Abstract
Background
To evaluate the clinical usefulness of thin-slice echo-planar imaging (EPI)-based diffusion-weighted imaging (DWI) with an on-console distortion correction technique, termed reverse encoding distortion correction DWI (RDC-DWI), in patients with non-functioning pituitary neuroendocrine tumor (PitNET)/pituitary adenoma.
Methods
Patients with non-functioning PitNET/pituitary adenoma who underwent 3-T RDC-DWI between December 2021 and September 2022 were retrospectively enrolled. Image quality was compared among RDC-DWI, DWI with correction for distortion induced by B0 inhomogeneity alone (B0-corrected-DWI), and original EPI-based DWI with anterior-posterior phase-encoding direction (AP-DWI). Susceptibility artifact, anatomical visualization of cranial nerves, overall tumor visualization, and visualization of cavernous sinus invasion were assessed qualitatively. Quantitative assessment of geometric distortion was performed by evaluation of anterior and posterior displacement between each DWI and the corresponding three-dimensional T2-weighted imaging. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and apparent diffusion coefficient values were measured.
Results
Sixty-four patients (age 70.8 ± 9.9 years [mean ± standard deviation]; 33 females) with non-functioning PitNET/pituitary adenoma were evaluated. In terms of susceptibility artifacts in the frontal and temporal lobes, visualization of left trigeminal nerve, overall tumor visualization, and anterior displacement, RDC-DWI performed the best and B0-corrected-DWI performed better than AP-DWI. The right oculomotor and right trigeminal nerves were better visualized by RDC-DWI than by B0-corrected-DWI and AP-DWI. Visualization of cavernous sinus invasion and posterior displacement were better by RDC-DWI and B0-corrected-DWI than by AP-DWI. SNR and CNR were the highest for RDC-DWI.
Conclusions
RDC-DWI achieved excellent image quality regarding susceptibility artifact, geometric distortion, and tumor visualization in patients with non-functioning PitNET/pituitary adenoma.
Relevance statement
RDC-DWI facilitates excellent visualization of the pituitary region and surrounding normal structures, and its on-console distortion correction technique is convenient. RDC-DWI can clearly depict cavernous sinus invasion of PitNET/pituitary adenoma even without contrast medium.
Key points
• RDC-DWI is an EPI-based DWI technique with a novel on-console distortion correction technique.
• RDC-DWI corrects distortion due to B0 field inhomogeneity and eddy current.
• We evaluated the usefulness of thin-slice RDC-DWI in non-functioning PitNET/pituitary adenoma.
• RDC-DWI exhibited excellent visualization in the pituitary region and surrounding structures.
• In addition, the on-console distortion correction of RDC-DWI is clinically convenient.
Graphical Abstract
Supplementary Information
The online version contains supplementary material available at 10.1186/s41747-024-00430-8.
Keywords: Artifacts, Diffusion magnetic resonance imaging, Echo-planar imaging, Neuroendocrine tumors, Pituitary neoplasms
Background
Diffusion-weighted imaging (DWI) is clinically useful for the evaluation of brain tumors [1]. In non-functioning pituitary neuroendocrine tumor (PitNET)/pituitary adenoma, DWI is valuable for assessing tumor consistency [2], prediction of Ki-67 expression [3], prediction of surgical outcome [4], and early detection of recurrence [5]. Echo-planar imaging-based DWI (EPI-DWI) is the most widely used DWI technique but is prone to geometric distortion along the phase-encoding direction due to B0 field inhomogeneity, especially in the pituitary region [6]. To overcome image degradation due to distortion in the pituitary region, various DWI techniques, including single-shot fast spin-echo DWI [7], line-scan DWI [8], periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER)/BLADE DWI [9, 10], reduced-field of view DWI [11], turbo-spin-echo (TSE) DWI [12, 13], and readout-segmented EPI-DWI [14, 15], have been developed. Nevertheless, the long acquisition time and low signal-to-noise ratio (SNR) of these techniques prevent their clinical application [10, 12, 16].
A popular strategy for correction of geometric distortion in EPI-DWI involves acquiring pairs of images with reverse phase-encoding directions, known as “blip-up blip-down” acquisition, to generate an undistorted image. This off-console technique, available as the “topup” tool in FMRIB Software Library−FSL (https://fsl.fmrib.ox.ac.uk/fsl/fslwiki/topup), can significantly improve extensive geometric distortion due to B0 field inhomogeneity using the b = 0 s/mm2 images with reverse phase-encoding directions [17, 18]. An off-console correction for distortion induced by eddy currents has also been reported [19]. However, off-console techniques such as these require external post-processing and are therefore complicated to use in clinical practice.
Distortion correction techniques for EPI-based DWI have recently been implemented in clinical MRI scanners [20]. A software package termed reverse encoding distortion correction DWI (RDC-DWI) has been developed [21, 22], which is a novel on-console correction technique for distortion due to B0 inhomogeneity and distortion induced by eddy currents. An on-console distortion correction technique of RDC-DWI is more convenient and clinically feasible compared with existing off-console distortion correction tools [17–20].
To the best of our knowledge, no previous study has evaluated the pituitary region on DWI employing an on-console distortion correction technique that uses pairs of images with reverse phase-encoding directions, such as RDC-DWI. This study aims to emphasize the clinical usefulness of thin-slice RDC-DWI by evaluating distortion, susceptibility artifact, and visualization of tumor and surrounding normal structures in patients with non-functioning PitNET/pituitary adenoma, in comparison with DWI with correction for distortion induced by B0 field inhomogeneity alone (B0-corrected-DWI) and original EPI-based DWI with anterior-posterior phase-encoding direction (AP-DWI).
