Large variation in radiation dose for routine abdomen CT: reasons for excess and easy tips for reduction

Rebecca Smith-Bindman 1,2,3,, Taewoon Kang 1, Philip W Chu 1, Yifei Wang 1, Carly Stewart 1, Marco Das 4, Phuong-Anh Duong 5, Luisa Cervantes 6, Ramit Lamba 7, Ryan K Lee 8, Fiona MacLeod 9, Nima Kasraie 10, Rebecca Neill 11, Pavlina Pike 12, Jodi Roehm 13, Sebastian Schindera 14, Robert Chung 15, Bradley N Delman 16, Cécile R L P N Jeukens 17, L Jay Starkey 18, Timothy P Szczykutowicz 19
PMCID: PMC10957641  PMID: 37735276

Abstract

Objective

To characterize the use and impact of radiation dose reduction techniques in actual practice for routine abdomen CT.

Methods

We retrospectively analyzed consecutive routine abdomen CT scans in adults from a large dose registry, contributed by 95 hospitals and imaging facilities. Grouping exams into deciles by, first, patient size, and second, size-adjusted dose length product (DLP), we summarized dose and technical parameters and estimated which parameters contributed most to between-protocols dose variation. Lastly, we modeled the total population dose if all protocols with mean size-adjusted DLP above 433 or 645 mGy-cm were reduced to these thresholds.

Results

A total of 748,846 CTs were performed using 1033 unique protocols. When sorted by patient size, patients with larger abdominal diameters had increased dose and effective mAs (milliampere seconds), even after adjusting for patient size. When sorted by size-adjusted dose, patients in the highest versus the lowest decile in size-adjusted DLP received 6.4 times the average dose (1680 vs 265 mGy-cm) even though diameter was no different (312 vs 309 mm). Effective mAs was 2.1-fold higher, unadjusted CTDIvol 2.9-fold, and phase 2.5-fold for patients in the highest versus lowest size-adjusted DLP decile. There was virtually no change in kV (kilovolt). Automatic exposure control was widely used to modulate mAs, whereas kV modulation was rare. Phase was the strongest driver of between-protocols variation. Broad adoption of optimized protocols could result in total population dose reductions of 18.6–40%.

Conclusion

There are large variations in radiation doses for routine abdomen CT unrelated to patient size. Modification of kV and single-phase scanning could result in substantial dose reduction.

Clinical relevance

Radiation dose-optimization techniques for routine abdomen CT are routinely under-utilized leading to higher doses than needed. Greater modification of technical parameters and number of phases could result in substantial reduction in radiation exposure to patients.

Key Points

• Based on an analysis of 748,846 routine abdomen CT scans in adults, radiation doses varied tremendously across patients of the same size and optimization techniques were routinely under-utilized.

• The difference in observed dose was due to variation in technical parameters and phase count. Automatic exposure control was commonly used to modify effective mAs, whereas kV was rarely adjusted for patient size. Routine abdomen CT should be performed using a single phase, yet multi-phase was common.

• kV modulation by patient size and restriction to a single phase for routine abdomen indications could result in substantial reduction in radiation doses using well-established dose optimization approaches.

Supplementary information

The online version contains supplementary material available at 10.1007/s00330-023-10076-6.

Keywords: Tomography, X-ray computed; Abdomen; Radiation dosage

Introduction

CT utilization has grown significantly in the last three decades, with an estimated 91 million scans performed in the United States in 2019 [1] and 90 million scans in the European Union in 2020 [2]. Concern over the corresponding increase in exposure to ionizing radiation has led to broad interest in radiation dose optimization and in avoiding unnecessary exams and non-indicated scan series [311]. Despite this attention, large variation in dose persists [9, 1215].

Extensive work has been published on dose reduction approaches. There have been meaningful technological developments, such as automatic exposure control (AEC) and iterative reconstruction [16], as well as educational efforts to encourage modification of technical parameters, such as the use of lower tube potential (kilovoltage, kV) and size-specific protocols [1720]. However, little is known to what extent radiology providers employ these techniques, or the impact on radiation dose when they are used. Understanding current practice could drive improvement.

Using observed CT data from a large radiation dose registry, we describe the frequency with which different technical parameters are used for routine abdomen CT and the impact on patient radiation dose when dose reduction strategies are or are not implemented.

Methods

Study population

Drawing from the University of California, San Francisco (UCSF) International CT Dose Registry (“registry” [13]), we retrospectively analyzed consecutive diagnostic CT scans performed from January 1, 2015 to October 21, 2020, in patients aged 18 years and older. The registry pools 100% of CT scans performed at imaging facilities from 27 healthcare organizations in 7 countries, all of which submitted data via Radimetrics© dose management software (Bayer HealthCare). The data include the four largest CT manufacturers (Table 1) [12, 13]. The UCSF Committee on Human Research approved this study with a waiver of informed consent. Collaborating institutions obtained local Institutional Review Board approval or relied on UCSF approval to contribute to the registry.

Table 1.

