Validation of 2D flow MRI for helical and vortical flows

Zia Mehmood 1, Hosamadin Assadi 1,2, Ciaran Grafton-Clarke 1,3, Rui Li 1,3, Gareth Matthews 1,3, Samer Alabed 4, Rebekah Girling 1, Victoria Underwood 1, Bahman Kasmai 1,3, Xiaodan Zhao 5, Fabrizio Ricci 6, Liang Zhong 5, Nay Aung 7, Steffen Erhard Petersen 8, Andrew J Swift 9, Vassilios S Vassiliou 1,3, João Cavalcante 10, Rob J van der Geest 11, Pankaj Garg 1,3,
PMCID: PMC10928773  PMID: 38458769

Abstract

Purpose

The main objective of this study was to develop two-dimensional (2D) phase contrast (PC) methods to quantify the helicity and vorticity of blood flow in the aortic root.

Methods

This proof-of-concept study used four-dimensional (4D) flow cardiovascular MR (4D flow CMR) data of five healthy controls, five patients with heart failure with preserved ejection fraction and five patients with aortic stenosis (AS). A PC through-plane generated by 4D flow data was treated as a 2D PC plane and compared with the original 4D flow. Visual assessment of flow vectors was used to assess helicity and vorticity. We quantified flow displacement (FD), systolic flow reversal ratio (sFRR) and rotational angle (RA) using 2D PC.

Results

For visual vortex flow presence near the inner curvature of the ascending aortic root on 4D flow CMR, sFRR demonstrated an area under the curve (AUC) of 0.955, p<0.001. A threshold of >8% for sFRR had a sensitivity of 82% and specificity of 100% for visual vortex presence. In addition, the average late systolic FD, a marker of flow eccentricity, also demonstrated an AUC of 0.909, p<0.001 for visual vortex flow. Manual systolic rotational flow angle change (ΔsRA) demonstrated excellent association with semiautomated ΔsRA (r=0.99, 95% CI 0.9907 to 0.999, p<0.001). In reproducibility testing, average systolic FD (FDsavg) showed a minimal bias at 1.28% with a high intraclass correlation coefficient (ICC=0.92). Similarly, sFRR had a minimal bias of 1.14% with an ICC of 0.96. ΔsRA demonstrated an acceptable bias of 5.72°—and an ICC of 0.99.

Conclusion

2D PC flow imaging can possibly quantify blood flow helicity (ΔRA) and vorticity (FRR). These imaging biomarkers of flow helicity and vorticity demonstrate high reproducibility for clinical adoption.

Trials registration number

NCT05114785.

Keywords: Magnetic Resonance Imaging, Aortic Diseases, Heart Failure, Aortic Valve Stenosis


WHAT IS ALREADY KNOWN ON THIS TOPIC

  • Despite some of the advances in knowledge and technology for in vivo assessment, there is limited literature on assessing aortic helical and vortical flows and their impact on clinical outcomes in various cardiovascular diseases. While capable, four-dimensional (4D) flow cardiovascular MR (CMR) is intricate and time-consuming to perform and not widely available. Consequently, its potential impact to link the association of aortic flow haemodynamics with various cardiovascular diseases is restricted. Previous studies have shown the relevance of these functional parameters of aortic flow and their association with reduced aortic conduit and reservoir function.

WHAT THIS STUDY ADDS

  • This study proposed a simplified approach for broader clinical assessment of helical and vortical flows in the aorta, which can be measured by two-dimensional (2D) phase-contrast (2D PC) CMR. 2D PC flow imaging can possibly measure aortic blood flow helicity and vorticity which appears to be comparable to 4D flow assessment.

HOW THIS STUDY MIGHT AFFECT RESEARCH, PRACTICE OR POLICY

  • Helicity and vorticity can possibly be quantified with 2D PC CMR, potentially allowing retrospective studies using large databases to infer population-level disease association of aortic flow haemodynamics in several cardiovascular diseases. This will enable not only simple clinical translation but also future studies to investigate the prognostic role of aortic flow using large databases.

Introduction

Aortic blood flow exhibits a complex and distinctive multidirectional pattern, influenced by the structure of the aortic valve, the shape and branching of the ascending aorta, as well as the compliance and elasticity of the aortic wall. Healthy individuals’ aortic blood flow has a characteristic spiral component defined as helicity along with a predominant laminar flow.1–3 Aortic vortical flow is a swirling or rotational movement of blood within the aorta which has been linked to increased wall stress, potentially leading to atherosclerosis, thrombus formation, increased wall stiffness and accelerating hypertension.4 5 Factors such as left ventricular function, aortic valve disease, aortic compliance and dilatation impact the helicity and vorticity of aortic blood flow. This could possibly impact aortic conduit and reservoir functions.6 7

Traditionally, quantitative blood flow analysis has been based on ECG-gated time-resolved two-dimensional (2D) phase contrast (PC) cardiovascular MR (CMR) for peak velocity, regurgitation fraction, stroke volume and shunt volumes.8 9 There are certain limitations with this modality, mainly to do with single directional velocity encoding, which does not provide direct information about helicity or vorticity of blood flow. four-dimensional (4D) flow CMR, on the other hand, allows comprehensive visual assessment and obtains blood flow analysis along three spatial dimensions, three velocity directions and time in a cardiac cycle.10 11 Functional haemodynamic parameters, such as helicity, vorticity and flow displacement (FD), can be derived retrospectively at any location within the thoracic aorta.12–16 However, 4D flow CMR remains complex, requiring long acquisition time and postprocessing, and, more importantly, is not widely accessible.

