Wearable devices for remote monitoring of hospitalized patients with COVID-19 in Vietnam

Nguyen Van Vinh Chau 1, Truong Ngoc Trung 1, Phan Nguyen Quoc Khanh Khanh 2, Phung Tran Huy Nhat 2,3, Hoang Minh Tu Van 2, Ho Bich Hai 2, Duong Bich Thuy 1, Nguyen Le Nhu Tung 1, Dao Bach Khoa 1, Tran Thi Dong Vien 1, Nguyen Van Hao 1,4, Pham Kieu Ngyuyet Oanh 1, Tran Dang Khoa 1, Nguyen Thanh Phong 1, Nguyen Thanh Nguyen 1, Julie Huynh 2, Timothy M Walker 2,5, Jennifer Van Nuil 2,5, Luu Phuoc An 2, Jacob McKnight 6, Le Mau Toan 1, Le Van Tan 2, Nguyen Thanh Dung 1, Nguyen Thanh Truong 1, C Louise Thwaites 2,5,a; OUCRU COVID Research Group
PMCID: PMC11004598  PMID: 38601327

Version Changes

Revised. Amendments from Version 1

We have added an additional section explaining the more nuanced differences between the systems and in the foot note to the table made it clear which departments which systems were used.

Abstract

Patients with severe COVID-19 disease require monitoring with pulse oximetry as a minimal requirement. In many low- and middle- income countries, this has been challenging due to lack of staff and equipment. Wearable pulse oximeters potentially offer an attractive means to address this need, due to their low cost, battery operability and capacity for remote monitoring. Between July and October 2021, Ho Chi Minh City experienced its first major wave of SARS-CoV-2 infection, leading to an unprecedented demand for monitoring in hospitalized patients. We assess the feasibility of a continuous remote monitoring system for patients with COVID-19 under these circumstances as we implemented 2 different systems using wearable pulse oximeter devices in a stepwise manner across 4 departments.

Keywords: Wearable monitoring, Pulse oximetry, COVID-19, Low-middle-income countries, resource-limited

Introduction

Monitoring peripheral oxygen saturation (SpO 2) is an essential component of management of moderate or severely ill patients with COVID-19, enabling timely institution of treatments such as oxygen or mechanical ventilation. Globally, the huge number of cases of COVID-19 has put health systems under enormous strain, and even simple monitoring tasks, such as SpO 2 observations, are challenging due to lack of equipment and trained staff. In many low and middle income countries (LMICs) these were already known to be constraints in providing critical care and the Hospital for Tropical Diseases Ho Chi Minh City, collaborating with Oxford University Clinical Research Unit, were already developing innovative ways of improving real-time monitoring of critically ill patients in the intensive care unit (ICU) 1 . Investigating the use of wearable monitoring systems formed part of this strategy as, in addition to lower cost, the wireless connectivity, small size and battery power offer considerable advantages for use in a resource-limited ICU where power supply and overcrowding are common issues 1 . A specific multidisciplinary team had identified and tested wearable pulse oximeters (measuring SpO 2) and electrocardiogram (ECG) devices for this purpose 2 . Using these selected wearables, previous work had demonstrated that data collection is feasible and that data quality collected from these devices was comparable to data derived from state-of-the-art bedside ICU monitors, however a large-scale deployment of wearable monitoring devices had not been performed 2 . Whilst particularly attractive in LMICs, the potential advantages of cost and automated data collection are also attractive in high income settings where wearable monitoring has been successfully used in post-surgical patients, and as a basis for early warning scores in ICUs 3, 4 .

The huge demand for basic monitoring (particularly of SpO 2) during the COVID-19 pandemic, led to several initiatives using wearable monitors in healthcare settings. The majority of those reported have been in high-income countries and in out-patient settings where patients’ self-reported intermittent values from wearable pulse oximeters have been used as means of triage or to facilitate early discharge of patients with relatively mild disease 57 . Successful automated monitoring of ambulatory patients with COVID-19 has been described using a system connecting wearable devices via bedside android tablets to clinician dashboards 7 . Use of such systems in ICUs is more limited 8, 9 . In high income settings, widespread availability of centralized bedside monitoring systems with wired-connectivity already provides staff with the means to remotely monitor patients’ vital signs continuously. However, in LMICs, where conventional pulse oximetry is often not available, simple low-cost systems have the potential for significantly improving patient management. Furthermore, implementation of pulse oximetry monitoring in resource-limited settings enables optimal Oxygen administration and has been shown to actually reduce overall Oxygen utilization 10 . Ensuring a sustainable Oxygen supply for patients with COVID-19 continues to be a problem in many LMICs, and increasing pulse oximetry monitoring capacity is integral to achieving this 11 .

