. 2024 Feb 21;10(5):e26147. doi: 10.1016/j.heliyon.2024.e26147

Enhancing ECG classification with continuous wavelet transform and multi-branch transformer

Chenyang Qiu 1, Hao Li 1,, Chaoqun Qi 1, Bo Li 1
PMCID: PMC10906304  PMID: 38434292

Abstract

Background

Accurate classification of electrocardiogram (ECG) signals is crucial for automatic diagnosis of heart diseases. However, existing ECG classification methods often require complex preprocessing and denoising operations, and traditional convolutional neural network (CNN)-based methods struggle to capture complex relationships and high-level time-series features.

Method

In this study, we propose an ECG classification method based on continuous wavelet transform and multi-branch transformer. The method utilizes continuous wavelet transform (CWT) to convert the ECG signal into time-series feature map, eliminating the need for complicated preprocessing. Additionally, the multi-branch transformer is introduced to enhance feature extraction during model training and improve classification performance by removing redundant information while preserving important features.

Results

The proposed method was evaluated on the CPSC 2018 (6877 cases) and MIT-BIH (47 cases) ECG public datasets, achieving an accuracy of 98.53% and 99.38%, respectively, with F1 scores of 97.57% and 98.65%. These results outperformed most existing methods, demonstrating the excellent performance of the proposed method.

Conclusion

The proposed method accurately classifies the ECG time-series feature map, which holds promise for the diagnosis of cardiac arrhythmias. The findings of this study are valuable for advancing the field of automatic ECG diagnosis.

Keywords: Arrhythmia, Multi-branch transformer, Continuous wavelet transform, Convolutional neural network, Time-series feature map

1. Introduction

The electrocardiogram (ECG) is a common method of cardiac monitoring and plays a crucial role in the diagnosis and monitoring of arrhythmias [[1], [2], [3], [4]]. However, the complexity and variability of ECG signals make the classification task extremely challenging. Traditional ECG classification methods rely on manual feature extraction and classifier design, which are prone to human error and require extensive expertise and experience [5]. As a result, there is a pressing need for more advanced and automated methods of ECG classification that can improve diagnostic accuracy and reduce the workload of medical professionals.

Continuous wavelet transform (CWT) [6] is a signal analysis technique widely used to transform ECG signals into 2D images, which can decompose ECG signals into wavelet coefficients of different scales and orientations, and then efficiently extract useful feature information by combining these coefficients into time-series feature map. Transformer [7] is a powerful deep learning tool that can adaptively learn important feature information and improve the classification ability of the model.

In recent years, with the development of deep learning technology, ECG classification methods based on deep learning have gradually received wide attention. For instance, Acharya et al. [8] proposed a deep convolutional neural network (DCNN) [9] ECG classification method, which can automatically detect cardiovascular diseases such as myocardial infarction. However, the method cannot learn from past decisions and features, which can lead to clusters being mixed with other classes and affecting classification results. To address this issue, Hannun et al. [10] introduced a long short-term memory (LSTM) network [11] ECG classification method that can also automatically detect arrhythmias, but requires data preprocessing. Recently, the transformer has been widely used in the field of ECG classification. For example, Ramkumar et al. [12] developed a bidirectional long-term and short-term memory network that utilizes an attention mechanism to adaptively learn important ECG features, enabling the classification of various arrhythmias. However, this method requires complex data preprocessing operations and consumes significant computational resources and time, which can affect the model's accuracy. Yang et al. [13] used a multi-view approach to fuse different lead features and used a multi-scale convolutional neural network to obtain ECGs of different scales with temporal characteristics. Lai et al. [14]collected a large number of wearable 12-lead ECG datasets and achieved real-time intelligent diagnosis through four data enhancement operations and a self-supervised learning classification framework. In addition, Han et al. [15] used the Gramian Angular Field (GAF) to map the original ECG signal into feature map, introduced a multi-instance learning (MIL) method to avoid information loss, and used a feature fusion method based on the attention mechanism to achieve accurate classification.

Although effective, the ECG classification methods mentioned above have certain limitations. Firstly, 1DCNN or LSTM-based [16,17] methods require signal preprocessing and manual feature selection, which can be prone to human error and cannot learn past decisions and features. Secondly, 2DCNN or transformer-based [[18], [19], [20], [21], [22]] methods, while not requiring preprocessing operations, struggle to extract complex relationships between time steps in the time-series feature map and high-level time-series features. To overcome these challenges, this paper proposes an ECG classification method that utilizes CWT and multi-branch transformer. This method converts the original ECG signal into time-series feature map using CWT, eliminating the need for complicated preprocessing. Additionally, the multi-branch transformer is introduced to enhance feature extraction and improve classification performance by removing redundant information while preserving important features. Experimental results demonstrate that the proposed method achieves excellent performance in ECG classification tasks.

This work makes a twofold contribution. Firstly, a multi-branch transformer-based ECG classification method is introduced. The method utilizes convolution to extract deep features of the image, and multi-branch transformer to extract global features of the image. This operation enables the accurate classification of various arrhythmias. Secondly, an enhanced multi-headed self-attention mechanism is used to compute only the useful features in the time-series feature map, which improves the efficiency while still focusing on the subtle changes of the signal.

The rest of this paper is structured as follows. The proposed method is described in Section 2. Section 3 describes the dataset used in this study with data enhancement and partitioning. The results of the study are analyzed in Section 4, and Section 5 discusses the results. Finally, conclusions and an outlook for future research are given in Section 6.

2. Methodologies

2.1. Overall architecture of ECG classification method

The overall flow of the proposed ECG classification method based on CWT and multi-branch transformer is shown in Fig. 1. Firstly, the obtained electrocardiogram signal is mapped into a time-series feature map through continuous wavelet transform. Then, the obtained image is data enhanced, and the Multi-branch Transformer algorithm is used for feature extraction and classification, ultimately obtaining different categories of arrhythmia.

