A systematic algorithm for large-bore arterial access closure after TAVI: the TAVI-MultiCLOSE study

Liesbeth Rosseel 1,2, Nicholas J Montarello 3, Philippe Nuyens 4, Gabriella Tirado-Conte 5, Angelo Quagliana 6, Kristoff Cornelis 7, Vincent Floré 8, Michael Rosseel 9, Gintautas Bieliauskas 10, Lars Sondergaard 11, Ole De Backer 12,*
PMCID: PMC10949328  PMID: 37982158

Abstract

Background

Despite transcatheter aortic valve implantation (TAVI) having become a routine procedure, access site bleeding and vascular complications are still a concern which contribute to procedure-related morbidity and mortality.

Aims

The TAVI-MultiCLOSE study aimed to assess the safety and efficacy of a new vascular closure algorithm for percutaneous large-bore arterial access closure following transfemoral (TF)-TAVI.

Methods

All consecutive TF-TAVI cases in which the MultiCLOSE vascular closure algorithm was used were prospectively included in a multicentre, observational study. This stepwise algorithm entails the reinsertion of a 6-8 Fr sheath (primary access) following the initial preclosure with one or two suture-based vascular closure devices (VCDs). This provides the operator with the opportunity to perform a quick and easy angiographic control and tailor the final vascular closure with either an additional suture- or plug-based VCD, or neither of these.

Results

Among 630 patients who underwent TF-TAVI utilising the MultiCLOSE algorithm, complete arterial haemostasis was achieved in 616 patients (98%). VCD failure occurred in 14 patients (2%), treated with either balloon inflation (N=1), covered stent (N=12) or surgical repair (N=1). Overall, this vascular closure approach resulted in a minor and major vascular complication rate of 2.2% and 0.6%, respectively. At 30 days, only one new minor vascular complication (0.2%) was noted. In-hospital and 30-day all-cause mortality rates were 0.2% and 1.0%, respectively.

Conclusions

Use of the MultiCLOSE vascular closure algorithm was demonstrated to contribute to an easy, safe, efficacious and durable vascular closure after TF-TAVI, resulting in a major vascular complication rate of less than 1%.

Introduction

Transcatheter aortic valve implantation (TAVI) has become the first-choice treatment for elderly patients with severe, symptomatic aortic stenosis1,2,3,4,5,6. Over the past years, there has been a large focus on the simplification and optimisation of the TAVI procedure, along with early patient discharge. One of the key conditions to implement such a practice is having complete control of the large-bore arterial access closure after TAVI.

Despite TAVI having become a routine procedure, access site bleeding and vascular complications related to the primary access are still a concern, as these are associated with procedure-related morbidity and mortality7,8. Meticulous computed tomography (CT) preprocedural planning and ultrasound-guided vascular puncture have contributed to a decrease in vascular complications related to TAVI. In addition, the availability of dedicated vascular closure devices (VCDs) has enabled the successful percutaneous closure of large-bore arteriotomies9,10. However, different closure techniques are used worldwide, with different success and failure rates11,12,13,14,15.

In this prospective, multicentre, observational study, we aimed to investigate the safety and efficacy of a systematic, stepwise approach for percutaneous large-bore arterial access closure following TAVI - referred to as the MultiCLOSE algorithm - which integrates a cascade of possible safety nets.

Methods

STUDY DESIGN AND PATIENT POPULATION

The TAVI-MultiCLOSE study is a prospective, multicentre, observational study which was conducted in three TAVI centres: one in Denmark and two in Belgium. All consecutive patients that underwent standard transfemoral (TF)-TAVI between May 2022 and June 2023 (Denmark) and September 2022 and June 2023 (Belgium) were included in the study. The only TF-TAVI patients that were excluded from the study were patients who had undergone intravascular lithotripsy-assisted TF-TAVI with a contralateral angiographic vascular closure control (N=18 in this study period). The study was approved by the local ethics committees and complied with the Declaration of Helsinki.

