Effects of Oil Pulling and Chlorhexidine Mouth Rinse on the Force Decay of Orthodontic Elastomeric Chains: A Comparative In Vitro Study

Keerthana Sivaraman 1,, Rajasekaran UB 1, Neetika Prabu 1, Arun Deepak 2, Nagaland T 3, Anusha Sreedharan 1
Editors: Alexander Muacevic, John R Adler
PMCID: PMC10909396  PMID: 38435192

Abstract

Background and objectives: Oil pulling is a traditional Indian folk remedy for maintaining oral hygiene among orthodontic patients. This study aimed to assess the effects of oil pulling and compare them with those of chlorhexidine (CHX) and distilled water on the force decay of elastomeric chains.

Methods: Twenty-one samples were tested in three groups. Each of these groups contained seven samples per group. The samples evaluated contained distilled water, 0.2% CHX, and sesame oil. The samples were dipped in various mouth rinses, and force degradation was measured using a dynamometer (dynamic universal testing machine, Instron 8801, Instron, Norwood, MA) during days 0, 1, 7, 14, 21, and 28. The significance level was considered at 1%.

Results: Force degradation was observed more in distilled water, followed by sesame oil, and less in CHX at the end of 28 days. Significant differences in values (p < 0.01) were found among the three groups in all the timelines in the study except on days 14 and 21.

Conclusion: Chlorhexidine showed the least amount of force decay, followed by oil pulling and distilled water. However, if oil pulling is practiced daily as a household remedy along with regular oral hygiene practices, it can save time and money and enhance general health.

Keywords: mouth rinse, force decay, elastomeric chains, oil pulling, chlorhexidine

Introduction

Individuals with orthodontic appliances must exercise extra care when maintaining oral hygiene because appliances increase the buildup of bacteria around brackets and bands [1, 2]. Orthodontists must make extra efforts to provide preventive education to each patient. For those who are unable to maintain good oral hygiene, chemical plaque management should be employed in addition to mechanical control [3]. Chlorhexidine (CHX) is one of the most potent and extensively researched antibacterial mouth rinses. Despite being considered the ‘gold standard’, it has several adverse effects associated with its prolonged use, such as impaired taste perception and tooth discoloration. It also affects the physical and mechanical characteristics of certain orthodontic components, such as the staining of modules and the force deterioration of elastics over time [4-6].

Commercial mouthwashes can facilitate antibiotic resistance. This notion has motivated the search for natural products to preserve dental health. Dr. Karach F. popularized the practice of oil pulling with edible oils in contemporary medicine [7]. Oil pulling is the traditional Ayurvedic practice of gargling with oil to prevent decay, foul breath, bleeding gums, and cracked lips, in addition to strengthening teeth, gums, and jaws. It is known to have therapeutic local as well as systemic effects and is considered an effective alternative to CHX for routine oral hygiene practices [6, 8]. The procedure of oil pulling involves swishing a measured volume of oil around the mouth for a period, forcing the oil in between all the teeth and around the mouth. Examples of organic oils that are used include sunflower oil, sesame oil, and coconut oil [9]. Sesame oil has the following advantages over CHX: no staining, no lingering aftertaste, and no allergy. Sesame oil is readily available in most homes and is five to six times more affordable than CHX [10].

The British Society of Periodontology states that "antiplaque agents like CHX are useful for managing acute periods when cleaning is difficult but not needed as a routine" [11]. Additionally, it must be noted that the use of CHX mouthwash is licensed only for 30 days of use [12]. Therefore, for patients receiving fixed orthodontic therapy, a safe, cost-effective, and frequently used substitute for CHX mouth rinse is needed. Though oil pulling is an obsolete procedure, it should be reinstated as a regular dental hygiene measure.

Elastomeric chains have been widely used in orthodontics since the 1960s because they do not require patient cooperation and are relatively hygienic, affordable, and simple to use [13, 14]. Elastic devices are important sources for the transmission of force to teeth but are not considered ideal because the force diminishes with activation time, oral media, and other dietary-related characteristics [15].

However, there is no scientific research in support of oil pulling that compares the effects of force degradation on elastomeric chains. Hence, the present study was designed to assess the effects of oil pulling and compare them with those of CHX and distilled water for one month.

Materials and methods

A laboratory study was conducted at the PSG Centre for Research and Consultancy, Coimbatore, India, for 28 days to test the force degradation of elastomeric chains using a dynamometer (dynamic universal testing machine, Instron 8801, Instron, Norwood, MA) (Figure 1).

Figure 1. Dynamometer (dynamic universal testing machine, Instron 8801, Instron, Norwood, MA).

Figure 1

The following armamentarium was used: distilled water, 0.2% CHX mouth rinse (Hexidrin brand), sesame oil (Idhayam brand), artificial saliva (Xerostat brand), and orthodontic elastomeric chains (short, American Orthodontics brand). This study did not involve the use of any animals, human data, or tissues; therefore, consent and ethical approval were not required.

Twenty-one samples were tested in three groups. There were seven samples per group. The samples evaluated were distilled water (Group 1: control), 0.2% CHX (Group 2: experimental group), and sesame oil (Group 3: experimental group) (Figure 2).

Figure 2. Control (distilled water) and experimental groups (chlorhexidine and sesame oil).

Figure 2

A customized wooden template was fabricated with seven pairs of supporting rods arranged row-wise, with 23.5 mm gap between each set of rods. Short elastomeric chains of five links each were pre-stretched 1.5 times the original length and fixed to the supporting rod (Figure 3).

Figure 3. Orthodontic elastomeric chains in a fabricated template.

Figure 3

These were immersed in the artificial saliva solution at a controlled temperature (Figure 4).

Figure 4. Samples immersed in artificial saliva.

Figure 4

The wooden template was dipped in the corresponding experimental and control solutions for one minute daily. These were then dipped in separate water baths for 10 seconds and placed back in the artificial saliva. The level of saliva in the template was verified every day to ensure the elastics were covered by this solution at all times. Six force measurements were taken during the experimental period at the following time intervals: days 0, 1, 7, 14, 21, and 28.

Statistical analysis

Data were entered in a Microsoft Excel sheet (Microsoft Corp., Redmond, WA), and statistical analysis was conducted using IBM SPSS software version 20.0 (IBM Corp., Armonk, NY). Mean and standard deviation were used to summarize the data. The effects of the groups (distilled water, CHX 0.2%, and sesame oil) and time (days 0, 1, 7, 14, 21, and 28) on force decay were analyzed using ANOVA and Tukey’s post hoc test. The significance level was considered at 1% (< 0.01).

Results

Factors influencing the force decay of elastomeric chains were time duration, type of elastomeric chain, and exposure to test solutions. The maximum force (0.572 kg) was observed in CHX, followed by sesame oil (0.460 kg), and the least force (0.435 kg) was observed in distilled water at the end of 28 days (Table 1).

Table 1. Force (Kg) measured for elastomeric chains with different mouth rinses at different time periods.

Time interval Distilled water mean (SD) Chlorhexidine mean (SD) Sesame oil mean (SD)
Day 0 0.728 (0.10) 0.827 (0.02) 0.715 (0.03)
Day 1 0.524 (0.07) 0.647 (0.00) 0.533 (0.00)
Day 7 0.490 (0.05) 0.610 (0.01) 0.518 (0.05)
Day 14 0.471 (0.12) 0.587 (0.05) 0.487 (0.09)
Day 21 0.455 (0.05) 0.586 (0.05) 0.482 (0.00)
Day 28 0.435 (0.05) 0.572 (0.06) 0.460 (0.01)

When the groups were compared with one another in the same period, no statistical difference was found on days 14 and 21 (p > 0.01). Statistically significant differences in force levels were noted during the initial hours, during 24 hours, on day seven, and after day 28 (p < 0.01) (Table 2).

Table 2. One-way ANOVA test among the groups at different time intervals.

*Distilled water (Group 1: control), chlorhexidine (Group 2: experimental group), and sesame oil (Group 3: experimental group). # p < 0.01: significant

Timeline Groups* Mean SD 95% confidence interval for mean p-value#
Lower bound Upper bound
Initial 1 .70100 .021517 .64755 .75445 0.001
2 .80500 .018028 .76022 .84978
3 .70433 .007506 .68569 .72298
24 hours 1 .51567 .022279 .46032 .57101 0.002
2 .63100 .027074 .56374 .69826
3 .51367 .023671 .45486 .57247
7 days 1 .43500 .054083 .30065 .56935 0.012
2 .59333 .037859 .49929 .68738
3 .53767 .036828 .44618 .62915
14 days 1 .45167 .045369 .33896 .56437 0.065
2 .56000 .052000 .43082 .68918
3 .45867 .051598 .33049 .58684
21 days 1 .43500 .030414 .35945 .51055 0.028
2 .57300 .066461 .40790 .73810
3 .45233 .045347 .33969 .56498
28 days 1 .42167 .019088 .37425 .46908 0.001
2 .57267 .006429 .55670 .58864
3 .46000 .010000 .43516 .48484

A sudden decline in the force levels during the initial 24 hours, followed by a gradual reduction of force over a period, was noted (Figure 5).

Figure 5. Force levels of various mouth rinses at each time interval.

Figure 5

Multiple group comparison was done using Tukey’s post hoc test. Statistically significant differences (p < 0.01) were noted in force levels when comparing Group 1 with Group 2, Group 2 with Groups 1 and 3, and Group 3 with Group 2 during the initial hours; when comparing Group 1 with Group 2, Group 2 with Groups 1 and 3, and Group 3 with Group 2 during 24 hours; when comparing Group 1 with Group 2 and Group 2 with Group 1 on the seventh day; and Group 1 with Group 2, Group 2 with Groups 1 and 3, and Group 3 with Group 2 on the 28th day. No significant differences (p > 0.01) were noted in force levels when multiple groups were compared on days 14 and 21; when Group 1 was compared with Group 3 and Group 3 was compared with Group 1 during the initial hours, during 24 hours, and on the day 28; and when Group 1 was compared with Group 3, Group 2 with Group 3, and Group 3 with Groups 1 and 2 on the day seven (Table 3).

Table 3. Tukey’s post hoc test; multiple group comparisons.

*p < 0.01: significant

Dependent variable (I) Group (J) Group Mean difference (I-J) p-value*  
Initial 1 2 -.104000* .001  
3 -.003333 1.000  
2 1 .104000* .001  
3 .100667* .001  
3 1 .003333 1.000  
2 -.100667* .001  
24 hours 1 2 -.115333* .004  
3 .002000 1.000  
2 1 .115333* .004  
3 .117333* .003  
3 1 -.002000 1.000  
2 -.117333* .003  
7 days 1 2 -.158333* .013  
3 -.102667 .084  
2 1 .158333* .013  
3 .055667 .508  
3 1 .102667 .084  
2 -.055667 .508  
14 days 1 2 -.108333 .112  
3 -.007000 1.000  
2 1 .108333 .112  
3 .101333 .141  
3 1 .007000 1.000  
2 -.101333 .141  
21 days 1 2 -.138000* .043  
3 -.017333 1.000  
2 1 .138000* .043  
3 .120667 .074  
3 1 .017333 1.000  
2 -.120667 .074  
28 days 1 2 -.151000* .000  
3 -.038333* .033  
2 1 .151000* .000  
3 .112667* .000  
3 1 .038333* .033  
2 -.112667* .000  

Discussion

Orthodontic appliances attached to tooth surfaces make it difficult to practice oral hygiene and act as additional bacterial plaque reservoirs. The enamel is demineralized as a result, leading to white spots, dental cavities, and gingivitis [1]. To prevent enamel demineralization, various mouth rinses are being prescribed. Out of these, chlorhexidine is a highly effective synthetic antibacterial agent [3]. Oil pulling can be used as an alternative method to maintain oral hygiene in orthodontic patients [9].

In the present study, elastomeric chain usage was considered to be 28 days because this matched the average amount of time between orthodontic consultations, as Motta et al. [16] and Pithon et al. [17] noted. The elastomeric chain segment was kept submerged in artificial saliva because force degradation in a humid medium is substantially greater than in a dry environment [18-20]. Short elastomeric chains were used to maintain a higher percentage of force over time [21].

The latex elastics in the present investigation displayed a force relaxation in the range of 22%-27% in the first 24 hours. Force levels at the end of 24 hours were 0.64 kg, 0.53 kg, and 0.52 kg in chlorhexidine (Group 2), sesame oil (Group 3), and distilled water (Group 1), respectively (Table 1). This result was similar to those of Pithon et al. [17], with a 20%-30% force decay. In contrast, Singh et al. [21] and Sam et al. [22] showed that the rates of force decay were 50% to 70% and 17% to 24%, respectively, for the first 24 hours. There was a large decrease in force during the first 24 hours, followed by mostly stable levels of force up to four weeks (Figure 5). This result is in line with those of Pithon et al. [17], Sufarnap et al. [23], Issa et al. [24], Samuels et al. [25], and Balhoff et al. [26].

However, results may vary depending on whether other biological components are present in the oral environment. Because this study was conducted in vitro under static conditions, the elastomeric chain’s performance could not replicate the degradation observed in vivo. Further studies in clinical settings, where the oral environment is varied because of dietary habits, microbial activity, different stretching conditions, and different brands of elastomeric chains, are necessary. This would help us better comprehend elastic materials’ physical characteristics in various clinical settings. Studies using different organic oils and herbal mouth rinses should be carried out in the future so that they can be used as an alternative to chlorhexidine in assessing the force decay of orthodontic elastomeric chains.

Conclusions

Chlorhexidine showed the least amount of force decay compared to oil and distilled water. Similarly, oil pulling had a significantly lower force decay on elastomeric chains compared to distilled water. It is thus concluded that, though CHX showed the least amount of force decay, it had certain adverse effects on prolonged usage. Hence, oil pulling can be recommended as a preventive oral hygiene practice with minimal adverse effects in fixed orthodontic patients because the treatment takes place over a longer duration. If oil pulling is practiced daily as a household remedy along with regular oral hygiene practices, it can save time and money and enhance general health.

The authors have declared that no competing interests exist.

Author Contributions

Concept and design:  Keerthana Sivaraman, Rajasekaran UB, Neetika Prabu, Arun Deepak, Anusha Sreedharan

Acquisition, analysis, or interpretation of data:  Keerthana Sivaraman, Nagaland T

Drafting of the manuscript:  Keerthana Sivaraman, Neetika Prabu, Arun Deepak, Nagaland T, Anusha Sreedharan

Critical review of the manuscript for important intellectual content:  Rajasekaran UB, Nagaland T, Anusha Sreedharan

Supervision:  Rajasekaran UB, Neetika Prabu, Arun Deepak, Anusha Sreedharan

Human Ethics

Consent was obtained or waived by all participants in this study

Animal Ethics

Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.