Routine abdomen and pelvis CT scans included in this report

n %
Total number of scans 748,846 100.0
Sex
 Men 321,648 43.0
 Women 426,840 57.0
 Non-binary or unknown 358 0.0
Age
 18–19 9,169 1.2
 20–29 67,695 9.0
 30–39 89,195 11.9
 40–49 110,103 14.7
 50–59 138,768 18.5
 60–69 142,551 19.0
 70–79 111,508 14.9
 80–89 63,881 8.5
 90–99 15,976 2.1
Manufacturer
 Canon 58,815 7.9
 General Electric 298,316 39.8
 Philips 146,937 19.6
 Siemens 244,778 32.7
Country
 Germany 6,565 0.9
 Israel 33,941 4.5
 Japan 8,732 1.2
 Netherlands 13,829 1.8
 Switzerland 8,963 1.2
 UK 11,515 1.5
 USA 665,301 88.8

For each CT scan, patient sex and age, effective patient diameter, CT category (reflecting the body region and image quality requirements [21]), scanner manufacturer and model, and the radiation dose metrics and technical parameters of the scan were extracted from the registry [12, 13, 22]. These analyses focus on scans obtained for routine abdomen CT, determined using information contained in the Digital Imaging and Communications in Medicine (DICOM) headers. Abdomen scans obtained for low-dose indications (e.g., suspected renal stones) or high-dose indications (e.g., cancer surveillance or assessment of acute intra-abdominal bleeding) are not included in this manuscript. A full list of indications included/excluded in routine abdomen CT, and the validation of the framework for assigning CT scans to categories demonstrating 90% accuracy compared with expert review, has been previously published [21]. Routine abdomen is one of the most common reasons patients undergo CT imaging, accounting for approximately 24% of all CT scans [21]. This category most closely aligns with the EUCLID abdomen, appendicitis category [12].

Radiation dose variables

Radiation dose metrics included the machine-reported volume computed tomography dose index (CTDIvol) and dose length product (DLP), and effective dose, which is calculated by Radimetrics and reflects future cancer risk. Radiation dose varies non-linearly by patient size, so that in general larger patients require asymmetrically higher doses than smaller patients to achieve sufficient image quality. To minimize the impact of patient size on variation across practices, we calculated size-adjusted CTDIvol and size-adjusted DLP by normalizing these metrics using the log-linear mixed regression between them and effective patient diameter (patient diameter, defined as the average diameter measured by Radimetrics on axial or scout images); the facility at which the scan was performed was included as a random effect. The following equations were used where the population median abdominal diameter = 303 mm:

ADJCTDIvol=CTDIvol×exp-EFFECTIVEPATIENTDIAMETER-303×0.007682
ADJDLP=DLP×exp(-EFFECTIVEPATIENTDIAMETER-303×0.008678)

This differs from size-specific dose estimate, which normalizes to obtain consistent dose per unit of tissue as reflected in a phantom [23, 24]. Our size normalization mitigates the effect of patient size on dose, using the relationship between size and dose observed in our dataset. The goal of our approach—in the context of comparing dose across protocols and hospitals—is to ensure the impact of patient size on doses is eliminated.

Patient size

Patients were divided by size in several ways to analyze the variation in radiation dose and technical parameters. First, patients were divided into deciles based on patient diameter in millimeters [21]. Second, because deciles may be too crude to measure changes in technical parameters that occur with change in patient diameter, as a second approach, patients were double-stratified into three size categories based on patient diameter, defined as small if they were less than or equal to the 25th percentile, medium if between 25 and 75th percentiles, and large if above the 75th percentile. Then within each size category, patients were divided into deciles (= 30 deciles).

Characterization of imaging protocols

We identified each protocol in the registry used for routine abdomen and pelvis indications (including all scans of the abdomen, pelvis, or combined abdomen and pelvis), defining a protocol as a unique combination of protocol name and specific scanner used. We determined whether each protocol was used in one, two, or all three patient size categories. To qualify, at least 20 patients in a size category had to have been imaged with a given protocol.

For each protocol, we used medians to summarize the radiation dose metrics and technical parameters of its constituent CT exams, including those directly reported by the scanner (CTDIvol, DLP, kV, mAs, collimation, pitch, and scan length) and those calculated (patient diameter, size-adjusted CTDIvol, size-adjusted DLP, effective dose, effective mAs, and number of phases). The scan length was defined as the total irradiated region and for multiple-phase studies was the average across irradiating events. For each exam, when there were multiple irradiating events, the DLP, size-adjusted DLP, ED, and number of phases were summed across the irradiating events, whereas the CTDIvol, size-adjusted CTDIvol, mAs, effective mAs, scan length, and pitch were averaged and weighted by scan length across irradiating events. Bolus scans were not included.

Statistical analysis

CT scans missing dose metrics or technical parameters were not included. In order to exclude outliers, we dropped scans with values less than the 0.1th percentile. Because scanner model may contribute to dose, we included only CT scans obtained on scanner models where at least 5 individual scanners of that model exist in the registry.

Each CT scan was assigned to one of the three patient size categories (small, medium, large). Descriptive statistics were examined for demographic variables, radiation dose metrics, and the technical parameters as they varied by patient size. Stratifying by decile of patient size, we calculated mean (and 95% CI calculated using bootstrapping) percent increase in radiation dose metrics and technical parameters per decile increase in size, and we calculated the percent increase (and 95% CI using bootstrapping) of the dose metrics and technical parameters in the highest compared with lowest deciles. We then repeated these analyses for each of the three patient size categories.

Protocols were next stratified by decile of size-adjusted DLP to illustrate the variation in dose. For each protocol, we calculated the median size-adjusted DLP, and then sorted and grouped into deciles of median size-adjusted DLP. Within each decile of size-adjusted DLP, we calculated the mean values (and 95% CI) of dose and technical parameters for the constituent protocols, and—to quantify variation in dose and parameters—we calculated the mean percent change (and 95% CI using bootstrapping) for each variable per decile increase in size-adjusted DLP and between the 1st (lowest dose) to 10th (highest dose) deciles. We then repeated these analyses for each of the three patient size categories.

