corneas.

Fig. 3.

(A) Schematic of the experimental setup. IOP is controlled using a water column. (B) Maps of the displacement field. (C) Surface displacement extracted from the surface of the cornea at 2, 4, and 6 kHz (bottom to top). (D) Fourier transform of the displacement. The two peaks correspond to A0 and S0 modes, respectively. (E) Corresponding dispersion relations of the A0 and S0 waves. The error bar represents three measurements on the same location of the porcine cornea. (F) Inter-sample variations from 7 samples. (G) Effect of the IOP on the phase velocities of the A0 mode at 10 kHz. (H) Effect of the IOP on the phase velocities of the S0 mode at 2 kHz. Dashed curves represent the theoretical fit.

The dependence of phase velocity on IOP was investigated by increasing the IOP from 5 mmHg up to 40 mmHg in increments of 5 or 10 mmHg. The deformation of the cornea, as measured from the OCT images, exhibited good agreement with our numerical simulations (Supplementary Fig. S5). Figure 3G illustrates the IOP-dependence of the A0 wave at 10 kHz, demonstrating a linear relationship. The velocity slope was found to be 0.12 m/s⋅mmHg or a 2.5% increase per mmHg at 15 mmHg. Figure 3H displays the IOP-dependence of the S0 wave at 2 kHz. The S0 wave velocity exhibited nonlinearity initially up to 10 mmHg, followed by a linear increase with larger IOPs. The slope at 15 mmHg was approximately 0.93 m/s⋅mmHg, or 3.3% increase per mmHg. The experimental data exhibited remarkable agreement with numerical simulation results based on the model with μ=9kPa and k1=45kPa. A slightly higher value of 500 was used for k2 to account for the nonlinearity of corneal tissues.

4.3. Human corneas in vivo

In the final phase of our study, we applied the OCE method to human eyes. Six healthy subjects with normal IOP in a range of 10–15 mmHg were recruited. Four subjects were in early 30s (31–34 years old), and two subjects were over 50s (53 and 62 years old). Details about the subjects can be found in Table 1. The study was conducted at the Massachusetts General Hospital (MGH) with approval from the Institutional Review Board (IRB) of MGH and the Mass General Brigham Human Research Office. A topical anesthetic was administered to the eye. Only the left eye of each subject was measured. The excitation contact probe used was spring-loaded to maintain a small, constant force of ~10 mN when in contact with the corneal surface (Fig. 4A). No noticeable corneal deformation was observed in OCT images. A complete scan from 6 to 16 kHz, with a 2 kHz interval, took 2.4 s. All subjects demonstrated cooperation by refraining from blinking and maintaining focus on a fixed target during the imaging scan.

Table 1.

In vivo human data

Subjects Age Gender Spherical diopters Corneal thickness (μm) Out-of-plane shear modulus G¯zx(kPa) In-plane tensile modulus E¯xx(kPa) E¯xx/4G¯zx
1 31 M −3.00 0.55 81 ± 3 3730 ± 57 11.5
2 34 F −1.50 0.50 93 ± 12 3060 ± 270 8.3
3 33 M −2.50 0.55 99 ± 7 3510 ±460 8.9
4 34 M −7.75 0.60 130 ± 10 3980 ± 500 7.7
5 62 M −3.00 0.55 70 ± 3 3400 ± 140 12.0
6 53 M −5.75 0.57 89 ± 2 6060 ± 870 22.4

Fig. 4. In vivo human measurement of the tensile and shear moduli using the S0 and A0 waves.

Fig. 4.

(A) Monitoring camera view of a human subject during measurement. (B) Displacement profiles at 6, 12, and 16 kHz for Subject 2. (C) Surface displacement. (D) Wavenumber domain Fourier transform result of the wave displacement. Arrows point to the A0 and S0 modes. (E) Dispersion curves measured from six healthy subjects. (F) Regression analysis of age dependence for the tensile and shear moduli measured at 16 kHz. Shaded regions represent 95% confidence intervals.

