Current status of transcatheter intervention for complex right ventricular outflow tract abnormalities

Yoshiyuki Kagiyama 1,2, Damien Kenny 1, Ziyad M Hijazi 3
PMCID: PMC10886730  PMID: 38404661

Abstract

Various transcatheter interventions for the right ventricular outflow tract (RVOT) have been introduced and developed in recent decades. Transcatheter pulmonary valve perforation was first introduced in the 1990s. Radiofrequency wire perforation has been the approach of choice for membranous pulmonary atresia in newborns, with high success rates, although complication rates remain relatively common. Stenting of the RVOT is a novel palliative treatment that may improve hemodynamics in neonatal patients with reduced pulmonary blood flow and RVOT obstruction. Whether this option is superior to other surgical palliative strategies or early primary repair of tetralogy of Fallot remains unclear. Transcatheter pulmonary valve replacement has been one of the biggest innovations in the last two decades. With the success of the Melody and SAPIEN valves, this technique has evolved into the gold standard therapy for RVOT abnormalities with excellent procedural safety and efficacy. Challenges remain in managing the wide heterogeneity of postoperative lesions seen in RVOT, and various technical modifications, such as pre-stenting, valve ring modification, or development of self-expanding systems, have been made. Recent large studies have revealed outcomes comparable to those of surgery, with less morbidity. Further experience and multicenter studies and registries to compare the outcomes of various strategies are necessary, with the ultimate goal of a single-step, minimally invasive approach offering the best longer-term anatomical and physiological results.

Introduction

Approximately 20% of patients with congenital heart disease (CHD) have multi-level right ventricular outflow tract (RVOT) abnormalities1 and the incidence of these specific types of CHD has been growing over the past three decades2. Until recently, only surgical repair or palliative therapy was available for complex RVOT abnormal lesions, such as pulmonary valve atresia, severe RVOT obstruction with tetralogy of Fallot, or postoperative pulmonary regurgitation or obstruction. However, with the evolution of radiofrequency perforation of the pulmonary valve, right ventricular outflow tract stenting, and transcatheter valve implantation, these patients can now receive transcatheter intervention as a less invasive and more effective alternative therapy. In this review, we discuss the current status of three representative transcatheter interventions for complex RVOT abnormalities.

Transcatheter pulmonary valve perforation

Background

Transcatheter perforation of the atretic membranous pulmonary valve was introduced in the 1990s as an alternative to surgical RVOT reconstruction for patients with pulmonary atresia and intact ventricular septum (PAIVS)1,3–6. Pulmonary circulation is dependent on the patent ductus arteriosus (PDA) in PAIVS patients, and the initial step of treatment for such patients is the establishment of adequate pulmonary blood flow (PBF). Perforation of the pulmonary valve is one of these options by creating RV to pulmonary artery (PA) continuity, with the ultimate goal of biventricular repair. Although hemodynamically favorable biventricular repair is the goal, it is difficult to achieve in patients with hypoplastic RV or coronary artery abnormalities. In the early days, because of the rarity of the condition and the morphological heterogeneity of PAIVS, the treatment strategy was not standardized and profoundly different between institutions7. With growing recognition about the risks of “aggressive” biventricular repair, currently approximately 40–60% of PAIVS patients receive this procedure to pursue biventricular repair7–9.

Preprocedural evaluation

The initial treatment strategy is selected based on factors such as the presence and extent of the ventriculocoronary connections, morphology of the RV and tricuspid valve (TV), and type of pulmonary atresia.

The existence and extent of the ventriculocoronary connections should be clarified before pulmonary valve perforation. Suprasystemic RV pressure from fetal life acts to maintain embryological ventriculocoronary connections, that is, sinusoidal communications with collision of conflicting blood flow from the aorta and RV inside the coronary artery (CA) cause potential distortion and stenosis of the coronary sinusoids1. In the presence of right ventricular dependent coronary artery circulation (RVDCC), ventriculocoronary connections with stenotic or obstructive antegrade CAs, reduction of blood flow or pressure in the RV may cause myocardial ischemia and death10. Therefore, patients with a broad area of myocardium supplied by the RVDCC are recognized as contraindications for valve perforation and subsequent RV decompression and are considered candidates for univentricular repair or heart transplantation11–13.

It is important to objectively evaluate RV size, diameter, and z-value of the TV on echocardiography5,14. The ratio of the TV to the mitral valve, RV inlet length z-score, and diastolic RV area z-score were also useful15. According to these studies, a TV z-score of over −2.5 SD is recognized as a good marker for biventricular repair and with TV z-score of −4.5 to −2.5 SD considered a borderline RV. In patients with borderline RV size, a sufficient increase in PBF immediately after successful valve perforation is not guaranteed, and some patients require an additional PBF source, such as a surgical shunt or stent into the PDA or RVOT8,16. However, it is well known that RV growth is favorable after pulmonary valve perforation and RV decompression17–19. Even if RV growth is insufficient to support full cardiac output due to the limited size or reduced compliance20, some patients procced to achieve a one-and-a-half ventricular repair by placing a bidirectional Glenn anastomosis. Therefore, in patients without RVDCC, an attempt at initial valve perforation may be worthwhile, even with a borderline RV8. Muscular-type pulmonary atresia with poorly developed infundibulum or severely dysplastic or Ebstein-like TV is a morphological contraindication for transcatheter perforation. Figure 1 shows the flowchart of the current treatment strategy for PAIVS.

Figure 1. Flowchart of current treatment strategy for PAIVS.

Figure 1.

Abbreviations: PBF, pulmonary blood flow; PDA, patent ductus arteriosus; RV, right ventricle; RVD, right ventricular decompression; TV, tricuspid valve.

Valve perforation and balloon dilation

This procedure (Figure 2) is usually performed under general anesthesia, with ductal patency maintained with prostaglandin E1 infusion. Prior to pulmonary valvular perforation, the nature of pulmonary atresia and RV hypoplasia, size of the pulmonary valve plate, infundibulum, RV, and PDA were evaluated using angiography. RVDCC should be excluded with RV angiography and/or aortography15. A 4 or 5 Fr Judkins right coronary catheter is maneuvered into the infundibulum via an antegrade approach. Location of the atretic pulmonary valve is confirmed with intermittent contrast injection and the catheter is placed to enface the center of the valve. Placing another angiographic or snare catheter into the main pulmonary artery across the PDA via a retrograde approach enables coaxial central alignment of the catheters.

Figure 2. Pulmonary valve perforation.

Figure 2.

(A) Preprocedural right ventricular outflow angiography shows an atretic pulmonary valve (black arrowheads) without antegrade flow. In frame B, a radiofrequency wire (arrow) perforated and passed the atretic pulmonary valve within a loop snare which was advanced retrogradely via the patent ductus arteriosus. The wire was snared and pulled into the descending aorta to establish a wire loop (C). An angioplasty balloon was inflated over the wire at the pulmonary valve (D). (E) Post Pulmonary angiography status post pulmonary valve perforation (E) and (F) post placement of the patent ductus arteriosus stent (white arrowheads).

Initially, transcatheter valve perforation equipment included a laser wire, the stiff end of a coronary wire, and radiofrequency (RF) wire. Currently, RF wire, which is designed to create a controlled hole in soft tissue by delivering radiofrequency (RF) energy, is widely utilized, with a high success rate of 91–94% and low complication rates21,22. A Nykanen RF wire (Baylis Medical, Montreal, Canada) is the most commonly used wire with 0.016” rounded tip and 0.024” wire covered with Teflon insulation22. To perforate the atretic valve, the traditional application of 3-10 Watts/sec for 2-5 s RF energy has been recommended23. However, a recent multicenter retrospective study reported that higher RF energy increased the risk of cardiac perforation24, therefore minimal optimal application of RF energy may be necessary. The high cost and low availability of RF generators and wires are other important considerations, particularly in developing countries.

With the refinement of chronic total occlusion (CTO) hardware technology, mechanical valve perforation using a CTO wire has been introduced as a technique with lower cost and risk of complications25–28. Because CTO wires are specially designed to penetrate occluded coronary arteries, they can provide better penetrability, straightforward pushability, and stability during perforation, which can reduce catheter displacement and subsequent cardiac perforation16. On the contrary, success rate in previous reports was as high as RF perforation26,27,29. The use of coronary microcatheters has been reported to enhance stabilization and increase success rates30.

Once the RVOT to PA continuity is established and confirmed with angiography, a microcatheter is advanced into the main PA and the RF wire is replaced by a 0.014” coronary wire. Thereafter, over this wire in the distal PA or the descending aorta through the PDA, serial balloon dilation of the pulmonary membrane to up to 140% of the valve annulus size or 7-8 mm diameter will be performed to achieve adequate PBF and RV decompression. To enhance trackability of the balloon system over the wire, many operators prefer to snare the wire in the descending aorta31.

Additional procedure or re-intervention

If the patient remains deeply cyanotic (saturation less than 80%), PDA stenting or surgical aorto-pulmonary shunt creation may be performed to augment the PBF. This could be performed in the same setting or several weeks after the initial valve perforation. Generally, the RV remains restrictive for a few weeks and it is difficult to predict which patients will require an additional source of PBF. Previous reports suggest that almost half of the patients required additional procedures to augment PBF after initial right ventricular decompression. Risk factors associated with failure to achieve a biventricular circulation include: small RV area, tricuspid valve hypoplasia, small RV to LV size ratio, and mild or lesser degrees of tricuspid valve regurgitation (Table 1)17,18,32,33.

Table 1. Recent studies which advocated pre-procedural predictors of final circulatory status.

Authors, year Petit 2017 Chen 2018 Maskatia 2018 * Yoldaş 2020
Patients number (group) Group assignment n 99 n 36, (26 / 5) BVR / not BVR n 81 n 31, (16 / 15) RVD success /not
Tripartite : Bipartite not shown 83% : 17% not shown 65% : 35%
TV annulus diameter, mm median 10.5 not shown not shown not shown
TV z-score median −0.86 mean −3.34 vs −6.86a median -0.64 mean -0.64 vs -1.18
RV area, cm 2 or area index median 2.0 not shown median 12.1 (/m2) mean 10.7 vs 8.8 (/m2)
TV/MV ratio median 0.85 median 0.80 not shown mean 0.84 vs 0.92
TR ≥ moderate 69% not shown 61% not shown
Procedural success rate 94% 92% 94% 48%
In-hospital mortality 0% 7% (3.5% by sepsis) 0% 19% (all by sepsis)
Require re-dilatation 64% not shown not shown 54%
Require additional PBF 45% 28% 46% 52%
Achievement of BVR 83% 84% 86% 48%
Factors associated with BVR achievement not advocated TV/MV ratio > 0.79 RV area index > 6 cm2/m2 TV/MV > 0.85 TV z-score > -1
Factors associated with failure to achieve BVR ≤mild TRb, lower RV area Bipartite / history of BAS not advocated not advocated

Notes.

Abbreviations
BVR
biventricular repair
PBF
pulmonary blood flow
RV
right ventricle
RVD
right ventricular decompression
RVOT
right ventricular outflow tract
TR
tricuspid regurgitation
TV
tricuspid valve
*

post hoc analysis data of Petit’s.

a

< − 2.5 SD 19%, −5 SD 50%.

b

not significant in its post hoc analysis.

Mortality and morbidity

Only a few studies have summarized the complication rates after this procedure. Perforation of the RVOT or PA is the most hazardous complication with prevalence of 10–19%6,24,34. Petit et al. reported that this complication was associated with the use of an RF wire and a greater degree of RF energy24. They also mentioned that arrhythmias were the most common complications, occurring in 61% of patients and requiring intervention in their cohort. Postprocedural major adverse cardiac events such as death, stroke, cardiopulmonary resuscitation, or mechanical circulatory support occurred in approximately 20% of patients in a recent multicenter study; nonetheless, this percentage was lower than that after single ventricular palliation, even when RVDCC patients were excluded from the analysis9.

Summary

Wire perforation to the atretic pulmonary valve is an established therapy with excellent results. However, indication remains controversial because of the morphological and physiological diversity of PAIVS. Both RF and CTO wires are effective; however, further development of knowledge and experience with CTO wires is necessary. Ultimately, it may be possible to provide transcatheter pulmonary valve replacement in these patients later in life, thereby providing complete percutaneous repair for this condition in some patients. Additionally, this technique may also be performed as palliative treatment for pulmonary atresia with ventricular septal defects in combination with RVOT stenting, particularly in the setting of complex MAPCA’s35,36.

Right ventricular outflow tract stenting

Background

The modified Blalock-Taussig-Thomas shunt (mBTTS) has largely been superseded by PDA stenting for duct-dependent pulmonary blood flow lesions as it has been shown to be associated with lower morbidity and better pulmonary arterial growth37–39. As an extension of this approach, forward PBF may reduce the risk of pulmonary over-circulation or diastolic run-off seen in “shunt physiology” with pulsatile flow optimizing PA growth40,41, RVOT stenting in neonates with symptomatic tetralogy has evolved42,43. Since its introduction, excellent outcomes, such as low mortality and favorable PA growth, have been reported42–45.

Indication

RVOT stenting is usually performed in neonates and small infants with symptomatic (increasing cyanosis or hypercyanotic spells) tetralogy of Fallot and its variants. It may be especially beneficial in high-risk groups, namely patients with low birth weight, small or non-confluent PAs, severe genetic disorders, or other extracardiac comorbidities39. There has been controversy about the superiority of RVOT stenting to PDA stenting, with the theoretical advantage of forward PBF mentioned above, while drawbacks of RVOT stenting relate to the impact of the stent on definitive surgical repair35.

