. 2024 Mar 6;7:278. doi: 10.1038/s42003-024-05973-5

Dynamic contrast optical coherence tomography (DyC-OCT) for label-free live cell imaging

Chao Ren 1,2, Senyue Hao 3, Fei Wang 1, Abigail Matt 1, Marcello Magri Amaral 1,4, Daniel Yang 5, Leyao Wang 5, Chao Zhou 1,2,3,
PMCID: PMC10918170  PMID: 38448627

Abstract

Dynamic contrast optical coherence tomography (DyC-OCT), an emerging imaging method, utilizes fluctuation patterns in OCT signals to enhance contrast, thereby enabling non-invasive label-free volumetric live cell imaging. In this mini review, we explain the core concepts behind DyC-OCT image formation and its system configurations, serving as practical guidance for future DyC-OCT users. Subsequently, we explore its applications in delivering high-quality, contrast-enhanced images of cellular morphology, as well as in monitoring changes in cellular activity/viability assay experiments.

Subject terms: Optical imaging, Image processing


Dynamic contrast optical coherence tomography (DyC-OCT), an emerging imaging method, utilizes fluctuation patterns in OCT signals to enhance contrast, thereby enabling non-invasive label-free volumetric live cell imaging.

Introduction

Optical coherence tomography (OCT)1, is an optical imaging modality which offers millimeter-depth volumetric optical biopsies of live biological samples within seconds2,3. OCT identifies depth-resolved back-scattered light, which arises from the inherent interaction of light with biological tissue, to generate label-free microtomographic images of biological specimens. Despite the prevailing focus of current OCT applications on structural and tissue-level imaging, its micron-level spatial resolution holds the potential for live cell imaging. However, the primary challenge lies in the low contrast resulting from the similarities in optical properties between cellular and subcellular features.

To overcome this problem, researchers have developed various contrast enhancement strategies. Contrast agents, such as microspheres4, magnetic microparticles5, and gold nanoparticles68, have been injected into samples to enhance their scattering and make targeted structures of interest stand out. An alternative label-free approach for enhancing cellular features involves the analysis of temporal OCT signals. The signal exhibits fluctuations corresponding to variations in the optical back-scattering properties in each voxel of the sample. These fluctuations reflect movements related to cellular motion, metabolic processes, and other cellular activities. OCT-angiography (OCT-A)911 and speckle variance OCT (SV-OCT)1214 first applied this label-free contrast analysis to highlight blood vessels and generate microvascular maps. In repeated images of the same tissue, the differences between repeated scans can distinguish areas where blood flows. Recently, this idea has been extended to live cell imaging, using a technique called dynamic contrast OCT (DyC-OCT)15. DyC-OCT highlights cell clusters or even individual cells to enhance the contrast of conventional OCT images, revealing the cellular morphology and dynamic information behind them.

This mini-review will cover the principles, implementations, and applications of DyC-OCT. We first provide a brief explanation of the sources of the DyC-OCT signal and detailed data processing procedures. Then, we summarize practical recommendations for data collection and selection of the most suitable system configuration. The scope of our review will extend to the most recent implementations of DyC-OCT, with an emphasis on advancements in morphological visualization and potential activity/viability assays. Furthermore, we will discuss the current challenges of DyC-OCT and propose potential solutions.

Principles of DyC-OCT

A healthy live cell experiences a variety of activities during its lifecycle, from inner metabolic processes, such as ATP-consuming process15,16, to outer cell movements. These ever-changing dynamics alter the detected OCT signal. The movement of scattering particles changes the scattering potential of any one voxel in the OCT volumetric data, causing fluctuations in the OCT signal intensity. By conducting repeated OCT scans, time-varying signal fluctuations are captured and then processed using algorithms to produce color-coded maps depicting DyC-OCT motion patterns. DyC-OCT is distinguished from traditional scattering-based OCT through its emphasis on dynamic fluctuations: the motions of microstructures are accented against the motionless regions in the OCT images, enhancing the image contrast.

Pattern extraction algorithms employed in DyC-OCT

Various algorithms have been developed to analyze OCT signal fluctuation patterns. Initial approaches, such as the logarithmic intensity variation (LIV)1719 and standard deviation (STD)15,2022 algorithms, utilize the principles in SV-OCT to measure the amplitudes of OCT signal fluctuations. These methods highlight areas of high activity in tissue structures, which exhibit greater fluctuation amplitudes, while reducing the visibility of stationary structures in DyC-OCT colormaps. As an example, Fig. 1a illustrates repeated OCT frames of an airway organoid, a human stem cell-derived 3D assembly that simulates respiratory tissues. The data was obtained using a spectral-domain OCT system, and the organoids were prepared in accordance with the STEMCELL airway organoid culturing protocol23. Figure 1.a1 shows the original OCT images, with repeated cross-sectional OCT scans at a fixed location. Each voxel has a temporal series showing its intensity changes. Figure 1.a2 is the zoom-in region marked by the red rectangle in Fig. 1.a1. For the three voxels highlighted in Fig. 1.a2, the OCT signal fluctuation patterns (e.g., amplitude, frequency) are readily distinguishable (Fig. 1.a3–a5). The STD-based DyC-OCT image (Fig. 1.b1) is generated by applying a jet colormap based on the STD value of each pixel. Figure 1.b2 is the same zoom-in region as Fig. 1.a2. Compared to original OCT images, STD-based DyC-OCT colormaps more effectively differentiate structures exhibiting active movements, thereby enhancing image contrast. Although effective, the STD algorithm fails to discern the speed of tissue dynamics. A subsequently development, OCT correlation decay speed (OCDS) analysis17,18, overcomes this by measuring the temporal dynamic motion pattern through a temporal decorrelation function (Fig. 1.c). Figure 1.c1–c3 display OCDS measurements from the three pixels indicated in Fig. 1.a2. By measuring the slope of each pixel’s autocorrelation curve (e.g., OCDS value), a direct quantitative assessment of motion dynamics is obtained. Colormaps can be generated using OCDS values (Fig. 1.c4–c5). OCDS-based DyC-OCT analysis can effectively distinguish tissue regions with varying dynamic speeds, offering direct motion dynamics measurements.

