Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2025 Apr 1.
Published in final edited form as: Magn Reson Med. 2023 Dec 19;91(4):1419–1433. doi: 10.1002/mrm.29944

Validation of Single Reference Variable Flip Angle (SR-VFA) dynamic T1 mapping with T2* correction using a novel rotating phantom

Michael A Malmberg 1,2,a), Henrik Odéen 2, Lorne W Hofstetter 3, J Rock Hadley 2, Dennis L Parker 2
PMCID: PMC10872756  NIHMSID: NIHMS1944922  PMID: 38115639

Abstract

Purpose:

To validate single reference variable flip angle (SR-VFA) dynamic T1 mapping with and without T2* correction against inversion recovery (IR) T1 measurements.

Methods:

A custom cylindrical phantom with 3 concentric compartments was filled with variably doped agar to produce a smooth spatial gradient of the T1 relaxation rate as a function of angle across each compartment. IR T1, VFA T1, and B1+ measurements were made on the phantom before rotation, and multi-echo stack-of-radial dynamic images were acquired during rotation via an MRI-compatible motor. B1+ corrected SR-VFA and SR-VFA-T2* T1 maps were computed from the sliding window reconstructed images and compared against rotationally-registered IR and VFA T1 maps to determine the percentage error.

Results:

Both VFA and SR-VFA-T2* T1 maps fell within 10% of IR T1 measurements for a low rotational speed, with a mean accuracy of 2.3% ± 2.6% and 2.8% ± 2.6% respectively. Increasing rotational speed was found to decrease the accuracy due to increasing temporal smoothing over ranges where the T1 change had a non-constant slope. SR-VFA T1 mapping was found to have similar accuracy as the SR-VFA-T2* and VFA methods at low echo times (~<2 ms), whereas accuracy degraded strongly with later echo times. T2* correction of the SR-VFA T1 maps was found to consistently improve accuracy and precision, especially at later echo times.

Conclusion:

SR-VFA-T2* dynamic T1 mapping was found to be accurate against reference IR T1 measurements within 10% in an agar phantom. Further validation is needed in mixed fat-water phantoms and in vivo.

Keywords: T1 mapping, Accuracy and Precision, Single reference variable flip angle, Quantitative Imaging, MRI-compatible motor, MRTI

1. Introduction

The longitudinal magnetization relaxation time constant, T1, is a fundamental time constant in magnetic resonance imaging (MRI) acquisitions that depends on the local tissue environment and allows MRI to produce excellent soft tissue image contrast. Quantitative T1 measurements have been used for a variety of clinical indications, such as hypoxia in tumors,1,2 liver diseases,3,4,5, pancreatitis,6,7 heart pathologies,8,9, 10,11,12 pulmonary disease,13 pulmonary cancer,14 and beyond.15,16,17 Furthermore, dynamic quantitative T1 mapping has been used to measure oxygenation in the brain18 and lungs,19 for local perfusion measurements using dynamic contrast enhanced imaging,20,21 and has been considered for T1-based thermometry in monitoring thermal therapies.22, 23, 24, 25 In these cases, high temporal resolution dynamic T1 mapping is desired to capture the dynamic processes (contrast perfusion, temperature change) being measured.

The widely accepted gold standard method for T1 mapping uses inversion recovery (IR) acquisitions, which are slow, even with fast spin-echo acquisitions. To obtain T1 measurements more efficiently, multiple different acquisition strategies have been developed, including the Look-Locker method26 and its derivatives such as MOLLI27 and shMOLLI,28 saturation recovery methods such as SAPPHIRE,29 and the variable flip angle (VFA) method.30 The VFA method is notable for being a fast T1 mapping method with good noise efficiency.31 However, it requires a minimum of two images per T1 map, and its accuracy is dependent on a wide range of factors, from flip angle choices and TRs32,33,34,35 to B1+ mapping36, 37 and adequate spoiling36,38,39. The acquisition time of the VFA method is also increased by the time required to reach a steady state after changing between flip angles.

A speed improvement for the VFA method, called the single-reference variable flip angle (SR-VFA) technique, measures T1 dynamically using subsequently acquired images at the high flip angle after a reference T1 map and baseline image at a low flip angle are acquired. By eliminating the need to alternate between flip angles, this method can generate dynamic and volumetric T1 maps at a high-temporal resolution. While the SR-VFA method has been evaluated in theory and simulation,35,40 its experimental accuracy has not been validated against gold standard IR T1 measurements.

The long acquisition time required for an IR T1 measurement makes it difficult to produce ground truth T1 measurements for any rapidly changing T1 situation, such as during contrast injection or focal heating. IR T1 measurements can only be obtained before and/or after dynamic T1 changes or at intermediate points if the T1 can be held constant. Further, it is difficult to obtain highly repeatable T1 values during these dynamic situations in which the T1 over time cannot be easily controlled.

To circumvent experimental challenges in evaluating dynamic T1 mapping protocols, we have developed an innovative rotating cylindrical agar phantom. The phantom was designed to have smooth variation in T1 values with respect to position by variably doping the agar along concentric rings. When the phantom is stationary, IR methods can be used to accurately measure the ground truth T1 distribution. When the phantom is rotating, each imaging voxel samples a signal that changes with the spatial variation in T1, thereby mimicking changes in T1 that might occur during, e.g., contrast agent passage or focal heating. The SR-VFA method can be used to perform dynamic T1 measurements during phantom rotation. By rotational registration of the IR T1 maps to their SR-VFA T1 counterparts, SR-VFA T1 mapping can be more accurately evaluated for accuracy against the gold standard, and inferences can be made on the relationship between the accuracy of dynamic T1 changes and the imaging temporal resolution.

This work tested the validity of the SR-VFA method of T1 mapping and its recently published T2* correction35 (hereafter referred to as the SR-VFA-T2* method) against VFA and gold standard IR T1 maps for both static and “dynamic” acquisitions in an agar phantom. This work will enable future validation studies in vitro and in vivo towards application in fast dynamic T1 mapping.

2. Theory

A brief overview of the theory of VFA, SR-VFA, and SR-VFA-T2* T1 mapping is given here for the benefit of the reader, with full derivations and discussions in corresponding referenced work.

2.1. Variable flip angle (VFA) T1 mapping

VFA T1 mapping employs 2 or more scans at differing flip angles, and fits the resulting magnitude image intensity values to the steady state spoiled gradient echo signal equation,30,41,42

S=M01-E1sinα1-E1cosαE2 [1],

where α is the flip angle (FA), E1=e-TR/T1, E2=e-TE/T2*, and M0 is the equilibrium magnetization. By alternating FAs, a quantitative T1 map can be generated every 2 image sets. Adequate spoiling and accurate knowledge of FAs are required to produce accurate T1 maps with this method.36,43,44,45 Additionally, a priori knowledge of the approximate T1 range of interest may be required for FA determination when SNR is limited, since the accuracy and precision of VFA T1 maps are influenced by the choice of FAs relative to the Ernst angle for each voxel.32,34,35

2.2. Single reference variable flip angle (SR-VFA) T1 mapping

SR-VFA T1 mapping extends the VFA method by applying the VFA method twice: once on a set of baseline images, and once on an image set consisting of one baseline image and one “dynamic” image, or an image taken at the new T1 that is to be measured. By combining the incorrect estimate of the “dynamic” T1 with the baseline T1 with an analytic correction, the true dynamic T1 can be obtained.40 This method relies on the same assumptions as the VFA method, with the additional requirement that TE << T2* and/or that no change in T2* occurs between baseline and dynamic images.

