Abstract
Background
This study aimed to compare the results of the Chronos binocular/monocular refraction system, that measures objective and subjective ocular refraction in one unit, to objective findings obtained from a conventional autorefractometer and a conventional subjective ocular refraction using a trial-frame in real space.
Methods
Twenty-eight healthy volunteers (21.2 ± 1.5 years old) were included in this study. Objective ocular refraction was measured using two tests: the Chronos binocular/monocular refraction system under binocular conditions and a conventional autorefractometer under monocular conditions. Subjective ocular refraction was measured using three tests: Chronos binocular/monocular refraction system under binocular, monocular conditions, and trial-frame in the real space under monocular conditions. The measurement distance was set to 5.0 m for each test. All ocular refractions were converted into spherical equivalents (SEs).
Results
The objective SE was significantly more negative with Chronos binocular/monocular refraction system under binocular condition (− 4.08 ± 2.76 D) than with the conventional autorefractometer under monocular condition (− 3.85 ± 2.66 D) (P = 0.002). Although, the subjective SE was significantly more negative with Chronos binocular/monocular refraction system under binocular condition (− 3.55 ± 2.67 D) than with the trial-frame in the real space under monocular condition (− 3.33 ± 2.75 D) (P = 0.002), Chronos binocular/monocular refraction system under monocular condition (− 3.17 ± 2.57 D) was not significantly different from that in trial-frame in real space under monocular condition (P = 0.33).
Conclusion
These findings suggest that the Chronos binocular/monocular refraction system, which can complete both objective and subjective ocular refraction tests in a single unit, is suitable for screening ocular refraction, although it produces slightly more myopic results. Furthermore, subjective ocular refraction testing accuracy in Chronos binocular/monocular refraction system can be equivalent to trial-frame in real-space testing by switching from binocular to monocular condition.
Keywords: Ocular refraction, Refractive error, Astigmatism, Autorefractometer, Visual acuity
Background
Autorefractometers are one of the most commonly used tools in eye care practice [1]. Almost all eye care practitioners base their subjective ocular refraction test on an autorefractometer to obtain the best-corrected visual acuity at distance, since the autorefraction accuracy approximates that of the subjective refraction [2, 3]. However, objective ocular refraction differs to some degree from subjective ocular refraction. This is because an autorefractometer evaluates the objective ocular refraction based on corneal and lens power and axial length [4], whereas subjective ocular refraction tests include other elements, such as depth of focus [5], cone density, and visual processing [6, 7].
In addition, subjective refraction aims to provide the most comfortable vision in daily life because different individuals use eyeglasses in different situations, such as using a computer [8], reading [9], and may prefer correction optimized to attain certain goals such as the prevention of progression of myopia [10]. Moreover, the subjective ocular refraction test in clinics generally examines the eyes individually, whereas individuals use their eyes in a binocular fashion. Ocular refraction under binocular conditions was also shown to be more myopic than that under monocular conditions, owing to convergence accommodation under binocular conditions [11]. Thus, the examiners may test the eyes under binocular conditions to provide the best refractive correction under these conditions. However, conventional refractive tests cannot simultaneously confirm the ocular refraction in either eye. Thus, the lack of binocular refraction methods may cause some patients to feel discomfort while wearing their eyeglasses [12].
The Chronos binocular/monocular refraction system (Chronos; Topcon Corp., Tokyo, Japan) allows objective and subjective refraction under binocular (binocular mode) and monocular (monocular mode) conditions to be performed on a single instrument and has been developed to overcome these issues. Objective and subjective ocular refraction tests using Chronos were designed to maintain binocular vision because the subjects were simultaneously looking through the left and right lens barrels with both eyes. Furthermore, ocular refractive tests under binocular vision are expected to be more relevant to daily vision than conventional refractive tests under monocular conditions.
However, the consistency between binocular testing using the Chronos binocular/monocular refraction system and conventional monocular refractive examination is unknown, and differences from conventional methods must be clearly defined if the Chronos binocular/monocular refraction system is to be used clinically. Therefore, this study aimed to compare the Chronos binocular/monocular refraction system with conventional methods of objective ocular refraction using a conventional autorefractometer, and subjective ocular refraction using a trial-frame in real space.
Methods
Participants
In total, 28 healthy volunteers (mean ± standard deviation, 21.2 ± 1.5 years old; range: 18–25 years) were included in this prospective study. All subjects underwent complete ophthalmological examinations, including measures of best-corrected distance visual acuity at 5.0 m, stereoacuity at 40 cm (Titmus Stereotest; Stereo Optical Co., Inc., Chicago, USA), close (33 cm) and distant (5.0 m) alternating prism cover tests to identify heterophoria, and fundus examinations. Patients with ocular disease and a history of ocular surgery were excluded from the study. The individuals with large phorias were not excluded in this study.
After explaining the nature of the study and possible complications, all subjects provided informed consent for inclusion in the identification of information/images in an online open-access publication. This study adhered to the tenets of the 2013 revised Declaration of Helsinki by the World Medical Association. The Institutional Review Board of Teikyo University approved the experimental protocol and consent procedures (approval no. 21–067).
