Abstract
Rationale
The measurement of the stable carbon and oxygen isotope ratio of (atmospheric) carbon dioxide (CO2) is a useful technique for the investigation and identification of the sources and sinks of the most abundant greenhouse gases by far. For this reason, we are presenting a measuring system here that enables a wide range of users to carry out stable isotope analysis of atmospheric CO2 using off‐the‐bench hardware and software.
Methods
The fully automated system uses cryogenic and gas chromatographic separation to analyse CO2 from 12‐mL whole air samples and consists of an autosampler, a Gasbench II (GB), a downstream cryo trap and a continuous flow gas interface feeding into a sector field mass spectrometer (GC Pal/GB/Cold Trap/ConFlo IV/DeltaV Plus). The evaluation of the system performance was based on the analysis of samples prepared from eight CO2 sources (four CO2 reference gases and four artificial air tanks).
Results
The overall measurement uncertainty (averaged single standard deviation (1σ) of measurement replicates from each CO2 source) in the determination of the carbon and oxygen isotope ratio was 0.04‰ and 0.09‰ (n = 24). Furthermore, we were able to show that the measurement data also allowed for the quantification of the CO2 mole fraction, with a precision of 1.2 μmol mol−1 in the analysis range of 400–500 μmol mol−1.
Conclusions
Our protocol provides a detailed description of the measurement set‐up and the analysis procedure, how raw data should be evaluated and gives recommendations for sample preparation and sampling to enable a fully automated whole air sample analysis. The quantification limit of CO2 mole fractions and measurement precision for carbon and oxygen isotope ratios of CO2 should meet the requirements of a wide range of users.
1. INTRODUCTION
The industrial era led to a change in the proportion of greenhouse gases in the atmosphere, with carbon dioxide (CO2) currently accounting for about 66% of the global warming potential (2021 global mean abundance of CO2: 415.7 ± 0.2 μmol mol−1). 1 To take mitigation measures, it is necessary to know not only the emission levels but also their sources and sinks. 2 The stable isotope ratios of carbon and oxygen in CO2 can be linked to the processes of its sources and sinks (e.g., burning of fossil fuels, photosynthesis, respiration, exchange reactions). The latter, (13C/12C, 18O/16O), can be determined using stable isotope ratio mass spectrometry (IRMS) or laser‐absorption spectroscopy (LAS). 3 As far as IRMS techniques are concerned, most published systems rely on hand‐made peripherals, high instrumental know‐how or do not allow fully automated sample analysis. See the current review of Manaj and Kim (2020). 4 Although these systems achieve high measurement accuracy and a high sample throughput, they have been individually adapted or optimised, which can be implemented only by experienced technical personnel and using specific components.
Here, we present a protocol on how to set up and operate an off‐the‐shelf continuous flow (CF)‐IRMS measurement system to analyse the carbon and oxygen stable isotope ratio of CO2 from whole air samples at ambient atmospheric mixing ratios. The aim of the presented measurement set‐up was to provide a fully automated system, including sample vessel preparation, sampling and sample analysis, which provides reasonable measurement performance and is available to a broad user community. The targeted measurement precision was <0.1‰ and <0.2‰ for the carbon and oxygen isotope ratio of CO2, respectively. Furthermore, it was tested whether the measurement set‐up would also allow for the estimation of the CO2 mole fraction in whole air samples.
The CF measurement set‐up presented here is based on the extraction of CO2 from 12 mL sample vessels using cryo‐focusing and gas chromatography. Therefore, a Gasbench II (GB; Thermo Fisher Scientific, Bremen, Germany) equipped with a 250‐μL sample loop and a subsequent Cold Trap was connected to a CF gas distribution system (ConFlo IV, Thermo Fisher Scientific, Bremen, Germany) managing the gas inlet into a sector field isotope ratio mass spectrometer (Delta V Plus, Thermo Fisher Scientific, Bremen, Germany). The measurement uncertainty was evaluated by the analysis of four distinct CO2 reference gases with assigned true isotope ratios mixed with synthetic air for the analysis. The measurement set‐up was then further used to calibrate the isotope ratios of CO2 in four artificial atmospheric air tanks to be used as working gases in atmospheric air monitoring campaigns. Furthermore, the impact of cryo‐focusing, chromatographic separation, water background levels and the open‐split sample dilution setting (‘blanking’) were evaluated and are discussed below.
2. MATERIALS AND METHODS
2.1. CF‐IRMS set‐up
The measurement set‐up consisted of a GC PAL Autosampler (CTC Analytics AG, Zwingen, Switzerland), a GB, a Dual Cold Trap (Thermo Fisher Scientific, Bremen, Germany), using the fused silica capillary trap only, a ConFlo IV and a Delta V Plus. The ensemble is shown in Figure 1. The autosampler was equipped with a heating unit (25°C) and a sample tray for 96 sample vials containing 12 mL Exetainer glass vials (Labco Limited, Lampeter, UK) and was only used to manoeuvre the double needle, which directs the sample gas from the sample vial to the GB at ~0.8 mL min−1. Alternatively, the set‐up could also be used with 20 mL headspace vials with crimped polytetrafluoroethylene (PTFE)‐coated butyl rubber septa, as presented in Leitner et al (2020). 5 , 6 At the GB (helium [He] inlet pressure of 80 kPa), the sample gas was first dried by passing through a Nafion™ membrane (purged continuously with a counter current stream of He), was then sent through a 250 μL sample loop and eventually vented from the system. At an initial transfer time of 160 s, the sample loop was further flushed for 60 s before switching the eight‐port valve from the ‘Load’ position to the ‘Inject’ position to transfer the sample loop volume to the Cold Trap, which had been immersed in a 3‐L liquid nitrogen (LN2) dewar already 10 s ahead. The sample loop was flushed for 50 s to ensure complete transfer of sample gas to the Cold Trap before switching the eight‐port valve back to the ‘Load’ position. The Cold Trap was released 10 s thereafter to transfer the cryo‐focused gases (CO2, N2O, water vapour) and residual air to a 25 m × 0.32 mm PoraPLOT Q GC‐column (Agilent Technologies GmbH, Vienna, Austria) to separate CO2 from the other air constituents. The gas chromatography (GC) outlet was linked to a second Nafion™ trap before entering the sample open split via the low‐flow capillary inside the gas distribution system (ConFlo IV), which then transferred the sample gas to the IRMS (Delta V Plus), as shown in Figure 1. To further dilute residual N2 and O2, the mass spectrometer (MS) capillary inside the ConFlo IV sample open split was connected only shortly before and during the retention times of CO2 from the 250‐μL gas samples. This procedure of loop filling, trapping and CO2 peak detection was repeated 10 times during the total analysis time of 30.2 min per sample vial. Detailed event timings are provided in Table S1 (supporting information). The monitoring of the CO2 sample gas peaks was performed together with CO2 working gas peaks (rectangular peaks, fixed mass 44 intensity of 5 V) added before and after the set of sample gas peaks (Figure 2). As the IRMS is also used to analyse samples by GC, a manual four‐port valve was placed in front of the low‐flow inlet of the ConFlo IV to switch between GB and GC.
FIGURE 1.

Instrumental set‐up for the analysis of 12‐mL whole air samples using a GC PAL autosampler and a Gasbench II (GB) connected to a Cold Trap, a ConFlo IV and a Delta V Plus.
FIGURE 2.

Chromatogram for the analysis of CO2 from a 12‐mL whole air sample (B1 air tank; CO2: 404 μmol mol−1) and the recording of the mass‐44 signal intensities. Rectangular peaks represent CO2 working gas peaks, and Gaussian‐shaped peaks represent CO2 sample gas obtained from 250‐μL sample gas aliquots.
The impact of temperature and pressure effects on the IRMS magnet performance, as pointed out by Ferretti et al (2000), 7 was taken care of by the use of an air conditioning system inside the lab maintaining the lab air temperature at 22°C.
2.2. Sample vial and air tank preparation
Gas was sampled in 12‐mL Exetainer glass vials (Labco Limited, Lampeter, UK) sealed with grey chlorobutyl septa and screw caps. Vials were prepared by flushing with synthetic air (SA; synthetic air 5.0 HC‐free, Messer Austria GmbH, Gumpoldskirchen, Austria), which refers to zero‐air, for 60 s at a flow of 250 mL min−1. Different settings of the inlet pressure, flush time and flush gas flow were tested, and we recommend exchanging the sample vial volume for at least 20 times to obtain CO2‐free vials. Flushing was performed using two G26 side‐bore needles piercing the septum of the Exetainer vial with the cap closed. For the preparation of isotope reference gas standards, 5 μL of four pure CO2 reference gases (R1, R2, R3, R4; ISO‐TOP, Messer Austria GmbH) and the CO2 working gas were added on top of the SA‐flushed vial using a 10‐μL gas‐tight syringe equipped with a G26 side‐bore needle (Hamilton Bonaduz AG, Bonaduz, Switzerland).
Whole air samples were mimed by using four 50‐L air tanks (7–10 MPa, Messer Austria GmbH), labelled B1, B2, B3 and B4, which had been prepared by adding pure CO2 to 50‐L SA tanks. The SA used to flush the Exetainer vials and that used to manufacture the air tanks was of the same quality and composition and consisted of a 79.5:20.5 mixture of N2 and O2 and the minor components CO2 (≤0.5 μmol mol−1), NOx (≤0.1 μmol mol−1) and H2O (≤5 μmol mol−1). The final isotopic composition of B1, B2, B3 and B4 was varied by the admixed amount of CO2 added from another two pure CO2 cylinders containing 13C‐depleted or 13C‐enriched CO2. The four air tanks obtained four distinct isotopic compositions and mole fractions of CO2, similar to the levels expected for atmospheric air (410–520 μmol mol−1, δ13C: −8 to −18‰). 8 , 9 , 10 , 11 , 12 , 13 , 14 , 15 , 16 The mole fraction of CO2 was 404, 404, 513 and 406 μmol mol−1 ± 2% absolute deviation, according to the Messer company, for B1, B2, B3 and B4, respectively. Whole air samples were prepared by flushing the Exetainer vials with tank air similar to the setting as that for SA‐flushing, but after the vials had been flushed with SA initially. For whole air sampling of atmospheric air, we refer to the sampling procedure presented in the recent publication of Leitner et al (2020 and 2023). 5 , 6
2.3. Referencing of carbon and oxygen isotope ratios of CO2
The stable carbon and oxygen isotope ratio of CO2 is reported in the δ‐notation (‰) and was referenced to the Vienna Peedee Belemnite (VPDB) scale for δ13C and δ18O values, anchored by isotope reference materials and internally cross‐checked (data not shown) by the analysis of CO2 liberated by H3PO4 acid digestion of NBS18 (δ13C: −5.014‰ ± 0.035 vs VPDB, δ18O: −23.2‰ ± 0.1 vs VPDB) 17 to be sure about the reference scale given for the isotope reference gases described below. In Brand et al, 18 there is a recommendation that carbonate reference materials, specifically NBS19, should not or not exclusively be used as δ18O reference anchors in the analysis of CO2 gases if these gases were obtained without acid digestion (e.g., CO2 in air). δ‐values were calculated as
where R is the ratio of the abundance of 13C to 12C and 18O to 16O of a sample (P) and a measurement standard (Std). 19
Four certified CO2 reference gases (R1, R2, R3, R4; ISO‐TOP, Messer Austria GmbH) with an assigned true δ13C value versus VPDB of −6.65 ± 0.13‰, −6.73 ± 0.13‰, −25.71 ± 0.20‰ and −39.00 ± 0.21‰ and assigned true δ18O value versus VPDB of −17.82 ± 0.20‰, −17.97 ± 0.20‰, −28.72 ± 0.15‰ and −30.34 ± 0.20‰ were used to determine the isotopic composition of the CO2 in four 50‐L air tanks, labelled B1, B2, B3, B4. The four 50‐L air tanks were prepared by admixing CO2 from two pure CO2 cylinders filled with a 13C‐depleted and a 13C‐enriched CO2 gas, not related to the four certified CO2 reference gases. The CO2 working gas (δ13C: −4.29‰ ± 0.13‰; δ18O: −12.27‰ ± 0.17‰; n = 7) of the IRMS was also referenced with the use of the certified CO2 reference gases, but via direct injection to a GC/IRMS set‐up as presented at Leitner et al. 6 Monitored working gas peaks, added before and after the sample gas peaks, indicated a single standard deviation (1σ) of δ13C‐ and δ18O‐values over the course of the presented measurements (n = 255) of 0.08‰ and 0.06‰, respectively.
2.4. Evaluation of CO2 measurement data
The measurement data presented include the analysis of three sample vials per air tank and per reference gas, the latter being added to vials filled with synthetic air. The sample vials were analysed in a single sequence run. Analysis of each sample vial comprised 10 CO2 gas peaks obtained by continuous filling and subsequent transfer of a 250‐μL sample loop volume. As a result of the dilution of the sample gas with the carrier gas (He), there was a decrease in the mass intensities (~60% between the first and tenth CO2 peaks). Therefore, raw data were first corrected for the δ 13C and δ18O non‐linearity effect because of the dependence of mass 44, 45 and 46 signal intensities. Non‐linearity corrected peaks of each sample vial were checked for outliers using a Grubbs Test 20 to then decide on the most optimal number of peaks to be used to calculate a sample’s mean value of δ 13C and δ18O. Then, δ‐values were normalised with the use of those samples, which had been prepared with the four reference gases, using linear regression and according to Paul et al. 21 Analysed samples comprised three sample replicates of each CO2 source (R1, R2, R3, R4, B1, B2, B3 and B4), analysed in a single sequence run. Normalised data were grouped by CO2 source to obtain a final calibrated mean value ± 1σ (measurement uncertainty) and combined uncertainty (u) for the air tanks (B1, B2, B3, B4). The initial raw data were obtained by the software Isodat (version 3.0) from Thermo Fisher Scientific, which is also used to operate the measurement system presented. For information on peak detection parameters and δ‐value calculation, see the Supporting Information.
3. RESULTS AND DISCUSSION
3.1. Chromatographic resolution
An exemplary chromatogram for the analysis of CO2 in 12‐mL whole air samples using the presented method is shown in Figure 2. It shows the intensity of mass 44 while analysing a sample vial containing 404 μmol mol−1 CO2 and rectangular working gas peaks before and after the 10 CO2 sample gas peaks. Sample gas peaks continuously decrease in time due to the mixing with the He carrier gas inside the sample vial. Each sample peak represents the CO2 mole fraction in 250‐μL sample gas and, at a CO2 mole fraction of 400 μmol mol−1, was initially at ~8.0 V and decreased to ~3.6 V within 10 sample gas peaks.
Mass intensities obtained from a blank‐sample measurement, that is, a sample vial purged with SA without the addition of pure CO2, are shown in Figure 3. The intensity of mass 44 ranged from 30 to 15 mV from the first to the tenth sample gas peaks and was considered negligible for evaluation (shown for peak 1:4 in plot A of Figure 3). Isobaric interference from residual N2 and O2 in the ion source cavity could not be identified from mass‐46 and mass‐28 scans, respectively, as shown in plot A (dotted line) and plot C (solid line) of Figure 3. The water background, which affects the mass 45 and hence the carbon isotope ratio, is shown in plot B of Figure 3 and was tested by analysing laboratory room air‐filled measurement vials. Despite the initially increasing mass‐18 intensity (1.8 to 3.5 V for the first to third peaks (scanned on Faraday collector cup 3 [3 1010 Ohm]) and subsequent constant intensity, the CO2 retention times were always found at constant mass 18 background intensities. Therefore, they were not affected by the evaporation of the water previously frozen out in LN2, as this only occurred ~20 s after each CO2 peak.
FIGURE 3.

Results from the measurement of a blank vial (filled with synthetic air [SA] only) shown in (A) and (C) and of a vial filled with lab air in (B). Vertical dot‐dashed lines indicate the position of CO2 sample gas peaks (retention time ± ½ the width). Mass scans comprised mass 44, 45 and 46 indicated by solid, dashed and dotted lines (A), mass 18 (B) and mass 28 (C).
3.2. Raw data processing
3.2.1. Non‐linearity correction and identification of outliers
Each sample analysis provided 10 carbon and oxygen isotope ratio values that should be corrected for the non‐linearity effect of the IRMS detector due to the decrease in mass intensities. The manufacturer recommends correcting the δ‐values using linear regression (first‐order polynomial function) of δ‐value and mass‐44 intensity, which must be generated from separate IRMS linearity measurements and should be performed before each measurement sequence. For δ13C, the slope from the linear regression was found to be 0.06‰ mV−1 (0.02‰ nA−1) with a coefficient of determination (R 2 ) of 0.91. It was found in our laboratory that regression using a second‐order polynomial function better described (R 2 : >0.99) the relation between δ13C and mass‐44 intensity (Table S2 and Figure S1 [supporting information]). Similarly, the non‐linearity effect in the oxygen isotope ratio did not indicate a linear dependency on the mass intensity and showed a fluctuation in δ18O‐values of ±0.02‰ at most and was therefore not taken into account (Figure S1). The circumstance of the observed non‐linearity effects could have been due to the condition of the ion source components or the low residual water content in the He gas and its influence on the ionisation and should therefore be treated individually. As a point of reference, it should be noted that the intensity of mass 18, scanned at the Faraday cup 3, was often less than 500 mV before the start of a measurement sequence, and this represents very ‘dry’ He.
A comparison of the δ13C single standard deviation per sample after non‐linearity correction, using first‐ and second‐order polynomial functions, as well as in comparison to raw δ13C‐values is shown in Figure S2 (supporting information) and illustrates that the use of a second‐order polynomial function was most effective to reduce the standard deviation of δ13C sample means.
After non‐linearity correction of the sample peaks, an outlier test of the δ13C and δ18O peak values of each sample was performed. Two outliers were found for carbon and three for oxygen within the data set of 240 values each (24 samples, 10 peaks each). These were distributed among individual samples and were considered random and non‐systematic outliers and therefore removed from the dataset.
3.2.2. Number of CO2 peaks retained from single sample measurements
In the next step, the optimal number of sample peaks (Figure 2) used in the data processing was determined based on the variation of the δ13C and δ18O single standard deviation (1σ) of the sample means (top plot) and means per CO2 source (bottom plot) shown in Figures S3 and S4 (supporting information). The analysis of individual samples revealed no significant change in the median 1σ values with expanding the number of included CO2 sample gas peaks (n) from n = 3 to n = 10 (Figure S3 [supporting information]). However, there was a reduction in the interquartile range of the 1σ values with an increasing number of consecutive CO2 peaks used. The variation in 1σ values showed a similar picture for data grouped by CO2 source as for individual sample vials. In principle, we recommend including all 10 sample gas peaks in the evaluation, as the lowest 1σ for gas sample replicates can be achieved in this way (Figure S3, S4 [supporting information]). If the measurement accuracy of replicate samples is not a priority, or if only single samples are analysed, one could reduce the number of sample gas aliquots analysed to five to increase sample throughput and one could expect a probably slightly larger 1σ of the single sample mean value. Although the analysis of the oxygen data showed significantly higher standard deviations (Figure S4 [supporting information]), the picture compared to the carbon data was the same. We therefore recommend the same procedure for evaluating the oxygen data as for the carbon data. The 1σ value of the means using 10 CO2 sample gas peaks was better than ±0.14‰ and ±0.07‰ for δ13C single sample vials and for CO2 source means and better than ±0.60‰ and ±0.20‰ for those of δ18O.
3.2.3. Normalisation and system performance
The δ‐values of 10 CO2 peaks per sample vial were averaged and denoted as mean values per sample vial. The mean values per sample vial of the CO2 reference gases (R1, R2, R3, R4) were used to normalise the δ‐values of the tank samples of B1, B2, B3, B4 by linear regression of the assigned true versus gathered δ‐values and according to Paul et al. 21 The linear regression parameters showed a slope of 1.008 and 1.000 and an R 2 better than 0.999 for δ‐values of carbon and oxygen, respectively. The mean values of the CO2 reference gas sample vial were checked by normalisation with the assigned true δ‐values of the remaining three reference gases. The respective regression parameters always had a slope of 1.008 for carbon and between 0.99 and 1.01 for oxygen with an R 2 better than 0.999.
Normalised sample vial means were grouped by CO2 source (B1, B2, B3, B4, R1, R2, R3, R4) to obtain a calibrated mean δ‐value and single standard deviation (1σ) for each CO2 source (n = 3). Results are presented in Table 1 and indicate a measurement uncertainty for the analysis of CO2‐source sample replicates (single standard deviation [1σ]) of less than 0.07‰ and 0.20‰ for carbon and oxygen isotope ratios, with an overall mean precision of 0.04 and 0.09‰, respectively. The four air tanks serve as calibration gases in the monitoring of urban CO2 inventories using LAS. 22 The specified 1σ represents the measurement uncertainty and uncertainty in the production of sample replicates. To include the uncertainty associated with the uncertainty of the δ‐values of the reference gases used for normalisation, the model of Meija and Chartrand 23 was applied to calculate a combined uncertainty (u) for the δ‐values of each air tank, which is also presented in Table 1.
TABLE 1.
Overview of the results obtained from the analysis of reference gases (R1, R2, R3, R4) admixed to synthetic air (SA) and samples from four air tanks (B1, B2, B3, B4).
| CO2 source | Name | δ13C | δ18O | CO2 mole fraction | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Assigned true | σ | Calibrated mean | 1σ | u | Assigned true | σ | Calibrated mean | 1σ | u | Calibrated | σ | Recovered mean | 1σ | ||
| Reference gas | R1 | −6.65 | 0.13 | −6.62 | 0.04 | −17.82 | 0.20 | −17.59 | 0.05 | ||||||
| Reference gas | R2 | −6.73 | 0.13 | −6.76 | 0.06 | −17.97 | 0.20 | −18.17 | 0.14 | ||||||
| Reference gas | R3 | −25.71 | 0.20 | −25.71 | 0.02 | −28.72 | 0.15 | −28.89 | 0.13 | ||||||
| Reference gas | R4 | −39.00 | 0.21 | −39.00 | 0.07 | −30.34 | 0.20 | −30.14 | 0.06 | ||||||
| Air tank | B1 | −14.65 | 0.03 | 0.13 | −18.57 | 0.05 | 0.2 | 404 | 8.1 | 404.6 | 1.1 | ||||
| Air tank | B2 | −10.08 | 0.06 | 0.10 | −16.17 | 0.20 | 0.4 | 404 | 8.1 | 403.3 | 0.4 | ||||
| Air tank | B3 | −17.56 | 0.03 | 0.13 | −20.14 | 0.03 | 0.2 | 513 | 10.3 | 513.0 | 1.2 | ||||
| Air tank | B4 | −7.60 | 0.02 | 0.12 | −14.91 | 0.07 | 0.3 | 406 | 8.1 | 405.9 | 0.9 | ||||
| 0.04 | 0.09 | ||||||||||||||
Note: Calibrated mean values (n = 3) represent normalised carbon and oxygen δ‐values (vs Vienna Peedee Belemnite [VPDB] using the assigned true values ±σ of R1, R2, R3, R4) ± the single standard deviation (1σ) representing the measurement uncertainty and the combined uncertainty (u) for the calibrated δ‐value means of B1, B2, B3 and B4. The calibrated mole fraction (± σ) of CO2 in B1, B2, B3 and B4 is presented versus the recovered mean mole fraction ±1σ.
In general, it is recommended to always follow the identical‐treatment‐principle 24 of samples and reference materials when analysing CO2 from gas samples to avoid ambiguities about the calibrated isotope ratios obtained. For reference materials, we refer to Brewer et al. 25 for a comprehensive review of appropriate reference materials to use when analysing gas samples (whole air samples), and to Ghosh et al. 26 on scale anchoring. Identical treatment also accounts for matrix effects, as discussed in Levitt. 27 With the method presented by us, the sample vials are flushed with synthetic air before either CO2 reference gas is admixed manually or tank air is flushed through. However, the remaining matrix (SA) in the vials was the same, as the SA used for flushing and making the air tanks was from the same manufacturer (Messer) and of the same quality. Furthermore, due to the slopes of the linear regression lines for the δ‐value normalisation, no matrix effect could be identified. In addition, the calibrated δ‐values of the reference gases did not indicate a larger offset for 13C‐depleted CO2, which Tu et al 28 had been reported.
Apart from maintaining the identical treatment, the evaluation of the measurement set‐up has tried to also tackle other analytical problems arising from the analysis of gas (air) samples. Due to an N2/O2 sample gas matrix, problems caused by ‘blanking’ (relative position of the MS and the sample capillary of the open split of the low‐flow sample) as discussed in Elsig and Leuenverger 29 must be prevented. Resulting isobaric interferences could not be identified according to the results presented in the section on chromatographic resolution. The problem of isotopic exchange with accumulated water carried over from the sample vials was investigated as follows. First, at evaluating the number of CO2 peaks from single sample measurements (Section 3.2.2), there was no indication of progressive change in the δ‐values of successive CO2 peaks. Second, with the determination of the maximum sample vial throughput before the LN2 trap has to be refilled. Therefore, a measurement sequence of sample vials filled with laboratory air was run. The automatic operation time limit was 23.7 h, and the drift of the δ‐values was 7.6 10−6‰ h−1 and 4.0 10−5‰ h−1 for δ13C‐ and δ18O, which also pointed towards minor water issues. Another factor, sample vial contamination (CO2 peaks obtained by analysis of sample vials flushed with SA only), also proved to be negligible (Figure 3). Finally, as the 1σ values did not depend on the magnitude of the calibrated δ‐values, memory effects (i.e., reminiscences of the previous sample that changed the isotopic composition of the measured sample) were not observed.
3.2.4. Calculation of the CO2 mole fraction
The air tank sample measurement data (B1, B2, B3, B4) was also used to back‐calculate the CO2 mole fraction, based on the correlation of the sum of the CO2 peak area (Vs), including peaks 1 to 10, and the calibrated CO2 mole fraction according to Messer AG. The results of the linear regression showed that the data points were correlated with an R 2 of 1.00 and gave a y‐intercept of −17 μmol mol−1 when extrapolated through the origin. The calculated recovered CO2 mole fractions were found to have a maximum deviation from the calibrated values of 0.7 μmol mol−1 and gave a single standard deviation of CO2 source sample replicates of better than 1.2 μmol mol−1 (Table 1). Defining the limit of quantification as 20 times the mass 44 background intensity (30–15 mV at peaks 1 to 10), which also ensures a signal intensity within IRMS detector non‐linearity, would yield a minimum CO2 mole fraction of ~57 μmol mol−1.
4. CONCLUSIONS
The presented protocol describes the design and operation of a measuring system for the determination of the stable carbon and oxygen isotope ratio and mole fraction of CO2 in air. Compared to already‐existing measuring systems, it is characterised by the sole use of readily available system components. The system improved the standard installation of the measuring system GB/ConFlo IV/Delta V Plus from Thermo Fisher Scientific by using a 250‐μL sample loop in combination with a downstream Cold Trap. As a result, when measuring 400 μmol mol−1 CO2 in the air, the system showed CO2 peak amplitudes of 8 V and excellent peak separation from other air constituents (N2, O2, water). There was no indication of memory or matrix effects, isobaric interference, ‘blanking’ issues, δ‐value drift or instrumental and sample vial contaminations with lab air.
Although the described system could not reach the measuring precision of a dual inlet system (δ13C: 0.01‰, δ18O: 0.02), it allows a reliable, fully automated sample analysis (~30 min each), using small sample volume flasks (12 mL). The system was found to be capable of providing accurate δ13C‐ and δ18O‐values and mole fractions of CO2 in the air with an averaged measurement precision (1σ) of 0.04‰, 0.09‰ and 1.2 μmol mol−1, respectively. The measurement accuracy achieved should therefore allow the isotope ratios obtained to be used in Keeling plot applications, 30 , 31 to study ecosystem respiration, to monitor urban CO2 emissions and to determine CO2 sources and sinks. From a more general point of view, it can be used to estimate exchange reactions between the ecosystem and the atmosphere. Finally, we also see the potential for use in the referencing/normalisation of secondary air reference tanks for field applications.
AUTHOR CONTRIBUTIONS
Simon Leitner: Conceptualisation (equal); formal analysis (lead); investigation (equal); methodology (equal); validation (lead); visualisation (lead); writing—original draft preparation (lead). Kathiravan Meeran: Investigation (equal); writing—review and editing (equal). Andrea Watzinger: Conceptualisation (equal); methodology (equal); writing—review and editing (equal).
PEER REVIEW
The peer review history for this article is available at https://www.webofscience.com/api/gateway/wos/peer-review/10.1002/rcm.9647.
Supporting information
Table S1. Time events of the Isodat isotope ratio mass spectrometry (IRMS) method entered in the time event tab. The Isodat configurator, therefore, needs to have the gas bench node added to the LF (low flow capillary) of the ConFlo IV interface, which has to be the source of the respective mass spectrometer (MS). The autosampler is added as AS2000 at the input of the gas bench node. The time event columns MS Capillary—ON and SamplDil 1—3 have to be added at the ConFlow IV interface events tab.
Table S2. Regression parameters for the δ‐value versus mass 44 amplitude isotope ratio mass spectrometry (IRMS) detector non‐linearity.
Figure S1. Plots of the isotope ratio mass spectrometry (IRMS)‐detector non‐linearity of the δ‐value and mass 44 amplitudes for carbon (top) and oxygen (bottom).
Figure S2. Comparison of the single standard deviation (1σ) of the δ13C values obtained from 10 carbon dioxide (CO2) peaks per sample dependent on the type of linearity correction applied to raw δ13C values. Points shaped as triangles show the difference between linearity corrected values using quadratic (x‐axis) and linear regression (y‐axis). Points shaped as rectangles show the comparison of 1σ calculated from raw δ13C values (y‐axis) versus quadratic regression corrected δ13C values (x‐axis).
Figure S3. Single standard deviation (1σ) of individual sample vials (top) and samples grouped by carbon dioxide (CO2)‐source (bottom) dependent on the number of CO2 measurement peaks used in the calculation of 1σ of carbon isotope ratios (δ13C) of CO2 in air.
Figure S4. Single standard deviation (1σ) of individual sample vials (top) and samples grouped by carbon dioxide (CO2)‐source (bottom) dependent on the number of CO2 measurement peaks used in the calculation of 1σ of the mean oxygen isotope ratio (δ18O) of CO2 in air.
ACKNOWLEDGEMENTS
We acknowledge financial support from the Austrian Science Fund (FWF) within grant P 30275 ‘How Dynamic Changes in Leaf Anatomy Affect Photosynthesis’ led by Daniel Tholen. This work was also funded by the Vienna Science and Technology Fund (WWTF) [10.47379/ESR20030] (‘Vienna Urban Carbon Laboratory’) led by Bradley Matthews.
Leitner S, Meeran K, Watzinger A. Stable isotope analysis of atmospheric CO2 using a Gasbench II‐Cold Trap‐IRMS setting. Rapid Commun Mass Spectrom. 2023;37(24):e9647. doi: 10.1002/rcm.9647
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
- 1. World Meteorological Organisation (MWO) . WMO Greenhouse Gas Bulletin No. 18: The State of Greenhouse Gases in the Atmosphere Based on Global Observations through 2021. WMO Greenhouse Gas Bulletin. 2022. Accessed June 29, 2023. https://library.wmo.int/doc_num.php?explnum_id=11352
- 2. IPCC. Climate Change . Mitigation of climate change summary for policymakers (SPM). Cambridge University Press; 2022. doi: 10.1017/9781009157926.001 [DOI] [Google Scholar]
- 3. Voglar GE, Zavadlav S, Levanič T, Ferlan M. Measuring techniques for concentration and stable isotopologues of CO2 in a terrestrial ecosystem: A review. Earth‐Science Rev. 2019;199:102978. doi: 10.1016/j.earscirev.2019.102978 [DOI] [Google Scholar]
- 4. Manaj S, Kim ST. Techniques for measuring carbon and oxygen isotope compositions of atmospheric CO2 via isotope ratio mass spectrometry. Rapid Commun Mass Spectrom. 2021;35(4):e8995. doi: 10.1002/rcm.8995 [DOI] [PubMed] [Google Scholar]
- 5. Leitner S, Feichtinger W, Mayer S, et al. UAV‐based sampling systems to analyse greenhouse gases and volatile organic compounds encompassing compound‐specific stable isotope analysis. Atmos Meas Tech. 2023;16(2):513‐527. doi: 10.5194/amt-16-513-2023 [DOI] [Google Scholar]
- 6. Leitner S, Hood‐Nowotny R, Watzinger A. Successive and automated stable isotope analysis of CO2, CH4 and N2O paving the way for unmanned aerial vehicle‐based sampling. Rapid Commun Mass Spectrom. 2020;34(24):1‐11. doi: 10.1002/rcm.8929 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Ferretti DF, Lowe DC, Martin RJ, Brailsford GW. A new gas chromatograph‐isotope ratio mass spectrometry technique for high‐precision, N2O‐free analysis of δ13C and δ18O in atmospheric CO2 from small air samples. J Geophys Res Atmos. 2000;105(D5):6709‐6718. doi: 10.1029/1999JD901051 [DOI] [Google Scholar]
- 8. Chamberlain SD, Ingraffea AR, Sparks JP. Sourcing methane and carbon dioxide emissions from a small city: Influence of natural gas leakage and combustion. Environ Pollut. 2016;218:102‐110. doi: 10.1016/j.envpol.2016.08.036 [DOI] [PubMed] [Google Scholar]
- 9. Djuricin S, Pataki DE, Xu X. A comparison of tracer methods for quantifying CO2 sources in an urban region. J Geophys Res Atmos. 2010;115(11):1‐13, D11303. doi: 10.1029/2009JD012236 [DOI] [Google Scholar]
- 10. Górka M, Lewicka‐Szczebak D. One‐year spatial and temporal monitoring of concentration and carbon isotopic composition of atmospheric CO2 in a Wrocław (SW Poland) city area. Appl Geochemistry. 2013;35:7‐13. doi: 10.1016/j.apgeochem.2013.05.010 [DOI] [Google Scholar]
- 11. Guha T, Ghosh P. Diurnal and seasonal variation of mixing ratio and δ13C of air CO2 observed at an urban station Bangalore. India Environ Sci Pollut Res. 2015;22(3):1877‐1890. doi: 10.1007/s11356-014-3530-3 [DOI] [PubMed] [Google Scholar]
- 12. Jasek A, Zimnoch M, Gorczyca Z, Smula E, Rozanski K. Seasonal variability of soil CO2 flux and its carbon isotope composition in Krakow urban area. Southern Poland Isotopes Environ Health Stud. 2014;50(2):143‐155. doi: 10.1080/10256016.2014.868455 [DOI] [PubMed] [Google Scholar]
- 13. Moore J, Jacobson AD. Seasonally varying contributions to urban CO2 in the Chicago, Illinois, USA region: Insights from a high‐resolution CO2 concentration and δ13C record. Elementa. 2015;3:1‐17. doi: 10.12952/journal.elementa.000052 [DOI] [Google Scholar]
- 14. Liu M, Pang L, Yang D, Wang J, Zhou Y. Kinetics study of gas pollutant adsorption and thermal desorption on silica gel. Appl Sci. 2017;7(6):609. doi: 10.3390/app7060609 [DOI] [Google Scholar]
- 15. Pataki DE, Bowling DR, Ehleringer JR, Zobitz JM. High resolution atmospheric monitoring of urban carbon dioxide sources. Geophys Res Lett. 2006;33(3):1‐5, L03813. doi: 10.1029/2005GL024822 [DOI] [Google Scholar]
- 16. Zimnoch M, Florkowski T, Necki JM, Neubert REM. Diurnal variability of δ13 and δ18O of atmospheric CO2 in the urban atmosphere of Kraków. Poland Isotopes Environ Health Stud. 2004;40(2):129‐143. doi: 10.1080/10256010410001670989 [DOI] [PubMed] [Google Scholar]
- 17. Stichler W. Interlaboratory comparison of new materials for carbon and oxygen isotope ratio measurements. International Atomic Energy Agency (IAEA); 1995. [Google Scholar]
- 18. Brand WA, Huang L, Mukai H, Chivulescu A, Richter JM, Rothe M. How well do we know VPDB? Variability of δ13C and δ18O in CO2 generated from NBS19‐calcite. Rapid Commun Mass Spectrom. 2009;23(6):915‐926. doi: 10.1002/rcm.3940 [DOI] [PubMed] [Google Scholar]
- 19. Coplen TB. Guidelines and recommended terms for expression of stable‐isotope‐ratio and gas‐ratio measurement results. Rapid Commun Mass Spectrom. 2011;25(17):2538‐2560. doi: 10.1002/rcm.5129 [DOI] [PubMed] [Google Scholar]
- 20. Grubbs FE. Procedures for detecting outlying observations in samples. Dent Tech. 1969;11(1):1‐21. doi: 10.1080/00401706.1969.10490657 [DOI] [Google Scholar]
- 21. Paul D, Skrzypek G, Fórizs I. Normalization of measured stable isotopic compositions to isotope reference scales – a review. Rapid Commun Mass Spectrom. 2007;21(18):3006‐3014. doi: 10.1002/rcm.3185 [DOI] [PubMed] [Google Scholar]
- 22. Meeran K, Matthews B, Leitner S, Sanden H, Chen J, Watzinger A. Tall tower measurements with laser isotope spectrometry to investigate urban CO2 emissions in Vienna. EGU General Assembly; 2023. doi: 10.5194/egusphere-egu23-13255 [DOI] [Google Scholar]
- 23. Meija J, Chartrand MMG. Uncertainty evaluation in normalization of isotope delta measurement results against international reference materials. Anal Bioanal Chem. 2018;410(3):1061‐1069. doi: 10.1007/s00216-017-0659-1 [DOI] [PubMed] [Google Scholar]
- 24. Werner RA, Brand WA. Referencing strategies and techniques in stable isotope ratio analysis 2001:501‐519. doi: 10.1002/rcm.258 [DOI] [PubMed]
- 25. Brewer PJ, Kim JS, Lee S, et al. Advances in reference materials and measurement techniques for greenhouse gas atmospheric observations. Metrologia. 2019;56(3):034006. doi: 10.1088/1681-7575/ab1506 [DOI] [Google Scholar]
- 26. Ghosh P, Patecki M, Rothe M, Brand WA. Calcite‐CO2 mixed into CO2‐free air: A new CO2‐in‐air stable isotope reference material for the VPDB scale. Rapid Commun Mass Spectrom. 2005;19(8):1097‐1119. doi: 10.1002/rcm.1886 [DOI] [Google Scholar]
- 27. Levitt NP. Sample matrix effects on measured carbon and oxygen isotope ratios during continuous‐flow isotope‐ratio mass spectrometry. Rapid Commun Mass Spectrom. 2014;28(21):2259‐2274. doi: 10.1002/rcm.7019 [DOI] [PubMed] [Google Scholar]
- 28. Tu KP, Brooks PD, Dawson TE. Using septum‐capped vials with continuous‐flow isotope ratio mass spectrometric analysis of atmospheric CO2 for Keeling plot applications. Rapid Commun Mass Spectrom. 2001;15(12):952‐956. doi: 10.1002/rcm.320 [DOI] [Google Scholar]
- 29. Elsig J, Leuenberger MC. 13C and 18O fractionation effects on open splits and on the ion source in continuous flow isotope ratio mass spectrometry. Rapid Commun Mass Spectrom. 2010;24(10):1419‐1430. doi: 10.1002/rcm.4531 [DOI] [PubMed] [Google Scholar]
- 30. Keeling CD. The concentration and isotopic abundances of atmospheric carbon dioxide in rural areas. Geochim Cosmochim Acta. 1958;13(4):322‐334. doi: 10.1016/0016-7037(58)90033-4 [DOI] [Google Scholar]
- 31. Ehleringer JR, Cook CS. Carbon and oxygen isotope ratios of ecosystem respiration along an Oregon conifer transect: Preliminary observations based on small‐flask sampling. Tree Physiol. 1998;18(8‐9):513‐519. doi: 10.1093/treephys/18.8-9.513 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1. Time events of the Isodat isotope ratio mass spectrometry (IRMS) method entered in the time event tab. The Isodat configurator, therefore, needs to have the gas bench node added to the LF (low flow capillary) of the ConFlo IV interface, which has to be the source of the respective mass spectrometer (MS). The autosampler is added as AS2000 at the input of the gas bench node. The time event columns MS Capillary—ON and SamplDil 1—3 have to be added at the ConFlow IV interface events tab.
Table S2. Regression parameters for the δ‐value versus mass 44 amplitude isotope ratio mass spectrometry (IRMS) detector non‐linearity.
Figure S1. Plots of the isotope ratio mass spectrometry (IRMS)‐detector non‐linearity of the δ‐value and mass 44 amplitudes for carbon (top) and oxygen (bottom).
Figure S2. Comparison of the single standard deviation (1σ) of the δ13C values obtained from 10 carbon dioxide (CO2) peaks per sample dependent on the type of linearity correction applied to raw δ13C values. Points shaped as triangles show the difference between linearity corrected values using quadratic (x‐axis) and linear regression (y‐axis). Points shaped as rectangles show the comparison of 1σ calculated from raw δ13C values (y‐axis) versus quadratic regression corrected δ13C values (x‐axis).
Figure S3. Single standard deviation (1σ) of individual sample vials (top) and samples grouped by carbon dioxide (CO2)‐source (bottom) dependent on the number of CO2 measurement peaks used in the calculation of 1σ of carbon isotope ratios (δ13C) of CO2 in air.
Figure S4. Single standard deviation (1σ) of individual sample vials (top) and samples grouped by carbon dioxide (CO2)‐source (bottom) dependent on the number of CO2 measurement peaks used in the calculation of 1σ of the mean oxygen isotope ratio (δ18O) of CO2 in air.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