MULTICLOSE VASCULAR CLOSURE ALGORITH

All three TAVI centres implemented the same MultiCLOSE vascular closure algorithm (Figure 1) in order to verify and obtain vascular closure following TF-TAVI. After ultrasound-guided arterial puncture of the primary access, a fluoroscopic image of the needle position at the arteriotomy site was stored. This stored fluoroscopic image is used as a reference image for the possible future deployment of a MANTA VCD (Teleflex) or positioning of a balloon or covered stent, if needed. Alternatively, one can premeasure the skin-to-arteriotomy distance at this stage by introducing the sheath of the Angio-Seal VCD (Terumo) (Figure 2). Next, preclosure with a single or double ProGlide or ProStyle (Abbott) was performed. After transcatheter aortic valve (TAV) deployment and removal of the delivery system, the large-bore introducer sheath was removed, keeping a 0.035” stiff guidewire in situ, while the suture knot(s) of the ProGlide/ProStyle were advanced a first time. Next, a short 6 Fr or 8 Fr sheath was introduced over the 0.035” stiff guidewire and the suture knot(s) were further tightened. The selection of a 6 Fr or 8 Fr sheath was left to the discretion of the TAVI operator and was guided by the degree of externally visible “oozing” or bleeding following initial vascular closure with the ProGlide/ProStyle. If no overt external bleeding was observed after introduction of the 6-8 Fr sheath, an angiography (approx. 5 ml contrast) through this short sheath was performed in order to detect residual extravasation beside the sheath or any other vascular complication. If no extravasation was observed (Moving image 1), either no additional VCD was used or a plug-based Angio-Seal (6 Fr or 8 Fr) was deployed after removal of the 6-8 Fr sheath, followed by a final tightening of the suture knot(s). In case of mild or moderate contrast extravasation at angiography (Moving image 2, Moving image 3), an Angio-Seal (mild) or additional ProGlide/ProStyle (moderate) was recommended, again followed by tightening of the suture knot(s) as the final step. The decision to use a 6 Fr or 8 Fr Angio-Seal was at the operator’s discretion and was guided by the Fr size of the short sheath utilised and/or degree of residual contrast extravasation beside the sheath.

Figure 1. MultiCLOSE algorithm for vascular access closure.

Figure 1

Novel and central to this vascular closure algorithm is the insertion of a 6 Fr or 8 Fr sheath following initial closure with one or two suture-based ProGlide/ProStyle(s). This provides the opportunity to perform a quick and easy angiographic control without the need to crossover from the contralateral femoral side or advance a catheter via the radial access, thereby saving time, contrast dye and radiation. A detailed description of this vascular closure algorithm can be read in the Methods section. TAVI: transcatheter aortic valve implantation

Figure 2. Use of the MANTA vascular closure device in the MultiCLOSE algorithm.

Figure 2

Technique A: This methodology should be used in case the skin-to-artery distance is unknown. Step 1: review the fluoroscopically stored image of the needle puncture site in relation to the femur head at the start of the TAVI procedure. Step 2: insert the MANTA sheath over the stiff guidewire and partially retract the dilator to permit fluoroscopic visualisation of the distal edge of the MANTA sheath. Position the distal edge of the MANTA sheath 2 cm cranial to the puncture site. Note the MANTA deployment depth in centimetres at the skin level. Fully reinsert the dilator and advance the MANTA sheath. Step 3: remove the dilator, introduce the collagen plug, retract the MANTA device to the deployment depth determined in Step 2, expose the anchor by rotating the deployment lever, and deploy the MANTA device as usual. Technique B: This methodology can be used when the operator is almost certain of using an Angio-Seal device for final vascular closure. Using the letters on the Angio-Seal sheath, the operator can measure the skin-to-artery distance. The MANTA deployment depth is between +1.5 and +2.0 cm in comparison to the measured skin-to-artery distance. TAVI: transcatheter aortic valve implantation

In case of overt major bleeding or major contrast extravasation following tightening of the ProGlide/ProStyle(s) (Moving image 4), the larger plug-based 18 Fr MANTA VCD could be instantly introduced on the stiff guidewire and deployed. In order to expose the anchor of the MANTA VCD at the correct depth, the operator can use one or both of the proposed methods in Figure 2. They could use either the initially stored fluoroscopic image of the needle’s arteriotomy site position or the depth measurement of the skin-to-arteriotomy distance, by means of the Angio-Seal sheath, in order to deploy the MANTA VCD at the correct depth.

Postprocedural protamine administration was left to the discretion of the operator, and a control vascular ultrasound was only performed when clinically indicated.

DATA COLLECTION

Baseline patient and procedural characteristics were prospectively collected in a dedicated electronic case report form. Procedural data encompassed the type and size of TAV and introducer sheath and the site of the primary and secondary access for TAVI. For the purpose of this study, the type and number of VCDs utilised were meticulously documented, including the Perclose ProGlide/ProStyle system, the Angio-Seal and MANTA VCDs. In addition, bailout strategies and any material (balloon, covered stent, surgical repair) used in case of unsatisfactory vascular closure were also documented.

ENDPOINTS AND DEFINITION

The primary endpoint was defined as in-hospital major vascular complications. Secondary endpoints included in-hospital minor vascular complications, 30-day minor and major vascular complications, VCD failure, and the need for endovascular (balloon and/or covered stent) or surgical intervention. All outcomes were defined according to the Valve Academic Research Consortium (VARC)-3 criteria16.

STATISTICAL ANALYSIS

Descriptive statistics were used to present baseline variables and primary and secondary endpoints. Continuous variables are reported as mean and standard deviation (±SD) in normal number distribution and as median and interquartile range (IQR) for skewed number distribution. Categorical variables are presented as counts and percentages.

