Reliability of an Instrumented Pressure Walkway for Measuring Walking and Running Characteristics in Young, Athletic Individuals

Alyx Jorgensen 1, Matthew McManigal 2, Austin Post 3, David Werner 4, Christopher Wichman 5, Matthew Tao 6, Elizabeth Wellsandt 7,
PMCID: PMC10987304  PMID: 38576831

Abstract

Background

Spatiotemporal parameters of gait are useful for identifying pathological gait patterns and presence of impairments. Reliability of the pressure-sensitive ZenoTM Walkway has not been established in young, active individuals without impairments, and no studies to this point have included running.

Purpose

The purposes of this study were to 1) determine if up to two additional trials of walking and running on the ZenoTM Walkway are needed to produce consistent measurements of spatiotemporal variables, and 2) establish test-retest reliability and minimal detectable change (MDC) values for common spatiotemporal variables measured during walking and running.

Study Design

Cross-Sectional Laboratory Study

Methods

Individuals (n=38) in this cross-sectional study walked and ran at self-selected comfortable speed on a pressure-sensitive ZenoTM Walkway. Twenty-one participants returned for follow-up testing between one and 14 days later. Intraclass correlation coefficients (ICCs) were used to assess reliability of spatiotemporal variable means using three, four, or five passes over the ZenoTM Walkway and to assess test-retest reliability of spatiotemporal variables across sessions.

Results

All variables showed excellent reliability (ICC > 0.995) for walking and running when measured using three, four, or five passes. Additionally, all variables demonstrated moderate to excellent test-retest reliability during walking (ICC: 0.732-0.982) and running (ICC: 0.679-0.985).

Conclusion

This study establishes a reliable measurement protocol of three one-way passes when using the ZenoTM Walkway for walking or running analysis. This is the first study to establish reliability of the ZenoTM Walkway during running and in young, active individuals without neuromusculoskeletal pathology.

Level of Evidence

3b

Keywords: gait, biomechanics, physical activity, healthy, test-retest reliability

INTRODUCTION

Objective human movement is traditionally studied in research laboratories with force platforms, motion capture systems, and other instrumented tools. The introduction of portable pressure walkways like the GAITRite® (CIR Systems, Inc, Franklin, NJ, USA) and ZenoTM Walkway (ProtoKinetics LLC, Havertown, PA, USA) has enabled objective assessment of gait and other movement parameters in clinical spaces. These walkways measure and collect data including spatiotemporal variables, which are useful in identifying pathological gait patterns,1–4 assessing changes in gait related to medical intervention or aging,5–14 and identifying sensorimotor impairments.15 Instrumented walkways have predominantly been used to assess older adults6,8,10–13,16,17 and those with known pathological gait patterns,2,4,5,7,18 but few studies exist that assess younger, more active individuals or include analysis of running in addition to walking.9,19

Although the ZenoTM Walkway has previously been used in research and clinical settings, standard protocols outlining the number of trials required to capture valid gait parameters have not been established. Studies that have measured gait parameters on instrumented walkways utilized anywhere from two to 10 walking trials.1,17–20 Besser et al.20 reported that, for a given spatiotemporal variable, five to 10 gait cycles were required for 90% of individuals in the study population to be within 5%-10% of their mean. However, this study was performed using the GAITRite® mat, and up to 24.2% difference in GAITRite® and ZenoTM measurements have been reported.16 Although instrumented walkways are generally understood to be reliable measurement tools,2,21–27 these reported differences indicate the need to assess the reliability of the ZenoTM Walkway independently. Sufficient data are required to establish an accurate representation of an individual’s gait pattern and to reduce error in identifying changes in gait over time. Furthermore, unnecessarily excessive data collection increases patient burden and fatigue. A standardized data collection protocol is needed so that researchers and clinicians can confidently identify when measured spatiotemporal characteristics fall outside the range of typical variability during walking or running on the ZenoTM Walkway.8,13,28

The purposes of this study were to 1) determine if up to two additional trials of walking and running on the ZenoTM Walkway are needed to produce consistent measurements of spatiotemporal variables, and 2) establish test-retest reliability and MDC values for common spatiotemporal variables measured during walking and running.

METHODS

This cross-sectional study included 38 participants (21 females; age: 24.2±3.5 years; BMI: 24.3±3.6 [range: 16.7-30.8]) recruited between July 2019 and January 2020. Twenty-one participants completed follow-up testing between one and 14 days later (12 females; age: 24.4±3.4 years; time between sessions: 5.9±2.9 days). Participants were included in this study if the following criteria were met: 15-35 years of age, no previous history of lower extremity injury or surgery during the prior six months, who participated in at least 50 hours per year of Level I (e.g. soccer, basketball, football) or II (e.g. tennis, skiing, dancing) cutting and pivoting sports.29 Exclusion criteria included recent history of hip, knee, or ankle pain or surgery during the prior six months, balance or neurological disorders, or current pregnancy. Testing was performed in the Clinical Movement Analysis Laboratory at the University of Nebraska Medical Center (UNMC). All participants provided written informed consent. This study was approved by the UNMC Institutional Review Board (IRB #0233-19-EP).

