QRS Detector Performance Evaluation Aware of Temporal Accuracy and Presence of Noise

Wojciech Reklewski 1, Marek Miśkowicz 1, Piotr Augustyniak 1,*
Editors: Steve Ling1, Georg Fischer1
PMCID: PMC10934794  PMID: 38475235

Abstract

Algorithms for QRS detection are fundamental in the ECG interpretive processing chain. They must meet several challenges, such as high reliability, high temporal accuracy, high immunity to noise, and low computational complexity. Unfortunately, the accuracy expressed by missed or redundant events statistics is often the only parameter used to evaluate the detector’s performance. In this paper, we first notice that statistics of true positive detections rely on researchers’ arbitrary selection of time tolerance between QRS detector output and the database reference. Next, we propose a multidimensional algorithm evaluation method and present its use on four example QRS detectors. The dimensions are (a) influence of detection temporal tolerance, tested for values between 8.33 and 164 ms; (b) noise immunity, tested with an ECG signal with an added muscular noise pattern and signal-to-noise ratio to the effect of “no added noise”, 15, 7, 3 dB; and (c) influence of QRS morphology, tested on the six most frequently represented morphology types in the MIT-BIH Arrhythmia Database. The multidimensional evaluation, as proposed in this paper, allows an in-depth comparison of QRS detection algorithms removing the limitations of existing one-dimensional methods. The method enables the assessment of the QRS detection algorithms according to the medical device application area and corresponding requirements of temporal accuracy, immunity to noise, and QRS morphology types. The analysis shows also that, for some algorithms, adding muscular noise to the ECG signal improves algorithm accuracy results.

Keywords: electrocardiogram (ECG), QRS detection, ECG interpretive software testing, biomedical signal processing, MIT-BIH Arrhythmia Database

1. Introduction

QRS detectors are widely used as a front end of various ECG processing chains; consequently, their usage is concerned virtually in any ECG-dedicated software. Moreover, from the signal processing viewpoint, they serve as an interface between a continuous series of samples captured by recording devices and a feature-based algorithm used for classification and detection purposes. Consequently, in several ECG software packages, the QRS detector and preceding filters are the unique procedures involving every sample of the record. Therefore, high accuracy of detection and low computational complexity are of key importance and are usually regarded as primary quality factors. Detection accuracy, however, becomes a less distinctive factor as many modern detectors approach 99.9%.

Mobile and wearable ECG data acquisition systems face an inherent conflict between autonomy time and usability. The time between battery recharge, power consumption, detection accuracy, size, and weight of the device have to be balanced by a compromise. To this end, two important quality aspects appear to produce a reliable estimate of the suitability of a particular QRS detection algorithm to an area of application. The first aspect is the temporal stability of the detection point sequence (that is, its independence from the QRS morphology type). With the QRS detector producing highly accurate detection points, one could avoid recalculations of more precise QRS peak positions and directly input the detection points to procedures such as HRV analysis or shape classifying. The second point is the noise immunity of the detection point sequence, which specifies how much noise (e.g., of muscular origin), present for example in wearable-based in-field ECG measurements, affect the precision of QRS detection and alter the medical findings produced by subsequent diagnostic procedures.

In this paper, we address the problem of the comprehensive evaluation of QRS detection performance using multidimensional criteria, including the temporal tolerance of QRS complex detection, robustness amid noise, and sensitivity to QRS morphology. The proposed methodology enables the selection of the QRS detectors for specific applications, such as clinical data analysis, long-term monitoring with mobile devices and wearables, fast results, or arrhythmia diagnosis. The analysis is demonstrated in the example of four selected QRS detection algorithms implemented in Python according to their specification in the relevant references [1,2,3,4]. The algorithms are tested on the MIT-BIH Arrhythmia Database (MIT-BIH AD) [5] with added muscular noise from the MIT-BIH Noise Stress Test Database (MIT-BIH NSTD) [6]. The tests were conducted on Dell Latitude E6400, Intel Core2Duo P8400, 2.26 GHz, and 4 GB RAM running with Debian 10.13. Implementation of the algorithms, test tools, and data processing were performed in Python 3.7.3. Plots were created in Jupyter Notebook (server v5.7.8 with Python 3.7.3 [GCC 8.3.0]).

The performance analysis shows that the QRS detectors demonstrate different sensitivities to the detection of temporal tolerance. Some algorithms maintain high detection accuracy, even for low values of temporal tolerance of QRS detection. The others exhibit good results only for high values of temporal tolerance (i.e., comparable to the QRS standard duration of 100 ms).

The QRS detection performance under extra noise is in general deteriorated. However, as shown in this paper, for some QRS morphologies, extra noise in the ECG signal can paradoxically improve QRS detection accuracy. This effect resembles the improvement of audio and video data by randomizing the quantization error known as dither [7,8].

The performance of the algorithms depends on QRS morphology, as each algorithm uses different ECG signal filtering methods and signal analysis in subsequent processing blocks. Some QRS morphologies are more problematic than others for each algorithm under analysis. In our tests, V-type QRS morphologies were the most problematic for algorithms under analysis.

The paper is organized as follows: Section 2 addresses related work; Section 3 describes the four algorithms under analysis, the database, and the preparation of test datasets; Section 4 presents results; Section 5 is the discussion; and Section 6 contains conclusions.

2. Related Work

Medical testing procedures are conventionally evaluated based on binary classification by calculating parameters such as TP, FN, FP, and TB. These parameters are also commonly used in the literature for performance evaluation and comparison of QRS detection algorithms [9]. True positive (TP) is the number of correctly detected R peaks, false negative (FN) is the number of omitted R-peaks, false positive (FP) is the number of places wrongly classified as R-peaks, and total beats (TB) is the number of annotated R-peaks in a database record. However, much less attention is paid to temporal detection accuracy. The numerical values of TP, FN, and FP depend on detector temporal tolerance (DTT), defined as the maximum allowed time difference between the algorithm detection points (R-peaks) and the corresponding annotations from the reference database [10]. A variety of temporal tolerance values are used in the literature, ranging from 60 ms to 160 ms, which sometimes results in the comparison of algorithms with different temporal resolutions. The problem of the sensitivity of the accuracy of QRS detection algorithms in the temporal resolution of the detection defined by DTT is examined in [10]. When the algorithm is tested with a certain high value of DTT, relatively distant locations of R-peak detection points and database annotation are successfully paired and counted as TP. Further testing with certain lower DTT values will result in the distance between detection points and database annotation exceeding the DTT value and, consequently, their pairing will be unsuccessful. Database annotation without paired detection points will be counted as FN, and algorithm detection without paired database annotation will be counted as FP. This way, the algorithm’s TP detection for higher values of DTT will be replaced by a pair of FN and FP detections for lower DTT values. As expected, the higher the DTT, the better the numerical results of the TP, FN, and FP. However, the deterioration rate of an algorithm’s accuracy depends on the given QRS detection algorithm. Some algorithms demonstrate slow degradation, while others suffer a quick performance drop with decreasing DTT values [10].

A substantial research effort has been dedicated to examining the robustness of QRS detectors’ performance against noise in ECG signal recordings; [11] reviews 38 major state-of-the-art techniques of QRS detection with comprehensive comparative analysis of techniques for ECG signal denoising and QRS detection. In their conclusion, the authors emphasize a need to invent computational techniques “to analyze the ECG signal with higher accuracy in all conditions”, which justifies the development of multidimensional methods to test algorithms’ performance, among other criteria, in various noise conditions and temporal accuracy requirements.

The performance results of the Pan–Tompkins QRS detection algorithm in noisy ambulatory ECG data with varying signal-to-noise ratios are presented in [12]. Two ECG databases are used for testing: the MIT-BIH NSTD [6] and the MIT-BIH AD [5]. The algorithm results for sensitivity (Se = TP/(TP + FN)) and positive predictivity (PPV or +P) (+P = TP/(TP + FP)) deteriorate from close to 100% for noise level with SNR = 24 dB to around 60% and 70% respectively for SNR = −6 dB. The analysis shows that the Pan–Tompkins algorithm needs improvements to achieve good detection performance for noisy signals.

In [13], the performance analysis of selected three well-known QRS detection algorithms is addressed: by Pan–Tompkins [4], WQRS [14], and by Hamilton [15] against the MIT-BIH AD and the noise-contaminated ECG signal with different levels of baseline wander (BW), muscle artifact (MA), and electrode motion (EM) artifact from the MIT-BIH NSTD. As shown in [13], noise and artifacts decreased the quality indices of algorithms from close to 100% for SNR = 12 dB to:

  • BW noise: Se = 95% for WQRS and +P = 62% for WQRS for SNR = −12 dB,

  • MA noise: Se = 83% for Hamilton and +P = 38% for WQRS for SNR = −12 dB,

  • and EM noise: Se = 65% for Hamilton and +P = 30% for WQRS for SNR = −12 dB.

The poorest performance was noted for ECG signals affected by EM artifacts.