Methods
Patient study
To investigate clinical usefulness of thin-slice RDC-DWI in the pituitary region, patients with non-functioning PitNET/pituitary adenoma were selected because non-functioning PitNET/pituitary adenoma is the most common pituitary tumor. Our institutional review board approved this retrospective study. Informed consent was waived due to its retrospective nature. A flowchart of the study design and patient inclusions is shown in Fig. 1. Eighty-nine consecutive patients (mean age, 68.1 ± 11.4 years [mean ± standard deviation]; 44 females) with non-functioning PitNET/pituitary adenoma who underwent MR imaging, including RDC-DWI with slice thickness of 1.2 mm and three-dimensional T2-weighted imaging (3D-T2WI), between December 2021 and September 2022 were included. Patients were excluded if they met the following exclusion criteria: (a) PitNET/pituitary adenoma with hemorrhage and (b) no apparent residual tumor after surgery.
Phantom study
To evaluate the reproducibility and accuracy of RDC-DWI, apparent diffusion coefficient (ADC) values were compared among the three DWIs (AP-DWI, B0-corrected-DWI, and RDC-DWI) and the theoretical values using a Quantitative Imaging Biomarkers Alliance−QIBA DWI phantom (Model 128, High Precision Devices, Inc., Boulder, USA). The phantom contains 13 vials filled with 30 mL of polymerpolyvinylpyrrolidone (PVP) in an aqueous solution, in six concentrations of PVP (0%, 10%, 20%, 30%, 40%, and 50%) corresponding to the theoretical ADC values specified by the manufacturer. The phantom was used with the temperature maintained between -0.20 and 0.20 °C (ice water bath).
MRI parameters
All studies were performed using a 3-T MR (Vantage Centurian, Canon Medical Systems Corporation, Otawara, Japan) with a 32-channel head coil. The same technical parameters were used for the phantom study and human study. The imaging parameters for thin-slice RDC-DWI and 3D-T2WI are shown in Table 1. In addition, the b = 0 s/mm2 images were scanned once for each phase-encoding direction and the b = 1,000 s/mm2 images were scanned eight times for each phase-encoding direction with simultaneous motion probing gradients (MPGs) in three orthogonal directions.
Table 1.
Parameter | RDC-DWI | 3D-T2WI |
---|---|---|
Image acquisition | Axial | Sagittal |
b value (s/mm2) | 0, 1,000 | − |
Repetition time (ms) | 3,850 | 2,800 |
Echo time (ms) | 65 | 80 |
Flip angle (degree) | 90 | 90 |
Field of view (mm2) | 220 × 220 | 220 × 220 |
Voxel size (mm3) | 0.7 × 0.7 × 1.2 | 0.34 × 0.34 × 0.8 |
Number of slices | 35 | 220 |
Band width (Hz/pixel) | 1,563 | 488 |
Number of acquisitions | 2 (1st, AP direction; 2nd, PA direction) | 1 |
Parallel imaging acceleration factor | 3 × | 3 × 2 |
Acquisition time (min:s) | 4:15 | 4:29 |
3D-T2WI Three-dimensional T2-weighted imaging, AP Anterior-posterior, DWI Diffusion-weighted imaging, PA Posterior-anterior, RDC-DWI Reverse encoding distortion correction DWI
RDC-DWI
RDC-DWI is an on-console distortion correction technique used in EPI-DWI. It corrects distortion induced by eddy currents by using the b = 1,000 s/mm2 images with reverse phase-encoding direction after application of MPGs in addition to correcting distortion due to B0 field inhomogeneity by using the b = 0 s/mm2 images with reverse phase-encoding directions.
Generation of AP-DWI, B0-corrected-DWI, and RDC-DWI
AP-DWI, B0-corrected-DWI, and RDC-DWI were generated on-console from the same raw data (Fig. 2). A shift map (B0) was generated using the b = 0 s/mm2 images with AP and PA phase-encoding directions. Shift maps (B0 + Eddy) were generated using the shift map (B0) and b = 1,000 s/mm2 images with AP and PA phase-encoding directions after application of MPGs. AP-DWI was generated using the b = 0 s/mm2 images with AP phase-encoding direction and b = 1,000 s/mm2 images with AP phase-encoding direction without distortion correction. B0-corrected-DWI was generated using the b = 0 s/mm2 images with AP and PA phase-encoding directions, b = 1,000 s/mm2 images with AP phase-encoding direction, and shift map (B0). RDC-DWI was generated using the b = 0 s/mm2 images with AP and PA phase-encoding directions, b = 1,000 s/mm2 images with AP and PA phase-encoding directions, and shift maps (B0 + Eddy) for three orthogonal MPG directions. In practice, the on-console RDC-DWI generation processes after image acquisition took approximately 2:10 min:s.