Fig. 1. Illustrations of the three DyC-OCT processing algorithms.

Fig. 1

a Repeated OCT frames of airway organoids and OCT time series of three pixels with different dynamics patterns. a1 Repeated OCT frames of airway organoids. a2 Zoomed-in view of the red ROI in (a1). a3a5 Normalized time series of OCT signals of the three regions labeled in (a2): (a3) a stationary region, (a4) an intensively moving region, and (a5) a mildly moving region. b STD based DyC-OCT colormap processing. b1 STD based color-coded DyC-OCT image of an airway organoid. b2 Zoomed-in view of the yellow ROI in (b1). c OCDS based DyC-OCT colormap processing. c1c3 Autocorrelation by time delay of OCT signals in (a3a5); the OCDSs are shown by the slope of the fitted lines of autocorrelations. c4 OCDS-based color-coded DyC-OCT image of an airway organoid. c5 Zoomed-in view of the yellow ROI in (c4). d PSD based DyC-OCT colormap processing. d1d3 PSDs of OCT signals in (a3a5); the green-shaded area represents 0.8–4 Hz, while red-shaded area covers 4–20 Hz. d4 PSD-based color-coded DyC-OCT image of an airway organoid. d5 Zoomed-in view of the yellow ROI in (d4). Cov: covariance; Var: variance; FFTt: fast Fourier transform on time dimension. I=IOCTx,z,ti,I=IOCT(x,z,ti+τj). Scale bar: 50 μm.

Both the STD and OCDS algorithms enhance dynamic contrast in OCT, requiring only around 10 repeated frames15,17,18. This facilitates rapid image acquisition and minimizes memory costs and processing time. An additional algorithm, utilizing fast Fourier transform (FFT) to analyze the power spectral density (PSD)16,2428 of each pixel, offers more comprehensive insights into tissue dynamics. Figure 1.d1–d3 demonstrate how PSDs of OCT signals from the pixels shown in Fig. 1.a2 distinctly reveal variations in fluctuation frequencies and intensities. To create a colormap, the integrated intensity within different frequency bands is color-coded2527; for example, signals in the 0.8–4 Hz range are shown in green, while signals in the 4–20 Hz range are shown in red. Furthermore, PSD can highlight dominant frequencies using a continuous colormap29,30. These colormaps can be overlaid on structural OCT images to provide a co-registered display of tissue dynamics (Fig. 1.d4–d5). However, while PSD analysis concurrently extracts dynamic signals across frequencies, it requires over 100 repeated OCT scans for accurate PSD measurements, substantially increasing the image acquisition time16,25,26. Supplementary Table 1 compares these three DyC-OCT processing algorithms on time complexity, extracted parameters, minimum repeated frames, advantages and drawbacks. Drawing upon established OCT technology, DyC-OCT processing algorithms can identify tissue structures and retrieve both the intensity and speed of their motion patterns. When tailoring DyC-OCT to examine specific biological samples or meet experimental aims, the OCT system setup and acquisition protocol are crucial. We will discuss the system requirements in the next section.

OCT modalities employed for DyC-OCT

The performance of DyC-OCT is heavily dependent on the choice of OCT hardware. Three types of OCT systems have been applied in DyC-OCT: full-field OCT (FF-OCT), swept-source OCT (SS-OCT), and spectral-domain OCT (SD-OCT). Different OCT modalities present trade-offs in imaging speed, resolution, and depth of penetration, making the selection of an appropriate imaging system and image acquisition parameters critical for optimal performance in specific applications.

FF-OCT31 generates OCT images from the en-face view using a two-dimensional (2D) camera. It offers a high isotropic resolution (~1 μm) to enable cellular imaging, but its penetration depth in scattering tissue is lower than that of the other two OCT modalities. As a result, FF-OCT is appropriate for samples with limited thickness (~300 μm) demanding superior resolutions, such as the retina organoids and explants22,24,32 or sparse cells28,30.

