Abstract
The future Ricochet experiment aims to search for new physics in the electroweak sector by measuring the Coherent Elastic Neutrino-Nucleus Scattering process from reactor antineutrinos with high precision down to the sub-100 eV nuclear recoil energy range. While the Ricochet collaboration is currently building the experimental setup at the reactor site, it is also finalizing the cryogenic detector arrays that will be integrated into the cryostat at the Institut Laue Langevin in early 2024. In this paper, we report on recent progress from the Ge cryogenic detector technology, called the CryoCube. More specifically, we present the first demonstration of a 30 eVee (electron equivalent) baseline ionization resolution (RMS) achieved with an early design of the detector assembly and its dedicated High Electron Mobility Transistor (HEMT) based front-end electronics with a total input capacitance of about 40 pF. This represents an order of magnitude improvement over the best ionization resolutions obtained on similar phonon-and-ionization germanium cryogenic detectors from the EDELWEISS and SuperCDMS dark matter experiments, and a factor of three improvement compared to the first fully-cryogenic HEMT-based preamplifier coupled to a CDMS-II germanium detector with a total input capacitance of 250 pF. Additionally, we discuss the implications of these results in the context of the future Ricochet experiment and its expected background mitigation performance.
Introduction
The recent first observation of Coherent Elastic Neutrino-Nucleus Scattering (CENNS) by the COHERENT collaboration has opened new avenues to search for physics beyond the standard model [1, 2]. The Ricochet collaboration [3, 4] is aiming for a CENNS measurement with percentage precision down to sub-100 eV nuclear recoil energies where signatures of such new physics may arise [5]. These include for instance the existence of sterile neutrinos and of new mediators that could be related to the long lasting dark matter problem, and the possibility of non-standard neutrino interactions that would dramatically affect our understanding of the electroweak sector.
The future Ricochet experiment will be deployed at the Institut Laue Langevin (ILL) within the H7 experimental site [3] to measure with high precision the CENNS process from reactor antineutrinos. To achieve its goal, the Ricochet experiment seeks to utilize a kg-scale cryogenic detector payload combining the CryoCube and the Q-Array, two cryogenic detector technologies consisting of 18-to-27 Ge and 9 Zn crystals, respectively. The cryogenic detectors will be located 8.8 m away from the 58 MW nominal thermal power reactor leading to a CENNS event rate of approximately 12.8 and 11.2 events/kg/day with a 50 eV energy threshold in the Ge and Zn target crystals, respectively. Both technologies are being optimized to combine a sub-100 eV energy threshold with particle identification capabilities to reject both the gamma-induced electronic recoil background, and the non-ionizing low energy excess observed in all low-energy threshold cryogenic detectors [6]. Above the threshold of particle identification, the neutron-induced nuclear recoils are expected to be the limiting background to the future Ricochet experiment. A recent characterization of this neutron background has shown that, despite the 15 m.w.e. artificial overburden at the ILL-H7 site, the Ricochet neutron background will be dominated by cosmogenic neutrons and that Ricochet’s expected CENNS signal-to-noise ratio should be around unity [3].
Similarly to the EDELWEISS experiment [7], particle identification with the CryoCube will be achieved thanks to the simultaneous measurement of phonon and ionization energies. The Ge crystals are equipped with a germanium Neutron Transmutation Dopped (NTD) phonon sensor and aluminum electrodes to collect signals from the ionization induced electron–hole pairs. Setting aside the low energy excess (see Sect. 6 for a dedicated discussion), to achieve the targeted CENNS sensitivity of the Ricochet experiment, the CryoCube phonon and ionization channels are designed to achieve 10 eV and 20 eVee (eV electron equivalent) baseline energy resolutions (RMS), respectively [8]. With such performance, the CryoCube discrimination energy threshold against electron recoils is expected to be on the order of 100 eV, allowing for a percentage precision CENNS measurement. While both the SuperCDMS [9] and EDELWEISS [10] collaborations have already demonstrated phonon energy resolutions on the (10) eV-scale with tens of gram target crystals, state-of-the-art ionization energy resolution would be the limiting factor for the CryoCube. At present, ionization resolutions of 220 eVee and 350 eVee, for EDELWEISS [11] and SuperCDMS [12] respectively, leading to keV-scale discrimination energy thresholds, are an order of magnitude greater than our Ricochet design requirement. It is worth mentioning that baseline ionization energy resolutions on the order of 30 eVee have already been achieved in large p-type point contact (PPC) Ge detectors operated at 77 K [13–16]. This was accomplished in part by minimizing the input capacitance to the preamplifier stage of their silicon Field Effect Transistors (FETs). The lower input capacitance is achieved by (1) reducing the capacitance of the electrode itself to ((1) pF), and (2) reducing the capacitance of the cabling between the FET and the electrode by reducing the distance between the two. The reduction of cabling capacitance is much more difficult to achieve with sub-100 mK cyogenic detectors with FET-based electronics operated at 100 K. Indeed, reducing the distance between the detectors and the FET to a few centimeters would impose an overwhelmingly large heat load to both the cryostat and the detectors preventing any simultaneous phonon measurement. Additionally, the highly asymmetric design of the PPC electrodes produces inhomogeneities in the electric field where some low field regions ( V/cm) may be affected by reduced charge collection efficiencies in the Ge crystal operated at 10 mK and at low voltage biases ( V). Such low voltage biases are required to keep significant particle discrimination capabilities in ensuring that the resulting Neganov–Trofimov–Luke phonon energy, created from drifting the charge carriers across the crystal [17, 18], is of the order of the recoil energy.