Fig. 1.

Fig. 1

The illustration of the proposed classification method. The figure shows the flowchart of the proposed classification method, including data preprocessing, data Enhancement, feature extraction, and classification.

2.2. Continuous wavelet transform

The wavelet transform can be divided into two forms: Continuous Wavelet Transform (CWT) [6] and Discrete Wavelet Transform (DWT) [23]. It is a multi-scale analysis technique that decomposes the original signal into approximate and detailed coefficients at varying scales. The approximate coefficients represent the low-frequency part of the signal, while the detail coefficients represent the high-frequency part of the signal. CWT is faster to compute than DWT, has better localization and multi-resolution characteristics, and is, therefore, used in this paper to convert ECG signals into time-series feature maps. The calculation of CWT is depicted in Eq. (1).

Ca,b=1|a|+x(t)ψ*(tba)dt (1)

where x(t) is the input signal, ψ(t) is the wavelet basis function, a and b are the scale and translation parameters, and * denotes the complex conjugate. Ca,b are the wavelet coefficients representing the components of the signal at scale a and position b. CWT has a better adaptive capability in terms of time-frequency resolution, which can better highlight the local feature of the actual ECG signal.

2.3. Multi-branch transformer-based ECG classification algorithm

The overall architecture of the multi-branch transformer-based ECG classification algorithm is illustrated in Fig. 2. The algorithm takes the original ECG signal as input and converts it into time-series feature map using CWT. In the feature extraction module, local features are extracted using improved CNN, and the extracted features are then weighted and fused by multi-branch transformer to enhance the classification performance. Finally, a fully connected layer maps the features of each arrhythmia class to obtain the final classification results. This algorithm not only has the transformer's ability to extract global features from images but also avoids its dependence on large datasets.

Fig. 2.

Fig. 2

Overall architecture of ECG classification algorithm. The figure shows the overall architecture of the proposed ECG classification algorithm, which consists of two main parts: the improved CNN and the multi-branch transformer.

2.3.1. Improved CNN

The proposed algorithm utilizes improved CNN for the local feature extraction module, comprising four convolutional blocks. Each block contains multiple convolutional layers and batch normalization layers. The inputs and outputs of the convolutional blocks are connected through residuals, which boosts network performance without increasing the parameter count. Fig. 2 illustrates the network structure.

In practical ECG classification tasks, it is important to consider time efficiency. To reduce the computational and storage overhead of the model, the improved CNN uses depthwise separable convolution (DSC) [24] instead of the standard convolutional layer. This approach can significantly decrease the number of parameters and computation of the model, leading to improved classification performance. The calculation process is demonstrated in Eq. (2).

F(x)=BN(ReLU(DSC(Conv(x)))) (2)

2.3.2. Multi-branch transformer

To improve the robustness and classification performance of the network in extracting ECG features, this paper adopts a cross structure comprising convolution and transformer blocks in the global feature extraction module. The multi-branch transformer structure can be seen in Fig. 2.

The multi-branch transformer layer enhances the input features for computation, while also establishing connections between different categories. It achieves this by connecting and interacting with each category to obtain EMHSA1(X). The final result is obtained by passing the original sequence through the second multi-headed attention module, and then EMHSA1(X) acts as the residual connection to get X0. For specific details on the calculation process, please refer to Eq. (3).

X0=EMHSA1(X)+EMHSA2(X)+X (3)

In the equation above, the outputs of the two branches are connected with residuals, which allows the modeling of the inputs at different levels and granularities. Finally, the feedforward neural network (FNN) [25] is added to obtain the final output of this transformer block. As shown in Eq. (4).

Y=LN(X0)+FNN(LN(X0)) (4)

The multi-branch transformer can capture various relationships in different branches and flexibly control the flow of information and interactions between them. This ability allows for a more comprehensive description of the interactions between sequences, improving the learning and generalization capabilities of the model.

2.3.3. Enhanced multi-headed self-attention mechanism

The traditional transformer employs a multi-headed self-attention mechanism [7] that captures only intra-sequence dependencies. It fails to consider inter-sequence or long-range dependencies across time steps and often requires considerable computational resources. Additionally, when dealing with time-series feature maps, it is inefficient and redundant to compute attention between each pixel in pairs. This is because the local regions surrounding each pixel share similar features.

To address the issues mentioned above, this paper proposes the integration of enhanced multi-headed self-attention (E-MHSA) into ECG classification networks. Unlike the traditional multi-headed self-attention mechanism, E-MHSA not only captures intra-sequence dependencies but also has the potential to handle inter-sequence and long-range dependencies in the input feature graph. Moreover, it exhibits high computational efficiency and requires minimal memory consumption. The architecture of E-MHSA is illustrated in Fig. 3.

Fig. 3.

Fig. 3

E-MHSA structure diagram. The figure shows the structure diagram of the proposed E-MHSA (Enhanced Multi-Head Self-Attention). The model consists of several layers of self-attention modules, where each module has multiple attention heads. The input to the model is a sequence of embeddings, which are transformed by the self-attention modules to capture the interactions between different parts of the sequence. In addition, the model also incorporates relative position coding to improve its performance. The proposed E-MHSA model has been shown to achieve state-of-the-art results on ECG classification tasks.

Suppose the input sequence is XRn×d,RwRw×d,RhRh×d, where n is the sequence length, and d is the feature dimension. A linear transformation of the input sequence yields Q=XWQ,K=XWK,V=XWV, where WQ,WK,WVRd×h is the weight matrix of the linear transformation, and h is the number of heads. Then, the weights of Q,K,V, respectively by the number of heads h are sliced, and the dimension of each head is d/h. First, calculate the attention score matrix SRh×n×n,CRh×n×n, see Eq. (5).