Procedure

Under general anesthesia, RVOT stenting (Figure 3) is performed from a femoral or jugular venous approach, although the subxiphoid hybrid approach may be considered in patients <2 kgs. Right ventricular angiography, with tube angulation of 30° right anterior oblique with 20° cranial tilt along with a straight lateral is performed to delineate the anatomy. A 0.014” wire is place in the distal branch pulmonary artery and may be “locked” in place with 3–4 clockwise turns. Some centers have shifted away from using longer sheaths because of the potential for hemodynamic instability, particularly in those with hypercyanotic spells. Usually, a stent size of 1–2 mm larger than the diameter of the RVOT in the diastolic phase is appropriate. The stent type may depend on the cath lab’s stocking options. Either a bare metal stent, or a covered stent to mitigate against neo-intimal proliferation, can be used42,43,45,46. If a long sheath is not used, angiography can be performed using a 4Fr pigtail catheter retrogradely placed in the RV from the aorta47. In smaller infants, transthoracic echocardiography can be used to position the stent. The RVOT should be fully covered with a single long or tandem stent to prevent recurrent progressive muscular obstruction42.

Figure 3. Right ventricular outflow stenting.

Figure 3.

Upper row: during stent deployment. Lower row: post stent deployment. Abbreviations: LAO, left anterior oblique; RAO, right anterior oblique.

Postprocedural outcome

There have been a limited number of large studies, and no randomized prospective studies examining the outcomes of RVOT stenting in infants. Currently, two large studies comparing RVOT stenting with surgical procedures from Toronto ( n = 42, including 9 TOF with pulmonary atresia)35 and Birmingham ( n = 60, all TOF)45 are available. According to the systematic review which included these two studies and other relatively small case series48, the reported successful stent placement rates were 93.6% (95% confidence interval = CI 89.6%–96.2%) and the overall improvement in oxygen saturation was 20.1% (95% CI 15.8%–25.3%). The Birmingham group also illustrated more symmetric branch PA growth in the RVOT stenting group (RPA z-score improved from −2.28 to −0.72, LPA z-score −2.08 to −0.05) compared with the mBTTS group (an increase of PA z-score in the RVOT stenting group was higher as 0.749 for the RPA and 0.599 for the LPA than that of the mBTTS group)45.

Procedural complications such as stent migration, tricuspid injury, bradyarrhythmias, and aortic deformation have been reported44,46,49,50. Although coronary abnormalities exist in 5–7% of TOF patients1, previous reports identified up to 10 patients with an abnormal coronary course; however, none of these demonstrated coronary ischemia related to RVOT stenting51. Mortality rates after RVOT stenting have been reported to be 2–5%35,44, similar to those after mBTTS48. According to the Toronto report, mortality after RVOT stenting (2/42) was similar to that after early (<3 months) surgical repair and elective (3-11 months) primary surgical repair (1/45). It should be noted that these studies may have encompassed selection bias, namely, the patients in the RVOT stenting groups in these studies had more severe status such as smaller PAs, multiple comorbidities35, prematurity, small-for-gestational-age, or gut/brain abnormalities44 than other groups. No direct comparative studies have been conducted on RVOT stenting and surgical repair in high-risk patients.

Bridging to the repair therapy

Re-intervention rates after RVOT stenting have been reported to be 41–56%, higher than those after mBTTS palliation44 or after surgical repair performed between 3-11 months (no difference between the RVOT stenting group and the early surgical groups)35. However, a further study from Toronto demonstrated that the risk of surgical reoperation was associated with early surgical repair, and not with the early catheter RVOT palliation52. Most of the reinterventions were transcatheter in nature due to RVOT obstruction (balloon dilation or repeat stent placement)35,44,50. Bare metal stent-induced tissue reaction and fibrous tissue ingrowth may have contributed to this obstruction. It is unclear whether drug-eluting stents or bioresorbable stents mitigate this reaction44,53.

The technical burden of definitive surgical repair following RVOT stenting is another concern. RVOT stents may require a transannular patch (TAP) repair procedure and/or relatively broad RVOT resection to retrieve the stent. In the Toronto report, none of the patients after RVOT stenting underwent valve-sparing repair35 whereas 41% of the early repair (TOF with pulmonary stenosis) group achieved valve-sparing surgery. However, there were differences in baseline pulmonary valvular size between the two groups (median z-score of −6.6 in the former group vs −4.4 in the latter group). Furthermore, there was no difference in the rates of valve-sparing surgery or pulmonary valvular size between the RVOT stenting group and the early repair for TOF with pulmonary atresia group. A similar result was shown in the Birmingham report, as a valve-sparing surgery rate of 7% after RVOT stenting was not significantly different from that of post-mBTTS patients (10%)44. Indeed, the Birmingham group moved to subvalvular stent placement to avoid the pulmonary valve; however, the rate of transannular patch placement remained high. Stent removal at the time of surgery may be influenced by the length of time the stent remains in place. Complete retrieval of the RVOT stent was reported in 95% of patients approximately 6 months after initial stent implantation, whereas in another report, only 47% of patients underwent complete retrieval at a median of 220 days following implantation53. It is unclear whether covered stent implantation reduces intimal ingrowth and ameliorates stent removal at the time of surgical repair.

Summary

Although RVOT stenting is a novel technique predominantly performed in high-risk patients, its safety and efficacy have been demonstrated. However, high reintervention rates and detrimental effects on the final surgical repair remain unresolved problems.

Transcatheter pulmonary valve replacement (TPVR)

Historical overview

After surgical repair of RVOT abnormalities, typically represented by TAP RVOT reconstruction for TOF or RV to PA conduit implantation for pulmonary atresia, most patients have RV remodelling secondary to pulmonary regurgitation (PR) or multi-level RVOT obstruction. These lesions are usually progressive, and consequent RV dilation or dysfunction is irreversible and leads to adverse clinical outcomes, such as heart failure, refractory tachyarrhythmias, or sudden cardiac death (SCD)54–56. Therefore, prophylactic repair of these lesions, such as pulmonary valve replacement, is essential. However, even a successful surgical valve replacement may not be definitive, and patients usually require several lifetime surgical valve replacements. Moreover, an increased number of open-heart surgeries are associated with morbidity and mortality57–59.

In 2000, Bonhoeffer et al. reported the first case of TPVR using a bovine internal jugular vein with a native valve mounted on a platinum iridium stent60, with the following case series demonstrating excellent procedural outcomes61,62. This technique led to the establishment of the Melody Transcatheter Pulmonary Valve (Medtronic Inc., Minneapolis, MN, USA)63, and the device received the European CE mark in 2006 and Food and Drug Administration (FDA) approval in the United States in 2010. Since then, with 20 years of cumulative experience, technical advances, and the introduction or invention of other novel valves, TPVR is currently a widely accepted, less invasive alternative repair treatment to SPVR, reducing the number of open-heart surgeries required during the patient’s lifetime.

Preprocedural evaluation

Currently, TPVR is performed using indications similar to those used in SPVR. For patients who have repaired CHD, such as TOF with cardiovascular symptoms attributable to significant PR64–66 or RVOT obstruction64,65, most clinical guidelines recommend pulmonary valve replacement (PVR). The European Society of Cardiology (ESC) guidelines specifically recommend performing TPVR in patients with non-native RVOT morphology65. However, the treatment indications for asymptomatic patients with significant PR or RVOT obstruction are controversial. Valve replacement is considered in cases of severe RV enlargement with a significantly elevated right ventricular end-diastolic volume (RVEDV) index to body surface area, right ventricular end-systolic volume (RVESV), or dysfunction of the RV on cardiac magnetic resonance imaging (CMR) (see Table 2).

Table 2. Recommendations for pulmonary valve replacement for repaired tetralogy of Fallot.

ACC / AHA 2018 ESC 2020 CCS 2022
Recommendation Indication for pulmonary valve replacement
Class I, or strong in CCS = symptomatic patients ≥moderate PR severe PR* and/or ≥ moderate RVOTOa TPVR is preferable than SPVR in ”non-native” RVOT patients ≥ severe PRb
Class IIa, or weak in CCS = asymptomatic patients ≥moderate PR with any 2 of RV enlargement ≥ mild LV or RV dysfunction stenosis RVSP ≥ 2/3 systemic objective exercise intolerance severe PR and/or RVOTO with any of progressive RV enlargement RVOTO: RVSP > 80 mmHg progressive RV dysfunction objective exercise intolerance severe PR with RV enlargement severe PR with any of concomitant lesion lesion require intervention sustained VT
Description
RV enlargement RVEDV ≥ 160 ml/m2 and/or RVESV ≥ 80 ml/m2 and/or RVEDV ≥ 2 x LVEDV RVEDV ≥ 160 ml/m2 and/or RVESV ≥ 80 ml/m2 and/or progression of TR ≥ moderate RVEDV ≥ 160 ml/m2 and/or RVESV ≥ 80 ml/m2

Notes.

Abbreviations
ACC
American College of Cardiology
AHA
American Heart Association
CCS
Canadian Cardiovascular Society
CMR
cardiac magnetic resonance imaging
ESC
European Society of Cardiology
LV
left ventricle
LVEDV
left ventricular end diastolic volume
PR
pulmonary regurgitation
RV
right ventricle
RVEDV
right ventricular end diastolic volume
RVESV
right ventricular end systolic volume
RVOTO
right ventricular outflow tract obstruction
RVSP
right ventricular systolic pressure
SPVR
surgical pulmonary valve replacement
TPVR
transcatheter pulmonary valve replacement
TR
tricuspid regurgitation
VT
ventricular tachycardia
*

Regurgitant fraction by CMR > 30–40%.

a

Peak velocity > 3 m/s.

b

Severe PR on echocardiography or a regurgitation fraction > 25% on CMR.

Initially, TPVR was indicated for patients aged ≥5 years and weighing ≥30 kg67 because of the large-calibre delivery system via the femoral vein; nevertheless, TPVR via femoral access is well tolerated for patients weighing <30 kg68. TPVR can be performed for a very small patient (e.g., <20 kg) with limited venous access, such as venous obstruction, via the transjugular, transsubclavian68,69, or transhepatic venous approach70, technical challenges remain, and hybrid pulmonary valve replacement can be more straightforward for a very small patient, particularly with complex anatomy of the RVOT71–73 or in a bailout situation74–76.

Understanding the anatomy of the valve landing zone with surgical history taking and multimodality imaging such as computed tomography (CT) or CMR is essential77,78. The shape of the RVOT, smallest diameter of the RVOT in systole, narrowest and largest diameter of the conduit, length of the RVOT or conduit, distance to PA bifurcation, anatomy and proximity of the CA, distance from the landing zone to the sternum, and calcification of the conduit or RVOT tissue should be assessed77,79.

Based on this information, a consensus from a multidisciplinary team, including interventionists, surgeons, ACHD specialists, and anesthetists, should be obtained before the procedure.

Preparation: RVOT rehabilitation, Valve in Valve, and native RVOT

Prior to the valve deployment, a full hemodynamic study, angiography in the RVOT, dilation of a compliant sizing balloon in the valve landing zone and simultaneous coronary angiography or aortography to rule out CA and aortic compression should be carried out64,65,80–83. Three-dimensional rotational angiography (3DRA) is useful for better visualization of RVOT morphology, CA or aortic compression during the balloon dilation test. Despite concerns about increased radiation exposure with 3DRA, it has been reported that the total radiation dose is no different to conventional bi-plane angiography84,85.

If RVOT obstruction exists prior to the valve implantation, RVOT rehabilitation and elimination of the obstructive lesion with a stent are essential. Elevated RV pressure has been reported to be an independent risk factor for death86,87 and ventricular tachyarrhythmia88. Furthermore, recent long-term follow-up studies of TPVR illustrate that the pressure gradient across the RVOT is a key risk factor for the development of valvular dysfunction or infective endocarditis (IE)89,90. Pre-stenting of the conduit prior to valve deployment is a common procedure for the Melody valve, as it helps provide a suitable landing zone and can reinforce radial strength to prevent frame fracture of the Melody valve91–95. To reduce prestent-related technical difficulties or risk of stent embolisation or displacement96, simultaneous stenting techniques, crimping of one or more bare metal stents over the valve, and deployment of the whole system in one step have been introduced97–99. Conduit injury is reported in approximately 5% of cases and can occur during RVOT rehabilitation, such as pre-stenting or TPVR, necessitating covered stent implantation100–102.

Valve-in-valve replacement in a surgically implanted bioprosthetic pulmonary valve (BPV) is effective and generally reported with less CA compression or frame fracture103–105, because the rigid frame of the BPV protects the new valve inside92,106. It is imperative to understand the accurate type and size of the previously implanted BPV, as well as the nominal diameter, which typically represents the outer diameter. The inner diameter varies depending on the BPV type107. If the inner diameter is insufficient to the estimated adequate pulmonary valvular size108, intentional frame fracture of the BPV can be performed with a high-pressure balloon up to the outer diameter or 1-2 mm larger than the BPV106. In such situations, consideration of CA compression is crucial.

Approximately three-quarters of patients with repaired CHD and RVOT abnormalities have a “native” outflow, which includes a patched RVOT anatomy. However, because the native RVOT usually has a dilated, complex, and heterogeneous morphology78,109 caused by the surgical scar and subsequent dyskinesis or aneurysmal changes in the RVOT110, it is not always straightforward to perform TPVR for native RVOT. There are five main types of RVOT anatomy: pyramidal, straight, funnel, convex, and hourglass shape78. The most common pyramidal shape has a dilated inflow, which is also the most difficult anatomy because of the lack of an adequate landing zone for the valve79. Technical attempts have been made, such as pre-stenting or surgical plication of the RVOT prior to the Melody valve and SAPIEN THV deployment110,111. To meet this demand, self-expanding alternatives such as the Venus P valve, Harmony valve, and Alterra Prestent have been introduced in the past decade.