SD-OCT and SS-OCT generate cross-sectional OCT images (B-scans) with each acquisition. SD-OCT uses a broadband light source in combination with a spectrometer to achieve a higher axial resolution (~1-2 μm) with an imaging depth of around 1 mm. SD-OCT is well suited for imaging thick tissues with layered structures or depth-resolved features, such as the tongue31, trachea16, and esophagus25. Using a high magnification objective lens, SD-OCT can also achieve ~1 μm lateral resolution with a reduced imaging depth of field27. SS-OCT33 uses a tunable broadband laser source, a high-speed balanced detector and a data acquisition card to acquire cross-sectional B-scan OCT frames. SS-OCT has a high imaging speed and detection depth that is suitable for imaging thick samples, such as tumor spheroids17,18. However, SS-OCT suffers from a low axial resolution of around 10 μm, limiting its ability to image cellular or subcellular structures, such as organelles and nuclei.

Table 1 summarizes the recent literature on DyC-OCT employing different OCT modalities, including information on the OCT system configuration (e.g., system hardware type, resolution, and scan rate), the samples imaged, signal processing algorithms, and the number of repeated frames. The foremost prerequisite for high-quality DyC-OCT analysis is the high spatial resolution: the system must resolve the target microstructure that accounts for added contrast after analysis. Secondly, to meet the Nyquist sampling criteria, the OCT frame or volume rate must be at least twice the frequency of the tissue dynamics being measured. Furthermore, to capture slow tissue dynamic changes, repeated three-dimensional (3D) OCT volumes can be acquired, rather than repeated 2D scans27, reducing the total acquisition time for 3D DyC-OCT. For samples exhibiting rapid dynamic changes, ultra-high speed OCT systems are needed to efficiently measure 3D DyC-OCT image stacks. We will discuss about OCT frame or volume rate selection criteria further in Outlook and Conclusions.

Table 1.

Summary of OCT system parameters, applied samples, and DyC-OCT algorithms used in recent publications

Frame/Volume Rate (Hz)a Resolution (μm) Samples Sample FOV (μm3) [DyC-OCT Image or Volume Size]b Total Scan Time (s) Algorithm NR[SR]c Author, Yearref.
Axial Lateral
FF-OCT: High axial and lateral resolutions; Low penetration depth and fixed FOV
138 1 1 mouse brain cortex, rat liver, and kidney 604 × 604 × 42 [-] 20 STD 10 [-] Apelian et al.15
100 0.5 mouse retina, macaque retina 1400 × 1400 [-] 1 FFT 100 [-] Thouvenin et al.32
100 7.7 1.7 macaque retina 100 × 100 × 100 [-] FFT 100 [-] Scholler et al.24
esophageal mucosa, duodenal mucosa 1300 × 1300 [-] 12 Quénéhervé et al.37
100 1.7 0.5 retinal organoids 320 × 320 [1440 × 1440] FFT 512 [-] Scholler et al.29
300 1 1 5 HeLa cells 3.3 FFT 1000 [-] Park et al.30
1 1 5 Nannizzia gypsea, Aspergillus fumigatus, Rhizopus arrhizus FFT Maldiney et al.41
100 1.7 0.5 ppRPE and hiRPE 320×320 [-] FFT 512 [-] Groux et al.22
140 photosynthetic cells FFT 200 [-] Bey et al.28
SD-OCT: High axial and lateral resolutions, deep penetration and flexible FOV
108 1 1 murine liver, mouse tongue 270 × 285 × 550 [500 × 500 × Nd] 695 FFT 100 [500] Münter et al.26
40 1 2 human esophageal and cervical biopsy [1000 × Nd] 25 FFT 200 [512] Leung et al.25
24 1.8 1.1 murine trachea wall tissue, murine liver 300 × 300 × 800 [500 × 500 × Nd] 41.7 FFT 100 [500 × 50] Münter et al.27
20 7.5 7.9 mouse bladder, mouse urothelium [1000 × Nd] 50 STD, OCDS 20 [1000] Xu et al.21
111 1.5 1.5 mouse trachea 500 × 500 × 800 [500 × 500 × Nd] 1200 FFT 150 [500] Kohlfaerber et al.16
4 3.8 4.9 alveolar organoids 1000 × 1000 × Z [512 × 128 × Nd] 60 LIV 32 [512 × 16] Morishita et al.43
112.5 2.1 3.9 human heart organoids 500 × Z [500 × Nd] FFT [500] Hao et al.42
SS-OCT: Deep penetration and flexible FOV; Low axial resolution
4 14 19 tumor spheroid 52.4 LIV, OCDS 65 [512 × 16] El-Sadek et al.17
4 14 18.1 tumor spheroid 1000 × 1000 × Z [512 × 128 × Nd] 52.4 LIV, OCDS 32 [512 × 16] El-Sadek et al.18

FFT fast Fourier transform, Amp amplitude, OCDS OCT correlation decay speed, FOV field of view, Z imaged tissue depth, Nd number of pixels in imaging depth. These two terms are not specified in literatures.

aFor FF-OCT, the acquired frames are fixed in size. For SD-OCT and SS-OCT, the frame/volume rates are specified with the size of repeated frames/volumes.

bFor DyC-OCT 3D volumes, FOV is shown in the form of (W × H × D). For DyC-OCT 2D images, FOV is shown in the form of (W × H). W width, H height and D depth.

cNR Number of repeated OCT frames or volumes acquired for DyC-OCT processing. SR: Data size of each repeated OCT frame or volume. For repeated volumes, the size is shown in the form of NA × NB. For repeated frames the size is NA. NA number of A-scans, and NB number of B-scans.

