Abstract
Purpose
This work proposed a convolutional neural network (CNN)‐based method trained with images acquired with electron density phantoms to reduce quantum noise for coronary artery calcium (CAC) scans reconstructed with slice thickness less than 3 mm.
Methods
A DenseNet model was used to estimate quantum noise for CAC scans reconstructed with slice thickness of 0.5, 1.0 and 1.5 mm. Training data was acquired using electron density phantoms in three different sizes. The label images of the CNN model were real noise maps, while the input images of the CNN model were pseudo noise maps. Image denoising was conducted by subtracting the CNN output images from thin‐sliced CAC scans. The efficacy of the proposed method was verified through both phantom study and patient study.
Results
By means of phantom study, the proposed method was proven effective in reducing quantum noise in CAC scans reconstructed with 1.5‐mm slice thickness without causing significant texture change or variation in HU values. With regard to patient study, calcifications were more clear on the denoised CAC scans reconstructed with slice thickness of 0.5, 1.0 and 1.5 mm than on 3‐mm slice images, while over‐smooth changes were not observed in the denoised CAC scans reconstructed with 1.5‐mm slice thickness.
Conclusion
Our results demonstrated that the electron density phantoms can be used to generate training data for the proposed CNN‐based denoising method to reduce quantum noise for CAC scans reconstructed with 1.5‐mm slice thickness. Because anthropomorphic phantom is not a necessity, our method could make image denoising more practical in routine clinical practice.
Keywords: CNN‐based denoising method, coronary artery calcium scan, partial volume effect
1. INTRODUCTION
Partial volume averaging is critical for object that are at the voxel resolution or less, so it is particularly problematic in CT scans for the diagnosis of coronary artery disease, such as coronary CT angiography (CCTA) and coronary artery calcium (CAC) scans. 1 , 2 , 3 In CCTA, images are usually reconstructed with a slice thickness of 0.5−0.6 mm, while the slice thickness used in routine CAC scoring is 3 mm. 4 , 5 , 6 The main coronary arteries are usually between 3 to 4 mm in diameter, so the visibility of coronary artery calcifications can be improved by using a thinner slice thickness. 7 , 8 , 9 , 10 , 11 , 12 However, quantum noise increases with reduced slice thickness. For conventional FBP images, the standard deviation (SD) in Hounsfield units (HU) due to Poisson noise is proportional to . 13 A higher mAs can be used to maintain the same image quality for thin slice reconstruction, but it would result in a higher radiation dose. Because it is imperative to keep radiation dose as low as reasonably achievable (ALARA), CT image denoising is an alternative to reduce the quantum noise due to thin slice reconstruction. Generally, the process of noise suppression is known as image denoising, which has been well studied over the past several decades. 14 , 15 Recently, convolutional neural network (CNN)‐based denoising methods, which attempt to learn a mapping function by optimizing a loss function on a training set that contains degraded‐clean image pairs, have been developed rapidly. 16 , 17 , 18 , 19 Hou et al. have reported that Densely connected convolutional network (DenseNet) can be used to reduce quantum noise in CAC scans reconstructed with 1.5‐mm slice thickness, where the model training was conducted by using datasets acquired with anthropomorphic thorax phantoms instead of actual patient scans to avoid any bias caused by partial volume averaging. 20 The anthropomorphic thorax phantoms could mimic the CT density and attenuation properties of human thorax with variable calcium inserts, but they are not widely available compared with the electron density phantom. In order to make image denoising more practical in routine clinical practice, this work proposed a CNN‐based method trained with images acquired with electron density phantoms to reduce quantum noise for CAC scans reconstructed with slice thickness less than 3 mm.
2. METHODS
2.1. CAC scan
All images were obtained from a 320‐detector row CT system (Aquilion One, Canon, Otawara, Japan). In our department, routine CAC scans are performed using tube current modulation at 120 kVp, where the image quality reference parameter was set as 55 to adjust tube current with a range of 40−300 mA. The gantry rotation speed was 0.35 s/rotation. Images were reconstructed using FBP with soft tissue kernel (FC12). All reconstructions employed a matrix size of 512 × 512 and 22‐cm display field‐of‐view. Phantom studies were carried out using gated protocols, wherein a cardiac simulator was employed to simulate an ECG signal with a mean heart rate of 60 bpm.