In [16], a new QRS detection method is proposed and validated, with different levels of baseline wander, muscle artifact, and electrode motion artifact as noise sources against MIT-BIH NSTD with the following processing blocks: first derivative, Hilbert transform envelope, wavelet transform, wavelet component reduction, signal reconstruction, and thresholding. The proposed QRS detection method achieves Se = 78.89% and +P = 75.25% for MIT-BIH NSTD and SNR = 0 dB.

The performance of three selected state-of-the-art QRS detection algorithms and the evaluation of the accuracy of their R-peak localization are included in [17]. The algorithms under analysis were the following: integrate and fire pulse train automaton [18], zero-crossing counts [19], and the knowledge-based method [20]. The authors propose a method to estimate the temporal accuracy of R-peak detection for normal and abnormal beats as well as a simple scheme to compensate for slackness introduced by the filtering part of the algorithms.

In opposition to the ubiquitous, conventional, TP, FN, and FP parameters mentioned at the beginning of this section, a novel QRS detection performance indicator, jitter with accuracy (JA), aimed at evaluating QRS detection algorithms under realistic noise scenarios, is proposed in [21]. The authors state that Se and +P metrics used to assess the quality of R-peak detection lose information value, where a high temporal tolerance of 100 ms or more [21] is used. Also, frequent use of the MIT-BIH AD, which is, according to the authors, artifact-free, leads to an overestimation of algorithm performance and unjustified reported Se and +P performance indicators significantly above 99%.

The relationship between QRS detection performance and database sampling frequency is examined in [22]. The analysis, carried out for the Hamilton algorithm [23] against the MIT-BIH database, shows that adapting the algorithm threshold parameters to sampling frequency optimizes the algorithm’s accuracy results.

In [24], the five selected multisignal heartbeat detectors are tested against 100 records from the training dataset of the PhysioNet/CinC Challenge 2014, with various noise levels added. The performance results and best-worse ranking of the detectors are reported.

The performance of 10 QRS detection algorithms against six internationally recognized ECG databases with various normal and abnormal beat types and various levels of noise and artifacts is reported in [25]. The tested algorithms were by Pan–Tompkins, Hamilton mean, Hamilton median, RS slope, sixth power, finite state machine (FSM), U3 transform, difference operation (DOM), ‘jqrs’, and optimized knowledge based (OKB). The overall results are reported for each algorithm and each database. The analysis shows that QRS detection results decrease significantly for poor signal-quality ECG signals for all tested algorithms.

A new R-peak detector based on neural networks is proposed in [26]. The algorithm performance does not deteriorate with low-quality or noisy ECG signals acquired from mobile electrocardiogram sensors, such as Holter monitors. The proposed QRS detector uses a 1-D self-organized operational neural network with generative neurons and offers lower computational complexity than conventional 1-D convolutional neural networks. The QRS detection results reported are a 99.10% F1 score (F1 = 2 · PPV · Se/(Se + PPV)), 99.79% Se, and 98.42% +P achieved on the China Physiological Signal Challenge-2020 dataset (CPSC-DB). The CPSC-DB database contains 1 026 095 beats, collected from arrhythmia patients, and includes real-world noise as well as artifacts from a wearable real-world Holter ECG device.

In [27], 10 QRS detection techniques published between 2020 and 2022 are compared based on the performance parameters: Se, PPV, F1 score, and DER (DER = (FN + FP)/TB).

A new R-peak detection technique based on visibility graph transformation, which maps a discrete time series to a graph by expressing each sample as a node and assigning edges between intervisible samples, is proposed in [28]. The proposed method is compared against two existing QRS detection methods on a noisy and sample-accurate University of Glasgow ECG Database [29] with two performance metrics: F1 score and root mean square of successive differences (RMSSD). The result of the first comparison is presented as a boxplot of the proposed F1 score method versus the SWT-based method [30] and matched filter detector [21]. There is an annotation that “the tolerance for deviation from the true R-peaks was 0%”, which corresponds to DTT = 0. In order to evaluate the RMSSD, a Wilcoxon signed-rank test between the estimated RMSSD values and the ground truth was performed and presented.

In [31], the authors present two (FastNVG and FastWHVG) computationally accelerated versions of the visibility graph transformation QRS detection method proposed in [28], together with an extended comparison with additional 7 QRS detectors from the literature.

ECG noise removal techniques are reviewed in [32]. The types and sources of noise are identified, and six major domains of denoising are subsequently explored. The techniques for denoising are presented and their performance is evaluated according to the following parameters: root-mean-square error, percentage-root-mean-square difference, and signal-to-noise ratio improvement.

3. Materials and Methods

To demonstrate the concept of evaluation of QRS detector performance aware of temporal accuracy, the presence of noise, and various QRS morphologies, we implemented four QRS detection algorithms and tested their accuracy, expressed by a true-positive-to-total-beats ratio (TP/TB). The tests are carried out for a range of DTT values and controlled mixing of muscular noise, which is often present in wearables applications. Obviously, it is desirable for QRS detectors to show good TP/TB scores, even for low DTT. Such detectors are particularly welcome, as R-peak location corrections are not necessary, and resynchronization of the heartbeat time series is not needed before further ECG processing steps.

The four algorithms selected from the QRS detection literature consist of three algorithms developed for mobile and wearables applications [1,2,3] and the algorithm broadly used as the reference in QRS detection literature [4]. Algorithms 1 and 2 belong to a group of algorithms based on digital filtering [9]. In addition, they can be classified into a group of “low” computational complexity using the subjective comparison with respect to computational load [9]. In contrast, Algorithm 3 does not use any digital filters and works on a different principle of operation based on a level-crossing sampling of the ECG signal. Algorithm 3 can be assigned to the group of “medium” computational load. Algorithm 4 is based on digital filters [9] and can be assigned to the group of “medium” computational load. Compared with Algorithms 1 and 2, Algorithm 4 uses more mathematical operations performed in multiple stages of signal processing.

Section 3.1, Section 3.2, Section 3.3 and Section 3.4 present the specification of QRS detection algorithms included in the study, while Section 3.5 contains the definition of test signals used to examine their immunity to noise in ECG signals.

3.1. Algorithm 1

Algorithm 1 has been designed to address the requirements of low-power and real-time operations for use in mobile and wearable applications. The algorithm’s two main processing blocks, preprocessing and dynamic thresholding, have been designed to minimize necessary computational resources and power consumption (Figure 1). The preprocessing block consists of three consecutive operations performed on the input digital ECG signal: differentiation, the moving window average, and squaring. The resultant feature signal is fed to the thresholding block, which works in sequence in three states. In State 1, the algorithm searches for the maximum value of the feature signal within a window of fixed length (260 ms). The time instant when the feature signal reaches its maximum is marked as an R-peak temporal location. State 2, following State 1, is a 200 ms wait from the R-peak detection. During State 3, the dynamic threshold is decreased exponentially with time until it reaches the value of the feature signal. State 3 is then terminated and the algorithm moves to State 1, the search for the maximum value of the feature signal. The threshold initial value in State 3 is adaptive and dependent on the average amplitude of all previously found R-peaks.

Figure 1.

Figure 1

Algorithm 1 block diagram based on [1].

3.2. Algorithm 2

Algorithm 2 has been designed to achieve the low computational complexity and high energy efficiency needed for mobile and portable applications. The preprocessing stage consists of two parallel signal processing paths (Figure 2). Both paths include the high-pass moving average filters but with different cutoff frequencies followed by a rectification operation. The moving average window lengths Nlong and Nshort define the cutoff frequencies of the filters. The outputs of the parallel processing paths are the inputs to the decision block. The preprocessed signal u[n], after high-pass filtering with a higher cutoff frequency (moving average window length Nshort), is used to decide when to start the QRS search window. During the search window of a fixed 200 ms length, the preprocessed signal of the other path y[n] after high-pass filtering with a lower cutoff frequency (moving average window length NLong) is analyzed in order to find its maximum value. The time instant of the maximum value of the y[n] signal is classified as an R-peak occurrence. The decision block of the algorithm works in three states: (1) identification of the search window, (2) detection of the R-peak by maximum y[n] value search, and (3) waiting state after R-peak occurrence. The threshold necessary to identify the search window is adaptive and its value is calculated based on the amplitude of the last R-peak and the previous threshold value.

Figure 2.

Figure 2

Algorithm 2 block diagram based on [2].

3.3. Algorithm 3

Algorithm 3 [3] is based on modified level-crossing sampling, which belongs to event-triggered sampling schemes [33]. The input to the algorithm is the analog ECG signal (instead of the digital ECG used for the other algorithms analyzed in this paper). In the level-crossing sampling, the analog input range is divided into 2M − 1 levels (where M is the sampling resolution) and the sample is taken only when the input signal crosses one of the levels. The level-crossing analog-to-digital converter (LC-ADC) used in this algorithm is modified by asymmetrical hysteresis. The sample is taken only when (a) the input signal crosses the sampling level in the same direction as the last sample taken or (b) the input signal crosses kl levels in the opposite direction (where kl is the hysteresis parameter). The LC-ADC outputs the samples that are non-uniformly spaced in time. By selecting the appropriate value of kl, analysis of sample clusters (Figure 3 signal change direction DVi, Token), and their timing (Figure 3 Time Data Dti), it is possible identify the input ECG signal peaks first and, in the next processing step, identify R-peaks. There are three main processing blocks: level-crossing analog-to-digital converter (LC-ADC), peak detector, and beat detector (Figure 3). The algorithm does not use any filters (unlike other algorithms analyzed in this paper) due to the properties of the level-crossing sampling scheme adopted in the LC-ADC with hysteresis. For the purposes of algorithm analysis in this paper, the input to the algorithm is a uniformly sampled ECG signal.