To identify which technical parameters contributed the most to observed between-protocols variation in dose, we combined the observed percent increases between deciles of the technical parameters with established understanding of parameter-dose relationship in medical physics. Specifically, for an exam in which a single technical parameter sees an X% increase with all other factors kept constant, the DLP will see the same X% increase for effective mAs, scan length, or number of phases, whereas the DLP will see an even greater increase [(1 + X)2.5–1]% for an X% increase in kV [25]. Using these established equations, for each technical parameter, we computed the expected increase in size-adjusted DLP per decile due to the observed change in the parameter. Technical parameters that contribute strongly to between-protocols dose variation should induce an expected DLP increase per decile similar to the observed size-adjusted DLP increase per decile.

Finally, to estimate the potential impact of widespread adoption of optimized protocols for routine abdomen CT, we first calculated the total dose to all patients in this study by summing the size-adjusted doses for all examinations. We then recalculated the total dose under two scenarios. First, we defined the target routine abdomen protocol dose as 433 mGy-cm. This is the median (achievable dose (AD)) for acute appendicitis in Europe using the EUCLID framework [26]. We then (1) summed all observed doses for patients who were scanned with protocols with a mean at or below 433 (their doses do not change), (2) summed all observed doses for remaining patients who were scanned with size-adjusted DLP at or below 433 (their doses also do not change), and (3) multiplied the remaining patients by 433 mGy-cm (to reflect their new optimized doses in this hypothetical calculation), and (4) summed the values produced by (1), (2), and (3) to produce a “post-optimization population total dose.” Second, we repeated this calculation using a target dose of 645 mGy-cm, which reflects the median dose of acute appendicitis in the USA using the EUCLID framework [26]. This value is similar to the median dose for abdomen CT reported by the American College of Radiology (AD = 615 mGy-cm for abdomen and pelvis with contrast and AD = 657 mGy-cm for abdomen and pelvis without contrast) [15]. We then estimated the total dose saved by computing the percent reduction between the observed and hypothetical post-optimization population total doses in both scenarios.

Results

A total of 748,846 routine abdomen CT scans were included (Table 1). There were more CT scans obtained in women (n = 426,840, 57%) and the number of scans increased with age and peaked in those aged 60–69 years (n = 142,551, 19%). Most scans were obtained on GE (n = 298,316, 39.8%) or Siemens (n = 244,778, 32.7%) scanners, and most were performed in the USA (n = 665,301, 88.8%).

The data included 1033 protocols collected from 95 facilities and 242 scanners. The mean number of scans per protocol was 725 (median = 159, range = 20–22, 733). About half of all protocols (n = 547, 53%) were not used in a size-specific fashion, meaning they were used for patients across all three size categories (Supplemental Table 1). Of protocols used selectively, 225 (21.8%) were used in patients from two size categories, while 261 (25.3%) were used in patients from a single category: 23 (2.2%) were used only in small patients; 172 (16.7%) in medium patients; and 66 (6.4%) in large patients.

Variation in radiation dose and technical parameters by patient size category

The mean values for patient diameter, radiation dose metrics, and technical parameters overall, and stratified by the three size categories, are provided in Table 2. As expected, the unadjusted CTDIvol and DLP values increased across the categories, more than doubling between small and large. Effective dose also increased steadily between categories. By contrast, size-adjusted CTDIvol and size-adjusted DLP did not change meaningfully by size category, demonstrating that the radiation dose variation attributed to patient size was eliminated through size adjustment.

Table 2.

Radiation dose metrics and technical parameters overall and stratified by patient size category

Variable (units) Patient size category
All patients Small Medium Large
n = 748,846 n = 165,816 n = 360,850 n = 222,180
Mean Std Dev Mean Std Dev Mean Std Dev Mean Std Dev
Patient diameter (mm) 309 (45) 252 (17) 301 (16) 363 (28)
CTDIvol (mGy) 13 (7.2) 8 (3.3) 12 (4.9) 20 (7.8)
DLP (mGy-cm) 798 (546) 447 (283) 701 (404) 1217 (629)
Effective dose (mSv) 13 (8) 9 (6) 12 (7) 16 (9)
mAs 181 (89) 120 (55) 166 (68) 251 (94)
Pitch* 1.0 (0.3) 1.0 (0.2) 1.0 (0.3) 1.0 (0.3)
Effective mAs 183 (95) 118 (54) 166 (67) 260 (106)
kV (kilovoltage) 119 (7.5) 116 (8.0) 118 (7.4) 121 (6.8)
Scan length (cm) 48 (9.2) 45 (8.8) 48 (9.0) 51 (9.2)
Slice thickness (mm) 3.7 (1.6) 3.7 (1.7) 3.7 (1.6) 3.7 (1.6)
Number of phases 1.3 (0.7) 1.3 (0.7) 1.3 (0.7) 1.3 (0.6)
Calculated size-adjusted variables
Size-adjusted CTDIvol (mGy) 12 (4.6) 11 (5.0) 12 (4.6) 12 (4.2)
Size-adjusted DLP (mGy-cm) 702 (389) 690 (448) 701 (388) 712 (338)

Abbreviations: CTDIvol volumetric computed tomography dose index, DLP dose length product, mAs milliampere-seconds, mGy milliGray, mSv millisieverts

*Table distance traveled in one 360° gantry rotation divided by beam collimation

Some technical parameters also increased across the size categories: e.g., mAs and effective mAs were approximately twice the value in large as small patients, which suggests automatic exposure control (AEC) may have been used. Notably, average kV and scan length were virtually unchanged with size category, while the number of phases as expected did not change (Table 2).