2.2. Electron density phantoms
The electron density phantoms (Model 062 M; CIRS, Norfolk, VA, USA), including CIRS_S (180 mm in diameter), CIRS_M (230 mm in diameter) and CIRS_L (320 mm × 270 mm), were used to generate data for model training (Figure 1). CIRS_S houses 9 rod inserts simulating lung, adipose, breast, muscle, liver, and bone with calcium hydroxyapatite (HA) density of 200, 800, and 1250 mg/cc. The rods were 30 mm in diameter and 50 mm in length. CIRS_M was constructed by covering CIRS_S with a 2.5‐cm‐thick bolus (Superflab Bolus; Radiation Products Design Inc, Albertville, MN, USA). CIRS_L consists of CIRS_S and an additional annulus made of soft tissue equivalent epoxy resin. CAC scan was first performed with a phantom housing all the rod inserts. Next, CAC scan was performed after removing two rods from the phantom. This process was repeated until all the rods marked with dashed circles in Figure 1 were all removed, so a total of 15 data sets were obtained for CNN training (3 phantoms × 5 scans). The volume CT dose index (CTDIvol) was 1.1, 2.0, 9.7 mGy for CIRS_S, CIRS_M and CIRS_L, respectively, which were resulted from a CAC scan with tube current of 40, 80, 300 mA, respectively.
FIGURE 1.

Electron density phantoms used for model training: (a) CIRS_S, (b) CIRS_M, (c) CIRS_L.
2.3. CNN model
The DenseNet model proposed by Tong et al. was used in this work to estimate quantum noise for CAC scans reconstructed with slice thickness of 0.5 mm (I0.5mm), 1.0 mm (I1.0mm) and 1.5 mm (I1.5mm). 21 The label image of the CNN model was the difference between 3‐mm sliced CAC scan (I3mm)) and Iz (z = 0.5, 1.0 or 1.5 mm), that is, the real noise map (). The input image of the CNN model was a pseudo noise map (), which was the difference between the original Iz and a smoothed Iz. To create the smoothed Iz, Iz was averaged every 2 × 2 pixels to reduce the matrix size from 512 × 512 to 256 × 256 and then interpolated by bi‐cubic interpolation to increase the matrix size from 256 × 256 to 512 × 512. The root mean square error (RMSE) was the loss function adopted to minimize the difference between CNN generated noise map () and . By utilizing RMSE as the loss function, the preference is given to achieving a high peak signal‐to‐noise ratio (PSNR). The filter weights for each layer were initialized using the MSRA (Microsoft Research Asia) filler technique, while all biases were initialized to zero. The models were trained using the Adam (adaptive moment estimation) optimizer with a mini‐batch size of 32, a learning rate of 0.0001, a momentum of 0.9, and a weight decay of 0.0001. The training datasets consisted of approximately 70 020 sub‐images, each measuring 25 × 25 pixels with a stride of 25 pixels. These sub‐images were shuffled by applying random permutation, serving as both input and label images. The CNN models were built by using Caffe (Convolutional Architecture for Fast Feature Embedding) CNN platform on an Ubuntu server. Figure 2 shows the workflow of the proposed CNN‐based denoising method, where was the denoising results for Iz, which was obtained by subtracting from Iz.
FIGURE 2.

Flowchart of the proposed CNN‐based denoising method (z = 0.5, 1.0 or 1.5 mm).
2.4. Anthropomorphic thorax phantoms
The anthropomorphic thorax phantoms (Thorax‐CCI, QRM GmbH, Möhrendorf, Germany), including QRM_S (300 mm × 200 mm), QRM_M (350 mm × 250 mm) and QRM_L (400 mm × 300 mm), were used to verify the efficacy of the proposed image denoising method (Figure 3). QRM_S consists of an anthropomorphic phantom body and a calibration insert, where the calibration insert contains nine cylindrical calcifications that are in three different sizes (1, 3, 5 mm in diameter and height) and three different HA densities (200, 400, and 800 mg/cm3). QRM_M consists of QRM_S and one extension ring, while QRM_L consists of QRM_S and two extension rings. The CTDIvol was 1.1, 1.4, 4.2 mGy for QRM_S, QRM_M and QRM_L, respectively, which were resulted from a CAC scan with tube current of 40, 50, 140 mA.
FIGURE 3.

Anthropomorphic thorax phantoms used for model validation.
2.5. Data analysis for phantom validation
RMSE and PSNR were calculated to quantify the difference between and (z = 0.5, 1.0 and 1.5 mm) over the whole image, defined as
| (1) |
where k is the number of voxels within the whole image;
| (2) |
where HUmax is the maximum HU value of the image. To investigate the quality and reliability of CT images within local sub‐regions, region of interest (ROI) analysis was conducted to calculate the mean and SD within ROIs. Two‐sample t‐test was used to compare HU values within ROIs measured from with those from Iz and . A p < 0.01 was used to determine statistical significance. In the case of electron density phantoms, a circular ROI was positioned at the center of the rod insert. Regarding the thorax phantoms, a threshold of 130 HU was employed to identify ROIs associated with calcification. In addition to the calcification ROIs, a background ROI was delineated on the calibration insert. Dice similarity coefficient was used to compare the calcification pixels in with those in Iz and . Due to the inability to detect 1‐mm‐diameter calcifications in 3‐mm slice images, these calcifications were excluded from data analysis.