As such, despite some of the advances in knowledge and technology for in vivo assessment, there is limited literature on assessing these functional parameters of aortic flow and their impact on clinical outcomes in various cardiovascular diseases. For broader clinical assessment of helical and vortical flows in the aorta and their clinical role investigation, simplified approaches are needed, which can be measured by both 2D PC and also by 4D flow CMR. This will allow not only simple clinical translation but also future studies to investigate the prognostic role of aortic flow using large databases. Therefore, the main objective of this study was to develop and validate pragmatic 2D PC methods to quantify both helicity and vorticity of blood flow in the aortic root.

Methods

Study cohort

For this proof-of-concept study, we identified patients from the PREFER-CMR registry (ClinicalTrials.gov: NCT05114785). We enrolled five cases of each aortic stenosis (AS) and heart failure with preserved ejection fraction (HFpEF) categories. The main inclusion criteria for patients were over 18 years of age and a confirmed diagnosis of AS or HFpEF by CMR. We selected HFpEF and AS patient groups because both these diseases are associated with aortic pathology and possible flow disturbances.17 18 The exclusion criteria were limited to any CMR contraindication. We also included five healthy controls (HCs) from previous research registries in this study. The main inclusion criteria for the HCs were >18 years of age and no prior history of cardiovascular disease.

Ethics approval and consent to participate

This study was conducted according to the principles outlined in the Declaration of Helsinki—Version 2013. The collection and management of data were approved by the National Research Ethics Service in the UK (21/NE/0149). A pragmatic opt-out informed consent was obtained from all patients included in the study.19

CMR protocol

CMR study was performed on a 1.5 Tesla Magnetom Sola Siemens system with a superconducting magnet (Siemens Healthineers AG, Erlangen, Germany). All patients were examined in the supine position, headfirst, using a respiratory sensor and ECG gating. Additionally, the scanner was equipped with an 18-channel biometric body coil.

The CMR protocol included baseline survey images and cines, gadolinium enhancement imaging and 4DF acquisition methods previously described by our group.20–25

For standard cines, we acquired 30 phases throughout the cardiac cycle. Other cine acquisition parameters include TR: 2.71, TE: 1.13, field of view (FOV): 360×289.3 mm2 with Phase FOV—80.4%, number of signal averages (NSA): 1, matrix: 224×180 (phase), bandwidth: 167.4 kHz, (930 Hz/Px), flip angle: 80, slice thickness: 8 mm and Grappa acceleration with a factor of 2.

For 4D flow acquisition, the initial velocity-encoding value (VENC) setting was 150–200 cm/s for all HCs and HFpEF cases. For patients with AS, the initial VENC was 400 cm/s. For 4DF, we acquired 30 phases throughout the cardiac cycle to keep the data consistent with cines. The acquired temporal resolution was 40 ms. Other 4D flow acquisition parameters include TR: 4.98, TE: 2.71, field of view (FOV): 200×256.3 mm2, NSA: 1, acquired voxel size=3×3×3 mm3, bandwidth: 31.616 kHz, (494 Hz/Px), flip angle: 5 and Grappa acceleration in the phase-encoding direction with a factor of 2 and slide direction of 1. The ECG was retrospectively gated with free breathing to avoid diastolic temporal blurring.

Image analysis

All image analyses were postprocessed with the in-house developed MASS research software (MASS; Version 2022-EXP, Leiden University Medical Center, Leiden, the Netherlands). A static reformatted plane was planned through the ascending aorta at the mid-main pulmonary artery level to generate a through-plane velocity encoded 2D PC data using 4D flow CMR data. This plane was treated as a 2D PC plane. Ascending aortic helical flow was defined as the flow swirling around the aortic centreline. Ascending aorta vortex flow was defined as any flow rotating on the long axis of the aorta near the inner curvature of the aortic root26 (figure 1).

Figure 1.

Figure 1

(A) Time-resolved 30 passed of two-dimensional (2D) phase contrast through-plane velocity mapping demonstrating systolic flow displacement (FD) (blue line from the centreline of the aorta at the mid-main pulmonary artery level) rotation during systolic phases. (B) The first rotational angle was recorded after the peak systole when the FDs is greater than 12% and has a stable value (green phase). (C) The second recorded phase to compute the rotational angle is at the end of the systole (red phase). (D) Concomitant, superimposed 2D vector using 4D flow CMR data was used to appreciate the direction of rotation when comparing it with the direction of rotation using FDs. CMR, cardiovascular MR.

