Abstract
Background
Microscopic imaging of cartilage is a key tool for the study and development of treatments for osteoarthritis. When cellular and sub‐cellular resolution is required, histology remains the gold standard approach, albeit limited by the lack of volumetric information as well as by processing artifacts. Cartilage imaging with the sub‐cellular resolution has only been demonstrated in the synchrotron environment.
Purpose
To provide a proof‐of‐concept demonstration of the capability of a laboratory‐based x‐ray phase‐contrast microscope to resolve sub‐cellular features in a cartilage sample.
Methods
This work is based on a laboratory‐based x‐ray microscope using intensity‐modulation masks. The structured nature of the beam, resulting from the mask apertures, allows the retrieval of three contrast channels, namely, transmission, refraction and dark‐field, with resolution depending only on the mask aperture width. An ex vivo equine cartilage sample was imaged with the x‐ray microscope and results were validated with synchrotron tomography and histology.
Results
Individual chondrocytes, that is, cells responsible for cartilage formation, could be detected with the laboratory‐based microscope. The complementarity of the three retrieved contrast channels allowed the detection of sub‐cellular features in the chondrocytes.
Conclusions
We provide the first proof‐of‐concept of imaging cartilage tissue with sub‐cellular resolution using a laboratory‐based x‐ray microscope.
Keywords: cartilage imaging, cellular imaging, dark‐field, phase‐contrast imaging, soft tissue imaging, x‐ray microscopy
1. INTRODUCTION
Imaging techniques play a pivotal role in both research and clinical practice for osteoarthritis (OA), including for diagnosis and monitoring of the disease, as well as studies in the etiology of the disease and the development of treatment. In vivo studies are typically performed with magnetic resonance imaging (MRI), 1 as x‐ray techniques typically lack the contrast required to visualize cartilage. X‐ray radiology however remains the main approach in the clinic for OA which is visualized as articular joint space narrowing that typically manifests in advanced disease. At the microscopic scale however, which is particularly relevant for assessing re‐cellularization and growth in tissue engineered models, histology is the most commonly used technique. While histology can be considered the gold standard for many biological and medical applications, it lacks the volumetric information available in tomographic imaging and restricts the visualization to a fixed anatomical orientation. Additionally, as a destructive technique, it is affected by processing (embedding and cutting) artifacts in the plane of dissection and, thus, observation. 2
Phase‐contrast x‐ray microscopy has the potential to overcome these limitations, by providing non‐destructive tomographic imaging with isotropic resolution in the micron length scale for samples in their native state. Phase‐contrast imaging relies on measuring the phase‐shift imparted by the sample to the x‐ray wave, as opposed to conventional x‐ray imaging which measures variations in x‐ray intensity (absorption contrast). Since changes in phase can be order of magnitude larger than changes in attenuation for soft tissue samples at relatively low x‐ray energies, an increased contrast for soft tissue is expected for phase‐contrast imaging compared to absorption imaging. Several studies have been published in the last two decades demonstrating the suitability of synchrotron‐based x‐ray phase‐contrast techniques for cartilage imaging, both in vivo 3 and ex vivo. 4 , 5 , 6 , 7 Examples of laboratory‐based cartilage imaging at macroscopic level exist for several phase‐contrast imaging methodologies, including analyzer‐based imaging, 8 grating interferometry, 9 propagation‐based imaging 10 and edge‐illumination. 11 These studies mainly demonstrate the capability of phase‐contrast imaging to distinguish between cartilage and bone tissues, due to the increased contrast arising from measuring phase shifts. However, images reported in the aforementioned studies are unable to capture structural information within the cartilage. This is attributed to insufficient imaging system resolution as opposed to the underlying physical principles of these phase‐based imaging systems. More recently, Horng et al. 12 reported on high‐resolution synchrotron‐based techniques for the visualization of cellular and sub‐cellular features in healthy and degenerated cartilage samples.
In this article, we report on the first proof‐of‐concept results of imaging ex vivo cartilage with sub‐cellular resolution using a laboratory‐based x‐ray phase‐contrast microscope. The laboratory results are validated with the gold standard synchrotron tomography, as well as histology. The complementarity of the retrieved contrast channels is also discussed with respect to cellular morphology.