Figure 4B presents representative displacement profiles obtained from Subject 1. At 6 kHz, only the A0 mode was excited, while at 12 kHz and above, a combination of the S0 and A0 mode was observed. Measured surface wave displacement profiles are shown in Fig. 4C, and the corresponding wavenumber domain plots are displayed in Fig. 4D. At 16 kHz, peaks appearing in the negative wavenumber domain are likely caused by the reflection of the S0 wave from the limbus and were therefore disregarded.

Figures 4E depict the measured phase velocity dispersion curves for the six healthy human subjects. Each data point and error bar represent the mean and standard deviation of two consecutive OCE scans performed at the same location. The wavelengths of the A0 and S0 waves are displayed in Appendix Table A1. At 12 kHz and above, the contact length (~0.5 mm) is notably larger than the half of the A0 wavelength, resulting in the suppression of the A0 wave. This is consistent with the observation that the S0 mode becomes more pronounced at and above 12 kHz. For Subject 6, the S0 mode was only observed at 16 kHz due to its long wavelengths at lower frequencies compared to the contact length.

The dispersion curves were fitted with the acoustoelastic model described in Section 2, assuming an IOP of 12 mmHg for all subjects. The measured shear and tensile modulus values are shown in Table 1. For Subjects 1 to 4 (31 – 34 yo), we obtained shear modulus Gzx in a range of 81 to 130 kPa, and tensile modulus Exx varying from 3.06 to 3.98 MPa. The anisotropy ratio Exx/4Gzx ranges from 7.7 to 11.5. Subjects 5 and 6 (62 and 53 yo) exhibited shear moduli (Gzx) of 70 and 89 kPa and tensile moduli (Exx) of 3.40 and 6.06 MPa, respectively. The ratios Exx/4Gzx were 12.0 and 22.4, higher than Subjects 1 to 4. Figure 4F shows the measured shear and tensile moduli as a function of the age. A linear regression analysis revealed a negative slope of −8.3 ±4.5 kPa per decade for shear modulus and a positive slope of 303 ± 206 kPa per decade for tensile modulus (95% confidence interval).

5. Discussion

In this study, we presented a guided-wave OCE system that enabled simultaneous excitation and detection of the S0 and A0 waves in corneas. While the relationship between the waves and elastic moduli has been known, to our knowledge, this is the first experimental study that measures tensile modulus directly from the S0 wave in the cornea. The measured wave dispersion and dependence on pre-stress (corneal tension) showed good agreement with our acoustoelastic theory based on the HGO constitutive model. By fitting the phase velocity dispersion curves of the S0 and A0 waves using the model, we obtained the in-plane tensile modulus and out-of-plane shear modulus of human corneas in vivo. From six normal subjects, the shear modulus G-zx ranged from 70 to 130 kPa and the tensile modulus E-xx ranged from 3.06 to 6.06 MPa.

In terms of the HGO model parameters, we determined that k1 ranged from 670 to 1420 kPa and k2 was set to 500, based on the data collected from the six subjects. While the HGO model is considered a reasonable choice for describing corneal behavior, ongoing efforts to develop more sophisticated constitutive models are underway [34, 35]. It would be interesting to incorporate these models into the acoustoelastic theory and derive parameters from our OCE data. However, this topic falls beyond the scope of the current study.

Comparing our results with previous mechanical measurements on ex vivo corneal tissues, tensile moduli have been reported in the range of 0.2 to 3 MPa [3638], while torsional tests applying shear stress along the corneal plane exhibited elastic moduli ranging from 3 to 50 kPa [39, 40]. Our in vivo results measured at ~ 10 kHz are slightly higher than these ex vivo mechanically measured values. We also compared our data with previous in vivo measurements reported in the literature (see Appendix Table A2). Some studies only reported group velocities [2, 15, 41, 42], from which it is difficult to quantify elastic moduli accurately without phase velocity dispersion information. Our measured A0 wave phase velocities were reasonably consistent with the reported values from previous OCE studies [13, 16]. Our measured tensile moduli were significantly larger than previously reported values up to 0.8 MPa using indentation and tonometer-based methods [8, 9]. We speculate that the indentation might cause eyeball movement and IOP change leading to an underestimation of tensile modulus.