Valve deployment

The valve system is delivered to the target area over a stiff guidewire in the distal PA, usually left PA is preferable with better trajectory and tracking (Figure 4). With stiffer delivery systems or dilated and complex anatomy of the RVOT and PA bifurcation, delivery of the valve can be challenging112. In such situations, one may consider the anchor balloon and buddy wire technique113 or utilize a kink-resistant, flexible long sheath such as GORE Dryseal (W. L. Gore & Associates, Inc, Delaware, USA) sheath112,114,115. The device preparation and deployment varies depending on the device (next section), and post-implant hemodynamic data and pulmonary angiograms are obtained.

Figure 4. Transcatheter pulmonary valve implantation.

Figure 4.

Upper row: for SAPIEN S3 valve, Lower row: for Venus P valve. From left to right: Valve deployment, pulmonary angiography post deployment, aortic angiography post deployment. Abbreviations: CAU, caudal angulation; LAO, left anterior oblique; RAO, right anterior oblique.

Balloon-expandable valve

The Melody valve (Medtronic Inc) is the first transcatheter implantable valve which consists of a trileaflet bovine jugular venous valve sutured inside a bare-metal Cheatham-platinum (CP) stent (NuMED, Inc., Hopkinton, New York) with gold-blazed welds. Currently, two valve sizes, 16 and 18 mm in diameter, are available, and the valve is delivered via the 22 Fr Medtronic Ensemble delivery system, which utilizes a balloon-in-balloon (NuMED, Inc., Hopkinton, New York) with 18, 20, and 22 mm outer balloons. Each valve size can be expanded and deployed up to 20 mm and 22 mm respectively63. The procedural success rate is 90–98% with excellent acute or mid-term results, achieving reduction of RVOT obstruction and excellent relief of PR89. Frame fracture, which leads to valve dysfunction or recurrent RVOT obstruction, is one of the biggest concerns; however, with the establishment of pre-stenting techniques91–94, the incidence of stent fracture has significantly decreased93.

SAPIEN XT and SAPIEN 3 (Edwards Lifesciences LLC, Irvine, California, USA) transcatheter heart valves (THVs) are other balloon-expandable valves which received CE marks in 2016 and 2021, respectively, and approval of the US FDA in 2016 and 2020, respectively. These THVs comprise integrated, unidirectional, trileaflet bovine pericardial tissue valves mounted within cobalt chromium stents and polyethylene terephthalate (PET) fabric cuffs.

The SAPIEN XT THV has three sizes of 23, 26, and 29 mm, and the SAPIEN 3 THV has an additional size of 20 mm. This wide range of sizes allows for implantation of SAPIEN THVs for various landing zone diameters, including native or patched RVOT’s116,117. The SAPIEN 3 THV also has a specific refined design of externally sealed PET skirts which may diminish paravalvular leaks and longer valve length for better stabilization118. The valve is crimped with an Edwards crimper and delivered via the Edwards Novaflex delivery system for SAPIEN XT or Edwards Commander Delivery system for SAPIEN 3. The characteristics of the three expandable valves are summarized in Table 3.

Table 3. Currently available balloon-expandable valves.

Melody SAPIEN XT SAPIEN 3
Appearance graphic file with name gcsp-2024-1-e202407-i001.jpg graphic file with name gcsp-2024-1-e202407-i002.jpg graphic file with name gcsp-2024-1-e202407-i003.jpg
Valve size, mm (balloon size*) 16 (18-20) 18 (18-22) 23 26 29 20 23 26 29
Valve height: unexpanded, mm 28 30 14.3 17.3 19.1 15.5 18 20 22.5
Appropriate RVOT diameter, mm ≤20 ≤22 18-22 21-25 24-27 16-19 18-22 21-25 24-28
Delivery sheath size, Fr 22 16, 18, 20 14, 16 (for 29 mm S3)
Suitable landing zone anatomy conduit (with prestent), bioprosthetic valve conduit, native RVOT, bioprosthetic valve for stenotic lesion longer than the valve, place prestent

Notes.

Abbreviations
RVOT
right ventricular outflow tract
*

Only different for the Melody valves.

The COMPASSION and COMPASSION S3 trials were prospective multicenter studies that demonstrated the safety and efficacy of SAPIEN XT THV and S3 THV prior to FDA approval. Implantation data from 69 patients and 3-year follow-up data from 57 patients were evaluated in the COMPASSION study. At 3 years, freedom from all-cause mortality, reintervention, and endocarditis were 98.4%, 93.7%, and 97.1%, respectively. The mean peak conduit gradient decreased from 37.5 mmHg to 17.8 mmHg and PR was mild or less in 91.1% of patients. Functional improvement in New York Heart Association (NYHA) functional class was observed in 93.5% of patients100.

In the COMPASSION S3 trial, 56 patients underwent implantation and 51 patients were followed up by one-year. Valve implantation was successful in 98.2% of the cases on the first attempt, and only one case required a second S3 THV. There was no immediate residual obstruction, defined as a peak gradient across the RVOT of ≥35 mmHg or ≥ moderate PR. One year follow-up results were excellent, with no mortality, IE, valve explant or reintervention, thrombosis, or frame fracture. The THV function on echo were favorable at one-year, mean RVOT gradient decreased from 28.0 mmHg to 14.2 mmHg and only one patient (2.1%) had a mean gradient ≥ 40 mmHg. 97.9% of patients had mild or less PR. Paravalvular leak was noted in 6.4%. Functional improvement in NYHA class was also reported and all patients remained less than class II (90.2% with class I and 9.8% with class II) at one year follow up119.

A multicenter retrospective study of 774 patients, which included 397 (51%) of patients with a native RVOT, demonstrated successful implantation in 97.4% patients and serious adverse events, including death in 2 cases, stent embolization or malposition, coronary or arterial compression, or tricuspid valve injury, in 67 (10%) patients, regardless of RVOT morphology116. The low profile of the Edwards Expandable Sheath (eSheath) set a minimum 14 Fr requirement, however it necessitates a slightly complicated maneuver of in vivo balloon loading to the eSheath and valve delivery without coverage119, which consequently may cause tricuspid valve injury. The prevalence of this has been reported as high as 6−7.5% during the valve passage118,120,121. Use of the Dryseal long sheath can decrease this complication115,122, and has become a standard technique in most centers. No frame fractures have been reported in the pulmonic position; hence, pre-stenting is not always necessary for SAPIEN THVs119,121. Subclinical leaflet thrombosis on CT is associated with increased gradients and adverse clinical events following transcatheter aortic valve replacement. This hypoattenuating leaflet thickening has been reported in post-TPVR patients with early valve failure123,124.

Self-expanding valves and a prestent system

The most important advantage of a self-expanding valve is the availability of large sizes to fit the dilated native RVOT. However, in the presence of concomitant RVOT stenosis, the use of a self-expanding valve may not provide a sufficient radial force. To ensure valve stability and reduce paravalvular leak, the deployed valve size is usually slightly larger in diameter than the narrowest diameter of the RVOT landing zone. The Harmony (Medtronic Inc.) transcatheter pulmonary valve (TPV) is a self-expanding valve for larger native RVOT’s, approved by the US FDA in 2021 after clinical trials under the FDA Early Feasibility Study (EFS) pathway and two subsequent studies, the Harmony pivotal study and Continued Access Study (CAS)125,126.

The valve has a 22 (TPV 22) or 25 (TPV 25) mm porcine pericardium valve with anti-mineralization treatment to minimize calcification which is sewn into a nickel-titanium (nitinol) frame covered with a polyethylene cloth. Both TPVs have asymmetrical hourglass shapes with larger proximal flaring to hold the enlarged RVOT. TPV 25 had a slightly shorter length of 51 mm, in comparison to 55 mm of TPV 22 and a wider base skirt of 54 mm (Table 4). Both TPVs are delivered via a 25 Fr coil loading delivery system which has good trackability. When delivered to the target area of the RVOT and unsheathed, the valve expands from its distal end. The valve can be released by rotating the handle of the delivery system. Fit analysis to evaluate the adequacy of Harmony valve placement is mandatory and is performed by utilizing ECG-gated CT angiography images with reconstruction in both systole and diastole. Early term results of the pivotal study and the CAS of 87 patients described favorable outcomes without procedure- or device-related death and reintervention-free survival in 98% of TPV 22 patients and in 91% of TPV 25 patients within one year127. The longevity of the valve is unclear; however, a 5-year follow-up data of 20 patients in EFS reported reintervention-free rates of 75% with two valves in valve TPVR’s, two surgical valve replacements, and one mid-term death128. Ventricular tachycardia (VT) was reported in 16% to 40% of patients with non-sustained VT and in 3% of patients with sustained VT, which may be related to device extension below the annular valve position127,129.

Table 4. Currently available (approved or investigational) self-expanding valves and the self-expanding prestent for native right ventricular outflow tract.

Harmony TPV Venus P-valve Pulsta valve PT valve Alterra stent
Appearance (outlet flair = upper in the right images) graphic file with name gcsp-2024-1-e202407-i004.jpg graphic file with name gcsp-2024-1-e202407-i005.jpg graphic file with name gcsp-2024-1-e202407-i006.jpg graphic file with name gcsp-2024-1-e202407-i007.jpg graphic file with name gcsp-2024-1-e202407-i008.jpg graphic file with name gcsp-2024-1-e202407-i009.jpg
Valve size, mm 22 25 16-36 (each 2) 18-28 (each 2) 20, 23, 26 SAPIEN S3 29
Middle waist , mm = valve size = valve size 28-46 (each 2) 22-32 (each 2) 38-54 27
Inlet flair, mm 41 54 +10 mm of waist +4 mm of waist 28-44 (each 4) 40 mm
Outlet flair, mm 32 43 same as inlet same as inlet same as inlet same as inlet
Total length mm 55 51 25, 30* 28, 31, 33, 38 38-54 49a
Coverage full covered full covered proximal only waist only full covered proximal only
Delivery system 25 Fr 25 Fr 22, 24 Fr 18, 20 Fr 21 Fr 16 Fr
Recapture 26 Fr DS 26 Fr DS unknown own systemb unknown own systemc

Notes.

Abbreviations
TPV
transcatheter pulmonary valve
DS
Gore® Dry seal sheath
*

Proximal covered length.

a

Proximal covered length = 30 mm.

b

Up to 30% of deployment.

c

Up to 50% of deployment.

The Venus P-valve (Venus MedTech, Hangzhou, China) is a self-expanding valve which received European CE mark in 2022. The valve is comprised of a porcine pericardium valve with proximal and distal flare-shaped nitinol self-expanding covered stents to fit the native RVOT. Only the distal flare portion is uncovered to facilitate blood flow into the branch pulmonary arteries. Currently available sizes range from 28 to 36 mm diameters in two mm increments, sized 2 to four mm larger than the balloon dilated RVOT diameter. Each valve diameter size has two available stent lengths of 25 and 30 mm. The delivery system is 22 or 24 Fr.

Previous reports have demonstrated challenges with delivery with previously placed left PA stents or sharply angled RVOT’s130131; however, utilizing the Dryseal sheath might provide more predictable delivery of the valve in such situations112. The early- and mid-term results of the Venus P-valve are excellent, with a 91–98% success rate with high freedom from valve dysfunction between 2–5 postprocedural years despite a 15–26% incidence of stent frame fractures130–132.

The Alterra Adaptive Prestent (Edwards Lifesciences) was designed for use as a docking adaptor for the 29 mm SAPIEN 3 THV within the RVOT. It comprises a self-expanding nitinol frame of symmetrically shaped inflow and outflow flaring with a PET fabric covering. Similar to the Venus-P valve, the distal outflow stent is uncovered to avoid peripheral PA obstruction. The central section is 27 mm in diameter, and both flared portions are 45 mm in diameter to provide firm holding to the RVOT. Although the total unconstrained device length is 48 mm, the completely covered length is only 30 mm which can be applied to shorter native RVOT’s. The optimal landing zone size is between 27 mm and 38 mm in diameter and ≥ 35 mm in length. The delivery system is designed to provide slow, controlled deployment and recapture within two attempts by utilizing a one-handed rotational knob133. An early feasibility study of this system demonstrated 100% successful deployment in 15 patients, with bleeding in four patients and one event of atrial fibrillation134.

Upcoming valves

Although currently not approved in either the U.S. or Europe, various novel valves are being evaluated in other countries. The Pulsta valve (TaeWoong Medical Co., South Korea) is a self-expanding valve system. Its unique feature is absence of the hourglass shape like other self-expanding systems. The valve comprises a porcine pericardium valve sutured on a knitted double-stranded Nitinol wire frame. Both the proximal and distal ends are uncovered and mildly flared four mm wider than the body of the valve, which is designed to avoid peripheral PA obstruction and to provide stable positioning. The valve is available in sizes of 18–28 mm with two mm increments. The length of the stent is 28–38 mm. The valve is crimped onto the 12 Fr delivery cable using the Heart Valve Crimper (Model RVS, Blockwise Engineering LLC). During mounting, the proximal frame is hooked to a “hook block” at the proximal part of the valve-loading area to provide controlled release and prevent abrupt valve jumping. The outer size of delivery sheath is 18-20 Fr135. An early- and mid-term report of Pulsta valve in 25 patients with significant PR described 100% procedural success and no mid-term valve stenosis or significant PR. The mean RVEDV index decreased from a mean of 169.7 to 126.9 ml/m2 after 6 months136.