2. MATERIALS AND METHODS
2.1. Sample preparation
A thin section ( mm3) of equine articular cartilage was excised post‐mortem from the metacarpophalangeal joint. Horse tissues were obtained following approval from the Royal Veterinary College Clinical Research Ethical Review Board (URN 2020 2017‐2). The sample was fixed in 4% paraformaldehyde and dehydrated in 70% ethanol. The sample was transferred to a 1‐mm diameter glass capillary (Hampton Research) and kept in an ethanol atmosphere.
2.2. Synchrotron imaging
The cartilage sample was imaged at Diamond Light Source (DLS) on the I13‐2 beamline. A so‐called “pink” polychromatic beam with mean energy of 25 keV was used to perform tomographic imaging of the sample. A Pco.edge 5.5 camera with effective pixel size of 0.8 µm was used and placed at 20 cm from the sample, while the source to sample distance was 200 m. The scan, performed with continuous sample rotation in the so‐called fly‐scan mode, consisted of 1800 projection images over 180 degrees for a total scan time of 5 min. Data were reconstructed using a filtered‐back‐projection algorithm following application of Paganin's phase retrieval 13 to all individual projections, implemented within the DLS Savu environment. 14
2.3. Laboratory imaging
Planar imaging of the cartilage sample was obtained with a laboratory‐based beam tracking (BT) system. 15 The set‐up 16 consists of a rotating Cu anode x‐ray source and a doubly‐curved multi‐layer monochromator, selecting the Cu Kα lines at 8 keV and focusing the beam into a 350‐µm focal spot. As schematically shown in Figure 1, the focused beam is structured into an array of beamlets by a 20‐ µm‐thick gold membrane featuring 2‐µm‐wide slits with a 20‐µm periodicity. The structured beamlets were imaged by a CMOS sensor, coupled to a scintillator and objective, with an effective pixel pitch of 1.1 µm and a field of view of 1.6× 1.6 mm2. A sample to detector distance of 15 mm was used, while the source to sample distance was 2.2 m. Changes in the shaped beamlets, with respect to a reference dataset without the sample in place, allow for the retrieval of sample properties: namely, transmission, refraction, and dark‐field or scattering. 15 The transmission and refraction contrast channels allow quantification of the imaginary and real part of the complex refractive index (), respectively, where λ is the x‐ray wavelength. Provided that the shaped beamlets are adequately sampled (i.e., a pixel size smaller than the beamlet projected width at the detector position is used) and that cross‐talk between beamlets can be kept below ≈ 20% (i.e., the projected focal spot size is smaller than the mask period), the system resolution is exclusively determined by the mask aperture width, rather than source size and detector pixel pitch as it is the case for other x‐ray imaging modalities. This resolution limit of imaging systems based on intensity‐modulation masks has been theoretically 17 and experimentally demonstrated. 18 A recent experiment, performed with the microscope described in this note and with a mask with smaller aperture width (1.5 µm), showed the increase in resolution expected from the aperture width reduction. 19 It is important to note that the resolution limit of the microscope discussed so far is only applicable to the direction of phase sensitivity, that is, orthogonally to the mask apertures. In the direction parallel to the mask slits, resolution is limited by the projected source size and detector pixel pitch. In fact, the one‐dimensional nature of the structured illumination of this set‐up creates a preferential direction both in terms of phase sensitivity and spatial resolution. This is particularly suited for imaging structures which have a strong directionality, for example, the nervous system and collagen. Isotropic resolution and phase sensitivity could be achieved through sample (or mask) rotation around the optical axis, or by adapting a two‐dimensional structured illumination like that shown in refs. [20, 21] to a microscopy system, which would however lead to longer exposure times due to the reduced mask open fraction. Transmission and refraction are associated with detection of individual sample features above the system resolution limit, that is, the mask aperture width. Dark‐field, 22 , 23 on the other hand, is sensitive to ensembles of sample features below the system resolution: while features are not individually resolved, the presence of the ensemble is highlighted.
FIGURE 1.

A schematic representation of the Beam Tracking (BT) principle. Profiles of shaped beamlets with (red) and without (blue) sample in place are shown, where changes in amplitude (T), position (R) and width (DF) of the beamlets are also highlighted.