Results

STUDY POPULATION

During the study period, a total of 630 patients underwent standard TF-TAVI: 489 patients were included from one high-volume Danish centre and another 70 and 71 patients were included from two Belgian centres. Baseline patient and procedural characteristics are summarised in Table 1. Of the total study cohort, 553 (88%) had TAVI performed via a right-sided transfemoral approach, and the radial artery was used as secondary access in 72 cases (12%). The relatively low use of the radial artery for introduction of the pigtail catheter was mainly due to the frequent use of a cerebral embolic protection device in this patient cohort.

Table 1. Baseline characteristics.

N=630
Patient characteristics
Age, years 79±7
Female 263 (42)
Arterial hypertension 462 (73)
Diabetes mellitus 132 (21)
Body mass index, kg/m2 26 [23.5-29.5]
Atrial fibrillation 222 (35)
Previous stroke 74 (12)
Coronary artery disease 255 (40)
Peripheral arterial disease 56 (9)
Chronic renal failure§ 19 (3)
Oral anticoagulation 213 (34)
EuroSCORE II, % 2.2 [1.3-3.7]
Procedural characteristics
Transcatheter aortic valve
Navitor (Abbott) 239 (38)
Evolut R/PRO(+) (Medtronic) 215 (34)

ACURATE neo2 (Boston Scientific)
145 (23)

SAPIEN 3/Ultra (Edwards Lifesciences)
31 (5)
Introducer sheath type/size
14 Fr expandable sheath 326 (52)
14 Fr sheath+exchanged for integrated sheath 192 (30)
16 Fr expandable sheath 13 (2)
16 Fr sheath+exchanged for integrated sheath 35 (6)
18 Fr sheath±exchanged for integrated sheath 44 (7)
20 Fr sheath±exchanged for integrated sheath 20 (3)
Secondary access
Contralateral femoral artery 558 (89)
Radial artery 72 (11)
Data are presented as mean±standard deviation, n (%) or median [interquartile range]. § estimated glomerular filtration rate <30 mL/min/1.73 m2. EuroSCORE: European System for Cardiac Operative Risk Evaluation

VASCULAR CLOSURE MANAGEMENT OF TAVI PRIMARY ACCESS SITE

In all 630 study patients, ultrasound-guided arterial puncture and preclosure with one (N=152; 24%) or two (N=478; 76%) suture-based ProGlide/ProStyle(s) was undertaken. Following removal of the large-bore introducer sheath, a 6 Fr or 8 Fr sheath was inserted in 623 patients (99%) in order to perform a control angiography. In 7 patients (1%), the ProGlide/ProStyle system failed resulting in overt major bleeding, for which the operator immediately switched to using a MANTA VCD. Following control angiography, no additional VCD, an additional Angio-Seal, an additional ProGlide/ProStyle, or MANTA VCD were used in 23 (4%), 585 (93%), 9 (1%) and 6 (1%) patients, respectively (Figure 3).

Figure 3. TAVI primary access site vascular closure management in the TAVI-MultiCLOSE study.

Figure 3

The vascular closure strategies and outcomes in the TAVI-MultiCLOSE study, including 630 TAVI patients treated by a percutaneous transfemoral approach and resulting in a major primary access site-related vascular complication rate of 0.6%. TAVI: transcatheter aortic valve implantation; VCD: vascular closure device

In 14 patients (2.2%), there was incomplete vascular closure following the MultiCLOSE algorithm. An endovascular bailout was successful in 13 of these patients (balloon: N=1; covered stent: N=12). In one patient, surgical vascular repair was needed; this was due to unsatisfactory vascular closure with 2 ProGlides and 1 Angio-Seal (Figure 3). In nearly all patients (N=628), protamine was administered at the end of the procedure.

PRIMARY AND SECONDARY ENDPOINTS

The primary endpoint occurred in 4 patients (0.6%); these patients had a VARC-3-defined in-hospital major vascular complication (Table 2). Major vascular complications occurred because of VCD failure leading to VARC type ≥2 bleeding, and final haemostasis was achieved by means of prolonged balloon inflation (N=1), covered stent placement (N=2) or surgical repair (N=1) (Supplementary Table 1).

Table 2. Clinical outcomes according to VARC-3 criteria.

N=630
Overall vascular complications
Minor vascular complication 14 (2.2)
Major vascular complication 4 (0.6)
Primary access site
Minor vascular complication 13 (2.1)
Major vascular complication 4 (0.6)
Secondary access site
Minor vascular complication 1 (0.2)
Major vascular complication 0
Access-related non-vascular complication
Minor vascular complication 0
Major vascular complication 0
New-onset vascular complication – discharge to 30 days
Minor vascular complication 1 (0.2)
Major vascular complication 0
Data are presented as n (%). VARC: Valve Academic Research Consortium

An additional 14 patients (2.2%) had minor vascular complications: 13 complications were at the primary access site and one pseudoaneurysm was at the secondary femoral access site (Supplementary Table 1). Only one patient presented with a new minor vascular complication at 30-day follow-up.