The Med-Zenith PT valve (Beijing Med-Zenith, China) is also a self-expanding valve with a porcine pericardial valve and a symmetrically shaped nitinol wire frame, and the valve is entirely covered with porcine pericardium. This is made for native RVOT and the clinical trial is still ongoing in China. Similar to other self-expanding systems, a preprocedural fit analysis is required. Twenty-two of the 37 patients who met the fit analysis criteria underwent implantation with the PT valve in a Chinese clinical trial. Valve implantation was successful in all patients without acute device embolization or malposition, coronary compression, ventricular arrhythmias, or tricuspid valve injury. RVEDV index decreased from mean of 181.6 to 123.4 ml/m2 after 1 year137.

Complication

Coronary artery compression, which usually involves the left main/left anterior descending branches, can be catastrophic and occurs in approximately 5% of patients during the balloon dilatation test138,139. Abnormal CA anatomy, altered CA position due to previous CA reimplantation or displacement of the aorta, or close distance between the CA and the landing zone of the valve are known substantial risks for this complication81,140. Therefore, it is important to delineate the origins and proximal courses of the coronary arteries before the procedure64; however, expected CA compressions cannot be reliably simulated, and balloon assessment is still essential. As mentioned previously, 3DRA is also helpful84. Aortic root compression is another important concern, and this complication has been documented83. This complication is more frequent with TPVR for native RVOT’s than for conduits, and significant deformation of the aortic root with aortic regurgitation is a contraindication for TPVR110,118.

Infective endocarditis is a widely recognized late complication that affects valve function and longevity. The annualized incidence of post-TPVR infective endocarditis with the Melody valve has been reported to be approximately 1.6–4%141–144 and the largest retrospective study to date included 2,476 patients and showed an IE incidence of 16.9% after TPVR after a median follow-up of 8 years145. However, there were no significant differences in the estimated survival free of endocarditis or survival free of valve replacement due to IE between the Melody TPVR group ( n = 241) and SPVR group ( n = 211) in a comparative study141. It is important to realize that IE after TPVR may be treatable with IV antimicrobial therapy, and explants of the valve may only be required if there is significant valve dysfunction146. Various risk factors of IE have been advocated, with another multicenter retrospective study of 845 post Melody TPVR patients revealing invasive RV-PA pressure gradients (per five mmHg) as an independent risk factor of IE (adjusted hazard ratio HR: 1.19, 95% CI 1.07–1.32; p =0.002)89. Despite the difficulty of direct comparison because of the different backgrounds of patients or institutions, several reports have suggested a low incidence of IE with SAPIEN THVs when compared with the Melody valve100,118,119,122, although this may be related to different subtypes of RVOT dysfunction.

Long term outcome

Currently, long-term outcomes are mostly available for the Melody valve and the SAPIEN valve. Jones et al. 90 reported 10-years follow up experience of 150 patients with the Melody valve, with a freedom from mortality rate of 90%, and freedom from any reintervention of 60%. They described age ≤21 years at time of implant (HR: 2.2, 95% CI 1.1–4.3; p =0.03), a primary indication of stenosis (HR: 2.7, 95% CI 1.3–5.4; p =0.006), the number of prior open-heart surgeries (HR: 1.3, 95% CI 1.0–1.8; p =0.06), higher post-implant RV-pulmonary artery peak-to-peak pressure gradient (HR: 1.1, 95% CI 1.0–1.1; p =0.006) as independent risk factors of reintervention. McElhinney et al. 147 also illustrated 8 years results looking at reintervention and survival after TPVR with both the Melody valve and the SAPIEN valve. There were 95/2476 (3.8%) deaths with 25% related to heart failure, 13% IE, and 7% related to procedural complications, respectively. The risk factors for death were age at TPVR (HR: 1.04 per year, 95% CI 1.03–1.06 per year; p¡0.001), a prosthetic valve in other positions (HR: 2.1, 95% CI 1.2–3.7; p 1/4 0.014), and an existing transvenous pacemaker/implantable cardioverter-defibrillator (HR: 2.1, 95% CI 1.3–3.4; P 1/4 0.004). In their cohort 258/2476 reinterventions were performed and risk factors for reintervention were age at TPVR age (HR: 0.95 per year, 95% CI 0.93–0.97 per year; p¡0.001), prior endocarditis (HR: 2.5, 95% CI 1.4–4.3; p 1/4 0.001), TPVR into a stented bioprosthetic valve (HR: 1.7, 95% CI 1.2–2.5; P 1/4 0.007), and post-implant gradient (HR: 1.4 per 10 mmHg, 95% CI 1.2–1.7 per 10 mmHg; p < 0.001). These reports suggested that TPVR is an effective alternative to SPVR with acceptable longevity.

Summary

Following the gground-breaking changes introduced by Bonhoeffer, TPVR has become one of the most important therapies for patients with congenital RVOT abnormalities. Consequently, it is now possible to provide complete percutaneous “repair” of complex RVOT abnormalities. Future advances in valve technology, including the potential for fetal valve replacement with bioengineered valves, may overcome issues surrounding somatic growth and valve degeneration. It is conceivable that percutaneous repair of the subtypes of tetralogy of Fallot may be feasible within a generation. Meanwhile, continued collaboration with our surgical colleagues to find less invasive and effective solutions for our patients is key.