At the outset of the pandemic a wearable pulse oximeter system was piloted in a field-hospital setting in Vietnam, using a screen sharing mobile application to access a bedside mobile device connected to a wearable pulse oximeter 12 ( Figure 1). Patients using this system were mobile and able to use this themselves (using the telephone for additional support). Following a surge in cases in Ho Chi Minh City June-August 2021, our team responded to a local need for expansion of monitoring capacity by implementing this in more severely-ill patients in high-dependency/ICU wards at the Hospital for Tropical Diseases. A second wearable-based pulse oximetry monitoring system was also implemented. This was similar in that it utilized a wearable pulse oximeter connected to a Bluetooth tablet but contained a bespoke user-interface and clinician dashboard capable of simultaneously displaying data from multiple patients. Both systems were capable of providing continuous SpO 2 and heart rate measurements, accessible both at the bedside and remotely.

Figure 1. Schematic representation of the monitoring systems used.

Figure 1.

Wearable pulse oximeters were connected with tablet/phones placed either by the bedside or nearby (outside room). These connected through either screen mirroring applications or direct wireless connections to enable remote viewing by medical staff in central staff workstations.

Here we describe the implementation process at the Hospital for Tropical Diseases, a large infectious disease hospital, during the COVID-19 pandemic, highlighting the key challenges that the COVID situation presented and potential benefits for future care of ICU patients in LMICs.

Methods

The Hospital for Tropical Diseases is am 800-bed tertiary referral hospital for infectious diseases, serving southern Vietnam (population 40 million). The Oxford University Clinical Research Unit has been hosted within the hospital for almost 30 years, aiming to work closely with HTD improving outcomes from patients with infectious diseases. At the outset of the pandemic (January 2021) a small number of isolation rooms were used to treat patients with COVID-19, however as the pandemic progressed increasing numbers of departments were given over to treating COVD-19 patients. By July 2021, full ICU facilities were available in 6 departments (4 of which were included in this project). A multidisciplinary team was involved in the project from the outset and was comprised of clinicians, managers, intensive care specialists, data scientists, industry experts and biomedical engineers. Health systems scientists were subsequently incorporated into the team. At the outset of the project clear responsibilities of each party were defined, including device management, data governance and ownership. Key hospital stakeholders defined the basic requirement for monitoring systems: namely the ability to continuously visualise individual patients’ SpO 2 remotely. Routine vital sign monitoring (intermittent manual measurements by bedside nursing staff) were measured and recorded in the hospital electronic and paper records as normal.

Face-to-face formal meetings involving all parties were often impossible due to social-distancing restrictions. A central communication system through one or two key individuals was established and electronic communication though email or encrypted social media groups was used to keep regular updates. An online ‘problem’ sheet was instituted at an early phase to allow issues to be communicated rapidly. Many of the workflows occurred in parallel due to the nature of the work.

The majority of the work focussed on institution of the new system (System A), as this had capability for monitoring more patients with access to a larger number of available compatible monitoring devices, at a time when monitoring devices were in short supply. Implementation and ongoing development of this system was also supported by an industry team. For example, the system was able to incorporate additional features specified by frontline workers such as a ‘traffic-light alarm’ system and adapt the clinician dashboard to staff requirements. Based on these user requirements defined by hospital stakeholders, a minimally viable product was created for System A ( Figure 1). This consisted of wearable pulse oximeters (Viatom Viatomtech, China) connected to a tablet in a 1-to-1 routine configuration, connected to a cloud server via 3G network allowing data visualization from all monitors on a computer screen. This system was piloted off-site in healthy people first then introduced into the ward environment where a further pilot was held using devices in a non-patient area. Following this, the prototype system was introduced into one department in a small number of patients following staff training.