In order to achieve the aperture‐driven resolution demonstrated in ref. [18], it is necessary to translate the sample by a distance equal to the mask period in the direction perpendicular to the mask slits, acquire images at each step, and recombine them (a process we refer to as dithering 24 ). The size of the dithering step is bound to be equal to or greater than the Nyquist frequency associated with the aperture width, to avoid aliasing. The sample was dithered in 20 steps of 0.95 µm, for an overall number of 20 planar images with an exposure time set to 10 s. For the retrieval of the three contrast channels, amplitude, mean and variance of each beamlet were extracted with and without the sample in place. Transmission (T), refraction (R) and dark‐field () were obtained from the measured beamlet parameters using the following relationships:
| (1) |
where the explicit dependence of T and R on β and δ is shown. Φ is the phase shift imparted to beam by the sample and the integration has to be extended over the depth of the object along the optical axis.
2.4. Histology
Due to limitations in performing standard histology processing for this sample arising from the dehydration occurring in ethanol, the sample was stained as a whole with Alcian blue and nuclear fast red dyes and imaged with an optical microscope (Leica DMi1, Leica Microsystems). Specifically, the sample was briefly hydrated in deionized water and then stained in 1% Alcian blue (pH 1.0) for 15 min, rinsed briefly in deionizsed water and then counterstained with nuclear red stain for 2 min. The samples were rinsed in water before microscopy imaging.
3. RESULTS AND DISCUSSION
A three‐dimensional representation of the tomogram acquired at DLS is shown in Figure 2a and compared with a similar view of the whole sample with histology stain (panel b). Magnified views of three regions of interest of the histology are reported in panels c–e. The uptake of the Alcian blue dye in the histology confirms the nature of the sample's tissue. Additionally, circular features visible in the histology can be identified as chondrocytes, due to their typical columnar alignment (marked with red arrowheads in the histology images), typical of deep cartilage layers. An analogous columnar arrangement for chondrocytes is highlighted in a reconstructed CT slice shown in panel f, matching the histology image of panel c. A feature that appeared to be the tidemark was visible (highlighted by green arrowheads in panel f) in the slice, identifying the transition between calcified and non‐calcified regions of the sample corresponding to the brighter (higher density) and darker (lower density) areas of the tomographic data, respectively. These findings are consistent with previously published literature, 25 showing how the columnar arrangement of the chondrocytes is prominent just above the tidemark.
FIGURE 2.

(a) Volumetric rendering of the cartilage sample from the DLS tomogram (scale bar 400 µm). Cut‐outs were extracted from the volume to allow the visualization of specific areas matching the histology. (b) Whole sample histological image, obtained by stitching two microscopy images due to a limited field of view. Specific areas, highlighted in the dashed boxes, are magnified in panels (c–e). Red arrowheads mark the columnar arrangement of chondrocytes. (f) Sagittal slice of the tomogram. Red arrowheads mark the columnar arrangement of chondrocytes. Green arrowheads identify the tidemark line. (g) Three‐dimensional visualization of segmented chondrocytes nuclei. (h) Distribution of nuclei diameter in the segmented chondrocytes.
Chondrocyte nuclei were segmented in the calcified area of the sample using threshold‐based and watershed segmentation algorithms in Avizo (Thermo Fisher Scientific, version 2022.1). A three‐dimensional rendering of the segmented nuclei is shown in Figure 2g), highlighting the expected columnar architecture. The distribution of the measured diameter of nuclei was measured (see Figure 2h), resulting in a mean nucleus size of 10.8 ± 0.1 µm, calculated over 1145 segmented cells, comparable with results reported in literature. 26 , 27
Figure 3 shows the planar images obtained using the laboratory microscope, for the transmission (), refraction and dark‐field contrast channels in panels a‐c. The transmission image reproduces the morphology of chondrocytes, including nucleus, cytoplasm and extra‐cellular matrix, in agreement with the DLS tomogram (Figure 3f). The refraction image highlights the edges of the cells' nuclei along the direction of sensitivity, that is, the one orthogonal to the mask apertures. The transmission image shows a higher contrast‐to‐noise ratio compared to refraction and dark After intensity‐based segmentation, the three contrast channels were fused to provide a single RGB image, with the three channels associated to transmission, absolute value of refraction and absolute value of dark‐field, respectively (panel d). The color rendering shows the complementarity of the three contrast channels. The refraction image was integrated 28 to obtain the phase image (see Equation 1), shown in panel e, giving a clearer picture of the capability of the system to resolve chondrocytes' nuclei (highlighted by red arrowheads).