2.3. Single reference variable flip angle T1 mapping with T2* correction (SR-VFA-T2*)

An in-depth simulation/mathematical analysis of the relation between T1 accuracy and the assumption of negligible T2* effects has been published previously.35 It was found that SR-VFA T1 maps suffer from a systematic bias, in which changes in T1 are underestimated due to changes in T2* weighting between the baseline and dynamic images. This bias is correctable by incorporating an estimate of T2* obtained from multi-echo acquisitions into the original SR-VFA T1 calculation.35, 40

3. Methods

3.1. Setup

Two phantoms were custom designed in SOLIDWORKS (Dassault Systèmes, Waltham, MA) and 3D-printed with PLA filament material. Both phantoms include 3 inner concentric rings, each filled with an agar solution that was doped to give T1 and T2 values that varied smoothly around the phantom in each ring (see Figure 1a,c). The dimensions of the rings are given in Table S1, and the 3D model files for both phantoms can be found at https://www.github.com/mikemalm6/SRVFA_Exp_Paper. Each of the three thick inner rings in the rotating phantom were set to contain a specific range of T1 and T2 values by varying the dopant amounts of CuSO4 (Copper sulfate pentahydrate, Acros Organics, Thermo Fisher Scientific, Waltham, MA) and MnCl2 (Manganese chloride tetrahydrate, Alfa Aesar, Thermo Fisher Scientific, Waltham, MA), similar to the concept used by Gopalan et al.,46 except the calibration of each solution’s T1 value was not performed. In the stationary phantom, the three rings were filled to have initial concentrations that yielded T1 values in the range of ~200–1500 ms. Due to diffusion of dopant between adjoining compartments and the geometry of the phantom, the T1 ranges for the innermost to outermost concentric rings were, ~300–900 ms, ~200–1300 ms, and ~200–1400 ms, respectively. The three thick rings in the rotating phantom, from innermost to outermost, included the following approximate T1 ranges: short (~150–500 ms), moderate (~350–1000 ms), and long (~750–2150 ms). Splitting the whole T1 range into parts in this way keeps the maximum ratio between two T1s in a ring at approximately three, which is the amount of T1 change over which the SR-VFA method was previously analyzed theoretically,35 and which provides a conservative estimate for the maximum T1 change expected during a focused ultrasound ablative procedure. The outermost thin ring was constructed to have a uniform concentration aimed to emulate similar relaxation characteristics to adipose tissue at body temperature.47 This region having a constant T1 and T2 relaxation rate could then be utilized as a reference during phantom rotation to more easily identify image artifacts. The concentrations of dopant used to create each phantom compartment is given in Tables S2 and S3.

Figure 1 –

Figure 1 –

Rotating phantom setup. (A) – 3D-printed phantom model with one outer ring and 3 thick concentric rings with spacers for the first agar filling inserted in alternating sections via slot guides in the print. Fiducial-marker is in the center. The high T1 region for each ring (i.e., with the lowest dopant concentration) is assigned to the largest section of each ring to maintain the desired T1 range for a longer period of time (weeks) despite diffusion. Diffusion is intended to smooth out the sharp T1 transition between sections. (B) – Rotating platform assembly with power cord and encased MRI-compatible motor with separate power box/grounding cable. Phantom is locked in place in the center of the platform via 4 aligning screws along the platform’s edge. (C) Rotating phantom filling procedure. CuSO4 and MnCl2 are mixed in custom amounts to produce 5×3 distinct “pre-agar” aqueous solutions, where yellow, red, and blue in the small beakers correspond with the outer, middle, and inner rings of the phantom respectively, and solutions are in order of decreasing T1 values (i.e., 1 = highest T1 in ring). Each solution is combined in a 1:3 ratio with hot 1% agar, mixed, allowed to cool to 50°C, and then poured into alternating sections of the phantom as shown above. After the agar sets in the fridge the remaining spacers are carefully removed and the resulting empty sections are filled with their respective agar mixes in the same fashion.

3.2. Filling process

The goal of the rotating phantom was to provide a smooth variation in T1 and T2 values with time during phantom rotation. To accomplish this, eight sections were filled in each thick ring with homogeneous doped agar mixtures to create a quasi-sinusoidal T1 pattern along each ring (see Figure 1c). Natural diffusion gradually smoothed out the sharp transitions between each section over several days. Dopant concentrations for each section were determined using relaxivities given by Thangavel et al.,48 and were adjusted so that a 3:1 mix of hot 1% w/v agar and the doped aqueous mix for each section would produce the desired dopant concentration in the resulting 0.75% w/v agar. Alternating sections of alternating rings of the phantom were filled with these mixtures, while in-between sections were blocked using removable 3D-printed spacers (see Figure 1a,c). After the spaced agar solutions had set, the spacers were removed and remaining sections were filled to produce continuous rings of variably doped agar. To prevent temperature-induced warping of the 3D-printed plastic, the agar mixtures had to cool to approximately 50 °C before pouring into each phantom compartment. Further cooling of the thin walls was done by filling the rings that weren’t being filled with the hot agar mix with cold water. The whole phantom was allowed to set in the fridge, and then to equilibrate to room temperature for 8+ hours before scanning. Figure 1c gives a visual description of the filling process for the rotating phantom, and Figure S1 gives a similar description for the stationary phantom.

3.3. Rotating platform

A custom MRI-compatible voltage-controlled electromagnetic motor49 was used to actuate a 3D-printed rotating platform. The platform used plastic spheres as ball-bearings to enable the rotation (see Figure 1b). The motor’s speed could be manually adjusted by changing the voltage input. In these experiments, the speed was controlled by switching between 1.25 V, 2.5 V, 3.7 V, and 7.4 V battery sources, thereby achieving rotational speeds at the MRI magnet’s isocenter of 1 rotation per ~254, 119, 83, and 41 seconds (Speeds 1, 2, 3 and 4), respectively. Thus, at the fastest speed, the time to cover the full T1 range within each ring was ~20 seconds, which is comparable to the temperature rise time during a focused ultrasound ablative procedure. The phantom described above was secured to the center of the rotating platform using plastic aligning screws.

3.4. Experiments

All experiments were performed on a Siemens Prisma Fit 3T MRI scanner (Siemens Healthcare, Erlangen, Germany). Turbo spin echo (TSE) IR measurements with an adiabatic slice-selective inversion pulse were used to determine reference standard T1 measurements in all cases. Detailed scan parameters are given in Table 1. Separate validation was performed of our TSE IR measurements against values from a calibrated NIST System Phantom (CaliberMRI, Boulder, CO) for the T1 values relevant to our custom build phantom (T1 > 120 ms, T2 > 14 ms). Accuracy (mean ± 2 SD) of the inversion recovery measurements used as reference standard for this work was found to be −0.43 ± 9.95 %, with a maximum deviation of 13.6%.