Thanks to their much lower intrinsic current noise, input capacitance and working temperature allowing for reduced cabling length and related stray capacitance, High Electron Mobility Transistors (HEMT), developed by the Center for Nanoscience and Nanotechnology (C2N) and commercialized by CryoHEMT [19], have been identified as a replacement for the standard FET to improve the ionization resolution of semiconducting cryogenic bolometers. A first fully cryogenic HEMT-based preamplifier has been successfully operated with a CDMS-II Ge detector, characterized by a total1 input capacitance of about 250 pF, and achieved an unprecedented 91 eVee (RMS) ionization resolution [20]. This result, while falling short of the Ricochet goals (see Sect. 6), demonstrated the interest of using HEMT-based preamplifiers in the context of cryogenic detectors, and paved the way to the ongoing CryoCube developments.
In this paper we report on the first demonstration of 30 eVee (RMS) ionization baseline energy resolution with Ricochet Ge cryogenic detectors operated at 15 mK. This result has been achieved with an early design of a sub-element of the CryoCube detector array, called MiniCryoCube, hosting three 38 g Ge detectors with their dedicated HEMT-based front-end electronics thermally anchored at 1 K. The paper is organized as follows: in Sect. 2 we present the experimental setup including the CryoCube holding structure and the Ge detector design; in Sect. 3 we detail the HEMT-based preamplifier; in Sects. 4 and 5 we present our data processing and results, respectively. Finally, we discuss the implication of these first results in the context of the low-energy excess mitigation in Sect. 6, and give our conclusions in Sect. 7.
Experimental setup
The first tests of charge preamplification with HEMTs presented hereafter were performed with three CryoCube prototype detectors: RED177, RED227, and RED237. Each detector consists of a 38 g Ge cylindrical crystal of 30 mm diameter and 10 mm height equipped with two planar electrodes and a 22 mm2 NTD sensor2 glued in the middle of the top electrode (see [8, 22] for more details about the CryoCube detector designs). It is worth mentioning that detectors with similar designs have already achieved phonon baseline energy resolutions between 18 and 30 eV, using the JFET-based EDELWEISS electronics [7], hence already nearing the Ricochet specifications on the phonon performance side [8, 10].
Figure 1 (left panel) shows a photo of the RED177 planar detector with its Ge crystal held in its copper casing. Though not read out in this work, the NTD sensor is both electrically and thermally connected to the flexible Kapton printed-circuit board (PCB) thanks to a total of six 25 m diameter gold wirebonds. The aluminum electrodes fully cover the flat top and bottom surfaces. In order to limit charge trapping on the lateral side, each electrode extends on that surface by 2 mm, hence limiting the height of the exposed bare Ge surface to 6 mm. The electrodes are electrically connected to the flexible PCB thanks to a total of eight 25 m diameter aluminum wirebonds. The detector holder has been designed to both minimize vibration induced noise and reduce electrical capacitance by ensuring a minimal distance between the aluminum electrodes and the copper casing of 3 mm [8]. From our simulations we expect a mutual capacitance between each electrode and the copper casing of 4.06 pF, and a mutual capacitance between the two electrodes of 10.86 pF [22].
The middle panel of Fig. 1 is a picture of the gold plated circular copper piece mounted on the bottom of the detector holder. It hosts the feedback and load resistors as well as the coupling capacitors (see Sect. 3), heat sinking them to 10 mK. Lastly, the detector and its 10 mK stage electronic components are electrically connected to their dedicated 1 K front-end electronics via a 100 m thick flexible Kapton PCB with unshielded constantan leads (see Fig. 1 left panel).