Variation in radiation dose and technical parameters by patient size deciles

Mean dose metrics and technical parameters by decile of patient diameter are shown in Table 3. Each decile includes between 74,883 and 74,888 CT scans. Average patient diameter ranged from 238 mm for patients in the first decile to 395 mm in the highest decile, reflecting an average 6% increase per decile (Table 3). Again, we observed the expected increase in unadjusted CTDIvol and DLP by size, reflecting an average increase of 15% and 16% per decile, respectively (0.2% and –0.54% per decile in the corresponding size-adjusted dose metrics). Similarly, effective mAs increased with patient size (12% per decile), yet again there was minimal (1%) change per decile in kV (116 in the 1st compared to 122 in the 10th reflecting an average percent increase per decile = 0.54% [95% CI 0.53–0.54%]). Figure 1 show the distribution of effective mAs and kV by patient decile, demonstrating a consistent increase in effective mAs with patient diameter but no change in kV.

Table 3.

Radiation dose metrics and technical parameters for CTs sorted by decile in patient size based on effective patient diameter. Each decile includes between 74,883 and 74,888 patients and reflects the mean for all constituent patients. The mean percent increase per decile increase in effective patient diameter, and the percent increase between the 1st and 10th deciles of diameter, is shown for all variables with the 95% confidence interval. Because these reflect mean values, the results may reflect a number that is implausible for an individual scan (e.g., kV = 116). For the percent change, the numbers are rounded to the nearest 1% unless estimate < 10%, where numbers show with 2 significant digits

Decile abdomen diameter Patient diameter (mm) CTDIvol (mGy) Size-adjusted CTDIvol (mGy) DLP (mGy-cm) Size-adjusted DLP (mGy-cm) Effective Dose (mSv) Pitch* Effective mAs Kilovoltage (kV) Scan length (cm) Number of phases
Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI)
graphic file with name 330_2023_10076_Figa_HTML.gif 1st 238 (238,238) 7 (7,7) 12 (12,12) 403 (401,405) 709 (705,712) 9 (9,9) 1.04 (1.03,1.04) 109 (109,110) 116 (116,116) 45 (45,45) 1.4 (1.3,1.4)
2nd 262 (262,262) 8 (8,8) 11 (11, 11) 475 (473, 477) 675 (672, 678) 9 (9,10) 1.03 (1.03, 1.03) 124 (124, 124) 117 (117, 117) 46 (46,46) 1.3 (1.3, 1.3)
3rd 277 (277,277) 9 (9,9) 11 (11,11) 539 (537, 541) 675 (672, 678) 10 (9,10) 1.03 (1.03, 1.03) 137 (136, 137) 117 (117, 117) 47 (47,47) 1.3 (1.3, 1.3)
4th 289 (289,289) 10 (10,10) 12 (12, 12) 608 (606, 611) 687 (684, 690) 11 (11,11) 1.02 (1.02, 1.03) 149 (149, 150) 118 (118, 118) 47 (47,47) 1.3 (1.3, 1.3)
5th 300 (300,300) 11 (11,12) 12 (12, 12) 682 (679, 685) 699 (696, 702) 12 (12,12) 1.02 (1.02, 1.02) 163 (163, 164) 118 (118, 118) 48 (48,48) 1.3 (1.3, 1.3)
6th 311 (311,311) 13 (13,13) 12 (12, 12) 764 (762, 767) 712 (709, 715) 12 (12,13) 1.02 (1.02, 1.02) 178 (177, 178) 119 (119, 119) 49 (49,49) 1.3 (1.3, 1.3)
7th 323 (323,323) 14 (14,14) 12 (12, 12) 865 (862, 868) 727 (725, 730) 13 (13,13) 1.02 (1.02, 1.02) 195 (195, 196) 119 (119, 119) 49 (49,49) 1.3 (1.3, 1.3)
8th 337 (337,337) 16 (16,16) 12 (12, 12) 989 (986, 993) 737 (734, 740) 14 (14,15) 1.01 (1.01, 1.01) 217 (217, 218) 120 (120, 120) 50 (50,50) 1.3 (1.3, 1.3)
9th 356 (356,356) 19 (19,19) 12 (12, 12) 1152 (1148, 1156) 728 (726, 731) 16 (16,16) 1.01 (1.00, 1.01) 248 (247, 249) 120 (120, 120) 51 (51,51) 1.3 (1.3, 1.3)
10th 395 (395,395) 24 (24,24) 12 (12, 12) 1500 (1495, 1505) 672 (669, 674) 19 (19,19) 0.99 (0.99, 0.99) 312 (311, 313) 122 (122, 122) 52 (52,52) 1.3 (1.3, 1.3)
Percent change 1st to 10th deciles (95% CI)** 66% (66%, 66%) 241% (239%, 242%) 1.5%(1.1%, 1.9%) 272% (270%, 274%) -5.3% (-5.8% -4.7%) 126% (124%, 127%) -4.3% (-4.5%, -4.0%) 186% (185%, 187%) 4.9% (4.9%, 5.0%) 17%(17%, 17%) -4.7% (-5.2%, -4.3%)
Mean percent increase per decile (95% CI)** 5.8% (5.8%, 5.8%) 15% (15%,1 5%) 0.20% (0.15%, 0.24%) 16% (16%, 16%) -0.54% (-0.60%, -0.48%) 9.6% (9.5%, 9.7%) -0.48% (-0.51%, -0.45%) 12% (12%, 13%) 0.54% (0.53%, 0.54%) 1.7% (1.7%, 1.8%) -0.53% (-0.59%, -0.48

Abbreviations: CI confidence interval, CTDIvol volumetric computed tomography dose index, DLP dose length product, mAs milliampere-seconds, mGy milliGray, mSv millisieverts

*Table distance traveled in one 360° gantry rotation divided by beam collimation

**95% bootstrap confidence intervals using 1,000 bootstrap samples of size 748,846

Fig. 1.