{Si,j=QiKjTd/hCi,j=(Rw+Rh)QiTd/h (5)

Then, the attention score matrix is subjected to softmax operation to obtain the attention matrix ARh×n×n, see Equation (6).

Ai,j=eSi,jk=1neSi,k+eCi,jk=1neCi,k (6)

Finally, the attention matrix A with the sliced V matrix is weighted and summed to obtain the output feature matrix ORn×d, see Eq. (7).

O=concat(head1,head2,,headh)WO (7)

In the above equation, the headi=j=1nAi,jVi,j represents the output of the first i output of the first attention head, and concat denotes the merging of the output from each attention head in the feature dimension, while WORd×d is the weight matrix of the output matrix.

E-MHSA exhibits improved performance and computational efficiency in handling long sequences and high-dimensional data by reducing the number of attention heads while ensuring accuracy. This allows for efficient processing of large amounts of data without sacrificing performance.

2.3.4. Relative position coding

To classify ECG time-series features, this paper utilizes relative position coding. This method offers superior accuracy in coding positional relationships, as well as greater robustness to time-series shifts. Additionally, it enhances generalization to new datasets and tasks while effectively modeling non-linear relationships. These benefits contribute to improved accuracy and robustness in classification results, particularly when coupled with E-MHSA. As illustrated in Fig. 3, R represents the relative position coding.

3. Experiment setup

3.1. Experimental dataset

In this study, we first convert the original ECG signal into time-series feature map using CWT, followed by feature extraction and classification to obtain the final results. The experiments were conducted using the China Physiological Signals Challenge 2018 (CPSC 2018) [26] and ECG recordings from the MIT-BIH [27], the latter were developed in collaboration with MIT and Beth Israel Medical Center. These datasets were used to train, test, and validate the robustness of our ECG classification algorithm. The arrhythmia categories in both datasets were independently annotated by two or more cardiac experts to ensure their authority. Table 1 shows the number of cases and class distribution for each dataset.

Table 1.

Assigning patients to categories. The letter G stands for group and the first row G1-G9 corresponds to the 9 categories in CPSC 2018 (Male 3699, Female 3178). The second row G1-G5 corresponds to the 5 categories in MIT-BIH (Male 25, Female 22).

Dataset G1 G2 G3 G4 G5 G6 G7 G8 G9 Total
CPSC 2018 918 1098 704 207 1695 574 653 826 202 6877
MIT-BIH 27 5 5 5 5 47
#patients 945 1103 709 212 1700 574 653 826 202 6924
#instances 2445 2603 2209 1712 3200 574 653 826 202 14,424

The CPSC 2018 dataset [Fig. 4 (a)] consists of 6877 12-lead ECG recordings, with durations ranging from 6 to 60 s and a sampling rate of 500 Hz. This dataset includes nine categories: Normal (N), Atrial fibrillation (AF), First-degree atrioventricular block (I-AVB), Left bundle branch block (LBBB), Right bundle branch block (RBBB), Premature atrial contraction (PAC), Premature ventricular contraction (PVC), ST-segment depression (STD), and ST-segment elevated (STE).

Fig. 4.

Fig. 4

Time-series feature map after CWT. (a) For CPSC 2018 dataset and (b) for MIT-BIH dataset. The figure shows the time-series feature maps obtained using continuous wavelet transform (CWT) for two different ECG datasets: CPSC 2018 and MIT-BIH. CWT is applied to the raw ECG signals to obtain a time-frequency representation, which captures both the temporal and spectral information of the signals. The resulting feature maps are then used as inputs to the proposed algorithm for ECG classification. The figure demonstrates the differences in the feature maps between the two datasets, highlighting the need for dataset-specific feature engineering in ECG analysis.

Furthermore, the MIT-BIH dataset [Fig. 4 (b)] includes 48 half-hour ECG recordings, including 23 arrhythmias and normal sinus rhythm, with a sampling rate of 360 Hz and two signal channels per recording. According to the ANSI/AAMI EC57 classification proposed by the Association for the Advancement of Medical Instrumentation (AAMI), the MIT-BIH arrhythmia database is divided into five categories: Normal (N), Supraventricular premature beats (S), Ventricular premature beats (V), Atrial fibrillation (F), and Unknown category (Q).

Fig. 4 displays the time-series feature maps obtained after the CWT transformation. Generally, information that is not easily obtained in the time domain can be obtained in the frequency domain. Although some 1D signal to 2D image methods can achieve this effect, CWT can more accurately capture the frequency domain information of the timing signal without losing the time domain information. Moreover, in ECG signals, some waveforms appear random in 1D signals but have more distinct features or show strong regionality in 2D images. Therefore, converting the original signal into time-series feature map using CWT is more conducive to feature extraction.

3.2. Dataset segmentation and data enhancement

The dataset used in this study is not evenly distributed, which is mainly due to patient privacy and the complex labeling task. To ensure that the neural network is not overfitting and to improve the overall performance of the classification algorithm, data enhancement is necessary to balance the number of normal and arrhythmia analogies. This paper employs various data augmentation techniques during the data preprocessing stage, such as brightness adjustment, rotation, horizontal flipping, scaling, and cropping, to increase the diversity and richness of the data, and enhance the adaptability of the model to different datasets. Moreover, the training set is approximately 8 times larger than the validation set, as illustrated in Fig. 5.

Fig. 5.

Fig. 5

Distribution of images across the 9 different classes. This figure depicts the distribution of images across the nine different classes in the CPSC 2018 dataset, which was utilized in the ECG classification task after data enhancement. The y-axis indicates the number of images, while the x-axis represents the nine different classes. The largest class is the RBBB class, consisting of 1695 images, whereas the smallest class is the N class, which comprises only 908 images. It is worth noting that class imbalance is a common issue in image classification tasks that can have a negative impact on the performance of the network. Therefore, to overcome this problem, appropriate data enhancement techniques and class weighting strategies are often employed.