FIGURE 3.

(a) Transmission, (b) refraction and (c) dark‐field retrieved images of the cartilage sample (scale bar 100 µm). (d) RGB image obtained by fusing images in panels (a–c) after intensity‐based segmentation. Red represents transmission, green the absolute value of refraction and blue scattering. (e) Integrated phase image. Red arrowheads mark chondrocytes' nuclei. (f) Volumetric rendering of the same area of the sample from the DLS tomogram. The yellow arrow in panel (a) identifies the chondrocyte shown in Figure 4.
Although sub‐cellular details are visible in all retrieved images, the transmission image (panel a) shows an increased contrast‐to‐noise ratio (CNR) compared to refraction and dark‐field (panels b–c), due to a higher background noise for the two phase‐based contrast channels arising from environmental instabilities over the timescale of the experiment. Some artifacts are visible in all retrieved images (a–c). Specifically, horizontal structured noise, oriented in the direction orthogonal to the mask slits, is visible in the retrieved images. This is due to the presence of absorbing bridges (or cross‐links) in the free‐standing mask supporting the open slits. Additionally, a brick‐like periodic noise is visible, which arises from the recombination of dithering steps in the presence of environmental vibrations that create a degree of inconsistency between successive steps. A magnified view of an individual chondrocyte (identified by a yellow arrow in Figure 3a) is provided in Figure 4, with relevant intensity profiles across the cell's nucleus. The transmission profile shows a variation in intensity, due to both variation in density between nucleus and cytoplasm and to the varying thickness of the cell. The differential refraction signal peaks at the nucleus' edges, highlighting a diameter of ≈ 10 µm, in agreement with the synchrotron‐based estimation. Finally, the dark‐field signal shows a positive value in the nuclear region, due to the presence of features below the resolution limit of the microscope.
FIGURE 4.

Detail of transmission (black), refraction (red) and dark‐field (blue) images for an individual chondrocyte. Intensity profiles for the channels are also shown. The dark‐field image is displayed at a different contrast level compared to Figure 3c to allow visualization of the cellular details.
4. CONCLUSIONS
This study provides the first proof‐of‐concept of imaging sub‐cellular features in cartilage with a laboratory‐based x‐ray microscope, validated with synchrotron data and gold standard histology. The presence of sub‐cellular structures in chondrocytes, such as nucleus and cytoplasm, can be revealed by making use of the complementarity of the retrieved contrast channels. It is important to note that, in this proof‐of‐concept work, only planar images were acquired with the laboratory‐based microscope. The suitability of the BT approach for tomographic reconstruction has been demonstrated before, 29 and we expect the planned additional developments of the microscope to support tomographic imaging. The main limitation in the current system preventing tomographic imaging is the number of dithering steps to be acquired at each rotation angle, with this number being proportional to the mask period. Free‐standing masks with a shorter period (7.5 µm to be compared with 20 µm for the mask used in this work) were recently fabricated 19 for the prototype microscope, drastically reducing the number of required dithering steps. Additionally, the use of alternative x‐ray optics, producing a smaller source focal spot, would allow reduction of the source to sample distance, leading to a higher flux and, thus, a reduction in total scan time. We expect the planned upgrades, combined with a cycloidal acquisition scheme, 30 to allow for tomographic scans to be acquired in a relatively short scan time (≈ 1 h). The newly fabricated mask features bridges (responsible for the structured noise in Figure 3) placed every 400 µm, rather than 50 µm as in the mask used in this work, will largely reduce the structured noise in the retrieved images. Further developments could include the capability to image thicker samples through the use of higher energy monochromatic beams, for example, Ga Kα (9.2 keV) or Mo Kα (17.5 keV), which would lead to a gain in accessible tissue thickness of a factors 2 and 10, respectively.
CONFLICT OF INTEREST STATEMENT
Dr J. Ferrara is a Rigaku employee; Rigaku has a potential interest in the commercial exploitation of the results presented in this abstract. Prof M. Endrizzi and Alessandro Olivo are named inventors on patents protecting the described technology. All other authors have no financial interests to disclose.