Fig. 1

Effective mAs and kV (25th, 50th, and 75th percentiles in distribution) by decile in effective patient diameter

We replicated this analysis separately by deciles within each of the three size categories with similar results (Table 4 and Supplemental Table 2). Across the categories, the average kV was 116 in the 1st decile of small patients (average diameter = 217 mm) and 123 in the 10th decile of large patients (average diameter = 423 mm). The average change in kV was 0.1–0.3% between deciles. This contrasts with effective mAs where the average change in effective mAs between deciles was 20-fold higher, 3.0–6.0% per decile. The magnitude of the changes in the other radiation dose metrics and technical parameters was greatest among large patients (Table 4). For example, unadjusted CTDIvol and DLP both increased by 4% per decile in small patients and 7% in large patients (Table 4).

Table 4.

The average percent increase in patient diameter, dose metrics, and technical parameters per decile increase in patient size within each of the three size categories (small, medium, large). Results shown to the nearest 0.1%

Patient Diameter (mm) CTDIvol (mGy) Size-adjusted CTDIvol (mGy) DLP (mGy-cm) Size-adjusted DLP (mGy-cm) Effective dose (mSv) Effective mAs Kilovoltage (kV) Scan length (cm)
All Patients 5.8% 14.7% 0.2% 15.8% -0.5% 9.6% 12.5% 0.5% 1.7%
Patients by size category
Small 2.5% 3.7% -1.1% 3.8% -1.5% 2.9% 3.0% 0.1% 1.0%
Medium 1.9% 5.2% 0.8% 5.7% 0.8% 3.1% 4.2% 0.2% 0.6%
Large 2.6% 6.6% -1.0% 7.1% -1.6% 5.3% 6.0% 0.3% 0.7%

Abbreviations: CTDIvol volumetric computed tomography dose index, DLP dose length product, mAs milliampere-seconds

Variation in technical parameters by radiation dose deciles

The 1033 unique protocols were next sorted by median size-adjusted DLP, and the average (mean) values of technical parameters and radiation dose metrics for each decile in size-adjusted DLP are shown in Table 5. The mean size-adjusted DLP increased 535% between the lowest and highest deciles. This difference was not driven by patient size, as diameter changed little across the deciles (range 309–321 mm). Rather, high size-adjusted DLPs resulted from acquisition techniques, most notably effective mAs and the number of phases. The effective mAs increased from 125 to 266 from the lowest to highest decile (113% increase), and the average number of phases more than doubled from 1.1 to 2.6 across deciles (149% increase). The fact that size-adjusted CTDIvol increased considerably across the deciles (166%), but nowhere close to the rate of adjusted DLP (535%), suggests that phase number is a strong driver of the change in DLP.

Table 5.

Routine abdomen and pelvis CT protocols, sorted by size-adjusted dose length product (DLP). There are 103 or 104 protocols in each decile, and there are a mean of 288–1381 CTs per protocol. For each decile, the mean radiation dose metrics and values for the technical parameters are provided. The mean percent change in each technical parameter and radiation dose metric for decile increase in size-adjusted DLP is provided, as is the percent change from the 1st to the 10th deciles in size-adjusted DLP. Lastly, the expected increase in DLP per decile due to observed variation in parameter (effective mAs, kV, scan length, and number of phases) is provided to quantify which protocols are the most important contributors to the observed variation in size-adjusted DLP

Size-adjusted DLP decile Patient diameter (mm) DLP (mGy-cm) Size-adjusted DLP (mGy-cm) CTDIvol (mGy) Size-adjusted CTDIvol (mGy) Effective dose (mSv) Effective mAs Kilovoltage (kV) Scan length (cm) Number of phases
Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI) Mean (95% CI)
graphic file with name 330_2023_10076_Figb_HTML.gif 1st 309 (302,315) 295 (275,315) 265 (254,275) 7 (6,7) 6 (6,7) 4 (4,5) 125 (114,135) 115 (113,117) 42 (39,44) 1.1 (1.0,1.1)
2nd 306 (302,310) 387 (369,405) 366 (361,370) 8 (8,9) 8 (7,8) 6 (6,6) 141 (131,151) 116 (114,118) 47 (45,49) 1.0 (1.0,1.1)
3rd 309 (305,313) 477 (458,497) 441 (437,445) 10 (9,10) 9 (9,10) 7 (7,7) 167 (153,181) 117 (115,118) 48 (46,49) 1.0 (1.0,1.0)
4th 306 (303,310) 518 (499,538) 496 (492,499) 10 (10,11) 10 (10,10) 8 (8,9) 164 (151,177) 117 (115,118) 48 (46,49) 1.1 (1.0,1.2)
5th 308 (304,312) 614 (588,640) 563 (560,567) 12 (12,13) 12 (11,12) 10 (9,10) 185 (172,197) 119 (117,120) 47 (45,48) 1.1 (1.0,1.2)
6th 309 (304,314) 695 (657,732) 630 (627,634) 13 (12,14) 12 (12,13) 11 (11,11) 181 (170,192) 118 (117,120) 46 (45,48) 1.2 (1.1,1.3)
7th 313 (308,319) 825 (772,877) 712 (707,717) 15 (14,16) 13 (13,14) 13 (12,13) 209 (194,224) 119 (117,120) 45 (43,47) 1.4 (1.3,1.5)
8th 312 (307,317) 943 (886,1000) 818 (810,826) 16 (15,18) 15 (14,15) 15 (14,15) 214 (199,230) 120 (119,121) 45 (44,47) 1.4 (1.3,1.6)
9th 321 (312,330) 1297 (1188,1405) 1015 (1001,1030) 20 (17,22) 15 (14,16) 19 (18,20) 296 (252,340) 121 (119,122) 47 (45,49) 1.7 (1.6,1.9)
10th 312 (304,320) 1892 (1753,2032) 1680 (1576,1785) 20 (18,23) 17 (16,19) 30 (28,32) 266 (229,303) 119 (119,120) 44 (42,45) 2.6 (2.4,2.8)
Percent change 1st to 10th decile (95% CI)1 1.0% (-0.4%, 1.7%) 541% (513%, 562%) 535% (531%, 547%) 194% (174%, 205%) 166% (160%,176%) 592% (561%, 610%) 113% (96%, 118%) 4.0% (3.7%, 5.0%) 4.6% (1.6%, 5.6%) 149% (141%, 154%)
Mean percent increase per decile (95% CI%)1 0.1% (0.0%, 0.2%) 23% (23%, 24%) 24% (24%, 24%) 13% (12%, 13%) 12% (11%, 12%) 25% (24%, 25%) 9.6% (8.5%, 9.7%) 0.4% (0.4%, 0.6%) 0.6% (0.3%, 0.8%) 12% (11%, 12%)
Expected increase in DLP per decile due to observed variation in parameter 10% 1% 1% 12%

