Abstract
Dynamic multiple light scattering (DMLS) has found numerous applications, including soft matter physics and biomedical optics. Yet biological tissues may have complex internal geometries, presenting a challenge for noninvasive measurements. Deciphering laminar dynamics is crucial to accurately interpret tissue or organ physiology. Seminal DMLS work noted that one can probe deeper layers indirectly by analyzing light fluctuations on shorter time scales. Recent technologies have enabled probing deeper layers directly by analyzing fluctuations at longer path lengths. The following question arises: are the indirect and direct approaches synergistic or redundant? Here, by adding an optical switch to path-length-filtered interferometric diffusing wave spectroscopy, we experimentally address this question in the context of a forearm occlusion study. We find that both approaches afford better distinction of light scattering dynamics in layered tissues than either approach alone. This motivates further development of methods that integrate both decorrelation time scale and light path length to probe layered tissues.
In biomedical optics, dynamic multiple light scattering (DMLS) from the blood gives rise to measurable fluctuations that can provide a quantitative index of blood flow [1]. This blood flow index (BFI) is measured by coherent detection in diffuse correlation spectroscopy (DCS) [1]. While DCS can noninvasively probe deep tissues, biological tissues are layered. One major challenge for DCS is to assess a deep layer in the presence of signals from intervening tissues. Meeting this challenge requires depth specificity: high sensitivity to deeper layers and low sensitivity to more superficial layers. Even though DCS is a continuous-wave (CW) technique and, as such, provides no direct path-length information, a seminal work showed that long light paths impact coherent light fluctuations at shorter time scales [2]. Thus analyzing the field autocorrelation at early time lags is an approach, albeit indirect, to probe deeper layers. Interestingly this approach is used sparingly in the field [3,4] and sometimes with negligible effect [5,6].
A more direct and intuitive approach to probe deeper layers is to select longer tissue paths [7]. Over the years, creative approaches have been applied to resolve the path length in DMLS of suspensions [8,9]. More recently, a number of technologies have emerged to directly provide the path length of coherent light fluctuations in vivo, including time–domain DCS [4,10,11], path-length-resolved (interferometric) DCS [12], interferometric NIRS [13], and interferometric diffusing wave spectroscopy (iDWS) with a TOF filter [14]. However, path-length-discriminated approaches are more complicated than simple CW approaches and, sometimes, photon-starved. The following questions arise: is it sufficient to use simpler CW techniques [15] and fit early decorrelation to assess deep flow [Fig. 1(a)]? Or is there a distinct benefit to detecting path-length-resolved light fluctuations [Fig. 1(b)]?
Fig. 1.

Two approaches to probe deeper layers in DMLS: (a) indirectly, by selecting early time scales in the autocorrelation, and (b) directly, by selecting longer paths. S, source; C, collector.
Answering these questions experimentally requires a TOF-discriminated approach with a signal-to-noise ratio to precisely measure early time scales where decorrelation is subtle. Here we employ path-length-filtered iDWS [14], which provides this set of characteristics. We realize a path-length filter by modulating the effective temporal coherence of the iDWS laser [Fig. 2]. We also introduce a pair of optical switches into the reference arm, alternating between longer paths and shorter paths at the heart rate. Critically, this system can discriminate short and long light paths quasi-simultaneously, enabling statistically powerful paired single-trial comparisons. Its many detection channels enable a high measurement SNR. To evaluate approaches to achieve depth specificity, we use a classic forearm occlusion protocol, with well-characterized laminar overshoot responses, characterized previously with continuous-wave (CW) DCS by increasing a source–collector (S–C) separation [16]. We analyze changes in BFI for long and short paths during both occlusion and hyperemic overshoot.
Fig. 2.