Bench testing and manual assessment

FD and its threshold

First, we developed methods to superimpose the FD angle on top of the 2D PC image for each phase in the cardiac cycle. FD was calculated using published methods.7 FD was calculated as the distance between the vessel centreline and the centre of the eccentric flow and was normalised for overall vessel size. It is presented as a percentage in this paper. In bench-testing, significant noise was noted in systolic FDs due to mainly minimal FD when there was mainly laminar flow in early systole. Importantly, in early systole, laminar flow resulted in FD being lower than 12% and rotating around randomly without any meaningful flow rotation. Hence, we used an FD threshold of 12% to circumvent this issue for all calculations. Additionally, we recorded FD in late systole (FDls) and diastole (FDd) using the same methods.

Manual tracking of rotational flow angle

The rotational flow was manually determined by mapping the angle of movement of FD from the peak systolic phase (FDspeak) till the end of systole (figure 1). To better appreciate the rotational flow and its direction, we used superimposed 2D vectors using 4D flow datasets. We then calculated the systolic rotational flow angle change (ΔsRA) from mid-systole till the end of systole (figure 2). The rotational direction was recorded as clockwise or counterclockwise (figure 1).

Figure 2.

Figure 2

(A) Aortic flow curve to identify peak systolic and end-systolic phases. (B) Rotational angle curve demonstrating where the start and end were recorded (orange dots). The start was recorded after normalisation from the baseline.

Manual mapping of vortical flow

Manual qualitative assessment of longitudinal vortical flow in the ascending aorta during systole was done using coronal cine with superimposed 2D velocity vectors with Doppler colour-coded signals of forward and backward flow. If, on visual assessment, significant vortical flow was noted, it was recorded (figure 3).

Figure 3.

Figure 3

(A) Vortical flow was visually identified in the long-axis aortic root views. During systole, at mid-main pulmonary artery level, both forward (blue) and backward (orange/red) flows are observed. (B) A reformatted cross-sectional phase contrast through plane demonstrating both forward and backward flows associated with longitudinal vorticity of blood flow near the inner curvature of the ascending aortic root. (C) Flow curves demonstrating the significant amount of backward flow compromising aortic conduit function.

Automated systolic rotational angle change

This was done without any visual input and solely using the rotational angle (RA) time-resolved curve of the FD line. Peak systolic phase was automatically registered as the peak flow rate on the flow curve. End systolic phase was determined where the downward slope of the descending systolic flow curve intersected with the x-axis (or no flow). Systolic RA change was determined from the point the flow angle stabilised after peak systole till the end of systole on the RA curve (figure 2).

Automated flow reversal ratio

Flow reversal ratio (FRR), systolic forward and systolic retrograde flows (SRF) were calculated as previously described in the literature.27 However, in this study, we quantified it in the total cardiac cycle and in the systolic phase only (sFRR)—from the start to the end of systole.

The equation to calculate the FRR was as follows:

FRR(%)=|Qretrograde(tsystole)|/Qforward(tsystole)×100

where Qforward (tsystole) and Qretrograde (tsystole) represent the forward and retrograde flows at peak systole flow.27

All CMR contour tracing and 4D flow analysis were performed by an EACVI level-III certified CMR expert (PG) with over 10 years of CMR experience and academic fellows (ZM and HA) under direct supervision. HA (3 years experience in advanced CMR) repeated the analysis for all cases using the same postprocessing protocol for reproducibility testing.

Statistical analysis

Data analyses were performed by using MedCalc (MedCalc Software, Ostend, Belgium, V.20.215). Data were treated as non-parametric due to small numbers. Continuous variables are expressed as median±IQR. Mann-Whitney U test was used to compare the independent variables. Agreement between visual and quantitative assessments and reproducibility analyses was done using several methods: Pearson’s correlation coefficient (r), Bland-Altman plots intraclass correlation coefficient (ICC) and C-statistics.

For comparing manual versus automated ΔsRA, we used Pearson’s correlation coefficient to assess the correlation and Bland-Altman plots to assess bias and agreement. For comparing categorical vortex visual analysis against FRR and other metrics of FD, we evaluated the area under the curve (AUC) using receiver operating characteristic curves. We used ICC to assess the correlation and Bland-Altman plots for reproducibility testing to assess bias and agreement. The significance threshold was set at p<0.05.

Results

Study population

Patient characteristics are summarised in table 1. A total of 15 cases (46% males) were included in this study. Of these, five were HCs, five were patients with HFpEF and five were AS patients. The median age of our cohort was 76 years (IQR: 56–81 years). The median height was 169 cm (IQR 162–178 cm) and the median weight was 81 kg (IQR 72–96 kg). Among the patients, 6 (60%) had hypertension, 60% had dyslipidaemia and 4 (40%) were diabetic. Other comorbidities included previous cerebrovascular accidents (2 (20%)) and myocardial infarction (1 (10%)). HCs were younger (52±12 years) compared with both HFpEF (80±12 years, p=0.02) and AS cohorts (78±7 years, p=0.02) with no cardiovascular risk factors.

Table 1.