In addition to this new system, the previously-piloted adaptive remote monitoring system (System B) was introduced in one department. The system, previously described, consisted of a wearable pulse oximeter (SmartCare, SmartCare Analytics, UK) connected to an android tablet via a Bluetooth connection (1-to-1 configuration). The pulse oximetry readings from several devices could be viewed in real-time on a remote computer screen via a screen mirror application which allowed mirroring of multiple devices, although not simultaneously. In addition to this method, ward staff could also use the departments’ close circuit television (CCTV) systems’ ‘zoom in’ function to view the bedside tablet’s display of pulse oximetry output. Thus, although both System A and B were able to display continuous heart rate and pulse oximetry, multiple patient data could be viewed using System A which also included additional alert data for abnormal values. System B however allowed detail waveform pulse oximetry data to be viewed and bedside tablets served as additional bedside monitoring displays which could be viewed by CCTV camera.

As ward staff were under extreme pressure, responding ‘ad hoc’ to technical issues such as battery run-out was not feasible, thus for both systems, each patient was allocated two wearable devices to allow constant monitoring (one device active and one being charged). These devices were changed at fixed times in the day irrespective of battery levels. Standard infection-control procedures were employed to clean devices between patients. Research team members were also directly involved in the implementation, supporting these activities. In addition, routine vital sign observations procedures were carried out as normal by clinical staff following hospital and Vietnamese Ministry of Health guidelines.

Results

Between August 3 rd and October 31 st 2021, the two monitoring systems were implemented in four dedicated departments for patients with COVID-19 at the Hospital for Tropical Diseases. These departments mainly received patients with severe and critical COVID-19 according to Vietnamese Ministry of Health Guidelines 13 , and all had ICU facilities for additional respiratory support (facemask oxygen, nasal oxygen, high flow nasal oxygen, non-invasive ventilation and invasive mechanical ventilation). Table 1 shows the total number of patients admitted to the four departments in which wearable monitoring was instituted between August and October 2021.

Table 1. Showing number of patients admitted to departments involved in the project August–October 2021.

Month Ward A Ward D Ward E A-ICU
August 104 126 189 37
September 72 83 107 33
October 22 75 94 13
Total 198 284 390 83

The new system (System A) was instituted in three of the departments and in one (Ward D), the previously-piloted (System B) was implemented. The phased scale up of this is shown in Figure 2 and the monthly cumulative admissions to the four departments over the same time period.

Figure 2. Figure showing stepwise scale up of wearable monitoring within hospital departments (Stacked bars).

Figure 2.

System A was introduced in Departments E (dark blue), AICU (grey) and A (light blue) with system B in Department D (green). Solid area represents cumulative new admissions to these 4 departments during the month.

The initial challenges encountered were practical issues setting up the basic infrastructure required to deploy the wearable systems, given the extreme clinical pressures at the time and limited availability of new equipment. During the set-up phase, the hospital completed infrastructure upgrades in several of the departments where the devices were deployed, for example ensuring oxygen supply, electric supply for equipment and physical infection control barriers. In parallel with this, any necessary hardware was installed, repurposing equipment such as monitors and laptop computers to reduce the need to purchase new equipment as much as possible.

System A was the first system introduced, with the first patients monitored within four weeks from the project outset. An initial pilot of five sets (each two devices), able to monitor five patients was scaled up to increase monitoring capacity for 20 patients within Ward E. The research team worked closely with the clinical team and the industry partner to troubleshoot technical problems as they arose. Following this phase, the system was introduced into two other departments: (Ward AICU and Ward A). System B was implemented in Ward D and provided capacity for 16 patients, with bedside tablets connected to a laptop allowing staff to remotely view monitoring data from outside the patient area.

We observed that the exact use of the wearable systems varied according to the ward. In the Wards A, D and E, the systems provided continuous monitoring in for individual patients who would otherwise receive intermittent manual measurements. In addition to using the clinician dashboard (System A), or laptop (System B), bedside tablets also served as equivalent of bedside monitors, enabling staff at the bedside to see continuous vital sign data. In AICU, where conventional bedside monitoring was already available, the wearables were used in patients in whom no remote-viewing facility was available, thus patients’ SpO 2 data could also be monitored from a central station outside of the patient areas.