Figure 1 (right panel) shows a fully integrated MiniCryoCube array hosting three planar detectors, namely RED177, RED227, and RED237, mounted on the 10 mK mixing chamber stage of the R &D cryostat at the Institute of Physics of the 2 Infinities (IP2I) in Lyon, France. The upper 1 K stage of the MiniCryoCube, only 5 cm above its lower 10 mK stage, is thermally anchored to the 1 K still plate thanks to a 15 mm diameter copper rod and a copper braid. The two stages of the MiniCryoCube are mechanically coupled to each other thanks to 1.6 mm thick X- and Z-shaped titanium alloy Ti-15-3-3-3 [23] support structures on the front/back and sides, respectively. With a set of heaters and thermometers located on the two MiniCryoCube stages, we measured that the total heat load from the 1 K stage on the 10 mK stage, only 5 cm below, is about 1 W. With a mixing chamber temperature set to 15 mK, the temperature at the detectors was found to be about 16 mK. Such thermal performance allows for an optimal operation of both the phonon and ionization channels of all three planar detectors simultaneously. Lastly, each HEMT-based 1 K preamplifier PCB is terminated by a 25-pin micro-D connector which interfaces with the readout cabling going directly to a 300 K feedthrough. To quantify the impact of the cabling impedance on the charge amplifier performance, all three detectors were connected via stainless steel shielded coaxial cables with either low-impedance Cu leads (RED177 and RED227) or high-impedance constantan leads (RED237).
The room temperature electronics consists of a BiLT-BN103 chassis equipped with low-noise DC sources BE2142 from iTEST and custom filters. The latter are used to bias the HEMT and the detector’s electrodes. The output signal is fed to a SR5184 amplifier from Signal Recovery, and low-pass filtered with a 8-order Bessel filter from KEMO with a cut-off frequency of 63 kHz. The signal streams are digitized at 100 or 200 kHz with a 16-bit National Instrument NI-6218 DAQ, and stored on disk for further data processing and analysis.
Prior to their installation in the cryostat the Ge detectors were exposed to an AmBe source, emitting neutrons per second, for about 114 h (RED177 and RED227) and 15 h (RED237) for 71Ge activation. This activation from neutron capture will lead to electron-capture decays from the K/L/M shells producing X-ray lines with summed energies of 10.37 keV, 1.30 keV, and 160 eV, respectively (see [24]). On top of being monoenergetic, these decays are uniformly distributed throughout the entire Ge crystal volume, allowing for a precise study of the detector response and calibration.
HEMT based common source preamplifier
The three HEMT-based preamplifiers considered in this work are in a standard common source configuration with a load resistor on the drain () [25]. The choice of using a common source scheme comes from its wide range of application. The latter is well suited to study the intrinsic properties of the transistor (e.g. bias curves, gain, and voltage noise), and leads to reasonable output gain of about 10 which is enough to be coupled to low-noise room temperature amplifiers [26]. The HEMTs and drain resistors are mounted on a custom PCB encapsulated in a copper chassis on the MiniCryoCube 1 K stage. Following our HEMT-based preamplifier model optimization [25], we only used 4.6 pF input capacitance HEMTs from CryoHEMT [19]. The dissipated power from each readout channel is given by:
1 |
with and the drain-source voltage and current bias, respectively. The standard HEMT bias operating point considered in the following of this work is { = 100 mV, = 300 A} leading to a dissipation power per HEMT channel of 120 W. In comparison to [20], our preamplifier design only uses one HEMT per ionization channel leading to significantly reduced heat load from 1 mW to 120 W per channel. With a maximum of 108 ionization channels in total for a fully instrumented CryoCube with 27 detectors, each equipped with up to 4 ionization channels [8, 22], we expect a total heat load of about 13 mW on the still stage, which can be handled by the Ricochet cryostat.
The 1 K HEMT-based common source preamplifier is connected to the 10 mK stage electronics which are located underneath the detector holder (see Fig. 1 middle panel). This stage encapsulates the total load and feedback resistors from Mini Systems Inc., and the 2 nF coupling capacitors. These components are glued on a 100 m thick flexible Kapton PCB, with copper leads, that goes around the copper casing to connect both the phonon and ionization selnsors to the readout electronics. According to capacitive measurements performed at room temperature using a precision RLC meter QuadTech 7600B and COMSOL® simulations, we found that the cabling parasitic capacitance from this flexible Kapton PCB should be between 15 pF and 25 pF, hence comparable to the detector’s mutual capacitances. As such the total input capacitance of our HEMT-based preamplifier is estimated to be about 40 pF per channel, hence about six times lower than what was achieved in [20]. According to our model, with this prototype cabling, we expect to achieve baseline energy resolutions between 27 and 35 eVee (RMS) considering a differential readout and data processing (see Sect. 4). Experimental efforts are ongoing to further reduce the parasitic capacitance from the flexible Kapton PCB down to only a few pF in order to achieve the targeted 20 eVee (RMS) baseline ionization resolution (see Sect. 7).