Abbreviations: CTDIvol volumetric computed tomography dose index, CI confidence interval, DLP dose length product, kV kilovoltage, mAs milliampere-seconds, mGy milliGray, mSv millisieverts

195% bootstrap confidence intervals using 1000 bootstrap samples of size 748,846

The contribution of each technical parameter to the observed 24% increase in size-adjusted DLP per decile is also shown in Table 5. For kV, the observed 0.4% increase in kV per decile is expected to result in only a small (1%) increase in size-adjusted dose. Because this is far smaller than the observed 24% change in size-adjusted dose, it suggests that kV was not a large contributor to between-protocols variation. Conversely, the larger observed increase in effective mAs (10%) and number of phases (12%) suggests these to be greater contributors to between-protocols dose variation, with number of phases being the most contributory.

This analysis was repeated, stratified within size category, producing similar results (Supplemental Table 3), with the range of size-adjusted DLP remarkably similar for all three size categories. For small patients, the size-adjusted DLP ranged from 259 to 1546 mGy-cm; medium patients 258–1705; and large patients 294–1549.

Impact of population dose optimization

The broad adoption of optimized protocol thresholds would result in population dose reductions of 40.0% (using a target of 433 mGy-cm) or 18.6% (using 645 mGy-cm) from current practice.

Discussion

We have described the technical parameters and dose metrics for almost 750,000 routine abdomen CT scans across diverse radiology practices. These show tremendous variation in radiation dose (535% in size-adjusted DLP) that is not driven by patient size but rather by how patients are scanned, particularly by manipulation of effective mAs and phase. The results also demonstrate that while automatic exposure control (AEC) is widely used to modify mAs, kV is almost never adjusted despite this being a demonstrated best practice [19, 27, 28]. Unindicated scan phases are the leading source of unnecessarily high radiation doses [29] and our results demonstrate high correlation of phase number with dose increases; protocols with the lowest size-adjusted doses used one scan phase, whereas those in the highest decile used an average of 2.6.

There are several approaches for dose optimization. One method is to develop protocols tailored to patient size, as it may be difficult for a single protocol to work across a large range of sizes even when AEC is available [1820]. In addition to adjusting mAs and kV, a well-optimized protocol would also adjust scan speed allowing dose adjustment for size. For example, if moving from a larger to a smaller patient protocol, in addition to using a lower kV and less mA, the helical pitch might increase and the tube rotation time decrease, which increases scan speed while lowering the maximum dose delivered [20]. However, we did not observe a systematic pattern of protocols being used in a size-specific manner, with only a quarter used in one size category. If this reflects a belief that multiple protocols are an administrative burden, approaches for simplified protocol management exist to ease this obstacle [30].

A second strategy is to tailor kV by patient size [31], which would lower doses particularly among smaller patients. Manufacturers have offered some scanners in recent years that make it easy to modify kV (such as Siemens’ Care KV, Canon’s SURE kV, and GE’s kV assist), but manual manipulation is and always has been available on all scanners. Yet we found kV is virtually never changed, even on machines that can automate kV selection. Reducing kV not only would reduce dose, but can also provide image quality advantages, such as improving contrast conspicuity in CT angiography, improved assessment of mural hyperenhancement in Crohn’s disease, and reducing contrast volume in a patient with chronic kidney disease. A third approach is to reduce unnecessary scan phases. While routine abdomen scans should be performed using a single-phase approach [21], we observed strong variation in number of phases, and more than half of scans used more than one phase.

The hypothetical dose optimization analyses demonstrate the magnitude of dose reduction that could be achieved (18.6–40.0%) if more practices adopted the optimized protocols widely used at some facilities. The target dose of 433 mGy-cm reflects a protocol around the 3rd decile in our distribution of size-adjusted DLP (where the mean effective mAs is 167, CTDIvol is 10 mGy) whereas the target dose of 645 mGy-cm reflects a target protocol around the 6th decile (mean effective mAs is 181 and CTDIvol 13 mGy).