Instrumentation

Spatiotemporal variables were measured using a ZenoTM Walkway with a sensor area 20 feet long and two feet wide with 0.4 square inch sensors. Spatiotemporal data was collected at 120 Hz with a resolution of 1.27 cm (0.5 in2).

Walking and Running Testing

Testing procedures included five trials each of walking and running at a self-selected comfortable speed. A trial was defined as a single pass in one direction along the full length of the ZenoTM Walkway. Participants were given approximately three meters of floor space before and after the length of the walkway to minimize effects of acceleration and deceleration on measured gait parameters. Speed was measured using a laser timing system (Brower Timing Systems, Draper, UT) placed at the beginning and end of the walkway. Self-selected speed was determined using three unrecorded trials of walking or running, respectively, during the baseline (T1) data collection. To ensure consistency during follow-up testing (T2), the laser timing system was used to maintain walking and running speed within ±5% of the self-selected speed measured during the baseline data collection. Participants wore their own athletic shoes and were instructed to wear the same pair of shoes for both testing sessions. Testing order was standardized, with walking completed first, followed by running. Walking and running testing procedures for the twenty-one participants who returned for follow-up reliability testing were identical to the baseline testing session. Data was collected and processed using ProtoKinetics Movement Analysis Software (PKMAS) (ProtoKinetics LLC, Havertown, PA). Right and left footfalls (i.e., footstrikes) were automatically labeled by PKMAS then confirmed by a research assistant. Variables of interest included stride length, stride width, step length, step time, stride time, stance time, swing time, percent stance phase, percent swing phase, single support time, total double support time, single support percent, total double support percent, speed, cadence, integrated pressure, and toe angle. Single support and double support variables were not analyzed for running.

Figure 1 illustrates each of the four spatial variables (stride length, stride width, step length, and toe angle). For toe angle, positive values indicate an externally rotated foot relative to the direction of progression. As defined by PKMAS software, step time and stride time were calculated using the first contact of two consecutive contralateral or ipsilateral footfalls, respectively. Stance time was defined as the time between initial and final contacts for a single footfall. Swing time was calculated as the difference between stride time and stance time. Percentage variables were calculated as a percent of one full gait cycle, defined as the time between the first contact of two consecutive ipsilateral footfalls (i.e., stride time). Single support time was defined as the time during which only one foot is in contact with the ground. Total double support time was defined as the period during which both feet were in contact with the ground during stance phase, including both the initial and terminal portions of double support during a single gait cycle. Cadence was calculated using the number of footfalls divided by the total ambulation time of a single trial and then converted to steps per minute. Integrated pressure was defined as the area under the relative pressure curve for a single footfall and was averaged across all footfalls of the same limb for each trial.

Figure 1. Schematic illustration of spatiotemporal variable definitions (step length, stride length, stride width, and toe angle).

Figure 1.

Statistical Analysis

Sample Size

An a priori power analysis was conducted using unpublished pilot data of 18 healthy, uninjured participants collected prior to this study. Each participant completed three trials of walking and running along the length of the ZenoTM Walkway. Mean and standard deviation (SD) of sample SDs for integrated pressure, step length, step width, and stance time were calculated for walking (integrated pressure: 7.40±1.80; step length: 2.10±0.65 cm; step width: 2.00±0.50 cm; stance time: 0.015±0.005 sec) and running (integrated pressure: 4.70±1.40; step length: 4.70±1.90 cm; step width: 2.20±0.70 cm; stance time: 0.012±0.003 sec). A total sample size of n=37 was required to achieve at least 80% power and a significance level of 0.05 for all gait variables during walking and running using an equivalence margin of ±20%. To determine test-retest reliability, a sample size of n=17 was determined sufficient to detect intraclass correlation of 0.85 (vs. a null intraclass correlation coefficient (ICC) of 0.5) with 80% power and a significance level of 0.05.