In our current study, we observed an upward trend of tensile modulus with age in vivo. This age-dependent stiffening is qualitatively consistent with previous results of ex vivo mechanical studies [3638]. This current study also observed a downward trend of shear modulus with age in vivo, consistently from our earlier finding [13]. The subjects above 50s exhibited 32% higher tensile modulus and 21% lower shear modulus (80 vs. 101 kPa) compared to the younger subjects (average age 33). This opposite trend between tensile and shear moduli is rather surprising and highly intriguing, although large-scale clinical studies are warranted not only to confirm this preliminary finding and but also to understand their implications. One plausible explanation for the opposite trend is that collagen fibers in the cornea lose elasticity and become stiffer with age, resulting in increased tensile modulus, while the interfibrillar matrix softens and diminishes shear modulus.

To capture the full dispersion curve of the S0 wave in human corneas, it is necessary to extend the frequency range beyond 16 kHz. Additionally, people with ocular hypertension and patients with high-tension glaucoma are expected to have even higher S0 wave speeds, necessitating higher frequencies for accurately characterizing the S0 waves. We have recently developed an ultrahigh-frequency OCE method [44] and plan to incorporate this technique in our future clinical studies. With further optimizations, the guided-wave OCE technique holds promise for comprehensive studies of corneal mechanics and its role in the management of corneal diseases and refractive surgeries.

Supplementary Material

1

Statement of Significance:

This article reports an innovative elastography technique using two guided elastic waves, demonstrating the measurement of both tensile and shear moduli in human cornea in vivo with unprecedented precision. This technique paves the way for comprehensive investigations into corneal mechanics and holds clinical significance in various aspects of corneal health and disease management.

Funding.

National Institutes of Health (NIH) via grants (R01-EY033356, P01-EY034857).

Appendix

Table A1.

Measured wavelengths of the A0 and S0 wave from six subjects.

Frequency (kHz) A0 wavelength (mm) S0 wavelength (mm)
6 1.54 ± 0.18 -
8 1.14 ± 0.15 -
10 0.89 ± 0.11 -
12 0.74 ± 0.10 3.93 ± 0.29
14 0.65 ± 0.07 2.98 ± 0.31
16 0.57 ± 0.05 2.63 ± 0.60

Table A2.

Previous in vivo measurements of corneal biomechanics

Refs Stimulus (Frequency) Method Subjects Results Model
Sit [12] Shaker (100 Hz) US Human (N = 20) Phase velocity, 1.82 ± 0.10 m/s Lamb wave
Nguyen [2] ARF (~800 Hz) US Porcine (N = 4) Group velocity, 6 ~ 8 m/s Lamb wave
Nguyen [41] ARF (~800 Hz) US Porcine (N = 4) Group velocity, ~1.9 m/s Lamb wave
Ramier [13] Contact probe (up to 16 kHz) OCE Human (N = 12) Phase velocity, 7.86 ± 0.75 m/s Rayleigh surface wave
Lan [15] Air-puff (~500 Hz) OCE Human (N = 18) Group velocity, ~3.5 m/s -
Jin [16] Air-puff (~500 Hz) OCE Human (N = 12) Phase velocity, 12.73 ± 1.46 m/s (>2 kHz) Scholte wave
Zvietcovich [45] ARF (2 kHz) OCE Rabbit (N = 4) Phase velocity, 3.75 ~ 5.72 m/s Lamb wave
Mekonnen [46] ARF (<3 kHz) OCE Mouse (N = 4) Phase velocity, 3.41 ± 0.52 m/s Lamb wave
Kirby [18] ARF (<3 kHz) OCE Rabbit (N = 5) Shear modulus 34 kPa to 261 kPa and tensile modulus 20 MPa to 44 MPa Lamb wave
Jin [47] ARF (~500 Hz) OCE Rabbit (N = 4) Phase velocity, 7.52 ± 1.11 m/s Rayleigh surface wave
Li [42] ARF (~500 Hz) OCE Rabbit (N = 4) Group velocity, ~6.2 m/s -
Kling [6] Air-puff (~30 Hz) Corvis ST Human (N = 9) Young’s modulus, ~0.40 MPa for posterior and ~1.52 MPa for anterior cornea Finite element analysis
Shih [7] Air-puff (~30 Hz) Corvis ST Human (N = 25) Young’s modulus, ~0.32 MPa Modal analysis
Pye [8] Applanation (Static) Applanation Tonometry Human (N = 100) Young’s modulus, 0.25 ± 0.1 MPa Orssengo-Pye model
Lam [9] Flat indenter (Static) Indentation Human (N = 29) Tangent modulus, 0.755 ± 0.159 MPa Thin shell model

ARF: acoustic radiation force, US: ultrasound elastography, OCE: optical coherence elastography

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Disclosures. The authors declare that there are no conflicts of interest related to this article.