2.6. Patient study
A retrospective analysis of patients who underwent CAC scans at our institution was conducted to evaluate the efficacy of the proposed method on real patients. Permission to access patient data was obtained following approval from the Institutional Human Research Ethics Committee. The Agatston score was calculated by multiplying the calcification area with a specific weighting factor, which is 1 for calcifications ranging from 130−199 HU, 2 for 200−299 HU, 3 for 300−399 HU and 4 for calcifications exceeding 400 HU. Therefore, the number of calcification pixels assigned for different weighting factors was measured on I3mm and (z = 0.5, 1.0, 1.5 mm) of clinical patients.
3. RESULTS
3.1. Validation with electron density phantoms
Figure 4 shows the RMSE and PSNR between and (z = 0.5, 1.0 and 1.5 mm) under different training iterations for CIRS_S, CIRS_M and CIRS_L. After conducting 80 000 iterations, no significant enhancement in RMSE and PSNR was observed. Therefore, for this study, the CNN model trained for 10 0000 iterations was utilized. Figure 5 demonstrates an axial plane of , , (z = 0.5, 1.0, 1.5 mm) and for CIRS_S. Figure 6 presents the corresponding outcomes for CIRS_M, while Figure 7 displays the results for CIRS_L. Table 1 summaries the analysis results of HU values for rod inserts in three electron density phantoms simulating various tissue materials on , and (z = 0.5, 1.0, 1.5 mm). The p‐values were acquired by conducting a two‐sample t‐test to assess the disparity in HU values between and the remaining images.
FIGURE 4.

RMSE (solid line, left axis) and PSNR (dashed line, right axis) between (a) and , (b) and , (c) and under different training iterations for CIRS_S, CIRS_M, CIRS_L (from top to bottom).
FIGURE 5.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for CIRS_S (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
FIGURE 6.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for CIRS_M (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
FIGURE 7.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for CIRS_L (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
TABLE 1.
Analysis results of HU values for rod inserts in CIRS_S, CIRS_M, CIRS_L.
|
|
|
|
|
|
|
|
||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean (HU) | SD (HU) | Mean (HU) | SD (HU) | p‐ Value | Mean (HU) | SD (HU) | p‐Value | Mean (HU) | SD (HU) | p‐ Value | Mean (HU) | SD (HU) | p‐ Value | Mean (HU) | SD (HU) | p ‐Value | Mean (HU) | SD (HU) | p‐ Value | |||||||||
| CIRS_S | Lung inhale | −829.59 | 22.87 | −828.43 | 42.77 | 0.20 | −829.10 | 37.27 | 0.55 | −829.92 | 30.86 | 0.65 | −828.58 | 19.06 | 0.07 | −829.39 | 20.35 | 0.73 | −830.24 | 21.01 | 0.26 | |||||||
| Lung exhale | −481.16 | 27.53 | −479.23 | 52.64 | 0.09 | −478.90 | 41.58 | 0.02 | −480.08 | 34.64 | 0.20 | −479.72 | 22.45 | 0.03 | −480.01 | 23.30 | 0.09 | −480.93 | 23.23 | 0.74 | ||||||||
| Adipose | −72.86 | 27.69 | −72.38 | 57.74 | 0.69 | −74.23 | 45.30 | 0.17 | −73.19 | 37.54 | 0.71 | −71.60 | 25.55 | 0.08 | −74.07 | 24.45 | 0.08 | −73.41 | 25.16 | 0.44 | ||||||||
| Breast | −25.93 | 27.80 | −25.80 | 52.30 | 0.91 | −24.65 | 42.30 | 0.18 | −25.45 | 35.28 | 0.57 | −25.75 | 27.05 | 0.80 | −25.68 | 24.33 | 0.72 | −25.97 | 24.80 | 0.95 | ||||||||
