Abstract
Purpose
Sodium MRI can be used to quantify tissue sodium concentration (TSC) in vivo; however, UTE sequences are required to capture the rapidly decaying signal. 2D MRI enables high in‐plane resolution but typically has long TEs. Half‐sinc excitation may enable UTE; however, twice as many readouts are necessary. Scan time can be minimized by reducing the number of signal averages (NSAs), but at a cost to SNR. We propose using compressed sensing (CS) to accelerate 2D half‐sinc acquisitions while maintaining SNR and TSC.
Methods
Ex vivo and in vivo TSC were compared between 2D spiral sequences with full‐sinc (TE = 0.73 ms, scan time ≈ 5 min) and half‐sinc excitation (TE = 0.23 ms, scan time ≈ 10 min), with 150 NSAs. Ex vivo, these were compared to a reference 3D sequence (TE = 0.22 ms, scan time ≈ 24 min). To investigate shortening 2D scan times, half‐sinc data was retrospectively reconstructed with fewer NSAs, comparing a nonuniform fast Fourier transform to CS. Resultant TSC and image quality were compared to reference 150 NSAs nonuniform fast Fourier transform images.
Results
TSC was significantly higher from half‐sinc than from full‐sinc acquisitions, ex vivo and in vivo. Ex vivo, half‐sinc data more closely matched the reference 3D sequence, indicating improved accuracy. In silico modeling confirmed this was due to shorter TEs minimizing bias caused by relaxation differences between phantoms and tissue. CS was successfully applied to in vivo, half‐sinc data, maintaining TSC and image quality (estimated SNR, edge sharpness, and qualitative metrics) with ≥50 NSAs.
Conclusion
2D sodium MRI with half‐sinc excitation and CS was validated, enabling TSC quantification with 2.25 × 2.25 mm2 resolution and scan times of ≤5 mins.
Keywords: compressed sensing, half‐sinc, 23Na MRI, sodium, UTE
1. INTRODUCTION
Tissue sodium concentration (TSC) is a potential quantitative biomarker of diseases that affect sodium homeostasis (e.g., chronic kidney disease 1 and hypertension 2 ). Sodium (23Na) MRI is a noninvasive method of quantifying TSC and can detect abnormal TSC in leg skin or muscle of patients with multiple sclerosis, 3 salt‐losing tubulopathies, 4 and muscular dystrophies. 5 However, 23Na MRI is limited by two main challenges: (i) poor SNR, due to low in vivo sodium concentrations and low intrinsic sensitivity; and (ii) rapid biexponential signal decay, which necessitates UTE sequences.
Typically, 3D sequences are used for 23Na MRI because the TE can be minimized using nonselective hard pulses. 6 , 7 , 8 , 9 However, in some applications (e.g., skin 10 or heart 11 ), 2D imaging may be preferable to enable higher in‐plane resolution, shorter scan times, or quicker image reconstruction. The main disadvantage of 2D imaging is longer TEs, due to the requirement for slice‐select and slice‐select‐rewinder gradients.
Half‐pulse RF excitations 12 can potentially overcome this but are rarely used in sodium imaging. 13 , 14 , 15 In this approach, two acquisitions are performed with the same readout gradients but half‐sinc excitations with slice‐select gradients of opposite polarity (Figure 1). The two acquisitions are subsequently added to give the same slice profile as the full‐sinc excitation. Although half‐sinc excitation results in a significant reduction in TE, overall scan time is doubled. This can be compensated by reducing the number of signal averages (NSAs), at a cost to SNR. Recently, compressed sensing (CS) has been shown to improve SNR in sodium imaging, 16 , 17 , 18 but only a small number of studies have applied it in the leg. 19 , 20
The focus of this study was to develop a high‐resolution 2D UTE–23Na MRI sequence with CS reconstruction for rapid and accurate TSC quantification in skeletal muscle and skin. Specifically, the aims were: (1) to develop a 2D UTE sequence with half‐sinc excitation for sodium imaging, (2) to optimize a CS reconstruction for accurate sodium quantification with reduced scan time, and (3) to validate the developed techniques ex vivo and in vivo by assessing TSC accuracy and image quality.
2. METHODS
2.1. Sequence design
We developed a 2D uniform‐density center‐out spiral sodium sequence using the spiral algorithm by Pipe and Zwart. 21 Sequence parameters were: FOV = 180 × 180 mm2, pixel size = 2.25 × 2.25 mm2, slice thickness = 30 mm, TR = 100 ms, spiral readout duration = 7.6 ms, 20 regularly spaced spirals for complete k‐space filling, and 150 NSAs.