Discussion

We have described the feasibility of implementing a wearable monitoring system in our hospital in a situation of exceptional healthcare system stress in Ho Chi Minh City, where there was a constant requirement for the system to respond to the evolving pandemic. Under these circumstances, our main aim was to rapidly set up a functioning system, that could complement current standard of care, using expertise and equipment available at the time. The urgent need for increased monitoring capacity meant that a ‘lean’ implementation approach was required – i.e. systems functioning with minimal requirements were introduced first and adapted over time. This was a big advantage in that it enabled rapid deployment and additional level of remote monitoring when there was a large unmet need. Our first wearable devices were deployed in patients in less than four weeks, from first conceptualization. Undoubtedly our teams’ previous experience and multidisciplinary expertise were enabling factors in achieving this, despite the new challenges of operating in COVID-19 context 14 .

As expected, challenges arose due to the novelty of the technology and speed at which it needed to be implemented. Our pre-pandemic experience, while useful for this, did not directly translate for the pandemic use. For example, support and technical staff were no longer easily able to access patients’ bedsides (or required full PPE to do so), and staff previously familiar with the devices had been located to other departments.

Similar to other LMICs, even outside of the pandemic, staff-to-patient ratios in Vietnamese hospitals are significantly lower than generally reported in high-income settings 15 . Accordingly, staff availability for training in operation and trouble-shooting of new devices may be limited. In the pandemic context, the constantly-changing capacity and structure of the city’s healthcare system and the implementation of specific care pathways necessitated frequent staff redeployment into and out of the departments described herein. As a consequence, additional time was required to train new staff in the systems, which under normal circumstances would have been less. The presence of dedicated research staff to support the project we believe was vital in enabling this ongoing training and education. Notably, however many of these staff had been redeployed from other research projects.

Although the features of battery power and blue tooth connectivity were attractive features for our immediate need, both were also aspects of the system requiring specific workflows to charge devices and maintain connectivity. In the UK, a similar wearable monitoring system was implemented in a high-dependency setting in a total of 59 patients between March and August 2020. Notably, in this context, there were significantly more staff available (Four to five nursing staff and three to four doctors for 19 patients) 7 . Staff responded to problems as they arose however in our setting, regular ‘maintenance’ schedules were found to be more effective ways of addressing these issues with fewer staff available and in view of the rapid turnover.

A limitation of this work is the lack of detailed operational data to more thoroughly describe the function of the deployed systems, for example of patient monitoring, and number of devices used at any one time. In the UK example above, the system was already developed prior to the pandemic and data metrics easily taken from this. As pandemic pressures ease, incorporating these metrics is an important step – both technically and clinically. Similarly, usability feedback from staff is also now an important consideration, but at the time, formal evaluation was limited by staff availability to both give and record feedback.

Conclusion and next steps

Our project shows how important pre-existing expertise and infrastructure were in rapidly implementing our solution. As we, hopefully, face a fading pandemic, the fundamental issues of providing high-quality low-cost critical care in LMICs remain, and central to this is the need to provide low-cost clinical monitoring. We aim to direct the learning and expertise gained in this project towards achieving this. Whilst our efforts in this project were directed at a specific pandemic-related solution, notably absent was capture and analysis of monitoring data. Recent advances in analytics (particularly artificial intelligence/ machine learning) offer exciting potential for decision support systems based on such data, and could be used to support patient care and improve patient outcomes in LMICs. Designing and implementing an effective monitoring solution with such capacity would have immense potential for care of critically ill patients in low-resource settings.

Funding Statement

This work was supported by the Foreign, Commonwealth & Development Office and Wellcome [225437/Z/22/Z]. Equipment supplied by iPARAMED company and donation by British Chamber of Commerce, Vietnam.

The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

[version 2; peer review: 3 approved, 1 approved with reservations]

Data availability

Data included in this manuscript are provided by the Hospital for Tropical Diseases and are available through OUCRU and HTD data sharing processes.