Image analysis: patient study
Qualitative evaluation
The image qualities of b = 1,000 s/mm2 images of the three DWIs (AP-DWI, B0-corrected-DWI, and RDC-DWI) were assessed for susceptibility artifact in the frontal and temporal lobes; anatomic visualization of the optic, oculomotor and trigeminal nerves; and overall tumor visualization using a 5-point Likert scale (0, very poor; 1, poor; 2, fair; 3, good; 4, excellent). Visualization of cavernous sinus invasion was assessed qualitatively using a 3-point Likert scale (0, poor; 1, fair; 2, good). Corresponding 3D-T2WI was used as the reference standard for assessing visualization of cavernous sinus invasion. The clinical course and surgical records were referred to when it was difficult to determine the extent of cavernous sinus invasion diagnosis using only 3D-T2WI. The criteria for image assessment are defined in Supplementary Figs. S1–S4. The evaluation was conducted by three board-certified neuroradiologists (S.Ok., S.Ot., and S.I., with 16, 13, and 8 years of experience in neuroradiology, respectively). The three DWIs were provided in random order, and each reader was blinded to the type of distortion correction method. The majority opinion of the raters was designated as the final value. If the three opinions differed, resolution was obtained by consensus.
Quantitative evaluation
Geometric distortion was determined as each of anterior and posterior displacement between each DWI and the corresponding 3D-T2WI in the sagittal plane using ITK-SNAP (http://www.itksnap.org/) (Fig. 3a–c) [23]. To calculate SNR, contrast-to-noise ratio (CNR), and ADC value, regions of interest (ROIs) were placed on the pons and on a solid portion of the PitNET/pituitary adenoma on the b = 1,000 s/mm2 image and the corresponding ADC map using ImageJ (https://imagej.nih.gov/ij/) (Supplementary Fig. S5). SNR was calculated as SIpons/SDpons, and CNR was calculated as (SIlesion − SIpons)/SDpons, where SIpons and SIlesion are mean signal intensity of the pons and the solid portion of the PitNET/pituitary adenoma, respectively, and SDpons is standard deviation of the pons. ROIs that including signal pileup artifacts were excluded for evaluating CNR and ADC value for PitNET/pituitary adenomas.
Evaluation of distortion and ROI measurements were performed by a board-certified radiologist (S.I.) and approved by another board-certified radiologist (S.Ok.).
Image analysis: phantom study
ROIs of area 212 mm2 were drawn on six different slices of the 13 vials with PVP concentrations of 0%, 10%, 20%, 30%, 40%, and 50% (Fig. 4a). The mean ADC values were measured on ADC maps for the three DWIs (AP-DWI, B0-corrected-DWI, and RDC-DWI) by a board-certified radiologist (S.I.) using ImageJ. Data of vials with 50% PVP were discarded because of air contamination (Fig. 4b–d).
Statistical analysis
Interrater reliability for image quality scores measured independently by the three radiologists was evaluated using Fleiss’ κ statistics [24]. The calculated κ statistic was interpreted as follows: ≤ 0.20, slight agreement; 0.21–0.40, fair agreement; 0.41–0.60, moderate agreement; 0.61–0.80, substantial agreement; and 0.81–1.00, almost perfect agreement. Image quality scores and measured displacements were compared among the three DWIs using the Friedman test followed by pairwise comparisons with Bonferroni correction because the results did not follow a normal distribution. SNR, CNR, and ADC values were also compared among the three DWIs using one-way repeated measures analysis of variance−ANOVA followed by pairwise comparisons with Bonferroni correction because these data followed a normal distribution. A p-value less than 0.050 was considered statistically significant.
Fleiss’ κ statistics and Friedman test were performed using RStudio Software (version 2022.12.0, RStudio, PBC, Boston, USA), and analysis of variance was performed using MedCalc version 20 (MedCalc Software, Ostend, Belgium).
Results
Patient study
Participants
Six patients with PitNET/pituitary adenoma with hemorrhage and 19 patients with no apparent residual tumor after surgery were excluded (Fig. 1). The remaining 64 patients (mean age, 70.8 ± 9.9 years [mean ± standard deviation]; 33 females) were evaluated and comprised 25 patients with unoperated clinically non-functioning PitNET/pituitary adenoma and 39 patients with residual non-functioning PitNET/pituitary adenoma after surgery. The demographics of all 64 participants are shown in Table 2.
Table 2.
Number | |
---|---|
Total patients | 64 |
Mean age ± standard deviation (years) | 70.8 ± 9.9 |
Male versus female | 31 versus 33 |
Unoperated clinically non-functioning PitNET/pituitary adenoma | 25 |
Residual non-functioning PitNET/pituitary adenoma after surgery | 39 |
Open surgery | 2 |
Endoscopic surgery | 32 |
Combined surgery | 5 |
Major medical history | |
72-year-old man with superficial hemosiderosis | 1 |
71-year-old woman with cerebral aneurysm clip | 1 |
65-year-old man with jugular foramen schwannoma | 1 |
Chiasma/optic nerve compression | 37 |
Chiasma compression | 27 |
Right optic nerve compression | 5 |
Left optic nerve compression | 5 |
PitNET Pituitary neuroendocrine tumor
Qualitative evaluation
Interrater reliability for qualitative evaluations was moderate to substantial agreement (κ = 0.59–0.78) (Supplementary Table S1). In 11 cases of cavernous sinus invasion, the diagnosis was determined with agreement of the three evaluators by also referring to the clinical course and surgical records because of difficulty using only 3D-T2WI.