Figure 3.

Figure 3

Algorithm 3 block diagram based on [3]. Peak detector output signal DVi is a two-bit signal where values 00 or 11 indicate the local peak in the ECG signal; that is, the sample that is taken when the input signal crosses kl sampling levels in the opposite direction to the sample taken previously. Token is a one-bit signal indicating with “1” the moment of sampling, and Dti is an 11-bit word readout of the counter to register information about sample time (required in non-uniform sampling).

3.4. Algorithm 4

Algorithm 4, developed by Pan and Tompkins and published in 1985, is the most widely referenced QRS detector. The adaptation of the original algorithm used in this study comes from [34]. In this adaptation, the decision block is simplified, whereas the preprocessing stage uses the Butterworth filter instead of the simple moving average filter. The original ECG signal preprocessing is retained and consists of a band pass filtering, differentiation, squaring, and moving window average [4] (Figure 4). The resulting preprocessed signal is fed to the decision block where, during State 1, its first local maximum that is higher than the detection threshold is marked as the R-peak (the condition for a local maximum is where the next and the previous sample values are smaller than the sample under analysis). Each local maximum smaller than the detection threshold modifies the value of the detection threshold in line with [4]. As soon as the R-peak is found, the algorithm enters the waiting state, State 2, which is 200 ms long, and thereafter resumes the local maximum search. The parallel analysis of filtered signals with a second threshold is not applied in this implementation of the original algorithm. The modification of the feature signal threshold based on a regular and irregular heart rate, as well as the searchback mechanism, is implemented.

Figure 4.

Figure 4

Algorithm 4 block diagram based on [4].

3.5. ECG Database and Test Dataset

In the MIT-BIH AD, the QRS morphology types [35] appear with the following number of cases: N (Normal)—75052, L (Left Bundle Branch Block)—8075, R (Right Bundle Branch Block)—7259, V (Ventricular Premature Beat)—7130, P (Paced)—7028, and A (Atrial Premature Beat)—2546. These six QRS morphology patterns were selected as the most frequent in the database (107090/109494 beats, i.e., 97.8%) as well as in real recordings expected in mobile patients. Records in the MIT-BIH AD are sampled at 360 Hz, which corresponds to the sampling interval of 2.7778 ms. Throughout this paper, we use the number of samples to describe the timeline.

Noise Pattern and Noisy Test Signals

In order to evaluate the immunity to noise of the QRS detectors under analysis, we have used the first channel signal from the MIT-BIH AD and three derivate datasets. Each of the three datasets was created by adding a noise signal from the MIT-BIH NSTD multiplied by three different scaling factors to the original MIT-BIH AD. Considering the wearable application as the most expected and the omnipresence of muscle artifacts, we decided to use a “muscle artifact” (MA) record from the MIT-BIH NSTD. From a practical viewpoint, the records were made with the same sampling parameters and the same length as the ECG signal. Moreover, the noise added is point-by-point trackable, allowing for a detailed comparison of the detector’s performance at each particular heartbeat. As we used original records from the MIT-BIH AD, the intrinsic noise already present in the data is out of our control. Consequently, the investigation of QRS detectors’ behavior in the presence of noise refers to “original” and not “noise-free” ECGs, and the relative signal-to-noise ratio (SNR) has been calculated based on the average power factor of the original record Ps (MIT-BIH AD) and added noise pattern Pn (MIT-BIH NSD record MA):

SNR=20logPsPn (1)

The power is calculated according to:

P=1N1i=0N1(xi+1xi)2 (2)

The following procedure has been applied to achieve the target relative SNR of 15, 7, and 3 dB. The mixing procedure for one record of the MIT-BIH AD starts with the calculation of SNR based on Equation (1), where Ps is the power factor for this MIT-BIH AD record and Pn is the power factor for the MIT-BIH NSTD record MA. The result is existing SNR (eSNR). Next, to calculate k (scaling factor), the target noise level (tSNR; for example, 3 dB) is input to Equation (3). The square root in Equation (3) is due to the fact that the amplitude ECGtest and ECGorig ratio is the square root of the power ratio. Once the scaling factor k is calculated, the test dataset ECGtest is calculated according to Equation (4), where MA is the MIT-BIH NSTD record MA. This procedure is repeated for all MIT-BIH AD records for all three (15, 7, 3 dB) noise-test datasets.

k=tSNReSNR (3)
ECGtest=ECGorig+k·MA  (4)

The process of adding noise to example record 121 of the MIT-BIH AD is illustrated in Figure 5 and Figure 6 below.

Figure 5.

Figure 5

The ECG (excerpt of record 121 from the MIT-BIH AD) and noise (excerpt of record MA (muscle artifact) from the MIT-BIH NSTD) before the mixing procedure.

Figure 6.

Figure 6

Figure 6

The ECG (excerpt of record 121 from the MIT-BIH AD) with (a) no added noise, (b) added noise for SNR = 15 dB, (c) added noise for SNR = 7 dB, and (d) added noise for SNR = 3 dB.

4. Results

The investigation of the QRS detection accuracy of four QRS detectors was performed separately for the six most frequent beat morphologies, five arbitrarily selected values of detection temporal tolerance, and four true-to-life levels of noise mixed with the database records. The amount of data to analyze and present is sizeable. Therefore, we present the following:

Table 1.

QRS detectors’ accuracy is expressed as the true-positive-to-total-beats ratio for various DTT values. The mean value and the standard deviation of the true-positive-to-total-beats ratio are computed for all six considered QRS morphology types and all four levels of added noise. The best results are highlighted in bold.

Algorithm
Detector
Temporal
Tolerance
[ms]
(Samples)
Algorithm 1 Algorithm 2 Algorithm 3 Algorithm 4
Mean Std Mean Std Mean Std Mean Std
8.33 (3) 27.54 21.93 65.75 19.74 8.26 9.52 14.30 6.36
47.22 (17) 92.53 9.27 87.77 11.16 80.66 11.68 52.85 14.17
86.11 (31) 96.79 4.44 98.66 1.62 88.43 6.25 74.93 8.83
125.00 (45) 98.89 1.02 99.55 0.44 92.22 4.72 98.52 1.33
163.89 (59) 99.25 0.89 99.71 0.36 94.80 4.45 99.46 0.55

Table 2.

QRS detectors’ accuracy is expressed as the true-positive-to-total-beats ratio for various heartbeat morphology types. The mean value and the standard deviation of the true-positive-to-total-beats ratio are computed for all five considered DTT values and all four levels of added noise. The best results are highlighted in bold.

Algorithm
Beat Type
Algorithm 1 Algorithm 2 Algorithm 3 Algorithm 4
Mean Std Mean Std Mean Std Mean Std
N 82.71 32.34 96.30 5.66 74.12 36.91 72.28 30.12
L 82.47 31.34 86.86 19.30 76.47 35.31 65.38 36.50
R 80.96 36.85 93.22 11.94 75.52 37.60 71.27 30.76
V 76.92 26.34 80.51 24.54 67.29 34.08 66.36 34.20
P 92.52 10.41 85.94 18.55 72.68 25.64 66.49 33.43
A 82.42 35.74 97.24 5.22 71.17 34.75 66.30 35.91

Table 3.

QRS detectors’ accuracy is expressed as the true-positive-to-total-beats ratio for various added noise levels. The mean value and the standard deviation of the true-positive-to-total-beats ratio are computed for all six considered QRS morphology types and all five considered DTT values. The best results are highlighted in bold.

Algorithm
Added Noise Level
Algorithm 1 Algorithm 2 Algorithm 3 Algorithm 4
Mean Std Mean Std Mean Std Mean Std
No noise added 83.72 30.88 90.68 16.80 77.12 35.90 62.03 36.60
SNR 15 dB 83.38 30.48 90.34 16.78 72.16 33.62 69.27 32.71
SNR 7 dB 82.78 30.08 89.86 16.93 70.47 33.00 70.43 31.67
SNR 3 dB 82.12 29.71 89.18 17.20 71.74 33.50 70.32 31.60

Figure 7.

Figure 7

Performance of Algorithm 1 in relation to QRS beat morphology (upper long horizontal axis), DTT values (in samples, bottom short horizontal axes), and added noise level (dot shape and color).

Figure 8.

Figure 8

Performance of Algorithm 2 in relation to QRS beat morphology (upper long horizontal axis), DTT values (in samples, bottom short horizontal axes), and added noise level (dot shape and color).

Figure 9.

Figure 9

Performance of Algorithm 3 in relation to QRS beat morphology (upper long horizontal axis), DTT values (in samples, bottom short horizontal axes), and added noise level (dot shape and color).

Figure 10.

Figure 10

Performance of Algorithm 4 in relation to QRS beat morphology (upper long horizontal axis), DTT values (in samples, bottom short horizontal axes), and added noise level (dot shape and color).