References

  • 1. Turner HC, Hao NV, Yacoub S, et al. : Achieving affordable critical care in low-income and middle-income countries. BMJ Glob Health. 2019;4(3):e001675. 10.1136/bmjgh-2019-001675 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Van HMT, Hao NV, Quoc KPN, et al. : Vital sign monitoring using wearable devices in a Vietnamese intensive care unit. BMJ Innov. 2021;7(Suppl 1):S7–11. 10.1136/bmjinnov-2021-000707 [DOI] [Google Scholar]
  • 3. Watkinson PJ, Pimentel MAF, Clifton L, et al. : Early detection of physiological deterioration in post-surgical patients using wearable technology combined with an integrated monitoring system: a pre- and post-interventional study. MedRxiv. 2020. 10.1101/2020.12.01.20240770 [DOI] [Google Scholar]
  • 4. Downey C, Randell R, Brown J, et al. : Continuous versus intermittent vital signs monitoring using a wearable, wireless patch in patients admitted to surgical wards: Pilot cluster randomized controlled trial. J Med Internet Res. 2018;20(12):e10802. 10.2196/10802 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. RCGP: RCGP paper on Virtual Wards, Silent Hypoxia and improving COVID outcomes.2020. Reference Source [Google Scholar]
  • 6. Vindrola-Padros C, Singh KE, Sidhu MS, et al. : Remote home monitoring (virtual wards) for confirmed or suspected COVID-19 patients: a rapid systematic review. EClinicalMedicine. 2021;37:100965. 10.1016/j.eclinm.2021.100965 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Santos MD, Roman C, Pimentel MAF, et al. : A Real-Time Wearable System for Monitoring Vital Signs of COVID-19 Patients in a Hospital Setting. Front Digit Health. 2021;3:630273. 10.3389/fdgth.2021.630273 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Steinhubl SR, Feye D, Levine AC, et al. : Validation of a portable, deployable system for continuous vital sign monitoring using a multiparametric wearable sensor and personalised analytics in an Ebola treatment centre. BMJ Glob Health. 2016;1(1):e000070. 10.1136/bmjgh-2016-000070 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Channa A, Popescu N, Skibinska J, et al. : The rise of wearable devices during the COVID-19 pandemic: A systematic review. Sensors (Basel). 2021;21(17):5787. 10.3390/s21175787 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Sutherland T, Moriau V, Niyonzima JM, et al. : The "Just Right" amount of oxygen. Improving oxygen use in a Rwandan emergency department. Ann Am Thorac Soc. 2019;16(9):1138–42. 10.1513/AnnalsATS.201811-763QI [DOI] [PubMed] [Google Scholar]
  • 11. Mahase E: Covid-19: Oxygen shortages two years into pandemic highlight pre-covid failures, says WHO. BMJ. 2022;376:o829. 10.1136/bmj.o829 [DOI] [PubMed] [Google Scholar]
  • 12. Chau NVV, Hai HB, Greeff H, et al. : Wearable remote monitoring for patients with COVID-19 in low-resource settings: Case study. BMJ Innov. 2021;7(Suppl 1):S12–5. 10.1136/bmjinnov-2021-000706 [DOI] [Google Scholar]
  • 13. Vietnam Ministry of Health: Guidelines for COVID-19 Management. Reference Source [Google Scholar]
  • 14. van Goor HMR, Eddahchouri Y, van Loon K, et al. : Can continuous remote vital sign monitoring reduce the number of room visits to patients suspected of COVID-19: A quasi-experimental study. Int J Nurs Stud. 2021;115:1038628. 10.1016/j.ijnurstu.2020.103868 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Finfer S, Hammond N, Tirupakuzhi BK, et al. : Critical Care Resources and Disaster Preparedness Survey 2020. MedRxiv. 2021;1–30. 10.1101/2021.04.08.21254902 [DOI] [Google Scholar]
Wellcome Open Res. 2024 Apr 13. doi: 10.21956/wellcomeopenres.21655.r70228

Reviewer response for version 2

Jonathan Clarke 1

Thank you for giving me the opportunity to review this article describing the implementation of two pathways to monitor patients with Covid-19 in hospitals in Vietnam using digital technologies. The article is well written overall, with a few typos identified and described below.

The authors acknowledge the limitations of the study in terms of their ability to evaluate the usability of the pathway or the patient benefit obtained from its implementation. As such, the article focusses instead on describing an intervention which may be of use to other health systems, or for which important lessons could be learned. With this in mind, I wonder if some of the discussion could be given over to clearly describing what the authors have learned from this experience that may be of use to others in a similar position. This set of recommendations based on their experience of a real-world rollout of a new technology at pace and amidst significant constraints are likely to be useful and of interest to a wider audience.

Several minor points for clarification in order to enable replication of the work or clear interpretation of the findings are described below.

Minor points:

Introduction:

- Please clarify that ‘Android’ refers to the operating system of the tablets

- Clarify what a ‘screen sharing mobile application’ is

Consider ensuring consistency in capitalisation of ‘oxygen’ across the article.