The strengths of this study are its large sample size, inclusion of diverse practices, and detailed technical parameter data. There are several limitations. First, we classified scans as being done for routine abdomen indications based on the study description and protocol name in the DICOM data, and we have shown this approach to be 90% accurate compared to expert chart review [21]. Nonetheless, some scans will have been misclassified as routine, but this is unlikely to impact the primary conclusions. Second, we calculated size-adjusted DLP based only on patient diameter and did not adjust for patient height, and taller patients may be represented more among those who received higher dose examinations. However, this is expected to have contributed only minimally to the larger variation in dose, as scan length changed only modestly across the observed deciles in dose. Third, we have only focused on a single indication for CT (albeit the most common indication); future studies should explore other reasons for imaging, such as oncologic care. Lastly, we did not assess image quality in this study, and thus cannot assess the impact of the observed dose variation on radiologists’ satisfaction with quality. However, all scans were obtained for routine care and were deemed adequate for diagnosis when they were obtained.

This work shows, in the actual practice of almost 100 imaging facilities, a large variation in radiation doses exceeding that required by patient size or scanning indication, as well as widespread failure to implement best practices beyond use of AEC. Broader adoption of evidence-based practices including using size-specific protocols, increasing manipulation of kV, and eliminating multiphase protocols would reduce this variation and improve the safety of routine abdominal imaging.

Supplementary information

Below is the link to the electronic supplementary material.

Abbreviations

AEC

Automatic exposure control

CTDIvol

Volume computed tomography dose index

DICOM

Digital Imaging and Communications in Medicine

DLP

Dose length product

kV

Kilovolt

mAs

Milliampere seconds

mGy

Milligray

Funding

This study received funding from the US National Institutes of Health (R01-CA181191) and the Patient-Centered Outcomes Research Institute (DI-2018C1-11375). Funders had no role in study design, collection, analysis, interpretation, and reporting of data; or decision to publish. The views in this article are solely the responsibility of the authors and do not necessarily represent the views of the Patient-Centered Outcomes Research Institute, its Board of Governors or Methodology Committee or other funders.

Declarations

Guarantor

The scientific guarantor of this publication is Dr. Rebecca Smith-Bindman.

Conflict of interest

Dr. Smith-Bindman is a co-founder of Alara Imaging, Inc., a company focused on improving the clinical and operational aspects of health systems including collecting and reporting radiation dose and image quality information associated with CT as part of payer-led quality programs. Alara Imaging played no role in any aspect of the paper, and the work does not overlap with Alara’s commercial activities. Dr. Smith Bindman has never received funding from Alara.

Statistics and biometry

Four biostatisticians included as coauthors on the manuscript contribute to the design and statistical analysis of the study including Dr. Kang, Mr. Chu, Dr. Wang, and Dr. Chung.

Informed consent

Written informed consent was not required for this study because the UCSF Committee on Human Research approved this study with a waiver of individual informed consent. Collaborating institutions obtained local Institutional Review Board approval or relied on the UCSF approval to contribute data to the registry and also waived the need for written informed consent.

Ethical approval

The UCSF Committee on Human Research approved this study. Collaborating institutions obtained local Institutional Review Board approval or relied on the UCSF approval to contribute data to the registry.

Study subjects

The data derive from The UCSF International CT dose registry, and some of the patients described in this report may have also been included in prior publications including the following referenced publications.

Bos D, Yu S, Luong J et al (2021) Diagnostic reference levels and median doses for common clinical indications of CT: findings from an international registry. Eur Radiol. https://doi.org/10.1007/s00330-021-08266-1

Smith-Bindman R, Wang Y, Chu P et al (2019) International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ 364:k4931

Smith-Bindman R, Yu S, Wang Y et al (2022) An image quality-informed framework for CT Characterization. Radiology 302:380-389

Chu PW, Yu S, Wang Y et al (2021) Reference phantom selection in pediatric computed tomography using data from a large, multicenter registry. Pediatr Radiol. https://doi.org/10.1007/s00247-021-05227-0

However, the aim of this work is to understand specific quality practices for routine abdomen CT and is unrelated to previous studies. There is no overlap in the methodology.