2.2. Full‐sinc and half‐sinc 2D excitation pulses
A single‐lobe, full‐sinc RF pulse was designed (flip angle = 90°, pulse duration = 600 μs), resulting in TE = 0.73 ms. Splitting this pulse for half‐sinc excitation (pulse duration = 300 μs) resulted in TE = 0.23 ms (Figure 1). Total scan time was 5:07 min:s and 10:15 min:s for 2D full‐sinc and half‐sinc acquisitions, respectively.
2.3. Imaging experiments
All imaging was performed on a multinuclear 3 T MRI scanner (Biograph mMR, Siemens Healthineers, Erlangen, Germany). A transmit/receive 23Na birdcage coil (Stark Contrast, Erlangen, Germany) was positioned around the tissue. Five calibration phantoms with nominal sodium concentrations of 10, 25, 50, 75, and 100 mM were placed on top of the tissue within the coil and image FOV. Phantoms were fixed in parallel foam sleeves, and care was taken to ensure they remained orthogonal to the imaging plane. Manual B0‐field shimming and flip‐angle calibration were performed separately for 2D and 3D acquisitions within a shim volume covering the tissue and calibration phantoms in the imaging volume.
2.3.1. Ex vivo imaging
In three ex vivo porcine samples, with one sample repeated twice, 2D sodium acquisitions with full‐sinc and half‐sinc excitation were compared to a reference 3D acquisition‐weighted stack‐of‐spirals sequence 22 with a nonselective excitation pulse (duration = 250 μs). The spiral trajectories, in‐plane FOV, in‐plane resolution, and TR were the same as the 2D sequences. Other 3D sequence parameters were slice thickness = 15 mm, slice partitions = 6, TE = 0.22 ms, NSAs = 120, total scan time = 24 min. Muscle TSC was quantified by averaging over two consecutive slices for sufficient SNR and to match the slice thickness of the 2D experiments.
2.3.2. In vivo imaging
Ten healthy volunteers were enrolled in the study (5 male, 5 female; mean age 30 ± 6 years) and provided written consent. This study was approved by the local research ethics committee (Ref. 15/0041). Sodium imaging was performed on the right calf using 2D acquisitions with full‐sinc and half‐sinc excitation and the same setup as the ex vivo experiment. TSC was quantified in the muscle and skin.
2.4. Image reconstruction
Conventional reconstruction for 2D and 3D acquisitions consisted of nonuniform fast Fourier transform (NUFFT) with density compensation. 23 , 24
In vivo 2D half‐sinc data with retrospectively reduced NSAs was also reconstructed using CS by solving the following minimization problem 25 :
(1) |
Where is the desired image, is the encoding matrix (which includes Fourier transform and sampling matrix), is the measured data, is a sparsifying transform, and is a regularization weighting parameter. The first term enforces data consistency, whereas the second term promotes sparsity in a transform domain.
Optimal regularization was investigated by reconstructing reduced NSAs data (15, 30, 40, 50, 60, 75) at different values of (0.0, 0.1, 0.3, 0.5, 0.7, 1.0, 1.3, 1.6, 3.2, 5.0), using a total variation transform. Optimal for a given NSAs was chosen by comparing the SD of pixels in a muscle region of interest (ROI) in TSC maps from CS (SDCS) with reference 150 NSAs NUFFT maps (SDRef), using SDCS/SDRef. Specifically, for a given NSAs, the optimal was taken where SDCS/SDRef was closest to 1.0 (i.e., the most similar pixel distribution to the reference map), minimizing noise (observed where SDCS/SDRef >1) and excess smoothing (observed where SDCS/SDRef <1).
All CS reconstructions were developed in‐house using Python (version 3.8.10) and TensorFlow MRI (version 0.21.0). 24
2.5. TSC quantification
ROIs were drawn in the center of each reference phantom (to avoid partial volume effects), and a linear calibration was performed (Supporting Information S1). TSC maps were computed in MatLab (R2022a, MathWorks, Inc., Natick, MA), and results are reported as apparent TSC (aTSC), 26 quantified as the mean value from ROIs drawn using ITK‐SNAP (version 3.8.0) 27 (Supporting Information S2).