Aim 1: Assessment of Additional Trials for Measurement Consistency

Statistical analysis was performed using IBM SPSS Statistics 26 (Armonk, NY). Two-way mixed effects intraclass correlation coefficients (ICC3,k) were used to assess absolute agreement between means of spatiotemporal variables calculated using three, four, or five passes over the ZenoTM Walkway.30 ICC values range from 0-1, with higher numbers indicating better agreement between variables. According to published literature, ICC values less than 0.5 indicate poor reliability, 0.5-0.75 indicates moderate reliability, 0.75-0.9 indicates good reliability, and values > 0.90 indicate excellent reliability.31

Aim 2: Establishing Test-Retest Reliability and Minimal Detectable Change

Two-way mixed effects intraclass correlation coefficients (ICC3,1) were used to assess absolute agreement of spatiotemporal variable means between testing sessions. For both walking and running, the average of each spatiotemporal variable during the first three trials was used in the analysis. MDC was calculated using the sample standard error of the mean (SEM), pooled standard deviation between both testing sessions (SDpooled), and ICC3,1:

MDC=1.96SEM2
where
SEM=SDpooled(1ICC)

The standard error of the mean provides information about precision, or repeatability, of a measurement. Bland-Altman plots (the difference between sessions vs. the mean between sessions) were used to visualize the agreement between sessions for each variable. Bland-Altman plots can be used to evaluate the amount of measurement bias. For example, if the absolute value of the difference between measurements tends to be greater than zero, or ±1 MDC in this analysis, it would suggest bias in the measurement.32,33 Limits of agreement were set as ± 1 MDC to visually assess the magnitude of difference between measurements and observe whether a majority of measurement differences fell within the range of the MDC (i.e. would be interpreted as a true change or not).

RESULTS

Aim 1: Assessment of Additional Trials for Measurement Consistency

This analysis included data collected in a single testing session for 38 participants (21 female; age: 24.2±3.5 years). During walking, the number of footfalls per trial for a single limb for the first three, four, and five one-way trials were 2.75±0.38, 2.73±0.38, and 2.73±0.39, respectively. During running, the number of footfalls for the first three, four, and five one-way trials were 1.58±0.49, 1.57±0.49, and 1.58±0.49, respectively. Mean (SD) values for each spatiotemporal variable calculated from all footfalls during either the first three, four, or five trials are provided in Table 1 (walking) and Table 2 (running). ICC values ranged from 0.995-1.000 for all variables during walking and running, indicating excellent reliability of spatiotemporal variables when using three, four or five passes over the ZenoTM Walkway.

Table 1. Descriptive statistics for spatiotemporal variables during walking as measured using three, four, or five one-way passes over the ZenoTM Walkway. Spatiotemporal values are presented as mean (SD).

Number of Walking Trials ICC3,k (95% CI)
Variable 3 4 5
Stride Length (cm) 156.86 (12.09) 156.81 (12.05) 156.82 (12.27) 0.999 (0.999-⁠1.000)
Stride Width (cm) 9.48 (2.94) 9.52 (2.93) 9.60 (2.88) 0.997 (0.996-0.999)
Step Length (cm) 78.46 (6.17) 78.46 (6.19) 78.44 (6.29) 0.999 (0.998-0.999)
Step Time (sec.) 0.53 (0.04) 0.53 (0.03) 0.53 (0.03) 0.999 (0.998-0.999)
Stride Time (sec.) 1.05 (0.07) 1.05 (0.06) 1.05 (0.06) 0.999 (0.998-1.000)
Stance Time (sec.) 0.66 (0.05) 0.66 (0.05) 0.66 (0.05) 0.999 (0.999-1.000)
Swing Time (sec.) 0.40 (0.02) 0.40 (0.02) 0.40 (0.02) 0.998 (0.996-0.999)
% Stance (% gait cycle) 62.08 (1.27) 62.11 (1.27) 62.10 (1.25) 0.997 (0.995-0.999)
% Swing (% gait cycle) 37.92 (1.27) 37.89 (1.27) 37.91 (1.25) 0.997 (0.996-0.999)
Single Support Time (sec.) 0.40 (0.02) 0.40 (0.02) 0.40 (0.02) 0.998 (0.996-0.999)
Double Support Time (sec.) 0.26 (0.04) 0.26 (0.04) 0.26 (0.04) 0.999 (0.998-0.999)
Single Support (% gait cycle) 38.02 (1.41) 38.01 (1.41) 38.00 (1.41) 0.998 (0.997-0.999)
Double Support (% gait cycle) 24.14 (2.53) 24.18 (2.52) 24.17 (2.52) 0.999 (0.998-0.999)
Integrated Pressure (p x sec.) 99.62 (23.28) 99.52 (23.12) 99.56 (23.18) 1.000 (1.000-1.000)
Toe Angle (degrees) 4.05 (4.29) 4.00 (4.27) 3.99 (4.18) 0.999 (0.998-0.999)
Speed (cm/sec) 149.55 (15.85) 149.33 (15.50) 149.29 (15.57) 0.999 (0.999-1.000)
Cadence (steps/min) 114.32 (6.85) 114.18 (6.75) 114.12 (6.68) 0.999 (0.998-0.999)

Intraclass correlation coefficient (ICC3,k) values indicate absolute agreement of spatiotemporal variables. ICC, intraclass correlation coefficient; CI, confidence interval; cm, centimeters; sec, seconds; %, percentage; p x sec, pressure times seconds; min, minute.