References

  • 1.Allen HD SR, Penny DJ, Cetta F, Feltes TF. Ninth edition Moss and Adams’ Heart Disease in Infants, Children, and Adolescents, Including the Fetus and Young Adult. Wolters Kluwer Health 2016 [Google Scholar]
  • 2.Liu Y, Chen S, Zuhlke L, Black GC, Choy MK, Li N, Keavney BD. Global birth prevalence of congenital heart defects 1970-2017: updated systematic review and meta-analysis of 260 studies. Int J Epidemiol. Apr 1. 2019;48(2):455–463. doi: 10.1093/ije/dyz009. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Rosenthal E, Qureshi SA, Chan KC, Martin RP, Skehan DJ, Jordan SC, Tynan M. Radiofrequency-assisted balloon dilatation in patients with pulmonary valve atresia and an intact ventricular septum. British Heart Journal. 1993;69(4):347. doi: 10.1136/hrt.69.4.347. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Justo RN, Nykanen DG, Williams WG, Freedom RM, Benson LN. Transcatheter perforation of the right ventricular outflow tract as initial therapy for pulmonary valve atresia and intact ventricular septum in the newborn. Catheterization and Cardiovascular Diagnosis. 1997;40(4):408–413. doi: 10.1002/(sici)1097-0304(199704)40:4<408::aid-ccd21>3.0.co;2-h. doi: 10.1002/(sici)1097-0304(199704)40:4<408::aid-ccd21>3.0.co;2-h. [DOI] [PubMed] [Google Scholar]
  • 5.Humpl T, Soderberg B, McCrindle BW, Nykanen DG, Freedom RM, Williams WG, Benson LN. Percutaneous balloon valvotomy in pulmonary atresia with intact ventricular septum: impact on patient care. Circulation. Aug 19. 2003;108(7):826–832. doi: 10.1161/01.CIR.0000084548.44131.D1. doi: [DOI] [PubMed] [Google Scholar]
  • 6.Hasan BS, Bautista-Hernandez V, McElhinney DB, Salvin J, Laussen PC, Prakash A, Geggel RL, Pigula FA. Outcomes of transcatheter approach for initial treatment of pulmonary atresia with intact ventricular septum. Catheter Cardiovasc Interv. Jan 1. 2013;81(1):111–118. doi: 10.1002/ccd.24288. doi: [DOI] [PubMed] [Google Scholar]
  • 7.Ashburn DA, Blackstone EH, Wells WJ, Jonas RA, Pigula FA, Manning PB, Lofland GK, Williams WG, McCrindle BW, Study CHS, members xx. Determinants of mortality and type of repair in neonates with pulmonary atresia and intact ventricular septum. Journal of Thoracic and Cardiovascular Surgery. 2004;127(4):1000–1007. doi: 10.1016/j.jtcvs.2003.11.057. doi: [DOI] [PubMed] [Google Scholar]
  • 8.Wright LK, Knight JH, Thomas AS, Oster ME, Louis JDSt, Kochilas LK. Long-term outcomes after intervention for pulmonary atresia with intact ventricular septum. Heart. 2019;105(13):1007–1013. doi: 10.1136/heartjnl-2018-314124. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Cheung EW, Mastropietro CW, Flores S, Amula V, Radman M, Kwiatkowski D, Puente BN, Buckley JR, Allen K, Loomba R, Karki K, Chiwane S, Cashen K, Piggott K, Kapileshwarkar Y, Gowda KMN, Badheka A, Raman R, Costello JM, Zang H, Iliopoulos I. Collaborative Research from the Pediatric Cardiac Intensive Care Society I. Procedural Outcomes of Pulmonary Atresia With Intact Ventricular Septum in Neonates: A Multicenter Study. Annals of Thoracic Surgery. 2023;115(6):1470–1477. doi: 10.1016/j.athoracsur.2022.07.055. doi: [DOI] [PubMed] [Google Scholar]
  • 10.Giglia TM, Mandell VS, Connor AR, Mayer JE, Lock JE. Diagnosis and management of right ventricle-dependent coronary circulation in pulmonary atresia with intact ventricular septum. Circulation. 1992;86(5):1516–1528. doi: 10.1161/01.CIR.86.5.1516. doi: [DOI] [PubMed] [Google Scholar]
  • 11.Louise Calder A, Peebles CR, Occleshaw CJ. The prevalence of coronary arterial abnormalities in pulmonary atresia with intact ventricular septum and their influence on surgical results. Cardiology in the Young. 2007;17(4):387–396. doi: 10.1017/S1047951107000893. doi: [DOI] [PubMed] [Google Scholar]
  • 12.Cleuziou J, Schreiber C, Eicken A, Hörer J, Busch R, Holper K, Lange R. Predictors for biventricular repair in pulmonary atresia with intact ventricular septum. Thorac Cardiovasc Surg. 2010;58 (6):339–344. doi: 10.1055/s-0030-1250101. doi: [DOI] [PubMed] [Google Scholar]
  • 13.Spigel ZA, Qureshi AM, Adachi I. Stratification of the right ventricle-dependent coronary circulation: relevance to treatment strategy. Journal of Thoracic and Cardiovascular Surgery. 2022;164(3):765–769. doi: 10.1016/j.jtcvs.2022.01.027. doi: [DOI] [PubMed] [Google Scholar]
  • 14.Daubeney PE, Blackstone EH, Weintraub RG, Slavik Z, Scanlon J, Webber SA. Relationship of the dimension of cardiac structures to body size: an echocardiographic study in normal infants and children. Cardiol Young. 1999;9(4):402–410. doi: 10.1017/s1047951100005217. doi: [DOI] [PubMed] [Google Scholar]
  • 15.Alwi M. Management algorithm in pulmonary atresia with intact ventricular septum. Catheter Cardiovasc Interv. 2006;67(5):679–686. doi: 10.1002/ccd.20672. doi: [DOI] [PubMed] [Google Scholar]
  • 16.Haddad RN, Hanna N, Charbel R, Daou L, Chehab G, Saliba Z. Ductal stenting to improve pulmonary blood flow in pulmonary atresia with intact ventricular septum and critical pulmonary stenosis after balloon valvuloplasty. Cardiology in the Young. 2019;29(4):492–498. doi: 10.1017/S1047951119000118. doi: [DOI] [PubMed] [Google Scholar]
  • 17.Petit CJ, Glatz AC, Qureshi AM, Sachdeva R, Maskatia SA, Justino H, Goldberg DJ, Mozumdar N, Whiteside W, Rogers LS, Nicholson GT, McCracken C, Kelleman M, Goldstein BH. Outcomes after decompression of the right ventricle in infants with pulmonary atresia with intact ventricular septum are associated with degree of tricuspid regurgitation: results from the congenital catheterization research collaborative. Circ Cardiovasc Interv. 2017;10(5) doi: 10.1161/CIRCINTERVENTIONS.116.004428. doi: [DOI] [PubMed] [Google Scholar]
  • 18.Maskatia SA, Petit CJ, Travers CD, Goldberg DJ, Rogers LS, Glatz AC, Qureshi AM, Goldstein BH, Ao J, Sachdeva R. Echocardiographic parameters associated with biventricular circulation and right ventricular growth following right ventricular decompression in patients with pulmonary atresia and intact ventricular septum: results from a multicenter study. Congenit Heart Dis. 2018;13(6):892–902. doi: 10.1111/chd.12671. doi: [DOI] [PubMed] [Google Scholar]
  • 19.Kotani Y, Kasahara S, Fujii Y, Eitoku T, Baba K, Otsuki S, Kuroko Y, Arai S, Sano S. A staged decompression of right ventricle allows growth of right ventricle and subsequent biventricular repair in patients with pulmonary atresia and intact ventricular septum. European Journal of Cardio-Thoracic Surgery. 2016;50(2):298–303. doi: 10.1093/ejcts/ezw124. doi: [DOI] [PubMed] [Google Scholar]
  • 20.Liang XC, Lam WW, Cheung EW, Wu AK, Wong SJ, Cheung YF. Restrictive right ventricular physiology and right ventricular fibrosis as assessed by cardiac magnetic resonance and exercise capacity after biventricular repair of pulmonary atresia and intact ventricular septum. Clinical Cardiology. 2010;33(2):104–110. doi: 10.1002/clc.20711. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Alwi M, Geetha K, Bilkis AA, Lim MK, Hasri S, Haifa AL, Sallehudin A, Zambahari R. Pulmonary atresia with intact ventricular septum percutaneous radiofrequency-assisted valvotomy and balloon dilation versus surgical valvotomy and Blalock Taussig shunt. Journal of the American College of Cardiology. 2000;35(2):468–476. doi: 10.1016/s0735-1097(99)00549-5. doi: [DOI] [PubMed] [Google Scholar]
  • 22.Rathgeber S, Auld B, Duncombe S, Hosking MC, Harris KC. Outcomes of Radiofrequency Perforation for Pulmonary Atresia and Intact Ventricular Septum: A Single-Centre Experience. Pediatric Cardiology. 2017;38(1):170–175. doi: 10.1007/s00246-016-1498-0. doi: [DOI] [PubMed] [Google Scholar]
  • 23.Benson LN, Nykanen D, Collison A. Radiofrequency perforation in the treatment of congenital heart disease. Catheter Cardiovasc Interv. 2002;56(1):72–82. doi: 10.1002/ccd.10213. doi: [DOI] [PubMed] [Google Scholar]
  • 24.Petit CJ, Qureshi AM, Glatz AC, Kelleman MS, McCracken CE, Ligon RA, Mozumdar N, Whiteside W, Khan A, Goldstein BH. Technical factors are associated with complications and repeat intervention in neonates undergoing transcatheter right ventricular decompression for pulmonary atresia and intact ventricular septum: results from the congenital catheterisation research collaborative. Cardiology in the Young. 2018;28(8):1042–1049. doi: 10.1017/S1047951118000756. doi: [DOI] [PubMed] [Google Scholar]
  • 25.Patil NC, Saxena A, Gupta SK, Juneja R, Mishra S, Ramakrishnan S, Kothari SS. Perforating the atretic pulmonary valve with CTO hardware: technical aspects. Catheter Cardiovasc Interv. 2016;88(5):E145–E150. doi: 10.1002/ccd.25760. doi: [DOI] [PubMed] [Google Scholar]
  • 26.Alwi M, Budi RR, Mood MC, Leong MC, Samion H. Pulmonary atresia with intact septum: the use of Conquest Pro coronary guidewire for perforation of atretic valve and subsequent interventions. Cardiology in the Young. 2013;23(2):197–202. doi: 10.1017/S1047951112000595. doi: [DOI] [PubMed] [Google Scholar]
  • 27.Bakhru S, Marathe S, Saxena M, Verma S, Saileela R, Dash TK, Koneti NR. Transcatheter pulmonary valve perforation using chronic total occlusion wire in pulmonary atresia with intact ventricular septum. Ann Pediatr Cardiol. 2017;10(1):5–10. doi: 10.4103/0974-2069.197065. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lefort B, Saint-Etienne C, Soule N, Ma I, Dion F, Chantepie A. Perforation of the atretic pulmonary valve using chronic total occlusion (CTO) wire and coronary microcatheter. Congenit Heart Dis. 2019;14(5):814–818. doi: 10.1111/chd.12812. doi: [DOI] [PubMed] [Google Scholar]
  • 29.Kamali H, Tanidir IC, Erdem A, Saritas T, Guzeltas A. The use of Chronic Total Occlusion (CTO) wires for perforation of atretic pulmonary valve; two centers experience. Pediatr Cardiol. 2021;42(5):1041–1048. doi: 10.1007/s00246-021-02578-1. doi: [DOI] [PubMed] [Google Scholar]
  • 30.Lefort B, Saint-Etienne C, Soulé N, Ma I, Dion F, Chantepie A. Perforation of the atretic pulmonary valve using chronic total occlusion (CTO) wire and coronary microcatheter. Congenital Heart Disease. 2019;14(5):814–818. doi: 10.1111/chd.12812. doi: [DOI] [PubMed] [Google Scholar]
  • 31.Lamers L, Garn B, Ellsworth E, Graziano JN. Decreased incidence of right-ventricular outflow tract complications using a retrograde snare technique for radiofrequency pulmonary valve perforation. Pediatr Cardiol. 2012;33(8):1275–80. doi: 10.1007/s00246-012-0295-7. doi: [DOI] [PubMed] [Google Scholar]
  • 32.Chen RHS, Ktc A, Chow PC, Yung TC, Cheung YF, Lun KS. Achieving biventricular circulation in patients with moderate hypoplastic right ventricle in pulmonary atresia intact ventricular septum after transcatheter pulmonary valve perforation. Congenit Heart Dis. 2018;13(6):884–891. doi: 10.1111/chd.12658. doi: [DOI] [PubMed] [Google Scholar]
  • 33.Yoldas T, Orun UA, Dogan V, Ozgur S, Kutsal A, Tak S, Dilli D. Transcatheter radiofrequency pulmonary valve perforation in newborns with pulmonary atresia/intact ventricular septum: Echocardiographic predictors of biventricular circulation. Echocardiography. 2020;37(8):1258–1264. doi: 10.1111/echo.14811. doi: [DOI] [PubMed] [Google Scholar]
  • 34.Chubb H, Pesonen E, Sivasubramanian S, Tibby SM, Simpson JM, Rosenthal E, Qureshi SA. Long-term outcome following catheter valvotomy for pulmonary atresia with intact ventricular septum. J Am Coll Cardiol. 2012;59(16):1468–76. doi: 10.1016/j.jacc.2012.01.022. doi: [DOI] [PubMed] [Google Scholar]
  • 35.Sandoval JP, Chaturvedi RR, Benson L, Morgan G, Van Arsdell G, Honjo O, Caldarone C, Lee KJ. Right ventricular outflow tract stenting in tetralogy of fallot infants with risk factors for early primary repair. Circ Cardiovasc Interv. 2016;9(12) doi: 10.1161/CIRCINTERVENTIONS.116.003979. doi: [DOI] [PubMed] [Google Scholar]
  • 36.Hribernik I, Jaber O, Bentham JR. Alternative hybrid approach to promote native pulmonary artery growth in pulmonary atresia VSD MAPCAs. JACC Case Rep. 2022;4(20):1366–1369. doi: 10.1016/j.jaccas.2022.07.022. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Bentham JR, Zava NK, Harrison WJ, Shauq A, Kalantre A, Derrick G, Chen RH, Dhillon R, Taliotis D, Kang SL, Crossland D, Adesokan A, Hermuzi A, Kudumula V, Yong S, Noonan P, Hayes N, Stumper O, Thomson JDR. Duct stenting versus modified blalock-taussig shunt in neonates with duct-dependent pulmonary blood flow: associations with clinical outcomes in a multicenter national study. Circulation. 2018;137(6):581–588. doi: 10.1161/CIRCULATIONAHA.117.028972. doi: [DOI] [PubMed] [Google Scholar]