- Re: ‘patients using this system were mobile’ - were they inpatients, or was this part of a remote / home monitoring programme?

- It may be useful to provide a little context re: Vietnam’s COVID-19 timeline / policies - i.e. its zero-Covid policies avoided significant numbers of cases until April 2021. I can imagine some readers getting confused around timelines as they differ from other countries.

Methods:

- Define HTD on first use

- Typo - ‘am’ / ‘an’

- As described in the methods - System A aggregated multiple patient readings through simultaneously receiving the screens of several separate tablets, rather than parsing the clinical data and then representing it in a different format?

- I agree with the suggestion of the previous reviewer to include System B in Figure 1 to clarify the differences between it and System A.

- What did the piloting on healthy individuals and a non-patient area involve? What did it aim to achieve, what was learned?

Results:

- Typo ‘continuous monitoring for individual patients’ rather than ‘continuous monitoring in for individual patients’

- Figure 2: Why was cumulative monthly admissions chosen to represent activity over the study period? Would either daily or weekly admissions or cumulative over the entire period be more useful? Daily or weekly admissions may give a clearer indication of changing activity over time than interpreting the gradient of a cumulative plot that resets half way through. The vertical bars represent the capacity of a unit at the time when its capacity changes. This is quite hard to interpret and would perhaps be better visualised as a line for each unit showing its capacity over time. This would allow ready comparison and avoid confusion about the capacity of a unit between the bars.

Is the rationale for developing the new method (or application) clearly explained?

Yes

Is the description of the method technically sound?

Yes

Are the conclusions about the method and its performance adequately supported by the findings presented in the article?

Yes

If any results are presented, are all the source data underlying the results available to ensure full reproducibility?

No source data required

Are sufficient details provided to allow replication of the method development and its use by others?

Partly

Reviewer Expertise:

Health policy, Health services research, Digital health, Public health

I confirm that I have read this submission and believe that I have an appropriate level of expertise to confirm that it is of an acceptable scientific standard.

Wellcome Open Res. 2024 Apr 9. doi: 10.21956/wellcomeopenres.21655.r64776

Reviewer response for version 2

Deepak Berwal 1

The abstract provides a clear overview of the study, outlining the need for wearable pulse oximeters in monitoring COVID-19 patients in low-resource settings. It effectively summarizes the study's objectives and methodology. However, it lacks specific findings or results. Adding a sentence about the key findings would enhance its completeness.

The methods section provides a detailed description of the study's setting, the two monitoring systems used, and the challenges faced during implementation. However, it lacks specific technical details about the wearable devices, which could be valuable for readers interested in replicating the study. Including information about the make and model of the devices, their technical specifications, and how they were integrated into the healthcare workflow would enhance this section.

There are a couple of articles like [1] and  Analysis of Reflectance Photoplethysmograph Sensors (zenodo.org). Authors can consider multiple factors like contact pressure, temperature, motion artifacts etc. to report the accuracy of the outcomes and compare the wearable device with already existing standard devices. The following two articles can be taken into consideration while doing this.

The results section provides a clear timeline of when the monitoring systems were implemented and in which departments. However, it lacks specific data on patient outcomes or the performance of the monitoring systems. Including information on the number of patients monitored, key findings, and any issues encountered during the monitoring period would make this section more informative.

The discussion section effectively summarizes the challenges faced during the implementation of the monitoring systems and the importance of pre-existing expertise and infrastructure. It also touches on the limitations of the study. However, it could benefit from a more in-depth discussion of the potential implications of this work, such as the scalability of wearable monitoring systems in LMICs and the need for data analytics for decision support.

Recommendation: I recommend revising the manuscript to include specific data on patient outcomes, the performance of the monitoring systems, and technical details about the wearable devices used. Additionally, consider discussing the broader implications of this work and its potential for scalability and data-driven decision support in LMICs.

Is the rationale for developing the new method (or application) clearly explained?

Partly

Is the description of the method technically sound?

Partly

Are the conclusions about the method and its performance adequately supported by the findings presented in the article?

Partly

If any results are presented, are all the source data underlying the results available to ensure full reproducibility?

Yes

Are sufficient details provided to allow replication of the method development and its use by others?