Due to the rapid decay of the sodium signal, TE can have a significant effect on aTSC accuracy. The sodium signal over time, t, can be described by: 28 , 29
(2) |
Where are the relative proportions of the short T2 () and long T2 () components. This equation can be used to model the effect of calibration phantom and tissue relaxation on signal amplitude, as well as to perform relaxation correction of the acquired signal (where t = TE). In silico modeling used normalized phantom and muscle signals, and the ratio of the two was used as a proxy for aTSC (with a ratio of 1 being most accurate). Because different tissue types and tissues affected by disease may not exhibit the same relaxation properties, it is to be expected that properties of calibration phantoms and tissues may not exactly match. However, relaxation correction can be performed if these parameters are known. Properties of the phantoms used for these experiments have been reported by Rot et al. 30 : = 0.7:0.3, = 4–6 ms, = 17–24 ms, and = 27–32 ms. However, measurement of individual tissue relaxation parameters was not feasible; hence, published values were used. Specifically, for demonstration, correction was performed on ex vivo data using literature values for muscle 31 : = 0.6:0.4, = 1.5–2.5 ms, = 15–30 ms, and = 12–25 ms 30 ; results are reported with and without relaxation correction and compared to modeled signals. Phantom and muscle relaxation properties are described further in Supporting Information S3.
2.6. Image quality assessment
In vivo quantitative and qualitative metrics were measured to assess the effect of reduced NSAs and CS on the quality of 2D half‐sinc images. Quantitative image quality was assessed using estimated SNR and edge sharpness (ES) measures. Calculation of SNR in images reconstructed using CS is nontrivial due to the uneven distribution of noise. Therefore, SNR was estimated in the images using mean signal intensity in an ROI in tissues of interest and SD of noise from a background ROI (Supporting Information S2). 32
ES was calculated by measuring the maximum gradient of pixel intensities across the 100 mM phantom border on aTSC maps (chosen as the phantom with the highest signal, limiting artificially high edges caused by noise). Pixel intensities were filtered using a Savitzky–Golay filter to remove the effect of noise (window width = 15 pixels, second‐order polynomial) before differentiation. ES was taken as the maximum gradient of the filtered pixel intensities (Supporting Information S2) and was calculated in four positions around the phantom, with the average value used for comparison. Estimated SNR and ES were computed in MatLab (MathWorks).
Qualitative image quality was assessed in two categories using a 5‐point Likert scale: ability to identify separate structures of the leg, such as skin and muscle (1 = very difficult, 2 = difficult, 3 = okay, 4 = easy, 5 = very easy), and perceptual noise (1 = very high, 2 = high, 3 = moderate, 4 = low, 5 = very low). Scoring was performed independently by two observers (v.m. and m.r.), and images were fully anonymized and presented in a random order.
2.7. Statistical analysis
In vivo full‐sinc and half‐sinc aTSC were compared using a paired t‐test. aTSC and image quality metrics from half‐sinc acquisitions with fewer NSAs were compared between NUFFT/CS reconstructed data and reference data using analysis of variance, followed by post hoc comparisons using Tukey's test. All data are presented as mean ± SD. Results were considered statistically significant for p < 0.05. All statistical analysis was performed using RStudio (Version 2022.7.2.576, RStudio, PBC, Boston, MA, USA).
3. RESULTS
23Na MRI data was successfully collected with full‐sinc and half‐sinc excitation, both ex vivo (in addition to the 3D sequence, N = 4) and in vivo (N = 10).
3.1. Comparison of full‐sinc and half‐sinc excitation
3.1.1. Ex vivo imaging
Figure 2A shows ex vivo images for all three acquisition sequences. aTSC in porcine muscle was calculated as 25 ± 1 mM from 3D and 2D half‐sinc data. However, 2D full‐sinc data resulted in lower aTSC values of 24 ± 1 mM (Figure 2B). When relaxation correction was performed, minimal changes were observed in aTSC for 3D (+0.1 mM) and half‐sinc data (+0.2 mM). However, the full‐sinc signal correction resulted in an increase of 2.0 mM in aTSC (Figure 2B), resulting in higher aTSC than 3D and half‐sinc data.
In silico modeling of sodium signals for the phantoms and muscle (Figure 2D) enabled the effect of different relaxation properties to be compared. At sequence‐specific TEs, the phantom/muscle signal ratio for the 3D, 2D full‐sinc, and half‐sinc acquisitions was 1.00, 0.93, and 0.99, respectively (Figure 2E). This translates to a median expected aTSC difference compared to the 3D data of 7.5% (range: 2.5%–24.0%) and 0.2% (range: 0.1%–0.3%) with the full‐sinc and half‐sinc, respectively, suggesting the bias observed in ex vivo full‐sinc data may be due to differences in TE.