(a) Path-length-filtered iDWS setup with optical switches allows the user to electronically choose between two TOF filter centers and also specify the TOF filter width. (b) As implemented, the system selects either long or short paths corresponding to the either the late or early TOF filter, respectively. S1, S2: MEMS fiber switches, ,: long and short paths.
The iDWS system [Fig. 2(a)] utilizes path-length discrimination [14] with added optical switching to select between two reference paths. An instantaneously narrow linewidth (~10 MHz) DBR laser at 852 nm, with a sample illumination power of approximately 70 mW, is current modulated by an RF sinusoid at 9.79 MHz and user-specified (variable) peak-to-peak current. Current modulation at this frequency induces optical frequency modulation via carrier density modulation, broadening the effective power spectrum and reducing the effective temporal coherence, as modeled with a Bessel-squared TOF filter [14], which is maximized for matched sample and reference path lengths [Fig. 2(b)].
Our previous path-length-filtered iDWS [14] was able to achieve either short or long path-length selectivity, but changing the filter center required swapping reference fibers. In this work, we take the technology one step further, incorporating switches to quasi-concurrently (to within a heartbeat) measure both short and long paths. The fiber optic switches are electronically controlled micro-electromechanical (MEMS) devices, which are quasi-synchronously (switching from early to late TOFs takes ~70 ms, whereas switching from late to early TOFs takes 15 ms) operated to select either reference path (long) or (short), which differ in their respective time delays. The switchable reference paths centered the TOF filter at either late (, ) or early (, ) sample TOFs [Fig. 2(b)]. The switches transition between the two paths at approximately the heart rate, so that in successive heart beats, the system selects long paths and then short paths . An analog output card (National Instruments) which synchronously controls both the RF signal generator and the switches such that switching reference paths is timed with RF voltages and , corresponding to the optimal filter widths, and , as defined below, for each respective path. The selection of long/short photon paths is equivalent to selection of late/early TOFs, as shown on a simulated time-of-flight distribution (DTOF) of the human forearm (, [16]) in Fig. 2(b). The first zero of the TOF filter is from the peak [Fig. 2(b)].
Are our early and late TOF filters able to achieve differential sensitivity to the skeletal muscle as hypothesized in Fig. 1(b)? For our long (short) path filter at 1 cm S–C separation, the mean/centroid of the filtered DTOF is estimated as 0.76 ns (0.18 ns), as compared to 0.25 ns for the unfiltered (CW) DTOF . Interestingly, a CW measurement at 2.7 cm (0.7 cm) achieves the same mean TOF as our late (early) TOF filter at 1 cm. Though a mean TOF is not a perfect proxy of penetration depth, it is well accepted that 2.7 cm and 0.7 cm S–C separations achieve differential sensitivity to the skeletal muscle [16], suggesting that our measurements do as well.
The correspondence between the RF voltage amplitude V and is obtained through a separate fiber Mach–Zehnder interferometer (Fig. S4 in Supplement 1 [14]). The interference fringes are recorded for each voltage from 0 to 10 , and an algorithm that measures phase evolution by counting peaks and valleys between stationary points gives a measure of . We found this method to yield a more accurate measure of the tuning range than the conventional Hilbert transform method [14]. Though our algorithm indicated a nonlinear relationship between the RF voltage V and , this relationship could be calibrated before the experiment.
Since our TOF-filtered approach does not provide direct access to optical path-length distribution, the optimal filter width was chosen to maximize the decay rate for the deep path and to minimize the autocorrelation decay rate for the superficial path . These criteria are based on the premise that the magnitude of the decay rate increases with the path length [2]. Thus, for which aims for greater path lengths, the filter yielding the highest decay rate is optimal, whereas for which aims for lesser path lengths, the filter yielding the lowest decay rate is optimal. Figure 3 shows determination of the optimal filter width for long and short paths, based on the rules discussed above. For long paths [Fig. 3(a)], the optimal filter width occurred at (corresponding to 1.64 ns peak to first zero of the Bessel filter), whereas for short paths [Fig. 3(c)], the optimal filter width occurred at (corresponding to 0.48 ns peak to first zero of the Bessel filter). Note that the optimal filter width is subject and location dependent in general. It is instructive to examine the effects of the filter on the light intensity as well. The optimal filter width for long paths corresponds nearly, but not exactly, to the minimum of the intensity, which arises when the zero of the mainlobe of the Bessel filter lies at the peak of the DTOF [Fig. 2(b)], thus rejecting early photons. The optimal filter width for short paths corresponds to a quasi-plateau of weakly decreasing intensity (at least within the voltage limit of the RF generator). As the early filter is centered near the DTOF peak, a lower-filtered intensity indicates filter narrowing and better rejecting late photons [Fig. 2(b)].
Fig. 3.

