Abstract
Optical mapping is a widely used tool to record and visualize the electrophysiological properties in a variety of myocardial preparations such as Langendorff‐perfused isolated hearts, coronary‐perfused wedge preparations, and cell culture monolayers. Motion artifact originating from the mechanical contraction of the myocardium creates a significant challenge to performing optical mapping of contracting hearts. Hence, to minimize the motion artifact, cardiac optical mapping studies are mostly performed on non‐contracting hearts, where the mechanical contraction is removed using pharmacological excitation–contraction uncouplers. However, such experimental preparations eliminate the possibility of electromechanical interaction, and effects such as mechano‐electric feedback cannot be studied. Recent developments in computer vision algorithms and ratiometric techniques have opened the possibility of performing optical mapping studies on isolated contracting hearts. In this review, we discuss the existing techniques and challenges of optical mapping of contracting hearts.
Keywords: action potential duration, motion artifacts, motion tracking, optical mapping, ratiometry, ventricular fibrillation
Abstract figure legend Combination of excitation ratiometry and motion tracking technique minimizes motion artifacts from optical action potentials (OAPs) in contracting hearts. The isolated contracting heart stained with voltage‐sensitive dye is excited with blue and green excitation wavelengths that are rapidly switched in time with each camera frame. The action potential modulated fluorescence emission is then captured on a single camera, separating the frames into odd and even generated optical data corresponding to blue and green excitation or vice versa. Performing motion tracking and computing the ratio between motion‐tracked videos significantly reduced the motion artifacts in OAPs as compared to raw signals. Created with biorender.com and published with permission.
Introduction
Cardiac optical mapping has been extensively used to study and characterize the electrophysiological properties of the heart and to map cardiac arrhythmias (Efimov et al., 2004; Herron et al., 2012). The optical nature of the recording enables high spatial and temporal resolution recordings of electrophysiological data. Cardiac action potentials (APs) and calcium transients in cardiac tissue can be recorded by loading fast‐responding voltage‐ and calcium‐sensitive dyes to cardiac tissue, independently or simultaneously (Choi & Salama, 2000; Herron et al., 2012; Uzelac et al., 2022). The dyes exhibit fluorescence, which is the absorption of high energy photons (lower wavelength) and the emission of low energy photons (higher wavelength). For an example of voltage‐sensitive dyes, the dye binds to the cardiac cell membrane and undergoes a spectral shift in its excitation and emission (fluorescence) spectrum in response to changes in transmembrane voltage. When the emitted light is appropriately filtered using a bandpass filter, this leads to changes in detected fluorescence intensity during a cardiac action potential that track changes in transmembrane voltage.
Figure 1 illustrates the basic principle of cardiac optical mapping used to record optical action potentials. The cardiac tissue is loaded with a voltage‐sensitive dye and then excited with light of wavelengths that lie within its excitation spectrum. The fluorescence emission of the dye that is modulated by transmembrane voltage is captured on a high‐resolution photodetector after passing through an emission filter. The commonly used photodetectors are charge‐coupled device (CCD), complementary metal oxide‐semiconductor (CMOS), and electron multiplying charge coupled device (EMCCD) cameras that have high quantum efficiencies. The emission filter transmits fluorescent light within a certain spectrum of wavelengths. This spectral region generally has a relatively high value of voltage‐dependent fractional change in fluorescence (∆F/F), such that action potential‐dependent intensity changes can be easily differentiated from the baseline fluorescence (F). However, the ∆F/F value of currently available voltage‐sensitive dyes is very small (Matiukas et al., 2007), with a maximum of around 15%, and hence visualization of action potentials usually requires post‐processing techniques such as normalization or background subtraction.
Although multiple voltage‐sensitive dyes with different values of ∆F/F are already available in the market, it is important to note the adverse effects of these dyes in addition to their benefits. For example, the most commonly used ratiometric potentiometric dye, di‐4‐ANEPPS, was observed to have effects on heart rate (Fialova et al., 2010; Janoušek et al., 2015) in rat, guinea pig and rabbit, conduction velocity (Larsen et al., 2012) in guinea pig, and ischaemia (Ronzhina et al., 2021) in rabbit. However, intracoronary injection of the dye di‐4‐ANBDQBS (JPW‐6033), designed for blood‐perfusion optical mapping (Matiukas et al., 2007), showed no signs of cardiac toxicity during continuous monitoring of ECG from baseline to 30 min after dye/solvent injection in pig models. Moreover, heart rate was stable throughout the recording period (Lee et al., 2019).
Measurements based on optical mapping have made important contributions to our understanding and observations of many electrophysiological properties that exhibit spatio‐temporal complexity in the heart (Bernus et al., 2007; Cherry & Fenton, 2008; Efimov et al., 1999; Mitrea et al., 2009; Pertsov et al., 1993). Moreover, optical mapping systems are easily adapted to study a wide variety of heart sizes and species (Banville & Grey, 2002; Chattipakorn et al., 2001; Fenton et al., 2009; Handa et al., 2021; Lang et al., 2011; Nanthakumar et al., 2007). Further details about this conventional optical mapping are not within the scope of this paper but can be found in multiple excellent reviews (Attin & Clusin, 2009; Efimov et al., 2004; O'Shea et al., 2020).
Optical mapping of non‐contracting hearts
Until recently, optical mapping studies involving contracting hearts were rarely performed due to the motion artifacts originating from mechanical motion of the cardiac tissue with respect to the detector. The sensitivity of optical mapping studies to motion is very high, because of the small ∆F/F signal, such that the minimal contractile motion of the heart can cause severe motion artifacts within the optical signal (Christoph et al., 2017; Svrcek et al., 2009). These motion artifacts preclude any reliable analysis of the repolarization properties of the myocardium, and in most cases, also prevent any accurate mapping of the propagation of the action potential wavefront.
Attempts have been made to minimize the motion of the heart without completely suppressing the mechanics, such as applying a moderate level of pressure onto the heart using movable pistons or glass walls (Chen et al., 2000; Laurita et al., 1998; Mandapati et al., 1998). However, these measures alone were often ineffective to eliminate motion artifacts and may cause local ischaemia due to the reduction of coronary flow and myocardial perfusion. The most common method to reduce motion artifacts from optical action potentials (OAPs) is by administering pharmacological agents to the heart or using low calcium concentrations (Salama & Morad, 1976). The pharmacological excitation–contraction drugs effectively remove any mechanical contraction and uncouple the electrical and mechanical activity of the heart, thus allowing the investigation of the electrophysiological properties of the heart without motion artifact. Three of the most used drugs are 2,3‐butanedione monoxime (BDM), also known as diacetyl monoxime, cytochalasin D (Cyto‐D) and blebbistatin. BDM and Cyto‐D are two non‐specific uncouplers such that inhibition of cardiac contractility requires relatively high concentrations. Though the exact mechanism of action of BDM is still unknown, it is a non‐competitive inhibitor of myofibrillar ATPase (Fedorov et al., 2007; Sellin & McArdle, 1994). Cyto‐D reduces or removes contraction by disrupting actin polymerization (Rueckschloss & Isenberg, 2001), but it is a toxic drug that requires much caution in its use. Blebbistatin, on the other hand, is a specific myosin II inhibitor of actin–myosin interaction and thus abolishes cardiac contraction even at low concentrations (Fedorov et al., 2007; Swift et al., 2021).