Applications of DyC-OCT

DyC-OCT applications can be divided into two types. The first utilizes dynamic information to highlight regions of high temporal variations, producing a morphological image with enhanced contrast. The second type interprets dynamics as an indicator of cell viability, with less concern for structures.

DyC-OCT to visualize cell and tissue morphology

Several approaches are available for visualizing cell and tissue morphology. Histology34, extensively used for disease diagnosis, examines stained and sectioned samples under a microscope in two dimensions. However, it involves tedious and invasive sample preparation methods, which induce inevitable distortion and deteriorate cells from their normal conditions. Fluorescence microscopy35 can visualize specific cell structures and physiology with the aid of exogenous or endogenous fluorophores. 2D wide-field epifluorescence36 imaging captures features readily, but has limited contrast and spatial resolution, while 3D fluorescence imaging modalities, such as confocal or two-photon microscopy, support depth scanning. However, the slow imaging speed (e.g., 5-10 frames per second) and restricted imaging depth (e.g., a few hundred microns) hinder imaging of large samples. As an alternative, DyC-OCT can achieve non-invasive, cell/tissue-specific, high-speed, and depth-resolvable dynamic live tissue imaging without the need for exogenous contrast agents.

DyC-OCT supports the identification of high dynamic structures with morphologies resembling cells and subcellular structures (nuclei and organelles) in various organs, such as mouse brain15, murine liver15,26,27, human esophageal tissues25, human gut explants37, as well as airway tissues16. DyC-OCT demonstrates a remarkable capability to reveal dynamic structures with distinguishable morphologies in different layers of retinal explants22,24,32. This matches the fact that retinal explants have different cell types by layers (Fig. 2a), which has been validated by histology38 and fluorescence images39. DyC-OCT also has impressive potential for cancer diagnoses, since it can detect dynamics in cancerous tissues, such as murine cancerous intestine15 and breast cancer40. Figure 2b compares OCT, DyC-OCT, and histology images of benign breast lobules and invasive breast tissue, where DyC-OCT dramatically enhances the contrast of OCT, and reports a high diagnostic accuracy40. DyC-OCT has also demonstrated the potential of fungal infection diagnosis to visualize 3D structures of Nannizzia gypsea, Aspergillus fumigatus, and Rhizopus arrhizus41. Overall, DyC-OCT can benefit disease diagnosis by enhancing cellular dynamic contrast to reveal changes in morphological features.

Fig. 2. Illustrations of DyC-OCT representative applications.

Fig. 2

a Morphology of retinal cells in different retinal layers23. (i) 3D volumetric DyC-OCT cellular morphology of a retinal explant. (ii) cross-sectional DyC-OCT morphology, showing retinal layers. (iii, iv, v, and vi) en face DyC-OCT cellular morphology of the inner nuclear layer (INL), outer nuclear layer (ONL), photoreceptor (PR) layer, and retinal pigment epithelium (RPE) layer. b Comparison of a benign breast lobule (i, OCT; iii, DyC-OCT; v, histology) and invasive breast cancer (ii, OCT; iv, DyC-OCT; vi, histology)40. c Necrosis of tumor spheroids revealed by DyC-OCT with increasing concentrations of Taxol (Paclitaxel)18, scale bar: 200 μm. (a), (b), and (c) are reproduced from Scholler et al.23, Yang et al.40, and El-Sadek et al.18 with permissions from publishers.

Another application of DyC-OCT is structure segmentation based on cell dynamics. For example, in the mouse tongue, DyC-OCT has successfully segmented the cornified layer, granular and spinous layers, basal layer, lamina propria, and muscle layer26. In retinal explants, the inner nuclear layer, outer nuclear layer, and photoreceptor layer have been clearly segmented by DyC-OCT22,24,32. In addition, DyC-OCT has improved accuracy in segmenting the urothelium in images of porcine urinary bladders21. For human heart organoids, our lab has reported the differentiation of cardiomyocytic beating clusters with two different beating frequencies using DyC-OCT42. The quick and reliable structure segmentation promotes the localization of clusters for quantitative assessment.

Another advantage of DyC-OCT is its ability to monitor morphological changes over time. For example, DyC-OCT has been used to monitor organoid development, such as retinal organoids29 and alveolar organoids43. DyC-OCT enables differentiation of cellular lifetime stages inside a single cell based on scattering particles’ movements in the organelles of HeLa cells30. Time-lapse DyC-OCT also shows the migration of immune cells inside airway tissues16. DyC-OCT provides a promising alternative to study the growth dynamics in organoids and tissue models, complementing existing microscopic imaging modalities.

DyC-OCT as cell activity/viability assays

DyC-OCT has been applied to assess cell activity in response to different types of stimuli. In studies of age-related macular degeneration (AMD), DyC-OCT has monitored the wound healing process of primary porcine and human stem cell-derived retinal pigment epithelium (RPE) cells after scratch assays to study degenerative disease evolution22. In studying chemical stimulations, DyC-OCT has revealed an increased beating frequency of ciliated cells in airway tissues under ATP-γ-S stimulation, demonstrating the function of ATP-γ-S in strengthening metabolic activities16. DyC-OCT has captured a signal of dynamics decay in hepatocytes in a rat liver treated with a glycolysis inhibitor, indicating that the inhibitor suppressed the hepatocytes’ activity15. In alveolar organoids, DyC-OCT has revealed changes in the size and shape of alveoli after bleomycin was applied43. In plant science, DyC-OCT was applied to evaluate the effects of LED illumination as well as iron and phosphate deficiency on photosynthetic cells under various environmentally relevant stress challenges28.