3.1.2. In vivo imaging
Similarly, in vivo aTSC was found to be significantly higher (p < 0.001) when quantified from half‐sinc compared to full‐sinc acquisitions (22 ± 2 and 19 ± 3 mM vs. 20 ± 2 and 17 ± 4 mM, for muscle and skin, respectively) (Figure 2C). This 10% difference is in agreement with the in silico model, suggesting the lower TE of half‐sinc excitation results in more accurate aTSC where relaxation correction is not feasible. Therefore, only half‐sinc data was carried forward for CS reconstruction (although it could be equally applied to full‐sinc data).
3.2. Optimization of compressed sensing on half‐sinc data
Reducing NSAs in in vivo half‐sinc data resulted in higher noise in NUFFT images (Figure 3A) (Supporting Information S4). Figure 4A,B (and Supporting Information S5) show the effect of CS with different regularization weighting on the distribution of aTSC within the muscle ROI. Although increasing reduces the width of the histogram of the muscle pixels (images become less noisy), very high results in excessive reduction in pixel variability, an over‐smoothed image, and a bias in muscle aTSC. The optimal values for different NSAs are presented in Figure 4C. Example images reconstructed with optimal across the range of NSAs and images reconstructed across the range are presented in Supporting Information S4 and S6, respectively.
CS reconstruction with optimal (CS‐) produced good image quality with as few as 50 NSAs (Figure 3) (Supporting Information S4). Therefore, images reconstructed with both 75 NSAs (halving of scan time compared to the 150 NSAs reference) and 50 NSAs were taken forward for muscle and skin aTSC quantification and image quality assessment.
3.3. aTSC quantification and image quality assessment
In vivo aTSC calculated within muscle and skin, as well as quantitative (estimated SNR and ES) and qualitative image quality metrics, are presented in Table 1. No significant differences in aTSC were observed between reference maps and maps with either 50 or 75 NSAs when reconstructed with the NUFFT or CS‐ reconstruction (p > 0.9).
TABLE 1.
75 NSAs | 50 NSAs | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Metric | Reference (150 NSAs) | NUFFT | CS‐ | p (Ref vs. NUFFT) | p (Ref vs. CS‐) | p (NUFFT vs. CS‐) | NUFFT | CS‐ | p (Ref vs. NUFFT) | p (Ref vs. CS‐) | p (NUFFT vs. CS‐) | ||
Accuracy | aTSC [mM] | Muscle | 21.5 ± 1.9 | 21.7 ± 2.0 | 22.1 ± 2.0 | 1.0 | 0.99 | 1.0 | 21.7 ± 2.0 | 22.2 ± 2.0 | 1.0 | 0.98 | 1.0 |
Skin | 18.8 ± 3.2 | 18.8 ± 3.3 | 18.2 ± 3.4 | 1.0 | 1.0 | 1.0 | 19.0 ± 3.2 | 18.1 ± 3.3 | 1.0 | 1.0 | 0.99 | ||
Quantitative image scores | Estimated SNR | Muscle | 7.0 ± 0.7 | 5.2 ± 0.5 | 7.5 ± 0.9 | <0.001 | 0.64 | <0.001 | 4.4 ± 0.4 | 6.5 ± 0.7 | <0.001 | 0.50 | <0.001 |
Skin | 6.1 ± 1.0 | 4.5 ± 0.7 | 6.1 ± 1.2 | 0.002 | 1.0 | 0.002 | 3.9 ± 0.5 | 5.3 ± 0.8 | <0.001 | 0.26 | 0.008 | ||
Edge sharpness [mM mm−1] | 12.7 ± 0.5 | 12.7 ± 0.5 | 12.7 ± 0.5 | 1.0 | 1.0 | 1.0 | 12.8 ± 0.5 | 12.7 ± 0.4 | 1.0 | 1.0 | 1.0 | ||
Qualitative image scores | Ability to identify separate structures of the leg | 3.2 ± 0.9 | 3.0 ± 0.7 | 3.8 ± 0.4 | 0.84 | 0.10 | 0.003 | 2.6 ± 0.5 | 3.5 ± 0.7 | 0.03 | 0.84 | <0.001 | |
Perceptual noise | 3.8 ± 0.4 | 3.1 ± 0.4 | 3.8 ± 0.4 | <0.001 | 1.0 | 0.002 | 2.6 ± 0.5 | 3.5 ± 0.7 | <0.001 | 0.26 | <0.001 |
Note: Reference = NUFFT reconstructed half‐sinc data with 150 NSAs (scan time = 10:15 min:s). 75 NSAs scan time = 5:07 min:s, 50 NSAs scan time = 3:25 min:s. Data is presented as mean ± SD.