In-hospital and 30-day all-cause mortality rates were 0.2% (N=1) and 1.0% (N=6), respectively. In-hospital mortality occurred as a result of coronary obstruction during a valve-in-valve procedure. None of the mortalities were attributed to an access site-related complication.

Discussion

In this prospective, multicentre study, an easy-to-implement algorithm for percutaneous closure of large-bore arterial access after TAVI was investigated. The MultiCLOSE algorithm consists of preclosure with 1 or 2 suture-based VCDs and angiographic control via the primary access site with a 6-8 Fr sheath to determine a tailored final vascular closure strategy. Final vascular closure can be obtained by the use of one additional suture- and/or plug-based VCD (Central illustration) or neither. Implementation of the MultiCLOSE algorithm was demonstrated to contribute to a safe and efficacious vascular closure after TAVI, resulting in a major vascular complication rate of less than 1%.

Central illustration. MultiCLOSE algorithm for closure of large-bore transfemoral access in TAVI.

Central illustration

The findings in the TAVI-MultiCLOSE study demonstrate the safety and efficacy of a new and easy-to-implement algorithm for the percutaneous closure of large-bore arterial access after TAVI, offering the TAVI operator a multitude of safety nets, resulting in a low vascular complication rate. TAVI: transcatheter aortic valve implantation; VCD: vascular closure device

Due to the decreasing insertion profiles of TAV delivery systems, meticulous preprocedural CT-based planning of the TAVI vascular access and ultrasound-guided arterial puncture, TAVI-related vascular complication rates have markedly decreased in the last decade17,18. A fully percutaneous TAVI approach has become the first-choice strategy due to superior results as compared to surgical approaches19,20. At the same time, patients with extensive calcific iliofemoral disease are nowadays also often treated by percutaneous TF-TAVI, as calcium modification techniques such as intravascular lithotripsy or orbital atherectomy have become available21,22. In any scenario, it is important to have a well-considered vascular access and closure strategy with adequate bailout options in order to limit vascular complications, which are associated with significant morbidity and mortality.

The MultiCLOSE vascular closure algorithm was quickly and easily adopted in the 3 centres treating a large, nearly all-comers TF-TAVI cohort and demonstrated high efficacy with low vascular complication rates (minor 2.2%, major 0.6%). This is comparable to and even better than the findings reported in the low-risk TAVI landmark trials5,6. In the PARTNER 3 Trial, the VARC-2 defined minor and major vascular complication rates were reported to be 4.2% and 2.0%, respectively; whereas in the Evolut Low Risk trial, the major vascular complication rate was 3.8% (minor vascular complications were not reported) - importantly, these outcomes were obtained in highly selected low-risk TAVI populations5,6. With the implementation of the MultiCLOSE algorithm, the goal was to reduce the incidence of major vascular complications to less than 1% in an all-comers TF-TAVI cohort. This goal was achieved in our study cohort.

The novelty and high efficacy of the MultiCLOSE algorithm is not due to the use of a novel (and often more expensive) VCD, but is due, rather, to the design of a stepwise and versatile algorithm for the percutaneous closure of the primary TAVI access, which can be tailored to the ad hoc requirements of each situation and which integrates a cascade of possible safety nets. The MultiCLOSE algorithm has several important components.

Firstly, a one-frame fluoroscopic image of the needle puncture site is stored immediately after ultrasound-guided arterial puncture. This fluoroscopic image, showing the exact location of the arteriotomy site in relation to the femur head, can be used as a reference image for the possible future deployment of a MANTA device (Figure 2) or for the positioning of a balloon or covered stent, if needed.