Data processing
Following each data acquisition, the data streams from the two ionization channels A (top electrode) and B (bottom electrode) are processed using the pipeline detailed in [27]. As the parasitic capacitance from the cabling is similar to that of the detector’s electrodes the internal noise sources from the HEMT preamplifier can propagate from one channel to the other. This results in some non-negligible fraction of the noise to be correlated between the two channels A and B, in addition to any correlated electromagnetic pick-up noise sources from the environment. As one electrode collects the holes and the other the electrons, the difference contains the summed signal amplitudes while the sum contains most of the observed noise from the two channels. The signals are therefore decorellated using an algorithm inspired by [28] about common noise sources subtraction. The result of this decorrelation is shown in Fig. 2, where the gray and black solid lines compare the noise power spectra of for RED227 before and after this decorellation procedure.
The red line on Fig. 2 is the prediction of our noise model adapted from [25] assuming a parasitic capacitance of 20 pF. The contributions of the total current, voltage and Johnson noise sources from the two ionization channels are also shown. Though a detailed study about the comparison between our HEMT-based preamplifier noise modelisation and the data is ongoing, we can already notice that our model is in reasonable agreement with the data, especially between 100 Hz and 30 kHz. We found that using this newly developed differential decorrelation processing method we improved the baseline energy resolutions by 10–20%. For the rest of this work we will only consider the decorrelated differential signal.
Results
In order to study the response of the three detectors and of these newly developed HEMT-based preamplifiers, continuous streams of 92, 75 and 65 h of data were accumulated on the detectors RED177, RED227 and RED237, respectively, from the 26th of January to the 5th of February 2023. The A and B electrodes were respectively biased at 2 V and −2 V, and the detector and preamplifer stages of the MiniCryoCube were stable at 17 mK and 1.135 K, respectively.
The resulting energy spectra, normalized to events/keVee/kg/day, from all three detectors are shown in Fig. 3. Each spectrum has been obtained after applying cuts to reject pile-up events and saturated traces but are not corrected for efficiency. In comparing the observed energy spectra from [10], using the same cryostat, crystal size, and shielding configuration, we expect an efficiency for all three detectors of about 50% above 300 eVee, i.e. 8-to-10- above the baseline noise.3 From Fig. 3 one can clearly see the X-ray lines at 1.30 and 10.37 keV emitted in the electron capture of 71Ge. These lines exhibit a non-Gaussian shoulder towards lower energies. This well-known feature is due to incomplete charge collection, and was quantified in [8] where we found an upper limit on the fraction of incomplete charge collection events of about 10% for the planar detector design. The experimental resolutions at 1.30 and 10.37 keV listed in Table 1 are the standard deviations of the Gaussian component obtained in least-square fits such as those shown in Fig. 3 (see inset panels), where the peaks are modeled as a Gaussian combined with an exponential tail. After subtracting quadratically the baseline energy resolutions from the observed line widths at 1.3 keV and 10.37 keV for all three detectors, we find residual energy dispersions between 9% and 20% larger than our expectations. The latter are computed assuming a standard Fano factor value of at 5.9 keV for Ge detectors operated at 77 K [29] and a mean energy to create an electron–hole pair of 3.0 eV [30, 31]. This could be explained by both the incomplete charge collection and the event-to-event gain variations expected with these common source preamplifiers which are not in a closed loop configuration [26]. In addition to the 71Ge K- and L-shell electron capture lines, the energy spectra shown in Fig. 3 exhibit a continuous component rising at the lowest energies. This increase in event rate below 500 eVee is attributed to the dominating neutron induced nuclear recoil background, as shown in [3]. Finally, the steep rise below 250 eVee, mostly visible for RED227 and RED237 which have lower resolutions than RED177, is due to noise induced triggers.
Table 1.
Detector | Baseline | 1.30 keV | 10.37 keV |
---|---|---|---|
RED177 | |||
RED227 | |||
RED237 |
Figure 4 (left panel) presents the energy distributions of noise traces for all three detectors (solid lines), accumulated over their continuous data acquisition periods, from which we derived the following time averaged baseline ionization energy resolutions (RMS) from a Gaussian fit (dashed curves): eVee (RED177), eVee (RED227), and eVee (RED237), see Table 1. This demonstrates for the first time that a 30 eVee ionization resolution has been achieved in a cryogenic bolometer operated at 17 mK (RED177), and that all three detectors in the array reached resolutions on the 30 eVee-scale. This also equals the performance of the best PPC ionization-only Ge detectors [13–16], operated at 77 K and thus with much relaxed heat load constraints. The worse energy resolution of RED237 is likely due to additional Johnson noise from its higher impedance constantan cabling, see Sect. 2, which reduced the gain of the RED237 HEMT preamplifier by a factor of two compared to the other detectors RED137 and RED227, further degrading its signal-to-noise ratio. Figure 4 (right panel) presents the time variation of the observed baseline energy resolution of the three detectors. Each data point corresponds to an averaged resolution value over two hours. The energy resolution of all three detectors was extremely stable over the entire acquisition time, with a dispersion of less than 1 eVee, and with no apparent time degradation. This result is particularly important in the context of the Ricochet experiment aiming at measuring the CENNS process with a high precision with year-long data acquisition time.