Table 1:

MRI acquisition parameters

Scan Type Parameter Value – Static Exp Value – Dynamic Exp
All Field-of-view 256 mm 400 mm
RF spoiling phase increment 117° (GRE scans)
50° (IR)
Bandwidth 1184 Hz/pix 501 Hz/pix
Slice thickness 5 mm (GRE),
10 mm (IR)
2 mm (GRE & IR)
Inversion recovery (turbo spin echo) TR 10.0 sec 8.0 sec
TI [25 75 150 400 900 1500 3500] ms [25 75 200 500 1000 2000 3000 7000] ms
Image size 128×128 192×192
ETL 14
AFI B1+ mapping (GRE) TR1 / TR2 30 ms / 150 ms
TEs 2.17 ms 3.0 ms
Image size 128×128×10 96×96×8
FA 60° 48°
VFA/SR-VFA (stack-of-radial GRE) TR 20 ms
TE [1.24, 2.80, 4.36] ms [1.94, 3.81, 5.68, 7.55] ms
Image size 128×128×10 192×192×8
FA [7°, 25°] [5.5°, 23°]
# Radial views / view increment 377 / ~68.7533°
# Repetitions 1 FA 5.5° - 2; FA 23° - 7
# Averages 1 FA 5.5° - 4; FA 23° - 1

3.4.1. Static experiment

Static scans were acquired to verify the SR-VFA method in the absence of dynamically changing T1 values. To do this, baseline TSE IR T1 measurements were performed, followed by pseudo-golden angle (PGA) stack-of-radial50 baseline measurements with FAs of 7° and 25°, and B1+ mapping with the Actual Flip angle Imaging (AFI) method.36, 52 Following this, PGA stack-of-radial measurements were repeated at each discrete 30° rotation from 0° to 360°, and B1+ maps were acquired after every 90° of rotation to ensure consistency of the B1+ field at each rotational interval.

3.4.2. Dynamic rotation experiment

Additional scans were acquired during rotation of the phantom to determine how accurately SR-VFA T1 mapping can capture true T1 even when T1 within a given voxel changes during the acquisition. As in the static experiment, TSE IR T1 scans and B1+ mapping with the AFI method were performed before rotation, in addition to baseline PGA stack-of-radial50 VFA measurements at FAs of 5.5° and 23°. FAs were independently chosen for each experiment to maximize precision in the SR-VFA method according to (35) for the range of T1 values present in the phantom and the TR used. PGA stack-of-radial measurements were chosen so that the k-space trajectory repeated itself after a fixed number of lines, thereby allowing for a trajectory-matched baseline comparison for each dynamic image.

Following this, additional radial measurements were acquired at 23° during rotation of the phantom. Rotational position was recorded before and after acquisition, as well as the elapsed rotation time, and then the rotating platform’s voltage was manually switched to acquire equivalent stack-of-radial measurements with the new rotation speed in a similar fashion. This was done for each of the 4 voltages (rotational speeds) mentioned previously. IR measurements were repeated after data acquisition for the first two rotation speeds was complete (halfway through the experiment), and again at the end to validate the phantom’s ground truth T1 consistency over the experiment’s duration. B1+ mapping was repeated before each dynamic rotation and again post-experiment to ensure B1+ consistency across the various rotational positions.

3.5. Simulation

Simulation was performed to determine the effects of our dynamic measurement setup on T1 accuracy for our measured IR T1 data. Simulated signal data was obtained via the spoiled gradient echo steady state signal equation at the FAs used in the dynamic rotation experiment at 1440 × 3 points, corresponding with the 0.25°-spaced ROI positions in each of the 3 rings of the IR T1 analysis (see next section). Echo times and T2* values were chosen to approximately match those from the dynamic rotation experiment.

To simulate rotation for the “dynamic” dataset for a single reference T1 map, the high FA signal data was filtered using a moving average over the number of points corresponding with the angle traversed by the phantom at each of the 4 measured speeds over the time it takes to acquire 55 radial views during the dynamic experiment. Since radial acquisitions cross the center of k-space with each view, each view contributes to the bulk contrast of the image, and thus to the T1. Then, as a simple approximation, the effect of rotation can be modeled as an average over the rotational angle corresponding with the time to fill “enough” of the center of k-space. To this end, 55 views were chosen, since the data were reconstructed using a sliding window with k-space weighted image contrast (KWIC)51 that uses a Fibonacci number of views in each KWIC section, and the section with 55 views covers a moderate portion of the center of frequency space. Using this algorithm with pseudo-golden angle acquisition, the k-space measurements near the center of k-space are acquired closely spaced in time with the time footprint increasing for measurements obtained farther from the center of k-space.

SR-VFA-T2* T1 values were calculated using these signals and compared against the reference IR dataset used in the imaging analysis. The percentage error was determined, and this simulated percentage error was compared against the actual percentage error observed in the data (see next section).

3.6. Image analysis

All images were reconstructed offline using in-house MATLAB code (MathWorks, Natick, MA) that utilized convolution gridding53 and iterative density compensation.54 All images contributing to T1 maps were Hamming-filtered in the frequency domain during reconstruction to limit the effects of noise on the images, and zero-fill interpolated by a factor of 2 to reduce partial-volume effects.55 IR T1 maps were reconstructed with the Reduced Dimension Nonlinear Least Squares algorithm.56 VFA T1 maps were computed on a voxel-by-voxel basis by fitting signal magnitude values to Equation 1.41 SR-VFA T1 maps were obtained via the process given by Svedin et al.,40 using the “baseline” images acquired at 0° rotation (static experiment) or the pre-rotation images (dynamic experiment) in conjunction with the high FA “dynamic” images acquired at each rotational angle (static experiment) or during rotation (dynamic rotation experiment). Four B1+ maps acquired during each experiment were averaged and applied during T1 calculation to account for FA errors.

Once all T1 maps were created, the average T1 was extracted from circle ROIs of radius 4 voxels at 1440 evenly spaced intervals (every ~0.25°) along the center of each thick concentric ring. These data were then registered to the same starting rotation angle. Following this registration, the stationary T1 maps’ ROI data (IR and VFA) underwent a simulated “rotation”, during which they were circularly shifted by the number of ROIs corresponding with the angle offset between time frames, as determined by the measured rotational speed and approximate start time of the rotation. Following this simulated rotation, the stationary and dynamic T1 maps’ ROI data could be compared directly to determine a percentage error. This ROI-based percentage error data was then grouped based on IR T1 values into ~20 ms bins, and the mean and standard deviation (SD) of T1 accuracy was determined for each bin. A visual description of this process is shown in Figure 2. To calculate the error in T1 percentage change as opposed to absolute T1, the T1 percentage change from baseline was calculated for each ROI, and the IR results were subtracted from the SR-VFA methods’ results.