Abbreviations: , regularization weighting parameter; aTSC, apparent tissue sodium concentration; CS‐, compressed sensing with optimized ; NSAs, number of signal averages; NUFFT, nonuniform fast Fourier transform.
Estimated SNR was similar in CS‐ images and reference images (p > 0.2), and both were significantly higher than NUFFT images (p ≤ 0.008). No significant difference in ES was observed between CS‐, NUFFT, and reference aTSC maps (p = 1.0).
Example images with assigned scores from the qualitative scoring are presented in Supporting Information S7. CS‐ images scored significantly higher than NUFFT images for perceptual noise (indicating less noise, p ≤ 0.002), but no difference was observed compared to reference images (p ≥ 0.2). Similarly, for the ability to identify separate structures of the leg, CS‐ images scored significantly higher than NUFFT images (p ≤ 0.003), whereas no significant difference was measured with the reference (p ≥ 0.1).
4. DISCUSSION
In this study, we validated the use of half‐sinc excitation for 2D spiral UTE 23Na MRI reconstructed using CS. The main findings of this study were: (i) half‐sinc excitation led to improved accuracy in aTSC, ex vivo and in vivo; (ii) CS reconstruction improved qualitative and quantitative image quality of reduced NSAs data in vivo; and (iii) aTSC quantification was unaffected with reduced NSAs, irrespective of reconstruction (CS or NUFFT).
Ex vivo aTSC calculated without relaxation correction from 2D half‐sinc data was equivalent to that obtained from reference 3D UTE acquisitions, whereas full‐sinc data resulted in lower aTSC. This is expected due to the longer TE of the full‐sinc acquisition and minor differences in relaxation between calibration phantoms and tissue, and was supported by in silico modeling. Performing relaxation correction led to a minor increase in aTSC for half‐sinc and 3D data (≤0.2 mM). However, the full‐sinc correction was much larger (+2.0 mM), which may represent an overcorrection because the true tissue T1 and T2 are unknown. Unfortunately, T1 and T2 mapping is extremely time consuming and not feasible in a clinical setting. 33 , 34 , 35 Consequently, the shorter TE of the half‐sinc acquisition, which minimizes T2 weighting and any resulting bias, is preferable to ensure greater aTSC quantification accuracy. For all techniques, good reproducibility was observed with a repeated measurement of one ex vivo sample (<0.4 mM difference in aTSC) (Figure 2B) and with calibration plots in all ex vivo samples (N = 4) (Supporting Information S1), with slightly better goodness of fit and reproducibility observed for half‐sinc data.
The main problem with half‐sinc acquisitions is the doubling of scan time; however, we have shown CS can be used to reconstruct images with fewer NSAs. Previous studies have investigated the use of CS in 3D 23Na MRI to remove undersampling artifacts, with early work showing underestimation of aTSC with CS reconstructions. 18 More recent studies have demonstrated aTSC accuracy, with an undersampling factor up to 4 in the brain 17 and leg. 20 In this study, we found undersampling of data (with full NSAs) resulted in more residual artifacts than reduced NSAs (with full sampling; Supporting Information S8).
A vital aspect of CS is choosing the correct regularization weighting, with previous studies choosing based on visual inspection or structural similarity index (SSIM). 17 In this study, we investigated methods based on pixel distribution in the aTSC maps, pixel distribution in the images, and SSIM (Supporting Information S9). Although the optimal regularization weighting was similar for all methods, matching pixel distribution in aTSC maps to the reference map was chosen because aTSC is the main clinical metric calculated from sodium images and SSIM has been shown to introduce contrast bias, 36 which may result in errors in aTSC quantification. Increasing above the optimal level resulted in images becoming over‐smoothed, and a bias and loss of variability in aTSC. Compared to standard NUFFT images, CS did not change muscle or skin aTSC quantification. Nevertheless, estimated SNR was found to be improved compared to NUFFT images, whereas ES remained equivalent. These results demonstrate the denoising effect of CS without introducing image blurring, which is important when quantifying aTSC in small or thin structures such as the skin.