| Muscle | 56.77 | 25.29 | 57.29 | 51.97 | 0.63 | 56.33 | 41.15 | 0.63 | 56.23 | 34.08 | 0.50 | 58.10 | 27.19 | 0.06 | 56.39 | 23.61 | 0.56 | 56.20 | 23.60 | 0.38 | ||||||||
| Liver | 71.83 | 24.31 | 71.71 | 50.61 | 0.91 | 72.93 | 42.26 | 0.23 | 72.22 | 34.00 | 0.62 | 72.91 | 22.06 | 0.08 | 72.75 | 23.95 | 0.15 | 72.08 | 23.24 | 0.69 | ||||||||
| Bone 200 mg/cc | 257.64 | 37.71 | 256.16 | 86.57 | 0.41 | 257.39 | 63.40 | 0.86 | 258.22 | 51.67 | 0.63 | 256.82 | 45.07 | 0.46 | 256.98 | 36.97 | 0.51 | 257.22 | 36.13 | 0.67 | ||||||||
| Bone 800 mg/cc | 1037.05 | 40.51 | 1035.98 | 69.16 | 0.48 | 1036.66 | 60.24 | 0.77 | 1036.88 | 50.74 | 0.89 | 1036.06 | 34.70 | 0.32 | 1036.62 | 33.58 | 0.66 | 1036.24 | 34.80 | 0.42 | ||||||||
| Bone 1250 mg/cc | 1577.16 | 42.04 | 1575.85 | 74.82 | 0.42 | 1576.10 | 63.77 | 0.46 | 1576.34 | 55.37 | 0.53 | 1577.75 | 38.77 | 0.58 | 1576.87 | 40.47 | 0.79 | 1576.14 | 41.08 | 0.36 | ||||||||
| CIRS_M | Lung inhale | −822.11 | 25.07 | −822.83 | 45.57 | 0.46 | −822.48 | 38.68 | 0.67 | −822.03 | 34.02 | 0.92 | −822.17 | 20.22 | 0.92 | −822.95 | 21.21 | 0.17 | −822.39 | 22.92 | 0.66 | |||||||
| Lung exhale | −467.80 | 27.58 | −465.65 | 56.58 | 0.07 | −467.15 | 43.44 | 0.50 | −467.10 | 36.02 | 0.41 | −467.44 | 25.06 | 0.60 | −468.91 | 23.07 | 0.10 | −468.13 | 23.95 | 0.63 | ||||||||
| Adipose | −68.26 | 32.70 | −69.11 | 77.37 | 0.59 | −69.31 | 53.95 | 0.38 | −68.22 | 45.87 | 0.97 | −69.14 | 32.24 | 0.31 | −69.50 | 28.29 | 0.13 | −69.40 | 30.10 | 0.17 | ||||||||
| Breast | −22.65 | 29.55 | −23.08 | 54.24 | 0.71 | −23.87 | 43.82 | 0.22 | −23.74 | 38.33 | 0.23 | −22.56 | 28.80 | 0.91 | −24.18 | 25.63 | 0.04 | −24.08 | 26.73 | 0.06 | ||||||||
| Muscle | 44.62 | 31.71 | 43.85 | 64.57 | 0.57 | 44.71 | 52.47 | 0.94 | 44.92 | 44.07 | 0.77 | 44.80 | 27.77 | 0.82 | 44.48 | 28.53 | 0.86 | 44.24 | 29.46 | 0.64 | ||||||||
| Liver | 73.83 | 28.49 | 72.73 | 53.99 | 0.34 | 73.37 | 46.03 | 0.65 | 73.00 | 39.17 | 0.37 | 73.70 | 23.38 | 0.86 | 73.10 | 24.42 | 0.30 | 72.92 | 25.66 | 0.21 | ||||||||
| Bone 200 mg/cc | 274.56 | 40.76 | 270.77 | 78.54 | 0.02 | 272.09 | 60.93 | 0.07 | 273.71 | 52.89 | 0.07 | 271.47 | 35.36 | * | 271.85 | 33.58 | 0.01 | 273.79 | 36.04 | 0.05 | ||||||||
| Bone 800 mg/cc | 955.10 | 40.85 | 957.29 | 89.16 | 0.23 | 955.01 | 66.13 | 0.95 | 955.68 | 56.79 | 0.66 | 957.78 | 43.44 | 0.02 | 955.45 | 37.76 | 0.74 | 954.91 | 39.23 | 0.86 | ||||||||
| Bone 1250 mg/cc | 1532.88 | 43.65 | 1531.80 | 74.32 | 0.50 | 1534.10 | 65.51 | 0.41 | 1532.20 | 57.97 | 0.62 | 1531.98 | 38.19 | 0.41 | 1534.40 | 39.55 | 0.17 | 1532.66 | 42.50 | 0.84 | ||||||||
| CIRS_L | Lung inhale | −792.05 | 29.97 | −793.54 | 55.86 | 0.21 | −792.75 | 44.81 | 0.49 | −792.91 | 39.34 | 0.36 | −792.95 | 28.41 | 0.25 | −793.41 | 26.14 | 0.07 | −793.37 | 27.51 | 0.09 | |||||||
| Lung exhale | −463.03 | 27.22 | −465.21 | 51.81 | 0.05 | −462.96 | 41.99 | 0.94 | −463.78 | 36.39 | 0.38 | −464.85 | 26.12 | 0.01 | −464.04 | 24.09 | 0.14 | −464.51 | 25.71 | 0.04 | ||||||||
| Adipose | −64.57 | 38.76 | −65.97 | 68.69 | 0.34 | −64.51 | 57.03 | 0.97 | −64.77 | 49.76 | 0.86 | −65.56 | 39.68 | 0.34 | −64.81 | 38.04 | 0.81 | −65.31 | 37.92 | 0.47 | ||||||||