Table 4.

Results for Algorithm 1, N-type morphology beats and DTT = 3 (8.33 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 1337 1176 1143 1123 −161 −194 −214
101 1860 1860 1853 1806 1766 −7 −54 −94
102 99 99 92 86 80 −7 −13 −19
103 2082 114 261 334 374 147 220 260
104 163 1 1 4 6 0 3 5
105 2526 2383 2305 2203 2128 −78 −180 −255
106 1507 47 122 158 182 75 111 135
108 1739 660 552 518 512 −108 −142 −148
112 2537 0 93 126 137 93 126 137
113 1789 0 29 117 148 29 117 148
114 1820 521 456 440 421 −65 −81 −100
115 1953 0 9 32 55 9 32 55
116 2302 1 51 139 178 50 138 177
117 1534 619 565 520 489 −54 −99 −130
119 1543 12 278 330 360 266 318 348
121 1861 1165 894 760 679 −271 −405 −486
122 2476 254 664 751 777 410 497 523
123 1515 3 77 59 79 74 56 76
200 1743 1 10 34 61 9 33 60
201 1625 16 66 100 137 50 84 121
202 2061 3 91 136 156 88 133 153
203 2529 280 320 360 371 40 80 91
205 2571 1394 1346 1290 1263 −48 −104 −131
208 1586 164 288 360 386 124 196 222
209 2621 1 1 4 10 0 3 9
210 2423 16 101 186 234 85 170 218
212 923 6 28 55 78 22 49 72
213 2641 8 33 101 149 25 93 141
215 3195 2 3 3 4 1 1 2
217 244 0 2 8 11 2 8 11
219 2082 8 111 210 250 103 202 242
220 1954 2 7 36 44 5 34 42
221 2031 0 12 39 80 12 39 80
222 2062 814 841 832 822 27 18 8
223 2029 0 9 40 77 9 40 77
228 1688 1204 1094 1015 972 −110 −189 −232
230 2255 0 0 2 3 0 2 3
231 314 265 219 226 229 −46 −39 −36
233 2230 191 346 431 470 155 240 279
234 2700 11 215 344 413 204 333 402
TOTAL 75,052 13,462 14,621 15,338 15,714 1159 1876 2252

Table 5.

Results for Algorithm 1, N-type morphology beats and DTT = 17 (47.22 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 2238 2238 2232 2224 0 −6 −14
101 1860 1860 1856 1820 1797 −4 −40 −63
102 99 99 98 95 92 −1 −4 −7
103 2082 2082 2082 2081 2077 0 −1 −5
104 163 163 163 163 161 0 0 −2
105 2526 2508 2510 2502 2477 2 −6 −31
106 1507 1504 1467 1425 1398 −37 −79 −106
108 1739 1666 1527 1459 1459 −139 −207 −207
112 2537 2536 2487 2343 2220 −49 −193 −316
113 1789 1788 1779 1770 1752 −9 −18 −36
114 1820 1778 1721 1659 1606 −57 −119 −172
115 1953 1952 1952 1930 1920 0 −22 −32
116 2302 2284 2280 2262 2232 −4 −22 −52
117 1534 1534 1437 1360 1319 −97 −174 −215
119 1543 1543 1490 1454 1428 −53 −89 −115
121 1861 1860 1694 1532 1403 −166 −328 −457
122 2476 2476 2471 2434 2376 −5 −42 −100
123 1515 1515 1462 1447 1437 −53 −68 −78
200 1743 1740 1738 1730 1725 −2 −10 −15
201 1625 1610 1595 1588 1586 −15 −22 −24
202 2061 2061 2037 2028 2021 −24 −33 −40
203 2529 2471 2455 2409 2382 −16 −62 −89
205 2571 2570 2570 2568 2563 0 −2 −7
208 1586 1579 1576 1572 1565 −3 −7 −14
209 2621 2621 2621 2621 2621 0 0 0
210 2423 2421 2417 2415 2405 −4 −6 −16
212 923 922 922 922 922 0 0 0
213 2641 2640 2640 2637 2637 0 −3 −3
215 3195 3194 3194 3194 3194 0 0 0
217 244 244 244 243 239 0 −1 −5
219 2082 2082 2069 2039 2032 −13 −43 −50
220 1954 1954 1954 1951 1929 0 −3 −25
221 2031 2031 2028 2020 2011 −3 −11 −20
222 2062 2048 2022 1983 1962 −26 −65 −86
223 2029 2029 2027 2019 2005 −2 −10 −24
228 1688 1643 1557 1508 1481 −86 −135 −162
230 2255 2255 2255 2254 2253 0 −1 −2
231 314 289 255 260 264 −34 −29 −25
233 2230 2228 2229 2226 2217 1 −2 −11
234 2700 2697 2699 2699 2699 2 2 2
TOTAL 75,052 74,715 73,818 72,854 72,091 −897 −1861 −2624

Table 6.

Results for Algorithm 2, N-type morphology beats and DTT = 3 (8.33 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 d SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 2236 2235 2229 2198 −1 −7 −38
101 1860 1856 1855 1856 1849 −1 0 −7
102 99 92 84 75 67 −8 −17 −25
103 2082 2080 2080 2079 2077 0 −1 −3
104 163 163 163 161 158 0 −2 −5
105 2526 2465 2444 2362 2267 −21 −103 −198
106 1507 1489 1489 1478 1471 0 −11 −18
108 1739 705 652 641 644 −53 −64 −61
112 2537 1637 1458 1266 1150 −179 −371 −487
113 1789 1787 1788 1788 1788 1 1 1
114 1820 521 505 475 455 −16 −46 −66
115 1953 1948 1947 1934 1916 −1 −14 −32
116 2302 2268 2222 2128 2048 −46 −140 −220
117 1534 1002 852 753 710 −150 −249 −292
119 1543 1542 1530 1455 1398 −12 −87 −144
121 1861 1609 1151 1002 903 −458 −607 −706
122 2476 2475 2316 2130 2014 −159 −345 −461
123 1515 1513 1494 1431 1362 −19 −82 −151
200 1743 1623 1575 1509 1446 −48 −114 −177
201 1625 1622 1622 1603 1575 0 −19 −47
202 2061 2061 2058 2038 1991 −3 −23 −70
203 2529 1810 1618 1497 1390 −192 −313 −420
205 2571 2567 2567 2549 2518 0 −18 −49
208 1586 1569 1568 1557 1529 −1 −12 −40
209 2621 2616 2616 2604 2586 0 −12 −30
210 2423 2417 2408 2355 2292 −9 −62 −125
212 923 922 922 921 920 0 −1 −2
213 2641 2637 2637 2616 2575 0 −21 −62
215 3195 1997 1960 1915 1901 −37 −82 −96
217 244 238 231 223 216 −7 −15 −22
219 2082 2077 2069 1992 1936 −8 −85 −141
220 1954 1939 1884 1852 1799 −55 −87 −140
221 2031 2030 2031 2020 2000 1 −10 −30
222 2062 2049 2037 2012 1964 −12 −37 −85
223 2029 1910 1873 1785 1709 −37 −125 −201
228 1688 1675 1672 1616 1560 −3 −59 −115
230 2255 215 301 371 411 86 156 196
231 314 314 314 314 314 0 0 0
233 2230 2223 2210 2135 2083 −13 −88 −140
234 2700 2696 2697 2696 2684 1 0 −12
TOTAL 75,052 66,595 65,135 63,423 61,874 −1460 −3172 −4721

Table 7.

Results for Algorithm 2, N-type morphology beats and DTT = 17 (47.22 ms), from the MIT-BIH AD. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 2236 2235 2232 2214 −1 −4 −22
101 1860 1856 1855 1856 1852 −1 0 −4
102 99 95 92 90 82 −3 −5 −13
103 2082 2080 2080 2080 2079 0 0 −1
104 163 163 163 163 162 0 0 −1
105 2526 2484 2479 2427 2378 −5 −57 −106
106 1507 1490 1492 1486 1485 2 −4 −5
108 1739 1610 1574 1556 1547 −36 −54 −63
112 2537 2536 2432 2270 2177 −104 −266 −359
113 1789 1787 1788 1788 1788 1 1 1
114 1820 1817 1811 1769 1723 −6 −48 −94
115 1953 1950 1950 1948 1938 0 −2 −12
116 2302 2275 2268 2228 2195 −7 −47 −80
117 1534 1534 1446 1381 1328 −88 −153 −206
119 1543 1542 1533 1480 1434 −9 −62 −108
121 1861 1860 1682 1579 1485 −178 −281 −375
122 2476 2475 2424 2339 2289 −51 −136 −186
123 1515 1513 1513 1491 1462 0 −22 −51
200 1743 1741 1739 1729 1716 −2 −12 −25
201 1625 1623 1623 1615 1604 0 −8 −19
202 2061 2061 2061 2049 2027 0 −12 −34
203 2529 2449 2344 2262 2188 −105 −187 −261
205 2571 2567 2567 2552 2525 0 −15 −42
208 1586 1569 1568 1563 1548 −1 −6 −21
209 2621 2620 2619 2618 2617 −1 −2 −3
210 2423 2418 2415 2394 2365 −3 −24 −53
212 923 922 922 921 921 0 −1 −1
213 2641 2638 2639 2634 2608 1 −4 −30
215 3195 3194 3194 3191 3187 0 −3 −7
217 244 244 243 242 240 −1 −2 −4
219 2082 2079 2077 2044 2007 −2 −35 −72
220 1954 1953 1951 1948 1939 −2 −5 −14
221 2031 2030 2031 2027 2015 1 −3 −15
222 2062 2055 2044 2021 1983 −11 −34 −72
223 2029 2028 2024 1999 1967 −4 −29 −61
228 1688 1677 1677 1635 1605 0 −42 −72
230 2255 2252 2252 2252 2252 0 0 0
231 314 314 314 314 314 0 0 0
233 2230 2224 2224 2190 2162 0 −34 −62
234 2700 2696 2697 2698 2689 1 2 −7
TOTAL 75,052 74,657 74,042 73,061 72,097 −615 −1596 −2560

Table 8.