  • 38.Glatz AC, Petit CJ, Goldstein BH, Kelleman MS, McCracken CE, McDonnell A, Buckey T, Mascio CE, Shashidharan S, Ligon RA, Ao J, Whiteside W, Wallen WJ, Metcalf CM, Aggarwal V, Agrawal H, Qureshi AM. Comparison between patent ductus arteriosus stent and modified blalock-taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow: insights from the congenital catheterization research collaborative. Circulation. 2018;137(6):589–601. doi: 10.1161/CIRCULATIONAHA.117.029987. doi: [DOI] [PubMed] [Google Scholar]
  • 39.Expert Consensus P, Miller JR, Stephens EH, Goldstone AB, Glatz AC, Kane L, Van Arsdell GS, Stellin G, Barron DJ, d’Udekem Y, Benson L, Quintessenza J, Ohye RG, Talwar S, Fremes SE, Emani SM, Eghtesady P. The American Association for Thoracic Surgery (AATS) 2022 Expert Consensus Document: management of infants and neonates with tetralogy of Fallot. J Thorac Cardiovasc Surg. 2023;165(1):221–250. doi: 10.1016/j.jtcvs.2022.07.025. doi: [DOI] [PubMed] [Google Scholar]
  • 40.Caspi J, Pettitt TW, Mulder T, Stopa A. Development of the pulmonary arteries after the norwood procedure: comparison between blalock-taussig shunt and right ventricular–pulmonary artery conduit. The Annals of Thoracic Surgery. 2008;86(4):1299–1304. doi: 10.1016/j.athoracsur.2008.06.016. doi: [DOI] [PubMed] [Google Scholar]
  • 41.Pruetz JD, Badran S, Dorey F, Starnes VA, Lewis AB. Differential branch pulmonary artery growth after the Norwood procedure with right ventricle–pulmonary artery conduit versus modified Blalock–Taussig shunt in hypoplastic left heart syndrome. The Journal of Thoracic and Cardiovascular Surgery. 2009/06/01/ 2009;137(6):1342–1348. doi: 10.1016/j.jtcvs.2009.03.019. doi: [DOI] [PubMed] [Google Scholar]
  • 42.Dohlen G, Chaturvedi RR, Benson LN, Ozawa A, Van Arsdell GS, Fruitman DS, Lee KJ. Stenting of the right ventricular outflow tract in the symptomatic infant with tetralogy of Fallot. Heart. 2009;95(2):142–7. doi: 10.1136/hrt.2007.135723. doi: [DOI] [PubMed] [Google Scholar]
  • 43.Stumper O, Ramchandani B, Noonan P, Mehta C, Bhole V, Reinhardt Z, Dhillon R, Miller PA, de Giovanni JV. Stenting of the right ventricular outflow tract. Heart. 2013;99(21):1603–8. doi: 10.1136/heartjnl-2013-304155. doi: [DOI] [PubMed] [Google Scholar]
  • 44.Quandt D, Ramchandani B, Penford G, Stickley J, Bhole V, Mehta C, Jones T, Barron DJ, Stumper O. Right ventricular outflow tract stent versus BT shunt palliation in Tetralogy of Fallot. Heart. 2017;103(24):1985–1991. doi: 10.1136/heartjnl-2016-310620. doi: [DOI] [PubMed] [Google Scholar]
  • 45.Quandt D, Ramchandani B, Stickley J, Mehta C, Bhole V, Barron DJ, Stumper O. Stenting of the right ventricular outflow tract promotes better pulmonary arterial growth compared with modified blalock-taussig shunt palliation in tetralogy of fallot-type lesions. JACC Cardiovasc Interv. 2017;10(17):1774–1784. doi: 10.1016/j.jcin.2017.06.023. doi: [DOI] [PubMed] [Google Scholar]
  • 46.Luxford JC, Adams PE, Roberts PA, Mervis J. Right ventricular outflow tract stenting is a safe and effective bridge to definitive repair in symptomatic infants with tetralogy of Fallot1. Hear Lung Circulation. 2023 doi: 10.1016/j.hlc.2023.02.010. doi: 10.1016/j.hlc.2023.02.010 PMID - 36964005. [DOI] [PubMed] [Google Scholar]
  • 47.Linnane N, Nasef MA, McMahon CJ, McGuinness J, McCrossan B, Oslizlok P, Walsh KP, Kenny D. Right ventricular outflow tract stenting in symptomatic infants without the use of a long delivery sheath. Catheter Cardiovasc Interv. 2021;98(2):E275–E281. doi: 10.1002/ccd.29708. doi: [DOI] [PubMed] [Google Scholar]
  • 48.Ghaderian M, Ahmadi A, Sabri MR, Behdad S, Dehghan B, Mahdavi C, Mansourian M, Shahsanaei F. Clinical outcome of right ventricular outflow tract stenting versus blalock-taussig shunt in tetralogy of fallot: a systematic review and meta-analysis. Curr Probl Cardiol. 2021;46(3):100643. doi: 10.1016/j.cpcardiol.2020.100643. doi: [DOI] [PubMed] [Google Scholar]
  • 49.Lee J, Sivalingam S, Alwi M. Stenting of right ventricular outflow tract in Tetralogy of Fallot with subarterial ventricular septal defect: a word of caution. Ann Pediatr Cardiol. 2017;10(3):281–283. doi: 10.4103/apc.APC_168_16. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Linnane N, Nasef MA, McMahon CJ, McGuinness J, McCrossan B, Oslizlok P, Walsh KP, Kenny D. Right ventricular outflow tract stenting in symptomatic infants without the use of a long delivery sheath. Catheter Cardio Inte. 2021;98(2):E275–E281. doi: 10.1002/ccd.29708. doi: 10.1002/ccd.29708 PMID - 33851761. [DOI] [PubMed] [Google Scholar]
  • 51.Afifi A, Mehta C, Bhole V, Chaudhari M, Khan NE, Jones TJ, Stumper O. Anomalous coronary artery in Tetralogy of Fallot-Feasibility of right ventricular outflow tract stenting as initial palliation. Catheter Cardiovasc Interv. 2022;100(1):105–112. doi: 10.1002/ccd.30223. doi: [DOI] [PubMed] [Google Scholar]
  • 52.Wilder TJ, Van Arsdell GS, Benson L, Pham-Hung E, Gritti M, Page A, Caldarone CA, Hickey EJ. Young infants with severe tetralogy of Fallot: early primary surgery versus transcatheter palliation. J Thorac Cardiovasc Surg. 2017;154(5):1692–1700. doi: 10.1016/j.jtcvs.2017.05.042. e2. doi: [DOI] [PubMed] [Google Scholar]
  • 53.Barron DJ, Ramchandani B, Murala J, Stumper O, de Giovanni JV, Jones TJ, Stickley J, Brawn WJ. Surgery following primary right ventricular outflow tract stenting for Fallot’s tetralogy and variants: rehabilitation of small pulmonary arteries. Eur J Cardiothorac Surg. 2013;44(4):656–62. doi: 10.1093/ejcts/ezt188. doi: [DOI] [PubMed] [Google Scholar]
  • 54.Nollert G, Fischlein T, Bouterwek S, Bohmer C, Klinner W, Reichart B. Long-term survival in patients with repair of tetralogy of Fallot: 36-year follow-up of 490 survivors of the first year after surgical repair. J Am Coll Cardiol. 1997;30(5):1374–83. doi: 10.1016/s0735-1097(97)00318-5. doi: [DOI] [PubMed] [Google Scholar]
  • 55.Gatzoulis MA, Balaji S, Webber SA, Siu SC, Hokanson JS, Poile C, Rosenthal M, Nakazawa M, Moller JH, Gillette PC, Webb GD, Redington AN. Risk factors for arrhythmia and sudden cardiac death late after repair of tetralogy of Fallot: a multicentre study. Lancet. 2000;356(9234):975–981. doi: 10.1016/S0140-6736(00)02714-8. doi: [DOI] [PubMed] [Google Scholar]
  • 56.Bokma JP, Geva T, Sleeper LA, Babu Narayan SV, Wald R, Hickey K, Jansen K, Wassall R, Lu M, Gatzoulis MA, Mulder BJ, Valente AM. A propensity score-adjusted analysis of clinical outcomes after pulmonary valve replacement in tetralogy of Fallot. Heart. 2018;104(9):738–744. doi: 10.1136/heartjnl-2017-312048. doi: [DOI] [PubMed] [Google Scholar]
  • 57.Dore A, Glancy DL, Stone S, Menashe VD, Somerville J. Cardiac surgery for grown-up congenital heart patients survey of 307 Consecutive operations from 1991 to 1994. The American Journal of Cardiology. 1997;80(7):906–913. doi: 10.1016/s0002-9149(97)00544-4. doi: 10.1016/s0002-9149(97)00544-4 PMID - 9382007. [DOI] [PubMed] [Google Scholar]
  • 58.Somerville J. Grown-up congenital heart disease–medical demands look back, look forward 2000. Thorac Cardiovasc Surg. 2001;49(1):21–6. doi: 10.1055/s-2001-9911. doi: [DOI] [PubMed] [Google Scholar]
  • 59.Vida VL, Berggren H, Brawn WJ, Daenen W, Di Carlo D, Di Donato R, Lindberg HL, Corno AF, Fragata J, Elliott MJ, Hraska V, Kiraly L, Lacour-Gayet F, Maruszewski B, Rubay J, Sairanen H, Sarris G, Urban A, Van Doorn C, Ziemer G, Stellin G. Risk of surgery for congenital heart disease in the adult: a multicentered European study. Ann Thorac Surg. 2007;83(1):161–8. doi: 10.1016/j.athoracsur.2006.07.045. doi: [DOI] [PubMed] [Google Scholar]
  • 60.Bonhoeffer P, Boudjemline Y, Saliba Z, Merckx J, Aggoun Y, Bonnet D, Acar P, Le Bidois J, Sidi D, Kachaner J. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet. 2000;356(9239):1403–5. doi: 10.1016/S0140-6736(00)02844-0. doi: [DOI] [PubMed] [Google Scholar]
  • 61.Bonhoeffer P, Boudjemline Y, Qureshi SA, Le Bidois J, Iserin L, Acar P, Merckx J, Kachaner J, Sidi D. Percutaneous insertion of the pulmonary valve. J Am Coll Cardiol. 2002;39(10):1664–9. doi: 10.1016/s0735-1097(02)01822-3. doi: [DOI] [PubMed] [Google Scholar]
  • 62.Khambadkone S, Coats L, Taylor A, Boudjemline Y, Derrick G, Tsang V, Cooper J, Muthurangu V, Hegde SR, Razavi RS, Pellerin D, Deanfield J, Bonhoeffer P. Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation. Aug 23. 2005;112(8):1189–1197. doi: 10.1161/CIRCULATIONAHA.104.523266. doi: [DOI] [PubMed] [Google Scholar]
  • 63.McElhinney DB, Hennesen JT. The Melody(R) valve and Ensemble(R) delivery system for transcatheter pulmonary valve replacement. Annals of the New York Academy of Sciences. 2013;1291(1):77–85. doi: 10.1111/nyas.12194. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Stout KK, Daniels CJ, Aboulhosn JA, Bozkurt B, Broberg CS, Colman JM, Crumb SR, Dearani JA, Fuller S, Gurvitz M, Khairy P, Landzberg MJ, Saidi A, Valente AM, Van Hare GF. 2018 AHA/ACC Guideline for the Management of Adults With Congenital Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. Journal of the American College of Cardiology. 2019;73(12):e81–e192. doi: 10.1016/j.jacc.2018.08.1029. doi: [DOI] [PubMed] [Google Scholar]
  • 65.Baumgartner H, De Backer J, Babu-Narayan SV, Budts W, Chessa M, Diller GP, Lung B, Kluin J, Lang IM, Meijboom F, Moons P, Mulder BJM, Oechslin E, Roos-Hesselink JW, Schwerzmann M, Sondergaard L, Zeppenfeld K, Group ESCSD. 2020 ESC Guidelines for the management of adult congenital heart disease. European Heart Journal. 2021;42(6):563–645. doi: 10.1093/eurheartj/ehaa554. doi: [DOI] [PubMed] [Google Scholar]
  • 66.Marelli A, Beauchesne L, Colman J, Ducas R, Grewal J, Keir M, Khairy P, Oechslin E, Therrien J, Muhll IFVonder, Wald RM, Silversides C, Barron DJ, Benson L, Bernier PL, Horlick E, Ibrahim R, Martucci G, Nair K, Poirier NC, Ross HJ, Baumgartner H, Daniels CJ, Gurvitz M, Roos-Hesselink JW, Kovacs AH, McLeod CJ, Mulder BJ, Warnes CA, Webb GD. Canadian Cardiovascular Society 2022 Guidelines for Cardiovascular Interventions in Adults With Congenital Heart Disease. Canadian Journal of Cardiology. 2022;38(7):862–896. doi: 10.1016/j.cjca.2022.03.021. doi: [DOI] [PubMed] [Google Scholar]
  • 67.Zahn EM, Hellenbrand WE, Lock JE, McElhinney DB. Implantation of the melody transcatheter pulmonary valve in patients with a dysfunctional right ventricular outflow tract conduit early results from the u.s. Clinical trial. Journal of the American College of Cardiology. 2009;54(18):1722–1729. doi: 10.1016/j.jacc.2009.06.034. doi: [DOI] [PubMed] [Google Scholar]
  • 68.Berman DP, McElhinney DB, Vincent JA, Hellenbrand WE, Zahn EM. Feasibility and Short-Term Outcomes of Percutaneous Transcatheter Pulmonary Valve Replacement in Small (<30 kg) Children With Dysfunctional Right Ventricular Outflow Tract Conduits. Circulation: Cardiovascular Interventions. 2014;7(2):142–148. doi: 10.1161/CIRCINTERVENTIONS.113.000881. doi: [DOI] [PubMed] [Google Scholar]
  • 69.Zampi JD, Berman DP, Bocks ML, Yu S, Zahn EM, Lu JC, Shaya JA, Armstrong AK. Factors associated with the internal jugular venous approach for Melody™ Transcatheter Pulmonary Valve implantation. Cardiology in the Young. 2016;26(5):948–956. doi: 10.1017/S1047951115001663. doi: [DOI] [PubMed] [Google Scholar]
  • 70.Coserria-Sanchez F, Iglesias-Lopez A, Gil-Ortega MVilla, Moruno-Tirado A, Garcia-Angleu F, Zunzunegui-Martinez JL. Percutaneous Pulmonary Melody Valve Implantation Through the Transhepatic Route in an 8-Year-old Patient. Rev Esp Cardiol (Engl Ed) 2018;71(10):861–862. doi: 10.1016/j.rec.2017.02.045. doi: [DOI] [PubMed] [Google Scholar]
  • 71.Holoshitz N, Ilbawi MN, Amin Z. Perventricular melody valve implantation in a 12 kg child. Catheter Cardiovasc Interv. 2013;82(5):824–7. doi: 10.1002/ccd.25044. doi: [DOI] [PubMed] [Google Scholar]
  • 72.Gupta A, Kenny D, Caputo M, Amin Z. Initial experience with elective perventricular melody valve placement in small patients. Pediatr Cardiol. 2017;38(3):575–581. doi: 10.1007/s00246-016-1550-0. doi: [DOI] [PubMed] [Google Scholar]
  • 73.Breatnach CR, McGuinness J, Ng LY, Franklin O, Redmond M, Nolke L, McMahon C, Oslizlok P, Walsh K, Kenny D. Procedural technique for hybrid pulmonary valve replacement in infants and small children. Eur J Cardiothorac Surg. 2021;59(4):823–830. doi: 10.1093/ejcts/ezaa410. doi: [DOI] [PubMed] [Google Scholar]
  • 74.Cubeddu RJ, Hijazi ZM. Bailout perventricular pulmonary valve implantation following failed percutaneous attempt using the Edwards Sapien transcatheter heart valve. Catheter Cardiovasc Interv. 2011;77(2):276–80. doi: 10.1002/ccd.22653. doi: [DOI] [PubMed] [Google Scholar]