Declaration of interests

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Supplemental document. See Supplement 1 for supporting content.

References

  • [1].Elsheikh A, Alhasso D, Mechanical anisotropy of porcine cornea and correlation with stromal microstructure, Experimental eye research 88(6) (2009) 1084–1091. [DOI] [PubMed] [Google Scholar]
  • [2].Nguyen T-M, Aubry J-F, Fink M, Bercoff J, Tanter M, In vivo evidence of porcine cornea anisotropy using supersonic shear wave imaging, Investigative ophthalmology & visual science 55(11) (2014) 7545–7552. [DOI] [PubMed] [Google Scholar]
  • [3].Eltony AM, Shao Peng, and Yun Seok-Hyun, Measuring mechanical anisotropy of the cornea with Brillouin microscopy, Nature communications 13 (2022) 1354. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Pitre JJ, Kirby MA, Li DS, Shen TT, Wang RK, O’Donnell M, Pelivanov I, Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE, Scientific reports 10 (2020) 12983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Luce DA, Determining in vivo biomechanical properties of the cornea with an ocular response analyzer, Journal of Cataract & Refractive Surgery 31(1) (2005) 156–162. [DOI] [PubMed] [Google Scholar]
  • [6].Kling S, Bekesi N, Dorronsoro C, Pascual D, Marcos S, Corneal viscoelastic properties from finite-element analysis of in vivo air-puff deformation, PloS one 9(8) (2014) e104904. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Shih PJ, Cao HJ, Huang CJ, Wang IJ, Shih WP, Yen JY, A corneal elastic dynamic model derived from Scheimpflug imaging technology, Ophthalmic and Physiological Optics 35(6) (2015) 663–672. [DOI] [PubMed] [Google Scholar]
  • [8].Pye DC, A clinical method for estimating the modulus of elasticity of the human cornea in vivo, Plos one 15(1) (2020) e0224824. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Lam AK, Hon Y, Leung LK, Lam DC, Repeatability of a novel corneal indentation device for corneal biomechanical measurement, Ophthalmic and Physiological Optics 35(4) (2015) 455–461. [DOI] [PubMed] [Google Scholar]
  • [10].Shao P, Eltony AM, Seiler TG, Tavakol B, Pineda R, Koller T, Seiler T, Yun S-H, Spatially-resolved Brillouin spectroscopy reveals biomechanical abnormalities in mild to advanced keratoconus in vivo, Scientific reports 9 (2019) 7467. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Zhang H, Asroui L, Tarib I, Dupps WJ, Scarcelli G, Randleman JB, Motion Tracking Brillouin Microscopy Evaluation of Normal, Keratoconic, and Post-Laser Vision Correction Corneas: Motion Tracking Brillouin Microscopy in Keratoconus and Laser Vision Correction, American Journal of Ophthalmology 254 (2023) 128–140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Sit AJ, Lin S-C, Kazemi A, McLaren JW, Pruet CM, Zhang X, In Vivo Non-Invasive Measurement of Young’s Modulus of Elasticity in Human Eyes: A Feasibility Study, Journal of glaucoma 26(11) (2017) 967. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Ramier A, Eltony AM, Chen Y, Clouser F, Birkenfeld JS, Watts A, Yun S-H, In vivo measurement of shear modulus of the human cornea using optical coherence elastography, Scientific reports 10 (2020) 17366. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Zvietcovich F, Birkenfeld J, Varea A, Gonzalez AM, Curatolo A, Marcos S, Multi-meridian wave-based corneal optical coherence elastography in normal and keratoconic patients, Investigative Ophthalmology & Visual Science, 2022, pp. 2380–A0183. [Google Scholar]
  • [15].Lan G, Aglyamov SR, Larin KV, Twa MD, In vivo human corneal shear-wave optical coherence elastography, Optometry and Vision Science 98(1) (2021) 58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Jin Z, Chen S, Dai Y, Bao C, Ye S, Zhou Y, Wang Y, Huang S, Wang Y, Shen M, In vivo noninvasive measurement of spatially resolved corneal elasticity in human eyes using Lamb wave optical coherence elastography, Journal of Biophotonics 13(8) (2020) e202000104. [DOI] [PubMed] [Google Scholar]