Apparatus
Chronos
This study used the Chronos binocular/monocular refraction system (Fig. 1). It was equipped with two auto refractometers based on the KR-800 automated refractometer (Topcon Corp, Tokyo, Japan) to measure the objective ocular refraction from − 25.00 to + 22.00 diopters (D) for spherical refractive error and up to 10.00 D of cylinder refractive error. The eye trackers detected each eye from the pupil and corneal reflections, positioned the auto-refractometers in front of each eye, and measured the ocular refraction in both eyes. The target fixation image was a house with a red roof.
The Chronos binocular/monocular refraction system uses several optical display systems for subjective refraction. The two optical systems project liquid-crystal images onto the subject’s right and left eyes and use the following four mechanisms to achieve a subjective refraction function:
Control of the convergence stimulus: The angle of measurement is adjusted by the three motors to realise a convergence stimulus at any distance from 25 cm to 6 m.
Control of accommodation stimulus: The spherical lens position is adjusted using one motor to realise accommodation stimuli at any distance from 25 cm to 6 m.
Control of spherical power: Spherical lens position is adjusted by one motor to realise any spherical power setting from + 18D to -18D.
Control of cylindrical power and cylindrical axis: The rotation angle of the two cylindrical lenses is adjusted by two motors to realise any cylindrical power and cylindrical axis setting from 0D to -8D [13, 14].
Chronos binocular/monocular refraction system can display optotypes on the two liquid crystal displays and occlude a non-tested eye using the console; however, the display for that eye does not turn off completely. The target (e.g. the Landolt ring) vanishes (Fig. 2A), whereas the peripheral frame remains to maintain binocular fusion (Fig. 2B).
The subjects fixated on the Landolt ring at a 5.0 m distance, which was adjusted optically, and the first ophthalmic lenses were selected automatically, based on the objective ocular refraction in both eyes. The examiner determined the subjective ocular refraction while examining the participant’s response using a Bluetooth tablet linked to Chronos binocular/monocular refraction system. The tablet device can change the status of the Chronos binocular/monocular refraction system, such as the measuring distance, spherical and cylindrical lens power, and target (Fig. 3).
Conventional auto refractometer
A KR-800 refractometer was used as the conventional autorefractometer. It can measure monocular objective ocular refraction over the same range as that measured by Chronos binocular/monocular refraction system.
Subjective ocular refraction
Subjective ocular refraction tests were performed in a well-lit room (600 lx), with both Chronos binocular/monocular refraction system and the trial-frame in real space using VC-60 (Takagi Co., Ltd., Nagano, Japan; Fig. 4), a standard visual acuity chart. The examiner presented the Landolt ring on a liquid-crystal display (resolution of 2560 × 1440 pixels), corresponding to the controller from a distance. The luminance of the backlight was 300 cd/m2.
Subjective ocular refraction was determined by combining the maximum plus or minimum minus spherical and cylindrical lenses necessary to provide a best-corrected visual acuity of − 0.176 logMAR (logarithm of the minimum angle of resolution) at an optical and natural distance of 5.0 m (with decimal acuity of 1.5).
The initial spherical lens was determined by adding + 1.00 D to the objective ocular refraction to prevent overcorrection, and the examiner (M.F.) confirmed that the participant was unable to clearly see the − 0.176 logMAR optotype with this lens. The examiner then added spherical lenses in increments of − 0.25D to obtain the highest visual acuity with spherical lenses alone. Additionally, the examiner used Jackson’s Cross Cylinder of ± 0.50 D and ± 0.25 D to determine the cylindrical lens after the single spherical lens obtained the highest visual acuity. Dot targets were used to determine the power and axes of the cylindrical lenses. The cylindrical lens axis was determined at 5° increments. After identifying the cylindrical lens, the examiner readjusted the spherical lens.
The procedure for subjective refraction test was as follows.
First, the subjective lens was selected by referring to the objective ocular refraction of both eyes. For the spherical lens, a + 1D lens was added to prevent overcorrection.
Showing the Landolt ring on the display, spherical lenses were added so that the subjects could obtain the highest visual acuity.
Showing dot targets, cylindrical lenses and axis were corrected using Jackson’s Cross Cylinder of ± 0.50 D and ± 0.25 D. The cylindrical lenses were added, and the axis was determined using 5° increments.
Showing the Landolt ring, spherical lenses were recorded so that the subjects could obtain their best-corrected visual acuity.
The procedure was the same for the subjective refraction test using Chronos binocular/monocular refraction system and the trial-frame in real space. However, the test using the Chronos binocular/monocular refraction system did not examine the peripheral frame of the eye to maintain binocular fusion. Although binocular balance is important in clinical practice, only the monocular visual acuity was evaluated in this study.
We were concerned that the results of the subjective refractive test would be affected by the objective ocular refraction results. Therefore, the subjects underwent the visual acuity test four times under four random conditions, as follows Fig. 5.