  • 75.Simpson KE, Huddleston CB, Foerster S, Nicholas R, Balzer D. Successful subxyphoid hybrid approach for placement of a Melody percutaneous pulmonary valve. Catheter Cardiovasc Interv. 2011;78(1):108–111. doi: 10.1002/ccd.22930. doi: [DOI] [PubMed] [Google Scholar]
  • 76.Berman DP, Burke R, Zahn EM. Use of a novel hybrid approach to salvage an attempted transcatheter pulmonary valve implant. Pediatric Cardiology. 2012;33(5):839–42. doi: 10.1007/s00246-012-0224-9. doi: [DOI] [PubMed] [Google Scholar]
  • 77.Chung R, Taylor AM. Imaging for preintervention planning: transcatheter pulmonary valve therapy. Circ Cardiovasc Imaging. 2014;7(1):182–9. doi: 10.1161/CIRCIMAGING.113.000826. doi: [DOI] [PubMed] [Google Scholar]
  • 78.Schievano S, Coats L, Migliavacca F, Norman W, Frigiola A, Deanfield J, Bonhoeffer P, Taylor AM. Variations in right ventricular outflow tract morphology following repair of congenital heart disease: implications for percutaneous pulmonary valve implantation. J Cardiovasc Magn Reson. 2007;9(4):687–95. doi: 10.1080/10976640601187596. doi: [DOI] [PubMed] [Google Scholar]
  • 79.Boutsikou M, Tzifa A. Noninvasive imaging prior to percutaneous pulmonary valve implantation. Hellenic J Cardiol. 2022;67:59–65. doi: 10.1016/j.hjc.2022.06.004. doi: [DOI] [PubMed] [Google Scholar]
  • 80.Lurz P, Coats L, Khambadkone S, Nordmeyer J, Boudjemline Y, Schievano S, Muthurangu V, Lee TY, Parenzan G, Derrick G, Cullen S, Walker F, Tsang V, Deanfield J, Taylor AM, Bonhoeffer P. Percutaneous pulmonary valve implantation: impact of evolving technology and learning curve on clinical outcome. Circulation. 2008;117(15):1964–72. doi: 10.1161/CIRCULATIONAHA.107.735779. doi: [DOI] [PubMed] [Google Scholar]
  • 81.Romans RA, Lu JC, Balasubramanian S, Whiteside W, Yu S, Aldoss OT, Armstrong AK, Boe BA, Balzer DT, Christensen JT, Jones TK, Keeshan B, McLennan D, Nicholson GT, Patel N, Salavitabar A, Shahanavaz S, Sullivan PM, Turner ME, Zampi JD. Cardiac magnetic resonance to predict coronary artery compression in transcatheter pulmonary valve implantation into conduits. JACC Cardiovasc Interv. 2022;15(9):979–988. doi: 10.1016/j.jcin.2022.02.047. doi: [DOI] [PubMed] [Google Scholar]
  • 82.Sridharan S, Coats L, Khambadkone S, Taylor AM, Bonhoeffer P. Images in cardiovascular medicine. Transcatheter right ventricular outflow tract intervention: the risk to the coronary circulation. Circulation. 2006;113(25):e934–5. doi: 10.1161/CIRCULATIONAHA.105.599514. doi: [DOI] [PubMed] [Google Scholar]
  • 83.Lindsay I, Aboulhosn J, Salem M, Levi D. Aortic root compression during transcatheter pulmonary valve replacement. Catheter Cardiovasc Interv. 2016;88(5):814–821. doi: 10.1002/ccd.26547. doi: [DOI] [PubMed] [Google Scholar]
  • 84.Pockett CR, Moore JW, El-Said HG. Three dimensional rotational angiography for assessment of coronary arteries during melody valve implantation: introducing a technique that may improve outcomes. Neth Heart J. 2017;25(2):82–90. doi: 10.1007/s12471-016-0931-6. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Avesani M, Kang SL, Jalal Z, Thambo JB, Iriart X. Renaissance of cardiac imaging to assist percutaneous interventions in congenital heart diseases: the role of three-dimensional echocardiography and multimodality imaging. Front Pediatr. 2022;10:894472. doi: 10.3389/fped.2022.894472. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Egbe AC, Taggart NW, Reddy YNV, Sufian M, Banala K, Vojjini R, Najam M, Osman K, Obokata M, Borlaug BA. Assessment and implications of right ventricular afterload in tetralogy of fallot. Am J Cardiol. 2019;124(11):1780–1784. doi: 10.1016/j.amjcard.2019.08.035. doi: [DOI] [PubMed] [Google Scholar]
  • 87.Latus H, Stammermann J, Voges I, Waschulzik B, Gutberlet M, Diller GP, Schranz D, Ewert P, Beerbaum P, Kuhne T, Sarikouch S. German competence network for congenital heart defects investigators. Impact of right ventricular pressure load after repair of tetralogy of fallot. J Am Heart Assoc. 2022;11(7):e022694. doi: 10.1161/JAHA.121.022694. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88.Valente AM, Gauvreau K, Assenza GE, Babu-Narayan SV, Schreier J, Gatzoulis MA, Groenink M, Inuzuka R, Kilner PJ, Koyak Z, Landzberg MJ, Mulder B, Powell AJ, Wald R, Geva T. Contemporary predictors of death and sustained ventricular tachycardia in patients with repaired tetralogy of Fallot enrolled in the INDICATOR cohort. Heart. 2014;100(3):247–53. doi: 10.1136/heartjnl-2013-304958. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 89.Nordmeyer J, Ewert P, Gewillig M, AlJufan M, Carminati M, Kretschmar O, Uebing A, Dahnert I, Rohle R, Schneider H, Witsenburg M, Benson L, Gitter R, Bokenkamp R, Mahadevan V, Berger F. Acute and mid-term outcomes of the post-approval MELODY Registry: a multicentre registry of transcatheter pulmonary valve implantation. Eur Heart J. 2019;40(27):2255–2264. doi: 10.1093/eurheartj/ehz201. doi: [DOI] [PubMed] [Google Scholar]
  • 90.Jones TK, McElhinney DB, Vincent JA, Hellenbrand WE, Cheatham JP, Berman DP, Zahn EM, Khan DM, Rhodes Jr JF, Weng S, Bergersen LJ. Long-term outcomes after melody transcatheter pulmonary valve replacement in the US investigational device exemption trial. Circ Cardiovasc Interv. 2022;15(1):e010852. doi: 10.1161/CIRCINTERVENTIONS.121.010852. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 91.Nordmeyer J, Khambadkone S, Coats L, Schievano S, Lurz P, Parenzan G, Taylor AM, Lock JE, Bonhoeffer P. Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation. Circulation. 2007;115(11):1392–7. doi: 10.1161/CIRCULATIONAHA.106.674259. doi: [DOI] [PubMed] [Google Scholar]
  • 92.Cabalka AK, Hellenbrand WE, Eicken A, Kreutzer J, Gray RG, Bergersen L, Berger F, Armstrong AK, Cheatham JP, Zahn EM, McElhinney DB. Relationships among conduit type, pre-stenting, and outcomes in patients undergoing transcatheter pulmonary valve replacement in the prospective north american and european melody valve trials. JACC Cardiovasc Interv. 2017;10(17):1746–1759. doi: 10.1016/j.jcin.2017.05.022. doi: [DOI] [PubMed] [Google Scholar]
  • 93.Cardoso R, Ansari M, Garcia D, Sandhu S, Brinster D, Piazza N. Pre-stenting for prevention of melody valve stent fractures: a systematic review and meta-analysis. Catheter Cardiovasc Interv. 2016;87(3):534–539. doi: 10.1002/ccd.26235. doi: [DOI] [PubMed] [Google Scholar]
  • 94.McElhinney DB, Cheatham JP, Jones TK, Lock JE, Vincent JA, Zahn EM, Hellenbrand WE. Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation. Circulation: Cardiovascular Interventions. 2011;4(6):602–614. doi: 10.1161/CIRCINTERVENTIONS.111.965616. doi: [DOI] [PubMed] [Google Scholar]
  • 95.Cools B, Brown S, Wevers M, Humbeeck JV, Boshoff D, Verdonckt C, Gewillig M. Right ventricle outflow tract pre-stenting: In vitro testing of rigidity and corrosion properties. Catheter Cardiovasc Interv. 2018;91(2):285–291. doi: 10.1002/ccd.27320. doi: [DOI] [PubMed] [Google Scholar]
  • 96.Karsenty C, Malekzadeh-Milani S, Fraisse A, Gewillig M, Bonnet D, Aldebert P, Ovaert C, Bouvaist H, Kempny A, Houeijeh A, Petit J, Hascoet S. Right ventricular outflow tract pre-stenting with AndraStent XXL before percutaneous pulmonary valve implantation. Arch Cardiovasc Dis. 2020;113(2):113–120. doi: 10.1016/j.acvd.2019.12.004. doi: [DOI] [PubMed] [Google Scholar]
  • 97.Heibel J, Reeves R, Alshawabkeh L, Justino H, El-Said H. Simultaneous Stenting With Edwards SAPIEN Transcatheter Pulmonary Valve Replacement. Journal of the Society for Cardiovascular Angiography & Interventions. 2023;2(1) doi: 10.1016/j.jscai.2022.100553. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Zablah JE, Dohlen G, Ystgaard MB, Morgan GJ. Piggyback mounting for stent and valve deployment during percutaneous pulmonary valve implantation. Catheter Cardiovasc Interv. 2022;100(4):606–611. doi: 10.1002/ccd.30391. doi: [DOI] [PubMed] [Google Scholar]
  • 99.Boudjemline Y. A new one-step procedure for pulmonary valve implantation of the melody valve: simultaneous pre-stenting and valve implantation. Catheter Cardiovasc Interv. 2018;91(1):64–70. doi: 10.1002/ccd.27332. doi: [DOI] [PubMed] [Google Scholar]
  • 100.Kenny D, Rhodes JF, Fleming GA, Kar S, Zahn EM, Vincent J, Shirali GS, Gorelick J, Fogel MA, Fahey JT, Kim DW, Babaliaros VC, Armstrong AK, Hijazi ZM. 3-Year outcomes of the edwards SAPIEN Transcatheter heart valve for conduit failure in the pulmonary position from the COMPASSION multicenter clinical trial. JACC Cardiovasc Interv. 2018;11(19):1920–1929. doi: 10.1016/j.jcin.2018.06.001. doi: [DOI] [PubMed] [Google Scholar]
  • 101.Delaney JW, Goldstein BH, Bishnoi RN, Bisselou KSM, McEnaney K, Minahan M, Ringel RE, Investigators P. Covered CP stent for treatment of right ventricular conduit injury during melody transcatheter pulmonary valve replacement. Circ Cardiovasc Interv. 2018;11(10):e006598. doi: 10.1161/CIRCINTERVENTIONS.118.006598. doi: [DOI] [PubMed] [Google Scholar]
  • 102.Bishnoi RN, Jones TK, Kreutzer J, Ringel RE. NuMED Covered Cheatham-Platinum Stent for the treatment or prevention of right ventricular outflow tract conduit disruption during transcatheter pulmonary valve replacement. Catheter Cardiovasc Interv. 2015;85(3):421–7. doi: 10.1002/ccd.25682. doi: [DOI] [PubMed] [Google Scholar]
  • 103.Gillespie MJ, Rome JJ, Levi DS, Williams RJ, Rhodes JF, Cheatham JP, Hellenbrand WE, Jones TK, Vincent JA, Zahn EM, McElhinney DB. Melody valve implant within failed bioprosthetic valves in the pulmonary position: a multicenter experience. Circ Cardiovasc Interv. 2012;5(6):862–70. doi: 10.1161/CIRCINTERVENTIONS.112.972216. doi: [DOI] [PubMed] [Google Scholar]
  • 104.Cabalka AK, Asnes JD, Balzer DT, Cheatham JP, Gillespie MJ, Jones TK, Justino H, Kim DW, Lung TH, Turner DR, McElhinney DB. Transcatheter pulmonary valve replacement using the melody valve for treatment of dysfunctional surgical bioprostheses: A multicenter study. J Thorac Cardiovasc Surg. 2018;155(4):1712–1724. doi: 10.1016/j.jtcvs.2017.10.143. doi: [DOI] [PubMed] [Google Scholar]
  • 105.Finch W, Levi DS, Salem M, Hageman A, Aboulhosn J. Transcatheter melody valve placement in large diameter bioprostheses and conduits: What is the optimal “Landing zone”? Catheter Cardiovasc Interv. 2015;86(5):E217–23. doi: 10.1002/ccd.25922. doi: [DOI] [PubMed] [Google Scholar]
  • 106.Shahanavaz S, Asnes JD, Grohmann J, Qureshi AM, Rome JJ, Tanase D, Crystal MA, Latson LA, Morray BH, Hellenbrand W, Balzer DT, Gewillig M, Love JC, Berdjis F, Gillespie MJ, McElhinney DB. Intentional fracture of bioprosthetic valve frames in patients undergoing valve-in-valve transcatheter pulmonary valve replacement. Circ Cardiovasc Interv. 2018;11(8):e006453. doi: 10.1161/CIRCINTERVENTIONS.118.006453. doi: [DOI] [PubMed] [Google Scholar]
  • 107.Bapat V, Mydin I, Chadalavada S, Tehrani H, Attia R, Thomas M. A guide to fluoroscopic identification and design of bioprosthetic valves: a reference for valve-in-valve procedure. Catheter Cardiovasc Interv. 2013;81(5):853–861. doi: 10.1002/ccd.24419. doi: [DOI] [PubMed] [Google Scholar]
  • 108.Capps SB, Elkins RC, Fronk DM. Body surface area as a predictor of aortic and pulmonary valve diameter. J Thorac Cardiovasc Surg. 2000;119(5):975–82. doi: 10.1016/s0022-5223(00)70092-4. doi: [DOI] [PubMed] [Google Scholar]
  • 109.Lurz P, Gaudin R, Taylor AM, Bonhoeffer P. Percutaneous pulmonary valve implantation. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2009:112–117. doi: 10.1053/j.pcsu.2009.01.011. doi: [DOI] [PubMed] [Google Scholar]
  • 110.Martin MH, Meadows J, McElhinney DB, Goldstein BH, Bergersen L, Qureshi AM, Shahanavaz S, Aboulhosn J, Berman D, Peng L, Gillespie M, Armstrong A, Weng C, Minich LL, Gray RG. Safety and feasibility of melody transcatheter pulmonary valve replacement in the native right ventricular outflow tract: a multicenter pediatric heart network scholar study. JACC Cardiovasc Interv. 2018;11(16):1642–1650. doi: 10.1016/j.jcin.2018.05.051. doi: [DOI] [PubMed] [Google Scholar]
  • 111.Carr K, Nijres BM, Windsor JJ, Nakamura Y, Karimi M, Ricci M, Aldoss O. Single-center experience of hybrid pulmonary valve replacement using left anterior thoracotomy with pulmonary artery plication in patients with large right ventricular outflow tract. J Am Heart Assoc. 2022;11(14):e026517. doi: 10.1161/JAHA.122.026517. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 112.I. Jones M, Murphy M, Rosenthal E, P. Walsh K, Kenny D, A. Qureshi S, Butera G. Use of the GORE® DrySeal Flex Introducer Sheath to Facilitate Implantation of the Transcatheter Venus P-valve. Congenital Heart Disease. 2021;16(3):197–203. doi: 10.32604/chd.2021.015222. doi: [DOI] [Google Scholar]