Mitigating the non-ionizing low-energy excess
The choice of a dual phonon-and-ionization readout for the CryoCube was initially motivated to reject the gamma induced electronic recoil and beta-decay related surface backgrounds [8]. However, as the phonon energy threshold of the bolometers improved, a new and overwhelming source of background of unknown origin has been identified. The latter, observed in all cryogenic experiments with a threshold below 100 eV, is the subject of ongoing intense worldwide investigations (see [6] for a detailed review). From the combined Ricochet-CryoCube [8] and EDELWEISS [10, 32] observations, this low-energy excess has been found to be non-ionizing. In [32] an upper limit on its ionization yield of (at 90% C.L.) has been derived. The rate of this excess, also referred to as “heat only”, has been measured to be between and events/day/kg/keV below 100 eV in different low-threshold cryogenic experiments [6]. As such, this background is about 5 orders of magnitude larger than the CENNS signal expected with the future Ricochet experiment at ILL [3]. If not mitigated efficiently, this “heat only” background could therefore jeopardize the CENNS sensitivity of the future Ricochet experiment.
This background can be rejected by requiring the presence of an ionization signal above a certain threshold. However, this comes with a loss in efficiency on the CENNS signal that depends on the ionization yield of nuclear recoils. For instance, a 3 to 6 ionization threshold cut leads to a rejection power ranging from about to [33]. Figure 5 shows the remaining CENNS event rate passing the ionization threshold (left y-axis), and the corresponding CENNS survival probability (right y-axis), as a function of the ionization baseline resolution. The calculations were done for rejection levels ranging from 3 to 6 and considering the standard Lindhard quenching factor model with [34]. Our results suggest that in the case of an overwhelming “heat only” excess, as currently observed in all cryogenic experiments, reducing the baseline ionization resolution is pivotal to the success of the future Ricochet experiment. Indeed, assuming a 4 ionization threshold corresponding to a rejection factor, as would be required from our observed “heat only” rate [8, 10, 32], the CENNS event rate increases from to 1 event/kg/day by reducing the baseline energy resolution from 91 eVee, as achieved in [20], to our targeted 20 eVee. With the already demonstrated baseline ionization resolutions presented in this work, the expected CENNS event rate surviving a 4 ionization threshold cut would be between 0.4 (RED177) and 0.2 (RED237) events/kg/day. Note that we obtain similar results when considering the recently measured CONUS quenching factor [35], and a significantly enhanced CENNS event rate by a factor of 3 when considering instead the measurement from [15]. The strong dependence of the Ricochet experiment’s sensitivity to the assumed nuclear recoil ionization yield highlights the need for precise measurements of this value near 100 eV recoil energy [36, 37], a task that should be facilitated with the major improvements in charge resolution achieved in this work.
Conclusion and outlook
In this paper we have presented the first ionization readout performance from three Ricochet germanium detector prototypes with their dedicated HEMT-based preamplifiers mounted in a CryoCube sub-element. All three detectors demonstrated 30 eVee-scale ionization energy resolutions with an estimated total input capacitance of 40 pF. This is comparable to the best resolutions achieved with ionization-only Ge detectors with few-pF input capacitance that operate with much less stringent heat load constraints [13–16]. This result corresponds to an improvement by a factor 7 and 11 with respect to previously achieved ionization resolutions from the EDELWEISS [21] and SuperCDMS [12] collaborations, both using cryogenic Ge detectors operated at 20 and 50 mK, respectively. It is also a factor of 3 better than the previously achieved 91 eVee resolution from the first cryogenic HEMT-based preamplifier that was characterized by a total input capacitance of 250 pF [20]. Additionally, our HEMT-based preamplifier dissipates about ten times less power per channel compared to [20], opening the possibility to readout 150 channels with heat sinking the cold electronics at 1 K, as required from the CryoCube detector assembly specifications.