Image quality scores and p-values are summarized in Table 3 and Supplementary Fig. S6. Scores for susceptibility artifacts in the frontal and temporal lobes were the best in RDC-DWI (median, 2.0; interquartile range, 2.0–2.0 and 2.0; 2.0–2.0, respectively) and significantly better in B0-corrected-DWI (2.0; 1.5–2.0 and 2.0; 2.0–2.0, respectively) than in AP-DWI (1.0; 1.0–1.0 and 1.0; 1.0–2.0, respectively). Scores for overall tumor visualization and anatomical visualization of the left trigeminal nerve were also the best in RDC-DWI (3.0; 3.0–4.0 and 4.0; 3.0–4.0, respectively) and were significantly better in B0-corrected-DWI (3.0; 3.0–3.0 and 3.0; 3.0–4.0, respectively) than in AP-DWI (2.0; 2.0–3.0 and 3.0; 3.0–4.0, respectively). Scores for anatomical visualization of the right oculomotor and right trigeminal nerves were significantly better in RDC-DWI (3.0; 2.0–3.0 and 4.0; 3.0–4.0, respectively) than in B0-corrected-DWI (3.0; 2.0–3.0 and 3.0; 3.0–3.0, respectively) and AP-DWI (3.0; 2.0–3.0 and 3.0; 3.0–3.0, respectively), but there was no significant difference between B0-corrected-DWI and AP-DWI. Scores for visualization of cavernous sinus invasion were significantly better in RDC-DWI (2.0; 2.0–2.0) and B0-corrected-DWI (2.0; 2.0–2.0) than in AP-DWI (1.0; 0.0–2.0), but there was no significant difference between RDC-DWI and B0-corrected-DWI. Score for anatomical visualization of the left optic nerve was better in B0-corrected-DWI (3.0; 3.0–3.0) than in AP-DWI (3.0; 3.0–3.0), but there was no significant difference between RDC-DWI (3.0; 3.0–3.0) and AP-DWI or between RDC-DWI and B0-corrected-DWI. Scores for anatomical visualization of the right optic and left oculomotor nerves showed no significant difference among the three DWIs. Representative images of the three DWIs are shown in Fig. 5 and Supplementary Fig. S7.
Table 3.
AP-DWI | B0-corrected-DWI | RDC-DWI | p-value | |
---|---|---|---|---|
Frontal artifact | 1.0 (1.0–1.0) | 2.0 (1.5–2.0) | 2.0 (2.0–2.0) | < 0.001a, b, c |
Temporal artifact | 1.0 (1.0–2.0) | 2.0 (2.0–2.0) | 2.0 (2.0–2.0) | < 0.001a, b, 0.001c |
Right optic nerve | 3.0 (3.0–3.0) | 3.0 (3.0–3.0) | 3.0 (3.0–3.0) | 0.220a, 0.290b, 1.000c |
Left optic nerve | 3.0 (3.0–3.0) | 3.0 (3.0–3.0) | 3.0 (3.0–3.0) | 0.003a, 0.093b, 1.000c |
Right oculomotor nerve | 3.0 (2.0–3.0) | 3.0 (2.0–3.0) | 3.0 (2.0–3.0) | 1.000a, 0.037b, 0.002c |
Left oculomotor nerve | 3.0 (2.0–3.0) | 3.0 (2.0–3.0) | 3.0 (3.0–3.0) | 1.000a, 0.146b, 0.071c |
Right trigeminal nerve | 3.0 (3.0–3.0) | 3.0 (3.0–3.0) | 4.0 (3.0–4.0) | 0.085a, < 0.001b, c |
Left trigeminal nerve | 3.0 (3.0–4.0) | 3.0 (3.0–4.0) | 4.0 (3.0–4.0) | 0.013a, < 0.001b, c |
Cavernous sinus invasion | 1.0 (0.0–2.0) | 2.0 (2.0–2.0) | 2.0 (2.0–2.0) | < 0.001a, b, 0.700c |
Tumor visualization | 2.0 (2.0–3.0) | 3.0 (3.0–3.0) | 3.0 (3.0–4.0) | < 0.001a, b, c |
Data are presented as the median (interquartile range) score. Note that susceptibility artifact in the frontal and temporal lobes; anatomic visualization of the optic, oculomotor, and trigeminal nerves; and overall tumor visualization are assessed using a 5-point Likert scale (0, very poor; 1, poor; 2, fair; 3, good; 4, excellent), and visualization of cavernous sinus invasion is assessed using a 3-point Likert scale (0, poor; 1, fair; 2, good)
aAP-DWI versus B0-corrected-DWI
bAP-DWI versus RDC-DWI
cB0-corrected-DWI versus RDC-DWI
AP Anterior-posterior, DWI Diffusion-weighted imaging, RDC-DWI Reverse encoding distortion correction DWI
Quantitative evaluation
Anterior sagittal displacement due to geometric distortion was the least in RDC-DWI (median, 2.8 mm; interquartile range, 1.8–4.7 mm) (p < 0.001) and significantly lower in B0-corrected-DWI (3.8 mm; 2.4–5.5 mm) than in AP-DWI (7.4 mm; 4.3–10.5 mm) (p < 0.001). Posterior sagittal displacement was significantly lower in RDC-DWI (2.0 mm; 1.4–3.5 mm) and B0-corrected-DWI (2.6 mm; 1.6–3.9 mm) than in AP-DWI (3.8 mm; 2.3–6.2 mm) (p < 0.001). The measured value of posterior sagittal displacement was the least in RDC-DWI. There was no significant difference between RDC-DWI and B0-corrected-DWI (p = 0.065) (Fig. 3d).