No

Reviewer Expertise:

Wearable devices, Healthcare systems, and Biomedical signals

I confirm that I have read this submission and believe that I have an appropriate level of expertise to confirm that it is of an acceptable scientific standard, however I have significant reservations, as outlined above.

References

  • 1. : SpO2 Measurement: Non-Idealities and Ways to Improve Estimation Accuracy in Wearable Pulse Oximeters. IEEE Sensors Journal .2022;22(12) : 10.1109/JSEN.2022.3170069 11653-11664 10.1109/JSEN.2022.3170069 [DOI] [Google Scholar]
Wellcome Open Res. 2024 Apr 9. doi: 10.21956/wellcomeopenres.21655.r75888

Reviewer response for version 2

Benjamin Smarr 1

This paper identifies limitations of wired infrastructure for monitoring patients during times of surging hospital load, as in COVID-19 peaks of infection. Based on this, they test the feasibility of using wireless devices to augment hospital monitoring capacity, using a single hospital in Ho Chi Minh City. I have very little to add as this is a well framed report.

This paper seeks to address specifically the use of wireless O2 monitors instead of wired monitors within a hospital setting. The paper is clearly written, though the introduction could be reduced if needed given the scope. It has a limited scope of assessing staff's ability to interact with wireless O2 sensor solutions, but it makes clear that it addresses the scope it sets. The language is easy to read and the Methods are interesting and may help others adopt these practices.

Is the rationale for developing the new method (or application) clearly explained?

Yes

Is the description of the method technically sound?

Yes

Are the conclusions about the method and its performance adequately supported by the findings presented in the article?

Yes

If any results are presented, are all the source data underlying the results available to ensure full reproducibility?

Yes

Are sufficient details provided to allow replication of the method development and its use by others?

Yes

Reviewer Expertise:

this field says "optional" but flags as required.

I confirm that I have read this submission and believe that I have an appropriate level of expertise to confirm that it is of an acceptable scientific standard.

Wellcome Open Res. 2023 May 16. doi: 10.21956/wellcomeopenres.19987.r56596

Reviewer response for version 1

Tamara Vagg 1

Thank you for the opportunity to review this manuscript. It is very well written (with the exception of one or two typos) and is a very interesting example of the practical application of digital health. The manuscript describes two systems that use a wearable device coupled with a visualisation platform to view continuous SpO2 and heart rate data in multiple wards (including ICU) during the COVID-19 pandemic. While I agree with the authors limitations, that some sort of metrics to support the feasibility of the systems would further contribute to this work – I feel that a strength of this work is the detail of the real-world deployment and integration which was driven by the necessity of the pandemic. As it is currently written, enough detail has been given so that it can be reproduced by others. And I hope that the authors would consider the flow diagram outlined in “Optional Suggestion”.

Minor Suggestions:

Figure 1: Would it be possible to expand to describe the nuanced differences between the two monitoring systems(A+B)?

Table 1: Would it be possible to expand Table to include which monitoring system was being used in each ward?

Optional Suggestion:

The stages of this system setup and deployment are very interesting. If it was allowed and agreeable by the authors, it would be fantastic to have some sort of flow diagram to describe the various phases (and paralleled workflows) and perhaps the different teams involved e.g ideation -> testing in field hospital -> infrastructure establishment -> pilot system x with ward x (n=x) -> further roll out etc.

Is the rationale for developing the new method (or application) clearly explained?

Yes

Is the description of the method technically sound?

Yes

Are the conclusions about the method and its performance adequately supported by the findings presented in the article?

Yes

If any results are presented, are all the source data underlying the results available to ensure full reproducibility?

Yes

Are sufficient details provided to allow replication of the method development and its use by others?

Yes

Reviewer Expertise:

Digital Health, Virtual Reality, Telehealth, Virtual Monitoring, mHealth, eHealth, Design, CF

I confirm that I have read this submission and believe that I have an appropriate level of expertise to confirm that it is of an acceptable scientific standard.

Wellcome Open Res. 2023 Jun 3.
C Louise Thwaites 1

Thank you for these comments. We will include these suggestions into a further version of this article.

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Data Availability Statement

    Data included in this manuscript are provided by the Hospital for Tropical Diseases and are available through OUCRU and HTD data sharing processes.


    Articles from Wellcome Open Research are provided here courtesy of The Wellcome Trust