Additional experiment: Chronos binocular/monocular refraction system under monocular vs. binocular conditions
As the default setting of the Chronos binocular/monocular refraction system is to test with both eyes open, it does not entirely match the conditions of a conventional subjective ocular refraction test conducted with a single eye separately. Thus, we conducted the subjective ocular refraction test on Chronos binocular/monocular refraction system under monocular conditions, such as conventional subjective ocular refraction, with one eye occluded.
The subject’s left eye was completely occluded with white gauze (Surgical Pad, Hakujuji Co., Ltd.) to assess the difference between the binocular and monocular conditions in the Chronos binocular/monocular refraction system. Subjective refraction in the additional experiment was initiated by adding + 1.00 D from the subjective spherical equivalents (SE) of the previous subjective ocular refraction test. The experimental environment and procedures for the Chronos binocular/monocular refraction system were the same as those used in the previous experiments.
Data acquisition conditions
In this study, the ocular refraction data were obtained under five conditions (Fig. 5; Table 1).
Table 1.
Objective testing | Subjective testing | |
---|---|---|
Test unit (Chronos binocular/monocular refraction system) | Binocular conditions |
• Binocular conditions • Monocular conditions |
Controls (monocular conditions) |
Conventional autorefractometer (KR-800) |
Trial-frame refraction (“real space”) |
The objective ocular refraction was measured using two tests: the Chronos binocular/monocular refraction system under binocular conditions and a conventional autorefractometer under monocular conditions.
The subjective ocular refraction was measured using three tests: Chronos binocular/monocular refraction system under binocular and monocular conditions and trial-frame in the real space under monocular conditions.
Data analysis
The objective and subjective ocular refractions were converted to SEs as follows:
where S and C are the spherical and cylindrical lens power, respectively. In this study, a cylindrical lens was used only in the negative format.
Objective and subjective astigmatism were converted from spherocylindrical to power vector notation by applying Fourier transformation using the following equations:
where α is an axis of a cylindrical lens, J0 indicates the cylinder lens power set at 90° and 180°, the positive values of J0 were with the rule (WTR) astigmatism, the negative values of J0 were against the rule astigmatism, and J45 is the cylinder lens power set at 45° and 135°, representing oblique astigmatism.
The simple conversion from the power vector notation to conventional notation is twice the square root of the sum of J02 and J452.
Statistical analysis
Bland–Altman analysis was performed to compare the objective ocular refractions in SE using J0 and J45 of the Chronos binocular/monocular refraction system under binocular conditions and KR-800 under monocular conditions, as well as the subjective ocular refractions in SE using J0 and J45 of the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular conditions.
After assessing the normality of variable distributions using Shapiro–Wilk tests, the fixed and proportional biases between the two apparatuses and the measurement conditions were analysed using paired t-tests and single linear regression analyses.
The chronosinocular/monocular refraction system maintains the binocular fusion. Therefore, the influence of fusional convergence was considered. The difference in the objective ocular refraction between the Chronos binocular/monocular refraction system under binocular conditions and the KR-800 under monocular conditions was calculated because of accommodation by convergence. The relationship between the degree of heterophoria and the differences in objective ocular refraction was analysed using Spearman’s rank correlation coefficients.
In an additional experiment, the differences between the Chronos binocular/monocular refraction system under monocular and binocular conditions and real space under monocular conditions were analysed using a paired t-test following the assessment of normality distribution using the Shapiro–Wilk test with Bonferroni correction.
IBM SPSS Statistics for Windows version 26 (IBM Corp., Armonk, NY, USA) was used to determine the significance of the differences, and P < 0.05 was considered statistically significant (Fig. 6).
This study achieved five ocular refraction results: two objective ocular refractions of the Chronos binocular/monocular refraction system under binocular conditions and KR-800 under monocular conditions, three subjective ocular refractions of the Chronos binocular/monocular refraction system under binocular conditions, monocular conditions, and a trial-frame in real space under monocular conditions. Bland–Altman analysis was performed to analyse the consistency of the Chronos binocular/monocular refraction system under binocular conditions and the conventional method, which measures monocularly, followed by evaluation of the difference between objective and subjective ocular refraction in the Chronos binocular/monocular refraction system under binocular conditions. The Spearman’s rank correlation coefficient was used to evaluate the correlation between objective ocular refraction by fusional convergence in the Chronos binocular/monocular refraction system under binocular conditions and KR-800 under monocular conditions. Finally, a paired t-test with Bonferroni correction was conducted to assess whether the subjective ocular refraction in the Chronos binocular/monocular refraction system under monocular conditions differed from the Chronos binocular/monocular refraction system under binocular conditions and the real space under monocular conditions.
Results
Characteristics of the subjects
The participants’ characteristics are listed in Table 2. The objective refraction was − 3.80 ± 2.68 D using the KR-800 under monocular condition. The angle of heterophoria was − 6.9 ± 5.9 prism diopter (PD) at near distance and − 3.3 ± 4.3 PD at far distance. All subjects had stereo acuity of 1.60 log arcsec (40 s equally).