Table 2. Descriptive statistics for spatiotemporal variables during running as measured using 3, 4, or 5 one-way passes over the ZenoTM Walkway. Spatiotemporal values are presented as mean (SD).

Number of Running Trials ICC3,k (95% CI)
Variable 3 4 5
Stride Length (cm) 236.17 (33.98) 236.59 (33.97) 236.82 (33.98) 0.999 (0.999-⁠1.000)
Stride Width (cm) 5.12 (4.43) 5.07 (4.32) 5.13 (4.30) 0.998 (0.997-0.999)
Step Length (cm) 118.06 (17.27) 118.41 (17.31) 118.43 (17.27) 0.999 (0.999-1.000)
Step Time (sec.) 0.38 (0.02) 0.38 (0.02) 0.38 (0.02) 0.998 (0.997-0.999)
Stride Time (sec.) 0.76 (0.05) 0.76 (0.05) 0.76 (0.05) 0.999 (0.999-1.000)
Stance Time (sec.) 0.28 (0.04) 0.28 (0.04) 0.28 (0.04) 0.999 (0.999-1.000)
Swing Time (sec.) 0.48 (0.04) 0.48 (0.04) 0.48 (0.04) 0.999 (0.998-0.999)
% Stance (% gait cycle) 36.23 (4.15) 36.24 (4.15) 36.37 (4.06) 0.999 (0.998-0.999)
% Swing (% gait cycle) 63.81 (4.14) 63.79 (4.14) 63.64 (4.06) 0.999 (0.998-0.999)
Integrated Pressure (p x sec.) 65.23 (14.70) 65.26 (14.76) 65.33 (14.77) 1.000 (0.999-1.000)
Toe Angle (degrees) 2.72 (5.37) 2.59 (5.46) 2.52 (5.48) 0.999 (0.998-0.999)
Speed (cm/sec) 313.60 (49.49) 314.11 (49.78) 314.27 (49.90) 1.000 (0.999-1.000)
Cadence (steps/min) 159.78 (10.06) 159.30 (9.90) 159.31 (9.93) 0.995 (0.992-0.997)

Intraclass correlation coefficient (ICC3,k) values indicate absolute agreement of spatiotemporal variables. ICC, intraclass correlation coefficient; CI, confidence interval; cm, centimeters; sec, seconds; %, percentage; p x sec, pressure times seconds; min, minute.

Aim 2: Establishing Test-Retest Reliability and Minimal Detectable Change

The mean of the first three trials for each spatiotemporal variable was used to assess test-retest reliability. Table 3 (walking) and Table 4 (running) show mean, ICC, and MDC values for each spatiotemporal variable. Bland-Altman plots of the mean value between T1 and T2 vs. the difference between T1 and T2 for stride length, stride width, step time, and step length are provided in Figure 2 (walking) and Figure 3 (running). Bland Altman plots for the remaining variables are available in Appendices A and B.

Table 3. Descriptive and reliability statistics during (a) walking and (b) running using three one-way passes over the ZenoTM Walkway at T1 and T2. Symbols indicate excellent, good, or moderate reliability.

T1 Mean T2 Mean SDpooled ICC3,1 (95% CI) SEM MDC (95%)
Stride Length (cm) 156.54 156.79 13.37 0.939 (0.856-⁠0.975)* 3.30 9.15
Stride Width (cm) 9.46 9.86 2.22 0.732 (0.456-⁠0.881) 1.15 3.18
Step Time (sec.) 0.52 0.52 0.03 0.929 (0.832-0.970)* 0.01 0.02
Step Length (cm) 78.16 78.40 6.97 0.908 (0.787-0.961)* 2.11 5.86
Stride Time (sec.) 1.04 1.04 0.06 0.925 (0.824-0.969)* 0.02 0.04
Stance Time (sec.) 0.65 0.65 0.04 0.920 (0.815-0.966)* 0.01 0.03
Swing Time (sec.) 0.39 0.39 0.02 0.861 (0.688-0.941) 0.01 0.02
% Stance (% gait cycle) 62.27 62.15 1.25 0.755 (0.489-0.893) 0.62 1.72
% Swing (% gait cycle) 37.74 37.88 1.24 0.766 (0.511-0.897) 0.60 1.67
Single Support Time (sec.) 0.40 0.39 0.02 0.850 (0.670-0.936) 0.01 0.02
Double Support Time (sec.) 0.25 0.25 0.03 0.874 (0.715-0.947) 0.01 0.03
Single Support (% gait cycle) 38.03 37.78 1.28 0.796 (0.568-0.911) 0.58 1.61
Double Support (% gait cycle) 24.30 24.37 2.39 0.809 (0.586-0.918) 1.05 2.90
Integrated Pressure (p x sec.) 95.98 95.74 19.31 0.982 (0.957-0.993)* 2.59 7.18
Toe Angle (degrees) 3.60 3.69 3.44 0.923 (0.820-0.968)* 0.95 2.65
Speed (cm/sec) 150.77 151.66 16.98 0.928 (0.832-0.970)* 4.56 12.63
Cadence (steps/min) 115.47 115.87 6.07 0.904 (0.780-0.960* 1.88 5.21