Baseline characteristics of the study population

HC (n=5) HFpEF (n=5) AS (n=5) P value* P value†
Age (years) 52±12 80±12 78±7 0.02 0.02
Gender (male) 2 (40) 2 (40) 3 (60) 1 0.55
Height (cm) 170±18 168±13 168±20 0.75 1
Weight (kg) 76±26 86±26 74±27 0.53 1
Diabetes mellitus 2 (40) 2 (40) 1
Hypertension 4 (80) 2 (40) 0.22
Myocardial infarction 1 (20) 2 (40) 0.51
Cerebrovascular accident 2 (40) 0 (0) 0.14
Dyslipidaemia 3 (60) 3 (60) 1

All data are presented in median±IQR or n (%).

*HC versus HFpEF.

†HC versus AS.

AS, aortic stenosis; HC, healthy controls; HFpEF, heart failure with preserved ejection fraction.

Healthy versus patients

All HCs (5 (100%)) had a clockwise rotation, while 2 (20%) of patients had an anticlockwise flow rotation, of which 1 had both rotations. Patients had a significantly higher FDspeak (19%±21%) and FDsavg (28.5%±7.5%) compared with HCs (all p=0.03). Similarly, other aortic flow parameters showed significant differences between patients and HCs, including FDlsavg (34.6%±7.7% vs 23%±11.2%, p=0.04) and SRF (15.6±14.2 mL vs 2.6±2.2 mL, p<0.0001). Moreover, healthy volunteers had statistically lower total FRR and sFRR when compared with patients (15%±14.3% vs 37.2%±13.9%) and (4.5%±3.7% vs 22.5±11.2%), subsequently. However, no statistical differences were observed between the two groups for RA (p=0.06), average diastolic FDdavg (p=0.06) and systolic forward flow (p=0.22). A comparison of 2D PC derived helicity and vorticity parameters of blood flow between HCs and patients with HFpEF and AS combined is summarised in table 2.

Table 2.

Comparison of helicity and vorticity parameters of blood flow between healthy cohorts and patients with HFpEF and AS combined

Healthy controls Patients P value
Number 5 10
Net aortic forward flow (mL) 58±28 58.4±16.9 0.86
Net aortic backward flow (mL) 1.5±5.2 6.2±4.2 0.16
Aortic maximum area (cm2) 9.2±3.4 11±5.9 0.09
Aortic minimum area (cm2) 6.6±2.2 8.6±2.8 0.13
Systolic forward flow (ml) 64±27.6 71.5±21.3 0.22
Markers of flow eccentricity
 Flow displacement in peak systole (%) 5±3 19±21 0.03
 Average systolic flow displacement (%) 19±4.7 28.5±7.5 0.03
 Average late systolic flow displacement (%) 23±11.2 34.6±7.7 0.04
 Average diastolic flow displacement (%) 30.5±21 40.1±16.4 0.59
Markers of flow vorticity
 Systolic retrograde flow (mL) 2.6±2.2 15.6±14.2 <0.0001
 Systolic flow reversal ratio (%) 4.5±3.7 22.5±11.2 <0.0001
 Total flow reversal ratio (%) 15±14.3 37.2±13.9 0.01
Marker of flow helicity
 Rotational angle (°) 7±17 35±31.6 0.06
 Clockwise rotation (righthand) 5 (100) 8 (80) 0.17

All data are presented as median±IQR and n (%).

AS, aortic stenosis; HFpEF, heart failure with preserved ejection fraction.

RA agreement

Manual ΔsRA demonstrated excellent association with semiautomated ΔsRA (r=0.99, 95% CI 0.9907 to 0.999, p<0.001). In addition, there was an excellent agreement with semiautomated ΔsRA (bias=0.5, limits of agreement=−20.5 to 21.4, p=0.86) (figure 4).

Figure 4.

Figure 4

(A) Scatter plot demonstrating an excellent association between manual and automated methods of rotational angle calculation. (B) Bland-Altman plots show minimal bias between the manual and automated rotational angles from the curve. 2D, two dimensions; PC, phase contrast; RA, rotational angle; LLA, lower limit of agreement; HLA, higer limit of agreement.

The agreement between manual assessment of rotational direction by investigation of 4D flow vectors and computed direction was in good agreement (kappa=0.76; 95% CI 0.32 to 1).

Vortical flow associations

As seen in table 3, patients with vortical flow had higher FDsavg (p=0.03), FDlsavg (p=0.02) and FDdavg (p=0.02). Furthermore, there was a significant difference between patients with no vortex present and patients with vortex for SRF (2.8±2.2 mL vs 14.9±16.1 mL, p=0.02), sFRR (3.8%±3.6% vs 20.5%±15%, p=0.01) and total FRR (16.9%±12.1% vs 37.1%±13.6%, p=0.02). Patients with vortical flow had a larger aortic maximum (11.2±3.8 cm2) and minimum areas (8.7±2.5 cm2) (p=0.04 and p=0.03, respectively). The two groups had no statistical differences in RA and systolic and net forward flows.

Table 3.