Table 2.
Subject | Age (years) | Sph (D) | Cyl (D) | Ax (°) | APCT (PD) | Stereo acuity | PD (mm) | |
---|---|---|---|---|---|---|---|---|
Near | Distant | |||||||
S1 | 23 | −6.50 | 0 | 180 | −16 | −16 | 40 | 62.5 |
S2 | 21 | −1.50 | 0 | 180 | −2 | −2 | 40 | 58 |
S3 | 21 | −1.75 | −1.00 | 90 | −8 | −4 | 40 | 61.5 |
S4 | 20 | −0.50 | −0.25 | 70 | −16 | −6 | 40 | 61 |
S5 | 21 | −2.25 | −0.50 | 170 | −4 | 0 | 40 | 65.5 |
S6 | 21 | −7.50 | −3.00 | 175 | −20 | −14 | 40 | 61 |
S7 | 19 | −4.75 | −2.00 | 175 | −6 | 0 | 40 | 62 |
S8 | 20 | −4.00 | −0.75 | 160 | −12 | −6 | 40 | 66.5 |
S9 | 23 | −7.75 | −1.00 | 180 | −8 | 0 | 40 | 61 |
S10 | 25 | −2.50 | −0.50 | 145 | 1 | −2 | 40 | 62 |
S11 | 21 | −7.75 | −1.00 | 175 | −14 | −8 | 40 | 59.5 |
S12 | 22 | −1.75 | 0 | 180 | 2 | −2 | 40 | 57.5 |
S13 | 21 | −6.25 | −1.00 | 170 | −8 | −4 | 40 | 61 |
S14 | 18 | −5.25 | −0.25 | 30 | −12 | −2 | 40 | 64.5 |
S15 | 19 | −5.50 | −0.75 | 165 | −4 | −1 | 40 | 61.5 |
S16 | 22 | 0 | 0 | 180 | −4 | 0 | 40 | 65 |
S17 | 22 | −1.00 | −0.25 | 100 | −6 | −2 | 40 | 59 |
S18 | 21 | −4.75 | −0.75 | 175 | −20 | −12 | 40 | 67 |
S19 | 21 | −3.75 | −1.50 | 180 | −4 | 0 | 40 | 59.5 |
S20 | 21 | −0.75 | −0.25 | 115 | −2 | −2 | 40 | 58 |
S21 | 24 | 0.25 | −0.25 | 180 | −8 | −4 | 40 | 61 |
S22 | 21 | −0.75 | −0.75 | 160 | 0 | 0 | 40 | 58 |
S23 | 21 | −5.25 | −0.75 | 170 | −6 | 0 | 40 | 69.5 |
S24 | 23 | −4.50 | −1.75 | 175 | −2 | −2 | 40 | 62.5 |
S25 | 22 | −4.75 | −0.50 | 165 | −4 | −2 | 40 | 61 |
S26 | 23 | −0.75 | −0.25 | 155 | 0 | 0 | 40 | 59.5 |
S27 | 21 | −5.00 | −1.25 | 175 | −4 | −2 | 40 | 60.5 |
S28 | 23 | −1.00 | −0.50 | 5 | −2 | 0 | 40 | 62.5 |
Sph, spherical power; Cyl, cylindrical power; Ax, cylindrical axis; APCT, alternative prism cover test; PD, pupillary distance
Objective ocular refraction: Chronos binocular/monocular refraction system under binocular condition) vs. KR-800 under monocular condition
The objective ocular refraction test in the Chronos binocular/monocular refraction system defaults to binocular conditions. Therefore, objective ocular refraction may differ from that of the conventional autorefractometer performed under monocular conditions. Therefore, we compared the objective ocular refractions of the Chronos binocular/monocular refraction system and KR-800 under binocular and monocular conditions.
The SE obtained with Chronos binocular/monocular refraction system under binocular condition was significantly more myopic (− 4.08 ± 2.76 D) than that of KR-800 under monocular condition (− 3.85 ± 2.66 D) (P = 0.002; Fig. 7A; Table 3). No proportional bias was observed between the two autorefractometers (R2 = 0.077, P = 0.153; Fig. 7A). The mean value of the differences between Chronos binocular/monocular refraction system under binocular condition and KR-800 under monocular condition was − 0.23 D and the 95% limits of agreement (LOA) ranged from − 0.93 to 0.47 D. The angle of heterophoria was significantly correlated with the difference in SE between the Chronos binocular/monocular refraction system under binocular conditions and KR-800 under monocular conditions without outlier data (without outliers, rs = − 0.389, P = 0.045; without outliers, rs = − 0.231, P = 0.163) (Fig. 8).
Table 3.