Figure 2:

Figure 2:

Dynamic rotation phantom experiment processing. Static phantom experiment processing was done similarly to the above, except since rotational position was not changing with time during acquisition, the 13 discrete rotation states (0 to 360° in 30° increments) replaces the time dimension on the x-axis. Additionally, in the static experiment, only IR measurements needed rotation via circular ROI shifting, as VFA measurements were acquired at every 30° increment. *Inputs to step 3 for the dynamic rotation phantom experiment were direct measurements, with the exception of rotation start time, which was estimated based on a start time within a few seconds of 2×377 stationary views.

4. Results

4.1. Accuracy and Precision

Figure 3 shows the binned T1 accuracy and precision of the SR-VFA, SR-VFA-T2*, and VFA methods for each ring of the phantom relative to IR T1 measurements in the static and dynamic rotation experiments. The first echo’s results are shown in both cases (TE = 1.24 ms and 1.94 ms, respectively), and Figure 3df’s results were taken from the lowest rotation speed acquisition (~254 sec/rev). The T1 accuracy of all methods fell approximately within ± 10%, with the stationary VFA method achieving slightly better accuracy and precision than the SR-VFA method. The T2* correction improved the accuracy and precision of the SR-VFA T1 measurements to bring the time-normalized precision to be comparable to the VFA method. Notably, the T2* correction most strongly improved the SR-VFA T1 accuracy and precision in the inner ring, in which the lowest T1 and T2 values were found. Oscillations are also seen in the accuracy plots for the dynamic single-reference T1 measurements (Figure 3d,e) that are not found in those for the stationary measurements (Figure 3a,b).

Figure 3:

Figure 3:

Accuracy of SR-VFA, SR-VFA-T2*, and dual flip angle VFA T1 mapping relative to inversion recovery T1 values for the static experiment (A-C) and dynamic rotation experiment (D-F). Data were obtained from the mean of T1 values within ROIs, as described in the Methods section. Precision is given as ± 2 SD for the VFA method, and ± √2 SD for the single reference methods to normalize for dynamic acquisition time per T1 map. Red horizontal lines at ± 10% are given as points of reference.

Figure 4 shows an overall comparison of the mean T1 accuracy and precision across all phantom rings and T1 values for each experiment, T1 mapping method, and echo time. With the SR-VFA method, the average absolute T1 percentage error and precision worsened with echo time and with rotation speed. The SR-VFA-T2* method improved the accuracy and precision in all cases and produced consistent T1 maps across the echo time dimension. At the lowest rotation speed (Speed 1), the SR-VFA-T2* method was comparable in accuracy and precision to the stationary VFA data. The static experiment showed similar results, with the SR-VFA-T2* T1 maps approaching the accuracy and precision of the VFA T1 maps. As in the other figures in this work, the SR-VFA and SR-VFA-T2* precision is given as sqrt(2) times the SD as opposed to 2 times the SD for the VFA method to account for the dynamic acquisition speed improvement afforded by the single reference methods.

Figure 4:

Figure 4:

Mean Absolute T1 % error and mean precision for the SR-VFA, SR-VFA-T2*, and VFA T1 mapping methods across the entire T1 range and all rings for the static and dynamic experiment. Precision is given as ± 2 SD for the VFA method, and ± √2 SD for the single reference methods to normalize for dynamic acquisition time per T1 map. (A) – Dynamic Rotation Experiment: Rotation speeds from lowest to highest are shown with increasing echo times in each group. One set of stationary VFA data was acquired before each rotation, and these data were all combined for the VFA column. (B) – Static Experiment. (A&B) – gray bar on the left of the SR-VFA-T2* groupings in each experiment indicates the combined echoes’ T1 map, as described in (35).

While this work is primarily focused on absolute T1 accuracy, the accuracy of percentage changes in T1 was also analyzed for the SR-VFA-T2* method relative to IR T1 values. It was found that for the range in which T1 changed by less than a factor of 3 (T1,dynamic / T1,BL ∈ [0.33 3]), the error in percentage T1 change (mean ± 2 SD of SR-VFA-T2* T1 percentage change minus IR T1 percentage change) was 0.88 ± 5.28% for the stationary experiment and 0.01 ± 6.38% for the dynamic rotation experiment at the lowest rotational speed. Additional details can be found in Figure S2.

4.2. SR-VFA vs. SR-VFA-T2*

Figures 5 and 6 gives the mean binned T1 accuracy of the SR-VFA method versus the SR-VFA-T2* method for each echo time in the static and dynamic rotation experiments respectively, grouped by phantom ring and by the relationship between the baseline (pre-rotation) T1 and the “dynamic” (post-rotation for static exp., during-rotation for dynamic exp.) T1. Each marker represents a different range of the ratio T1,dynamic T1,baseline.

Figure 5:

Figure 5:

SR-VFA mean T1 measurement accuracy and precision before (top row) and after T2* correction (bottom row) for the static experiment. Each column is a different echo time, (TE = [1.24, 2.8, 4.36] ms). The overall accuracy is improved after application of the T2* correction. Different marker types indicate the relationship between the dynamic mean T1 (T1,dyn) and the baseline mean T1 (T1,BL) for a given ROI.

Figure 6:

Figure 6:

SR-VFA mean T1 measurement accuracy and precision before (top row) and after T2* correction (bottom row) for the dynamic rotation experiment. Each column is a different echo time, (TE = [1.94, 3.81, 5.68, 7.55] ms). In all cases, the accuracy is improved after application of the T2* correction. Different marker types indicate the relationship between the dynamic mean T1 (T1,dyn) and the baseline mean T1 (T1,BL) for a given ROI.

Figure 6 also indicates that the SR-VFA method underestimates changes in T1 relative to the T1 at the baseline acquisition. For points at which T1,dynamic T1,baseline < 75%, as shown by the dot markers in Figure 6, the SR-VFA method overestimated the true T1 values by up to ~30%. When T1,dynamic T1,baseline > 133%, as shown by the open circle markers in Figure 6, the SR-VFA method underestimated the true T1 values by up to ~10%. Echo time had only a minor effect on the T1 accuracy when T1,dynamic T1,baseline was closer to 1. In both experiments accuracy decreased with increasing echo time. Additionally, the bias increased with echo time most strongly in the inner ring, where both T1 and T2 values are the smallest. Similarly, when the T2* correction was applied, the inner ring experienced the largest change, though accuracy improved to varying degrees across the entire T1 range. As shown in Figures 46, even just two later echoes can be sufficient to provide a T2* estimate sufficient for correction, though the T2* correction improves with a better estimate of T2* (e.g., by using more echoes).