  • [17].Li Y, Zhu J, Chen JJ, Yu J, Jin Z, Miao Y, Browne AW, Zhou Q, Chen Z, Simultaneously imaging and quantifying in vivo mechanical properties of crystalline lens and cornea using optical coherence elastography with acoustic radiation force excitation, APL photonics 4(10) (2019) 106104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Kirby MA, Regnault G, Pelivanov I, O’Donnell M, Wang RK, Shen TT, Elastography for Quantification of Corneal Anisotropic Elasticity: In Vivo Rabbit Study, Translational Vision Science & Technology 12(3) (2023) 15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Li G-Y, Feng X, Yun S-H, In vivo optical coherence elastography reveals spatial variation and anisotropy of corneal stiffness, arXiv preprint arXiv:2307.04083 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].DelMonte DW, Kim T, Anatomy and physiology of the cornea, Journal of Cataract & Refractive Surgery 37(3) (2011) 588–598. [DOI] [PubMed] [Google Scholar]
  • [21].Meek KM, Quantock AJ, The use of X-ray scattering techniques to determine corneal ultrastructure, Progress in retinal and eye research 20(1) (2001) 95–137. [DOI] [PubMed] [Google Scholar]
  • [22].Ruberti JW, Sinha Roy A, Roberts CJ, Corneal biomechanics and biomaterials, Annual review of biomedical engineering 13 (2011) 269–295. [DOI] [PubMed] [Google Scholar]
  • [23].Kotecha A, What biomechanical properties of the cornea are relevant for the clinician?, Survey of ophthalmology 52(6) (2007) S109–S114. [DOI] [PubMed] [Google Scholar]
  • [24].Gasser TC, Ogden RW, Holzapfel GA, Hyperelastic modelling of arterial layers with distributed collagen fibre orientations, Journal of the royal society interface 3(6) (2006) 15–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Pandolfi A, Cornea modelling, Eye and Vision 7 (2020) 2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Pandolfi A, Holzapfel GA, Three-dimensional modeling and computational analysis of the human cornea considering distributed collagen fibril orientations, (2008). [DOI] [PubMed]
  • [27].Destrade M, Saccomandi G, Waves in nonlinear pre-stressed materials, Springer Science & Business Media 2007. [Google Scholar]
  • [28].Li G-Y, He Q, Mangan R, Xu G, Mo C, Luo J, Destrade M, Cao Y, Guided waves in pre-stressed hyperelastic plates and tubes: Application to the ultrasound elastography of thin-walled soft materials, Journal of the Mechanics and Physics of Solids 102 (2017) 67–79. [Google Scholar]
  • [29].Han ZL, Li JS, Singh M, Wu C, Liu CH, Raghunathan R, Aglyamov SR, Vantipalli S, Twa MD, Larin KV, Optical coherence elastography assessment of corneal viscoelasticity with a modified Rayleigh-Lamb wave model, J. Mech. Behav. Biomed. Mater 66 (2017) 87–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Ramier A, Tavakol B, Yun S-H, Measuring mechanical wave speed, dispersion, and viscoelastic modulus of the cornea using optical coherence elastography, Opt Express 27(12) (2019) 16635–16649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Li G-Y, Feng X, Ramier A, Yun S-H, Supershear surface waves reveal prestress and anisotropy of soft materials, Journal of the Mechanics and Physics of Solids 169 (2022) 105085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Song S, Huang Z, Wang RK, Tracking mechanical wave propagation within tissue using phase-sensitive optical coherence tomography: motion artifact and its compensation, J Biomed Opt 18(12) (2013) 121505. [DOI] [PubMed] [Google Scholar]