3.3. Study environment and parameter settings

The study discussed in this paper utilized PyCharm as the integrated development environment and the deep learning framework Pytorch. A detailed summary of the hardware configuration and software environment used in the study is presented in Table 2. The input image size was set to 224 × 224, with a fixed learning rate of 0.0001 used during training. The training process was conducted for 50 rounds (Epoch), with each training batch size set at 32, and the optimization algorithm used was adaptive gradient descent (Adam).

Table 2.

Experimental environment.

Hardware or Software Version or Model
Operating System Windows 11
CPU Intel Core i7 12,700
GPU NVIDIA GeForce RTX 3070Ti
Graphics Memory 8G
Operating Memory 16G
Integrated Development Environment PyCharm 2022.1
Programming Languages Python 3.9
Deep Learning Framework Pytorch 1.11
CUDA CUDA 11.4

3.4. Evaluation indicators

In the experiment, the algorithm is evaluated using Accuracy (Acc), Precision (Pre), Recall (Rec), and F1-score as evaluation indicators. The formulas for calculating these metrics are provided in Eqs. (8), (9), (10), (11).

Acc=TP+TNTP+FN+FP+TN (8)
Pre=TPTP+FP (9)
Rec=TPTP+FN (10)
F1=2×Pre×RecPre+Rec (11)

where TP, TN, FP, and FN represent the number of true positives, true negatives, false positives, and false negatives predicted by the model for positive and negative classes. Moreover, a higher F1 score indicates better classifier performance and a score closer to 1 is considered optimal.

4. Results

4.1. Comparison study of different wavelet transform

In this section, we first used DWT and CWT to map the original electrocardiogram signals into time-series feature maps and tested the performance of the classifier on the CPSC 2018 dataset. From Table 3, it can be seen that CWT can transform signals at any time scale, providing richer frequency information, and is suitable for ECG signal processing and multi-scale analysis.

Table 3.

The comparison of Discrete Wavelet Transform (DWT) and Continuous Wavelet Transform (CWT).

Indicators (%) DWT CWT
Acc 98.42 98.53
Pre 98.01 98.19
Rec 96.58 96.95
F1 97.22 97.57

4.2. Comparison study of different models

In this experimental section, we validate the proposed ECG classification algorithm on the CPSC 2018 arrhythmia public dataset. To demonstrate the superiority of our approach, we compare it not only with Vgg16 [28] and ResNet50 [29], which use a convolutional neural network as the infrastructure but also with several vision transformer-based classification algorithms, namely Vision transformer (ViT) [30], Data-efficient image transformer (DeiT) [31], Pooling-based vision transformer (PiT) [32] and Swin transformer (SwinT) [33]. It is worth noting that all models were trained without the use of pre-trained weights. Fig. 6 presents the comparison results of the different algorithms.

Fig. 6.

Fig. 6

Comparison studies of different models. The figure shows a comparison of the performance of different ECG classification tasks. The y-axis represents the evaluation metric used to measure the performance, and the x-axis represents the different algorithms being compared. This figure shows the advantages of our algorithm.

Based on Fig. 6, Fig. 7, it is evident that the proposed algorithm delivers superior performance across all evaluation metrics, particularly in accuracy, with an impressive score of 98.53%, which is approximately 2.5% higher than the other models on average. Notably, the Vgg16 and ResNet50 models performed relatively well, with accuracies of 97.01% and 97.60%, respectively. On the other hand, the transformer family models exhibited average performance, with DeiT and ViT recording accuracy scores of 94.73% and 95.64%, respectively. Furthermore, our proposed algorithm outperformed PiT and SwinT, which achieved accuracy rates of 96.80% and 97.48%, respectively. In terms of precision, recall, and F1 score, our algorithm delivered superior results of 98.19%, 96.95%, and 97.57%, respectively, compared to other models. These findings highlight the remarkable performance advantage of our proposed algorithm for ECG classification problems, providing accurate and reliable diagnostic support for heart diseases in the field of clinical diagnosis and monitoring.

Fig. 7.

Fig. 7

Confusion matrix of the CPSC 2018 dataset. This figure shows the performance of the proposed method on the CPSC 2018 dataset.

4.3. Ablation study

To validate the efficiency of the algorithm proposed, a range of ablation experiments have been carried out in this section, utilizing the enhanced CNN, multi-branch transformer (MT), and E-MHSA. These experiments have all been conducted on the CPSC 2018 dataset.

In Fig. 8, we compare the classification performance of different parts, with the vertical axis representing the classification accuracy. The results demonstrate that using the improved CNN leads to a 0.34% increase in classification performance. When incorporating the multi-branch transformer, there is a further improvement of 0.12%. Moreover, by replacing the original MHSA with E-MHSA in the multi-branch transformer, the accuracy reaches the highest point of 98.53%. These findings suggest that the proposed algorithm is effective in extracting crucial point information from the time-series feature map, leading to a superior classification performance.

Fig. 8.

Fig. 8

Ablation studies of various parts. The figure shows the results of ablation studies, which involve systematically removing or disabling various parts of the proposed algorithm to evaluate their impact on performance. The y-axis represents the evaluation metric used to measure the performance. The x-axis represents the different ablation conditions being compared. The figure demonstrates the relative contributions of different parts of the algorithm, allowing researchers to identify the most important components and optimize their design.