Comparison of helicity and vorticity parameters of blood flow between patients with and without vorticity

No vortex present Vortex present P value
Number 4 11
Net aortic forward flow (mL) 60.5±26.1 58±18.9 0.85
Net aortic backward flow (mL) 2.1±5.6 5.7±4.6 0.34
Average systolic flow displacement (%) 18.5±3.9 28.3±7.4 0.03
Average late systolic flow displacement (%) 20.5±7.2 34.3±7.3 0.02
Average diastolic flow displacement (%) 25.7±6.3 42±11.9 0.02
Rotational angle (°) 10.7±20.3 34±43.2 0.28
Systolic forward flow (mL) 64.5±23.4 71.3±18.1 0.47
Systolic retrograde flow (mL) 2.8±2.2 14.9±16.1 0.02
Systolic flow reversal ratio (%) 3.8±3.6 20.5±15 0.01
Total flow reversal ratio (%) 16.9±12.1 37.1±13.6 0.02
Aortic maximum area (cm2) 9.4±3.4 11.2±3.8 0.04
Aortic minimum area (cm2) 6.5±2.1 8.7±2.5 0.03

All data are presented as median±IQR and n (%).

Vortical flow agreement

The FDlsavg, a marker of flow eccentricity, demonstrated an AUC of 0.909, p<0.001 for visual vortex flow presence near the inner curvature of ascending aortic root. A threshold of >24.6% for FDlsavg had a sensitivity of 91% and a specificity of 100% for visual vortex presence (figure 5). Similarly, SRF demonstrated an AUC of 0.909, p<0.001 for visual vortex flow presence near the inner curvature of the ascending aortic root. A threshold of >5 mL for SRF had a sensitivity of 82% and specificity of 100% for visual vortex presence (figure 5).

Figure 5.

Figure 5

(A) Receiver operator curve demonstrating excellent agreement between visual assessment of longitudinal vortex flow in ascending aorta and systolic flow reversal ratio (sFRR) percentage. (B, C) Receiver operator curve showing good correlation between visual assessment of longitudinal vortex flow in ascending aorta and averaged flow displacements in diastolic phases and also the retrograde systolic flow, which is happening directly as a result of the vortex. AUC, area under the curve.

However, sFRR demonstrated an AUC of 0.955, p<0.001 for visual vortex flow presence near the inner curvature of ascending aortic root. A threshold of >8% for sFRR had a sensitivity of 82% and specificity of 100% for visual vortex presence (figure 5).

Reproducibility

Measures of interobserver variability are provided in table 4. The intraclass correlation between all parameters remained good (between 0.87 and 0.99). The bias for flow indices was minimal (0–4 mL). FDsavg showed a minimal bias at 1.28% with a high ICC of 0.92. Similarly, sFRR had a minimal bias of 1.14% with an ICC of 0.96.

Table 4.

Interobserver repeatability analysis for vortical and helical flow parameters by 2D PC CMR 

Bias ICC P value
Net aortic forward flow (mL) 1.84 0.98 0.57
Net aortic backward flow (mL) 0.53 0.97 0.35
Average systolic flow displacement (%) 1.28 0.92 0.45
Average late systolic flow displacement (%) 2.01 0.90 0.44
Average diastolic flow displacement (%) 3.43 0.87 0.36
Rotational angle (°) 5.72 0.99 0.5
Systolic forward flow (mL) 3.19 0.99 0.26
Systolic retrograde flow (mL) 2.28 0.95 0.25
Systolic flow reversal ratio (%) 1.14 0.96 0.33
Total flow reversal ratio (%) 0.99 0.95 0.39
Aortic maximum area (cm2) 0.92 0.94 0.17
Aortic minimum area (cm2) 0.64 0.89 0.20

Bias has been computed using Bland-Altman plot analysis. Absolute values of bias are presented.

CMR, cardiovascular MR; 2D, two dimensions; ICC, intraclass correlation coefficient; PC, phase contrast.

Discussion

In this study, we demonstrated that both helicity and vorticity can possibly be quantified using 2D PC CMR. Change in systolic RA change (ΔsRA), a marker of helicity of blood flow, demonstrated excellent agreement between manual and semiautomated methods, yielding a Pearson’s correlation coefficient of 0.99 and a bias of 0.95. In addition, sFRR demonstrated excellent diagnostic agreement with visual assessment of systolic vortical flow around the inner curvature of the aortic root. Even diastolic FD, a marker of flow eccentricity and SRF were associated with vortical flow in the ascending aortic root.

Systolic flow reversal: a marker of vorticity

Systolic flow should be predominantly anterograde, with any retrograde flow impairing aortic root conduit function, as captured by an increase in FRR. The relevance of FRR has been shown in patients with bicuspid aortic valve (BAV)27 and AS previously.28 Barker et al 27 demonstrated that all BAV patients with AS had FRR >10%. Our study did not investigate BAV, but in all patients with HFpEF and AS, we noted similar higher systolic FRR with a median value of 22.5%. Ha et al 28 retrospectively analysed haemodynamic characteristics of ascending aorta in 80 patients with severe AS using 2D PC CMR and showed significantly higher FRR in patients with aortic root dilatation (>4 cm). The study also showed that subsequent aortic dilatation was associated with higher aortic flow skewness. The major strength of these studies is that they did not use 4D flow CMR and demonstrated the clinical relevance of aortic flow haemodynamics assessed by 2D PC only.