Objective testing | Subjective testing | |
---|---|---|
Test unit (Chronos binocular/monocular refraction system) |
Binocular conditions: –4.08 ± 2.76 D |
• Binocular conditions:–3.55 ± 2.67 D • Monocular conditions:–3.17 ± 2.57 D |
Controls (monocular conditions) |
Conventional autorefractometer (KR-800): –3.85 ± 2.66 D |
Trial-frame in real space –3.33 ± 2.75 D |
The J0 of Chronos binocular/monocular refraction system under binocular condition (0.37 ± 0.43 D) was significantly more positive than that of KR-800 under monocular condition (0.29 ± 0.39 D) (P = 0.005; Fig. 7B). A proportional bias was not found between the two autorefractometers (R2 = 0.095, P = 0.111; Fig. 7B’). The mean value of the differences between Chronos binocular/monocular refraction system under binocular condition and KR-800 under monocular condition was − 0.09 D and the 95% LOA ranged from − 0.20 D − 0.38 D.
The J45 of Chronos binocular/monocular refraction system under binocular condition (− 0.02 ± 0.11 D) was significantly more positive than that of KR-800 under monocular condition (− 0.08 ± 0.10 D) (P = 0.003; Fig. 7C). No proportional bias was observed between the two autorefractometers (R2 = 0.029, P = 0.385; Fig. 7C). The mean value of the differences between Chronos binocular/monocular refraction system under binocular condition and KR-800 under monocular condition was − 0.06 D and the 95% LOA ranged from − 0.14 D − 0.26 D.
Red and blue boxplots with dots indicate the objective SE in the Chronos binocular/monocular refraction system and the KR-800 (A–C). The solid and dashed red lines indicate the average and 95% limits of agreement in the differences between Chronos binocular/monocular refraction system and KR-800 (A’–C’). The SE (A), J0 (B), and J45 (C) differed significantly between the two systems and KR-800. No proportional bias was observed between the two devices.
The angle of heterophoria was not significantly correlated with difference of SE between Chronos binocular/monocular refraction system and KR-800 without outlier (rs = − 0.389, P = 0.045) and with outlier (rs = − 0.231, P = 0.163).
Subjective refraction test: Chronos binocular/monocular refraction system under binocular condition vs. conventional trial-frame real space under monocular condition
The subjective ocular refraction test in the Chronos Binocular/Monocular Refraction System defaults to binocular conditions. A Landolt ring was presented to the subject’s eye, and a blank was shown to the other eye. However, black frames that maintain binocular fusion remain in both eyes. Therefore, subjective ocular refraction between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular conditions may differ.
The subjective SE from Chronos binocular/monocular refraction system under binocular condition (− 3.55 ± 2.67 D) was significantly more myopic than that from trial-frame in real space under monocular condition (− 3.33 ± 2.75 D) (P = 0.002; Fig. 9A; Table 3). No proportional bias was observed between the two examinations (R2 = 0.070, P = 0.174; Fig. 9A). The mean value of the differences between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular condition was − 0.22 D. The 95% LOA ranged from − 0.87 − 0.43 D.
The J0 was not significantly different between Chronos binocular/monocular refraction system under binocular condition (0.23 ± 0.34 D) and trial-frame in real space under monocular condition (0.24 ± 0.39 D) (P = 0.61; Fig. 9B). No proportional bias was observed between the two examinations (R2 = 0.137, P = 0.053; Fig. 9B). The mean value of the differences between Chronos binocular/monocular refraction system under binocular condition and trial-frame in real space under monocular condition was − 0.01 D, and the 95% LOA ranged from − 0.29 − 0.26 D.
The J45 was not significantly different between Chronos binocular/monocular refraction system under binocular condition (− 0.05 ± 0.19 D) and trial-frame in real space under monocular condition (− 0.07 ± 0.15 D) (P = 0.41; Fig. 9C). No proportional bias was observed between the two examinations (R2 = 0.093, P = 0.115; Fig. 9C). The mean value of the differences between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular condition was 0.02 D, and the 95% LOA ranged from − 0.27 − 0.32 D.
The red and blue boxplots with dots indicate the objective SE in the Chronos binocular/monocular refraction system under binocular conditions, and the trial-frame in real space under monocular conditions (A–C). The solid and dashed red lines indicate the average and 95% limits of agreement of the difference between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular conditions (A’–C’). Only SE (A) was significantly and negatively greater in the Chronos binocular/monocular refraction system than in the trial-frame in real space. No proportional bias is observed between the two devices.
Difference between objective and subjective chronos test
In general, objective ocular refraction measured by autorefractometer is more negative than subjective ocular refraction [15]. We examined whether similar results occurred in the Chronos binocular/monocular refraction system.
The SE and J0 were significantly more myopic in the objective test (SE, − 4.08 ± 2.76 D; J0, 0.37 ± 0.43 D) than in the subjective test (SE, − 3.55 ± 2.67 D; J0, 0.23 ± 0.34 D) in Chronos binocular/monocular refraction system under binocular condition (P < 0.001). J45 did not significantly differ between the objective (− 0.02 ± 0.11 D) and subjective (− 0.05 ± 0.19 D) tests under binocular condition (P = 0.145).