Next, a short 6-8 Fr sheath is reinserted over a stiff guidewire following removal of the large-bore sheath and advancement of the initially deployed ProGlide/ProStyle(s). This particular approach confers several advantages. 1) It allows a safe and easy ipsilateral angiographic control through the 6-8 Fr sheath. In comparison to a more traditional approach with control angiography from a contralateral femoral or radial secondary access, this strategy is easier and helps to save time, contrast and radiation exposure. 2) The operator has better tactile feedback when advancing and pushing the ProGlide/Prostyle knot(s) towards the arterial vessel wall thereby touching the small-calibre sheath; this prevents unintentional vascular damage and is particularly useful in the challenging subset of obese patients. 3) Based on the angiography, the MultiCLOSE algorithm enables a tailored final vascular closure strategy depending on the degree of contrast extravasation beside the 6-8 Fr sheath. In a majority of cases (93%), an additional Angio-Seal was used in order to achieve instant, complete haemostasis. The efficacy of such a combined suture- and plug-based vascular closure after TAVI has been shown previously by Costa et al (2021), who reported minor and major vascular complication rates of 2.4% and 1.6%, respectively, in a propensity score-matched population14. However, instead of an upfront combined strategy of both a suture- and plug-based VCD, the MultiCLOSE algorithm allows for a more versatile and situation-tailored approach in which either no additional VCD (in case of “pinching” or stenosis of the artery) or an additional Angio-Seal, ProGlide/ProStyle, or MANTA VCD can be used, thus adjusting to the degree of residual contrast extravasation at control angiography. In case of a challenging vascular closure, the MultiCLOSE algorithm also offers a cascade of safety nets: the operator can first opt to deploy an additional ProGlide/ProStyle in case of moderate contrast extravasation, followed by a new ipsilateral control angiography and final vascular closure with a plug-based VCD. 4) Finally, by reinserting a 6-8 Fr sheath, the stiff guidewire can be kept in place without interference with the sutures. This guarantees vascular access in case of complete VCD failure and the subsequent need for MANTA or reinsertion of a large-bore sheath.

Using the MultiCLOSE algorithm, successful vascular closure with complete haemostasis was achieved in 616 patients (98%). In the remaining 14 patients (2%), there was need for an additional bailout strategy due to ongoing bleeding, which mostly consisted of placement of a covered stent (N=12) at the level of the TAVI primary access site. Only one patient required surgical vascular repair; vascular access via the primary access site was not possible, and the lesion could not be crossed via a secondary contra- or ipsilateral approach.

Taken together, we can conclude that the MultiCLOSE vascular closure algorithm is easy to implement and that it contributes to a safe and efficacious percutaneous vascular closure after TAVI, resulting in a low major vascular complication rate (<1%). The use of percutaneous bailout strategies is relatively rare; however, it is strongly recommended to be prepared for this eventuality by measuring the arterial puncture site’s dimensions at the preprocedural CT analysis and having the correctly sized balloons and stents available in the intervention room. It is of critical importance to have zero tolerance for residual bleeding or “oozing” at completion of the procedure, especially when seeking to adopt an early discharge policy after TAVI. In this study, we demonstrate that the implementation of the MultiCLOSE vascular closure algorithm helps to obtain instant, reliable and durable complete haemostasis following percutaneous TF-TAVI.

Finally, this study does not provide an answer to the question of whether vascular preclosure is more efficient when using one or two ProGlide/ProStyle(s). However, we did observe an increased use of only one preclosure device over the time period of this study. The decision to use one or two ProGlide/ProStyle(s) was left to the discretion of the operator and is a rather subjective decision. In patients with a straightforward vascular access (non-obese, single puncture, ≤mild artery calcifications) and good tactile feedback during deployment of a first ProGlide/ProStyle, most operators nowadays will not use a second preclosure device. In Supplementary Figure 1, we demonstrated similar vascular complication rates when using single versus double ProGlide/ProStyle preclosure. However, patient selection bias cannot be excluded, and therefore, future studies will be needed to shed light on this topic.

Limitations

This was a single-arm, observational study conducted in three hospitals; to determine the true clinical benefit of the MultiCLOSE algorithm for the reduction of TAVI-related vascular complications, a randomised controlled study is needed. Patients with severe peripheral arterial disease undergoing TF-TAVI were excluded from this study, as these patients were treated by intravascular lithotripsy-assisted TF-TAVI with use of a contra- or ipsilateral safety wire. However, this study did include all other consecutively treated TF-TAVI patients (95-100% of all-comers) in whom this MultiCLOSE vascular closure algorithm was used, and there were no missing data at follow-up.

Conclusions

The findings in this real-world TAVI-MultiCLOSE study demonstrate the safety and efficacy of a new and easy-to-implement algorithm for percutaneous closure of large-bore arterial access after TF-TAVI, offering the TAVI operator multiple levels of safety nets, resulting in a major vascular complication rate of less than 1%.

Impact on daily practice

Vascular complications after large-bore transfemoral (TF) access in transcatheter aortic valve implantation (TAVI) are associated with significant morbidity and mortality. The novel, stepwise MultiCLOSE vascular closure algorithm involves the reinsertion of a 6-8 Fr sheath following an initial closure with one or two suture-based vascular closure devices (VCDs). This provides the opportunity to perform a quick and easy angiographic control and tailor the final vascular closure with either an additional suture- or plug-based VCD, or neither of these. Use of the MultiCLOSE algorithm was demonstrated to contribute to an easy, safe, efficacious and durable vascular closure after TF-TAVI with multiple levels of safety nets, resulting in a major vascular complication rate <1%.