  • [33].Feng X, Li G-Y, Ramier A, Eltony AM, Yun S-H, In vivo stiffness measurement of epidermis, dermis, and hypodermis using broadband Rayleigh-wave optical coherence elastography, Acta Biomaterialia 146 (2022) 295–305. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Liu T, Shen M, Li H, Zhang Y, Mu B, Zhao X, Wang Y, Changes and quantitative characterization of hyper-viscoelastic biomechanical properties for young corneal stroma after standard corneal cross-linking treatment with different ultraviolet-A energies, Acta biomaterialia 113 (2020) 438–451. [DOI] [PubMed] [Google Scholar]
  • [35].Wang S, Chester SA, Multi-physics modeling and finite element formulation of corneal UV cross-linking, Biomechanics and Modeling in Mechanobiology 20(4) (2021) 1561–1578. [DOI] [PubMed] [Google Scholar]
  • [36].Elsheikh A, Wang D, Brown M, Rama P, Campanelli M, Pye D, Assessment of corneal biomechanical properties and their variation with age, Current eye research 32(1) (2007) 11–19. [DOI] [PubMed] [Google Scholar]
  • [37].Wollensak G, Spoerl E, Seiler T, Stress-strain measurements of human and porcine corneas after riboflavin–ultraviolet-A-induced cross-linking, Journal of Cataract & Refractive Surgery 29(9) (2003) 1780–1785. [DOI] [PubMed] [Google Scholar]
  • [38].Cartwright NEK, Tyrer JR, Marshall J, Age-related differences in the elasticity of the human cornea, Investigative ophthalmology & visual science 52(7) (2011) 4324–4329. [DOI] [PubMed] [Google Scholar]
  • [39].Petsche SJ, Chernyak D, Martiz J, Levenston ME, Pinsky PM, Depth-dependent transverse shear properties of the human corneal stroma, Investigative ophthalmology & visual science 53(2) (2012) 873–880. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40].Hatami-Marbini H, Viscoelastic shear properties of the corneal stroma, Journal of biomechanics 47(3) (2014) 723–728. [DOI] [PubMed] [Google Scholar]
  • [41].Nguyen T-M, Aubry J-F, Touboul D, Fink M, Gennisson J-L, Bercoff J, Tanter M, Monitoring of cornea elastic properties changes during UV-A/riboflavin-induced corneal collagen cross-linking using supersonic shear wave imaging: a pilot study, Investigative ophthalmology & visual science 53(9) (2012) 5948–5954. [DOI] [PubMed] [Google Scholar]
  • [42].Li Y, Moon S, Chen JJ, Zhu Z, Chen Z, Ultrahigh-sensitive optical coherence elastography, Light: Science & Applications 9 (2020) 58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [43].Petroll WM, Miron-Mendoza M, Mechanical interactions and crosstalk between corneal keratocytes and the extracellular matrix, Experimental eye research 133 (2015) 49–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [44].Feng X, Li G-Y, Yun S-H, Ultra-wideband optical coherence elastography from acoustic to ultrasonic frequencies, arXiv preprint arXiv:2211.10534 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [45].Zvietcovich F, Nair A, Singh M, Aglyamov SR, Twa MD, Larin KV, In vivo assessment of corneal biomechanics under a localized cross-linking treatment using confocal air-coupled optical coherence elastography, Biomedical Optics Express 13(5) (2022) 2644–2654. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46].Mekonnen T, Lin X, Zevallos-Delgado C, Singh M, Aglyamov SR, Coulson-Thomas V, Larin KV, Longitudinal assessment of the effect of alkali burns on corneal biomechanical properties using optical coherence elastography, Journal of Biophotonics 15(8) (2022) e202200022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [47].Jin Z, Khazaeinezhad R, Zhu J, Yu J, Qu Y, He Y, Li Y, Alvarez-Arenas TEG, Lu F, Chen Z, In-vivo 3D corneal elasticity using air-coupled ultrasound optical coherence elastography, Biomedical Optics Express 10(12) (2019) 6272–6285. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES