Abstract
Introduction
As a result of the insufficient ocular anatomical parameters used to customize implantable collamer lens (ICL), many patients still cannot achieve a suitable vault after ICL implantation surgery. This study analyzed the characteristics of a new anatomical parameter crystalline lens rise (CLR) in a population with high myopia and explored the influence of CLR on the vault after ICL implantation.
Methods
Patients (298 eyes) with high myopia who underwent ICL implantation were enrolled to study CLR characteristics. Postoperatively, patients (159 eyes) were divided into five groups according to the value of CLR (A, CLR ≤ − 150; B, − 150 < CLR ≤ 0; C, 0 < CLR < 150; D, 150 ≤ CLR < 300; E, CLR ≥ 300 μm), and to investigate the correlation between CLR and vault.
Results
In the 298 eyes, the CLR had a normal distribution (P = 0.35) and the mean CLR was 67.93 ± 150.66 μm. Ninety-nine eyes (33.22%) had a CLR ≤ 0 μm, of which 20 eyes (6.71%) had a CLR ≤ − 150 μm; 199 eyes (66.78%) had a CLR > 0 μm, of which 20 eyes (6.71%) had a CLR ≥ 300 μm. In 159 eyes, the CLR was negatively correlated with the vault at 1 day (R = − 0.497, P < 0.001), 3 months (R = − 0.505, P < 0.001), and 6 months (R = − 0.505, P < 0.001) postoperatively. At 6 months, the vault of group A was statistically significantly different compared to groups B–E (all P < 0.05), and that of group E was statistically significantly different compared to groups A–D (all P < 0.001). The remaining groups did not show statistically significant differences (all P > 0.05).
Conclusion
The CLR had a normal distribution in the high myopia population, and 13.42% of the CLR values were extreme (CLR ≤ − 150 μm or CLR ≥ 300 μm). A larger ICL diameter than that recommended by the manufacturer should be considered when the CLR is ≥ 300 μm and a smaller ICL diameter should be considered when the CLR is ≤ − 150 μm.
Keywords: Crystalline lens rise, Intraocular collamer lens, Vault, Refractive surgery, High myopia
Key Summary Points
| Why carry out this study? |
| More anatomical parameters of eyes are needed to improve the accuracy of lens size selection. |
| Previous studies have shown that the crystalline lens rise (CLR) value could have an influence on the postoperative vault. |
| This study aimed to analyze the characteristics of a new anatomical parameter crystalline lens rise in a population with high myopia and explore the influence of CLR on the vault after implantable collamer lens (ICL) implantation. |
| What was learned from the study? |
| CLR in the high myopia population had a normal distribution. The mean CLR value was 67.93 ± 150.66 μm (range − 350 to 480 μm; and 95% confidence interval 50.76–85.11 μm). |
| CLR was negatively correlated with the postoperative vault. When CLR was from − 150 to 300 μm, it has little effect on the vault. When CLR ≥ 300 μm, the achieved vault is lower; when CLR ≤ − 150 μm, the achieved vault is higher. |
| To achieve the ideal postoperative vault, if the CLR is ≥ 300 μm, a larger ICL diameter should be considered, and if the CLR is ≤ − 150 μm, a smaller ICL diameter or an intraoperative ICL rotation to a certain angle might be good choices. |
Introduction
Intraocular collamer lens (ICL) implantation is currently the mainstream intraocular surgical method used to correct refractive errors in the clinic and has the advantages of reversibility, a wide range of correction degrees, and high postoperative visual quality. The vault is defined as the vertical distance between the posterior surface of the ICL and the anterior surface of the crystalline lens, which is an important indicator for evaluating the safety of ICL implantation surgery, and its ideal range is 0.25–0.75 mm [1, 2]. The postoperative vault is affected by ocular anatomical parameters, ICL diameter, and their interaction [3]. An abnormal postoperative vault is the leading cause of complications. Previous studies have found that if the ICL size is selected only on the basis of white-to-white (WTW) and anterior chamber depth (ACD), as usual, more than 20% of patients do not achieve the ideal vault after surgery [4–6]. As a result of the complexity of intraocular anatomy, and despite surgeons comprehensively considering the sulcus-to-sulcus (STS) distance [7, 8], some patients still experience an abnormal vault after surgery [9], and 2.6% require ICL replacement [10–12]. The accurate selection of the appropriate diameter of the ICL to prevent an abnormal vault remains a problem. Related studies have found a link between crystalline lens rise (CLR) and vault [4, 13–16]; therefore, this study further explored the distribution characteristics of CLR in high myopia populations and the effect of CLR on vault, with the aim of providing more reference indicators to improve the safety of ICL implantation after surgery.
Methods
Study Design
This prospective study included two parts. In the first part, all patients (298 eyes) who underwent ICL V4c implantation at the Refractive Center of the First Affiliated Hospital of Zhengzhou University between June 2020 and December 2022 were included to measure preoperative CLR and analyze its distribution characteristics in a high myopia population. In the second part, after surgery, patients (159 eyes) who were followed up regularly for more than 6 months were included to investigate the vault. They were further divided into five groups according to the value of CLR (A, CLR ≤ − 150; B, − 150 < CLR ≤ 0; C, 0 < CLR < 150; D, 150 ≤ CLR < 300; E, CLR ≥ 300 μm) to explore the links between preoperative CLR and postoperative vault. All patients underwent implantation using the ICL model recommended by the manufacturer. All of the patients were included in the monocular group. The right eye was enrolled if both eyes satisfied the inclusion requirements. Patients who did not meet the ICL implantation criteria were excluded. This study was conducted in compliance with the tenets of the Declaration of Helsinki, and the study protocol was approved by the Ethics Committee of the First Affiliated Hospital of Zhengzhou University (2023-KY-1354). All patients provided informed consent after receiving a detailed explanation of the surgical procedure before the operation.
Anterior Segment Optical Coherence Tomography (AS-OCT) Imaging
Anterior segment imaging was performed using anterior segment optical coherence tomography (AS-OCT) (Visante). The parameters measured by AS-OCT were defined as follows: The vault was defined as the vertical distance between the posterior surface of the ICL and the anterior surface of the crystalline lens. The ACD was defined as the distance between the posterior apex of the cornea and the front apex of the crystalline lens. Angle-to-angle (ATA) diameter was defined as the distance between the angle recesses on the nasal and temporal sides. CLR was defined as the vertical distance between the anterior pole of the crystalline lens and the horizontal line that joins the two anterior chamber angle crypts (Fig. 1) [17]. When the line was above the anterior pole of the crystalline lens, the CLR value was recorded as positive (CLR > 0 μm), otherwise as negative (CLR < 0 μm). To improve the accuracy of the measurement, the small absolute value of the CLR was calculated by subtracting the vertical central distance between the corneal endothelium and the anterior pole of the crystalline lens from the vertical central distance between the corneal endothelium and the horizontal line joining the two anterior chamber angle crypts. Each measurement was repeated three times by the same physician under the same mesopic conditions, and the average value was taken.
Fig. 1.

Image of CLR, ATA, and ACD measured by AS-OCT. CLR crystalline lens rise, ATA angle-to-angle, ACD anterior chamber depth, AS-OCT anterior segment optical coherence tomography, CCT Central corneal thickness
Surgical Procedure
After topical anesthesia, a 3.0-mm temporal corneal incision was made and an ophthalmic viscosurgical device (OVD) was injected into the anterior chamber. When the ICL was placed in the anterior chamber, a special crystalline tweezer was used to adjust the four haptics in the sulcus. The left OVD was completely removed from the anterior chamber using a balanced salt solution. Intraocular pressure was normal. Postoperatively, tobramycin and dexamethasone eye drops and antibiotics were administered topically.
Postoperative Follow-up
Postoperative measurements, including automated refraction, uncorrected distance visual acuity, intraocular pressure (IOP), slit-lamp microscopy, and AS-OCT, were performed 1 day, 3 months, and 6 months postoperatively. Symptomatic measures were provided during follow-up, as necessary, according to the patient’s condition.
Statistical Analysis
All statistical analyses were performed using RStudio (Version 4.3.0), and the quantitative data were presented as mean ± standard deviation. Pearson’s correlation analysis was performed to investigate the correlation between CLR and postoperative vault. Repeated-measures analysis of variance was performed to compare the vault at different postoperative time points (1 day, 3 months, and 6 months). A one-way analysis of variance was performed to compare vaults in the different groups (A–E). P < 0.05 was considered statistically significant.
Results
Preoperative Baseline Data
In the first part, 298 patients (298 eyes) were included to investigate the characteristics of CLR in the high myopia population, including 132 men (44.30%) and 166 women (55.70%). The mean age was 24.73 ± 4.75 years (range 18–38 years), and the mean spherical equivalent (SE) was − 9.95 ± 2.61 D (range − 6.00 to − 18.00 D).
In the second part, 159 eyes (159 patients) were included to investigate the effect of CLR on vault in the high myopia population, which included 61 men (38.36%) and 98 women (61.64%). The mean age was 24.70 ± 4.79 years (range 18–38 years), and the mean SE was − 10.09 ± 2.53 D (range − 6.25 to − 17.25 D). Table 1 presents the descriptive data of the samples.
Table 1.
Preoperative characteristics
| Characteristic | First part | Second part |
|---|---|---|
| N = 298 eyes/298 patients | N = 159 eyes/159 patients | |
| Age (years) | 24.73 ± 4.75 (18, 38) | 24.70 ± 4.79 (18, 38) |
| Gender (M/F) | 132/166 | 61/98 |
| Sphere (D) | − 9.36 ± 2.53 (− 6.00, − 18.00) | − 9.60 ± 2.41 (− 6.00, − 17.00) |
| Cylinder (D) | − 1.18 ± 1.17 (0, − 5.75) | − 1.02 ± 1.00 (0, − 5.00) |
| SE (D) | − 9.95 ± 2.61 (− 6.00, − 18.00) | − 10.09 ± 2.53 (− 6.25, − 17.25) |
| WTW (mm) | 11.53 ± 0.36 (10.78, 12.41) | 11.52 ± 0.37 (10.89, 12.40) |
| CLR (μm) | 67.93 ± 150.66 (− 350, 480) | 68.52 ± 177.46 ( − 320, 480) |
SE spherical equivalent, WTW white-to-white, CLR crystalline lens rise
Distribution of CLR in High Myopia Population
In this part, we confirmed that the CLR had a normal distribution in the 298 eyes (P = 0.345) using the Shapiro–Wilk test (Fig. 2). The mean CLR value was 67.93 ± 150.66 μm (range − 350 to 480 μm; and 95% confidence interval 50.76–85.11 μm).
Fig. 2.

The CLR had a normal distribution in the high myopia population. CLR crystalline lens rise
Further statistical tests revealed that of the 298 eyes, 20 (6.71%), 79 (26.51%), 108 (36.24%), 71 (23.83%), and 20 (6.71%) eyes had a CLR ≤ − 150, − 150 to 0, 0–150, 150–300, and ≥ 300 μm, respectively (Table 2).
Table 2.
Distribution of CLR in the high myopia population (298 eyes)
| CLR (μm) | Eyes | Percentage (%) | Cumulative percentage (%) |
|---|---|---|---|
| Less than − 150 | 20 | 6.71 | 6.71 |
| − 150 to 0 | 79 | 26.51 | 33.22 |
| 0–150 | 108 | 36.24 | 69.46 |
| 150–300 | 71 | 23.83 | 93.29 |
| More than 300 | 20 | 6.71 | 100 |
CLR crystalline lens rise
Effect of CLR on Vault in High Myopia Population
In this part, the vault of the 159 eyes at postoperative 1 day, 3 months, and 6 months was 0.66 ± 0.23 mm, 0.58 ± 0.20 mm, and 0.56 ± 0.19 mm, respectively (Table 3, Fig. 3). The vault on postoperative 1 day was significantly different from that at 3 and 6 months (all P < 0.05). However, there were no statistically significant differences between 3 and 6 months (P = 0.71).
Table 3.
Changes in the vault after ICL implantation (mean ± standard deviation, 159 eyes)
| Time | Vault (mm) |
|---|---|
| 1 day | 0.66 ± 0.23 |
| 3 months | 0.58 ± 0.20 |
| 6 months | 0.56 ± 0.19 |
| F | 156.48 |
| P | < 0.001 |
ICL implantable collamer lens
Fig. 3.

Time course of vault changes in eyes undergoing ICL implantation. The vault decreased gradually with time and stabilized at 3 months after operation. ICL implantable collamer lens
CLR showed a negative correlation with postoperative vault at 1 day, 3 months, and 6 months after surgery, according to Pearson’s correlation analysis, with a correlation coefficient of − 0.497 (P < 0.001), − 0.505 (P < 0.001), and − 0.505 (P < 0.001), respectively (Fig. 4).
Fig. 4.

The vault was negatively correlated with CLR at all time points of postoperative follow-up. CLR crystalline lens rise
The vault at 6 months postoperatively was selected to further analyze the differences between the different CLR groups (A–E). Table 4 and Fig. 5 show the vaults at 6 months postoperatively in the different CLR groups. Statistically significant differences were observed between the subgroups (F = 24.00; P < 0.001). After Bonferroni comparisons were performed, group A was significantly different from groups B–E (all P < 0.05), and group E was significantly different from groups A–D (all P < 0.001). However, there were no statistically significant differences between groups B, C, and D (all P > 0.05).
Table 4.
Comparison of the vault in different groups at 6 months ICL implantation (mean ± standard deviation, 159 eyes)
| Group | Eyes | Vault (mm) | 95% CI |
|---|---|---|---|
| A | 20 | 0.75 ± 0.19 | (0.66, 0.83) |
| B | 29 | 0.58 ± 0.18 | (0.51, 0.65) |
| C | 64 | 0.59 ± 0.17 | (0.55, 0.63) |
| D | 26 | 0.52 ± 0.13 | (0.47, 0.58) |
| E | 20 | 0.33 ± 0.11 | (0.28, 0.38) |
| F | 24.00 | ||
| P | < 0.001 |
A, CLR ≤ − 150 μm; B, − 150 < CLR ≤ 0 μm; C, 0 < CLR < 150 μm; D, 150 ≤ CLR < 300 μm; E, CLR ≥ 300 μm
CLR crystalline lens rise, ICL implantable collamer lens, CI confidence interval
Fig. 5.

The difference of vault at 6 months among different groups. CLR crystalline lens rise. A, CLR ≤ − 150 μm; B, − 150 < CLR ≤ 0 μm; C, 0 < CLR < 150 μm; D, 150 ≤ CLR < 300 μm; E, CLR ≥ 300 μm
Discussion
Studies have shown that approximately 20% of patients with ICL customization based on WTW and ACD cannot achieve a perfect vault after the operation [18]. Although the diameter of the sulcus is included as one of the criteria for selecting the ICL model, many patients still require secondary surgery for ICL replacement or even ICL removal [19]. In 2005, Baïkoff et al. proposed the use of CLR as a safety standard for crystalline surgery [17].
CLR was defined as the vertical distance between the anterior pole of the crystalline lens and the horizontal line joining the two anterior chamber angle crypts, which directly reflected the relationship between the anterior pole of the crystalline lens and the anterior chamber. CLR is negatively correlated with ACD [20, 21]. It is well understood that the more convex the crystalline is, the more space it occupies in the anterior chamber, which might result in a lower vault than predicted. Fully understanding the distribution characteristics of CLR in the high myopia population is of great significance for the preoperative selection of the ICL diameter and evaluation of the postoperative vault after ICL surgery. Therefore, as one of the factors that influence the postoperative vault, CLR should not be ignored in the selection of ICL.
In this study, we observed CLR in 298 patients with high myopia. The CLR was normally distributed, which is consistent with the findings of Gonzalez-Lopez et al. [22]. That study also showed most patients with high myopia had the anterior pole of the crystalline lens located above the line that joins the two anterior chamber angle crypts, but the corresponding proportion of this group of patients in this study (66.8%) was lower than that reported by Gonzalez-Lopez et al. (86.5%) [22]. First, the CLR values for the different lighting conditions were different. When the light intensity changed, the CLR value also changed. Second, it is also possible that the instruments used to measure the CLR values were different. The ability to recognize anatomical structures also differed for different instruments.
We know from our experiments that CLR is negatively correlated with the postoperative vault, which is consistent with previous studies [18, 23, 24]. The authors reported that the mean CLR of the low vault group (< 100 μm) was higher than that of the high vault group (> 750 μm) [23]. Wang et al. [25] found that the correlation coefficient between CLR and the 1-year postoperative vault was − 0.509, which was close to the correlation coefficient ( − 0.505) between CLR and the 6-month postoperative vault in this study. The fact that crystalline lens could have a crucial influence on vault was also confirmed by Qi et al. [26]. On the basis of the results of studies by Trancón et al. [18] and Nakamura et al. [4], it could be observed that for every 100-μm increase in CLR, the vault decreased by 37–40 μm.
Previous studies have shown that the vault stabilizes 3 months postoperatively [24, 27]. The results of our study also did not show changes in the vault between 3 and 6 months after surgery. Therefore, we chose the vault 6 months after surgery to analyze the relationship between the CLR and the vault. Given that there was no clear grouping standard for CLR measured by optical coherence tomography, to further study how CLR affected the postoperative vault, we exploratively divided the subjects into five groups according to CLR: A, CLR ≤ − 150; B, − 150 < CLR ≤ 0; C, 0 < CLR < 150; D, 150 ≤ CLR < 300; and E, CLR ≥ 300 μm. The results showed that group A was significantly different from groups B–E (all P < 0.05), and group E was significantly different from groups A–D (all P < 0.001). There were no statistically significant differences between groups B, C, and D (all P > 0.05). In other words, this further study found that when the CLR was from − 150 to 300 μm, it had little effect on the postoperative vault. However, when the CLR was greater than 300 μm (CLR ≥ 300 μm), the distance between the intraocular lens and the crystalline lens decreased, and the postoperative vault was expected to be relatively lower; when the CLR was less than − 150 μm (CLR ≤ − 150 μm), the distance between the intraocular lens and the crystalline lens increased, and the postoperative vault was expected to be relatively higher.
A limitation of this study is the short follow-up period. In subsequent studies, we will further explore the relationship between the crystalline lens rise and the vault in the late postoperative period.
Conclusion
Our study groups the research data, for the first time, according to the CLR (A, CLR ≤ − 150; B, − 150 < CLR ≤ 0; C, 0 < CLR < 150; D, 150 ≤ CLR < 300; E, CLR ≥ 300 μm) to explore the effect of preoperative CLR on postoperative vault. The obtained results suggest the CLR was normally distributed. When the CLR was from − 150 to 300 μm, it had little effect on the postoperative vault. However, when the CLR was greater than 300 μm (CLR ≥ 300 μm), the postoperative vault was expected to be relatively lower; when the CLR was less than − 150 μm (CLR ≤ − 150 μm), the postoperative vault was expected to be relatively higher. The extreme value of the CLR should be considered when calculating the required lens size for the ICL. If the CLR is ≥ 300 μm, a diameter of ICL larger than that recommended by the manufacturer should be considered, and if the CLR is ≤ − 150 μm, a smaller diameter of ICL or an intraoperative ICL rotation at a certain angle may be good choices.
Acknowledgements
We thank all study participants for their involvement.
Author Contributions
Design and conduct of study (Ziyu Zhou, Xiaoyu Zhao); Collection (Ziyu Zhou, Wenxin Xue, Xiaohang Jiao), Management (Jing Yang, Weiqun Wang, Yanhui Bai), Analysis (Ziyu Zhou, Yanhui Bai), Interpretation of data (Ziyu Zhou, Yanhui Bai). Preparation (Ziyu Zhou, Xiaoyu Zhao), review or approval of manuscript (Ziyu Zhou); and responsibility for the integrity of the entire study and manuscript (Ziyu Zhou). All authors read and approved the final manuscript.
Funding
This study was supported by the Key R & D and Promotion Project of Henan Science and Technology Department (Strategic Promotion of Innovative R and D: Grant No. 202102310073). The funders had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication. The Rapid Service Fee was funded by the authors.
Data Availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
Declarations
Conflict of Interest
Ziyu Zhou, Xiaoyu Zhao, Xiaohang Jiao, Wenxin Xue, Jing Yang, Weiqun Wang, Yanhui Bai declare no conflicts of interests.
Ethical Approval
This study was conducted in compliance with the tenets of the Declaration of Helsinki, and the study protocol was approved by the Ethics Committee of the First Affiliated Hospital of Zhengzhou University (2023-KY-1354). All patients provided informed consent after receiving a detailed explanation of the surgical procedure before the operation.
Footnotes
Ziyu Zhou and Xiaoyu Zhao contributed equally and should be considered as joint first authors.
References
- 1.Güell JL, Morral M, Kook D, Kohnen T. Phakic intraocular lenses part 1: historical overview, current models, selection criteria, and surgical techniques. J Cataract Refract Surg. 2010;36(11):1976–1993. doi: 10.1016/j.jcrs.2010.08.014. [DOI] [PubMed] [Google Scholar]
- 2.Choi KH, Chung SE, Chung TY, Chung ES. Ultrasound biomicroscopy for determining visian implantable contact lens length in phakic IOL implantation. J Refract Surg. 2007;23(4):362–367. doi: 10.3928/1081-597X-20070401-08. [DOI] [PubMed] [Google Scholar]
- 3.Nam SW, Lim DH, Hyun J, Chung ES, Chung TY. Buffering zone of implantable collamer lens sizing in V4c. BMC Ophthalmol. 2017;17(1):260. doi: 10.1186/s12886-017-0663-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Implantable collamer lens sizing method based on swept-source anterior segment optical coherence tomography. Am J Ophthalmol. 2018;187:99–107. doi: 10.1016/j.ajo.2017.12.015. [DOI] [PubMed] [Google Scholar]
- 5.Lee DH, Choi SH, Chung ES, Chung TY. Correlation between preoperative biometry and posterior chamber phakic Visian implantable collamer lens vaulting. Ophthalmology. 2012;119(2):272–277. doi: 10.1016/j.ophtha.2011.07.047. [DOI] [PubMed] [Google Scholar]
- 6.Zhu QJ, Pan H, Yuan Y. Study on the early changes of vaulting after posterior chamber phakic intraocular lens implantation. Chin J Ocular Trauma Occup Eye Dis. 2022;44(2):118–122. [Google Scholar]
- 7.Zhu QJ, Chen WJ, et al. Prediction of the vaulting after posterior chamber intraocular lens implantation. Chin J Ophthalmol. 2021;57(7):519–525. doi: 10.3760/cma.j.cn112142-20201222-00837. [DOI] [PubMed] [Google Scholar]
- 8.Wang J, Wang ZY, Zhang Q, Tao LM. Comparative analysis of preoperative predicted vaults by 50 MHz full-scale ultrasound biomicroscopy and postoperative real vaults in phakic posterior chamber implantable collamer lens. J Clin Ophthalmol. 2021;29(1):45–49. [Google Scholar]
- 9.Packer M. Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens. Clin Ophthalmol. 2016 doi: 10.2147/OPTH.S111620. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Pesando PM, Ghiringhello MP, Di Meglio G, Fanton G. Posterior chamber phakic intraocular lens (ICL) for hyperopia: ten-year follow-up. J Cataract Refract Surg. 2007;33(9):1579–1584. doi: 10.1016/j.jcrs.2007.05.030. [DOI] [PubMed] [Google Scholar]
- 11.Zeng QY, Xie XL, Chen Q. Prevention and management of collagen copolymer phakic intraocular lens exchange: causes and surgical techniques. J Cataract Refract Surg. 2015;41(3):576–584. doi: 10.1016/j.jcrs.2014.06.036. [DOI] [PubMed] [Google Scholar]
- 12.Rayner SA, Bhikoo R, Gray T. Spherical implantable collamer lenses for myopia and hyperopia: 126 eyes with 1-year follow up. Clin Exp Ophthalmol. 2010;38(1):21–26. doi: 10.1111/j.1442-9071.2010.02192.x. [DOI] [PubMed] [Google Scholar]
- 13.Schmidinger G, Lackner B, Pieh S, Skorpik C. Long-term changes in posterior chamber phakic intraocular collamer lens vaulting in myopic patients. Ophthalmology. 2010;117(8):1506–1511. doi: 10.1016/j.ophtha.2009.12.013. [DOI] [PubMed] [Google Scholar]
- 14.Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Optimization of implantable collamer lens sizing based on swept-source anterior segment optical coherence tomography. J Cataract Refract Surg. 2020;46(5):742–748. doi: 10.1097/j.jcrs.0000000000000134. [DOI] [PubMed] [Google Scholar]
- 15.Cerpa Manito S, Sánchez Trancón A, Torrado Sierra O, Baptista AM, Serra PM. Biometric and ICL-related risk factors associated to sub-optimal vaults in eyes implanted with implantable collamer lenses. Eye Vis. 2021;8(1):26. doi: 10.1186/s40662-021-00250-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Gonzalez-Lopez F, Mompean B, Bilbao-Calabuig R, Vila-Arteaga J, Beltran J, Baviera J. Dynamic assessment of light-induced vaulting changes of implantable collamer lens with central port by swept-source OCT: pilot study. Transl Vis Sci Technol. 2018;7(3):4. doi: 10.1167/tvst.7.3.4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Baïkoff G, Bourgeon G, Jodai HJ, Fontaine A, Lellis FV, Trinquet L. Pigment dispersion and Artisan phakic intraocular lenses: crystalline lens rise as a safety criterion. J Cataract Refract Surg. 2005;31(4):674–680. doi: 10.1016/j.jcrs.2004.09.034. [DOI] [PubMed] [Google Scholar]
- 18.Trancón AS, Manito SC, Sierra OT, Baptista AM, Serra PM. Determining vault size in implantable collamer lenses: preoperative anatomy and lens parameters. J Cataract Refract Surg. 2020;46(5):728–736. doi: 10.1097/j.jcrs.0000000000000146. [DOI] [PubMed] [Google Scholar]
- 19.Reinstein DZ, Lovisolo CF, Archer TJ, Gobbe M. Comparison of postoperative vault height predictability using white-to-white or sulcus diameter-based sizing for the Visian implantable collamer lens. J Refract Surg. 2013;29(1):30–35. doi: 10.3928/1081597X-20121210-02. [DOI] [PubMed] [Google Scholar]
- 20.Matsuki T, Hirose F, Ito S, Hata M, Hirami Y, Kurimoto Y. Influence of anterior segment biometric parameters on the anterior chamber angle width in eyes with angle closure. J Glaucoma. 2015;24(2):144–148. doi: 10.1097/IJG.0000000000000090. [DOI] [PubMed] [Google Scholar]
- 21.Yan PS, Lin HT, Wang QL, Zhang ZP. Anterior segment variations with age and accommodation demonstrated by slit-lamp-adapted optical coherence tomography. Ophthalmology. 2010;117(12):2301–2307. doi: 10.1016/j.ophtha.2010.03.027. [DOI] [PubMed] [Google Scholar]
- 22.Gonzalez-Lopez F, Bilbao-Calabuig R, Mompean B, Luezas J, Ortega-Usobiaga J, Druchkiv V. Determining the potential role of crystalline lens rise in vaulting in posterior chamber phakic collamer lens surgery for correction of myopia. J Refract Surg. 2019;35(3):177–183. doi: 10.3928/1081597X-20190204-01. [DOI] [PubMed] [Google Scholar]
- 23.Mansoori T, Agraharam SG. Reverse pupillary block after retinal detachment surgery in an eye with toric implantable collamer lens. Int Ophthalmol. 2019;39(3):703–710. doi: 10.1007/s10792-018-0848-8. [DOI] [PubMed] [Google Scholar]
- 24.Du GP, Huang YF, Wang LQ, Ge M. Pentacam anterior segment analysis system to measure changes in arch height and its effect on postoperative visual outcome after ICL – a 3-year follow-up. Paper presented at: Proceedings of the 12th National Conference on Cornea and Ocular Surface Diseases and the Fourth National Conference on Corneal Refractive Surgery; May 01, 2012; Szechwan, China.
- 25.Wang J, Wang ZY, Xu T, Tao LM. Effect of crystalline lens rise on the vault after implantation of phakic posterior chamber implantable collamer lens. Int J Ophthalmol. 2021;21(2):377–381. [Google Scholar]
- 26.Qi MY, Chen Q, Zeng QY. The effect of the crystalline lens on central vault after implantable collamer lens implantation. J Refract Surg. 2017;33(8):519–523. doi: 10.3928/1081597X-20170602-02. [DOI] [PubMed] [Google Scholar]
- 27.Du GP, Huang YF, Wang LQ, et al. Changes in objective vault and effect on vision outcomes after implantable collamer lens implantation: 1-year follow-up. Eur J Ophthalmol. 2012;22(2):153–160. doi: 10.5301/EJO.2011.8359. [DOI] [PubMed] [Google Scholar]