4.3. Averaging effect – acquisition speed vs accuracy

A comparison between the simulation and measured data for the SR-VFA-T2* method is shown in Figure 7. The moving average effect of radial acquisitions produces an alternating over- and under-estimation at locations in the phantom where the concentration of dopants has not diffused sufficiently for a smooth T1 transition. This produces the oscillations shown in Figures 3 and 6, as proven by the matching T1 positions of the oscillations in Figures 7b and 7c. Increasing rotation speed increases the T1 smoothing effect, and thus increases the amplitude of the oscillations in Figures 7b and 7c. After taking the average difference between the measured and simulated SR-VFA-T2* T1 percentage error measurements, a residual trace remains in Figure 7d that bears strong resemblance to the VFA mean T1 accuracy traces found in Figure 3f.

Figure 7:

Figure 7:

Simulation of dynamic rotation effects on SR-VFA-T2* T1 accuracy. Simulated signals were generated by inputting measured IR T1 data (solid black traces, plot A) into the steady state spoiled gradient echo signal equation and applying moving averages to simulate phantom rotation (see Methods section 3.5). SR-VFA-T2* T1 values were calculated from these signals. Colors denote different phantom rings, and line styles in plots A-C correspond to different rotational speeds. (A) - Mean T1 trace from ROIs around each phantom ring from inversion recovery data and SR-VFA-T2* simulated data at each rotational speed. (B & C) – Binned mean % error of SR-VFA-T2* T1 values at each rotational speed vs inversion recovery T1 values from the simulated data (B) and measured data from the dynamic rotation experiment (C). (D) – Difference plot between lines in B and C averaged over all speeds. The moving average effect of radial acquisitions combined with the nonlinear T1 gradient in the phantom is shown to produce the oscillations seen in single reference methods’ T1 accuracy.

4.4. Measurement Stability

Figures 8 and 9 show the precision of the B1+ mapping and signal across rotational positions respectively. The B1+ precision over the voxels covering the center of each ring was found to be (mean ± 2 SD of the histograms’ data in Figure 8) 0.014 ± 0.011 (max 0.031) in the stationary experiment and 0.012 ± 0.010 (max 0.028) in the dynamic rotation experiment. The worse precision found in the middle ring of the static experiment can also be seen in the slightly worse T1 precision for the VFA method’s middle ring (Figure 3c). In addition, the bright spots (higher SD) in the inner ring for the dynamic rotation experiment coincide with the positions of the low T1 portion of the ring across the four B1+ maps acquired.

Figure 8:

Figure 8:

2x SD of B1+ across the 4 B1+ maps acquired in the static (A) and dynamic rotation (B) phantom experiment. B1+ maps were acquired at [0°, 90°, 180°, 270°] in the static experiment (A) and at [60°, 149.5°, 237°, 303.3°] in the dynamic rotation experiment (B). Each histogram in (A) and (B) is normalized as a probability density function. Note the overall higher SD in the middle ring of the static experiment and the inner ring of the dynamic experiment.

Figure 9:

Figure 9:

Signal variation across successive rotations of the phantom during the high flip angle scan for Speed 3. ~4 full rotations are included, with signal data taken from a single circular ROI at the top of each ring. (A) – Precision of signal vs rotational position; (B) – Coefficient of variation box-and-whisker plot for each ring. Signal from one rotation to another is highly reproducible, with higher coefficient of variation in the outer rings due to higher T1 values in the outer rings, which produces lower signal values in the GRE acquisition.

5. Discussion

This study is the first, to the authors’ knowledge, to validate a T1 measurement method by direct comparison of dynamic quantitative T1 maps to gold standard IR T1 maps. This comparison was made possible by our design of a variably-doped cylindrical phantom combined with motorized rotation during image acquisition. In this way, accuracy could be compared simultaneously for a broad range of T1 values experiencing a broad range of T1 changes from their baseline values.

The rotating cylinder concept used in this work provides a repeatable model to validate dynamic T1 measurements against the gold standard. T1 values measured on a run-to-run basis are less repeatable during a contrast injection or focal heating due to changing tissue properties or difficulty of precise control of contrast injection or heating versus time at every point. However, the design concept used in this work allowed for repeatable highs and lows in signal and T1 over acquisition time and allowed for these to be captured at speeds varying between ~0.24 and 1.5 rotations per minute.

It was shown that SR-VFA T1 accuracy degrades with increasing echo time, such that changes in T1 from the baseline T1 value are underestimated. However, these inaccuracies were found to be correctable through the application of the T2* correction published previously, as shown in Figures 5 and 6.35 Additionally, it was shown that a substantial correction can be performed even with two later echoes, such that the resulting T1 is nearly as accurate as the Ts’s obtained without correction at the early echo times. Thus, the SR-VFA-T2* T1 mapping method, with only a single dynamic image per T1 map, could produce T1 values within 10% of gold standard IR measurements. Similar accuracy was obtained for changes in T1 (see Figure S2). Further, this study found that with the SR-VFA-T2* method, the combined echoes’ estimate of T1 had approximately equivalent accuracy and precision to that of the earliest echo used in the correction. The lack of improvement after combining echoes could be due to the correlation present between different echo times’ corrected T1 maps,35 and decreasing SNR with increasing echo time.

In the rotation experiment, we found that temporal averaging caused a predictable over- and under-estimation of true T1 at positions where the T1 changed with a non-constant slope. The oscillations in the T1 error curves in Figure 3 appear to be highly repeatable and appear to be caused by the fact that there is a rapidly changing slope of T1 values around each ring in the cylinder, as indicated by the T1 variation shown in Figure 7. In implementing a radial sliding window KWIC reconstruction, the measurements at the center of k-space are effectively averaged over a short time interval. When the T1 value changes with a non-constant slope during that interval, the calculated T1 from the resulting images may not match the true T1 value, the T1 at the center of the averaging window. This effect manifested as oscillations in the T1 error curves in Figures 3df, 6, and 7, and was visible across all rotation speeds, as shown in Figure 7. This temporal averaging oscillatory effect was simulated and removed in Figure 7bd, and the strong similarity between Figure 7d and the stationary VFA mean T1 accuracy traces in Figure 3f possibly indicates that any other bulk-motion-induced artifacts in our experiments were small in comparison with the effects of temporal blurring. As expected, increasing the speed of rotation had a negative impact on the T1 accuracy and precision, as shown in Figure 4 and Figure 7. We note that this oscillatory effect is intrinsic to our method of phantom preparation. However, this importantly indicates that for situations where T1 may change with non-constant slope over the measurement window, especially the time window for the acquisition of the center of k-space, accurate dynamic T1 mapping requires that either 1) the scan trajectory and acquisition parameters be chosen to shrink the measurement window until the T1 changes over that time can be reasonably approximated as linear, else 2) other compensatory techniques that incorporate prior knowledge about the dynamics of the T1 changes are used. Condition 1 was approximately met for the lowest rotation speed (see Figure 7a, solid-colored lines), and thus the SR-VFA-T2* method’s T1 accuracy and time-normalized precision were comparable to the stationary VFA T1 measurement for this case. With the use of a faster KWIC-compatible trajectory than the stack-of-radial measurements used in this work, accuracy in the higher rotation speeds would likely have been improved.