Decay rate and filtered intensity as a function of filter width, parameterized by , for long [(a) and (b)] and short [(c) and (d)] paths. Optimal filter locations, defined by the criteria discussed in the text, are denoted by stars. Shading denotes standard deviation across three consecutive filter sweeps.
We put the system to use in a classic forearm occlusion protocol [Fig. 4(a)], which comprised 3 states: 20 s baseline, 30 s occlusion, and 20 s release where a hyperemic peak is observed [Fig. 4(b)]. Processing steps include: (1) parsing of the switch data in order to effectively segregate the data collected for paths or and processing them independently; (2) rolling mean subtraction, camera background removal, and pixel binning; (3) computing the interferometric autocorrelation after reference autocorrelation subtraction; (4) estimating through appropriate fitting (double exponential fit up to 20 lags for late TOFs and quadratic fit up to 10 lags for the early TOFs to capture the curvature effectively), leading to as a function of lag for baseline (averaged over ~11 heartbeats), occlusion (averaged over ~17 heartbeats), and peak hyperemia (averaged over one heartbeat), where a heartbeat is ~0.8 s [Fig. 4(c)]; and (5) calculating the BFI relative to baseline for paths and across states.
Fig. 4.

(a) Forearm occlusion protocol for system validation. (b) BFI relative to baseline (rBFI) for each path calculated for . (c) Autocorrelation curves for long and short paths at baseline (averaged ~11 heartbeats), occlusion (averaged ~17 heartbeats), and peak overshoot (averaged single heartbeat). (d) Autocorrelation curves for long paths overlaid with rescaled short paths curves from (c) with the same legend as (c); green dotted box: zoom (S, source; C, collector).
The curves were found to be concave down at early lags. Such negative curvature is not expected for pure Brownian motion, as is typically assumed in DCS [1]. Therefore, to account for curvature at early lags, we fit to a model proposed by Carp et al. [17,18], which postulates an intrinsic time scale for randomization, . Fits to the semi-infinite DCS model were biased compared to this model. Specifically,
| (1) |
where
| (2) |
Here is the autocorrelation lag time, is the DTOF with being the time-of-flight filter, and is the BFI. We fit for and for every curve to obtain a time course as well as for baseline (averaged ~11 heartbeats for our datasets), occlusion (averaged ~17 heartbeats), and peak hyperemia (single heartbeat). Fitting was performed over a range given by . This approach enables comparisons independent of the autocorrelation decay rate. rBFI is taken as with respect to baseline. When rBFI is determined with [Fig. 4(b)], long paths show higher overshoot than the short paths. This suggests the efficacy of longer photon paths in achieving better depth sensitivity, as deeper muscle tissue shows higher overshoots [16].
The curvature of was highlighted by rescaling the -axis for short paths so that the crossing overlaps with that of long paths [Fig. 4(d)]. While the reasons for this deviation from the ideal Brownian model are beyond the study scope, empirically we can account for this downward curvature with the parameter , as mentioned above.
Figures 5(a) and 5(b) show rBFI during occlusion and peak overshoot, respectively, as a function of , clearly demonstrating higher rBFI at early lags during a peak overshoot, and the opposite trend during occlusion. Figures 5(c) and 5(d) show the rBFI time course for long and short paths, respectively. For long paths, the highest peak rBFI is observed for the earliest lags, whereas the opposite is true for short paths [Fig. 5(c)]. This suggests that even with a path-length filter [Fig. 2(b)], early (10–100 μs) fluctuation time scales must be accessed to probe layered tissues with better depth sensitivity.
Fig. 5.

rBFI during occlusion (a) and peak overshoot (b), as a function of , for both long and short paths. (c) and (d) rBFI time course as a function of , for long paths and short paths , respectively.
Figure 6 shows the results of switching between short path filtering and no filtering (CW). CW and short path filtering show systematic differences [Figs. 6(a)–6(b)]. A larger overshoot in CW mode is evident in Figs. 6(c)–6(d). This result parallels Figs. 5(c) and 5(d), where longer paths show a larger overshoot.
Fig. 6.

Comparison of continuous-wave (CW) and short paths. rBFI during occlusion (a) and peak overshoot (b), as a function of , for both CW and short paths. (c) and (d) rBFI time course as a function of threshold, for CW and short paths, respectively.
Having established that short path filtering is beneficial, we performed 8 trials of forearm occlusion with quasi-concurrent switching between long and short paths. We fit to Eq. (2) for different to obtain rBFI as a function of [Figs. 7(a) and 7(b)]. For peak hyperemic overshoot [Fig. 7(b)], rBFI peaks at later lags tend to be similar for both paths. When early lags are selected, however, rBFI for long paths exceeds that for short paths. Such differences can only be appreciated when considering different time scales (lags) and path lengths. While these results parallel those reported in [4], our work benefits from a high autocorrelation SNR afforded by >100 parallel channels, which enables higher fidelity curves to observe the effects of continuously varying fitting range. Moreover, this work is distinct in the use of a model with a physical basis [18], whereas [4] follows an empirical bi-exponential fitting approach.
Fig. 7.