One of the main advantages of 2D imaging over 3D imaging is the ability to achieve higher in‐plane resolution and/or shorter scan times. All sequences used in this work have a higher spatial resolution (2.25 mm2) than typical sodium imaging (3–4 mm2), 37 , 38 but the half‐sinc 2D sequence has a significantly reduced acquisition time (≈10 min, with 150 NSAs) compared to the 3D sequence (24 mins). It should be noted that if images were acquired with a more conventional resolution (3–4 mm2), and equivalent SNR, the scan time for the 2D half‐sinc sequence could be reduced to ≈3 min (with 150 NSAs), and the 3D to ≈∼7.5 min.
4.1. Limitations
For the CS optimization, a 150 NSAs NUFFT 2D half‐sinc image was used as a reference. Although this image was visibly noisy, it provided the best estimate of in vivo aTSC values while reasonably limiting volunteer scan time. For clinical studies, true aTSC within tissue is not easily measured; however, in the future, preclinical studies may be conducted for full validation of aTSC accuracy. 39
Accurate quantification of skin aTSC can be challenging due to its thin structure and partial volume effects caused by low spatial resolution. Despite increased spatial resolution compared to typical sodium imaging at 3 T, 37 , 38 skin aTSC was still lower than values reported at higher field strengths. 40 It should also be noted that drawing ROIs directly onto sodium images may introduce bias; it may be more accurate to draw ROIs on registered high‐resolution proton images. Exact registration, however, can be challenging.
Another limitation of this method is the thick slices used. This is reasonable in the calf, where anatomical structure changes are small along the slice axis, and enables increased SNR and consequently higher in‐plane resolution. However, it may not be suitable for other applications. In this case, slice thickness can be reduced, and in‐plane resolution decreased to maintain SNR. Furthermore, this study only acquired a single 2D slice through the calf. For imaging other anatomy, such as kidney or brain, full coverage of the organ may be required, thus necessitating the use of 2D multi‐slice or 3D methods.
5. CONCLUSION
2D 23Na MRI using half‐sinc excitation and CS enables muscle and skin aTSC quantification in vivo in the calf within a reasonable scan time. Reducing scan time to ≤5 min will improve patient comfort and compliance and enables inclusion of 23Na MRI as part of a comprehensive quantitative MRI protocol while keeping the overall scan time short. By improving image quality and reducing the likelihood of motion artifacts while maintaining high resolution, this technique has the potential to improve diagnostic accuracy and monitoring of patients with diseases affecting sodium homeostasis.
FUNDING INFORMATION
r.r.b. is funded by the University College London Hospitals (UCLH) Biomedical Research Centre (BRC) (grant BRC870/CM/JS/10132). j.m.t. and j.a.s. are funded by the UK Research and Innovation (UKRI) Future Leaders Fellowship (grant MR/S032290/1). s.r. is supported by the Engineering and Physical Sciences Research Council (EPSRC)‐funded University College London (UCL) Centre for Doctoral Training in Intelligent, Integrated Imaging in Healthcare (i4health) (grant EP/S021930/1), and the Department of Health's National Institute for Health and Care Research (NIHR)‐funded BRC at UCLH. b.s.s. is supported by Wings for Life (grant 169111). c.g.w.k. receives funds from Horizon2020 (Research and Innovation Action Grants Human Brain Project [SGA3]) (grant 945539), the BRC (grant BRC704/CAP/CGW), the Medical Research Council (MRC) (grant MR/S026088/1), Ataxia UK and the Rosetrees Trust (grants PGL22/100041 and PGL21/10079). S.B.W. is supported by Kidney Research UK (grant RP_017_20190306).
CONFLICT OF INTEREST STATEMENT
c.g.w.k. is a shareholder in Queen Square Analytics Ltd.