Chronos binocular/monocular refraction system under monocular vs. chronos binocular/monocular refraction system under binocular vs. trial-frame in real space under monocular conditions
By default, the Chronos binocular/monocular refraction system performs tests under binocular conditions, which is not the same as the monocular condition used in traditional subjective ocular refraction tests. To ensure accurate results, we conducted a subjective ocular refraction test using the Chronos binocular/monocular refraction system under monocular conditions, similar to a conventional test in which one eye was occluded.
The subjective SE was significantly more hyperopic for Chronos binocular/monocular refraction system under the monocular condition (− 3.17 ± 2.57 D) than under the binocular condition (− 3.55 ± 2.67 D D) (P < 0.001; Fig. 9; Table 3). The subjective SE did not significantly differ between Chronos binocular/monocular refraction system under the monocular condition and the real space under monocular condition (− 3.33 ± 2.75 D) (P = 0.33; Fig. 10).
The Chronos binocular/monocular refraction system subjective refraction tested under binocular conditions resulted in a significantly more negative spherical equivalent than the Chronos binocular/monocular refraction system subjective refraction tested under monocular conditions and the trial-frame refraction tested in real space and under monocular conditions.
***: P < 0.001, Paired t-test with Bonferroni correction.
**: P = 0.006, Paired t-test with Bonferroni correction.
Discussion
Overview
Objective and subjective ocular refraction were measured using the Chronos binocular/monocular refraction system. The SE in the objective and subjective ocular refractions with Chronos binocular/monocular refraction system under binocular condition changed to myopic values of − 0.23 D and − 0.22 D compared to KR-800 and trial-frame in real space under monocular condition. These findings indicate that the Chronos binocular/monocular refraction system under binocular conditions resulted in refractive errors that were more myopic in both objective and subjective refraction tests under monocular conditions. Therefore, under binocular conditions, the Chronos binocular/monocular refraction system is unsuitable for situations requiring accurate refraction, such as preoperative examinations and eyeglass prescriptions. In contrast, the Chronos binocular/monocular refraction system is suitable for screening under binocular conditions because it can complete both objective and subjective ocular refraction tests in a single unit.
The subjective ocular refraction of the Chronos binocular/monocular refraction system under monocular conditions was significantly more hyperopic than that under binocular conditions, and was not significantly different from that of the trial-frame in real space under monocular conditions. These findings suggest that subjective ocular refraction in the monocular condition provides less myopia results than the binocular condition, and that the monocular condition results are comparable to the trial-frame in real space.
Objective ocular refraction
The Chronos binocular/monocular refraction system was equipped with two automated refractometers based on a KR-800 conventional autorefractometer. However, the objective SE was significantly more myopic with the Chronos binocular/monocular refraction system under binocular conditions than with the KR-800 under monocular conditions (Fig. 7A and A’). The subjects maintained binocular fusion using Chronos to measure objective ocular refraction. Conversely, the KR-800 measures objective ocular refraction without fusional convergence because it uses monocular measurements, and convergence induces accommodation [11]. Hence, we considered accommodation to be induced by fusional convergence by using Chronos.
For objective astigmatism, J0 and J45 were significantly more positive with the Chronos binocular/monocular refraction system under binocular conditions than with the KR-800 under monocular conditions (Fig. 7B and B’, 7 C, and 7 C’). An increase in J0 indicates that the WTR becomes stronger, and an increase in J45 indicates that the power of the ocular refraction shifts from 135° to 45°. Our findings are consistent with those of Porrill et al., who demonstrated that both eyes were inverted during convergence [16]. These findings support our hypothesis that accommodation is induced by fusional convergence, using the Chronos binocular/monocular refraction system. However, previous studies reported that accommodation alters astigmatism [17]. Therefore, measuring the amplitude of fusional convergence may be necessary to test our hypothesis in future studies.
Subjective ocular refraction
Regarding subjective ocular refraction, the subjective SE obtained with the Chronos binocular/monocular refraction system under binocular condition was significantly more myopic than that obtained trial-frame in real space under monocular condition (Fig. 9A and A′). We believe that subjective ocular refraction under binocular conditions shifted to myopia owing to accommodation related to convergence. Therefore, the results of the subjective ocular refraction test may depend on those of the objective ocular refraction test [18]. To minimize the bias of the objective ocular refraction in the subjective ocular refraction test, we mixed the objective ocular refraction measured by the Chronos binocular/monocular refraction system under binocular condition and KR-800 under the monocular condition. In this study, the measurement distance of subjective ocular refraction was the same between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular conditions.