Although the excitation–contraction uncouplers are designed to suppress contraction while preserving the electrical activity, BDM and Cyto‐D were associated with electrophysiological changes in multiple studies (Banville & Grey, 2002; Kettlewell et al., 2004; Lee et al., 2001; Pitruzzello et al., 2007; Riccio et al., 1999). Blebbistatin, instead, has been considered the electromechanical uncoupler of choice since it was introduced in 2003 (Straight et al., 2003) and is considered to have minimal effects on cardiac electrophysiology (Dou et al., 2007; Fedorov et al., 2007, 2009; Fenton et al., 2008; Hansen et al., 2018; Jou et al., 2010; Lou et al., 2012). Contrastingly, a few studies have recently reported effects of blebbistatin on cardiac electrophysiology (Brack et al., 2013; Kappadan et al., 2020; Lee et al., 2019). The reason behind this disparity could be differences in the metabolic demand and oxygen consumption of contracting hearts as compared to hearts administered with blebbistatin (Kuzmiak‐Glancy et al., 2015; Swift et al., 2021). Therefore, the observed differences may not be the direct effects of blebbistatin but may be an indirect consequence of the altered metabolic state of the heart when the contraction is inhibited using an excitation–contraction uncoupler. To date, no studies have shown clear effects of blebbistatin on ion channels responsible for cardiac repolarization. Furthermore, incorrect usage of blebbistatin may lead to blebbistatin precipitation and alterations of cardiac electrophysiology (Swift et al., 2012) and may account for some of the observed electrophysiological changes following blebbistatin administration (Swift et al., 2021).
Even though the excitation–contraction uncouplers allow recording of optical action potentials without motion artifacts, crucial information contained in cardiac contractility and coupled cardiac electromechanics is lost. Moreover, electromechanical coupling is bidirectional such that electrical excitation of the cardiac cell membrane causes mechanical contraction via excitation–contraction coupling (ECC) and mechanical stretch on the myocardium can cause electrical excitation through stretch activated channels via a process known as mechano‐electric feedback (Franz, 1996, 2000; Taggart, 1996; Zabel et al., 1996), and such effects cannot be studied in a non‐contracting heart.
Optical mapping in contracting hearts
Given the significant limitations of conducting optical mapping experiments in non‐contracting hearts paralysed by excitation–contraction uncouplers, there is currently significant interest in developing robust methods to perform cardiac optical mapping in contracting hearts. In this next section, we highlight the three factors that contribute to motion artifacts in the optical action potential and describe novel strategies that facilitate optical mapping of the contracting heart.
Factors contributing to motion artifacts
Optical mapping studies involving beating hearts are susceptible to motion artifacts, as shown in Fig. 2 Aa. Optical action potential recorded from contracting hearts are characterized by severe distortion of the baseline and repolarization phase (the fast depolarization phase of the action potential is often largely unaffected). Administration of blebbistatin accordingly eliminates such artifacts (Fig. 2 Ab). Optical action potentials from contracting hearts are not only affected by the contractile motion of the heart, but the artifact is also affected by the changes in homogeneity of illumination and dye loading. Inhomogeneous illumination and spatio‐temporal variation of dye concentration can diminish the quality of recorded OAPs while the heart contracts (Bachtel et al., 2011; Knisley et al., 2000), but importantly, also contribute to motion artifact. The artifact can be caused by differences in fluorescence intensity across different heart regions with inhomogeneous dye concentration and inhomogeneous illumination. As the heart contracts, a camera pixel that was previously recording signals from one region of the heart will now be recording signals from another region, which can have very different fluorescence intensity. These variations in fluorescence can be larger than the intensity change due to action potentials, thus distorting the OAP signal while the heart moves. For example, baseline fluorescence (during diastole) of calcium‐sensitive dyes is less as compared to voltage‐sensitive dyes, resulting in lesser motion artifacts in optical calcium transients in comparison with optical action potential. Temporal variation of dye concentration due to photobleaching will also cause motion artifacts by affecting the baseline fluorescence. As an example, the morphologies of OAPs and motion artifacts are completely different at two loci (Fig. 2 B, pixel a and pixel b) on the left ventricular surface of a contracting rabbit heart even though the pixel displacements due to the contractile motion are similar. This indicates the possibility of having additional factors contributing to the motion artifacts (Fig. 2 C). Furthermore, contraction of cardiomyocytes results in higher membrane density per pixel resulting in a higher signal amplitude and could potentially contribute to motion artifacts in the OAPs.
It is also important to note that motion artifacts observed in optical mapping studies with different voltage‐sensitive dyes can also be different as the baseline fluorescence (F) and voltage sensitive signal amplitude (∆F/F) can vary depending on the dye being used. For example, very slow dye washout and signal decay was reported for near infrared dyes di‐4‐ANBDQPQ and di‐4‐ANBDQBS in comparison with blue‐green excitation dye di‐4‐ANEPPS (Matiukas et al., 2007). Additionally, light–tissue interactions such as light scattering and absorption are different for different excitation wavelengths and hence may also contribute to motion artifacts.
Hence, the motion artifacts arising from the contracting hearts are due to a combination of all these factors and any robust motion artifact removal technique should address these issues carefully.
Techniques to reduce motion artifacts
Motion tracking
The loss of correspondence between the camera pixel and a specific area of myocardial tissue throughout a cardiac cycle, induced by contraction, is a serious challenge that causes signal distortion (Fig. 2). Adopting motion tracking through the different frames of videos has been shown to minimize motion artifacts in multiple studies. For example, Rohde et al. (2005) performed image registration on optical data obtained from rabbit hearts to correct motion artifacts due to global motions but not accounting for local motions. In addition, tracking the motion of the heart using ring‐shaped polyethylene markers attached to the epicardial surface has been used to perform simultaneous optical mapping of action potentials and wall motion in isolated perfused pig hearts (Bourgeois et al., 2011). Furthermore, dot‐shaped markers and additional geometry cameras were used to reconstruct the epicardial geometry in three dimensions (3D) (Zhang et al., 2016). One major limitation of performing motion tracking using markers is the restriction of spatial resolution of the recording, and interpolation of the data may be required between the markers. To overcome this issue, two‐dimensional (2D) marker‐free motion tracking that is capable of tracking motion and non‐rigid deformation, based on optical flow, has been employed and the efficacy of motion tracking has been demonstrated using synthetic and experimental data (Christoph & Luther, 2018), thus making it possible to perform optical mapping on freely contracting hearts (Christoph & Luther, 2018; Kappadan et al., 2020). Unlike other image registration techniques used for optical mapping data (Khwaounjoo et al., 2015; Rohde et al., 2005), this technique uses locally normalized, contrast enhanced videos for motion tracking. This is a pre‐processing step to avoid accidental tracking of electrical activation instead of contractile motion itself. Since motion tracking of optical data can provide information on electrical and mechanical deformations of the heart, it was utilized to study the relationship between the turbulent electrical activity and elasto‐mechanical patterns during ventricular fibrillation (Christoph et al., 2018). As an extension, 3D deforming of ventricular wall during sinus rhythm was reconstructed using 2D motion tracking and a multi‐camera system, thereby allowing panoramic optical mapping of beating hearts (Christoph et al., 2017).
In this section, below, we describe the basic principle of marker‐free 2D motion tracking (Fig. 3). A normal optical mapping recording consists of a series of frames. Each frame is an image with intensity values proportional to the recorded fluorescence intensity. Between the frames, the recorded intensity changes due to motion of the heart and the changes in the action potential, the latter accounting for a small fractional change (∆F/F = 3−10% for ANEP dyes). The motion tracking algorithm computes the geometrical transformation between the reference (R) and test image (T) (Fig. 3 A) by registering the test image onto the reference image, such that the difference between test and reference image is minimum. For this, an image frame is selected as the reference image and the geometrical transformation is computed between the reference and test images. For paced data, the reference image is selected as the frame just before the pacing stimulus is delivered, and for Ventricular Fibrillation (VF) data, the reference image is any arbitrary frame. Each test image must be registered with the reference image separately. An important characteristic of this motion‐tracking algorithm is that it does not require any markers attached to the heart to assist motion tracking but instead tracks the motion using intensity values between a pair of images (test and reference).