Table 4 lists the SNR, CNR, and ADC values. SNR was the highest in RDC-DWI and was significantly higher in AP-DWI (p < 0.001) than in B0-corrected-DWI (p = 0.001). After excluding 26 patients in whom ROIs contained signal pileup artifacts, CNR and ADC values of the solid portion of PitNET/pituitary adenomas were calculated in 38 patients. Among the three DWIs, CNR was the highest in RDC-DWI, and CNR was significantly higher in RDC-DWI than in B0-corrected-DWI (p < 0.001). There was no statistically significant difference in CNR between B0-corrected-DWI and AP-DWI (p = 1.000), or between RDC-DWI and AP-DWI (p = 0.094). There was no significant difference in ADC values of the pons or the solid portion of PitNET/pituitary adenomas among the three DWIs.
Table 4.
AP-DWI | B0-corrected-DWI | RDC-DWI | p-value | |
---|---|---|---|---|
SNR | 11.7 ± 2.4 | 11.1 ± 1.9 | 14.5 ± 3.2 | 0.001a, < 0.001b, c |
CNR | 4.5 ± 4.4 | 4.8 ± 2.4 | 6.1 ± 2.8 | 1.000a, 0.094b, < 0.001c |
ADCpons (×10-6 mm2/s) | 679.5 ± 46.2 | 674.5 ± 47.8 | 679.4 ± 35.9 | 0.779a, 1.000b, 0.959c |
ADCPitNET/pituitary adenoma (×10-6 mm2/s) | 695.7 ± 141.8 | 701.3 ± 138.4 | 711.9 ± 134.5 | 1.000a, 0.359b, 0.309c |
Data are presented as the mean ± standard deviation
aAP-DWI versus B0-corrected-DWI
bAP-DWI versus RDC-DWI
cB0-corrected-DWI versus RDC-DWI
ADC Apparent diffusion coefficient, ADCPitNET/pituitary adenoma ADC value for the solid portion of pituitary neuroendocrine tumor (PitNET)/pituitary adenoma, ADCpons ADC value for the pons, AP Anterior-posterior, CNR Contrast-to-noise ratio, DWI Diffusion-weighted imaging, SNR Signal-to-noise ratio, PitNET Pituitary neuroendocrine tumor, RDC-DWI Reverse encoding distortion correction DWI
Phantom study
The b = 1,000 s/mm2 images on RDC-DWI showed the fewest artifacts and least distortion among the three DWIs (AP-DWI, B0-corrected-DWI, and RDC-DWI) (Fig. 4b, c, and d). ADC values measured in each of the five PVP concentrations showed no significant difference among the three DWIs for concentrations of 0%, 30%, and 40% (Table 5). Statistically significant differences in ADC values were found at PVP concentrations of 10% and 20%; however, ADC values at these two concentrations on RDC-DWI were the closest to the theoretical ADC values provided by the manufacturer [25].
Table 5.
%PVP | AP-DWI | B0-corrected-DWI | RDC-DWI | Ground truth | p-value |
---|---|---|---|---|---|
0% | 1,128.4 ± 28.3 | 1,126.3 ± 30.4 | 1,123.4 ± 18.6 | 1,090 ± 9 | 0.763a, 0.196b, 1.000c |
10% | 848.3 ± 2.9 | 850.1 ± 8.1 | 832.0 ± 1.2 | 825 ± 8 | 1.000a, < 0.001b, c |
20% | 619.3 ± 9.1 | 624.7 ± 10.4 | 606.1 ± 10.6 | 598 ± 7 | 0.002a, < 0.001b, c |
30% | 427.8 ± 19.1 | 430.9 ± 23.4 | 428.4 ± 19.1 | 396 ± 5 | 0.299a, 1.000b, 0.674c |
40% | 256.0 ± 13.49 | 264.2 ± 23.4 | 259.3 ± 9.8 | 237 ± 6 | 0.165a, 0.057b, 0.896c |
Data are presented as the mean ± standard deviation. The 50% PVP data were discarded because of air contamination
aAP-DWI versus B0-corrected-DWI
bAP-DWI versus RDC-DWI
cB0-corrected-DWI versus RDC-DWI
ADC Apparent diffusion coefficient, AP Anterior-posterior, DWI Diffusion-weighted imaging, PVP Polymerpolyvinylpyrrolidone, RDC-DWI Reverse encoding distortion correction DWI
Discussion
The present study demonstrated that thin-slice EPI-based RDC-DWI with a novel on-console distortion correction technique enabled excellent visualization of the pituitary region and surrounding normal structures. Overall, RDC-DWI outperformed the other DWIs (B0-corrected-DWI and AP-DWI), exhibiting the best image quality in terms of susceptibility artifacts, geometric distortion, visualization of cranial nerves and cavernous sinus invasion, and overall tumor visualization. The excellent visualization on RDC-DWI results from its characteristic image acquisition and distortion correction. The on-console distortion correction technique of RDC-DWI is based on the b = 1,000 s/mm2 images with reverse phase-encoding directions after application of MPGs in three orthogonal directions, in addition to the b = 0 s/mm2 images with reverse phase-encoding directions. Therefore, RDC-DWI achieved combined correction of distortion due to B0 field inhomogeneity and of eddy current-induced distortion. Moreover, the on-console distortion correction of RDC-DWI is more feasible in clinical practice than the existing off-console distortion correction tools [17–20].