To maintain binocular fusion, the peripheral frame of the eye was not examined in the subjective refraction test using the Chronos binocular/monocular refraction system under binocular conditions (Fig. 3). We occluded the subject’s left eye to avoid binocular fusion when using the Chronos binocular/monocular refraction system in an additional experiment. The subjective SE under monocular conditions showed a hyperopic shift relative to that under binocular conditions. Furthermore, the subjective SE did not significantly differ between the Chronos binocular/monocular refraction system under monocular conditions and the trial-frame in real space under monocular conditions (Fig. 10). These findings support our hypothesis that convergence accommodation through fusional convergence works by using a Chronos binocular/monocular refraction system under binocular conditions. Moreover, with the Chronos binocular/monocular refraction system, an image from one eye can be removed via the console. Therefore, we can set the condition of the subjective refraction test within an error of 0.25 D between the conventional monocular test and the binocular test as required, which is closer to the trial-frame in real-space vision.
However, the subjective J0 and J45 were not significantly different between the Chronos binocular/monocular refraction system under binocular conditions and the trial-frame in real space under monocular conditions (Fig. 9A and A’, 9B, and 9B’). We assume this was due to Jackson’s cross-cylinder method. Moreover, the astigmatism axis was determined in 5° increments. For example, if the axis of astigmatism is changed from 0° to 5° and the astigmatism power is increased by − 0.25 D, J0 changes by 0.11 D. Although the incyclodeviation (this is intorsion) also occurs during the subjective refraction test similar with the objective refraction test, Chronos binocular/monocular refraction system under binocular condition can determine subjective ocular refraction with similar accuracy as the trial-frame in real space under monocular condition, if the astigmatic axis changed in 5° increments in Jackson’s Cross Cylinder.
In this study, we compared objective and subjective ocular refractions in the Chronos binocular/monocular refraction system under binocular and monocular conditions and a trial-frame in real space under monocular conditions; however, eye movements during the examinations could not be evaluated. Particularly, the results of the present study suggest that fusional convergence occurs under binocular conditions when using Chronos binocular/monocular refraction system (Fig. 8). Furthermore, there were no participants with esophoria, despite the random recruitment of participants in this study. Therefore, to test our hypothesis, it was necessary to simultaneously measure eye movements during ocular refraction tests under non-cycloplegic and cycloplegic conditions to confirm whether fusional convergence is induced by the peripheral frame. Thus, in future studies, we aim to include an eye tracking system to Chronos binocular/monocular refraction system to assess the consistency of eye movements and ocular refraction upon recruiting subjects with esophoria, exophoria, and orthophoria.
Conclusion
The objective and subjective SEs were significantly more negative when using the Chronos binocular/monocular refraction system under binocular conditions than when using the trial-frame in real space under monocular conditions. The subjective SE under monocular conditions using the Chronos binocular/monocular refraction system was comparable to that of a trial-frame in real space under monocular conditions. These findings suggest that the Chronos binocular/monocular refraction system, capable of completing objective and subjective ocular refraction testing with a single unit, is suitable for screening ocular refraction, although it produces slightly more myopic results. Furthermore, the subjective ocular refraction testing accuracy in the Chronos binocular/monocular refraction system can be made equivalent to a trial-frame in real-space testing by switching from binocular to monocular conditions.
Acknowledgements
The authors are grateful to Editage for helpful English proofreading.
Abbreviations
- SE
spherical equivalents
- D
diopters
- logMAR
logarithm of the minimum angle of resolution
- WTR
with the rule
- LOA
limits of agreement
Author contributions
M.H. conceived the project. M. H., T. Y., A. H., Y. H., and M. S. designed the experiments. M.F. performed the experiments. M.F. and M.H. analysed data. M. F., M. H., and K. M. wrote the manuscript. All the authors have reviewed and approved the final manuscript. T.Y., A.H., Y.H., and M.S. were not involved in the experiments, data analyses, or manuscript writing.
Funding
This work was supported by Grants-in-Aid for Early-Career Scientists, Scientific Research (A) and (B), Challenging Exploratory Research; the Japan Society for the Promotion of Science (22K18231, 20K04271, 22H00539, and 19K21783 [all M.H.]); the Nakatani Foundation (M.H.); the Inamori Foundation (M.H.); the Japan Medical Herb Association Research Grant Program (M.H.); Clinical Research Promotion Foundation (2021) (M.H.); the Terumo Life Science Foundation (22-III 1009) (M.H.), and Intensive Support for Young Promising Researchers, Japan Agency for Medical Research and Development (23he0422031j0001 to M.H.).
Data availability
All data related to this study are stated in the paper.
Declarations
Ethics approval and consent to participate
After explaining the nature of the study and possible complications, all participants provided informed consent. This study adhered to the tenets of the 2013 revised Declaration of Helsinki of the World Medical Association. The Institutional Review Board of Teikyo University approved the experimental protocol and consent procedures (approval no. 21–067).
Consent for publication
Not applicable.
Competing interests
The author M.H., is financial support to Topcon Corp. for this study performed. T.Y., A.H., Y.H., and M.S. are employees of the Topcon Corporation. As described in the Authors’ Contributions section, they only contributed to setting up the study design. To avoid bias, the experiments, data analysis, and writing of this paper were conducted independently at Teikyo University. T.Y., A.H., Y.H., and M.S. were not involved in the experiments, data analyses, or manuscript writing. All other authors declare no competing interests.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Gurnani B, Kaur K. Autorefractors. StatPearls. edn. Treasure Island (FL); 2023.