| Breast | −1.10 | 30.99 | −1.87 | 60.90 | 0.55 | −2.85 | 49.94 | 0.11 | −2.02 | 41.96 | 0.35 | −1.29 | 33.41 | 0.82 | −3.25 | 30.31 | 0.01 | −2.43 | 30.31 | 0.10 | ||||||||
| Muscle | 54.83 | 38.07 | 54.39 | 68.21 | 0.76 | 55.93 | 58.08 | 0.40 | 54.40 | 50.90 | 0.72 | 55.77 | 33.43 | 0.33 | 55.54 | 33.89 | 0.46 | 53.51 | 35.79 | 0.18 | ||||||||
| Liver | 78.95 | 34.16 | 79.10 | 57.33 | 0.91 | 79.79 | 51.67 | 0.47 | 79.25 | 43.50 | 0.77 | 79.93 | 28.65 | 0.25 | 78.84 | 30.72 | 0.89 | 78.35 | 31.28 | 0.49 | ||||||||
| Bone 200 mg/cc | 289.87 | 44.79 | 290.68 | 87.91 | 0.66 | 288.80 | 65.10 | 0.47 | 288.44 | 57.64 | 0.30 | 291.68 | 47.79 | 0.14 | 288.43 | 41.48 | 0.21 | 287.64 | 42.96 | 0.06 | ||||||||
| Bone 800 mg/cc | 924.29 | 42.09 | 922.85 | 73.96 | 0.37 | 924.56 | 57.23 | 0.84 | 923.94 | 50.75 | 0.78 | 922.84 | 38.76 | 0.18 | 923.68 | 36.31 | 0.56 | 922.71 | 37.81 | 0.14 | ||||||||
| Bone 1250 mg/cc | 1437.49 | 48.52 | 1437.21 | 85.16 | 0.88 | 1437.29 | 70.14 | 0.90 | 1438.30 | 62.09 | 0.58 | 1436.52 | 42.36 | 0.42 | 1435.63 | 41.62 | 0.12 | 1436.71 | 45.38 | 0.53 | ||||||||
Abbreviations: HU, Hounsfield units; SD, standard deviation.
3.2. Validation with anthropomorphic thorax phantoms
Figure 8 demonstrates an axial plane of , , (z = 0.5, 1.0, 1.5 mm) and for QRM_S. Figure 9 presents the corresponding outcomes for QRM_M, while Figure 10 displays the results for QRM_L. The RMSE and PSNR between and for QRM phantoms are shown in Supplement 1. Figure 11 shows the Dice similarity coefficients calculated by comparing the calcifications pixels in with those in and (z = 0.5, 1.0, 1.5 mm) for calcifications in three anthropomorphic thorax phantoms with density of , , . Table 2 summaries the analysis results of HU values within background and calcification ROIs on , and (z = 0.5, 1.0, 1.5 mm) for three electron density phantoms. The p‐values were acquired by conducting a two‐sample t‐test to assess the disparity in HU values between and the remaining images.
FIGURE 8.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for QRM_S (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
FIGURE 9.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for QRM_M (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
FIGURE 10.

Axial plane of (a) , (b) , (c) , (d) , (e) , (f) , (g) , (h) , (i) , (j) for QRM_L (window level/window width = 0/1000 HU for (a)−(d), (h)−(j) and 0/340 HU for (e)−(g)). HU, Hounsfield units.
FIGURE 11.

Dice similarity coefficients calculated by comparing calcification pixels in with those in , , , , , for calcifications in QRM_S, QRM_M and QRM_L (from top to bottom) with density of (a) , (b) , (c) .
TABLE 2.
Analysis results of HU values within background and calcification ROIs for QRM_S, QRM_M, QRM_L.
|
|
|
|
|
|
|
|
||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean (HU) | SD (HU) | Mean (HU) | SD (HU) | p‐Value | Mean (HU) | SD (HU) | p‐Value | Mean (HU) | SD (HU) | p‐Value | Mean (HU) | SD (HU) | p‐Value | Mean (HU) | SD (HU) | p‐ Value | Mean (HU) | SD (HU) | p‐ Value | |||||||||
| QRM_S | Background | 41.65 | 22.77 | 41.70 | 48.73 | 0.90 | 41.72 | 37.89 | 0.83 | 41.74 | 30.89 | 0.75 | 41.61 | 19.94 | 0.83 | 41.05 | 19.12 | 0.01 | 41.20 | 19.72 | 0.04 | |||||||
|
|