4.4. Robustness study

The study employed the same classification algorithm, parameter settings, and evaluation metrics as the previous paper. It is worth noting that all models were trained without using pre-trained weights. The comparison results of different algorithms are presented in Fig. 9. As depicted in Fig. 9, Fig. 10, the proposed algorithm outperformed other algorithms on the MIT-BIH public ECG dataset with an accuracy rate of 99.38%, which is considerably higher than that of other algorithms. Furthermore, it exhibited the best results concerning precision, recall, and F1 score. Consequently, the proposed algorithm demonstrated excellent performance and robustness on various datasets and effectively improved the accuracy and efficiency of ECG signal classification.

Fig. 9.

Fig. 9

Robustness studies of the proposed algorithm. The figure shows the results of robustness studies, which involve testing the performance of our proposed algorithm on the MIT-BIH ECG benchmark dataset. The y-axis represents the evaluation metric used to measure the performance, the x-axis represents the different algorithms being compared. The figure demonstrates the ability of the proposed algorithm to maintain high performance, indicating its robustness and generalizability.

Fig. 10.

Fig. 10

Confusion matrix of the MIT-BIH dataset. This figure shows the performance of the proposed method on the MIT-BIH dataset.

4.5. Comparison of different models with and without noises

We conducted a thorough analysis of the model's ability to handle noise in ECG signals. To address the need for extended sequence data, we selected the MIT-BIH dataset, where each record spans 1800 s. Table 4 illustrates that, in the presence of noise in the original signal, both convolution and Transformer-based algorithms exhibit a declining performance trend. Nevertheless, our algorithm sustains an accuracy of 98.19% in such situations. Subsequently, following the denoising of the original signal, our algorithm exhibits a substantial performance enhancement, reaching an accuracy of 99.38%, representing a 1.19% increase compared to the pre-denoising stage. Consequently, our algorithm showcases exceptional performance in managing extended sequences and noisy ECG signals.

Table 4.

Comparison of different models with and without noises.

Methods Noises
Without Noises
F1 (%) Acc (%) Rec (%) Pre (%) F1 (%) Acc (%) Rec (%) Pre (%)
Vgg16 [28] 91.26 96.13 91.19 91.33 91.81 96.23 91.90 91.93
ResNet50 [29] 91.31 96.43 91.83 91.46 92.39 96.26 92.10 91.93
ViT [30] 89.07 90.87 89.17 89.17 90 91.73 90.03 90.13
DeiT [31] 89.56 90.53 89.70 89.70 90.38 91.78 90.57 90.57
PiT [32] 91.28 96.59 91.33 91.33 92.19 96.48 92.33 92.33
SwinT [33] 97.09 98.14 96.49 97.58 97.92 98.53 97.93 98.07
Proposed method 97.85 98.19 97.95 98.09 98.65 99.38 98.89 98.96

5. Discussion

In this research, we utilized the CWT method to transform the original one-dimensional ECG signal into time-series feature map. Additionally, Table 5 lists other relevant studies [[34], [35], [36], [37], [38], [39]]. Although some of these methods were not specifically intended for ECG signals, they all address the classification of time-series signals and thus have some reference value. The CWT method used in this paper achieved the highest classification accuracy of 99.38%, which is superior to other methods. By using CWT to convert ECG signals into time-series feature maps, the signal features can be more comprehensively described, leading to improved classification accuracy, generalization performance, and classifier robustness.

Table 5.

Nine methods review. The methods include: Short Time Fourier Transform (STFT), Markov Transition Field (MTF), Gramian Angular Field (GAF), Recurrence Plots (RP), Motif Difference Field (MDF), Relative Position Matrix (RPM), Superlet Transform (SLT), Finite Difference Method (FDM). The Classifiers include: Denoised Diffusion Probabilistic Model (DDPM), Normal Cloud Representation CNN (NCR CNN).

Literature Methods Classifier Datasets Accuracy (%)
Kim et al. [34] STFT CNN CinC 2017 99.26
Adib et al. [35] MTF DDPM MIT-BIH 98.00
Ahmad et al. [36] GAF CNN MIT-BIH 98.40
Mathunjwa et al. [37] RP DCNN MIT-BIH 98.36
Zhang et al. [38] MDF CNN TwoLeadECG 98.96
Hssayni et al. [39] RPM NCR CNN TwoLeadECG 97.42
Tripathi et al. [40] SLT CNN MIT-BIH 96.20
Kauppinen et al. [41] FDM CNN MIT-BIH 98.01
Proposed method CWT CNN + MT MIT-BIH 99.38

To efficiently learn features in temporal spectrum images, this paper proposes a deep learning algorithm based on a multi-branch transformer for classifying arrhythmias. Using the nine methods outlined in Table 5, the algorithm extracts time-series feature maps, which are then used for feature extraction and classification. Our algorithm achieves a remarkable classification accuracy of up to 99.38%, as well as the highest F1 score, recall, and accuracy of 98.65%, 98.89%, and 98.96%, respectively, as demonstrated in Fig. 11. These results demonstrate the feasibility of our research solution and the effectiveness of the multi-branch transformer in extracting useful features and accurately classifying arrhythmias.

Fig. 11.

Fig. 11

Comparison of the nine methods. The figure shows, on the MIT-BIH, the performance of our algorithm after converting 1D ECG signals to time-series feature maps using nine different methods. The y-axis represents the evaluation metrics used to measure the performance and the x-axis represents the different methods being compared. The figure shows the advantages of the CWT method.