Application of motion tracking to contracting hearts can produce notable reduction in motion artifacts as indicated by the difference image (test minus reference) in Fig. 3 Aa. The difference image before motion tracking (R − T) is characterized by significant motion artifacts whereas substantial reduction of motion artifacts is visible in the difference image after motion tracking (R − Tt). The computed displacement vectors (Fig. 3 Ab) represent the direction of the contractile motion. Further details about motion tracking algorithms and their validation can be found in the work of Christoph & Luther (2018) as well as in Lebert et al. (2022). Figure 3 B shows the application of 2D motion tracking where the motion amplitude (pixel displacement) of the heart during fixed frequency pacing and ventricular fibrillation is displayed as motion amplitude maps. As expected, motion amplitude of the contracting heart (no blebbistatin) decreases with increase in blebbistatin concentration. For this heart, contraction was inhibited substantially at a blebbistatin concentration of 2.8 μM as indicated by a mean motion amplitude value close to zero (0.43 pixels) and in general a blebbistatin concentration <5 μM completely suppressed the contraction. When the contracting heart is in ventricular fibrillation (VF motion), the mean motion amplitude value is 1.02 pixels (∼0.2 mm), indicating that contraction amplitude can be very small during VF, even without blebbistatin, and thus optical mapping has been used without uncouplers in those cases (Valderrábano et al., 2001). In some species, such as alligators (Herndon et al., 2021; Jensen et al., 2018) and frogs, blebbistatin does not suppresses contraction, and in those cases, other drugs, low calcium Tyrode solution or motion tracking needs to be used, this last option shown in Fig. 3 C.
Even though marker‐free motion tracking enables performance of optical mapping experiments on isolated contracting hearts, it is important to verify the accuracy of the technique to avoid tracking error and data misinterpretation. The study by Christoph & Luther (2018) validated the robustness and accuracy of 2D motion tracking using experimental and synthetically generated optical mapping videos. They achieved considerable reduction (75–80%) in motion artifacts while comparing the tracked data with simulated ground‐truth data. Furthermore, validation of the technique was performed using microelectrode recording as shown in Fig. 3 D.
Excitation and emission ratiometry
Ratiometric optical mapping techniques (Bachtel et al., 2011; Knisley et al., 2000) are effective for minimizing motion artifact from contracting hearts. Unlike conventional optical mapping, ratiometric techniques record two fluorescence signals that exhibit different changes in fluorescence (∆F). Taking the ratio between these two signals leads to attenuation of artifacts that are in common. Moreover, this technique diminishes the impact of variations in dye concentrations and illumination artifacts from OAP recordings, which is not addressed by motion tracking algorithms alone (Fig. 2 C).
There are two principal ratiometry techniques: (i) excitation ratiometry, which uses dual excitation wavelengths to excite the voltage‐sensitive fluorescent dye and a single emission band to collect the fluorescent light, which is modulated by transmembrane voltage; and (ii) emission ratiometry, which uses a single excitation wavelength to excite the dye and dual emission bands to collect the fluorescence emission. Excitation ratiometry is relatively easy to implement and can remove common artifacts in OAPs originating from inhomogeneity in dye concentrations across the heart. On the other hand, emission ratiometry requires two detectors that are aligned to image the same field of view of the heart, making it more difficult to implement. However, emission ratiometry can be used to remove inhomogeneous dye loading and illumination artifacts as they are common in OAPs that are recorded on the two detectors.
Excitation ratiometry using pulsed LEDs was first used to record intracellular calcium concentration [Ca2+]i in cultured rat cardiomyocytes (Fukano et al., 2006, 2008). However, the earliest application of excitation ratiometry to map transmembrane potential in cardiac muscle was demonstrated by Bachtel et al. (2011). They utilized blue (450 ± 10 nm) and cyan (505 ± 15 nm) LEDs to excite Di‐4‐ANEPPS (1‐(3‐sulfonatopropyl)‐4‐[β‐[2‐(di‐n‐butylamino)‐6‐naphthyl] vinyl] pyridinium betaine) and a single bandpass emission filter (585 ± 20 nm) to collect the fluorescence emission from the Langendorff‐perfused swine hearts. Similarly, emission ratiometry has been used to reduce motion artifacts in multiple studies (Brandes et al., 1992; Hooks et al., 2001; Hortigon‐Vinagre et al., 2016; Knisley et al., 2000; Rohr & Kucera, 1998). For example, Knisley et al. (2000) used two photodetectors to minimize motion artifacts from isolated rabbit hearts loaded with di‐4‐ANEPPS and excited by a laser beam of wavelength 488 nm. They observed a reduction in photobleaching‐induced drifts and motion artifacts in optical signals while taking the ratio between intensities of two fluorescence emission bands (green and red). Moreover, Hortigon‐Vinagre et al. (2016) successfully applied this method to di‐4‐ANEPPS‐loaded human induced pluripotent stem cell‐derived cardiomyocytes (hiPSC‐CM). Excitation was stimulated with LEDs (470 ± 10 nm) and the fluorescence emission was captured using two photomultipliers at 510−560 nm and 590−650 nm.
Figure 4 shows the principle and application of excitation and emission ratiometry to remove motion artifacts due to the variations in illumination and dye loading. Fig. 4 A shows the schematic of excitation ratiometry by utilizing the ratiometric property of di‐4‐ANEPPS. The dye undergoes voltage‐dependent shifts in its excitation and emission spectrum. Excitation of the dye with blue and green wavelengths produces action potential‐dependent fluorescent changes (∆F) in opposite directions. Thus, the artifacts that are common in the fluorescence emission of blue and green channels cancel out when taking the ratio between corresponding fluorescence intensities. Example time traces of optical action potentials as depicted in Fig. 4 C show the reduction in dye loading and illumination artifacts (see limitations of optical mapping in contracting hearts). Figure 4 B depicts the principle of emission ratiometry, where voltage‐dependent fluorescence emission from two emission bands is recorded using two cameras for a single excitation light. The ratio of the recorded intensities from these cameras can reduce motion artifacts due to dye loading and illumination differences. Calculating the ratio of fluorescence intensities for green and red emission bands minimizes motion artifacts (Fig. 4 D). It is to be noted that the observed reduction in motion artifacts in Fig. 4 C and Fig. 4 D are with use of ratiometry technique alone (without motion tracking), and the motion artifacts are not removed completely. This is because ratiometry only corrects for changes in base fluorescence between adjacent moving regions but cannot correct for the interference of action potentials measured between the regions as they move. Therefore, this technique only works if there is little dispersion (i.e. change in AP shape and duration) in the tissue. However, if large dispersion occurs in small distances, such as during discordant alternans obtained at fast pacing, ratiometry alone will give incorrect AP information. This is even more important for calcium measurements as calcium is not spatially coupled, allowing for large variations at very small distances (Greene et al., 2022; Shiferaw et al., 2005).
Combination of motion tracking and ratiometry
Previous optical mapping studies in contracting hearts show that neither motion tracking nor ratiometry alone can remove the motion artifacts completely. To minimize the motion artifacts originating from multiple factors as shown in Fig. 2 C, motion tracking and emission/excitation ratiometry techniques were combined in a few studies. For example, Seo et al. (2010) used fiducial markers to track cardiac tissue deformation and emission ratiometry for action potential mapping. They studied the relationship between strain heterogeneity and initiation of arrhythmia in rabbit ventricular tissue preparations and whole hearts. Bourgeois et al. (2011) used excitation ratiometry for optical mapping and ring‐shaped markers for tracking epicardial motion in swine hearts, and epicardial strain was computed using the tracked position of the markers. As an extension to Bourgeois et al.’s method, Zhang et al. (2016) performed motion tracking using dot‐shaped fiducial markers attached to the epicardium and hence were able to enhance the spatial resolution by orders of magnitude. Excitation ratiometry was also used here to minimize motion artifacts further. Additionally, they tracked the motion in three dimensions using binocular imaging with additional video cameras and quantified epicardial deformation by computing finite strain fields. More details about electromechanical optical mapping is provided by Nesmith et al. (2020).