- 2.Hastings GD, Marsack JD, Nguyen LC, Cheng H, Applegate RA. Is an objective refraction optimised using the visual Strehl ratio better than a subjective refraction? Ophthalmic Physiol Opt. 2017;37(3):317–25. doi: 10.1111/opo.12363. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Rossato M, Nart A, Messina G, Favro F, Rossato V, Rrutja E, Biancalana V. The Refraction Assessment and the electronic trial Frame Measurement during Standing or sitting position can affect Postural Stability. Int J Environ Res Public Health. 2022;19(3):1558. doi: 10.3390/ijerph19031558. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Lei Y, Chen X, Cheng M, Li B, Jiang Y, Xu Y, Wang X. Comparisons of objective and subjective refraction with and without cycloplegia using binocular wavefront optometer with autorefraction and retinoscopy in school-age children. Graefe’s Archive Clin Experimental Ophthalmol. 2023;261(5):1465–72. doi: 10.1007/s00417-022-05936-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Green DG, Powers MK, Banks MS. Depth of focus, eye size and visual acuity. Vis Res. 1980;20(10):827–35. doi: 10.1016/0042-6989(80)90063-2. [DOI] [PubMed] [Google Scholar]
- 6.Li KY, Tiruveedhula P, Roorda A. Intersubject variability of foveal cone photoreceptor density in relation to eye length. Invest Ophthalmol Vis Sci. 2010;51(12):6858–67. doi: 10.1167/iovs.10-5499. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Rossi EA, Roorda A. The relationship between visual resolution and cone spacing in the human fovea. Nat Neurosci. 2010;13(2):156–7. doi: 10.1038/nn.2465. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Rosenfield M. Computer vision syndrome: a review of ocular causes and potential treatments. Ophthalmic Physiol Opt. 2011;31(5):502–15. doi: 10.1111/j.1475-1313.2011.00834.x. [DOI] [PubMed] [Google Scholar]
- 9.Katz JA, Karpecki PM, Dorca A, Chiva-Razavi S, Floyd H, Barnes E, Wuttke M, Donnenfeld E. Presbyopia– A Review of Current Treatment options and emerging therapies. Clin Ophthalmol. 2021;15:2167–78. doi: 10.2147/OPTH.S259011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Yazdani N, Sadeghi R, Ehsaei A, Taghipour A, Hasanzadeh S, Zarifmahmoudi L, Heravian Shandiz J. Under-correction or full correction of myopia? A meta-analysis. J Optom. 2021;14(1):11–9. doi: 10.1016/j.optom.2020.04.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kersten D, Legge GE. Convergence accommodation. J Opt Soc Am. 1983;73(3):332–8. doi: 10.1364/JOSA.73.000332. [DOI] [PubMed] [Google Scholar]
- 12.Marmamula S, Bhoopalan D, Kumbham TR, Yelagondula VK, Keeffe J. Prevalence, pattern, and compliance with spectacles use among the elderly in homes for the aged in South India: the Hyderabad Ocular Morbidity in Elderly Study (HOMES) Indian J Ophthalmol. 2023;71(1):263–7. doi: 10.4103/ijo.IJO_884_22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993;19(4):524–33. doi: 10.1016/S0886-3350(13)80617-7. [DOI] [PubMed] [Google Scholar]
- 14.Tamura M, Moriguchi Y, Yeh SY, Matsumoto A, Shibutani M, Asao T, Mino T, Nakanishi M, Kubota A, Akiba M. Sensorless astigmatism correction using a variable cross-cylinder for high lateral resolution optical coherence tomography in a human retina. Appl Opt. 2021;60(30):9553–9. doi: 10.1364/AO.441646. [DOI] [PubMed] [Google Scholar]
- 15.Yeung IY, Mantry S, Cunliffe IA, Benson MT, Shah S. Correlation of Nidek OPD-Scan objective refraction with subjective refraction. J Refract Surg. 2004;20(5 Suppl):S734–736. doi: 10.3928/1081-597X-20040903-22. [DOI] [PubMed] [Google Scholar]
- 16.Porrill J, Ivins JP, Frisby JP. The variation of torsion with vergence and elevation. Vis Res. 1999;39(23):3934–50. doi: 10.1016/S0042-6989(99)00102-9. [DOI] [PubMed] [Google Scholar]
- 17.Radhakrishnan H, Charman WN. Age-related changes in static accommodation and accommodative miosis. Ophthalmic Physiol Opt. 2007;27(4):342–52. doi: 10.1111/j.1475-1313.2007.00484.x. [DOI] [PubMed] [Google Scholar]
- 18.Kemchoknatee P, Sunlakaviset P, Khieokhoen N, Srisombut T, Tangon D. A Comparison of Autorefraction and Subjective Refraction in an Academic Optometry Clinic. Cureus 2023. [DOI] [PMC free article] [PubMed]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data related to this study are stated in the paper.