Nevertheless, even though it is plausibly clear that physiological FRR should correspond to vortical flow in the ascending root, a clear threshold has yet to be established. This study utilises 4D Flow CMR for qualitatively and quantitatively validating these aortic physiological flow parameters to 2D PC methods. Importantly, we showed that an sFRR cut-off of >8% was 82% sensitive and 100% specific for visually present vortex during systole. This adds to the growing body of evidence that increased FRR is plausibly associated with reduced aortic conduit function and should be further investigated as a potential metric for improving risk stratification of patients in aortopathy, aortic valvulopathy and other cardiovascular conditions.

Systolic RA change: a marker of helical flow

A study by Frydrychowicz et al demonstrated aortic flow to be predominantly right-handed helical flow by 4D flow CMR streamline visualisation in 62 healthy subjects.29 These findings are consistent with our manual and semiautomated analysis and in agreement with clockwise rotation in 2D. Previous studies have investigated ascending aortic helical flow by three-dimensional (3D) streamlines using 4D flow CMR and have shown that a rotation of <180° is considered normal.30 31 The clinical significance of the increased helical flow remains poorly understood. Physiologically, it is plausible to speculate that a significant increase in helicity of the blood flow in the aorta will increase the overall distance the blood needs to travel during systole and, hence, compromise systolic force in the periphery. This could affect distal perfusion and potentially cause ischaemia. Bürk et al have shown that helical aortic flow increases in patients with a dilated ascending aorta31 or aortic valve disease.32 Even though our study is not directly quantifying the same angle in 3D of the streamlines but instead is quantifying the time-resolved RA in a 2D through-plane, we have demonstrated a similar finding that helicity of blood flow increases in patients not only with AS but also with HFpEF. However, the reference range of the systolic RA change (ΔsRA) investigated in our study remains to be determined in healthy populations. The methods we propose can potentially allow broader clinical translation and investigation of both physiological and pathological aortic flow in several cardiovascular diseases.

Limitations

There are several limitations to the present study. First, the number of controls and patients is insufficient to infer any population-level disease association with flow patterns and their quantification. Second, we have used a different method to validate the helicity of blood flow compared with what has been used in the literature before. The main reason for doing so was to use a time-resolved approach and have better agreement with the rotation in 2D. Third, this study only tests complex flow haemodynamics at the level of ascending aorta (mid-main pulmonary artery level). Hence, the results of this work should be cautiously judged at other levels in the aortic root and other large vessels, for example, the main pulmonary artery. Finally, our work did not evaluate these methods in bicuspid aortic valve patients, and future studies need to evaluate how reliable these methods are in that patient population.

Conclusion

3D PC flow imaging can possibly quantify blood flow helicity (ΔRA) and vorticity (FRR). These imaging biomarkers of flow helicity and vorticity demonstrate high reproducibility for clinical adoption. Further research is needed to validate these findings in larger, more diverse patient populations.

Footnotes

Twitter: @HosamAssadi, @NA

Contributors: ZM: data curation, data collection, formal analysis, methodology, visualisation, writing-original draft preparation, writing-review and editing. HA: data curation, data collection, formal analysis, statistical analysis, visualisation, writing-original draft preparation, writing review and editing. RL: data collection, writing-original draft preparation, writing-review and editing. GM: data collection, writing-original draft preparation, writing-review and editing. CG-C: data collection, writing-original draft preparation, writing-review and editing. SA: data collection, critical contextual review, independent review and editing. RG: resources, independent review and editing. VU: resources, independent review and editing. XZ: conceptualisation, critical contextual review, independent review and editing. BK: data curation, data collection, critical contextual review. FR: conceptualisation, critical contextual review, independent review and editing. LZ: conceptualisation, critical contextual review, independent review and editing. NA: conceptualisation, critical contextual review, independent review and editing. SEP: conceptualisation, critical contextual review, independent review and editing. AJS: funding acquisition, study design, methodology, critical contextual review, writing review and editing. VSV: conceptualisation, critical contextual review, independent review and editing. JC: conceptualisation, critical contextual review, independent review and editing. RJvdG: software solution, validation, methodology, critical contextual review, independent review and editing. PG: conceptualisation, supervision, data curation, funding acquisition, resources, formal analysis, statistical analysis, investigation, methodology, critical contextual review, visualisation, writing-original draft preparation, writing review and editing. All authors listed have contributed sufficiently to the project to be included, and all those qualified to be authors are listed in the author byline. All authors read and approved the final manuscript. PG is the guarantor of the data and this study.

Funding: PG and AJS are funded by Wellcome Trust Clinical Research Career Development Fellowships (220703/Z/20/Z and 205188/Z/16/Z). For the purpose of Open Access, the authors have applied a CG-C BY public copyright licence to any Author Accepted Manuscript version arising from this submission.

Disclaimer: The funders had no role in study design, data collection, analysis, publishing decision, or manuscript preparation.

Competing interests: PG is a clinical advisor for Pie Medical Imaging and Medis Medical Imaging and has done consultancy for Edwards Life Sciences and Anteris. All other authors have no competing interests to declare.

Provenance and peer review: Not commissioned; externally peer reviewed.