243.41 | 52.49 | 248.56 | 63.12 | 0.46 | 256.17 | 67.28 | 0.08 | 256.56 | 62.94 | 0.06 | 219.73 | 49.86 | * | 230.21 | 55.76 | 0.04 | 238.67 | 55.99 | 0.46 | ||||||||
|
|
372.45 | 168.43 | 382.48 | 163.82 | 0.51 | 377.56 | 165.63 | 0.74 | 373.24 | 166.89 | 0.96 | 375.87 | 152.54 | 0.82 | 367.49 | 155.05 | 0.74 | 363.73 | 159.81 | 0.56 | ||||||||
|
|
611.27 | 383.27 | 618.06 | 385.29 | 0.83 | 622.14 | 387.07 | 0.73 | 606.71 | 382.79 | 0.89 | 617.07 | 357.76 | 0.85 | 628.07 | 375.05 | 0.59 | 606.67 | 372.55 | 0.88 | ||||||||
| QRM_M | Background | 40.26 | 21.77 | 40.07 | 46.79 | 0.64 | 40.11 | 36.25 | 0.63 | 40.32 | 29.71 | 0.84 | 40.20 | 19.33 | 0.80 | 39.69 | 18.63 | 0.01 | 39.81 | 19.04 | 0.04 | |||||||
|
|
231.04 | 56.89 | 251.64 | 68.11 | 0.01 | 247.50 | 56.87 | 0.02 | 237.02 | 54.13 | 0.39 | 219.56 | 47.16 | 0.08 | 222.99 | 47.79 | 0.22 | 222.49 | 48.42 | 0.19 | ||||||||
|
|
368.88 | 146.18 | 382.59 | 148.77 | 0.33 | 382.48 | 146.64 | 0.33 | 380.65 | 147.63 | 0.40 | 366.74 | 143.43 | 0.88 | 366.37 | 139.73 | 0.85 | 365.81 | 140.46 | 0.82 | ||||||||
|
|
591.05 | 353.91 | 605.40 | 353.69 | 0.63 | 603.92 | 356.67 | 0.67 | 602.45 | 355.67 | 0.71 | 594.72 | 329.81 | 0.90 | 590.47 | 340.61 | 0.98 | 593.69 | 346.41 | 0.93 | ||||||||
| QRM_L | Background | 47.33 | 21.33 | 47.30 | 46.47 | 0.92 | 47.57 | 34.47 | 0.40 | 47.26 | 29.27 | 0.80 | 47.42 | 18.85 | 0.65 | 47.06 | 17.62 | 0.17 | 46.86 | 18.80 | 0.02 | |||||||
|
|
225.72 | 42.59 | 238.26 | 60.70 | 0.06 | 234.70 | 47.89 | 0.12 | 234.17 | 45.74 | 0.13 | 213.66 | 53.49 | 0.05 | 213.23 | 44.72 | 0.03 | 225.27 | 42.37 | 0.93 | ||||||||
|
|
348.57 | 144.18 | 367.84 | 141.24 | 0.15 | 367.14 | 145.46 | 0.18 | 363.57 | 142.65 | 0.27 | 354.33 | 137.49 | 0.67 | 351.53 | 141.17 | 0.83 | 353.13 | 139.03 | 0.73 | ||||||||
|
|
566.63 | 320.64 | 581.63 | 326.44 | 0.59 | 580.33 | 328.22 | 0.62 | 581.83 | 326.25 | 0.58 | 563.23 | 307.02 | 0.90 | 570.16 | 319.13 | 0.90 | 568.83 | 318.48 | 0.94 | ||||||||
Abbreviation: HU, Hounsfield units; ROI, region of interest; SD, standard deviation.
* The asterisk indicates a p‐value < 0.01.
3.3. Evaluation with patient scans
Figure 12 demonstrates an axial plane of and (z = 0.5, 1.0, 1.5 mm) for three individual patients. The CTDIvol was 2.6, 5.2, 7.6 mGy for patient #1 (body mass index (BMI) = 27 kg/m2), patient #2 (BMI = 31.5 kg/m2), patient #3 (BMI = 38.2 kg/m2), respectively, which were resulted from a CAC scan with tube current of 120, 210, 300 mA. The pixels of the calcifications indicted by yellow arrow were classified into 4 HU groups, and the results were shown in Figure 13. For the first patient, the total number of calcification pixels detected on , , , were 59, 95, 86, 65, respectively. For the second patient, the total number of calcification pixels detected on , , , were 75, 91, 96, 88, respectively. For the third patient, the total number of calcification pixels detected on , , , were 83, 105, 99, 88, respectively. An axial plane of , , and for three individual patients are shown in Supplement 2–4.
FIGURE 12.

Axial plane (window level/window width = 0/1000 HU) of (a) , (b) , (c) , (d) for patient #1, patient #2 and patient #3 (from top to bottom). HU, Hounsfield units.
FIGURE 13.

Comparing the number of calcification pixels for (a) patient #1, (b) patient #2, (c) patient #3.