Cell viability assays are important measurements to characterize cellular activity in response to external physical and chemical stimulations. Current methods rely on the introduction of luminescent probes that can be used to characterize cell membrane permeability, ATP synthetization, and protease activities, which are specific indicators of cellular viability. DyC-OCT measures cellular dynamics, which may be used as an alternative to measure cell viability. A recent study combined DyC-OCT with supervised machine learning models to quantitatively and in real-time assess the HeLa cell viability when keeping the cells at room temperature without CO2 supply for a few hours44.

DyC-OCT’s analytical potential extends to clinical applications, including drug screening17,18. Specifically, DyC-OCT was integrated into cancer research to assess the impacts of two anti-cancer drugs, Paclitaxel and Blebbistatin, on mammary organoids. This analysis revealed statistically significant distinctions between the time- and concentration-dependent reactions of the two drugs (Fig. 2c). In another study comparing Paclitaxel and the active metabolite of irinotecan (SN-38), DyC-OCT successfully captured the necrotic process of the cancer cells in human-derived tumor spheroids. These studies showed the promise of using DyC-OCT for drug screening.

Outlook and conclusions

DyC-OCT has exhibited its potential to capture dynamic processes in various biological samples, the timescale of which contributes to the choice of hardware and imaging specifications. Molecular dynamics such as DNA replication and RNA transcription can happen in milliseconds45, while cellular differentiation and tissue development can take hours or longer45. It is important to choose an imaging protocol that matches the frame/volume rate of the OCT imaging system with the expected timescale/frequency of the dynamic activity. The frame/volume rate should be at least two times higher than the frequency of the activities to satisfy the Nyquist sampling criterion. Frame/volume rate can be changed by altering the exposure time or number of A-scans or B-scans in an image stack, while keeping into consideration the spatial sampling requirements. Cross-sectional OCT scanning speed restricts the frame rate to capturing events that occur on the millisecond scale, such as tissue metabolic process46. Slow activities happening in a fraction of second, such as beating of the heart organoid47 and ciliated cells in airway tissue or organoids, can be readily captured by DyC-OCT. For even slower processes that happen in minutes, hours or days, such as cell division, embryo development or tissue regeneration, time-lapse 3D DyC-OCT can be obtained using repeated 3D OCT scans.

Researchers have made impressive progress in DyC-OCT. However, a few challenges remain open. Box 1 summarized the key progress and opening challenges of DyC-OCT. Validation of observed imaging features is one of the main challenges in DyC-OCT research. DyC-OCT signals are not explicitly tied to any subcellular compartment (e.g., nucleus, cytoplasm) or specific molecules (e.g., DNA, fluorescence protein). They originate from temporal changes in activities which can vary between different cell and tissue types. As a result, without co-registered images from other biological imaging technologies as validation, interpretation of DyC-OCT signals can be challenging. Although both histology25,48 and fluorescence microscopy18,20,48 have been applied for comparison with DyC-OCT results, co-registration between the modalities remains a challenge, typically due to different imaging orientation between 3D OCT volumes and 2D imaging modalities48,49. The difficulties in co-registering also bring obstacles in precise validation based on quantitative matching, which requires pixel-wise registration between images from DyC-OCT and other modalities. The lack of validation prevents DyC-OCT from being applied to quantitatively analyze tissue morphology and cellular activities. Multi-modal imaging platforms that combine DyC-OCT with modalities like confocal, light-sheet, or two-photon microscopy enable co-registered imaging of the same sample. This approach not only cross-validates DyC-OCT results but also yields complementary information. For example, simultaneous calcium fluorescence and DyC-OCT imaging of beating heart organoids can enhance our understanding of cardiac electromechanical coupling42. Moreover, correlating DyC-OCT signals with established biological activities aids in interpreting DyC-OCT data and linking it to specific dynamic events. Researchers have observed OCT fluctuation pattern changes following stimulations known to alter activities, like ATP-γ-S16, inhibitors15,43 and cancer drugs18,48. These findings reinforce DyC-OCT’s validity and provide insights into dynamic activities in biological systems.

Another challenge is that despite fast OCT B-scan acquisition, DyC-OCT imaging, especially 3D DyC-OCT, requires dozens or hundreds of repeated frames scanning from all B-scan locations, prolonging the imaging acquisition time. To speed up DyC-OCT acquisition, one potential solution is to use parallel OCT imaging, such as space-division multiplexing OCT (SDM-OCT)50 or line-field parallel SS-OCT51, which acquire OCT images from different B-scan locations simultaneously. Reducing acquisition time not only enhances the sampling rate of 3D DyC-OCT but also opens up possibilities for in vivo applications, where fast acquisition is essential to minimize motion artifacts.