In Sect. 6, we investigated the implication of reaching such low ionization resolutions on the mitigation of the limiting low-energy excess (so-called “heat only”). We found that by imposing an ionization threshold cut with our targeted 20 eVee ionization resolution, this overwhelming background could be reduced by about while retaining a CENNS signal event rate of 1 event/kg/day. This study also highlighted the crucial need for the Ricochet experiment to improve by about an order of magnitude the ionization performance previously achieved by the EDELWEISS and SuperCDMS collaborations to secure its CENNS sensitivity.
Though the HEMT-based common source preamplifiers considered here allowed us to first study the HEMT properties and achieve a 30 eVee-scale ionization resolution, they do not fully comply with the CryoCube readout specifications. Due to their low-gain (10) and their required low-noise amplifiers at room temperature, the collaboration has developed optimized HEMT-based preamplifiers that will be tested in the coming months [26]. The increased gain of these preamplifiers will lead to a lower sensitivity to environmental noise with a gain of 100. Additionally, they will be compatible with our newly developed room-temperature amplifiers and digitizer which will allow us to read out the phonon and ionization channels of these detectors simultaneously. Lastly, using this dedicated CryoCube readout electronics together with reduced parasitic capacitance flexible PCB, currently under development, we expect to further improve our ionization resolution down to the targeted 20 eVee.
Such ionization performance, combined with already demonstrated 20 eV-scale baseline phonon energy resolution [10], will allow to reduce the particle identification threshold down to the 100 eVnr-scale (eV nuclear recoil), hence one order of magnitude lower than that achieved in current dual-readout cryogenic dark matter experiments such as EDELWEISS [11], SuperCDMS [12], and CRESST [38]. Finally, this will simultaneously allow for (1) a precise CENNS measurement with Ricochet, (2) the possibility to directly measure the nuclear recoil ionization yield down to 100 eVnr, and (3) further characterizations and understanding the origin of the overwhelming low-energy excess affecting all low-threshold cryogenic experiments, drastically limiting their CENNS and low-mass dark matter sensitivities [6].
Acknowledgements
This project received funding from the European Research Council (ERC) under the European Union’s Horizon 2020 research and innovation program under Grant Agreement ERC-StG-CENNS 803079, the French National Research Agency (ANR) within the project ANR-20-CE31-0006, the LabEx Lyon Institute of Origins (ANR-10-LABX-0066) of the Université de Lyon, within the Plan France2030, and the NSF under Grant PHY-2209585 and PHY-2013203. We are grateful to Laurent Couraud and Antonella Cavanna for the HEMT production and development which is supported in part by the French network RENATECH. A portion of the work carried out at MIT was supported by DOE QuantISED award DE-SC0020181 and the Heising-Simons Foundation. This work is also partly supported by the Ministry of Science and Higher Education of the Russian Federation.
Data Availability Statement
This manuscript has no associated data or the data will not be deposited. [Authors’ comment: The data from this study will not be deposited but can be available from the corresponding author on reasonable request.]
Footnotes
The total input capacitance is the sum of the capacitances from the detector, the cabling and the HEMT.
The NTD sensors used for the CryoCube detector array have been retrieved from the EDELWEISS-III experiment [21] and modified to match the CryoCube specifications.
It is worth mentioning that a dedicated study about our detection efficiency, including the dual phonon-ionization readout and extended to lower energies, is currently ongoing and will be presented in a future work.
Contributor Information
Ricochet Collaboration, Email: j.billard@ipnl.in2p3.fr
Ricochet Collaboration:
C. Augier, G. Baulieu, V. Belov, L. Bergé, J. Billard, G. Bres, J. -. L. Bret, A. Broniatowski, M. Calvo, A. Cazes, D. Chaize, M. Chala, M. Chapellier, L. Chaplinsky, G. Chemin, R. Chen, J. Colas, E. Cudmore, M. De Jesus, P. de Marcillac, L. Dumoulin, O. Exshaw, S. Ferriol, E. Figueroa-Feliciano, J.-B. Filippini, J. A. Formaggio, S. Fuard, J. Gascon, A. Giuliani, J. Goupy, C. Goy, C. Guerin, E. Guy, P. Harrington, S. A. Hertel, M. Heusch, Z. Hong, J.-C. Ianigro, Y. Jin, A. Juillard, D. Karaivanov, S. Kazarcev, J. Lamblin, H. Lattaud, M. Li, A. Lubashevskiy, S. Marnieros, N. Martini, D. W. Mayer, J. Minet, A. Monfardini, F. Mounier, V. Novati, E. Olivieri, C. Oriol, L. Ovalle Mateo, P. K. Patel, E. Perbet, H. D. Pinckney, D. V. Poda, D. Ponomarev, F. Rarbi, J.-S. Real, T. Redon, F. C. Reyes, A. Robert, S. Rozov, I. Rozova, S. Scorza, B. Schmidt, Ye. Shevchik, T. Soldner, J. Stachurska, A. Stutz, L. Vagneron, W. Van De Pontseele, F. Vezzu, L. Winslow, E. Yakushev, and D. Zinatulina
References
- 1.COHERENT Collaboration, D. Akimov et al., Observation of Coherent Elastic Neutrino-Nucleus Scattering. Science 357(6356), 1123–1126 (2017). arXiv:1708.01294 [DOI] [PubMed]
- 2.COHERENT Collaboration, D. Akimov et al., First Measurement of Coherent Elastic Neutrino-Nucleus Scattering on Argon. Phys. Rev. Lett. 126(1), 012002 (2021). arXiv:2003.10630 [DOI] [PubMed]
- 3.Ricochet Collaboration, C. Augier et al., Fast neutron background characterization of the future Ricochet experiment at the ILL research nuclear reactor. Eur. Phys. J. C 83(1), 20 (2023). arXiv:2208.01760
- 4.Ricochet Collaboration, C. Augier et al., Ricochet Progress and status, in 19th International Workshop on Low Temperature Detectors, 11 (2021). arXiv:2111.06745
- 5.Billard J, Johnston J, Kavanagh BJ. Prospects for exploring new physics in coherent elastic neutrino-nucleus scattering. JCAP. 2018;11:016. doi: 10.1088/1475-7516/2018/11/016. [DOI] [Google Scholar]
- 6.Adari P, et al. EXCESS workshop: descriptions of rising low-energy spectra. SciPost Phys. Proc. 2022;9:001. doi: 10.21468/SciPostPhysProc.9.001. [DOI] [Google Scholar]
- 7.EDELWEISS Collaboration, E. Armengaud et al., Performance of the EDELWEISS-III experiment for direct dark matter searches. JINST 12(08), P08010 (2017). arXiv:1706.01070
- 8.RICOCHET Collaboration, T. Salagnac et al., Optimization and performance of the CryoCube detector for the future RICOCHET low-energy neutrino experiment, in 19th International Workshop on Low Temperature Detectors, 11 (2021). arXiv:2111.12438
- 9.SuperCDMS Collaboration, I. Alkhatib et al., Light dark matter search with a High-resolution athermal phonon detector operated above ground. Phys. Rev. Lett. 127, 061801 (2021). arXiv:2007.14289 [DOI] [PubMed]
- 10.EDELWEISS Collaboration, E. Armengaud et al., Searching for low-mass dark matter particles with a massive Ge bolometer operated above-ground. Phys. Rev. D 99(8), 082003 (2019). arXiv:1901.03588
- 11.EDELWEISS Collaboration, L. Hehn et al., Improved EDELWEISS-III sensitivity for low-mass WIMPs using a profile likelihood approach. Eur. Phys. J. C 76(10), 548 (2016). arXiv:1607.03367
- 12.SuperCDMS Collaboration, R. Agnese et al., Search for low-mass weakly interacting massive particles with SuperCDMS. Phys. Rev. Lett. 112(24), 241302 (2014). arXiv:1402.7137 [DOI] [PubMed]
- 13.CDEX Collaboration, L.T. Yang et al., Limits on light WIMPs with a 1 kg-scale germanium detector at 160 eVee physics threshold at the China Jinping Underground Laboratory. Chin. Phys. C 42(2), 023002 (2018). arXiv:1710.06650
- 14.GeN Collaboration, I. Alekseev et al., First results of the GeN experiment on coherent elastic neutrino-nucleus scattering. Phys. Rev. D 106(5), L051101 (2022). arXiv:2205.04305
- 15.J.I. Collar, A.R.L. Kavner, C.M. Lewis, Germanium response to sub-keV nuclear recoils: a multipronged experimental characterization. Phys. Rev. D 103(12), 122003 (2021). arXiv:2102.10089
- 16.CONUS Collaboration, H. Bonet et al., Constraints on elastic neutrino nucleus scattering in the fully coherent regime from the CONUS experiment. Phys. Rev. Lett. 126(4), 041804 (2021). arXiv:2011.00210 [DOI] [PubMed]
- 17.Neganov B, Trofimov V. Calorimetric method measuring ionizing radiation. Otkryt. Izobret. 1985;146:215. [Google Scholar]