T1, baseline timepoint; T2, follow-up timepoint; SDpooled , pooled standard deviation; ICC, intraclass correlation coefficient; CI, Confidence Interval; SEM, standard error of the mean; MDC, minimal detectable change; cm, centimeters; sec, seconds; %, percentage; p x sec, pressure times seconds; min, minute. *Excellent, Good, Moderate.31

Table 4. Descriptive and reliability statistics during running using three one-way passes over the ZenoTM Walkway at T1 and T2. Symbols indicate excellent, good, or moderate reliability.

T1 Mean T2 Mean SDpooled ICC3,1 (95% CI) SEM MDC (95%)
Stride Length (cm) 248.12 248.31 32.58 0.928 (0.832-⁠0.970)* 8.74 24.24
Stride Width (cm) 5.38 5.39 4.56 0.879 (0.726-⁠0.949) 1.66 4.61
Step Time (sec.) 0.37 0.37 0.02 0.731 (0.445-0.881) 0.01 0.03
Step Length (cm) 124.29 123.97 16.58 0.934 (0.844-0.972)* 4.26 11.81
Stride Time (sec.) 0.74 0.74 0.05 0.761 (0.495-0.896) 0.02 0.06
Stance Time (sec.) 0.26 0.27 0.04 0.694 (0.392-0.862) 0.02 0.06
Swing Time (sec.) 0.48 0.48 0.04 0.708 (0.410-0.870) 0.02 0.05
% Stance (% gait cycle) 34.99 35.39 3.02 0.817 (0.608-0.921) 1.29 3.58
% Swing (% gait cycle) 65.05 64.61 3.01 0.820 (0.614-0.922) 1.28 3.54
Integrated Pressure (p x sec.) 62.24 62.39 12.64 0.968 (0.923-0.987)* 2.26 6.27
Toe Angle (degrees) 2.03 1.63 5.68 0.761 (0.497-0.896) 2.78 7.77
Speed (cm/sec) 335.30 336.47 45.87 0.985 (0.964-0.994)* 5.62 15.57
Cadence (steps/min) 162.52 163.52 10.68 0.679 (0.358-0.856) 6.05 16.78

T1, baseline timepoint; T2, follow-up timepoint; SDpooled , pooled standard deviation; ICC, intraclass correlation coefficient; CI, Confidence Interval; SEM, standard error of the mean; MDC, minimal detectable change; cm, centimeters; sec, seconds; %, percentage; p x sec, pressure times seconds; min, minute. *Excellent, Good, Moderate.31

Figure 2. Sample Bland-Altman plots for spatiotemporal variables during walking.

Figure 2.

The dashed red lines and shaded areas indicate limits of agreement equal to ±1 MDC. The dashed black line indicates the mean difference for all participants between the first and second testing sessions. The solid black line is included as a reference and indicates no difference between testing sessions. Plots for all variables are available in Appendix A. T1, baseline timepoint; T2, follow-up timepoint; cm, centimeters; sec, seconds

Figure 3. Sample Bland-Altman plots for spatiotemporal variables during running.

Figure 3.

The dashed red lines and shaded areas indicate limits of agreement equal to ±1 MDC. The dashed black line indicates the mean difference for all participants between the first and second testing sessions. The solid black line is included as a reference and indicates no difference between testing sessions. Plots for all variables are available in Appendix B. T1, baseline timepoint; T2, follow-up timepoint; cm, centimeters; sec, seconds

ICC values during walking ranged from 0.732-0.982. Variables with excellent test-retest reliability (ICC > 0.90) during walking on the ZenoTM Walkway were step length, stride length, step time, stride time, stance time, integrated pressure, toe angle, speed, and cadence. Variables with good reliability (ICC 0.75-0.90) were swing time, percent stance, percent swing, single support time, double support time, single support percent, and double support percent. Stride width (ICC=0.732) was the least reliable variable and the only walking variable with moderate reliability (ICC 0.5-0.75).

For running, ICC values ranged from 0.683-0.939. Variables with excellent test-retest reliability during running on the ZenoTM Walkway included step length, stride length, integrated pressure, and speed. Variables with good reliability included stride width, step time, stride time, percent stance, percent swing, and toe angle. Variables with moderate reliability during running included stance time, swing time, and cadence. No variables demonstrated poor test-retest reliability during running.