ACKNOWLEDGMENTS
Research reported in this publication was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award Number R01EB028829. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. Additional support was received from the EPSRC (grants EP/T005408/1, EP/P023231/1 and EP/M028100/1), the Wellcome Trust 221367/Z/20/Z, and the European Union's Horizon 2020 research and innovation programme under Grant Agreement No. 777222. AO was supported by the Royal Academy of Engineering under their “Chairs in Emerging Technologies” scheme. SS is a UKRI EPSRC Doctoral Prize Fellow (EP/T517793/1). The authors gratefully acknowledge the provision of beamtime MG31421‐1 on ID13‐2 at the Diamond Light Source and the assistance of the beamline scientists there. Use of equine tissue was approved by the Royal Veterinary College Clinical Research Ethical Review Board (URN 2020 2017‐2). This work is dedicated to our late friend and colleague Prof Paul F. McMillan, who kickstarted this collaborative endeavour through his insatiable scientific curiosity and enthusiasm for cross‐disciplinary research. He is, and will be, sorely missed.
Esposito M, Astolfo A, Cipiccia S, et al. Technical note: Cartilage imaging with sub‐cellular resolution using a laboratory‐based phase‐contrast x‐ray microscope. Med Phys. 2023;50:6130–6136. 10.1002/mp.16599
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
- 1. Hayashi D, Roemer FW, Guermazi A. Imaging of osteoarthritis‐recent research developments and future perspective. Br J Radiol. 2018;91:20170349. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Savvidis S, Gerli MFM, Pellegrini M, et al. Monitoring tissue engineered constructs and protocols with laboratory‐based x‐ray phase contrast tomography. Acta Biomater. 2022;141:290‐299. [DOI] [PubMed] [Google Scholar]
- 3. Bennett EE, Kopace R, Stein AF, Wen H. A grating‐based single‐shot x‐ray phase contrast and diffraction method for in vivo imaging. Med Phys. 2010;37:6047‐6054. _eprint: https://onlinelibrary.wiley.com/doi/pdf/10.1118/1.3501311 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Horng A, Brun E, Mittone A, et al. Cartilage and soft tissue imaging using X‐rays: propagation‐based phase‐contrast computed tomography of the human knee in comparison with clinical imaging techniques and histology. Invest Radiol. 2014;49:627‐634. [DOI] [PubMed] [Google Scholar]
- 5. Coan P, Bamberg F, Diemoz PC, et al. Characterization of osteoarthritic and normal human patella cartilage by computed tomography X‐ray phase‐contrast imaging: a feasibility study. Invest Radiol. 2010;45:437‐444. [DOI] [PubMed] [Google Scholar]
- 6. Majumdar S, Issever AS, Burghardt A, et al. Diffraction enhanced imaging of articular cartilage and comparison with micro‐computed tomography of the underlying bone structure. Eur Radiol. 2004;14:1440‐1448. [DOI] [PubMed] [Google Scholar]
- 7. Li J, Zhong Z, Connor D, Mollenhauer J, Muehleman C. Phase‐sensitive X‐ray imaging of synovial joints. Osteoarthritis Cartilage. 2009;17:1193‐1196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Muehleman C, Li J, Connor D, Parham C, Pisano E, Zhong Z. Diffraction‐enhanced imaging of musculoskeletal tissues using a conventional X‐ray tube. Acad Radiol. 2009;16:918‐923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Stutman D, Beck TJ, Carrino JA, Bingham CO. Talbot phase‐contrast x‐ray imaging for the small joints of the hand. Phys Med Biol. 2011;56:5697‐5720. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Lee YS, Heo E‐A, Jun HY, et al. Articular cartilage imaging by the use of phase‐contrast tomography in a collagen‐induced arthritis mouse model. Acad Radiol. 2010;17:244‐250. [DOI] [PubMed] [Google Scholar]
- 11. Marenzana M, Hagen CK, Borges PDN, et al. Visualization of small lesions in rat cartilage by means of laboratory‐based x‐ray phase contrast imaging. Phys Med Biol. 2012;57:8173. IOP Publishing. [DOI] [PubMed] [Google Scholar]