Since a low rotation speed is analogous to slow contrast injection or heating, both of which may be unacceptable, it is clear that fast, efficient trajectories will be needed in addition to the speed improvements afforded by SR-VFA T1 mapping. In the case of focused ultrasound ablation, where T1 changes in fat over the course of a ~20 second sonication could be used in combined PRF/T1 thermometry, a faster sequence than that used in this work would be required. However, this work indicates that given sufficient time resolution, SR-VFA-T2* T1 mapping can produce accurate, precise T1 values from a single dynamic image per dynamic T1 map.

Besides the effects of rotation just described, B1+ inaccuracies could account for most of the remaining T1 inaccuracy, due to our assumption of constant B1+ despite phantom rotation and due to the use of the AFI method with TR2 close to the minimum T1 being measured. For evaluation of the first assumption, Figure 8 shows the SD of the measured B1+ field throughout each experiment. It indicates that the B1+ maps remained approximately constant, with the precision between the four B1+ maps acquired at different positions being (mean ± 2 SD of the histograms’ data in Figure 8) 0.014 ± 0.011 for the static experiment and 0.012 ± 0.010 for the rotation experiment over the center of each phantom ring. A 1% error in B1+ produces approximately a 2% error in T1,43 so thus the error associated with noise in our B1+ mapping and/or slight B1+ changes with phantom rotation could account for errors in T1 accuracy as large as ~3–5% in the static experiment and ~2–5% in the dynamic rotation experiment. In regards to the accuracy of our AFI B1+ mapping implementation, we found a notable decrease in T1 accuracy at the lowest of T1 values (~150 ms) in the dynamic experiment, and 150 ms was used as the TR2 in our AFI B1+ mapping implementation. The AFI method is known to begin to underestimate B1+ when T1 is equal to or shorter than TR2,52 and thus it is possible that B1+ inaccuracy was systematically greater in this region, leading to a larger than normal overestimation of T1. This could be the reason for the bright spots on the inner ring for the dynamic rotation experiment, given that the low T1 values’ positions during each B1+ map acquisition line up well with the bright spots’ locations. However, even in these regions, T1 errors remained within ± ~10% of inversion recovery measurements. while the B1+ field was approximately constant over the course of these experiments, respiratory and other motion during in-vivo experiments could cause there to be a different B1+ field between the baseline and dynamic acquisitions. The effects of this kind of B1+ variation have not yet been studied and warrant further investigation in in vivo studies.

During the rotational registration of the ROIs for the SR-VFA to VFA and IR T1 comparisons for the dynamic rotation experiment, the time of beginning rotation was adjusted to minimize the error between IR T1 and SR-VFA T1 values. This was done due to the difficulty of determining the precise start time of rotation relative to the acquired k-space lines. Because of this, it is possible there is a systematic measurement error in opposite directions for increasing versus decreasing T1 values within the phantom’s rotation in this study’s favor. However, the similarly high level of agreement between the VFA, SR-VFA-T2*, and IR T1 measurements in the static experiment implies that this adjustment of rotation time was close to accurate.

Though this study validated SR-VFA-T2* T1 mapping in an agar phantom against IR T1 mapping directly, the in vivo case will likely not have such a direct comparison, and will have the added difficulty of mixed fat and water voxels. The phantom in this work did include a large portion of the physiological range of T1 in body tissues, but since VFA T1 mapping is sensitive to partial volume effects when fat and water are both present in imaged voxels, the derived SR-VFA and SR-VFA-T2* T1 mapping methods are also sensitive to these. Future studies and in vivo work will require fat-water separation techniques such as Dixon,57 IDEAL,58 or others59 to achieve optimal results.

In summary, this work represents the beginning of validation for the SR-VFA method in aqueous materials. Further study is needed for validation of the SR-VFA methods in fat and other non-aqueous tissues such as bone.

Supplementary Material

Supinfo

Acknowledgements

This work was funded by NIH grants R01 EB028316, R03 EB029204, R21 EB033638, and S10 OD018482. We also express gratitude to Robb Merrill for helpful conversations about 3D-printing.

Footnotes

Conflict of Interest

DLP is a co-inventor of the SR-VFA T1 method and, through the University of Utah, holds IP on this method (US10830853B1). LWH has a financial interest in the company 4D Surgical which develops MRI-compatible actuators and robotic systems and holds IP related to the MRI-compatible motor.