4. DISCUSSION
Several deep learning methods have been used to reduce the quantum noise in CT imaging, including discriminative models, generative models, and hybrid models. 16 , 17 , 18 , 19 The DenseNet model used in this work was based on the discriminative approach that represents bottom‐up execution to define a decision boundary that divides the feature space into areas containing feature vectors belong to the same class. 22 The discriminative models have been demonstrated to be able to reduce the streaking artifacts tremendously while preserving the fine structures in the processed CT images. 17 , 18 , 19 However, it has also been reported that the discriminative models would cause over‐smooth changes in the image due to the least squares objective function. 16 , 17 Based on our experience, when the electron density phantoms were used to acquire datasets for model training, the CNN output images were quite blurry if DenseNet was trained with (z = 0.5, 1.0, 1.5 mm) as input and as label. In our opinion, this phenomenon may be due to the lack of high‐resolution insert in the electron density phantoms. Usually, model training was conducted by using datasets acquired from similar scenarios that is encountered in the denoising task. Hence, the degradation in spatial resolution was not observed in the denoising method trained with images acquired with the anthropomorphic thorax phantoms. Divel et al. have proposed an image domain noise insertion method for CT images, in which white noise was filtered, scaled and added to the image to simulate low‐dose scans. 23 Their work inspired us to start from the following aspects: (1) it may be possible to conduct image denoising by subtracting quantum noise from images; (2) quantum noise contains less anatomical details compared to CT images. Hence, the workflow shown in Figure 2 was proposed to reduce quantum noise in thin‐sliced CAC scans, where the input and label of DenseNet were and (z = 0.5, 1.0, 1.5 mm), respectively.
For the validation with electron density phantoms, shows the highest statistical fluctuations, followed by and . Based on naked eye observation, there was a noticeable difference in overall texture between and , but the difference was less obvious between and and was barely discernable between and . With regard to the results of ROI analysis, the SD values measured from were lower than those from Iz (z = 0.5, 1.0, 1.5 mm), while the largest difference in mean value between Iz and was 3.09 HU. No statistically significant difference was found in CIRS_S and CIRS_L. As for CIRS_M, statistically significant difference was found when comparing the HU values of bone insert with HA density of 200 mg/cc between and .
For the validation with anthropomorphic thorax phantoms, shows the highest statistical fluctuations, followed by and . Based on naked eye observation, the difference in background texture was noticeable between and , which was less obvious between and and was barely discernable between and . The calcifications with density of and identified on (z = 0.5, 1.0 and 1.5 mm) have little differences in terms of shape and sharpness when compared to those identified on . As for the calcifications with density of , calcification blur was observed in , which was less obvious in and was barely discernable in . The Dice similarity coefficients for calcifications with density of obtained by comparing and were usually higher than those obtained by comparing and or comparing and . This phenomenon was also found in the calcifications with density of and . With regard to the results of ROI analysis, the SD values measured from were lower than those from Iz (z = 0.5, 1.0, 1.5 mm). No statistically significant difference was found in any QRM phantom at the background region. As for the calcifications, there was no statistically significant difference in QRM_M and QRM_L, while statistically significant difference was observed in QRM_S when comparing the HU values of calcifications with density of identified on to those on .
By means of phantom studies, the proposed method was validated effective in reducing quantum noise in I1.5mm without causing significant texture change or variation in HU values. In order to further confirm this finding, the proposed method underwent evaluation using actual patient data. Based on naked eye observation, over‐smooth changes were observed at the background region in . This phenomenon was less obvious in , while no serious texture changes were observed in . The calcifications were more clear on (z = 0.5, 1.0, 1.5 mm) than on , which can be confirm by the number of detected calcification pixels. Overall, this work demonstrated that the electron density phantoms can be used to generate training data for the proposed CNN‐based denoising method to reduce quantum noise for CAC scans reconstructed with 1.5‐mm slice thickness. Because anthropomorphic phantom is not a necessity, our method could make image denoising more practical in routine clinical practice. However, more clinical trials are necessary to provide evidence for the effectiveness of the proposed method.
5. CONCLUSION
This work proposed a CNN‐based method trained with images acquired with electron density phantoms to reduce quantum noise for CAC scans reconstructed with slice thickness less than 3 mm. By means of phantom studies, the proposed method was proven effective in reducing quantum noise in I1.5mm without causing significant texture change or variation in HU values. With regard to patient study, the calcifications were more clear on (z = 0.5, 1.0, 1.5 mm) than on , while over‐smooth changes were not observed in . Our results demonstrated that the electron density phantoms can be used to generate training data for the proposed CNN‐based denoising method to reduce quantum noise for CAC scans reconstructed with 1.5‐mm slice thickness. Because anthropomorphic phantom is not a necessity, our method could make image denoising more practical in routine clinical practice.
AUTHOR CONTRIBUTIONS
Ching‐Ching Yang: Conceptualization; methodology; software; validation; formal analysis; investigation; resources; data curation; writing; visualization; supervision. Kuei‐Yuan Hou: Formal analysis; resources; data curation; writing; project administration.
CONFLICT OF INTEREST STATEMENT
The authors declare no conflicts of interest.
Supporting information
Supporting Information
ACKNOWLEDGMENTS
This work was supported by a grant from the Kaohsiung Medical University Research Foundation (KMU‐M113022).