In the initial studies on classifying ECG signals, most of the methods used were based on one-dimensional signals [8,12,17], and a combination of multiple features, including temporal and frequency features, were empirically extracted. These methods required filtering the original signal, which resulted in the loss of significant information and some limitations. However, with the development of deep learning techniques, it has become possible to learn useful features from the original ECG signal, leading to the emergence of automatic arrhythmia classification based on deep learning, which is currently a popular research area [17]. One-dimensional convolutional neural networks have demonstrated good results in arrhythmia classification, but these methods only analyze the morphological features of ECG recordings, neglecting frequency and energy distribution information. To address these limitations, several studies have used the fusion of one-dimensional signals and time-series feature maps [[42], [43], [44]], allowing for encoding of key points of the original signal, but still requiring complex preprocessing and feature loss. In contrast, our study converts the original 1D signal directly into time-series feature map and employs an end-to-end approach for direct feature extraction of the time-series feature map, thereby improving classification accuracy while preserving the original features to a great extent. This approach differs from previous studies presented in Table 6.

Table 6.

Literature review.

Literature Features Set Classifier Results (%)
Rec Acc
Acharya et al. [8] R-peaks detection and11-layer deep neural network Convolutional neural network 95.49 95.22
Ramkumar et al. [12] Dual tree complex wavelet transform Auto- Encoder and Bidirectional long short-term memory 99.43 97.15
Yang et al. [13] Multi-view approach Multi-scale convolutional neural network 95.47
Lai et al. [14] Deep neural network 87.30 96.90
Han et al. [15] Gramian angular field and multi-instance learning Convolutional neural network and attention mechanism
Liang et al. [17] Convolutional neural network- recurrent neural network 96.20 95.15
Vijayakumar et al. [42] Denoising filter Decision tree 94.35 96.50
Mazaheri et al. [43] Denoising filter Feed forward neural network 98.87 98.75
Le et al. [45] Fusion transformer encoder 98.29
Che et al. [46] Difference method and wavelet transform Convolutional neural network and transformer 87.80
Wang et al. [47] Convolutional neural network and transformer 97.40
Meng et al. [48] Baseline drift Lightweight fussing transformer 94.47 99.32
R Singh et al. [49] Maximal overlap discrete wavelet transform Bi-directional Long Short-Term Memory 95.40
Proposed method Continuous wavelet transform Multi-branch transformer and enhanced multi-headed self-attention 98.89 99.38

In recent studies on ECG signal classification tasks using transformer (Table 6), a combined approach of CNN and transformer has become the mainstream method for automatic ECG classification [[45], [46], [47], [50],49,51]. These approaches utilize the self-attention mechanism of the transformer to extract spatial information from images and convert them into sequential form for processing, thereby eliminating the need for manual feature extraction. However, these methods only employ a single transformer in their network structure, which limits the ability to fully exploit the correlation between time-series ECG signals and ignores important features such as nonlinearity and temporality.

The novel deep neural network proposed in Ref. [13], based on multi-view learning, integrates multi-scale convolutional blocks and coordinate attention modules to acquire high-quality electrocardiogram (ECG) features. It demonstrates excellent performance when dealing with long-term ECG signal records. In Ref. [14], self-collected ECG signal records were utilized for self-supervised learning to extract information from massive ECG data. Leveraging deep neural networks, this approach maintains high sensitivity and specificity in clinical testing. Reference [15] employed Graph Attention Filtering (GAF) to transform the raw ECG signals into images containing spatial-domain-related information between heartbeats, which were then combined with the original signals. Furthermore, a multi-instance learning (MIL) method was introduced to address data imbalance issues in long-term ECG signals. By utilizing multi-modal inputs, this model better allocates weights between instances and focuses more on information-rich instances.

Some studies have attempted to address this limitation by using a multi-branch transformer to implement the ECG classification task [44,48]. However, they have utilized the multi-headed self-attention mechanism (MHSA) of the original transformer for adaptive weighting of the entire image, which leads to poor computational efficiency. Since ECG signals are highly structured data in time-series feature maps, only a small fraction of the upper pixels contain useful information (Fig. 4). Pairwise attention computation between all pixels is therefore highly inefficient and redundant. In contrast, our study proposes a multi-branch transformer network with an enhanced multi-headed self-attention mechanism (Fig. 2, Fig. 3) that extracts only the relevant features from the image and focuses on subtle changes in the signal. This approach reduces computation and improves classification accuracy (Table 6), making it more suitable for realistic application scenarios. Compared to the methods mentioned above, our approach is more advantageous and efficient.

Although this study has yielded positive outcomes, it is essential to acknowledge its limitations. One of the primary constraints is the limited size of the experimental dataset, which may not be fully representative of the intricate nature of real-world scenarios. Furthermore, due to the concern for patient privacy, the scope of data collection had to be restricted, which has potentially resulted in a narrower focus of the study. Therefore, to validate the performance of the proposed algorithm, it is imperative to conduct further testing on larger and more diverse datasets. This will not only enhance the reliability of the study but also provide a more comprehensive understanding of the algorithm's efficacy in different contexts.

6. Conclusion

This study found that out of the nine commonly used methods for converting ECG signals into time-series feature maps, CWT exhibits the most robust encoding ability for signals, thereby facilitating feature extraction for classification algorithms. Moreover, the implementation of a multi-branch transformer and enhanced multi-headed self-attention mechanism leads to improved algorithm performance and generalization capabilities. Generally, this study has successfully demonstrated the automatic classification of cardiac arrhythmias, which could assist in the diagnosis and treatment of cardiovascular diseases while reducing physicians' workload.

In future work, we will further explore how to better handle the noise in the ECG signal while retaining more ECG signal characteristics. And use the data in the ECG cloud platform to continue training the model so that it can learn more arrhythmia features and improve the generalization performance of the model in the face of new long-term data.

Funding statement

Hao Li was supported by the Yunnan Province Major Science and Technology Projects [202202AE090019].

Data availability statement

Data will be made available on request.

Additional information

No additional information is available for this paper.

CRediT authorship contribution statement

Chenyang Qiu: Writing – review & editing, Writing – original draft, Visualization, Validation, Software, Resources, Project administration, Data curation. Hao Li: Supervision, Resources, Project administration. Chaoqun Qi: Software, Resources, Data curation. Bo Li: Data curation.