Results for Algorithm 3, N-type morphology beats and DTT = 3 (8.33 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 202 197 168 156 −5 −34 −46
101 1860 16 32 32 36 16 16 20
102 99 0 0 1 1 0 1 1
103 2082 2 3 2 1 1 0 −1
104 163 8 9 12 8 1 4 0
105 2526 35 37 45 56 2 10 21
106 1507 0 1 0 2 1 0 2
108 1739 2 6 17 28 4 15 26
112 2537 58 165 208 239 107 150 181
113 1789 0 0 0 0 0 0 0
114 1820 244 224 228 211 −20 −16 −33
115 1953 0 0 0 1 0 0 1
116 2302 0 2 3 8 2 3 8
117 1534 679 530 490 488 −149 −189 −191
119 1543 0 0 1 1 0 1 1
121 1861 0 32 66 89 32 66 89
122 2476 0 1 7 11 1 7 11
123 1515 0 0 0 0 0 0 0
200 1743 0 0 0 10 0 0 10
201 1625 0 2 4 5 2 4 5
202 2061 0 1 3 2 1 3 2
203 2529 39 62 82 101 23 43 62
205 2571 29 37 42 47 8 13 18
208 1586 89 90 79 80 1 −10 −9
209 2621 2 1 5 7 −1 3 5
210 2423 0 7 12 15 7 12 15
212 923 0 0 1 1 0 1 1
213 2641 0 0 2 1 0 2 1
215 3195 1 4 12 17 3 11 16
217 244 0 0 0 1 0 0 1
219 2082 2 2 2 4 0 0 2
220 1954 0 1 1 1 1 1 1
221 2031 0 1 2 4 1 2 4
222 2062 630 662 607 587 32 −23 −43
223 2029 0 1 6 6 1 6 6
228 1688 6 20 27 32 14 21 26
230 2255 0 0 0 1 0 0 1
231 314 0 0 0 0 0 0 0
233 2230 0 0 2 2 0 2 2
234 2700 0 0 0 1 0 0 1
TOTAL 75,052 2044 2130 2169 2261 86 125 217

Table 9.

Results for Algorithm 3, N-type morphology beats and DTT = 17 (47.22 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-TYPE beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 2239 2104 1906 1829 −135 −333 −410
101 1860 1853 1697 1530 1452 −156 −323 −401
102 99 99 88 82 77 −11 −17 −22
103 2082 2082 1954 1759 1690 −128 −323 −392
104 163 159 156 146 138 −3 −13 −21
105 2526 2440 2166 1999 1987 −274 −441 −453
106 1507 1500 1331 1227 1202 −169 −273 −298
108 1739 1570 1256 1149 1160 −314 −421 −410
112 2537 2534 2101 2079 2014 −433 −455 −520
113 1789 1789 1684 1556 1539 −105 −233 −250
114 1820 1393 1432 1356 1277 39 −37 −116
115 1953 1953 1618 1535 1545 −335 −418 −408
116 2302 2277 1859 1884 1950 −418 −393 −327
117 1534 1532 1236 1208 1242 −296 −324 −290
119 1543 1543 1168 1194 1225 −375 −349 −318
121 1861 1857 1436 1375 1345 −421 −482 −512
122 2476 2474 1948 1991 2012 −526 −483 −462
123 1515 1515 1158 1147 1164 −357 −368 −351
200 1743 1711 0 0 1408 −1711 −1711 −303
201 1625 1621 1609 1497 1406 −12 −124 −215
202 2061 2058 1974 1799 1723 −84 −259 −335
203 2529 2453 2236 2078 2042 −217 −375 −411
205 2571 2567 2456 2239 2174 −111 −328 −393
208 1586 1559 1429 1325 1307 −130 −234 −252
209 2621 2607 2562 2389 2329 −45 −218 −278
210 2423 2393 2350 2164 2004 −43 −229 −389
212 923 916 862 799 746 −54 −117 −170
213 2641 2632 2346 2221 2208 −286 −411 −424
215 3195 3187 3156 3003 2916 −31 −184 −271
217 244 244 217 198 199 −27 −46 −45
219 2082 2082 1724 1671 1659 −358 −411 −423
220 1954 1953 1642 1534 1537 −311 −419 −416
221 2031 2027 1955 1754 1726 −72 −273 −301
222 2062 1665 1712 1613 1530 47 −52 −135
223 2029 2028 1740 1670 1619 −288 −358 −409
228 1688 1649 1540 1381 1336 −109 −268 −313
230 2255 2245 2089 1904 1831 −156 −341 −414
231 314 314 296 254 246 −18 −60 −68
233 2230 2228 1980 1836 1839 −248 −392 −389
234 2700 2699 2540 2346 2251 −159 −353 −448
TOTAL 75,052 73,647 64,807 60,798 60,884 −8840 −12,849 −12,763

Table 10.

Results for Algorithm 4, N-type morphology beats and DTT = 3 (8.33 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 376 522 608 616 146 232 240
101 1860 303 434 500 516 131 197 213
102 99 16 31 23 21 15 7 5
103 2082 197 301 411 448 104 214 251
104 163 18 23 29 29 5 11 11
105 2526 2178 2021 1778 1626 −157 −400 −552
106 1507 396 432 476 480 36 80 84
108 1739 96 209 232 223 113 136 127
112 2537 7 155 131 105 148 124 98
113 1789 199 289 370 415 90 171 216
114 1820 71 138 137 132 67 66 61
115 1953 4 84 157 196 80 153 192
116 2302 64 337 365 385 273 301 321
117 1534 142 308 313 292 166 171 150
119 1543 756 796 667 588 40 −89 −168
121 1861 663 467 153 87 −196 −510 −576
122 2476 2123 1734 1383 1141 −389 −740 −982
123 1515 7 74 147 153 67 140 146
200 1743 1 20 52 70 19 51 69
201 1625 27 117 185 228 90 158 201
202 2061 11 48 127 204 37 116 193
203 2529 114 210 221 227 96 107 113
205 2571 561 741 788 730 180 227 169
208 1586 314 440 441 408 126 127 94
209 2621 71 186 304 355 115 233 284
210 2423 54 329 400 464 275 346 410
212 923 514 511 499 490 −3 −15 −24
213 2641 772 1714 1635 1392 942 863 620
215 3195 3 24 90 129 21 87 126
217 244 0 14 19 22 14 19 22
219 2082 75 466 563 537 391 488 462
220 1954 7 74 161 196 67 154 189
221 2031 45 200 312 355 155 267 310
222 2062 543 607 600 593 64 57 50
223 2029 10 73 107 97 63 97 87
228 1688 336 543 504 504 207 168 168
230 2255 0 2 11 28 2 11 28
231 314 106 111 116 126 5 10 20
233 2230 14 209 321 337 195 307 323
234 2700 706 872 953 997 166 247 291
TOTAL 75,052 11,900 15,866 16,289 15,942 3966 4389 4042

Table 11.

Results for Algorithm 4, N-type morphology beats and DTT = 17 (47.22 ms), from the MIT-BIH AD. Improvements in results are marked in bold. Records 107, 109, 111, 118, 124, 207, 214, and 232 are not shown, as there are no N-type morphology beats in those records.