DISCUSSION

The purposes of this study were to 1) determine if up to two additional trials of walking and running on the ZenoTM Walkway are needed to produce consistent measurements of spatiotemporal variables, and 2) establish test-retest reliability and MDC values for common spatiotemporal variables measured during walking and running. The results of the first aim suggest that three one-way passes over the walkway are sufficient to collect reliable walking and running data in young, active individuals, when compared with using four or five one-way passes. This information can serve as a guideline for efficient and effective use of a ZenoTM Walkway in clinical and research settings. Fewer trials may reduce patient or participant fatigue and the burden of time on both the patient and clinician/researcher collecting the data.

Use of the ZenoTM Walkway and other instrumented pressure walkways has historically been limited to older adults or populations with known gait pattern abnormalities. The results of this study support its use in younger, more active populations and with higher-level movements like running. The use of pressure walkways with these movements would allow a deeper understanding of running without the time and cost of three-dimensional motion capture. Results from the second aim demonstrate that the ZenoTM Walkway is a reliable measurement tool for spatiotemporal characteristics of walking and running in a relatively young (15-35 years old) population, with most variables showing good to excellent reliability. Stride width, percent stance, and percent swing were the least reliable variables during walking with ICCs of 0.732-0.766, although they still show moderate to good reliability when measured by a ZenoTM Walkway. Although these values showed lower ICCs, they each demonstrated relatively small magnitude SEMs, supporting the precision and reliability of the measurements.34 Participants in this trial demonstrated comparable walking gait parameters to previously published data collected on the ZenoTM Walkway in samples of a similar age range.9 Minimal detectable change values indicate the amount of expected variability between testing sessions for each variable during walking or running. Variations larger than the MDC indicate true changes in gait pattern. The MDC values reported in this study are unique to the ZenoTM Walkway and may be useful in clinical and research settings to discern subtle but legitimate changes in gait parameters longitudinally or in response to intervention that would not otherwise be perceived. Further investigation is needed to determine whether these MDC values correspond to clinically-meaningful changes in spatiotemporal characteristics of walking and running.

This is the first study to quantify spatiotemporal parameters of running when using the ZenoTM Walkway. All spatiotemporal variables included in this analysis showed moderate to excellent test-retest reliability during running. Temporal variables including cadence, stride time, step time, and stance time were the least reliable with ICCs ranging from 0.683-0.771. In general, reliability decreased during running compared to walking. This reduction in reliability during running could be a result of measurement error at higher velocities. Despite showing reduced reliability, differences in measured spatiotemporal variables between sessions may still fall within the range of normal variability and not affect the larger goal of assessing for clinically-relevant changes in running characteristics.

Limitations of this study include the narrow inclusion criteria such that results may only be generalized to young, active individuals and not to older, active adults and sedentary young people. Future studies should aim to assess the reliability of specific high-level movements like hopping to increase the utility of the ZenoTM Walkway for young, athletic populations.

CONCLUSION

The findings of this study recommend a reliable measurement protocol of three one-way passes over the ZenoTM Walkway for assessment of walking or running. During both tasks, spatiotemporal variables of gait show moderate to excellent test-retest reliability when measured at least one day apart. The ZenoTM Walkway is a reliable measurement tool for gait analysis and can be used in the assessment and management of young, active populations without neuromusculoskeletal impairment or injury.

Conflicts of interest

None

Supplementary Material

Appendix A
Appendix B

Funding Statement

This work was supported by the Nebraska Bankers Association and the University of Nebraska Medical Center. The funding sources had no involvement in study design, data collection, analysis, interpretation, manuscript writing, or decision to submit the article for publication.