Yang CC, Hou KY. A CNN‐based denoising method trained with images acquired with electron density phantoms for thin‐sliced coronary artery calcium scans. J Appl Clin Med Phys. 2024;25:e14287. 10.1002/acm2.14287
REFERENCES
- 1. Barrett JF, Keat N. Artifacts in CT: recognition and avoidance. Radiographics. 2004;24(6):1679‐1691. [DOI] [PubMed] [Google Scholar]
- 2. Kalisz K, Buethe J, Saboo SS, Abbara S, Halliburton S, Rajiah P. Artifacts at Cardiac CT: physics and solutions. Radiographics. 2016;36(7):2064‐2083. [DOI] [PubMed] [Google Scholar]
- 3. Šprem J, de Vos BD, Lessmann N, et al. Coronary calcium scoring with partial volume correction in anthropomorphic thorax phantom and screening chest CT images. PLoS One. 2018;13(12):e0209318. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Jacobs JE, Boxt LM, Desjardins B, Fishman EK, Larson PA, Schoepf J. ACR practice guideline for the performance and interpretation of cardiac computed tomography (CT). J Am Coll Radiol. 2006;3(9):677‐685. [DOI] [PubMed] [Google Scholar]
- 5. Halpern EJ. Clinical applications of cardiac CT angiography. Insights into Imaging. 2010;1(4):205‐222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Serruys PW, Hara H, Garg S, et al. Coronary computed tomographic angiography for complete assessment of coronary artery disease: JACC state‐of‐the‐art review. J Am Coll Cardiol. 2021;78(7):713‐736. [DOI] [PubMed] [Google Scholar]
- 7. Dodge JT Jr, Brown BG, Bolson EL, Dodge HT. Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation. 1992;86(1):232‐246. [DOI] [PubMed] [Google Scholar]
- 8. Mühlenbruch G, Thomas C, Wildberger JE, et al. Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in vivo. Invest Radiol. 2005;40(11):695‐699. [DOI] [PubMed] [Google Scholar]
- 9. Fukumoto W, Nagaoka M, Higaki T, et al. Measurement of coronary artery calcium volume using ultra‐high‐resolution computed tomography: a preliminary phantom and cadaver study. European journal of radiology open. 2020;7:100253. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Urabe Y, Yamamoto H, Kitagawa T, et al. Identifying small coronary calcification in non‐contrast 0.5‐mm slice reconstruction to diagnose coronary artery disease in patients with a conventional zero coronary artery calcium score. J Atheroscler Thromb. 2016;23(12):1324‐1333. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. van der Bijl N, de Bruin PW, Geleijns J, et al. Assessment of coronary artery calcium by using volumetric 320‐row multi‐detector computed tomography: comparison of 0.5 mm with 3.0 mm slice reconstructions. Int J Cardiovasc Imaging. 2010;26(4):473‐482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Aslam A, Khokhar US, Chaudhry A, et al. Assessment of isotropic calcium using 0.5‐mm reconstructions from 320‐row CT data sets identifies more patients with non‐zero Agatston score and more subclinical atherosclerosis than standard 3.0‐mm coronary artery calcium scan and CT angiography. J Cardiovasc Comput Tomogr. 2014;8(1):58‐66. [DOI] [PubMed] [Google Scholar]
- 13. Boas F, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012;4. [Google Scholar]
- 14. Fan L, Zhang F, Fan H, Zhang C. Brief review of image denoising techniques. Vis Comput Ind Biomed Art. 2019;2(1):7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Diwakar M, Kumar M. A review on CT image noise and its denoising. Biomed Signal Process Control. 2018;42:73‐88. [Google Scholar]
- 16. Kulathilake KASH, Abdullah NA, Sabri AQM, Lai KW. A review on deep learning approaches for low‐dose computed tomography restoration. Complex & Intelligent Systems. 2023;9(3):2713‐2745. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Huang L, Jiang H, Li S, Bai Z, Zhang J. Two stage residual CNN for texture denoising and structure enhancement on low dose CT image. Comput Methods Programs Biomed. 2020;184:105115. [DOI] [PubMed] [Google Scholar]
- 18. Jiang X, Jin Y, Yao Y. Correction to: low‐dose CT lung images denoising based on multiscale parallel convolution neural network. The Visual Computer. 2021;37:1‐1. [Google Scholar]
- 19. Shiri I, Akhavanallaf A, Sanaat A, et al. Ultra‐low‐dose chest CT imaging of COVID‐19 patients using a deep residual neural network. Eur Radiol. 2021;31(3):1420‐1431. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Hou KY, Yang CC. Investigating the feasibility of using DenseNet to improve coronary calcification detection in CT. Acad Radiol. 2022;30(8):1600‐1613. [DOI] [PubMed] [Google Scholar]
- 21. Tong T, Li G, Liu X, Gao Q. Image super‐resolution using dense skip connections. In: Paper presented at: 2017 IEEE International Conference on Computer Vision (ICCV); 22–29 Oct. 2017. 2017.
- 22. Shrestha A, Mahmood A. Review of deep learning algorithms and architectures. IEEE Access. 2019;7:53040‐53065. [Google Scholar]
- 23. Divel SE, Pelc NJ. Accurate image domain noise insertion in CT images. IEEE Trans Med Imaging. 2020;39(6):1906‐1916. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information