In a slightly different approach, we previously used a combination of marker‐free motion tracking and excitation ratiometry on isolated rabbit hearts (Kappadan et al., 2020) (Figs 5 and 6). In that study, we measured optical action potential duration (APD50 and APD70) from Langendorff‐perfused contracting hearts and systematically studied APD restitution and ventricular fibrillation dynamics in contracting hearts and compared the properties to that of blebbistatin‐uncoupled hearts.
Figure 5 shows the flowchart of our method to perform motion correction technique on Langendorff‐perfused contracting hearts. Motion correction involves ratiometry and motion tracking. More specifically, we show the step‐by‐step processes (Fig. 5 A–E) involved in the excitation ratiometry, marker‐free motion tracking, and the combination of these techniques. In short, we capture optical data for blue and green excitation wavelengths (250 frames each) and perform motion tracking separately from the videos corresponding to each excitation wavelength. The excitation light is switched between camera frames so that odd frames record fluorescent light excited with blue light and the even frames record fluorescent light excited with green excitation light (or vice versa). The ratio between the tracked data (tracked blue and tracked green) is used for further analysis.
The combination of motion tracking and ratiometry (motion correction) significantly reduced motion artifacts and enhanced the quality of OAPs as compared to either of the methods alone (Fig. 6 A). Furthermore, motion correction techniques can be applied on the entire epicardial surface and properties such as APDs can be measured from a contracting heart with minimal motion artifacts. Figure 6 B shows an example of motion correction applied on a contracting rabbit heart. Single pixel time traces from different locations of the contracting heart show severe distortions in optical action potentials due to motion artifacts (dashed lines), whereas motion correction substantially reduced motion artifacts (continuous lines). As depicted in Fig. 6 B, the morphology of action potential signals heavily varies spatially on the contracting heart, indicating spatial variations in motion artifacts. This implies that the factors contributing to motion artifacts (Fig. 2 C) are different at these locations. That is, either motion amplitude, or dye concentration or illumination is different at these locations or combinations of these factors are different.
The main benefit of adopting both motion tracking and ratiometry techniques (motion correction) is the elimination of distortions of the repolarization phase of the action potential, which consequently impacts measurements of APDs and cardiac restitution properties. APD80 maps, for example, in non‐corrected hearts show scattered values of APD across the myocardium, while APD maps on a motion‐corrected heart show minimal distortion of APD values across the myocardium (Fig. 7 B). By using either of the methods alone, residual motion artifacts are clearly visible in the APD maps. Thus, combination of motion tracking and ratiometry is highly recommended for studies involving the repolarization phase of the action potential. On the other hand, the rapid upstrokes of action potentials are minimally affected due to motion artifacts, as shown in the activation time maps (Fig. 7 A). Activation time for each pixel location is calculated as the time corresponding to the steepest segment of the action potential upstroke (dF/dt max), which can be assigned relatively accurately even in the context of motion artifact (Laughner et al., 2012).
The motion correction technique can also be applied to study the fibrillatory dynamics in contracting hearts. As the amplitude of contractile motion during VF is relatively small, the application of motion correction technique can be highly successful in minimizing motion artifacts. Hence, techniques such as phase analysis can be directly used to visualize VF activation patterns. For example, Christoph et al. (2018) used phase analysis to visualize and characterize VF in motion‐tracked pig and rabbit hearts ex vivo. Moreover, they computed mechanical phase singularity based on strain rate and showed co‐localization of electrical and mechanical phase singularities. We recently compared phase singularity dynamics during VF in isolated contracting rabbit hearts using (i) motion tracking and (ii) combination of motion tracking and ratiometry and showed good agreement between phase maps (Kappadan et al., 2020). This is due to the relatively small motion of the heart during VF. We also investigated the spatial distribution of dominant frequency during VF using Fourier transform‐based power spectra analysis, in contracting hearts.
It is also important to discuss the feasibility of ratiometric and marker‐free motion tracking techniques to map different animal species and cardiac chambers. The ratiometric optical mapping approach has been successfully used to minimize motion artifacts from OAPs of isolated rabbit (Dumas & Kinisley, 2005; Kappadan et al., 2020; Knisley et al., 2000) and pig (Bachtel et al., 2011; Bourgeois et al., 2011; Lee et al., 2019) whole hearts. Ratiometry was also employed to record action potentials from neonatal rat ventricular myocytes (Chowdhury et al., 2018) and human induced pluripotent stem cell‐derived cardiomyocytes (hiPSC‐CM) (Hortigon‐Vinagre et al., 2016). Moreover, ratiometry utilizing the calcium sensitive dye Indo‐1 was used to suppress motion artifacts from intracellular Ca2+ concentrations in isolated rat hearts. The marker‐free motion tracking has been already used in isolated rabbit (Christoph & Luther, 2018; Christoph et al., 2018, 2017) and pig (Christoph et al., 2018) hearts and a combination of marker‐free motion tracking and ratiometry was applied to isolated rabbit hearts (Kappadan et al., 2020). The motion tracking in these studies was performed on the optical mapping videos of whole heart (atria and ventricles) with a focus on ventricles during pacing and fibrillation. Recently, GPU‐accelerated motion tracking has been successfully applied to hiPSC‐CM and isolated hearts of mouse, rabbit and pig (Lebert et al., 2022). However, to our knowledge, no studies have performed motion correction specific to highly trabeculated endocardium or complex atrial structure.
Limitations of optical mapping in the contracting heart
Next to the significant benefits, optical mapping with contracting hearts still possesses some limitations. Most importantly, the oxygen‐carrying capacity of commonly used crystalloid perfusates for ex vivo heart preparations is lower than whole blood (Chen et al., 1987; Gauduel et al., 1985). The metabolic demand and oxygen consumption of the contracting hearts are higher than their electromechanically uncoupled counterparts (Kuzmiak‐Glancy et al., 2015). As a result, the contracting hearts are at an increased risk of myocardial hypoxaemia. However, this is not only a limitation of optical mapping but also of the isolated contracting heart experiments in general. Nevertheless, myocardial oxygenation is a key factor in optical mapping of the contracting hearts, and it should be considered carefully while performing experiments that may cause inadequacy of oxygen supply, such as during high frequency electrical stimulation and recording of the cardiac restitution curve. In such cases, electrophysiological recordings from contracting hearts are also compromised and this can be treated by mechanical arrest using blebbistatin.
Other limitations are mostly related to motion tracking and ratiometry. Despite adopting 2D (i.e. in‐plane up‐and‐down and side‐to‐side) motion tracking to reduce the impact of motion artifacts, 2D motion tracking cannot account for out‐of‐plane (i.e. back‐and‐forth) motions of higher amplitudes, and therefore the capacity of motion tracking algorithms is limited.
As excitation ratiometry involves two different excitation wavelengths that lie within the excitation spectrum of the dye (for example, blue and green), the light intensity variations across the heart tissue with these wavelengths may not be the same. Hence, illumination‐related motion artifacts are not common in the recorded OAPs and ratiometry will not remove illumination artifacts completely. Therefore, special care is required during the experiment to arrange these excitation light sources to achieve (near) homogeneous intensity across the tissue in the ratio image. The light sources have to be adjusted while looking at the ratio image and the recording can be started once the ratio image shows homogeneous intensity values. Another shortcoming of excitation ratiometry is the loss of temporal resolution of the recording by a factor of 2. Since the recorded camera frames are split into two, corresponding to the blue and the green excitation wavelengths, the temporal resolution of the action potential signals with the excitation ratiometry is half of the camera resolution. For example, 500 fps camera resolution can only produce 250 fps of ratiometric video data.