Methodology

• Prospective

• Observational

• Multicenter study

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.IMV (2020) 2019 CT Market Outlook Report CT contrast media utilization reaches a crossroads. IMV Medical Information Division, Des Plaines, p 2020
  • 2.Eurostat-Statistics Explained (2022) Healthcare resource statistics – technical resources and medical technology. Table 3 Use of Imaging Equipment. Available via https://ec.europa.eu/eurostat/statistics-explained/index.php?title=Healthcare_resource_statistics_-_technical_resources_and_medical_technology&oldid=452031#Use_of_medical_technology. Accessed 29 Sep 2022
  • 3.Amis ES, Jr, Butler PF, Applegate KE, et al. American College of Radiology white paper on radiation dose in medicine. J Am Coll Radiol. 2007;4:272–284. doi: 10.1016/j.jacr.2007.03.002. [DOI] [PubMed] [Google Scholar]
  • 4.European Commission (2014) Medical Protection No 180. Medical Radiation Exposure of the European Population. Part 1/2. Publications Office of the European Union, Luxembourg
  • 5.Frija G, Damilakis J, Paulo G, Loose R, Vano E, European Society of Radiology (ESR) Cumulative effective dose from recurrent CT examinations in Europe: proposal for clinical guidance based on an ESR EuroSafe Imaging survey. Eur Radiol. 2021;31:5514–5523. doi: 10.1007/s00330-021-07696-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Frija G, Hoeschen C, Granata C, et al. ESR EuroSafe Imaging and its role in promoting radiation protection - 6 years of success. Insights Imaging. 2021;12:3. doi: 10.1186/s13244-020-00949-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Goske MJ, Applegate KE, Boylan J, et al. Image Gently(SM): a national education and communication campaign in radiology using the science of social marketing. J Am Coll Radiol. 2008;5:1200–1205. doi: 10.1016/j.jacr.2008.06.007. [DOI] [PubMed] [Google Scholar]
  • 8.International Atomic Energy Agency Report of a consultation on justification of patient exposures in medical imaging. Radiat Prot Dosimetry. 2009;135:137–144. doi: 10.1093/rpd/ncp107. [DOI] [PubMed] [Google Scholar]
  • 9.Tsapaki V, Damilakis J, Paulo G, et al. CT diagnostic reference levels based on clinical indications: results of a large-scale European survey. Eur Radiol. 2021;31:4459–4469. doi: 10.1007/s00330-020-07652-5. [DOI] [PubMed] [Google Scholar]
  • 10.Vano E, Miller DL, Martin CJ, et al. ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP. 2017;46:1–144. doi: 10.1177/0146645317717209. [DOI] [PubMed] [Google Scholar]
  • 11.Vano E, Loose R, Frija G, et al. Notifications and alerts in patient dose values for computed tomography and fluoroscopy-guided interventional procedures. Eur Radiol. 2022;32:5525–5531. doi: 10.1007/s00330-022-08675-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Bos D, Yu S, Luong J, et al. Diagnostic reference levels and median doses for common clinical indications of CT: findings from an international registry. Eur Radiol. 2021 doi: 10.1007/s00330-021-08266-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Smith-Bindman R, Wang Y, Chu P, et al. International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ. 2019;364:k4931. doi: 10.1136/bmj.k4931. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kanal KM, Butler PF, Chatfield MB, et al. U.S. diagnostic reference levels and achievable doses for 10 pediatric CT examinations. Radiology. 2022;302:164–174. doi: 10.1148/radiol.2021211241. [DOI] [PubMed] [Google Scholar]
  • 15.Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. U.S. diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology. 2017;284:120–133. doi: 10.1148/radiol.2017161911. [DOI] [PubMed] [Google Scholar]
  • 16.Kubo T, Ohno Y, Seo JB, et al. Securing safe and informative thoracic CT examinations-progress of radiation dose reduction techniques. Eur J Radiol. 2017;86:313–319. doi: 10.1016/j.ejrad.2016.10.012. [DOI] [PubMed] [Google Scholar]
  • 17.Lira D, Padole A, Kalra MK, Singh S. Tube potential and CT radiation dose optimization. AJR Am J Roentgenol. 2015;204:W4–10. doi: 10.2214/AJR.14.13281. [DOI] [PubMed] [Google Scholar]
  • 18.Mayo-Smith WW, Hara AK, Mahesh M, Sahani DV, Pavlicek W. How I do it: managing radiation dose in CT. Radiology. 2014;273:657–672. doi: 10.1148/radiol.14132328. [DOI] [PubMed] [Google Scholar]
  • 19.Yu L, Li H, Fletcher JG, McCollough CH. Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys. 2010;37:234–243. doi: 10.1118/1.3264614. [DOI] [PubMed] [Google Scholar]
  • 20.Szczykutowicz TP, Bour RK, Rubert N, Wendt G, Pozniak M, Ranallo FN. CT protocol management: simplifying the process by using a master protocol concept. J Appl Clin Med Phys. 2015;16:228–243. doi: 10.1120/jacmp.v16i4.5412. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Smith-Bindman R, Yu S, Wang Y, et al. An image quality-informed framework for CT characterization. Radiology. 2022;302:380–389. doi: 10.1148/radiol.2021210591. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Chu PW, Yu S, Wang Y, et al. Reference phantom selection in pediatric computed tomography using data from a large, multicenter registry. Pediatr Radiol. 2021 doi: 10.1007/s00247-021-05227-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.American Association of Physicists in Medicine Report No. 204 - Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. American Association of Physicists in Medicine, College Park, MD [DOI] [PubMed]
  • 24.McCollough C, Bakalyar DM, Bostani M, et al. Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: the report of AAPM Task Group 220. AAPM Rep. 2014;2014:6–23. [PMC free article] [PubMed] [Google Scholar]
  • 25.Hsieh J. Analytical models for multi-slice helical CT performance parameters. Med Phys. 2003;30:169–178. doi: 10.1118/1.1533750. [DOI] [PubMed] [Google Scholar]
  • 26.Bos D, Yu S, Luong J, et al. Diagnostic reference levels and median doses for common clinical indications of CT: findings from an international registry. Eur Radiol. 2022;32:1971–1982. doi: 10.1007/s00330-021-08266-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Kalender WA, Deak P, Kellermeier M, van Straten M, Vollmar SV. Application- and patient size-dependent optimization of X-ray spectra for CT. Med Phys. 2009;36:993–1007. doi: 10.1118/1.3075901. [DOI] [PubMed] [Google Scholar]
  • 28.Yu L, Liu X, Leng S, et al. Radiation dose reduction in computed tomography: techniques and future perspective. Imaging Med. 2009;1:65–84. doi: 10.2217/iim.09.5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Guite KM, Hinshaw JL, Ranallo FN, Lindstrom MJ, Lee FT., Jr Ionizing radiation in abdominal CT: unindicated multiphase scans are an important source of medically unnecessary exposure. J Am Coll Radiol. 2011;8:756–761. doi: 10.1016/j.jacr.2011.05.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Szczykutowicz TP, Rubert N, Belden D, et al. A Wiki-based solution to managing your institution's imaging protocols. J Am Coll Radiol. 2016;13:822–824. doi: 10.1016/j.jacr.2015.10.021. [DOI] [PubMed] [Google Scholar]
  • 31.Hsieh J, Society of Photo-optical Instrumentation E (2022) Computed tomography : principles, design, artifacts, and recent advances, Fourth edition. edn. SPIE, Bellingham, Washington (1000 20th St. Bellingham WA 98225–6705 USA)

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from European Radiology are provided here courtesy of Springer

RESOURCES