Data availability statement

Data are available on reasonable request. The datasets generated and analysed during the current study are not publicly available. Access to the raw images of patients is not permitted since specialised postprocessing imaging-based solutions can identify the study patients in the future. Data are available from the corresponding author on reasonable request.

Ethics statements

Patient consent for publication

Not applicable.

Ethics approval

This study was conducted according to the principles outlined in the Declaration of Helsinki - —Version 2013. The collection and management of data were approved by the National Research Ethics Service in the UK (21/NE/0149). A pragmatic opt-out informed consent was obtained from all patients included in the study.

References

  • 1. Tse KM, Chang R, Lee HP, et al. A computational fluid dynamics study on geometrical influence of the aorta on haemodynamics. Eur J Cardiothorac Surg 2013;43:829–38. 10.1093/ejcts/ezs388 [DOI] [PubMed] [Google Scholar]
  • 2. Burris NS, Hope MD. 4D flow MRI applications for aortic disease. Magn Reson Imaging Clin N Am 2015;23:15–23. 10.1016/j.mric.2014.08.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Kilner PJ, Yang GZ, Mohiaddin RH, et al. Helical and retrograde secondary flow patterns in the aortic arch studied by three-directional magnetic resonance velocity mapping. Circulation 1993;88:2235–47. 10.1161/01.cir.88.5.2235 [DOI] [PubMed] [Google Scholar]
  • 4. Gimbrone MA, Topper JN, Nagel T, et al. Endothelial dysfunction, hemodynamic forces, and atherogenesis. Ann N Y Acad Sci 2000;902:230–9. 10.1111/j.1749-6632.2000.tb06318.x [DOI] [PubMed] [Google Scholar]
  • 5. Malek AM, Alper SL, Izumo S. Hemodynamic shear stress and its role in atherosclerosis. JAMA 1999;282:2035–42. 10.1001/jama.282.21.2035 [DOI] [PubMed] [Google Scholar]
  • 6. Guala A, Dux-Santoy L, Teixido-Tura G, et al. Wall shear stress predicts aortic dilation in patients with bicuspid aortic valve. JACC Cardiovasc Imaging 2022;15:46–56. 10.1016/j.jcmg.2021.09.023 [DOI] [PubMed] [Google Scholar]
  • 7. Burris NS, Sigovan M, Knauer HA, et al. Systolic flow displacement correlates with future ascending aortic growth in patients with bicuspid aortic valves undergoing magnetic resonance surveillance. Invest Radiol 2014;49:635–9. 10.1097/RLI.0000000000000064 [DOI] [PubMed] [Google Scholar]
  • 8. Vejpongsa P, Xu J, Quinones MA, et al. Differences in cardiac remodeling in left-sided valvular regurgitation: implications for optimal definition of significant aortic regurgitation. JACC Cardiovasc Imaging 2022;15:1730–41. 10.1016/j.jcmg.2022.05.006 [DOI] [PubMed] [Google Scholar]
  • 9. Nayak KS, Nielsen J-F, Bernstein MA, et al. Cardiovascular magnetic resonance phase contrast imaging. J Cardiovasc Magn Reson 2015;17:71. 10.1186/s12968-015-0172-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Juffermans JF, Minderhoud SCS, Wittgren J, et al. Multicenter consistency assessment of valvular flow quantification with 21 automated valve tracking in 4D flow CMR. JACC Cardiovasc Imaging 2021;14:1354–66. 10.1016/j.jcmg.2020.12.014 [DOI] [PubMed] [Google Scholar]
  • 11. Garg P, Swift AJ, Zhong L, et al. Assessment of mitral valve regurgitation by cardiovascular magnetic resonance imaging. Nat Rev Cardiol 2020;17:298–312. 10.1038/s41569-019-0305-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Ebel S, Dufke J, Köhler B, et al. Automated quantitative extraction and analysis of 4D flow patterns in the ascending aorta: an intraindividual comparison at 1.5 T and 3 T. Sci Rep 2020;10:2949. 10.1038/s41598-020-59826-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Meuschke M, Köhler B, Preim U, et al. Semi-automatic vortex flow classification in 4D PC-MRI data of the aorta. Computer Graphics Forum 2016;35:351–60. 10.1111/cgf.12911 [DOI] [Google Scholar]
  • 14. Tyszka JM, Laidlaw DH, Asa JW, et al. Three-dimensional, time- resolved (4D) relative pressure mapping using magnetic resonance imaging. J Magn Reson Imaging 2000;12:321–9. [DOI] [PubMed] [Google Scholar]
  • 15. Sigovan M, Hope MD, Dyverfeldt P, et al. Comparison of four-dimensional flow parameters for quantification of flow eccentricity in the ascending aorta. J Magn Reson Imaging 2011;34:1226–30. 10.1002/jmri.22800 [DOI] [PubMed] [Google Scholar]
  • 16. Ayaon-Albarran A, Fernandez-Jimenez R, Silva-Guisasola J, et al. Systolic flow displacement using 3d magnetic resonance imaging in an experimental model of ascending aorta aneurysm: impact of rheological factors. Eur J Cardiothorac Surg 2016;50:685–92. 10.1093/ejcts/ezw132 [DOI] [PubMed] [Google Scholar]