Due to the aforementioned limitations of motion tracking and ratiometry, the combined motion correction technique is unable to eliminate motion artifacts completely. Accurate measurements of APD90 across all pixels on the heart are still difficult to achieve and it is likely that motion‐compensated ratiometric OAPs yet contain residual motion artifacts. However, incorporation of motion tracking in 3D can potentially help to reduce the residual artifacts. Moreover, replacing excitation ratiometry with emission ratiometry will in principle solve the issues related to illumination and temporal resolution as it uses only one light source instead of two and hence can minimize motion artifacts further.
Basic requirements and recommendations for optical mapping of the contracting heart
Here we provide some basic recommendations for performing optical mapping experiments on contracting hearts. From our experience, combining motion tracking and ratiometry can minimize motion artifacts significantly and we recommend using combined techniques for accurate measurement of APD and restitution properties. However, ratiometry or motion tracking alone may be sufficient to compute activation time and activation maps as the action potential upstroke is less impacted by motion. For arrhythmia analysis, such as ventricular fibrillation, where there is reduced motion of the heart, motion tracking may be sufficient (for instance, see Figure 12 from Kappadan et al., 2020), but the combined method is recommended for higher accuracy.
Ratiometric optical mapping of cardiac action potential requires ratiometric voltage‐sensitive dyes. The most commonly used dye for ratiometric measurements is di‐4‐ANEPPS, but studies have been performed using other voltage‐sensitive dyes such as di‐4‐ANBDQBS, di‐4‐ANEQ(F)PTEA) and di‐8‐ANEPPS (Chowdhury et al., 2018; Lee et al., 2019). Excitation ratiometry with di‐4‐ANEPPS requires excitation light with blue and green (or cyan) wavelengths that produces opposite polarity of ∆F, and a single emission filter to collect the fluorescence emission corresponding to both the excitation light. For example, we used blue (460 ± 5 nm) and green (540 ± 12.5 nm) wavelengths for excitation and the fluorescence emission was captured using an EMCCD camera Evolve 128 (128 × 128 pixels, 16‐bit dynamic range), Photometrics, Arizona, USA equipped with a bandpass emission filter (590 ± 55 nm) (Kappadan, 2021; Kappadan et al., 2020). It is to be noted that for excitation ratiometry to work well, the excitation light intensities on the heart should be adjusted to achieve a near homogeneous illumination for both wavelengths. For emission ratiometry, we used four blue (470 ± 20 nm) excitation LEDs (Cairn Research, Faversham, UK) and the fluorescence emission is captured on two orthogonally placed CMOS cameras (128 × 80 pixels) using green (535 ± 35 nm, Chroma Technology), Vermont, USA and red (>620 nm, Chroma Technology) emission filters. The cameras are carefully aligned to look at the same field of view. Other possible combinations of optics for excitation and emission ratiometry are discussed in the Ratiometry section. 2D motion tracking of the cardiac tissue requires motion tracking (Matlab, The MathWorks, Natick, MA, USA) and visualization (C++) algorithms that can be performed using a single CPU with a computation time of approximately 1 min per video image. However, the computation time can be reduced significantly by using parallel computing and it was recently demonstrated that optical mapping with numerical motion tracking can be performed in real‐time (Lebert et al., 2022).
Summary
This review focuses on the existing techniques for performing optical mapping of isolated contracting hearts. The emergence of computer vision algorithms and numerical motion tracking techniques opens the possibility of performing optical mapping studies in contracting hearts. Motion tracking techniques are highly effective in reducing motion artifacts due to the relative motion between the heart and the camera, whereas ratiometry helps to minimize dye loading and illumination artifacts. Combining these methodologies has shown promise in improving the quality of optically recorded electrophysiological signals and properties, such as action potential duration and cardiac restitution properties. The novel techniques reviewed here can be used to overcome the limitations of conventional optical mapping with paralysed hearts, provide insight into the mechanisms underlying mechano‐electrical coupling and cardiac arrhythmia, as well as create experimental conditions involving mechanics to study the effect of myocardial ischaemia and energetics. However, it is important to note that isolated contracting heart experiments can also be compromised in certain situations owing to the higher oxygen and metabolic demands. In such conditions, excitation–contraction uncoupling using blebbistatin may provide more accurate results and can be preferred. However, on the contrary, there could be effects such as ischaemia underestimation while using blebbistatin due to lower metabolic and oxygen demands. Therefore, there is no single ideal experimental technique to use in all scenarios. Hence, the decision on whether to use contracting heart or EC uncoupled heart depends on the specific research goals. For measurements involving metabolism and cardiac electromechanics including mechano‐electric feedback, we recommend using contracting heart optical mapping.
Additional information
Competing interests
The authors have no conflicts to disclose.
Author contributions
The experiments were performed in the laboratories of S.L. and F.S.N. F.S.N. and V.K. designed the research. V.K. performed the experiments. V.K. and A.S. prepared the figures and drafted the manuscript. A.S., U.P., S.L., I.U., F.F., N.S.P., J.C. and F.S.N. contributed to experiments and data acquisition. F.S.N., J.C., N.S.P., U.P., S.L., F.F. and I.U. provided critical comments and contributed to the revision of the manuscript. All authors have read and approved the final version of this manuscript and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. All persons designated as authors qualify for authorship, and all those who qualify for authorship are listed.
Funding
This study was supported by the British Heart Foundation (RG/16/3/32175 and RG/F/22/110078 to F.S.N., N.S.P. and V.K.), Imperial NIHR Biomedical Research Centre funding (to F.S.N.), European Union's Horizon 2020 through the project Advanced BiomEdical OPTICAL Imaging and Data Analysis (BE‐OPTICAL) under grant agreement number 675512 (U.P. and S.L.), German Centre for Cardiovascular Research (DZHK e.V.), German Federal Ministry of Education and Research (BMBF, project FKZ 031A147, Go‐Bio), the German Research Foundation (DFG, Collaborative Research Centers SFB 1002, Projects B05 and C03 and SFB 937, Project A18) and the Max Planck Society (to S.L.), NIH 1R01HL143450‐01 (to F.F.), NSF‐FDA‐203 789 and NSF 1 762 553 (to F.F.).
Supporting information
Acknowledgements
The authors would like to thank Danya Agha‐Jaffar and Marion Kunze for technical assistance.
Biographies
Vineesh Kappadan is a Research Associate at the National Heart and Lung Institute, Imperial College London. He is currently investigating cardiac arrhythmia dynamics in isolated contracting hearts using optical mapping. He was awarded his PhD by University of Göttingen, Germany, for his work on ratiometric optical mapping and marker‐free motion tracking of contracting hearts. He completed his master's degree in Applied Optics from Indian Institute of Technology Delhi, and his main area of interests are optics and its application to biology.
Fu Siong Ng is a Clinical Senior Lecturer in Cardiac Electrophysiology at Imperial College London, and a Consultant Cardiologist. He is a Clinician Scientist and currently leads a programme of research into arrhythmogenic mechanisms alongside performing invasive ablation procedures and implanting pacemakers and defibrillators in patients with heart rhythm disorders. He is also the Programme Director for the intercalated BSc in Cardiovascular Sciences at Imperial College London and Divisional Research Lead at Chelsea and Westminster Hospital NHS Foundation Trust.
Handling Editors: Laura Bennet & Nordine Helassa
The peer review history is available in the Supporting Information section of this article (https://doi.org/10.1113/JP283683#support‐information‐section).