Supporting information
Baker RR, Muthurangu V, Rega M, et al. 2D sodium MRI of the human calf using half‐sinc excitation pulses and compressed sensing. Magn Reson Med. 2024;91:325‐336. doi: 10.1002/mrm.29841
REFERENCES
- 1. Friedrich A, Linz P, Nagel AM, et al. Hemodialysis patients with cardiovascular disease reveal increased tissue Na+ deposition. Kidney Blood Press Res. 2022;47:185‐193. [DOI] [PubMed] [Google Scholar]
- 2. Kopp C, Linz P, Dahlmann A, et al. 23Na magnetic resonance imaging‐determined tissue sodium in healthy subjects and hypertensive patients. Hypertension. 2013;61:635‐640. [DOI] [PubMed] [Google Scholar]
- 3. Huhn K, Linz P, Pemsel F, et al. Skin sodium is increased in male patients with multiple sclerosis and related animal models. Proc Natl Acad Sci USA. 2021;118:e2102549118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Evans RDR, Antonelou M, Sathiananthamoorthy S, et al. Inherited salt‐losing tubulopathies are associated with immunodeficiency due to impaired IL‐17 responses. Nat Commun. 2020;11:4368. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Gerhalter T, Gast LV, Marty B, et al. 23Na MRI depicts early changes in ion homeostasis in skeletal muscle tissue of patients with Duchenne muscular dystrophy. J Magn Reson Imaging. 2019;50:1103‐1113. [DOI] [PubMed] [Google Scholar]
- 6. Nagel AM, Laun FB, Weber M, Matthies C, Semmler W, Schad LR. Sodium MRI using a density‐adapted 3D radial acquisition technique. Magn Reson Med. 2009;62:1565‐1573. [DOI] [PubMed] [Google Scholar]
- 7. Boada FE, Shen GX, Chang SY, Thulborn KR. Spectrally weighted twisted projection imaging: reducing T2 signal attenuation effects in fast three‐dimensional sodium imaging. Magn Reson Med. 1997;38:1022‐1028. [DOI] [PubMed] [Google Scholar]
- 8. Lu A, Atkinson IC, Claiborne TC, Damen FC, Thulborn KR. Quantitative sodium imaging with a flexible twisted projection pulse sequence. Magn Reson Med. 2010;63:1583‐1593. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Gurney PT, Hargreaves BA, Nishimura DG. Design and analysis of a practical 3D cones trajectory. Magn Reson Med. 2006;55:575‐582. [DOI] [PubMed] [Google Scholar]
- 10. Dyke JP, Meyring‐Wösten A, Zhao Y, Linz P, Thijssen S, Kotanko P. Reliability and agreement of sodium (23Na) MRI in calf muscle and skin of healthy subjects from the US. Clin Imaging. 2018;52:100‐105. [DOI] [PubMed] [Google Scholar]
- 11. Konstandin S, Schad LR. Two‐dimensional radial sodium heart MRI using variable‐rate selective excitation and retrospective electrocardiogram gating with golden angle increments. Magn Reson Med. 2013;70:791‐799. [DOI] [PubMed] [Google Scholar]
- 12. Pauly J, Conolly S, Nishimura D, Macovski A. Slice‐selective excitation for very short T2 species. In Proceedings of the 8th Annual Meeting of SMRM, Amsterdam, The Netherlands, 1989. p. 28.
- 13. Wu C, Johnston LA, Blunck Y. 2D ultra‐short TE imaging for 23Na MRI using half‐pulse excitation. In Proceedings of the 31st Annual Meeting of ISMRM, London, UK, 2022. p. 1282.
- 14. Wittrich ML, Nagel AM, Schmitter S, Bachert P, Ladd ME, Kratzer FJ. 2D 23Na imaging using half‐VERSE pulses. In Proceedings of the 31st Annual Meeting of ISMRM, London, UK, 2022. p. 1277.
- 15. Konstandin S, Nagel AM, Heiler PM, Schad LR. Two‐dimensional radial acquisition technique with density adaption in sodium MRI. Magn Reson Med. 2011;65:1091‐1097. [DOI] [PubMed] [Google Scholar]
- 16. Chen Q, Shah NJ, Worthoff WA. Compressed sensing in sodium magnetic resonance imaging: techniques, applications, and future prospects. J Magn Reson Imaging. 2022;55:1340‐1356. [DOI] [PubMed] [Google Scholar]
- 17. Blunck Y, Kolbe SC, Moffat BA, Ordidge RJ, Cleary JO, Johnston LA. Compressed sensing effects on quantitative analysis of undersampled human brain sodium MRI. Magn Reson Med. 2020;83:1025‐1033. [DOI] [PubMed] [Google Scholar]
- 18. Madelin G, Chang G, Otazo R, Jerschow A, Regatte RR. Compressed sensing sodium MRI of cartilage at 7T: preliminary study. J Magn Reson. 2012;214:360‐365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Utzschneider M, Behl NGR, Lachner S, et al. Accelerated quantification of tissue sodium concentration in skeletal muscle tissue: quantitative capability of dictionary learning compressed sensing. MAGMA. 2020;33:495‐505. [DOI] [PubMed] [Google Scholar]
- 20. Utzschneider M, Müller M, Gast LV, et al. Towards accelerated quantitative sodium MRI at 7 T in the skeletal muscle: comparison of anisotropic acquisition‐ and compressed sensing techniques. Magn Reson Imaging. 2021;75:72‐88. [DOI] [PubMed] [Google Scholar]
- 21. Pipe JG, Zwart NR. Spiral trajectory design: a flexible numerical algorithm and base analytical equations. Magn Reson Med. 2014;71:278‐285. [DOI] [PubMed] [Google Scholar]
- 22. Qian Y, Zhao T, Zheng H, Weimer J, Boada FE. High‐resolution sodium imaging of human brain at 7 T. Magn Reson Med. 2012;68:227‐233. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Pipe JG, Menon P. Sampling density compensation in MRI: rationale and an iterative numerical solution. Magn Reson Med. 1999;41:179‐186. [DOI] [PubMed] [Google Scholar]
- 24. Montalt‐Tordera J, Steeden J, Muthurangu V. TensorFlow MRI: a library for modern computational MRI on heterogenous systems. In Proceedings of the 31st Annual Meeting of ISMRM, London, UK, 2022. p. 2769.