References

  1. Brach Jennifer S., Studenski Stephanie, Perera Subashan, VanSwearingen Jessie M., Newman Anne B. Gait & Posture. 3. Vol. 27. Elsevier BV; Stance time and step width variability have unique contributing impairments in older persons; pp. 431–439. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Roche Bastien, Simon Anne-Laure, Guilmin-Crépon Sophie, Boizeau Priscilla, Andriss Béatrice, Alberti Corinne, Presedo Ana, Ilharreborde Brice, Husson Isabelle. Gait & Posture. Vol. 66. Elsevier BV; Test-retest reliability of an instrumented electronic walkway system (GAITRite) for the measurement of spatio-temporal gait parameters in young patients with Friedreich’s ataxia; pp. 45–50. [DOI] [PubMed] [Google Scholar]
  3. Guo Guodong, Guffey Keegan, Chen Wenbin, Pergami Paola. Neuroinformatics. 1. Vol. 15. Springer Science and Business Media LLC; Classification of normal and pathological gait in young children based on foot pressure data; pp. 13–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Monticone Marco, Ambrosini Emilia, Fiorentini Roberta, Rocca Barbara, Liquori Valentina, Pedrocchi Alessandra, Ferrante Simona. Gait & Posture. 4. Vol. 40. Elsevier BV; Reliability of spatial–temporal gait parameters during dual-task interference in people with multiple sclerosis. A cross-sectional study; pp. 715–718. [DOI] [PubMed] [Google Scholar]
  5. Song Melissa, Lieberman Abraham, Fife Terry, Nielsen Megan, Hayden Sharon, Sabbagh Marwan, Shi Jiong. Acta Neurologica Scandinavica. 4. Vol. 139. Hindawi Limited; A prospective study on gait dominant normal pressure hydrocephalus; pp. 389–394. [DOI] [PubMed] [Google Scholar]
  6. Virmani Tuhin, Gupta Harsh, Shah Jesal, Larson-Prior Linda. Gait & Posture. Vol. 65. Elsevier BV; Objective measures of gait and balance in healthy non-falling adults as a function of age; pp. 100–105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Urquhart Donna M., Morris Meg E., Iansek Robert. Archives of Physical Medicine and Rehabilitation. 6. Vol. 80. Elsevier BV; Gait consistency over a 7-day interval in people with Parkinson's disease; pp. 696–701. [DOI] [PubMed] [Google Scholar]
  8. Hausdorff Jeffrey M., Rios Dean A., Edelberg Helen K. Archives of Physical Medicine and Rehabilitation. 8. Vol. 82. Elsevier BV; Gait variability and fall risk in community-living older adults: a 1-year prospective study; pp. 1050–1056. [DOI] [PubMed] [Google Scholar]
  9. McKay Marnee J., Baldwin Jennifer N., Ferreira Paulo, Simic Milena, Vanicek Natalie, Wojciechowski Elizabeth, Mudge Anita, Burns Joshua. Gait & Posture. Vol. 58. Elsevier BV; Spatiotemporal and plantar pressure patterns of 1000 healthy individuals aged 3–101 years; pp. 78–87. [DOI] [PubMed] [Google Scholar]
  10. Verlinden V.J.A., van der Geest J.N., Hoogendam Y.Y., Hofman A., Breteler M.M.B., Ikram M.A. Gait & Posture. 4. Vol. 37. Elsevier BV; Gait patterns in a community-dwelling population aged 50 years and older; pp. 500–505. [DOI] [PubMed] [Google Scholar]
  11. Herssens Nolan, Verbecque Evi, Hallemans Ann, Vereeck Luc, Van Rompaey Vincent, Saeys Wim. Gait & Posture. Vol. 64. Elsevier BV; Do spatiotemporal parameters and gait variability differ across the lifespan of healthy adults? A systematic review; pp. 181–190. [DOI] [PubMed] [Google Scholar]
  12. Hars Mélany, Herrmann François R., Trombetti Andrea. Gait & Posture. 4. Vol. 38. Elsevier BV; Reliability and minimal detectable change of gait variables in community-dwelling and hospitalized older fallers; pp. 1010–1014. [DOI] [PubMed] [Google Scholar]
  13. Marigold Daniel S., Patla Aftab E. Gait & Posture. 4. Vol. 27. Elsevier BV; Age-related changes in gait for multi-surface terrain; pp. 689–696. [DOI] [PubMed] [Google Scholar]
  14. Paterson K. L., Lythgo N. D., Hill K. D. Age and Ageing. 6. Vol. 38. Oxford University Press (OUP); Gait variability in younger and older adult women is altered by overground walking protocol; pp. 745–748. [DOI] [PubMed] [Google Scholar]
  15. Kirkland Megan C., Wadden Katie P., Ploughman Michelle. Disability and Rehabilitation. 8. Vol. 44. Informa UK Limited; Bipedal hopping as a new measure to detect subtle sensorimotor impairment in people with multiple sclerosis; pp. 1544–1555. [DOI] [PubMed] [Google Scholar]
  16. Vallabhajosula Srikant, Humphrey Sarah K., Cook Alys J., Freund Jane E. Journal of Geriatric Physical Therapy. 3. Vol. 42. Ovid Technologies (Wolters Kluwer Health); Concurrent validity of the Zeno Walkway for measuring spatiotemporal gait parameters in older adults; pp. E42–E50. [DOI] [PubMed] [Google Scholar]
  17. Hollman John H., McDade Eric M., Petersen Ronald C. Gait & Posture. 1. Vol. 34. Elsevier BV; Normative spatiotemporal gait parameters in older adults; pp. 111–118. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Chen Alice, Kirkland Megan C., Wadden Katie P., Wallack Elizabeth M., Ploughman Michelle. Gait & Posture. Vol. 78. Elsevier BV; Reliability of gait and dual-task measures in multiple sclerosis; pp. 19–25. [DOI] [PubMed] [Google Scholar]