References

  • 1.Bluemke E, Bertrand A, Chu KY, et al. Using Variable Flip Angle (VFA) and Modified Look-Locker Inversion Recovery (MOLLI) T1 mapping in clinical OE-MRI. Magnetic Resonance Imaging. 2022;89:92–99. doi: 10.1016/j.mri.2022.03.001 [DOI] [PubMed] [Google Scholar]
  • 2.Featherstone AK, O’Connor JPB, Little RA, et al. Data-driven mapping of hypoxia-related tumor heterogeneity using DCE-MRI and OE-MRI. Magnetic Resonance in Medicine. 2018;79(4):2236–2245. doi: 10.1002/mrm.26860 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Li Z, Sun J, Hu X, et al. Assessment of liver fibrosis by variable flip angle T1 mapping at 3.0T. Journal of Magnetic Resonance Imaging. 2016;43(3):698–703. doi: 10.1002/jmri.25030 [DOI] [PubMed] [Google Scholar]
  • 4.Tadimalla S, Wilson DJ, Shelley D, et al. Bias, Repeatability and Reproducibility of Liver T1 Mapping With Variable Flip Angles. Journal of Magnetic Resonance Imaging. 2022;56(4):1042–1052. doi: 10.1002/jmri.28127 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Cassinotto C, Feldis M, Vergniol J, et al. MR relaxometry in chronic liver diseases: Comparison of T1 mapping, T2 mapping, and diffusion-weighted imaging for assessing cirrhosis diagnosis and severity. European Journal of Radiology. 2015;84(8):1459–1465. doi: 10.1016/j.ejrad.2015.05.019 [DOI] [PubMed] [Google Scholar]
  • 6.Tirkes T, Lin C, Fogel EL, Sherman SS, Wang Q, Sandrasegaran K. T1 mapping for diagnosis of mild chronic pancreatitis. Journal of Magnetic Resonance Imaging. 2017;45(4):1171–1176. doi: 10.1002/jmri.25428 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Wang M, Gao F, Wang X, et al. Magnetic resonance elastography and T1 mapping for early diagnosis and classification of chronic pancreatitis. Journal of Magnetic Resonance Imaging. 2018;48(3):837–845. doi: 10.1002/jmri.26008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Karamitsos TD, Piechnik SK, Banypersad SM, et al. Noncontrast T1 Mapping for the Diagnosis of Cardiac Amyloidosis. JACC: Cardiovascular Imaging. 2013;6(4):488–497. doi: 10.1016/j.jcmg.2012.11.013 [DOI] [PubMed] [Google Scholar]
  • 9.Puntmann VO, Peker E, Chandrashekhar Y, Nagel E. T1 Mapping in Characterizing Myocardial Disease. Circulation Research. 2016;119(2):277–299. doi: 10.1161/CIRCRESAHA.116.307974 [DOI] [PubMed] [Google Scholar]
  • 10.Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M. T1 Mapping: Basic Techniques and Clinical Applications. JACC: Cardiovascular Imaging. 2016;9(1):67–81. doi: 10.1016/j.jcmg.2015.11.005 [DOI] [PubMed] [Google Scholar]
  • 11.Jellis CL, Kwon DH. Myocardial T1 mapping: modalities and clinical applications. Cardiovasc Diagn Ther. 2014;4(2):126–137. doi: 10.3978/j.issn.2223-3652.2013.09.03 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Schelbert EB, Messroghli DR. State of the Art: Clinical Applications of Cardiac T1 Mapping. Radiology. 2016;278(3):658–676. doi: 10.1148/radiol.2016141802 [DOI] [PubMed] [Google Scholar]
  • 13.Jobst BJ, Triphan SMF, Sedlaczek O, et al. Functional Lung MRI in Chronic Obstructive Pulmonary Disease: Comparison of T1 Mapping, Oxygen-Enhanced T1 Mapping and Dynamic Contrast Enhanced Perfusion. PLOS ONE. 2015;10(3):e0121520. doi: 10.1371/journal.pone.0121520 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Yan Q, Yi Y, Shen J, et al. Preliminary study of 3 T-MRI native T1-mapping radiomics in differential diagnosis of non-calcified solid pulmonary nodules/masses. Cancer Cell International. 2021;21(1):539. doi: 10.1186/s12935-021-02195-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Bernhardt BC, Fadaie F, Vos de Wael R, et al. Preferential susceptibility of limbic cortices to microstructural damage in temporal lobe epilepsy: A quantitative T1 mapping study. NeuroImage. 2018;182:294–303. doi: 10.1016/j.neuroimage.2017.06.002 [DOI] [PubMed] [Google Scholar]
  • 16.Vrenken H, Geurts JJG, Knol DL, et al. Whole-Brain T1 Mapping in Multiple Sclerosis: Global Changes of Normal-appearing Gray and White Matter. Radiology. 2006;240(3):811–820. doi: 10.1148/radiol.2403050569 [DOI] [PubMed] [Google Scholar]
  • 17.Müller A, Jurcoane A, Kebir S, et al. Quantitative T1-mapping detects cloudy-enhancing tumor compartments predicting outcome of patients with glioblastoma. Cancer Medicine. 2017;6(1):89–99. doi: 10.1002/cam4.966 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Haddock B, Larsson HBW, Hansen AE, Rostrup E. Measurement of brain oxygenation changes using dynamic T1-weighted imaging. NeuroImage. 2013;78:7–15. doi: 10.1016/j.neuroimage.2013.03.068 [DOI] [PubMed] [Google Scholar]
  • 19.Arnold JFT, Fidler F, Wang T, Pracht ED, Schmidt M, Jakob PM. Imaging lung function using rapid dynamic acquisition of T1-maps during oxygen enhancement. Magn Reson Mater Phy. 2004;16(5):246–253. doi: 10.1007/s10334-004-0034-z [DOI] [PubMed] [Google Scholar]
  • 20.Sanz-Requena R, Martí-Bonmatí L, Pérez-Martínez R, García-Martí G. Dynamic contrast-enhanced case-control analysis in 3T MRI of prostate cancer can help to characterize tumor aggressiveness. European Journal of Radiology. 2016;85(11):2119–2126. doi: 10.1016/j.ejrad.2016.09.022 [DOI] [PubMed] [Google Scholar]
  • 21.Rudisch A, Kremser C, Judmaier W, Zunterer H, DeVries AF. Dynamic contrast-enhanced magnetic resonance imaging: a non-invasive method to evaluate significant differences between malignant and normal tissue. European Journal of Radiology. 2005;53(3):514–519. doi: 10.1016/j.ejrad.2004.06.002 [DOI] [PubMed] [Google Scholar]
  • 22.Todd N, Diakite M, Payne A, Parker DL. In vivo evaluation of multi-echo hybrid PRF/T1 approach for temperature monitoring during breast MR-guided focused ultrasound surgery treatments. Magnetic Resonance in Medicine. 2014;72(3):793–799. doi: 10.1002/mrm.24976 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Todd N, Diakite M, Payne A, Parker DL. Hybrid proton resonance frequency/T1 technique for simultaneous temperature monitoring in adipose and aqueous tissues. Magnetic Resonance in Medicine. 2013;69(1):62–70. doi: 10.1002/mrm.24228 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Zhang L, Armstrong T, Li X, Wu HH. A variable flip angle golden-angle-ordered 3D stack-of-radial MRI technique for simultaneous proton resonant frequency shift and T1-based thermometry. Magnetic Resonance in Medicine. 2019;82(6):2062–2076. doi: 10.1002/mrm.27883 [DOI] [PubMed] [Google Scholar]
  • 25.Hey S, de Smet M, Stehning C, et al. Simultaneous T1 measurements and proton resonance frequency shift based thermometry using variable flip angles. Magnetic Resonance in Medicine. 2012;67(2):457–463. doi: 10.1002/mrm.22987 [DOI] [PubMed] [Google Scholar]
  • 26.Look DC, Locker DR. Time Saving in Measurement of NMR and EPR Relaxation Times. Review of Scientific Instruments. 2003;41(2):250–251. doi: 10.1063/1.1684482 [DOI] [Google Scholar]
  • 27.Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magnetic Resonance in Medicine. 2004;52(1):141–146. doi: 10.1002/mrm.20110 [DOI] [PubMed] [Google Scholar]