Supplementary data

Supplementary Table 1

Vascular complications according to VARC-3 criteria.

Supplementary Figure 1

Outcomes with single versus double ProGlide/ProStyle preclosure.

Moving image 1

No contrast extravasation with direct right femoral angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (49.9KB, mp4)
Moving image 2

Mild contrast extravasation with direct left femoral angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (46KB, mp4)
Moving image 3

Moderate contrast extravasation and mild narrowing of the right common femoral artery with direct angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (56.1KB, mp4)
Moving image 4

Severe contrast extravasation and mild narrowing of the left common femoral artery with direct angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (32.8KB, mp4)

Acknowledgments

Conflict of interest statement

G. Bieliauskas, L. Sondergaard, and O. De Backer received institutional research grants and consulting fees from Abbott. G. Tirado-Conte holds a research-training contract with ‘Rio Hortega’ (CM21/00091) from the Spanish Ministry of Science and Innovation (Instituto de Salud Carlos III). The other authors have no conflicts of interest to declare.

Abbreviations

CT

computed tomography

TAV

transcatheter aortic valve

TAVI

transcatheter aortic valve implantation

TF

transfemoral

VCD

vascular closure device

Contributor Information

Liesbeth Rosseel, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark; Cardiology Department, Algemeen Stedelijk Ziekenhuis, Aalst, Belgium.

Nicholas J. Montarello, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Philippe Nuyens, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Gabriella Tirado-Conte, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Angelo Quagliana, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Kristoff Cornelis, Cardiology Department, Algemeen Ziekenhuis Maria Middelares, Ghent, Belgium.

Vincent Floré, Cardiology Department, Algemeen Ziekenhuis Maria Middelares, Ghent, Belgium.

Michael Rosseel, Cardiology Department, Algemeen Stedelijk Ziekenhuis, Aalst, Belgium.

Gintautas Bieliauskas, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Lars Sondergaard, Consultant Cardiologist, Abbott, Santa Clara, CA, USA.