References
- Attin, M. , & Clusin, W. T. (2009). Basic concepts of optical mapping techniques in cardiac electrophysiology. Biological Research for Nursing, 11(2), 195–207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bachtel, A. D. , Gray, R. A. , Stohlman, J. M. , Bourgeois, E. B. , Pollard, A. E. , & Rogers, J. M. (2011). A novel approach to dual excitation ratiometric optical mapping of cardiac action potentials with di‐4‐ANEPPS using pulsed LED excitation. IEEE Transactions on Bio‐Medical Engineering, 58(7), 2120–2126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Banville, I. , & Gray, R. A. (2002). Effect of action potential duration and conduction velocity restitution and their spatial dispersion on alternans and the stability of arrhythmias. Journal of Cardiovascular Electrophysiology, 13(11), 1141–1149. [DOI] [PubMed] [Google Scholar]
- Bernus, O. , Mukund, K. S. , & Pertsov, A. M. (2007). Detection of intramyocardial scroll waves using absorptive transillumination imaging. Journal of Biomedial Optics, 12(1), 014035. [DOI] [PubMed] [Google Scholar]
- Bourgeois, E. B. , Bachtel, A. D. , Huang, J. , Walcott, G. P. , & Rogers, J. M. (2011). Simultaneous optical mapping of transmembrane potential and wall motion in isolated, perfused whole hearts. Journal of Biomedial Optics, 16(9), 096020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brack, K. E. , Narang, R. , Winter, J. , & Ng, G. A. (2013). The mechanical uncoupler blebbistatin is associated with significant electrophysiological effects in the isolated rabbit heart. Experimental Physiology, 98(5), 1009–1027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brandes, R. , Figueredo, V. M. , Camacho, S. A. , Massie, B. M. , & Weiner, M. W. (1992). Suppression of motion artifacts in fluorescence spectroscopy of perfused hearts. American Journal of Physiology, 263(3 Pt 2), H972–980. [DOI] [PubMed] [Google Scholar]
- Chattipakorn, N. , Banville, I. , Gray, R. A. , & Ideker, R. E. (2001). Mechanism of ventricular defibrillation for near‐defibrillation threshold shocks: A whole‐heart optical mapping study in swine. Circulation, 104(11), 1313–1319. [DOI] [PubMed] [Google Scholar]
- Chen, J. , Mandapati, R. , Berenfeld, O. , Skanes, A. C. , & Jalife, J. (2000). High‐frequency periodic sources underlie ventricular fibrillation in the isolated rabbit heart. Circulation Research, 86(1), 86–93. [DOI] [PubMed] [Google Scholar]
- Chen, V. , Chen, Y. H. , & Downing, S. E. (1987). An improved isolated working rabbit heart preparation using red cell enhanced perfusate. The Yale Journal of Biology and Medicine, 60(3), 209–219. [PMC free article] [PubMed] [Google Scholar]
- Cherry, E. M. , & Fenton, F. H. (2008). Visualization of spiral and scroll waves in simulated and experimental cardiac tissue. New Journal of Physics, 10(12), 125016. [Google Scholar]
- Choi, B. R. , & Salama, G. (2000). Simultaneous maps of optical action potentials and calcium transients in guinea‐pig hearts: Mechanisms underlying concordant alternans. The Journal of Physiology, 529(1), 171–188. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chowdhury, R. A. , Tzortzis, K. N. , Dupont, E. , Selvadurai, S. , Perbellini, F. , Cantwell, C. D. , Ng, F. S. , Simon, A. R. , Terracciano, C. M. , & Peters, N. S. (2018). Concurrent micro‐ to macro‐cardiac electrophysiology in myocyte cultures and human heart slices. Scientific Reports, 8(1), 6947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Christoph, J. , Chebbok, M. , Richter, C. , Schröder‐Schetelig, J. , Bittihn, P. , Stein, S. , Uzelac, I. , Fenton, F. H. , Hasenfuß, G. , Gilmour, R. F. , & Luther, S. (2018). Electromechanical vortex filaments during cardiac fibrillation. Nature, 555(7698), 667–672. [DOI] [PubMed] [Google Scholar]
- Christoph, J. , & Luther, S. (2018). Marker‐free tracking for motion artifact compensation and deformation measurements in optical mapping videos of contracting hearts. Frontiers in Physiology, 9, 1483. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Christoph, J. , Schröder‐Schetelig, J. , & Luther, S. (2017). Electromechanical optical mapping. Progress in Biophysics and Molecular Biology, 130(Pt B), 150–169. [DOI] [PubMed] [Google Scholar]
- Dou, Y. , Arlock, P. , & Arner, A. (2007). Blebbistatin specifically inhibits actin‐myosin interaction in mouse cardiac muscle. American Journal of Physiology. Cell Physiology, 293(3), C1148–C1153. [DOI] [PubMed] [Google Scholar]
- Dumas, J. H. , & Kinisley, S. B. (2005). Two‐photon excitation of di‐4‐ANEPPS for optical recording of action potentials in rabbit heart. Annals of Biomedical Engineering, 33(12), 1802–1807. [DOI] [PubMed] [Google Scholar]
- Efimov, I. R. , Nikolski, V. P. , & Salama, G. (2004). Optical imaging of the heart. Circulation Research, 95(1), 21–33. [DOI] [PubMed] [Google Scholar]
- Efimov, I. R. , Sidorov, V. , Cheng, Y. , & Wollenzier, B. (1999). Evidence of three‐dimensional scroll waves with ribbon‐shaped filament as a mechanism of ventricular tachycardia in the isolated rabbit heart. Journal of Cardiovascular Electrophysiology, 10(11), 1452–1462. [DOI] [PubMed] [Google Scholar]
- Fedorov, V. V. , Lozinsky, I. T. , Sosunov, E. A. , Anyukhovsky, E. P. , Rosen, M. R. , Balke, C. W. , & Efimov, I. R. (2007). Application of blebbistatin as an excitation‐contraction uncoupler for electrophysiologic study of rat and rabbit hearts. Heart Rhythm, 4(5), 619–626. [DOI] [PubMed] [Google Scholar]
- Fedorov, V. V. , Schuessler, R. B. , Hemphill, M. , Ambrosi, C. M. , Chang, R. , Voloshina, A. S. , Brown, K. , Hucker, W. J. , & Efimov, I. R. (2009). Structural and functional evidence for discrete exit pathways that connect the Canine Sinoatrial node and atria. Circulation Research, 104(7), 915–923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fenton, F. H. , Cherry, E. M. , & Kornreich, B. G. (2008). Termination of equine atrial fibrillation by quinidine: An optical mapping study. Journal of Veterinary Cardiology, 10(2), 87–103. [DOI] [PubMed] [Google Scholar]
- Fenton, F. H. , Luther, S. , Cherry, E. M. , Otani, N. F. , Krinsky, V. , Pumir, A. , Bodenschatz, E. , & Gilmour, R. F. (2009). Termination of atrial fibrillation using pulsed low‐energy far‐field stimulation. Circulation, 120(6), 467–476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fialova, K. , Kolarova, J. , Provaznik, I. , & Novakova, M. (2010). Comparison of voltage‐sensitive dye di‐4‐ANNEPS effects in isolated hearts of rat, guinea pig, and rabbit. In 2010 Computing in Cardiology, pp. 565–568.