  • 28.Piechnik SK, Ferreira VM, Dall’Armellina E, et al. Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. Journal of Cardiovascular Magnetic Resonance. 2010;12(1):69. doi: 10.1186/1532-429X-12-69 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Weingärtner S, Akçakaya M, Basha T, et al. Combined saturation/inversion recovery sequences for improved evaluation of scar and diffuse fibrosis in patients with arrhythmia or heart rate variability. Magnetic Resonance in Medicine. 2014;71(3):1024–1034. doi: 10.1002/mrm.24761 [DOI] [PubMed] [Google Scholar]
  • 30.Fram EK, Herfkens RJ, Johnson GA, et al. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magnetic Resonance Imaging. 1987;5(3):201–208. doi: 10.1016/0730-725X(87)90021-X [DOI] [PubMed] [Google Scholar]
  • 31.Deoni SCL, Rutt BK, Peters TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magnetic Resonance in Medicine. 2003;49(3):515–526. doi: 10.1002/mrm.10407 [DOI] [PubMed] [Google Scholar]
  • 32.Deoni SCL, Peters TM, Rutt BK. Determination of optimal angles for variable nutation proton magnetic spin-lattice, T1, and spin-spin, T2, relaxation times measurement. Magnetic Resonance in Medicine. 2004;51(1):194–199. doi: 10.1002/mrm.10661 [DOI] [PubMed] [Google Scholar]
  • 33.Dietrich O, Freiermuth M, Willerding L, Reiser MF, Peller M. Flip angle-optimized fast dynamic T1 mapping with a 3D gradient echo sequence. Magn Reson Med. 2015;73(3):1158–1163. doi: 10.1002/mrm.25199 [DOI] [PubMed] [Google Scholar]
  • 34.Schabel MC, Morrell GR. Uncertainty in T1 mapping using the variable flip angle method with two flip angles. Phys Med Biol. 2008;54(1):N1–N8. doi: 10.1088/0031-9155/54/1/N01 [DOI] [PubMed] [Google Scholar]
  • 35.Malmberg MA, Odéen H, Parker DL. Effects of T2* on accuracy and precision of dynamic T1 measurements using the single reference variable flip angle method: a simulation study. Medical Physics. 2022;49(4):2396–2412. doi: 10.1002/mp.15476 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Yarnykh VL. Optimal radiofrequency and gradient spoiling for improved accuracy of T1 and B1+ measurements using fast steady-state techniques. Magnetic Resonance in Medicine. 2010;63(6):1610–1626. doi: 10.1002/mrm.22394 [DOI] [PubMed] [Google Scholar]
  • 37.Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB. On the accuracy of T1 mapping: Searching for common ground. Magnetic Resonance in Medicine. 2015;73(2):514–522. doi: 10.1002/mrm.25135 [DOI] [PubMed] [Google Scholar]
  • 38.Preibisch C, Deichmann R. Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magnetic Resonance in Medicine. 2009;61(1):125–135. doi: 10.1002/mrm.21776 [DOI] [PubMed] [Google Scholar]
  • 39.Baudrexel S, Nöth U, Schüre JR, Deichmann R. T1 mapping with the variable flip angle technique: A simple correction for insufficient spoiling of transverse magnetization. Magnetic Resonance in Medicine. 2018;79(6):3082–3092. doi: 10.1002/mrm.26979 [DOI] [PubMed] [Google Scholar]
  • 40.Svedin BT, Payne A, Parker DL. Simultaneous proton resonance frequency shift thermometry and T1 measurements using a single reference variable flip angle T1 method. Magnetic Resonance in Medicine. 2019;81(5):3138–3152. doi: 10.1002/mrm.27643 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Brookes JA, Redpath TW, Gilbert FJ, Murray AD, Staff RT. Accuracy of T1 measurement in dynamic contrast-enhanced breast MRI using two- and three-dimensional variable flip angle fast low-angle shot. Journal of Magnetic Resonance Imaging. 1999;9(2):163–171. doi: [DOI] [PubMed] [Google Scholar]
  • 42.Wang HZ, Riederer SJ, Lee JN. Optimizing the precision in T1 relaxation estimation using limited flip angles. Magnetic Resonance in Medicine. 1987;5(5):399–416. doi: 10.1002/mrm.1910050502 [DOI] [PubMed] [Google Scholar]
  • 43.Cheng HLM, Wright GA. Rapid high-resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity. Magnetic Resonance in Medicine. 2006;55(3):566–574. doi: 10.1002/mrm.20791 [DOI] [PubMed] [Google Scholar]
  • 44.Wang J, Mao W, Qiu M, Smith MB, Constable RT. Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities. Magnetic Resonance in Medicine. 2006;56(2):463–468. doi: 10.1002/mrm.20947 [DOI] [PubMed] [Google Scholar]
  • 45.Wang J, Qiu M, Kim H, Constable RT. T1 Measurements incorporating flip angle calibration and correction in vivo. Journal of Magnetic Resonance. 2006;182(2):283–292. doi: 10.1016/j.jmr.2006.07.005 [DOI] [PubMed] [Google Scholar]
  • 46.Gopalan K, Tamir JI, Arias AC, Lustig M. Quantitative anatomy mimicking slice phantoms. Magn Reson Med. 2021;86(2):1159–1166. doi: 10.1002/mrm.28740 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Baron P, Deckers R, Knuttel FM, Bartels LW. T1 and T2 temperature dependence of female human breast adipose tissue at 1.5 T: groundwork for monitoring thermal therapies in the breast. NMR in Biomedicine. 2015;28(11):1463–1470. doi: 10.1002/nbm.3410 [DOI] [PubMed] [Google Scholar]
  • 48.Thangavel K, Saritaş EÜ. Aqueous paramagnetic solutions for MRI phantoms at 3 T: A detailed study on relaxivities. Turk J Elec Eng & Comp Sci. 2017;25(3):2108–2121. [Google Scholar]
  • 49.Hofstetter LW, Hadley JR, Merrill R, Pham H, Fine GC, Parker DL. MRI-compatible electromagnetic servomotor for image-guided medical robotics. Commun Eng. 2022;1(1):1–10. doi: 10.1038/s44172-022-00001-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Svedin BT, Payne A, Bolster BD, Parker DL. Multiecho pseudo-golden angle stack of stars thermometry with high spatial and temporal resolution using k-space weighted image contrast. Magn Reson Med. 2018;79(3):1407–1419. doi: 10.1002/mrm.26797 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Song HK, Dougherty L. k-Space weighted image contrast (KWIC) for contrast manipulation in projection reconstruction MRI. Magnetic Resonance in Medicine. 2000;44(6):825–832. doi: [DOI] [PubMed] [Google Scholar]
  • 52.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magnetic Resonance in Medicine. 2007;57(1):192–200. doi: 10.1002/mrm.21120 [DOI] [PubMed] [Google Scholar]
  • 53.Beatty PJ, Nishimura DG, Pauly JM. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Transactions on Medical Imaging. 2005;24(6):799–808. doi: 10.1109/TMI.2005.848376 [DOI] [PubMed] [Google Scholar]
  • 54.Pipe JG, Menon P. Sampling density compensation in MRI: Rationale and an iterative numerical solution. Magnetic Resonance in Medicine. 1999;41(1):179–186. doi: [DOI] [PubMed] [Google Scholar]
  • 55.Du YP, Parker DL, Davis WL, Cao G. Reduction of partial-volume artifacts with zero-filled interpolation in three-dimensional MR angiography. Journal of Magnetic Resonance Imaging. 1994;4(5):733–741. doi: 10.1002/jmri.1880040517 [DOI] [PubMed] [Google Scholar]
  • 56.Barral JK, Gudmundson E, Stikov N, Etezadi-Amoli M, Stoica P, Nishimura DG. A robust methodology for in vivo T1 mapping. Magnetic Resonance in Medicine. 2010;64(4):1057–1067. doi: 10.1002/mrm.22497 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Dixon WT. Simple proton spectroscopic imaging. Radiology. 1984;153(1):189–194. doi: 10.1148/radiology.153.1.6089263 [DOI] [PubMed] [Google Scholar]
  • 58.Glover GH. Multipoint dixon technique for water and fat proton and susceptibility imaging. Journal of Magnetic Resonance Imaging. 1991;1(5):521–530. doi: 10.1002/jmri.1880010504 [DOI] [PubMed] [Google Scholar]
  • 59.Ma J Dixon techniques for water and fat imaging. Journal of Magnetic Resonance Imaging. 2008;28(3):543–558. doi: 10.1002/jmri.21492 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supinfo

RESOURCES