Ole De Backer, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

References

  1. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Brown DL, Block PC, Guyton RA, Pichard AD, Bavaria JE, Herrmann HC, Douglas PS, Petersen JL, Akin JJ, Anderson WN, Wang D, Pocock S PARTNER Trial Investigators. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. 2010;363:1597–607. doi: 10.1056/NEJMoa1008232. [DOI] [PubMed] [Google Scholar]
  2. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR, Williams M, Dewey T, Kapadia S, Babaliaros V, Thourani VH, Corso P, Pichard AD, Bavaria JE, Herrmann HC, Akin JJ, Anderson WN, Wang D, Pocock SJ PARTNER Trial Investigators. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364:2187–98. doi: 10.1056/NEJMoa1103510. [DOI] [PubMed] [Google Scholar]
  3. Leon MB, Smith CR, Mack MJ, Makkar RR, Svensson LG, Kodali SK, Thourani VH, Tuzcu EM, Miller DC, Herrmann HC, Doshi D, Cohen DJ, Pichard AD, Kapadia S, Dewey T, Babaliaros V, Szeto WY, Williams MR, Kereiakes D, Zajarias A, Greason KL, Whisenant BK, Hodson RW, Moses JW, Trento A, Brown DL, Fearon WF, Pibarot P, Hahn RT, Jaber WA, Anderson WN, Alu MC, Webb JG PARTNER 2 Investigators. Transcatheter or Surgical Aortic-Valve Replacement in Intermediate-Risk Patients. N Engl J Med. 2016;374:1609–20. doi: 10.1056/NEJMoa1514616. [DOI] [PubMed] [Google Scholar]
  4. Reardon MJ, Van Mieghem, Popma JJ, Kleiman NS, Søndergaard L, Mumtaz M, Adams DH, Deeb GM, Maini B, Gada H, Chetcuti S, Gleason T, Heiser J, Lange R, Merhi W, Oh JK, Olsen PS, Piazza N, Williams M, Windecker S, Yakubov SJ, Grube E, Makkar R, Lee JS, Conte J, Vang E, Nguyen H, Chang Y, Mugglin AS, Serruys PW, Kappetein AP SURTAVI Investigators. Surgical or Transcatheter Aortic-Valve Replacement in Intermediate-Risk Patients. N Engl J Med. 2017;376:1321–31. doi: 10.1056/NEJMoa1700456. [DOI] [PubMed] [Google Scholar]
  5. Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, Kapadia SR, Malaisrie SC, Cohen DJ, Pibarot P, Leipsic J, Hahn RT, Blanke P, Williams MR, McCabe JM, Brown DL, Babaliaros V, Goldman S, Szeto WY, Genereux P, Pershad A, Pocock SJ, Alu MC, Webb JG, Smith CR PARTNER 3 Investigators. Transcatheter Aortic-Valve Replacement with a Balloon-Expandable Valve in Low-Risk Patients. N Engl J Med. 2019;380:1695–705. doi: 10.1056/NEJMoa1814052. [DOI] [PubMed] [Google Scholar]
  6. Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, Bajwa T, Heiser JC, Merhi W, Kleiman NS, Askew J, Sorajja P, Rovin J, Chetcuti SJ, Adams DH, Teirstein PS, Zorn GL, Forrest JK, Tchétché D, Resar J, Walton A, Piazza N, Ramlawi B, Robinson N, Petrossian G, Gleason TG, Oh JK, Boulware MJ, Qiao H, Mugglin AS, Reardon MJ Evolut Low Risk Trial Investigators. Transcatheter Aortic-Valve Replacement with a Self-Expanding Valve in Low-Risk Patients. N Engl J Med. 2019;380:1706–15. doi: 10.1056/NEJMoa1816885. [DOI] [PubMed] [Google Scholar]
  7. Raju S, Eisenberg N, Montbriand J, Cusimano RJ, Feindel C, Ouzounian M, Horlick E, Osten M, Tsang W, Roche-Nagle G. Vascular Complications and Procedures Following Transcatheter Aortic Valve Implantation. Eur J Vasc Endovasc Surg. 2019;58:437–44. doi: 10.1016/j.ejvs.2019.03.014. [DOI] [PubMed] [Google Scholar]
  8. Van Kesteren, van Mourik, Vendrik J, Wiegerinck EMA, Henriques JPS, Koch KT, Wykrzykowska JJ, de Winter, Piek JJ, van Lienden, Reekers JA, Vis MM, Planken RN, Baan J. Incidence, Predictors, and Impact of Vascular Complications After Transfemoral Transcatheter Aortic Valve Implantation With the SAPIEN 3 Prosthesis. Am J Cardiol. 2018;121:1231–8. doi: 10.1016/j.amjcard.2018.01.050. [DOI] [PubMed] [Google Scholar]
  9. Vincent F, Spillemaeker H, Kyheng M, Belin-Vincent C, Delhaye C, Piérache A, Denimal T, Verdier B, Debry N, Moussa M, Schurtz G, Porouchani S, Cosenza A, Juthier F, Pamart T, Richardson M, Coisne A, Hertault A, Sobocinski J, Modine T, Pontana F, Duhamel A, Labreuche J, Van Belle. Ultrasound Guidance to Reduce Vascular and Bleeding Complications of Percutaneous Transfemoral Transcatheter Aortic Valve Replacement: A Propensity Score-Matched Comparison. J Am Heart Assoc. 2020;9:e014916. doi: 10.1161/JAHA.119.014916. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Noori VJ, Eldrup-Jørgensen J. A systematic review of vascular closure devices for femoral artery puncture sites. J Vasc Surg. 2018;68:887–99. doi: 10.1016/j.jvs.2018.05.019. [DOI] [PubMed] [Google Scholar]
  11. Barbanti M, Capranzano P, Ohno Y, Gulino S, Sgroi C, Immè S, Tamburino C, Cannata S, Patanè M, Di Stefano, Todaro D, Di Simone, Deste W, Gargiulo G, Capodanno D, Grasso C, Tamburino C. Comparison of suture-based vascular closure devices in transfemoral transcatheter aortic valve implantation. EuroIntervention. 2015;11:690–7. doi: 10.4244/EIJV11I6A137. [DOI] [PubMed] [Google Scholar]
  12. Sorropago G, Singh G, Sorropago A, Sole A, Rossi J, Tolva VS, Stabile E, Scalise F. A new Percutaneous technique for effective vascular Access Site closure in patients undergoing Transfemoral aortic valve implantation and thoraco-abdominal aortic aneurysm rEpair: the PASTE study. EuroIntervention. 2018;14:e1278–85. doi: 10.4244/EIJ-D-18-00245. [DOI] [PubMed] [Google Scholar]