In addition to reduction in susceptibility artifacts and distortion, RDC-DWI produced high-contrast images and had the highest SNR among the three DWIs (AP-DWI, B0-corrected-DWI, and RDC-DWI) even with slice thickness of 1.2 mm, which was achieved by two-fold number of excitations, combining images with AP and PA phase-encoding directions. Measurement of precise SNR in DWI after application of distortion correction is complicated; in the present study, however, RDC-DWI provided images with high SNR, whereas in previous studies, no DWI technique has achieved SNR higher than that of conventional EPI-DWI [8, 10, 12, 26]. Furthermore, compared with previously reported DWI techniques with distortion reduction, slice thickness of 1.2 mm in RDC-DWI resulted in higher spatial resolution with distortion correction and enabled observation from multiple directions using multiplanar reconstruction of the pituitary lesion and surrounding normal structures [7, 8, 10–13].
We also assessed the accuracy of RDC-DWI for visualizing cavernous sinus invasion by PitNET/pituitary adenoma, which has not been evaluated previously on DWI. Scores for visualization of cavernous sinus invasion were significantly better for RDC-DWI and B0-corrected-DWI than for AP-DWI. Based on our results, RDC-DWI and B0-corrected-DWI might visualize cavernous sinus invasion well even without contrast medium [27–30]. With regard to postoperative cases in particular, RDC-DWI and B0-corrected-DWI can contribute to early detection and visualization of residual PitNET/pituitary adenoma in the cavernous sinus [5, 31]. Occasionally, residual PitNET/pituitary adenoma in the cavernous sinus can be difficult to assess on T2-weighted or contrast-enhanced T1-weighted images after surgery due to postoperative change [32]; however, RDC-DWI is better able to visualize residual PitNET/pituitary adenoma compared with other sequences because the venous pool of the cavernous sinus appears as a signal void on DWI.
Among the three DWIs, RDC-DWI showed the best visualization of both the trigeminal and right oculomotor nerves. This finding indicates that RDC-DWI had fewer susceptibility artifacts than B0-corrected-DWI, for the reason that susceptibility artifact is strongest where the cranial nerves pass near the paranasal sinuses. Anatomical visualization of the optic nerves was not superior on RDC-DWI, probably because the optic chiasma or optic nerves were compressed in more than half of the studied cases, and there might have been minimal air space between the optic nerves and the PitNET/pituitary adenomas.
The acquisition time for thin-slice RDC-DWI was 4:15 min:s. Scan times were also long in previously reported advanced DWI techniques for the pituitary region, such as line-scan DWI [8], periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER)/BLADE DWI [9, 10], turbo-spin-echo (TSE) DWI [12, 13], and readout-segmented EPI-DWI [14, 15], although parameters such as slice thickness, number of slices, field of view, and number of excitations were different from RDC-DWI. Scan time is longer for RDC-DWI than conventional EPI-DWI but is clinically acceptable considering the clinical usefulness of RDC-DWI in the pituitary region, as shown in this study. Distortion correction techniques with deep learning reconstruction might have potential to reduce scan time and improve image quality, although its usefulness and advantages remain unclear [33].
Our study has several limitations. First, we included images in both pre- and post-operative states. RDC-DWI seemed to have the best visualization both preoperatively and postoperatively, although we did not conduct a comparison. Second, we used 3D-T2WI as the reference in evaluating visualization of cavernous sinus invasion. Cavernous sinus invasion is generally diagnosed with contrast-enhanced T1WI [27, 28, 30, 32]; however, many of the present patients did not undergo contrast-enhanced MRI. We referred to the clinical course and surgical records when it was difficult to determine the diagnosis using only 3D-T2WI. Finally, we included relatively considerable number of patients; however, more pre- and postoperative patients may be needed to generalize our results. Further evaluation in other pituitary lesions is also desirable. In addition, in our phantom study, the 50% PVP data were discarded due to air contamination. The other phantom study on RDC-DWI seems more reliable [22].
In conclusion, thin-slice RDC-DWI provided excellent image quality in terms of distortion, susceptibility artifact, and visualization of tumor and surrounding normal structures. RDC-DWI achieved accurate and high-contrast visualization with high spatial resolution, using an on-console combined correction for distortion due to B0 field inhomogeneity and eddy current-induced distortion. Thin-slice RDC-DWI facilitates visualization of the pituitary region and surrounding normal structures even without contrast medium.
Supplementary Information
Acknowledgements
We are grateful to Mr. Yuichiro Sano and Mr. Nobuyasu Ichinose for their technical support.
We did not use LLMs for our manuscript.
Abbreviations
- 3D-T2WI
Three-dimensional T2-weighted imaging
- ADC
Apparent diffusion coefficient
- AP
Anterior-posterior
- CNR
Contrast-to-noise ratio
- DWI
Diffusion-weighted imaging
- EPI
Echo-planar imaging
- MPG
Motion probing gradient
- PA
Posterior-anterior
- PitNET
Pituitary neuroendocrine tumor
- PVP
Polymerpolyvinylpyrrolidone
- RDC-DWI
Reverse encoding distortion correction DWI
- ROI
Region of interest
- SNR
Signal-to-noise ratio
Authors’ contributions
Study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting, SI, SOk, YF; approval of final version of submitted manuscript, all authors; agrees to ensure any questions related to the work are appropriately resolved, all authors; literature research, SI, SOk; clinical studies, SI, SOk; statistical analysis, SI, SOk; and manuscript editing, all authors.
Funding
This work was supported by JSPS KAKENHI Grant Numbers 21K20834, 21K15826, and 22K07746; by ISHIZUE 2023 of Kyoto University and by The Kyoto University Foundation.
Availability of data and materials
Data generated or analyzed during the study are available from the corresponding author by request.