  19. Paterson Kade L., Hill Keith D., Lythgo Noel D., Maschette Wayne. Archives of Physical Medicine and Rehabilitation. 12. Vol. 89. Elsevier BV; The reliability of spatiotemporal gait data for young and older women during continuous overground walking; pp. 2360–2365. [DOI] [PubMed] [Google Scholar]
  20. Representation of temporal spatial gait parameters using means in adults without impairment. Besser M., Kmieczak K., Schwartz L., Snyderman M., Wasko J., Selby-Silverstein L. 1999Gait Posture. 9:113. [Google Scholar]
  21. Parati Monica, Ambrosini Emilia, De Maria Beatrice, GALLOTTA Matteo, DALLA VECCHIA Laura A., FERRIERO Giorgio, FERRANTE Simona. European Journal of Physical and Rehabilitation Medicine. 3. Vol. 58. Edizioni Minerva Medica; The reliability of gait parameters captured via instrumented walkways: a systematic review and meta-analysis; pp. 363–377. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Cutlip Robert G, Mancinelli Corrie, Huber Frances, DiPasquale James. Gait & Posture. 2. Vol. 12. Elsevier BV; Evaluation of an instrumented walkway for measurement of the kinematic parameters of gait; pp. 134–138. [DOI] [PubMed] [Google Scholar]
  23. Ade Vanessa, Schalkwijk Dale, Psarakis Michael, Laporte Michael D., Faras Timothy J., Sandoval Remi, Najjar Fadi, Stubbs Peter W. PLOS ONE. 7. Vol. 13. Public Library of Science (PLoS); Between session reliability of heel-to-toe progression measurements in the stance phase of gait; p. e0200436. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. McDonough Andrew L., Batavia Mitchell, Chen Fang C., Kwon Soonjung, Ziai James. Archives of Physical Medicine and Rehabilitation. 3. Vol. 82. Elsevier BV; The validity and reliability of the GAITRite system's measurements: A preliminary evaluation; pp. 419–425. [DOI] [PubMed] [Google Scholar]
  25. van Uden Cornelis J T, Besser Marcus P. BMC Musculoskeletal Disorders. Vol. 5. Springer Science and Business Media LLC; Test-retest reliability of temporal and spatial gait characteristics measured with an instrumented walkway system (GAITRite®) p. 13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Bilney Belinda, Morris Meg, Webster Kate. Gait & Posture. 1. Vol. 17. Elsevier BV; Concurrent related validity of the GAITRite® walkway system for quantification of the spatial and temporal parameters of gait; pp. 68–74. [DOI] [PubMed] [Google Scholar]
  27. Menz Hylton B, Latt Mark D, Tiedemann Anne, Mun San Kwan Marcella, Lord Stephen R. Gait & Posture. 1. Vol. 20. Elsevier BV; Reliability of the GAITRite® walkway system for the quantification of temporo-spatial parameters of gait in young and older people; pp. 20–25. [DOI] [PubMed] [Google Scholar]
  28. Stergiou Nicholas, Decker Leslie M. Human Movement Science. 5. Vol. 30. Elsevier BV; Human movement variability, nonlinear dynamics, and pathology: is there a connection? pp. 869–888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Grindem Hege, Eitzen Ingrid, Snyder-Mackler Lynn, Risberg May Arna. British Journal of Sports Medicine. 9. Vol. 48. BMJ; Online registration of monthly sports participation after anterior cruciate ligament injury: a reliability and validity study; pp. 748–753. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Shrout Patrick E., Fleiss Joseph L. Psychological Bulletin. 2. Vol. 86. American Psychological Association (APA); Intraclass correlations: uses in assessing rater reliability. pp. 420–428. [DOI] [PubMed] [Google Scholar]
  31. Koo Terry K., Li Mae Y. Journal of Chiropractic Medicine. 2. Vol. 15. Elsevier BV; A guideline of selecting and reporting intraclass correlation coefficients for reliability research; pp. 155–163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Giavarina Davide. Biochemia Medica. 2. Vol. 25. Croatian Society for Medical Biochemistry and Laboratory Medicine; Understanding Bland Altman analysis; pp. 141–151. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. Hergott Colleen G., Bolgla Lori A., Waller Jennifer L., Dowling Aaron, Ezzell Kennedy, Graves Corley, Peed William. Cardiopulmonary Physical Therapy Journal. 4. Vol. 33. Ovid Technologies (Wolters Kluwer Health); TUG-10: A modification of the timed up and go test for aerobic assessment in older adults; pp. 157–165. [DOI] [Google Scholar]
  34. Denegar Craig R., Ball Donald W. Journal of Sport Rehabilitation. 1. Vol. 2. Human Kinetics; Assessing the reliability and precision of measurement: an introduction to intraclass correlation and standard error of measurement; pp. 35–42. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Appendix A
Appendix B

Articles from International Journal of Sports Physical Therapy are provided here courtesy of North American Sports Medicine Institute