- Franz, M. R. (1996). Mechano‐electrical feedback in ventricular myocardium. Cardiovascular Research, 32(1), 15–24. [PubMed] [Google Scholar]
- Franz, M. R. (2000). Mechano‐electrical feedback. Cardiovascular Research, 45(2), 263–266. [DOI] [PubMed] [Google Scholar]
- Fukano, T. , Shimozono, S. , & Miyawaki, A. (2006). Fast dual‐excitation ratiometry with light‐emitting diodes and high‐speed liquid crystal shutters. Biochemical and Biophysical Research Communications, 340(1), 250–255. [DOI] [PubMed] [Google Scholar]
- Fukano, T. , Shimozono, S. , & Miyawaki, A. (2008). Development of microscopic systems for high‐speed dual‐excitation ratiometric Ca2+ imaging. Brain Cell Biology, 36(1–4), 43–52. [DOI] [PubMed] [Google Scholar]
- Gauduel, Y. , Martin, J. L. , Teisseire, B. , Duruble, M. , & Duvelleroy, M. (1985). Hemodynamic and metabolic responses of the working heart in relation to the oxygen carrying capacity of the perfusion medium. General Physiology and Biophysics, 4(6), 573–587. [PubMed] [Google Scholar]
- Greene, D. , Kaboudian, A. , Wasserstrom, J. A. , Fenton, F. H. , & Shiferaw, Y. (2022). Voltage‐mediated mechanism for calcium wave synchronization and arrhythmogenesis in atrial tissue. Biophysical Journal, 121(3), 383–395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Handa, B. S. , Li, X. , Baxan, N. , Roney, C. H. , Shchendrygina, A. , Mansfield, C. A. , Jabbour, R. J. , Pitcher, D. S. , Chowdhury, R. A. , Peters, N. S. , & Ng, F. S. (2021). Ventricular fibrillation mechanism and global fibrillatory organization are determined by gap junction coupling and fibrosis pattern. Cardiovascular Research, 117(4), 1078–1090. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hansen, B. J. , Li, N. , Helfrich, K. M. , Abudulwahed, S. H. , Artiga, E. J. , Joseph, M. E. , Mohler, P. J. , Hummel, J. D. , & Fedorov, V. V. (2018). First in vivo use of high‐resolution near‐infrared optical mapping to assess atrial activation during sinus rhythm and atrial fibrillation in a large animal model. Circ Arrhythm Electrophysiol, 11(12), e006870. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herndon, C. , Astley, H. C. , Owerkowicz, T. , & Fenton, F. H. (2021). Defibrillate you later, Alligator: Q10 scaling and refractoriness keeps alligators from fibrillation. Integrative Organismal Biology, 3(1), obaa047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herron, T. J. , Lee, P. , & Jalife, J. (2012). Optical imaging of voltage and calcium in cardiac cells & tissues. Circulation Research, 110(4), 609–623. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hooks, D. A. , LeGrice, I. J. , Harvey, J. D. , & Smaill, B. H. (2001). Intramural multisite recording of transmembrane potential in the heart. Biophysical Journal, 81(5), 2671–2680. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hortigon‐Vinagre, M. P. , Zamora, V. , Burton, F. L. , Green, J. , Gintant, G. A. , & Smith, G. L. (2016). The use of ratiometric fluorescence measurements of the voltage sensitive dye Di‐4‐ANEPPS to examine action potential characteristics and drug effects on human induced pluripotent stem cell‐derived cardiomyocytes. Toxicological Sciences off Journal Society of Toxicology, 154(2), 320–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Janoušek, O. , Ronzhina, M. , Hejč, J. , Olejníčková, V. , Stračina, T. , Fialová, K. , Nováková, M. , Provazník, I. , & Kolářová, J. (2015). The effect of voltage‐sensitive dye di‐4‐ANEPPS on heart rate variability in Langendorff‐perfused isolated rabbit heart. In 2015 Computing in Cardiology Conference (CinC), pp. 1049–1052.
- Jensen, B. , Boukens, B. J. , Crossley, D. A. , Conner, J. , Mohan, R. A. , van Duijvenboden, K. , Postma, A. V. , Gloschat, C. R. , Elsey, R. M. , Sedmera, D. , Efimov, I. R. , & Christoffels, V. M. (2018). Specialized impulse conduction pathway in the alligator heart. eLife, 7, e32120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jou, C. J. , Spitzer, K. W. , & Tristani‐Firouzi, M. (2010). Blebbistatin effectively uncouples the excitation‐contraction process in zebrafish embryonic heart. Cellular Physiology and Biochemistry, 25(4–5), 419–424. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kappadan, V. (2021). Ratiometric fluorescence imaging and marker‐free motion tracking of Langendorff perfused beating rabbit hearts. Advance online publication. 10.53846/goediss-8384 [DOI]
- Kappadan, V. , Telele, S. , Uzelac, I. , Fenton, F. , Parlitz, U. , Luther, S. , & Christoph, J. (2020). High‐resolution optical measurement of cardiac restitution, contraction, and fibrillation dynamics in beating vs. blebbistatin‐uncoupled isolated rabbit hearts. Frontiers in Physiology, 11, 464. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kettlewell, S. , Walker, N. L. , Cobbe, S. M. , Burton, F. L. , & Smith, G. L. (2004). The electrophysiological and mechanical effects of 2,3‐butane‐dione monoxime and cytochalasin‐D in the Langendorff perfused rabbit heart. Experimental Physiology, 89(2), 163–172. [DOI] [PubMed] [Google Scholar]
- Khwaounjoo, P. , Rutherford, S. L. , Svrcek, M. , LeGrice, I. J. , Trew, M. L. , & Smaill, B. H. (2015). Image‐based motion correction for optical mapping of cardiac electrical activity. Annals of Biomedical Engineering, 43(5), 1235–1246. [DOI] [PubMed] [Google Scholar]
- Knisley, S. B. , Justice, R. K. , Kong, W. , & Johnson, P. L. (2000). Ratiometry of transmembrane voltage‐sensitive fluorescent dye emission in hearts. American Journal of Physiology. Heart and Circulatory Physiology, 279(3), H1421–H1433. [DOI] [PubMed] [Google Scholar]
- Lang, D. , Sulkin, M. , Lou, Q. , & Efimov, I. R. (2011). Optical mapping of action potentials and calcium transients in the mouse heart. Journal of Visualized Experiments, 13(55), 3275. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Larsen, A. P. , Sciuto, K. J. , Moreno, A. P. , & Poelzing, S. (2012). The voltage‐sensitive dye di‐4‐ANEPPS slows conduction velocity in isolated guinea pig hearts. Heart Rhythm, 9(9), 1493–1500. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Laughner, J. I. , Ng, F. S. , Sulkin, M. S. , Arthur, R. M. , & Efimov, I. R. (2012). Processing and analysis of cardiac optical mapping data obtained with potentiometric dyes. American Journal of Physiology. Heart and Circulatory Physiology, 303(7), H753‐H765. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Laurita, K. R. , Girouard, S. D. , Akar, F. G. , & Rosenbaum, D. S. (1998). Modulated dispersion explains changes in arrhythmia vulnerability during premature stimulation of the heart. Circulation, 98(24), 2774–2780. [DOI] [PubMed] [Google Scholar]
- Lebert, J. , Ravi, N. , Kensah, G. , & Christoph, J. (2022). Real‐time optical mapping of contracting cardiac tissues with GPU‐accelerated numerical motion tracking. Frontiers in Cardiovascular Medicine, 9, 787627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lee, M. H. , Lin, S. F. , Ohara, T. , Omichi, C. , Okuyama, Y. , Chudin, E. , Garfinkel, A. , Weiss, J. N. , Karagueuzian, H. S. , & Chen, P. S. (2001). Effects of diacetyl monoxime and cytochalasin D on ventricular fibrillation in swine right ventricles. American Journal of Physiology. Heart and Circulatory Physiology, 280(6), H2689‐H2696. [DOI] [PubMed] [Google Scholar]