Although classical algorithms, such as LIV, STD and FFT, can extract the dynamic pattern from OCT signal fluctuations, their image processing steps are laborious. Alternatively, machine-learning or deep-learning-based models can extract dynamic features to generate DyC-OCT images and reveal dynamic activities not resolvable by classical algorithms. Learning-based time series analysis algorithms, such as long short-term memory (LSTM)52 networks, could be employed to extract more informative dynamic patterns.

In summary, DyC-OCT leverages high-resolution OCT imaging to capture unique cellular dynamics and activities in live samples, without requiring tissue labeling. Various algorithms have been developed to identify distinct motion patterns, thus providing dynamic insights into biological tissues. DyC-OCT’s application extends to dynamic image contrast enhancement in ex vivo tissue samples, 3D organoids, and other in vitro models. Future advancements in multi-modal imaging systems, high-speed and parallel OCT imaging, combined with cutting-edge image processing algorithms, promise to fully unveil the capabilities of this emerging imaging technology.

Box 1 Key progress and open challenges.

Key Progress

  1. DyC-OCT provides noninvasive, high-resolution, contrast-enhanced imaging by utilizing diverse pattern extraction algorithms across common OCT modalities.

  2. DyC-OCT discerns cellular microstructures in ex vivo tissues and in vitro organoids. This technique has been utilized in disease diagnosis, structure segmentation, and developmental monitoring.

  3. DyC-OCT provides quantitative information on dynamic biological activities and has been employed as potential cellular activity/viability assays.

Open Challenges and Opportunities

  1. Interpreting features in DyC-OCT images is challenging without rigorous validation through established methods, owing to the indirect relationship between DyC-OCT signals and the dynamic biological activities of cells and tissues.

  2. To date, DyC-OCT applications have primarily focused on ex vivo living tissues. In vivo applications face challenges due to prolonged image acquisition times and sample motions. These issues could potentially be addressed with high-speed parallel imaging OCT systems.

  3. Learning-based algorithms may be able to effectively extract dynamic features for DyC-OCT, thereby improving efficiency and revealing dynamic activities that classical algorithms cannot resolve.

Supplementary information

Supplementary Table 1 (106.7KB, pdf)

Acknowledgements

The authors wish to express their gratitude to Mr. James Ballard for his constructive feedback on the manuscript. This work is supported by the National Institutes of Health (NIH) grants R01EB025209, 1R01HL156265, and 1R21EB03268401A1 awarded to C.Z., and the National Science Foundation (NSF) Graduate Research Fellowship (GRF) awarded to A.M.

Author contributions

C.R. and C.Z. conceived the outline of this mini review, with inputs from S.H., F.W., A.M., and M.A.; C.R. performed the literature research and wrote the first draft; F.W. and C.R. generated the table and figures; D.Y. and L.W. developed the airway organoids; all authors thoroughly reviewed and revised the manuscript; C.Z. supervised and guided the project.

Peer review

Peer review information

Communications Biology thanks the anonymous reviewers for their contribution to the peer review of this work. Primary Handling Editor: Manuel Breuer.

Competing interests

C.Z. is an Editorial Board Member for Communications Biology, but was not involved in the editorial review of, nor the decision to publish this article. All other authors declare no competing interests.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

The online version contains supplementary material available at 10.1038/s42003-024-05973-5.