Declaration of competing interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

  • 1.Serhani M.A., El Kassabi H T., Ismail H., et al. ECG monitoring systems: review, architecture, processes, and key challenges. Sensors. 2020;20(6):1796. doi: 10.3390/s20061796. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Li H., Lin Z., An Z., et al. Automatic electrocardiogram detection and classification using bidirectional long short-term memory network improved by Bayesian optimization. Biomed. Signal Process Control. 2022;73 [Google Scholar]
  • 3.Denysyuk H.V., Pinto R.J., Silva P.M., et al. Algorithms for automated diagnosis of cardiovascular diseases based on ECG data: a comprehensive systematic review. Heliyon. 2023 doi: 10.1016/j.heliyon.2023.e13601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Murat F., Yildirim O., Talo M., et al. Application of deep learning techniques for heartbeats detection using ECG signals-analysis and review. Comput. Biol. Med. 2020;120 doi: 10.1016/j.compbiomed.2020.103726. [DOI] [PubMed] [Google Scholar]
  • 5.Jiang Y., Liu L., Guo X., Jiang W., et al. A comprehensive survey of deep learning based ECG classification. IEEE Access. 2021;9:65017–65033. [Google Scholar]
  • 6.Wang Y., Li H., et al. A review on electrocardiogram signal processing based on wavelet transform. Front. Physiol. 2020;11:899. [Google Scholar]
  • 7.Vaswani A., Shazeer N., Parmar N., et al. Attention is all you need. Adv. Neural Inf. Process. Syst. 2017;30 [Google Scholar]
  • 8.Acharya U.R., Fujita H., Oh S.L., et al. Application of deep convolutional neural network for automated detection of myocardial infarction using ECG signals. Inf. Sci. 2017;415:190–198. [Google Scholar]
  • 9.LeCun Y., Bottou L., Bengio Y., et al. Gradient-based learning applied to document recognition. Proc. IEEE. 1998;86(11):2278–2324. [Google Scholar]
  • 10.Hannun A.Y., Rajpurkar P., Haghpanahi M., et al. Cardiologist-level arrhythmia detection and classification in ambulatory electrocardiograms using a deep neural network. Nat. Med. 2019;25(1):65–69. doi: 10.1038/s41591-018-0268-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Graves A. Long short-term memory. Superv. Seq. Label. Recurr. Neural Networks. 2012:37–45. [Google Scholar]
  • 12.Ramkumar M., Kumar R.S., Manjunathan A., et al. Auto-encoder and bidirectional long short-term memory based automated arrhythmia classification for ECG signal. Biomed. Signal Process Control. 2022;77 [Google Scholar]
  • 13.Yang S., Lian C., Zeng Z., et al. A multi-view multi-scale neural network for multi-label ECG classification. IEEE Trans. Emerg. Top. Comput. Intell. 2023 [Google Scholar]
  • 14.Lai J., Tan H., Wang J., et al. Practical intelligent diagnostic algorithm for wearable 12-lead ECG via self-supervised learning on large-scale dataset. Nat. Commun. 2023;14(1):3741. doi: 10.1038/s41467-023-39472-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Han H., Lian C., Zeng Z., et al. Multimodal multi-instance learning for long-term ECG classification. Knowl. Base Syst. 2023;270 [Google Scholar]
  • 16.Srivastava G., Chauhan A., Kargeti N., et al. ApneaNet: a hybrid 1DCNN-LSTM architecture for detection of Obstructive Sleep Apnea using digitized ECG signals. Biomed. Signal Process Control. 2023;84 [Google Scholar]
  • 17.Liang H., Lu Y. A CNN-RNN unified framework for intrapartum cardiotocograph classification. Comput. Methods Progr. Biomed. 2023;229 doi: 10.1016/j.cmpb.2022.107300. [DOI] [PubMed] [Google Scholar]
  • 18.Barzekar H., Patel Y., Tong L., et al. MultiNet with transformers: a model for cancer diagnosis using images. arXiv preprint arXiv:2301.09007. 2023 [Google Scholar]
  • 19.Liu H., Cui S., Zhao X., et al. Detection of obstructive sleep apnea from single-channel ECG signals using a CNN-transformer architecture. Biomed. Signal Process Control. 2023;82 [Google Scholar]
  • 20.Anjum N., Sathi K.A., Hossain M.A., et al. A temporal transformer-based fusion framework for morphological arrhythmia classification. Computers. 2023;12(3):68. [Google Scholar]
  • 21.Li Y., Li K., Chen J., et al. Pilot stress detection through physiological signals using A transformer-based deep learning model. IEEE Sensor. J. 2023 [Google Scholar]
  • 22.Zhao Y., Ren J., Zhang B., et al. An explainable attention-based TCN heartbeats classification model for arrhythmia detection. Biomed. Signal Process Control. 2023;80 [Google Scholar]
  • 23.Shensa M.J. The discrete wavelet transform: wedding the a trous and Mallat algorithms. IEEE Trans. Signal Process. 1992;40(10):2464–2482. [Google Scholar]
  • 24.Chollet F. Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition. 2017. Xception: deep learning with depthwise separable convolutions; pp. 1251–1258. [Google Scholar]
  • 25.Bebis G., Georgiopoulos M. Feed-forward neural networks. IEEE Potentials. 1994;13(4):27–31. [Google Scholar]
  • 26.Liu F., Liu C., Zhao L., et al. An open access database for evaluating the algorithms of electrocardiogram rhythm and morphology abnormality detection. J. Med. Imaging Health Inform. 2018;8(7):1368–1373. [Google Scholar]
  • 27.Moody G.B., Mark R.G. The impact of the MIT-BIH arrhythmia database. IEEE Eng. Med. Biol. Mag. 2001;20(3):45–50. doi: 10.1109/51.932724. [DOI] [PubMed] [Google Scholar]