Shaded plot (mean ± std. err) summarizing rBFI for different thresholds for long versus short paths during occlusion (a) and peak hyperemic overshoot (b). Results show a synergistic interaction of path length and fluctuation time scale. Note that the occlusion values are averaged over a 17-heartbeat time window, and as such, do not represent the minimal observed BFI.
An interesting additional finding in this study was the variation of . We observe that for , for long paths is always lower than that for short paths during all phases. The mean (std error) values of for long paths are 2.68 μs (0.87 μs) for baseline, 34.45 μs (8.10 μs) for occlusion, and 0.84 μs (0.08 μs) for overshoot, while for short paths, they are 10.47 μs (2.01 μs) for baseline, 69.21 μs (9.94 μs) for occlusion, and 1.79 μs (0.81 μs) for overshoot. We observe that for occlusion is higher than that for baseline, and that for peak is lower than during baseline, irrespective of the chosen paths or , but it has to be noted that the values for overshoot estimated through fitting are much lower than the smallest lag measurable. Since the explanation for this behavior would entail further investigation into factors influencing the ballistic-to-diffusion regime, we refrain from analyzing in detail as it is beyond the study scope.
In conclusion, we have experimentally demonstrated the merits of applying path-length filtering and short fluctuation time scale analysis together to probe deep tissue layers, demonstrating clear-cut improvements over either approach alone. This can pave the way for the synergistic use of these techniques involving path-length resolution and short lag correlators such as TD-DCS or iNIRS to obtain better depth sensitivity in critical care neuromonitoring or transabdominal fetal oximetry applications.
Funding.
National Institutes of Health (EB029747, EB032840, EY031469).
Footnotes
Disclosures. The authors declare no conflict of interest.
Data availability.
Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.
REFERENCES
- 1.Durduran T, Choe R, Baker WB, et al. , Rep. Prog. Phys 73, 076701 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Pine DJ, Weitz DA, Chaikin PM, et al. , Phys. Rev. Lett 60, 1134 (1988). [DOI] [PubMed] [Google Scholar]
- 3.Verdecchia K, Diop M, Lee A, et al. , Biomed. Opt. Express 7, 3659 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Samaei S, Sawosz P, Kacprzak M, et al. , Sci. Rep 11, 1817 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Wu M, Chan S-T, Mazumder D, et al. , Neurophotonics 8, 015001 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Selb J, Boas D, Chan S-T, et al. , Neurophotonics 1, 015005 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Torricelli A, Contini D, Pifferi A, et al. , NeuroImage 85, 28 (2014). [DOI] [PubMed] [Google Scholar]
- 8.Yodh AG, Kaplan PD, and Pine DJ, Phys. Rev. B 42, 4744 (1990). [DOI] [PubMed] [Google Scholar]
- 9.Bizheva KK, Siegel AM, and Boas DA, Phys. Rev. E 58, 7664 (1998). [Google Scholar]
- 10.Sutin JDB, Zimmerman B, Tyulmankov D, et al. , Optica 3, 1006 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Pagliazzi M, Sekar SKV, Colombo L, et al. , Biomed. Opt. Express 8, 5311 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Safi AM, Moka S, Harrah M, et al. , in Biophotonics Congress 2021, OSA Technical Digest (Optica Publishing Group, 2021), p. BTh1B.6. [Google Scholar]
- 13.Kholiqov O, Zhou W, Zhang T, et al. , Nat. Commun 11, 391 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Zhao M, Zhou W, Aparanji S, et al. , Optica 10, 42 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Shang Y and Yu G, Appl. Phys. Lett 105, 133702 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Yu G, Durduran T, Lech G, et al. , J. Biomed. Opt 10, 024027 (2005). [DOI] [PubMed] [Google Scholar]
- 17.Berne BJ and Pecora R, Dynamic Light Scattering: with Applications to Chemistry, Biology, and Physics (Courier Corporation, 2000). [Google Scholar]
- 18.Carp SA, Roche-Labarbe N, Franceschini M-A, et al. , Biomed. Opt. Express 2, 2047 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.