MIT-BIH Arrhythmia
Database, Normal—N-Type Beats
No Noise Added SNR = 15 dB SNR = 7 dB SNR = 3 dB SNR = 15 dB vs. No Noise Added SNR = 7 dB vs. No Noise Added SNR = 3 dB vs. No Noise Added
Record TB TP TP TP TP
100 2239 1920 1889 1909 1851 −31 −11 −69
101 1860 1332 1359 1381 1361 27 49 29
102 99 20 45 37 36 25 17 16
103 2082 1339 1480 1554 1602 141 215 263
104 163 58 85 96 93 27 38 35
105 2526 2372 2281 2093 2001 −91 −279 −371
106 1507 1500 1473 1438 1399 −27 −62 −101
108 1739 457 891 929 933 434 472 476
112 2537 1117 1503 1545 1462 386 428 345
113 1789 645 958 1107 1185 313 462 540
114 1820 579 873 920 923 294 341 344
115 1953 393 768 956 1017 375 563 624
116 2302 1388 1745 1744 1721 357 356 333
117 1534 409 857 907 905 448 498 496
119 1543 1508 1408 1266 1211 −100 −242 −297
121 1861 1036 1123 979 873 87 −57 −163
122 2476 2459 2205 2038 1958 −254 −421 −501
123 1515 313 736 864 859 423 551 546
200 1743 678 952 1040 1074 274 362 396
201 1625 964 1131 1184 1183 167 220 219
202 2061 895 1216 1333 1353 321 438 458
203 2529 1277 1649 1668 1692 372 391 415
205 2571 2018 2159 2137 2087 141 119 69
208 1586 1127 1271 1267 1216 144 140 89
209 2621 957 1354 1582 1656 397 625 699
210 2423 1026 1515 1622 1649 489 596 623
212 923 856 842 848 829 −14 −8 −27
213 2641 1558 2161 2180 2083 603 622 525
215 3195 922 1300 1521 1651 378 599 729
217 244 55 135 145 149 80 90 94
219 2082 726 1442 1536 1503 716 810 777
220 1954 1106 1277 1351 1372 171 245 266
221 2031 739 1235 1379 1438 496 640 699
222 2062 1870 1840 1795 1754 −30 −75 −116
223 2029 1220 1368 1359 1350 148 139 130
228 1688 1375 1406 1285 1242 31 −90 −133
230 2255 722 813 928 1009 91 206 287
231 314 291 278 278 283 −13 −13 −8
233 2230 1190 1458 1504 1483 268 314 293
234 2700 1943 2209 2270 2254 266 327 311
TOTAL 75,052 42,360 50,690 51,975 51,700 8330 9615 9340

4.1. Statistics of the QRS Detectors’ Performance

With four independent variables (detector number, DTT, QRS morphology type, and added noise level), one can build a total of 16 statistical analyses. We selected the three most representative statistical analyses for independent studies of the detectors’ performance and vulnerability to DTT, QRS morphology type, and added noise level. Table 1 summarizes statistics on TP/TB for each algorithm (columns) and each tolerance window length (rows) for various DTT values, while the mean value and standard deviation are calculated for all QRS morphology types and added noise levels.

Table 2 presents statistics on TP/TB for each QRS morphology type (rows), while the mean value and standard deviation are calculated for all DTT values and noise levels. This table represents the detection quality and its independence from the QRS morphology, which is unknown at the time of detection.

Table 3 presents statistics on TP/TB for each algorithm (columns) and each noise level tested (rows), while the mean value and standard deviation are calculated for all QRS morphology types and DTT values. This table helps to explore which algorithm is the best and which gives the most stable detection results in the presence of noise.

4.2. Plots of the Detectors’ Performance

To provide deeper insight into detectors’ performance, we also use plots (displayed in Figure 7, Figure 8, Figure 9 and Figure 10) that examine the properties of each studied algorithm independently.

5. Discussion

5.1. Influence of Noise

While planning the experimental work, the expectation was that, with increasing levels of noise the TP/TB results would deteriorate. When we analyze the average TP/TB results in Table 3, we can conclude that the expectations are fulfilled for Algorithm 1, Algorithm 2, and Algorithm 3 but not for Algorithm 4.

The average TP/TB (Table 3) results for increased levels of noise are in the following range for no added noise to maximum added noise SNR = 3 dB, respectively:

  • Algorithm 1—83.72% and 82.12%,

  • Algorithm 2—90.68% and 89.18%,

  • Algorithm 3—77.12% and 71.74%,

  • and Algorithm 4—62.03% and 70.43%; there is no deterioration, but an improvement of 8.4% with added noise.

For Algorithm 1, the improvement in average TP/TB with increased level of added noise has not been observed (Table 3). Turning the analysis of Algorithm 1 to not-averaged TP/TB data (Figure 7), we can observe that, for DTT 8.33 ms (three samples) QRS morphology types N, L, R, and V, there is an improvement in TP/TB with increased added noise level for several records. Looking into details (Table 4) for QRS morphology of N type, DTT 8.33 ms (three samples), the TP/TB results for added noise 15 dB, 7 dB, and 3 dB are improved compared with records with no added noise for the following records: 103, 106, 112, 113, 115, 117, 119, 122, 123, 200, 201, 202, 203, 208, 210, 213, 219, 220, 221, 233, and 234 (Table 4). It constitutes results for over 50% of the records (21 from 40 records with N-type QRS morphology). The biggest improvement is registered for record 122. For greater values of DTT and N-type beats with this algorithm, there are no improvements in TP/TB (Figure 7); detailed results for DTT 47.22 ms (17 samples) are shown in Table 5.

For Algorithm 2, an improvement in average TP/TB with an increased level of added noise has not been observed (Table 3). As can be seen in the detailed data for N-type QRS morphology and DTT 8.33 ms (three samples) in Table 6, there is only one case of TP/TB improvement, for record 230. The improvement in TP results for no added noise versus added noise of 15, 7, and 3 dB are 86, 156, and 196, respectively.

For Algorithm 3, the improvement in average TP/TB with an increased level of added noise has not been observed (Table 3). When analyzing Algorithm 3 results for not-averaged TP/TB data (Figure 9), the improvement of TP/TB with an increased level of added noise is revealed for QRS morphology type V for all values of DTT. Analysis of the results for N-type QRS morphology (Table 8 and Table 9) reveals that, for DTT 8.33 ms (three samples), there is improvement in totals, whereas for DTT 47.22 ms (17 samples), there is no improvement in totals (only in a few selected records).

For Algorithm 4, the improvement in average TP/TB with an increased level of added noise has been observed (Table 3). Looking into details (Table 10) for QRS morphology of N-type, DTT 8.33 ms (three samples), the TP/TB results for added noise 15 dB, 7 dB, and 3 dB are improved for 35 out of 40 records. The improvement is not revealed for records 105, 119, 121, 122, and 212. For DTT 47.22 ms (17 samples), the improvement is revealed for 30 out of 40 records. The improvement is not revealed for the following records: 100, 105, 106, 119, 121, 122, 212, 222, 228, and 231.

Why do TP/TB results for Algorithm 4 not deteriorate with increased levels of noise? When we look into detailed data in Table 10, TP/TB results for N-type QRS morphology, and DTT 8.33 ms (three samples), the improvement in TP/TB with increasing levels of noise can be observed for 30 out of 40 records. Similar improvements of results are revealed for DTT 47.22 ms (17 samples) and can be observed in Table 11. Why do TP/TB results improve with added noise? Our hypothesis is that it is related to Algorithm 4 higher computational complexity being reflected in more calculations and processing blocks in computation. Thus, the hypothesis is that adding noise to the input ECG signal reduces friction between the blocks of computation. The phenomena of reduction of error from sticky moving parts in mechanical computers used to perform navigation and bomb trajectory calculations are cited as the first observation and purposeful use of dither [36]. In analog-to-digital conversion, dither—purposeful distortion to the input signal—causes digitization error to behave well statistically (dithered quantization) [7,8].

In summary, by adding muscular noise, we expected a deterioration of the detectors’ performance. Surprisingly, our results show that this is not always the case. For Algorithm 4, there is general improvement visible in averaged TP/TB data and confirmed in detailed data. For other algorithms, general improvement in averaged TP/TB has not been observed. Still, the phenomena of TP/TB improvement with added noise are present, although on a smaller scale for other algorithms (namely for specific QRS morphologies, DTT values, or individual records).

5.2. Influence of Detector Time Tolerance DTT

While planning the experimental work, the expectation was that increasing DTT would improve the TP/TB results. This assumption was additionally supported by results of one-dimensional analysis of DTT influence on TP in [10].

When we analyze the results in Table 1, we can see that results confirm the expectation for all algorithms and all DTT values under analysis. The average TP/TB results calculated over all added noise and QRS morphologies achieved for the lowest DTT 8.33 ms (three samples) and highest DTT 163.89 ms (59 samples), respectively, are:

  • Algorithm 1—27.54% and 99.25%,

  • Algorithm 2—65.75% and 99.71%,

  • Algorithm 3—8.26% and 94.80%,

  • and Algorithm 4—14.30% and 99.46%.

The following are additional observations from the analysis of plots (Figure 7, Figure 8, Figure 9 and Figure 10) for all algorithms:

  • Algorithm 1 for QRS morphologies N, L, R, P, and A, with the exception of DTT 8.33 ms (three samples), reveals TP/TB of more than 90%. For QRS morphology type V, the TP/TB results are below 90% for all DTT ≤ 125 ms (45 samples). For DTT 8.33 ms (three samples), for all QRS morphologies, the TP/TB results are below 30%.

  • Algorithm 2 results of TP/TB for all DTT and for QRS morphology N and A are above 80%. For L-, R-, V- and P-type QRS morphologies, the TP/TB results are above 90% for DTT > 86.11 ms (31 samples).

  • Algorithm 3 TP/TB results for DTT 8.33 ms and for all QRS morphologies are below 30%, and for other higher DTT values, there is a clear deterioration of TP/TB results with decreasing DTT. For all QRS morphologies TP/TB is above 90% only for DTT 125 ms and 163.89 ms (45 and 59 samples).