  13. Gheorghe L, Brouwer J, Mathijssen H, Nijenhuis VJ, Rensing BJ, Swaans MJ, Chan Pin, Heijmen RH, De Kroon, Sonker U, Van der, Ten Berg. Early Outcomes After Percutaneous Closure of Access Site in Transfemoral Transcatheter Valve Implantation Using the Novel Vascular Closure Device Collagen Plug-Based MANTA. Am J Cardiol. 2019;124:1265–71. doi: 10.1016/j.amjcard.2019.07.030. [DOI] [PubMed] [Google Scholar]
  14. Costa G, Valvo R, Picci A, Criscione E, Reddavid C, Motta S, Strazzieri O, Deste W, Giuffrida A, Garretto V, Cannizzaro MT, Inserra C, Veroux P, Giaquinta A, Sgroi C, Tamburino C, Barbanti M. An upfront combined strategy for endovascular haemostasis in transfemoral transcatheter aortic valve implantation. EuroIntervention. 2021;17:728–35. doi: 10.4244/EIJ-D-20-01125. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Gmeiner JMD, Linnemann M, Steffen J, Scherer C, Orban M, Theiss H, Mehilli J, Sadoni S, Peterß S, Joskowiak D, Hagl C, Tsilimparis N, Curta A, Maurus S, Doldi PM, Löw K, Haum M, Roden D, Hausleiter J, Massberg S, Rizas K, Deseive S, Braun D. Dual ProGlide versus ProGlide and FemoSeal for vascular access haemostasis after transcatheter aortic valve implantation. EuroIntervention. 2022;18:812–9. doi: 10.4244/EIJ-D-22-00311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. VARC-3 WRITING Généreux P, Piazza N, Alu MC, Nazif T, Hahn RT, Pibarot P, Bax JJ, Leipsic JA, Blanke P, Blackstone EH, Finn MT, Kapadia S, Linke A, Mack MJ, Makkar R, Mehran R, Popma JJ, Reardon M, Rodes-Cabau J, Van Mieghem NM, Webb JG, Cohen DJ, Leon MB. Valve Academic Research Consortium 3: updated endpoint definitions for aortic valve clinical research. Eur Heart J. 2021;42:1825–57. doi: 10.1093/eurheartj/ehaa799. [DOI] [PubMed] [Google Scholar]
  17. Sherwood MW, Xiang K, Matsouaka R, Li Z, Vemulapalli S, Vora AN, Fanaroff A, Harrison JK, Thourani VH, Holmes D, Kirtane A, Pineda AM, Peterson ED, Rao SV. Incidence, Temporal Trends, and Associated Outcomes of Vascular and Bleeding Complications in Patients Undergoing Transfemoral Transcatheter Aortic Valve Replacement: Insights From the Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapies Registry. Circ Cardiovasc Interv. 2020;13:e008227. doi: 10.1161/CIRCINTERVENTIONS.119.008227. [DOI] [PubMed] [Google Scholar]
  18. Winter MP, Bartko P, Hofer F, Zbiral M, Burger A, Ghanim B, Kastner J, Lang IM, Mascherbauer J, Hengstenberg C, Goliasch G. Evolution of outcome and complications in TAVR: a meta-analysis of observational and randomized studies. Sci Rep. 2020;10:15568. doi: 10.1038/s41598-020-72453-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Costa G, Bieliauskas G, Fukutomi M, Ihlemann N, Søndergaard L, De Backer O. Feasibility and safety of a fully percutaneous transcatheter aortic valve replacement program. Catheter Cardiovasc Interv. 2021;97:E418–24. doi: 10.1002/ccd.29117. [DOI] [PubMed] [Google Scholar]
  20. Drafts BC, Choi CH, Sangal K, Cammarata MW, Applegate RJ, Gandhi SK, Kincaid EH, Kon N, Zhao DX. Comparison of outcomes with surgical cut-down versus percutaneous transfemoral transcatheter aortic valve replacement: TAVR transfemoral access comparisons between surgical cut-down and percutaneous approach. Catheter Cardiovasc Interv. 2018;91:1354–62. doi: 10.1002/ccd.27377. [DOI] [PubMed] [Google Scholar]
  21. Sawaya FJ, Bajoras V, Vanahaverbeke M, Wang C, Bieliauskas G, Søndergaard L, De Backer. Intravascular Lithotripsy-Assisted Transfemoral TAVI: The Copenhagen Experience and Literature Review. Front Cardiovasc Med. 2021;8:739750. doi: 10.3389/fcvm.2021.739750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Quagliana A, Montarello NJ, Vanhaverbeke M, Willemen Y, Campens L, Søndergaard L, De Backer. Orbital atherectomy to facilitate transfemoral transcatheter aortic valve implantation in patients with calcified iliofemoral arteries: a case series. Eur Heart J Case Rep. 2023;7:ytad310. doi: 10.1093/ehjcr/ytad310. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Table 1

Vascular complications according to VARC-3 criteria.

Supplementary Figure 1

Outcomes with single versus double ProGlide/ProStyle preclosure.

Moving image 1

No contrast extravasation with direct right femoral angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (49.9KB, mp4)
Moving image 2

Mild contrast extravasation with direct left femoral angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (46KB, mp4)
Moving image 3

Moderate contrast extravasation and mild narrowing of the right common femoral artery with direct angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (56.1KB, mp4)
Moving image 4

Severe contrast extravasation and mild narrowing of the left common femoral artery with direct angiography through a 6 Fr sheath over a 0.035” stiff guidewire after removal of the large-bore introducer sheath.

Download video file (32.8KB, mp4)