- 25. Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182‐1195. [DOI] [PubMed] [Google Scholar]
- 26. Stobbe RW, Beaulieu C. Residual quadrupole interaction in brain and its effect on quantitative sodium imaging. NMR Biomed. 2016;29:119‐128. [DOI] [PubMed] [Google Scholar]
- 27. Yushkevich PA, Piven J, Hazlett HC, et al. User‐guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006;31:1116‐1128. [DOI] [PubMed] [Google Scholar]
- 28. Mirkes CC, Hoffmann J, Shajan G, Pohmann R, Scheffler K. High‐resolution quantitative sodium imaging at 9.4 Tesla. Magn Reson Med. 2015;73:342‐351. [DOI] [PubMed] [Google Scholar]
- 29. Rooney WD, Springer CS Jr. A comprehensive approach to the analysis and interpretation of the resonances of spins 3/2 from living systems. NMR Biomed. 1991;4:209‐226. [DOI] [PubMed] [Google Scholar]
- 30. Rot S, Oliver‐Taylor A, Golay X, Solanky B, Gandini Wheeler‐Kingshott CAM. Towards standardising quantification in 23Na‐MRI by synthetic polyacrylamide gel phantoms. In Proceedings of the 31st Annual Meeting of ISMRM, London, UK, 2022. p. 1281.
- 31. Madelin G, Lee J, Regatte RR, Jerschow A. Sodium MRI: methods and applications. Prog Nucl Magn Reson Spectrosc. 2014;79:14‐47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal‐to‐noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26:375‐385. [DOI] [PubMed] [Google Scholar]
- 33. Gast LV, Völker S, Utzschneider M, et al. Combined imaging of potassium and sodium in human skeletal muscle tissue at 7 T. Magn Reson Med. 2021;85:239‐253. [DOI] [PubMed] [Google Scholar]
- 34. Blunck Y, Josan S, Taqdees SW, et al. 3D‐multi‐echo radial imaging of 23Na (3D‐MERINA) for time‐efficient multi‐parameter tissue compartment mapping. Magn Reson Med. 2018;79:1950‐1961. [DOI] [PubMed] [Google Scholar]
- 35. Kratzer FJ, Flassbeck S, Schmitter S, et al. 3D sodium (23Na) magnetic resonance fingerprinting for time‐efficient relaxometric mapping. Magn Reson Med. 2021;86:2412‐2425. [DOI] [PubMed] [Google Scholar]
- 36. Kotevski Z, Mitrevski P. Experimental comparison of PSNR and SSIM metrics for video quality estimation. In: Davcev D, Gómez J, eds. ICT Innovations 2009. Springer; 2010:357‐366. [Google Scholar]
- 37. Christa M, Weng AM, Geier B, et al. Increased myocardial sodium signal intensity in Conn's syndrome detected by 23Na magnetic resonance imaging. Eur Heart J Cardiovasc Imaging. 2019;20:263‐270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Wang P, Deger MS, Kang H, Ikizler TA, Titze J, Gore JC. Sex differences in sodium deposition in human muscle and skin. Magn Reson Imaging. 2017;36:93‐97. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Kopp C, Linz P, Wachsmuth L, et al. 23Na magnetic resonance imaging of tissue sodium. Hypertension. 2012;59:167‐172. [DOI] [PubMed] [Google Scholar]
- 40. Linz P, Santoro D, Renz W, et al. Skin sodium measured with 23Na MRI at 7.0 T. NMR Biomed. 2015;28:54‐62. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.