- Lee, P. , Quintanilla, J. G. , Alfonso‐Almazán, J. M. , Galán‐Arriola, C. , Yan, P. , Sánchez‐González, J. , Pérez‐Castellano, N. , Pérez‐Villacastín, J. , Ibañez, B. , Loew, L. M. , & Filgueiras‐Rama, D. (2019). In vivo ratiometric optical mapping enables high‐resolution cardiac electrophysiology in pig models. Cardiovascular Research, 115(11), 1659–1671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lou, Q. , Li, W. , & Efimov, I. R. (2012). The role of dynamic instability and wavelength in arrhythmia maintenance as revealed by panoramic imaging with blebbistatin vs. 2,3‐butanedione monoxime. American Journal of Physiology. Heart and Circulatory, 302(1), H262–H269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mandapati, R. , Asano, Y. , Baxter, W. T. , Gray, R. , Davidenko, J. , & Jalife, J. (1998). Quantification of effects of global ischemia on dynamics of ventricular fibrillation in isolated rabbit heart. Circulation, 98(16), 1688–1696. [DOI] [PubMed] [Google Scholar]
- Matiukas, A. , Mitrea, B. G. , Qin, M. , Pertsov, A. M. , Shvedko, A. G. , Warren, M. D. , Zaitsev, A. V. , Wuskell, J. P. , Wei, M. , Watras, J. , & Loew, L. M. (2007). Near infrared voltage sensitive fluorescent dyes optimized for optical mapping in blood‐perfused myocardium. Heart Rhythm, 4(11), 1441–1451. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mitrea, B. G. , Wellner, M. , & Pertsov, A. M. (2009). Monitoring intramyocardial reentry using alternating transillumination. Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009, 4194–4197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nanthakumar, K. , Jalife, J. , Massé, S. , Downar, E. , Pop, M. , Asta, J. , Ross, H. , Rao, V. , Mironov, S. , Sevaptsidis, E. , Rogers, J. , Wright, G. , & Dhopeshwarkar, R. (2007). Optical mapping of Langendorff‐perfused human hearts: Establishing a model for the study of ventricular fibrillation in humans. American Journal of Physiology. Heart and Circulatory Physiology, 293(1), H875–H880. [DOI] [PubMed] [Google Scholar]
- Nesmith, H. W. , Zhang, H. , & Rogers, J. M. (2020). Optical mapping of electromechanics in intact organs. Experimental Biology and Medicine, 245(4), 368–373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- O'Shea, C. , Kabir, S. N. , Holmes, A. P. , Lei, M. , Fabritz, L. , Rajpoot, K. , & Pavlovic, D. (2020). Cardiac optical mapping – State‐of‐the‐art and future challenges. International Journal of Biochemistry & Cell Biology, 126, 105804. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pertsov, A. M. , Davidenko, J. M. , Salomonsz, R. , Baxter, W. T. , & Jalife, J. (1993). Spiral waves of excitation underlie reentrant activity in isolated cardiac muscle. Circulation Research, 72(3), 631–650. [DOI] [PubMed] [Google Scholar]
- Pitruzzello, A. M. , Krassowska, W. , & Idriss, S. F. (2007). Spatial heterogeneity of the restitution portrait in rabbit epicardium. American Journal of Physiology. Heart and Circulatory Physiology, 292(3), H1568–H1578. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Riccio, M. L. , Koller, M. L. , & Gilmour, R. F. (1999). Electrical restitution and spatiotemporal organization during ventricular fibrillation. Circulation Research, 84(8), 955–963. [DOI] [PubMed] [Google Scholar]
- Rohde, G. K. , Dawant, B. M. , & Lin, S.‐F. (2005). Correction of motion artifact in cardiac optical mapping using image registration. IEEE Transactions on Bio‐Medical Engineering, 52(2), 338–341. [DOI] [PubMed] [Google Scholar]
- Rohr, S. , & Kucera, J. P. (1998). Optical recording system based on a fiber optic image conduit: Assessment of microscopic activation patterns in cardiac tissue. Biophysical Journal, 75(2), 1062–1075. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ronzhina, M. , Stracina, T. , Lacinova, L. , Ondacova, K. , Pavlovicova, M. , Marsanova, L. , Smisek, R. , Janousek, O. , Fialova, K. , Kolarova, J. , Novakova, M. , & Provaznik, I. (2021). Di‐4‐ANEPPS modulates electrical activity and progress of myocardial ischemia in rabbit isolated heart. Frontiers in Physiology, 12, 667065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rueckschloss, U. , & Isenberg, G. (2001). Cytochalasin D reduces Ca2+ currents via cofilin‐activated depolymerization of F‐actin in guinea‐pig cardiomyocytes. The Journal of Physiology, 537(2), 363–370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuzmiak‐Glancy, S. , Jaimes, R. 3rd , Wengrowski, A. M. , & Kay, M. W. (2015). Oxygen demand of perfused heart preparations: How electromechanical function and inadequate oxygenation affect physiology and optical measurements. Experimental Physiology, 100(6), 603–616. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Salama, G. , & Morad, M. (1976). Merocyanine 540 as an optical probe of transmembrane electrical activity in the heart. Science, 191(4226), 485–487. [DOI] [PubMed] [Google Scholar]
- Sellin, L. C. , & McArdle, J. J. (1994). Multiple effects of 2,3‐butanedione monoxime. Pharmacology & Toxicology, 74(4–5), 305–313. [DOI] [PubMed] [Google Scholar]
- Seo, K. , Inagaki, M. , Nishimura, S. , Hidaka, I. , Sugimachi, M. , Hisada, T. , & Sugiura, S. (2010). Structural heterogeneity in the ventricular wall plays a significant role in the initiation of stretch‐induced arrhythmias in perfused rabbit right ventricular tissues and whole heart preparations. Circulation Research, 106(1), 176–184. [DOI] [PubMed] [Google Scholar]
- Shiferaw, Y. , Sato, D. , & Karma, A. (2005). Coupled dynamics of voltage and calcium in paced cardiac cells. Physical Review. E, Statistical, Nonlinear, and Soft Matter Physics, 71(2), 021903. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Straight, A. F. , Cheung, A. , Limouze, J. , Chen, I. , Westwood, N. J. , Sellers, J. R. , & Mitchison, T. J. (2003). Dissecting temporal and spatial control of cytokinesis with a myosin II Inhibitor. Science, 299(5613), 1743–1747. [DOI] [PubMed] [Google Scholar]
- Svrcek, M. , Rutherford, S. , Chen, A. Y. , Provaznik, I. , & Smaill, B. (2009). Characteristics of motion artifacts in cardiac optical mapping studies. Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009, 3240–3243. [DOI] [PubMed] [Google Scholar]
- Swift, L. M. , Asfour, H. , Posnack, N. G. , Arutunyan, A. , Kay, M. W. , & Sarvazyan, N. (2012). Properties of blebbistatin for cardiac optical mapping and other imaging applications. Pflugers Archiv: European Journal of Physiology, 464(5), 503–512. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Swift, L. M. , Kay, M. W. , Ripplinger, C. M. , & Posnack, N. G. (2021). Stop the beat to see the rhythm: Excitation‐contraction uncoupling in cardiac research. American Journal of Physiology. Heart and Circulatory Physiology, 321(6), H1005–H1013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Taggart, P. (1996). Mechano‐electric feedback in the human heart. Cardiovascular Research, 32(1), 38–43. [PubMed] [Google Scholar]
- Uzelac, I. , Crowley, C. J. , Iravanian, S. , Kim, T. Y. , Cho, H. C. , & Fenton, F. H. (2022). Methodology for cross‐talk elimination in simultaneous voltage and calcium optical mapping measurements with Semasbestic wavelengths. Frontiers in Physiology, 13, 812968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valderrábano, M. , Lee, M.‐H. , Ohara, T. , Lai, A. C. , Fishbein, M. C. , Lin, S.‐F. , Karagueuzian, H. S. , & Chen, P.‐S. (2001). Dynamics of intramural and transmural reentry during ventricular fibrillation in isolated swine ventricles. Circulation Research, 88(8), 839–848. [DOI] [PubMed] [Google Scholar]
- Zabel, M. , Koller, B. S. , Sachs, F. , & Franz, M. R. (1996). Stretch‐induced voltage changes in the isolated beating heart: Importance of the timing of stretch and implications for stretch‐activated ion channels. Cardiovascular Research, 32(1), 120–130. [PubMed] [Google Scholar]
- Zhang, H. , Iijima, K. , Huang, J. , Walcott, G. P. , & Rogers, J. M. (2016). Optical mapping of membrane potential and epicardial deformation in beating hearts. Biophysical Journal, 111(2), 438–451. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.