Declarations
Ethics approval and consent to participate
Institutional Review Board approval was obtained (Kyoto University Graduate School and Faculty of Medicine, Ethics Committee, reference number R3706). Written informed consent was not required for this study because of its retrospective nature. Written informed consent was waived by the Institutional Review Board.
Consent for publication
Not applicable.
Competing interests
Hiroki Kondo and Rimika Imai are the employees of Canon Medical Systems Corporation. Koji Fujimoto, Tsuneo Saga, Kanae Kawai Miyake, and Hitomi Numamoto belong to the endowed chair of Industry-Academia Collaboration Project between Kyoto University and Canon Medical Systems Corporation. The remaining authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Provenzale JM, Mukundan S, Barboriak DP. Diffusion-weighted and perfusion MR imaging for brain tumor characterization and assessment of treatment response. Radiology. 2006;239:632–649. doi: 10.1148/radiol.2393042031. [DOI] [PubMed] [Google Scholar]
- 2.Pierallini A, Caramia F, Falcone C, et al. Pituitary macroadenomas: preoperative evaluation of consistency with diffusion-weighted MR imaging--initial experience. Radiology. 2006;239:223–231. doi: 10.1148/radiol.2383042204. [DOI] [PubMed] [Google Scholar]
- 3.Tamrazi B, Pekmezci M, Aboian M, Tihan T, Glastonbury CM. Apparent diffusion coefficient and pituitary macroadenomas: pre-operative assessment of tumor atypia. Pituitary. 2017;20:195–200. doi: 10.1007/s11102-016-0759-5. [DOI] [PubMed] [Google Scholar]
- 4.Boxerman JL, Rogg JM, Donahue JE, Machan JT, Goldman MA, Doberstein CE. Preoperative MRI evaluation of pituitary macroadenoma: imaging features predictive of successful transsphenoidal surgery. AJR Am J Roentgenol. 2010;195:720–728. doi: 10.2214/AJR.09.4128. [DOI] [PubMed] [Google Scholar]
- 5.Hassan HA, Bessar MA, Herzallah IR, Laury AM, Arnaout MM, Basha MAA. Diagnostic value of early postoperative MRI and diffusion-weighted imaging following trans-sphenoidal resection of non-functioning pituitary macroadenomas. Clin Radiol. 2018;73:535–541. doi: 10.1016/j.crad.2017.12.007. [DOI] [PubMed] [Google Scholar]
- 6.Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med. 1995;34:65–73. doi: 10.1002/mrm.1910340111. [DOI] [PubMed] [Google Scholar]
- 7.Kunii N, Abe T, Kawamo M, Tanioka D, Izumiyama H, Moritani T. Rathke’s cleft cysts: differentiation from other cystic lesions in the pituitary fossa by use of single-shot fast spin-echo diffusion-weighted MR imaging. Acta Neurochir (Wien) 2007;149:759–769. doi: 10.1007/s00701-007-1234-x. [DOI] [PubMed] [Google Scholar]
- 8.Suzuki C, Maeda M, Hori K, et al. Apparent diffusion coefficient of pituitary macroadenoma evaluated with line-scan diffusion-weighted imaging. J Neuroradiol. 2007;34:228–235. doi: 10.1016/j.neurad.2007.06.007. [DOI] [PubMed] [Google Scholar]
- 9.Mahmoud OM, Tominaga A, Amatya VJ, et al. Role of PROPELLER diffusion-weighted imaging and apparent diffusion coefficient in the evaluation of pituitary adenomas. Eur J Radiol. 2011;80:412–417. doi: 10.1016/j.ejrad.2010.05.023. [DOI] [PubMed] [Google Scholar]
- 10.Yiping L, Hui L, Kun Z, Daoying G, Bo Y. Diffusion-weighted imaging of the sellar region: a comparison study of BLADE and single-shot echo planar imaging sequences. Eur J Radiol. 2014;83:1239–1244. doi: 10.1016/j.ejrad.2014.03.011. [DOI] [PubMed] [Google Scholar]
- 11.Wang M, Liu H, Wei X, et al. Application of reduced-FOV diffusion-weighted imaging in evaluation of normal pituitary glands and pituitary macroadenomas. AJNR Am J Neuroradiol. 2018;39:1499–1504. doi: 10.3174/ajnr.A5735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Khant ZA, Azuma M, Kadota Y, et al. Evaluation of pituitary structures and lesions with turbo spin-echo diffusion-weighted imaging. J Neurol Sci. 2019;405:116390. doi: 10.1016/j.jns.2019.07.008. [DOI] [PubMed] [Google Scholar]
- 13.Kamimura K, Nakajo M, Fukukura Y, et al. Intravoxel incoherent motion in normal pituitary gland: initial study with turbo spin-echo diffusion-weighted imaging. AJNR Am J Neuroradiol. 2016;37:2328–2333. doi: 10.3174/ajnr.A4930. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Su CQ, Zhang X, Pan T, et al. Texture analysis of high b-value diffusion-weighted imaging for evaluating consistency of pituitary macroadenomas. J Magn Reson Imaging. 2020;51:1507–1513. doi: 10.1002/jmri.26941. [DOI] [PubMed] [Google Scholar]
- 15.Rutland JW, Loewenstern J, Ranti D, et al. Analysis of 7-tesla diffusion-weighted imaging in the prediction of pituitary macroadenoma consistency. J Neurosurg. 2020;134:771–779. doi: 10.3171/2019.12.JNS192940. [DOI] [PubMed] [Google Scholar]