References

  • 1.Huang D, et al. Optical coherence tomography. Science. 1991;254:1178–1181. doi: 10.1126/science.1957169. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Klein T, Huber R, High-speed OCT. light sources and systems [Invited] Biomed. Opt. Express. 2017;8:828–859. doi: 10.1364/BOE.8.000828. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Grulkowski, I. et al. In Optical Coherence Tomography: Technology and Applications (eds. Drexler, W. & Fujimoto, J. G.) 319–356 (Springer International Publishing, 2015). 10.1007/978-3-319-06419-2_11.
  • 4.Lee TM, et al. Engineered microsphere contrast agents for optical coherence tomography. Opt. Lett. 2003;28:1546–1548. doi: 10.1364/OL.28.001546. [DOI] [PubMed] [Google Scholar]
  • 5.Oldenburg AL, Gunther JR, Boppart SA. Imaging magnetically labeled cells with magnetomotive optical coherence tomography. Opt. Lett. 2005;30:747–749. doi: 10.1364/OL.30.000747. [DOI] [PubMed] [Google Scholar]
  • 6.Agrawal A, et al. Quantitative evaluation of optical coherence tomography signal enhancement with gold nanoshells. J. Biomed. Opt. 2006;11:041121. doi: 10.1117/1.2339071. [DOI] [PubMed] [Google Scholar]
  • 7.Zhou C, et al. Photothermal optical coherence tomography in ex vivo human breast tissues using gold nanoshells. Opt. Lett. 2010;35:700–702. doi: 10.1364/OL.35.000700. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kavalaraki A, Spyratou E, Kouri MA, Efstathopoulos EP. Gold nanoparticles as contrast agents in ophthalmic imaging. Optics. 2023;4:74–99. doi: 10.3390/opt4010007. [DOI] [Google Scholar]
  • 9.de Carlo TE, Romano A, Waheed NK, Duker JS. A review of optical coherence tomography angiography (OCTA) Int. J. Retin. Vitr. 2015;1:5. doi: 10.1186/s40942-015-0005-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Jia Y, et al. Split-spectrum amplitude-decorrelation angiography with optical coherence tomography. Opt. Express. 2012;20:4710–4725. doi: 10.1364/OE.20.004710. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Pujari A, et al. Optical coherence tomography angiography in neuro-ophthalmology: current clinical role and future perspectives. Surv. Ophthalmol. 2021;66:471–481. doi: 10.1016/j.survophthal.2020.10.009. [DOI] [PubMed] [Google Scholar]
  • 12.Grishina OA, Wang S, Larina IV. Speckle variance optical coherence tomography of blood flow in the beating mouse embryonic heart. J. Biophotonics. 2017;10:735–743. doi: 10.1002/jbio.201600293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Luisi JD, Lin JL, Ameredes BT, Motamedi M. Spatial-temporal speckle variance in the en-face view as a contrast for optical coherence tomography angiography (OCTA) Sensors. 2022;22:2447. doi: 10.3390/s22072447. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Markowitz O, Schwartz M, Minhas S, Siegel DM. Speckle-variance optical coherence tomography: a novel approach to skin cancer characterization using vascular patterns. Dermatol. Online J. 2016;22:13030/qt7w10290r. doi: 10.5070/D3224030620. [DOI] [PubMed] [Google Scholar]
  • 15.Apelian C, Harms F, Thouvenin O, Boccara AC. Dynamic full field optical coherence tomography: subcellular metabolic contrast revealed in tissues by interferometric signals temporal analysis. Biomed. Opt. Express. 2016;7:1511–1524. doi: 10.1364/BOE.7.001511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Kohlfaerber T, et al. Dynamic microscopic optical coherence tomography to visualize the morphological and functional micro-anatomy of the airways. Biomed. Opt. Express. 2022;13:3211. doi: 10.1364/BOE.456104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.El-Sadek IA, et al. Optical coherence tomography-based tissue dynamics imaging for longitudinal and drug response evaluation of tumor spheroids. Biomed. Opt. Express. 2020;11:6231–6248. doi: 10.1364/BOE.404336. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.El-Sadek IA, et al. Three-dimensional dynamics optical coherence tomography for tumor spheroid evaluation. Biomed. Opt. Express. 2021;12:6844. doi: 10.1364/BOE.440444. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mukherjee P, et al. Label-free metabolic imaging of non-alcoholic-fatty-liver-disease (NAFLD) liver by volumetric dynamic optical coherence tomography. Biomed. Opt. Express. 2022;13:4071. doi: 10.1364/BOE.461433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Thouvenin O, Fink M, Boccara AC. Dynamic multimodal full-field optical coherence tomography and fluorescence structured illumination microscopy. J. Biomed. Opt. 2017;22:026004. doi: 10.1117/1.JBO.22.2.026004. [DOI] [PubMed] [Google Scholar]
  • 21.Xu Z, Zhu H, Wang H. Segmentation of the urothelium in optical coherence tomography images with dynamic contrast. J. Biomed. Opt. 2021;26:086002. doi: 10.1117/1.JBO.26.8.086002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Groux K, et al. Dynamic full-field optical coherence tomography allows live imaging of retinal pigment epithelium stress model. Commun. Biol. 2022;5:1–11. doi: 10.1038/s42003-022-03479-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Tan Q, Choi KM, Sicard D, Tschumperlin DJ. Human airway organoid engineering as a step toward lung regeneration and disease modeling. Biomaterials. 2017;113:118–132. doi: 10.1016/j.biomaterials.2016.10.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Scholler J, et al. Probing dynamic processes in the eye at multiple spatial and temporal scales with multimodal full field OCT. Biomed. Opt. Express. 2019;10:731. doi: 10.1364/BOE.10.000731. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Leung HM, et al. Imaging intracellular motion with dynamic micro-optical coherence tomography. Biomed. Opt. Express. 2020;11:2768–2778. doi: 10.1364/BOE.390782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Münter M, et al. Dynamic contrast in scanning microscopic OCT. Opt. Lett. 2020;45:4766–4769. doi: 10.1364/OL.396134. [DOI] [PubMed] [Google Scholar]