  • 113.Shah RR, Poommipanit P, Law MA, Amin Z. Anchor balloon, buddy wire, and wire and sheath techniques to deploy percutaneous pulmonary valves in tetralogy of fallot patients. Catheter Cardiovasc Interv. 2018;92(5):915–920. doi: 10.1002/ccd.27022. doi: [DOI] [PubMed] [Google Scholar]
  • 114.Butera G, Hansen JH, Jones MI. Tricuspid regurgitation complicating SAPIEN 3 valve implantation in pulmonary position. Catheter Cardiovasc Interv. 2019;94(6):894. doi: 10.1002/ccd.28083. doi: [DOI] [PubMed] [Google Scholar]
  • 115.Kenny D, Morgan GJ, Murphy M, AlAlwi K, Giugno L, Zablah J, Carminati M, Walsh K. Use of 65 cm large calibre Dryseal sheaths to facilitate delivery of the Edwards SAPIEN valve to dysfunctional right ventricular outflow tracts. Catheter Cardiovasc Interv. 2019;94(3):409–413. doi: 10.1002/ccd.28409. doi: [DOI] [PubMed] [Google Scholar]
  • 116.Shahanavaz S, Zahn EM, Levi DS, Aboulhousn JA, Hascoet S, Qureshi AM, Porras D, Morgan GJ, Bauser Heaton H, Martin MH, Keeshan B, Asnes JD, Kenny D, Ringewald JM, Zablah JE, Ivy M, Morray BH, Torres AJ, Berman DP, Gillespie MJ, Chaszczewski K, Zampi JD, Walsh KP, Julien P, Goldstein BH, Sathanandam SK, Karsenty C, Balzer DT, McElhinney DB. Transcatheter pulmonary valve replacement with the sapien prosthesis. J Am Coll Cardiol. 2020;76(24):2847–2858. doi: 10.1016/j.jacc.2020.10.041. doi: [DOI] [PubMed] [Google Scholar]
  • 117.Sivakumar K, Chandrasekaran R, Hijazi ZM. Unique challenges posed by a dysfunctional native right ventricular outflow tract for percutaneous pulmonary valve implantation using SAPIEN-S3 valve. Ann Pediatr Cardiol. 2022;15(2):175–179. doi: 10.4103/apc.apc_86_21. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 118.Sinha S, Aboulhosn J, Asnes J, Bocks M, Zahn E, Goldstein BH, Zampi J, Hellenbrand W, Salem M, Levi D. Initial results from the off-label use of the SAPIEN S3 valve for percutaneous transcatheter pulmonary valve replacement: a multi-institutional experience. Catheter Cardiovasc Interv. 2019;93(3):455–463. doi: 10.1002/ccd.27973. doi: [DOI] [PubMed] [Google Scholar]
  • 119.Lim DS, Kim D, Aboulhosn J, Levi D, Fleming G, Hainstock M, Sommer R, Torres AJ, Zhao Y, Shirali G, Babaliaros V. Congenital pulmonic valve dysfunction treated with SAPIEN 3 Transcatheter heart valve (from the COMPASSION S3 Trial) Am J Cardiol. 2023;190:102–109. doi: 10.1016/j.amjcard.2022.12.010. doi: [DOI] [PubMed] [Google Scholar]
  • 120.Faccini A, Butera G. Tricuspid regurgitation as a complication of Edwards Sapien XT valve implantation in pulmonary position a problem to deal with. Catheter Cardiovasc Interv. 2018;91(5):927–931. doi: 10.1002/ccd.27527. doi: [DOI] [PubMed] [Google Scholar]
  • 121.Morgan GJ, Sadeghi S, Salem MM, Wilson N, Kay J, Rothman A, Galindo A, Martin MH, Gray R, Ross M, Aboulhosn JA, Levi DS. SAPIEN valve for percutaneous transcatheter pulmonary valve replacement without pre-stenting: a multi-institutional experience. Catheter Cardiovasc Interv. 2019;93(2):324–329. doi: 10.1002/ccd.27932. doi: [DOI] [PubMed] [Google Scholar]
  • 122.Houeijeh A, Batteux C, Karsenty C, Ramdane N, Lecerf F, Valdeolmillos E, Lourtet-Hascoet J, Cohen S, Belli E, Petit J, Hascoet S. Long-term outcomes of transcatheter pulmonary valve implantation with melody and SAPIEN valves. Int J Cardiol. 2023;370:156–166. doi: 10.1016/j.ijcard.2022.10.141. doi: [DOI] [PubMed] [Google Scholar]
  • 123.Riahi M, Blanke P, Webb J, Carere RG. Early leaflet thrombosis complicating transcatheter implantation of a Sapien 3 valve in a native right ventricular outflow tract. Catheter Cardiovasc Interv. 2018;92(5):925–929. doi: 10.1002/ccd.27183. doi: [DOI] [PubMed] [Google Scholar]
  • 124.Hammadah M, Han BK, de Oliveira Nunes M, Aboulhosn JA, Zahn EM, Babaliaros V, Daniels MJ, Cavalcante JL, Lesser JR, Garcia S. Hypoattenuated leaflet thickening after transcatheter pulmonary valve replacement with the SAPIEN 3 valve. JACC Cardiovasc Imaging. 2021;14(10):2047–2048. doi: 10.1016/j.jcmg.2021.04.023. doi: [DOI] [PubMed] [Google Scholar]
  • 125.Bergersen L, Benson LN, Gillespie MJ, Cheatham SL, Crean AM, Hor KN, Horlick EM, Lung TH, McHenry BT, Osten MD, Powell AJ, Cheatham JP. Harmony feasibility trial: acute and short-term outcomes with a self-expanding transcatheter pulmonary valve. JACC Cardiovasc Interv. 2017;10(17):1763–1773. doi: 10.1016/j.jcin.2017.05.034. doi: [DOI] [PubMed] [Google Scholar]
  • 126.Gillespie MJ, Benson LN, Bergersen L, Bacha EA, Cheatham SL, Crean AM, Eicken A, Ewert P, Geva T, Hellenbrand WE, Hor KN, Horlick EM, Jones TK, Mayer J, McHenry BT, Osten MD, Powell AJ, Zahn EM, Cheatham JP. Patient selection process for the harmony transcatheter pulmonary valve early feasibility study. Am J Cardiol. 2017;120(8):1387–1392. doi: 10.1016/j.amjcard.2017.07.034. doi: [DOI] [PubMed] [Google Scholar]
  • 127.Gillespie Matthew J, McElhinney Doff B, Jones Thomas K, Levi Daniel S, Asnes J, Gray Robert G, Cabalka Allison K, Fujimoto K, Qureshi Athar M, Justino H, Bergersen L, Benson Lee N, Haugan D, Boe Brian A, Cheatham John P. One-year outcomes in a pooled cohort of harmony transcatheter pulmonary valve clinical trial participants. JACC Cardiovasc Interv. 2023 doi: 10.1016/j.jcin.2023.03.002. doi: [DOI] [PubMed] [Google Scholar]
  • 128.Gillespie MJ, Bergersen L, Benson LN, Weng S, Cheatham JP. 5-Year outcomes from the harmony native outflow tract early feasibility study. JACC Cardiovasc Interv. 2021;14(7):816–817. doi: 10.1016/j.jcin.2021.01.046. doi: [DOI] [PubMed] [Google Scholar]
  • 129.Simmons MA, Elder RW, Shabanova V, Hellenbrand W, Asnes J. Ventricular arrhythmias immediately following transcatheter pulmonary valve implantation: a cause for concern? Catheter Cardiovasc Interv. 2018;91(5):920–926. doi: 10.1002/ccd.27454. doi: [DOI] [PubMed] [Google Scholar]
  • 130.Morgan G, Prachasilchai P, Promphan W, Rosenthal E, Sivakumar K, Kappanayil M, Sakidjan I, Walsh KP, Kenny D, Thomson J, Koneti NR, Awasthy N, Thanopoulos B, Roymanee S, Qureshi S. Medium-term results of percutaneous pulmonary valve implantation using the Venus P-valve: international experience. EuroIntervention. 2019;14(13):1363–1370. doi: 10.4244/EIJ-D-18-00299. doi: [DOI] [PubMed] [Google Scholar]
  • 131.Sivakumar K, Sagar P, Qureshi S, Promphan W, Sasidharan B, Awasthy N, Kappanayil M, Suresh PV, Koneti NR. Outcomes of Venus P-valve for dysfunctional right ventricular outflow tracts from Indian Venus P-valve database. Ann Pediatr Cardiol. 2021;14(3):281–292. doi: 10.4103/apc.APC_175_20. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Zhou D, Pan W, Jilaihawi H, Zhang G, Feng Y, Pan X, Liu J, Yu S, Cao Q, Ge J. A self-expanding percutaneous valve for patients with pulmonary regurgitation and an enlarged native right ventricular outflow tract: one-year results. EuroIntervention. 2019;14(13):1371–1377. doi: 10.4244/EIJ-D-18-00715. doi: [DOI] [PubMed] [Google Scholar]
  • 133.Zahn EM, Chang JC, Armer D, Garg R. First human implant of the Alterra Adaptive Prestent(TM): a new self-expanding device designed to remodel the right ventricular outflow tract. Catheter Cardiovasc Interv. 2018;91(6):1125–1129. doi: 10.1002/ccd.27581. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 134.Shahanavaz S, Balzer D, Babaliaros V, Kim D, Dimas V, Veeram Reddy SR, Leipsic J, Blanke P, Shirali G, Parthiban A, Gorelick J, Zahn EM. Alterra adaptive prestent and SAPIEN 3 THV for congenital pulmonic valve dysfunction: an early feasibility study. JACC Cardiovasc Interv. 2020;13(21):2510–2524. doi: 10.1016/j.jcin.2020.06.039. doi: [DOI] [PubMed] [Google Scholar]
  • 135.Kim GB, Song MK, Bae EJ, Park EA, Lee W, Lim HG, Kim YJ. Successful feasibility human trial of a new self-expandable percutaneous pulmonary valve (pulsta valve) implantation using knitted nitinol wire backbone and trileaflet alpha-gal-free porcine pericardial valve in the native right ventricular outflow tract. Circ Cardiovasc Interv. 2018;11(6):e006494. doi: 10.1161/CIRCINTERVENTIONS.118.006494. doi: [DOI] [PubMed] [Google Scholar]
  • 136.Lee SY, Kim GB, Kim SH, Jang SI, Choi JY, Kang IS, Kim YH. Mid-term outcomes of the Pulsta transcatheter pulmonary valve for the native right ventricular outflow tract. Catheter Cardiovasc Interv. 2021;98(5):E724–E732. doi: 10.1002/ccd.29865. doi: [DOI] [PubMed] [Google Scholar]
  • 137.Shang X, Dong N, Zhang C, Wang Y. The clinical trial outcomes of Med-Zenith PT-valve in the treatment of patients with severe pulmonary regurgitation. Front Cardiovasc Med. 2022;9:887886. doi: 10.3389/fcvm.2022.887886. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 138.Morray BH, McElhinney DB, Cheatham JP, Zahn EM, Berman DP, Sullivan PM, Lock JE, Jones TK. Risk of coronary artery compression among patients referred for transcatheter pulmonary valve implantation: a multicenter experience. Circ Cardiovasc Interv. 2013;6(5):535–42. doi: 10.1161/CIRCINTERVENTIONS.113.000202. doi: [DOI] [PubMed] [Google Scholar]
  • 139.Fraisse A, Assaidi A, Mauri L, Malekzadeh-Milani S, Thambo JB, Bonnet D, Iserin L, Mancini J, Boudjemline Y. Coronary artery compression during intention to treat right ventricle outflow with percutaneous pulmonary valve implantation: incidence, diagnosis, and outcome. Catheter Cardiovasc Interv. 2014;83(7):E260–8. doi: 10.1002/ccd.25471. doi: [DOI] [PubMed] [Google Scholar]
  • 140.Rinaldi E, Sadeghi S, Rajpal S, Boe BA, Daniels C, Cheatham J, Sinha S, Levi DS, Aboulhosn J. Utility of CT Angiography for the Prediction of Coronary Artery Compression in Patients Undergoing Transcatheter Pulmonary Valve Replacement. World J Pediatr Congenit Heart Surg. 2020;11(3):295–303. doi: 10.1177/2150135120905670. doi: [DOI] [PubMed] [Google Scholar]
  • 141.Georgiev S, Ewert P, Eicken A, Hager A, Horer J, Cleuziou J, Meierhofer C, Tanase D. Munich comparative study: prospective long-term outcome of the transcatheter melody valve versus surgical pulmonary bioprosthesis with up to 12 years of follow-up. Circ Cardiovasc Interv. 2020;13(7):e008963. doi: 10.1161/CIRCINTERVENTIONS.119.008963. doi: [DOI] [PubMed] [Google Scholar]
  • 142.Van Dijck I, Budts W, Cools B, Eyskens B, Boshoff DE, Heying R, Frerich S, Vanagt WY, Troost E, Gewillig M. Infective endocarditis of a transcatheter pulmonary valve in comparison with surgical implants. Heart. 2015;101(10):788–93. doi: 10.1136/heartjnl-2014-306761. doi: [DOI] [PubMed] [Google Scholar]
  • 143.Sadeghi S, Wadia S, Lluri G, Tarabay J, Fernando A, Salem M, Sinha S, Levi DS, Aboulhosn J. Risk factors for infective endocarditis following transcatheter pulmonary valve replacement in patients with congenital heart disease. Catheter Cardiovasc Interv. 2019;94(4):625–635. doi: 10.1002/ccd.28474. doi: [DOI] [PubMed] [Google Scholar]
  • 144.Georgiev S, Ewert P, Tanase D, Hess J, Hager A, Cleuziou J, Meierhofer C, Eicken A. A low residual pressure gradient yields excellent long-term outcome after percutaneous pulmonary valve implantation. JACC Cardiovasc Interv. 2019;12(16):1594–1603. doi: 10.1016/j.jcin.2019.03.037. doi: [DOI] [PubMed] [Google Scholar]
  • 145.McElhinney DB, Zhang Y, Aboulhosn JA, Morray BH, Biernacka EK, Qureshi AM, Torres AJ, Shahanavaz S, Goldstein BH, Cabalka AK, Bauser-Heaton H, Georgiev S, Berger F, Millan-Iturbe O, Peng LF, Armstrong AK, Levi DS, Fronczak-Jakubczyk A, Sondergaard L, Anderson JH, Schranz D, Jones TK, Cheatham JP, Schubert S, Ewert P. Multicenter study of endocarditis after transcatheter pulmonary valve replacement. J Am Coll Cardiol. 2021;78(6):575–589. doi: 10.1016/j.jacc.2021.05.044. doi: [DOI] [PubMed] [Google Scholar]
  • 146.Davtyan A, Guyon PW, El-Sabrout HR, Ponder R, Ramchandar N, Weber R, Zayed W, Ratnayaka K, Nigro JJ, Moore JW, Bauser-Heaton H, Alshawabkeh L, Reeves RR, Levi D, Aboulhosn J, Justino H, Bradley J, El-Said HG. Selective valve removal for melody valve endocarditis: practice variations in a multicenter experience. Pediatr Cardiol. 2022;43(4):894–902. doi: 10.1007/s00246-021-02801-z. doi: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 147.McElhinney DB, Zhang Y, Levi DS, Georgiev S, Biernacka EK, Goldstein BH, Shahanavaz S, Qureshi AM, Cabalka AK, Bauser-Heaton H, Torres AJ, Morray BH, Armstrong AK, Millan-Iturbe O, Peng LF, Aboulhosn JA, Ruzyllo W, Berger F, Sondergaard L, Schranz D, Cheatham JP, Jones TK, Ewert P, Schubert S. Reintervention and Survival After Transcatheter Pulmonary Valve Replacement. J Am Coll Cardiol. 2022;79(1):18–32. doi: 10.1016/j.jacc.2021.10.031. doi: [DOI] [PubMed] [Google Scholar]