- 16.Fu Q, Kong XC, Liu DX, et al. Clinical comparison of single-shot EPI, readout-segmented EPI and TGSE-BLADE for diffusion-weighted imaging of cerebellopontine angle tumors on 3 tesla. Magn Reson Imaging. 2021;84:76–83. doi: 10.1016/j.mri.2021.09.009. [DOI] [PubMed] [Google Scholar]
- 17.Andersson JL, Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage. 2003;20:870–888. doi: 10.1016/S1053-8119(03)00336-7. [DOI] [PubMed] [Google Scholar]
- 18.Morgan PS, Bowtell RW, McIntyre DJ, Worthington BS. Correction of spatial distortion in EPI due to inhomogeneous static magnetic fields using the reversed gradient method. J Magn Reson Imaging. 2004;19:499–507. doi: 10.1002/jmri.20032. [DOI] [PubMed] [Google Scholar]
- 19.Andersson JLR, Sotiropoulos SN. An integrated approach to correction for off-resonance effects and subject movement in diffusion MR imaging. Neuroimage. 2016;125:1063–1078. doi: 10.1016/j.neuroimage.2015.10.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Sjöholm T, Kullberg J, Strand R, Engström M, Ahlström H, Malmberg F. Improved geometric accuracy of whole body diffusion-weighted imaging at 1.5T and 3T using reverse polarity gradients. Sci Rep. 2022;12:11605. doi: 10.1038/s41598-022-15872-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Ueda T, Ohno Y, Shinohara M, et al. Reverse encoding distortion correction for diffusion-weighted MRI: efficacy for improving image quality and ADC evaluation for differentiating malignant from benign areas in suspected prostatic cancer patients. Eur J Radiol. 2023;162:110764. doi: 10.1016/j.ejrad.2023.110764. [DOI] [PubMed] [Google Scholar]
- 22.Numamoto H, Fujimoto K, Miyake KK et al (2023) Evaluating reproducibility of the ADC and distortion in diffusion-weighted imaging (DWI) with reverse encoding distortion correction (RDC). Magn Reson Med Sci. 10.2463/mrms.mp.2023-0102 [DOI] [PubMed]
- 23.Yushkevich PA, Piven J, Hazlett HC, et al. User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006;31:1116–1128. doi: 10.1016/j.neuroimage.2006.01.015. [DOI] [PubMed] [Google Scholar]
- 24.Fleiss J. Measuring nominal scale agreement among many raters. Psychol Bull. 1971;76:378–382. doi: 10.1037/h0031619. [DOI] [Google Scholar]
- 25.Keenan KE, Carnicka S, Gottlieb SC, Stupic KF (2017) Assessing changes in MRI measurands incurred in a scanner upgrade: is my study comprised?. In: Proceedings of the ISMRM 25th Annual Meeting, Honolulu, 22−24 April 2017. P. 3922
- 26.Okuchi S, Fushimi Y, Yoshida K, et al. Comparison of TGSE-BLADE DWI, RESOLVE DWI, and SS-EPI DWI in healthy volunteers and patients after cerebral aneurysm clipping. Sci Rep. 2022;12:17689. doi: 10.1038/s41598-022-22760-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Knosp E, Steiner E, Kitz K, Matula C. Pituitary adenomas with invasion of the cavernous sinus space: a magnetic resonance imaging classification compared with surgical findings. Neurosurgery. 1993;33:610–617. doi: 10.1227/00006123-199310000-00008. [DOI] [PubMed] [Google Scholar]
- 28.Cottier JP, Destrieux C, Brunereau L, et al. Cavernous sinus invasion by pituitary adenoma: MR imaging. Radiology. 2000;215:463–469. doi: 10.1148/radiology.215.2.r00ap18463. [DOI] [PubMed] [Google Scholar]
- 29.Cao L, Chen H, Hong J, Ma M, Zhong Q, Wang S. Magnetic resonance imaging appearance of the medial wall of the cavernous sinus for the assessment of cavernous sinus invasion by pituitary adenomas. J Neuroradiol. 2013;40:245–251. doi: 10.1016/j.neurad.2013.06.003. [DOI] [PubMed] [Google Scholar]
- 30.Micko AS, Wöhrer A, Wolfsberger S, Knosp E. Invasion of the cavernous sinus space in pituitary adenomas: endoscopic verification and its correlation with an MRI-based classification. J Neurosurg. 2015;122:803–811. doi: 10.3171/2014.12.JNS141083. [DOI] [PubMed] [Google Scholar]
- 31.Ko CC, Chen TY, Lim SW, Kuo YT, Wu TC, Chen JH. Prediction of recurrence in solid nonfunctioning pituitary macroadenomas: additional benefits of diffusion-weighted MR imaging. J Neurosurg. 2019;132:351–359. doi: 10.3171/2018.10.JNS181783. [DOI] [PubMed] [Google Scholar]
- 32.Kim M, Kim HS, Kim HJ, et al. Thin-slice pituitary MRI with deep learning-based reconstruction: diagnostic performance in a postoperative setting. Radiology. 2021;298:114–122. doi: 10.1148/radiol.2020200723. [DOI] [PubMed] [Google Scholar]
- 33.Hu Z, Wang Y, Zhang Z, et al. Distortion correction of single-shot EPI enabled by deep-learning. Neuroimage. 2020;221:117170. doi: 10.1016/j.neuroimage.2020.117170. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data generated or analyzed during the study are available from the corresponding author by request.