  • 27.Münter M, et al. Microscopic optical coherence tomography (mOCT) at 600 kHz for 4D volumetric imaging and dynamic contrast. Biomed. Opt. Express. 2021;12:6024–6039. doi: 10.1364/BOE.425001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Bey H, et al. Dynamic Cell Imaging: application to the diatom Phaeodactylum tricornutum under environmental stresses. Eur. J. Phycol. 2022;0:1–11. [Google Scholar]
  • 29.Scholler J, et al. Dynamic full-field optical coherence tomography: 3D live-imaging of retinal organoids. Light Sci. Appl. 2020;9:140. doi: 10.1038/s41377-020-00375-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Park S, et al. Quantitative evaluation of the dynamic activity of HeLa cells in different viability states using dynamic full-field optical coherence microscopy. Biomed. Opt. Express. 2021;12:6431. doi: 10.1364/BOE.436330. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Beaurepaire E, Boccara AC, Lebec M, Blanchot L, Saint-Jalmes H. Full-field optical coherence microscopy. Opt. Lett. 1998;23:244–246. doi: 10.1364/OL.23.000244. [DOI] [PubMed] [Google Scholar]
  • 32.Thouvenin O, et al. Cell motility as contrast agent in retinal explant imaging with full-field optical coherence tomography. Invest. Ophthalmol. Vis. Sci. 2017;58:4605–4615. doi: 10.1167/iovs.17-22375. [DOI] [PubMed] [Google Scholar]
  • 33.Choma MA, Hsu K, Izatt JA. Swept source optical coherence tomography using an all-fiber 1300-nm ring laser source. J. Biomed. Opt. 2005;10:044009. doi: 10.1117/1.1961474. [DOI] [PubMed] [Google Scholar]
  • 34.Alturkistani HA, Tashkandi FM, Mohammedsaleh ZM. Histological stains: a literature review and case study. Glob. J. Health Sci. 2016;8:72–79. doi: 10.5539/gjhs.v8n3p72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Sanderson MJ, Smith I, Parker I, Bootman MD. Fluorescence microscopy. Cold Spring Harb. Protoc. 2014;2014:pdb.top071795. doi: 10.1101/pdb.top071795. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Webb DJ, Brown CM. Epi-fluorescence microscopy. Methods Mol. Biol. Clifton NJ. 2013;931:29–59. doi: 10.1007/978-1-62703-056-4_2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Quénéhervé L, et al. Full-field optical coherence tomography: novel imaging technique for extemporaneous high-resolution analysis of mucosal architecture in human gut biopsies. Gut. 2021;70:6–8. doi: 10.1136/gutjnl-2020-321228. [DOI] [PubMed] [Google Scholar]
  • 38.Tsang, S. H. & Sharma, T. In Atlas of Inherited Retinal Diseases (eds. Tsang, S. H. & Sharma, T.) 3–5 (Springer International Publishing, 2018). 10.1007/978-3-319-95046-4_1.
  • 39.Grünert U, Martin PR. Cell types and cell circuits in human and non-human primate retina. Prog. Retin. Eye Res. 2020;78:100844. doi: 10.1016/j.preteyeres.2020.100844. [DOI] [PubMed] [Google Scholar]
  • 40.Yang H, et al. Use of high-resolution full-field optical coherence tomography and dynamic cell imaging for rapid intraoperative diagnosis during breast cancer surgery. Cancer. 2020;126:3847–3856. doi: 10.1002/cncr.32838. [DOI] [PubMed] [Google Scholar]
  • 41.Maldiney T, et al. Dynamic full-field optical coherence tomography as complementary tool in fungal diagnostics. J. Med. Mycol. 2022;32:101303. doi: 10.1016/j.mycmed.2022.101303. [DOI] [PubMed] [Google Scholar]
  • 42.Hao S, et al. Dual-modality imaging system for monitoring human heart organoids beating in vitro. Opt. Lett. 2023;48:3929–3932. doi: 10.1364/OL.493824. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Morishita R, et al. Label-free intratissue activity imaging of alveolar organoids with dynamic optical coherence tomography. Biomed. Opt. Express. 2023;14:2333–2351. doi: 10.1364/BOE.488097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Park S, et al. Label-free, non-invasive, and repeatable cell viability bioassay using dynamic full-field optical coherence microscopy and supervised machine learning. Biomed. Opt. Express. 2022;13:3187–3194. doi: 10.1364/BOE.452471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Cui, Y. & Irudayaraj, J. In Epigenetics and Gene Expression in Cancer, Inflammatory and Immune Diseases (eds. Stefanska, B. & MacEwan, D. J.) 135–154 (Springer, 2017). 10.1007/978-1-4939-6743-8_9.
  • 46.Shamir, M., Bar-On, Y., Phillips, R. & Milo, R. SnapShot: timescales in cell biology. Cell164, 1302–1302.e1 (2016). [DOI] [PubMed]
  • 47.Lewis-Israeli YR, et al. Self-assembling human heart organoids for the modeling of cardiac development and congenital heart disease. Nat. Commun. 2021;12:5142. doi: 10.1038/s41467-021-25329-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Abd El-Sadek I, et al. Label-free drug response evaluation of human derived tumor spheroids using three-dimensional dynamic optical coherence tomography. Sci. Rep. 2023;13:15377. doi: 10.1038/s41598-023-41846-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Plekhanov AA, et al. Histological validation of in vivo assessment of cancer tissue inhomogeneity and automated morphological segmentation enabled by optical coherence elastography. Sci. Rep. 2020;10:11781. doi: 10.1038/s41598-020-68631-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Zhou C, Alex A, Rasakanthan J, Ma Y. Space-division multiplexing optical coherence tomography. Opt. Express. 2013;21:19219–19227. doi: 10.1364/OE.21.019219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Fechtig DJ, et al. Line-field parallel swept source MHz OCT for structural and functional retinal imaging. Biomed. Opt. Express. 2015;6:716–735. doi: 10.1364/BOE.6.000716. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Van Houdt G, Mosquera C, Nápoles G. A review on the long short-term memory model. Artif. Intell. Rev. 2020;53:5929–5955. doi: 10.1007/s10462-020-09838-1. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Table 1 (106.7KB, pdf)

Articles from Communications Biology are provided here courtesy of Nature Publishing Group

RESOURCES