  • Algorithm 4 TP/TB results for all QRS morphologies are above 90% only for DTT 125 ms and 163.89 ms (45 and 59 samples). Results for other values of DTT clearly deteriorate well below 80% with decreasing DTT value.

In summary, decreasing DTT causes deterioration of TP/TB with varying ratios dependent on algorithm, QRS morphology, and added noise level.

5.3. Influence of QRS Morphology

While planning the experimental work, the expected result was that, depending on the algorithm, certain QRS morphologies are more difficult for precise R-peak detection than other QRS morphologies. During the analysis of average TP/TB results in Table 2, we observed that the aforementioned expectation is fulfilled for all algorithms.

On closer analysis of the range of TP/TB results from worst to best in Table 2, we can observe the following range of TP/TB results for algorithm and QRS morphology, respectively:

  • Algorithm 1 from 76.92% for V to 92.52% for P,

  • Algorithm 2 from 80.51% for V to 97.24% for A,

  • Algorithm 3 from 67.29% for V to 76.47% for L,

  • and Algorithm 4 from 65.38% for L to 72.28% for N.

We can conclude that, for the four algorithms under analysis, the most difficult in precise R-peak location was V-type QRS morphology for three algorithms and L-type for one algorithm. In terms of the easiest QRS morphology type for precise R-peak location, it is different for each algorithm under analysis and yields P-, A-, L-, and N-types of QRS morphology.

Visual analysis of the TP/TB results displayed in plots (Figure 7, Figure 8, Figure 9 and Figure 10) in a direction to reveal any visual similarities in locations of the data points for various QRS morphologies is presented below for algorithm and QRS morphology, respectively:

  • Algorithm 1 for N, R, and A—similar locations; for L, V, and P, each type—different locations,

  • Algorithm 2 for N, R, and A—similar locations; for L, V, and P—similar locations,

  • Algorithm 3 for N, L, and R—similar locations; for V, P, and A, each type—different locations,

  • and Algorithm 4 for N, R, and A—similar locations; for L, V, and P—similar locations.

In summary, the most difficult and easiest QRS morphology type for accurate R-peak detection by a given detection algorithm is specific for that algorithm. The worst and best TP/TB results for all levels of noise in ECG signal and all DTT values under analysis are also specific for each algorithm.

5.4. Comparison of the Algorithms Studied

It is evident that a high-performance QRS detector should demonstrate a high TP/TB ratio even with low DTT. Therefore, a series of QRS detection instants may reliably be used by following procedures such as heart rate variability. In this aspect, all atrial premature, blocked, paced, and ventricular beats are not considered as representative to cardiac cycle control from the autonomous nervous system. Consequently, the precision of N-type QRS detection is more important than other types of beat morphology. At the same time, the high stability of the detection point sequence (that is, its independence from QRS morphology) is a desired algorithm feature. Lastly, an algorithm’s detection robustness to noise is very important, especially for mobile and wearable applications, as the level of noise is high and fluctuates depending on the environment and human activity.

Algorithm 1 and Algorithm 2 demonstrate good robustness to noise in ECG signal (Table 3, Figure 7 and Figure 8), whereas the QRS detection accuracy for Algorithm 3 and Algorithm 4 for noisy ECG signal is significantly worse. Additionally, the detection accuracy strongly drops with noise level, which makes the Algorithm 3 and Algorithm 4 not suitable for implementation in wearable ECG devices.

Algorithm 2 has a good detection accuracy even for low temporal tolerances (DTT = 3), except for V-type beats it has a TP/TB score over 50%, and for DTT = 17, it grows over 93% except for types L and V. This is a clinically acceptable detection ratio for remote cardiac rhythm detection and, as our results show, can be achieved with a temporal accuracy of 47.22 ms (17 samples).

Algorithm 1 and Algorithm 3 work well for DTT ≥ 47.22 ms (17 samples); TP/TB exceeds 90%, except for QRS types V for Algorithm 1 and Algorithm 3, and P for Algorithm 3, but its performance drops for low temporal tolerance (DTT 8.33 ms, three samples). Compared with Algorithm 1, Algorithm 2, and Algorithm 3, which were developed in the last decade, Algorithm 4, developed in 1985 and broadly referenced in the literature on QRS detection, is more sensitive to noise (Figure 10) and QRS morphology. It achieves satisfactory detection accuracy only for large DTT values.

Algorithm 1 (Figure 7) shows perfect noise immunity for V-type QRS morphology and high noise immunity for other beat types. For low noise levels, it has similar scores to Algorithm 3 (Figure 9), except for P morphology, where it is, again, significantly better.

Adding noise improves the results of Algorithm 3 in V-type morphology, and for high DTT also in N-type QRS morphology. This is also observed in Algorithm 4 (Figure 10), where, in most cases, red dots (SNR = 3 dB) are not the lowest data points in the plot.

5.5. Limitations of the Study and Future Research

A lack of computational complexity analysis may be considered as the main limitation of our study. Although the number of elementary operations for each algorithm and the growth of resource demand related to input size can be easily determined, preferences in this aspect depend on the target platform. The hardware architecture (e.g., multicore processors, vector data processing) and machine representation of data determine principal usability factors, such as maximum time of autonomy or the necessary size of battery, particularly in wearable recorders.

Another limitation of our study is the use of the MIT-DIH Arrhythmia Database, which is relatively sparsely sampled (sampling frequency 360 Hz) and known for annotation errors. In fact, the position of the maximum of the signal depends on the ECG lead used (each lead “sees” the spatial electrical phenomenon of heartbeat from a different viewpoint) and the true QRS maximum (i.e., precise R-peak) position can only be estimated from vectorcardiography. Moreover, in all cases, the sampling process is in no way synchronized with heart action. Consequently, the maximum falls between samples, and calculation of its true position requires interpolation. Nevertheless, we assume that possible annotation errors are rare and equally distributed before and after the true positions of R-peaks. This may result in an overestimation of the standard deviation results, but with no effect on the mean results of the algorithms.

The above two limitations will be addressed in future research, as will be the analysis of the sources of improvement in TP/TB results under conditions of added noise, as discussed in Section 5.1.

6. Conclusions

This study focused on the performance evaluation of QRS detectors aware of temporal accuracy and the presence of noise. Contrarily to most authors, satisfied with detection correctness as the sole quality indicator, we propose multidimensional criteria, including

  • accuracy (i.e., TP/TB),

  • precision of detection point location, i.e., TP/TB dependence on the DTT,

  • sensitivity to noise,

  • and sensitivity to QRS morphology.

To show the multidimensional method for QRS detectors’ evaluation of temporal accuracy, we compared three modern QRS detection algorithms and a well-known, commonly referenced Pan–Tompkins algorithm originating from the pioneer age of computerized electrocardiography.

The main scientific contribution of this work lies in testing QRS detectors in multiple dimensions, including various time tolerance values (DTT), which determines the detection statistics expressed by true positive, false positive, and false negative detection cases. Algorithms that demonstrate good accuracy with low values of DTT are the most precise algorithms. Output from such algorithms is immediately usable for subsequent ECG processing procedures such as HRV analysis or QRS morphology classification. Moreover, we observed that the accuracy of QRS detection depends on QRS morphology. Tests performed for various levels of added muscular noise revealed that there are significant differences between algorithms with respect to their noise robustness. High robustness amid noise (stable detection accuracy in noisy ECG recordings) is important in mobile and wearable applications in unstable conditions (e.g., in motion).

Author Contributions

Conceptualization, W.R. and P.A.; methodology, W.R. and P.A.; software, W.R. and M.M.; validation, P.A.; formal analysis, M.M.; investigation, W.R. and P.A.; resources, W.R. and P.A.; data curation, W.R.; writing—original draft preparation, W.R. and P.A.; writing—review and editing, W.R., M.M. and P.A.; visualization, W.R.; supervision, P.A.; project administration, M.M.; funding acquisition, P.A. and M.M. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable, no ethical issues, only public database used as study materials.

Informed Consent Statement

Not applicable, no human-acquired data, only public database used as study materials.

Data Availability Statement

Not applicable, data used are already available as public database.

Conflicts of Interest

The authors declare no conflicts of interest. The funders had no role in the design of the study, in the collection, analyses, or interpretation of data, in the writing of the manuscript, or in the decision to publish the results.

Funding Statement

This research was funded by AGH University of Krakow in 2023 and 2024 as research project no. 16.16.120.773, and by the National Science Centre of Poland under research project DEC-2018/31/B/ST7/03874.