- 18.P.N. Luke, Voltage-assisted calorimetric ionization detector. J. Appl. Phys. 64(6858), 6858 (1988)
- 19.Dong Q, Liang Y-X, Ferry D, et al. Ultra-low noise high electron mobility transistors for high-impedance and low-frequency deep cryogenic readout electronics. Appl. Phys. Lett. 2014;105:013504–013504. doi: 10.1063/1.4887368. [DOI] [Google Scholar]
- 20.A. Phipps, A. Juillard, B. Sadoulet, B. Serfass, Y. Jin, A HEMT-based cryogenic charge amplifier with sub-100 eVee ionization resolution for massive semiconductor dark matter detectors. Nucl. Instrum. Methods A 940, 181–184 (2019). arXiv:1611.09712
- 21.EDELWEISS Collaboration, L. Hehn et al., Improved EDELWEISS-III sensitivity for low-mass WIMPs using a profile likelihood approach. Eur. Phys. J. C 76(10), 548 (2016). arXiv:1607.03367
- 22.D. Misiak, Développements de nouveaux détecteurs cryogéniques bas seuils pour la recherche de matiére noire légére et la physique des neutrinos de basse énergie (Theses, Université de Lyon, Feb., 2021)
- 23.Wikus P, Hertel S, Leman S, et al. The electrical resistance and thermal conductivity of ti 15v–3cr-3sn-3al at cryogenic temperatures. Cryogenics. 2011;51(1):41–44. doi: 10.1016/j.cryogenics.2010.10.010. [DOI] [Google Scholar]
- 24.EDELWEISS Collaboration, Q. Arnaud et al., First germanium-based constraints on sub-MeV Dark Matter with the EDELWEISS experiment. Phys. Rev. Lett. 125(14), 141301 (2020). arXiv:2003.01046 [DOI] [PubMed]
- 25.A. Juillard et al., Low-noise HEMTs for coherent elastic neutrino scattering and low-mass dark matter cryogenic semiconductor detectors. J. Low Temp. Phys. 199(3-4), 798–806 (2019). arXiv:1909.02879
- 26.Ricochet Collaboration, G. Baulieu et al., HEMT-based 1 K front-end electronics for the heat and ionization Ge CryoCube of the future Ricochet CENS Experiment. J. Low Temp. Phys. 209(3-4), 570–580 (2022). arXiv:2111.10308
- 27.RICOCHET Collaboration, J. Colas, J. Billard, S. Ferriol, J. Gascon, T. Salagnac, Development of data processing and analysis pipeline for the RICOCHET experiment. arXiv:2111.12856
- 28.B. Allen, W.-S. Hua, A. C. Ottewill, Automatic cross talk removal from multichannel data. arXiv:gr-qc/9909083
- 29.Lowe B . Measurements of fano factors in silicon and germanium in the low-energy x-ray region. Nucl. Instrum. Methods A. 1997;399(2):354–364. doi: 10.1016/S0168-9002(97)00965-0. [DOI] [Google Scholar]
- 30.EDELWEISS Collaboration, A. Benoit et al., Measurement of the response of heat-and-ionization germanium detectors to nuclear recoils. Nucl. Instrum. Methods A 577, 558–568 (2007). arXiv:astro-ph/0607502
- 31.Knoll G. Radiation Detection and Measurement. New York: Wiley; 2010. [Google Scholar]
- 32.EDELWEISS Collaboration, E. Armengaud et al., Search for sub-GeV dark matter via the Migdal effect with an EDELWEISS germanium detector with NbSi transition-edge sensors. Phys. Rev. D 106(6), 062004 (2022). arXiv:2203.03993
- 33.Particle Data Group Collaboration, R.L. Workman et al., Review of Particle Physics, Sec. 40 Statistics. PTEP2022, 083C01 (2022)
- 34.J. Lindhard, Influence of crystal lattice on motion of energetic charged particles. Kongel. Dan. Vidensk. Selsk., Mat.-Fys. Medd. 34(14) (1965)
- 35.A. Bonhomme et al., Direct measurement of the ionization quenching factor of nuclear recoils in germanium in the keV energy range. Eur. Phys. J. C 82(9), 815 (2022). arXiv:2202.03754
- 36.CRAB, NUCLEUS Collaboration, H. Abele et al., Observation of a nuclear recoil peak at the 100 eV scale induced by neutron capture. arXiv:2211.03631 [DOI] [PubMed]
- 37.SuperCDMS Collaboration Collaboration, M.F. Albakry, I. Alkhatib, D. Alonso, et al., First measurement of the nuclear-recoil ionization yield in silicon at 100 ev. Phys. Rev. Lett. 131, 091801 (2023) [DOI] [PubMed]
- 38.CRESST Collaboration, A.H. Abdelhameed et al., First results from the CRESST-III low-mass dark matter program. Phys. Rev. D 100(10), 102002 (2019). arXiv:1904.00498
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
This manuscript has no associated data or the data will not be deposited. [Authors’ comment: The data from this study will not be deposited but can be available from the corresponding author on reasonable request.]