- 12. Horng A, Stroebel J, Geith T, et al. Multiscale X‐ray phase contrast imaging of human cartilage for investigating osteoarthritis formation. J Biomed Sci. 2021;28:42. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Paganin D, Mayo SC, Gureyev TE, Miller PR, Wilkins SW. Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object. J Microsc. 2002;206:33‐40. _eprint: https://onlinelibrary.wiley.com/doi/pdf/10.1046/j.1365‐2818.2002.01010.x [DOI] [PubMed] [Google Scholar]
- 14. Atwood RC, Bodey AJ, Price SWT, Basham M, Drakopoulos M. A high‐throughput system for high‐quality tomographic reconstruction of large datasets at diamond light source. Philos Trans R Soc, Ser A. 2015;373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Vittoria FA, Kallon GKN, Basta D, et al. Beam tracking approach for single‐shot retrieval of absorption, refraction, and dark‐field signals with laboratory x‐ray sources. Appl Phys Lett. 2015;106:224102. [Google Scholar]
- 16. Esposito M, Massimi L, Buchanan I, Ferrara JD, Endrizzi M, Olivo A. Test and optimization of a multi‐modal phase‐based x‐ray microscope for soft tissue imaging. in Medical Imaging 2022: Physics of Medical Imaging, vol 12031. SPIE; 2022:1‐7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Diemoz PC, Vittoria FA, Olivo A. Spatial resolution of edge illumination X‐ray phase‐contrast imaging. Opt Express OE. 2014;22:15514‐15529. [DOI] [PubMed] [Google Scholar]
- 18. Esposito M, Massimi L, Buchanan I, Ferrara JD, Endrizzi M, Olivo A. A laboratory‐based, low‐energy, multi‐modal x‐ray microscope with user‐defined resolution. Appl Phys Lett. 2022;120:234101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Makarova OV, Divan R, Moldovan N, et al. Freestanding high‐aspect‐ratio gold masks for low‐energy, phase‐based x‐ray microscopy. Nanotechnology. 2022;34:045301. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Navarrete‐Leon C, Doherty A, Savvidis S, et al. X‐ray phase‐contrast microtomography of soft tissues using a compact laboratory system with two‐directional sensitivity. 2023;10(7):880‐887. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Lioliou G, Roche i Morgó O, Marathe S, et al. Cycloidal‐spiral sampling for three‐modal x‐ray CT flyscans with two‐dimensional phase sensitivity. Sci Rep. 2022;12:21336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Endrizzi M, Diemoz PC, Millard TP, et al. Hard X‐ray dark‐field imaging with incoherent sample illumination. Appl Phys Lett. 2014;104:024106. [Google Scholar]
- 23. Pfeiffer F, Bech M, Bunk O, et al. Hard‐X‐ray dark‐field imaging using a grating interferometer. Nat Mater. 2008;7:134‐137. [DOI] [PubMed] [Google Scholar]
- 24. Hagen CK, Diemoz PC, Endrizzi M, Olivo A. The effect of the spatial sampling rate on quantitative phase information extracted from planar and tomographic edge illumination x‐ray phase contrast images. J Phys D: Appl Phys. 2014;47:455401. [Google Scholar]
- 25. McIlwraith CW, Frisbie DD, Kawcak CE, Fuller CJ, Hurtig M, Cruz A. The OARSI histopathology initiative – recommendations for histological assessments of osteoarthritis in the horse. Osteoarthritis Cartilage. 2010;18(3):S93‐105. [DOI] [PubMed] [Google Scholar]
- 26. Guilak F, Tedrow JR, Burgkart R. Viscoelastic properties of the cell nucleus. Biochem Biophys Res Commun. 2000;269:781‐786. [DOI] [PubMed] [Google Scholar]
- 27. Jadin KD, Wong BL, Bae WC, et al. Depth‐varying density and organization of chondrocytes in immature and mature bovine articular cartilage assessed by 3d imaging and analysis. J Histochem Cytochem. 2005;53:1109‐1119. [DOI] [PubMed] [Google Scholar]
- 28. Massimi L, Buchanan I, Astolfo A, Endrizzi M, Olivo A. Fast, non‐iterative algorithm for quantitative integration of X‐ray differential phase‐contrast images. Opt Express, OE. 2020;28:39677‐39687. Optical Society of America. [DOI] [PubMed] [Google Scholar]
- 29. Vittoria FA, Endrizzi M, Kallon GK, et al. Multimodal phase‐based X‐ray microtomography with nonmicrofocal laboratory sources. Phys Rev Appl. 2017;8:064009. [Google Scholar]
- 30. Hagen CK, Vittoria FA, Morgó ORI, Endrizzi M, Olivo A. Cycloidal computed tomography. Phys Rev Appl. 2020;14:014069. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.