  • 17. Abe H, Caracciolo G, Kheradvar A, et al. Contrast echocardiography for assessing left ventricular vortex strength in heart failure: a prospective cohort study. Eur Heart J Cardiovasc Imaging 2013;14:1049–60. 10.1093/ehjci/jet049 [DOI] [PubMed] [Google Scholar]
  • 18. von Knobelsdorff-Brenkenhoff F, Karunaharamoorthy A, Trauzeddel RF, et al. Evaluation of aortic blood flow and wall shear stress in aortic stenosis and its association with left ventricular remodeling. Circ: Cardiovascular Imaging 2016;9. 10.1161/CIRCIMAGING.115.004038 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Vellinga A, Cormican M, Hanahoe B, et al. Opt-out as an acceptable method of obtaining consent in medical research: a short report. BMC Med Res Methodol 2011;11:40. 10.1186/1471-2288-11-40 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Garg P, Crandon S, Swoboda PP, et al. Left ventricular blood flow kinetic energy after myocardial infarction - insights from 4D flow cardiovascular magnetic resonance. J Cardiovasc Magn Reson 2018;20:61. 10.1186/s12968-018-0483-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Barker N, Zafar H, Fidock B, et al. Age-associated changes in 4D flow CMR derived tricuspid valvular flow and right ventricular blood flow kinetic energy. Sci Rep 2020;10:9908. 10.1038/s41598-020-66958-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Garg P, van der Geest RJ, Swoboda PP, et al. Left ventricular thrombus formation in myocardial infarction is associated with altered left ventricular blood flow energetics. Eur Heart J Cardiovasc Imaging 2019;20:108–17. 10.1093/ehjci/jey121 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Assadi H, Uthayachandran B, Li R, et al. Kat-ARC accelerated 4D flow CMR: clinical validation for transvalvular flow and peak velocity assessment. Eur Radiol Exp 2022;6:46. 10.1186/s41747-022-00299-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Garg P, Broadbent DA, Swoboda PP, et al. Acute infarct extracellular volume mapping to quantify myocardial area at risk and chronic infarct size on cardiovascular magnetic resonance imaging. Circ Cardiovasc Imaging 2017;10:e006182. 10.1161/CIRCIMAGING.117.006182 [DOI] [PubMed] [Google Scholar]
  • 25. Assadi H, Grafton-Clarke C, Demirkiran A, et al. Mitral regurgitation quantified by CMR 4D-flow is associated with microvascular obstruction post reperfused ST-segment elevation myocardial infarction. BMC Res Notes 2022;15:181. 10.1186/s13104-022-06063-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Bogren HG, Buonocore MH. 4D magnetic resonance velocity mapping of blood flow patterns in the aorta in young vs. Elderly normal subjects. J Magn Reson Imaging 1999;10:861–9. [DOI] [PubMed] [Google Scholar]
  • 27. Barker AJ, Lanning C, Shandas R. Quantification of hemodynamic wall shear stress in patients with bicuspid aortic valve using phase-contrast MRI. Ann Biomed Eng 2010;38:788–800. 10.1007/s10439-009-9854-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Ha H, Koo HJ, Lee JG, et al. Association between flow skewness and aortic dilatation in patients with aortic stenosis. Int J Cardiovasc Imaging 2017;33:1969–78. 10.1007/s10554-017-1196-x [DOI] [PubMed] [Google Scholar]
  • 29. Frydrychowicz A, Berger A, Munoz Del Rio A, et al. Interdependencies of aortic arch secondary flow patterns, geometry, and age analysed by 4-dimensional phase contrast magnetic resonance imaging at 3 Tesla. Eur Radiol 2012;22:1122–30. 10.1007/s00330-011-2353-6 [DOI] [PubMed] [Google Scholar]
  • 30. Oechtering TH, Sieren MM, Hunold P, et al. Time-resolved 3-dimensional magnetic resonance phase contrast imaging (4D flow MRI) reveals altered blood flow patterns in the ascending aorta of patients with valve-sparing aortic root replacement. J Thorac Cardiovasc Surg 2020;159:798–810. 10.1016/j.jtcvs.2019.02.127 [DOI] [PubMed] [Google Scholar]
  • 31. Bürk J, Blanke P, Stankovic Z, et al. Evaluation of 3d blood flow patterns and wall shear stress in the normal and dilated thoracic aorta using flow-sensitive 4D CMR. J Cardiovasc Magn Reson 2012;14:84. 10.1186/1532-429X-14-84 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Lorenz R, Bock J, Barker AJ, et al. 4D flow magnetic resonance imaging in bicuspid aortic valve disease demonstrates altered distribution of aortic blood flow helicity. Magn Reson Med 2014;71:1542–53. 10.1002/mrm.24802 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data are available on reasonable request. The datasets generated and analysed during the current study are not publicly available. Access to the raw images of patients is not permitted since specialised postprocessing imaging-based solutions can identify the study patients in the future. Data are available from the corresponding author on reasonable request.


Articles from Open Heart are provided here courtesy of BMJ Publishing Group

RESOURCES