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.Gutiérrez-Rivas R., Garcia J.J., Marnane W.P., Hernández A. Novel real-time low-complexity QRS complex detector based on adaptive thresholding. IEEE Sens. J. 2015;15:6036–6043. doi: 10.1109/JSEN.2015.2450773. [DOI] [Google Scholar]
  • 2.Reklewski W., Heryan K., Miśkowicz M., Augustyniak P. Real time ECG R-peak detection by extremum sampling; Proceedings of the 2020 IEEE 6th International Conference on Event-Based Control, Communication, and Signal Processing (EBCCSP); Krakow, Poland. 23–25 September 2020; pp. 1–7. [Google Scholar]
  • 3.Ravanshad N., Rezaee-Dehsorkh H., Lotfi R., Lian Y. A level-crossing based QRS-detection algorithm for wearable ECG sensors. IEEE J. Biomed. Health Inform. 2013;18:183–192. doi: 10.1109/JBHI.2013.2274809. [DOI] [PubMed] [Google Scholar]
  • 4.Pan J., Tompkins W.J. A real-time QRS detection algorithm. IEEE Trans. Biomed. Eng. 1985;3:230–236. doi: 10.1109/TBME.1985.325532. [DOI] [PubMed] [Google Scholar]
  • 5.Moody G.B., Mark R.G. The MIT-BIH arrhythmia database on CD-ROM and software for use with it. IEEE Comput. Cardiol. Proceed. 1990;2:185–188. [Google Scholar]
  • 6.Moody G., Muldrow W., Mark R. The MIT-BIH noise stress test database. Comput. Cardiol. 1984;11:381–384. [Google Scholar]
  • 7.Carbone P., Narduzzi C., Petri D. Dither signal effects on the resolution of nonlinear quantizers. IEEE Trans. Instrum. Meas. 1994;43:139–145. doi: 10.1109/19.293410. [DOI] [Google Scholar]
  • 8.Gray R.M., Stockham T.G. Dithered quantizers. IEEE Trans. Inf. Theory. 1993;39:805–812. doi: 10.1109/18.256489. [DOI] [Google Scholar]
  • 9.Kohler B.U., Hennig C., Orglmeister R. The principles of software QRS detection. IEEE Eng. Med. Biol. Mag. 2002;21:42–57. doi: 10.1109/51.993193. [DOI] [PubMed] [Google Scholar]
  • 10.Heryan K., Reklewski W., Szaflarski A., Ordowski M., Augustyniak P., Miśkowicz M. Sensitivity of QRS Detection Accuracy to Detector Temporal Resolution; Proceedings of the 2021 IEEE Computing in Cardiology (CinC); Brno, Czech Republic. 13–15 September 2021; pp. 1–4. [Google Scholar]
  • 11.Tripathi P.M., Kumar A., Komaragiri R., Kumar M. A review on computational methods for denoising and detecting ECG signals to detect cardiovascular diseases. Arch. Comput. Methods Eng. 2022;29:1875–1914. doi: 10.1007/s11831-021-09642-2. [DOI] [Google Scholar]
  • 12.Fariha MA Z., Ikeura R., Hayakawa S., Tsutsumi S. Journal of Physics: Conference Series. Volume 1532. IOP Publishing; Bristol, UK: 2020. Analysis of Pan-Tompkins algorithm performance with noisy ECG signals; p. 012022. [Google Scholar]
  • 13.Mohd Apandi Z.F., Ikeura R., Hayakawa S., Tsutsumi S. An analysis of the effects of noisy electrocardiogram signal on heartbeat detection performance. Bioengineering. 2020;7:53. doi: 10.3390/bioengineering7020053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zong W., Moody G.B., Jiang D. A robust open-source algorithm to detect onset and duration of QRS complexes. Comput. Cardiol. 2003;30:737–740. [Google Scholar]
  • 15.Hamilton P. Open Source ECG Analysis. Comput. Cardiol. 2002;1:101–104. [Google Scholar]
  • 16.D’Aloia M., Longo A., Rizzi M. Noisy ECG signal analysis for automatic peak detection. Information. 2019;10:35. doi: 10.3390/info10020035. [DOI] [Google Scholar]
  • 17.Gradl S., Leutheuser H., Elgendi M., Lang N., Eskofier B.M. Temporal correction of detected R-peaks in ECG signals: A crucial step to improve QRS detection algorithms; Proceedings of the 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Milan, Italy. 25–29 August 2015; pp. 522–525. [DOI] [PubMed] [Google Scholar]
  • 18.Nallathambi G., Prıncipe J.C. Integrate and fire pulse train automaton for QRS detection. IEEE Trans. Biomed. Eng. 2014;61:317–326. doi: 10.1109/TBME.2013.2282954. [DOI] [PubMed] [Google Scholar]
  • 19.Kohler B.U., Hennig C., Orglmeister R. QRS Detection Using Zero Crossing Counts. Progress Biomed. Res. 2003;8:138–145. [Google Scholar]
  • 20.Elgendi M. Fast QRS detection with an optimized knowledge-based method: Evaluation on 11 standard ECG databases. PLoS ONE. 2013;8:e73557. doi: 10.1371/journal.pone.0073557. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Porr B., Macfarlane P.W. A new QRS detector stress test combining temporal jitter and accuracy (JA) reveals significant performance differences amongst popular detectors. bioRxiv. 2023:722397. doi: 10.1101/722397. [DOI] [Google Scholar]
  • 22.Gusev M., Domazet E. Communications in Computer and Information Science, Proceedings of the ICT Innovations 2018, Engineering and Life Sciences: 10th International Conference, ICT Innovations 2018, Ohrid, Macedonia, 17–19 September 2018. Springer International Publishing; Berlin/Heidelberg, Germany: 2018. Optimizing the impact of resampling on QRS detection; pp. 107–119. [Google Scholar]
  • 23.Hamilton P.S., Tompkins W.J. Quantitative investigation of QRS detection rules using the MIT/BIH arrhythmia database. IEEE Trans. Biomed. Eng. 1986;12:1157–1165. doi: 10.1109/TBME.1986.325695. [DOI] [PubMed] [Google Scholar]
  • 24.Vollmer M. Noise resistance of several top-scored heart beat detectors; Proceedings of the 2017 Computing in Cardiology (CinC); Rennes, France. 24–27 September 2017; Piscataway, NJ, USA: IEEE; pp. 1–4. [Google Scholar]
  • 25.Liu F., Liu C., Jiang X., Zhang Z., Zhang Y., Li J., Wei S. Performance analysis of ten common QRS detectors on different ECG application cases. J. Healthc. Eng. 2018;2018:9050812. doi: 10.1155/2018/9050812. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Gabbouj M., Kiranyaz S., Malik J., Zahid M.U., Ince T., Chowdhury M.E., Khandakar A., Tahir A. IEEE Transactions on Neural Networks and Learning Systems. IEEE; Piscataway, NJ, USA: 2022. Robust peak detection for holter ECGs by self-organized operational neural networks. [DOI] [PubMed] [Google Scholar]
  • 27.Revanth N., Bennet M.A. Detection of QRS Complexes in ECG Signal; Proceedings of the 2023 IEEE International Conference on Integrated Circuits and Communication Systems (ICICACS); Raichur, India. 24–25 February 2023; Piscataway, NJ, USA: IEEE; pp. 1–5. [Google Scholar]
  • 28.Koka T., Muma M. Fast and Sample Accurate R-Peak Detection for Noisy ECG Using Visibility Graphs; Proceedings of the 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC); Glasgow, UK. 11–15 July 2022; Piscataway, NJ, USA: IEEE; pp. 121–126. [DOI] [PubMed] [Google Scholar]
  • 29.Howell L., Porr B. High precision ECG Database with Annotated R Peaks, Recorded and Filmed Under Realistic Conditions. 2018. [DOI]
  • 30.Kalidas V., Tamil L. Real-time QRS detector using Stationary Wavelet Transform for Automated ECG Analysis; Proceedings of the 2017 IEEE 17th International Conference on Bioinformatics and Bioengineering (BIBE); Washington, DC, USA. 23–25 October 2017; pp. 457–461. [Google Scholar]
  • 31.Emrich J., Koka T., Wirth S., Muma M. Accelerated Sample-Accurate R-Peak Detectors Based on Visibility Graphs; Proceedings of the 2023 31st European Signal Processing Conference (EUSIPCO); Helsinki, Finland. 4–8 September 2023; Piscataway, NJ, USA: IEEE; pp. 1090–1094. [Google Scholar]
  • 32.Chatterjee S., Thakur R.S., Yadav R.N., Gupta L., Raghuvanshi D.K. Review of noise removal techniques in ECG signals. IET Signal Process. 2020;14:569–590. doi: 10.1049/iet-spr.2020.0104. [DOI] [Google Scholar]
  • 33.Miskowicz M. Event-based Control and Signal Processing. CRC Press; Boca Raton, FL, USA: 2016. Reducing communication by event-triggered sampling; pp. 37–58. [Google Scholar]
  • 34.Porr B., Howell L., Stournaras J., Nir Y. Popular ECG R peak detectors written in python, (1.3.3) Zenodo; Geneva, Switzerland: 2023. [DOI] [Google Scholar]
  • 35.Goldberger A., Amaral L.A., Glass L., Hausdorff J.M., Ivanov P.C., Mark R.G., Mietus J.E., Moody G.B., Peng C.-K., Stanley H.E. PhysioBank, PhysioToolkit, and PhysioNet: Components of a new research resource for complex physiologic signals. Circulation. 2000;101:e215–e220. doi: 10.1161/01.CIR.101.23.e215. [DOI] [PubMed] [Google Scholar]
  • 36.Pohlmann K.C. Principles of Digital Audio. McGraw-Hill Professional; New York, NY, USA: 2000. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable, data used are already available as public database.