References

  • 1.Influences of bracket bonding on mutans streptococcus in plaque detected by real time fluorescence-quantitative polymerase chain reaction. Ai H, Lu HF, Liang HY, Wu J, Li RL, Liu GP, Xi Y. https://pubmed.ncbi.nlm.nih.gov/16336838/ Chin Med J (Engl) 2005;118:2005–2010. [PubMed] [Google Scholar]
  • 2.Long-term fluoride release from resin-reinforced orthodontic cements following recharge with fluoride solution. dos Santos RL, Pithon MM, Vaitsman DS, Araújo MT, de Souza MM, Nojima MG. Braz Dent J. 2010;21:98–103. doi: 10.1590/s0103-64402010000200002. [DOI] [PubMed] [Google Scholar]
  • 3.Effectiveness of 0.50% and 0.75% chlorhexidine dentifrices in orthodontic patients: a double-blind and randomized controlled trial. Oltramari-Navarro PV, Titarelli JM, Marsicano JA, Henriques JF, Janson G, Lauris JR, Buzalaf MA. Am J Orthod Dentofacial Orthop. 2009;136:651–656. doi: 10.1016/j.ajodo.2008.01.017. [DOI] [PubMed] [Google Scholar]
  • 4.A two-year randomized clinical trial of chlorhexidine varnish on dental caries in Chinese preschool children. Du MQ, Tai BJ, Jiang H, Lo EC, Fan MW, Bian Z. J Dent Res. 2006;85:557–559. doi: 10.1177/154405910608500615. [DOI] [PubMed] [Google Scholar]
  • 5.Sodium fluoride and chlorhexidine effect in the inhibition of mutans streptococci in children with dental caries: a randomized, double-blind clinical trial. Lobo PL, de Carvalho CB, Fonseca SG, de Castro RS, Monteiro AJ, Fonteles MC, Fonteles CS. Oral Microbiol Immunol. 2008;23:486–491. doi: 10.1111/j.1399-302X.2008.00458.x. [DOI] [PubMed] [Google Scholar]
  • 6.Effect of different 1% chlorhexidine varnish regimens on mutans streptococci levels in saliva and dental biofilm. Ribeiro LG, Hashizume LN, Maltz M. https://pubmed.ncbi.nlm.nih.gov/19024254/ Am J Dent. 2008;21:295–299. [PubMed] [Google Scholar]
  • 7.Effect of coconut oil in plaque related gingivitis - a preliminary report. Peedikayil FC, Sreenivasan P, Narayanan A. Niger Med J. 2015;56:143–147. doi: 10.4103/0300-1652.153406. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Oil pulling as an adjunct to improve oral health in orthodontic patients: a clinicomicrobial study. Kolhe SA, Patani S, Gulve N, Pawar R, Dhope SV, Gajeshwar H. Int J Orthod Rehabil. 2019;10:152–155. [Google Scholar]
  • 9.Oil pulling for maintaining oral hygiene - a review. Shanbhag VK. J Tradit Complement Med. 2017;7:106–109. doi: 10.1016/j.jtcme.2016.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Effect of oil pulling on plaque induced gingivitis: a randomized, controlled, triple-blind study. Asokan S, Emmadi P, Chamundeswari R. Indian J Dent Res. 2009;20:47–51. doi: 10.4103/0970-9290.49067. [DOI] [PubMed] [Google Scholar]
  • 11.British Society of Periodontology. The Good Practitioner’s Guide to Periodontology. Liverpool, GBR: British Society of Periodontology; 2016. The good practitioner’s guide to periodontology; pp. 10–14. [Google Scholar]
  • 12.Current uses of chlorhexidine for management of oral disease: a narrative review. Brookes ZL, Bescos R, Belfield LA, Ali K, Roberts A. J Dent. 2020;103:103497. doi: 10.1016/j.jdent.2020.103497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.In vivo comparison of force decay between injection molded and die-cut stamped elastomers. Bousquet JA Jr, Tuesta O, Flores-Mir C. Am J Orthod Dentofacial Orthop. 2006;129:384–389. doi: 10.1016/j.ajodo.2005.09.002. [DOI] [PubMed] [Google Scholar]
  • 14.Elastomeric chain force decay in artificial saliva: an in vitro study. Halimi A, Azeroual MF, Doukkali A, El Mabrouk K, Zaoui F. Int Orthod. 2013;11:60–70. doi: 10.1016/j.ortho.2012.12.007. [DOI] [PubMed] [Google Scholar]
  • 15.Study of the degradation of force generated by synthetic orthodontic elastics [Article in Portugese] Araujo FBC, Ursi WJS. Rev Dent Press Ortodon Ortop Facial. 2006;11:52–61. [Google Scholar]
  • 16.In vitro evaluation of force delivered by elastomeric chains. da Motta AFJ, Cury-Saramago AD, Nojima LI. https://www.scielo.br/j/dpjo/a/dKQqxdJv9fxWvSdtn4FxY9r/?format=pdf&lang=en Dental Press J Orthod. 2011;16:0–8. [Google Scholar]
  • 17.Does chlorhexidine in different formulations interfere with the force of orthodontic elastics? Pithon MM, Santana DA, Sousa KH, Farias IM. Angle Orthod. 2013;83:313–318. doi: 10.2319/061312-493.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Calibration of force extension and force degradation characteristics of orthodontic latex elastics. Kanchana P, Godfrey K. Am J Orthod Dentofacial Orthop. 2000;118:280–287. doi: 10.1067/mod.2000.104493. [DOI] [PubMed] [Google Scholar]
  • 19.Relaxation of orthodontic elastomeric chains and modules in vitro and in vivo. Ash JL, Nikolai RJ. J Dent Res. 1978;57:685–690. doi: 10.1177/00220345780570050301. [DOI] [PubMed] [Google Scholar]
  • 20.Mechanical and biological comparison of latex and silicone rubber bands. Hwang CJ, Cha JY. Am J Orthod Dentofacial Orthop. 2003;124:379–386. doi: 10.1016/s0889-5406(03)00564-x. [DOI] [PubMed] [Google Scholar]
  • 21.Elastics in orthodontics: a review. Singh VP, Pokharel PR, Pariekh K, et al. Health Renaissance. 2012;10:49–56. [Google Scholar]
  • 22.Effect of chlorhexidine mouth rinse in force decay of closed, short, and long elastomeric chain - an in vitro study. Sam RJ, Mishra V, Yadav A, Yadav D, Joshi D, Martina K. Indian J Orthod Dentofacial Res. 2022;8(4):237–244. [Google Scholar]
  • 23.Effect of sodium fluoride in chlorhexidine mouthwashes on force decay and permanent deformation of orthodontic elastomeric chain. Sufarnap E, Harahap KI, Terry T. PJoD. 2021;33:74–80. [Google Scholar]
  • 24.The effects of zinc-containing mouthwashes on the force degradation of orthodontic elastomeric chains: an in vitro study. Issa AR, Kadhum AS, Mohammed SA. Int J Dent. 2022;2022:3557317. doi: 10.1155/2022/3557317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.A comparison of the rate of space closure using a nickel-titanium spring and an elastic module: a clinical study. Samuels RHA, Orth M, Rudge SJ, Mair LH. https://www.ajodo.org/article/S0889-5406(05)81798-6/fulltext. Am J Orthod Dentofacial Orthop. 1993;103:464–467. doi: 10.1016/S0889-5406(05)81798-6. [DOI] [PubMed] [Google Scholar]
  • 26.Force decay of elastomeric chains - a mechanical design and product comparison study. Balhoff DA, Shuldberg M, Hagan JL, Ballard RW, Armbruster PC. J Orthod. 2011;38:40–47. doi: 10.1179/14653121141227. [DOI] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.