  • 28.Simonyan K., Zisserman A. Very deep convolutional networks for large-scale image recognition. arXiv preprint arXiv:1409.1556. 2014 [Google Scholar]
  • 29.He K., Zhang X., Ren S., et al. Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition. 2016. Deep residual learning for image recognition; pp. 770–778. [Google Scholar]
  • 30.Dosovitskiy A., Beyer L., Kolesnikov A., et al. An image is worth 16x16 words: transformers for image recognition at scale. arXiv preprint arXiv:2010.11929. 2020 [Google Scholar]
  • 31.Touvron H., Cord M., Douze M., et al. Training data-efficient image transformers & distillation through attention. International conference on machine learning. PMLR. 2021:10347–10357. [Google Scholar]
  • 32.Heo B., Yun S., Han D., et al. Proceedings of the IEEE/CVF International Conference on Computer Vision. 2021. Rethinking spatial dimensions of vision transformers; pp. 11936–11945. [Google Scholar]
  • 33.Liu Z., Lin Y., Cao Y., et al. Proceedings of the IEEE/CVF International Conference on Computer Vision. 2021. Swin transformer: hierarchical vision transformer using shifted windows; pp. 10012–10022. [Google Scholar]
  • 34.Kim D.H., Lee G., Kim S.H. An ECG stitching scheme for driver arrhythmia classification based on deep learning. Sensors. 2023;23(6):3257. doi: 10.3390/s23063257. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Adib E., Fernandez A., Afghah F., et al. Synthetic ECG signal generation using probabilistic diffusion models. arXiv preprint arXiv:2303.02475. 2023 [Google Scholar]
  • 36.Ahmad Z., Tabassum A., Guan L., et al. ECG heartbeat classification using multimodal fusion. IEEE Access. 2021;9:100615–100626. [Google Scholar]
  • 37.Mathunjwa B.M., Lin Y.T., Lin C.H., et al. ECG recurrence plot-based arrhythmia classification using two-dimensional deep residual CNN features. Sensors. 2022;22(4):1660. doi: 10.3390/s22041660. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Zhang Y., Chen X. Motif difference field: a simple and effective image representation of time series for classification. arXiv preprint arXiv:2001.07582. 2020 [Google Scholar]
  • 39.Hssayni E.H., Joudar N.E., Ettaouil M. A deep learning framework for time series classification using normal cloud representation and convolutional neural network optimization. Comput. Intell. 2022;38(6):2056–2074. [Google Scholar]
  • 40.Tripathi P.M., Kumar A., Kumar M., et al. Multilevel classification and detection of cardiac arrhythmias with high-resolution superlet transform and deep convolution neural network. IEEE Trans. Instrum. Meas. 2022;71:1–13. [Google Scholar]
  • 41.Kauppinen P., Hyttinen J., Laarne P., et al. A software implementation for detailed volume conductor modelling in electrophysiology using finite difference method. Comput. Methods Progr. Biomed. 1999;58(2):191–203. doi: 10.1016/s0169-2607(98)00084-4. [DOI] [PubMed] [Google Scholar]
  • 42.Vijayakumar T., Vinothkanna R., Duraipandian M. Fusion based feature extraction analysis of ECG signal interpretation–a systematic approach. J. Artif. Intell. 2021;3(1):1–16. [Google Scholar]
  • 43.Mazaheri V., Khodadadi H. Heart arrhythmia diagnosis based on the combination of morphological, frequency and nonlinear features of ECG signals and metaheuristic feature selection algorithm. Expert Syst. Appl. 2020;161 [Google Scholar]
  • 44.Li J., Li M., Zhang J., et al. IEEE; 2021. ECG classification with dual-branch transformer-based network; pp. 1693–1696. (2021 IEEE 18th International Symposium on Biomedical Imaging (ISBI)). [Google Scholar]
  • 45.Le M.D., Rathour V.S., Truong Q.S., et al. 2021 IEEE EMBS International Conference on Biomedical and Health Informatics (BHI) IEEE; 2021. Multi-module recurrent convolutional neural network with transformer encoder for ECG arrhythmia classification; pp. 1–5. [Google Scholar]
  • 46.Che C., Zhang P., Zhu M., et al. Constrained transformer network for ECG signal processing and arrhythmia classification. BMC Med. Inf. Decis. Making. 2021;21(1):1–13. doi: 10.1186/s12911-021-01546-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Wang D., Qiu L., Zhu W., et al. Inter-patient ECG characteristic wave detection based on convolutional neural network combined with transformer. Biomed. Signal Process Control. 2023;81 [Google Scholar]
  • 48.Meng L., Tan W., Ma J., et al. Enhancing dynamic ECG heartbeat classification with lightweight transformer model. Artif. Intell. Med. 2022;124 doi: 10.1016/j.artmed.2022.102236. [DOI] [PubMed] [Google Scholar]
  • 49.Singh R., Rajpal N., Mehta R. 2021. An Empiric Analysis of Wavelet-Based Feature Extraction on Deep Learning and Machine Learning Algorithms for Arrhythmia Classification. [Google Scholar]
  • 50.Hu R., Chen J., Zhou L. A transformer-based deep neural network for arrhythmia detection using continuous ECG signals. Comput. Biol. Med. 2022;144 doi: 10.1016/j.compbiomed.2022.105325. [DOI] [PubMed] [Google Scholar]
  • 51.Singh R., Rajpal N., Mehta R. Non-invasive Single Channel integration model for fetal ECG extraction and sustainable fetal healthcare using wavelet framework. Multimed. Tool. Appl. 2022:1–27. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data will be made available on request.